effects of sample holders on activated carbon …
TRANSCRIPT
The Pennsylvania State University
The Graduate School
Department of Energy and Mineral Engineering
EFFECTS OF SAMPLE HOLDERS ON ACTIVATED CARBON SORBENT
TUBES WHEN MEASURING ORGANIC VAPOR CONCENTRATIONS
A Thesis in
Energy and Mineral Engineering
by
Brian S. Marpoe
2010 Brian S. Marpoe
Submitted in Partial Fulfillment
of the Requirements
for the Degree of
Master of Science
August 2010
ii
The thesis of Brian Marpoe was reviewed and approved* by the following:
William A. Groves
Associate Professor of Industrial Health and Safety
Thesis Advisor
R. Larry Grayson
Professor of Energy and Mineral Engineering
Mark S. Klima
Associate Professor of Mineral Processing and Geo-Environmental Engineering
Yaw D. Yeboah
Professor and Department Head of Energy and Mineral Engineering
*Signatures are on file in the Graduate School
iii
ABSTRACT
A study was conducted to examine whether there are significant differences
between organic vapor concentrations measured using activated carbon sorbent tubes
with three different sample holder configurations: no sample holder (open tube), and
using SKC- and Buck-brand sample holders. A two-level fractional factorial
experimental design was used with the following factors and levels: Vapor (low boiling
point, high boiling point), Pump Type (pulsating, continuous), Exposure Profile (variable,
constant), Flow Rate (30 mL/min, 200 mL/min), Duration (30 min, 80 min), and Sample
Placement (mannequin, free-hanging). Two of each sample holder configuration (six
total) were placed in an exposure chamber and a dynamic test-atmosphere generation
system was used to prepare atmospheres containing approximately 12 to 15 ppm n-
hexane (low b.p. vapor: 69º C) or m-xylene (high b.p. vapor: 139 ºC) with exposure
profiles and sampling conducted according to the run sheet generated from the
experimental design. A total of 24 runs were completed with six samples collected per
run, yielding 144 samples which were analyzed by GC/FID.
Concentration results for each pair of SKC- and Buck-brand sample holders were
averaged and normalized by dividing by the average result for the open-tube sampler
from the same run to eliminate the effect of daily variation in chamber concentrations.
The resulting ratio of sample tube holder and open tube concentrations was used as the
response variable. Results of ANOVA using the General Linear Model (MINITAB
Release 14) identified Flow Rate (p=0.006) and Sample Holder (p=0.021) as significant
iv
factors. However, the magnitude of the effect was small with the mean concentration
ratio ranging from 0.99 -1.04 for the Flow Rate effect, and from 0.99-1.03 for the Sample
Holder effect. These results suggest that exposure assessment errors resulting from use
of sorbent tube sample holders for organic vapor monitoring are relatively small and not
likely to be of practical importance.
v
TABLE OF CONTENTS
List of Figures................................................................................................................vii
List of Tables................................................................................................................. .ix
Acknowledgements..........................................................................................................x
Chapter 1 Introduction ................................................................................................. ...1
Chapter 2 Background ................................................................................................. ...4
2.1 Review of Literature ...................................................................................... ...4
2.1.1 History of Industrial Hygiene .............................................................. ...4
2.1.2 Personal Sampling Methods and Techniques ...................................... ...7
2.1.3 Solid Sorbent Gas and Vapor Samplers .............................................. .12
2.1.4 Error Associated With Sampling ......................................................... .17
2.2 Overall Benefits of the Study ........................................................................ .19
2.3 Specific Aims ................................................................................................. .20
Chapter 3 Methods ....................................................................................................... .21
3.1 Preliminary Test Atmosphere Generation Tests ............................................ .21
3.2 Equipment and Instrumentation ..................................................................... .23
3.2.1 SKC and Buck Sample Holders .......................................................... .23
3.2.2 Continuous and Pulsating Pumps ........................................................ .25
3.2.3 Photo Ionization Detector (PID) .......................................................... .27
3.2.4 Exposure Chamber and Gas Generation System ................................. .28
vi
3.3 Experimental Design ..................................................................................... .30
3.4 Experimental Procedure ................................................................................. .33
Chapter 4 Results and Discussion ................................................................................ .36
4.1 Individual Factor Results Using Measured Concentrations .......................... .36
4.2 Individual Factor Results Using a Ratio of Measured Concentrations.......... .44
4.3 Overall Sample Holder Results ..................................................................... .48
4.4 Comparing Sampled Concentration vs. Measured PID Concentration ......... .50
Chapter 5 Conclusions ................................................................................................. .53
References .................................................................................................................... .54
Appendix A Raw Data ................................................................................................ .56
vii
LIST OF FIGURES
Figure 1.1: SKC Sample Holder....................................................................................2
Figure 2.1: NIOSH Design of Charcoal Tube (Crisp, 1980) ....................................... 15
Figure 2.2: A.P. Buck Sample holder (left), open tube sample (center), and SKC
Sample Holder (right) ................................................................................... 16
Figure 3.1: SKC Sample Holder .................................................................................. 24
Figure 3.2: Adjustable Buck Sample Holder ............................................................... 25
Figure 3.3: Pulsating Pump (left), DryCal, and Continuous Pump (right) .................. 26
Figure 3.4: MiniRAE 2000 PID ................................................................................... 27
Figure 3.5: Water Bath With Gas Washing Bottle ...................................................... 29
Figure 3.6: Test Atmosphere Generation System ........................................................ 29
Figure 3.7: Actual Chamber Used in the Study ........................................................... 30
Figure 3.8: Mannequin and Free Hanging Stand ......................................................... 34
Figure 3.9: Constant Concentration Exposure Profile for Vapors ............................... 34
Figure 3.10: Variable Concentration Exposure Profile for Vapors ............................. 35
Figure 4.1: ANOVA Results for Concentration (PPM) as a Response Variable......... 38
viii
Figure 4.2: Interaction Plot for Vapor, Pump Type, and Duration Factors ................. 39
Figure 4.3: Measured PID Concentration for Run 4L In Experimental Design. ......... 40
Figure 4.4: Concentration Rise of Both Vapors in the Experimental Chamber...........42
Figure 4.5: Rate of Decrease in Vapor Concentration ................................................. 43
Figure 4.6: ANOVA Results for Ratio of Concentrations as a Response Variable ..... 45
Figure 4.7: Interaction Plot for Flow Rate and Sample Holder ................................... 47
Figure 4.8: Interval Plot Comparing Means of Measured Concentration and PID
Concentration.............................................................................................51
ix
LIST OF TABLES
Table 2.1: The Use of Solid Sorbents in Air Sampling from 1960-1980 ................... 14
Table 2.2: Required Accuracy For Specific Concentration Levels ............................. 18
Table 3.1: Chemical Properties and Exposure Limits for n-Hexane and m-Xylene ... 22
Table 3.2: Parameters For Test Atmosphere Generation ............................................. 23
Table 3.3: Effects Examined in Multi-Factorial Design .............................................. 31
Table 3.4: Study Design Matrix ................................................................................... 32
Table 4.1: Analysis of Variance Results for Concentration......................................... 37
Table 4.2: Analysis of Variance for Concentration with Interaction Terms ................ 379
Table 4.3: Analysis of Variance Results for Ratio of Concentrations ......................... 40
Table 4.4: Analysis of Variance Results for Ratio of Concentrations with
Interaction Terms ........................................................................................ 456
Table 4.5: Ratio's of Open Holder and Sample Holders..............................................469
Table 4.6: T-test Comparing Measured Concentration vs. PID Concentration...........50
Table 4.7: One-Sample T-test Comparing Ratio's of Measured and PID
Concentrations.............................................................................................51
x
ACKNOWLEDGEMENTS
I would like to express my sincere gratitude and thanks to my advisor Dr. William
Groves whose advice, encouragement, and support throughout my stay here at Penn State
from the initial to final level helped me to develop an understanding of the subject.
Support for this project by the National Institute for Occupational Safety and
Health (NIOSH) is gratefully acknowledged.
Lastly, I would like to thank my parents, Harold and Deb, for helping to support
me financially and emotionally, not only through my collegiate career, but throughout my
life.
1
Chapter 1
Introduction
When air sampling is needed to test for organic gases or vapors in a work place
the most common type of sampling uses a sampling train that consists of a pump and
carbon sorbent tube. The sorbent tube is filled with activated carbon where vapors can be
adsorbed. The sample can then be desorbed from the carbon using carbon disulfide (CS2)
into a solution and analyzed using gas chromatography. Gas chromatography coupled
with a Flame Ionization Detector (FID) is used to evaluate the solution so the sample
concentration can be calculated. When opening the glass sorbent tube before sampling,
the tips have to be broken off each end, leaving behind sharp glass edges. These glass
edges could easily harm a worker during personal sampling if they are not careful.
Sample holders have been designed to protect workers from the glass edges of carbon
sorbent tubes by placing the sample holder directly over the sorbent tube and still
allowing it to sample with no hindrances. There are different brands and types of sample
holders, but all are very similar. They usually are a plastic casing that screws directly
over the sample tube, and has a clip that allows it to be placed on a worker’s collar or
shirt as seen in Figure 1.1.
2
Figure 1.1: SKC Sample Holder
A previous study was done that involved sampling vapors in a chamber while a
subject was performing a physical test to evaluate a physiologic sampling pump (PSP)
(Lin et al., 2008; Lee et al., 2009; Lin et al., 2010) . From this study, the air samples
were taken using a sample holder or using an open sorbent tube. Although the purpose of
the experiment was to examine the performance of the PSP and its custom-designed
sample holder, results were also obtained using a traditional sampling pump with and
without a sample holder. These results suggested possible variations and a significant
difference, on the order of 10%, between sampled vapor concentrations with and without
a sample holder. This proposed the question, "Are sample holders affecting sampled
concentrations?" To address this problem, the National Institute for Occupational Safety
and Health (NIOSH) funded this research project to determine if sample holders were
affecting the measured concentrations, and if any underlying factors played a role in the
different concentrations.
3
The project focused on experimentally examining whether using three different
sample holder configurations would cause significant differences (>10%) between
concentrations measured using carbon sorbent tubes for organic vapors. The three types
of sample holder configurations were: a) no sample holder, b) using a SKC brand sample
holder, c) using an A.P. Buck brand sample holder. The project also aimed to identify
and characterize significant factors affecting concentrations such as the chemical and
physical properties of vapors, type of sampling pump, exposure profile, sampling flow
rate, and if the sample holder was placed on a mannequin or placed on a free-hanging
sampling stand. Lastly, if significant effects were identified, evaluating mechanisms that
governs the effects would be analyzed.
A fractional factorial experimental design was used to incorporate all of these
factors into 24 experiments. These experiments involved sampling organic vapors in a
chamber using sampling trains under the conditions given in the experimental design. A
vapor generation system put a known concentration of the vapor through the chamber,
while six samples were taken using different sample holders (two of each type) to
determine the differences in each measured concentration. From the previous study it
was hypothesized that sample holders were causing differences in measured
concentrations.
4
Chapter 2
Background
A review of the literature relevant to the history and background of this project is
presented in this chapter. An overview of the experiment done before this project that
helped in development of the experimental design is also described. Lastly, the overall
objectives and specific aims for this study will be discussed.
2.1 Review of Literature
This section presents a review of current literature on the history of gas and vapor
sampling, measuring with carbon sorbent tubes, and associated error while sampling. It
should be noted that not much research has occurred with sample holder designs and
uses.
2.1.1 History of Industrial Hygiene
Hippocrates recognized and recorded the very earliest signs of occupational
disease during the 4th century B.C by discovering the existence of lead toxicity in the
mining industry. Ulrich Ellenbog published a pamphlet on occupational diseases in 1473,
and Bernadino Ramazzini wrote the first full treatise on the same subject in 1700 (Cullis
and Firth, 1981), but very little concern was put forth to protect the health of workers
until the 19th century. It was not until 1833 that the Factory Acts of Great Britain were
passed because of the rapid increase of industrial diseases and accidents happening in
5
Great Britain. These acts were aimed more toward providing the workers with
compensation for accidents rather than preventing and controlling the causes, but are
considered the first effective legislative acts in industry that required concern for the
workers (Leidel, Busch, and Lynch, 1977).
In 1908, the United States passed a compensation act (Federal Employers'
Liability Act; FELA) for injured employees of any interstate railroad, and in 1911 the
initial state worker's compensation laws were agreed upon by the government. Since
worker's compensation is administered on a state-by-state basis, individual states slowly
created their own worker's compensation laws, and by 1948 every state had some form of
worker's compensation to take care of injured employees. Employers started to become
more aware of health and safety incident rates over the next 20 years. The increase of
awareness with the ever-growing industrial type jobs led to the passing of multiple health
and safety acts in the 1960's. The largest movement made for the health and protection of
workers and prevention of diseases came from The Federal Coal Mine Health and Safety
Act of 1969 (Leidel, Busch, and Lynch, 1977). This act helped develop and set the basis
of what United States hygienists believe is the most important piece of federal legislation:
The Occupational Safety and Health Act of 1970 (Perkins, 1997).
