z-axis geometric efficiency in ct - impactscan.org · z-axis geometric efficiency in ct •...
TRANSCRIPT
WC 2003 August 2003
Sue Edyvean, Nicholas Keat ImPACT* (Imaging Performance Assessment of CT Scanners)LondonUK
*An MHRA Evaluation centre for the UK Department of Health
(Medicines and Healthcare productsRegulatory Agency)
www.impactscan.org
Comparison of Definitions of Geometric Efficiency in Computed Tomography Scanners
WC 2003 August 2003
Z-axis geometric efficiency in CT
• Characterises the extent of the radiation beam that is used for image creation (along the z-axis)
z-axis
WC 2003 August 2003
Geometric efficiency (g.eff.)
• Single slice – Usually little wasted dose,
as all of beam is used forimaging, except for narrow slices where sometimes post-patientcollimation is used
WC 2003 August 2003
Geometric efficiency (g.eff.)
• Multi slice– Usually more unused dose, as an
even irradiation of slices is required– Penumbral region of the beam
therefore can’t be used for imaging
WC 2003 August 2003
• IEC 60601- 2- 44 Ed. 1 1999: Original definition– “the full width at half maximum of the sensitivity profile
expressed as percentage of the full width at half maximum of the dose profile.”
• IEC 60601- 2- 44 Ed. 2 Am.1 2003: New definition – “the integral of the dose profile along the z-direction,
integrated over the range subtended by the detector elements used during acquisition, expressed as a percentage of the total integral of the dose profile in the z-direction”
• g.eff. to be displayed on scan console when less than 70%
Definitions for z-axis geometric efficiency (g.eff.)
WC 2003 August 2003
• Old definition:– Ratio of the measured imaged width to the measured
dose profile width (add imaged widths for multi-slice)
• New definition:– Integral of dose profile over range of nominal slice width
divided by integral of whole dose profile
Definitions – paraphrased
measured imaged width
dose profile width
integral ofdose profile overrange of nominal slice
nominal slice width
z-axis integral of whole dose profile
add -> total imaged width total nominal acquisition width(nominal total collimation)
WC 2003 August 2003
Purpose of this study
• Identify problems associated with old g.eff.• Compare two calculation approaches for new g.eff.• Compare values between old and new
WC 2003 August 2003
Methods of measurement – dose profiles
• Kodak X-Omat V Radiotherapy film• Scanned at iso-centre• Read out with scanning micro-densitometer• Optical density profile converted to dose profile
dose profile
0
1
2
3
4
5
6
7
8
0 5 10 15 20 25 30 35 40
distance along z-axis
dose
fwhm
WC 2003 August 2003
Methods of measurement – imaged slice width
• Angled metal plates scanned in axial mode• Width projected into image → Z-sensitivity profile
WC 2003 August 2003
Problems with old g.eff.
• Different measurement conditions (air vs phantom)• For multi slice scanners
– Many image width combinations for one beam width• Many measurements needed
– e.g. Philips Mx8000 four slice scanner
20
16
10
4
1
Beam width (mm)
2 x 104 x 520
2 x 816
1 x 102 x 54 x 2.510
1 x 42 x 24 x 14
1 x 12 x 0.51
Image width combinationsNo. x slice (mm)
Beam width (mm)
WC 2003 August 2003
Problems with old g.eff. (continued)
• Different focal spots– Potentially doubles the number of values
• Sixteen slice scanners – narrow slices– Need accurate measurement for imaged widths ~ < 1 mm– Not many existing test objects will do that
ImPACT’s ‘thin slice’ tool50 µm Titanium plates8° angle to scan plane
WC 2003 August 2003
Potential problems for calculation of new g.eff.
• Dose profile– ..‘integrated over the range subtended by detector
elements used’ ( � nominal total acquisition width)– Position of dose profile relative to detectors is not defined– Shape asymmetric � assume a reference position
