fluor 3

Upload: suradip-das

Post on 08-Apr-2018

216 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/7/2019 fluor 3

    1/6

    Application of fluorescence lifetime imaging(FLIM) in latent finger mark detection

    L.K. Seah *, P. Wang, V.M. Murukeshan, Z.X. Chao

    School of Mechanical and Aerospace Engineering, Nanyang Technological University,

    50 Nanyang Avenue, Singapore 639798, Singapore

    Received 14 April 2005; received in revised form 1 August 2005; accepted 24 August 2005

    Available online 22 September 2005

    Abstract

    Fluorescence lifetime imaging (FLIM) in frequency domain enables the mapping of the spatial distribution of fluorescence

    lifetimes of a specimen. It has been extensively applied in biology. In this paper, a theoretical analysis for the fluorescence

    lifetime determination of latent finger mark samples is described, which is followed by the feasibility study of using FLIM in

    frequency domain for latent finger marks detection. Preliminary experiments are carried out with latent finger marks treated with

    a fluorescent powder on two different substrates. The resulting fluorescence lifetime image of finger mark revealed a good

    contrast, and was able to detect the latent finger marks clearly.

    # 2005 Elsevier Ireland Ltd. All rights reserved.

    Keywords: Fluorescence lifetime imaging; Frequency domain; Latent finger mark; Homodyne

    1. Introduction

    Finger marks are one of the most valuable types of

    physical evidence and hence its detection plays a significant

    role in criminal investigation and forensic science. There are

    three types of finger marks that occur at a crime scene:

    visible, plastic and latent. Latent finger mark is the most

    difficult to be detected. The existing detection techniques for

    latent finger marks have their own limitations. Most con-

    ventional detection methods (physical and chemical) cannot

    detect older marks [1,2]. Laser induced fluorescence detec-

    tion of latent finger mark was initially explored in 1976 and

    involves the detection of fluorescence intensity, color and

    lifetime [9]. In general, conventional methods for detectionof latent finger marks take the first two properties into

    consideration. However, the existing techniques based on

    fluorescence filtering fail when there is strong fluorescence

    emission from the background, and are ineffective when the

    emission wavelengths from the finger mark and that of

    background fall in the close wavelength range. In this

    context, FLIM method would be beneficial.

    Fluorescence lifetime is the average decay time of the

    fluorescence emitted by a molecule after excitation with a

    short laser pulse. Fluorescence lifetime imaging (FLIM)

    has received considerable attention and has been widely

    used in biophysics and medical diagnosis [36]. FLIM

    could obtain the unique and quantitative information

    available from dynamic fluorescence measurements, and

    possibly distinguish several fluorescence species with

    dissimilar lifetimes even though they may have overlap-

    ping spectra.

    This paper deals with determination of average fluores-cence lifetimes of latent finger mark sample on a pixel-by-

    pixel basis using homodyne detection method. A number of

    images of the sample are collected at different phase delays

    relative to the excitation light. A subsequent fit of fluores-

    cence intensity on a pixel-by-pixel basis yields the fluores-

    cence lifetime distribution image of the latent finger mark,

    which is independent of the fluorescence intensity.

    www.elsevier.com/locate/forsciintForensic Science International 160 (2006) 109114

    * Corresponding author. Tel.: +65 6790 4824; fax: +65 6795 4632.

    E-mail address: [email protected] (L.K. Seah).

    0379-0738/$ see front matter # 2005 Elsevier Ireland Ltd. All rights reserved.

    doi:10.1016/j.forsciint.2005.08.018

  • 8/7/2019 fluor 3

    2/6

    2. Principle and theory of fluorescence lifetime

    measurement

    2.1. Fluorescence lifetime measurement in frequency

    domain

    The measurement of fluorescence lifetime is based on the

    cross-correlation method introduced by Spencer and Weber

    [7] (Fig. 1). The lifetime resolved fluorescence experiment is

    carried out by modulating the excitation light sinusoidally at

    a high frequency (typically on order of 10100 MHz) and

    determining the demodulation and phase of high frequency

    modulated fluorescence emission.

    The fluorescence lifetime can be calculated by measuring

    the phase shift and demodulation of the fluorescence emis-

    sion relative to the phase and demodulation depth of the

    excitation light.

