improved fluoroimmunoassays using the dye alexa fluor 647 with the raptor, a fiber optic biosensor
TRANSCRIPT
Improved fluoroimmunoassays using the dye Alexa Fluor 647
with the RAPTOR, a fiber optic biosensor
George P. Anderson*, Nandan L. Nerurkar
Center for Bio/Molecular Science and Engineering, U.S. Naval Research Laboratory, 4555 Overlook Avenue S.W. Code 6900,
Washington, DC 20375-5438, USA
Received 9 May 2002; received in revised form 18 July 2002; accepted 22 July 2002
Abstract
The performance of the fluorescent dye Alexa Fluor 647 (AF647) was explored as an alternative to Cy5 for immunoassays
on the RAPTOR, a fiber optic biosensor. The RAPTOR performs sandwich fluoroimmunoassays on the surface of small
polystyrene optical waveguides for analyte detection. Fluorescence and immunoassay data were examined at various dye-to-
protein (D/P) ratios for both Cy5 and Alexa Fluor 647. Primarily, due to the self-quenching characteristics of Cy5, Alexa Fluor
647 is substantially more effective in fluoroimmunoassays, yielding over twice the signal for any given analyte concentration.
Alexa Fluor 647 can be attached to antibodies at higher ratios, D/P= 6, before self-quenching begins to limit the dye’s
effectiveness. Furthermore, while Alexa Fluor 647 becomes quenched at high dye-to-protein ratios, D/P= 9, the net
fluorescence yield reaches a maximum, as opposed to Cy5-labeled proteins, which become nearly nonfluorescent at high
labeling ratios, D/Pz 6. The limitations of Cy5 were elucidated with an immunoassay for ricin, while the advantages of Alexa
Fluor 647 were demonstrated in both direct binding assays as well as in a sandwich immunoassay for staphylococcal
enterotoxin B.
D 2002 Elsevier Science B.V. All rights reserved.
Keywords: Alexa Fluor 647; Cy5; Biosensor; Immunoassay; Fluorophore
1. Introduction
Alexa Fluor 647 (AF647) is a relatively new
fluorescent dye, which has peak excitation at 650
nm and peak emission at 665 nm. It has a molecular
weight of f 1300 and an extinction coefficient of
203,000 cm � 1 M � 1. These properties are similar to
the more commonly employed dye Cy5, whose max-
imum excitation is 649 nm, maximum emission is 670
nm, has a molecular weight of 975, and an extinction
coefficient of 250,000 cm� 1 M � 1. AF647 had been
purported to be more photostable than Cy5 dye
(Panchuk-Voloshina et al., 1999), but has not been
extensively examined for use in fluoroimmunoassays.
While there are a number of fluorescent dyes in the
Alexa Fluor series, only the AF647 was tested, being
the one most compatible for use with the current
excitation/emission requirements (635 nm/>665 nm)
0022-1759/02/$ - see front matter D 2002 Elsevier Science B.V. All rights reserved.
PII: S0022 -1759 (02 )00327 -7
* Corresponding author. Tel.: +1-202-404-6033; fax: +1-202-
767-9594.
E-mail address: [email protected]
(G.P. Anderson).
www.elsevier.com/locate/jim
Journal of Immunological Methods 271 (2002) 17–24
of the RAPTOR fiber optic biosensor (Fig. 1; Ander-
son and Rowe-Taitt, 2001; Anderson et al., 2000;
King et al., 2000). The RAPTOR is a portable,
automated biosensor that, like its predecessor the
Analyte 2000, is useful for on-site analysis of food,
water, or clinical samples for biological contaminants
(Anderson et al., 1998; Cao et al., 1995; DeMarco et
al., 1999; Ligler et al., 1993; Tempelman et al., 1996).
Thus, we were interested in determining what, if any,
advantage lies in using this new dye to prepare the
fluorescent immuno-reagents for the RAPTOR.
The quenching characteristics of Cy5 at various
dye-to-protein (D/P) ratios have been described
previously by Gruber et al. (2000). This self-
quenching is thought to be caused by formation
of Cy5 dimers; the conjecture being that hydro-
phobic interactions between Cy5 molecules result in
nonrandom binding to the protein surface, and once
immobilized, these deleterious interactions are only
enhanced.
