aqua r.o. simply maintenance and user manual

15
Aqua R.O. Simply 1/15 AQUA R.O. SIMPLY MAINTENANCE AND USER MANUAL Distributed by: 37 Tannery Lane #06-08 Tannery House Singapore 347790 Tel: +65 6741 2994 • Fax: +65 6741 2995 S.I.A.T.A Asia Pacific Pte Ltd www.siata.com.sg

Upload: others

Post on 16-Oct-2021

4 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: AQUA R.O. SIMPLY MAINTENANCE AND USER MANUAL

Aqua R.O. Simply – 1/15

AQUA R.O. SIMPLY MAINTENANCE AND USER MANUAL

Distributed by:

37 Tannery Lane#06-08 Tannery HouseSingapore 347790Tel: +65 6741 2994 • Fax: +65 6741 2995

S.I.A.T.A Asia Pacific Pte LtdDistributed by:

www.siata.com.sg

Page 2: AQUA R.O. SIMPLY MAINTENANCE AND USER MANUAL

Aqua R.O. Simply – 2/15

CE COMPLIANCE DECLARATION

Manufacturer S.I.A.T.A. S.r.l.

Declares that the following material:

PN RO-SIMP/05

Description AQUA R.O. SIMPLY

conforms to the essential requirements of the following DIRECTIVES:

Electromagnetic Compatibility 89/336/CEE, 93/68/CEE Low Voltage 73/23/CEE, 93/68/CEE

compliance with the following regulations was verified:

EN 50081-1 General Regulations regarding Emission - part 1: residential, commercial, and light industry environments

EN 50082-1 General Regulations regarding Immunity - part 1: residential, commercial, and light industry environments

S.I.A.T.A. S.r.l. has a Quality System which conforms to the requirements of regulation: ISO 9001/UNI EN ISO 9001-ed. 1994 (Certificate No. 95.022 SGS ICS )

Managing Director LUIGI FERRALI

Page 3: AQUA R.O. SIMPLY MAINTENANCE AND USER MANUAL

Aqua R.O. Simply – 3/15

Table of Contents 1 – BASIC FEATURES ..................................................................................................................................................... 4 2 – TECHNICAL DATA .................................................................................................................................................... 5 3 – LED AND KEY MEANINGS ...................................................................................................................................... 6 4 – VIEWING STATES OF THE DISPLAY ..................................................................................................................... 6 5 – GENERAL INFORMATION ....................................................................................................................................... 7

5.1 – PACKAGING AND STORAGE ..................................................................................................................................... 7 5.2 – INSTALLATION ........................................................................................................................................................ 7 5.3 – ELECTRICAL CONNECTION ....................................................................................................................................... 7 5.4 – PROTECTIVE DEVICES.............................................................................................................................................. 7

6 – USE INSTRUCTIONS ................................................................................................................................................. 8 6.1 – START UP ................................................................................................................................................................ 8 6.2 – PROGRAMMING OF THE WORK PARAMETERS. .......................................................................................................... 8

6.2.1 – Cleaning programming and operation. .......................................................................................................... 9 6.2.2 – Conductivity control operation. .................................................................................................................... 10 6.2.3 – Pressure control operation. .......................................................................................................................... 10 6.2.4 – Inhibit signal operation. ............................................................................................................................... 11

6.3 – THE FLOATS AND SUPPLY STARTUP. ...................................................................................................................... 11 6.4 – THE MAINTENANCE PHASE. ................................................................................................................................... 12 6.5 – THE OPERATIONAL STATES. .................................................................................................................................. 12

7 – What to do if…. .......................................................................................................................................................... 13 7.1 - ... AQUA R.O. SIMPLY DOES NOT COME ON? .......................................................................................................... 13 7.3 - ... THERE IS NO POWER TO THE ELECTRIC UTILITY TERMINALS? ............................................................................. 13 7.4 - ... AQUA R.O. SIMPLY IS NOT WORKING PROPERLY ? ............................................................................................. 13

Attachment A .................................................................................................................................................................... 14 Attachment B .................................................................................................................................................................... 14

Page 4: AQUA R.O. SIMPLY MAINTENANCE AND USER MANUAL

Aqua R.O. Simply – 4/15

1 – BASIC FEATURES Aqua R.O. Simply is dedicated to the realization of desalination systems which operate by reverse osmosis. The system is controlled by conductivity, pressure, and level sensors. Aqua R.O. Simply is equipped with a buffer battery which retains saved work parameters in the case of a power supply failure. Aqua R.O. Simply is equipped with EEPROM memory where programmed data may be saved for over 10 years. Aqua R.O. Simply, like the entire line of SIATA controllers, conforms with CEE Directives and is produced in the SIATA Plant in Montespertoli which operates with a Quality System certified according to regulation

ISO 9001 / UNI EN ISO 9001.

