winflow4 , análisis de cianuro - analisis intrumental

Upload: yean-carlos

Post on 13-Apr-2018

244 views

Category:

Documents


1 download

TRANSCRIPT

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    1/101

    WinFLOW Software V4.2

    Operators Manual

    151 Graham Road P.O. Box 9010 College Station, Texas 77842-9010

    Telephone (979) 690-1711 FAX (979) 690-0440 www.oico.com [email protected]

    Flow SolutionIVSystem

    Flow Solution3100 System

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    2/101

    ii

    Notice

    The information contained in this document may be revised without notice.

    OI Analytical shall not be liable for errors contained herein or for incidental or

    consequential damages in connection with the furnishing, performance, or use of this

    material.

    No part of this document may be reproduced, photocopied, or translated to another

    language without the prior written consent of OI Analytical.

    Rev. 2.3 July 2007

    OI Analytical Part Number A001965

    Printed in the U.S.A.

    Publication 134800707

    Copyright 20052007 OI Analytical

    Adobe and Acrobat Reader are registered trademarks of Adobe Systems, Inc.

    Windows is a registered trademark of Microsoft Corporation.CNSolution, Flow Solution, and WinFLOW are trademarks or registered trademarks of OI Analytical.

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    3/101

    iii

    Limited Warranty

    OI Analytical warrants each OI Analytical manufactured product against defects in

    materials and workmanship under normal use and service for a period of one year.

    Equipment installed by OI Analytical is warranted from the installation date; all other

    equipment is warranted from the ship date. If purchaser schedules or delays installation

    more than 90 days after delivery, then the warranty period starts on the 91st day from

    date of shipment. This warranty extends only to the original purchaser. OI Analyticalwill, at its option, repair or replace equipment that proves to be defective during the

    warranty period, provided the equipment is returned to OI Analytical at the expense of

    the purchaser.

    Consumables, expendables, and parts are warranted for 30 days and are not covered

    under extended warranties or service contracts.

    OI Analytical warrants for a period of one year from the date of delivery: (i) the

    Software, when installed and used with an OI Analytical recommended hardware

    configuration, will perform in substantial conformance with the documentation

    supplied with the Software; and (ii) the physical media on which the Software is

    furnished will be free from defects in materials and workmanship under normal use.

    This warranty shall not apply to defects originating from, but not limited to, the

    following:

    Improper maintenance or operation by the purchaser;

    Purchaser supplied accessories or consumables;

    Modification or misuse by the purchaser;

    Operation outside the products environmental and electrical specifications;

    Software, interfacing, parts, or supplies not supplied by OI Analytical;

    A computer not meeting the minimum specifications recommended by OI

    Analytical;

    Improper or inadequate site preparation;

    Purchaser induced contamination or leaks.

    THE FOREGOING WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES,

    EXPRESSED OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANYWARRANTY OF MERCHANTABILITY, FITNESS, OR ADEQUACY FOR ANY

    PARTICULAR PURPOSE OR USE. OI ANALYTICAL SHALL NOT BE LIABLE

    FOR ANY SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES,

    WHETHER IN CONTRACT, TORT, OR OTHERWISE.

    Any service requests or questions should be directed to the OI Analytical Customer

    Support Center at (800) 336-1911 or (979) 690-1711.

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    4/101

    iv

    Safety Information

    For operator safety, pay special attention to the WARNINGand CAUTIONstatements

    throughout the manual.

    A WARNINGindicates a condition or possible situation that could result in

    physical injury to the operator.

    A CAUTIONindicates a condition or possible situation that could damage ordestroy the product or the operators work.

    Warnings and precautions in this manual or on the instrument must be followed during

    operation, service, and repair of the instrument. Failure to follow these warnings and

    precautions violates the safety design standards and intended use of the instrument.

    OI Analytical is not liable for the operators failure to comply with warnings and

    precautions.

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    5/101

    v

    Table of Contents

    Notice .............................................................................................................................. ii

    Limited Warranty ........................................................................................................ iii

    Safety Information ....................................................................................................... iv

    Chapter 1: Introduct ion ........................................................1Features .......................................................................................................................... 1

    Computer Requirements .............................................................................................. 2

    Security Settings for WindowsXP Operating Systems...........................................2

    Chapter 2: Instal lat ion ..........................................................4Installing WinFLOW.................................................................................................... 4

    Security Settings for WindowsXP Operating Systems ........................................... 5

    Chapter 3: Configurat ion .....................................................7Configuring WinFLOW ............................................................................................... 7

    Configuring the RA Sampler ....................................................................................... 9

    Vial Depths..............................................................................................................10

    Dilutor Setup ........................................................................................................... 12

    Wash and Diluent ....................................................................................................13

    Rack Placement .......................................................................................................14

    Rack Editor..............................................................................................................15

    Configuring the 3360 Sampler................................................................................... 18

    Vial Depths..............................................................................................................19

    Dilutor Setup ........................................................................................................... 20

    Wash and Diluent ....................................................................................................21

    Rack Placement .......................................................................................................21

    Peck Probe during Rinse .........................................................................................22

    Configuring the 3090 Sampler................................................................................... 22

    Vial Depths..............................................................................................................23

    Wash and Diluent ....................................................................................................24

    Rack Placement .......................................................................................................25

    Peck Probe during Rinse .........................................................................................25

    Chapter 4: Introduction to WinFLOW ...............................26Starting WinFLOW .................................................................................................... 26

    Select View Dialog Box .......................................................................................... 26

    Tools Menu................................................................................................................... 27Preferences .............................................................................................................. 28

    Configuration...........................................................................................................28

    Events Scheduler .....................................................................................................29

    Instrument Control...................................................................................................30

    Pump Timer.............................................................................................................30

    Window Menu ............................................................................................................. 30

    About Menu ................................................................................................................. 31

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    6/101

    vi

    Shortcut Keys .............................................................................................................. 31

    Icon Ribbon.................................................................................................................. 31

    System Security ........................................................................................................... 33

    WinFLOW Quick Start .............................................................................................. 34

    Chapter 5: Method Editor ...................................................36General Description .................................................................................................... 36

    Method Editor Icon Ribbon ..................................................................................... 37

    Opening a Method ...................................................................................................37

    Creating a Method ...................................................................................................38

    Saving a Method......................................................................................................38

    Flow Control Settings Screen ..................................................................................... 38

    Instrument................................................................................................................39

    Sampling Mode .......................................................................................................39

    CCV/Drift Fail.........................................................................................................41

    Vial Types ............................................................................................................... 41

    Off-scale (HI) Samples............................................................................................41

    Insert Baselines........................................................................................................42

    Pump Control...........................................................................................................42Detector/Channel Settings Screen ............................................................................. 43

    Channels .................................................................................................................. 43

    Name .......................................................................................................................43

    Methodology ........................................................................................................... 44

    CAS Number ........................................................................................................... 44

    Detector Settings .....................................................................................................44

    Channel Subtractions...............................................................................................45

    Signal Filter and Marking Settings Screen............................................................... 46

    Smoothing ............................................................................................................... 47

    Baseline Marking.....................................................................................................48

    Inverted Chemistry..................................................................................................48

    Peak Marking...........................................................................................................48Calibration/Quantitation Settings Screen................................................................. 50

    Calibration............................................................................................................... 51

    Quantitation............................................................................................................. 51

    Flagging...................................................................................................................52

    Overlimit Samples...................................................................................................52

    Timed Events Editor Screen....................................................................................... 52

    B (Basis of Timing).................................................................................................53

    Time.........................................................................................................................53

    Device......................................................................................................................53

    # (Selector) .............................................................................................................. 54

    State .........................................................................................................................54

    Calibrants Table Editor Screen ................................................................................. 55Calibrants.................................................................................................................55

    Independent Calibration Verification ......................................................................56

    Continuing Calibration Verification........................................................................56

    Spike........................................................................................................................56

    Stock Calibrants.......................................................................................................57

    Drift Correction .......................................................................................................57

    Verifying Method Settings .......................................................................................... 57

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    7/101

    vii

    Chapter 6: Sample Table....................................................58Cup Number ............................................................................................................ 59

    Sample Names.........................................................................................................60

    Sample Type............................................................................................................60

    Replicate..................................................................................................................61

