Page 1: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS

Version 2016 / 07

Stator All-In-One

Page 2: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 2/52

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS

ATESTEO GmbH

Konrad-Zuse-Str. 3

52477 Alsdorf

T +49 (0) 2404 9870-0

F +49 (0) 2404 9870-59

[email protected]

www.atesteo.com

Page 3: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52

Dear customer,

before leaving our company every unit is tested by extensive function

and quality examinations, which guarantee that the system complies

with the stated specifications. Nevertheless should there be any

problem, please contact us.

Before shipping a system, the serial numbers of each component of

your configuration is registered by our company, so that an individual

and short-term support can be guaranteed. It is understood that we will

inform you about innovations and modifications of the system.

Guarantee

ATESTEO takes over the Guarantee for the legislative prescribed time.

All the reparations will be done in this time without calculation

executed. Any damage, which is caused by equipment usage, cannot be

as guarantee case valid.

Page 4: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 4/52

F1iS / F2iS

F1i / F2i

FLFM1iS

FLFM1eS

Page 5: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 5/52

Content

1 Description ...................................................................................................................................... 7

2 Safety Instructions .......................................................................................................................... 8

3 Introduction ..................................................................................................................................... 9

3.1 One-channel telemetry (FM) F1i/F2i .............................................................................................. 10

3.2 Dual -Telemetry DT (FM/ FM) F1i/F2i ............................................................................................ 11

4 Mechanical Installation ................................................................................................................. 13

4.1 Part list ........................................................................................................................................... 13

4.1.1 Part list F1iS/F2iS ........................................................................................................................... 13

4.1.2 Part list F1i/F2i ................................................................................................................................ 13

4.1.3 Part list FLFM1iS............................................................................................................................. 13

4.1.4 Part list FLFM1eS ........................................................................................................................... 13

4.2 Installation F1iS / F2iS ................................................................................................................... 14

4.2.1 Mounting the Stator ......................................................................................................................... 15

4.2.2 Mounting of the foot base ............................................................................................................... 16

4.2.3 Mounting with the foot base ............................................................................................................ 20

4.2.4 Mounting the Optional Magnetic Speed Detection ......................................................................... 22

4.2.5 Recommended fastening torques for screws ................................................................................. 23

4.3 Installation F1i/F2i torquemeter ..................................................................................................... 24

4.3.1 Mounting the Stator ......................................................................................................................... 26

4.3.2 Recommended starting torques for screws .................................................................................... 26

4.4 Installation FLFM1iS ...................................................................................................................... 27

4.4.1 Mounting the Stator ......................................................................................................................... 28

4.4.2 Mounting the torquemeter ............................................................................................................... 29

4.4.3 Mounting the optical speed detection ............................................................................................. 29

4.4.4 Mounting the stator with a foot base ............................................................................................... 30

4.4.5 Mounting the torquemeter ............................................................................................................... 33

4.4.6 Recommended fastening torques for screws ................................................................................. 33

4.4.7 Mounting the optical speed detection ............................................................................................. 35

4.1 Installation FLFM1eS ...................................................................................................................... 36

4.1.3 Mounting the Stator ......................................................................................................................... 37

4.1.4 Mounting the optical speed detection ............................................................................................. 38

4.1.5 Recommended fastening torques for screws ................................................................................. 38

4.1.6 Electronics ....................................................................................................................................... 39

4.1.7 Components earthing ...................................................................................................................... 39

5 Speed measuring system ............................................................................................................. 40

5.1 Speed measuring system F1iS / F2iS............................................................................................ 40

5.1.1 Inductive Sensor ............................................................................................................................. 40

5.1.2 Magnetic Sensor (F1iS: 1000 increments / F2iS:1448 increments) ............................................... 41

5.1.3 Adjustment of the right distance:..................................................................................................... 42

5.2 Speed measuring system F1/F1i (480 increments): ...................................................................... 43

