lp capacitive brochure r6 · 2020. 7. 30. · 8.00/7.97 0.315/0.314 51 2.01 133 5.25 15 0.6 5.1...
TRANSCRIPT
PositionDisplacementHigh Speed
High ResolutionCustomizable Solutions
Off the Shelf
CAPACITIVESENSORS
HIGH PRECISION NON-CONTACT SENSORS
CAPACITIVE SENSORS
Capacitive Sensor Diagram
Elite Series (CPL490, CPL190/290)Best performance; adjustments and indicators
CPL230, CPL350No user adjustments or indicators
CPA100 Precision Analog ProximityAdjustable gain and zero, IndicatorsSetpoint output
RangeDependent on Sensing Area
Mounting• Cylindrical:3 mm, 5 mm, 8 mm, 3/8” 18 mm
• Rectangular: 20 mm, 45 mm
PROBE
Electric Field (AC)
Driver
Target
Selections
Prob
eSe
lections
DriverExcitation volt age (AC)Signal processingCalibrationOutput voltage/data
CustomSensor DesignWeare glad toworkwith you on custom probes andelectronics for your specific application. Over half of ourorders are customized, this includes:
• CustomProbe ShapesAndDesigns
• CustomCable Lengths And Connectors
• VacuumCompatible Probes
• CustomRanges
• Custom Electronics
lionprecision.comOur Lion Precision Technical Library provides a high level oftechnical detail about all of our products and technologiesincluding all manuals and drawings of the probes.
Our Capacitive Sensing Theory of Operation TechNoteis viewed thousands of times amonth.
Need assistance? Please contact Lion Precision at 651.484.6544 or [email protected]
ELITE SERIESCPL500 CPL490 CPL190/290 CPL230 CPL350 CPA100
Product Selection Guide CPL591/592 CPL490 CPL190/290 CPL230 CPL350 CPA100
STANDARD
TECH
NICALSP
ECIFICAT
IONS
Typical Resolution*(% F.S. rms) @ 15 kHz 0.004% 0.0007% 0.003% 0.004% 0.004% 0.03% @
Midrange
Nonlinearity* 0.1% 0.2% 0.2% 0.5% 0.2% 0.2%
Dual Range(Sensitivity) CPL290
User Adjustments
Adjustable Gain
Adjustable Offset
Range Indicator
Setpoint/ SwitchedOutput
Channels Per Package 1 - 8 1 - 3 1 - 8 1 - 6 1 1
Selectable Bandwidth(Internal Dip-Switch)
0.1, 1, 10, 15kHz
1, 10, 15, 50kHz
0.1, 1, 10, 15kHz
0.1, 1, 10, 15kHz
0.1, 1, 10, 15kHz
Maximum Bandwidth 15 kHz 50 kHz 15 kHz 15 kHz 15 kHz 15 kHz
CustomizationAvailable
Digital Output
* Typical Specifications dependent on probe and range
Product SelectionGuide
Selecting the Right Technology
*Specifications listed on the following page are for probeswith standard 2meter cables.Listed specifications do not applywhen customization is included. Performance ofcustomized productwill depend on specific options.
Probe VacuumCompatibility (to 10-6 Torr)ProbeHigh-VacuumCompatibility (to 10-12 Torr)
Options requiring custom calibration and deviation from listed specifications:
• CustomRanges
• Nonflat or Undersized Target Shape
• CustomCable Length
• Probe Extension Cables (2m, 3m, 4m; Probes calibratedwith extension cablesmust beusedwith them)
CAPACITIVE DRIVER
CPL590Capacitive Driver• Digital Outputs
• USB, SPI and EtherCAT
• Small 2U Size
• Up to 4 Selectable Ranges on 1 Probe
• Best Linearity
• Higher Stability
• Digital Adjustment
SpecificationsResolution1: 0.0005%@100Hz
0.003%@15 kHz
Selectable Bandwidth: 100Hz, 1, 10, 15kHz
Linearity² : <0.1% F.S. typical
Max Drift: 0.04% F.S./°C
Operating Temp: 4-50°C
Output: ±10V, 0Ω 10mAmax
Rear-Panel: SPI, EtherCAT1. Dependent on probe, range, and bandwidth. See next page for details.