    Dilution and Weight Factors ...................................................................................61

    Vial Type.................................................................................................................61

    Comments................................................................................................................61Sample Table Menu Options ...................................................................................... 62

    File Menu ................................................................................................................ 62

    Edit Menu................................................................................................................ 62

    Tools and Window Menus.......................................................................................63

    Creating a Sample Table............................................................................................. 63

    Synchronization Peak (SYNC)................................................................................63

    Carryover Correction (CO)......................................................................................63

    Calibrants.................................................................................................................63

    Calibrants To Be Diluted.........................................................................................63

    Unknowns To Be Diluted........................................................................................64

    Stock Solutions........................................................................................................64

    Incremental Sensitivity Drift Correction .................................................................65Total Sensitivity Drift Correction............................................................................65

    Independent Calibration Verification ......................................................................65

    Continuing Calibration Verification........................................................................65

    Spiked Samples .......................................................................................................66

    Blanks and Baseline Samples..................................................................................66

    Autodilution ................................................................................................................. 67

    Import and Export Functions .................................................................................... 68

    Exporting a Table ....................................................................................................68

    Importing a Table ....................................................................................................68

    Chapter 7: Data Collection .................................................69Initiating Data Collection ........................................................................................... 69

    Monitoring the Baseline.............................................................................................. 72

    Starting a Sample Analysis......................................................................................... 73

    Peak Plot Screen......................................................................................................73

    Calibration Plot Screen............................................................................................73

    Calibration Results Screen ...................................................................................... 74

    Peak Table Results Screen.......................................................................................75

    Adding Samples to a Sample Table.........................................................................75

    Modifying a Sample Table During a Run ................................................................. 76

    Stopping a Run ............................................................................................................ 76

    Operator-Stopped Run.............................................................................................76

    WinFLOW-Stopped Run.........................................................................................76

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    8/101

    viii

    Chapter 8: Results and Analys is .......................................77Loading a Data File..................................................................................................... 77

    Data Processing Tools ................................................................................................. 78

    Data Analysis Screen...............................................................................................78

    Grid On/Off Icon .....................................................................................................78

    Peak Editor Icon ......................................................................................................78

    Recalculate Icon ......................................................................................................78

    Delete/Reprocess Peaks Icon...................................................................................79Menu Options .......................................................................................................... 80

    Peak Plot Screen.......................................................................................................... 82

    Magnifying the Peak Plot ........................................................................................ 82

    Using the Cursor For Peak Identification................................................................83

    Changing a Peak State (Add, Delete, Toggle Baseline Point).................................83

    Calibration Plot Screen............................................................................................... 84

    Magnifying the Calibration Plot..............................................................................84

    Using the Cursor for Calibration Point Identification .............................................84

    Removing Outlier Calibration Points ...................................................................... 84

    Peak Results Table Screen.......................................................................................... 85

    Peak Results Table Fields........................................................................................85

    Calibration Results Table Screen............................................................................... 87Calibrant ID.............................................................................................................87

    Calibration Coefficients...........................................................................................87

    Correlation Coefficient............................................................................................87

    Carryover.................................................................................................................87

    Drift Correction .......................................................................................................87

    Reports ......................................................................................................................... 88

    Exporting a Report ..................................................................................................... 89

    Chapter 9: Troubleshooting...............................................90Instrument Errors....................................................................................................... 90

    WinFLOW Errors....................................................................................................... 91Reporting Errors ......................................................................................................92

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    9/101

    WinFLOW Software Operators Manual: Chapter 1 1

    Rev. 2.3

    Ch a p t e r 1 Introduction

    The WinFLOW software is a Windows-based program designed to offer a high

    level of automation when used with any of the OI Analytical Flow Solution

    systems,as well as many other automated chemistry analyzers.

    Features

    Automatic startup and shutdown enhances system productivity.*

    Pause option permits run and sample table modification anytime during a run.

    Sensitivity drift correction eliminates the need to fully recalibrate after ICV/CCV

    failure.

    Operator-defined dilution factors add operator flexibility in autodilution mode.*

    Sample dilution before the run reduces analysis time.

    Automatic working calibration standard preparation from stock solutions reduces

    operator workload.*

    Manual injection mode permits system operation without an autosampler.

    Peak editing improves peak height and area marking.

    Post-run peak height, area, and digital signal processing algorithms enhance

    analyte detectability.

    Mean and RSD values add important QA/QC data to reports.

    Options provide customized reporting capabilities.

    Complete French, German, and Spanish language support provides enhanced

    usability.

    *This feature requires specific hardware for full system functionality.

    The WinFLOW software package includes the following:

    CD-ROM containing the following:

    WinFLOW Software

    Hardware Operators Manual (one of the following):

    WinFLOWSoftware Operators Manual

    Flow SolutionIV Operators Manual

    Flow SolutionIVOperators Manual

    Flow Solution3000 Operators Manual

    Flow Solution3100 Operators Manual

    CNSolutionFS 3100 Operators Manual

    Adobe

    Acrobat Reader

    WinFLOWsoftware serial number (labeled on the outside of the CD-ROM case)

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    10/101

    Introduction

    WinFLOW Software Operators Manual: Chapter 1 2

    Rev. 2.3

    Computer Requirements

    The following table contains minimum and recommended computer requirements for

    successfully installing and using the WinFLOWsoftware.

    NOTE: Using WinFLOWsoftware requires a working knowledge of the Windows

    2000 or WindowsXP Pro operating system.

    NOTE: Flow SolutionIII and Flow SolutionIV systems larger than three channels

    or Flow Solution3000 systems larger than two channels require the

    recommended computer requirements.

    NOTE: Power management features (excluding the monitor) can cause shutdowns

    and errors during analysis.

    Securi ty Settings

    for WindowsXPOperatingSystems

    When using the WinFLOWsoftware with a WindowsXP Pro operating system, set

    permissions to Everyonefor full control of all WinFLOWsoftware features. The

    default setting for the WindowsXP Pro operating system is NO users. Therefore,

    change permissions to Everyone after installing WinFLOWsoftware on the

    computer.

    1. Enable Full Control to Everyone by right-clicking Startat the screens lower left

    corner.

    2. Click Explore.

    3. Find the Folder named Flow_4.

    Table 1.1. Computer requirements

    Parameter Minimum Recommended

    Type IBM compatible IBM compatible

    Processor speed 1 GHz 3 GHz

    Memory 512 MB 1 GB or greater

    Free disk space 500 MB 1 GB or greater

    Operating system Windows2000 or XP

    Pro

    WindowsXP Pro

    Graphics SVGA XGA

    COM port One available RS-232 or

    USB-to-RS-232 converter

    One available RS-232

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    11/101

    Introduction

    WinFLOW Software Operators Manual: Chapter 1 3

    Rev. 2.3

    NOTE: WinFLOW normally installs in this folder. If the software installed in a

    different location, go to that folder.

    4. Right-click the Flow_4 folder and select Properties.

    5. Under the Securitytab, click the Addbutton.

    6. In the Select Users, Computers, or Groups dialog box, type the word Everyone

    under the Enter the object names to select sectionand click OK(see Figure 1.1).

    Figure 1.1. Permissions dialog box

    7. Click the checkbox under Allownext to Full Controlto enable full control.

    8. Click Applyand OKto exit the screen.

    http://-/?-http://-/?-
  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    12/101

    WinFLOW Software Operators Manual: Chapter 2 4

    Rev. 2.3

    Ch a p t e r 2 Installation

    This chapter presents step-by-step procedures for properly installing WinFLOW.

    Installing WinFLOW

    The WinFLOW CD contains WinFLOW software, the WinFLOW software manual,

    and all instrument operation manuals. Install WinFLOW using the following

    instructions:

    1. From the Windows environment, insert the WinFLOW CD into the available CD-

    ROM drive. The WinFLOW Setup Disc screen appears (Figure 2.1).

    Figure 2.1. WinFLOW Setup Disc screen

    If the WinFLOW Setup Disc screen does not automatically appear after inserting

    the CD:

    Click the Windows Startbutton.

    Select Run.

    Typex:\setup,wherexrepresents the letter assigned to the CD-ROM drive.

    Click OKto display the WinFLOW Setup Disc screen.

    2. Select the appropriate language and click Installto begin the installation process.

    Follow the instructions on the screen.

