finite element modeling of the mechanical effects of the uv laser ablation of polymer coatings
TRANSCRIPT
www.elsevier.com/locate/apsusc
Available online at www.sciencedirect.com
Applied Surface Science 254 (2008) 3531–3539
Finite element modeling of the mechanical effects of the UV
laser ablation of polymer coatings
Ioannis N. Koukoulis a, Christopher G. Provatidis a,*, Savas Georgiou b
a Department of Mechanical Engineering, National Technical University of Athens, Zografou Campus, GR-15773 Athens, Greeceb Institute of Electronic Structure and Laser, Foundation for Research and Technology-Hellas, 71110 Heraklion, Crete, Greece
Received 4 July 2007; received in revised form 23 November 2007; accepted 24 November 2007
Available online 3 December 2007
Abstract
Laser ablation constitutes the basis of a number of techniques aiming at the processing and diagnosis of polymeric coatings on a variety of
substrates. In all these applications, however, the issue is raised about the mechanical effects of the procedure on the substrate integrity. To this end,
we employ finite element modeling for simulating the mechanical effects of UV laser ablation on a polymer specimen, with particular emphasis on
the structural modifications that may be induced at areas away from the ablation spot. The cylindrical specimen consists of a poly(methyl-
methacrylate) (PMMA) film on a silica substrate. The analysis shows that stresses of high enough amplitude may propagate to distances far away
from the irradiated spot and may induce deleterious mechanical deformations (e.g., cracks or delaminations). The dependence of the distribution of
the tensile stresses on the thickness of the two components, as well as on size of the ablation spot area, is examined. Finally, the possibility of
growth of pre-existing defects is shown. The results are overall in very good agreement with experimental observations.
# 2007 Elsevier B.V. All rights reserved.
Keywords: UV laser ablation; Finite element modeling
1. Introduction
Laser ablation has evolved into a most important technique
for the structuring and microstructuring of materials in a wide
range of technological fields and industrial sectors [1]. In
particular, UV ablation provides the means of structuring
polymers and biopolymers with minimal, if any, deleterious
side effects. Thus, UV ablation is extensively employed in
microelectronics for drilling, structuring a post-operation
removal of polymers, in biology for scaffolding polymer
substrates for micro-arrays. On the very same phenomena relies
the use of lasers for photorefractive keratectomy as well as for
the laser restoration of painted artworks. Ablation is also often
employed as the means for producing high-amplitude ultra-
sound pulses for diagnostic purposes.
* Corresponding author at: Department of Mechanical Engineering, National
Technical University of Athens, 9 Iroon Polytechniou Avenue, Zografou
Campus, GR-15773 Athens, Greece. Tel.: +30 210 7721520;
fax: +30 210 7722347.
E-mail address: [email protected] (C.G. Provatidis).
0169-4332/$ – see front matter # 2007 Elsevier B.V. All rights reserved.
doi:10.1016/j.apsusc.2007.11.041
Given the importance of these applications, the crucial
question arises about the nature and extent of any side effects
that may compromise the integrity of the substrate in the short-
or-long run. The thermal and chemical effects of UVablation on
polymers have been investigated in detail and generally
appropriate laser irradiation conditions for their minimization
have been established. In contrast, the study of the plausible
mechanical effects of UV ablation has been rather limited.
There are three plausible sources of such impact, namely the
expansion of gaseous products formed within the irradiated
volume, the thermoelastic expansion of the laser-irradiated
volume and the back-momentum exerted by the ejected
material. Ablation has been well demonstrated to result in the
development of stress waves of even up to �1 GPa amplitude
[1–5]. It can be expected that during propagation through the
specimen, these high-amplitude waves may induce structural
modifications at areas away from the ablation spot. Thus, in
contrast to the photochemical effects, which are confined to the
laser irradiated area, the photomechanical effects of UV
ablation can be much more delocalized.
In most cases, structure formation is limited within the
irradiated area. There are, however, a few reports of
I.N. Koukoulis et al. / Applied Surface Science 254 (2008) 3531–35393532
modifications induced by UV ablation at areas away from the
irradiation spot. For instance, upon ablation of skin, damage to
layers 400 mm below the optical penetration depth was detected
through optical examination [6]. Similarly, cellular damage and
tissue tearing between the lamellar structures have been
reported for corneas upon ablation with ArF excimer irradiation
[7–9]. In a different type of application, specifically in the laser
restoration of artworks, growth of pre-existing defects has been
indicated following laser-assisted varnish removal [10,11].
To investigate further this possibility, holographic inter-
ferometry has been used for examining the structural
modifications effected upon excimer laser (l = 193 nm and
248 nm) ablation of poly(methylmethacrylate) (PMMA) and
polystyrene films cast on suprasil substrates [12,13]. Fringe
patterns, indicative of local delaminations, were clearly
resolved at various regions over the full extent of the substrate.
Importantly, no such changes were detected in the irradiation
below the ablation threshold, even after extensive irradiation
(�1000 pulses). It is also interesting that for well-annealed
(i.e., initially largely defect-free) samples, defects are minor
upon ablation with few laser pulses, but they grow rapidly in
size in the subsequent irradiation pulses.
