mg8g/mg8e paramagnetic oxygen analyzers · a magnetic wind oxygen analyzer is characterized by...
TRANSCRIPT
TechnicalInformation
Contents1. Introduction...................................................................................................22. TypesandApplicationsofOxygenAnalyzers..........................................3
2.1 ParamagneticOxygenAnalyzers....................................................................... 32.1.1 MagneticProportionalFlowRateMethod............................................ 4
2.1.2 MagneticWindMethod......................................................................... 4
2.1.3 MagneticForceMethod(DumbbellType)............................................ 6
2.1.4 MagneticForceMethod(PressureSensorType)................................. 6
2.2 ElectrochemicalOxygenAnalyzers................................................................... 72.2.1 ZirconiaSystem.................................................................................... 7
2.2.2 ElectrodeSystem.................................................................................. 9
2.3 AbsorptionSpectroscopyOxygenAnalyzer.....................................................102.3.1 TunableDiodeLaserSpectroscopy(TDLS)Method......................... 10
2.4 ApplicationsofOxygenAnalyzersbyMeasurementPrinciple.................... 112.5 OxygenAnalyzerSelectionFlowchart............................................................ 12
3. MG8G/MG8EFeatures.............................................................................. 134. MG8G/MG8ESystemConfigurationandSensorConstruction........... 14
4.1 SystemConfiguration....................................................................................... 144.2 SuitableApplicationsforMG8G/MG8EParamagneticOxygenAnalyzers.. 154.3 SensorConstruction......................................................................................... 16
5. EffectsofInterferenceGas....................................................................... 195.1 MeasurementofOxygen(O2)inthePresenceofInterferenceGases........ 195.2 DataonEffectsofInterferenceGases............................................................. 205.3 InterferenceGasCompensation...................................................................... 21
6. Characteristics(MG8ETestData)............................................................ 226.1 ResponseCharacteristic................................................................................... 226.2 FlowCharacteristics.......................................................................................... 226.3 AttitudeError...................................................................................................... 236.4 EffectsofAtmosphericPressureandLong-termDrift.................................. 23
RevisionInformation........................................................................................... 24
MG8G/MG8EParamagneticOxygenAnalyzers
TI11P03A05-01E
TI11P03A05-01E©CopyrightSep.2009(YK)2ndEditionOct.2012(YK)
YokogawaElectricCorporation2-9-32,Nakacho,Musashino-shi,Tokyo,180-8750JapanTel.:81-422-52-5617 Fax.:81-422-52-6792
2
AllRightsReserved.Copyright©2009,YokogawaElectricCorporation TI11P03A05-01E
1.Introduction
1. IntroductionTheMG8G/MG8EParamagneticOxygenAnalyzerscanmeasuretheoxygenconcentrationinagasmixturewithhighprecision.Theydothisbasedontheprinciplethatoxygenisaparamagneticmaterial(magneticallysusceptible).Unlikezirconiaoxygenanalyzers,theMG8G/MG8Eanalyzerscanalsomeasure the oxygen concentration in a flammable gas mixture.
The MG8G/MG8E Paramagnetic Oxygen Analyzers use a magnetic proportional flow rate sensor that was developedbyYokogawatoimprovetheperformanceofitsmagneticwindsensors.Thishasthefollowingfeatures:
• Cleanauxiliarygas(N2) flows past the sensor in the detection unit without coming into contact with the samplegas,ensuringlong-termmeasurementstabilityevenwhencorrosiveorcontaminatedgasismeasured.
• Thermistors with high sensitivity and fast response times directly detect changes in the flow rate of the auxiliarygas,achievinga90%responsewithin3seconds.
• Therearenomovingparts,ensuringexcellentresistancetovibrationandshock.
• Aninterference-gascompensationfunctionensuresaccuratemeasurementsevenwhenflammablebackgroundgaseswithmagneticproperties(havingamuchlowerlevelofmagneticsusceptibilitycomparedtooxygen)arepresent.
• Excellentindicatingstabilityinthevicinityofzeromakesitsuitableforlow-concentrationmeasurements(e.g.forsafetycontrol).
Thistechnicalinformationdocument(TI11P03A05-01E)waspreparedwiththepurposeofhelpingusersunderstandandmakethebestuseofthesefeaturesoftheMG8G/MG8EParamagneticOxygenAnalyzers.Itdescribesthemeasurementmethodsusedbyvarioustypesofoxygenanalyzersaswellastheirfeaturesandapplications,thedesignandoperationoftheMG8G/MG8EParamagneticOxygenAnalyzers,andtheeffectsandcharacteristicsofinterferencegases.
Oct.30,2012-00
TI11P03A05-01E
32.TypesandApplicationsofOxygenAnalyzers
Oct.30,2012-00
2. TypesandApplicationsofOxygenAnalyzers
Oxygen,whichmakesupabout21%oftheair,isanessentialelementincombustionandcellularrespiration.Sinceoxygenisinvolvedinvarioustypesofchemicalprocesses,themeasurementandcontroloftheoxygenconcentrationareveryimportantforcontrollingtheseprocesses.
Therearevariousmethodsformeasuringtheoxygenconcentration,andthemajorityofthesefallintothetwofollowingroughcategories:(1)methodsthatusetheparamagneticpropertyofoxygen,(2)methodsthatusetheelectrochemicalpropertyofoxygen,and(3)methodsthatusetheabsorptionspectroscopyofoxygen.
2.1 ParamagneticOxygenAnalyzersFigure2.1comparestheparamagneticpropertiesofoxygenandothergases,andonlyoxygenexhibitsvery high magnetic susceptibility; most other gases have a relatively weak attraction to magnetic fields.
Nitrogenmonoxide(NO)alsoexhibitsafairlystrongmagneticsusceptibility;however,eveninexhaustgasesitsconcentrationisapproximatelyonly100ppm,whichissolowthatitsparamagneticpropertycanbeignored.
