TE Taylor Enterprises, Inc. www.variation.com
1
Validation Report for Excel Spreadsheet:
STAT-18 - Skewness Kurtosis Normality Tests
Report Number: TE-17-1
Approvals:
___________________________________ September 1, 20177
Dr. Wayne A. Taylor Date Study Director
___________________________________ September 1, 2017
Ann B. Taylor Date President
TE Taylor Enterprises, Inc. www.variation.com
2
Validation Protocol for Excel Spreadsheet: STAT-18 - Skewness Kurtosis Normality Tests
Report Number: TE-17-1
1.0 Introduction The book Statistical Procedures for the Medical Device Industry by Dr. Wayne Taylor (2017) includes several Excel Spreadsheets that can be downloaded from www.variation.com/procedures. One of these spreadsheets is STAT-18 – Skewness Kurtosis Normality Tests – Protected.xls. This spreadsheet is password protected to prevent changes. The first tab displays the information below. The version being validated is Revision 1 with a date of 8/30/2017.
For those purchasing the book Statistical Procedures for the Medical Device Industry, this file contains a spreadsheet for performing the Skewness Kurtosis All and Specific normality tests. Further information about the use of this spreadsheet is provided in the book.
STATISTICAL PROCEDURES FOR THE MEDICAL DEVICE INDUSTRY
STAT-18: Normality Testing and Transformations
Dr. Wayne A. Taylor
www.variation.com/proceduresBook Link:
Revision 1: 8/30/2017
Copyright © 2017 Taylor Enterprises, Inc.All rights reserved.
This spreadsheet performs two normality tests. The first test is the skewness-kurtosis all general normality test. It is designed to detect all types of departures from normality. This normality test is described in STAT-18, Appendix A of the book. This test is included in the R software package under the name D’Agostino Chi2 Omnibus test.
The second test is the skewness-kurtosis specific normality test. It is designed to detect specific departures from normality that can cause false acceptances of a variable sampling plan. This normality test is described in STAT-18, Appendix C of the book. It is a combination of the D’Agostino Z3 Skewness and D’Agostino Z4 Kurtosis tests. It is designed to reject for tails longer than the normal distribution as evidenced by skewness towards a specification limit or high kurtosis. Formulas and references for both tests are provided in STAT-18, Appendix C of the book.
A key property of the spreadsheet is that tabs are independent of each other so tabs can be copied and the copied tab performs correctly and tabs can be deleted without affecting other tabs.
TE Taylor Enterprises, Inc. www.variation.com
3
2.0 Protocol Protocol Number: TE-17-1, Validation Protocol for Excel Spreadsheet: STAT-18 - Skewness Kurtosis Normality Tests was written to validate this spreadsheet. It can be found in Appendix A. It lists the requirements, test cases and a matrix linking the tests to the requirements demonstrating complete coverage of the requirements.
3.0 Testing The test script was executed on two machines. The first one had Windows 10 and Excel 2016 installed. The results were saved in the file STAT-18 -Windows 10. It is included as Appendix B.
The second machine had Windows 7 and Excel 2007 installed. The results were saved in the file STAT-18 -Windows 7. It is included as Appendix C.
4.0 Test Results All tests passed on both machines as shown in Appendices B and C. The validation has been successively completed.
TE Taylor Enterprises, Inc. www.variation.com
Appendix A
Validation Protocol for Excel Spreadsheet:
STAT-18 - Skewness Kurtosis Normality Tests
Protocol Number: TE-17-1
TE Taylor Enterprises, Inc. www.variation.com
1
Validation Protocol for Excel Spreadsheet:
STAT-18 - Skewness Kurtosis Normality Tests
Protocol Number: TE-17-1
Approvals:
___________________________________ August 30, 2017
Dr. Wayne A. Taylor Date Study Director
___________________________________ August 30, 2017
Ann B. Taylor Date President
TE Taylor Enterprises, Inc. www.variation.com
2
Validation Protocol for Excel Spreadsheet: STAT-18 - Skewness Kurtosis Normality Tests
Protocol Number: TE-17-1
1.0 Introduction The book Statistical Procedures for the Medical Device Industry by Dr. Wayne Taylor (2017) includes several Excel Spreadsheets that can be downloaded from www.variation.com/procedures. One of these spreadsheets is STAT-18 – Skewness Kurtosis Normality Tests – Protected.xls. This spreadsheet is password protected to prevent changes. The first tab displays the information below. The version being validated is Revision 1 with a date of 8/30/2017.