Prior to 1970, very few states were regulating health and safety in the workplace.
Injury, accident, and death statistics were available, and people started to realize the
losses from every day work places. In 1968, the Secretary of Labor William Wirtz stated
that there will be 55 dead, 8,500 disabled, and 27,200 hurt every day in America's work
places (Mintz, 2002). The Occupational Safety and Health Act was developed in 1970 to
help govern the health and safety of workers in the private sector and federal government
6
in the United States. The Act's main goal was to keep the workplace safe by keeping it
free from any recognized hazards, such as excessive noise levels, unsanitary conditions,
toxic chemicals, and mechanical dangers. Taken directly from the purpose in the
preamble to the Act it states:
"To assure safe and healthful working conditions for working men and
women; by authorizing enforcement of the standards developed under
the Act; by assisting and encouraging the states in their efforts to assure
safe and healthful working conditions; by providing for research,
information, education, and training in the field of occupational safety
and health; and for other purposes." (Occupational Safety and Health
Act of 1970)
This act created a new administrative unit in the Department of Labor, the Occupational
Safety and Health Administration (OSHA). OSHA was to serve and regulate the work
places throughout the United States and ensure the safety of all employees. OSHA also
served to create the standards, rules, and regulations that every industry and work place
would have to comply with. The OSHA standards that became important to industrial
hygienists were the Permissible Exposure Limits (PEL) designed to control chemical
substance concentrations to keep workers safe (Perkins, 1997). In 1974, along with the
National Institute for Occupational Safety and Health (NIOSH), OSHA began putting
together tables of approximately 400 chemical substances. Health standards for each
chemical substance were created and included areas such as:
Measurement of employee exposure
Medical surveillance
7
Methods of compliance
Handling and use of liquid substances
Employee training
Recordkeeping
Sanitation and housekeeping (Leidel, Busch, and Lynch, 1977)
These standards could be followed using appropriate sampling and analytical methods for
measuring employee exposures that would be later developed by OSHA.
2.1.2 Personal Sampling Methods and Techniques
Personal sampling is the best way to monitor what and how much of a
contaminant an employee is being exposed to. Industrial hygiene programs' have one
main goal, and it is to accurately evaluate employees' occupational exposures to airborne
contaminants using sampling techniques (Leidel, Busch, and Lynch, 1977). OSHA has
developed many guidelines and rules to sample workers and work places for harmful
aerosols, gases, vapors, or dusts. Even though there are guidelines for sampling, there is
no generally accepted standardized procedure for air quality testing. Employers base
their monitoring responsibilities on several factors:
1) Industries will put forth different efforts and approaches to air monitoring
depending on the type of work done. A chemical company that deals with
toxic chemicals daily will put more effort into an air monitoring program
than a plant putting together automobile parts.
8
2) The safety manager or health professional in the company will have
different views on responsibility from company to company.
3) Air quality monitoring requires manpower, capital equipment,
laboratories, and a reasonably high level of expertise that a company will
have to afford. (Cullis and Firth, 1981)
In 1974, NIOSH and OSHA put together a book of guidelines for air sampling in
response for the need of standardized methods. One of the important requirements for
each air sampling method under consideration was to provide results on average within
±25% of the (true) concentration 95% of the time (Kennedy, Fischbach, Song, Eller, and
Shulman, 1996). These types of confidence intervals take into account many areas of
error with sampling and will be discussed later in the chapter.
Industrial hygiene analytical methods are directed towards personal sampling of a
worker. To prepare a method for personal sampling it is best to ask oneself the following
six questions: what, where, how long, when, who, and how many to sample? When
answering the question what to sample for, the response should be anything that may
cause harm or sickness to a worker or the surrounding environment in the workplace.
Where to sample is a difficult question that is important with this study which is
examining the use of sample holders. Studies have shown that there is bias whether or
not the sample is taken from the dominant or non-dominant side of the worker, and
similar studies have shown opposite results. Either way, hygienists still attach the
sampler on a worker's shirt or collar in the breathing zone, and if multiple samples are
needed it is suggested to alternate sides to eliminate any bias. The breathing zone is
defined by Perkins to be anywhere around the nose or mouth. How long to sample
9
should be based on what is being sampled. Does the hazard produce acute or chronic
effects? A time weighted average (TWA) over eight hours is the common answer, which
can then be compared to the occupational exposure limit (OEL). When to sample is
another difficult question. Hygienists are most likely to pick the worst-case sampling
time to determine if workers are being overexposed to the hazard and it is usually what is
done to ensure the safety of the worker. NIOSH developed a technique in 1977 called the
Recommended Employee Exposure Determination and Measurement Strategy, which
helped indentify who to sample in the work place (Perkins, 1997). The strategy
suggested choosing the worst-case exposure situation and sampling that worker. This
was to ensure that if the sample results come back in compliance there would be no need
for further sampling. Lastly, how many samples should be taken has had lengthy
considerations. Initially, large sample sizes were recommended to ensure enough
employees were being sampled, but today hygienists must consider adequate statistical
design of their sampling strategy to determine the number of samples needed (Perkins,
1997). All of these questions need to be answered to produce a good monitoring
program.
When developing a monitoring program, it is first helpful to understand the
reasons for sampling for air contaminants and how the resulting data will be used.
Reasons for air monitoring is given by Cullis (Cullis and Firth, 1981), and they are as
follows:
1) To protect the worker from sickness, impaired health, discomfort, and
inefficiency due to working conditions;
10
2) To fulfill the company's responsibilities under the local Health and
Safety legislation;
3) To ensure compliance with air quality standards;
4) To establish an ongoing record of employee exposure levels;
5) To indicate areas and operations where respiratory protection may be
necessary; and
6) To provide data for epidemiological surveys, where medical records are
analyzed with exposure data.
Cullis states that the main goal of an air monitoring program is to evaluate the average
exposure and range of exposures for every job situation in the company. Sample results
that are exceeding the recommended OEL's are a guide for safety professionals to
develop a plan to improve the conditions. Whether it means re-designing the work
practices, using a new type of engineering control, or the addition of personal protective
equipment (PPE); monitoring programs help to minimize worker's exposure (Cullis and
Firth, 1981). There are different personal monitoring techniques that can be used to
identify airborne contaminants.
There are diverse types of devices used to sample organic gases and vapors in a
work place. These types of instruments include dosimeters, chemical stain devices, and
time weighted average (TWA) sampling devices. This study focused on TWA sampling
devices for gases and vapors. TWA instruments sample for a known sampling time with
a known flow rate. Once the amount of contaminant sampled for is determined by
analysis, an average concentration can be determined by knowing the mass of the
contaminant and the amount of volume that passed through the TWA device. Different
11
types of TWA samplers include slow-filling containers, evacuated metal containers,
impingers, bubblers, and vapor adsorption tubes (Cullis and Firth, 1981).
It is less difficult to sample for gases and vapors rather than aerosols in an
occupational environment because at standard temperatures and pressures they follow the
ideal gas law and mix freely with air. This allows the gas or vapor to reach equilibrium
fairly quickly. Since gases and vapors are often used together, it should be noted there
are differences between the two. Gases are substances that at standard conditions (70°F
and 1 atm) are in their normal physical state. Vapors are gaseous forms of a substance
that under standard conditions may exist as a solid or liquid. Both gases and vapors are
collected with the same air sampling device and methods. The two techniques used to
sample for gases and vapors require laboratory analysis of collected samples, or use a
direct-reading instrument that is capable of sampling a volume of air (Peach and Carr,
1986). The two techniques are very different, and both have their pros and cons. A
direct-reading instrument can get immediate concentrations; while sampling using a
charcoal tube requires time between measurements and obtaining the results. However,
there are very few gases and vapors that OSHA allows Photo Ionization Detectors (PID's)
to be used for compliance measurements, so traditional laboratory analysis is most often
used to test a worker's safety. Both techniques are used in this study, but the main focus
is the method used that needs laboratory analysis, and more specifically solid sorbent gas
samplers.
12
2.1.3 Solid Sorbent Gas and Vapor Samplers
Gas and vapor adsorption tubes are the most widely used, and probably the most
convenient method for collecting TWA samples. Before 1936, filtration techniques were
being used to measure aerosols such as asbestos, coal dust, and silica in industrial
atmospheres, but gases and vapors also were causing diseases among workers. The
adsorption properties of activated carbon, silica gel, and alumina were documented in the
mid-1930's. Early attempts tried using gravimetric techniques to analyze these samples
but were unsuccessful due to water vapor and other contaminants also being adsorbed.
Initially, scientists could not determine ways to separate the gas or vapor from other
materials on the sample. This in turn led to additional weight on the samples, so the mass
could not be determined of the contaminant, which hindered calculating the TWA
concentrations. In 1952, a separation technique, gas chromatography (GC), became
available and by 1958, experiments were being done in conjunction with charcoal
sorbents and gas chromatography analysis (Perkins, 1997). The development of the small
battery-operated pump around 1960 helped push the use of charcoal to sample for gases
and vapors on a worker. The pump allowed for samples to be taken in the worker's
breathing zone and over an entire work shift (Rappaport and Kupper, 2008). Finally in
1964, carbon disulfide (CS2) was discovered to remove contaminants from the charcoal,
and not interfere with the analysis of other compounds in the GC column. Industrial
Hygienists could then unload a charcoal tube into a vial, and add a known amount of CS2
to prepare an injection sample for GC analysis. After this discovery was made, charcoal
became the solid sorbent of choice for sampling gases and vapors (Perkins, 1997).
13
The principle of adsorption is to pull a known volume of air at a constant flow
rate over the charcoal in a glass tube. As the air passes through the tube, the molecules of
gas can be adsorbed onto the charcoal while being chemically unchanged. The chemical
can then be desorbed using a liquid solvent and analyzed (Peach and Carr, 1986). To act
as an effective collecting medium a sorption material must fulfill several requirements:
1) It must be able to trap small concentrations of contaminant from a stream
of air, and retain it until an analysis can be made;
2) A convenient efficient method of desorption must be available;
3) The capacity of the sorbent within the sampler tube must be large enough
to retain an amount of contaminant sufficient to facilitate its analysis;
4) The sorbent must be such that no chemical change occurs to the
contaminant while stored; and
5) The sorbent should adsorb the contaminant in the presence of other
contaminants (Crisp, 1980).
Solid sorbents other than charcoal can be used to adsorb vapors and gases. Some
examples of such sorbents are porous polymers, gas-liquid chromatography
compositions, silica gel, clathrate silicates, and alumina (Crisp, 1980). These other
materials were introduced because activated carbon cannot adsorb some materials when
another sorbent can. For example, carbon does not bind well with alcohols or other polar
compounds in which case another sorbent would be used. Active charcoal and carbon are
the primary ones used as sorbents, but other applicable sorbents are listed in Table 2.1.
14
Table 2.1: The Use of Solid Sorbents in Air Sampling from 1960-1980 (Crisp, 1980)
Type of Sorbent Number of Compounds
Reported Adsorbed
Active Charcoal and Carbon 139
Porous Polymers 131
Silica Gel 84
Loaded Chromatographic Columns 43
Molecular Sieves 6
Alumina 6
Miscellaneous 3
The charcoal can be obtained from a variety of materials ranging from coconut
shells, coal, wood, peat, cotton, and Douglas fir (Levin and Carleborg, 1987). The main
advantage of using charcoal is that it has an extensive internal surface area. This area can
be as large as 10,000 sq. ft. per gram of material, which greatly enhances its adsorption
capacity (Peach and Carr, 1986). For industrial work, two types of charcoal have been
used most often. Coconut charcoal is the sorbent recommended by NIOSH and is
utilized the most, but charcoal made from petroleum is used the second highest (Perkins,
1997).
The first standardized miniature adsorbent tube was developed by NIOSH and the
design can be seen in Figure 2.1 (Anderson, Levin, Lindahl, and Nilsson, 1984).
15
Figure 2.1: NIOSH Design of Charcoal Tube (Crisp, 1980)
NIOSH produced the charcoal tube to be made from glass 7-cm long, 4-mm i.d., and 5-
mm o.d., and made it so that both ends are sealed until sampling begins. There are two
sections of charcoal in the tube. The primary section contains 100 mg, and the second
section is 50 mg of charcoal. Each section is analyzed separately and the back section is
used to indicate breakthrough. To open the tube for sampling, each tip is broken off and
attached accordingly to a sampling pump with a rubber hose. Sample holders are placed
over the tube for two reasons. The first reason is ensure worker safety from the sharp
glass edges on the end. The second reason is the sample holder allows the charcoal tube
to be easily attached to the worker with a clip directly in the breathing zone, yet not
interfering with basic job tasks. After sampling, the charcoal tube should be sealed
effectively with plastic caps that come with the tubes, and then stored according to the
volatility and reactivity of the contaminant sampled until analysis can be done (Perkins,
1997).
Figure 2.2 shows two types of sample holders and an open tube. From the image
it is easy to see that sample holders are very similar. Other types of sample holders are
also available, but all operate the same way. Depending on the manufacturer of the
16
sample holder, a needle valve may be included on the sample holder to adjust the flow
rate, which also can be seen in Figure 2.2 on the open tube in the middle.
