interphalangeal arthrodesis of the toe with a new ... · rocher h., guillo s., averous c.,...
TRANSCRIPT
ROCHER H., GUILLO S., AVEROUS C., CERMOLACCE C., DETERME P., DIEBOLD P.-F.
Interphalangeal Arthrodesis of the Toe with a New Radiolucent Intramedullary Implant (TOEGRIP®)
DIEBOLD P.-F., ROCHER H., , DETERME P., CERMOLACCE C., GUILLO S., AVEROUS C., LEIBER WACKENHEIN F.
Interphalangeal Arthrodesis
of the Toe with a New Radiolucent
Intramedullary Implant
ROCHER H., GUILLO S., AVEROUS C., CERMOLACCE C., DETERME P., DIEBOLD P.-F.
Interphalangeal Arthrodesis of the Toe with a New Radiolucent Intramedullary Implant (TOEGRIP®)
Interphalangeal Arthrodesis of the Toe with a
New Radiolucent Intramedullary Implant
My disclosure is in the Final AOFAS Mobile App.
I have no potential conflicts with this presentation.
DIEBOLD P.-F.
ROCHER H., GUILLO S., AVEROUS C., CERMOLACCE C., DETERME P., DIEBOLD P.-F.
Interphalangeal Arthrodesis of the Toe with a New Radiolucent Intramedullary Implant (TOEGRIP®)
Background: Interphalangeal arthrodesis is a very common surgical treatment of
rigid hammer toe and claw toe deformities. The K-wires habitually used in this
procedure are sometimes complicated by pin tract infection, migration, discomfort
and breakage. The aim of this study is to evaluate the results of an interphalangeal
arthrodesis with a new radiolucent angulated intramedullary implant.
Methods: 377 implants were placed in 297 patients between October 1st 2011 and
October 1st 2012. In this study 157 patients had more than one year follow up and
142 patients were reviewed. The operation technique is explained in detail.
Results: This intramedullary device offers a good immediate mechanical stability,
adequate deformity correction, and a high rate of consolidation (83%). The
satisfaction rate of patients in this series was 94%.
Conclusions: This study demonstrates good results with a new generation of
radiolucent implants.
ABSTRACT
ROCHER H., GUILLO S., AVEROUS C., CERMOLACCE C., DETERME P., DIEBOLD P.-F.
Interphalangeal Arthrodesis of the Toe with a New Radiolucent Intramedullary Implant (TOEGRIP®)
This retrospective study with one year follow-up reviews 180 implants in 142 patients.
Mean age at the time of surgery was 61 years (range 28-85 years).
There were 128 patients with a stiff and flext PIP and 14 patients with a too long second toe
in front of the first toe pathology. Surgeries involved 142 second toe (79%), 30 third toe
(17%), 7 fourth toe (4%) and only in one case the fifth toe. In 70% of the patients, surgical
treatment of associated pathology of the first ray (hallux valgus, hallux rigidus) was needed.
Eleven cases were a revision surgery after a failed operation. A rigid PIP joint was the
indication of surgery in 90% of the cases. For 10% the joint was flexible and the joint
arthrodesis was dictated by an excessive toe length.
The 297 patients were operated by senior surgeons trained in foot surgery. Surgeries involved 142 second toe (79%), 30 third toe (17%), 7 fourth toe (4%) and only in one case the fifth toe. In 70% of the patients, surgical treatment of associated pathology of the first ray (hallux valgus, hallux rigidus) was needed. Eleven cases were a revision surgery after a failed operation. A rigid PIP joint was the indication of surgery in 90% of the cases. For 10% the joint was flexible and the joint arthrodesis was dictated by an excessive toe length.
MATERIAL AND METHODS
ROCHER H., GUILLO S., AVEROUS C., CERMOLACCE C., DETERME P., DIEBOLD P.-F.