F0201.ai
Relative values of various gases assuming the bulk magnetic susceptibility of oxygen is 100
+21.6
Cl2CO2
NH3
CH4
N2O
N2
+6.2
+43.8
H2
C2H2
C2H4
+100
–0.128
–0.613
–0.575
–0.575
–0.376
–0.113
–0.37
–0.85
–0.42
NO2
Air
NO
O2
Figure2.1 RelativeBulkMagneticSusceptibilitiesofGasses
Paramagneticoxygenanalyzersusethefollowingmethods
(1) Magnetic proportional flow rate method(2) Magneticwindmethod(3) Magneticforcemethod(dumbbelltype)(4) Magneticforcemethod(pressuresensortype)
4
TI11P03A05-01E
2.TypesandApplicationsofOxygenAnalyzers
Oct.30,2012-00
2.1.1 MagneticProportionalFlowRateMethod
Exhaust gasSample + auxiliary gas
Thermistors
Magnet
Auxiliary gasSample gas
B
A
F0202.ai
Figure2.2 MagneticProportionalFlowRateSensor
AsshowninFigure2.2,asamplegasisintroducedfromthesensor’ssamplegasinletanddividedintotwostreamsinthering-shapedgasflowpath.Anauxiliarygasisintroducedfromtheauxiliarygasinletanddividedintotwostreams,AandB,whichflowtowardsthering-shapedgasflowpath.Athermistorisinstalled in each of these streams to determine their flow rates, and a magnet generates a magnetic field in streamB.
When a sample gas contains oxygen, the oxygen is drawn into the magnetic field, thereby decreasing the flow rate of auxiliary gas in stream B. The difference in the flow rates for streams A and B is proportional to theoxygenconcentrationofthesamplegas.Thisdifferenceisdeterminedandoutputbythethermistors.TheMG8G/MG8EParamagneticOxygenAnalyzersusethismethod.
A magnetic proportional flow rate oxygen analyzer is characterized by fast response and strong resistance tovibrationandshock.Furthermore,thethermistorsdonotcomeincontactwiththesamplegas,therebyensuringlong-termmeasurementstabilityevenwhencorrosiveorcontaminatedgasismeasured.
2.1.2 MagneticWindMethodThismethodwasdevelopedalongtimeagoandusesmanydifferenttypesofmeasurementcells.
Figure2.3showstwocylindricalchamberswithopposingconicallyshapedmagneticpolesthathaveflattopsandareseparatedbyaglass-coatedring-shapedheatingwireelement.Thechamberontheleftisthemeasurementchamber.Whenasamplegasisintroducedintothischamber,oxygenisdrawntowardtheheatingwire,wherethemagneticfieldisstrongest.Heatedbythewire,theoxygenlosesitsmagneticsusceptibilityandispushedupwardsbycoolergascomingfromthelowerpartofthechamber.Inthisway,aso-calledmagneticwindisgeneratedinthemeasurementchamber,anditsintensityisproportionaltotheoxygenconcentrationinthesamplegas.
Thetemperatureoftheheatingwireisrelatedtothecoolingeffectofthemagneticwind,andtothethermalconductivity, density, specific heat, and viscosity of the gas surrounding the heating wire.
Compensationisperformedwiththereferencechamber,whichisidenticalinconstructiontothemeasurement chamber, with the exception that no magnetic field is created. The two heating wires form a Wheatstonebridge,andtheresistancechangeisdetectedasachangeinthevoltage.
TI11P03A05-01E
52.TypesandApplicationsofOxygenAnalyzers
Oct.30,2012-00
Heatingelement
Sample gas
Measurementchamber
Reference chamber
Magneticwind
Magnet
Magnet
Heatingelement
Naturalconvection
Bridgeammeter
F0203.ai
vol%O2
Figure2.3 ExampleofConstructionofMagneticWindOxygenAnalyzer (withCylindricalMeasurementChamber)
Figure2.4showsameasurementcellwitharing-shapedgaspath.Athinglasstubewithaheatingwirewoundarounditprovidesadirectpaththroughthecenterofthecircularpath.Themagneticpoleofapermanentmagnetisplacedtotheleftofthecenterofthistube.Asamplegascontainingoxygenisattracted to the strongest part of the magnetic field generated by this magnet. When the oxygen is heated bythetube’sheatingwire,itlosesitsmagnetizationandispushedoutbycoolergasenteringfromtheleftsideofthetube;thus,amagneticwindpassingfromlefttorightisgeneratedinthetube.
Amagneticwindoxygenanalyzerischaracterizedbystrongresistancetovibrationandshock;ontheotherhand,sinceitsdetectionmethodreliesonthethermalconductivityofgas,itissusceptibletorapidchangesintheambienttemperatureandchangesinthegascomposition.Furthermore,measurementscanbeaffectedbythecontaminationofthesamplegasthroughcontactwiththesensorunit.
N
S
P O
M
mA
vol%O2
Sample gas
O2
Bridgeammeter
F0204.ai
Figure2.4 ExampleofConstructionofMagneticWindOxygenAnalyzer (withRing-shapedMeasurementChamber)
6
TI11P03A05-01E
2.TypesandApplicationsofOxygenAnalyzers
Oct.30,2012-00
2.1.3 MagneticForceMethod(DumbbellType)Figure2.5showsadumbbell-shapedobjectwithlowmagneticsusceptibilitythatissuspendedfromaplatinumorquartzwirewithinamagneticfield.Whenasamplegascontainingoxygenisguidedtothevicinityofthedumbbell,theoxygenisattractedtothepointofmaximummagneticfieldstrength,causingtheballsonthedumbbelltodeflectslightlyinanoppositedirection.Thetwistingofthesuspensionwireisdetected using a light source and a reflector that is attached to the center of the wire. The signal causes a current to flow through an excitation coil to correct the deflection. The current is proportional to the oxygen concentration.