2. Dependent on probe and range. See next page for details.
4.00 ”101.6 mm
8.12”206.25 mm
3.25”82.55mm
Export LicenseMay require an export license to some countries.
Need assistance? Please contact Lion Precision at 651.484.6544 or [email protected]
CPL590Probes and Ranges
Selecting the Right Technology
C5SShapeC = CylindricalR = Rectangular
Size in mmC: DiameterR: Longest Side
Body StyleBlank = LongS = ShortR = Right Angle
Sensing Area
TARGET
NEAR GAP
RANGETARGET
Sensing AreaDiametermm
Measurement Range Resolution1 @Bandwidth Available Body Sizes
Range TypeRangeμmmils
Near Gapμmmils
1 kHznmμin
10 kHznmμin
15 kHznmμin
Linearity%FS
Models Body Sizes
0.5
Fine 100.4
200.8
0.140.006
0.560.022
0.700.028 1.0 C3S
C3RC5SC5RC5
Standard 502.0
502.0
0.700.028
4.20.17
5.60.22 0.50
Extended 803.0
602.4
1.40.056
7.00.28 – 0.50
0.8Fine 25
1.0753.0
0.700.028
1.70.070
2.10.084 0.50 C3S
C3RC5SC5RC5Standard 100
4.01004.0
1.40.056
4.90.20
7.00.28 0.50
2.0
Ultrafine 100.4
200.8
0.110.004
0.210.008
0.350.014 1.0
C5SC5RC5C8SC8RC8
Fine 502.0
753.0
0.420.017
0.840.034
1.40.056 0.30
Standard 25010.0
1255.0
0.560.060
5.60.22
7.00.28 0.30
Extended 50020.0
1255.0
4.20.17
110.44
140.56 0.30
3.2
Fine 502.0
1255.0
0.560.022
1.40.056
2.20.073 0.30
C8SC8RC8
Standard 50020.0
25010
4.20.17
8.40.37
140.56 0.30
Extended 125050.0
25010
210.84
281.1
421.7 0.30
5.6
Fine 502.0
2259.0
0.560.022
1.10.44
1.80.072 0.30
C9.5SC9.5RC9.5R20
Standard 50020.0
50020
4.20.17
9.80.39
140.56 0.30
Extended 200080.0
25010
140.56
281.1
421.7 0.30
13
Fine 200080
200080
421.7
494.2
562.2 0.50
C18Standard 3200125
200080
562.2
702.8
843.4 0.50
Extended 5000200
3000120
1405.6
1807.3
2108.4 0.50
19Standard 2500
1005000200
1003.9
1005.0
1405.6 0.50
R45Extended 6000
2503000120
1706.7
1009.0
25010 1.0
21Standard 8000
3005000200
1405.6
1807.3
2108.4 0.50
C25Extended 12500
5005000200
25010
32013
35014 0.50
Resolution values are RMS. Peak-to-peak values are typically 8-10 times greater than the RMS values.In high EMI conditions (10 V/m) output DC level may shift and noise may rise to 0.2 VRMS (1.3% resolution).
CPL590Probe Measurement
Ranges and Resolutions
CAPACITIVE DRIVER
CPL490Elite Series• Highest resolution
• Widest bandwidth (50 kHz)• Five-element range indicator
• Zero adjust
• Front-panel BNC analog output
• Differential output to National Instruments 68-pinconnector
• Uses 2ndGeneration Probes
SpecificationsResolution1: 0.0004%@1 kHz
0.0007%@15 kHz
0.002%@50 kHz
Selectable Bandwidth: 1, 10, 15, 50 kHz
Linearity2: <0.2% F.S. typical
MaxDrift: 0.02-0.04% F.S./°C
Operating Temp: 15-40°C
Front-Panel BNC: ±10V, 0Ω 10mAmax
Rear-Panel National Inst: ±10V, 0ΩDifferential
Multiple Channels: Up to 3 per enclosure(Contact Lion Precision formore than 3)
1. Dependent on probe, range, and bandwidth. See next page for details.
2. Dependent on probe and range. See next page for details.
Listed specifications assume a twometer probe cable: Flat measurementarea diameter at least 1.3 times larger than the Sensing Areawith nocustomizations.
The CPL490 uses 2ndGeneration probes.