Page 5: AQUA R.O. SIMPLY MAINTENANCE AND USER MANUAL

Aqua R.O. Simply – 5/15

2 – TECHNICAL DATA

Power supply voltage 230 Vac 10% Mains frequency 50 Hz 3% Absorbed power 1200 VA Operating temperature 0° C – 55° C Protection class IP65 Protection regarding spurious disturbances ICE 65 Trial class 4 ( 3000 V ) Container dimensions 240 mm x 185 mm x 115 mm Overall weight 1.5 Kg

Fig. 1

Fig. 1 Hidden key

Page 6: AQUA R.O. SIMPLY MAINTENANCE AND USER MANUAL

Aqua R.O. Simply – 6/15

3 – LED AND KEY MEANINGS

Tables 1 and 2 describe the respective meanings of the LEDs and keys located on the front panel of the controller as shown in fig. 1.

Tab. 1 – LED operation LED Meaning

SERVICE On during supply. STAND-BY On while awaiting conditions for supply startup.

INHIBIT On when the inhibit input is activated or proves to be closed. ALARM On when an alarm condition is activated.

Tab. 2 – Key operation Key Description

START STOP

If kept pressed at startup, access to programming is permitted. In work parameter programming, it increases the value being modified.

During service, it starts or stops supply.

DISPLAY During programming, it permits passage to subsequent data. During service, it changes the data shown on the display.

HIDDEN KEY

Located under the words Inhibit and Alarm (see fig.1), this allows the controller to be reset, requiring programming to be read from eeprom memory.

4 – VIEWING STATES OF THE DISPLAY

The display permits the visualization of the various operational states and/or data by means of the Display key.

Tab. 3 – Data viewable on the display

Display Operational state / Data visualized

Display S t - 0 The operational state.

Display 0 1 0 0 The conductivity value read by the probe.

Display 0.1.8 0 The set point value.

Display 0.0 0 0 The number of operation hours of the pump.

Page 7: AQUA R.O. SIMPLY MAINTENANCE AND USER MANUAL

Aqua R.O. Simply – 7/15

5 – GENERAL INFORMATION Installation of the controller must be carried out by qualified personnel; the installation procedures must be carried out with the device unplugged. 5.1 – Packaging and storage Packaging consists of a cardboard box equipped with a product identification label containing the following information: name of manufacturer and manufacturer PN RO-SIMP/05; the quick reference guide for installation and programming can be found inside the box. No special procedures are necessary for unpacking. The device must be stored in an environment having the following characteristics: temperature between 0°C and +55°C; relative humidity between 30 % and 95 %. 5.2 – Installation The device consists of a box in PVC with a Plexiglas cover which may be opened manually to access the keyboard and a cover which may be removed via two screws which permits access to the external connection terminal board. The box is certified and has an IP65 protection grade. That is, Aqua R.O. Simply is protected against infiltration of dust and direct jets of water. This protection is effective when the cover is closed and the cable tubes are a correct size in relation to the cable being used and are properly assembled. If these conditions are not complied with, the protection grade becomes IP40. That is, the device is not protected against infiltration of dust or water. The back panel of the box is equipped with special mechanisms for wall-mounted installation. 5.3 – Electrical connection The controller must be electrically connected according to that indicated in diagram CE0017 shown in attachment A. The device provides 230Vac outputs with the following signals :

Service pump; Service valve; Cleaning valve; Intake valve;

Attention!! Before any installation and/or maintenance operation, make sure that the controller is disconnected from the power supply. The controller's power supply must conform to that prescribed by the Low Voltage directive (73/23/CEE). Aqua R.O. Simply is equipped with a terminal for ground connection. 5.4 – Protection devices The device is protected against overcurrents by a 6.3 A delayed mains fuse mounted on the front panel and by a 0.6 A quick blow fuse inside the controller. It is furthermore protected against power supply disturbances through special filters mounted inside the controller. A special Autoreset circuit automatically intervenes if Aqua R.O. Simply operation shuts down after particularly strong disturbances.

