1-D Flat Fields for COS G130M and G160M

Tom AkeTIPS

17 June 2010

Page 2: 1-D Flat Fields for COS G130M and G160M Tom Ake TIPS 17 June 2010

Status of COS FUV Flat Fields

• SMOV Program and Results

• CALCOS Processing

• Generation of 1-D Flats Through Spectral Iteration

• 1-D Flat Field Evaluation and Achievable S/N

• Caveats and Plans

Page 3: 1-D Flat Fields for COS G130M and G160M Tom Ake TIPS 17 June 2010

SMOV Program and Results

• SMOV 11491 mapped science region of

detector with WD0320-539

• 5 cross-dispersion positions with G130M,

and 2 each with G160M and G140L

• Different cenwaves and FP-POS settings

helped separate spectral and detector features

QuickTime™ and a decompressor

are needed to see this picture.

QuickTime™ and a decompressor

are needed to see this picture.

• 2-D flats were made for each grating and segment

• Flats removed prominent dips due to grid wire shadowing, but induced some structure due to low S/N

• 1-D correction is somewhat better than 2-D

Page 4: 1-D Flat Fields for COS G130M and G160M Tom Ake TIPS 17 June 2010

CALCOS Processing

• CALCOS was designed to apply a 2-D flat field prior to spectral extraction. A unity flat is currently implemented in the pipeline.

• Grid wire shadows are the largest FPN features (20% deep, every 840 pixels)

• When CALCOS coadded different FP-POS exposures into an X1DSUM spectrum, features were reduced indepth, but appeared in more places

• For 4 FP-POS steps, 30% of pixels were affected by grid wires

• Until we have a flat field, changed CALCOS SDQFLAG keyword to ignore grid wires when creating X1DSUM spectra

QuickTime™ and a decompressor

are needed to see this picture.

Page 5: 1-D Flat Fields for COS G130M and G160M Tom Ake TIPS 17 June 2010

Generation of 1-D Flats

• Spectral iteration of X1D extracted spectra used to create 1-D flats– Technique had been developed for GHRS

– Requires data taken at different grating settings (cenwave and/or FP-POS position)

– Iterate between wavelength and pixel space in merging and correcting data sets

– Solves simultaneously for the stellar spectrum and underlying fixed pattern noise

• Each grating processed separately since spectra fall at different cross-dispersion locations

PID Program Target Grating Cenwave FP-POS 11491 SMOV Flat Field WD0320-539 G130M 1291,1309 1,3 11494 SMOV High S/N WD0947+857 G130M 1309 1, 2, 3, 4 11897 C17 Sensitivity WD0947+857 G130M 1291,1309,1327 3 11491 SMOV Flat Field WD0320-539 G160M 1600 1, 2, 3, 4 11494 SMOV High S/N WD1057+719 G160M 1600 1, 2, 3, 4 11897 C17 Sensitivity WD1057+719 G160M 1577,1589,1600,

1611,1623 3

Page 6: 1-D Flat Fields for COS G130M and G160M Tom Ake TIPS 17 June 2010

Spectral Iteration Example - WD

• Internal calibration system consists of two deuterium lamps illuminating a flat field calibration aperture (FCA)

– Light takes nearly the same optical path as an external target– Only the science areas of the detectors are illuminated, not the wavelength

calibration region – FCA (X=1750 µm, Y= 750 µm) is larger than the PSA (700 µm diameter) – Aperture mechanism moves in both dispersion and cross-dispersion directions

• External flat field calibration exposures were taken through the PSA during thermal vacuum tests in 2003 and 2006

– Preserved internal lamp– Allowed characterization of illumination angle dependence between PSA and

FCA

QuickTime™ and a decompressor

are needed to see this picture.

Page 7: 1-D Flat Fields for COS G130M and G160M Tom Ake TIPS 17 June 2010

Spectral Iteration Example - Busy Spectrum

QuickTime™ and a decompressor

are needed to see this picture.

Page 8: 1-D Flat Fields for COS G130M and G160M Tom Ake TIPS 17 June 2010

Final 1-D Flat Fields - G130M

QuickTime™ and a decompressor

are needed to see this picture.

Page 9: 1-D Flat Fields for COS G130M and G160M Tom Ake TIPS 17 June 2010

Final 1-D Flat Fields - G160M

QuickTime™ and a decompressor

are needed to see this picture.

Page 10: 1-D Flat Fields for COS G130M and G160M Tom Ake TIPS 17 June 2010

Flat Field Evaluation

QuickTime™ and a decompressor

are needed to see this picture.

G130M Segment Flat Field Statistics Segment A With Grid Wires No Grid Wires Data Set FLAT RESID FLAT RESID Total 0.050 0.036 11897 0.062 0.034 0.052 0.034 11494 0.072 0.046 0.064 0.046 11491 0.050 0.016 0.037 0.016 Segment B With Grid Wires No Grid Wires Data Set FLAT RESID FLAT RESID Total 0.052 0.035 11897 0.060 0.027 0.046 0.027 11494 0.063 0.033 0.051 0.032 11491 0.052 0.014 0.035 0.013

Consistency check performed by dividing final 1-D flat into each contributing flat • Grid wire shadows are nicely corrected• Detector dead spots leave residuals since spectra were taken at different Y position. These regions were never expected to be correctable and are flagged by CALCOS• Long wavelength end of segment A (X>11000) shows either misalignment of flats or low S/N effects

Page 11: 1-D Flat Fields for COS G130M and G160M Tom Ake TIPS 17 June 2010

Signal-to-Noise Achieved

• Distribution of P-flat variations give maximum S/N without a flat field

– Histogram of variations in each NUV stripe fit with Gaussian profile

– Widths indicate S/N (= 1/) ~ 50 per resel can be obtained without a flat

QuickTime™ and a decompressor

are needed to see this picture.

• Maximum S/N for single grating setting (~14 per pixel) reached at ~700 counts/pixel

• Current CALCOS X1DSUM ignoring grid wires improves global S/N by smoothing FPN

• Flat fielding increases S/N close to Poisson noise for single exposures. With 4 FP-POS steps, S/N=45 per pixel possible.

• Caveat - using same data to evaluate as what went into the flats, although different targets, various cenwaves, and multiple FP-POS steps were averaged

Page 12: 1-D Flat Fields for COS G130M and G160M Tom Ake TIPS 17 June 2010

Conclusions and Plans

• 1-D flats show promise. Need to check against more data

• Current flats cannot be used to correct old data since the flux calibration was created without flat fielding

• Need to investigate why long wavelength side of segment A is so noisy

• G140L still to be studied. Criteria for iteration convergence may need revision since spectrum covers only part of the detector segments