5.87 ”149.0 mm
3.75”95.25mm
8.50”215.9 mm
Export LicenseMay require an export license to some countries.
Need assistance? Please contact Lion Precision at 651.484.6544 or [email protected]
CPL490Probes and Ranges
SecondGeneration Probes• The CPL490 uses 2ndGeneration Capacitive Probeswhich include electronics in the probe housing.
• The probes aremounted by the 8mmdiameter probe body extending from the larger housing.
• Twomodels are available differing only in the sensing area diameter and associatedmeasurement ranges.
2G-C8-0.5: 0.5 mm sensing area2G-C8-1.2: 1.2 mm sensing area
Standard Cable2 m ±25 mm
672.64
150.6
0-100-0.4
mminch
170.67
Clamp MountingDiameter8.00/7.97
0.315/0.314
512.01
1335.25
150.6
5.10.20
4.00.166.00.24
0.5 mmSensing Area1.2 mmSensing AreaHeat Shrink
Strain Relief
Sensing Area
TARGET
NEAR GAP
RANGETARGET
CPL490Probe Measurement
Ranges and Resolutions
Sensing AreaDiametermm(ProbeModel) Range Type
Rangeμmmils
Near Gapμmmils
1 kHznmμin
10 kHznmμin
15 kHznmμin
50 kHznmμin
ProbeMax. Drift% FS/°C
0.5(2G-C8-0.5)
Fine 100.4
200.8
0.050.002
0.070.003
0.090.004
0.260.010 0.04
Standard 502.0
251.0
0.170.007
0.270.011
0.350.014
1.00.040 0.02
Extended 1004.0
502.0
0.380.015
0.800.31
1.00.039
3.30.14 0.03
1.2(2g-C8-1.2)
Fine 502.0
251.0
0.150.006
0.200.008
0.220.009
0.630.25 0.02
Standard 1004.0
502.0
0.330.013
0.400.016
0.520.021
1.70.065 0.02
Ultrafine 2008.0
1004.0
0.680.027
1.00.040
1.30.050
3.80.15 0.02
Range is determined by the sensing area diameter. The larger the diameter, the larger the range.Flat target surface must be 1.3 times larger that the sensing area diameter.
CAPACITIVE DRIVER
CPL190/CPL290Elite Series• High resolution five-element range indication
• Coarse/Fine zero adjust with disable
• Front-panel BNC analog output
• Differential output to National Instruments68-pin connector
SpecificationsResolution1: 0.0005%@100 kHz
0.003%@15 kHz
Selectable Bandwidth: 100 Hz, 1, 10, 15 kHz
Linearity2: <0.2% F.S. typical
MaxDrift: 0.04% F.S./°C
Operating Temp: -50°C
Front-Panel BNC: ±10V, 0Ω 10mAmax
Rear-Panel National Inst: ±10V, 0ΩDifferential1. Dependent on probe, range, and bandwidth. See next page for details.
2. Dependent on probe and range. See next page for details.
Listed specifications assume a twometer probe cable; Flat measurementarea diameter at least 1.3 times larger than the Sensing Areawith nocustomizations.
Different probe body styles/sizes are available for each Sensing area.
Export LicenseMay require an export license to some countries.