For those purchasing the book Statistical Procedures for the Medical Device Industry, this file contains a spreadsheet for performing the Skewness Kurtosis All and Specific normality tests. Further information about the use of this spreadsheet is provided in the book.
STATISTICAL PROCEDURES FOR THE MEDICAL DEVICE INDUSTRY
STAT-18: Normality Testing and Transformations
Dr. Wayne A. Taylor
www.variation.com/proceduresBook Link:
Revision 1: 8/30/2017
Copyright © 2017 Taylor Enterprises, Inc.All rights reserved.
This spreadsheet performs two normality tests. The first test is the skewness-kurtosis all general normality test. It is designed to detect all types of departures from normality. This normality test is described in STAT-18, Appendix A of the book. This test is included in the R software package under the name D’Agostino Chi2 Omnibus test.
The second test is the skewness-kurtosis specific normality test. It is designed to detect specific departures from normality that can cause false acceptances of a variable sampling plan. This normality test is described in STAT-18, Appendix C of the book. It is a combination of the D’Agostino Z3 Skewness and D’Agostino Z4 Kurtosis tests. It is designed to reject for tails longer than the normal distribution as evidenced by skewness towards a specification limit or high kurtosis. Formulas and references for both tests are provided in STAT-18, Appendix C of the book.
A key property of the spreadsheet is that tabs are independent of each other so tabs can be copied and the copied tab performs correctly and tabs can be deleted without affecting other tabs.
TE Taylor Enterprises, Inc. www.variation.com
3
2.0 Requirements Table 1 lists the user requirements to be validated.
Table 1: User Requirements for STAT-18 - Skewness Kurtosis Normality Tests
ID Description
1
Input of Data:
a. Data can be typed or pasted into the data column of the spreadsheet
b. A header or name for the data can be typed into the top cell of the data column, which is displayed at the top of the summary data section
c. The type of specification limit can be typed into the specification cell of the spreadsheet
d. Only the data column, including header, and the type of specification cell can be changed
2
Data Checking:
a. Sample size is zero when there are no valid values. Otherwise it is the number of valid values
b. Average is blank when there are no valid values. Otherwise it is the average of the valid values.
c. Standard deviation is blank when there are 1 or fewer valid values. Otherwise it is the standard deviation of the valid values.
d. Skewness is blank when there are 2 or fewer valid values or the standard deviation is 0. Otherwise it is skewness of the valid values.
e. Kurtosis is blank when there are 3 or fewer valid values or the standard deviation is 0. Otherwise it is the kurtosis of the valid values.
3
Skewness Kurtosis All test:
a. If there are less than 15 valid values or the standard deviation is 0, no p-value is displayed and an error message is displayed
b. Otherwise, the p-value is displayed
c. If the p-value is displayed and is > 0.05, a message that the data passes SK All normality test is displayed
d. If the p-value is displayed and is ≤ 0.05, a message that the data fails SK All normality test" is displayed
4
Skewness Kurtosis Specific test:
a. If there are less than 15 valid values, the standard deviation is 0 or the type of specification limit is not a “T”, “U” or “L”, an error message is displayed
b. Otherwise, the pass/fail conclusion is displayed along with the type of specification limit.
5 Tabs:
a. Tabs are independent of each other so tab can be copied and all other tab deleted with tab results remaining the same.
TE Taylor Enterprises, Inc. www.variation.com
4
3.0 Test Environment All testing will be performed at the facilities of Taylor Enterprises, Inc. (TEI), Libertyville, IL USA
All individuals performing the testing described below shall have a thorough knowledge of STAT-018 and associated spreadsheet and be proficient in its use.
The spreadsheet does not include any macros. All functionality is through equations including the standard functions included with Excel that have remained unchanged since early versions of Excel. The purpose of this document is to describe the validation of this spreadsheet across a range of systems. The validation tests described in the following sections shall performed on a newer system with Windows 10 / Excel 2016 and an older system with Windows 7 / Excel 2007.