We first set out to confirm those results and to
investigate how using Cy5-labeled antibody at various
D/P ratios impacted the performance of the RAPTOR.
In addition, we wanted to test Cy5-labeled antibody at
its optimal D/P ratio to provide a frame of reference
for judging the effectiveness of AF647. The primary
goal of this study was to examine if another dye,
namely AF647, could significantly increase the over-
all fluorescent signal generated during both direct and
sandwich immunoassays, and thereby improve the
overall limits of detection.
2. Materials and methods
2.1. Buffers and reagents
Ricin was purchased from Sigma (St. Louis, MO).
Antiricin antibodies were provided by Naval Medical
Research Command, Silver Spring, MD. Staphylo-
coccal enterotoxin B (SEB) and affinity-purified
sheep anti-SEB IgG were purchased from Toxin
Technology (Sarasota, FL). RAPTOR wash buffer
consisted of phosphate buffer (Sigma, 8.3 mM, pH
7.3) containing 0.05% (v/v) Triton X-100 (TX-100)
and 0.01% (w/v) sodium azide (PBT). The fluores-
cently labeled antibodies were diluted into a blocking
buffer, which consisted of PBT plus 1 mg/ml casein, 1
mg/ml bovine serum albumin (BSA), and 10 mM
sucrose.
2.2. Waveguide and coupon preparation
Capture antibodies were immobilized onto fiber
optic waveguides by passive adsorption. Injection-
molded, polystyrene fiber optic waveguides (Resear-
ch International, Woodinville, WA) were first black-
ened at their distal ends to prevent reflection of
excitation light. Waveguides were then placed into
capillary tubes (100 Al, cut to 4 cm length) pre-
filled with 38 Al of the appropriate antibody sol-
ution, in general, 100 Ag/ml IgG in 0.1 M sodium
carbonate buffer (pH 9.6). After overnight incuba-
tion at 4 jC, the waveguides were rinsed with
distilled water and glued into the disposable cou-
pons (visible in Fig. 1, lower left). All coupons
were made with three identical waveguides coated
in with captured antibody, and a fourth serving as a
control.
2.3. Preparation of Cy5-labeled rabbit antiricin IgG
Rabbit antiricin IgG was labeled with bisfunctional
NHS-ester Cy5 dye (Amersham, New Jersey). One
vial of Cy5 dye was diluted to 100 Al with water. The
dye was portioned out into 10, 30, and 60 Al aliquots,and each reacted with 300 Ag of antibody in 50 mM
Fig. 1. RAPTOR fiber optic biosensor with optical waveguide
coupon in the foreground.
G.P. Anderson, N.L. Nerurkar / Journal of Immunological Methods 271 (2002) 17–2418
sodium tetraborate, 40 mM NaCl, pH 9.0. The reac-
tions proceeded for 30 min at room temperature in the
dark. Subsequently, labeled protein was separated
from unincorporated dye by size-exclusion chroma-
tography.
2.4. Preparation of AF647-labeled IgG
Alexa Fluor 647 NHS ester (1 mg) was pur-
chased from Molecular Probes (Eugene, OR). The
dye was dissolved in 500 Al of anhydrous acetoni-
trile and divided into 20 Al aliquots, then dried
using an Eppendorf spin vac. These dye aliquots
were sealed under nitrogen and stored at � 20 jCwith desiccant until use.
AF647-conjugated antibodies were prepared at
various D/P ratios. One aliquot of AF647 dye
(f 40 Ag) was first dissolved with 10 Al of DMSO,
then diluted to a final volume of 100 Al with water.
Different volumes of dyes (10, 30, or 60 Al) were
added to 300 Ag IgG (2 mg/ml final concentration) in
50 mM sodium tetraborate, 40 mM NaCl, pH 9.0. The
reaction progressed for 30 min at room temperature in
the dark. Then the labeled protein was separated from
unincorporated dye by size-exclusion chromatogra-
phy. The absorbance at 280 and 650 nm for each
antibody was determined and the D/P ratios were
calculated. This protocol produced goat IgG labeled
at D/P ratios of 2.5, 6.3, and 8.4, respectively. Other
antibodies were labeled with AF647 in a similar
manner; however, the amount of dye required to
obtain similar D/P ratios varied from antibody to
antibody. Most antibodies required f 120 Ag dye/
mg protein to obtain D/P ratios near 6. The fluores-
cence of the labeled antibodies was measured using a
Perkin Elmer LS3 Fluorimeter with a 0.25-cm path-
length cuvette.
