nomenclature - brunel university london · web viewintense efforts have been made to introduce this...
TRANSCRIPT
Journal of Medical Devices - ASME
Numerical and Experimental Simulations of the Wireless Energy Transmission and
Harvesting by a Camera PillElizabeth ShumbayawondaCentre of Biomedical EngineeringUniversity of Surrey Guildford, Surrey GU2 7XH, UKe-mail: [email protected]
Ali A. SalifuDepartment of Mechanical Engineering SciencesUniversity of Surrey Guildford, Surrey GU2 7XH, UKe-mail: [email protected]
Constantina Lekakou1
Department of Mechanical Engineering SciencesUniversity of Surrey Guildford, Surrey GU2 7XH, UKe-mail: [email protected]
John P. CosmasDepartment of Electronic and Computer Engineering Brunel University Uxbridge UB8 3PH, UKe-mail: [email protected]
ABSTRACT
This paper investigates the energy transmitted to and harvested by a camera pill travelling along the
gastrointestinal tract. It focuses on the transmitted electromagnetic (EM) energy in the frequency range of
0.18 to 2450 MHz and compares it to the mechanical energy due to the motion of the pill and the force
exerted from the intestine in its peristalsis onto the pill, and the electrochemical energy due to the change
of pH along the path of the pill. A comprehensive multilayer EM power transmission model is constructed
and implemented in a numerical code, including power attenuation through each layer and multi-
1
Journal of Medical Devices - ASME
reflections at material interfaces. Computer simulations of EM power transmission through a multilayer
abdomen to a pill travelling in the intestine are presented for the human abdominal cavity as well as
phantom organs and phantom environments, coupled with corresponding experimental studies using
these phantom components and environments. Two types of phantom abdomen are investigated: a
ballistic gel and a multilayer duck breast. Phantom small intestine involves gelatin gel layers with
embedded phantom chyme. Due to limitations related to the energy safety limit of skin exposure and
energy losses in the transmission through the abdomen and intestines, inductive range frequencies are
recommended which may yield energy harvesting of 10-50 mWh during 8 hours of pill journey,
complemented by about 10 mWh of mechanical energy and 10 mWh of electrochemical energy
harvesting, in addition to about 330 mWh typically stored in the coin batteries of a camera pill.
2
Journal of Medical Devices - ASME
INTRODUCTION
A camera pill is the most common type of an electrofunctional pill which
combines the idea of “Fantastic Voyage”[1] with the great potential of disease diagnosis
and treatment in difficult-to-reach parts of the body, such as the inner folds of the
intestine. Intense efforts have been made to introduce this technology for the
gastrointestinal (GI) tract [2-4], which entails a pill journey that may last up to 12 hours.
There are currently several commercially available camera pills, for example the
PillCam® capsule by Given Imaging, the EndoCapsule® by Olympus, the CapsoCam® by
CapsoVision, the MiroCam® by IntroMedic, the OMOM® by Jianshan and the Norika® by
RF System [5-6]. These commercial pills are of a general cylindrical shape of 9-11 mm
diameter, 23-32 mm length and 2-4 g mass, and most have one video camera operating
at a frame rate of 2-6 fps [5-6]. An in-depth study of the specifications of each of the
above commercial camera pills reveals that their high-power consuming parts, operating
intermittently, include 4-6 LEDs operating at about 15 mW for each LED, the CMOS that
may consume 2 mW operating at 2 fps, and the transceiver that may operate at 15 mW
in data transmission mode. Current commercial capsules have coin-type silver oxide
batteries [7] of about 330 mWh in total that may last for about 8 hours, corresponding
to a time averaged power of 41 mW. The big challenge is to develop on-board energy
harvesting technology to prolong the battery life so that the camera pill conducts a
comprehensive investigation through the whole length of the GI tract. Continuously
harvested energy would also offer the opportunity to reduce the battery size or even
eliminate the whole battery or extend the pill functions. Different types of energies may
3
Journal of Medical Devices - ASME
be suitable for harvesting, some already in the environment such as electromagnetic
(EM) energy, others produced by the pill such as kinetic energy, while others may be
generated by the pill interactions with the walls and the fluid of the GI track.
Electromagnetic (EM) energy harvesting involves wireless energy in a wide range
of frequencies. At first thought, EM energy harvesting seems very attractive due to the
potential large amounts of EM energy in the atmosphere at megahertz level (radio,
GSM) and gigahertz level (Bluetooth, WiFi); however, the energy transmitted through
the human body should be constrained by the energy density safety limit, Elim (in mWh
cm-2) reaching the external body surface area, given by:
E lim=S lim t lim /60 (1)
where Slim and tlim are the power density and average time limit, respectively, depending
on the transmitted frequency as presented in Table 1 [8]. Table 1 shows that the
highest safety limit is for frequencies below 1.34 MHz.
Electromagnetic induction is used typically for long-term continuous
communication processes usually at the frequencies of 13.56 MHz or 28 MHz. It would
require a coil in the pill and an external energizing coil, which should maintain alignment
for maximum energy transfer; this has been tested successfully for distances up to 1 m
[9]. Pan et al [10] tested an inductive transmission system at 181 kHz to transmit 150
mW in the air for which they estimated a specific absorption rate (SAR) of less than
0.329 W kg-1 [11] that was lower than the basic short-term limit of 10 W kg-1 set by the
4
Journal of Medical Devices - ASME
International Commission on Non-Ionizing Radiation Protection (ICNIRP) or the
occupational exposure limit of 0.4 W kg-1 based on the whole-body average. However,
the long term environmental exposure of the general public has a whole-body average
limit of 0.08 W kg-1.
On the other hand, RF communication is much more effective with the Medical
Implant Communication System (MICS) band specified at 403-405 MHz and a power
limit of 25 W in the air, and the Industrial Scientific and Medical (ISM) bands at 433.92
MHz, 915 MHz, 2.4 GHz and 5.8 GHz [12, 13]. However, such frequencies may suffer
high absorption loss in their transmission through the different layers of the body to the
camera pill. At the same time, such RF frequencies are also used for the data
communication from the camera pill to the external receiver, hence, a comprehensive
analysis of the transmission losses would offer double benefits.
In past analyses of EM energy and power transmission, power reflection at
middle-layer interfaces is generally neglected [14] or homogeneous simulating liquids
have been used in experimental studies, although it has been recognized that standing-
wave effects due to impedance mismatch at interfaces increase the SAR level [15, 16].
The major aim of this study is to conduct a theoretical and experimental analysis
of the electromagnetic energy transmission losses though the abdominal cavity to an
electrofunctional pill. The theoretical research focuses on the construction of a
comprehensive numerical model implemented into a computer code that includes a
multilayer material and the transmission and reflection terms at all material interfaces.