Thismethodischaracterizedbyawidedynamicrangeandisnotinfluencedmuchbythepresenceofbackgroundgases.Thedownsideisthatithaslessresistancetovibrationandmechanicalshockandissusceptibletocontaminationandcorrosion.
Suspending wire
Dumbbell
Excitationcoil
Magnetic poleReflector
Photo sensors
Lightsource Vol%O2
F0205.ai
Figure2.5 MagneticForceMethod(dumbbelltype)
2.1.4 MagneticForceMethod(PressureSensorType)Whentwogasescomeintocontactinamagneticfield,adifferentialpressureisgeneratedthatisproportionaltothedifferenceinthemagneticsusceptibilitiesofthegases.Providingthattheoxygencontentoftheauxiliarygas(referencegas)remainsataconstantlevel,theoxygenconcentrationofthesamplegascanbedeterminedbymeansofthedifferentialpressure.Sincealternatingcurrentallowsforeasierdetectionandamplificationofsignals,amagneticfieldiscreatedusinganelectromagnetthatisintermittentlyexcited.Acondensermicrophonesensorormicroflowsensorthatisusedininfraredgasanalyzersisusedtodetectsmallvoltages.Figure2.6illustratesthismethod.
Althoughthismethodrequiresanauxiliarygas,itisaffectedrelativelylittlebybackgroundgasesandhasafastresponsetime.
Electromagnet Restrictor
Sample gas
Auxiliary gas
Small pressure sensor
F0206.ai
P
Figure2.6 MagneticForceMethod(pressuresensortype)
TI11P03A05-01E
72.TypesandApplicationsofOxygenAnalyzers
Oct.30,2012-00
2.2 ElectrochemicalOxygenAnalyzers2.2.1 ZirconiaSystem
(1) OxygenConcentrationCellAzirconiaceramicthathasbeenstabilized,forexample,bycalciumoxide(CaO)oryttriumoxide(Y2O3),isasolidelectrolytethatisconductiveonlytooxygenionswhenitisheatedtoahightemperature.
Figure2.7showsazirconiaelementwithaporousplatinumelectrodeattachedoneachside.Whentheelectrodesareexposedtoagasthatcontainsoxygen,thefollowingreactionsoccurbetweentheelectrodes,withthezirconiaelementservingasaseparator:
P1side(cathode): O2+4e-->2O2-
P2side(anode): 2O2--->O2+4e
ThismeansthatoxygenmoleculesgainelectronsandbecomeoxygenionsatelectrodeP1,givingitahigherpartialoxygenconcentration.TheseionstravelthroughthezirconiaelementtoelectrodeP2,creatinganoxygenconcentrationcell.Here,electronsarereleasedtoformoxygenmolecules,givingitalowerpartialoxygenconcentration.Thesereactionsbetweentheelectrodesgenerateanelectromotiveforce(E),whichisrepresentedbythefollowingNernstequation:
E=- RT ln P24F P1
R:gasconstant
T:absolutetemperature
F:Faradayconstant
WhentheP1sideisexposedtoareferencegas(e.g.air)andtheP2sideisexposedtoasamplegas,theoxygenconcentrationinthesamplegascanbedeterminedbymeasuringtheelectromotiveforcethatisgenerated.
Gas of high oxygen concentration
P1 Electrode (cathode)
Zirconia element
Electrode (anode)P2
Gas of low oxygen concentration
F0207.ai
Figure2.7 OxygenConcentrationCell
TheZR402/ZR202ZirconiaOxygenAnalyzers,theAV550GAveragingConverter,theZS8Explosion-proofOxygenAnalyzerandtheOX400LowConcentrationOxygenAnalyzerusethismethod.
8
TI11P03A05-01E
2.TypesandApplicationsofOxygenAnalyzers
Oct.30,2012-00
(2) LimitingCurrentSystemAsshowninthefigure2.8,iftheflowofoxygenintothecathodeofazirconiaelementheatedtohightemperatureislimited,thereappearsaregionwherethecurrentbecomesconstantevenwhentheappliedvoltageisincreased.Thislimitedcurrentisproportionaltotheoxygenconcentration.
F0208.ai
Gas diffusion hole
Diffusionchamber
Anode Current
Zirconia, or solid electrolyte
O2
O2 Cathode
Figure2.8 LimitingCurrentSystem
TheOX100/OX102OxygenAnalyzersusethismethod.
TI11P03A05-01E
92.TypesandApplicationsofOxygenAnalyzers
Oct.30,2012-00
2.2.2 ElectrodeSystemFigure2.9showstheinteriorofanelectrodeoxygenanalyzer.Ajelly-likeelectrolyteisappliedtothegoldcathode and silver anode, over which a thin Teflon membrane is stretched that is only permeable to oxygen. Whenavoltageof0.5Vto0.8Visappliedbetweentheelectrodes,apolarographiclimitingcurrentthatisproportionaltotheoxygenconcentrationcanbedetected.
Agalvaniccelloxygenanalyzerusesanelectrodesystemwiththiskindofseparatingmembrane.Withthismethod,anelectrode(cathode)madeofanoblemetalsuchasgoldorsilverandanelectrode(anode)madeofabasemetalsuchasleadarelocatedintheelectrolyte,whichisincontactwiththeairthroughtheseparatingmembrane.Thisseparatingmembraneisonlypermeabletooxygen.Thepermeatedoxygencausesanoxidationreactionatthecathodeandgeneratesapotentialwhenanadditionalresistorisconnectedtobothelectrodes.Theoxygenconcentrationcanbedeterminedbymonitoringthegeneratedcurrent.
Thedownsideofthismethodisthatthelifeofthesensorcellislimitedbybeingincontactwiththeairevenwhenmeasurementisnotbeingperformed,soperiodicreplacementisrequired.Anotherproblemisthatanegativedriftoccurswhentheanalyzerisoperatedcontinuously.Therefore,thisanalyzerisnotsuitableforcontinuousmeasurementsoveralongperiodoftime.However,itiscompactandcostsrelativelylittle,soitisusedasaportableanalyzer.