Here, we turn to the use of finite element method (FEM) in
order to model stress propagation in the substrate upon UV laser
ablation and examine to what extent they may result in
mechanical modifications. A number of methods, including
analytical approaches [1], molecular dynamics simulations [5],
finite element methods [14], or even combinations of them [15],
have been employed for modeling UV ablation. Generally, in
these studies, the emphasis has been on assessing the relative
contributions of thermal and photomechanical (stress-genera-
tion) processes to the laser-induced material ejection. In contrast,
the present study concerns the stress distribution along the radial
direction on the area of the PMMA film. We show first that
neither heat diffusion away from the laser irradiation area nor
thermoelastic stress generation are sufficient to account for
structural effects away from the irradiation spot. Instead, the only
viable source for effecting structural modifications is through the
mechanical pressure that is exerted during ablation and
subsequently propagates within the sample. The implications
of these findings for the structural modifications that may be
effected upon laser ablation of thin polymer films are discussed.
2. Description of the model
2.1. Finite element formulation
Stress propagation in continua is described by the general
equations of equilibrium governing the linear dynamic
Table 1
Finite element models and their geometrical specifications
Model
1 2 3
PMMA film thickness (mm) 50 50 100
Silica substrate thickness (mm) 1 2 1
response of a system of finite elements, which is given by [16]:
MUðtÞ þ CUðtÞ þKUðtÞ ¼ FðtÞ (1)
where M, C and K are the mass, damping, and stiffness
matrices; F(t) is the vector of externally applied loads that
depends on the time t; and U, U, and U are the displacement,
velocity, and acceleration vectors of the finite element assem-
blage. If Rayleigh damping is assumed, the damping matrix C
is written in the form:
C ¼ aMþ bK (2)
where a and b are the so-called Rayleigh constants.
2.2. Description of geometry and finite element model
In this work, Eq. (1) was solved using the commercial
software FEM code ANSYS 10.0. A two-dimensional
axisymmetric model of the specimens is used, as this is a
common geometry of processed samples in applications. For
the simulations the specimens have dimensions typical of the
samples employed in the holographic experiments [12]—film
thickness in the 50–100 mm range, the quartz substrate
thickness of 1–2 mm and a radius of 25 mm, was developed.
The upper and lower bound were used, leading to four finite
element models (Table 1) that consist of about 7000–14,000
axisymmetric four-noded elements and about equal number of
nodes, the number depending on the dimensions of the PMMA
film as well as of the substrate. For the film of 50 mm and
100 mm thickness, the element divisions were 3 and 6,
respectively. Also, for the substrate of 1 mm and 2 mm
thickness, the element divisions were 8 and 16, respectively.
For both of them, along the radial direction of 25 mm, the
element divisions were 625. The aspect ratio of the rectangular
elements was 1:2.35 for the PMMA film and 1:3.13 for the
silica substrate, values that give good results (a ratio up to 1:4
can be considered reliable). A satisfactory element length
should be lmin/10, where lmin represents the shortest
wavelength of interest. In the case of irradiation/heating with
nanosecond pulses, the generated acoustic waves are of very
high bandwidths (up to the frequency of the laser pulse);
however, it is the lower wave modes that carry most energy. The
time step constitutes a compromise between the requirement
for sufficient time resolution and reasonable computational
times. A time step shorter than the Tmin/p, where Tmin
represents the period of the highest frequency, was considered
appropriate [16]. Additionally, two more FEM models based on
the geometrical dimensions of model 3 were analysed; model 5
assumes a small elliptical inclusion while model 6 considers a
laser spot diameter of double size.
4 5 6
100 100 + small void 100 + double Ø spot
2 1 1
Fig. 1. Loading schemes of pressure as a function of time.
I.N. Koukoulis et al. / Applied Surface Science 254 (2008) 3531–3539 3533
In all cases reported in this work, we assume a rigid binding
between the nodes in the polymer and the ones in the silica, for
otherwise delamination at the polymer/suprasil interface
behind the ‘‘irradiated area’’ occurs, thus complicating the
examination of the stress propagation further away. Never-
theless, this corresponds to the experimental results where
indeed for non-annealed (i.e., weakly adhering) films, laser
ablation is found to result almost exclusively in the
delamination of a circular area around the spot, with reduced
defect formation farther away.
2.3. Loading conditions
Although the FEM can analyze quite complex problems,
some additional simplifying assumptions were made here so as
to have a reliable but sufficiently simplified model to permit
direct physical understanding of the involved mechanisms. The
first simplifying assumption is the neglect of the temperature
distribution away from the irradiated spot after a laser pulse.
The second one is the neglect of the thermoelastically generated
acoustic wave due to the sudden thermal expansion of the spot.
The validity of these assumptions is examined in Section 3.
