rapid enhanced chemiluminescent enzyme immunoassay

7
AnalyticaChimicaActa, 266(1992)193-199 193 ElsevierSciencePublishersBV,Amsterdam Rapidenhancedchemiluminescentenzymeimmunoassay forferritinmonitoredusinginstantphotographicfilm AngelaO'Toole DepartmentofClinicalEndocrinology,Women'sHospita4Birmingham(UK) LarryJKncka DepartmentofPathologyandLaboratoryMedicine,Universityof Pennsylvania, Philadelphia,PA19104(USA) GaryHGThorpeandThomasPWhitehead WolfsonAppliedTechnologyLaboratory,WolfsonResearchLaboratories,QueenElizabethMedicalCentre,Edgbaston, BirminghamB1S2TH(UK) (Received14thJanuary1992,revisedmanuscriptreceived24thFebruary1992) Abstract Arapidchemiluminescentenzymeimmunometncassay(EIA)forserumfemtinhasbeen developedwhich combinesahalophenol-enhancedchemilummescentassayfortheperoxidase labelwithphotographicdetection (20000ASAinstantfilm)oflightemissionAssayswereperformedusingamicrotitreplateasthesolidsupport anda speciallydesignedcameralummometertomeasurelightemissionfromindividualantibodycoatedwells Theassay operatesasathresholdtestaboveacertainserumfemtmconcentrationthefilmwasfullyorpartiallyexposedwhilst belowthislevelthefilmremainedunexposedTheeffectofincubationtimeandtemperaturewereinvestigated and tworapidmimunoassaysweredevelopedInone,a10-mmincubationwithsampleandwithconjugatedetected 1ng (10µg1 -1 ) offerntm,whilst in theother,3-mmincubationsatambienttemperaturedetected40ng(400lAgI - ') of femtinTheseassayswerereliable,producedapermanentphotographicrecordoftheresultsandthefastest assay wascompletedinlessthan10min Keywords Chenulummescence,Enzymaticmethods,Immunoassay,Cameralummometer,Ferntin,Serumfemtm Thechemilummescentassayforhorseradish peromdasebasedonthe6-hydroxybenzothiazole orhalophenol-enhancedlummol-peroxidereac- tionisbothrapidandsensitiveandcanbeused asareplacementfortheconventionalcolonmet- ncassaysforperoxidaselabels [1-5] Correspondenceto LJKncka,DepartmentofPathologyand LaboratoryMedicine,HospitaloftheUniversityofPennsylva- nia,WilliamPepperLaboratory,3400SpruceStreet,Philadel- phia,PA19104-4283(USA) 0003-2670/92/$050001992 - ElsevierSciencePublishersBVAllrightsreserved Anadvantageofanenhancedchemilummes- centassayforperoxidaseisthatitisrapid(1min ascomparedtotypically,between15and30nun foracolorimetricassay)thusthetunetakento quantitateperoxidaselabelsistrivialincompari- sonwithincubationtimeswithsampleandwith conjugatematypicalenzymeunmunoassaySince itisunlikelythatasignificantlyfasterchemilumi- nescentperoxidaseassaywillbedeveloped,any furtherreductioninthetunetocompletea chemilummescentenzymeimmunoassaycanonly

Upload: an-pro

Post on 07-Sep-2015

216 views

Category:

Documents


0 download

DESCRIPTION

Rapid Enhanced Chemiluminescent Enzyme Immunoassay

TRANSCRIPT

  • Analytica ChimicaActa, 266 (1992) 193-199

    193

    Elsevier Science Publishers B V, Amsterdam

    Rapid enhanced chemiluminescent enzyme immunoassay

    for ferritin monitored using instant photographic film

    Angela O'Toole

    Department of Clinical Endocrinology, Women's Hospita4 Birmingham (UK)

    Larry J Kncka

    Department of Pathology and Laboratory Medicine, University of Pennsylvania, Philadelphia, PA 19104 (USA)

    Gary H G Thorpe and Thomas P Whitehead

    Wolfson Applied Technology Laboratory, Wolfson Research Laboratories, Queen Elizabeth Medical Centre, Edgbaston,