2.5. RAPTOR design and operation
Insertion of the assay coupon into the RAPTOR
coupon compartment aligns all necessary optical paths
and engages all fluidic connections required for sam-
ple analysis. A pneumatic pump moves buffer, air,
fluorescent reagent (from on-board reservoirs), or
sample within the system. Serpentine channels in the
coupon provide a common path across the waveguide
surfaces. Bubble detectors, which monitor liquid-to-
air interfaces, control introduction of sample and
fluorescent reagent.
During the two-step sandwich immunoassay, sam-
ple is flowed over the four fiber optic waveguides
mounted in the assay coupon. Antigen present in the
sample binds to the fiber optic waveguide coated with a
corresponding capture antibody; unbound material is
washed away by a brief rinse with PBT. Fluorescently
labeled antibody is next introduced and binds to the
antibody–antigen complexes on the waveguide sur-
face, completing the sandwich assay. This tracer
reagent is maintained at a suitable temperature in an
onboard thermal storage module and recovered after
each assay cycle, allowing multiple sequential analyses
to be performed. Excitation light from four 5 mW
Sanyo laser diodes (635 nm) within the RAPTOR is
focused into the fiber optic waveguides. An evanescent
wave is created along each waveguide, exciting the
fluorescent emission of specifically bound fluoro-
phore-labeled antibodies. The portion of the fluores-
cence captured by the optical waveguide is collimated
by the waveguide’s molded lens and focused onto a
photodiode using a ball lens, chosen for its light-
gathering power and short focal length. A long-pass
dichroic filter (665 nm) rejects reflected laser light.
2.6. Assay procedure
Two RAPTOR assay formats were used, the sand-
wich assay, for which the instrument was designed,
and a simple direct binding assay, where the ligand
itself is fluorescently labeled. The sandwich assay
consists of multiple steps, which are performed auto-
matically by the RAPTOR. Immediately after loading
the coupon into the instrument prior to analyte chal-
lenge, the RAPTOR automatically initiates a 5-min
baseline protocol. The waveguides are briefly rinsed
with PBT (for hydration), and are then incubated for
90 s with fluorescently labeled antibody. Afterwards,
this tracer reagent is returned to its compartment, the
waveguides are rinsed with PBT, and an initial back-
ground wash value is determined. This background
value represents the nonspecific binding of the fluo-
rescent antibody to the waveguide surface and must
be determined before samples are analyzed.
To analyze a sample, it is first loaded into the
sample port using a 1-ml syringe equipped with a
blunt-tipped needle or a transfer pipette. Then the
G.P. Anderson, N.L. Nerurkar / Journal of Immunological Methods 271 (2002) 17–24 19
sample is flowed into the coupon and is incubated
with the waveguides for 7 min. A subsequent wash
with PBT eliminates unbound material; the RAPTOR
automatically flushes the sample port with PBT at the
same time, thus preventing sample carry-over. Next,
the coupon is cleared with air to prevent dilution of
the incoming tracer reagent. Fluorophore-labeled anti-
body reagent is flowed into the coupon and incubated
for 90 s to interrogate the amount of antigen bound to
the waveguides. The tracer reagent is returned to its
reservoir for reuse and a final PBT wash is completed.
The entire standard assay cycle, including all wash
steps, is completed in 10 min.
In direct binding assays, no tracer reagent was used
since the samples contained the fluorescent target,
which was bound directly to the capture antibodies
on the waveguide. For the direct assay, the fluorescent
sample was incubated with the waveguides for the
entire 10 min, with binding being monitored inter-
mittently, and a final reading taken after the wave-
guides were washed.
The rate of fluorescence increase during incubation
with the labeled antibody is automatically calculated
(‘‘assay rate’’), and a final reading is taken to determine
the increase in fluorescence due to antibody bound
during the entire assay cycle (‘‘wash delta’’). Since the
two parameters parallel one another, we only consid-
ered wash deltas for these experiments. Assay data are
stored by the RAPTOR and can be downloaded
through a serial port for quantitative analysis.