    NOTE: Do not use more than eight characters or spaces in the directory names.

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    13/101

    Installation

    WinFLOW Software Operators Manual: Chapter 2 5

    Rev. 2.3

    3. After WinFLOW installs, click OK. The Configure WinFLOW dialog box appears

    (Figure 2.2). Proceed to Chapter 3, Configuration on page 7for details on

    configuring WinFLOW.

    Figure 2.2. Configure WinFLOW dialog box

    Security Settings for WindowsXP Operating Systems

    When using WinFLOW with a Windows XP operating system, set permissions to

    Everyonefor full control of all of WinFLOW software features. The default setting for

    a Windows XP operating system is NO users. Therefore, change permissions toEveryone after installing WinFLOW on the computer.

    1. Enable Full Control to Everyone by right-clicking Startat the screens lower left

    corner.

    2. Click Explore.

    http://-/?-http://-/?-
  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    14/101

    Installation

    WinFLOW Software Operators Manual: Chapter 2 6

    Rev. 2.3

    3. Find the Folder named Flow_4.

    NOTE: WinFLOW normally installs in this folder. If the software installed in a

    different location, go to that folder.

    4. Right-click the Flow_4 folder and select Properties.

    5. Click the Securitytab and add the group Everyone if not already created

    (Figure 2.3).

    Figure 2.3. Permissions dialog box

    6. Click the checkbox under Allownext to Full Controlto allow everyone full control

    permissions.

    7. Click Applyand OKto exit the screen.

    http://-/?-http://-/?-
  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    15/101

    WinFLOW Software Operators Manual: Chapter 3 7

    Rev. 2.3

    Ch a p t e r 3 Configuration

    After installing WinFLOW, configure communications to the analyzer and autosampler.

    Configuring WinFLOW

    1. Access the Configuration Editor by selecting FLOW config 4.2from the Start

    menu. Click Programs, WinFLOW, and then the FLOW config icon (or go to

    Tools Configuration in WinFLOW). The Configure WinFLOW dialog box appears

    (see Figure 3.1).

    Figure 3.1. Configure WinFLOW dialog box

    2. Under Configure Software, select the appropriate language for WinFLOW to

    display for its titles, buttons, labels, etc. The language options are English, French,

    Dutch, and Spanish.

    http://-/?-http://-/?-
  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    16/101

    Configuration

    WinFLOW Software Operators Manual: Chapter 3 8

    Rev. 2.3

    3. Click the Edit Facility Informationbutton to enter a facility name and department

    to display on hard-copy reports.

    4. Select the Create Debug Filescheckbox to generate text files that assist OI

    Analytical support personnel with troubleshooting software and performance

    problems.

    5. Select the communication port (Com Port), Sampler Control, and the analyzer

    type (Platform) to use. Only available communication ports appear as choices.

    6. Select the appropriate Sampler Control.

    a. Select Standaloneif using an autosampler not listed or if no autosampler is

    used. If using an autosampler not listed, the autosamplers timing cycle must

    match WinFLOW. The autosampler and data collection must initiate at the

    same time.

    b. Select RA to use vials and vial racks provided with the RA Sampler. The RA

    Sampler supports any combination of three trays: 21, 24, 40, 60, or 90

    positions; the RA sampler also supports the Autodilutor.

    c. Select 3360to use vials and vial racks provided with the Model 3360 Sampler.

    The Model 3360 supports any combination of four 60-position and 90-position

    racks; the Model 3360 also supports the Autodilutor.

    d. Select 3090to use vials and vial racks provided with the Model 3090 Sampler;

    the Model 3090 supports either one 60-position or one 90-position rack.

    e. Click the Configure Samplerbutton to set the sampler to utilize up to four

    different types of sample cups, configure any dilutor in use, specify the desiredrack options, and define alternate locations for wash and diluent fluids.

    7. Check the Auto 502 check box, if appropriate:

    If using a Flow Solution IV or IVsystem, check the Auto 502checkbox to control

    the Precision Pump through the Instrument Control dialog box or the Auto Events

    Scheduler (see Chapter 4, Tools Menu on page 27).

    If using an Flow Solution 3100 system, select the Auto 502checkbox to control the

    Precision Pump.

    The CNSolution FS 3100 uses the Model 3090 Autosampler. Flow Solution 3100

    systems use either the Model 3090 or Model 3360 Autosampler. Flow Solution III

    and IV systems use any of the samplers, including the RA, Model 3090, and Model

    3360. Run upgraded systems without one of the listed autosamplers in the standal-

    one sampler control mode, provided the autosampler being used is capable of inde-

    pendent operation.

    NOTE: Configuring WinFLOW every time the software opens is not necessary.

    Change WinFLOW configuration settings only if altering the instrument

    configuration.

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    17/101

    Configuration

    WinFLOW Software Operators Manual: Chapter 3 9

    Rev. 2.3

    8. After selecting the checkbox to control the Precision Pump, use the Pump Turns

    pulldown menu to select the pump direction (when viewed from the front of the

    pump). Leftindicates the negative pump direction (the pump pushes the liquids

    clockwise), while Rightindicates the positive direction (counter-clockwise).

    Most Flow Solution IV customers will select Right. Most Flow Solution 3100 cus-

    tomers will select Left.

    Configuring the RA Sampler

    1. Select the RA sampler in the Configure WinFLOW dialog box.

    2. Click Configure Sampler. The RA Configuration dialog box appears as shown in

    Figure 3.2.

    Figure 3.2. RA Configuration dialog box

    http://-/?-http://-/?-
  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    18/101

    Configuration

    WinFLOW Software Operators Manual: Chapter 3 10

    Rev. 2.3

    Vial Depths Using a Traveling Wash Reservoi r (TWR)

    Adjust the depth that the sample probe travels into a sample cup.

    1. Click Vial Depthsin the RA Configuration dialog box. The Vial Depths dialog box

    appears (see Figure 3.3).

    Figure 3.3. Vial Depths dialog box

    2. Enter the appropriate vial depths (in millimeters) in the corresponding boxes. The

    maximum probe depth in the sample vial is approximately 95 mm.

    Using a Static Wash Reservoir (No TWR)

    Adjust the depth that the sample probe travels into a sample cup.

    Due to variations of individual sampler setups, no exact default depths are available,

    although starting depths are provided in Table 3.1 on page 11. Instead, the process iscompleted by trial and error where the optimal probe positions are determined for each

    vial.

    Any of the Vial Type numbers can be assigned to any of the vials available (2 mL,

    4 mL, 8 mL, and 12 mL). Follow the steps below to configure the vial depths for each

    Vial Type.

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    19/101

    Configuration

    WinFLOW Software Operators Manual: Chapter 3 11

    Rev. 2.3

    1. Click Vial Depthsin the RA Configuration dialog box. The Vial Depths dialog box

    appears (see Figure 3.3).

    Figure 3.4. Vial Depths dialog box

    2. Enter the appropriate vial depths (in millimeters) in the corresponding boxes. Ifusing a static wash reservoir (no TWR), the range depends on the setup probe

    position in the autosampler arm and the wash reservoir type being used; the range is

    generally larger than 95 mm.

    To get the maximum probe depth into the vial, set the probe at its uppermost posi-

    tion and adjust the needle so that the tip of the needle barely clears the lip of the

    wash reservoir. Lock the needle into position with a set screw.

    Figure 3.5. Probe needle in relation to the wash reservoir

    a. Determine which vial size will be used the most (from 2 mL, 4 mL, 8 mL, and

    12 mL). This value normally designates Vial Type 1.

    b. Verify Vial Type 1is set as the Default Vial Type in the Method Editor (see

    Chapter 5, Flow Control Settings Screen on page 38).

    Table 3.1. Vial Depth starting values

    Vial Size

    (mL)Rack Size

    Approx imate

    Starting Value (mm)

    2 90-position 65

    4 60-position with shelf 75

    8 90-position 100

    12 60-position 100

    Needle

    WashReservoir

    Just clear the

    reservoir

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    20/101

    Configuration

    WinFLOW Software Operators Manual: Chapter 3 12

    Rev. 2.3

    c. Open the WinFLOW software and create a Sample Table with a cup position

    adjacent to the position that contains a vial of Vial Type 1.

    d. Send the sampler to the position chosen in step 1 by starting a run. Check the

    depth of the probe tip to determine whether to raise, lower, or leave the tip.