Computer simulations of the wireless EM energy transmission through a multilayer
5
Journal of Medical Devices - ASME
abdominal cavity to the pill will be conducted across a broad spectrum of wave
frequencies and the harvested EM energy will be estimated. Furthermore, experimental
studies of power transmission through phantom organs and combined phantom organ
environments will be carried out to investigate the accuracy of their representation of
the abdominal cavity environment in EM power transmission applications. The results
of the RF energy harvesting will be compared to other possible types of energy
harvesting by a camera pill.
THEORETICAL MODELLING OF THE EM ENERGY TRANSMISSION AND HARVESTING
The analysis of the EM energy harvesting comprises the construction of a
numerical model of the EM energy transmission to the pill from an external EM
radiation point source. This is followed by computer simulations of EM energy
transmission through multilayers to the camera pill located in the small intestine, at
different depths, using literature data about the dimensions of the multilayers in the
human abdomen and their material properties. These simulations are then repeated to
assess phantom abdomens and up to four layers of phantom intestines where
experiments were carried out to determine the EM energy transmission properties of
these phantom abdomens and phantom intestine.
The theoretical model considers electromagnetic energy harvesting from an
external point source to a camera pill travelling in the GI tract. Fig. 1 illustrates a
sequence of six biomaterial layers for the radiated EM energy to reach the pill located in
the intestine.
6
Journal of Medical Devices - ASME
The total power, po, emitted from the point EM power source may suffer a first
loss as is transmitted through the air, resulting in a power density, P1,ent, entering the
skin layer at a distance, Rair, from the EM point source, given by the equation:
P1 , ent=poe
−2α air Rair
4 πRair2
(2)
where air is the attenuation constant of air. The power density, P1,ent, entering the skin is
specified by the safety constraints given in Table 1, depending on the harvested or
transmitted EM energy frequency, and equation (1).
In the EM energy transmission across the multiple layers, a general numerical
scheme has been established of energy balance at the interfaces between any layer i
and its neighboring layers i-1 and i+1, as well as energy attenuation in its transmission
through each layer i, as illustrated in Fig. 2. The energy balance at each layer interface i-
1/i considers that the total power density arriving at the exit of layer i-1, Pi-1,ex, is split
into reflected and transmitted power densities, Pi-1,ex rf and Pi-1,ex tf, respectively:
Pi−1 , ex=Pi−1 , ex r f+P i−1, ex t f (3)
The power density at the start of layer i, Pi,start, is equal to the sum of power density
transmitted from the i-1 layer and the attenuated reflected energy from the exit of layer
i:
7
Journal of Medical Devices - ASME
Pi , start=Pi−1 , ex t f+Pi , exr f e−2 αi Δri ( Ri +Δr i)2
R i2 (4)
where ri is the thickness of layer i. That Pi,start energy then is further attenuated in its
transmission to the exit of layer i, according to the equation:
Pi , ex=Pi , start e−2 αi Δri
R i2
( Ri +Δr i)2
(5)
Equations (3)-(5) are then applied to all layers in the EM energy transmission path,
comprising a system of highly interdependent equations. For this reason, a C++
computer algorithm was written with an iterative numerical scheme following the
Gauss-Seidel method for systems of algebraic equations.
For each layer i the attenuation constant, reflection and transmission coefficients
depend on the layer material or medium and the EM wave frequency, f. The attenuation
constant, i, for layer i is given by:
α i=2πf
c [ μr ,i εr , i'
2 (√1+( εr , i' '
εr , i' )
2
−1)]1/2
(6)
where c is the speed of light, r,i is the material relative magnetic permeability, and ε r ,i'
and ε r ,i' '
are the real and imaginary components of the material relative complex
8
Journal of Medical Devices - ASME
permittivity, respectively. The following relations have been used for the reflection and
transmission of the EM energy, respectively, at the interface between layers i-1 and i:
r f ,i−1/i=|ρi−1/ i2 |
t f , i−1/ i=1−|ρi−1/ i2 | (7)
The reflection coefficient, i-1/i, at the interface between layers i-1 and i can be
calculated using the following relations [17]:
ρi−1/i=ηi−ηi−1
ηi+ηi−1
ηi=( μ i
εi )1/2
ε i=εi'− j εi
' '
ε i'=ε∞ , i+
ε 1, i
1+( ff 1 , i )
2+
ε2, i
1+( ff 2 , i )
2
εi' '=
σ i
2 πf (8)
where i and i are the intrinsic impedance and conductivity of the material of layer i.
Using the first three equations of system (8) and also the assumption that for all
biological tissues, i = 1, the following relation has been derived for the reflection
coefficient i-1/i:
9
Journal of Medical Devices - ASME
ρi−1/i = A+ jB- 2(C+jD)√E-jFG
whereA=εi−1
¿ +εi−1¿ −εi
¿−εi¿
B=−(εi−1' −εi
' )( εi−1' ' +εi
' ' )+(εi−1' ' −εi
' ' )( εi−1' +εi
' )C+ jD=( εi−1
' −εi' )+ j( εi−1
' ' −εi' ' )
E− jF=(εi−1' εi
' −εi−1' ' εi
' ' )− j ( εi−1' εi
' '+εi−1' ' εi
' )
G=( εi−1' −εi
' )2+( εi−1' ' −εi
' ' )2(9)
EXPERIMENTAL DESCRIPTION: MATERIALS AND PROCEDURES
Experimental studies of EM power transmission through phantom abdomen
components and environments were conducted using a hp 8714C Spectrum Analyzer
(300 kHz – 3 GHz) for the RF frequencies of 0.3, 13.3, 28.3, 433.92, 915, 1800 and 2450
MHz. The phantom bio-components comprised: gelatin gel phantom intestines filled
with phantom chyme; phantom muscle from ballistic gel; duck breast multilayer part.
The phantom small intestines were flat gelatin layers of 7 mm thickness with
embedded phantom chyme. Collagen is a major material of the extracellular matrix of
the human intestine walls, especially in the muscular layer [18]. Gelatin has been
employed as low-cost collagen substitute in tissue engineering [19] and, hence, been
also adopted in the current investigations. Porcine gelatin type A [supplied by Sigma
Aldrich, UK] was used in this study. A 4 wt% gelatin solution in water was cast in a tray,
left in the fridge overnight, and crosslinked in a 2 %w/v glutaraldehyde solution for 2 hr
following the procedure in [20] to yield a gel case of about 30-35 mm thickness, which
was washed with deionized water. The case was filled with a layer of phantom chyme
representing the chyme of human intestines consisting of the semi-fluid mass of
digested food. The phantom chyme was made using crushed nuts soaked in lemon juice
10
Journal of Medical Devices - ASME
for 12 hr (to simulate the digestion process in the stomach) and then washed using
deionized water to pH = 7 (considered as a representative average pH value in the small
intestine [21]). Finally, the phantom chyme was covered with a top gel layer of 30-35
mm thickness, yielding composite layers of phantom intestine medium consisting of
crosslinked gelatin gel with embedded phantom chyme.