2.7. Comparing fluorescent dye capabilities
Comparing Cy5 to AF647 was most efficiently
achieved by testing RAPTOR detection of analyte
concentrations beginning at 1 ng/ml (0.1 ng/ml for
SEB assays). Consecutive assays were then performed
on the same waveguides with increasing concentra-
tions of antigen up to 1000 ng/ml. As a precursor to
the series, all tests began with a blank sample (PBT)
prior to the lowest antigen concentration.
3. Results
The standard protocol for the RATOR was to
prepare Cy5-labeled antibodies at a D/P ratio of 2 to
4. Initial tests examined if signal enhancement could
be achieved by pushing the number of Cy5 dyes per
antibody even higher. Fluorescence measurements
performed for Cy5 incorporated at various ratios onto
rabbit antiricin IgG showed that fluorescence dec-
reased significantly with the addition of Cy5 mole-
cules in excess of the optimal range of 2 to 4 (Fig. 2).
These same rabbit antiricin IgGs, plus an additional
IgG prepared later with a Cy5 D/P ratio of 2.7, were
tested with the RAPTOR to determine the effect of
quenching on the fiber optic immunoassays (Fig. 3).
Not surprisingly, the immunoassay results mirrored
the fluorescence yields. However, despite exhibiting
measurable fluorescence, albeit small, the antibodies
labeled at D/P= 9.2 produced virtually no signal in the
immunoassay. To determine if this poor performance
was due to quenching or loss of activity due to
overlabeling (McCormack et al., 1996), we retested
the same waveguides with the antibody tracer labeled
with a D/P of 1.9. Since these waveguides had been
exposed to 1000 ng/ml of ricin, one would expect an
increase of around 5000 pA had the previous labeled
antibody been inactivated. However, the three wave-
guides in this experiment only had an average increase
of 40 pA, indicating that they had bound the IgG with
the 9.2 D/P and had few additional binding sites for
the D/P 1.9 tracer. Thus, the highly labeled antibodies
were not inhibited from binding to the analyte, but
Fig. 2. Relative fluorescence of Cy5-labeled rabbit antiricin IgG at
constant antibody concentration of 10 Ag/ml (1 ml samples), excited
at 630 nm. D/P: 1.9, solid line; 6.5, dotted line; 9.2, dashed line.
G.P. Anderson, N.L. Nerurkar / Journal of Immunological Methods 271 (2002) 17–2420
rather, the low fluorescent yield was largely the effect
of self-quenching.
Evidence of Cy5 dimer formation was apparent
from the spectroscopic data. The absorbance spectra
for goat IgG antibodies labeled with Cy5 at three
D/P ratios were compared (Fig. 4). Free Cy5 dye
molecules have an absorption maximum at 650 nm,
with a shoulder at 610 nm. As the antibodies’ D/P
ratio increases, there is a very significant increase
in the relative intensity of the 610 to 650 nm
absorption ratio. This increase in absorbance at
600 nm has been attributed to nonfluorescent dimer
formation (Gruber et al., 2000). The legend for Fig.
4 gives two D/P values for each sample. The first
D/P value given was calculated based only on the
650 nm adsorption. This is followed by an adjusted
dye-to-protein ratio calculated when the anomalous
dye adsorption at 600 nm was also added in as if it
were also 650 nm adsorption.
Fig. 5 shows fluorescence data for goat IgG anti-
bodies at three D/P ratios of AF647 dye. Spectra were
collected similar to the Cy5-labeling data shown in
Fig. 2. Fluorescence increases with the addition of
AF647 molecules to antibody, up to a D/P ratio of 6,
after which fluorescent yield varied little with addi-
tional dye molecules bound to the antibody. The
fluorescence of an antibody labeled with an average
of 3.6 Cy5 molecules each is shown for comparison.
One can note that significant self-quenching of the
Cy5 has already occurred.
To determine the fluorescent yield of AF647 mol-
ecules as the D/P ratio increases, antibodies labeled at
increasing ratios were measured in samples of fixed
dye quantity. This setup permits easy comparison of
their relative quantum yields (Fig. 6), and illustrates
Fig. 4. UV absorbance of Cy5-labeled goat IgG at various ratios,
compared to free dye for reference. D/P: 5, adj 10.5, solid line; 4,
adj 7, dotted line; 3 adj 5.3, dashed line; free Cy5 dye, dashed-doted
line.