Page 8: AQUA R.O. SIMPLY MAINTENANCE AND USER MANUAL

Aqua R.O. Simply – 8/15

6 – USE INSTRUCTIONS

6.1 – Startup

The machine is turned on using the general ON (I) - OFF (0) switch located on the front panel. This switch operates both power conductors. Power supply voltage must be 230 Vac – 50 Hz. Verify power supply voltage by consulting the adhesive labels located on the connection cover.

6.2 – Programming the work parameters.

In order to access work parameter programming, the controller must be turned off and then on again by keeping the Start / Stop key located on the panel pressed. The display will show four lines which indicate access to the programming procedure. From now until the end of the procedure, the Start / Stop key functions in Advance. That is, it permits the value being modified to be increased, whereas the Display key permits passage to subsequent data. The following table lists the programming steps:

Tab. 4 – Programming sequence

Display Operational state / Data visualized

Startup - - - - By turning on the controller with the Start / Stop key pressed, programming is accessed.

Display 0 t 0 3 Cleaning time. ( 0 – 99 )

Display 1 t 3 0 Intervention time in case conductivity is superior to the set point. ( 2 – 99 sec.)

Display 2 t 0 5 Intervention time in case of an alarm signal from the pressure input. ( 2 – 99 sec.)

Display 3 t 0 0 Intervention time in case of a signal activated by the inhibit input. ( 0 – 10 sec.)

Display 4 t 3 0 The time period between two cleanings. ( 1 – 99 )

Display 0 A t 0 The pump is activated (1) or deactivated (0) during cleaning.

Display 1 A t 0 The time period between two cleanings is expressed in hours (1) or in minutes (0).

Display 2 A t 0 Cleaning after supply is activated (1) or deactivated (0).

Display 3 A t 0 Cleaning time is expressed in minutes (1) or in seconds (0).

Display 0 H t 1 Cleaning is carried out with the intake valve open (0) or closed (1).

Display 1 H t 1 Pressure alarm input control is activated (1) or deactivated (0).

Display 2 H t 1 Conductivity value control is activated (1) or deactivated (0).

Display 3 H t 1 Cleaning during supply is activated (1) or deactivated (0).

Display 0 C t 1 Reserved.

Display 1 C t 1 Cleaning before supply is activated (1) or deactivated (0).

Page 9: AQUA R.O. SIMPLY MAINTENANCE AND USER MANUAL

Aqua R.O. Simply – 9/15

Display 2 C t 1 Not used.

Display 3 C t 1 Not used.

Display 0.1.8 0 Conductivity threshold value (set point). ( 0000 – 2999 )

Display A t 0 0 Reserved.

Display Programming is saved in static memory (eeprom).

After programming, we recommend pressing the hidden reset key in order to require programming to be read from the static memory (eeprom).

6.2.1 – Cleaning programming and operation.

By cleaning we mean the procedure used by the controller for removing concentration residues from the membrane. Cleaning programming consists of choosing operation and interval times and the procedure for carrying out the operation.

The following can be described in reference to tab. 4 in par. 6.2:

0t03 is the cleaning time, programmable from 0 to 99. This time may be expressed in minutes or in seconds based on the contents of the 3At flag, 1 to have the time in minutes, 0 to have it in seconds.

4t30 is the time period between two cleanings, programmable from 0 to 99. This time may be expressed in hours or in minutes based on the contents of the 1At flag, 1 to have the time in hours, 0 to have it in minutes.

1Ct is the flag which controls cleaning before supply. If programmed with a value of 1, cleaning is carried before supply startup with a duration time equal to the periodic cleaning time. If programmed with a value of 0, cleaning is not carried out.

2At is the flag which controls cleaning after supply. If programmed with a value of 1, cleaning will be carried out when the operator presses the Start / Stop key to stop supply or when the level inputs signal the presence of water in the collection tank. If programmed with a value of 0, cleaning will not be carried out.

3Ht is the flag which controls cleaning during supply. If programmed with a value of 1, when the time period between two cleanings is up, supply will stop and cleaning begins. When cleaning has finished supply starts up again. It will stop again when the programmed time period is up. If programmed with a value of 0, cleaning will not be carried out.