dose profile
0
1
2
3
4
5
6
7
8
0 5 10 15 20 25 30 35 40
distance along z-axis
dose
WC 2003 August 2003
Two approaches for new g.eff. calculation
• Could either– Centre integration range on centre of dose profile
• Define as halfway between fwhm
– Position of range which gives maximum geometric efficiency
dose profile
0
1
2
3
4
5
6
7
8
0 5 10 15 20 25 30 35 40
distance along z-axis
dose
dose profile
0
1
2
3
4
5
6
7
8
0 5 10 15 20 25 30 35 40
distance along z-axis
dose
A. B.
WC 2003 August 2003
Two approaches for new g.eff. calculation
dose profile
0
1
2
3
4
5
6
7
8
0 5 10 15 20 25 30 35 40
distance along z-axis
dose
dose profile
0
1
2
3
4
5
6
7
8
0 5 10 15 20 25 30 35 40
distance along z-axis
dose
• Comparison of two techniques– Maximum difference = 0.04 % (ss), 0.6 % (ms)– Mean difference = 0.1% (ms)
2. Maximum g.eff.1: Centred on dose profile
A. B.
WC 2003 August 2003
Results
• Analysed data from 13 scanners (all manufacturers)– 4 single slice– 5 four slice– 1 eight slice– 3 sixteen slice
• Total of 123 different slice width combinations– Old g.eff.
• Compared new and old g.eff.
WC 2003 August 2003
Single slice - old g.eff. (imaged width / dose width)
– Most scanners have values of approximately 100%– Values higher than 100% probably due to scatter– Low for narrow slices using post patient collimation
0
20
40
60
80
100
120
0 1 2 3 4 5 6 7 8 9 10Slice width (mm)
Geo
met
ric
Eff
icie
ncy
(%)
Scanner AScanner BScanner CScanner D
70%
WC 2003 August 2003
Single slice - new g.eff. (dose in nominal width/whole profile)
– g.eff. values lower than 100% – Radiation falling outside the nominal width– Non rectangular dose profiles as widths gets thinner
0
20
40
60
80
100
0 1 2 3 4 5 6 7 8 9 10Slice width (mm)
Geo
met
ric
Eff
icie
ncy
(%)
Scanner AScanner BScanner CScanner D
70%
WC 2003 August 2003
Slice width (mm) 1 slice 2 slice 3 slice 4 slice 6 slice 12 slice
0.6 71%
1 106%
0.75 77%
1.5 77% 88%
3 88% 76%
4.5 77% 87%
5 95%
6 87%
9 77% 88%
10 96%
Multi slice - old g.eff. (imaged width / dose width)
• Unused penumbra – values less than 100%• Many values for each scanner
– e.g. Siemens Sensation 16 (small focus only)Imaged width
(mm)
WC 2003 August 2003
Slice width (mm) 1 slice 2 slice 3 slice 4 slice 6 slice 12 slice
0.6 71%
1 106%
0.75 77%
1.5 77% 88%
3 88% 76%
4.5 77% 87%
5 95%
6 87%
9 77% 88%
10 96%
Multi slice - old g.eff. (continued)
• Same beam width– Different values
Imaged width (mm)
9 mm beam
18 mm beam
WC 2003 August 2003
Multi slice - new g.eff. (dose in nominal width/whole profile)
• Values for each beam width only• Unused penumbra – values less than 100%
dose within nominal width equals total dose
dose within nominal width less than total dose
Four slice scanners only
0
20
40
60
80
100
0 2 4 6 8 101214161820222426283032Collimation (mm)
Geo
met
ric
Eff
icie
ncy
(%)
Scanner AScanner BScanner CScanner D
70%
WC 2003 August 2003
0
20
40
60
80
100
120
0 20 40 60 80 100 120Geometric efficiency (old) (%)
Geo
met
ric
Eff
icie
ncy
(new
) (%
)
Multi slice - new g.eff. versus old g.eff.
• Mean ratio = 0.99, but standard deviation = 10%• Biggest differences at narrow collimations
– total measured imaged width matches dose width, but both greater than nominal
• Some valuesnow below70% 2 x 1 mm
2 x 0.63 mm
2 x 0.63 mm
70% old
70% new
WC 2003 August 2003
Conclusions
• Old g.eff. definition– Worked well enough for single slice systems– Drawback is many slice width combinations for multi slice
• New g.eff. does not use imaged width; only the dose profile and nominal width– Two calculation techniques are comparable– Results for single slice are all generally lower
• spread of dose outside of nominal width not always identified before
– Multi-slice new g.eff. values similar to old values• except for some thin slices where both imaged slice width and dose
width are > nominal� Some values now below 70% threshold
WC 2003 August 2003
Sue Edyvean, Nicholas Keat ImPACT* (Imaging Performance assessment of CT Scanners)LondonUK
*An MHRA Evaluation centre for the UK Department of Health
(Medicines and Healthcare productsRegulatory Agency)
www.impactscan.org
Comparison of Definitions of Geometric Efficiency in Computed Tomography Scanners