Interphalangeal Arthrodesis of the Toe with a New Radiolucent Intramedullary Implant (TOEGRIP®)
The implant : Toegrip
ROCHER H., GUILLO S., AVEROUS C., CERMOLACCE C., DETERME P., DIEBOLD P.-F.
Interphalangeal Arthrodesis of the Toe with a New Radiolucent Intramedullary Implant (TOEGRIP®)
A mid-dorsal incision of the proximal interphalangeal (PIP) joint is performed, removing a roughly wedge-shaped piece of the hyperkeratosis. The extensor apparatus is then incised. It is very important to completely free the last centimeter of the first phalanx (P1) and the base of the second phalanx (P2). The preparation must release the collateral ligaments, medial and lateral and the plantar and dorsal capsule. The condyles of the first phalanx (P1) are cut with an oscillating saw to avoid any breakage. The cutting line followed the angle of the implant. P1 is prepared first. A 2 mm pin is used to mark the center of the intramedullar canal. Preparation is continued using reamers with increasing size until an optimal cortical fit is found. The last reamer defines the final size of the implant. The angle of the implant is defined by the positions of the adjacent toes. The longer part of the implant is inserted into the proximal phalanx with the appropriate holder. The holder is removed and the middle phalanx is positioned over the shorter part of the implant to ensure a close contact between the two bone sections. We designed a special “fork” extremely helpful for the reduction. Bone contact is checked by freeing the extensor apparatus, if necessary, and avoiding any rotation of the toe. The extensor tendon is then sutured, followed by the skin. A tenotomy at the level of the MP joint may also be performed to prevent a toe-to-ground contact insufficiency, tested by a push up test.
SURGICAL TECHNIQUE
ROCHER H., GUILLO S., AVEROUS C., CERMOLACCE C., DETERME P., DIEBOLD P.-F.
Interphalangeal Arthrodesis of the Toe with a New Radiolucent Intramedullary Implant (TOEGRIP®)
After one year 142 patients were available for follow up. 133 patients (94%) were satisfied, 137 patients (97%) had no pain and 5 patients (3%) reported a moderate pain. The VAS score was O.1 (0 to 3) really improved from the pre operating statement. 10% of the toes had some swelling, this was significant in 1% of the toes. On the one year x-ray control, 83% of the PIP arthrodesis were completely fused
We noted a radiolucency around the implant in 5% of the cases. The function of the toes was estimated normal by 97% of the patients with a correct pulp ground contact. All the patients estimated that the toes were fitting better in the shoewear with no limitation in walking ability. There was no recurrency of the claw toe. The axis of the toe was estimated as good in 97% sagitally or frontally.
RESULTS
ROCHER H., GUILLO S., AVEROUS C., CERMOLACCE C., DETERME P., DIEBOLD P.-F.
Interphalangeal Arthrodesis of the Toe with a New Radiolucent Intramedullary Implant (TOEGRIP®)
fibrous pseudarthrosis aligned (17%) IPP fused (83%)
ROCHER H., GUILLO S., AVEROUS C., CERMOLACCE C., DETERME P., DIEBOLD P.-F.
Interphalangeal Arthrodesis of the Toe with a New Radiolucent Intramedullary Implant (TOEGRIP®)
The intramedullary implants offer several advantages. The absence of exposed hardware minimizes the risk of digital infection. No removal of hardware is needed which decreases anxiety and improves satisfaction. Because hardware is retained, there is far less recurrence of the hammertoe deformity or other misalignments (25). Some breakages were noted on x-rays which are moreover at the origin of patients’ complaints. Although Alvine and Garvin reported a fusion rate of 97% in a series of seventy-three toes, they had an 87% patient satisfaction (26). Ellington published the results of a 2-piece intramedullary fusion device (24). They reported a complication rate of 55,3%, including nonunions, with 3 hardware failures and 3 intraoperative fractures. The revision rate was 7,9%. An intramedullary implant decreased the possible malrotation position of the toe which is not controlled by a pin fixation (34). No infections developed in this toegrip serie. No breakage of the implant occurred. None of the patients had major complaints about transfer metatarsalgia, floppiness of the toes. None of the patients demonstrated malrotation.