0At is the flag which controls pump startup during cleaning. If programmed with a value of 1, cleaning will be carried out with the pump on.

0Ht is the flag which controls the intake electrovalve during cleaning. If programmed with a value of 1, cleaning will be carried out with the electrovalve closed. Otherwise it will be open.

Page 10: AQUA R.O. SIMPLY MAINTENANCE AND USER MANUAL

Aqua R.O. Simply – 10/15

In summary, Aqua R.O. Simply carries out 4 types of cleaning:

- Periodic cleaning during stand-by which cannot be deactivated; - Cleaning before every supply, activated when flag 1Ct is on 1; - Periodic cleaning during supply, activated when flag 3Ht is on 1; - Cleaning at the end of every supply, activated when flag 2At is on 1;

The length of such cleanings is programmed in 0txx, time expressed in seconds if 3At is on 0 or in minutes if 3At is on 1. Periodic cleanings begin when the time programmed in 4txx runs out, expressed in minutes if 1At is on 0, or in hours if 1At is on 1.

Whatever type of cleaning amongst those indicated above is being carried out, it will be carried out with the pump off if 0At is on 0 or with the pump on if 0At is on 1. Furthermore, the intake valve will be off if 0Ht is on 1, or will be on if 0Ht is on 0.

For periodic cleaning during stand-by nothing must be done, it is already activated. Simply program the times as previously indicated. For cleaning before supply as well, simply program the 1Ct flag on 1. The times are the same as those for periodic cleaning. For periodic cleaning during supply as well simply program the 3Ht flag on 1. The times are the same as those for periodic cleaning. For cleaning after supply as well, simply program the 2At flag on 1. The times are the same as those for periodic cleaning. Obviously pre-supply cleaning, periodic cleaning during supply, and cleaning after supply may be activated individually. They may all be activated, only one or two may be activated, or none of them may be activated.

6.2.2 – Conductivity control operation.

During supply, Aqua R.O. Simply carries out a control of the quality of the water by means of a conductivity probe. The value read by the probe is compared with a threshold value called the set point (value 0.1.80 in tab. 4 of par. 6.2) and if it proves to be superior, the service electrovalve is immediately closed. At this point, the time programmed in 1txx and expressed in seconds begins counting down. If the conductivity value descends below the set point value before the intervention time runs out, the service electrovalves reopen and supply continues; if the intervention time runs out and the conductivity value is still superior to the set point value, the controller is shut down by an alarm in order to avoid polluting the water of the utility. To reset the system, the operator must intervene and press the Start / Stop key. If the controller is started up again when the conductivity value is still superior to the set point value, the system will shut down again when the intervention time runs out.

All of this is carried out only if the 2Ht flag is on 1.

6.2.3 – Pressure control operation.

Aqua R.O. Simply provides an input for the connection of a pressure indicator device. Whatever type of instrument used (gauges which read high or no pressure), it must close the contacts of the controller when an alarm condition is present. When the controller reads that the alarm input is closed, the time programmed in 2txx and expressed in seconds begins to count down. When the time is up, the controller is shut down by an alarm in order to prevent damage to the system.

Page 11: AQUA R.O. SIMPLY MAINTENANCE AND USER MANUAL

Aqua R.O. Simply – 11/15

To reset the system, the operator must intervene and press the Start / Stop key. If the controller is started up again with the pressure alarm input still closed, the system will shut down again when the intervention time runs out.

All of this is carried out only if the 1Ht flag is on 1.

6.2.4 – Inhibit signal operation.

Aqua R.O. Simply provides a second input, the inhibit input, for the connection of alarm devices. Unlike the previous signals, this input is always activated and when an alarm condition has disappeared, the controller automatically resets operation. The presence of an inhibit signal is indicated by a special led which lights up on the panel. If the signal appears during stand-by, the led lights up and all controller operations are shut down including both cleaning and supply startup in automatic or manual. If the signal appears during cleaning or supply, the operation in progress is stopped and the controller lights up the led on the panel. When the signal disappears, the shutdown procedure starts up again.

No matter what type of device is connected to the inhibit input, it must close the terminals when an alarm condition is present.

The intervention time may be programmed using the 3txx parameter. This may be from 0 to 10 seconds; when it is on 0, controller shutdown immediately occurs when the inhibit signal appears. Otherwise, it must run for the time indicated before the controller will be stopped.

