advance optima -- high performance measurement technology ... teledyne’s model 3000mb paramagnetic...

14
Advance Optima -- High Performance Measurement Technology Oxygen Analyzer Module Magnos 16/17 Part of the Advance Optima product family, with all advantages Magnos 16 Magnetomechanical measuring principle Short T 90 time Freely selectable measuring ranges Highly suppressed measuring ranges High accuracy and stability Monitoring of the measuring chamber flow rate Calibration without bottled test gas TÜV approval Magnos 17 Thermomagnetical measuring principle Robust measuring cell The Advance Optima features one housing, a single system controller module, one user interface and multiple analyzer modules. All these analyzer modules have the same mechanical and electrical interfaces and plug directly into the standard housing. Other than the analyzer modules, all components are standard throughout the system. Oxygen Analyzer Modules Magnos 16 and Magnos 17 The measuring principles for the Magnos 16 and 17 are based on the specific paramagnetic behaviour of oxygen. These analyzers are characterized by a high measurement accuracy and stability. In order to cover a broad scope of applications, ABB Analytical offers the Magnos with two different measuring techniques.

Upload: others

Post on 18-Apr-2020

4 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Advance Optima -- High Performance Measurement Technology ... Teledyne’s Model 3000MB Paramagnetic Percent Oxygen Analyzer is a versatile, microprocessor based instrument specifically

Advance Optima -- High Performance Measurement Technology

Oxygen Analyzer Module Magnos 16/17

■ Part of the Advance Optima product family, with all advantages

Magnos 16■ Magnetomechanical measuring principle■ Short T90 time■ Freely selectable measuring ranges■ Highly suppressed measuring ranges■ High accuracy and stability■ Monitoring of the measuring chamber flow rate■ Calibration without bottled test gas■ TÜV approval

Magnos 17■ Thermomagnetical measuring principle■ Robust measuring cell

The Advance Optima features one housing, a singlesystem controller module, one user interface and multipleanalyzer modules. All these analyzer modules have thesame mechanical and electrical interfaces and plug directlyinto the standard housing. Other than the analyzer modules,all components are standard throughout the system.

Oxygen Analyzer Modules Magnos 16 and Magnos 17

The measuring principles for the Magnos 16 and 17 arebased on the specific paramagnetic behaviour of oxygen.These analyzers are characterized by a high measurementaccuracy and stability. In order to cover a broad scope ofapplications, ABB Analytical offers the Magnos with twodifferent measuring techniques.

Page 2: Advance Optima -- High Performance Measurement Technology ... Teledyne’s Model 3000MB Paramagnetic Percent Oxygen Analyzer is a versatile, microprocessor based instrument specifically

• Advanced paramagnetic sensor with barometric pressure compensation and PID temperature control for enhanced stability

• Exceptional performance and price value

• Proven Teledyne microprocessor-based platform

• Designed for common air separation unit, medical air, and medical oxygen applications

Teledyne's Model 3010MA Paramagnetic Percent Oxygen Analyzer is a versatile, microprocessor-based instrument

specifically designed for oxygen purity analysis. The 3010MA is a precise, cost effective instrument capable of measuring percent oxygen in specific ranges as required by the air separation and medical gas qualification industries. Drawing on the solid performance of Teledyne's highly successful 3000 series control platform, the 3010MA provides the user with a wide variety of features while maintaining cost effectiveness.

PRINCIPLE OF MEASUREMENTThe physical property distinguishing oxygen from other gases is its paramagnetism that is significantly greater than that of other common gases. Consequently, the molecules of oxygen are attracted by strong magnetic fields -- a phenomena which can be used for detection purposes.

A paramagnetic sensor consists of two spheres arranged in the form of a dumbbell suspended in a symmetrical,

P a r a m a g n e t i c O x y g e n A n a l y z e r

TELEDYNEANALYTICAL INSTRUMENTS

non-uniform magnetic field. When the surrounding gas contains oxygen, the dumbbell spheres rotate out of the magnetic field by the relatively strong attraction of oxygen.