Fig. 3. Fluorescent signal increase produced by various ratios of
Cy5-labeled rabbit antiricin IgG on the RAPTOR, upon challenge
with 1 Ag/ml ricin. The average response of three separate
waveguides is plotted.
Fig. 5. Relative fluorescence of AF647-labeled goat IgG antibodies
at multiple dye to protein ratios. Measurements at constant antibody
concentration of 10 Ag/ml, excited at 630 nm. D/P: 8.4, solid line;
6.3, dotted line; 2.5, dashed line. A Cy5-labeled antibody is also
shown, D/P= 3.6, dashed-dotted line.
G.P. Anderson, N.L. Nerurkar / Journal of Immunological Methods 271 (2002) 17–24 21
the amount of self-quenching that the dye exhibits as
more label is added per antibody. Notably, antibodies
labeled at D/P= 2.5 are equal in intensity to free dye
of the same label concentration, meaning virtually no
quenching occurs. As the D/P ratio is increased, the
intensity of fluorescence contributed by each individ-
ual dye molecule decreases. However, as noted in Fig.
5, each AF647 molecule contributes more to the total
fluorescence than it detracts via increased quenching,
thus producing a larger net signal, up to at least a D/P
ratio of 6.
The AF647-labeled antibodies with varying D/P
ratios were tested on RAPTOR both in direct binding
assays and in the more applicable sandwich assays in
order to gauge the performance of AF647 in fluo-
roimmunoassays. The direct assays were useful for
analysis because the surface is already saturated with
binding molecules, and therefore, a rapid and consis-
tently robust fluorescent signal is generated. However,
since the actual test format is the sandwich immuno-
assay, it is vital to ensure that results obtained in the
direct assay correspond to the utilized assay method-
ology. Fig. 7A and B shows the signal generated for
direct binding assays in which samples of 100 and
1000 ng of AF647-labeled goat IgG at various D/P
ratios are bound by rabbit anti-goat-coated optic
waveguides. Similar assays were performed for Cy5-
labeled antibodies, which are optimally labeled (D/
P= 3.6) to provide a direct comparison. Since AF647
has reduced self-quenching, the dye can be added to
antibodies in a higher ratio resulting in fluorescent
signal that peaks in performance at approximately 6
dyes per protein. At this D/P ratio, the signal derived
in the direct binding assay increased an average of
40% over that produced by an optimal amount of Cy5
dye.
A sandwich immunoassay, the standard test for-
mat of the RAPTOR, was performed to confirm
results from the direct binding assays and fluores-
cence data. Sheep anti-SEB IgG labeled at three D/
P ratios with AF647 and at D/P= 3.6 for Cy5, was
used as tracer reagent for the detection of SEB at
various concentrations. Most conclusive is the sig-
nal generated at the two highest concentrations
Fig. 7. Direct binding assays: (A) Wash deltas for AF647-labeled
goat IgG samples (100 ng/ml) at various D/P ratios, black circles.
Cy5 is represented by the open diamond at D/P= 3.6. (B) Same as A
except the goat IgG sample concentration was 1000 ng/ml. Error
bars represent the standard error of the mean of five separate
waveguides.
Fig. 6. Relative fluorescence of AF647-labeled antibodies at
multiple dye-to-protein ratios for constant dye concentration (O.D.
650 nm= 0.1). D/P: 2.5, solid line; 4.0, dashed line; 6.3, narrow
dashed line, and 8.4, dashed-dotted line. Free AF647 dye is also
shown, dotted line.
G.P. Anderson, N.L. Nerurkar / Journal of Immunological Methods 271 (2002) 17–2422
tested, 10 and 100 ng (Fig. 8). Here, the same
trends observed in the direct binding assays are
repeated, except that for this assay, the antibody
optimally labeled with Cy5 is out-performed by
antibody labeled with AF647 at all tested ratios.