Page 6: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 6/52

5.2.1 Adjustment of speed measuring for torquemeter F1/F1i: ............................................................... 43

5.2.2 Adjustment of the right distance:..................................................................................................... 43

5.3 Speed measuring system FLFM1iS, FLFM1eS ............................................................................. 45

5.3.1 Optical speed detection .................................................................................................................. 45

5.3.2 Setup the optical speed .................................................................................................................. 46

5.3.3 Speed measuring adjustment ......................................................................................................... 48

6 EC Manufacturer’s Declaration: .................................................................................................. 50

Page 7: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 7/52

1 Description

The iS torque measurement systems are representing a further development of the new generation of

torquemeters with integrated evaluation unit. With the exception of a 24VDC power supply, no external

components are required for operation.

High-end temperature compensation guarantees very good stability and repeatability of the output signals.

The standard model is equipped with an inductive one track speed measurement system.

For your convenience the torquemeters of the iS series are interchangeable with the same stator.

Specifications

Power Supply 24V DC max. 1A

Sample Rate Torque 1000 Samples / Second

Sample Rate Speed f > 1000 Hz 1000 Samples / Second

Sample Rate Speed f < 1000 Hz f / 2 Samples / Second

Lowest Frequency, which can be

measured

5Hz

(the output for frequencies <5Hz is 0Hz)

Analog Output Range selectable 0..5V, 0..10V, -5..+5V, -10..+10V

Analog Output Signal Resolution 16 bit

Analog Output Impedance 50 Ohm

Optional Current Output (Torque) selectable 4..20mA, 0..20mA

Filter Torque: 1st-order IIR-Filter with 6 fixed cut-off

frequencies

Speed: Moving Averager with adjustable filter depth

CAN Interface CAN2B

Identifier free adjustable

max. 1MBaud

max. 1000 messages/channel/second

Serial port RS232, 19200 Baud, 8 Data Bit, No Parity Bit,

1 Stop Bit, No Protocol

Frequency outputs RS422

Torque

Inductive speed sensor

Magnetic speed sensor (optional)

Page 8: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 8/52

2 Safety Instructions

Before setting-up operation, maintenance or additional handling of the measurement system, it

is important to review the following:

Review any safety procedures including the operation manual.

Check that all protective devices are installed and functioning properly.

Note on additional standards:

Low Voltage Directive 73/23/EWG, Electromagnetic Compatibility

Directive 89/336/EWG and the harmonized standards

DIN EN 292-1 Safety of machinery

DIN EN 292-2 Safety of machinery

Maintenance and inspections on the electrical equipment are to be executed by trained

personal. Non-designated use and modifications of the measurement system will make the

EG-Conformity declaration invalid.

Page 9: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 9/52

3 Introduction

In this manual you will find all steps which are necessary to start-up the ATESTEO torque and speed

measurement system.

This manual is usable for the following types of torque meters:

Torquemeter F1iS

Torquemeter F2iS

Torquemeter F1i

Torquemeter F2i

Torquemeter FLFM1iS

Torquemeter FLFM1eS

All measurement systems work contactless and are maintenance-free. The data transmission is realized by a

frequency modulated infrared transmitter. The power of the rotating electronic module works inductive.

Page 10: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 10/52

3.1 One-channel telemetry (FM) F1i/F2i

Features:

frequency output proportional to torque 60 kHz 20 kHz

analogue output [V] proportional to torque with 1000 readings/s

frequency output proportional to speed

analogue output [V] proportional to speed

shunt – calibration

Zero adjustment automatically via remote control

serial Interface RS232

CAN 2B interface

Page 11: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 11/52

T

D

(temperature values)

16 bit

EEPROM

D D

F

F

A A

+V

IR-TRANSMISSION 1

IR-TRANSMISSION 2

+V

3.2 Dual -Telemetry DT (FM/ FM) F1i/F2i

The overall system is created to integrate a second amplifier and a second IR-telemetry during the

production of the torque meter. This second amplifier amplifies the signal from the strain gage bridge with a

very high accuracy. The result is a second range of highly precise measurement of small torques.