Need assistance? Please contact Lion Precision at 651.484.6544 or [email protected]
CPL190/CPL290Probes and Ranges
Sensing Area
TARGET
NEAR GAP
RANGETARGET
C5SShapeC = CylindricalR = Rectangular
Size in mmC: DiameterR: Longest Side
Body StyleBlank = LongS = ShortR = Right Angle
CPL190/290Probe Measurement
Ranges and Resolutions
SensingArea
Diametermm
Measurement Range Resolution1 @Bandwidth Available Body Sizes
RangeType
Rangeμmmils
NearGapμmmils
100 Hznmμin
1 kHznmμin
10 kHznmμin
15 kHznmμin
Linearity% F.S. Models Body Sizes
0.5
Fine 100.4
200.8
0.060.003
0.140.006
0.400.016
0.500.020 0.25 C3S
C3RC5SC5RC5
Standard 502.0
502.0
0.300.012
0.500.020
3.00.12
4.00.16 0.25
Extended 803.0
602.4
0.500.02
1.00.040
5.00.20 – 0.25
0.8Fine 25
1.0753.0
0.200.008
0.500.020
1.20.050
1.50.060 0.15 C3S
C3RC5SC5RC5Standard 100
4.01004.0
0.500.020
1.00.040
3.50.14
5.00.20 0.15
2.0
Ultrafine 100.4
200.8
0.050.002
0.080.003
0.150.006
0.250.010 0.15
C5SC5RC5C8SC8RC8
Fine 502.0
753.0
0.200.008
0.300.012
0.600.024
1.00.040 0.15
Standard 25010.0
1255.0
0.80.032
1.00.040
4.00.16
5.00.20 0.10
Extended 50020.0
1255.0
1.50.060
3.00.12
8.00.32
100.40 0.15
3.2
Fine 502.0
1255.0
0.250.010
0.40.016
1.00.042
1.60.048 0.20
C8SC8RC8
Standard 50020.0
25010
2.00.08
3.00.12
6.00.24
100.40 0.15
Extended 125050.0
25010
100.40
150.60
200.80
301.2 0.20
5.6
Fine 502.0
2259.0
0.30.012
0.40.016
0.80.032
1.30.052 0.20
C9.5SC9.5RC9.5R20
Standard 50020.0
50020
2.50.100
3.00.12
7.00.28
100.40 0.15
Extended 200080.0
25010
7.00.28
100.40
200.80
301.2 0.20
13
Fine 200080
200080
200.80
301.2
351.4
401.6 0.50
C18Standard 3200125
200080
301.2
401.6
502.0
602.4 0.50
Extended 5000200
3000120
753.0
1004.0
1305.2
1506.0 0.50
19Standard 2500
1005000200
502.0
702.8
903.6
1004.0 0.30
R45Extended 6000
2503000120
903.6
1204.8
1606.4
1807.2 1.0
21Standard 8000
3005000200
753.0
1004.0
1305.2
1506.0 0.50
C25Extended 12500
5005000200
1305.2
1807.2
2309.2
25010 0.50
Resolution values are RMS. Peak-to-peak values are typically 8-10 times greater than the RMS values.In high EMI conditions (10 V/m) output DC level may shift and noise may rise to 0.2 VRMS (1% resolution).
CAPACITIVE DRIVER
CPL230Compact,Multi-Channel Driver• Separate power and signal connectors
• OEMand embedded application
• Small Size, high-density package
• No user adjustments
• ±5V single-ended output, ±10V differential output
CPL350Compact, Single-Channel Driver• OEMand embedded applications
• ±10V single-ended
• BNC output
• ±10V differential D-Sub output
• No user adjustments
• Small size
• High-density package
SpecificationsResolution1: 0.0007%@100 kHz
0.004%@15 kHz
Selectable Bandwidth: 100 Hz, 1, 10, 15 kHz
Linearity2: <0.5% F.S. typical
Probe Drift: 0.04% F.S./°C
Operating Temp: 4-50°C
Output: ±10 V, Differential
±5 V, Single-Ended
Input Power3: ±15 VDC, 500mAmax1. Dependent on probe, range, and bandwidth. See next page for details.
2. Dependent on probe and range. See next page for details.
3. External power supply included. See next page for details.
Listed specifications assume a twometer probe cable; Flat measurementarea diameter at least 1.3 times larger than the Sensing Areawith nocustomizations.
5.87 ”149.09mm
2.37 ”60.19mm
4.50 ”114.3mm
8.50 ”215.9mm
2.12 ”53.90mm
4.0 ”101.6mmCPL230
CPL350
Export LicenseMay require an export license to some countries.