Figure 2.2: A.P. Buck Sample Holder (left), Open Tube Sample (center), and SKC
Sample Holder (right)
The analysis of solvent gas and vapors is typically done using gas
chromatography. Once the contaminant has been adsorbed onto charcoal, it can be
desorbed using a suitable solvent and analyzed using the gas chromatograph. Typically,
when measuring for a gas or vapor in a workplace other chemicals will be adsorbed onto
the charcoal also. What makes the gas chromatograph so helpful is the ability to
simultaneously separate and measure all the compounds in the solvent mixture (White,
Taylor, Mauer, and Kupel, 1970). NIOSH recommends flame ionization or electron
capture gas chromatography. Flame Ionization Detectors (FID) are a type of gas detector
that uses a flame to ionize the solution passing through the GC, while an Electron
Capture Detector (ECD) uses a radioactive beta particle emitter in place of the flame to
detect molecules by measuring the change in current between the collector anode and
17
cathode in the sample solution. The current is changed because electrons are being
captured, and these signals are what translates into the concentrations being measured.
Although complete recovery of the analyte from the sampler is desired, at a minimum,
the estimated recovery of the analyte from the collection medium should be greater than
or equal to 75% for concentrations equivalent to sampling 0.1, 0.5, 1.0, and 2.0 times the
exposure limit (Kennedy, Fischbach, Song, Eller, and Shulman, 1996). The disadvantages
of using GC analysis are the solvent desorption of the contaminants is not always 100%
and carbon disulfide is extremely toxic and flammable (Peach and Carr, 1986).
2.1.4 Error Associated With Sampling
During sampling and analysis for gases and vapors there is error introduced in
various steps of the methods. NIOSH's Guidelines for Air Sampling and Analytical
Method Development and Evaluation states that for any method to be considered valid, it
has to evaluate whether, on average, over a concentration range of 0.1-2.0 times the
exposure limit, the method will provide a result that is within ±25% of the true
concentration 95% of the time with the absolute bias being less than 10% (Kennedy,
Fischbach, Song, Eller, and Shulman, 1996). For example, if an experiment had a known
concentration of a contaminant to be 100 ppm, and an industrial hygienist were to
measure between 75 ppm and 125 ppm 95% of the time, then that method could be
considered acceptable. Exposure assessment consists of an evaluation of the method,
pre-sampling procedures, post-sampling procedures, and analysis of the sample. The pre-
and post-sampling procedures, along with the analysis of the sample have uncertainties
18
associated with them. NIOSH published a table in their Sampling Strategy Manual that
can be seen in Table 2.2. The table shows the required accuracy for a method at different
levels of concentrations. If the concentration is lower than the action level, a method
only needs to be 50% accurate.
Table 2.2: Required Accuracy For Specific Concentration Levels (Leidel, Busch, and
Lynch, 1977)
Concentration Required
Accuracy
Above Permissible Exposure Limit ±25%
At or below the Permissible Exposure
Limit and above the Action Level ±35%
At or below the Action Level ±50%
Crisp (1980) suggested seven criteria for a successful sampling procedure:
1) Accuracy and precision of combined analytical and collection procedures for
concentrations ranging from 0.1-2.0 TLV to be within ±16% of true
concentrations at the 95% confidence level;
2) Total recovery efficiency at least 75%;
3) Bias between total recovery and desorption efficiency less than ±10%;
4) Capability of taking 10-15 min samples at ceiling and excursion levels;
5) Minimum sampling time of 1 hour, preferred 4-8 hours for TWA;
6) Stored samples should compare within ±10% relative to initial samples after
being stored for 14 days; and
7) The flow rate of the sampling pump should be known with an accuracy of
±5%.
19
The value NIOSH derived for pump and calibration error is commonly assumed
to be 5% (Perkins, 1997). Other sources of error include loss of sample to the walls of
tubes and sample holders through adsorption. When first beginning to sample for an
organic gas or vapor, some of the molecules will adsorb to the walls of the rubber hose
until equilibrium is reached. Lastly, a comprehensive survey of several compounds over
the range of 0.6x TLV to its 15 minute short term exposure limit (STEL) was done, and it
concluded that total relative error was 12.7% split between analytical and sampling error
(Crisp, 1980).
2.2 Overall Benefits of the Study
The results of this study would indicate whether or not sample holders when used
to measure gases and vapors with carbon sorbent tubes were causing differences in
measured concentrations. It also would help show the effects of flow rate, pump type,
exposure profile, vapor properties, and if the sample holder is mounted on a mannequin
or free hanging stand, on measured concentrations. If the results determined that
measured concentrations had differences greater than 10% with and without a sample
holder, possible future research would be to provide models for adjustments that could be
made to adjust the concentrations.
20
2.3 Specific Aims
The specific aims of this study were as follows:
1.) Examine whether there are significant differences (>10%) between
concentrations measured using activated carbon sorbent tubes with three
different sample holder configurations: a) no sample holder - charcoal tube
facing upward with end open to ambient atmosphere, b) using an SKC-brand
sample holder, and c) using an A.P. Buck-brand sample holder.
2.) Characterize the nature of any significant effects by examining different
vapors spanning a range of chemical/physical properties (e.g., polarity, BP,
VP), different types of sampling pumps (pulsating stroke-counter type vs.
continuous type pumps), different vapor exposure profiles (variable vs.
relatively constant), and different sampling flow rates (200 mL/min vs. 30
mL/min).
3.) Propose and evaluate mechanisms governing any identified effects.
21
Chapter 3
Methods
This section outlines the instrumentation and procedure used in this study. Based
on a previous study using the vapor generation chamber, it was decided to use the same
conditions in the chamber for exposure profiles. The conditions had to be modified
slightly, and a preliminary test atmosphere generation test was done with the chamber to
ensure that both vapors being used were approximately the same concentration averaged
over time. The experimental setup included continuous and pulsating pumps to measure
the concentrations, and a PID to measure a reference concentration in the chamber, which
could be later used as a comparison. There were a total of 24 sample runs, which
combined all of the factors with one another. All of the samples were collected and then
analyzed using a gas chromatograph (GC) with a flame ionization detector (FID).
3.1 Preliminary Test Atmosphere Generation Tests
To compare the sampled concentrations between the two types of vapors being
used (n-hexane and m-xylene), preliminary tests were conducted to configure the vapor
generation system so that both vapors had the same average concentration over the
sampled time. To keep the experiment safe, 15 PPM was chosen as the target
concentration for both vapors, which was well below OSHA's PEL's (Table 3.1). To
produce these concentrations, changes in the water bath temperature and flow rate
through the gas washing bottle had to be adjusted.
22
Table 3.1: Chemical Properties and Exposure Limits for n-Hexane and m-Xylene
Property n-Hexane m-Xylene
Molecular Formula C6H14 C8H10
Molecular Weight 86.18 g/mol 106.16 g/mol
Vapor Pressure 124 mm Hg 8 mm Hg
Boiling Point 69 °C 139 °C
OSHA PEL 500 ppm 100 ppm
ACGIH TLV 50 ppm 100 ppm
The constant exposure profile for m-xylene was tested first. For a constant
exposure profile the vapor generation system was turned on at a specific setting, and the
concentration was allowed to reach equilibrium in the chamber. A PID was placed in the
chamber, visible through a window, to monitor the concentration. Initially, the
temperature and flow rate were used from the previous study to determine a baseline
concentration. The first concentration was found to be 35 PPM, so a reduction in the
flow rate was done, and trials continued until 15 PPM was reached. Since the variable
exposure profile consisted of the gas generation system being turned on for 10 minutes
and then off for 10 minutes; 20 minute trial runs were done so that the average PID
reading was 15 PPM. The same procedure was done with n-hexane.
It was determined that for m-xylene a water bath temperature of 55°C would be
used for both constant and variable exposure profiles. A flow rate of 1.85 L/min was
used for the constant exposure profile, while a flow rate of 4.07 L/min was turned on and
23
off for the variable profile. Since n-hexane is more volatile than xylene, lower
temperatures and flow rates were needed to reach the 15 PPM concentration. The
temperature of the water bath containing n-hexane was 27°C, just above room
temperature. The flow rate used for the constant exposure profile was 0.150 L/min, and
0.415 L/min for the variable exposure profile. The flow rates were adjusted using
LabVIEW and will be discussed later in the procedure.
Table 3.2: Parameters for Test Atmosphere Generation
Parameter m-xylene n-hexane
Water Bath Temperature 55 °C 27 °C
Flow Rate for Constant
Exposure Profile
1.85 L/min 0.150 L/min
Flow Rate for Variable
Exposure Profile
4.07 L/min 0.415 L/min
3.2 Equipment and Instrumentation
The following section will describe all of the equipment and instrumentation
used in this study. There were no human subjects used in this study.
3.2.1 SKC and Buck Sample Holders
The two types of sample holders chosen for this project were SKC- and A.P.
Buck-brand sample holders. The holders consisted of an adjustable flow holder where
the carbon sorbent tube was attached, and a plastic cover was used to go over the tube.
24
The adjustable flow holder used a needle valve to better regulate the flow passing through
the sorbent tube. One can use the sample pump to acquire approximately the target flow
rate, and make further adjustments using the needle valve to obtain the exact flow rate
needed.
The SKC holder (SKC Single Adjustable Low Flow Tube Holder #224-26-01)
was combined with a constant pressure controller (SKC Low Flow CPC #224-26CPC-10)
to be used for low-flow applications on the continuous pumps. The protective tube cover
(SKC Type B Protective Tube Cover #224-29B) can be seen in Figure 3.1. Each SKC
holder was labeled and used in both continuous and pulsating pump experiments.
Figure 3.1: SKC Sample Holder
Two types of Buck Holders were used because of the limitations with the flow
adjustments when combined with the pulsating pump. The valve inside the adjustable
low-flow sample holder (A.P. Buck Inc #APB-109030) would not operate correctly at
low flow rates with the pulsating pump, so a similar non-adjustable low-flow holder had
to be used (A.P. Buck Inc #APB-109032). Both sample holders looked and operated the
25
same; the only difference was one could not be adjusted and had to be used with the
pulsating pumps. Figure 3.2 shows an image of the adjustable buck sample holder.
Figure 3.2: Adjustable Buck Sample Holder
Activated carbon sorbent tubes (SKC #226-01) were used throughout the entire
study. Each tube was pre-labeled with the date, experiment run number, and a number
showing where it was positioned in the chamber. The location in the chamber indicates
where on the mannequin or free-hanging sampler the sample was hung. An example
label would be 9-6, indicating that the sorbent tube was used in experiment 9, and
positioned farthest to the right. The sorbent tubes were stored after sampling in a freezer
until analysis could be done.
3.2.2 Continuous and Pulsating Pumps
Six of each type of sampling pump (continuous, pulsating) were used for
measuring the concentrations in the chamber. The two models of pumps used were the
continuous pump (SKC AirLite #110-100) and pulsating pump (SKC 222 Series Pump
26
#222-3). Each pump was labeled to make it easier when pre- and post-calibrating the
pumps.
Calibrating the pumps was done using a primary standard before and after each
experimental run with a DryCal electronic flowmeter (BIOS DryCal DC-Lite DCL-M).
The pulsating pumps had re-chargeable batteries, which were fully charged after every
run. The continuous pumps operated on 4 AA batteries, which were swapped out after
approximately 250 minutes of operation. Figure 3.3 shows both pumps along with the
DryCal used in the study.
Figure 3.3: Pulsating Pump (left), DryCal, and Continuous Pump (right)
Initially, each pulsating pump was also calibrated using a bubble buret and stop
watch combination to determine the flow rate (mL/stroke). When calibrating the
pulsating pumps for trial runs, the beginning counter strokes were recorded along with
the final counter strokes after the experiment, and the volume was calculated. These
27
volumes were then compared with the volume calculated using the DryCal flow rates. In
every experiment both volumes sampled were within 5% and thereafter the flow rates
were determined using the DryCal only.
3.2.3 Photo Ionization Detector (PID)
A PID (MiniRAE 2000) was used to record chamber concentrations at one-second
intervals during experimental runs. The PID was calibrated before each run according to
manufacturer's instructions using 100 ppm isobutylene span gas, and then a correction
factor was used to adjust for the two vapors being used. A 10.6 eV lamp was used in the
PID, which gave correction factors of 4.3 and 0.43 for m-xylene and n-hexane,
respectively. After every run, the PID was connected to a computer, and a plot of
concentration versus time was displayed and saved. Details on each run were also saved
including the highest and average concentrations. An image of the PID used throughout
this study is shown in Figure 3.4.
Figure 3.4: MiniRAE 2000 PID
28
3.2.4 Exposure Chamber and Gas Generation System
The exposure chamber and gas generation system were the same used in the
previous study. The gas generation system was controlled using a program in LabVIEW
(Student Version 8.2) written by Brandon Lin (PhD - Industrial Engineering). The
designed program used a mass flow controller to control the flow rates being passed
through the gas bubbler. The vapors leaving the bubbler would then enter a dilution
stream to pass through the test chamber. The program allowed the user to enter voltages,
which related to specific flow rates that could be turned on for appropriate durations of
time. This is how the constant and variable exposure profiles were obtained.