Consequently, the often necessary replacement of the torque sensor for the highly precise measurement of

less torques can be dropped. This second measuring system also includes a temperature compensation and

a shunt calibration as in the first one.(see illustration).

To exploit the total measuring preciseness of the minor measuring range one must

consider, to stop and unload the power train after each measuring cycle, which takes place

under a high torque load. Afterwards the systems needs to be reset to "0" otherwise the

hysteresis figures, recorded in the spring body, would overlap

the more sensitive second measuring channel.

With the Dual –Telemetry system (FM/FM) it is possible to measure with one torquemeter high and small

torques with a high accuracy.

Page 12: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 12/52

Features:

2 x frequency output proportional to torque 60 kHz 20 kHz

frequency output proportional to speed

analogue output [V] proportional to torque with 1000 readings/s

analogue output [V] proportional to speed

shunt – calibration

Zero adjustment automatically via remote control

System-parameter via RS232.

CAN 2B interface

Page 13: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 13/52

4 Mechanical Installation

4.1 Part list

4.1.1 Part list F1iS/F2iS

The complete measuring equipment consists of the following parts:

Torquemeter (Rotor)

Stator All-In-One (Evaluation unit)

Connectors 16 pole male / 12 pole female

4.1.2 Part list F1i/F2i

Complete measuring equipment consists of following parts:

Torquemeter (Rotor)

Stator – top and lower part

Connectors 10/ 12 pole

4.1.3 Part list FLFM1iS

The complete measuring equipment consists of the following parts:

Torquemeter (Rotor)

Stator All-In-One (Evaluation unit)

Connectors 16 pole male / 12 pole female

4.1.4 Part list FLFM1eS

Torquemeter (Rotor)

Stator FLFM1eS

TCU 2 (Torque Control Unit)

Page 14: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 14/52

4.2 Installation F1iS / F2iS

The following figure shows the correct offset between the end faces of the dynamometer-flange and the

dynamometer-housing.

F1iS / F2iS

Type Nominal data S [mm]

F1iS +1 ... -2 … -5 without magnetic speed encoder

F2iS -1 … -4 … –7 without magnetic speed encoder

F1iS -1,5 …-2 … -2,5 with magnetic speed encoder

F2iS -3,5 ... -4 … –4,5 with magnetic speed encoder

Height of centring for the stator

Type A [mm]

F1iS less then: 2

F2iS less then: 4

Housing

E-dyno

Flange E-dyno

+ -

Page 15: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 15/52

4.2.1 Mounting the Stator

If it is possible, mount the stator of the torquemeter with the electronic-housing aligned to a 9 o'clock

position. In this case no liquid can concentrate in the housing. As a precaution the electronic components are

sealed with protective paint.

1. 2.

3. 4.

Page 16: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 16/52

4.2.2 Mounting of the foot base

Example for a complete mounted system:

Page 17: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 17/52

52,5

7,5

215

(60) 100

Adapter flange

Angle sheet for foot mounting base

4xM6

Page 18: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 18/52

4x bolt DIN912 (Hexagon socket) M6

Page 19: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 19/52

Torquemeter - flange

8x bolt DIN912 (Hexagon socket) M12x40 (Length depends on thickness of customers adapter flanges)

Page 20: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 20/52

4.2.3 Mounting with the foot base

Example for a complete mounted system:

Page 21: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 21/52

58

12

260

(70)

4xM8 100

Angle sheet for foot

mounting base

Adapter flange

Page 22: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 22/52

4.2.4 Mounting the Optional Magnetic Speed Detection

See manual electrical installation…

6.