    Birmingham B1S 2TH (UK)

    (Received 14th January 1992, revised manuscript received 24th February 1992)

    Abstract

    A rapid chemiluminescent enzyme immunometnc assay (EIA) for serum femtin has been developed which

    combines a halophenol-enhanced chemilummescent assay for the peroxidase label with photographic detection

    (20000 ASA instant film) of light emission Assays were performed using a microtitre plate as the solid support and a

    specially designed camera lummometer to measure light emission from individual antibody coated wells The assay

    operates as a threshold test above a certain serum femtm concentration the film was fully or partially exposed whilst

    below this level the film remained unexposed The effect of incubation time and temperature were investigated and

    two rapid mimunoassays were developed In one, a 10-mm incubation with sample and with conjugate detected 1 ng

    (10 g 1 -1 ) of ferntm, whilst in the other, 3-mm incubations at ambient temperature detected 40 ng (400 lAg I- ') of

    femtin These assays were reliable, produced a permanent photographic record of the results and the fastest assay

    was completed in less than 10 min

    Keywords Chenulummescence, Enzymatic methods, Immunoassay, Camera lummometer, Ferntin, Serum femtm

    The chemilummescent assay for horseradish

    peromdase based on the 6-hydroxybenzothiazole

    or halophenol-enhanced lummol-peroxide reac-

    tion is both rapid and sensitive and can be used

    as a replacement for the conventional colonmet-

    nc assays for peroxidase labels [1-5]

    Correspondence to L J Kncka, Department of Pathology and

    Laboratory Medicine, Hospital of the University of Pennsylva-

    nia, William Pepper Laboratory, 3400 Spruce Street, Philadel-

    phia, PA 19104-4283 (USA)