Two types of phantom abdomen were fabricated and tested. The first phantom
abdomen was based on ballistic gel as proposed by [22] and [23], using a 10 wt% gelatin
solution to prepare the ballistic gel, with starting material porcine gelatin type A
[supplied by Sigma Aldrich, UK]. A gel thickness of 23 mm thickness (excluding the
intestine part) was found the most suitable to match the predicted power losses of
power transmission through the human abdomen. The second phantom abdomen
comprised a duck breast part (supplied from a food supermarket) trimmed to a
thickness of 20 mm and consisting of visible muscle, fat and skin layers; two different
duck breast parts were tested, coming from different suppliers and food supermarkets.
Each phantom component was tested for the transmission of an inputted power
at different frequencies. Then combined abdomen environments were tested including
the phantom abdomen (ballistic gel or duck breast) and layers of phantom intestines as
presented in Fig. 3(a) and (b). A custom-made rig from Perspex® was used to house the
tested medium or environment in each test, consisting of a bottom and top Perspex®
plate of 2 mm each and Perspex® spacers to maintain homogeneous thickness for the
soft tested media and vertical orientation and alignment of the transmitting and
receiving probes, as presented in Fig. 3. Initial tests yielded the transmitted power losses
11
Journal of Medical Devices - ASME
experienced by each Perspex® plate (top and bottom) at each tested frequency. Hence,
these transmission losses due to the Perspex® housing were excluded from the
calculated experimental results of the transmitted power through the phantom media
to be presented in the experimental results section.
NUMERICAL PREDICTIONS AND EXPERIMENTAL RESULTS
Computer Simulations of EM Energy Transmission through Multilayers to the Camera
Pill in the Small Intestine
First of all, data was collected from the literature [15, 24-25] about the
dimensions and material properties of each layer of human body in the abdomen and
abdominal cavity in the EM energy transmission line from an external point source to a
camera pill travelling in the small intestine, as presented in Fig. 1. This data and the
attenuation constant and reflection coefficient of each layer, calculated according to
relations (6) and (9) respectively, are presented in Table 2.
Computer simulations of the EM power transmission were carried out using our
algorithm, which implemented an iterative solution of equations (3)-(5) using the Gauss-
Seidel method. A power source of Pstart = 1 mW cm-2 of different frequencies in the range
of 0.181-5800 MHz (as reported in Table 2) was used at the external skin surface. Power
transmission to the camera pill was predicted for pill locations at different depths from
the skin surface, i.e. at the end of the abdomen, and after 1, 2, 3, 4 and 5 layers of small
intestine. The results are presented in Fig. 4. The abdomen consisted of five layers
comprising the skin, fat, muscle, peritoneum and omentum layers of thickness and
12
Journal of Medical Devices - ASME
properties as presented in Table 2. The internal reflections at layer interfaces increased
significantly the efficiency of power transmission through the abdomen despite the
relatively large thickness of the fat and omentum layers. As a result, power transmitted
through the abdomen experienced no net losses in the EM induction frequency range of
0.181-28 MHz, was reduced to about 20% and 10% at the RF frequencies of 433.92 and
915 MHz, respectively, and to 1% and 0.1% at the RF frequencies of 2450 and 5800 MHz,
respectively. Deeper locations of the pill further reduced the transmitted power by 10 to
100 times approximately as the depth increased from 1 to 5 layers of the small intestine.
The energy safety limit is another important factor that would ultimately affect
the possible amount of energy to be harvested by an electrofunctional pill. Table 1
presents the formulae for the power density limit as a function of frequency, classified
into different frequency ranges by FCC. These power density limits at different
frequencies have been calculated in Table 2 and apply for a safety time limit of 30 min.
Considering an average pill journey of 8 hr, a power density, Pstart = Slim/16, has been used
at the external skin surface, which varies as a function of frequency as displayed in Fig.
5. The computer simulations of the power transmission through the abdomen and at
different depths inside the abdomen were repeated with Pstart at the safety limit
according to the EM power frequency and the results are presented in Fig. 5. In this
case, it can be seen that at 0.181 MHz over 60 mWh and 8-12 mWh can be harvested by
the pill during 8 hr at a location just behind the abdomen and in the intestines,
respectively, which may add 0.2-1.2 hr of pill operation, considering that the current pill
batteries of 330 mWh in total last for 8 hr. Due to the imposed stricter power safety
13
Journal of Medical Devices - ASME
limit at frequencies above 1.34 MHz (Table 1), the energy that can be harvested at 13.5
and 28 MHz falls to 0.6 mWh and 0.015 mWh for 8 hr behind the abdomen,
respectively, and to 0.04-0.1 mWh and 0.007-0.02 mWh in the intestines, respectively,
despite the low transmission losses at these frequencies (as seen in Fig. 4). The
harvested energy becomes lower at 0.027 mWh for 8 hr just behind the abdomen (and
even less in deeper layers of the small intestine) at the RF frequencies of 433 MHz and
915 MHz and falls further to 0.006 mWh for 8 hr behind the abdomen at 2450 MHz.
Experimental and Simulation Studies of RF Power Transmission through Phantom
Abdomen Environments
The experimental results of the EM power transmission through a single layer
and multiple layers of phantom small intestine (made from gelatin with embedded
chyme) are presented in Fig. 6 for a probe transmitted power of 3.162 mW. Computer
simulations of this RF power transmission were carried out, starting with the single small
intestine layer to determine the physical parameters for this medium from the
numerical fit of the predictions with the experimental results. In this manner, the
attenuation constant of the phantom small intestine was fitted to the value of int = 1.7,
13, 20, 83, 90, 95 and 99 m-1 at the frequencies of 0.3, 13.5, 28, 433, 915, 1800 and 2450
MHz, respectively. These values then were used in the computer simulations of the EM
power transmission through multiple layers of phantom small intestine and the
predictions were in relatively good agreement with the corresponding experimental
results as shown in Fig. 6, given also the experimental error which reaches a maximum
14
Journal of Medical Devices - ASME
±14%. However, the attenuation constant of these phantom small intestines is higher
than that of the human intestines, as presented in Table 2, and changes less at high
frequencies than for the human intestines. Human intestines are composed mainly of
collagen and elastin composite gel and gelatin is a collagen derivative. However, it is
unclear what type of chyme, if any, is associated with the properties of the small
intestine included in Table 2, and different types of chyme could bring significant
differences in the electromagnetic properties of the composite material.
