Oxygen Analyzer Module Magnos 16/17
■ Part of the Advance Optima product family, with all advantages
Magnos 16■ Magnetomechanical measuring principle■ Short T90 time■ Freely selectable measuring ranges■ Highly suppressed measuring ranges■ High accuracy and stability■ Monitoring of the measuring chamber flow rate■ Calibration without bottled test gas■ TÜV approval
Magnos 17■ Thermomagnetical measuring principle■ Robust measuring cell
The Advance Optima features one housing, a singlesystem controller module, one user interface and multipleanalyzer modules. All these analyzer modules have thesame mechanical and electrical interfaces and plug directlyinto the standard housing. Other than the analyzer modules,all components are standard throughout the system.
Oxygen Analyzer Modules Magnos 16 and Magnos 17
The measuring principles for the Magnos 16 and 17 arebased on the specific paramagnetic behaviour of oxygen.These analyzers are characterized by a high measurementaccuracy and stability. In order to cover a broad scope ofapplications, ABB Analytical offers the Magnos with twodifferent measuring techniques.
• Advanced paramagnetic sensor with barometric pressure compensation and PID temperature control for enhanced stability
• Exceptional performance and price value
• Proven Teledyne microprocessor-based platform
• Designed for common air separation unit, medical air, and medical oxygen applications
Teledyne's Model 3010MA Paramagnetic Percent Oxygen Analyzer is a versatile, microprocessor-based instrument
specifically designed for oxygen purity analysis. The 3010MA is a precise, cost effective instrument capable of measuring percent oxygen in specific ranges as required by the air separation and medical gas qualification industries. Drawing on the solid performance of Teledyne's highly successful 3000 series control platform, the 3010MA provides the user with a wide variety of features while maintaining cost effectiveness.
PRINCIPLE OF MEASUREMENTThe physical property distinguishing oxygen from other gases is its paramagnetism that is significantly greater than that of other common gases. Consequently, the molecules of oxygen are attracted by strong magnetic fields -- a phenomena which can be used for detection purposes.
A paramagnetic sensor consists of two spheres arranged in the form of a dumbbell suspended in a symmetrical,
P a r a m a g n e t i c O x y g e n A n a l y z e r
TELEDYNEANALYTICAL INSTRUMENTS
non-uniform magnetic field. When the surrounding gas contains oxygen, the dumbbell spheres rotate out of the magnetic field by the relatively strong attraction of oxygen.
A light beam focused on a mirror attached to the dumbbell reflects asymmetrically onto two photo diodes resulting in a voltage shift. The voltage difference produces a current used to drive the dumbbells back to the original position. The current flow required to maintain the null position is directly proportional to the oxygen concentration and is shown linearly in volume percent oxygen in the display. Properly maintained paramagnetic sensors last for years with
little or no upkeep.
STABILITY BUILT-INParamagnetic sensors, operating under suppressed, high-purity conditions, are natively subject to swings in their output based on changes in the surrounding barometric pressure. To properly compensate for such atmospheric pressure changes, Teledyne has incorporated an absolute pressure sensor in our system design.
In addition, to ensure an enhanced, stable performance, Teledyne maintains the paramagnetic sensor in its own isolated, heated compartment. A PID controller is used to properly control and maintain the temperature within the sensor compartment to optimize the analysis.
The analysis section is mounted directly beside the 3010MA control unit allowing for a panel-saving, single 19" rack mount configuration.
FLEXIBILITYThe 3010MA offers three dynamic, user-configurable ranges plus auto-ranging. The instrument is linear on all three ranges, eliminating the need to recalibrate while switching between ranges. The analyzer comes standard with an isolated 4-20 mADC output for oxygen concentration and range identification (optional).
Additionally, a bi-directional RS-232C serial communication interface provides for remote monitoring and control of span and zero functions. Teledyne provides auto-calibration capabilities as a standard feature in the 3000 series platform.