An image of the water bath and gas washing bottle used can be seen in Figure 3.5.
Figure 3.6 is a schematic that shows how the chamber operated. The chemical being
used would be filled in a gas washing bottle and placed in the water bath set to the pre-
determined temperature. The mass flow controller, operated by LabVIEW, would send a
known flow rate through the gas washing bottle pushing the vapor into a diluted stream
of fresh air. The vapor was pushed through the chamber using a blower, and an exhaust
fan on the other side of the chamber would pull the air out into a hood.
29
Figure 3.5: Water Bath with Gas Washing Bottle
The frame of the chamber was made from wood and was 4.5 ft x 6 ft x 9 ft (243
ft3). It was sealed tightly with plastic vinyl wrap to minimize leakage from the chamber.
The supply and exhaust blowers were adjusted to keep the chamber under negative
pressure. This ensured that if there were any leaks the vapor would still stay in the
Dilution Air
Supply Blower
Compressed
Air
Dynamic Test Atmosphere
Generation System
Exposure Chamber
Exhaust Blower
Figure 3.6: Test Atmosphere Generation System
30
chamber. Inside the chamber was a light to see when setting up the sample trains and a
small stand to set the pumps on during operation. Figure 3.7 shows an image of the
actual chamber used.
Figure 3.7: Actual Chamber Used in the Study
3.3 Experimental Design
A multi-factorial analysis design was used for this study to account for the
different levels of each contributing factor in the experiment. This design was developed
in consultation with James Slaven (Biostatistics and Epidemiology Branch, HELD,
NIOSH). There were five effects, and each effect was broken into two levels as seen in
Table 3.3.
31
Table 3.3: Effects Examined in Multi-Factorial Design
Effect / Factor Level 1 ( - ) Level 2 ( + )
Vapor Low Boiling Point
(n-hexane)
High Boiling Point
(m-xylene)
Pump Type Pulsating Continuous
Exposure Profile Variable Constant*
Flow Rate Low (~30 mL/min) High (~200 mL/min)
Mannequin Yes No (free hanging)
*For constant exposure profile experiments short and long term samples were collected
A resolution V design was used to yield results so that all of the main effects or
two-factor interactions would not be aliased with any other main effect or two-factor
interaction. This type of design gives a study-design matrix with 16 experimental runs
noted in Table 3.4.
The last factor examined was the type of sample holder used. The three types of
sample holders were an open sorbent tube, a SKC brand sample holder, and a Buck brand
sample holder. To study the effect of each sample holder, two duplicate sampling units
for each of the three sample holder types were used in every experimental run resulting in
six measured concentrations per run. For example, Run 1 from Table 3.4 would include
n-hexane, a pulsating pump, variable concentration exposure profile, a low sampling flow
rate, and mounting the sampler on a mannequin. This experimental run will then have
six samplers, two each for the three sample holder types. The study design matrix is
given in Table 3.4, but the actual run order was randomized to transform any systematic
effects that could not be controlled.
32
Table 3.4: Study Design Matrix
Run Vapor Pump Exposure Flow Mannequin Long and Short
Duration Samples
1 - - - - + no
2 + - - - - no
3 - + - - - no
4 + + - - + no
5 - - + - - yes
6 + - + - + yes
7 - + + - + yes
8 + + + - - yes
9 - - - + - no
10 + - - + + no
11 - + - + + no
12 + + - + - no
13 - - + + + yes
14 + - + + - yes
15 - + + + - yes
16 + + + + + yes
When the exposure profile was constant, an additional factor, duration, was
included as a blocking effect. A regular run, whether it is variable or constant exposure,
is 81-minutes long. To identify whether sampling time affects the concentrations
measured with and without a sample holder, short sampling time data was collected just
for the constant exposure profile. The short experimental runs were 33-minutes long, and
there were 8 of these, which resulted in a combined 24 experimental runs for the entire
project.
33
3.4 Experimental Procedure
A test-atmosphere generation system and exposure chamber were used for the
evaluation of the sample holders and are depicted in Figure 3.6. The vapor was generated
using a gas washing bottle containing either n-hexane or m-xylene in a water bath kept at
a constant temperature. The temperature for each vapor was determined by trial runs to
get both vapor concentrations approximately 15 ppm. The mannequin or free hanging
sampler was placed in the exposure chamber according to the planned sampling designs.
The six different sample holders were also randomly placed two inches apart forming a
straight line across the mannequin or free hanging sampler for each scenario. The sample
holders were randomized by assigning each sample train an ID and using a random
number generator to choose the order they were placed on the mannequin or free hanging
stand. Figure 3.8 shows the mannequin and free hanging stand used with the six
sampling trains attached. The sample holders were oriented on the mannequin as they
would be in an industrial sampling scenario, and on each stand the level of air entry was
adjusted to be approximately four feet off the ground.
34
Figure 3.8: Mannequin and Free Hanging Stand
For constant exposure profiles the bubbler was turned on with a set flow rate and
allowed to run for 80 minutes for long runs and 30 minutes for short runs. A typical
exposure profile is presented in Figure 3.9. The variable exposure profile was generated
by turning on and off the compressed air through the bubbler every ten minutes to
produce a concentration profile similar to that shown in Figure 3.10.
Figure 3.9: Constant Concentration Exposure Profile for Vapors
0
4
8
12
16
0 10 20 30 40 50 60 70 80
Co
nce
ntr
atio
n (
PP
M)
Time (min)
35
Figure 3.10: Variable Concentration Exposure Profile for Vapors
Both types of air sampling pumps were pre- and post-calibrated using a Drycal for
every experiment. After the pumps were pre-calibrated for each run, a calibrated PID
was placed in the chamber to record the vapor concentration profile, which could be later
plotted for analysis and compared to sample results. The pumps were each given a
specific sample holder and a labeled carbon sorbent tube that was placed on each pump
connected by a rubber hose. The sample trains were then placed in the chamber and
either attached to the mannequin as they would be in real practice or placed on the free
hanging stand. An electronic thermometer was also placed in the chamber, and the
beginning and ending temperature and humidity was recorded. The selected test
atmosphere and pumps were then turned on for the duration of the experimental run.
Calibration standards and quality control samples were prepared and analyzed
with each set of experimental samples. The sorbent tubes were analyzed on site using a
Gas Chromatograph and Flame Ionization Detector. All statistical tests were compared
to a alpha value of 0.05. Once the results were obtained, a data summary table was sent
to James Slaven (Biostatistician, NIOSH) for statistical analysis.
0
5
10
15
20
25
30
0 20 40 60 80
Co
nce
ntr
atio
n (
PP
M)
Time (min)
36
Chapter 4
Results and Discussion
Sample holders are used every day by Industrial Hygienists when taking personal
samples in a workplace. No research has been presented on whether or not these sample
holders could be affecting measured concentrations. The purpose of this study was to
evaluate whether or not sample holders, when operating under specific conditions, will
give significantly different measured concentrations. The study first examined the
individual factors affecting the measured concentrations. Secondly, a test using a ratio of
the sample holder results combined with the open sample tube concentrations was used to
determine any differences in sample holder concentrations. Lastly, the study looks at
how a PID measuring real-time concentrations compares to actual measured samples.
4.1 Individual Factor Results Using Measured Concentrations
Individual effects on concentration were analyzed using the multi-factorial design.
Each effect had two levels, excluding the sample holder and positioning in the chamber,
which had three and six levels, respectively. The mean concentration for each level was
compared to the opposite levels (e.g. high vs. low flow rate) using an Analysis of
Variance. Table 4.1 summarizes the ANOVA results for every factor using the
measured concentrations. The three variables that gave significant results were the
duration of the experiment, the type of pump used, and the vapor that was used. All three
factors had p-values less than 0.0001.
37
Table 4.1: Analysis of Variance Results for Concentration (significant factors in bold)
Source DF Seq SS Adj SS Adj MS F p
Vapor 1 52.032 52.032 52.032 35.51 0.000
Pump Type 1 29.666 29.666 29.266 20.25 0.000
Exposure Profile 1 30.329 0.920 0.920 0.63 0.429
Flow Rate 1 0.039 0.039 0.039 0.03 0.871
Mannequin 1 1.392 1.392 1.392 0.95 0.331
Duration 1 58.064 58.064 58.064 39.63 0.000
Holder 2 5.474 4.880 2.440 1.67 0.193
Position 5 0.262 0.262 0.052 0.04 0.999
Error 130 190.465 190.465 1.465
Total 143 367.723
It can be noted that the sample holders had no significant difference when testing
the concentrations. Figure 4.1 is a main effects plot that shows how the concentrations
for each factor compare to one another. It shows the mean concentration for each level
and compares it to the opposite level. For example, when looking at the vapor factor, the
mean concentration of n-hexane was lower than the mean concentration of m-xylene.
38
Me
an
of
Co
nce
ntr
ati
on
PP
M
Xy leneHexane
14.4
13.6
12.8
PulsatingC ontinuous V ariableC onstant
LowHigh
14.4
13.6
12.8
YesNo 83 Min33 Min
SKCO pen TubeBuck
14.4
13.6
12.8
654321
V apor Pump Ty pe Exp. Profile
F low Rate Mannequin Duration
Holder Position
Main Effects Plot (data means) for Concentration PPM
Figure 4.1: ANOVA Results for Concentration (PPM) as a Response Variable
Another ANOVA was then done using the concentrations, but this time including
interaction terms using the three significant factors from Table 4.1. These results along
with an interaction plot can be seen in Table 4.2 and Figure 4.2. The only interaction
term that was significant was the vapor*duration term with a p-value equal to 0.000. An
interaction plots is a graph that measure the interaction between the means of two factors,
with the second factor held constant. Parallel lines in an interaction plot indicates no
interaction, while the larger the slopes of the two lines become, the higher the degree of
interaction.
39
Table 4.2: Analysis of Variance for Concentration with Interaction Terms
Source DF Seq SS Adj SS Adj MS F p
Vapor 1 52.032 52.032 52.032 90.87 0.000
Pump Type 1 29.666 29.666 29.266 21.80 0.000
Exposure Profile 1 30.329 0.920 0.920 0.94 0.335
Flow Rate 1 0.039 0.039 0.039 0.04 0.843
Mannequin 1 1.392 1.392 1.392 1.42 0.236
Duration 1 58.064 58.064 58.064 59.12 0.000
Holder 2 5.474 4.880 2.440 2.48 0.087
Position 5 0.262 0.262 0.052 0.05 0.998
Vapor*Pump Type 1 0.397 0.397 0.397 0.40 0.526
Vapor*Duration 1 63.019 63.019 63.019 64.17 0.000
Pump Type*Duration 1 2.326 2.326 2.326 2.37 0.126
Error 127 190.465 124.723 0.982
Total 143 367.723
VaporVapor
15.0
13.5
12.0
Pump T ypePump T ype
DurationDuration
83 Min33 Min
PulsatingC ontinuous
15.0
13.5
12.0
Xy leneHexane
15.0
13.5
12.0
Vapor
Hexane
Xylene
Pump Type
Continuous
Pulsating
Duration
33 Min
83 Min
Interaction Plot (data means) for Concentration PPM
Figure 4.2: Interaction Plot for Vapor, Pump Type, and Duration Factors
40
For the duration results, the longer the run was the higher the mean concentration
was in the chamber. This effect most likely happened because the concentration in the
chamber would slowly drift higher the longer the sample ran, and this drifting could not
be controlled. The concentrations were attempted to be balanced out using the LabVIEW
software and slowly decreasing the flow rate through the bubbler as time progressed, but
doing this only added small fluctuations in the concentration profile. During the
preliminary test runs, the profiles were designed to average 15 ppm over the sampling
time. Figure 4.3 shows the PID concentration in the chamber for a long run (81 min).
Over the 81-minute experiment, the vapor concentration continues to rise until the end of
the experiment when the gas generation system was turned off, yet it yielded an average
concentration of 15.6 PPM.
Figure 4.3: Measured PID Concentration for Run 4L in Experimental Design.
The graph shows the constant drift in concentration as time progressed.
0
2
4
6
8
10
12
14
16
18
20
0 10 20 30 40 50 60 70 80
Concentr
ation (
PP
M)
Time (Minutes)
41
The pump type had a significant effect on the concentration, resulting in the
continuous pump giving an overall higher mean concentration than the pulsating pump.
At first, differences in flow rates between the pumps were thought to be causing this, but
flow rates were shown to provide almost the exact same mean concentration. Evidence
for this pump type effect could not be determined.