Upon laser ablation, a relatively small amount of material (in
the systems studied here, �1–7 mm) is ejected, but because of
the high velocities, the exerted back-momentum/recoil results
in a compressive wave of very high values [17]. For irradiation
with typical nanosecond excimer-laser pulses, the pressure
depends on the radiant exposure [3], ranging from 104 Pa at
fluences close to the ablation threshold to 109 Pa at fluences at
which intense plasma is formed; however, in the present
examination, a value of 1 GPa was used. For the simulations, a
pressure was applied uniformly and perpendicularly on the area
of ablation spot (for an irradiated area much larger than the
optical penetration depth, the stress waves develop mainly in
the axial direction, whereas the stress in the other two directions
can be neglected). Boundary restriction of displacement was set
on the periphery of the specimen.
The typical excimer laser pulse is �30 ns, but most studies
indicate that material ejection and thus the exerted recoil
continues up to 100 ns and even 1 ms [1]. Various loading
schemes have been examined; three of them are presented in
Fig. 1, where the rate of the pressure increase is double the rate
of the pressure decrease. The actual loading scheme during
laser ablation is expected to be much more complex than the
ones depicted in Fig. 1, but nevertheless the simplified ones
should capture the essential features of the laser-induced
pressure. The pressure is assumed to be uniform above a
circular laser spot of diameter d = 1 mm while the case of
doubling this diameter is also presented (model 6, Table 1). The
total time of solution was set 300 ms, time enough to observe
the highest values on stresses.
2.4. Material properties
For silica, typical static mechanical properties were used
[18]; it is clarified that it has a linear and independent from
strain-rate response to stresses up to 1 GPa [19]. On the other
hand, for PMMA, as common for thermoplastic materials, the
mechanical properties depend sensitively and in a complex
manner on the temperature and the loading rate/strain rate [20].
For instance the elastic modulus and the yield stress are reduced
by half upon a temperature change from 20 8C to 60 8C [21–
22]. The Young modulus varies from 1720 MPa [23] to
4000 MPa [24] for intermediate strain rates ðe ¼ 18:6 s�1Þ, but
it reaches a value of �17,000 MPa for compression at high
strain rates [25]. The yield stress for compression is reported to
be 28 MPa [25] at low strain rates, but 120 MPa at high rates,
whereas for tension, it is reported to be around 30 MPa at low
strain rate and 70 MPa at high rates [20,24]. Concerning the
Poisson ratio, reported values range from �0.28 [26] to �0.39
[27]. The reasons for this variability are related to the influence
of the polymerization degree, of the method of production and
plausibly of the environmental conditions during sample
storage. Molecular weight also affects the fatigue resistance
[28,29]. For the simulations, the values assigned to the material
properties of PMMA were the averaged ones of those found in
literature. Finally, the elastic response of the material is
assumed to be linear, which can be justified by the fact that the
investigation concerns areas far from the spot, where the strain
rates and stresses are relatively low. The typical material
properties needed for a mechanical analysis are the elastic
modulus (E) and the Poisson ratio (n), however, more are
requested for the transient and thermal or thermoelastic
analyses. The extra material properties are the density (r),
the heat capacity (Cp), the thermal conductivity, the constant of
linear thermal expansion and the damping. The used values for
the thermal parameters are shown in Table 2.
Concerning damping, a literature survey did not reveal any
damping data for silica; also, a relevant publication [30] does
not use damping in a finite element model. On the other hand,
acoustic damping coefficient was reported for PMMA
(�10�3 Np/cm at 20 kHz, increasing to �0.88 Np/cm at
10 MHz [31]); damping is estimated to become significant
only after �100 cm and few cm for low- and high-frequency
(>10 MHz) waves, respectively. Given that the radius of
the specimen under consideration is small enough (25 mm), the
ablation-induced stress can reflect 100 times within the
polymer/substrate before it is dissipated.
Table 2
Material properties
Material property
PMMA Silica
Young’s modulus, E (GPa) 3.2 75
Poisson ratio, n 0.35 0.17
Density, r (kg/m3) 1190 2200
Specific heat capacity, Cp (J/kg 8C) 2000 (25–250 8C) 3000 (250 + 8C) 750
Thermal conductivity (W/(m K)) 0.22 1.38
Thermal expansion (mm/(m K)) 80
Damping, b 3.5E � 10
I.N. Koukoulis et al. / Applied Surface Science 254 (2008) 3531–35393534
In order to introduce the above-mentioned attenuation
measurements in our six finite element models, it was decided
to calculate adequately reliable damping coefficients (cf.
Eq. (2)). To this purpose, these experiments were reproduced in
the virtual environment of ANSYS 10.0. Due to restrictions of
the software used (ANSYS 10.0) related to the existence of two
different material sets, the constant a was considered to be zero
while the other constant b (cf. Eq. (2)) had to obtain a single
value. Therefore, wave propagation through these specimens
was accurately represented using another finite element mesh
from which the averaged Rayleigh constant b, being equivalent
to the measured attenuation (in Ref. [31]) for the first two
frequencies of interest (6 MHz and 10 MHz), was estimated by
back-calculation as shown in Table 2.