    0003-2670/92/$05 00 0 1992 - Elsevier Science Publishers B V All rights reserved

    An advantage of an enhanced chemilummes-

    cent assay for peroxidase is that it is rapid (1 min

    as compared to typically, between 15 and 30 nun

    for a colorimetric assay) thus the tune taken to

    quantitate peroxidase labels is trivial in compari-

    son with incubation times with sample and with

    conjugate m a typical enzyme unmunoassay Since

    it is unlikely that a significantly faster chemilumi-

    nescent peroxidase assay will be developed, any

    further reduction in the tune to complete a

    chemilummescent enzyme immunoassay can only

  • 194

    be achieved by reducing incubation tunes For

    sandwich immunoassays such reductions are pos-

    sible but have not been extensively investigated

    For example, Yolken and Leister [6] showed that

    in an assay for H influenza type b polysaccharide

    it was possible to reduce incubation times with

    sample and with conjugate to as little as 10 min

    This assay was, however, less sensitive (by a fac-

    tor of 10) than an assay using 2-h incubations It

    was also possible to eliminate one of the mcuba-

    tions by combining the sample and conjugate in a

    single 10-min incubation

    A further advantage of the enhanced chemilu-

    mmescent peroxidase assay is that the magnitude

    of the light emission is sufficiently high to make

    possible its detection using photographic film

    The combination of chemiluminescent assays

    and photographic detection of light has formed

    the basis of some very simple and rapid assays for

    analytes such as ATP, NADH and primary bile

    acids [7-9] These operate as threshold tests

    above a certain analyte concentration the film is

    exposed whilst below this concentration it re-

    mains unexposed

    The major objective of this study was to com-

    bine together the enhanced chemilummescent as-

    say for peroxidase, an enzyme immunoassay, and

    photographic detection of light emission to pro-

    duce a rapid and simple chemilummescent en-

    zyme immunoassay which would operate as a

    threshold test An enzyme immunoassay for serum

    ferritin was chosen for study since diagnostically

    it is important to know if a patient has levels

    either lower than 10 g 1 -1 (iron deficiency

    anaemia) or higher than 400 g 1 -1 (iron over-

    load) Thus two immunoassays were constructed

    in one the threshold was set such that the film

    was only exposed when specimens with ferritin

    concentrations greater than 10 Ag 1 -1 were as-

    sayed, whilst in the other assay the threshold for

    exposure of the film was set at 400 g 1 -1

    p-Iodophenol was chosen as the enhancer since it

    is particularly effective in enhancing the light

    emission from the lummol-peroxide-peroxidase

    reaction Initially assay conditions were optimised

    using a conventional lummometer which used a

    photomultiplier tube to detect light emission The

    assays were then transferred to a camera lumi-

    A O'Toole et al /Anal Chun Acta 266 (1992) 193-199

    nometer which used instant photographic film as

    the light detector [9-11]

    EXPERIMENTAL

    Hydrogen peroxide (30%, w/v) was purchased

    from Sigma (Poole) p-Iodophenol was obtained

    from Aldrich (Gillingham) An antiferritin-per-

    oxidase conjugate was supplied by Dako Im-

    munoglobulms (Copenhagen)

    Luminol

    This was purified and converted to its sodium

    salt by recrystalization from hot sodium hydrox-

    ide (1 M) and decolorization with charcoal [12]

    Measurement of light emission

    Photomultiplter tube A lummometer designed

    and built in the laboratory based on a side-window

    photomultiplier tube (EMI Type 9781A, 94 AA

    lumen') having facilities for automatic injection

    of reagents into reaction cuvettes was used to

    measure light emission [13]

    Photographic film Polaroid Land film Type

    612 (ASA 20000) (Polaroid, Cambridge, MA)

    served as the light detector The camera lumi-

    nometer consists of a light-tight box mounted on

    top of a Polaroid film camera back Inside the

    light-tight compartment is a mask into which fits

    a 10 x 6 well piece of a microtitre plate This

    serves to isolate light emission from individual

    wells of the microtitre plate The mask is sepa-

    rated from the Polaroid film by a shutter With-

    drawal of the shutter allows the mask to drop

    onto the film thereby producing a contact print of

    glowing solutions contained in individual wells of

    the microtitre plate housed in the mask At the

    end of the exposure period the shutter is in-

    serted, thus terminating the exposure of the film,

    and the instant film is developed Degree of

    exposure of the film is proportional to the inten-

    sity of the light emission

    Solid support

    Polyvinylchlonde (PVC) M24 microtitre plates

    (Dynatech, Billmghurst) were coated with the IgG

    fraction of anti-human ferntin (Dako) by adding

  • A O'Took et aL /Anal Chum. Acta 266 (1992) 193-199

    100 l of a solution of the antibody (10 g 1 - ')

    diluted 1 m 2000 with carbonate-bicarbonate

    buffer (0 05 M, pH 9 5) to each well and mcubat-

    ing the plate at 4C overnight

    Horseradish peroxidase assay

    Colonmetnc end-point The substrate solution

    comprised a mixture of hydrogen peroxide (25

    mM) and 2,2'-azino-di-(3-ethylbenzthiazoline

    sulphonate) diammomum salt (ABTS, Boeh-

    ringer, Mannheim) 3 0 mM in sodium acetate

    buffer (0 1 M, pH 4 5) containing sodium dihy-

    drogen phosphate (0 05 M) Substrate solution

    (250 l) was added at timed intervals to each well

    and colour allowed to develop for 0 5 h A sample

    (200 l) of solution from each well was diluted

    with 1 ml of PBS [phosphate buffer (0 015 M, pH

    7 2)-sodium chloride (0 15 M)] and the ab-

    sorbance measured at 415 rim

    Enhanced chemilummescent end-point The

    substrate solution comprised sodium lummol (25

    mg) in 50 ml Tris buffer (0 1 M, pH 8 5) contain-

    ing 31 l hydrogen peroxide (30%, w/v) and

    p-iodophenol (13 6, 27 2, or 40 9 M) in dimethyl

    sulphoxide (DMSO) Substrate solution (200 l)

    was added to each well and the light emission

    recorded using a luminometer

    Ferntin standards These were prepared from a

    sample of purified human spleen ferritin as de-

    scribed previously [14]