The vapor factor concentrations may have differed due to the higher volatility of
n-hexane when compared with m-xylene. Since n-hexane has a higher volatility than m-
xylene, one would believe that n-hexane would reach equilibrium faster in the chamber
resulting in higher concentrations. However, the interaction plot above shows n-hexane
has a lower average concentration than m-xylene. The time it took for both vapor
concentrations to reach approximate equilibrium has been plotted in Figure 4.4. It is
difficult to tell actually when equilibrium was reached because of the constant drift in the
chamber, but from the graphs it can be noted that around six minutes into n-hexane's run
the concentration started flattening out. M-xylene's concentration started to flatten out
after approximately eight minutes. Both experiments were 81-minute runs, and the data
shows that n-hexane does indeed reach equilibrium faster than m-xylene. These results
would lead one to believe that n-hexane should have a higher average concentration than
m-xylene, which was not the case. The decay of concentration was then examined to see
how fast each vapor exited the chamber once the vapor generation system was turned off.
42
Figure 4.4: Concentration Rise of Both Vapors in the Experimental Chamber
During phases in the experiment when the gas generation system would be turned
off, the PID showed the concentration of hexane would drop faster than xylene. This
could cause n-hexane to have a much lower overall mean concentration, and would
explain the significant differences between the two concentrations. Figure 4.5 shows
how fast the concentration of n-hexane and m-xylene would decrease after the vapor
generation system was turned off. N-hexane would reach approximately 0 ppm about
two minutes faster than m-xylene. Volatility is one possible explanation for the
differences in vapor concentrations.
0
2
4
6
8
10
12
14
0 5 10
Co
nce
ntr
atio
n (
PP
M)
Time (Minutes)
m-Xylene
0
2
4
6
8
10
12
14
16
0 5 10
Co
nce
ntr
atio
n (
PP
M)
Time (Minutes)
n-Hexane
43
Figure 4.5: Rate of Decrease in Vapor Concentration
Lastly, when looking at the interaction terms from Table 4.2 the Vapor*Duration
interaction term was significantly different. This could also support the above statements
indicating a difference in volatility and time a sample was in the chamber. This
difference could be from each vapor taking longer to reach maximum in a chamber
(Figure 4.4), the vapor taking longer to dissipate from the chamber (Figure 4.5), or a
combination of both.
The significant differences when looking at just the concentrations of the vapors
is hard to designate whether they are significant, because the vapor generation system
differed from day-to-day experiments. All experiments were done during the summer
months when temperature and humidity were fluctuating, and this resulted in the chamber
producing different concentrations ranging from 12-16 PPM each day. It was decided to
use a ratio of concentrations for each separate day to further analyze the results using the
open tube sampler as a reference tube.
0
5
10
15
20
25
30
35
0 50 100 150 200
Co
nce
ntr
atio
n (
PP
M)
Time (Seconds)
Xylene
Hexane
44
4.2 Individual Factor Results Using a Ratio of Measured Concentrations
After noting these significant results, each effect was then tested using a ratio of
concentrations between the SKC or A.P. Buck sample holder and the open sample tube.
By using this ratio, the open sample tube becomes the reference tube concentration to
which the other two sample holders can be compared. By doing this, it negates the day-
to-day variations in the chamber concentrations. All factors, including the sample
holders, were then analyzed using an ANOVA test. Table 4.3 shows the results of the
ANOVA test, indicating flow rate and the sample holders to have significant differences
with p-values equal to 0.006 and 0.021, respectively. A main effects plot, Figure 4.6,
shows the graphical results of the ANOVA comparing the two levels of each factor.
Using this ratio method, the factors that were significant when looking at just the
concentrations are no longer significant, and this method shows that sample holders may
be causing differences in measured concentrations.
45
Table 4.3: Analysis of Variance Results for Ratio of Concentrations (significant factors in
bold)
Source DF Seq SS Adj SS Adj MS F p
Vapor 1 0.001 0.001 0.001 0.25 0.618
Pump Type 1 0.006 0.006 0.006 2.27 0.139
Exposure Profile 1 0.006 0.008 0.008 3.04 0.089
Flow Rate 1 0.023 0.023 0.023 8.43 0.006
Mannequin 1 0.000 0.000 0.000 0.08 0.782
Duration 1 0.002 0.002 0.002 0.85 0.361
Holder 1 0.015 0.015 0.015 5.77 0.021
Error 40 0.107 0.107 0.003
Total 47 0.160
Me
an
of
Ra
tio
Xy leneHexane
1.05
1.00
0.95
PulsatingC ontinuous V ariableC onstant
30 mL/min200 mL/min
1.05
1.00
0.95
YesNo 80 min30 min
SKCBuck
1.05
1.00
0.95
V apor Pump Ty pe Exposure Profile
F low Rate Mannequin Duration
Holder
Main Effects Plot (fitted means) for Ratio
Figure 4.6: ANOVA Results for Ratio of Concentrations as a Response Variable
46
Interaction terms between the two significant factors were also examined and can
be seen in Table 4.4. In this case, the interaction between flow rate and the type of sample
holder was not significant, but an interaction plot can also be seen in Figure 4.7 to
compare the two factors.
Table 4.4: Analysis of Variance Results for Ratio of Concentrations with Interaction
Terms
Source DF Seq SS Adj SS Adj MS F p
Vapor 1 0.001 0.001 0.001 0.25 0.618
Pump Type 1 0.006 0.006 0.006 2.27 0.139
Exposure Profile 1 0.006 0.008 0.008 3.04 0.089
Flow Rate 1 0.023 0.023 0.023 8.43 0.006
Mannequin 1 0.000 0.000 0.000 0.08 0.782
Duration 1 0.002 0.002 0.002 0.85 0.361
Holder 1 0.015 0.015 0.015 5.77 0.021
Flow Rate*Holder 1 0.003 0.003 0.003 1.26 0.269
Error 39 0.104 0.104 0.003
Total 47 0.160
47
Flow RateFlow Rate
HolderHolder
SKCBuck
1.06
1.04
1.02
1.00
0.98
30 mL/min200 mL/min
1.06
1.04
1.02
1.00
0.98
Flow Rate
200 mL/min
30 mL/min
Holder
Buck
SKC
Interaction Plot (data means) for Ratio
Figure 4.7: Interaction Plot for Flow Rate and Sample Holder
The flow rate was a significant factor in both sample holder cases and was
believed to be caused from a decrease in flow for the high flow rate experiments. In both
sample holder cases, the higher flow rate resulted in smaller ratios when compared to the
low flow rate. When pre- and post-calibrating for the low flow rate the pumps seemed to
stay consistent throughout the entire experiment. During high flow rate runs, it was noted
that sometimes the post calibration flow rate would drop 2% to 5%. Fluctuations of flow
rates of less than 5% is allowable according to NIOSH's methods. These differences in
flow rates could lead to smaller measured concentrations, resulting in the average flow
rate being smaller than it was intended to be. This drop in flow rate may have resulted
from batteries losing charge during the experiment. The pulsating pumps had
rechargeable battery packs that could be charged to full at the end of every day. The
48
continuous pumps used four AA batteries and would be swapped out every 3 runs. It is
possible that during the third run, diminishing battery lives began to affect the flow rate
of the pump.
Lastly, the ratio ANOVA test did show a significant difference between sample
holders, and this was one of the main points of this research. A closer look using a T-test
study will examine these differences in the next section.
4.3 Overall Sample Holder Results
The following results determined if either of the two sample holders were
measuring a significantly different concentration when compared to the open sample
holder. This was done by taking the ratio between the open holder concentration and the
other two sample holder concentrations. A T-test was performed to see if the
concentration ratios were one, meaning no difference in concentrations. Table 4.6 gives
the mean ratio of concentrations, standard deviation, and T-test p-value, which was
compared to a significant value of 0.05. The p-value for the ratio between open holder
and buck holder was 0.0382, indicating that the ratio of concentrations differ significantly
enough from one. The mean value of the ratio was 1.02, showing that the Buck holder
usually gave a larger concentration than the open holder. The opposite is true for the
mean ratio between the open holder and SKC holder. It had a smaller value than one, but
the p-value was 0.0651, indicating insignificance.
49
Table 4.5: Ratio of Open and Sample Holder
SKC Holder/Open Holder Buck Holder/Open Holder
Mean Ratio 0.99 1.02
Standard Deviation 0.04 0.08
T-test p-value 0.0651 0.0382
This first set of data shows that the Buck sample holder is in fact giving larger
readings than an open-faced sorbent tube. This table did not take into account any other
factors besides the sample holder used. Even though the Buck holder did give a
significant difference in concentration, the difference of 2% is small when compared to
other sources of error associated when sampling for vapors. NIOSH’s Occupational
Exposure Sampling Strategy Manual estimates the error associated with pump and
calibration is 5%. There is also error introduced when using the Gas Chromatograph and
Flame Ionization Detector that varies from 5% to 10% depending on the equipment used.
Overall, a method of measuring air concentration should be within ±25% of the true
concentration 95% of the time. There is error built into these sampling strategies, and the
small 2% from these results is not large enough to state that sample holders caused the
differences in measured concentrations.
The statistical results do say that the difference is significant, but in real world
industrial hygiene applications a difference of more than 10% would have been around
the significance needed to indicate that sample holders were having an impact on
measured concentrations.
50
4.4 Comparing Sampled Concentration vs. Measured PID Concentration
The data collected during this study gives a good opportunity to test the
performance of a PID. Depending on the PID, error may be introduced by temperature,
humidity, other gases present, and calibration error. Two tests were done to determine if
the measured PID concentrations were equal to the sample concentration results to
examine the PID accuracy. The first test performed was a two sample T-test comparing
the measured concentrations with the PID concentrations. Table 4.7 shows the results of
the test, and Figure 4.8 gives an interval plot to show the range of concentrations. The p-
value was 0.000, indicating that the means of the two concentrations significantly
differed.
Table 4.6: T-test Comparing Measured Concentration vs. PID Concentration
N Mean St. Dev. T-Value p-value
Measured Concentration 144 13.49 ppm 1.60 ppm 4.50 0.000
PID Concentration 144 14.34 ppm 1.59 ppm
51
Co
nce
ntr
ati
on
(P
PM
)
PID Conc.Concentration
15.0
14.5
14.0
13.5
13.0
Interval Plot of Concentration, PID Conc.95% CI for the Mean
Figure 4.8: Interval Plot Comparing Means of Measured Concentration and PID
Concentration
However, just as when studying the sample holders, day-to-day variations in the
vapor chamber concentrations could be affecting the means of both measured and PID
concentrations. A T-test was performed on the ratio (ratio = measured concentration/PID
concentration) of the two concentrations to examine if the PID and measured
concentrations were equal. Results of the T-test can be seen in Table 4.8. The 95%
confidence interval was equal to 0.93 to 0.97, and a p-value equal to 0.0001. This
indicates the ratio significantly differs from 1, and the PID concentrations are generally
larger than the sample concentration.
Table 4.7: One-Sample T-test Comparing Ratios of Measured and PID Concentrations
Variable N Mean St. Dev. 95% C.I.
Ratio of
Concentrations 144 0.95 0.13 0.93 - 0.97
52
Even though significantly different results have been shown for both T-tests
above, the PID typically measures within 5% of the measured sample concentration.
Since PID's are becoming more and more popular for sampling for gas and vapors in the
workplace, this result is promising that they are performing with respectable accuracy.
53
Chapter 5
Conclusions
The objective of this project was to study the effects sample holders have on gas
and vapor sampling with activated carbon sorbent tubes, and to identify factors that may
produce differences in concentrations when compared to a sample train without a sample
holder.
The results suggest that sample holders do affect measured concentrations, but
have such a small effect when considering the amount of error and uncertainty already
associated with the methods of sampling for vapors and gases. The A.P. Buck sample
holder measured a +2% difference in sample concentration when compared to an open
sorbent tube, while the SKC sample holder had a -1% difference. It cannot be
determined if this error came from the sample holder or is just part of the built-in error of
sampling and analytical error. It can be concluded that using a sample holder to protect a
worker from getting injured will not make a noticeable difference in the actual
concentration sampled.
Effects such as flow rate and pump type showed significant results when sampling
with sample holders, but again the difference is minimal and will not practically affect
the final concentrations being measured. It should also be noted that PID devices, such
as the MiniRAE 2000, give good representation to the actual concentrations of the air
being sampled.
54
References
Anderson, K., Levin, J.O., Lindahl, R., Nilsson, C.A. (1984). Influence of Air Humidity on
Sampling Efficiency of Some Solid Adsorbents Used for Sampling Organics from Work-
Room Air. Chemosphere , 13 (3), 437-444.
Crisp, S. (1980). Solid Sorbent Gas Samplers. The Annals of Occupational Hygiene , 47-76.
Cullis, C. F., Firth, J. G. (1981). Detection and Measurement of Hazardous Gases.
London: Heinemann.
Guenier, J., and Muller, J. (1984). Sampling of Gaseous Pollutants on Activated Charcoal with
900 mg Tubes. The Annals of Occupational Hygiene , 28 (1), 61-75.
Kennedy, E. R., Fischbach, T. J., Song, R., Eller, P. M., Shulman, S. A. (1996). Summary of
the NIOSH Guidelines for Air Sampling and Analytical Method Development and
Evaluation. The Analyst , 121, 1163-1169.