3. Results
3.1. Temperature conduction
A major concern in modeling the results of UV ablation
concerns the influence of the high temperature rise within the
irradiated area. For PMMA at 248 nm, it is generally accepted
that a thermal mechanism dominates [1,32]. Temperatures (at
the polymer surface) have been estimated to be 800–1000 K at
the end of the laser pulse. Subsequently the substrate
temperature along the irradiation axis (assumed perpendicular
to the substrate) scales with depth z (from the film surface) and
time t as [33]:
Tðz; tÞ ¼ T0 þaeffFLASER
2Cp
expða2effDthtÞ
�expð�aeffzÞerfc
�aeff
ffiffiffiffiffiffiffiffiDthtp
� z
2ffiffiffiffiffiffiffiffiDthtp
�
þ expðaeffzÞerfc
�aeff
ffiffiffiffiffiffiffiffiDthtp
þ z
2ffiffiffiffiffiffiffiffiDthtp
��
where T0 = 300 K, aeff the effective absorption coefficient
(experimentally estimated to be �1000 cm�1), FLASER the
laser fluence, Cp the heat capacity at constant pressure, and
Dth the heat diffusivity. The equation neglects heat losses due to
the thermal decomposition of the polymer and any material
desorption; for this reason, we have limited simulations to
fluences close and below the ablation thresholds. On the other
hand, the temperature evolution on the radial direction is simply
Tðz; tÞ ¼ T0 þ ðaeffFLASER=C pÞ expð�x2e=DthtÞ, with xe denot-
ing the distance from the irradiatiated area along the radial
direction. Estimations by these formula show that heat diffusion
towards the bulk and radially is quite slow compared with the
time scales of stress propagation. This is confirmed by FEM
analyses of the temperature distribution due to heat conduction
after a laser pulse. (In these FEM analyses, an initial tempera-
ture of 1000 K was assumed at the spot for a depth of 8 mm and
then the thermal conduction is solved through time.) Further-
more, since a laser repetition rate of 1 Hz is assumed, the time
between successive laser pulses is much smaller than the heat
relaxation time ð1=a2effDthÞ so that the temperature within the
laser irradiated spot and of the sample has returned to the initial
value before irradiation with subsequent laser pulse.
3.2. Thermal expansion
The pressure rise developed in a substrate upon isochoric
heating with a heating (square pulse) can be estimated by
DP ¼ ðbaeffFLASER=rkTcVÞð1� e�u=uÞ where b is the thermal
expansion coefficient, aeff as previously the effective absorp-
tion coefficient, CV the heat capacity at constant volume, kT the
isothermal compressibility and u = tpulse/tac where tac is the
time required for an acoustic wave to traverse the irradiated
volume. The factor within the parentheses accounts for the
decrease in the pressure due to wave propagation out of the
irradiated volume. Based on typical properties for polymers,
DP for nanosecond excimer laser pulses is estimated to be
�0.1 MPa. This order of magnitude is confirmed by a transient
FEM thermoelastic analysis assuming the temperature to rise
from a usual room temperature to 1000 K in the time duration
of the laser pulse (30 ns). It is noted that if the finite element
model did not consider the melt of the PMMA (520 K), the
pressure would hardly approach the value of 1 MPa.
3.3. Stress waves
Since the influence of the temperature rise can be neglected,
the following presentation focuses on the FEM results of the
stress propagation due to the mechanical force generated by the
back momentum/recoil.
Assuming simple elastic wave theory and negligible thermal
effects, the impulsive load applied at the top of the specimen
close to its center, induces two types of elastic waves, i.e., one
Fig. 3. Radial distribution of max S1 for models 3 and 4 (PMMA film: 100 mm;
substrate: 1 and 2 mm, respectively).
I.N. Koukoulis et al. / Applied Surface Science 254 (2008) 3531–3539 3535
dilatational (P-wave, c1) of velocity and one shear (S-wave, c2)
of velocity for PMMA and for silica C1PMMA = 1839 m/s
and C2PMMA = 751 m/s whereas C1Silica = 5848 m/s and C2Sili-
ca = 3688 m/s. In addition, Rayleigh waves travel along the free
surface and the interface [33].
Moreover, it is well known that for a single P wave incident,
two waves, P and SV, are reflected. This phenomenon is
referred to as mode conversion. Similarly, for a single SV wave
incident, two waves, SVand P, are reflected. In contrast, the SH
wave reflects as itself, with no mode conversion, quite
analogous to acoustic waves reflects [33].
Within this context, based on the acoustic impedances of the
two materials, reflectivity at the polymer/quartz interface is
estimated to be 0.6; reflectivity at the free boundaries is
assumed to be �1, where the minus sign means incident
compressive wave is reflected as a tensile wave. Under-
standably, the process gets quite complicated and cannot be
solved analytically. Furthermore, the applied force causes
bending of the substrate, which further modifies the propagat-
ing stress waves. Under these circumstances, the application of
the finite element method becomes necessary.
3.4. Overview of finite element results
Several FE models have been studied in order to address the
influence of the polymer and suprasil substrate thickness, as
well as of the ablation spot area, on the stress propagation. In
addition, a calculation on one model that includes a small void
has been performed. Only the results for the first pulse are
presented, since further FEM modeling showed no significant
dependence of the estimated stress propagation/distribution on
the reduction of the PMMA film thickness within the ablated
area that occurs with successive laser pulses.
