    Life Products LP-6,

    LP 6 Plus and LP 10Home Ventilator

    ByBryce Younger

  • 8/12/2019 LP10 LP6 Descriptions

    2/15

    2

    Classification Electrically poweredMicroprocessor

    controlled

    Single circuit

    Rotary drive pistonModes LP 6

    OFF, BatteryCharge, A/C,SIMV, and

    Pressure Limitedmodes

    Sine flow wave in allmodes.

    PEEP possible by

    adding PEEP valveto exhalation valve.

  • 8/12/2019 LP10 LP6 Descriptions

    3/15

    3LP 10

  • 8/12/2019 LP10 LP6 Descriptions

    4/15

    4

    Modes on LP 6 Plus and LP 10 Standby A/C: pressure limited or

    not (LP 10)

    SIMV: machine breathspress limited or not (LP10)

    Pressure Cycle Set VT I.T. controls flowrate

    Hi pressure alarmsetting endsinspiration butwithout audible

    alarm. Airwaypressure isindependent of HPalarm setting duringinspiration.

    Pressure Limit Control(LP 10 only) does not

    end inspiration.

    Pressure Limit Control

  • 8/12/2019 LP10 LP6 Descriptions

    5/15

    5

    Classification In SIMV, patient must

    breathe through the

    circuit, humidifier and

    then through the piston

    intake valve unless one-

    way valve teed in nearpatient wye.

    Internal resistance is 1.0

    cm H20/L/sec.

    Oxygen is added between

    the ventilator and thehumidifier or via reservoir

    bag added to piston

    intake valve for 40% or

    more oxygen

    concentrations.

  • 8/12/2019 LP10 LP6 Descriptions

    6/15

    6

    Controls Vt = 100 - 2200 ml Rate = 1 - 38 BPM

    Inspiratory Time = .5 - 5.5sec

    Sensitivity -10 to +10 cm H20

    High Pressure Alarm Limit25 to 100 cm H20

    Internal pressure relief at100 cm H20.

    LP 10 has Pressure Limit

    Control

  • 8/12/2019 LP10 LP6 Descriptions

    7/15

    7

    Modes A/C Mode

    Set

    Tidal volume

    Inspiratory time If the inspiratory time

    is set too long, asystem error alarmwill be activated.Otherwise, inspiratorytime essentiallydetermines the flow

    rate of the set tidalvolume.

    Rate

    LP 10 breaths can bePressure Limited This pressure limit doesnot end inspiration.

    LP 6

    LP 10

  • 8/12/2019 LP10 LP6 Descriptions

    8/15

    8

    Modes cont. Pressure Limited Mode (LP 6)

    The High Pressure

    Alarm/Limit control limits

    pressure wi thoutan alarm

    and does not end

    inspiration. The set tidalvolume may notbe delivered

    if inadequate inspiratory

    time is set. If inspiratory

    time too long will get System

    Error Alarm. Airway pressure developed

    during inspiration is

    independent of alarm/limit

    setting.

  • 8/12/2019 LP10 LP6 Descriptions

    9/15

    9

    Modes cont. SIMV Mode

    Mandatory breaths givenaccording to the Rate controland patient allowed tobreathe through ventilator orone-way valve, if added, forspontaneous breaths. If theSIMV rate is set < 6 BPM andthe patient fails to initiate anassisted SIMV breath within20 seconds, the apnea alarm

    activates and the ventilatorswitches to Backupventilation at a rate of 10BPM.

    LP 10 machine breaths canbe Pressure Limited.

  • 8/12/2019 LP10 LP6 Descriptions

    10/15

    10

    Alarms Low Pressure/Apnea

    Low pressure is the primarydisconnect alarm except whenrate set < 6 BPM. Apnea alarmactivates if no patient effortsensed within 20 seconds and

    the rate is < 6 BPM. Automatically switches to 10

    BPM at the same tidal volumeand inspiratory time. Alarmmust be manually reset.

    Functional in all modes.

  • 8/12/2019 LP10 LP6 Descriptions

    11/15

    11

    Alarms Low Power

    alarms when batterypower drops to 10 volts.There is no alarm whenpower source switchesfrom internal battery to

    AC. There is an alarmwhen power switchesfrom AC to internal andmust be reset.

    High Pressure

    Ends inspiration in A/Cand SIMV modes. Setsthe pressure limit whichdoes not end inspirationin Pressure Limited Modefor LP 6.

  • 8/12/2019 LP10 LP6 Descriptions

    12/15

    12

    Alarms

    Setting error

    activates ifinspiratory time

    setting incompatiblewith set tidal volumeand respiratory rate.Ends inspiration.

    Power switch over

    indicates whetherAC, internal, orexternal battery inuse.

  • 8/12/2019 LP10 LP6 Descriptions

    13/15

    13

    Indicators

    Battery Test - shows charge level underneath

    airway pressure.

    Breathing effort - indicates assisted breaths. Power - indicates whether AC power or

    battery charge, external battery, or internal

    battery in use. Battery requires 2 hours of

    charge for every 1 hour of use.

  • 8/12/2019 LP10 LP6 Descriptions

    14/15

    14

    O2 Bleed in

    Can bleed in oxygen between ventilator and

    humidifier up to 40%. 40% and over must add

    reservoir bag and oxygen to the back at thepiston intake filter.

    VTset = VTdesired (1 - FIO2)

    .79

    LPM = (VTdesired - VTset) BPM

    1000

  • 8/12/2019 LP10 LP6 Descriptions

    15/15

    15

    Circuit

    Circuit

    Requires exhalation valve and proximal

    airway pressure line. Clean circuit withsoapy water, rinse, and then soak in 1/3

    cup vinegar per cup of water for 30

    minutes every 24 hr.