Silver electrode (+)
Electrolyte jelly
Gold electrode (-)ThermistorTeflon membraneTeflon membrane F0209.ai
Figure2.9 SeparatingMembraneElectrodeforMeasuringOxygenConcentration
10
TI11P03A05-01E
2.TypesandApplicationsofOxygenAnalyzers
Oct.30,2012-00
2.3 AbsorptionSpectroscopyOxygenAnalyzer2.3.1 TunableDiodeLaserSpectroscopy(TDLS)Method
OperationalPrincipleTDLSmeasurementsare basedonabsorptionspectroscopy.TDLSAnalyzeroperatesbymeasuringtheamountoflaserlightthatisabsorbed(lost)asittravelsthroughthegasbeingmeasured.InthesimplestformaTDLSanalyzerconsistsofalaserthatemitsinfraredlight,opticallensestofocusthelaserlightthroughthegastobemeasuredandthenontoadetector,thedetector,andelectronicsthatcontrolthelaserandtranslatethedetectorsignalintoasignalrepresentingthegasconcentration.Gasmoleculesabsorblightatspecificwavelengths,calledabsorptionlines.ThisabsorptionfollowstheBeer-LambertLaw.
I=I0•e-E・G・L
whereIistheradiationintensityafterabsorption
I0istheinitialradiationintensity
E is the extinction coefficient
Gisthegasconcentration
andListhepathlengthofthemeasurementarea
UsingaTunableDiodeLaserasalightsourceforspectroscopyhasthefollowingbenefits:• Sensitivity.ApplicationDependant.Sub-PPMinsomeapplications.
• Selectivity.Thenarrowlinewidthofthelaserisabletoresolvesingleabsorptionlines.Thisprovidesmorechoicesofaparticularpeaktouseformeasurement,usuallyallowingoneisolatedpeaktobeused.
• Power.Diodelasershavepowerrangingfrom0.5mWto20mW.Also,beinghighlycoherentthisallowsmeasurementinopticallythickenvironments(highparticulateloading).
• Monochromatic. No dispersive element (filter, etc.) required. Light source itself is selective.
• Tunable.Wavelengthcanbesweptacrosstheentireabsorptionfeature,thisallowsresonant(peak)andnonresonant(baseline)measurementduringeveryscan.Bymeasuringthebaselineandpeak,power at the detector can fluctuate rapidly by large amounts without affecting the measurement. This isusefulforhighparticulateapplications.
TheTDLS200andtheTDLS220(ExtractiveType)TDLSAnalyzersusethismethod.
Photo detector(Photo diode)Laser diode
Optical path
Laser beam
Process gas stream F0210.ai
Photo detector(Photo diode)Laser diode
Optical path
Laser beam
Process gas stream F0210.ai
TI11P03A05-01E
112.TypesandApplicationsofOxygenAnalyzers
Oct.30,2012-00
2.4 ApplicationsofOxygenAnalyzersbyMeasurementPrinciple
Table 2.1 shows which oxygen analyzer types are suitable for specific applications and industries.
Table2.1OxygenAnalyzerApplications,byAnalyzerType
Analyzertype Zirconia Magnetic Laser
Yokogawa’smodelZR202GZR22G
+ZR402G
ZR202SZR22S
+ZR402G
OX400OX100OX102
MG8G MG8E TDLS200
Application:Boiler/combustionfurnaceAllindustrytypes A BElectricpower A BOil/petrochemical A A BPulp/paper A B
Application:HeatingfurnaceOil/petrochemical A A BIronandsteel,nonferrousmetal A B BCeramic A B B
Application:ProcessqualityandsafetycontrolOil/petrochemical A APulp/paper C BOther(Ironandsteeletc.) B B A B
RemarksCombustion control, (C: humidity measurement)
Explosion-proof type
Built into device Quality control, combustion control
Explosion-proof type, safety and quality control
Combustion,safety control
A:MostsuitableB/C:Suitable
12
TI11P03A05-01E
2.TypesandApplicationsofOxygenAnalyzers
Oct.30,2012-00
2.5 OxygenAnalyzerSelectionFlowchart
*1
For semiconductor or electronic component manufacturing process
For combustioncontrol or
process control
With samplingDirect in-situ
Explosion-proofconstruction
required?
Explosion-proofconstruction
required?
YES or NO
Oxygen Analyzer
Flammable gas included?
NO
NO
YES
YES
OX400 or OX100 or OX102
MG8ETDLS200
Explosion-proof typeNote 2
Note 1
Paramagnetic Oxygen AnalyzersLaser Analyzers
MG8G
General-purpose type
Application
Sampling method
ZR22S+ZO21P+ZR402G
ZR22G+ZR402Gor ZR202G
ZR22G+ZO21P+ZR402G
ZR22G+AV550G
ZR22S+ZR402Gor ZR202S
Note 1: If a flammable gas makes up 0.5 % or less of the gas mixture, select “NO.”Note 2: All of these instruments must be ordered with the custom order process.Note 3: The ZR402G converter must not be installed in a hazardous area.Note 4: Max. of 8 points for measurement and averaging.Note 5: Optical path averagingNote 6: The ZO21P is a high temperature probe adapter.
Averaging measurement ?
NO
NONO
YES
Explosion-proofconstruction required?
YES
YES
YES
NO
Zirconia Oxygen Analyzers
Gas temperature 700°C or less?
Gas temperature 700°C or less?
Laser Analyzers
TDLS200
Note 3 Note 3, Note 6 Note 6
Note 2
Note 4Multi-points measurementand averaging Note 5
Line averaging
Averaging method
*1
Note 2
TI11P03A05-01E
133.MG8G/MG8EFeatures
3. MG8G/MG8EFeatures
uHighlyReliableandEasytoMaintain•Thesensorisdesignedsothatitdoesnotcomeintocontactwiththesamplegas(seesection4.3).Long-
termmeasurementstabilityisensuredevenwhencorrosiveorcontaminatedgasismeasured.