4.0 Test Script For each system run the test script in Appendix A. The test script includes instructions for documenting the system used, the downloading of the spreadsheet and test cases to run.
Make a copy of Appendix A in a separate Word document for each system. Copy the evidence of each test case into this word file. All test cases should be run by the same person on a single day. When done save the file, print it, and then sign and date the printed copy.
The first two items in the test script provide evident the system, software and spreadsheet use. They are not linked to the requirements. The remaining test cases are designed to cover all the requirements. A matrix showing the linkage is provided in Appendix B.
Appendix C provides evidence of the correct results for the test cases. Averages, standard deviations, skewness and kurtosis tests are calculated using Minitab, version 18 (www.minitab.com). The p-value for SK All (called omnibus), ZS, two-sided p-value for skewness and ZK are calculated using the R software package, version 3.2.4 (cran.r-project.org). The other p-values are calculated using the following equations in Excel:
Skewness upper one-sided p-value =NORMDIST(-ZS,0,1,TRUE)
Skewness lower one-sided p-value =NORMDIST(ZS,0,1,TRUE)
Kurtosis upper one-sided p-value =NORMDIST(-ZK,0,1,TRUE)
TE Taylor Enterprises, Inc. www.variation.com
5
Appendix A: Test Script
1. Document System Used
1.1. Capture a screen shot showing the computer make and model
<Insert screen shot here>
1.2. Capture a screen shot showing the operating system including version number
<Insert screen shot here>
1.3. Capture a screen shot showing the version of Excel used
<Insert screen shot here>
2. Download Spreadsheet and Open
2.1. Download the spreadsheet from http://www.variation.com/procedures/SK_normality_tests.html. Download the protected version. Save it to the hard drive. Capture a screen shot showing the file downloaded.
<Insert screen shot here>
2.2. Double click the spreadsheet to open it. Capture a screen shot of the section of the first tab showing the version and date.
<Insert screen shot here>
Acceptance Criteria: The following is displayed “Revision 1: 8/30/2017”
Pass or Fail: XXXX Initials: XXX Date: XXX
3. Input of Data When Normality Test Cannot Be Performed
3.1. Double click on at least 10 non-yellow cells on each of the three sheets (Req. 1d).
<Insert screen shot here>
Acceptance Criteria: Each time a message should be displayed that the cell is protected.
Pass or Fail: XXXX Initials: XXX Date: XXX
3.2. Go to the Skewness Kurtosis Norm Tab. Change cell D7 to “1” (Req. 1c).
<Insert screen shot here>
Acceptance Criteria: The cell should change to “1”.
Pass or Fail: XXXX Initials: XXX Date: XXX
TE Taylor Enterprises, Inc. www.variation.com
6
3.3. Change cell B11 to “Diameter” (Req. 1b).
<Insert screen shot here>
Acceptance Criteria: Both B11 and D11 should change to “Diameter”. The sample size should be 0. The average, SD, skewness, kurtosis and p-value should be blank. The conclusions should indicate there is insufficient data.
Pass or Fail: XXXX Initials: XXX Date: XXX
3.4. Change cell B12 to “A” (2-4 for invalid values. n=0)
<Insert screen shot here>
Acceptance Criteria: The sample size should be 0. The average, SD, skewness, kurtosis and p-value should be blank. The conclusions should indicate there is insufficient data.
Pass or Fail: XXXX Initials: XXX Date: XXX
3.5. Change cell B14 to “1” (Req. 2-4 for n=1)
<Insert screen shot here>
Acceptance Criteria: The sample size and average should be 1. The standard deviation, skewness, kurtosis and p-value should be blank. The conclusions should indicate there is insufficient data.
Pass or Fail: XXXX Initials: XXX Date: XXX
3.6. Change cell B16 to “1” (Req. 2-4 for n=2)
<Insert screen shot here>
Acceptance Criteria: The sample size should be 2. The average should be 1. The standard deviation should be 0. The skewness, kurtosis and p-value should be blank. The conclusions should indicate there is insufficient data.
Pass or Fail: XXXX Initials: XXX Date: XXX
3.7. Change cell B17 to “1” (Req. 2-4 for n=3 and SD=0)
<Insert screen shot here>
Acceptance Criteria: The sample size should be 3. The average should be 1. The standard deviation should be 0. The skewness, kurtosis and p-value should be blank. The conclusions should indicate there is insufficient data.
