Radiance Substrates HRP Substrates for Chemiluminescent Westerns Long Protocol for Catalog Numbers AC2203 Radiance ECL Sample, sufficient for 70 cm 2 membrane AC2204 Radiance ECL, 500 ml, sufficient for 5000 cm 2 membrane AC2100 Radiance Q Sample, sufficient for 70 cm 2 membrane AC2101 Radiance Q, 150 ml, sufficient for 1500 cm 2 membrane AC2102 Radiance Plus Sample, sufficient for 70 cm 2 membrane AC2103 Radiance Plus, 150 mL, sufficient for 1500 cm 2 membrane DC0013-003

Upload: others

Post on 19-Apr-2020

17 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Radiance Substrates - Azure Biosystems · • Radiance ECL Luminol/enhancer solution 3.5 ml • Radiance Peroxide Chemiluminescent Detection Reagent 3.5 ml AC2204 Radiance ECL Chemiluminescent

Radiance SubstratesHRP Substrates for Chemiluminescent Westerns

Long Protocol for Catalog Numbers

AC2203 Radiance ECL Sample, sufficientfor70cm2 membrane

AC2204 Radiance ECL, 500 ml, sufficientfor5000cm2 membrane

AC2100 Radiance Q Sample, sufficientfor70cm2 membrane

AC2101 Radiance Q, 150 ml, sufficientfor1500cm2 membrane

AC2102 Radiance Plus Sample, sufficientfor70cm2 membrane

AC2103 Radiance Plus, 150 mL, sufficientfor1500cm2 membrane

DC0013-003

Page 2: Radiance Substrates - Azure Biosystems · • Radiance ECL Luminol/enhancer solution 3.5 ml • Radiance Peroxide Chemiluminescent Detection Reagent 3.5 ml AC2204 Radiance ECL Chemiluminescent

Important InformationThe following instructions are for use with the Radiance enhanced chemiluminescent substrate kits, catalog numbers AC2100, AC2101, AC2102, AC2103, AC2203, and AC2204. Please see the Kit Contents section for details.

Storage InformationThe Radiance reagents are stable at room temperature for at least one year. For more information, see the Shipping and Storage Conditions section on page 3.

Warnings and Precautions•TheRadiancereagentsareforresearchuseonly.•Alwaysweargloveswhenhandlingmembranesandreagents.•RefertoMSDSforadditionalsafetyinformation.•Theproductisguaranteedtobefreeofmanufacturerdefect,and

to function as described when the enclosed protocol is followed by properly trained personnel. Please see the Warranty section for more information.

1

Page 3: Radiance Substrates - Azure Biosystems · • Radiance ECL Luminol/enhancer solution 3.5 ml • Radiance Peroxide Chemiluminescent Detection Reagent 3.5 ml AC2204 Radiance ECL Chemiluminescent

2

Table of ContentsSection Page

1. Kit Contents 3

2. Shipping and Storage Conditions 3

3. AdditionalMaterialsRequired 4

4. Background 4

5. Western Blotting 5

6. OverviewoftheProtocolforChemiluminescentWesternBlots 7

7. QuickProtocol 8

8. Detailed Protocol 9

9. Troubleshooting and FAQ 12

10. References 13

11. Related Products 13

12. Warranty 14

13. User Notes 14

Page 4: Radiance Substrates - Azure Biosystems · • Radiance ECL Luminol/enhancer solution 3.5 ml • Radiance Peroxide Chemiluminescent Detection Reagent 3.5 ml AC2204 Radiance ECL Chemiluminescent

1. Kit Contents

AC2203 RadianceECLChemiluminescentHRPSubstrate,Samplesize,sufficientfor70cm2 membrane •RadianceECLLuminol/enhancersolution 3.5ml •RadiancePeroxideChemiluminescentDetectionReagent 3.5ml