    NOTE: For 8-mL and 12-mL tubes:If the probe goes over the vial, but doesn't godown, the depth is set deeper than the sampler can go. Decrease the depth

    until the probe will lower. This value is the maximum depth for any vial; the

    maximum possible depth is 100mm.

    e. If the depth needs adjustment, close WinFLOW and open Configure Winflow.

    Make the appropriate depth changes and click OK.

    f. Open Winflow and repeat steps dthrough euntil the optimal depth is reached.

    g. Repeat steps athrough ffor each Vial Type.

    NOTE: If only using one vial type, only set the depth for the default Vial Type.

    NOTE: Vial types other than the default must be designated with their type number in

    the Vialcolumn of the Sample Table for each non-default vial position.

    Dilutor Setup If using the Autodilutor with the RA Sampler, click Dilutor Setup in the RAConfiguration dialog box. The Dilutor Setup dialog box appears (see Figure 3.6).

    Figure 3.6. Dilutor Setup dialog box

    By selecting Use Dilutor, Syringe Size and Syringe Speed settings become available

    for input.

    Use the pull-down menu to select the syringe size. The default syringe size is 2,500 L.

    The available syringe sizes include the following: 1,000 L, 2,500 L, 5,000 L, and

    10,000 L.

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    21/101

    Configuration

    WinFLOW Software Operators Manual: Chapter 3 13

    Rev. 2.3

    Syringe size and the application determine proper syringe speed. The speed ranges from

    25 for larger syringes to 50 for smaller syringes. The default speed values for each

    syringe size are shown in Table 3.2.

    Wash and Diluent OI Analytical offers two RA Sampler probe arm configurations. Specify which type touse by clicking Wash and Diluentin the RA Configuration dialog box. The Wash and

    Diluent dialog box appears (Figure 3.7).

    Figure 3.7. Wash and Diluent dialog box

    1. Choose a Wash Source:

    TWR(traveling wash reservoir)

    Reservoir(fixed reservoir)

    Cup Number(any specified cup location)

    Most applications recommend using the TWR as the wash source. This configura-

    tion allows a two-stage rinsing process.

    Use the fixed reservoir as the recommended alternative to the TWR. Using a speci-

    fied cup or vial location as the wash source is possible, but not recommended due to

    volume limitations.

    Table 3.2. Recommended Values for Syringe Sizes

    Syringe Size(L)

    RecommendedSpeed

    Recommended CupSize

    Rack Size forRecommended Cup

    1,000 50 2-mL 90-position

    2,500 35 4-mL 60-position with shelf

    5,000 25 8-mL 90-position

    10,000 22 12-mL 60-position

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    22/101

    Configuration

    WinFLOW Software Operators Manual: Chapter 3 14

    Rev. 2.3

    2. Choose a Diluent Source:

    Reservoir

    Cup Number

    3. Click OKto close the Wash and Diluent dialog box.

    Rack Placement Setup the racks being used with the RA Sampler by clicking Rack Placement in theRA Configuration dialog box.The Rack Placement dialog box appears (Figure 3.8).

    Figure 3.8. Rack Placement dialog box

    The RA Samplers default configuration uses three sample racks and one standards

    rack. Use up to six sample racks:

    The preprogrammed sample rack sizes are 21, 24, 40, 60, 90, Incubtr2, and

    Incubtr3.

    The default sample rack size is 90.

    The available standard rack sizes are 10 and 20.

    The default standards rack size is 20.

    Click OKto close the Rack Placement dialog box.

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    23/101

    Configuration

    WinFLOW Software Operators Manual: Chapter 3 15

    Rev. 2.3

    Rack Editor The Rack Editor configures the RA Sampler to use virtually any size or type of rack.Program the rack size and type by clicking Custom Rack Editorfrom the RA

    Configuration dialog box. The Rack Editor dialog box appears (Figure 3.9).

    Figure 3.9. Rack Editor dialog box

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    24/101

    Configuration

    WinFLOW Software Operators Manual: Chapter 3 16

    Rev. 2.3

    To use racks other than those provided with the RA Sampler, measure the rack and vial

    slot dimensions and the distance between each vial slot. Enter these measurements into

    the corresponding locations in the Rack Editor. When saved, custom racks display as an

    option in the Rack Placement screen. Create both sample and standard racks using the

    following steps.

    1. Click Add New Type. Type the custom rack name and click OK.

    2. Select Add Cups. The Add Cups dialog box appears (see Figure 3.10).

    Figure 3.10. Add Cups dialog box

    3. Enter the following information in the corresponding fields:

    4. Click OK. Click Close to exit the Rack Editor dialog box.

    First Cup #: Number between 1 and 99

    Cup Wid (mm): Cup width in millimeters

    Number of Cups: Number of rows (left to right) and number of

    columns (back to front)

    First Cup Position (mm): Distance in millimeters from the standard

    trays upper left corner (closest to the samplerbody) to the middle of the first cup

    Inter-Cup Spacing (mm): Distance from the cup-to-cup center in

    millimeters

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    25/101

    Configuration

    WinFLOW Software Operators Manual: Chapter 3 17

    Rev. 2.3

    5. Click Rack Placementin the RA Sampler Configuration dialog box to access the

    Rack Placement dialog box (Figure 3.7).

    Figure 3.11. Rack Placement dialog box

    4. Select the custom rack name just created for the appropriate rack placement. After

    selecting a custom rack, the Align Rack dialog box appears as shown in Figure

    3.12(if the dialog box does not appear, click the Alignbutton beside the custom

    rack).

    Figure 3.12. Align Rack dialog box

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    26/101

    Configuration

    WinFLOW Software Operators Manual: Chapter 3 18

    Rev. 2.3

    6. Click OKto accept the default settings or enter new values to adjust the offset. The

    RA Sampler rack is now ready and displays the custom rack as shown in Figure

    3.13(if the custom rack screen does not appear, click OK to close the Rack

    Placement dialog box and the Custom Rack Editor button).

    Figure 3.13. Rack Editor dialog box displays a custom RA Sampler rack ready for use

    Configuring the 3360 Sampler

    1. Select 3360in the Configure WinFLOW dialog box.

    2. Click Configure Sampler. The 3090/3360 Configuration dialog box appears(Figure 3.14).

    Figure 3.14. 3090/3360 Configuration dialog box

    http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-
  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    27/101

    Configuration

    WinFLOW Software Operators Manual: Chapter 3 19

    Rev. 2.3

    Vial Depths Adjust the depth that the sample probe travels into a sample cup.

    1. Click Vial Depths. The Vial Depths dialog box appears (Figure 3.15).

    Figure 3.15. Vial Depths dialog box

    Enter the desired vial depths (in millimeters) in the corresponding boxes. The depths

    correspond to the depth of the needle when used with the following cups:

    Other settings may be used if other types of cups are used in the sampler. This

    information is utilized in the sample table, to allow the operator to include multiple cup

    types in a single run.

    Table 3.3. Vial Depths

    Cup TypeDefault Depth

    (mm)

    Cup Dimensions

    (mm)

    Cup Capacity

    (mL)

    1 150 13 x 100 8

    2 93 12 x 25 2

    3 113 12 x 30 4

    4 150 16 x 100 12

    http://-/?-http://-/?-
  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    28/101

    Configuration

    WinFLOW Software Operators Manual: Chapter 3 20

    Rev. 2.3

    Dilutor Setup If using the Autodilutor with the 3360 Sampler, click Dilutor Setupin the 3090/3360Configuration dialog box. The Dilutor Setup dialog box appears (Figure 3.16).

    Figure 3.16. Dilutor Setup dialog box

    By selecting Use Dilutor, Syringe Size and Syringe Speed settings become available

    for input.

    Use the pull-down menu to select the syringe size. The default syringe size is 2,500 L.

    Syringe size and the application determine proper syringe speed. Table 3.4shows the

    recommended speeds for each syringe size, as well as the cup size that is typically

    utilized for each syringe size.