Model 3010MP a r a m a g n e t i c O x y g e n A n a l y z e r
SPECIFICATIONSRanges: 3 user selectable ranges of 95 - 100%, 96 - 100%
and 98 - 100%; auto-ranging with ID output
Accuracy: ± 0.045% O2 absolute at constant temperature (once thermal equilibrium has been achieved)
± 0.05% O2 over the entire operating temperature range (once thermal equilibrium has been achieved)
Response time: 90% of full scale in < 10 seconds at 77° F (25° C)
Linearity: 0.1% oxygen
Zero drift: ±0.5% oxygen per month
Span drift: ±0.5% oxygen per month
Repeatability: ±0.10% oxygen
Operating temp: 68 to 86° F (20 to 30° C)
Outputs: Two 0-1 VDC and two 4-20 mADC isolated (concentration and range ID)
Alarms: 1 x system-failure alarm contact to detect power failure;
2 x adjustable concentration threshold alarms with fully programmable set-points.
Analysis display: 5 digit red LED, 3/5" high numerals
Menu display: 20 character, 2 line vacuum fluorescent
Digital interface: Full duplex RS-232C communications port
Power requirements: 90 - 130 or 200 - 240 VAC at 47 / 63 Hz (user selected on rear of control unit)
Oxygen sensor: Long-life paramagnetic detector
Sample connections: User specified 1/4" or 6 mm fittings
Area classification: General purpose, indoor environment
Mounting: 19" rack mount
Dimensions: 8.7" H x 19" W x 12.23" L
Weight: 35 lbs (16 Kg)
MODEL 3010MA PARAMAGNETIC OXYGEN ANALYZER
APPLICATIONS• Air separation Bulk gas plants
• Medical air / medical oxygen qualifications
ADVANTAGES• Linearity of analysis across three user-selectable ranges
• Temperature controlled / isolated sensor for long term stability
• Atmospheric Pressure Compensation for high-end, accurate performance
• Auto-ranging to follow process upsets
• Auto-calibration electronics standard
• Exceptional value at a favorable price
STANDARD FEATURES• Proven, robust paramagnetic sensor
• Three user-selectable ranges plus cal range
• Signal output: isolated 4-20mADC
• Programmable auto-ranging
• Two fully adjustable concentration alarm set points with programmable relay functions, Form C contacts, 3A resistive
• Calibration contact span / zero, Form A normally open contacts, 3A resistive
• Self diagnostics with Form C failure alarm contacts
• Full duplex RS-232C communication link
• Five digit oxygen concentration LED display
• 2 x 20 alphanumeric vacuum fluorescent display for set up and diagnostics
• Switch selectable power supply: 90 - 130 or 200 - 240 VAC at 47 / 63 Hz
• Remote calibration digital inputs
© 2005 Teledyne Analytical Instruments, A Teledyne Technologies Company.All rights reserved. Printed in the USA. 11/05LD
WarrantyInstrument is warranted for 1 year against defects in materialor workmanshipNOTE: Specifications and features will vary with application. The above are established and validated during design, but are not to be construed as test criteria for every product. All specifications and features are subject to change without notice.
TELEDYNE ANALYTICAL INSTRUMENTSA Teledyne Technologies Company16830 Chestnut StreetCity of Industry, California 91748, USA
TEL: 626-934-1500 or 888-789-8168FAX: 626-934-1651 EMAIL: [email protected]
www.teledyne-ai.com
• Microprocessor based• Exceptional performance, price value• Advanced paramagnetic sensor• Designed for the air separation industry
Teledyne’s Model 3000MB Paramagnetic Percent Oxygen Analyzer is a versatile, microprocessor based instrument specifically designed to detect oxygen in a variety of gases. The 3000MB is a precise, cost effective instrument capable of measuring percent oxygen in as low as the 0-1% range. Drawing on the solid performance of Teledyne’s highly successful 3000 series control platform, the 3000MB provides the user with a wide variety of features and options to satisfy virtually any application.
PRINCIPLE OF MEASUREMENTThe physical property distinguishing oxygen from other gases is its paramagnetism which is significantly greater than that of common gases. Consequently the molecules of oxygen are attracted by strong magnetic fields which can be used in detection. A paramagnetic sensor consists of two spheres arranged in the form of a dumbbell suspended in a symmetrical, non-uniform magnetic field. When the surrounding gas contains oxygen, the dumbbell spheres rotate out of the magnetic field by the relatively strong attraction of oxygen.A light beam focused on a mirror attached to the dumbbell reflects asymmetrically onto two photo diodes resulting in a voltage shift. The voltage difference produces a current
Teledyne Analytical Instrumentsused to drive the dumbbells back to the original position. The current flow required to maintain the null position is directly proportional to the oxygen concentration and is shown linearly in volume percent oxygen in the display. Properly maintained paramagnetic sensors last for years with little or no attention making them ideal for critical measurements in the low percent range.