Pass or Fail: XXXX Initials: XXX Date: XXX
3.8. Change cell B18 to “1” (Req. 2-4 for n=4 and SD=0)
<Insert screen shot here>
Acceptance Criteria: The sample size should be 4. The average should be 1. The standard deviation should be 0. The skewness, kurtosis and p-value should be blank. The conclusions should indicate there is insufficient data.
Pass or Fail: XXXX Initials: XXX Date: XXX
TE Taylor Enterprises, Inc. www.variation.com
7
3.9. Copy cell B18 to clipboard. Select cells B12 to B211 and then select paste “1” (Req. 1a, 2-4 for n=200 and SD=0)
<Insert screen shot here>
Acceptance Criteria: The data column should be filled with “1”. The sample size should be 200. The average should be 1. The standard deviation should be 0. The skewness, kurtosis and p-value should be blank. The conclusions should indicate there is insufficient data.
Pass or Fail: XXXX Initials: XXX Date: XXX
4. Input of Data When Normality Test Can Be Performed
4.1. Clear Cells B12:B211. Then paste the following into cells B12:B15. (Req. 2-4 for n=4 and SD>0).
0.1257 0.1249 0.1257 0.125
<Insert screen shot here>
Acceptance Criteria: The sample size should be 4. The average should round to 0.12532. The standard deviation should round to 0.00043. The skewness should round to -0.04558. The kurtosis should round to -5.73743. The p-value should be blank. The conclusions should indicate there is insufficient data.
Pass or Fail: XXXX Initials: XXX Date: XXX
4.2. Paste the following into cells B16:B25 (Req. 2-4 for n=14 and SD>0).
0.1264
0.1238
0.1254
0.1248
0.1253
0.1234
0.1262
0.1262
0.1269
0.1245
<Insert screen shot here>
Acceptance Criteria: The sample size should be 14. The average should round to 0.12530. The standard deviation should round to 0.00099. The skewness should round to -0.366071.
TE Taylor Enterprises, Inc. www.variation.com
8
The kurtosis should round to -0.264172. The p-value should be blank. The conclusions should indicate there is insufficient data.
Pass or Fail: XXXX Initials: XXX Date: XXX
4.3. Type 0.1252 into cell B26 (Req. 3-4 for n>=15, Normal data, Invalid type of specification).
<Insert screen shot here>
Acceptance Criteria: The sample size should be 15. A p-value should be given. The conclusions for the SK All test should be it passes. The conclusion for the SK Spec test should be incorrect type of specification.
Pass or Fail: XXXX Initials: XXX Date: XXX
4.4. Change cell D7 to “T” and paste cells D10:29 of Example Data Sets tab into cells B12:31 (Req. 3-4 for n>=15, normal data, two-sided specification)
<Insert screen shot here>
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.8047. The conclusions for the SK All test should be it passes. The conclusion for the SK Spec test should be it passes for a two-sided specification. Cell J17 should round to -0.659 (ZS). Cell J18 should round to 0.5099 (p-value 2-sided skewness). Cell J29 should round to -0.0156 (ZK). Cell J31 should round to 0.5062 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: XXXX Initials: XXX Date: XXX
4.5. Change cell D7 to “U” (Req. 3-4 for n>=15, normal data, upper specification)
<Insert screen shot here>
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.8047. The conclusions for the SK All test should be it passes. The conclusion for the SK Spec test should be it passes for an upper specification. Cell J17 should round to -0.659 (ZS). Cell J19 should round to 0.7450 (p-value positive skewness). Cell J29 should round to -0.0156 (ZK) Cell J31 should round to 0.5062 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: XXXX Initials: XXX Date: XXX
4.6. Change cell D7 to “L” (Req. 3-4 for n>=15, normal data, lower specification)
<Insert screen shot here>
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.8047. The conclusions for the SK All test should be it passes. The conclusion for the SK Spec test should be it passes for a lower specification. Cell J17 should round to -0.659 (ZS). Cell J20 should round to 0.2550 (p-value negative skewness). Cell J29 should round to -0.0156 (ZK).