    Table 3.4. Recommended Values for Syringe Sizes

    Syringe Size

    (L)

    Recommended

    Speed

    Recommended Cup

    Size and Type

    Rack Size for

    Recommended Cup

    2,500 23 4-mL type 38-mL type 1

    12 mL type 4

    60-position with shelf or90-position

    http://-/?-http://-/?-http://-/?-http://-/?-
  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    29/101

    Configuration

    WinFLOW Software Operators Manual: Chapter 3 21

    Rev. 2.3

    Wash and Diluent To provide an alternate source of wash or diluent solution, click the Wash and Diluentbutton. The Wash and Diluent dialog box appears (Figure 3.17).

    Figure 3.17. Wash and Diluent dialog box

    1. Choose Reservoirto draw wash/diluent from the rinse station.

    2. Click Cup Numberand supply a cup number and cup type if an alternate solution

    is needed.

    Using a specified cup or vial location as the wash source is possible, but not recom-

    mended due to volume limitations.

    Rack Placement Setup the racks being used with the 3360 Sampler by clicking Rack Placementin the3090/3360 Configuration dialog box. The Rack Placement dialog box appears (Figure

    3.18).

    Figure 3.18. Rack Placement dialog box

    http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-
  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    30/101

    Configuration

    WinFLOW Software Operators Manual: Chapter 3 22

    Rev. 2.3

    The 3360 Sampler uses four sample racks and one standards rack.

    The preprogrammed sample rack sizes are 60 and 90.

    The default sample rack size is 90.

    The available standard rack sizes are 10 and 12.

    The default standards rack size is 10.

    Peck Probe duringRinse

    Check this checkbox to rinse the sampler probe by pecking. Pecking is where the probe

    is plunged into and out of the rinse position three times between samples. This rinse

    method is typically used with SFA.

    Leave this box unchecked in order to rinse the probe by simply lowering it into the rinse

    position. This rinse method is typically used with FIA.

    Configuring the 3090 Sampler

    1. Select 3090 in the Configure WinFLOW dialog box.

    2. Click Configure Sampler. The 3090/3360 Configuration dialog box appears

    (Figure 3.19).

    Figure 3.19. 3090/3360 Configuration dialog box

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    31/101

    Configuration

    WinFLOW Software Operators Manual: Chapter 3 23

    Rev. 2.3

    Vial Depths Adjust the depth that the sample probe travels into a sample cup.

    1. Click Vial Depthsin the 3090/3360 Configuration dialog box. The Vial Depths

    dialog box appears (Figure 3.20).

    Figure 3.20. Vial Depths dialog box

    2. Enter the appropriate vial depths (in millimeters) in the corresponding boxes. The

    vial depths correspond to the depth of the needle when used with the following

    cups:

    Other settings may be used if other types of cups are used in the sampler. This

    information is utilized in the sample table to allow the operator to include multiple cup

    types in a single run.

    3. Click OKto close the Vial Depths dialog box.

    Table 3.5. Vial Depths

    Cup TypeDefault Depth

    (mm)

    Cup Dimensions

    (mm)

    Cup Capacity

    (mL)

    1 150 13 x 100 8

    2 93 12 x 25 2

    3 113 12 x 30 4

    4 150 16 x 100 12

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    32/101

    Configuration

    WinFLOW Software Operators Manual: Chapter 3 24

    Rev. 2.3

    Wash and Diluent To provide an alternate source of wash or diluent solution, click the Wash and Diluentbutton. The Wash and Diluent dialog box appears (Figure 3.21).

    Figure 3.21. Wash and Diluent dialog box

    1. Select a Wash Source:

    a. Choose Reservoirto draw wash from the rinse station.

    b. Click Cup Numberand supply a cup number and cup type if an alternate

    solution is needed.

    Using a specified cup or vial location as the wash source is possible, but not

    recommended due to volume limitations.

    2. Choose a Diluent Source.

    a. Choose Reservoirto draw diluent from the rinse station.

    b. Click Cup Numberand supply a cup number and cup type if an alternate

    solution is needed.

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    33/101

    Configuration

    WinFLOW Software Operators Manual: Chapter 3 25

    Rev. 2.3

    Rack Placement Setup the racks being used with the 3090 Sampler by clicking Rack Placementin the3090/3360 Configuration dialog box. The Rack Placement dialog box appears (Figure

    3.22).

    Figure 3.22. Rack Placement dialog box

    The 3090 uses a single sample rack and one standards rack.

    The supported sample rack sizes are 60 and 90.

    The default sample rack size is 90.

    The only supported standard rack sizes is a 10-position rack.

    Peck Probe during

    Rinse

    Check this checkbox to rinse the sampler probe by pecking. Pecking is where the probe

    is plunged into and out of the rinse position three times between samples. This rinsemethod is typically used with SFA.

    Leave this box unchecked in order to rinse the probe by simply lowering it into the rinse

    position. This rinse method is typically used with FIA.

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    34/101

    WinFLOW Software Operators Manual: Chapter 4 26

    Rev. 2.3

    Ch a p t e r 4 Introduction to

    WinFLOW

    Starting WinFLOW

    1. Double-click the WinFLOW icon or select WinFLOWfrom the Start menu.

    2. Enter the serial number included in the WinFLOW documentation.

    NOTE: WinFLOW only requests the serial number the first time the software starts.

    3. If password protection is enabled for program startup, enter the user name and

    password, and click OK. For more information on password protection see

    System Security in this chapter.

    NOTE: All instrument modules must be properly powered on and initialized for

    online operation.

    Select View DialogBox

    WinFLOW displays the Select View dialog box on startup (Figure 4.1).

    Figure 4.1. Select View dialog box

    The Select View dialog box allows easy, single-click access to WinFLOWs four

    primary options. It also provides an icon to exit WinFLOW. Table 4.1outlines

    WinFLOWs four main options.

    http://-/?-http://-/?-
  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    35/101

    Introduction to WinFLOW

    WinFLOW Software Operators Manual: Chapter 4 27

    Rev. 2.3

    When the Select View dialog box appears, the main menu bar displays File, Tools, and

    Window menus, which can also be accessed from any WinFLOW screen. The File

    menu also provides an Exit option.

    Tools Menu

    The Toolsmenu provides access to screens that establish WinFLOW preferences,

    configure instruments, and set pump speeds, valve states, and heater temperatures.

    Figure 4.2. Tools Menu Preferencesaccesses the Preferences dialog box.

    Configurationaccesses the Configuration dialog box.

    Auto Event Scheduleraccesses the Event Scheduler dialog box.

    Instrument Controlaccesses the Instrument Control dialog box.

    Pump Timerdisplays the number of hours the pumps are used.

    Table 4.1. WinFLOW main options

    Main Option Select ViewIcon

    ShortcutKey

    Small Icon Function

    Collect Data F2 Initiates data collection following

    sample table and method selection.

    Sample Table F3 Opens a table displaying sample

    cup number, name, type, replicate,

    dilution, and weight factors.

    Method Editor F4 Sets data collection parameters and

    calibrant and QC information.

    Data Analysis F5 Accesses results from any

    previously saved run. Contains

    peak plot, peak table, calibration

    plot, and calibration data tables.

    Exit F6 N/A Exits WinFLOW.

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    36/101

    Introduction to WinFLOW

    WinFLOW Software Operators Manual: Chapter 4 28

    Rev. 2.3

    Preferences Set global WinFLOW preferences such as default views and languages using thePreferences dialog box (Figure 4.3).

    Figure 4.3. Preferences dialog box

    Configuration The Configuration dialog box (Figure 4.4) provides information about the instrument

    platform, such as the autosampler installed with the system, the firmware version, andthe number of communicating detectors (channels), valves, and pumps.

    The Configuration dialog box allows the operator to determine if the analyzer, as well

    as each installed component, communicates with the computer. Selecting Configure

    when the analyzer or a component loses communication to re-establish a link between it

    and the computer.

    Security Sets WinFLOW to request a password on startup. For more

    information see System Security on page 33in this

    chapter.

    Save Views Sets WinFLOW to display any screen on startup that

    remained open when WinFLOW closed.

    NOTE: Save Views does not bring up results and analysis files left open when

    exiting.

    Verify Close Sets WinFLOW to automatically prompt to verify theoperators intention to exit the program.

    Re-enable Helpers Implements various warning screens that can be disabled

    as operators gain experience with WinFLOW.

    Language Select from English, French, German, or Spanish. Contact

    your local OI Analytical distributor for more information

    about other available languages.