FLEXIBILITYThe 3000MB offers three dynamic, user-configurable ranges plus autoranging, and can be calibrated in any range. The instrument is linear on all three ranges, eliminating the need to recalibrate while switching between ranges. The analyzer comes standard with an isolated 4-20 mADC output for oxygen concentration and range identification. Additionally, a bi-directional RS-232C serial communication interface provides for remote monitoring and control of span and zero functions. Teledyne provides auto-calibration capabilities as a standard feature in the 3000 series platform.
APPLICATIONS• Monitoring inert gas blanketing• Air separation and liquefaction• Chemical reaction monitoring• Emissions monitoring• Petrochemical process control• Quality assurance• Gas analysis certification
THE LEADER IN OXYGENThe 3000MB can be ordered as a standard unit or as part of a larger analytical system. With over 50 years of analytical instrument experience, Teledyne supplies sensors, custom engineered analyzers and complete monitoring systems to satisfy unique applications. From portables to on-line units, in microprocessor based or analog configurations, no other manufacturer comes close to matching Teledyne’s breadth of instrumentation.
ADVANTAGES• Linearity of analysis across three user-selectable ranges• Autoranging to follow process upsets• Auto-calibration electronics standard
3000
MB
Paramagnetic Oxygen Analyzer
3000
MB
Paramagnetic Oxygen Analyzer
Operating temp: 32 to 113°F (0 to 45°C)Outputs: Two 0-1 VDC & two 4-20 mADC isolated (concentration and range ID)Alarms: One system-failure alarm contact to detect power failure; two adjustable concentration threshold alarms with fully programmable setpointsAnalysis display: 5 digit red LED, 3/5” high numeralsMenu display: 20 character, 2 line vacuum fluorescentDigital interface: Full duplex RS-232C communications portPower requirements: Universal AC input ranges, 100 to 240 VAC, 50 / 60 Hz, 70 watts maxOxygen sensor: ParamagneticSample connections: User specified 1/4” or 6mm fittingsArea classification: General purpose, non-hazardous Optional: FM approved for Class 1, Div 2 (no purge ) Y purged for Class 1 Div 1 areas Dimensions: 11.81” W x 20.28” H x 8.9” D
Model 3000MB Sample Gas Conditioning RequirementsSample inlet pressure: 5 to 20 psig (regulate to fixed inlet pressure)Sample flow: 1 to 2 SCFHSample dewpoint: Non-condensing at ambient temperatureSample temperature: 0 to 50°CDetector wetted parts: Glass, steel, gold, viton, acrylic glass, epoxy resin
For complete turnkey systems, please contact Teledyne or your local representative
MODEL 3000MB PARAMAGNETIC OxYGEN ANALYzER
STANDARD FEATURES• Proven, robust paramagnetic sensor• Three user-selectable ranges plus cal range• Signal and Range ID output: isolated 4-20 mADC• Programmable autoranging• Two fully adjustable concentration alarm set points with programmable relay functions, Form C contacts, 3A resistive• Calibration contact span/zero, Form A normally open contact, 3A resistive• Self diagnostics with Form C failure alarm contacts• Full duplex RS-232C communication link• Five digit oxygen concentration LED display• 2x20 alphanumeric vacuum fluorescent display for set up and diagnostics• Brass sample passages and fittings• Universal power supply: 100-240 VAC / 50-60 Hz• Remote calibration digital inputs
OPTIONS-C Electronically operated cal/zero valves
SPECIFICATIONSRanges: 3 user selectable ranges; typical ranges 0-5% to 0-100%; autoranging with ID output (lower ranges available)Calibration range: 0-25%Accuracy: ± 1% of full scale at constant temperature ± 5% of full scale over operating temperature range on factory default analysis ranges, once thermal equilibrium has been achievedResponse time: 90% of full scale in < 10 seconds at 77°F (25°C) Linearity: 0.1% O2zero drift: 0.05% O2 / 1 monthSpan drift: 1% FS / 1 monthRepeatability: 0.1% / FS
© 2005 Teledyne Analytical Instruments, A Teledyne Technologies Company.All rights reserved. Printed in the USA. 10/05LD
WarrantyInstrument is warranted for 1 year against defects in materialor workmanshipNOTE: Specifications and features will vary with application. The above are established and validated during design, but are not to be construed as test criteria for every product. All specifications and features are subject to change without notice.