AC2204RadianceECLChemiluminescentHRPSubstrate,sufficientfor5000 cm2 membrane •RadianceECLLuminol/enhancersolution 250ml •RadiancePeroxideChemiluminescentDetectionReagent 250ml

AC2100RadianceChemiluminescentHRPSubstrate,Samplesize,sufficientfor70cm2 of membrane surface •RadianceQLuminol/enhancersolution 3.5ml •RadiancePeroxideChemiluminescentDetectionReagent 3.5ml

AC2101RadianceChemiluminescentHRPSubstrate,sufficientfor1500cm2 of membrane surface •RadianceQLuminol/enhancersolution 75ml •RadiancePeroxideChemiluminescentDetectionReagent 75ml

AC2102RadiancePlusChemiluminescentHRPSubstrate,Samplesize,sufficientfor70cm2 membrane •RadiancePlusLuminol/enhancersolution 3.5ml •RadiancePeroxideChemiluminescentDetectionReagent 3.5ml

AC2103 RadiancePlusChemiluminescentHRPSubstrate,sufficientfor1500 cm2 membrane •RadiancePlusLuminol/enhancersolution 75ml •RadiancePeroxideChemiluminescentDetectionReagent 75ml

3

Page 5: Radiance Substrates - Azure Biosystems · • Radiance ECL Luminol/enhancer solution 3.5 ml • Radiance Peroxide Chemiluminescent Detection Reagent 3.5 ml AC2204 Radiance ECL Chemiluminescent

2. Shipping and Storage ConditionsProductmaybeshippedatanyambienttemperature.Noextratemperaturecontrolorinsulationisrequired.Allshippingmethods(groundandexpress)areacceptable.TheRadiancereagentsarestableat room temperature for at least one year. Accidental freezing does not significantlyaffecttheperformance,butmultiplefreezing-thawingcyclesare not recommended.

3. Additional Materials Required• ElectrophoresisapparatusandbuffersforSDS-PAGE

• Tankandbuffersforelectrophoretictransferofproteinsfromgeltomembrane

• NitrocelluloseorPVDFmembrane,cuttosizeofgel.AllmembraneproductsavailablefromAzure(seetheRelatedProductssection)arecompatible with Radiance.

• Washingbuffer(PBS-TorTBS-T).Forbestresults,useAzureBlot WashingBuffer(seetheRelatedProductssection)

•Blockingbuffer

• Primaryantibodycompatiblewithyourapplication

• Secondaryantibody,conjugatedtoHorseradishperoxidase(HRP)correspondingtoyourprimaryantibody(seetheRelatedProductssection)

• CCD-baseddetectionsystem,orfilm(seetheRelatedProductssection)

4. BackgroundRadiancechemiluminescenthorseradishperoxidase(HRP)substratesarespeciallyformulatedforCCDimagingandtoprovideexcellentperformance with Azure’s cSeries imaging systems. Radiance substrates produce strong, long-lasting signal that is linear with respect to protein load.

Radiance ECL is an enhanced chemiluminescent substrate specially developed for routine Western blotting. The strong signal and low background produces cleaner blots, allowing for more meaningful semi-quantitativecomparisons.RadianceECLalsoproducessignalthatislonglasting,sorepeatexposurescanbeperformedwithoutfearoflosingdata.

4

Page 6: Radiance Substrates - Azure Biosystems · • Radiance ECL Luminol/enhancer solution 3.5 ml • Radiance Peroxide Chemiluminescent Detection Reagent 3.5 ml AC2204 Radiance ECL Chemiluminescent

5

5. Western BlottingWestern blotting is a protein analysis tool that has become commonplace in the molecular biology and protein chemistry laboratory. The principle of chemiluminescent Western blotting is shown in Figure 1. The general protocol, including the role of Radiance Substrates, can be seen on page7.Proteinsareseparatedbysizeviaelectrophoresis,andthentransferred electrophoretically from the gel to a membrane support, usuallynitrocelluloseorPVDF.Thismembranecontainingthetransferredproteins is commonly referred to as a blot. The location of a protein of interest is detected on the blot by applying the primary antibody, which binds to the protein. The primary antibody bound to the blot is then visualized using a secondary antibody that binds to the primary antibody. The secondary antibody is labeled in some way to make it detectable, such as with a radioactive isotope or an enzyme that can be detected by its activity.

