Overview of Pressure Measurement and Calibration

Page 2: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

2

Course Objectives

Describe elements used in pressure measurement

Perform procedures for calibration of pressure

Describe instruments to be calibrated and select test equipment based on datasheets

Page 3: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

3

Page 4: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

4

Page 5: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

5

Page 6: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

6

Page 7: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

7

Page 8: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

8

Page 9: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

9

Page 10: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

10

Page 11: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

11

Page 12: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

12

Page 13: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

13

Page 14: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

14

Page 15: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

15

Page 16: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

16

Page 17: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

17

Page 18: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

18

Page 19: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

19

Page 20: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

20

Page 21: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

21

Page 22: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

22

Page 23: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

23

Page 24: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

24

Page 25: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

25

Page 26: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

26

The 5 Pressure Types

Gauge Vacuum Compound Differential Absolute

Page 27: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

27

Gauge Pressure

Because gauge pressure gauges allow the surrounding ambient pressure to effect both sides of the sensing element, the effects of barometric pressure are essentially negated. Therefore, a "gauge pressure" gauge with an open inlet port will start with the pointer at zero, which means that the gauge is indicating "no pressure in excess of barometric." Gauge pressure dial gauges usually position the zero point at approximately 7:00 on the dial and rotate in the clockwise direction.

Page 28: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

28

Vacuum Vacuum gauges measure negative pressures, i.e.

the removal of atmospheric pressure. Using "gauge pressure zero" as the starting point, the gauge will indicate the "vacuum level" in positive numbers as more pressure is removed. Vacuum gauges usually position zero at 5:00 on the dial and rotate counter clockwise. Most vacuum gauges are rated to a full scale of 30 In Hg or 15 psi. Remember that since "vacuum" is simply the removal of atmospheric pressure, the highest level of vacuum that can be achieved on a given day is equal to the barometric pressure.

Page 29: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

29

Vacuum (Cont.)

(You can only remove what's there to remove in the first place!). Users often comment that they cannot get the gauge to read "all the way down to 30 In Hg, therefore the gauge is not working properly ." In most cases, the reason that they cannot achieve a 30 In Hg vacuum reading is that the existing barometric pressure is less than 30 In Hg, thus a 30 In Hg vacuum reading is unachievable. Vacuum gauges will not tell you how far you are from a complete vacuum; they will only tell you how far you are from “gauge pressure zero”.

Page 30: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

30

Compound Pressure

Starting at "gauge pressure zero," a compound gauge simply combines the vacuum indication of the straight vacuum gauge with the gauge pressure indication of a gauge pressure gauge. The position of the zero is dependent upon the full scale rating of the pressure side. The pointer will travel in the counterclockwise direction for vacuum indications, and clockwise for pressure indications.

Page 31: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

31

Differential Pressure

Starting at "gauge pressure zero," a differential pressure gauge simply measures the difference between 2 input pressures. A differential dial gauge looks like a gauge pressure gauge, except it has a second inlet port, with one port marked “hi” and the other marked “lo”. The line that connects to the higher pressure side of the application is always connected to the “hi” port, while the lower pressure line is connected to the “lo” port.

Page 32: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

32

Absolute Pressure Unlike "gauge pressure," an absolute

gauge does not allow ambient pressure to affect both sides of the sensing element. To achieve a true absolute reading, an absolute pressure gauge must have the atmosphere removed from around one side of the sensing element (referred to as an "evacuated reference"). The result is that an absolute gauge with an open inlet port will indicate the barometric pressure (which is usually between 14 and 15 psi).

Page 33: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

33

Absolute Pressure (cont)

This will allow the user to either apply pressure causing the gauge to read above the barometric pressure value, or remove pressure (pull vacuum) causing the gauge to read below the barometric reading. "Absolute zero" should be achievable under full vacuum, since the starting point is the actual barometric reading, and a full vacuum will remove all of this existing pressure. Thus, unlike standard “vacuum” gauges, the absolute gauge will indicate how far the pressure is from a complete vacuum. 

Page 34: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

34

Absolute Pressure (Cont)

On absolute dial gauges, the zero point will be positioned at 7:00 on the dial and the pointer will rotate clockwise when pressure is applied, and counterclockwise when pressure is

Page 35: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

35

Bourdon Tube Gage Cal ACCURACY: PROCEDURES/DEFINITIONS Accuracy inspection – Readings at

approximately five points equally spaced over the dial should be taken, both upscale and downscale, before and after lightly rapping the gauge to remove friction. A pressure standard with accuracy at least 4 times greater than the accuracy of the gauge being tested is recommended.

Page 36: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

36

Equipment – A finely regulated pressure supply will be required. It is critical that the piping system associated with the test setup be leak tight. The gauge under test should be positioned as it will be in service to eliminate positional errors due to gravity.