TELEDYNE ANALYTICAL INSTRUMENTSA Teledyne Technologies Company16830 Chestnut StreetCity of Industry, California 91748, USA
TEL: 626-934-1500 or 888-789-8168FAx: 626-934-1651 EMAIL: [email protected]
www.teledyne-ai.com
• Microprocessor based• Exceptional performance, price value• Advanced paramagnetic sensor• Designed for air separation industry
Teledyne’s Model 3000MA Paramagnetic Percent Oxygen Analyzer is a versatile, microprocessor based instrument specifically designed to detect oxygen in a variety of gases. The 3000MA is a precise, cost effective instrument capable of measuring percent oxygen in as low as the 0-1% range. Drawing on the solid performance of Teledyne’s highly successful 3000 series control platform, the 3000MA provides the user with a wide variety of features and options to satisfy virtually any application.
PRINCIPLE OF MEASUREMENTThe physical property distinguishing oxygen from other gases is its paramagnetism which is significantly greater than that of common gases. Consequently the molecules of oxygen are attracted by strong magnetic fields which can be used in detection. A paramagnetic sensor consists of two spheres arranged in the form of a dumbbell suspended in a symmetrical, non-uniform magnetic field. When the surrounding gas contains oxygen, the dumbbell spheres rotate out of the magnetic field by the relatively strong attraction of oxygen.A light beam focused on a mirror attached to the dumbbell reflects asymmetrically onto two photo
Teledyne Analytical Instruments30
00M
A30
00M
A
Paramagnetic Oxygen AnalyzerParamagnetic Oxygen Analyzer
diodes resulting in a voltage shift. The voltage difference produces a current used to drive the dumbbells back to the original position. The current flow required to maintain the null position is directly proportional to the oxygen concentration and is shown linearly in volume percent oxygen in the display. Properly maintained paramagnetic sensors last for years with little or no attention making them ideal for critical measurements in the low percent range.
FLEXIBILITYThe 3000MA offers three dynamic, user configurable ranges plus autoranging, and can be calibrated in any range. The instrument is linear on all three ranges, eliminating the need to recalibrate while switching between ranges. The analyzer comes standard with an isolated 4-20 mADC output for oxygen concentration and range identification. Additionally, a bi-directional RS-232C serial communication interface provides for remote monitoring and control of span and zero functions. Teledyne provides auto-calibration capabilities as a standard feature in the 3000 series platform.
APPLICATIONS• Monitoring inert gas blanketing• Air separation and liquefaction• Chemical reaction monitoring• Emissions monitoring• Petrochemical process control• Quality assurance• Gas analysis certification
THE LEADER IN OXYGENThe 3000MA can be ordered as a standard unit or as part of a larger analytical system. With over 50 years of analytical instrument experience, Teledyne supplies sensors, custom engineered analyzers and complete monitoring systems to satisfy unique applications. From portables to on-line units, in microprocessor based or analog configurations, no other manufacturer comes close to matching Teledyne’s breadth of instrumentation.