RadianceECLisdesignedtobecompatiblewithbothX-rayfilmdetectionandCCDimagingthoughthelimiteddynamicrangeoffilmwillmakeresultingdatalessquantitative.

Radiance Q produces strong signal that is linear over a broad range of proteinconcentrations,allowinguserstoaccuratelyquantifyproteinbands. Additionally, no substrate depletion at high protein loads, allows the user to take full advantage of the linear range of the CCD detection method,forthemostaccuratequantitationacrossahighdynamicrange.RadianceQisalsocompatiblewithX-rayfilmdetection,thoughthelimiteddynamicrangeoffilmwillmakeresultingdatalessquantitative.

Radiance Plus is most suitable for detection of low-abundance proteins, when amounts of available primary antibodies are very limited and high dilution factors are desired, or when primary antibodies have relatively lowbindingconstants.RadiancePlusisalsocompatiblewithX-rayfilmdetection,thoughthelimiteddynamicrangeoffilmwillmakeresultingdatalessquantitative.RadiancePlusproducesastrong,long-livedsignal, which, combined with very low background levels, allows for long exposuretimesenablingthedetectionoflow-abundanceproteins.

Page 7: Radiance Substrates - Azure Biosystems · • Radiance ECL Luminol/enhancer solution 3.5 ml • Radiance Peroxide Chemiluminescent Detection Reagent 3.5 ml AC2204 Radiance ECL Chemiluminescent

6

Figure 1. The principle of chemiluminescent Western blotting.

Secondary AntibodyHRP Conjugate

Primary Antibody

Oxidized Products

Light

Proteins transferred to the membrane

BlockerMembrane

Substrate

HRP

Figure 2. Chemiluminescence of luminol.

NH2

NH

NH

Luminol

HRP + H2O2

NH2

+ N2 + light

Since 1988, enhanced chemiluminescence or ECL(1) has become oneof the most common detection methods in Western blotting(2). In thismethod,thesecondaryantibodyisconjugatedtotheenzymeHorseradishperoxidase(1,2). Once bound to the membrane, the secondary antibody is detected by incubating the blot with a solution containing an HRPsubstrate that generates a light-emitting product after reaction with HRP(Figures1,2).ThechemiluminescentsignalcanbedetectedbyexposingtheblottoX-rayfilm,orbyimagingwithaCCDcamera.

Radiance Substrates are enhanced chemiluminescent substrates developed for CCD imaging. They produce low background and bright, long-lasting signal.

Page 8: Radiance Substrates - Azure Biosystems · • Radiance ECL Luminol/enhancer solution 3.5 ml • Radiance Peroxide Chemiluminescent Detection Reagent 3.5 ml AC2204 Radiance ECL Chemiluminescent

7

6. Overview of the Protocol for Chemiluminescent Western Blots

Transferproteins from gel to membrane

Blocktomasknonspecificproteinbindingsitesonmembrane

Primary antibodybinds to protein of interest

Washtoremoveexcessantibody

Electrophoresis to separate proteins in sample

Secondary antibodybinds to primary antibody

Washtoremoveexcessantibody

Substrate (Radiance)substrate reacts with HRP bound to secondary antibody

to create luminescent signal

ImagedetectluminescentsignalwithCCDcameraorfilm

Page 9: Radiance Substrates - Azure Biosystems · • Radiance ECL Luminol/enhancer solution 3.5 ml • Radiance Peroxide Chemiluminescent Detection Reagent 3.5 ml AC2204 Radiance ECL Chemiluminescent