Page 23: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 23/52

4.2.5 Recommended fastening torques for screws

F1iS

Rated Torque

Nm

PCD

130 mm

Quality Quantity Torque

Nm

up to 2.500 8x M12 12.9 8 135

up to 2.100 8x M12 10.9 8 115

up to 1.400 8x M12 8.8 8 77

F2iS

PCD

196 mm

up to 15.000 16x M16x1,5

fine thread

12.9 16 380

up to 10.000 16x M16 12.9 16 330

up to 9.000 16x M16 10.9 16 280

up to 6.000 16x M16 8.8 16 190

Page 24: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 24/52

4.3 Installation F1i/F2i torquemeter

Order of assembly F1i / F2i torquemeter:

1 Statorring

2 Torquemeter

3 Stator top part

Page 25: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 25/52

The following figure shows the correct offset between the end faces of the dynamometer-flange and the

dynamometer-housing.

F1i and F2i

Typ Nominal data S [mm]

F1 / F1i -1,5 .. 0 .. 0,2

F2 / F2i -1,5 .. 0 .. 0,5

Housing

E-dyno

Flange E-dyno

+ -

Page 26: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 26/52

4.3.1 Mounting the Stator

If it is possible place the stator of the torque meter F1i / F2i so that the housing of the electronic is adjusted

at the left side. In this case no liquid can be collect in the stator. The hole electronic is filled with a potting

component and can’t damaged by liquid.

4.3.2 Recommended starting torques for screws

F1/F1i

Rated Torque

Nm

PCD Quality Quantity Torque

Nm

up to 2.500 8x M12 12.9 8 135

up to 2.100 8x M12 10.9 8 115

up to 1.400 8x M12 8.8 8 77

F2/F2i

up to 15.000 16x M16x1,5

fine thread

12.9 16 380

up to 10.000 16x M16 12.9 16 330

up to 9.000 16x M16 10.9 16 280

up to 6.000 16x M16 8.8 16 190

Page 27: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 27/52

4.4 Installation FLFM1iS

Order of assembly: FLFM1iS torquemeter

1.Stator

2.Torquemeter

The following figure shows the correct offset between the end faces of the dynamometer-flange and the

dynamometer-housing.

FLFM1iS

Type Nominal data S [mm]

FLFM1iS +7,7 ... +8,0 … +8,5

Housing E-dyno

Flange E-dyno

+ -

S

Page 28: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 28/52

4.4.1 Mounting the Stator

The stator of the torquemeter must be mounted with the electronic-housing aligned to a 9 o'clock position. In

this case no liquid can concentrate in the housing. As a precaution the electronic components are sealed

with protective paint.

Page 29: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 29/52

4.4.2 Mounting the torquemeter

4.4.3 Mounting the optical speed detection

Page 30: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 30/52

4.4.4 Mounting the stator with a foot base

By using the optical speed detection it is not recommended, to mount the stator through foot mounting base.

The alignment (rotor – stator) must be very exact. Otherwise it is possible, that there arise unwanted contact

between rotor and stator.

Example for a complete mounted system:

Page 31: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 31/52

4x bolt DIN912 (Hexagon socket) M8

0,5

59,5

(60)

189

±0,05

Angle sheet for foot mounting base 4xM8

Adapter flange

100

Page 32: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 32/52

Page 33: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 33/52

4.4.5 Mounting the torquemeter

4.4.6 Recommended fastening torques for screws

8x bolt DIN912 (Hexagon socket)

M10x30 (Length depends on thickness of customers adapter flanges)

Page 34: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 34/52

FLFM1iS

Rated Torque

Nm

PCD

75 mm

Quality Quantity Torque

Nm

Up to 500 8x M10 8.8 8 50

Up to 800 8x M10 10.9 8 70

Up to 1.000 8x M10 12.9 8 80

Page 35: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 35/52

4.4.7 Mounting the optical speed detection

Page 36: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 36/52

4.1 Installation FLFM1eS

Order of assembly FLFM1eS torquemeter:

1.Stator

2.Torquemeter

3. Electronics

The following figure shows the correct offset between the end faces of the dynamometer-flange and the

dynamometer-housing.