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    37/101

    Introduction to WinFLOW

    WinFLOW Software Operators Manual: Chapter 4 29

    Rev. 2.3

    Figure 4.4. Configuration dialog box

    Events Scheduler The Events Scheduler dialog box (Figure 4.5) sets what action(s) the pump(s) take(s)when any of the specified events occur, such as the end of an analysis.

    Figure 4.5. Events Scheduler dialog box

    Add Selects the event, time, and action.

    Edit Edits any previously chosen event.

    Remove Removes any previously chosen event.

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    38/101

    Introduction to WinFLOW

    WinFLOW Software Operators Manual: Chapter 4 30

    Rev. 2.3

    Instrument Control Manually set pump speeds, valve states, and heater temperatures from the InstrumentControl dialog box (Figure 4.6).

    NOTE: The heater temperature settings do not control the Power Base cartridge

    heaters.

    Figure 4.6. Instrument Control dialog box

    Pump Timer The Pump Timer dialog box (Figure 4.6) indicates the number of hours the pump(s)run. Use this feature to determine when consumables, such as pump tubes and

    membranes, need replacing. Reset the pump timer at any time.

    Figure 4.7. Pump Timer dialog box

    Window Menu

    The Windowmenu contains options for displaying open windows within WinFLOW.

    The Window menu lists all open windows. Access any open window from the Window

    menu by selecting the name or entering the number associated with the window.

    Figure 4.8. Window menu Cascadeoverlaps open windows.

    Tileplaces open windows side by side.

    Arrange Iconsplaces minimized WinFLOW screens along the bottom the Select View

    dialog box.

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    39/101

    Introduction to WinFLOW

    WinFLOW Software Operators Manual: Chapter 4 31

    Rev. 2.3

    About Menu

    The Aboutmenu contains information about software and firmware versions, as well as

    provides technical support access numbers.

    Shortcut Keys

    WinFLOW contains several keyboard shortcuts. These function keys provide an

    additional method for locating and enabling various software activities. Table 4.2

    describes the available function keys.

    These function keys duplicate the Select Viewicons, which are also available from the

    icon ribbon or the pulldown menu. Therefore, WinFLOW provides multiple avenues for

    accessing the most frequently used software features.

    Icon Ribbon

    The following list details the various icons and their purposes.

    Table 4.2. Shortcut keys

    Function Key Description

    F2 Opens the Data Collection dialog box.

    F3 Opens the Sample Table dialog box.

    F4 Opens the Method Editor dialog box.

    F5 Opens the Data Analysis dialog box.

    F12 Exits WinFLOW.

    Opens the Data Collection screen.

    Opens the Sample Table screen.

    Opens the Method Editor screen.

    Opens the Data Analysis screen.

    Initiates data display.

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    40/101

    Introduction to WinFLOW

    WinFLOW Software Operators Manual: Chapter 4 32

    Rev. 2.3

    Resets to the baseline views beginning, which deletes all data collected.

    Fast forward icon, which initiates data collection.

    Pauses data collection.

    Stops data collection.

    Zeros the baseline signal on the peak plot.

    Removes or restores the grid lines in the peak plot and calibration plot.

    Peak plot update lock freezes the baseline update used during data

    collection.

    Activates the peak editor cross.

    Deletes or reprocesses peaks.

    Recalculates results.

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    41/101

    Introduction to WinFLOW

    WinFLOW Software Operators Manual: Chapter 4 33

    Rev. 2.3

    System Security

    System security reduces unauthorized WinFLOW software access or use. It provides

    simple, password-protected access to WinFLOW and some of its features. Accessing

    the Security screen alwaysrequires a password.

    If the system password is lost, contact OI Analytical Customer Support at(800) 336-1911 or (979) 690-1711.

    1. Select Preferencesfrom the Toolsmenu to access the Preferences dialog box

    (Figure 4.3). Select the Security checkbox. The password dialog box appears

    (Figure 4.9).

    Figure 4.9. Password dialog box.

    2. Enter the user name and password. Click OKto return to the Preferences dialogbox.

    NOTE: Use the default user name and password when first activating the Security

    option. The default user name is systemand the default password is

    manager. Neither entry is case-sensitive.

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    42/101

    Introduction to WinFLOW

    WinFLOW Software Operators Manual: Chapter 4 34

    Rev. 2.3

    3. Click Set Privilegesto edit, delete, or add a new user name and password. The

    Security dialog box appears (Figure 4.10).

    Figure 4.10. Security dialog box

    4. Select Edit, Delete, or Add, and follow the instructions presented to edit, delete, or

    add a user name and password.

    WinFLOW Quick Start

    NOTE: The following instructions assume all modules and hydraulic connections are

    installed.

    1. Turn on all Flow Solution modules and the computer.

    2. Start the WinFLOW software.

    3. Verify all tubing connections are made. Initiate pump(s) rotation by pressing the

    pumps Runbutton.

    4. Click the Method Editor icon. Verify the method is configured for the analysis to

    be performed. Save any changes to the appropriate file.

    5. Click the Sample Tableicon. Alter an existing table or create a new sample table.

    Save any changes to the appropriate file.

    6. Click the Data Collecticon. Select a method and sample table.

    7. Enter an operator name, operator ID, and filename for the run results.

    8. Click the Play icon and monitor the baseline. Once the baseline stabilizes with

    startup solution, switch to reagents and again monitor the baseline until it stabilizes

    (typically 1530 minutes).

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    43/101

    Introduction to WinFLOW

    WinFLOW Software Operators Manual: Chapter 4 35

    Rev. 2.3

    9. While the baseline stabilizes, load the autosampler trays according to the sample

    table.

    10. Click the Fast Forward icon when the baseline stabilizes.

    11. Monitor the run until the SYNC peak elutes and is successfully marked. Allow the

    run to continue.

    If the operator completes the run or the run completes automatically, the data saves

    automatically to the result file defined in step 7.

    NOTE: Select the Stopicon to stop the run. Save results once the completed run

    displays.

    12. Review the results.

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    44/101

    WinFLOW Software Operators Manual: Chapter 5 36

    Rev. 2.3

    Ch a p t e r 5 Method Editor

    The WinFLOW Method Editor controls how WinFLOW collects and interprets data.

    General Description

    Open the Method Editor using one of the following ways:

    Click the Method Editor icon from the Select View dialog box,

    Click the Method Editor icon on the main icon ribbon, or

    Press [F4].

    Depending on the installed hardware, the Method Editor includes screens that modify

    all method parameters necessary for running an analysis. Use the following six screenswith flow analysis hardware:

    Flow Control Settings

    Detector/Channel Settings

    Signal Filter and Marking Settings

    Calibration/Quantitation Settings

    Timed Events Editor

    Calibrants Table

    This chapter discusses each of these screens and their associated menus. After setting

    parameters properly, these screens do not require modifying during future analyses for

    specific analytes.

    Locate each screens title at the top of the window. The window bar also contains the

    open methods name. Access each screen using one of the following ways:

    Click the screens icon in the Method Editor icon ribbon,

    Click the Method Editor icon ribbons advance or reverse arrow until the desired

    screen appears, or

    Go to the Method Editors Page menu, which lists each of the screens separatelyalong with their associated quick key functions.

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    45/101

    Method Editor

    WinFLOW Software Operators Manual: Chapter 5 37

    Rev. 2.3

    Method Editor IconRibbon

    Opening a Method A method automatically opens when accessing the Method Editor. Open a previouslysaved method or an installed method using the following steps:

    1. Go to theFilemenu and select Open.

    2. Select the method to open, then click OK.

    NOTE: If the methods platform is not the same as the current platform, a dialog box

    appears warning that other method options need updating to match the

    platform selected. If selecting Yes, WinFLOW adapts some method

    parameters to allow WinFLOW to work on the current platform.

    CAUTION:

    Always use Save Aswhenaltering an existing

    method so the original

    does not become

    overwritten.

    3. After the method loads, modify it if necessary. See Creating a Method on page 38in this chapter.

    Advances to the next screen.

    Accesses the Flow Control Settings screen.

    Accesses the Detector/Channel Settings screen.

    Accesses the Signal Filter/Marking Settings screen.

    Accesses the Calibration/Quantitation Settings screen.

    Accesses the Timed Events Editor screen.