AVANTAGES
ROCHER H., GUILLO S., AVEROUS C., CERMOLACCE C., DETERME P., DIEBOLD P.-F.
Interphalangeal Arthrodesis of the Toe with a New Radiolucent Intramedullary Implant (TOEGRIP®)
Lesser toe deformities are very common and the surgical management of these deformities are comprised between 28% and 46% of all forefoot surgery (28). PIP joint arthrodesis is a good solution for the treatment of toe PIP deformities. The implant used in this series represents a new generation of intramedullar devices offering good immediate mechanical stability, adequate deformity correction, and a high rate of consolidation with no short-term or long-term tolerance issues. In addition it can be easily used in revision surgery.
CONCLUSION
ROCHER H., GUILLO S., AVEROUS C., CERMOLACCE C., DETERME P., DIEBOLD P.-F.
Interphalangeal Arthrodesis of the Toe with a New Radiolucent Intramedullary Implant (TOEGRIP®)
1. Scheck M. Etiology of acquired hammertoe deformity Clin Orthop 123-63, 1977. 1. Mickle KJ, Munro BJ, Lord SR, Menz HB, Steele JR, ISB Clinical Biomechanics Award 2009: toe weakness
and deformity increase the risk of falls in older people. Clin Biomech (Bristol, Avon) 10:787-791, 2009. 1. Edwards WH, Beischer AD, Interphalangeal joint arthrodesis of the lesser toes. Foot Ankle Clin 1:43-48,
2002. 1. Femino JE, Mueller K. Complications of lesser toe surgery. Clin Orthop Relat Res. 391:72-88, 2001. 1. Harmonson JK, Harkless LB. Operative procedures for the correction of hammertoe, claw toe, and mallet
toe: a littérature review. Clin Podiatr Med Surg. 2:211-220, 1996. 1. Alvine FG. Amputation surgery – the immediate fit prostheses. S D J Med. 7:7-11, 1980. 1. Coughlin MJ. Lesser toe deformities. Orthopedics 1:63-75, 1987.
REFERENCES (1/2)
ROCHER H., GUILLO S., AVEROUS C., CERMOLACCE C., DETERME P., DIEBOLD P.-F.
Interphalangeal Arthrodesis of the Toe with a New Radiolucent Intramedullary Implant (TOEGRIP®)
8. Konkel KF, Menger AG, Retzlaff SA. Hammer toe correction using an absorbable intramedullary pin. Foot Ankle Int. 8:916-20, 2007.
8. Lamm BM, Ribeiro CE, Vlahovic TC, Fiorilli A, Bauer GR,Hillstrom HJ. Lesser proximal interphalangeal
joint arthordesis: a retrospective analysis of the ped-in-hole and end-to-end procedures. J Am Podiatr Med Assoc. 7:331-6, 2001.
8. Lehman DE, Smith RW. Treatment of symptomatic hammertoe with a proximal interphalangeal joint
arthrodesis. Foot Ankle Int. 9:535-41, 1995. 8. Pietrzak WS, Lessek TP, Perns SV. A bioabsorbable fixation implant for use in proximal interphalangeal
joint (hammertoe) arthrodesis: Biomechanical testing in a synthetic bone substrate. J Foot Ankle Surg. 4555:288-94, 2006.
8. Shaw AH, Alvarez G. The use of digital implants for the correction of hammertoe deformity and their
potential complications and management. J Foot Surg. 1:63-74, 1992. 8. Taylor RG. The treatment of claw toes by multiple transfers of flexor into extensor tendons. J Bone Joint
Surg Br. B4:539-42, 1951. 8. Baig a, Geary N. fution rate and patient satisfaction in proximal interphalangeal joint fusion of the minor
toes using kirschner wire fixation. The foot. 6:120-1, 1996.
REFERENCES (2/2)