FLFM1iS

Type Nominal data S [mm]

FLFM1eS 7,7 ... 8,0 … 8,5

Housing E-dyno

Flange E-dyno

+ -

S

S

Page 37: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 37/52

4.1.3 Mounting the Stator

The stator of the torquemeter must be mounted with the electronic-housing aligned to a 9 o'clock position. In

this case no liquid can concentrate in the housing. As a precaution the electronic components are sealed

with protective paint.

8x bolt DIN912 (Hexagon socket)

M10x40 (Length depends on thickness of customers adapter flanges)

1. 2.

3

.

4

.

Page 38: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 38/52

4.1.4 Mounting the optical speed detection

4.1.5 Recommended fastening torques for screws

FLFM1eS

Rated Torque

Nm

PCD

75 mm

Quality Quantity Torque

Nm

Up to 500 8x M10 8.8 8 50

Up to 800 8x M10 10.9 8 70

Up to 1.000 8x M10 12.9 8 80

Page 39: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 39/52

4.1.6 Electronics

4.1.7 Components earthing

Please take care of the TCU2 housing, it has to have the same potential with the stator ring. The machine

parts are often lacquered, so we advice to set up additional electrical connection between the both

components

Page 40: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 40/52

5 Speed measuring system

5.1 Speed measuring system F1iS / F2iS

5.1.1 Inductive Sensor

The inductive speed sensor is equipped as standard and supplies one track with 60 increments per round at

the torquemeter F1iS and 120 increments per round at the torquemeter F2iS. It is located at the inner side of

the stator ring.

The air-gap between rotor and stator is factory calibrated to an operational distance of 0.5 up to 2.5 mm and

needs no readjustment.

The speed-sensor signals are provided as RS422 signals and as processed values for the analog outputs

and CAN messages. The RS422 signals can be accessed by connector X750.

Inductive

Sensor

Page 41: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 41/52

5.1.2 Magnetic Sensor (F1iS: 1000 increments / F2iS:1448 increments)

The magnetic speed sensor is available as an optional high-resolution speed acquisition providing 2 tracks

with 1000 increments per round at F1iS and accordingly 1448 increments per round at F2iS

and a 90° phase shift, thus giving the capability for detecting the rotational direction. It is located on a

mounting bracket placed above the electronic compartment. The magnetic speed sensor consists of a

sensor module which is connected via a 7 pole connector to the stator electronics.

The speed-sensor signals are provided as RS422 signals and as processed values for the analog outputs

and CAN messages. The RS422 signals can be accessed by connector X750, output pins 3 and 4 for track 1

and pins 5 and 6 for track 2 (refer to chapter „Electrical specifications“).

Page 42: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 42/52

5.1.3 Adjustment of the right distance:

To adjust right distance lossen 2 screws like shown in the picture below. The distance should be ideally

0,5mm.

F1iS -> 0,5mm (tolerance ±0,4mm)

F2iS -> 0,5mm (tolerance ±0,4mm)

Page 43: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 43/52

5.2 Speed measuring system F1/F1i (480 increments):

5.2.1 Adjustment of speed measuring for torquemeter F1/F1i:

After each assembly of the torquemeter housing at a new machine the distance between the speed module

and the impulse generator wheel must be checked and if necessary adjust.

The optimum air gap between the speed module and the impulse generator wheel is 0.3mm.

The air gap tolerance is ±0,2mm. Additional adjustments are not required.

F1iS -> 0,3mm (tolerance ±0,2mm)

5.2.2 Adjustment of the right distance:

At first unscrew the internal hexagon screw (fixing screw).

Now it is possible to move the speed module radial so that the distance between the speed module and the

impulse generator wheel can vary.

With the special shim plate the difference can be adjust to the nominal dimension. If the difference is correct,

fix the screw and the adjustment is done.