A light beam focused on a mirror attached to the dumbbell reflects asymmetrically onto two photo diodes resulting in a voltage shift. The voltage difference produces a current used to drive the dumbbells back to the original position. The current flow required to maintain the null position is directly proportional to the oxygen concentration and is shown linearly in volume percent oxygen in the display. Properly maintained paramagnetic sensors last for years with

little or no upkeep.

STABILITY BUILT-INParamagnetic sensors, operating under suppressed, high-purity conditions, are natively subject to swings in their output based on changes in the surrounding barometric pressure. To properly compensate for such atmospheric pressure changes, Teledyne has incorporated an absolute pressure sensor in our system design.

In addition, to ensure an enhanced, stable performance, Teledyne maintains the paramagnetic sensor in its own isolated, heated compartment. A PID controller is used to properly control and maintain the temperature within the sensor compartment to optimize the analysis.

The analysis section is mounted directly beside the 3010MA control unit allowing for a panel-saving, single 19" rack mount configuration.

FLEXIBILITYThe 3010MA offers three dynamic, user-configurable ranges plus auto-ranging. The instrument is linear on all three ranges, eliminating the need to recalibrate while switching between ranges. The analyzer comes standard with an isolated 4-20 mADC output for oxygen concentration and range identification (optional).

Additionally, a bi-directional RS-232C serial communication interface provides for remote monitoring and control of span and zero functions. Teledyne provides auto-calibration capabilities as a standard feature in the 3000 series platform.

Model 3010MP a r a m a g n e t i c O x y g e n A n a l y z e r

Page 3: Advance Optima -- High Performance Measurement Technology ... Teledyne’s Model 3000MB Paramagnetic Percent Oxygen Analyzer is a versatile, microprocessor based instrument specifically

SPECIFICATIONSRanges: 3 user selectable ranges of 95 - 100%, 96 - 100%

and 98 - 100%; auto-ranging with ID output

Accuracy: ± 0.045% O2 absolute at constant temperature (once thermal equilibrium has been achieved)

± 0.05% O2 over the entire operating temperature range (once thermal equilibrium has been achieved)

Response time: 90% of full scale in < 10 seconds at 77° F (25° C)

Linearity: 0.1% oxygen

Zero drift: ±0.5% oxygen per month

Span drift: ±0.5% oxygen per month

Repeatability: ±0.10% oxygen

Operating temp: 68 to 86° F (20 to 30° C)

Outputs: Two 0-1 VDC and two 4-20 mADC isolated (concentration and range ID)

Alarms: 1 x system-failure alarm contact to detect power failure;

2 x adjustable concentration threshold alarms with fully programmable set-points.

Analysis display: 5 digit red LED, 3/5" high numerals

Menu display: 20 character, 2 line vacuum fluorescent

Digital interface: Full duplex RS-232C communications port

Power requirements: 90 - 130 or 200 - 240 VAC at 47 / 63 Hz (user selected on rear of control unit)

Oxygen sensor: Long-life paramagnetic detector

Sample connections: User specified 1/4" or 6 mm fittings

Area classification: General purpose, indoor environment

Mounting: 19" rack mount

Dimensions: 8.7" H x 19" W x 12.23" L

Weight: 35 lbs (16 Kg)

MODEL 3010MA PARAMAGNETIC OXYGEN ANALYZER

APPLICATIONS• Air separation Bulk gas plants

• Medical air / medical oxygen qualifications

ADVANTAGES• Linearity of analysis across three user-selectable ranges

• Temperature controlled / isolated sensor for long term stability

• Atmospheric Pressure Compensation for high-end, accurate performance

• Auto-ranging to follow process upsets

• Auto-calibration electronics standard

• Exceptional value at a favorable price

STANDARD FEATURES• Proven, robust paramagnetic sensor

• Three user-selectable ranges plus cal range

• Signal output: isolated 4-20mADC

• Programmable auto-ranging

• Two fully adjustable concentration alarm set points with programmable relay functions, Form C contacts, 3A resistive

• Calibration contact span / zero, Form A normally open contacts, 3A resistive

• Self diagnostics with Form C failure alarm contacts

• Full duplex RS-232C communication link

• Five digit oxygen concentration LED display

• 2 x 20 alphanumeric vacuum fluorescent display for set up and diagnostics

• Switch selectable power supply: 90 - 130 or 200 - 240 VAC at 47 / 63 Hz

• Remote calibration digital inputs

© 2005 Teledyne Analytical Instruments, A Teledyne Technologies Company.All rights reserved. Printed in the USA. 11/05LD

WarrantyInstrument is warranted for 1 year against defects in materialor workmanshipNOTE: Specifications and features will vary with application. The above are established and validated during design, but are not to be construed as test criteria for every product. All specifications and features are subject to change without notice.