TE Taylor Enterprises, Inc. www.variation.com
9
Cell J31 should round to 0.5062 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: XXXX Initials: XXX Date: XXX
4.7. Change cell D7 to “T” and cell B12:B31 to the values below (Req. 3-4 for n>=15, positive kurtosis data, two-sided specification)
0.1 -1 0.2 0.2 -2.9 2.6 0.1 1.1 -0.9 -0.2 -0.6 -0.3 1.3 0.4 0
-0.3 -0.1 -0.2 0.1 -1.4
<Insert screen shot here>
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.1057. The conclusions for the SK All test should be it passes. The conclusion for the SK Spec test should be it fails for a two-sided specification. Cell J17 should round to -0.1734 (ZS). Cell J18 should round to 0.8624 (p-value 2-sided skewness). Cell J29 should round to 2.1131 (ZK). Cell J31 should round to 0.0173 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: XXXX Initials: XXX Date: XXX
4.8. Change cell D7 to “U” (Req. 3-4 for n>=15, normal data, upper specification)
<Insert screen shot here>
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.1057. The conclusions for the SK All test should be it passes. The conclusion for the SK Spec test should be it fails for an upper specification. Cell J17 should round to -0.1734 (ZS).
TE Taylor Enterprises, Inc. www.variation.com
10
Cell J19 should round to 0.5688 (p-value positive skewness). Cell J29 should round to 2.1131 (ZK). Cell J31 should round to 0.0173 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: XXXX Initials: XXX Date: XXX
4.9. Change cell D7 to “L” (Req. 3-4 for n>=15, normal data, lower specification)
<Insert screen shot here>
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.1057. The conclusions for the SK All test should be it passes. The conclusion for the SK Spec test should be it fails for a lower specification. Cell J17 should round to -0.1734 (ZS). Cell J20 should round to 0.4312 (p-value negative skewness). Cell J29 should round to 2.1131 (ZK). Cell J31 should round to 0.0173 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: XXXX Initials: XXX Date: XXX
4.10. Change cell D7 to “T” and cell B12:B31 to the values below (Req. 3-4 for n>=15, negative kurtosis data, two-sided specification)
-1.1 -0.5 -0.4 1.1 -1 1.2 -0.8 1.2 -1.2 -1 0.6 -1.1 1.2 -1.2
1 -0.8 -1.1
1 1.2 1.2
<Insert screen shot here>
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.0000 The conclusions for the SK All test should be it fails. The conclusion for the SK Spec test should be it passes for a two-sided
TE Taylor Enterprises, Inc. www.variation.com
11
specification. Cell J17 should round to 0.3238 (ZS). Cell J18 should round to 0.7461 (p-value 2-sided skewness). Cell J29 should round to -5.0793 (ZK) Cell J31 should round to 1.0000 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: XXXX Initials: XXX Date: XXX
4.11. Change cell D7 to “U” (Req. 3-4 for n>=15, normal data, upper specification)
<Insert screen shot here>
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.0000. The conclusions for the SK All test should be it fails. The conclusion for the SK Spec test should be it passes for an upper specification. Cell J17 should round to 0.3238 (ZS). Cell J19 should round to 0.3730 (p-value positive skewness). Cell J29 should round to -5.0793 (ZK). Cell J31 should round to 1.0000 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: XXXX Initials: XXX Date: XXX
4.12. Change cell D7 to “L” (Req. 3-4 for n>=15, normal data, lower specification)
<Insert screen shot here>
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.0000. The conclusions for the SK All test should be it fails. The conclusion for the SK Spec test should be it passes for a lower specification. Cell J17 should round to 0.3238 (ZS). Cell J20 should round to 0.6270 (p-value negative skewness). Cell J29 should round to -5.0793 (ZK). Cell J31 should round to 1.0000 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: XXXX Initials: XXX Date: XXX