A ballistic gel mass of thickness of 23 mm was used to match best the total
transmitted power ratio (Pex/Pstart) of the real abdomen across the high frequencies (433-
2450 MHz) of the investigated frequency spectrum, where the largest transmission
losses occurred according to the predicted RF power transmission through the human
abdomen in Fig. 4. The experimental data for the ballistic gel phantom abdomen
presented in Fig. 6 seem to be in relative agreement with the predictions for the real
abdomen (Fig. 4) at the frequencies of 433-2450 MHz, within the experimental error of
±10%. However, they differ by about 30% at the low frequencies of 13 and 28 MHz,
with the ballistic gel phantom abdomen exhibiting higher transmission losses. Computer
simulations were used to fit the attenuation constant of this ballistic gel abdomen to bg
= 6, 8, 11, 39.4, 60, 86 and 100 m-1 at the frequencies of 0.3, 13.5, 28, 433.92, 915, 1800
and 2450 MHz, respectively, and the results of these predictions are also presented in
Fig. 6. Further computer simulations were conducted for RF transmission through
environments of ballistic gel abdomen and 1 or 2 layers of phantom small intestine, with
good agreement with the experimental data of the latter case after the numerical fit of
15
Journal of Medical Devices - ASME
the reflection coefficient from the ballistic gel to the phantom small intestine to r f,bg/int =
0.4 for all investigated frequencies in the range of 0.3 – 2450 MHz.
Fig. 7 presents the corresponding experimental data and predictions for a duck
breast phantom abdomen of a thickness of 20 mm. The duck breast is a multilayer
material incorporating visible layers of skin, fat and muscle and the experimental data
presented in Fig. 7 had a high experimental error of ±30%. The experimental data of the
total transmitted power ratio (Pex/Pstart) in Fig. 7 differed from the predictions of Pex/Pstart
through the human abdomen in Fig. 4 by +4% (higher in Fig.7) at 0.3 MHz, -48% at 13.5
MHz, -32% at 28 MHz, +244% at 433 MHz, -28% at 915 MHz and -21% at 2450 MHz.
Computer simulations were used to fit the attenuation constant of the duck breast to
the values of db = 3.4, 13, 15, 24, 63, 95 and 111 m-1 at the frequencies of 0.3, 13.5, 28,
433, 915, 1800 and 2450 MHz, respectively. EM power transmission was also measured
and simulated through environments of the duck breast phantom abdomen and 1, 2, 3
and 4 layers of phantom small intestine, with all results presented in Fig. 7. Generally
good agreement is observed between experimental results and predictions for the
combined duck breast-intestine phantom environments with the reflection coefficient
from the duck breast to the phantom small intestine fitted to rf,db/int = 0.4 in the range of
0.3-2450 MHz.
DISCUSSION
This discussion section will start with an analysis of the variability and
repeatability of the data of transmitted RF power through the human abdomen and
16
Journal of Medical Devices - ASME
abdominal cavity environment (predictions of numerical simulations) or the
environment of the phantom organs fabricated or investigated in this study
(experimental data and predictions). It will then proceed to the results of the RF energy
harvesting from this study and compare them to estimates of other forms of energy
harvesting for a camera pill.
The human abdomen and abdominal cavity is a complex environment and while
it is well represented by the layer diagram in Fig. 1, the thickness and thickness
distribution, composition and density of each layer vary enormously between people,
also depending on age and various conditions such as obesity and diabetes. The range
of waist circumference values may be used as an indication of the cumulative variability
and is considered to be between 60 and 120 cm [26]. A major factor for this variability is
the amount of visceral fat (also known as visceral adipose tissue) that is accumulated
between the organs in the abdominal cavity (in addition to the omentum layer in Fig. 1)
and in this case between the intestine length folds as is evidenced in MRI scans [27]. The
amount of visceral fat between the intestine folds varies enormously between people
and is not included in the human abdominal cavity environment model in this study as is
depicted in Fig.1. The intestine has a total length of 8±1.3 m, as reported by Hounnou et
al [28], and folds in its length so the power transmission to the pill might be through up
to 3 intestine parts in line for most of the length of the intestine [27]; up to 5 intestine
parts can be seen in an abdominal MRI image cross-section [27] from the skin to the
spine but are usually staggered and only in some parts of the abdominal cavity may be
found all in-line. Hence, the predicted power transmission and harvesting by a camera
17
Journal of Medical Devices - ASME
pill in the human intestine in Fig. 4 and Fig. 5 may be considered through the abdomen
and 1, 2, or 3 intestines. In fact, the pill is located inside an intestine part rather than at
its external wall, so the power is transmitted through the abdomen and 0.5, 1.5 or 2.5
intestines to reach the pill. Furthermore, it is unclear whether the properties of the
small intestine compiled from the literature [24] were measured for specimens of
intestine full with chyme (digested food) and from what type of food this chyme
originated. Given these uncertainties, on the assumption of the pill been a quarter of
the time in each part of the small intestine, i.e. after the abdomen or after the abdomen
and 1 or 2 or 3 intestines, and a starting power density at the outer skin surface equal to
the safety limit for the transmitted RF frequency, the following values of harvested
energy can be estimated after 8 hours of pill journey for different RF frequencies: 23,
0.21, 0.04, 0.01, 0.01, 0.002, 0.002 mWh for 0.18, 13.5, 28, 433, 915, 2450 and 5800
MHz, respectively.
With regards to the ballistic gel phantom abdomen, this is a single-type of
material with relatively low experimental error (±10%) in the transmitted power
measurements (Fig. 6) with regards to variability across the specimen surface (10
measurements) or between specimens (3 different specimens). The power losses
through the ballistic gel phantom abdomen are similar to those through the human
abdomen at the RF frequencies of 433-2450 MHz, whereas the ballistic gel experiences
lower transmitted power than the human abdomen at low frequencies, below 30 MHz
possibly due to the lack of reflected power at interfaces given that it has no different
layers.
18
Journal of Medical Devices - ASME
The duck breast phantom abdomen, on the other hand, is a complex material
system with visible skin, fat and muscle layers. As with the human abdomen and
abdominal cavity, the thickness and thickness distribution, composition and density of
each layer may vary across the same breast part and also between individual ducks: the
experimental data of the transmitted power through the duck breast in Fig. 7 exhibited
an experimental error of ±30% across the surface of the same breast part, due not only
to material layer variability but also to the shape of the original breast piece which
although it was compressed to a homogeneous thickness in the Perspex® housing, its
thicker part in the middle was compressed more and, hence, there was density
variation. The variability of the experimental data between two different duck breasts
from different suppliers rose to ±70%. However, the data of power transmitted through
the duck breast agreed with those through the human abdomen at the low frequency of
0.181 MHz where losses were low and the transmitted power was reinforced by
reflections at the interfaces between layers, given that both human abdomen and duck
breast phantom abdomen are multilayer materials.
It has been thought to compare the value of RF energy harvesting by a camera
pill to other energies possible to be harvested by a camera pill, including mechanical and
electrochemical energy.