TELEDYNE ANALYTICAL INSTRUMENTSA Teledyne Technologies Company16830 Chestnut StreetCity of Industry, California 91748, USA

TEL: 626-934-1500 or 888-789-8168FAX: 626-934-1651 EMAIL: [email protected]

www.teledyne-ai.com

Page 4: Advance Optima -- High Performance Measurement Technology ... Teledyne’s Model 3000MB Paramagnetic Percent Oxygen Analyzer is a versatile, microprocessor based instrument specifically
Page 5: Advance Optima -- High Performance Measurement Technology ... Teledyne’s Model 3000MB Paramagnetic Percent Oxygen Analyzer is a versatile, microprocessor based instrument specifically
Page 6: Advance Optima -- High Performance Measurement Technology ... Teledyne’s Model 3000MB Paramagnetic Percent Oxygen Analyzer is a versatile, microprocessor based instrument specifically
Page 7: Advance Optima -- High Performance Measurement Technology ... Teledyne’s Model 3000MB Paramagnetic Percent Oxygen Analyzer is a versatile, microprocessor based instrument specifically
Page 8: Advance Optima -- High Performance Measurement Technology ... Teledyne’s Model 3000MB Paramagnetic Percent Oxygen Analyzer is a versatile, microprocessor based instrument specifically

• Microprocessor based• Exceptional performance, price value• Advanced paramagnetic sensor• Designed for the air separation industry

Teledyne’s Model 3000MB Paramagnetic Percent Oxygen Analyzer is a versatile, microprocessor based instrument specifically designed to detect oxygen in a variety of gases. The 3000MB is a precise, cost effective instrument capable of measuring percent oxygen in as low as the 0-1% range. Drawing on the solid performance of Teledyne’s highly successful 3000 series control platform, the 3000MB provides the user with a wide variety of features and options to satisfy virtually any application.

PRINCIPLE OF MEASUREMENTThe physical property distinguishing oxygen from other gases is its paramagnetism which is significantly greater than that of common gases. Consequently the molecules of oxygen are attracted by strong magnetic fields which can be used in detection. A paramagnetic sensor consists of two spheres arranged in the form of a dumbbell suspended in a symmetrical, non-uniform magnetic field. When the surrounding gas contains oxygen, the dumbbell spheres rotate out of the magnetic field by the relatively strong attraction of oxygen.A light beam focused on a mirror attached to the dumbbell reflects asymmetrically onto two photo diodes resulting in a voltage shift. The voltage difference produces a current

Teledyne Analytical Instrumentsused to drive the dumbbells back to the original position. The current flow required to maintain the null position is directly proportional to the oxygen concentration and is shown linearly in volume percent oxygen in the display. Properly maintained paramagnetic sensors last for years with little or no attention making them ideal for critical measurements in the low percent range.

FLEXIBILITYThe 3000MB offers three dynamic, user-configurable ranges plus autoranging, and can be calibrated in any range. The instrument is linear on all three ranges, eliminating the need to recalibrate while switching between ranges. The analyzer comes standard with an isolated 4-20 mADC output for oxygen concentration and range identification. Additionally, a bi-directional RS-232C serial communication interface provides for remote monitoring and control of span and zero functions. Teledyne provides auto-calibration capabilities as a standard feature in the 3000 series platform.

APPLICATIONS• Monitoring inert gas blanketing• Air separation and liquefaction• Chemical reaction monitoring• Emissions monitoring• Petrochemical process control• Quality assurance• Gas analysis certification

THE LEADER IN OXYGENThe 3000MB can be ordered as a standard unit or as part of a larger analytical system. With over 50 years of analytical instrument experience, Teledyne supplies sensors, custom engineered analyzers and complete monitoring systems to satisfy unique applications. From portables to on-line units, in microprocessor based or analog configurations, no other manufacturer comes close to matching Teledyne’s breadth of instrumentation.