4.13. Change cell D7 to “T” and cell B12:B31 to the values below (Req. 3-4 for n>=15, positive skewness data, two-sided specification)
-0.7 -0.4 -0.4 -0.6 1.8 -0.4 -0.4 1.1 1.5 -0.4 0.3
TE Taylor Enterprises, Inc. www.variation.com
12
-0.7 -0.8 -0.8 -0.8 -0.8 -0.8 -0.8 -0.3 0.7
<Insert screen shot here>
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.0150. The conclusions for the SK All test should be it fails. The conclusion for the SK Spec test should be it fails for a two-sided specification. Cell J17 should round to 2.6691 (ZS). Cell J18 should round to 0.0076 (p-value 2-sided skewness). Cell J29 should round to 1.1275 (ZK). Cell J31 should round to 0.1298 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: XXXX Initials: XXX Date: XXX
4.14. Change cell D7 to “U” (Req. 3-4 for n>=15, normal data, upper specification)
<Insert screen shot here>
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.0150. The conclusions for the SK All test should be it fails. The conclusion for the SK Spec test should be it fails for an upper specification. Cell J17 should round to 2.6691 (ZS). Cell J19 should round to 0.0038 (p-value positive skewness). Cell J29 should round to 1.1275 (ZK). Cell J31 should round to 0.1298 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: XXXX Initials: XXX Date: XXX
4.15. Change cell D7 to “L” (Req. 3-4 for n>=15, normal data, lower specification)
<Insert screen shot here>
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.0150. The conclusions for the SK All test should be it fails. The conclusion for the SK Spec test should be it passes for a lower specification. Cell J17 should round to 2.6691 (ZS). Cell J20 should round to 0.9962 (p-value negative skewness). Cell J29 should round to 1.1275 (ZK). Cell J31 should round to 0.1298 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: XXXX Initials: XXX Date: XXX
TE Taylor Enterprises, Inc. www.variation.com
13
4.16. Change cell D7 to “T” and cell B12:B31 to the values below (Req. 3-4 for n>=15, negative skewness data, two-sided specification)
0.7 0.4 0.4 0.6 -1.8 0.4 0.4 -1.1 -1.5 0.4 -0.3 0.7 0.8 0.8 0.8 0.8 0.8 0.8 0.3 -0.7
<Insert screen shot here>
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.0150. The conclusions for the SK All test should be it fails. The conclusion for the SK Spec test should be it fails for a two-sided specification. Cell J17 should round to -2.6691 (ZS). Cell J18 should round to 0.0076 (p-value 2-sided skewness). Cell J29 should round to 1.1275 (ZK). Cell J31 should round to 0.1298 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: XXXX Initials: XXX Date: XXX
4.17. Change cell D7 to “U” (Req. 3-4 for n>=15, normal data, upper specification)
<Insert screen shot here>
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.0150. The conclusions for the SK All test should be it fails. The conclusion for the SK Spec test should be it passes for an upper specification. Cell J17 should round to -2.6691 (ZS). Cell J19 should round to 0.9962 (p-value positive skewness). Cell J29 should round to 1.1275 (ZK).
TE Taylor Enterprises, Inc. www.variation.com
14
Cell J31 should round to 0.1298 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: XXXX Initials: XXX Date: XXX
4.18. Change cell D7 to “L” (Req. 3-4 for n>=15, normal data, lower specification)
<Insert screen shot here>
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.0150. The conclusions for the SK All test should be it fails. The conclusion for the SK Spec test should be it fails for a lower specification. Cell J17 should round to -2.6691 (ZS). Cell J20 should round to 0.0038 (p-value negative skewness). Cell J29 should round to 1.1275 (ZK). Cell J31 should round to 0.1298 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: XXXX Initials: XXX Date: XXX