6.3 – The floats and supply startup.

The main instrument for supply startup is the level control of the water in the collection tank. This is carried out by means of two floats, one for low level and one for high level, directly connected to the controller on terminal board J2. The floats must be the normally closed type. That is, the contacts must prove to be closed when water is present.

Supply startup takes place when the low level input opens and the high level is already open; this means that if the low level input proves to be open while the high level input is closed (for example in the case of a broken float), supply will not begin.

The signal of the floats has priority over the Start / Stop key. If it is pressed when the floats indicate the presence of water and the system is therefore in stand-by, supply will startup and immediately stop. Similarly, if it is pressed when the floats indicate a lack of water and the system is therefore being supplied, the system will stop and immediately start up again.

If one does not wish to use any level control, terminals 1 and 9 of the high level must remain open while terminals 2 and 10 must remain closed by means of a jumper. This way, supply startup and shutdown are controlled exclusively by the Start / Stop key.

If one wishes to use only one float for level control of the collection tank, terminals 9 and 10 must be closed by means of a jumper and the float must be connected between the abovementioned jumper and ground available on both terminals 1 and 2.

The drawing shown in attachment B illustrates the various connection possibilities just described and the logic sequence of the float signals during the productive cycle of the controller.

Page 12: AQUA R.O. SIMPLY MAINTENANCE AND USER MANUAL

Aqua R.O. Simply – 12/15

6.4 – The maintenance phase.

The maintenance phase is activated by turning off and restarting the controller by keeping the Display key pressed. During this phase, the pump and drainage and cleaning valves are activated, whereas the service valve remains off. Signals from any connected sensors and the conductivity value are ignored. This phase is indicated on the display by the wording St-C (see par. 6.5).

To exit from maintenance, simply press the Start / Stop key.

6.5 – The operational states.

The Display key permits changing the parameter shown on the display. Table 3 illustrates which data are shown by the controller. The first of these, the operational state communicates what the controller is doing in that moment. The table below is a complete list of the operational states of the Aqua R.O. Simply:

Tab. 5 – List of operational states

Display Operational state

S t - 0 Stand-by. The controller is waiting to begin supply.

S t - 1 Delay time before pump activation.

S t - 2 Supply in progress.

S t - 3 End of supply request.

S t - 4 Cleaning after supply activated.

S t - 5 Reserved.

S t - 6 Passage phase between cleaning and stand-by.

S t - 7 Inhibit signal activated.

S t - 8 Delay time between pump and intake valve activation.

S t - 9 High conductivity alarm.

S t - H Pressure alarm.

S t - A Cleaning during supply in progress.

S t - C Maintenance phase.

S t - d Pre-supply cleaning in progress.

Page 13: AQUA R.O. SIMPLY MAINTENANCE AND USER MANUAL

Aqua R.O. Simply – 13/15

7 – What to do if….

Some basic operations for resolving small problems which can arise while using Aqua R.O. Simply are indicated below. If these suggestions do not manage to solve the problems, please call SIATA service assistance.

7.1 - ... Aqua R.O. Simply does not come on ?

1. Verify that the power plug is properly inserted into the outlet;2. Verify the outlet is properly powered;3. Verify that the power conductors are correctly inserted into the terminals and that they are properly tightened;4. Check the integrity of the 6.3 A delayed fuse located on the panel.5. If autoreset does not intervene after 10 seconds, press the hidden reset key (see par. 2 and 3)

7.3 - ... there is no power to the electric utility terminals?

1. Aqua R.O. Simply directly supplies power to the electric utilities. If the outputs do not receive power, check the6.3 A fuse located on the panel.

7.4 - ... Aqua R.O. Simply is not working properly ?

1. Verify proper programming of the controller;2. Press the hidden reset key (see par. 2 and 3).

Page 14: AQUA R.O. SIMPLY MAINTENANCE AND USER MANUAL

Aqua R.O. Simply – 14/15

Attachment A

Page 15: AQUA R.O. SIMPLY MAINTENANCE AND USER MANUAL

Aqua R.O. Simply – 15/15

Attachment B

Distributed by:

37 Tannery Lane#06-08 Tannery HouseSingapore 347790Tel: +65 6741 2994 • Fax: +65 6741 2995

S.I.A.T.A Asia Pacific Pte LtdDistributed by:

www.siata.com.sg