8

7. Quick ProtocolFor additional information, see the detailed protocol which follows.

Step User Notes

1. Prepare your protein blot

2. Block membrane for 1 hour at room temperature(RT)

3. Incubate blot with primary antibody for one hour at RT with gentle agitation

4. Wash blot:•1xquickly•1x15min,with1ml/cm2 membrane•3x5min,withatleast0.5ml/cm2

membrane each time

5. Incubate blot with secondary antibody for one hour at RT with gentle agitation

6. Wash blot:•3x5min,withatleast0.5ml/cm2

membrane each time

7.MixRadiancecomponents1:1andplace0.1ml/cm2 on blot for 2 minutes

8.Drainexcessreagent

9. Cover damp blot with plastic wrap and imagebyCCDcameraorexposuretoX-rayfilm

Page 10: Radiance Substrates - Azure Biosystems · • Radiance ECL Luminol/enhancer solution 3.5 ml • Radiance Peroxide Chemiluminescent Detection Reagent 3.5 ml AC2204 Radiance ECL Chemiluminescent

9

8. Detailed Protocol

Step Notes

1. Prepare a protein blot

1.1. Separate the protein sample(s)viaelectrophoresis

1.2. Transfer proteins to membrane

•Prewetmembraneintransferbuffer,andassemble transfer sandwich according to tank manufacturer’s instructions.

•Dot-blotsorslotblotscan also be detected with Radiance.

•Anyelectrophoresissystemandbuffer,suchas Laemmli, is compatible with Radiance.

•Awetortanktransfermethodispreferred,though semi-dry methods should also be compatible.WehavefoundthatthebuffersystemdevelopedbyBoltetall(3)workswell.

•BothnitrocelluloseandPVDFmembranesarecompatible with Radiance Substrates.

•IfusingPVDF,firstwetmembranewitha1minincubationin100%MeOHfollowedbywaterfor~5minandthentransferbufferfor5-10min.

•Forslotblotapplications,nitrocelluloseismuchmoreconvenientthanPVDFbecauseitismoredifficulttoavoidbubbleswithPVDF.

2. Block membrane

•Incubatetheblotinablockingbufferwithgentleagitation for 1 hour at room temperature(RT).Use0.2to0.5mlofblockingbufferper cm2 of blot to provide adequateblocking.

•Blockingmasksnon-specificproteinbindingsites on the membrane, reducing background andincreasingthespecificityofbindingoftheprimary antibody to the protein of interest.

•Theoptimalblockingbufferwilldependinpart on the nature of the antigen of interest, andonthequalityoftheprimaryantibody.Common blocking agents including non-fat dry milk have been found to be compatible with Radiance.

•10to20mlisusuallysufficientforatypical7x9cmmini-blot.

3. Incubate blot with primary antibody

•Diluteprimaryantibodyinblockingbuffer.

•Incubateblotwithprimaryantibody solution for 1 hour at RT with gentle agitation.

•Optimalprimaryantibodydilutionsmustbedetermined empirically.

•ForCCDimaging,werecommendprimaryantibody dilutions from 1:1000 to 1:10,000. A good initial dilution is 1:5000.

Page 11: Radiance Substrates - Azure Biosystems · • Radiance ECL Luminol/enhancer solution 3.5 ml • Radiance Peroxide Chemiluminescent Detection Reagent 3.5 ml AC2204 Radiance ECL Chemiluminescent

10

8. Detailed Protocol, continued

Step Notes

3. Incubate blot with primary antibody, continued

•Iftheblotwillbeimagedonfilm,use2–5xless primary antibody than for CCD imaging. Forexample,if1:1000dilutionoftheprimaryantibody was optimal for CCD detection, 1:5000issuitableforfilmdetection.

•Antibodycanbeaddedtoadishandplacedonashaker,orasmallervolume(5-10ml)canbe used by sealing the blot into a bag and placing it on a rotary platform.