Mechanical energy can be generated from the motion of the pill, the intestine
peristalsis and vibrations generated by the pill locomotion. The small intestine
undergoes peristalsis at a rate f = 11 contractions per minute to move chyme at an
average speed of 1 cm s-1 while the large intestine undergoes peristalsis at a rate f = 1-3
19
Journal of Medical Devices - ASME
contractions per minute to move chyme at an average speed of 0.083 cm s -1 [29].
Following the force analysis and external area function of a capsule pill proposed by
Woo et al [30], a force F = 1.22-1.42 N was estimated as exerted by the intestinal walls
on the pill [30] which is assumed for a collision duration t = 1 s while two collisions are
assumed per pulse. Assuming a pill velocity, v, equal to that of the interstitial fluid, an
energy conversion factor = 0.5 and a transducer conversion efficiency t = 0.5, the
following power may be harvested from the peristalsis of the small intestine:
p=2 fΔtFv ηηt=1 .1 mW (10)
Considering a total pill journey of 8 hours through the small intestine, the maximum
total harvested energy from intestinal peristalsis would be about 9 mWh which is 2.7%
of the 330 mWh usually stored by the coin batteries of the camera pill. Adding to this
energy the kinetic energy from the pill motion and pill rotation at each collision with the
intestine wall for a pill of a mass m = 4 g, radius of 5.5 mm and length of 23 mm, a
gyroscope efficiency factor of 0.5, and also = 0.5 and t = 0.5 as previously, a total
energy of 0.7x10-4 mWh was estimated as harvested from the kinetic energy of the
moving pill, which is negligible.
Energy may also be harvested by the change of pH as the pill travels through the
GI tract [31], which may bring a measurable voltage change in a pH sensor (about 33 mV
per unit of pH change [32]), which several types of commercial diagnostic pills actually
carry. Considering a charge current of 110 mA for charging the battery and an efficiency
20
Journal of Medical Devices - ASME
factor = 0.5, estimates of the present study for the mean power and harvested energy
are presented in Table 3 for the duration of the pill journey in different parts of the GI
tract with changing pH [33]. The total electrochemical energy harvesting due to the
change of pH in the GI tract is estimated to be 9.5 mWh which is about 2.9 % of the 330
mWh usually stored by the coin batteries of the camera pill.
CONCLUSIONS
A comprehensive multilayer EM power transmission model was developed and
implemented in this study, taking into account multi-reflections at material interfaces.
This model approach led to accurate efficiency predictions of wireless power
transmission through the abdominal cavity to an electrofunctional pill. The predicted
EM energy that can be harvested by the pill in the small intestine is below 1 mW cm -2 in
the EM frequency range above 10 MHz, due to the energy safety limit of skin exposure.
For low frequencies below 1.34 MHz the high allowed safe energy value of 100 mW cm -2
for 30 min results in transmitted and harvested EM energy of 10-50 mWh depending on
pill location during its journey of 8 hours in the GI tract in the abdominal cavity. Such
harvested energy could be added to a mechanical energy of about 10 mWh harvested
by the pill colliding with the intestine in its peristalsis and an energy of about 10 mWh
harvested by the pill from pH changes in the GI tract, reaching a total harvested energy
of 30-80 mWh, which may supplement the energy of 330 mWh stored in typical camera
pill batteries and prolong the active operation of a camera pill by 44 to 116 min.
21
Journal of Medical Devices - ASME
Both a duck breast and a ballistic gel of 23 mm thickness may represent a human
multilayer abdominal wall in EM power transmission, excluding the intestinal part or
other internal organs. The ballistic gel is a single-type material with low variability and
high repeatability in terms of its RF transmission properties. It matches a standard
human abdomen (Table 2) well at the RF frequencies of 433-2450 MHz but experiences
lower overall transmitted power at frequencies below 30 MHz, most possibly due to the
fact that it is not a multilayer material and lacks reflected power at interfaces. For this
reason, the duck breast, being a multilayer material, matched the human abdomen at
low frequencies. On the other hand, there was substantial variability between different
specimens of duck breast originating from different duck sources, as in fact is also the
case with human specimens. Gelatin gel layers of 7 mm thickness with embedded
phantom chyme exhibited higher attenuation constant than a standard human small
intestine (Table 2) although it is not known whether the literature data for the human
intestine was for intestines containing chyme and what type of chyme.
In conclusion, the real value of a comprehensive numerical model, including
multi-reflections at material interfaces, has been demonstrated in both the predictions
of the computational simulations and determining appropriate experimental phantom
models representing the human abdomen environment at different RF frequencies.
22
Journal of Medical Devices - ASME
NOMENCLATURE
CMOS Complementary metal-oxide semiconductor
EM Electromagnetic
GI Gastrointestinal
GSM Global System for Mobile Communications
ISM Industrial Scientific and Medical (band)
LED Light emitting diode
MICS Medical Implant Communication System
RF Radio frequency
SAR Specific absorption rate
c Speed of light
Elim Energy density safety limit
f Frequency
F Force
i-1, i, i+1 Layers of abdomen
m Mass
p Power
po Power emitted from a point EM power source
P1,ent Power density entering the skin layer
23
Journal of Medical Devices - ASME
Pi,ex Power density arriving at the exit of layer i
Pex Power density at the exit of the tested phantom organ assembly
Pi,start Power density at the start of layer i
Pstart Power density at the start of the skin layer
rf Reflection fraction of power
rf,bg/int Power reflection fraction at the interface between ballistic gel and
intestine
rf,db/int Power reflection fraction at the interface between duck breast and
intestine
rf,i-1/i Power reflection fraction at the interface between layers i-1 and i
Rair Distance between the EM point source and the skin
Ri Distance between the EM point source and layer i
Slim Power density limit
tf Transmission fraction of power
tf,i-1/i Power transmission fraction at the interface between layers i-1 and i
tlim Average time limit
v Velocity
VpH Open circuit voltage generated by a pH difference
Greek
symbols
24
Journal of Medical Devices - ASME
air Attenuation constant of air
bg Attenuation constant of ballistic gel
db Attenuation constant of duck breast
i Attenuation constant of layer i
int Attenuation constant of intestine
ri Thickness of layer i
t Time duration
ε r ,i' Real component of the material relative complex permittivity of layer i
ε r ,i' ' Imaginary component of the material relative complex permittivity of
layer i
Energy conversion factor
i Intrinsic impedance of the material of layer i
t Transducer conversion efficiency
r,i Material relative magnetic permeability of layer i
i Conductivity of the material of layer i
25
Journal of Medical Devices - ASME
REFERENCES
[1] Kleiner, H., 1966, “Fantastic voyage”, Science Fiction Film.