In order to present average values, the calculated stresses
refer to mid-thickness of the PMMA film. Although a typical
diagram of the maximum stress versus time consists of both
tensile and compressive values, the analysis (Figs. 2–5 and 7)
focuses on the tensile stress, because it is much more efficient
Fig. 2. Radial distribution of max S1 for models 1 and 2 (PMMA film: 50 mm;
substrate: 1 and 2 mm, respectively).
than the compressive one in resulting in material yielding [2].
The value that indicates the tensile stress is the first principal
stress (S1) which is calculated at every single point by
determining a special set of coordinate system so that the shear
stress components vanish and only the normal stresses remain
(from which the higher one is the S1) [34]; clearly, the
orientation of this system generally changes from point to point.
At every point the maximum value of S1 (max S1) through all
the time instants was obtained; it is remarkable that in all cases
tested the calculated max S1 was found to be positive, a fact that
denotes tensile stress. In general, the results for a radius of less
than 1 mm are not presented because this area is affected
considerably by the concentrated load, as well as is not an issue
of this study since delocalized effects are examined.
The distribution of max S1 is presented for the first four
models in Figs. 2 and 3, which also present the dependence of
the max S1 on substrate (suprasil) thickness for two specific
thicknesses of the PMMA film. Moreover, Fig. 4 compares
models 1 and 3 which have the same substrate thickness but
Fig. 4. Influence of the thickness of PMMA film on the radial distribution of
max S1 for models 1 and 3 (PMMA film: 50 mm and 100 mm respectively;
substrate thickness: 1 mm).
Fig. 5. Comparison of radial distribution of max S1 for model 5 and for pressure
duration 300 ns (case 1), 600 ns (case 2) and 1000 ns (case 3). Fig. 7. Comparison of radial distribution of max S1 for models 3 and 6 (spot
diameter 2 mm).
I.N. Koukoulis et al. / Applied Surface Science 254 (2008) 3531–35393536
different PMMA film thickness. Also, Fig. 5 illustrates the
influence of a pre-existing small void in the PMMA film at a
radius of 10 mm, model 5, for the three above-mentioned cases
of loading; the small void has an elliptical shape with the major
axis equal to 10 mm and the minor to 1 mm. The analyses of
model 5 using different loading schemes prescribe the role of
the loading duration as well as the role of a pre-existing defect
that probably exists in the mass of polymer film.
A point at a radius of 10 mm has been chosen as being far
enough from the irradiated area. At this point the displacement
in the axial direction for the models 1–4 as well as for model 6 is
shown in Fig. 6. From this figure, the 1st natural frequency for
the models with 1 mm thick substrate is seen to be f1 = 4240 Hz
and for those with 2 mm is f2 = 8475 Hz. The estimated natural
frequencies are in good agreement with the experimental ones
[35].
A comparison of the max S1 distribution between the model
6 which has larger spot size and model 3 is presented in Fig. 7
and of the stress raise factor in Fig. 8.
Fig. 6. Displacement of a node at 10 mm from the spot as a function of time for
models 1–4 and 6.
An energy aspect during loading is presented for the whole
model of PMMA film and silica substrate in Fig. 9 as well as
separately for each component in Fig. 10. The values are the
total mechanical energy (elastic and kinetic) that has been given
to the model through the work of the mechanical force.
4. Discussion
According to both analytical models and the finite-element
analysis, neither heat diffusion nor thermoelastic-generated
stress can contribute by any way in structural deformations
even at a short distance away from the irradiation spot,
nevertheless on the spot they may. Thus, the experimentally
observed effects in the holographic experiments on polymers as
well as on the tissues must be ascribed to stresses propagating
through the irradiated medium as a result of the recoil exerted
by the ejecta. Understandably, the geometry of the sample plays
an important role on the distribution of the propagating stress.
Depending on the relative thickness of the materials and also on
their absolute values, there is a maximum superposition of the
Fig. 8. Stress ratio when the diameter spot size is double/single.
Fig. 9. The total mechanical energy for the whole model at t1 = 200 ns (peak of
pressure value) and the t2 = 600 ns (end of loading).
I.N. Koukoulis et al. / Applied Surface Science 254 (2008) 3531–3539 3537
various traveling waves at specific locations, and these
locations are expected to be most prone to failure. Close to
the irradiation spot, the waveforms exhibit several oscillations
due to multiple reflections of the waves. Further away, the
amplitudes decrease and oscillations are hardly discernible, as a
result of diffraction resulting in the distortion of the stress profile.
In the present case, diffraction is due to the finite size of the laser-
affected area (A). At distances d beyond the diffraction length
defined as Zf = A/(4lac), lac is the wavelength of the acoustic
wave, the initially planar acoustic wave starts becoming
spherical into a cone with apex angle 2# ¼ 2 arcsinðd=2zfÞ. In
the studied samples consisting of a thin layer on a substrate with
different mechanical properties, wave propagation deviates
somewhat from this simple formula.
The models with the same film thickness exhibit a
dependence on the substrate thickness as seen in Figs. 2 and
3. In the former figure the differences are quite large though in
the latter the differences are concentrated in a radius of 10 mm;
further away the results are almost the same. The observation of
those results indicates the importance of the substrate thickness
and is shown clearly by the aspect of energy. The models with
Fig. 10. The total mechanical energy for the substrate and the film at the t1
thicker substrate gain less energy at the end of loading time, in
particular the energy that the substrate will gain is a function of
its own thickness (in the range of polymer thickness which is
investigated) (Figs. 9 and 10).