4. Wash blot to remove excess primary antibody

•1xquickly•1x15min,with0.7ml/cm

membrane•3x5min,withatleast0.3ml/cm2 membrane each time.

•Forbestresults,useAzureBlotWashingBuffer(AC2113)whichisoptimizedforchemiluminescentaswellasfluorescentblots.PBS-T or TBS-T are also compatible with Radiance.

•Werecommendwashingorrockingblotsina clean dish on a shaker to provide gentle agitation.

•Forexample,astandard7x9membranerequires:~50mlofwashingsolutionforthe15min wash; and ~20 ml of washing solution for 5 min washes.

5. Incubate blot with secondary antibody

•Dilutesecondaryantibodyinblockingbuffer.

•Incubateblotwithsecondary antibody solution for 1 hour at RT with gentle agitation.

•Optimalsecondaryantibodydilutionsmustbedetermined empirically.

•Werecommendsecondaryantibodydilutionsof 1:5,000 to 1:20,000. A good initial dilution is 1:10,000.

•Iftheblotwillbeimagedonfilm,use2–5xlesssecondary antibody than for CCD imaging. 1:50,000 dilution is a good starting point for filmdetection.

•Seealsonotesforstep3.

6. Wash blot to remove excess secondary antibody

•3x5min,withatleast0.3ml/cm2 membrane each time.

•Seenotesforstep4.

Page 12: Radiance Substrates - Azure Biosystems · • Radiance ECL Luminol/enhancer solution 3.5 ml • Radiance Peroxide Chemiluminescent Detection Reagent 3.5 ml AC2204 Radiance ECL Chemiluminescent

11

8. Detailed Protocol, continued

Step Notes

7. Incubate blot with Radiance Substrate

•Mixcomponents1and2ina1:1ratioinsufficientamounts to obtain at least 0.1ml/cm2 of the blot and add to the blot.

•Itisbettertopreparetheworkingsolutionjustbeforeuse.However,mixedreagent is stable for several hours at RT.

•Allowsubstratetoreactwith blot for 2 minutes.

•Becarefulnottotouchorputpressureontheblotasthiscanresultinnon-specificbackground.

•Useonlyplasticforceps,notmetal;metalforceps damage the blocked surface, creating new adsorption sites. Also, traces of metal may act as a catalyst for non-enzymatic substrateoxidation,resultinginveryhighbackground.

•Theminimalamountofworkingreagentis0.1ml/cm2.Forexample,fora7x9cmblot,thisminimalvolumeis7x9x0.1=6.3ml.

•Ifusingtheminimalamountofworkingreagent, incubation may be done without agitation.Makesurethemembranesurfaceislevelsoadequatereagentisheldbysurfacetension.

•Incubationmayalsobedonewithgentleagitationinatrayjustslightlylargerthanthemembrane. Increase the reagent volume as necessary to ensure the membrane is adequatelycoveredwithreagent.

8. Drain excess reagent

•Removeexcesssubstrateviacapillaryactionby touching a KimWipe® or other absorbent material to the edge of the blot.

9. Image blot

•Whileblotisdamp,coverwith transparent plastic wrap and either place blot inCCDimager,orexposeblottofilm.

•Werecommendtryingthreeexposures;30 sec, 2 min, and 5 min.

•Theblotcanbeimagedandre-imagedforseveralhours;70%oftheinitialsignalwillremain after 60 minutes, and substantial signal will remain after 8-10 hours.

Page 13: Radiance Substrates - Azure Biosystems · • Radiance ECL Luminol/enhancer solution 3.5 ml • Radiance Peroxide Chemiluminescent Detection Reagent 3.5 ml AC2204 Radiance ECL Chemiluminescent

12

Problem Possible SoutionsHigh background •Reduceprimaryantibodyconcentrationbyincreasing

the dilution factor.•Tryadifferentblockingbuffer.•Tryashorterexposuretime.•Increasewashingtime.