4. Discussion
Fluorimetry data shown in Fig. 5 display the relative
intensity of goat IgG antibodies labeled with AF647 at
various dye-to-protein ratios, as well as a Cy5-labeled
antibody for comparison. Notably, when one attaches
multiple labels to an antibody, the net fluorescence
yield per antibody is much larger for AF647 than for
Cy5. Even at a dye-to-protein ratio of 2.5, AF647
fluoresced more intensely than antibodies labeled with
Cy5 at the higher ratio of 3.6. This difference becomes
even more dramatic if one looks instead at the max-
imum fluorescence yield generated by AF647-labeled
antibodies, i.e. a D/P ratio of 6, which is clearly more
fluorescent than Cy5-labeled antibodies that only
becomes less fluorescent with additional dye mole-
cules, as shown in Fig. 2. Clearly, the improvement
with AF647 is due to reduced self-quenching, since
fluorescence from AF647-labeled proteins approaches
an upper limit, having the same intensity at its optimum
D/P ratio of 6 as at 9 dyes per protein. By comparing
fluorescence yields for samples prepared at a constant
dye concentration (Fig. 6), it is possible to visualize the
net contribution of each dye molecule to the total
fluorescence of the antibody as a whole. Ideally, if no
quenching was encountered, all samples would have
identical curves. However, Fig. 6 illustrates that as the
amount of AF647 dye on a protein is increased, each
dye molecule contributes less to the overall signal
produced by the labeled antibody. At 2.5 dyes per
antibody, the fluorescence is identical to free AF647
dye, where it is assumed there is little to no quenching.
With each additional dye molecule added to the anti-
body, the yield per dye molecule is reduced. Despite
this trend, each additional label still contributes more to
fluorescent signal than it mitigates from the antibody as
a whole, until a threshold of 6 to 7 dyes per protein is
passed, at which point added fluorescence normalizes
with quenching, which results in the upper limit.
Hence, one may conclude that AF647 can be added
to an antibody in greater quantity than Cy5 to increase
fluorescent yield.
RAPTOR translates this fluorescence data into
applicable evidence of AF647’s superiority to Cy5
in fluoroimmunoassays. Juxtaposing first Cy5 and
AF647 in a direct binding assay, such as that
between goat IgG and rabbit anti-goat IgG (Fig.
7), provides a simple comparison of the two dyes.
Comparing the signal increases from Cy5-labeled
goat IgG at a D/P ratio of 3.6, which is in the
optimal range, to AF647-labeled goat IgG at multi-
ple ratios, one can gauge their relative effectiveness.
Using the AF647 dye, the assay signal increased in
direct proportion with the D/P ratio reaching a
maximum at a D/P ratio of 6, consistent with the
fluorescence results (Fig. 5). More importantly, in
direct assays, the AF647 dye (D/P= 6) improved
yields over Cy5 by an average of 50% and 40%,
Fig. 8. Sandwich assays: (A) Wash deltas for varied dye-to-protein
ratios in the detection of 10 ng SEB with AF647-labeled sheep anti-
SEB IgG reagent, black circles. Cy5-labeled sheep anti-SEB IgG is
represented in the above graph by an open diamond at D/P= 2.6. (B)
Results for detection of 100 ng SEB. Error bars represent the
standard error of the mean of at least five separate waveguides.
G.P. Anderson, N.L. Nerurkar / Journal of Immunological Methods 271 (2002) 17–24 23
respectively, at 100 and 1000 ng/ml of goat IgG.
Given that significant variation occurred between
individual assays largely due to waveguide-to-wave-
guide variations, these results were repeated numer-
ous times (n= 6) as were the sandwich immuno-
assay results (n = 7) to improve our confidence in
the data.
After observing promising results for the direct
immunoassay, we examined if AF647 would also
improve the performance of a sandwich immuno-
assay for SEB. Again, the signal is compared at
various ratios of AF647 to antibody and the same
antibody labeled with Cy5 at a D/P ratio of 2.6.
Unlike the direct binding assay, here the average
signal for all the Alexa Fluor-labeled antibodies
was greater than the Cy5-labeled antibody. The
combined average signal increases for the Alexa
Fluor-labeled antibodies at D/P ratios of 4.8 and
7.8, compared to the Cy5-labeled antibody (D/
P= 2.6), were 93% and 136% greater at 10 and 100
ng/ml SEB, respectively. Thus, two different fluo-
roimmunoassays using an AF647-labeled antibody
were shown to give improved fluorescent signals.