    If the excitation light intensity is modulated as a sinu-

    soidal function of time Ev(t) with a frequency f, the fluor-escence emission intensity is a sinusoidal function with the

    same frequency [8]:

    Evt E0;v1 ME;vsinv t fE;v (1)

    Fvt F0;v1 MF;vsinv t fF;v: (2)

    where E0,v is the time-independent offset, ME,v the relative

    modulation of the excitation light, v the radial frequency

    (v = 2pf)andfE,v is the arbitrary phase lag of the excitation.

    The symbols F0,v, M0,v, MF,v and fF,v of Eq. (2) are similar

    but refer to the fluorescence emission. The total phase lag of

    the fluorescence emission relative to the excitation light Df

    and modulation Mcan be expressed by

    Df fF;v fE;v (3)

    MMF;v

    ME;v(4)

    For a single lifetime, the lifetime can be determined from

    both Df and M:

    tf 1

    vtanDf (5)

    tM1

    v

    ffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi1

    M2 1

    r(6)

    where tf and tMare the lifetimes calculated from Df and M,

    respectively. If they aredifferent, it can be concluded that the

    system is heterogeneous. In this paper, the average lifetime

    was calculated by Eq. (5).

    The appropriate modulation frequencies for measuring

    fluorescence lifetimes, which fall in the range of 110 ns, are

    20100 MHz. It is impossible to directly measure the phase

    and demodulation of the time-varying signals with such high

    frequency. Therefore, the high frequency signal is trans-

    formed to either DC (homodyne measurement) or very low

    frequencies (heterodyne measurement), hence the phase and

    demodulation can easily be measured. In this paper, homo-

    dyne method was used as it is relatively simple compared to

    heterodyne method.

    2.2. Homodyne detection

    The principle of the homodyne measurement can be

    introduced easily for the simplest case of a perfectly sinu-soidal excitation and a single fluorescence lifetime. For

    homodyne detection, the source and detector frequencies

    are the same. The intensity is proportional to the sine

    function of the phase difference between the detector and

    the emission. To acquire the entire phase and modulation

    information, it is necessary to change the phase of the

    detecting device relative to that of the excitation light. In

    most of the case, the gain of the detection is modulated

    according to

    Gvt G0;v1 MG;vsinv t fG;v: (7)

    where G0,v is the time-independent offset, MG,v the relative

    modulation of the gain and fG,v is the arbitrary phase lag ofthe gain.

    Then the detected intensity is given by

    F0t F0;01 MF;0sinf0G;v fF;v; (8)

    where F0;0 F0;vG0;v; MF;0 12MF;vMG;v andf

    0G;v

    fG;v p2:

    To obtain Df and M, a reference measurement must be

    made by measuring a reference compound with known

    fluorescence lifetime (tR), the detected signal has the similar

    expression

    R0t R0;01 MR;0sinf0G;v fR;v (9)

    where R0;0 R0;vG0;v andMR;0 12 MR;vMG;v:Then Df and Mare defined as

    Df fF;v fs;v fF;v fR;v tan1vtR (10)

    MMF;v

    MS;v

    MF;v

    MR;v

    ffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi1 vtR

    2q (11)

    The subscript Sdenotes scattered light whose effective

    lifetime is zero.

    L.K. Seah et al. / Forensic Science International 160 (2006) 109114110

    Fig. 1. Phase shift Df and demodulation Mof the fluorescence

    signal with respect to the excitation light. The modulation

    M= (ACem/DCem)/(ACex/DCex).

  • 8/7/2019 fluor 3

    3/6

    2.3. Data analysis

    The data are taken as a series of subimages collected at

    equidistant phase steps. For Eqs. (10) and (11), the key is to

    get the value ofMF,v and fF,v. For a convenient analysis,

    Eq. (8) is redefined as:

    yt;i a11 a2sinxi a3 (12)

    where a1 = F0,0, a2 = MF,0, a3 = fF,v and xi f0G;v;i,

    where f0G;v;i is the different phase value of the gain. The

    goal of curve fitting is to find the value ofa1, a2 and a3 that

    will make the calculated results yt,i closest to the detected

    signals yd,i [8]. Eq. (12) is a non-linear relationship, but a

    linear expression could be obtained through transformation:

    yt;i c1 sin xi c2 cos xi c3 (13)

    where c1 a1a2 cos a3; c2 a1a2 sin a3 and c3 a1:In order to do the linear curve fit, Eq. (13) was regarded as

    multi-variable function. The images include the informationofyd,i and xi. After the process of multi-variable curve fit, the

    values of coefficients c1, c2 and c3, are obtained. The values

    ofa1, a2 and a3, are calculated by

    a1 c3

    a2

    ffiffiffiffiffiffiffiffiffiffiffiffiffiffiffic21 c

    22

    pc3

    a3 tanc2

    c1

    b1 > 0

    a3 tanc2

    c1

    p b1 < 0

    8>>>:

    (14)

    Through the same process, a2,R and a3,R are calculated.