Method – ASME B40.1 recommends that known pressure (based on the reading from the pressure standard used) be applied to the gauge under test. Readings including any error from the nominal input pressure, are then taken from the gauge under test. The practice of aligning the pointer of the gauge under test with a dial graduation and then reading the error from the master gauge (“reverse reading”) can result in inconsistent and misleading data and should NOT be used.

Page 37: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

37

Calibration chart – After recording all of the readings it is necessary to calculate the errors associated with each test point using the following formula: ERROR in percent = 100 times (TRUE VALUE minus READING) ÷ RANGE. Plotting the individual errors (Figure 1) makes it possible to visualize the total gauge characteristic. The plot

should contain all four curves: upscale – before rap; upscale – after rap; downscale – before rap; downscale – after rap. Rap means lightly tapping the gauge before reading to remove friction as described in ASME B40.1.

Page 38: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

38

Page 39: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

39

Referring to Figure 1, several classes of error may be seen: Zero – An error which is approximately equal over the entire scale. This error can be manifested when either the gauge is dropped or overpressured and the bourdon tube takes a permanent set. This error may often be corrected by simply repositioning the pointer. Except for test gauges, it is recommended that the pointer be set at midscale pressure to “split” the errors

Page 40: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

40

Page 41: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

41

Span – A span error exists when the error at full scale pressure is different from the error at zero pressure. This error is often proportional to the applied pressure. Most Ashcroft gauges are equipped with an internal, adjusting mechanism with which the user can correct any span errors which have developed in service.

Linearity – A gauge that has been properly spanned can still be out of specification at intermediate points if the response of the gauge as seen in Figure 1 on page 6 is not linear. The Ashcroft Duragauge® is equipped with a rotary movement feature which permits the user to minimize this class of error.

Page 42: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

42

Hysteresis – Some bourdon tubes have a material property known as hysteresis. This material characteristic results in differences between the upscale and downscale curves. This class of error can not be eliminated by adjusting the gauge movement or dial position. in readings before and after lightly tapping.

Friction – This error is defined as the difference the gauge case at a check point. Possible causes of friction are burrs or foreign material in the movement gearing, “bound” linkages between the movement and the bourdon tube, or an improperly tensioned hairspring. If correcting these potential causes of friction does not eliminate excessive friction error, the movement should be replaced.

  

Page 43: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

43

6.1 Calibration – Rotary Movement Gauges Inspect gauge for accuracy. Many times gauges are

simply “off zero” and a simple pointer adjustment using the micrometer pointer is adequate. If inspection shows the gauge warrants recalibration to correct span and/or linearity errors, proceed as follows: a. Remove ring, window and, if solid front case, the rear closure assembly.

b. Pressurize the gauge once to full scale and back to zero.

c. Refer to Figure 2 for a view of a typical Ashcroft system assembly with component parts identified.

d. For solid front gauges, adjust the micrometer pointer so that it rests at the true zero position. For open front gauges the pointer and dial must also be disassembled and the pointer should then be lightly pressed onto the pinion at the 9:00 o’clock position.

Page 44: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

44

e. Apply full scale pressure and note the magnitude of the span error. With open front gauges, ideal span (270 degrees) will exist when at full scale pressure the pointer rests exactly at the 6:00 o’clock position.

f. If the span has shifted significantly (span error greater than 10%), the gauge should be replaced because there may be some partial corrosion inside the bourdon tube which could lead to ultimate failure. If the span error exceeds 0.25%, loosen the lower link screw and move the lower end of the link toward the movement to increase span or away to decrease span. An adjustment of 0.004 inch will change the span by approximately 1%. This is a repetitive procedure which often requires more than one adjustment of the link position and the subsequent rechecking of the errors at zero and full scale pressure.

Page 45: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

45

g. Apply midscale pressure and note error in reading. Even

though the gauge is accurate at zero and full scale, it may be inaccurate at the midpoint. This is called linearity error and is minimized by rotating the movement. If the error is positive, the movement should be rotated counter clockwise. Rotating the movement one degree will change this error by approximately 0.25%. Rotating the movement often affects span and it should be subsequently rechecked and readjusted if necessary according to step 6.1e and 6.1f.

h. While recalibrating the gauge, the friction error – difference

in readings taken with and without rap – should be noted. This error should not exceed the basic accuracy of the gauge. If the friction error is excessive, the movement should be replaced. One possible cause of excessive friction is improper adjustment of the hairspring. The hairspring torque,or tension, must be adequate without being excessive. The hairspring should also be level, unwind evenly (no turns rubbing) and it should never tangle.

Page 46: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

46

NOTES: 1 For operation of test gauge external zero

reset, refer to Figure 32 For test gauge calibration procedure,refer

to Figure 46.3 Positive Pressure Ranges – a. Remove dial and lightly press pointer onto

pinion at 9:00 o’clock position. b. Apply full scale pressure and rotate span

block as shown in Figure 5a until pointer rests at 6:00 o’clock position.