In the comparison between the models that have the same
substrate thickness, i.e., between models 1 and 3 and between 2
and 4, there are similar results. However, because of the
dependence on PMMA thickness, the differences in the max S1
distribution (Fig. 3) between the models 3 and 4 are much
smaller than models 1 and 2, although they have the same
geometrical differences (Fig. 2).
This can be justified by the fact of different gain of energy as
shown in Figs. 9 and 10. The models with the same film
thickness contain the same energy as the pressure takes its peak
value but when the application of the load stops the mechanical
energy that remains in the specimen is less in the models with
2 mm substrate compared to those with 1 mm substrate. The
energy is dispersed to the substrate or reduced due to the
negative work of the external force. This energy aspect can give
an explanation for the stress results. Those with the same film
thickness take the same energy but the substrate thickness is the
one that defines how much energy remains at the end of loading.
The energy on the substrate at the end of loading depends only
on its own thickness.
Additionally, the energy at the film at the models 3–4 is
almost the same in contrast with the models 1–2 where the
energy that remains within the film is a function of the substrate
thickness; the film of 50 mm absorbs less mechanical energy
than the film of 100 mm (Fig. 9) at the time of the peak value of
applied pressure (1 GPa).
Concerning the values and the distribution of max S1, we can
distinguish two main cases. The first refers to a relatively thick
substrate (2 mm: models 2 and 4) where within the thicker film
significantly higher values appear. The second refers to a
relatively thin substrate (1mm: models 1 and 3) where within
the thicker film slightly higher values appear (Fig. 4). The
values are somewhat similar for the area with a radius larger
than 10 mm. As a rule of thumb, the results show that stress
= 200 ns (peak of pressure value) and the t2 = 600 ns (end of loading).
I.N. Koukoulis et al. / Applied Surface Science 254 (2008) 3531–35393538
amplitude scales with the film thickness and inversely with the
silica thickness.
Alterations in geometrical characteristics change the
dynamic behavior of the specimen. As shown in Fig. 6, the
vertical displacement reflects the movement of the substrate.
Indeed, the maximum displacement is observed at �50 ms for
the 1 mm silica substrate, and at 25 ms for the 2 mm one, which
is four and two times longer, relatively, than the time that takes
for the slowest elastic wave to reach at the same point.
Understandably, its amplitude is determined by the quartz
thickness. The amplitude is�8.5 mm for the models with 1-mm
thick substrate and�2.2 mm for those with the 2-mm thickness,
as expected from the fact that the initial substrate bending
affected by a given force is inverse proportional to its thickness.
The stress rise is due to a combination of the stress waves in
the film and in the substrate, as well as of the bending of the
substrate. It is remarkable that the highest value is observed at
time 67 ms, therefore, suggesting that structural modifications
may be initiated at times well after the 30 ns excimer laser
pulse.
Concerning the influence of the irradiated area, expectably,
doubling the radius of the ablation spot (all other geometrical
parameters kept the same), and correspondingly the exerted
force (constant pressure), results in much higher maximum
tensile stresses (Fig. 7). Yet, the stress raise factor is not four
throughout the sample (as would be expected from a linear
static problem) but it varies with the location, evidently due to
the different superposition of the traveling waves. For the
particular models presented here, the stress ratio (Fig. 8) has a
large variation up to a radius of �7 mm around the irradiated
spot taking values from �1.5 to 5 and a smaller one, about 3,
further away. Although the amplitude of stresses is not four
times higher, the vertical displacement is so (e.g., in Fig. 6,
�8.5 mm for model 3 versus 33.7 mm for model 6). This
difference in the scaling of the displacements versus that of
stresses further shows that the vertical displacements are mainly
determined by the substrate bending as well as the total force
applied.
We also observe that the time of loading is essential. The
influence is not only at the area of the void but also at the
distribution of the stresses values all over the radial direction.
The rise of the stress values is much higher between cases 1 and
2 than between cases 2 and 3, which means that the max S1
stress distribution is inversely proportional to the duration of
the loading case in a non-linear way. The explanation for this
influence is that when the time is long the specimen has the time
to move up as the pressure diminishes and the pressure takes
back a part of the given energy through the negative mechanical
work. In contrast, for short loading times, as the specimen does
not have the time to move up, the pressure continues to push it
down, resulting thus in the transfer of a higher mechanical work
even if it has lower value. Understandably there is a threshold of
time reduction where the applied load gives the major of energy
and if the time of enforcement gets a lower value will give less
due to lack of time in this case.