ADVANTAGES• Linearity of analysis across three user-selectable ranges• Autoranging to follow process upsets• Auto-calibration electronics standard

3000

MB

Paramagnetic Oxygen Analyzer

3000

MB

Paramagnetic Oxygen Analyzer

Page 9: Advance Optima -- High Performance Measurement Technology ... Teledyne’s Model 3000MB Paramagnetic Percent Oxygen Analyzer is a versatile, microprocessor based instrument specifically

Operating temp: 32 to 113°F (0 to 45°C)Outputs: Two 0-1 VDC & two 4-20 mADC isolated (concentration and range ID)Alarms: One system-failure alarm contact to detect power failure; two adjustable concentration threshold alarms with fully programmable setpointsAnalysis display: 5 digit red LED, 3/5” high numeralsMenu display: 20 character, 2 line vacuum fluorescentDigital interface: Full duplex RS-232C communications portPower requirements: Universal AC input ranges, 100 to 240 VAC, 50 / 60 Hz, 70 watts maxOxygen sensor: ParamagneticSample connections: User specified 1/4” or 6mm fittingsArea classification: General purpose, non-hazardous Optional: FM approved for Class 1, Div 2 (no purge ) Y purged for Class 1 Div 1 areas Dimensions: 11.81” W x 20.28” H x 8.9” D

Model 3000MB Sample Gas Conditioning RequirementsSample inlet pressure: 5 to 20 psig (regulate to fixed inlet pressure)Sample flow: 1 to 2 SCFHSample dewpoint: Non-condensing at ambient temperatureSample temperature: 0 to 50°CDetector wetted parts: Glass, steel, gold, viton, acrylic glass, epoxy resin

For complete turnkey systems, please contact Teledyne or your local representative

MODEL 3000MB PARAMAGNETIC OxYGEN ANALYzER

STANDARD FEATURES• Proven, robust paramagnetic sensor• Three user-selectable ranges plus cal range• Signal and Range ID output: isolated 4-20 mADC• Programmable autoranging• Two fully adjustable concentration alarm set points with programmable relay functions, Form C contacts, 3A resistive• Calibration contact span/zero, Form A normally open contact, 3A resistive• Self diagnostics with Form C failure alarm contacts• Full duplex RS-232C communication link• Five digit oxygen concentration LED display• 2x20 alphanumeric vacuum fluorescent display for set up and diagnostics• Brass sample passages and fittings• Universal power supply: 100-240 VAC / 50-60 Hz• Remote calibration digital inputs

OPTIONS-C Electronically operated cal/zero valves

SPECIFICATIONSRanges: 3 user selectable ranges; typical ranges 0-5% to 0-100%; autoranging with ID output (lower ranges available)Calibration range: 0-25%Accuracy: ± 1% of full scale at constant temperature ± 5% of full scale over operating temperature range on factory default analysis ranges, once thermal equilibrium has been achievedResponse time: 90% of full scale in < 10 seconds at 77°F (25°C) Linearity: 0.1% O2zero drift: 0.05% O2 / 1 monthSpan drift: 1% FS / 1 monthRepeatability: 0.1% / FS

© 2005 Teledyne Analytical Instruments, A Teledyne Technologies Company.All rights reserved. Printed in the USA. 10/05LD

WarrantyInstrument is warranted for 1 year against defects in materialor workmanshipNOTE: Specifications and features will vary with application. The above are established and validated during design, but are not to be construed as test criteria for every product. All specifications and features are subject to change without notice.