    The Mand Df are obtained by simple calculation:

    Ma2;F

    a2;R

    1ffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi1 vtR

    2q (15)

    Df a3;F a3;R tan1vtR; (16)

    where Fand R refer to the sample and the reference signal,

    respectively.

    3. Experimental setup and procedure

    The schematic diagram of the experimental set up is

    shown in Fig. 2. A CW Argon laser (Coherent Innova 90 C)

    is modulated using an electro optic modulator (Conoptics

    EOM, Model 350-210) capable of operating in the range of

    0100 MHz. A power amplifier is used as the power source

    to the modulator, which is modulated by a sinusoidal signal

    output from function generator 1 (Agilent, Model 33250A).

    The modulated laser light is coupled into a single mode

    optical fiber (OZ optics) to illuminate the sample treated

    with fluorescence powders. An intensified charge coupled

    device camera (ICCD, Lavison, Pico Star HR) collects the

    resulting fluorescence emission from the sample. Function

    generator 2 (Agilent, Model 33250A) is used as the signalsource to modulate the gain of ICCD. The two function

    generators aresynchronized. A long pass optical filter is used

    to cut off the excitation-scattered light from the sample. The

    ICCD is connected with a camera interface board and

    controlled by a computer.

    By changing the phase of the function generator 2, the

    phase shift of the detecting device relative to the exciting

    light is changed. In order to have a good fit, 10 images are

    obtained in one period equal spaced out of 368 Then the

    different phase step of the gain is defined as:

    X 0;p=5; 2p=5; 3p=5; 4p=5;p; 6p=5; 7p=5; 8p=5; 9p=5

    After 10 sample images and 10 reference images are

    collected at equidistant phase steps, they are used to calcu-

    late the lifetime by FLIM program, which is implemented in

    MATLAB. The program is outlined schematically in Fig. 3.

    First a2,R and a3,R are obtained by calculating the reference

    data. After importing the source data, for every pixel whose

    L.K. Seah et al. / Forensic Science International 160 (2006) 109114 111

    Fig. 2. Schematic diagram of the experiment setup.

  • 8/7/2019 fluor 3

    4/6

    intensity is within the threshold, the intensity values of

    equidistant phase steps are fitted to a sine wave to yield

    the demodulation factor a2,Fand the phase shift factor a3,F.

    In order to process the curve fitting procedure, Eq. (13) is

    transformed to

    Y c1

    sinX c2

    cosX c3

    (17)

    The function in MATLAB toolbox is used to obtain the

    coefficients. Eq. (14) is used to obtain the value ofa1,F, a2,F

    and a3,F. Then MandDf arecalculated to decide the lifetime

    through Eqs. (15) and (16). After every pixel is processed,

    the maximum and minimum values are set as the value range

    to output the lifetime image.

    In order to simulate a condition of having very close

    emission wavelengths for finger mark and background

    fluorescence, fresh finger marks from a single person are

    deposited on fluorescing color paper and treated with blitz-

    green powder (Lightning Powder Company) that is applied

    L.K. Seah et al. / Forensic Science International 160 (2006) 109114112

    Fig. 3. Block diagram of FLIM program.

    Fig. 4. The intensity image (a) and fluorescence lifetime image (b) of the blitz-green-treated finger mark samples on green fluorescence paper

    substrate.

  • 8/7/2019 fluor 3

    5/6

    by means of a magneticbrush.The emissionwavelength of the

    fluorescing color paper is 516 nm and its lifetime is 3.1 ns,

    while those of blitz-green powder are 523 nm and 8.96 nm.

    Experiment was also carried out on the finger mark

    treated with blitz-green fluorescent powder on smooth

    calendar paper. The background has two major fluores-

    cence components with lifetime value 13.42 ns and

    2.56 ns. Lifetime value of finger mark sample treated

    with blitz-green powder is 8.96 ns, which is shorter than

    that of the major fluorescence component (13.42 ns) from

    the background

    4. Result and discussion

    Selecting a 700 700 pixel region, the intensity imageof the finger mark on green fluorescence paper treated

    with blitz-green powder is shown in Fig. 4a. The contrast

    of the latent finger mark image is found to be poor and

    hence could not identify the ridge details. This is due to

    that the emission wavelength of green fluorescence paper

    (516 nm) is merged with that of the latent finger mark

    treated with blitz-green fluorescent powder (523 nm).