Speed modul

Air gap

3x

fixing screws

Page 44: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 44/52

ATESTEO- shim plate for speed module F1/F1i

Torquemeter F1

For a first check turn the torquemeter by hand.

shim plate

Page 45: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 45/52

5.3 Speed measuring system FLFM1iS, FLFM1eS

5.3.1 Optical speed detection

The optical speed sensor is available as an optional high-resolution speed acquisition providing 2 tracks with

alternate 240, 360 or 400 increments per round and a 90° phase shift, thus giving the capability for detecting

the rotational direction. It is located at the inner side of the stator-ring.

The speed-sensor signals are provided as RS422 signals and as processed values for the analog outputs

and CAN messages. The RS422 signals can be accessed by connector X750 (refer to chapter „Electrical

specifications“).

Optical speed disc

Page 46: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 46/52

5.3.2 Setup the optical speed

With every modification (transformation) of the measuring flange housing into a new machine the optical

speed sensing has to be newly balanced. This is important to do, because changing the distance between

speed disk and cover of receiver side causes a change of signal amplitude. Moreover, the electrical

properties can vary in different modules.

To setup the signal processing board you need to use a oscilloscope.

For balancing do the following:

Remove the cover of the signal processing unit.

To carry out the adjustment procedure the measuring points have to be measured with oscilloscope.

At MP1 till MP4 there will be measured 4 voltage signals from the IR receiver (transistor).

With the speed there are sinusoidal signals, wherein MP1 to MP2 and MP3 to MP4 have 180° phase

displacement and the both pairs have 90° to each other phase displacement.

For the basic settings it will be sufficient to turn the measuring flange by hand during the measurement.

The aim of the setting is that all speed signals provide output sinusoidal signals with approximately equal

amplitude in the range of about 0,8V – 6V.

With a potentiometer TRM100 the electric current will be set, which flows through the IR transmitter diodes.

With the current rising the amplitude of the AC voltage increases and the DC voltage of the speed signals

decreases (MP1 till MP4).

The current has to be so set, so that a minimal voltage would be in MP1 till MP4 0,8V till 1V. The sinusoidal voltage signal must not be cut off in upper and lower max. point.

GND

MP1

MP2

MP3

MP4

TRM100

Page 47: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 47/52

The amplitude has to be so set, so that the sending current could be so high as it

is possible. The sinusoidal voltage signals of the speed signals have then the

smallest DC offset.

Page 48: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 48/52

5.3.3 Speed measuring adjustment

The optical speed mask position at the speed module is factory adjustment and needs no readjustment

Additional adjustment is not required.

Speed Module for Torquemeter FLFM1iS, FLFM1eS

Optical

speed mask 4xIR transmitting diodes

Page 49: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 49/52

The optimum position of the speed measuring system will be determined by positioning rotor to stator.

Following masses have to be observed.

For a first check turn the torquemeter by hand.

A: axial displacement

Rotor-Stator

B: axial deflection

without optical speed sensing 28... 28,3...29,5 ±0,3 mm

with optical speed sensing 28,3...28,5...29 ±0,3 mm

Page 50: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 50/52

6 EC Manufacturer’s Declaration:

Page 51: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 51/52

Page 52: Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS · Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 3/52 Dear customer, before leaving our company every unit

ATESTEO GmbH

Mechanical Installation F1iS/F2iS, F1i/F2i, FLFM1iS, FLFM1eS 52/52

Sie möchten mehr über unsere Produkte, Lösungen und Services aus den Bereichen Messsysteme,

Fahrzeugausrüstung und Aktuatoren erfahren? Dann rufen Sie uns einfach an unter +49 (0) 2404 9870570

oder mailen Sie uns an [email protected]. Ihr persönlicher ATESTEO Ansprechpartner ist gern für

Sie da.

Want to learn more about our products, solutions and services in the fields of measurement systems, vehicle

equipment and actuators? Then please call us under +49 (0) 2404 9870 570 or email us at

[email protected]. Your personal ATESTEO contact is always at your disposal.