Need assistance? Please contact Lion Precision at 651.484.6544 or [email protected]
CPL230/350Probes and Ranges
C5SShapeC = CylindricalR = Rectangular
Size in mmC: DiameterR: Longest Side
Body StyleBlank = LongS = ShortR = RightAngle
CPL230/350Probe Measurement
Ranges and Resolutions
Sensing Area
TARGET
NEAR GAP
RANGETARGET
Sensing AreaDiametermm
Measurement Range Resolution1 @Bandwidth Available Body Sizes
RangeType
Rangeμmmils
Near Gapμmmils
100 Hznmμin
1 kHznmμin
10 kHznmμin
15 kHznmμin
Linearity Models Body Sizes
0.5
Fine 100.4
200.8
0.080.003
0.140.006
0.560.022
0.700.028 1.0 C3S
C3RC5SC5RC5
Standard 502.0
502.0
0.420.017
0.700.028
4.20.17
5.60.22 0.50
Extended 803.0
602.4
0.700.028
1.40.056
7.00.28 – 0.50
0.8Fine 25
1.0753.0
0.280.011
0.700.028
1.70.070
2.10.084 0.50
C3SC3RC5SC5RC5
Standard 1004.0
1004.0
0.700.030
1.40.056
4.90.20
7.00.28 0.50
2.0
Ultrafine 100.4
200.8
0.070.003
0.110.004
0.210.008
0.350.014 1.0
C5SC5RC5C8SC8RC8
Fine 502.0
753.0
0.280.011
0.420.017
0.840.034
1.40.056 0.30
Standard 25010.0
1255.0
1.10.04
1.40.060
5.60.22
7.00.28 0.30
Extended 50020.0
1255.0
2.10.084
4.20.17
110.44
140.56 0.30
3.2
Fine 502.0
1255.0
0.350.014
0.560.022
1.40.056
2.20.073 0.30
C8SC8RC8
Standard 50020.0
25010
2.80.11
4.20.17
8.40.37
140.56 0.30
Extended 125050.0
25010
140.56
210.84
281.1
421.7 0.30
5.6
Fine 502.0
2259.0
0.420.017
0.560.022
1.10.44
1.80.072 0.30
C9.5SC9.5RC9.5R20
Standard 50020.0
50020
3.50.14
4.20.17
9.80.39
140.56 0.30
Extended 200080.0
25010
9.80.39
140.56
280.1.1
421.7 0.30
13
Fine 200080
200080
281.1
421.7
492.0
562.2 0.50
C18Standard 3200125
200080
421.7
562.2
702.8
843.4 0.50
Extended 5000200
3000120
1004.0
1405.6
1807.3
2108.4 0.50
19Standard 2500
1005000200
702.8
1003.9
1005.0
1405.6 0.50
R45Extended 6000
2503000120
1305.2
1706.7
1009.0
25010 1.0
21Standard 8000
3005000200
1004.0
1405.6
1807.3
2108.4 0.50
C25Extended 12500
5005000200
1807.3
25010
32013
35014 0.50
Resolution values are RMS. Peak-to-peak values are typically 8-10 times greater than the RMS values.In high EMI conditions (10 V/m) output DC level may shift and noise may rise to 0.2 VRMS (1.3% resolution).
CAPACITIVE DRIVER
CPA100PrecisionAnalog Proximity Switch• Lower Cost
• 0-10 V nonlinear analog output
• Setpoint/Switched output
• Adjustable gain and zero
• Five-element range indicator
• Remote gain and zero adjustment connections
• DIN rail case
• Multiple channel sync
SpecificationsResolution1: CPA100: 0.03%RMS@15 kHz
CPA 100e: 0.3μmRMSor higher@ 15 kHz
Bandwidth: 15 kHz
Linearity: Nonlinear
MaxDrift: Driver: 0.2% F.S./°C
Probe: 0.05% F.S./°C
Analog Output: 0-10V, 0Ω, 15mAmax.
Setpoint: Fixed at 5 V (mid-range)
Setpoint Output: Solid state switch closure
On state: 2.5Ω, 100mAmax
Off state leakage: 10μAmax
Operating Temp: 4-50°C
Onput Power2: 15-24 VDC, 2.5W1. Typical at midpoint. Dependent on probe, range, and bandwidth.See next page for details.
2. External power supply not included – available as accessory (p.3)
Listed specifications assume a twometer probe cable; Flat measurementarea diameter at least 1.3 times larger than the Sensing Areawith nocustomizations.
Different probe body styles/sizes are available for each Sensing area.
Export LicenseMay require an export license to some countries.