    EM for serum ferntan Ferritin standards (100

    l) or serum samples (100 l) were incubated in

    the antibody coated wells for 1 h at room temper-

    ature The wells were then emptied and washed 3

    times using PBS-Tween (phosphate buffer 0 015

    M, pH 7 2, 0 15 M sodium chloride containing 0 5

    ml 1 - ' Tween 20) A 1 500 dilution of an antifer-

    ritin-peroxidase conjugate (100 l) in PBS-

    BSA-Tween (2 g 1 - ' BSA) was added to each

    well and incubation continued for 1 h at room

    temperature The wells were washed and dried as

    described above Bound peroxidase activity was

    then determined using either the colorimetric or

    chemiluminescent assay procedure

    In order to measure light emission using the

    photomultiplier tube based lummometer, individ-

    ual wells were cut from the plate and lodged in

    the bottom of a plastic cuvette (Sarstedt, Leices-

    195

    ter) Substrate was then pipetted directly into the

    well and the light emission measured

    Modified chemdummescent EIAs for ferntm

    Incubation tunes The effect of varying the

    incubation period was studied by assaying ferntm

    standards (0-100 g 1 - ') where the incubation

    period with sample and with conjugate were re-

    duced to 10, 5, 3 and 1 min Timing errors in the

    assays with short incubations were minimized by

    using a multi-pipette (8 channel Titertek Flow

    Laboratories, Irvine) to dispense samples and

    reagents

    Incubation temperature The above series of

    experiments were also carried out at 37C in

    order to assess the influence of incubation tem-

    perature on the assay

    Enhancer concentration Total and peak light

    emission in a chemiluminescent assay for peroxi-

    dase depend upon the concentration of the en-

    hancer (p-iodophenol) Thus the influence of dif-

    ferent concentrations of p-iodophenol (13 6, 27 2,

    40 9 M) in the substrate solution on the detec-

    tion limit for ferritin was investigated

    Kinetics of light emission The time taken to

    reach peak light emission in the halophenol-en-

    hanced chemilummescent peroxidase assay is de-

    pendent on the peroxidase concentration Thus

    the kinetics of light emission were studied by

    assaying a series of ferritin standards (0-100 g

    1 - ') and monitoring light emission for 5-10 min

    Precision studies Assessment of analytical im-

    precision is problematical for assays which pro-

    duce semi-quantitative results as in the case of

    the luminescent EIA for ferritm with photo-

    graphic detection of light emission Thus an esti-

    mate of the reliability of the assay was deter-

    mined by repeatedly analysing the series of fer-

    ritin standards and scoring the results for each

    standard (exposure of the film) on a detected/

    undetected basis

    Clinical specimens A series of serum speci-

    mens representing a range of ferntm concentra-

    tions and which had been previously assayed us-

    ing a Fernzyme kit (Abbott Laboratories, North

    Chicago, IL) were reassayed using the chemilumi-

    nescent EIA (photographic detection of light)

    The ferritin standards used in the Fernzyme kit

  • 196

    and the standards prepared from human spleen

    were cross-analysed in order to determine the

    comparability of the ferntin values assigned to

    the two sets of standards

    RESULTS

    Colonmetnc EIA for femtin

    An EIA based on PVC MTPs as the solid

    support was required since only these plates

    would fit in the camera lumiometer Thus a

    colonmetric EIA for ferntm based on polystyrene

    MTPs described earlier [12] was adapted to PVC

    plates Acceptable analytical performance was

    achieved on plates coated with a 500-fold dilution

    of antiferritin and two 1-h incubations at room

    temperature with sample and conjugate respec-

    tively This assay formed the basis of the rapid

    chemiluminescent EIA described in the following

    sections

    Chemiluminescent EIA for ferntm using a pho-

    tomultiplier tube to detect light emission

    The feasibility of performing luminescent

    EIA's with short incubation tunes was initially

    investigated using a conventional lummometer to

    detect light emission, since a photomultiplier tube

    TABLE 1

    Enhanced chemiluminescent enzyme immunoassay for serum femtin using a photomultipher tube to detect the light emission The