    Fig. 4b shows the corresponding average lifetime image

    of latent finger mark. The fluorescence powder has longer

    lifetime than the background fluorescence of the green

    paper. It can be seen that lifetime image effectivelysuppress the background fluorescence so as to give a clear

    ridge details of the finger mark

    Fig. 5 shows the intensity image and fluorescence life-

    time image of the finger mark treated with blitz-green

    powder on postcard. Due to the strong multiple fluorescence

    emissions from the background, the latent finger mark

    cannot be identified under normal imaging. As mentioned

    in the previous case, in the fluorescence lifetime image the

    background can be subdued successfully despite the back-

    ground fluorescence is having a higher intensity emission.

    For the case that the background has longer lifetime, a

    negative image will be obtained compared to Figs. 4b and 5b.

    The finger mark can also be detected in that case.

    5. Conclusion

    FLIM has been widely used in biology and biomedical

    fields. In this paper, lifetime imaging of latent finger marks

    on two different substrates were carried out. An approach

    that involved the calculation of total phase lag and demo-

    dulation factor is used to determine the lifetimes pixel by

    pixel. The results show that the FLIM can detect the latent

    finger mark whose emission wavelength falls in the close

    wavelength range with that of the background. The ridge

    details of latent finger mark can be identified clearly. The use

    of lifetime imaging yields results that are independent of the

    fluorescence intensity. FLIM technique is a general approach

    for latent finger mark detection and is only limited by the

    lifetime difference between the substrate and the finger

    mark. Hence, as long as there is a lifetime difference that

    can be detected, this technique can be applied.

    Acknowledgements

    The authors gratefully acknowledge the support and fund-ing from The Enterprise Challenge (TEC), Singapore and

    Academic Research Fund (AcRF), Nanyang Technological

    University. The authors express their sincere gratitude to

    Singapore Police Force for their collaboration in this project.

    References

    [1] E.R. Menzel, Fingerprint Detection with Lasers, second ed.,

    Marcel Dekker, New York, 1999.

    L.K. Seah et al. / Forensic Science International 160 (2006) 109114 113

    Fig. 5. The intensity image (a) and fluorescence lifetime image (b) of the blitz-green-treated finger mark samples on postcard substrate.

  • 8/7/2019 fluor 3

    6/6

    [2] H.C. Lee, R.E. Gaensslen, Advances in Fingerprint Technology,

    Elsevier, New York, 1991.

    [3] R. Sanders, H.C. Gererritsen, Fluorescence lifetime imaging of

    free calcium in single cells, Bioimaging 2 (1994) 131

    138.

    [4] D. Elson, S. Webb, J. Siegel, K. Suhling, D. Davis, J. Lever, D.Phillips, A. Wallace, P. French, Biomedical applications of

    fluorescence lifetime imaging, Opt. Photon. News 11 (2002)

    2732.

    [5] K. Dowling, M.J. Dayel, Hyde SCW, French PMW, M.J. Lever,

    J.D. Hares, High resolution time-domain fluorescence lifetime

    imaging for biomedical applications, J. Mod. Opt. 46 (2) (1999)

    199209.

    [6] M. Elangovan, R.N. Day, A. Periasamy, Nanosecond fluores-

    cence resonance energy transfer-fluorescence lifetime imaging

    microscopy to localize the protein interactions in a single living

    cell, J. Microsc. 205 (2002) 314.

    [7] R.D. Spencer, G. Weber, Measurements of subnanosecond

    fluorescence lifetime with a cross-correlation phase fluorometer,Ann. N. Y. Acad. Sci. 158 (1969) 361376.

    [8] Gadella TWJ, R.M. Clegg, T.M. Jovin, Fluorescence lifetime

    imaging microscopy pixel-by-pixel analysis of phase-modula-

    tion data, Bioimaging 2 (1994) 135159.

    [9] L.K. Seah, U.S. Dinish, S.K. Ong, Z.X. Chao, V.M. Muruke-

    shan, Time-resolved imaging of latent fingermarks with nano-

    second resolution, Opt. Laser Tech. 36 (5) (2004) 371376.

    L.K. Seah et al. / Forensic Science International 160 (2006) 109114114