    Accesses the Calibrants Table screen.

    Returns to the previous screen.

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    46/101

    Method Editor

    WinFLOW Software Operators Manual: Chapter 5 38

    Rev. 2.3

    Creating a Method The fastest way to create a method consists of altering an existing method and thenusing Save Asto save the new method under a new filename. WinFLOW installs with

    the following six preset methods: single and dual methods for the Flow Solution 3000

    (FS 3000) and the Flow Solution IV (FS IV), and a three-channel method for the FS IV.

    Use these methods as templates for creating new methods. Most parameters in the

    preset methods do not need changing when creating a new method for a given

    chemistry.

    Saving a Method When altering a method and the changes have not been previously saved, the Save Asdialog box automatically appears when another method opens.

    NOTE: Saving a modified method as its original name overwrites the original settings

    in that method. Save As modifies a method under a different name to maintain

    the original settings.

    Flow Control Settings ScreenSelect an instrument, activate channels, enable the baseline verification function, set the

    sampling mode, set the default vial type (when using the Model 3090, Model 3360, or

    RA Sampler), and define pump speeds from the Flow Control Settings screen

    (Figure 5.1). Access the Flow Control Settings screen by clicking its icon from the

    Method Editor icon ribbon, by going to the Pagemenu and selecting Flow Control, or

    by pressing [Shift]+[F1].

    Figure 5.1. Flow Control Settings screen

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    47/101

    Method Editor

    WinFLOW Software Operators Manual: Chapter 5 39

    Rev. 2.3

    Instrument Specify one of three instrument platform groups:

    FS III/IV/3100

    FS 3000

    Microbes-Coli

    If the platform changes, a warning message displays stating that parameters need

    updating to reflect the new platform.

    NOTE: Changing the instrument selected does not automatically change the hardware

    configuration. Although methods for any platform can be created at any time,

    a method cannot be used unless WinFLOW is configured to function with the

    current hardware (see Chapter 3, Configuration).

    Sampling Mode Set Sampling Mode to one of two options:

    Use Single Sample mode by performing the following steps:

    1. Select Single Sample as the Sampling Mode in the Flow Control Settings screen.

    2. Proceed with data collection as described in Chapter 7, Data Collection on

    page 69.

    At the point of data collection, the Single Sample Mode dialog box appears

    (Figure 5.2).

    Figure 5.2. Single Sample Mode dialog box

    Uninterrupted Allows operation with an integrated autosampler. All samples in

    a sample table analyze in a single, uninterrupted run.

    Single Sample Allows flow injection analysis (FIA) operation with or without

    an autosampler, or continuous flow analysis operation with an

    autosampler. Each sample loads and injects when the operator

    chooses.

    Interval Currently unavailable.

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    48/101

    Method Editor

    WinFLOW Software Operators Manual: Chapter 5 40

    Rev. 2.3

    3. For FIA operation without an autosampler, manually place the sample needle into

    the sample or standard solution and click Sample. The sample or standard aspirates

    into the sample loop for the time specified in the timed events table (Figure 5.8)

    (see Timed Events Editor Screen on page 52in this chapter). When the loop

    valve rotates to the inject position, place the sample needle into the wash source.

    For FIA or segmented flow analysis (SFA)-continuous flow operation with anautosampler, click Samplewhen the next sample standard is ready. The

    autosampler automatically places the sample needle into the sample or standard vial

    and extracts the sample or standard for the time specified in the timed events table

    (Figure 5.8). The sample needle then returns to the wash source.

    Figure 5.2displays after each samples cycle time completes until all the samples in

    the sample table are processed.

    The operator can also use a blank table each time a sample or standard aspirates.

    1. Select Generate Blank Tablein the Run Setup dialog box, which appears when

    data collection initiates (Figure 5.3).

    Figure 5.3. Selecting the single sample blank table option

    2. When the Single Sample Mode dialog box appears (Figure 5.2), click Edit. A blank

    table appears.

    3. Fill in the next sample or standard and close the window. WinFLOW prompts to

    save the table. Saving the table automatically begins the analysis.

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    49/101

    Method Editor

    WinFLOW Software Operators Manual: Chapter 5 41

    Rev. 2.3

    CCV/Drift Fail Automatic recalibration options require deciding on a system action following a failedcontinuing calibration verification (CCV). For more information on CCVs, see

    Continuing Calibration Verification on page 56in this chapter.

    Vial Types Set Vial Types when using the Model 3090, Model 3360, or RA Samplers. WinFLOWcan use four operator-defined vial types within each analysis. Define each vial types

    depth in the WinFLOW Configuration dialog box (see Chapter 3, Configuration on

    page 7). Set the default vial type to represent the majority of the operator-defined vial

    types used.

    Off-scale (HI)Samples

    Set autodilution options when using the Model 3360 or RA Samplers with the

    Autodilutor. Enable autodilution by selecting Enable Autodilution. Once autodilution

    enables, the following options become available:

    Flag Only Places an F in the results tables flag column.

    Pause Run Flags the calibrant and aborts the analysis.

    Recalibrate Flags the calibrant and reruns the calibration section. Sampling

    resumes at the cup after the last successful CCV.

    Dilution Factors The Model 511 Autodilutor contains hardware-fixed

    dilution factors.

    The Autodilutor allows selecting a continuously variable

    set of dilution factors. Enter up to five dilution factors,

    each separated by commas. The Autodilutor uses the first

    entered dilution factor for any sample that goes over thehighest calibration standard concentration. The autodilutor

    uses subsequent operator-defined dilution factors if the

    diluted sample concentration is still too high. Although the

    autodilutor has a 1:24,000-step resolution, dilution factors

    greater than 200 should not be used.

    CCV Frequency Chooses the frequency of running CCVs with the diluted

    samples (e.g., run a CCV every 10 samples). Designate the

    CCV in the sample table after the dilution cups (see

    Figure 6.1 on page 58and Table 6.2 on page 60).

    Number of

    Followers

    Selects the number of followers to be run. A follower is the

    sample(s) after an over-limit (HI) sample. Therecommended setting is two followers.

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    50/101

    Method Editor

    WinFLOW Software Operators Manual: Chapter 5 42

    Rev. 2.3

    Insert Baselines Insert wash baseline (WBL) points at regular intervals. A WBL point forces thebaseline through that point. When using the RA, Model 3090, and Model 3360

    samplers in a segmented system, the WBL sample is taken from the wash reservoir

    instead of a sample cup. For a FIA system, the sample probe stays in the wash, but the

    valve does not inject, and carrier solution determines the baseline. Enable baseline

    verification by selecting Enable under Insert Baselines, then set the desired First

    position, Interval, and Duration.

    When enabled, the baseline verification operates on all the activated channels. Baseline

    verification points, which are designated WBL, do not appear in the sample table, but

    they do appear in the sequence table preview.

    Pump Control Control pump speed when operating the Precision Pump with the FS 3100, FS 3000,and CNSolution analyzers. Set these values in 1% increments.

    Automated Precision Pump Control Opt ions

    Refer to the method for the operating speed of the pump for the individual application.

    Operate the Precision Pump at 40% speed in a counterclockwise direction

    (40 setting) for most applications. Typically use a 10% slow speed (10 setting).

    Automated FS 3000 and CNSolution 3000 Pump Control Opt ions

    The FS 3000 and CNSolution 3000 pump typically operates at 100% full speed in aclockwise direction with the pump speed switch in the low position. A common slow

    speed setting is 50%, which means the pump turns at full speed 50% of the time and

    turns off 50% of the time. For a 30-second cycle time and a 50% slow speed setting, the

    pump operates for 15 seconds and then stops for 15 seconds.

    Start Position Provides the sample table location where the first baseline

    verification point occurs. Check the sequence table preview in the

    Data Collection screen to verify correct placement.

    Interval Sets the number of samples between baseline verifications (250).

    Duration Sets the number of baseline verifications to run consecutively (15).

    Run Speed Sets the desired pump speed during normal operation.

    Slow Speed Sets the desired pump speed used when the Pump Slow command

    activates either manually or automatically. For manual operation,

    see Chapter 4, Instrument Control on page 30. For automatic

    operation, see Chapter 4, Events Scheduler on page 29.