Lee, Larry, Flemmer, Michael, Lee, Eun Gyang, Harper, Martin, Lin, Ming-I, Groves, William,
Frievalds, Andris, Slaven, James. A Novel Physiologic Sampling Pump Capable of Rapid
Response Breathing. J. Environ. Monit., 2009, 11, 1020 - 1027, DOI: 10.1039/b816699d.
Leidel, N. A., Busch, K. A., Lynch, J. R. (1977). NIOSH. Occupational Exposure Sampling
Strategy Manual. Cincinnati: Superintendent of Documents.
Levin, J. O., Carleborg, L. (1987). Evaluation of Solid Sorbents for Sampling Ketones in Work-
Room Air. The Annals of Occupational Hygiene , 31 (1), 31-38.
Lin, Ming-I, Groves, William A., Freivalds, Andris., Lee, Eun Gyang, Harper, Martin, Slaven,
James E., and Lee, Larry. Exposure Assessment by Physiologic Sampling Pump -
Prediction of Minute Ventilation Using A Portable Respiratory Inductive
Plethysmograph System, J. Environ. Monit., 2008, 10, 1179 - 1186, DOI:
10.1039/b806292.
55
Lin, Ming-I, Groves, William A., Freivalds, Andris., Lee, Larry, Lee, Eun Gyang, , Slaven,
James E., and Harper, Martin. Laboratory Evaluation of a Physiologic Sampling Pump
(PSP). J. Environ. Monit., 2010, DOI: 10.1039/b923986c.
Mintz, Benjamin W. "History of the Federal Occupational Safety and Health
Administration."Fundamentals of Industrial Hygiene. By Barbara A. Plog and Patricia
Quinlan. Itasca: National Safety Council, 2002. 825-26.
Occupational Safety and Health Act of 1970: the OSH Act. Des Plaines, IL: U.S. Dept. of Labor,
Occupational Safety and Health Administration, Office of Training and Education, 1987.
Peach, M. J., Carr, W. G. (1986). Air Sampling and Analysis for Gases and Vapors. In
Occupational Respiratory Diseases (pp. 41-68).
Perkins, J. L. (1997). Modern Industrial Hygiene. New York: Van Nostrand Reinhold.
Rappaport, S. M., Kupper, L. L. (2008). Quantitative Exposure Assessment. El Cerrito, CA:
Stephen Rappaport.
White, L. D., Taylor, D. G., Mauer, P. A., Kupel, R. E. (1970). A Convienient Optimized Method
of the Analysis of Selected Solvent Vapors in the Industrial Atomsphere. American
Industrial Hygiene Association Journal , 31 (2), 225-232.
56
Appendix A
Raw Data
The following pages are the data sheets recorded for every experiment.
57
Experiment #__1__
Date ___7/02/09____
Vapor: Hexane
Pump Type: Pulsating / Continuous
Exposure: Constant / Variable
Pump Flow Rate: Low / High
Mannequin: Yes / No
Sampling Time: Short / Long
Pump ID & Sample Holder Type
CA
SKC 2
IM
Buck 1
CM
Open 2
83
SKC 1
IA
Open 1
82
Buck 2
Pre-Cal Flowrate
29.5 29.9 29 28.5 30.8 29.7
Post-Cal Flowrate
29.5 30 28.9 28.4 31.1 29.6
Initial Stroke Count
793082 95358 757270 626104 56931 845116
Final Stroke Count
798687 100453 762600 630896 62345 850146
Time Sampled 81:15
Charcoal Tube # 1 2 3 4 5 6
PID Average Concentration: ____16 PPM_____
Notes: Temperature: 70°F
Humidity: 17%
58
Experiment #__2__
Date ____7/06/09____
Vapor: ___Xylene____
Pump Type: Pulsating / Continuous
Exposure: Constant / Variable
Pump Flow Rate: Low / High
Mannequin: Yes / No Sampling Time:
Short / Long
Pump ID & Sample Holder
Type
83 SKC 2
CM Open 2
IM Buck 2
CA Open 1
IA SKC 1
82 Buck 1
Pre-Cal Flowrate
28.1 28.9 29.6 29.3 31 29.5
Post-Cal Flowrate
28.1 29 30 29.4 31.1 29.6
Initial Stroke Count
631138 762874 100989 799122 62581 850796
Final Stroke Count
635868 768198 106081 804717 67973 856001
Time Sampled 81:19
Charcoal Tube # 1 2 3 4 5 6
PID Average Concentration: ____16.9 PPM___
Notes: Temperature: 73°F
Humidity: 16%
59
Experiment #_3S_
Date ____7/07/09____
Vapor: ____Hexane____
Pump Type: Pulsating / Continuous
Exposure: Constant / Variable
Pump Flow Rate: Low / High
Mannequin: Yes / No
Sampling Time: Short / Long
Pump ID & Sample Holder
Type
B SKC 1
E Buck 2
D Open 2
A SKC 2
C Buck 1
F Open 1
Pre-Cal Flowrate 196.3 201.9 200 201.3 198.5 202.1
Post-Cal Flowrate
203.2 206.5 194.8 189.3 197.8 202.2
Initial Stroke Count
Final Stroke Count
Time Sampled 33:16
Charcoal Tube # 1S 2S 3S 4S 5S 6S
PID Average Concentration: ____12.1 PPM____
Notes: Temperature: 72°F
Humidity: 16%
60
Experiment #_3L_
Date ____7/07/09____
Vapor: ____Hexane____
Pump Type: Pulsating / Continuous
Exposure: Constant / Variable
Pump Flow Rate: Low / High
Mannequin: Yes / No
Sampling Time: Short / Long
Pump ID & Sample Holder
Type
B SKC 1
E Buck 2
D Open 2
A SKC 2
C Buck 1
F Open 1
Pre-Cal Flowrate 198.6 200 200.7 201.2 200.8 198.5
Post-Cal Flowrate
200 199.1 199.6 199 200.1 199.2
Initial Stroke Count
Final Stroke Count
Time Sampled 83:50
Charcoal Tube # 1L 2L 3L 4L 5L 6L
PID Average Concentration: ____13.7 PPM___
Notes: Temperature: 73°F
Humidity: 22%
61
Experiment #_4S_
Date ____7/13/09____
Vapor: ___Hexane____
Pump Type: Pulsating / Continuous
Exposure: Constant / Variable
Pump Flow Rate: Low / High
Mannequin: Yes / No
Sampling Time: Short / Long
Pump ID & Sample Holder
Type
D SKC 2
C Buck 1
B Open 2
E Buck 2
F SKC 1
A Open 1
Pre-Cal Flowrate 30.4 30 30.5 31.7 28.9 29.1
Post-Cal Flowrate 30.3 30.4 30.8 32.6 29 28.9
Initial Stroke Count
Final Stroke Count
Time Sampled 33:07
Charcoal Tube # 1S 2S 3S 4S 5S 6S
PID Average Concentration: ____13.1 PPM___
Notes: Temperature: 72°F
Humidity: 16%
62
Experiment #_4L_
Date ____7/13/09____
Vapor: ___Hexane____
Pump Type: Pulsating / Continuous
Exposure: Constant / Variable
Pump Flow Rate: Low / High
Mannequin: Yes / No
Sampling Time: Short / Long
Pump ID & Sample Holder
Type
D SKC 2
C Buck 1
B Open 2
E Buck 2
F SKC 1
A Open 1
Pre-Cal Flowrate 30.3 30.4 30.8 32.6 29 28.9
Post-Cal Flowrate 30.1 29.9 30.7 33.3 28.7 28.6
Initial Stroke Count
Final Stroke Count
Time Sampled 84:00
Charcoal Tube # 1L 2L 3L 4L 5L 6L
PID Average Concentration: ___15.6 PPM____
Notes: Temperature: 73°F
Humidity: 17%
63
Experiment #__5__
Date ___7/14/09____
Vapor: ____Xylene____
Pump Type: Pulsating / Continuous
Exposure: Constant / Variable
Pump Flow Rate: Low / High
Mannequin: Yes / No
Sampling Time: Short / Long
Pump ID & Sample Holder
Type
C Buck 1
B SKC 2
D Open 2
A Open 1
F SKC 1
E Buck 2
Pre-Cal Flowrate 30.8 31.8 28.6 29 32 29.4
Post-Cal Flowrate 29.6 31.3 28.1 28.2 31.9 29.1
Initial Stroke Count
Final Stroke Count
Time Sampled 81:05
Charcoal Tube # 1 2 3 4 5 6
PID Average Concentration: ___15.9 PPM___
Notes: Temperature: 75°F
Humidity: 15%
64
Experiment #__6__
Date ___7/14/09____
Vapor: ____Hexane____
Pump Type: Pulsating / Continuous
Exposure: Constant / Variable
Pump Flow Rate: Low / High
Mannequin: Yes / No
Sampling Time: Short / Long
Pump ID & Sample Holder
Type
IA SKC 1
82 Buck 1
83 SKC 2
IM Buck 2
CA Open 2
CM Open 1
Pre-Cal Flowrate
201.4 201.4 201.4 200.8 197.5 201.7
Post-Cal Flowrate
197.9 199.4 200 200.9 197.6 199.6
Initial Stroke Count
69041 857021 637153 107052 807196 769923
Final Stroke Count
105091 893224 674477 141719 843858 806197
Time Sampled 81:02
Charcoal Tube #
1 2 3 4 5 6
PID Average Concentration: ____16.8 PPM___
Notes: Temperature: 72°F
Humidity: 16%
65
Experiment #__7__
Date ____7/15/09_____
Vapor: ____Hexane____
Pump Type: Pulsating / Continuous
Exposure: Constant / Variable
Pump Flow Rate: Low / High
Mannequin: Yes / No
Sampling Time: Short / Long
Pump ID & Sample Holder
Type
B SKC 1
C Buck 1
E Buck 2
F Open 1
D Open 2
A SKC 2
Pre-Cal Flowrate 31.5 30.2 29.7 31.6 29.3 28.8
Post-Cal Flowrate 31.4 29.6 29.5 31.3 28.8 28.2
Initial Stroke Count
Final Stroke Count
Time Sampled 81:00
Charcoal Tube # 1 2 3 4 5 6
PID Average Concentration: ___16.1 PPM___
Notes: Temperature: 72°F
Humidity: 16%
66
Experiment #__8__
Date ___7/16/09____
Vapor: ____Xylene____
Pump Type: Pulsating / Continuous
Exposure: Constant / Variable
Pump Flow Rate: Low / High
Mannequin: Yes / No
Sampling Time: Short / Long
Pump ID & Sample Holder
Type
82 Buck 2
83 Open 1
IM Buck 1
CA SKC 1
IA Open 2
CM SKC 2
Pre-Cal Flowrate
200.4 199.7 201.4 201.2 200.3 200.4
Post-Cal Flowrate
199.2 195.5 201 198.4 200.3 200.5
Initial Stroke Count
894377 676013 142747 845211 107261 808282
Final Stroke Count
930586 714374 177585 882956 144223 844842
Time Sampled 81:14
Charcoal Tube #
1 2 3 4 5 6
PID Average Concentration: ___15.5 PPM____
Notes: Temperature: 75°F
Humidity: 35%
67
Experiment #__9__
Date ____7/16/09_____
Vapor: ____Xylene____
Pump Type: Pulsating / Continuous
Exposure: Constant / Variable
Pump Flow Rate: Low / High
Mannequin: Yes / No
Sampling Time: Short / Long
Pump ID & Sample Holder
Type
C Buck 1
F Open 1
D Open 2
B SKC 1
E Buck 2
A SKC 2
Pre-Cal Flowrate 200.4 198.5 199.8 201.1 200.2 200.6
Post-Cal Flowrate
200.2 201.5 199 202.1 199.9 198.7
Initial Stroke Count
Final Stroke Count
Time Sampled 81:15
Charcoal Tube # 1 2 3 4 5 6
PID Average Concentration: ___14.5 PPM____
Notes: Temperature: 73°F
Humidity: 18%
68
Experiment #_10S_
Date ___7/20/09____
Vapor: ____Xylene____
Pump Type: Pulsating / Continuous
Exposure: Constant / Variable
Pump Flow Rate: Low / High
Mannequin: Yes / No
Sampling Time: Short / Long
Pump ID & Sample Holder
Type
IM Buck 2
CA SKC 2
CM SKC 1
82 Buck 1
IA Open 2
83 Open 1
Pre-Cal Flowrate 31.7 31.5 30.3 30.4 28.5 31.4
Post-Cal Flowrate 31.8 31.7 30.4 30.3 28.3 31.5
Initial Stroke Count
178085 883952 845358 931343 145151 715538