•Thelackofanymovingpartsensuresexcellentresistancetovibrationandshock.
•TheMG8Eisexplosion-proof(ExdIIBT4X).
uHighStabilityandFastResponse•Themainpartofthesensorunitincludingthemeasurementcellisinahousingwhereaconstant
temperatureismaintained.Thisminimizesanyeffectonthesamplegasbytheambienttemperatureandalsopreventsdraining,therebyassuringstablemeasurement.
•Highlysensitivethermistorswithfastresponsetimesdirectlydetecttheflowrateofauxiliarygases,ensuringa90%responsewithinthreeseconds.
•Excellentindicatingstabilityinthevicinityofzeromakesthesensorsuitableforlowconcentrationmeasurement(e.g.forsafetycontrol).
•Aninterference-gascompensationfunctionensuresaccuratemeasurementevenwhenparamagneticbackground gases (including flammable gases) are present.
uConverterwithaWealthofConvenientFunctions•Easycalibration
Allyouneedtodotocalibratetheunitisintroduceacalibrationgaswithazerooxygenconcentrationandpressthecalibrationbutton,thenrepeattheprocedurewiththegascontainingapre-determinedoxygen.Tosavetime,automaticcalibrationmodecanbeselected.Furthermore,anoutputfunctionthatdrivesasolenoidvalvetoswitchbetweenzeroandspanisprovidedasastandardfeature.
•Richself-diagnosticfunctions
Cell,measurementunittemperature,analogunit,digitalunit,andmemoryfailurescanbedetected.Ifafailureisdetected,theFAILlamplightsandamessageisdisplayed.Furthermore,iftheauxiliarygaspressurefallsbelowthesetpoint,contactoutputisactivated.
Oct.30,2012-00
14
TI11P03A05-01E
4.MG8G/MG8ESystemConfigurationandSensorConstruction
Oct.30,2012-00
4. MG8G/MG8ESystemConfigurationandSensorConstruction
4.1 SystemConfiguration
Figure 4.1 shows the basic system configuration of the MG8G/MG8E Paramagnetic Oxygen Analyzers.
Thepre-treatmentunitvariesdependingonthedeviceandapplication,soallpre-treatmentunitshavetobecustom-ordered.
F0401.ai
Analog output (4 to 20 mA DC)Contact output(abnormal)(maintenance)(range selection answerback)(Hi / Lo alarm)
Gas outlet
Filter
Flowmeter
MG8G/MG8E
Needle valve
Needle valve
Pressurereducing valve
Pressurereducingvalve
Pressuremeter
Pressure reducing valve
Auxiliary gas N2
Zero gas N2
Span gas O2+N2
Samplegas
Water feedWater drainSteam inlet
Steam outletPower supply
Pre-treatmentunit
or Instrument air
P
Figure4.1 MG8G/MG8EOxygenAnalyzerSystemConfiguration
TI11P03A05-01E
154.MG8G/MG8ESystemConfigurationandSensorConstruction
Oct.30,2012-00
4.2 SuitableApplicationsforMG8G/MG8EParamagneticOxygenAnalyzers
TheMG8GParamagneticOxygenAnalyzerisforgeneraluseandtheMG8EParamagneticOxygenAnalyzer is for explosion-proof applications. The MG8E has been certified by the TIIS* to meet the Ex d IIB T4Xstandard.
* Technology Institution of Industry Safety, Japan
ThefollowingtableshowstheapplicablecriteriaoftheMG8GandMG8E.
Table4.1ApplicabilityCriteriaofMG8ParamagneticOxygenAnalyzers(InstallationEnvironmentandMeasuredGas)
MG8OxygenAnalyzer
ApplicableRange
InstallationSite SampleGas
HazardousArea
Non-hazardous
Area*2
ClassAandBhazardousgases*1
orMixedgasesoflessthan4%hydrogen
Mixedgasesof4to100%hydrogen
ClassChazardousgas*1,
excludinghydrogen*3
Atmosphere SampleGas Atmosphere SampleGas Atmosphere SampleGas
MG8Eusedasexplosion-proof(Ex d IIB T4X*4)
0-1to25%O2(Notapplicablefor21-25%O2)
OK OK OK OK NA NA NA NA
MG8Eusedasnon-explosion-proof 0-1to25%O2 NA OK NA OK NA OK NA NA
MG8Gusedasnon-explosion-proof 0-5to25%O2 NA OK NA OK NA NA NA NA
NA:Notapplicable
*1: Refer to the Users Guide to Installing Explosion-proof Electrical Apparatus at Plants, issued by the Technology Institution of IndustrialSafety,Japan.
*2: The definition of the non-hazardous area is followed by the description in the Users Guide to Installing Explosion-proof ElectricalApparatusatPlants,issuedbytheTechnologyInstitutionofIndustrialSafety,Japan:Asanon-hazardousareaisconsidered a place where no occurrence of explosive gas atmospheres is guaranteed by the foreperson and confirmed by awrittendocument.
*3: Acetylene, carbon disulfide, hydrogen, and ethyl nitrate.*4: Ex d IIB T4X (a)Structure:Explosion-proof (b) Scope of area: Industrial sites and hazardous areas in office buildings. May not be used in hazardous locations in
mines. (c)Scopeofsamplegasorvapor: (c-1)ClassAandBhazardousgasesorvapor (c-2)Gasorvaporwithignitiontemperatureof135°Corgreater (c-3) Hydrogen concentration must be below 4%. Not applicable for gases containing acetylene, carbon disulfide
andethylnitrate. (d)Operatingconditions (d-1)Beforeopeningthecover,removepowerandmakesureofnon-hazardousatmospheres. (d-2)Donotuseformeasuringoxygenconcentrationofgasesotherthanthosecontainingairoroxygen
equivalent to or less than air, or those mixed with flammable gas or vapor.