From the standpoint of applications, it is important to know
if these stresses propagating into a sample upon laser ablation
can result in damage. Assuming brittle fracture, failure occurs
when the S1 reaches the uniaxial tension strength. Since for
PMMA, this quantity is less than 15 MPa, it would appear that
the stress field remains within safe conditions, at least for a
radius larger than 5 mm. Of course, the stress required for
failure is sharply decreased in the presence of defects/cracks. A
pre-existing defect can result in a region of a very high
concentration of stresses and after some loading cycles, a crack
of high enough size may develop, as the experimental results
indicate [11,12]. The FEM model indicates that the existence of
the void inside the PMMA film results in stresses of �31 MPa
for case 1, �53 MPa for case 2 and �120 MPa for case 3
(Fig. 5). For case 3, the stresses are high enough to cause crack
and crack propagation eventually after some cycles, but even
for the other two cases there is a possibility of reaching the yield
stress, depending on the mechanical properties of the film.
These are in qualitative correspondence with Griffith’s theory
of brittle fracture, according to which a planar sharp edge crack
in a linear homogeneous, elastic solid generates a stress field
around its tip described by the relation sðr; uÞ ¼ ðKI=ffiffiffiffiffiffiffiffi2prp
Þ fwhere r and u are the polar coordinates of a reference system
with the crack tip in the origin, f is a trigonometric function and
KI ¼ sffiffiffiffiffiffipap ffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi
secðpa=wÞp
is the stress-intensity factor depen-
dent on geometry and load characteristic (s is the stress in the
undisturbed region, a is the crack size and w is the width of the
specimen). Fracture will appear when KI � KIC where KIC is the
fracture toughness of the material, which for PMMA is
1.65 MPa m1/2.
Interestingly, the computational findings indicate a rise of
the stresses in the polymer close to the edge of the specimen
(Figs. 2, 3 and 10) which can be ascribed to the ‘‘accumulation’’
of higher stresses there due to the boundary restrictions. Indeed,
in the experiments, a high number of cracks were found to be at
the specimen edge, although this was not the usual case for
delaminations.
Though the position of voids/defects can be easily specified
in models, this is not the case experimentally. In fact, for the
case of thin polymer coatings/films on quartz, there may be
residual stresses upon film casting [36], and given the weak
adhesion strength between the two materials, delamination can
happen, even in the absence of pre-existing defects. Experi-
mentally, the degree of residual stresses and of defects is
controlled by the extent of sample annealing, but of course this
is applied to the whole sample and does not provide any control
on the localization of the defect/void. Thus, it was considered
not worth the effort to try to model the actual spatial distribution
of the laser-induced voids/defects. Furthermore, it should be
noted that experimentally, defects are observed only after a
number of laser pulses.
5. Conclusions
A series of FE analyses has shown the importance of the
mechanical parameters that influence the area away from the
ablation spot. It is shown that the raise in the temperature
concerns only the ablation area and for the rest can be
disregarded. Another issue is the geometrical parameters which
I.N. Koukoulis et al. / Applied Surface Science 254 (2008) 3531–3539 3539
determine the bending which are mainly the thickness and the
flexural modulus of the silica substrate. The thickness of the
PMMA film is also very determinant due to the reliance with the
energy that the film gains and is very much affected by the
loading scheme.
Finally an already existing defect can cause a significant rise
in the stresses and drive its expansion. The size of the spot as
well as the laser fluence are important because they are the
factors which produce the value of total force. It is also
important the flexural modulus of the substrate because defines
pretty much the total bending. The remaining stresses, partially
small adhesion between the two materials or a pre-existence of
a small void can help on the appearance of small defects which
rise with an additional load through laser ablation.
Acknowledgement
The work was supported by PENED 2001 administered by
the Greek Ministry of Industry.
References
[1] D. Bauerle, Laser Processing and Chemistry, Springer, Berlin, 2000
(Chapters 12 and 13).
[2] G. Paltauf, P.E. Dyer, Photomechanical processes and effects in ablation,
Chem. Rev. 103 (2003) 487–518.
[3] A.D. Zweig, V. Venugopalan, T.F. Deutsch, Stress generated in polyimide
by excimer-laser irradiation, J. Appl. Phys. 74 (6) (1993) 4181.
[4] S. Siano, R. Pini, R. Salimbeni, Appl. Phys. Lett. 74 (1999) 1233.
[5] L.V. Zhigilei, E. Leveugle, B.J. Garrison, Y.G. Yingling, M.I. Zeifman,
Computer simulations of laser ablation of molecular substrates, Chem.
Rev. 103 (2003) 321–347.
[6] S. Watanabe, T.J. Flotte, D.J. McAuliffe, S.L. Jacques, J. Invest. Dermatol.
90 (1988) 761.
[7] A.G. Doukas, D.J. McAuliffe, T.J. Flotte, Ultrasound Med. Biol. 19 (1993)
137.
[8] S.L. Jacques (Ed.), Proceedings of Laser–Tissue Interaction IX, SPIE
Proc. Ser. 3254SPIE, Washington, 1998, and references therein.
[9] C.B. Gabrielly, E. Pacella, S. Abdolrahimzadeh, F. Regine, R. Mollo,
Ophthalmic Surg. Lasers 30 (1999) 442, and references therein.
[10] S. Georgiou, V. Zafiropulos, D. Anglos, C. Balas, V. Tornari, C. Fotakis,
Appl. Surf. Sci. 127–129 (1998) 738.
[11] V. Tornari, D. Fantidou, V. Zafiropulos, N.A. Vainos, C. Fotakis, SPIE
3411 (1998) 420.