35mmDIN Rail Mount
4.4”112mm
0.9”23mm
3.9”100mm
Need assistance? Please contact Lion Precision at 651.484.6544 or [email protected]
CPA100Probes and Ranges
SensingArea
Diametermm
Measurement Range Midpoint Resolution1
@15 kHzAvailable Body Sizes
RangeType
Rangeμmmils
Near Gapμmmils
CPA100μmμin
CPA100eμmμin
Models Body Sizes
0.5Standard 50
2.0502.0
552.2
40016 C3S
C3RC5SC5RC5Extended 80
3.0602.4
1004.0
40016
0.8 Standard 1004.0
1004.0
552.2
40016
C3SC3R
C5SC5RC5
2.0Standard 250
10.01255.0
401.6
40016 C5S
C5RC5
C8SC8RC8Extended 500
20.01255.0
552.2
40016
3.2Standard 500
20.025010
853.4
40016 C8S
C8RC8Extended 1250
50.025010
1305.2
40016
5.6Standard 500
20.050020
1405.5
40016 C9.5S
C9.5RC9.5Extended 2000
80.025010
1706.7
80032
13Standard 3200
125200080
100040
100040
C18Extended 5000
2003000120
160063
160063
19Standard 2500
1005000200
80032
80032
R45Extended 6000
2503000120
110044
110044
21Standard 8000
3005000200
120047
120047
C25Extended 12500
5005000200
160063
160063
The CPA100 does not require an export license.
TARGET
NEAR GAP
RANGETARGET
C5SShapeC = CylindricalR = Rectangular
Size in mmC: DiameterR: Longest Side
Body StyleBlank = LongS = ShortR = RightAngle
Sensing Area
CPA100Probe Measurement
Ranges and Resolutions
CAPACITIVE PROBES
ProbeModel Numbers• Probemodel numbers are a combination of the BodyModelnumber and Sensing AreaDiameter in mm (e.g. C5S-08 or R45-19).
• Cad Files available atwww.lionprecision.com/technical-library
Size
/Sha
pe
Bod
yMod
el
Mechanical
Sensing AreaSensingArea
Diameter(mm)
Measurement Ranges (by Driver Model)
CPL190, CPL290CPL230, CPL350
μmmils
CPA100μmmils
3mm
Cylin
drical C3S
0.5 10, 50, 800.4, 2.0, 3.0
50, 802.0, 3.0
0.8 25, 1001.0, 4.0
1004.0
C3R
0.5 10, 50, 800.4, 2.0, 3.0
50, 802.0, 3.0
0.8 25, 1001.0, 4.0
1004.0
5mm
Cylin
drical
C5
0.5 10, 50, 800.4, 2.0, 3.0
50, 802.0, 3.0
0.8 25, 1001.0, 4.0
1004.0
2.0 10, 50, 250, 5000.4, 2.0, 10.0, 20.0
250, 50010.0, 20.0
C5S
0.5 10, 50, 800.4, 2.0, 3.0
50, 802.0, 3.0
0.8 25, 1001.0, 4.0
1004.0
2.0 10, 50, 250, 5000.4, 2.0, 10.0, 20.0
250, 50010.0, 20.0
C5R
0.5 10, 50, 800.4, 2.0, 3.0
50, 802.0, 3.0
0.8 25, 1001.0, 4.0
1004.0
2.0 10, 50, 250, 5000.4, 2.0, 10.0, 20.0
250, 50010.0, 20.0
20mm
Rectan
gle
R20 5.6 50, 500, 20002.0, 20.0, 80.0
200, 200020.0, 80.0
C5SShapeC = CylindricalR = Rectangular
Size in mmC: DiameterR: Longest Side
Body StyleBlank = LongS = ShortR = RightAngle
3 mm +0/-0.03 mm (0.118” +0/-0.001”)
3 mm +0/-0.03 mm (0.118” +0/-0.001”)
5.00/4.97 mm (0.1965/0.1955”)
5.00/4.97 mm (0.1965/0.1955”)
6.5 mm (0.26”)
4 mm (0.16”)
1 mm (0.04”) 17 mm (0.67”)
5 mm (0.20”)
3 mm (0.12”)1 mm (0.04”)
11 mm (0.43”)
4 mm (0.16”)
4 mm (0.16”)
4 mm (0.16”)
1 mm (0.04”)
1 mm (0.04”)
1 mm (0.04”)
27 mm (1.06”)
14 mm (0.55”)
14 mm (0.55”)
15 mm (0.59”)
20 mm (0.787”)
20 mm (0.787”)
1.8 mmØ
5mm (0.197”)16 mm (0.630”)
1.8 mm3.2 mmØ
Standard cable length: 2meters. Operating and storage temperature range: 4-50°C. Available as vacuum compatible.