16
TI11P03A05-01E
4.MG8G/MG8ESystemConfigurationandSensorConstruction
Oct.30,2012-00
4.3 SensorConstructionThe MG8G/MG8E Paramagnetic Oxygen Analyzers use a magnetic proportional flow rate sensor that was developedbyYokogawa.Thissensorhashighsensitivityandthesamefastresponsetimeofazirconiasensor,anditsconstructionprovideshighresistancetovibrationandshock.
Inaddition,thesensor’sthermistorsdonotcomeintodirectcontactwiththesamplegas,therebyensuringlong-termmeasurementstabilityevenwhencorrosiveorcontaminatedgasismeasured.
(1) ConstructionandOperatingPrincipleFigure4.2showstheconstructionofthering-shapedsensor.
• AsamplegasentersfromthesamplegasinletandisdividedintostreamsAandA’inthering-shapedgaspath.
• AnauxiliarygasentersfromtheauxiliarygasinletandisdividedintostreamsBandB’,whichthenpassthroughathermistorthatispositionedineachofthegaspaths.Thesethermistorsdetecttheflow rates of the auxiliary gasses.
• Theauxiliarygasstreamsthenenterthering-shapedpathatpointsDandE,wheretheymeetwiththe sample gasses and flow in direction C toward an outlet.
• AstrongmagneticfieldiscreatedatpointD,wherestreamBoftheauxiliarygasentersthering-shapedpath.Ifthesamplegascontainsoxygen,theoxygenisattractedtoandconcentratedinthemagnetic field at point D. The oxygen thus restricts the flow of auxiliary gas at point D, which alters the ratio of flow rate B to flow rate B’. This change in the ratio is proportional to the oxygen concentration, anditisdetectedwiththethermistorsandoutput.
Sample gas + auxiliary gas Auxiliary gas
Sample gas
Magnet
Magnet
Auxiliary gas inlet
Sensorthermistors
Sample gas inlet
Samplegas
Auxiliary gas
Magnet
Sample gas + auxiliary gas outlet
Sample gas + auxiliary gas
F0402.ai
B
B‘
CD
E
A
A‘
Figure4.2 MagneticProportionalFlowRateSensor
TI11P03A05-01E
174.MG8G/MG8ESystemConfigurationandSensorConstruction
Oct.30,2012-00
(2) SensorUnitGasPathTheMG8G/MG8EParamagneticOxygenAnalyzersachievestablemeasurementbyavarietyofmeans.(SeeFigure4.3.)
• Themainpartofthesensorunit,includingthecell,iscontainedwithinahousing.Insidethishousingaconstanttemperatureismaintainedtoensurethatthegastemperatureisnotaffectedbychangesintheambienttemperature.
• Inordertostabilizethetemperaturewithinthishousing,anintegralcomputationfunctionhasbeenadded.
• Measureshavealsobeentakentopreventfluctuatingindicationscausedbychangesinthesampleand auxiliary gas flow rates.
(a)BypassUnit(onlyavailablewiththeMG8E)The horizontal axis represents the flow rate of the sample gas entering the paramagnetic oxygen analyzer, andtheverticalaxisrepresentstheflowrateofthesamplegasenteringthecell.Ifthesamplegasentersthe paramagnetic oxygen analyzer at a slow rate, all the gas flows into the cell. If the rate exceeds 235 ml/min, the excess portion flows into the bypass unit.
(b)SampleGasPre-heatingUnit(onlyavailablewiththeMG8E)Thesensorunithousingiskeptataconstant55°C.Thetemperatureofthesamplegasenteringthehousingisusuallylessthan55°C.Anychangeinthesamplegastemperaturecanchangethemeasurementcelltemperatureandaffectitsoutput.
Beforethesamplegasentersthemeasurementcell,theMG8EParamagneticOxygenAnalyzerpassesitthroughapre-heatingunitinordertobringitclosetothetemperaturemaintainedinsidethesensorunithousing.
(c)StabilizingFlowRateofAuxiliaryGasTheauxiliarygasisnitrogen(N2).TheMG8G/MG8EParamagneticOxygenAnalyzersdeterminetheoxygen concentration using the ratio of the flow rates for this nitrogen gas; therefore, it is essential to keep the flow rate constant. This is done by means of capillary restriction resistance.
(d)PressureSwitch(onlyavailablewiththeMG8E)If the auxiliary gas flow is disrupted while the paramagnetic oxygen analyzer is operating, the output signal overshootsthescale.Topreventthisfromoccurring,apressureswitchontheprimarysideoftheauxiliarygaslineregulatoractivatesanalarmtonotifytheoperatorthatpreventiveactionshouldbetaken.
(e)TemperatureControlforSensorUnitHousing(onlyavailablewiththeMG8E)Toguardagainstdisturbancessuchasvoltagefluctuations,anintegralcomputationfunctionhasbeenaddedthatmakesitpossibletocontrolthetemperaturetowithin0.005°C.Furthermore,themeasurementcelltemperatureiscontrolledduringthewarm-upphasebycapturingthecelltemperature(summingofthesignalsfromtheleftandrightthermistors)andincreasingtheheatertemperatureifthecelltemperatureislow,thengraduallyloweringtheheatertemperatureasthecelltemperaturerises.
(f)CompensationforAtmosphericPressureError(onlyavailablewiththeMG8E)Equippedwithanatmosphericpressure-compensationsensorasstandard,atmosphericpressureerrorcanbecompensated.
18
TI11P03A05-01E
4.MG8G/MG8ESystemConfigurationandSensorConstruction
Oct.30,2012-00
(3) BlockDiagramofOxygenAnalyzerGasPathinSensorUnit
F0403.ai
Explosion-proofhousing wall
Sensor unithousing
Bypass unit
Capillary tube
Sample gas pre-heating unit
Regulator
Pressure switch
Pressure gauge
Flame arrestor
Flame arrestor0 to 200 kPa
Sample gas inlet Auxiliary gas (N2) inlet Gas outlet
Note: A pressure sensor for compensating for atmospheric pressure changes is installed in the electrical circuit.