ADVANTAGES • Linearity of analysis across three user-selectable ranges • Autoranging to follow process upsets • Auto-calibration electronics standard
Operating temp: 32 to 113°F (0 to 45°C)Outputs: Two 0-1 VDC & two 4-20 mADC isolated (concentration and range ID)Alarms: One system-failure alarm contact to detect power failure; two adjustable concentration threshold alarms with fully programmable setpoints.Analysis display: 5 digit red LED, 3/5” high numeralsMenu display: 20 character, 2 line vacuum fluorescentDigital interface: Full duplex RS-232C communications portPower requirements: Universal AC input ranges, 100 to 240 VAC, 50 / 60 Hz, 70 watts maxOxygen sensor: ParamagneticSample connections: User specified 1/4” or 6mm fittingsArea classification: General purpose flush mountedDimensions: 6.96” H x 8.7” W x 12.2” L (control unit)
Sample Gas Conditioning RequirementsSample inlet pressure: 5 to 20 psig (regulate to fixed inlet pressure)Sample flow: 1 to 2 SCFHSample dewpoint: Non-condensing at ambient temperatureSample temperature: 0 to 50°CDetector wetted parts: Glass, steel, gold, viton, acrylic glass, epoxy resin
For complete turnkey systems, please contact Teledyne or your local representative
MODEL 3000MA PARAMAGNETIC OxyGEN ANALyzER
© 2005 Teledyne Analytical Instruments, A Teledyne Technologies Company.All rights reserved. Printed in the USA. 08/05LD
STANDARD FEATURES• Proven, robust paramagnetic sensor• Three user-selectable ranges plus cal range• Signal and Range ID output: isolated 4-20 mADC• Programmable autoranging• Two fully adjustable concentration alarm set points with programmable relay functions, Form C contacts, 3A resistive• Calibration contact span/zero, Form A normally open contact, 3A resistive• Self diagnostics with Form C failure alarm contacts• Full duplex RS-232C communication link• Five digit oxygen concentration LED display• 2x20 alphanumeric vacuum fluorescent display for set up and diagnostics• Brass sample passages and fittings• Universal power supply: 100-240 VAC / 50-60 Hz• Remote calibration digital inputs
OPTIONS- C Electronically operated cal/zero valves
SPECIFICATIONSRanges: 3 user selectable ranges; typical ranges 0-5% to 0-100%; autoranging with ID output (lower ranges available)Calibration range: 0-25%Accuracy: ±1% of full scale at constant temperature; ±5% of full scale over operating temperature range on factory default analysis ranges, once thermal equilibrium has been achievedResponse time: 90% of full scale in < 10 seconds at 77°F (25°C) Linearity: 1% O2zero drift: 1% O2 / 1 monthSpan drift: 1% FS / 1 monthRepeatability: 0.1% / FS
WarrantyInstrument is warranted for 1 year against defects in material or workmanshipNOTE: Specifications and features will vary with application. The above are established and validated during design, but are not to be construed as test criteria for every product. All specifications and features are subject to change without notice.
TELEDYNE ANALYTICAL INSTRUMENTSA Teledyne Technologies Company16830 Chestnut StreetCity of Industry, California 91748, USA
TEL: 626-934-1500 or 888-789-8168FAx: 626-934-1651 EMAIL: [email protected]
www.teledyne-ai.com
Explosion Proof ParamagneticOxygen Analyzer
MODEL 3020MExplosion Proof Paramagnetic
Oxygen Analyzer
TELEDYNEANALYTICAL INSTRUMENTS
Built for Reliability and Performance
• Fully explosion proof enclosure
• Microprocessor based
• Exceptional performance, price value
• Advanced paramagnetic sensor
• Designed for petrochemical industry
Teledyne’s Model 3020M Paramagnetic Percent Oxygen Analyzer is a versatile, microprocessor based instrument specifically designed to detect oxygen in a variety of gases. The 3020M is a precise, cost effective instrument capable of measuring percent oxygen in as low as the 0-1% range. Drawing on the solid performance of Teledyne’s highly successful 3000 series control platform, the 3020M provides the user with a wide variety of features and options to satisfy virtually any application.
PRINCIPLE OF MEASUREMENTThe physical property distinguishing oxygen from other gases is its paramagnetism which is significantly greater than that of common gases. Consequently the molecules of oxygen are attracted by strong magnetic fields which can be used in detection.
A paramagnetic sensor consists of two spheres arranged in the form of a dumbbell suspended in a symmetrical, non-uniform magnetic field. When the surrounding gas contains oxygen, the dumbbell spheres rotate out of the magnetic field by the relatively strong attraction of oxygen.