[2] Toennies, J. L., Tortora, G., Simi, M., Valdastri, P., and Webster III, R. J., 2010, “Swallowable medical devices for diagnosis and surgery: the state of the art”, Proc. IMechE Part C: J. Mechanical Engineering Science, 224, pp. 1397-1414
[3] Pan, G., and Wang, L., 2012, “SwallowableWireless Capsule Endoscopy: Progress and Technical Challenges”, Gastroenterology Research and Practice, 2012, Article ID 841691, 9 pp
[4] Hale, M. F., Sidhu, R., and McAlindon, M. E., 2014, “Capsule endoscopy: Current practice and future directions”, World J. Gastroenterol., 20(24), pp. 7752-7759
[5] Koprowski, R., 2015, “Overview of technical solutions and assessment of clinical usefulness of capsule endoscopy”, BioMed. Eng. OnLine, 14(111), pp. 1-23
[6] Barkin, J.A., and Barkin, J.S., 2017, “Video capsule endoscopy: Technology, reading, and troubleshooting”in “Evaluation of the Small Bowel”, Editor: Gerson, L.B., Gastrointestinal Endoscopy Clinics of North America, Elsevier, Philadelphia, Penn, pp. 17-19
[7] Qureshi, W.A., 2004, “Current and future applications of the capsule camera”, Nature Reviews Drug Discovery, 3, pp. 447-450
[8] Cleveland, Jr., R. F., Sylvar, D. M., and Ulcek, J. L., 1997, “Evaluating Compliance with FCC Guidelines for Human Exposure to Radiofrequency Electromagnetic Fields”, Federal Communications Commission Office of Engineering & Technology, Washington, D.C., OET Bulletin 65, Edition 97-01
[9] Wang, L., Johannessen, E. A., Hammond, P. A., Cui, L., Reid, S. W. J., Cooper, J. M., and Cumming, D. R. S., 2005, “A programmable microsystem using system-on-chip for real-time biotelemetry”, IEEE Trans. Biomedical Engineering, 52(7), pp. 1251-1260
[10] Pan, G., Xin, W., Yan, G., and Chen, J., 2011, “A video wireless capsule endoscopy system powered wirelessly: design, analysis and experiment”, Meas. Sci. Technol., 22, 065802, 9 pp
[11] Xin, W., Yan, G., and Wang, W., 2010, “Study of a wireless power transmission system for an active capsule endoscope”, Int. J. Med Robotics Comput. Assist. Surg., 6, pp. 113–122
26
Journal of Medical Devices - ASME
[12] Yuce, M. R., Dissanayake, T., and Keong, H. C., 2009, “Wireless Telemetry for Electronic Pill Technology”, in Sensors, 2009 IEEE, Christchurch, UK, pp. 1433 – 1438
[13] Higgins H., 2006, “Wireless Communication” in Body Sensor Networks, G-Z. Yang, Ed. London: Springer, pp. 117-143
[14] Kurup, D., Vermeeren, G., Tanghe, E. , Joseph, W., and Martens, L., 2015, “In-to-Out Body Antenna-Independent Path Loss Model for Multilayered Tissues and Heterogeneous Medium”, Sensors, 15, pp. 408-421
[15] Christ, A., Klingenböck, A., Samaras, T., Goiceanu, C., and Kuster, N., 2006, “The Dependence of Electromagnetic Far-Field Absorption on Body Tissue Composition in the Frequency Range From 300 MHz to 6 GHz”, IEEE Trans. Microwave Theory and Techn., 54(5), pp.2188-2195
[16] Nguyen, C., and Han, J., 2014, “Chapter 2: System Analysis” in “Time-Domain Ultra-Wideband Radar, Sensor and Components: Theory, Analysis and Design”, New York, Springer, pp.7-23
[17] Sadiku, M. N. O., 1994,"Elements of Electromagnetics”, New York: Saunders College Publishing
[18] Graham, M. F., Diegelmann, R. F., Elson, C. O., Lindblad, W. J., Gotschalk, N., Gay, S., and Gay, R., 1988, “Collagen Content and Types in the Intestinal Strictures of Crohn’s disease”, Gastroenterology, 94, pp. 257-265
[19] Elsayed, Y., Lekakou, C., Labeed, F., and Tomlins, P., 2016, “Smooth muscle tissue engineering in crosslinked electrospun gelatin scaffolds”, J. Biomed. Mat. Res. Part A, 104(1), pp. 313–321
[20] Elsayed, Y., Lekakou, C., and Tomlins, P., 2014, “Monitoring and modelling of oxygen transport through un-crosslinked and crosslinked gelatine gels”, Polymer Testing, 40, pp. 106-115
[21] Liu, L., Yao, W.D., Rao, Y.F., Lu, X.Y., and Gao, J.Q., 2017, “pH-Responsive carriers for oral drug delivery: challenges and opportunities of current platforms”, Drug Delivery, 24(1), pp. 569-581
[22] Lee, S. H., Lee, J., Yoon, Y. J., Park, S., Cheon, C., Kim, K., and Nam, S., 2011, “A Widebrand Spiral Antenna for Ingestible Capsule Endoscope Systems: Experimental Results in a Human Phantom and a Pig”, IEEE Trans. Biomed. Eng., 58(6), pp. 1734-1741
[23] Cronin, D. S., and Falzon, C, 2011, “Characterization of 10% Ballistic Gelatine to Evaluate Temperature, Aging and Strain rate Effects”, Experimental Mechanics, 51(7),
27
Journal of Medical Devices - ASME
pp. 1197-1203
[24] Gabriel, C., 1996, "Compilation of the dielectric properties of body tissues at RF and microwave frequencies", Brooks Air Force Base, report no. AL/OE-TR-1996-0037
[25] Tural’chuk, P. A. , Vendik, O. G., and Vendik, I. B., 2015, “Propagation of Electromagnetic Waves in Biological Media:Refraction at Interfaces”, Techn. Physics Letters, 41(3), pp. 270–272
[26] Després, J. P., Lemieux I., and Prud'homme D., 2001, “Treatment of obesity: need to focus on high risk abdominally obese patients”, BMJ., 322(7288), pp. 716-20.