TELEDYNE ANALYTICAL INSTRUMENTSA Teledyne Technologies Company16830 Chestnut StreetCity of Industry, California 91748, USA

TEL: 626-934-1500 or 888-789-8168FAx: 626-934-1651 EMAIL: [email protected]

www.teledyne-ai.com

Page 10: Advance Optima -- High Performance Measurement Technology ... Teledyne’s Model 3000MB Paramagnetic Percent Oxygen Analyzer is a versatile, microprocessor based instrument specifically

• Microprocessor based• Exceptional performance, price value• Advanced paramagnetic sensor• Designed for air separation industry

Teledyne’s Model 3000MA Paramagnetic Percent Oxygen Analyzer is a versatile, microprocessor based instrument specifically designed to detect oxygen in a variety of gases. The 3000MA is a precise, cost effective instrument capable of measuring percent oxygen in as low as the 0-1% range. Drawing on the solid performance of Teledyne’s highly successful 3000 series control platform, the 3000MA provides the user with a wide variety of features and options to satisfy virtually any application.

PRINCIPLE OF MEASUREMENTThe physical property distinguishing oxygen from other gases is its paramagnetism which is significantly greater than that of common gases. Consequently the molecules of oxygen are attracted by strong magnetic fields which can be used in detection. A paramagnetic sensor consists of two spheres arranged in the form of a dumbbell suspended in a symmetrical, non-uniform magnetic field. When the surrounding gas contains oxygen, the dumbbell spheres rotate out of the magnetic field by the relatively strong attraction of oxygen.A light beam focused on a mirror attached to the dumbbell reflects asymmetrically onto two photo

Teledyne Analytical Instruments30

00M

A30

00M

A

Paramagnetic Oxygen AnalyzerParamagnetic Oxygen Analyzer

diodes resulting in a voltage shift. The voltage difference produces a current used to drive the dumbbells back to the original position. The current flow required to maintain the null position is directly proportional to the oxygen concentration and is shown linearly in volume percent oxygen in the display. Properly maintained paramagnetic sensors last for years with little or no attention making them ideal for critical measurements in the low percent range.

FLEXIBILITYThe 3000MA offers three dynamic, user configurable ranges plus autoranging, and can be calibrated in any range. The instrument is linear on all three ranges, eliminating the need to recalibrate while switching between ranges. The analyzer comes standard with an isolated 4-20 mADC output for oxygen concentration and range identification. Additionally, a bi-directional RS-232C serial communication interface provides for remote monitoring and control of span and zero functions. Teledyne provides auto-calibration capabilities as a standard feature in the 3000 series platform.

APPLICATIONS• Monitoring inert gas blanketing• Air separation and liquefaction• Chemical reaction monitoring• Emissions monitoring• Petrochemical process control• Quality assurance• Gas analysis certification

THE LEADER IN OXYGENThe 3000MA can be ordered as a standard unit or as part of a larger analytical system. With over 50 years of analytical instrument experience, Teledyne supplies sensors, custom engineered analyzers and complete monitoring systems to satisfy unique applications. From portables to on-line units, in microprocessor based or analog configurations, no other manufacturer comes close to matching Teledyne’s breadth of instrumentation.

ADVANTAGES • Linearity of analysis across three user-selectable ranges • Autoranging to follow process upsets • Auto-calibration electronics standard

Page 11: Advance Optima -- High Performance Measurement Technology ... Teledyne’s Model 3000MB Paramagnetic Percent Oxygen Analyzer is a versatile, microprocessor based instrument specifically

Operating temp: 32 to 113°F (0 to 45°C)Outputs: Two 0-1 VDC & two 4-20 mADC isolated (concentration and range ID)Alarms: One system-failure alarm contact to detect power failure; two adjustable concentration threshold alarms with fully programmable setpoints.Analysis display: 5 digit red LED, 3/5” high numeralsMenu display: 20 character, 2 line vacuum fluorescentDigital interface: Full duplex RS-232C communications portPower requirements: Universal AC input ranges, 100 to 240 VAC, 50 / 60 Hz, 70 watts maxOxygen sensor: ParamagneticSample connections: User specified 1/4” or 6mm fittingsArea classification: General purpose flush mountedDimensions: 6.96” H x 8.7” W x 12.2” L (control unit)

Sample Gas Conditioning RequirementsSample inlet pressure: 5 to 20 psig (regulate to fixed inlet pressure)Sample flow: 1 to 2 SCFHSample dewpoint: Non-condensing at ambient temperatureSample temperature: 0 to 50°CDetector wetted parts: Glass, steel, gold, viton, acrylic glass, epoxy resin

For complete turnkey systems, please contact Teledyne or your local representative