A light beam focused on a mirror attached to the dumbbell reflects asymmetrically onto two photo diodes resulting in a voltage shift. The voltage difference produces a current used to drive the dumbbells back to the original
position. The current flow required to maintain the null position is directly proportional to the oxygen concentration and is shown linearly in volume percent oxygen in the display. Properly maintained paramagnetic sensors last for years with little or no attention making them ideal for critical measurements in the low percent range.
FLEXIBILITYThe 3020M offers three dynamic, user-configurable ranges plus autoranging, and can be calibrated in any range. The instrument is linear on all three ranges, eliminating the need to recalibrate while switching between ranges. The analyzer comes standard with an isolated 4-20 mADC output for oxygen concentration and range identification. Additionally, a bi-directional RS-232C serial communication interface provides for remote monitoring and control of span and zero functions. Teledyne provides auto-calibration capabilities as a standard feature in the 3000 series platform.
APPLICATIONSThis instrument is designed specifically for the detection of percent oxygen in petrochemical applications for process control, quality assurance, and safety related purposes. Typical petrochemical plants requiring the 3020M are:
• Ethyl Benzene / Styrene Monomer (EB / SM)
• Pure Terephthalic Acid (PTA)
• Caprolactum
• Phenol / Cumene
• Polycarbonate
• Polyethylene
• Polystyrene
• Ethylene Oxide / Ethylene Glycol (EO / EG)
• Methyl Methacrylate (MMA)
• Ammonia / Urea Fertilizer
• Refinery FCC Unit and flare gas applications
• Marine Vapor Recovery (MVR)
THE LEADER IN OXYGENThe 3020M can be ordered as a standard unit or as part of a larger analytical system. With over 50 years of analytical instrument experience, Teledyne supplies sensors, custom engineered analyzers and complete monitoring systems to satisfy unique applications. From portables to on-line units, in microprocessor based or analog configurations, no other manufacturer comes close to matching Teledyne’s breadth of instrumentation.
ADVANTAGES• Linearity of analysis across three user-selectable ranges
• Autoranging to follow process upsets
• Auto-calibration electronics standard
• No purge gas requirements
MODEL 3020M PARAMAGNETIC OXYGEN ANALYZER
© 2005 Teledyne Analytical Instruments, A Teledyne Technologies Company.All rights reserved. Printed in the USA. 11/05LD
WarrantyInstrument is warranted for 1 year against defects in materialor workmanshipNOTE: Specifications and features will vary with application. The above are established and validated during design, but are not to be construed as test criteria for every product. All specifications and features are subject to change without notice.
TELEDYNE ANALYTICAL INSTRUMENTSA Teledyne Technologies Company16830 Chestnut StreetCity of Industry, California 91748, USA
TEL: 626-934-1500 or 888-789-8168FAX: 626-934-1651 EMAIL: [email protected]
www.teledyne-ai.com
Zero drift: 0.05% O2 / 1 Month
Span drift: 1% FS / 1 Month
Repeatability: 0.1% / FS
Operating temp: 32 to 113°F (0 to 45°C)
Outputs: Two 0-1 VDC & two 4-20 mADC isolated (concentration and range ID)
Alarms: One system-failure alarm contact to detect power failure; two adjustable concentration threshold alarms with fully programmable setpoints.