Page 47: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

47

c. Fully exhaust pressure and check that pointer still is at 9:00 o’clock position. If not repeat steps 1 and 2 until span is correct

d. Remove pointer and reassemble dial and dial screws (finger tight).

e. Lightly press pointer onto pinion. f. Check accuracy at full scale. If error

exceeds 1% return to step 1, otherwise proceed.

g. Check accuracy at midscale. If error exceeds 1% slide dial left or right to compensate.

h. Continue at * below

Page 48: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

48

Vacuum range – a. Using a pencil, refer to dial and mark the 0 and 25

inch of Hg positions on the case flange. b. Remove the dial. c. Apply 25 inches of Hg vacuum. d. Lightly press pointer onto pinion carefully aligning it

with the 25 inch of Hg vacuum mark on case flange. e. Release vacuum fully. f. Note agreement of pointer to zero mark on case

flange. g. If span is high or low, turn span block as shown in

Figure 5b h. Repeat steps 4 through 8 until span is correct. i. Remove the pointer.

Page 49: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

49

j. With 25 inches of Hg vacuum applied, reassemble dial, dial screws (finger tight) and pointer.

k. Apply 15 inches of Hg vacuum and note accuracy of indication. If required, slide dial left or right to reduce error to 1% maximum.

l. Continue at * below. *Now complete calibration of the gauge as follows:

a. Firmly tighten dial screws. b. Firmly tap pointer onto pinion, using brass back-up

tool from Ashcroft kit 1205T if gauge has rear blow-out plug. If gauge has top fill hole no back-up is required.

c. Recheck accuracy at zero, midscale and fullscale points (Figures 5a & 5b ).

d. Reassemble window, gasket and ring.  

Page 50: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

50

Page 51: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

51

Page 52: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

52

Page 53: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

53

Page 54: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

54

Page 55: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

55

Page 56: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

56

Page 57: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

57

Page 58: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

58

Page 59: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

59

Millivolt Output Pressure TransducersTransducers with millivolt output are normally the most economical pressure transducers. The output of the millivolt transducer is nominally around 30mV. The actual output is directly proportional to the pressure transducer input power or excitation. If the excitation fluctuates, the output will change also. Because of this dependence on the excitation level, regulated power supplies are suggested for use with millivolt transducers. Because the output signal is so low, the transducer should not be located in an electrically noisy environment. The distances between

the transducer and the readout instrument should also be kept relatively short.

Pressure Transducer/Transmitter

Pressure transducers are generally available with three types of electrical output; millivolt, volt and 4-20mA. Below is a summary of the outputs and when they are best used.

Page 60: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

60

Voltage Output Pressure TransducersVoltage output transducers include integral signal conditioning which provide a much higher output than a millivolt transducer. The output is normally 0-5Vdc or 0-10Vdc. Although model specific, the output of the transducer is not normally a direct function of excitation. This means unregulated power supplies are often sufficient as long as they fall within a specified power range. Because they have a higher level output these transducers are not as susceptible to electrical noise as millivolt transducers and can therefore be used in much more industrial environments.

Page 61: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

61

4-20 mA Output Pressure TransducersThese types of transducers are also known as pressure transmitters. Since a 4-20mA signal is least affected by electrical noise and resistance in the signal wires, these transducers are best used when the signal must be transmitted long distances. It is not uncommon to use these transducers in applications where the lead wire must be 1000 feet or more.

Page 62: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

62

Transducer/Transmitter Calibration Calibrations of Transducers/Transmitters

are performed with the use of a pressure standard and the appropriate output monitoring device.

Transducers/Transmitters have a maximum range (varying by construction) but are usually scaled lower to set the psi/output signal ratio to obtain a more desirable resolution.

Page 63: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

63

In most cases the desired output will be 4-20mA. Calibration adjustments will be made at zero and span.

Knowing the fullscale pressure of the device under test will come in handy when adjusting for 20mA. Zero of course is just zero psi although the output in this case will have a value of 4mA.

Page 64: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

64

Alternating and adjusting between these settings will eventually tune the unit. As there is some interaction between the adjustments.

Some of the newer Rosemount transmitters will have pushbuttons for zero and span making it easier to tune. In most instances the pots will be labeled.

Page 65: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

65

Prior to performing any calibration the selecting the correct standards is necessary.

In selecting the standards it is important to consider

1) the range of pressure over which the test will be performed.

2) the range of temperature over which the test will be performed

3) The Duration of test.

4) Degree of accuracy of the Unit under test.

Calibrators

Page 66: Pressure Overview Of Measurement And Calibration

Biometrix Corporation (800)-890-8909

66

Documentation The documentation used to record these test

should conform to the approved methodology of the customer being serviced.

Having the previous history generated on the unit being tested will help ensure conformity.

Generating documents of a similar instrument model from the customer’s inventory will also prevent inconsistencies