MODEL 3000MA PARAMAGNETIC OxyGEN ANALyzER

© 2005 Teledyne Analytical Instruments, A Teledyne Technologies Company.All rights reserved. Printed in the USA. 08/05LD

STANDARD FEATURES• Proven, robust paramagnetic sensor• Three user-selectable ranges plus cal range• Signal and Range ID output: isolated 4-20 mADC• Programmable autoranging• Two fully adjustable concentration alarm set points with programmable relay functions, Form C contacts, 3A resistive• Calibration contact span/zero, Form A normally open contact, 3A resistive• Self diagnostics with Form C failure alarm contacts• Full duplex RS-232C communication link• Five digit oxygen concentration LED display• 2x20 alphanumeric vacuum fluorescent display for set up and diagnostics• Brass sample passages and fittings• Universal power supply: 100-240 VAC / 50-60 Hz• Remote calibration digital inputs

OPTIONS- C Electronically operated cal/zero valves

SPECIFICATIONSRanges: 3 user selectable ranges; typical ranges 0-5% to 0-100%; autoranging with ID output (lower ranges available)Calibration range: 0-25%Accuracy: ±1% of full scale at constant temperature; ±5% of full scale over operating temperature range on factory default analysis ranges, once thermal equilibrium has been achievedResponse time: 90% of full scale in < 10 seconds at 77°F (25°C) Linearity: 1% O2zero drift: 1% O2 / 1 monthSpan drift: 1% FS / 1 monthRepeatability: 0.1% / FS

WarrantyInstrument is warranted for 1 year against defects in material or workmanshipNOTE: Specifications and features will vary with application. The above are established and validated during design, but are not to be construed as test criteria for every product. All specifications and features are subject to change without notice.

TELEDYNE ANALYTICAL INSTRUMENTSA Teledyne Technologies Company16830 Chestnut StreetCity of Industry, California 91748, USA

TEL: 626-934-1500 or 888-789-8168FAx: 626-934-1651 EMAIL: [email protected]

www.teledyne-ai.com

Page 12: Advance Optima -- High Performance Measurement Technology ... Teledyne’s Model 3000MB Paramagnetic Percent Oxygen Analyzer is a versatile, microprocessor based instrument specifically

Explosion Proof ParamagneticOxygen Analyzer

MODEL 3020MExplosion Proof Paramagnetic

Oxygen Analyzer

TELEDYNEANALYTICAL INSTRUMENTS

Built for Reliability and Performance

• Fully explosion proof enclosure

• Microprocessor based

• Exceptional performance, price value

• Advanced paramagnetic sensor

• Designed for petrochemical industry

Teledyne’s Model 3020M Paramagnetic Percent Oxygen Analyzer is a versatile, microprocessor based instrument specifically designed to detect oxygen in a variety of gases. The 3020M is a precise, cost effective instrument capable of measuring percent oxygen in as low as the 0-1% range. Drawing on the solid performance of Teledyne’s highly successful 3000 series control platform, the 3020M provides the user with a wide variety of features and options to satisfy virtually any application.

PRINCIPLE OF MEASUREMENTThe physical property distinguishing oxygen from other gases is its paramagnetism which is significantly greater than that of common gases. Consequently the molecules of oxygen are attracted by strong magnetic fields which can be used in detection.

A paramagnetic sensor consists of two spheres arranged in the form of a dumbbell suspended in a symmetrical, non-uniform magnetic field. When the surrounding gas contains oxygen, the dumbbell spheres rotate out of the magnetic field by the relatively strong attraction of oxygen.

A light beam focused on a mirror attached to the dumbbell reflects asymmetrically onto two photo diodes resulting in a voltage shift. The voltage difference produces a current used to drive the dumbbells back to the original

position. The current flow required to maintain the null position is directly proportional to the oxygen concentration and is shown linearly in volume percent oxygen in the display. Properly maintained paramagnetic sensors last for years with little or no attention making them ideal for critical measurements in the low percent range.