[27] Mantatzis, M., Milousis, M., Katergari, S., Delistamatis, A., Papachristou, D. N., and Prassopoulos, P., 2014, “Abdominal Adipose Tissue Distribution on MRI and Diabetes“, Acad. Radiol., 21, pp. 667–674
[28] Hounnou, G., Destrieux, C., Desmé, J., Bertrand, P., and Velut, S., 2002, “Anatomical study of the length of the human intestine”, Surgical and Radiologic Anatomy”, 24(5), pp. 290–294
[29] Pocock, G., and Richards, C. D., 2009, “The Large Intestine”, in The Human Body, an Introduction for the Biomedical and Health Science, New York: Oxford University Press, p. 528
[30] Woo, S. H., Kim, T. W., Mohy-Ud-Din, Z., Park, Il Y., and Cho, J-H., 2011, “Small intestinal model for electrically propelled capsule endoscopy”, BioMedical Engineering OnLine, 10(108), DOI: 10.1186/1475-925X-10-108, 20 pp
[31] He, C., Kiziroglou, M. E., Yates, D., and Yeatman, E., 2010, “MEMS Energy Harvester for Wireless Biosensors”, in Proc. IEEE Conf. Micro-electro-mechanical Systems, Hong Kong, pp. 172-175
[32] Guillaumin, J. G., Sadowski, D., Kaler, K., and Mintchev, M., 2007, “Ingestible capsule for impedance and pH monitoring in the esophagus”, IEEE Trans, Biomed. Engng, 54(12), pp. 2231-2236
[33] Zarate, N., Mohammed, S. D., O'Shaughnessy, E., Newell, M., Yazaki, E., Williams, N. S., Lunniss, P. J., Semler, J. R., and Scott, S. M., 2010, “Accurate localization of a fall in pH within the ileocecal region: validation using a dual-scintigraphic technique”, American J. Physiology - Gastrointestinal and Liver Physiology, 299(6), pp. G1276–G1286
28
Journal of Medical Devices - ASME
Figure Captions List
Fig. 1 Electromagnetic (EM) energy transmission from an external point source
to an electrofunctional pill travelling in the intestine via the sequence of
multiple material layers of the abdominal cavity
Fig. 2 General EM energy transmission scheme across the general layer i and its
neighboring layers i-1 and i+1, indicating transmitted and reflected
power densities at layer interfaces, and attenuated power densities
across the layer i
Fig. 3 Phantom abdomen environments in a custom-made housing rig from
Perspex® used in the experimental studies of RF power transmission: (a)
ballistic gel phantom abdomen and phantom small intestines with
embedded chyme; (b) phantom abdomen of duck breast with skin and
phantom small intestines with embedded chyme.
Fig. 4 Computer simulation predictions of the harvested power from a power
source of 1 mW cm-2 at skin surface to the camera pill located at different
depths (at the end of the abdomen or in different intestine layers) as a
function of the radiation wave frequency
Fig. 5 Computer simulation predictions of the harvested power from a power
source of safety limit power density at skin surface to the camera pill
located at different depths (at the end of the abdomen or in different
intestine layers) as a function of the radiation wave frequency
Fig. 6 Results of transmitted power at different frequencies in the range of 433-
29
Journal of Medical Devices - ASME
2450 MHz from corresponding experiments of wireless EM energy
transmission through different phantom environment components (1-4
phantom small intestines with embedded chyme; phantom abdomen
from ballistic gel) and phantom environments (ballistic gel phantom
abdomen and 1-2 phantom small intestines with embedded chyme). (All
data points include % error bars derived from the maximum error
between three repeat experiments using new materials). Lines of
predictions with the following parameter fitted values: phantom
intestines (ri = 7 mm): int = 1.7 m-1 (at 0.3 MHz), int = 13 m-1 (at 13.5
MHz), int = 20 m-1 (at 28 MHz), int = 83 m-1 (at 433 MHz), 90 m-1 (at 915
MHz), 95 m-1 (at 1800 MHz), 99 m-1 (at 2450 MHz); rf,int = 0; ballistic gel
phantom abdomen (ri = 23 mm): bg = 6 m-1 (at 0.3 MHz), bg = 8 m-1 (at
13.5 MHz), bg = 11 m-1 (at 28 MHz), bg = 39.4 m-1 (at 433 MHz), 60 m-1
(at 915 MHz), 86 m-1 (at 1800 MHz), 100 m-1 (at 2450 MHz); rf,bg = 0.40
(0.3-2450 MHz).
Fig. 7 Results of transmitted power at different frequencies in the range of 0.3-
2450 MHz from corresponding experiments of wireless EM energy
transmission through phantom abdomen from duck breast and phantom
environments (phantom abdomen from duck breast and 1-4 phantom
small intestines with embedded chyme). (All data points include % error
bars derived from the maximum error between 3 repeat experiments
using new materials and 10 different points in the area of each
30
Journal of Medical Devices - ASME
specimen). Lines of predictions with the following parameter fitted
values: phantom intestines (ri = 7 mm): int = 1.7 m-1 (at 0.3 MHz), int =
13 m-1 (13.5 MHz), int = 20 m-1 (28 MHz), int = 83 m-1 (433 MHz), 90 m-1
(915 MHz), 95 m-1 (1800 MHz), 99 m-1 (2450 MHz); rf,int = 0; duck breast
phantom abdomen (ri = 20 mm): db = 3.4 m-1 (at 0.3 MHz), db = 13 m-1
(13.5 MHz), db = 15 m-1 (28 MHz), db = 24 m-1 (433 MHz), 63 m-1 (915
MHz), 95 m-1 (1800 MHz), 111 m-1 (2450 MHz); rf,db = 0.40 (0.3-2450
MHz).
31
Journal of Medical Devices - ASME
Table Caption List
Table 1 FCC Electromagnetic Energy Safety Limits for Uncontrolled Exposure for
an Average Time Limit tlim of 30 minutes
Table 2 Data and Parameters Calculated for Each Layer in the Wireless Energy
Transmission Line to an Electrofunctional Pill from an External EM Energy
Source, for Different EM Wave Frequencies
Table 3 Parameters and Estimated Energy Harvested from the pH Change in
Different Regions of the GI Tract
32
Journal of Medical Devices - ASME
Fig. 1. Electromagnetic (EM) energy transmission from an external point source to an
electrofunctional pill travelling in the intestine via the sequence of multiple material
layers of the abdominal cavity
33
Journal of Medical Devices - ASME
Fig. 2. General EM energy transmission scheme across the general layer i and its
neighboring layers i-1 and i+1, indicating transmitted and reflected power densities at
layer interfaces, and attenuated power densities across the layer i
34
Journal of Medical Devices - ASME
Fig. 3. Phantom abdomen environments in a custom-made housing rig from Perspex®
used in the experimental studies of RF power transmission: (a) ballistic gel phantom
abdomen and phantom small intestines with embedded chyme; (b) phantom abdomen
of duck breast with skin and phantom small intestines with embedded chyme.
