Non‐targeted Lipidomic Analysis by Direct Infusion Mass Spectrometry

Jianzhong Chen, PhDAssistant Professor, School of Optometry, UAB

1

Page 2: Non targeted Lipidomic Analysis by Direct Infusion Mass Spectrometry€¦ · References on Contaminants 1. Keller BO, Jie Suib, Alex B. Youngc, Randy M. Whittal, Interferences and

Lipidomics: A subset of Metabolomics

2https://en.wikipedia.org/wiki/Lipidomics

Page 3: Non targeted Lipidomic Analysis by Direct Infusion Mass Spectrometry€¦ · References on Contaminants 1. Keller BO, Jie Suib, Alex B. Youngc, Randy M. Whittal, Interferences and

Lipids: Eight Categories

3Fahy E. et. al., J. Lipid Res. 2005, 839-861

Page 4: Non targeted Lipidomic Analysis by Direct Infusion Mass Spectrometry€¦ · References on Contaminants 1. Keller BO, Jie Suib, Alex B. Youngc, Randy M. Whittal, Interferences and

Mass Spectrometric Approaches for Lipidomic Analysis

• Targeted at one or several specific lipid species

• multiple reaction monitoring (MRM) 

• Targeted at one specific lipid class/subclass 

• precursor ion scan 

• Neutral loss scan

• Non‐targeted analysis of all lipid classes 

• MS/MS (Identification) combined with high resolution MS (Quantification)

• SWATH (Sequential Window Acquisition of all Theoretical fragment‐ion spectra)

4

Page 5: Non targeted Lipidomic Analysis by Direct Infusion Mass Spectrometry€¦ · References on Contaminants 1. Keller BO, Jie Suib, Alex B. Youngc, Randy M. Whittal, Interferences and

Non‐targeted Lipidomic Analysis• Advantages

• Comprehensive

• Rapid

• Big picture

• Challenges

• Relatively low sensitivity

• Full scan

• Neutral lipids

• More severe interference peaks

• Complexed data analysis

• More severe interference peaks

• Overlapping5

Page 6: Non targeted Lipidomic Analysis by Direct Infusion Mass Spectrometry€¦ · References on Contaminants 1. Keller BO, Jie Suib, Alex B. Youngc, Randy M. Whittal, Interferences and

Electrospray‐Q‐TOF Mass Spectrometer

http://www.astbury.leeds.ac.uk/facil/MStut/mstutorial.htm

Page 7: Non targeted Lipidomic Analysis by Direct Infusion Mass Spectrometry€¦ · References on Contaminants 1. Keller BO, Jie Suib, Alex B. Youngc, Randy M. Whittal, Interferences and

Bis(2‐ethylhexyl) adipate 101 detergent101 detergent

Dibutyl phthalate

101 detergent

Non‐targeted Mass Spectrometry Analysis of Lipids Before optimization

After optimization

Analyte peak

7

Page 8: Non targeted Lipidomic Analysis by Direct Infusion Mass Spectrometry€¦ · References on Contaminants 1. Keller BO, Jie Suib, Alex B. Youngc, Randy M. Whittal, Interferences and

Optimize Instrument Condition

• Electrospray probe position

• Desolvation temperature

• Flow rate

• MS profile

• Other parameters

• heating gas, nebulization gas, cone voltage/decluster voltage

8

Page 9: Non targeted Lipidomic Analysis by Direct Infusion Mass Spectrometry€¦ · References on Contaminants 1. Keller BO, Jie Suib, Alex B. Youngc, Randy M. Whittal, Interferences and

Interference Peaks in Mass Spectrometric Analysis

• Contaminants from plastics (additive, polymer)

• Multiple adducts formation: H+, NH4+, Na+, K+

• Non‐covalent adduct formation

• homo/hetero lipid dimers; 

• between lipid and impurity 

• In‐source dissociation

• Solvent degradation: CHCl3  HCl

• Carryover: previous runs, glassware, calibrants

9

Page 10: Non targeted Lipidomic Analysis by Direct Infusion Mass Spectrometry€¦ · References on Contaminants 1. Keller BO, Jie Suib, Alex B. Youngc, Randy M. Whittal, Interferences and

Common Contaminants in Samples• Plasticizers:

phthalates

esters of aliphatic dicarboxylic acids

• Polymers:

• Detergents:

• Ingredients in cosmetics and hair conditioners

44.03 DaPEG

58.04 DaPPG 

74.02 Dasilicone rubber

44.03 Da

Distearyldimethylammonium chloride m/z 550.63

Triton X‐100

10

Page 11: Non targeted Lipidomic Analysis by Direct Infusion Mass Spectrometry€¦ · References on Contaminants 1. Keller BO, Jie Suib, Alex B. Youngc, Randy M. Whittal, Interferences and