    assays were performed at room temperature

    a D = Detected (light emission twice the blank value), ND = not detected

    A OToole et al. /Anal. Chum Acta 266 (1992) 193-199

    is more sensitive than photographic film Only

    standards with high femtin concentrations were

    detected using shorter incubations (two 10-mm,

    two 5-min, two 3-mm, or two 1-mm) and these

    results are summarized in Table 1 Increasing the

    concentration of p-iodophenol up to 40 9 M

    increased peak light emission and hence the sen-

    sitivity of the assay For example, using 27 2 M

    p-iodophenol the 15 Ag 1 -1 ferritin standard was

    detectable whilst using 409 M the 10 g 1 -1

    standard was detectable above background [i e ,

    the signal was twice the background (blank) light

    emission] (Table 1) High concentrations of p-

    iodophenol were not studied, since they are

    known to inhibit light emission in a chemilumi-

    nescent peroxidase assay [2] Thus in subsequent

    studies using the camera lummometer a p-

    iodophenol concentration of 40 9 M was used in

    the peroxidase assay

    The time taken to reach peak light emission

    varied with femtin concentration The slowest

    kinetics (time to peak light emission, ca 3 min)

    were observed with the 10 g 1 -1 femtin stan-

    dard Thus in subsequent studies using the cam-

    era lummometer, the influence of a delay be-

    tween the initiation of the luminescent reaction

    and detection of light emission was investigated

    Since this study was directed primarily towards

    Ferritin

    (g l- ')

    Assay result

    a

    100 D D D D D D D D

    50 D D D D D D D ND

    30 D D D D D D D ND

    20 D D D D D D ND ND

    15 D D D D D D ND ND

    10 D ND D D ND D ND ND

    Incubation time (min)

    Sample 60 10 10 10 5 5 3 1

    Conjugate 60 10 10 10 5 5 3 1

    p-Iodophenol (M) 272 136 272 409 272 409 272 272

  • A O'Toole et al /Anal Chun Acta 266 (1992) 193-199

    the use of photographic film as a detector, the

    influence of a delay on the sensitivity of the

    chemilummescent femtin assay using a photo-

    multiplier tube as detector was not studied

    Chemiluminescent EM for ferntin using photo-

    graphic film to detect light emission

    The various steps in the assay are listed in

    Table 2 and typical examples of results from a

    luminescent EIA for ferritm carried out using the

    camera lummometer are shown in Fig 1 The

    blank, which consisted of p-iodophenol-lumi-

    nol-peroxide, was not detectable The exposure

    latitude of the film is such that generally, a graded

    response was not seen and assessment of results

    was on a detected (film exposed) undetected (film

    unexposed) basis hence this assay worked as a

    threshold test

    Assay speed and sensitivity

    The best combination of speed and sensitivity

    was achieved in the assay using 10-mm mcuba-

    tions with sample and with conjugate This assay

    allowed detection of the 10 g 1 - i ferritin stand-

    ard (1 ng in the well) and took ca 25 min to

    complete (Table 3) Detection of higher concen-

    trations of ferritm was not problematical and a

    400-g 1 -i standard could be detected using 3-

    min incubations with sample and conjugate at

    ambient temperature (data not shown)

    Precision

    In a limited study of precision the assay was

    found to be reliable, e g, in nine consecutive

    TABLE 2

    Flow chart for an enhanced chemilummescent EIA with pho-

    tographic detection of light emission using a camera lumi-

    nometer

    (a)

    197

    Fig 1 Photographs of light emission from enhanced chemilu-

    minescent EIAs for serum ferntin using a camera lummome-

    ter (a) Placement of serum ferntin standards (g l -1 ) on the

    microtiter plate Results are for successive 10-mm incubations

    with the standard and conjugate, (b) no delay between initia-

    tion and measurement of light emission, and (c) a 30-s delay

    analyses of the ferritin standards (10-min incuba-

    tions) all of the standards produced similar de-

    grees of exposure of the photographic film Un-

    like other assays a different detector (photo-

    graphic film) is used for each of the series of

    assays which comprise the precision study, thus

    minor variations in film sensitivity will contribute

    towards any imprecision observed

    Temperature

    The rate of antigen antibody reactions in-

    creases with temperature, thus sensitivity in sand-

    wich immunoassays can be improved by perform-

    Step Time (nun)