Analysis display: 5 digit red LED, 3/5" high numerals
Menu display: 20 character, 2 line vacuum fluorescent
Digital interface: Full duplex RS-232C communications port
Power requirements: Universal AC input ranges, 100 to 240 VAC, 50 / 60 Hz, 70 watts max
Oxygen sensor: Paramagnetic
Sample connections: User specified 1/4" or 6mm fittings
Enclosure: Explosion proof U/L and CSA listed housing for Class I, Div 1, Group B, C & D service; NEMA 4/7 rated with baked powder epoxy finish for corrosion resistance
Mounting: Bulkhead mount
Dimensions: 15.25" W x 25.63" H x 11" D (387.35 mm x 651 mm x 279.4 mm)
Sample GaS ConditioninG RequiRementSSample inlet pressure: 5 to 20 psig (regulate to fixed inlet pressure)
Sample flow: 1 to 2 SCFH
Sample dewpoint: Non-condensing at ambient temperature
Sample temperature: 0 to 50°C
Detector wetted parts: Glass, steel, gold, viton, acrylic glass, epoxy resin
For complete turnkey systems, please contact Teledyne or your local representative
StandaRd FeatuReS• Proven, robust paramagnetic sensor
• Explosion proof NEMA 4/7 rated enclosure (Class I, Div 1, Group B)
• Three user-selectable ranges plus cal range
• Signal and Range ID output: isolated 4-20 mADC
• Programmable autoranging
• Two fully adjustable concentration alarm set points with programmable relay functions, Form C contacts, 3A resistive
• Calibration contact span/zero, Form A normally open contact, 3A resistive
• Self diagnostics with Form C failure alarm contacts
• Full duplex RS-232C communication link
• Five digit oxygen concentration LED display
• 2x20 alphanumeric vacuum fluorescent display for set up and diagnostics
• Stainless steel sample passages and fittings
• Universal power supply: 100-240 VAC / 50-60 Hz
• Remote calibration digital inputs
optionS-C Electronically operated cal/zero valves
-F Flame arrestors for Class I, Div 1, Group C/D service
-G Flame arrestors for Class I, Div 1, Group C/D service with cal valves
-H Flame arrestors for Group B (hydrogen) service
-I Flame arrestors for Group B service with cal valves
SpeCiFiCationSRanges: 3 user selectable ranges; typical ranges
0-5% to 0-100%; autoranging with ID output (lower ranges available)
Calibration range: 0-25%
Accuracy: ±1% of full scale at constant temperature; ±5% of full scale over operating temperature range on factory default analysis ranges, once thermal equilibrium has been achieved.
Response time: 90% of full scale in < 10 seconds at 77°F (25°C)
Linearity: 0.1% O2
Technical Data
Measuring Principle:Paramagnetic behaviour of oxygen
Magnos 16
Sample Component:Oxygen (O2)
Measuring Ranges:❚ Smallest measuring range: 0...1 Vol.-% O2
❚ Largest measuring range: 0...100 Vol.-% O2
❚ Measuring range switching ratio: ≤ 1:100
❚ Max. measuring range suppression 1:100, e. g. 99...100 Vol.-% O2
Calibration:❚ Zero- and end-point calibration with nitrogen (N2)
and air or sample gas mixture
❚ Single-point calibration with room air
❚ Automatic calibration via built-in pneumatic module or external valves
Magnos 17
Sample Component:Oxygen (O2) in flue gases or in nitrogen (N2)
Measuring Ranges:❚ Smallest measuring range: 0...3 Vol.-% O2
❚ Largest measuring range: 0...100 Vol.-% O2
❚ Measuring range switching ratio: ≤ 1:33
Calibration:❚ Zero-point calibration with oxygen-free process gas
or substitute gas
❚ End-point calibration with process gas having a known oxygen concentration or with substitute gas
❚ Automatic calibration via built-in pneumatic module or external valves
Magnos 16 – based on the magnetomechanical measuringprinciple the analyzer is ideally suited for oxygen purity meas-urement, process gas measurements and inert gas blanketing.In emission measurements the Magnos 16 fulfills the specialrequirements, e.g. TA-Luft (German Federal Air Purity)Regulations, German 13. and 17. BImSchV Regulations. Theability to freely select measuring ranges and measure highlysuppressed ranges means that the analyzer can be easilyadapted to specific tasks. Even measurements for safety areno problem – monitoring the sample flow rate through themeasuring chamber always ensures that the current oxygenconcentration is being measured. Mounting the analyzer in aflame-proof housing (EEx d) allows use in hazardous areas,zone 1. The standard unit is certified for Class I, Division 2hazardous areas. Another advantage of the analyzer moduleis its short T90 response time. Calibration of the zero point isonly required once a month using air or nitrogen.
Magnos 17 – based on the thermomagnetical measuringprinciple the analyzer is used mainly in the analysis of fluegas including mixed fuel boilers, incinerators, cement fluegas and roasting gas. The robust measuring cell means theMagnos 17 is especially insensitive to vibrations and shocks.
The range of oxygen analyzer modules is complimentedby an electrochemical cell.
ABB AutomationAnalytical DivisionStierstädter Straße 5, D-60488 Frankfurt a. M., http://www.abb.com/analytical 50/24-04 EN 10.99