Probes are not damaged by contactwith the sensing tip.
Range is determined by the probe Sensing Area diameter – the larger the diameter, the larger the range.
To hold specifications, flat target surface diametermust be 1.3 times larger than the sensing area diameter.
Measurement Ranges and other performance specifications are dependent on the selected drivermodel.
Need assistance? Please contact Lion Precision at 651.484.6544 or [email protected]
Size
/Sha
pe
Bod
yMod
elMechanical Sensing
AreaDiameter(mm)
Measurement Ranges (by Driver Model)
CPL190, CPL290CPL230, CPL350
μmmils
CPA100μmmils
8mm
Cylin
drical
C8
2.0 10, 50, 250, 5000.4, 2.0, 10.0, 20.0
250, 50010.0, 20.0
3.2 50, 500, 12502.0, 20.0, 50.0
500, 125020.0, 50.0
C8S
2.0 10, 50, 250, 5000.4, 2.0, 10.0, 20.0
250, 50010.0, 20.0
3.2 50, 500, 12502.0, 20.0, 50.0
500, 125020.0, 50.0
C8R
2.0 10, 50, 250, 5000.4, 2.0, 10.0, 20.0
250, 50010.0, 20.0
3.2 50, 500, 12502.0, 20.0, 50.0
500, 125020.0, 50.0
9.5m
m(3/8”)
Cylin
drical
C9.5 5.6 50, 500, 20002.0, 20.0, 80.0
500, 200020.0, 80.0
C9.5S 5.6 50, 500, 20002.0, 20.0, 80.0
500, 200020.0, 80.0
C9.5R 5.6 50, 500, 20002.0, 20.0, 80.0
500, 200020.0, 80.0
18mm
Cylin
drical
C18 13 2000, 3200, 500080, 125, 2000
3200, 5000125, 200
25mm
Cylin
drical
C25 21 8000, 12500300, 500
8000, 12500300, 500
45mm
(1.75”
)Cylindrical
R45 19 2500, 6000100, 250
2500, 6000100, 250
8.00/7.97mm (0.315/0.314”)
8.00/7.97mm (0.315/0.314”)
9.53/9.50mm (0.375/0.374”)
9.53/9.50mm (0.375/0.374”)
9.53/9.50mm(0.375/0.374”)
17.91 ± 0.025mm (0.705 ± 0.001”)
17.91 ± 0.025mm (0.705 ± 0.001”)
8.00/7.97mm(0.315/0.314”)
6 mm (0.24”)
6 mm (0.24”)
6 mm (0.24”)
1 mm (0.04”)
1 mm (0.04”)
1 mm (0.04”)
1 mm (0.04”)
1 mm (0.04”)
1 mm (0.04”)
8 mm (0.31”)
8 mm (0.31”)
57.2 mm (2.25”)19.1 mm (0.75”)
44.5 mm(1.75”)
38.1 mm(1.5”)
12.5 mm (0.5”) 63.5 mm (2.5”)
M4x.074x
15.8 mm(0.62”)
25 mm(0.99”)
8 mm (0.32”)
8 mm (0.32”)
8 mm (0.32”)
39 mm (1.54”)
17 mm (0.67”)
17 mm (0.67”)
17 mm (0.67”)
17 mm (0.67”)
35 mm (1.38”)
35 mm (1.38”)
39 mm (1.54”)
7166 4th Street N. Oakdale, MN 55128T: 651.484.6544 F: [email protected] division of Motion Tech Automation
All Rights Reserved - August - 2019 ©2020 Lion Precision - Specifications Subject to Change Without Notice.
CAPACITIVE SENSORS
Sensor
Sensor
Sensor Sensor
Sensor
Sensor
Sensor
Position Displacement
Deformation
Thickness
Part Sorting
Runout, Eccentricity
Vibration
Typical Applications
COMMITMENT TO SERVICELion Precision’s commitment to service is unsurpassed in the industry. We partner with our customersto ensure their success by providing optimized sensing solutions. Contact us today and let us solve yourdifficult measurement problems.