5. Conclusion
The focus of this study has been to evaluate the
ability of Alexa Fluor 647 to perform as a fluo-
rescent label for antibodies in sandwich fluoroim-
munoassays for analyte detection in biosensors such
as the RAPTOR. It can be concluded that AF647 is
an effective replacement in assays that have com-
monly used Cy5. Due to the limited quenching
characteristics of AF647, the dye can be added in
greater quantity to an antibody, while still remain-
ing fluorescent. This is obviously a quality critical
to improve detection limits in fluoroimmunosensors.
This reduced quenching also translates into a much
broader range of acceptable D/P ratios. Such a
quality in fluorophores used to label proteins is
invaluable because the labeling process is only
approximate and often can yield unwanted D/P
ratios. Still, if care is taken to achieve optimal D/
P ratios with Alexa Fluor 647, one can significantly
increase the fluorescent signal produced by an
antibody over that achievable with Cy5 in fluo-
roimmunoassays.
Acknowledgements
We thank Lisa Shriver-Lake and Charles Patterson
for their helpful suggestions. This work was funded in
part by the Defense Threat Reduction Agency and by
the U.S. Marine Corps. The views expressed here are
those of the authors and do not represent the opinions
of the U.S. Navy, the U.S Department of Defense, or
the U.S. Government.
References
Anderson, G.P., Rowe-Taitt, C.A., 2001. Water quality monitoring
using an automated portable fiber optic biosensor: RAPTOR.
Proc. SPIE 4206, 58.
Anderson, G.P., King, K.D., Cao, L.K., Jacoby, M., Ligler, F.S.,
Ezzell, J., 1998. Quantifying serum anti-plague antibody using
a fiber optic biosensor. Clin. Diagn. Lab. Immunol. 5, 609.
Anderson, G.P., King, K.D., Gaffney, K.L., Johnson, L.H., 2000.
Multianalyte detection using the fiber optic biosensor. Biosens.
Bioelectron. 14, 771.
Cao, L.K., Anderson, G.P., Ligler, F.S., Ezzell, J., 1995. Detection
of Yersinia pestis F1 antigen by a fiber optic biosensor. J. Clin.
Microbiol. 33, 336.
DeMarco, D.R., Saaski, E.W., McCrae, D.A., Lim, D.M., 1999.
Rapid detection of Escherichia coli O157:H7 in ground beef
using a fiber-optic biosensor. J. Food Prot. 62, 711.
Gruber, H.J., Hahn, C.D., Kada, G., Riener, C.K., Harms, G.S.,
Ahrer, W., Dax, T.G., Knaus, H., 2000. Anomalous fluorescence
enhancement of Cy3 and Cy3.5 versus anomalous fluorescence
loss of Cy5 and Cy7 upon covalent linking to IgG and non-
covalent binding to avidin. Bioconjug. Chem. 11, 696.
King, K.D., Vanniere, J.M., LeBlanc, J.L., Bullock, K.E., Anderson,
G.P., 2000. Automated fiber optic biosensor for multiplexed
immunoassays. Environ. Sci. Technol. 34, 2845.
Ligler, F.S., Golden, J.G., Shriver-Lake, L.C., Ogert, R.A., Wijesu-
ria, D., Anderson, G.P., 1993. Fiber-optic biosensor for the de-
tection of hazardous materials. ImmunoMethods 3, 122.
McCormack, T., O’Keeffe, G., MacCraith, B., O’Kennedy, R.,
1996. Assessment of the effects of increased fluorophore label-
ing on the binding ability of an antibody. Anal. Lett. 29, 953.
Panchuk-Voloshina, N., Haugland, R.P., Bishop-Stewart, J., Bhal-
gat, M.K., Millard, P.J., Mao, F., Leung, W.Y., Haugland, R.P.,
1999. Alexa dyes, a series of new fluorescent dyes that yield
exceptionally bright, photostable conjugates. J. Histochem. Cy-
tochem. 47, 1179.
Tempelman, L.A., King, K.D., Anderson, G.P., Ligler, F.S., 1996.
Quantitating staphylococcal enterotoxin B in diverse media us-
ing a portable fiber optic biosensor. Anal. Biochem. 223, 50.
G.P. Anderson, N.L. Nerurkar / Journal of Immunological Methods 271 (2002) 17–2424