35
Journal of Medical Devices - ASME
Fig. 4. Computer simulation predictions of the harvested power from a power source of
1 mW cm-2 at skin surface to the camera pill located at different depths (at the end of
the abdomen or in different intestine layers) as a function of the radiation wave
frequency
36
Journal of Medical Devices - ASME
Fig. 5. Computer simulation predictions of the harvested power from a power source of
safety limit power density at skin surface to the camera pill located at different depths
(at the end of the abdomen or in different intestine layers) as a function of the radiation
wave frequency
37
Journal of Medical Devices - ASME
Fig. 6. Results of transmitted power at different frequencies in the range of 433-2450
MHz from corresponding experiments of wireless EM energy transmission through
different phantom environment components (1-4 phantom small intestines with
embedded chyme; phantom abdomen from ballistic gel) and phantom environments
(ballistic gel phantom abdomen and 1-2 phantom small intestines with embedded
chyme). (All data points include % error bars derived from the maximum error between
three repeat experiments using new materials). Lines of predictions with the following
parameter fitted values: phantom intestines (ri = 7 mm): int = 1.7 m-1 (at 0.3 MHz), int
= 13 m-1 (at 13.5 MHz), int = 20 m-1 (at 28 MHz), int = 83 m-1 (at 433 MHz), 90 m-1 (at 915
MHz), 95 m-1 (at 1800 MHz), 99 m-1 (at 2450 MHz); rf,int = 0; ballistic gel phantom
abdomen (ri = 23 mm): bg = 6 m-1 (at 0.3 MHz), bg = 8 m-1 (at 13.5 MHz), bg = 11 m-1
(at 28 MHz), bg = 39.4 m-1 (at 433 MHz), 60 m-1 (at 915 MHz), 86 m-1 (at 1800 MHz), 100
m-1 (at 2450 MHz); rf,bg = 0.40 (0.3-2450 MHz).
38
Journal of Medical Devices - ASME
Fig. 7. Results of transmitted power at different frequencies in the range of 0.3-2450
MHz from corresponding experiments of wireless EM energy transmission through
phantom abdomen from duck breast and phantom environments (phantom abdomen
from duck breast and 1-4 phantom small intestines with embedded chyme). (All data
points include % error bars derived from the maximum error between 3 repeat
experiments using new materials and 10 different points in the area of each specimen).
Lines of predictions with the following parameter fitted values: phantom intestines (ri =
7 mm): int = 1.7 m-1 (at 0.3 MHz), int = 13 m-1 (13.5 MHz), int = 20 m-1 (28 MHz), int =
83 m-1 (433 MHz), 90 m-1 (915 MHz), 95 m-1 (1800 MHz), 99 m-1 (2450 MHz); rf,int = 0; duck
breast phantom abdomen (ri = 20 mm): db = 3.4 m-1 (at 0.3 MHz), db = 13 m-1 (13.5
MHz), db = 15 m-1 (28 MHz), db = 24 m-1 (433 MHz), 63 m-1 (915 MHz), 95 m-1 (1800
MHz), 111 m-1 (2450 MHz); rf,db = 0.40 (0.3-2450 MHz).
39
Journal of Medical Devices - ASME
Table 1. FCC Electromagnetic Energy Safety Limits for Uncontrolled Exposure for an
Average Time Limit tlim of 30 minutes
Frequency Range (MHz)
Power Density* Safety Limit, Slim (mW cm-2)
(*Plane-wave equivalent)
0.3 – 1.34 1001.34 – 30. 180/f2
30 - 300 0.2300 - 1500 f/15001500 –1.0x105 1.0
40
Journal of Medical Devices - ASME
41
TABLE 2. Data and Parameters Calculated for Each Layer in the Wireless Energy Transmission Line to an
Electrofunctional Pill from an External EM Energy Source, for Different EM Wave Frequencies
Dimension or Property of Layer i
Layer 1Skin
Layer 2Fat
Layer 3Muscle
Layer 4Peritoneum
Layer 5Omentum
Layer 6Small Intestine
ri (mm) 3.3 30 8.7 0.0001 15 70.181 MHzSlim (mW cm-2)ε r ,i
'
ε r ,i' '
i (m-1)
|ρi−1/i2 |
1001107.50391.0980.005
59.7992439.0440.131
|ρ skin / fat2 |=0.
6638.26737708.3270.477
|ρ fat /mus2 |=0.957
108.999198620.0031.195
|ρmus/ per2 |=0.
59.7992439.0440.131
|ρ per/om2 |=
0.969
11848.95461126.0840.602
|ρ fat /int2 |=0.783
13.56 MHzSlim (mW cm-2)ε r ,i
'
ε r ,i' '
i (m-1)
|ρi−1/i2 |
0.979285.253315.5292.379
11.82540.2361.103
|ρ skin / fat2 |=0.355
138.434832.7095.338
|ρ fat /mus2 |=
0.809
108.2632656.61310.149
|ρmus/ per2 |=0
11.82540.2361.103
|ρ per/om2 |=
0.784
362.6521836.9347.808
|ρ fat /int2 |=0.799
28 MHzSlim (mW cm-2)ε r ,i
'
ε r ,i' '
i (m-1)
|ρi−1/i2 |
0.230161.083213.7404.284
8.33921.2011.577
|ρ skin / fat2 |=0.403
94.466420.7377.615
|ρ fat /mus2 |=0.736
106.0631294.49714.331
|ρmus/ per2 |=0.
8.33921.2011.577
|ρ per/om2 |=
0.737
197.723950.38611.537
|ρ fat /int2 |=0.781
433.92 MHzSlim (mW cm-2)ε r ,i
'
ε r ,i' '
i (m-1)
|ρi−1/i2 |
0.28946.06129.09318.658
5.5661.7263.288
|ρ skin / fat2 |=0.467
56.86633.35019.354
|ρ fat /mus2 |=0.390
70.63193.62243.920
|ρmus/ per2 |
=0.533
5.5661.7263.288
|ρ per/om2 |=0.356
65.26479.59439.467
|ρ fat /int2 |=0.401
915 MHzSlim (mW cm-2)ε r ,i
'
ε r ,i' '
i (m-1)
|ρi−1/i2 |
0.6141.32917.12525.031
5.4601.0104.126
|ρ skin / fat2 |=0.712
54.99718.62523.752
|ρ fat /mus2 |=0.584
68.60747.51652.251
|ρmus/ per2 |
=0.431
5.4601.0104.126
|ρ per/om2 |=0.520
59.38842.68250.277
|ρ fat /int2 |=0.370
2450 MHzSlim (mW cm-2)ε r ,i
'
ε r ,i' '
i (m-1)
|ρi−1/i2 |
1.038.00710.74244.302
5.2800.7678.345
|ρ skin / fat2 |=0.893
52.72912.75744.782
|ρ fat /mus2 |=0.715
66.24325.36978.646
|ρmus/ per2 |=0.
5.2800.7678.345
|ρ per/om2 |=0.728
54.42423.27879.292
|ρ fat /int2 |=0.932
5800 MHzSlim (mW cm-2)ε r ,i
'1.035.11411.520
4.9550.908
48.48515.377
60.46924.296
4.9550.908
48.67220.905
Journal of Medical Devices - ASME
42
Table 3. Parameters and Estimated Energy Harvested from the pH Change in Different
Regions of the GI Tract
Location pH Time (min)
VpH
(mV)Power(mW)
Energy (mWh)
Stomach-to-duodenum
5 32 165 9 4.85
Duodenum-to-distal small intestine
0.5 113 16.5 0.9 1.70
Distal small intestine-to-cecum
1.5 39 49.5 2.7 1.76
Cecum-to-descending colon
1 39 33 1.8 1.17
TOTAL ENERGY (mWh)
9.47