Final Stroke Count
180283 886229 847557 933447 147030 717706
Time Sampled 33:00
Charcoal Tube # 1S 2S 3S 4S 5S 6S
PID Average Concentration: ____16.3 PPM___
Notes: Temperature: 73°F
Humidity: 16%
69
Experiment #_10L_
Date ____7/20/09____
Vapor: ____Xylene____
Pump Type: Pulsating / Continuous
Exposure: Constant / Variable
Pump Flow Rate: Low / High
Mannequin: Yes / No
Sampling Time: Short / Long
Pump ID & Sample Holder
Type
IM Buck 2
CA SKC 2
CM SKC 1
82 Buck 1
IA Open 2
83 Open 1
Pre-Cal Flowrate 31.8 31.7 30.4 30.3 28.3 32
Post-Cal Flowrate 31.9 31.9 30.2 30.3 28.4 31.8
Initial Stroke Count
180397 886354 847703 933614 147146 717957
Final Stroke Count
186014 892174 853290 938987 151926 723474
Time Sampled 84:00
Charcoal Tube # 1L 2L 3L 4L 5L 6L
PID Average Concentration: ___14.8 PPM___
Notes: Temperature: 73°F
Humidity: 16%
70
Experiment #_11S_
Date ____7/22/09____
Vapor: ____Xylene_____
Pump Type: Pulsating / Continuous
Exposure: Constant / Variable
Pump Flow Rate: Low / High
Mannequin: Yes / No
Sampling Time: Short / Long
Pump ID & Sample Holder
Type
E Buck 2
D Open 1
B Open 2
F SKC 2
A SKC 1
C Buck 1
Pre-Cal Flowrate 198.9 200.9 199.8 200.2 199.3 198.1
Post-Cal Flowrate
197.9 198.6 201 198.1 197.1 197.4
Initial Stroke Count
Final Stroke Count
Time Sampled 33:00
Charcoal Tube # 1S 2S 3S 4S 5S 6S
PID Average Concentration: __12.0 PPM___
Notes: Temperature: 73°F
Humidity: 23%
71
Experiment #_11L_
Date ___7/22/09____
Vapor: ____Xylene____
Pump Type: Pulsating / Continuous
Exposure: Constant / Variable
Pump Flow Rate: Low / High
Mannequin: Yes / No
Sampling Time: Short / Long
Pump ID & Sample Holder
Type
E Buck 2
D Open 1
B Open 2
F SKC 2
A SKC 1
C Buck 1
Pre-Cal Flowrate 201.1 200.4 201.1 199.7 200.4 199
Post-Cal Flowrate
196.3 200 199.9 198.6 198.3 187.4
Initial Stroke Count
Final Stroke Count
Time Sampled 84:55
Charcoal Tube # 1L 2L 3L 4L 5L 6L
PID Average Concentration: ___12.3 PPM____
Notes: Temperature: 71°F
Humidity: 25%
Pump C had a drop in flow rate of more than 5%
72
Experiment #_12S_
Date ___7/23/09___
Vapor: ___Xylene_____
Pump Type: Pulsating / Continuous
Exposure: Constant / Variable
Pump Flow Rate: Low / High
Mannequin: Yes / No
Sampling Time: Short / Long
Pump ID & Sample Holder
Type
B Open 2
D SKC 1
C Buck 1
F Open 1
E Buck 2
A SKC 2
Pre-Cal Flowrate 28.1 32.3 29 28.9 30.3 31.1
Post-Cal Flowrate 28.3 32.5 28 28.6 29.4 30.5
Initial Stroke Count
Final Stroke Count
Time Sampled 33:00
Charcoal Tube # 1S 2S 3S 4S 5S 6S
PID Average Concentration: ___13.3 PPM___
Notes: Temperature: 72°F
Humidity: 23%
73
Experiment #_12L_
Date ____7/23/09____
Vapor: ____Xylene____
Pump Type: Pulsating / Continuous
Exposure: Constant / Variable
Pump Flow Rate: Low / High
Mannequin: Yes / No
Sampling Time: Short / Long
Pump ID & Sample Holder
Type
B Open 2
D SKC 1
C Buck 1
F Open 1
E Buck 2
A SKC 2
Pre-Cal Flowrate 28.3 32.5 28.2 28.6 28.3 30.5
Post-Cal Flowrate 28.3 32.5 28.2 28.2 28.1 30.2
Initial Stroke Count
Final Stroke Count
Time Sampled 83:00
Charcoal Tube # 1L 2L 3L 4L 5L 6L
PID Average Concentration: ___14.8 PPM___
Notes: Temperature: 73°F
Humidity: 32%
74
Experiment #_13S_
Date ___7/27/09____
Vapor: ____Hexane____
Pump Type: Pulsating / Continuous
Exposure: Constant / Variable
Pump Flow Rate: Low / High
Mannequin: Yes / No
Sampling Time: Short / Long
Pump ID & Sample Holder
Type
IA SKC 2
82 Buck 1
CA Open 2
83 SKC 1
CM Open 1
IM Buck 2
Pre-Cal Flowrate 30.9 30.3 28.7 30.6 29.3 29.8
Post-Cal Flowrate 31.6 31.2 28.9 31.4 29.1 30
Initial Stroke Count
152552 939314 892544 723824 990427 186439
Final Stroke Count
154770 941509 894596 726082 992719 188584
Time Sampled 34:00
Charcoal Tube # 1S 2S 3S 4S 5S 6S
PID Average Concentration: ___14.8 PPM___
Notes: Temperature: 70°F
Humidity: 24%
75
Experiment #_13L_
Date ___7/27/09____
Vapor: ____Hexane____
Pump Type: Pulsating / Continuous
Exposure: Constant / Variable
Pump Flow Rate: Low / High
Mannequin: Yes / No
Sampling Time: Short / Long
Pump ID & Sample Holder
Type
IA SKC 2
82 Buck 1
CA Open 2
83 SKC 1
CM Open 1
IM Buck 2
Pre-Cal Flowrate 32 31.4 28.9 31.9 29.1 30
Post-Cal Flowrate
31.3 30.6 28.9 31.8 28.4 30.7
Initial Stroke Count
154877 941624 894661 726254 992833 188691
Final Stroke Count
160237 946940 899655 731758 998390 193900
Time Sampled 83:15
Charcoal Tube # 1L 2L 3L 4L 5L 6L
PID Average Concentration: ___12.8 PPM____
Notes: Temperature: 72°F
Humidity: 22%
76
Experiment #_14_
Date ____7/28/09____
Vapor: ____Hexane____
Pump Type: Pulsating / Continuous
Exposure: Constant / Variable
Pump Flow Rate: Low / High
Mannequin: Yes / No
Sampling Time: Short / Long
Pump ID & Sample Holder
Type
E Buck 2
C Buck 1
B SKC 2
D Open 2
F SKC 1
A Open 1
Pre-Cal Flowrate 198.4 200.1 198.6 199.5 199.5 199.3
Post-Cal Flowrate
197 198.3 198.6 199.1 199 198
Initial Stroke Count
Final Stroke Count
Time Sampled 81:15
Charcoal Tube # 1 2 3 4 5 6
PID Average Concentration: __14.4 PPM___
Notes: Temperature: 73°F
Humidity: 22%
77
Experiment #_15S_
Date ____7/29/09____
Vapor: ____Xylene____
Pump Type: Pulsating / Continuous
Exposure: Constant / Variable
Pump Flow Rate: Low / High
Mannequin: Yes / No
Sampling Time: Short / Long
Pump ID & Sample Holder
Type
IM Buck 2
CA Open 1
82 Buck 1
IA Open 2
CM SKC 2
83 SKC 1
Pre-Cal Flowrate 200.1 200.2 198 199.5 199.4 201.5
Post-Cal Flowrate
199.7 197.5 197.2 198.1 198.4 197.7
Initial Stroke Count
196167 900882 948176 161788 999038 733008
Final Stroke Count
210027 916636 962508 176894 1014317 749067
Time Sampled 33:15
Charcoal Tube # 1S 2S 3S 4S 5S 6S
PID Average Concentration: ___15.2 PPM____
Notes: Temperature: 73°F
Humidity: 32%
78
Experiment #_15L_
Date ___7/29/09___
Vapor: ____Xylene____
Pump Type: Pulsating / Continuous
Exposure: Constant / Variable
Pump Flow Rate: Low / High
Mannequin: Yes / No
Sampling Time: Short / Long
Pump ID & Sample Holder
Type
IM Buck 2
CA Open 1
82 Buck 1
IA Open 2
CM SKC 2
83 SKC 1
Pre-Cal Flowrate 200.5 200.8 199.1 198.7 198 198.4
Post-Cal Flowrate
200.8 202 201.3 200.1 200.1 198
Initial Stroke Count
210796 917573 963464 177416 15074 750232
Final Stroke Count
246131 956808 1000086 215449 53529 789281
Time Sampled 83:30
Charcoal Tube # 1L 2L 3L 4L 5L 6L
PID Average Concentration: ___12.5 PPM___
Notes: Temperature: 73°F
Humidity: 31%
79
Experiment #_16S_
Date ___7/30/09___
Vapor: ____Hexane____
Pump Type: Pulsating / Continuous
Exposure: Constant / Variable
Pump Flow Rate: Low / High
Mannequin: Yes / No
Sampling Time: Short / Long
Pump ID & Sample Holder
Type
83 SKC 2
82 Buck 2
CM SKC 1
IA Open 2
IM Buck 1
CA Open 1
Pre-Cal Flowrate
199 199.4 200.8 200 200.4 200.3
Post-Cal Flowrate
197.2 199.1 200.9 200.1 200.8 198.8
Initial Stroke Count
790590 962 54975 216883 246797 957563
Final Stroke Count
806819 15665 71002 232224 260960 973370
Time Sampled 33:15
Charcoal Tube # 1S 2S 3S 4S 5S 6S
PID Average Concentration: __12.0 PPM____
Notes: Temperature: 72°F
Humidity: 24%
80
Experiment #_16L_
Date ____7/30/09____
Vapor: ____Hexane____
Pump Type: Pulsating / Continuous
Exposure: Constant / Variable
Pump Flow Rate: Low / High
Mannequin: Yes / No
Sampling Time: Short / Long
Pump ID & Sample Holder
Type
83 SKC 2
82 Buck 2
CM SKC 1
IA Open 2
IM Buck 1
CA Open 1
Pre-Cal Flowrate 199 199 200.8 198.8 200.8 198.8
Post-Cal Flowrate 192.4 198.3 198.8 197.9 200.8 196.9
Initial Stroke Count
807462 15948 71750 232871 261429 973717
Final Stroke Count
846906 52728 111079 270994 296948 1012814
Time Sampled 83:30
Charcoal Tube # 1L 2L 3L 4L 5L 6L
PID Average Concentration: ___12.8 PPM___
Notes: Temperature: 72°F
Humidity: 24%
81
Table A1: Raw Data With Individual Sample Results and Sampler Positions on Mannequin or Stand
Run Conc.
(ppm) Vapor Pump Type
Exp.
Profile Flow Rate Mannequin Duration Holder Position
PID Conc.
(ppm) Temp. °F RH %
1 14.6 Hex Pulse Var Low No 83 Min SKC 1
16.0 70 17
1 14.1 Hex Pulse Var Low No 83 Min Buck 2
1 13.9 Hex Pulse Var Low No 83 Min Open 3
1 14.2 Hex Pulse Var Low No 83 Min SKC 4
1 14.0 Hex Pulse Var Low No 83 Min Open 5
1 13.5 Hex Pulse Var Low No 83 Min Buck 6
2 13.7 Xyl Pulse Var Low Yes 83 Min SKC 1
16.9 73 16
2 12.5 Xyl Pulse Var Low Yes 83 Min Open 2
2 14.8 Xyl Pulse Var Low Yes 83 Min Buck 3
2 13.5 Xyl Pulse Var Low Yes 83 Min Open 4
2 14.4 Xyl Pulse Var Low Yes 83 Min SKC 5
2 13.3 Xyl Pulse Var Low Yes 83 Min Buck 6
3S 11.1 Hex Cont Const High Yes 33 Min SKC 1
12.1 72 16
3S 11.4 Hex Cont Const High Yes 33 Min Buck 2
3S 10.6 Hex Cont Const High Yes 33 Min Open 3
3S 10.5 Hex Cont Const High Yes 33 Min SKC 4
3S 11.2 Hex Cont Const High Yes 33 Min Buck 5
3S 11.6 Hex Cont Const High Yes 33 Min Open 6
3L 13.7 Hex Cont Const High Yes 83 Min SKC 1
13.7 72 16
3L 14.2 Hex Cont Const High Yes 83 Min Buck 2
3L 14.2 Hex Cont Const High Yes 83 Min Open 3
3L 14.2 Hex Cont Const High Yes 83 Min SKC 4
3L 14.6 Hex Cont Const High Yes 83 Min Buck 5
3L 14.9 Hex Cont Const High Yes 83 Min Open 6
82
Table A1: Raw Data With Individual Sample Results and Sampler Positions on Mannequin or Stand (cont.)
Run Conc.
(ppm) Vapor Pump Type
Exp.
Profile Flow Rate Mannequin Duration Holder Position
PID Conc.