FLEXIBILITYThe 3020M offers three dynamic, user-configurable ranges plus autoranging, and can be calibrated in any range. The instrument is linear on all three ranges, eliminating the need to recalibrate while switching between ranges. The analyzer comes standard with an isolated 4-20 mADC output for oxygen concentration and range identification. Additionally, a bi-directional RS-232C serial communication interface provides for remote monitoring and control of span and zero functions. Teledyne provides auto-calibration capabilities as a standard feature in the 3000 series platform.

APPLICATIONSThis instrument is designed specifically for the detection of percent oxygen in petrochemical applications for process control, quality assurance, and safety related purposes. Typical petrochemical plants requiring the 3020M are:

• Ethyl Benzene / Styrene Monomer (EB / SM)

• Pure Terephthalic Acid (PTA)

• Caprolactum

• Phenol / Cumene

• Polycarbonate

• Polyethylene

• Polystyrene

• Ethylene Oxide / Ethylene Glycol (EO / EG)

• Methyl Methacrylate (MMA)

• Ammonia / Urea Fertilizer

• Refinery FCC Unit and flare gas applications

• Marine Vapor Recovery (MVR)

THE LEADER IN OXYGENThe 3020M can be ordered as a standard unit or as part of a larger analytical system. With over 50 years of analytical instrument experience, Teledyne supplies sensors, custom engineered analyzers and complete monitoring systems to satisfy unique applications. From portables to on-line units, in microprocessor based or analog configurations, no other manufacturer comes close to matching Teledyne’s breadth of instrumentation.

ADVANTAGES• Linearity of analysis across three user-selectable ranges

• Autoranging to follow process upsets

• Auto-calibration electronics standard

• No purge gas requirements

Page 13: Advance Optima -- High Performance Measurement Technology ... Teledyne’s Model 3000MB Paramagnetic Percent Oxygen Analyzer is a versatile, microprocessor based instrument specifically

MODEL 3020M PARAMAGNETIC OXYGEN ANALYZER

© 2005 Teledyne Analytical Instruments, A Teledyne Technologies Company.All rights reserved. Printed in the USA. 11/05LD

WarrantyInstrument is warranted for 1 year against defects in materialor workmanshipNOTE: Specifications and features will vary with application. The above are established and validated during design, but are not to be construed as test criteria for every product. All specifications and features are subject to change without notice.

TELEDYNE ANALYTICAL INSTRUMENTSA Teledyne Technologies Company16830 Chestnut StreetCity of Industry, California 91748, USA

TEL: 626-934-1500 or 888-789-8168FAX: 626-934-1651 EMAIL: [email protected]

www.teledyne-ai.com

Zero drift: 0.05% O2 / 1 Month

Span drift: 1% FS / 1 Month

Repeatability: 0.1% / FS

Operating temp: 32 to 113°F (0 to 45°C)

Outputs: Two 0-1 VDC & two 4-20 mADC isolated (concentration and range ID)

Alarms: One system-failure alarm contact to detect power failure; two adjustable concentration threshold alarms with fully programmable setpoints.

Analysis display: 5 digit red LED, 3/5" high numerals

Menu display: 20 character, 2 line vacuum fluorescent

Digital interface: Full duplex RS-232C communications port

Power requirements: Universal AC input ranges, 100 to 240 VAC, 50 / 60 Hz, 70 watts max

Oxygen sensor: Paramagnetic

Sample connections: User specified 1/4" or 6mm fittings

Enclosure: Explosion proof U/L and CSA listed housing for Class I, Div 1, Group B, C & D service; NEMA 4/7 rated with baked powder epoxy finish for corrosion resistance

Mounting: Bulkhead mount

Dimensions: 15.25" W x 25.63" H x 11" D (387.35 mm x 651 mm x 279.4 mm)

Sample GaS ConditioninG RequiRementSSample inlet pressure: 5 to 20 psig (regulate to fixed inlet pressure)

Sample flow: 1 to 2 SCFH

Sample dewpoint: Non-condensing at ambient temperature

Sample temperature: 0 to 50°C

Detector wetted parts: Glass, steel, gold, viton, acrylic glass, epoxy resin

For complete turnkey systems, please contact Teledyne or your local representative

StandaRd FeatuReS• Proven, robust paramagnetic sensor

• Explosion proof NEMA 4/7 rated enclosure (Class I, Div 1, Group B)