Measurement Cell
P
100 to600ml/min
235 ml/min
300 to 800 ml/min 35 ml/min 335 to 835 ml/min
Figure4.3BlockDiagramofMG8EGasPathinSensorUnit
F0404.ai
Flow
rate
of g
as fl
owin
g in
to c
ell (
ml/m
in)
Setting range of sample gas flow rate
Flow rate of sample gas flowing into MG8E (ml/min)
240
220
200
180
160
140
120
100100 200 300 400 500 600 700 800 900 10000
Figure4.4FlowCharacteristicsofMG8EBypassUnit
TI11P03A05-01E
195.EffectsofInterferenceGas
Oct.30,2012-00
5. EffectsofInterferenceGas
5.1 MeasurementofOxygen(O2)inthePresenceofInterferenceGases
ThefollowingtableindicateswhentheMG8Eanalyzercanbeusedtomeasureoxygen(O2)inthepresenceofinterferencegases.
No. Typeofindustry InterferencegascomponentsMeasuringrange(%)
Remarks0–1 0–2 0–5
1Petrochemical
H2: 35%, CO2: 7%, N2: B A A A Note 12 C3: 0 – 5%, C4: 5 – 10%, N2: B — A A Note 1, off gas3 Oil H2: 10 – 40%, C1: 20 – 30%, C2: 15 – 30%, N2: B A A A Note 1, to flare stack4
Chemical
VCM (C2H3C1) + Air A A A5 C4H10 + Air — A A Note 26 CH4: 50%, C2H4: 25%, CO2: 4%, Ar: 15% A A A7 C4H10: 20%, O2: 6%, N2: B A A A8 H2 +O2 A A A Note 1, electrolytic plant9
Iron and steelH2: 75%, N2: 25% + Air — A A Note 1
10 (H2: 45%, CH4: 17%, CO: 16%, CO2: 4%) + Air — A A Note 111 Electric power (H2: 30%, CO2: 36%, CH4: 4%) + Air — A A Note 1, fuel cell
12 Environmental sewage CH4: 55%, CO2: 39% + Air A A A Digestion gas
13 Nuclear power He + O2 A A A Note 2A:Measurementispossiblewhentheeffectoftheinterferencegasiswithin±0.02%O2or±1%ofspan,whicheverishigher.Note 1: Auxiliary gas flow rate: 55 ml/minNote2:PleasecontactuswhenC4H10is50vol%ormore,orwhennuclearpower(He+O2)istobemeasured.
TestingMethod(1)Adjustthezerobalancewiththeinterferencegasandnitrogen(N2)gas.
(2)PerformcalibrationwiththeO2/N2gasandobtainthedeviationfromtheO2/interferencegas.
CAUTIONTheinformationintheabovetableappliesonlytoclassAandBhazardousgasesorvaporsandtogasmixtures that are less than 4% hydrogen gas, in which cases the analyzer must be used as specified in the explosion-proof specification.
Theanalyzercannotbeusedtomeasuretheoxygenconcentrationingasesorvaporscontainingacetylene,carbondisulfide,orethylenenitrate.Furthermore,theexplosion-proofcertificationdoesnotapplytoNos.1,3,8,9,10,and11intheabovetable.
20
TI11P03A05-01E
5.EffectsofInterferenceGas
Oct.30,2012-00
5.2 DataonEffectsofInterferenceGasesThefollowingdataonamixtureofinterferencecomponentsandO2isfromananalyzerthathasbeencalibratedwithN2andO2+N2.
Theoxygenconcentrationcanbemeasuredwithin±2%ofspan.
O2+CO2(92%)+N2(8%)
Range:0to1vol%O2
Auxiliary gas flow rate: 35 ml/min
2 3 4 5
11 20 43234 5
6 7 8 9 101
O2(0.89%)+CO2(92.0%)+N2(7.11%)
N2
O2(0.89%)+N2(99.11%)
O2(0.59%)+CO2
(92.0%)+N2(7.41%)
1%(5V)
O2(0.59%)+N2(99.41%)
0%(1V)
O2=0.59%
O2=0.89%
F0501.ai
3 min
Figure5.1DataExampleofEffectsofInterferenceGases
TI11P03A05-01E
215.EffectsofInterferenceGas
Oct.30,2012-00
5.3 InterferenceGasCompensationMG8G/MG8EParamagneticOxygenAnalyzersmakeuseoftheparamagneticpropertyofoxygen(magneticsusceptibility,i.e.theabilitytobecomemagnetizedinthepresenceofanexternallyappliedmagneticfield)tomeasuretheoxygenconcentration.Gasesotherthanoxygenmayalsoexhibitthischaracteristic,althoughtheywillnotbenearlyasmagneticallysusceptibleasoxygen.Accordingly,processgaseswithvaryinglevelsofmagneticsusceptibilitycaninterferewithandcauseerrorsinthemeasurementoftheoxygenconcentration.
Errorssuchasthesethatarecausedbytheparamagneticpropertyofprocessgasesmaybesignificantwithlowoxygenconcentrationsintherangeof0%to1%;however,theMG8G/MG8EParamagneticOxygenAnalyzersareabletocanceloutthisinterferenceerrorbasedonthedifferenceindensityofasamplegasandareferencegas.AsshowninFigure5.2,asamplegasisdividedintostreamsAandA’inthering-shapedgaspath.Anauxiliarygas(N2) enters at the center of the sensor unit and flows left and right instreamsBandB’.AmagneticfieldisappliedatpointDwheretheauxiliarygasentersthering-shapedpath.Eveniftheauxiliarygasdoesnotcontainoxygen,thepresentofparamagneticgasessuchasN2willaffecttheratiooftheauxiliarygasflowratesinstreamsBandB’,resultinginanerror.Thiserrorcanbecompensatedbychangingthecellangle(cellattitude).