    Incubate with sample 1-30

    Wash < 1

    Incubate with conjugate 1-30

    Wash < 1

    Add chemiluminescent substrate < 1

    Delay period 0-2

    Expose film 025

    Develop film 0 5

  • 198

    A O'Toole et al /Anal Chun. Acta 266 (1992) 193-199

    TABLE 3

    Enhanced chemilummescent enzyme immunoassay for serum femtin using a photographic film to detect the light emission The

    assay was performed at room temperature and p-todophenol (40 9 AM) was used as the enhancer

    a

    D = Detected (film exposed), F = famt (film partially exposed), ND = not detected

    mg the assay at elevated temperatures Increasing

    the incubation temperature from room tempera-

    ture to 37C improved the sensitivity of the assays

    using very short incubation times (e g , 2 x 3 min)

    (Table 4)

    Clinical studies

    In a batch of 31 specimens the rapid enhanced

    chemilununescent EIA for serum ferritm (thresh-

    old, 400 g 1 -1) correctly identified all specimens

    with ferritin concentrations above 400 g 1 -1

    TABLE 4

    Effect of coating and incubation temperature on the perform-

    ance of an enhanced chemilummescent enzyme immunoassay

    for serum femtin p-Iodophenol (409 AM) was used as the

    enhancer and instant photographic film was used to detect the

    light emission The incubation times with sample and with

    conjugate were each 3 min

    a

    For definitions see Table 3

    Likewise the EIA with a threshold set at 10 g

    1 -1 correctly identified all specimens with ferritin

    concentrations above 10 g 1 -1 Initially, some

    difficulties were encountered because of differ-

    ences between the values for the ferntm stan-

    dards used in the enhanced chemilununescent

    EIA and in the commercial ferntm EIA used to

    assay the specimens Cross-analysis of standards

    however, established a conversion factor for com-

    parison of results from the two assays

    DISCUSSION

    This study has established the enhanced

    chemilummescent EIA for ferntm using instant

    photographic film as a viable alternative to simi-

    lar assays monitored colorimetrically The assay is

    sensitive (10 g 1 -1 ), a permanent record of the

    results is obtained and these can easily be read

    visually In its present format the assay operates

    as a threshold test and assay conditions can be

    adjusted so that only samples above a certain

    ferntin concentration will produce sufficient light

    to expose the instant film The chemiluminescent

    detection method is much quicker than conven-

    tional colorimetric assays (less than 1 mm for

    exposure and film development versus 15-30 min

    for incubation and measurement of absorbance in

    a colorimetric assay) and the total assay time can

    Femtm (g I - t) Assay result

    100 D D D D D D D D D D D D

    50 D D D D D D D ND D D D ND

    30 D D D D D D D ND D D ND ND

    20 D D D DD D D ND ND D ND ND

    15 D D D D NDF D ND ND ND ND ND

    10 D F D D ND ND F ND ND ND ND ND

    Incubation time (mm)