• Three user-selectable ranges plus cal range

• Signal and Range ID output: isolated 4-20 mADC

• Programmable autoranging

• Two fully adjustable concentration alarm set points with programmable relay functions, Form C contacts, 3A resistive

• Calibration contact span/zero, Form A normally open contact, 3A resistive

• Self diagnostics with Form C failure alarm contacts

• Full duplex RS-232C communication link

• Five digit oxygen concentration LED display

• 2x20 alphanumeric vacuum fluorescent display for set up and diagnostics

• Stainless steel sample passages and fittings

• Universal power supply: 100-240 VAC / 50-60 Hz

• Remote calibration digital inputs

optionS-C Electronically operated cal/zero valves

-F Flame arrestors for Class I, Div 1, Group C/D service

-G Flame arrestors for Class I, Div 1, Group C/D service with cal valves

-H Flame arrestors for Group B (hydrogen) service

-I Flame arrestors for Group B service with cal valves

SpeCiFiCationSRanges: 3 user selectable ranges; typical ranges

0-5% to 0-100%; autoranging with ID output (lower ranges available)

Calibration range: 0-25%

Accuracy: ±1% of full scale at constant temperature; ±5% of full scale over operating temperature range on factory default analysis ranges, once thermal equilibrium has been achieved.

Response time: 90% of full scale in < 10 seconds at 77°F (25°C)

Linearity: 0.1% O2

Page 14: Advance Optima -- High Performance Measurement Technology ... Teledyne’s Model 3000MB Paramagnetic Percent Oxygen Analyzer is a versatile, microprocessor based instrument specifically

Technical Data

Measuring Principle:Paramagnetic behaviour of oxygen

Magnos 16

Sample Component:Oxygen (O2)

Measuring Ranges:❚ Smallest measuring range: 0...1 Vol.-% O2

❚ Largest measuring range: 0...100 Vol.-% O2

❚ Measuring range switching ratio: ≤ 1:100

❚ Max. measuring range suppression 1:100, e. g. 99...100 Vol.-% O2

Calibration:❚ Zero- and end-point calibration with nitrogen (N2)

and air or sample gas mixture

❚ Single-point calibration with room air

❚ Automatic calibration via built-in pneumatic module or external valves

Magnos 17

Sample Component:Oxygen (O2) in flue gases or in nitrogen (N2)

Measuring Ranges:❚ Smallest measuring range: 0...3 Vol.-% O2

❚ Largest measuring range: 0...100 Vol.-% O2

❚ Measuring range switching ratio: ≤ 1:33

Calibration:❚ Zero-point calibration with oxygen-free process gas

or substitute gas

❚ End-point calibration with process gas having a known oxygen concentration or with substitute gas

❚ Automatic calibration via built-in pneumatic module or external valves

Magnos 16 – based on the magnetomechanical measuringprinciple the analyzer is ideally suited for oxygen purity meas-urement, process gas measurements and inert gas blanketing.In emission measurements the Magnos 16 fulfills the specialrequirements, e.g. TA-Luft (German Federal Air Purity)Regulations, German 13. and 17. BImSchV Regulations. Theability to freely select measuring ranges and measure highlysuppressed ranges means that the analyzer can be easilyadapted to specific tasks. Even measurements for safety areno problem – monitoring the sample flow rate through themeasuring chamber always ensures that the current oxygenconcentration is being measured. Mounting the analyzer in aflame-proof housing (EEx d) allows use in hazardous areas,zone 1. The standard unit is certified for Class I, Division 2hazardous areas. Another advantage of the analyzer moduleis its short T90 response time. Calibration of the zero point isonly required once a month using air or nitrogen.

Magnos 17 – based on the thermomagnetical measuringprinciple the analyzer is used mainly in the analysis of fluegas including mixed fuel boilers, incinerators, cement fluegas and roasting gas. The robust measuring cell means theMagnos 17 is especially insensitive to vibrations and shocks.

The range of oxygen analyzer modules is complimentedby an electrochemical cell.

ABB AutomationAnalytical DivisionStierstädter Straße 5, D-60488 Frankfurt a. M., http://www.abb.com/analytical 50/24-04 EN 10.99