No or low signal •Checkthatcorrectprimaryantibodyused.•Checkthatsecondaryantibodyrecognizesprimary(forexampleiftheprimaryisarabbitantibody,thatthesecondaryisgoat-anti-rabbit).

White spots withinbands

•Improvetransfer,makingsuretoremoveanybubblesbetween the gel and the membrane.

Speckled background •Filtersecondaryantibody.•Filterblockingandwashingbuffers.•Ensurethatthelaboratoryenvironmentisclean,to

minimize dust, debris or any other particles that might come in contact with the blot. Cover the dish during incubation or washing steps.

•Usenon-powderedgloves,orswitchtoadifferentkind of gloves. We recommend powder-free nitrile gloves or polyethylene gloves.

9. Troubleshooting & FAQ, continued

Westernblottingcanrequiresubstantialoptimizationduetothemultiplesteps involved. The correct amount of protein to load on the gel and thebest dilutions of primary and secondary antibodies must be determinedempirically.Somecommonquestionsareaddressedbelow:

Page 14: Radiance Substrates - Azure Biosystems · • Radiance ECL Luminol/enhancer solution 3.5 ml • Radiance Peroxide Chemiluminescent Detection Reagent 3.5 ml AC2204 Radiance ECL Chemiluminescent

10. References1.ThorpeGH,KrickaLJ,MoseleySB,WhiteheadTP,Phenolsasenhancersofthechemiluminescenthorseradishperoxidase-luminolhydrogenperoxidereaction:applicationinluminescence-monitored enzyme immunoassays. Clin Chem. 1985Aug;31(8):1335-41.

2.LeongMM,FoxGR.,Enhancementofluminol-basedimmunodotandWestern blotting assays by iodophenol. Anal Biochem.1988Jul;172(1):145-50.

3.BoltM.W.,MahoneyP.A,High-efficiencyblottingofproteinsofdiverse sizes following sodium dodecyl sulfate-polyacrylamide gel electophoresis.AnalBiochem.1997May1;247(2):185-192.

11. Related ProductsCatalog Number Product Size

AC2113 AzureBlotWashingBuffer 500ml

AC2114 Goat-anti-rabbitHRP-conjugated 500μl secondary antibody

AC2115 Goat-anti-mouseHRP-conjugated 500μl secondary antibody

AC2105 LowFluorescenceWesternMembrane(PVDF) 10 sheets 7x9cm

AC2106 NitrocelluloseTransferMembrane0.45μm 10 sheets 7x9cm

AC2107 NitrocelluloseTransferMembrane0.22μm 10 sheets 7x9cm

13

Page 15: Radiance Substrates - Azure Biosystems · • Radiance ECL Luminol/enhancer solution 3.5 ml • Radiance Peroxide Chemiluminescent Detection Reagent 3.5 ml AC2204 Radiance ECL Chemiluminescent

12. WarrantyThis product is warranted to be free of defects of material or workmanship,andtoperformasdescribedinthepublishedspecificationswhen stored according to the documentation included with the product, and used according to the accompanying instruction manual by appropriately trained personnel. If the product is found to have a defectuponfirstuseandwithin30daysofshipment,theproductmaybereplaced.Thiswarrantyextendsonlytotheoriginalpurchaserofthe product. There is no obligation to replace the product as a result of misuse, improper storage, or negligence of the buyer.

13. User Notes

14

Page 16: Radiance Substrates - Azure Biosystems · • Radiance ECL Luminol/enhancer solution 3.5 ml • Radiance Peroxide Chemiluminescent Detection Reagent 3.5 ml AC2204 Radiance ECL Chemiluminescent

Copyright © 2014-2018 Azure Biosystems Inc. All rights reserved. The Azure logo is a trademarks of the Company. All other trademarks, service marks and tradenames appearing in this brochure are the property of their respective owners.

www.azurebiosystems.com • [email protected]