Forexample,ifcarbondioxide(CO2),whichhasalowermagneticsusceptibilitythannitrogen(N2),passesthroughthemeasurementcell,theanalyzerwillreadanegativevalue.IfthecellistiltedasshowninFigure5.3,theflowrateoftheauxiliarygastowardstreamB’willincreaseduetothehigherdensityoftheCO2.Thiswillchangetheflowrate,therebycancelingoutthenegativedeviation.Thechangeintheauxiliarygasflowrateduetothemagneticsusceptibilityofthegasiscancelledoutbychangingthemeasurementcellangle,whichaltersthedensityandcausesachangeintheauxiliarygasflowrate.Inthisway,theinterferenceerrorcanbecompensatedfor.
Priortoshipmentfromthefactory,eachMG8G/MG8EParamagneticOxygenAnalyzerreceivesafinaltuninginwhichthemeasurementcellangle(attitude)isadjustedbasedonthemagneticcharacteristicsanddensityofthecustomer’ssamplegas.Thiscellangleisstoredintheanalyzer’smemory.
Wheninstallingtheanalyzer,rotatetheadjustmentknobuntilthebubbleinthebuilt-inbubblelevelisatthecenterposition.Thisbringsthemeasurementcelltothecorrectangle.
A‘
A
B
B‘
CD
E
Auxiliary gas inlet
Sensorthermistors
Sample gas inlet
Samplegas
Auxiliary gas
Sample gas + auxiliary gas outlet
Sample gas + auxiliary gas
Magnet
F0502.ai
Figure5.2DiagramofMeasurementCell
F0503.ai
Auxiliary gas inlet
Sample gas stream A
Magnet
Magnet
Sample gas inlet
Sensor thermistor
Auxiliary gas stream B
Sample gas stream A’
Auxiliary gas stream B’
Angleadjustment
Figure5.3ViewofMeasurementCellCross-section
22
TI11P03A05-01E
6.Characteristics(MG8ETestData)
Oct.30,2012-00
6. Characteristics(MG8ETestData)
6.1 ResponseCharacteristicResponsetime:90%responsewithin6seconds(fromthegasinletoftheoxygenanalyzer)
Responsecheckdata: Range:0%to1vol%O2
Zerogas:N2(99.999%),Spangas:N2+O2(0.978%)
F0601.ai
Paramagnetic Oxygen Analyzer
Paramagnetic oxygen analyzer
Exhaust gasSwitching cock
Zero gasSpan gas
Samplegasflowrate(ml/min)
Zero ➞ Span Span ➞ ZeroDead time(sec) 90% response(sec) Dead time(sec) 90% response(sec)
300 3.2 1.6 3.3 1.5600 2.9 1.5 2.6 1.5900 2.2 1.6 2.2 1.4
Note:Theanalyzerreadingwillreacha100%responsein10seconds(Yokogawa’spreviousmodel:30to50seconds)
6.2 FlowCharacteristicsSpecifications of flow characteristic testing
Range:0to1vol%O2
Auxiliary gas flow rate: 35 ml/min
Zerogas:N2,Spangas:O2(0.975%)+N2
<Testresult:±0.2%ofspan>
F0602.ai
(ml/min)
(ml/min)
700
610
350
510610
500
300
600
4.880
0.9968
1.0008
0.99710.9973
4.874
4.886
4.882
1% 0%(5 V) (1 V)
(0)
(0)(0)
Flow rate Output
Flow rate Output
(+0.20% of span)
(-0.20% of span)
(+0.10% of span)
(+0.10% of span)
<Span> <Zero>
1 min
(V DC)
(V DC)
TI11P03A05-01E
236.Characteristics(MG8ETestData)
Oct.30,2012-00
6.3 AttitudeErrorTestingthatshowshowchangesintheattitudeoftheoxygenanalyzeraffecttheindicatedvalue
Range:0to2vol%O2
Zerobalance:CO(65%)+H2(4.5%)+CO2andN2
Samplegas: Zerogas:N2,Spangas:O2(1.61%)+N2
Sample gas flow rate: 600 ml/min
Auxiliary gas flow rate: 35 ml/min
Attitudechange(1°increaseinangle)Front Back Left Right
Zero gas -0.02% of span -0.001% of span -0.137% of span -0.095% of spanSpan gas -0.04% of span -0.25% of span -0.216% of span
*In the case of 0 to 1% O2range: Vicinityofzero:0.2%/1°ofspan(Yokogawa’spreviousmodel:about10%/1°ofspan) Vicinityofspan:0.4%/1°ofspan
6.4 EffectsofAtmosphericPressureandLong-termDrift
1050
0 168 336 504
0 168 336 504
5
4
3
2
1
0
-1
-2
-3
-4
-5
5
4
3
2
1
0
-1
-2
-3
-4
-5
1000
950
900
850
800
1050
1000
950
900
850
800
Effects of Atmospheric Pressure
Specification: ±1% of span/10 hPa
Specification: ±2% of span
0.12% of span/10 hPa
Long-term Drift
Zero pointSpan point
Atm
osph
eric
pre
ssur
e (h
Pa)
Atm
osph
eric
pre
ssur
e (h
Pa)
Pressure
Zero point
Span point
Pressure
Span
Span
Zero
Zero
1 week
1 week
Time (hr)
Time (hr)
Error (%)
Error (%)
1st week: 1.2% of span2nd week: 0.5% of span3rd week: 0.3% of span
F0603.ai
24
TI11P03A05-01E
RevisionInformationTitle : MG8G/MG8EParamagneticOxygenAnalyzersManualnumber: TI11P03A05-01E
Sep.2009/1stEditionNewlypublished
Oct.2012/2ndEditionOverallrevision
Oct.30,2012-00