(ppm) Temp. °F RH %
4S 12.0 Hex Cont Const Low No 33 Min SKC 1
13.1 72 16
4S 13.8 Hex Cont Const Low No 33 Min Buck 2
4S 10.8 Hex Cont Const Low No 33 Min Open 3
4S 12.6 Hex Cont Const Low No 33 Min Buck 4
4S 10.5 Hex Cont Const Low No 33 Min SKC 5
4S 11.3 Hex Cont Const Low No 33 Min Open 6
4L 12.6 Hex Cont Const Low No 83 Min SKC 1
15.6 73 17
4L 14.5 Hex Cont Const Low No 83 Min Buck 2
4L 13.3 Hex Cont Const Low No 83 Min Open 3
4L 14.9 Hex Cont Const Low No 83 Min Buck 4
4L 12.6 Hex Cont Const Low No 83 Min SKC 5
4L 13.0 Hex Cont Const Low No 83 Min Open 6
5 15.4 Xyl Cont Var Low No 83 Min Buck 1
15.9 75 15
5 13.1 Xyl Cont Var Low No 83 Min SKC 2
5 13.2 Xyl Cont Var Low No 83 Min Open 3
5 13.3 Xyl Cont Var Low No 83 Min Open 4
5 13.4 Xyl Cont Var Low No 83 Min SKC 5
5 15.3 Xyl Cont Var Low No 83 Min Buck 6
6 14.2 Hex Pulse Var High Yes 83 Min SKC 1
16.8 72 16
6 14.2 Hex Pulse Var High Yes 83 Min Buck 2
6 14.1 Hex Pulse Var High Yes 83 Min SKC 3
6 14.3 Hex Pulse Var High Yes 83 Min Buck 4
6 14.3 Hex Pulse Var High Yes 83 Min Open 5
6 14.4 Hex Pulse Var High Yes 83 Min Open 6
83
Table A1: Raw Data With Individual Sample Results and Sampler Positions on Mannequin or Stand (cont.)
Run Conc.
(ppm) Vapor Pump Type
Exp.
Profile Flow Rate Mannequin Duration Holder Position
PID Conc.
(ppm) Temp. °F RH %
7 13.9 Hex Cont Var Low Yes 83 Min SKC 1
16.1 72 16
7 16.8 Hex Cont Var Low Yes 83 Min Buck 2
7 15.9 Hex Cont Var Low Yes 83 Min Buck 3
7 13.6 Hex Cont Var Low Yes 83 Min Open 4
7 13.5 Hex Cont Var Low Yes 83 Min Open 5
7 13.5 Hex Cont Var Low Yes 83 Min SKC 6
8 13.2 Xyl Pulse Var High No 83 Min Buck 1
15.5 75 35
8 13.2 Xyl Pulse Var High No 83 Min Open 2
8 12.9 Xyl Pulse Var High No 83 Min Buck 3
8 12.8 Xyl Pulse Var High No 83 Min SKC 4
8 13.6 Xyl Pulse Var High No 83 Min Open 5
8 13.2 Xyl Pulse Var High No 83 Min SKC 6
9 14.5 Xyl Cont Var High Yes 83 Min Buck 1
14.5 73 18
9 15.0 Xyl Cont Var High Yes 83 Min Open 2
9 14.3 Xyl Cont Var High Yes 83 Min Open 3
9 14.2 Xyl Cont Var High Yes 83 Min SKC 4
9 14.2 Xyl Cont Var High Yes 83 Min Buck 5
9 13.8 Xyl Cont Var High Yes 83 Min SKC 6
10S 15.1 Xyl Pulse Const Low No 33 Min Buck 1
16.3 73 16
10S 15.6 Xyl Pulse Const Low No 33 Min SKC 2
10S 15.2 Xyl Pulse Const Low No 33 Min SKC 3
10S 14.3 Xyl Pulse Const Low No 33 Min Buck 4
10S 15.3 Xyl Pulse Const Low No 33 Min Open 5
10S 15.6 Xyl Pulse Const Low No 33 Min Open 6
84
Table A1: Raw Data With Individual Sample Results and Sampler Positions on Mannequin or Stand (cont.)
Run Conc.
(ppm) Vapor Pump Type
Exp.
Profile Flow Rate Mannequin Duration Holder Position
PID Conc.
(ppm) Temp. °F RH %
10L 13.1 Xyl Pulse Const Low No 83 Min Buck 1
14.8 73 16
10L 13.5 Xyl Pulse Const Low No 83 Min SKC 2
10L 13.4 Xyl Pulse Const Low No 83 Min SKC 3
10L 13.0 Xyl Pulse Const Low No 83 Min Buck 4
10L 13.6 Xyl Pulse Const Low No 83 Min Open 5
10L 13.1 Xyl Pulse Const Low No 83 Min Open 6
11S 13.8 Xyl Cont Const High No 33 Min Buck 1
12.0 73 23
11S 14.3 Xyl Cont Const High No 33 Min Open 2
11S 13.4 Xyl Cont Const High No 33 Min Open 3
11S 13.3 Xyl Cont Const High No 33 Min SKC 4
11S 13.6 Xyl Cont Const High No 33 Min SKC 5
11S 14.2 Xyl Cont Const High No 33 Min Buck 6
11L 15.1 Xyl Cont Const High No 83 Min Buck 1
12.3 71 25
11L 14.8 Xyl Cont Const High No 83 Min Open 2
11L 14.8 Xyl Cont Const High No 83 Min Open 3
11L 13.6 Xyl Cont Const High No 83 Min SKC 4
11L 14.1 Xyl Cont Const High No 83 Min SKC 5
11L 14.5 Xyl Cont Const High No 83 Min Buck 6
12S 13.8 Xyl Cont Const Low Yes 33 Min Open 1
13.3 72 23
12S 13.6 Xyl Cont Const Low Yes 33 Min SKC 2
12S 14.5 Xyl Cont Const Low Yes 33 Min Buck 3
12S 14.1 Xyl Cont Const Low Yes 33 Min Open 4
12S 14.1 Xyl Cont Const Low Yes 33 Min Buck 5
12S 13.9 Xyl Cont Const Low Yes 33 Min SKC 6
85
Table A1: Raw Data With Individual Sample Results and Sampler Positions on Mannequin or Stand (cont.)
Run Conc.
(ppm) Vapor Pump Type
Exp.
Profile Flow Rate Mannequin Duration Holder Position
PID Conc.
(ppm) Temp. °F RH %
12L 16.6 Xyl Cont Const Low Yes 83 Min Open 1
14.8 73 32
12L 16.0 Xyl Cont Const Low Yes 83 Min SKC 2
12L 17.0 Xyl Cont Const Low Yes 83 Min Buck 3
12L 16.3 Xyl Cont Const Low Yes 83 Min Open 4
12L 16.5 Xyl Cont Const Low Yes 83 Min Buck 5
12L 16.3 Xyl Cont Const Low Yes 83 Min SKC 6
13S 9.2 Hex Pulse Const Low Yes 33 Min SKC 1
14.8 70 24
13S 9.0 Hex Pulse Const Low Yes 33 Min Buck 2
13S 9.5 Hex Pulse Const Low Yes 33 Min Open 3
13S 9.4 Hex Pulse Const Low Yes 33 Min SKC 4
13S 9.4 Hex Pulse Const Low Yes 33 Min Open 5
13S 9.4 Hex Pulse Const Low Yes 33 Min Buck 6
13L 11.7 Hex Pulse Const Low Yes 83 Min SKC 1
12.8 72 22
13L 11.4 Hex Pulse Const Low Yes 83 Min Buck 2
13L 12.1 Hex Pulse Const Low Yes 83 Min Open 3
13L 12.0 Hex Pulse Const Low Yes 83 Min SKC 4
13L 12.6 Hex Pulse Const Low Yes 83 Min Open 5
13L 12.0 Hex Pulse Const Low Yes 83 Min Buck 6
14 15.2 Hex Pulse Var High No 83 Min Buck 1
14.4 73 22
14 15.0 Hex Pulse Var High No 83 Min Buck 2
14 14.6 Hex Pulse Var High No 83 Min SKC 3
14 15.3 Hex Pulse Var High No 83 Min Open 4
14 15.3 Hex Pulse Var High No 83 Min SKC 5
14 15.7 Hex Pulse Var High No 83 Min Open 6
86
Table A1: Raw Data With Individual Sample Results and Sampler Positions on Mannequin or Stand (cont.)
Run Conc.
(ppm) Vapor Pump Type
Exp.
Profile Flow Rate Mannequin Duration Holder Position
PID Conc.
(ppm) Temp. °F RH %
15S 12.4 Xyl Pulse Const High Yes 33 Min Buck 1
15.2 73 32
15S 12.4 Xyl Pulse Const High Yes 33 Min Open 2
15S 12.8 Xyl Pulse Const High Yes 33 Min Buck 3
15S 13.0 Xyl Pulse Const High Yes 33 Min Open 4
15S 13.1 Xyl Pulse Const High Yes 33 Min SKC 5
15S 12.9 Xyl Pulse Const High Yes 33 Min SKC 6
15L 14.2 Xyl Pulse Const High Yes 83 Min Buck 1
12.5 73 31
15L 14.2 Xyl Pulse Const High Yes 83 Min Open 2
15L 14.0 Xyl Pulse Const High Yes 83 Min Buck 3
15L 14.2 Xyl Pulse Const High Yes 83 Min Open 4
15L 14.2 Xyl Pulse Const High Yes 83 Min SKC 5
15L 13.7 Xyl Pulse Const High Yes 83 Min SKC 6
16S 10.9 Hex Pulse Const High No 33 Min SKC 1
12.0 72 24
16S 11.4 Hex Pulse Const High No 33 Min Buck 2
16S 11.3 Hex Pulse Const High No 33 Min SKC 3
16S 11.1 Hex Pulse Const High No 33 Min Open 4
16S 11.3 Hex Pulse Const High No 33 Min Buck 5
16S 11.3 Hex Pulse Const High No 33 Min Open 6
16L 13.4 Hex Pulse Const High No 83 Min SKC 1
12.8 72 24
16L 13.6 Hex Pulse Const High No 83 Min Buck 2
16L 13.4 Hex Pulse Const High No 83 Min SKC 3
16L 13.7 Hex Pulse Const High No 83 Min Open 4
16L 13.7 Hex Pulse Const High No 83 Min Buck 5
16L 13.4 Hex Pulse Const High No 83 Min Open 6
87
Table A2: Run Table With Results For Concentration Ratio Run Vapor Pump
Type Exp. Prof. Flow Rate Mannequin Holder Duration Conc.
Ratio
1 Hex Pulse Var Low No SKC Long 1.03
2 Xyl Pulse Var Low Yes SKC Long 1.08
3 Hex Cont. Con High Yes SKC Short 0.97
4 Hex Cont. Con High Yes SKC Long 0.96
5 Hex Cont. Con Low No SKC Short 1.02
6 Hex Cont. Con Low No SKC Long 0.96
7 Xyl Cont. Var Low No SKC Long 1.00
8 Hex Pulse Var High Yes SKC Long 0.99
9 Hex Cont. Var Low Yes SKC Long 1.01
10 Xyl Pulse Var High No SKC Long 0.97
11 Xyl Cont. Var High Yes SKC Long 0.96
12 Xyl Pulse Con Low No SKC Short 1.00
13 Xyl Pulse Con Low No SKC Long 1.01
14 Xyl Cont. Con High No SKC Short 0.97
15 Xyl Cont. Con High No SKC Long 0.94
16 Xyl Cont. Con Low Yes SKC Short 0.99
17 Xyl Cont. Con Low Yes SKC Long 0.98
18 Hex Pulse Con Low Yes SKC Short 0.98
19 Hex Pulse Con Low Yes SKC Long 0.96
20 Hex Cont. Var High No SKC Long 0.96
21 Xyl Pulse Con High Yes SKC Short 1.02
22 Xyl Pulse Con High Yes SKC Long 0.98
23 Hex Pulse Con High No SKC Short 0.99
24 Hex Pulse Con High No SKC Long 0.99
1 Hex Pulse Var Low No Buck Long 0.99
2 Xyl Pulse Var Low Yes Buck Long 1.08
3 Hex Cont. Con High Yes Buck Short 1.02
4 Hex Cont. Con High Yes Buck Long 0.99
5 Hex Cont. Con Low No Buck Short 1.19
6 Hex Cont. Con Low No Buck Long 1.12
7 Xyl Cont. Var Low No Buck Long 1.16
8 Hex Pulse Var High Yes Buck Long 0.99
9 Hex Cont. Var Low Yes Buck Long 1.21
10 Xyl Pulse Var High No Buck Long 0.97
11 Xyl Cont. Var High Yes Buck Long 0.98
12 Xyl Pulse Con Low No Buck Short 0.95
13 Xyl Pulse Con Low No Buck Long 0.98
14 Xyl Cont. Con High No Buck Short 1.01
15 Xyl Cont. Con High No Buck Long 1.00
16 Xyl Cont. Con Low Yes Buck Short 1.03
17 Xyl Cont. Con Low Yes Buck Long 1.02
18 Hex Pulse Con Low Yes Buck Short 0.97
19 Hex Pulse Con Low Yes Buck Long 0.95
20 Hex Cont. Var High No Buck Long 0.97
21 Xyl Pulse Con High Yes Buck Short 0.99
22 Xyl Pulse Con High Yes Buck Long 0.99
23 Hex Pulse Con High No Buck Short 1.01
24 Hex Pulse Con High No Buck Long 1.01