[12] A. Bonarou, L. Antonucci, V. Tornari, S. Georgiou, C. Fotakis, Holo-
graphic interferometry for the structural diagnostics of UV laser ablation
of polymer substrates, Appl. Phys. A 73 (2001) 647–651.
[13] A. Athanassiou, E. Andreou, A. Bonarou, V. Tornari, D. Anglos, S.
Georgiou, C. Fotakis, Examination of chemical and structural modifica-
tions in the UV ablation of polymers, Appl. Surf. Sci. 197–198 (2002)
757–763.
[14] J.C. Conde, F. Lusquinos, P. Gonzalez, J. Serra, B. Leon, A. Dima, L.
Cultrera, D. Guido, A. Zocco, A. Perrone, Finite element analysis of the
initial stages of the laser ablation process, Thin Solid Films 453–454
(2004) 323–327.
[15] J.A. Smirnova, L.V. Zhigilei, B.J. Garrison, A combined molecular
dynamics and finite element method technique applied to laser induced
pressure wave propagation, Comput. Phys. Commun. 118 (1999) 11–16.
[16] K.J. Bathe, Finite Element Procedures in Engineering Analysis, Prentice-
Hall, NJ, 1982.
[17] P.E. Dyer, R. Srinivasan, Nanosecond photoacoustic studies on ultraviolet
laser ablation of organic polymers, Appl. Phys. Lett. 48 (6) (1986) 445–
447.
[18] http://www.matweb.com.
[19] R. Ochoa, T.P. Swiler, J.H. Simmons, Molecular dynamics studies of
brittle failure in silica: effect of thermal vibrations, J. Non-Cryst. Solids
128 (1991) 57–68.
[20] W. Chen, F. Lu, M. Cheng, Tension and compression tests of two polymers
under quasistatic and dynamic loading, Polym. Test. 21 (2002) 113–121.
[21] V. Jardret, P. Morel, Viscoelastic effects on the scratch resistance of
polymers: relationship between mechanical properties and scratch proper-
ties at various temperatures, Prog. Org. Coat. 48 (2003) 322–331.
[22] E.M. Arruda, M.C. Boyce, R. Jayachandran, Effects of strain rate,
temperature and thermomechanical coupling on the finite strain deforma-
tion of glassy polymers, Mech. Mater. 19 (1995) 193–212.
[23] Q.H. Shah, H. Homma, Fracture criterion of materials subjected to
temperature gradient, Int. J. Pres. Ves. Pip. 62 (1995).
[24] H. Wu, G. Ma, Y. Xia, Experimental study of tensile properties of PMMA
at intermediate strain rate, Mater. Lett. 58 (2004) 3681–3685.
[25] Z. Li, J. Lambros, Strain rate effects on the thermomechanical behavior of
polymers, Int. J. Solids Struct. 38 (20) (2001) 3549–3562.
[26] X.F. Yao, W. Xu, M.Q. Xu, K. Arakawa, T. Mada, K. Takahashi, Experi-
mental study of dynamic fracture behavior of PMMA with overlapping
offset-parallel cracks, Polym. Test. 22 (2003) 663–670.
[27] H. Wada, M. Seika, T.C. Kennedy, C.A. Calder, Investigation of loading
rate and plate thickness effects on dynamic fracture toughness of PMMA,
Eng. Fract. Mech. 54 (6) (1996) 805–811.
[28] R.P. Kusy, A.R. Greenberg, Influence of molecular weight on the dynamic
mechanical properties of poly(methyl methacrylate), J. Therm. Anal. 18
(1980) 117–126.
[29] P. Prentice, Influence of molecular weight on the fracture of poly(methyl
methacrylate) (PMMA), Polymer 24 (1983) 334–350.
[30] T.D. Bennett, L. Li, Modeling laser texturing of silicate glass, J. Appl.
Phys. 89 (2) (2001) 942–950.
[31] J.R. Assay, D.L. Lamberson, A.H. Guenther, Pressure and temperature
dependence of the acoustic velocities in polymethacrylate, J. Appl. Phys.
40 (4) (1969) 1768–1783.
[32] G. Bounos, A. Kolloch, T. Stergiannakos, E. Varatsikou, S. Georgiou,
Assessment of the thermal and structural changes in the nanosecond
irradiation of doped PMMA and PS at 308 nm, 248 nm and 193 nm via the
examination of dopant-deriving product formation, J. Appl. Phys. 98
(2005) 084317.
[33] K.F. Graff, Wave motion in elastic solids, Dover, New York, 1975.
[34] L.E. Malvern, Introduction to the mechanics of a continuous medium,
Pentice-Hall, Inc., Engelwood, NJ, 1969.
[35] E. Esposito, L. Scalise, V. Tornari, Measurement of stress waves in
polymers generated by UV laser ablation, Opt. Lasers Eng. 38 (2002)
207–215.
[36] I. McCulloch, H.T. Man, K. Song, H. Yoon, Mechanical failure in thin-film
nonlinear optical polymers: structure and processing issues, J. Appl.
Polym. Sci. 53 (5) (1994) 665–676.