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    51/101

    Method Editor

    WinFLOW Software Operators Manual: Chapter 5 43

    Rev. 2.3

    Detector/Channel Settings Screen

    The Detector/Channel Settings screen (Figure 5.4) sets parameters for each detector

    channel individually. Update information for a particular channel by selecting the

    channels line in the Channels list box. The information on the screen pertains only to

    the highlighted channel. Only available channels appear in the Detector/Channel

    Settings screen. Access the Detector/Channel Settings screen by clicking its icon fromthe Method Editor icon ribbon, by going to the Pagemenu and selectingDetector/

    Channel, or by pressing [Shift]+[F2].

    Figure 5.4. Detector/Channel Settings screen

    Channels The Channelslistbox displays all the available channels. The number of availablechannels depends on the number of installed detectors. A channel presents as available

    following successful communication with a detector.

    Activate a channel by double-clicking the channel name. A number sign (#) sign

    appears to the left of the activated channel name (Figure 5.4). Deactivate a channel by

    double-clicking the channel name again. Method parameters only apply to active

    channels. Therefore, channels must be activated before moving to other Method Editor

    screens.

    NOTE: If opening a method that designates an unavailable channel, an X appears

    on the channel names left. Re-establish contact with the channel by going tothe Main menu and selecting Setup and Configure. If unsuccessful, the

    systems exits and restarts WinFLOW.

    Name Designate a nameto describe the chemistry being run. WinFLOW uses the channeldesignation (e.g., Channel 1) as the default name.

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    52/101

    Method Editor

    WinFLOW Software Operators Manual: Chapter 5 44

    Rev. 2.3

    Methodology Designate a methodology that describe the specific chemistry being run.

    CAS Number Enter the Chemical Abstract Service (CAS) number for the standard used in theselected channel, if applicable.

    Detector Settings

    Mode Sets the detector mode as Photometric, Amperometric, ISE

    (ion-selective electrode), Fluorometric, or Matrix(seawater

    matrix). Select Photometric for the Expanded Range(ER)

    detector. Select Amperometric for the Amperometric detector

    and for the analog-to-digital (A/D) converter. Use ISE for all ion-

    selective probes such as pH or fluoride. Use Fluorometric with

    the fluorometric plug-in detector.

    Acquisition Rate Determines the data collection rate in Hz (or points per second).

    Options include 1, 2, 3, 5, or 10 Hz. The 1-Hz default valueusually suffices. When running very low analyte levels,

    increasing the acquisition rate can result in a sensitivity increase.

    When operating multiple channels, set the acquisition rate to 1

    Hz so the computer does not overload with data.

    Cell Potential Set the cell potential in amperometric mode using the detectors

    10-turn precision potentiometer. Setting cell potential via

    software is a future option.

    A-D Gain Sets the analog-to-digital gain for the sample and reference.

    Options include 1, 2, 4, 8, 16, 32, 64, 128, or Auto. Set Sample to

    Autofor all photometric detectors. The photometric detectorautomatically sets the optimum gain based on the amount of light

    reaching the photodiode. Set the amperometric detector to a fixed

    gain of 1.

    Set Reference to Auto for all photometric detectors. The

    photometric detector automatically sets the optimum gain based

    on the amount of light reaching the photodiode. The

    amperometric detector does not have a reference gain.

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    53/101

    Method Editor

    WinFLOW Software Operators Manual: Chapter 5 45

    Rev. 2.3

    ChannelSubtractions

    Subtract peak concentration, peak height, or area in one channel from peak

    concentration, peak height, or area in another channel. For example, nitrate-plus-nitrite

    analysis typically occurs on one channel and nitrite-only analysis on another. Use the

    Channel Subtractions function for automatically calculating nitrate concentration by

    subtracting nitrite concentration from nitrate-plus-nitrite concentration. Enable Channel

    Subtractions on only one of the two channels because peak concentrations in one

    channel subtract from peak concentrations in another channel. For example, channel 1peak concentrations (nitrite) subtract from channel 2 concentrations (nitrate-plus-

    nitrite). Therefore, enable Channel Subtractions on channel 1 and not channel 2.

    Subtracted data (nitrate) appears with other data in the peak results table.

    NOTE: When using Channel Subtractions, OI Analytical recommends using

    concentrations rather than peak height or area.

    Subtracted From

    Channel

    Selects the channel from which the current channel subtracts.

    In the nitrate-plus-nitrite example, the Subtract From Channel

    is the nitrate-plus-nitrite channel (channel 2).

    Subtraction Name Designates the subtracted datas name. For example, if nitrite

    peak concentrations subtract from nitrate-plus-nitrite peak

    concentrations, name the subtracted data Nitrate.

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    54/101

    Method Editor

    WinFLOW Software Operators Manual: Chapter 5 46

    Rev. 2.3

    Signal Filter and Marking Settings Screen

    Set smoothing, baseline marking, and peak marking options for each channel

    individually from the Signal Filter and Marking Settings screen (Figure 5.5).

    Access the Signal Filter and Marking Settings screen by clicking its icon from the

    Method Editor icon ribbon, by going to the Pagemenu and selecting Signal Filter/Peak Marking, or by pressing [Shift]+[F3].

    Figure 5.5. Signal Filter and Marking Settings screen

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    55/101

    Method Editor

    WinFLOW Software Operators Manual: Chapter 5 47

    Rev. 2.3

    Smoothing Select one of three software-based signal processing algorithms to reduce noise,improve precision, and ease integrating difficult peaks (small size relative to noise).

    Choose None, Moving Average, Olympic, or Savitzky-Golay. Select one of these

    algorithms in advance. If selecting None, access these algorithms after the run

    completes and apply them in the results file.

    Moving Average Selects a simple algorithm for filtering where the average of an

    input set is used to generate one output value. This algorithm

    tends to reduce peak height maxima due to the lower values on

    either side of the peak. Peak area measurements remain largely

    unaffected. Use a maximum of nine data points and one

    iteration for best results with this algorithm.

    Olympic Selects a variation of the moving average algorithm. It

    eliminates high and low values and averages the remaining

    values. This algorithm very effectively screens out spikes (high

    frequency noise). Like moving average, the olympic algorithm

    tends to reduce peak height. Peak area measurements remain

    largely unaffected. Use seven data points (11 maximum) and

    one iteration for best results with this algorithm when acquiring

    data at 1 Hz.

    Savitzky-Golay Selects an algorithm that multiplies each point by a weighting

    factor and then averages the points in a data set. The factors,

    derived from second- and third-degree polynomials, weight the

    center of the data point set and progressively lessen the

    weighting to the points further out. Unlike the other algorithms,

    this setting minimizes peak height maxima distortions. Peak

    area measurements also remain unaffected. Use several data

    points and multiple iterations for best results with this

    algorithm when acquiring data at 1 Hz.

    Points Defines the number of input data points processed to generate

    each output data point. Options include 3, 5, 7, 9, 11, 13, and

    15. The data acquisition rate is a very important parameter in

    determining the optimum number of points for a given peak

    width. The default setting uses five data points, which is

    adequate for most applications. Use a higher value when

    processing extremely wide peaks.

    Iterations Sets the repetition number that the algorithm treats the output

    of one operation (filtering algorithm) as the input for the next

    operation. It increases the degree of filtering (smoothing) whilekeeping the filter width (points) optimized for the peak width.

  • 7/23/2019 winflow4 , anlisis de cianuro - analisis intrumental

    56/101

    Method Editor

    WinFLOW Software Operators Manual: Chapter 5 48

    Rev. 2.3

    Baseline Marking Baseline Markingconsists of the following three parameters:

    InvertedChemistry

    Select Inverted Chemistry when the chemistry being run measures a decrease in

    response instead of an increase. Select this option only if the analytical method for the

    chemistry used states that it is an inverse chemistry.

    NOTE: This feature can only be used with certain detector types.

    Peak Marking

    Lead Time Sets the time in seconds before the SYNC peak, which is used

    as the initial baseline point. WinFLOW determines the

    baseline lead after marking the SYNC peak by counting

    backward from the SYNC peak marker. Be sure to set the

    baseline lead time longer than half the SYNC peaks width so

    WinFLOW does not mark the baseline on the rising slope of

    the SYNC peak itself. As a general rule, set the baseline lead

    time to 15 seconds more than the peak width. The def