Contamination from Detergent

Detergent Triton

44 Da

44 Da44 Da

44 Da

44 Da

44 Da44 Da

44.03 Da

11

Page 12: Non targeted Lipidomic Analysis by Direct Infusion Mass Spectrometry€¦ · References on Contaminants 1. Keller BO, Jie Suib, Alex B. Youngc, Randy M. Whittal, Interferences and

Contamination from Silicone Rubber in LC/MS

74.02 Da

74

74

12

Page 13: Non targeted Lipidomic Analysis by Direct Infusion Mass Spectrometry€¦ · References on Contaminants 1. Keller BO, Jie Suib, Alex B. Youngc, Randy M. Whittal, Interferences and

Multiple adduct formation

Chen JZ, et. al, IOVS, 2013, 54: 5730‐5753.  13

Page 14: Non targeted Lipidomic Analysis by Direct Infusion Mass Spectrometry€¦ · References on Contaminants 1. Keller BO, Jie Suib, Alex B. Youngc, Randy M. Whittal, Interferences and

Identification of lipids by MS/MS

Chen JZ, et. al, IOVS, 2010, 54: 5730‐5753. 

cholesteryl ester in meibum24:1/Chol

Chen J. et al., IOVS, 2010, 51(12), 6220-6231

cholesteryl nervonate

369.4389.4

[C27H45]+

[M+Na]+

[C23H45COOH+Na]+

14

Page 15: Non targeted Lipidomic Analysis by Direct Infusion Mass Spectrometry€¦ · References on Contaminants 1. Keller BO, Jie Suib, Alex B. Youngc, Randy M. Whittal, Interferences and

Quantification of lipids by MS

Chen JZ, et. al, IOVS, 2013, 54: 5730‐5753. 

No spike

▲▲

▲▲ ▲

▲▲

▲▲WE36:1

WE36:2 WE40:1CE22:0

CE24:1

Spiked with 11 lipid standards(10 nM each)

WE38:1

WE38:0

WE40:0

CE18:2CE18:1CE18:0

15

Page 16: Non targeted Lipidomic Analysis by Direct Infusion Mass Spectrometry€¦ · References on Contaminants 1. Keller BO, Jie Suib, Alex B. Youngc, Randy M. Whittal, Interferences and

Chen JZ, et. al, IOVS, 2013, 54: 5730‐5753. 

Manual Comparison of Relative Peak Intensities

Chen J. et al., IOVS, 2013, 54(8), 5730-5753

α,ω Type II diester(18:1/32:1/18:1)

ω Type I-St diester(18:1/32:1/cholesteryl)

16

Page 17: Non targeted Lipidomic Analysis by Direct Infusion Mass Spectrometry€¦ · References on Contaminants 1. Keller BO, Jie Suib, Alex B. Youngc, Randy M. Whittal, Interferences and

Diesters down‐regulated in Dry Eye Meibum

Chen J. et al., IOVS, 2013, 54(8), 5730-5753

α,ω Type II diester(18:1/32:1/18:1)

ω Type I-St diester(18:1/32:1/cholesteryl)

17

Page 18: Non targeted Lipidomic Analysis by Direct Infusion Mass Spectrometry€¦ · References on Contaminants 1. Keller BO, Jie Suib, Alex B. Youngc, Randy M. Whittal, Interferences and

SWATH: Full MS/MS Archive of Every Compound in Sample

18

Q1 Q2

Fast Q1 precursor selection step-wise through mass range

CID Fragmentation Direct infusion, flow injection, and lipid-class targeted LC techniques

Collection of High resolution MS/MS

https://www.uab.edu/proteomics/metabolomics/symposium/symoposium_dec2012.php

Page 19: Non targeted Lipidomic Analysis by Direct Infusion Mass Spectrometry€¦ · References on Contaminants 1. Keller BO, Jie Suib, Alex B. Youngc, Randy M. Whittal, Interferences and

• Minimizing contaminants/interference peaks is particularly important for non‐targeted lipidomic analysis.

• It is important to confirm the identity of lipids by MS/MS before quantifying lipids by MS.

• SWATH appears to be a promising method for non‐targeted lipidomic analysis.

Summary

19

Page 20: Non targeted Lipidomic Analysis by Direct Infusion Mass Spectrometry€¦ · References on Contaminants 1. Keller BO, Jie Suib, Alex B. Youngc, Randy M. Whittal, Interferences and

References on Contaminants

1. Keller BO, Jie Suib, Alex B. Youngc, Randy M. Whittal, Interferences and contaminants encounterd in modern mass spectrometry, Analytica Chimica Acta, 2008, 627: 71‐81

2. Ende M, Spiteller G, Contaminants in mass spectrometry, Mass Spectrometry Review, 1982, 1: 29‐62  