    Sample 60 10 10 10 5 5 5 3 3 3 3 1

    Conjugate 60 10 10 10 5 5 5 3 3 3 3 1

    Delay time (s) 0 0 30 120 0 30 120 0 0 0 120 0

    Exposure time (s) 30 30 30 15 15 15 15 15 120 300 15 300

    Femtin (g I - t) Assay result

    a

    100 D D D D

    50 ND D D D

    30 ND ND D D

    20 ND ND F D

    10 ND ND ND ND

    Delay time (s) 0 120 0 120

    Exposure time (s) 15 15 15 15

    Coating temperature (C) 4 4 4 4

    Incubation temperature (C) 20 20 37 37

  • A O'Toole et al /Anal Chum Acta 266 (1992) 193-199

    be further reduced by shortening the incubation

    times with sample and with conjugate This is,

    however, only achieved at the expense of assay

    sensitivity for example a photographic assay us-

    mg two 10-min incubations will detect 1 ng of

    ferntin whilst an assay using two 1-min incuba-

    tions can only detect 10 ng of ferritm

    In immunoextraction assays incubation times

    are usually an order of magnitude longer than the

    time taken to dispense samples or reagents This

    ensures that, for large batches, the effect of any

    discrepancies in timing between dispensing sam-

    ple or reagents for the first and last assay tube

    are minimal As incubation periods are short-

    ened, the time taken to dispense reagents eventu-

    ally becomes significant compared to the incuba-

    tion time, and timing errors can only be mini-

    mized by limiting the size of the batch of speci-

    mens tested In this study only small batches of

    specimens were tested and timing errors further

    minimized by using an eight channel multiple

    pipette It is however, possible to simultaneously

    dispense reagents to all wells of a microtitre plate

    using a 96-place multiple dispenser and this type

    of device should facilitate the development of

    rapid assays with large batches of samples

    Instant photographic film has proved a very

    convenient and sensitive detection method for

    light emission in the enhanced chemiluminescent

    EIA for ferritin Further improvements in sensi-

    tivity would be obtained by thermo-kinetic en-

    hancement of the immunological reaction and by

    using a faster film, but to-date, 20 000 ASA is the

    fastest instant film available An advantage of the

    combination of an enhanced chemiluminescent

    assay and photographic detection of light is that

    the final step in the assay can be carried out in a

    simple instrument (camera luminometer) which

    requires no power supply This is a totally chemi-

    cal assay in the peroxidase assay, chemical en-

    ergy is transformed into light emission which is

    then detected by the light-induced chemical reac-

    tions m the photographic film emulsion These

    advantages suggest that enhanced chemilummes-

    cent EIAs monitored using instant film would be

    ideal for rapid clinical laboratory and extra-

    laboratory screening tests which do not require

    high analytical sensitivity

    REFERENCES

    1 G H G Thorpe, L J Kncka, E Gillespie, S Moseley, R

    Amess, N Baggett and T P Whitehead, Anal Biochem,

    145 (1985) 96

    2 G H G Thorpe, L J Kncka and T P Whitehead, Clm

    Chem, 31 (1985) 913

    3 G H G Thorpe and L J Kncka, Methods Enzymol , 133

    (1986) 331

    4 L J Kricka and GHG Thorpe, in K. Van Dyke and R

    Van Dyke (Eds ), Luminescence Immunoassay and Molec-

    ular Applications, CRC Press, Boca Raton, FL, 1990, p

    77

    5 G H G Thorpe, L A Williams, L J Kricka, T P White-

    head, H Evans and D R Stanworth, J Immunol Meth-

    ods, 79 (1985) 57

    6 R H Yolken and F J Leister, J Clm Microbiol, 13 (1981)

    738

    7 T J N Carter, T P Whitehegd and L J Kricka, Talanta, 29

    (1982)529

    8 K Green, L J Kncka, G H G Thorpe and T P White-

    head, Talanta, 31 (1983) 173

    9 L J Kricka and G H G Thorpe, Methods Enzymol , 133

    (1986) 404

    10 R A Bunce, G H G Thorpe, J E C Gibbons, P R Killeen,

    G Ogden, L J Kricka and T P Whitehead, Analyst, 110

    (1985) 657

    11 T P Whitehead, GHG Thorpe, L J Kncka, JEC Gib-

    bons and R A Bunce, US Patent, 4879097 (1984)

    12 G Ham, R Belcher, L J Kncka and T J N Carter, Anal

    Lett, 12 (1979) 535

    13 R A Bunce, T J N Carter, L J Kricka, T P Whitehead

    and J H Kennedy, Brit Pat, 2025609A (1980)

    14 S Lmpisarn, L J Kncka and T P Whitehead, Ann Clin

    Biochem, 18 (1981) 48

    199

    page 1page 2page 3page 4page 5page 6page 7