4.19. Change cell D7 to “T” and cell B12:B211 to the values below (Req. 3-4 for n=200, normal data, two-sided specification)
2.1 0.1 -0.5 0.1 -0.6
1 -0.4 0.1 -2.4 -1
0.6 0.1 -0.3 2.3 -1
0.3 -0.3
2 -0.9 0.1 0.8 -1.5
1 2.1 0.7 -0.1 -0.4
TE Taylor Enterprises, Inc. www.variation.com
15
0.6 -1
-0.5 0.4 0.2 -0.1 -0.3 -2.5 -0.2 0.3 -0.8 -0.2 -3
1.9 0.7 -0.2 0.4 -3.4 0.9 -1.8 0.8 0
-0.2 0
-0.1 0.8 -1.1 1.4 2.5 0.5 -0.1 -1.8 0.9 0.1 0.2 1.3 0.1 -0.1 0.7 -0.9 0.3 -0.2 -0.1 1.3
TE Taylor Enterprises, Inc. www.variation.com
16
2.3 0.5 0.6 1.1 1
-2.2 2.5 1.6 1.2 1.7 -0.9 2.1 0.6 0.6 0.3 1.1 0.5 -0.9 0.7 -0.4 -0.7 0.9 0.1 0.5 -0.2 -1.3 -1
-2.3 1.3 -0.5 -0.3 1.9 1
-0.2 -0.1 1.2 -0.7 2.1 1.1 1.4 -0.5
2 -0.3 0.3
TE Taylor Enterprises, Inc. www.variation.com
17
-0.2 1.3 1.1 -1.3 -0.6 -0.8 -0.2 1.2 -0.7 -2.7 0.1 -1.1 -0.8 -0.8 -0.3 1.1 0
-0.5 1.6 0.5 0.5 -1.6 2.4 -0.3 0.6 -0.5
0 0.8 1.9 -0.4 0.3 -0.2 1.6 -0.5 -0.8 -1.2 -1.6 0.4 1.3 0.4 -1.4 -0.6 -0.5 -0.4
TE Taylor Enterprises, Inc. www.variation.com
18
0.2 1.4 -1.5 1.3 -0.2 -0.1 -0.8 0.7 -0.3 -1.2 0.5 -0.7 -0.7 0.7 -0.3 -1
0.5 0.6 1.6 1 1
0.7 -0.1 -0.9 -0.1
1 0.7 0.1 0.6 0.2 1.6 -0.5 -0.7 -1.1 0.4 -0.6 1.9 -0.7 -0.2 -0.5 -0.7
<Insert screen shot here>
TE Taylor Enterprises, Inc. www.variation.com
19
Acceptance Criteria: The sample size should be 200. The average should round to 0.11150. The standard deviation should round to 1.07923. The skewness should round to -0.196780. The kurtosis should round to 0.386000. The p-value should round to 0.2651. The conclusions for the SK All test should be it passes. The conclusion for the SK Spec test should be it passes for a two-sided specification. Cell J17 should round to -1.1567 (ZS). Cell J18 should round to 0.2474 (p-value 2-sided skewness). Cell J29 should round to 1.1478 (ZK). Cell J31 should round to 0.1255 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: XXXX Initials: XXX Date: XXX
4.20. Change cell D7 to “U” (Req. 3-4 for n=200, normal data, upper specification)
<Insert screen shot here>
Acceptance Criteria: The sample size should be 200. The p-value should round to 0.2651. The conclusions for the SK All test should be it passes. The conclusion for the SK Spec test should be it passes for an upper specification. Cell J17 should round to -1.1567 (ZS). Cell J19 should round to 0.8763 (p-value positive skewness). Cell J29 should round to 1.1478 (ZK). Cell J31 should round to 0.1255 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: XXXX Initials: XXX Date: XXX
4.21. Change cell D7 to “L” (Req. 3-4 for n=200, normal data, lower specification)
<Insert screen shot here>
Acceptance Criteria: The sample size should be 200. The p-value should round to 0.2651. The conclusions for the SK All test should be it passes. The conclusion for the SK Spec test should be it passes for a lower specification. Cell J17 should round to -1.1567 (ZS). Cell J20 should round to 0.1237 (p-value negative skewness). Cell J29 should round to 1.1478 (ZK). Cell J31 should round to 0.1255 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: XXXX Initials: XXX Date: XXX
5. Independence of tab
5.1. Copy the Skewness Kurtosis Tests tab. Delete all tab but the copy.
<Insert screen shot here>
Acceptance Criteria: The sample size should be 200. The average should round to 0.11150. The standard deviation should round to 1.07923.
TE Taylor Enterprises, Inc. www.variation.com
20
The skewness should round to -0.196780. The kurtosis should round to 0.386000. The p-value should round to 0.2651. The conclusions for the SK All test should be it passes. The conclusion for the SK Spec test should be it passes for a lower specification. Cell J17 should round to -1.1567 (ZS). Cell J20 should round to 0.1237 (p-value negative skewness). Cell J29 should round to 1.1478 (ZK). Cell J31 should round to 0.1255 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: XXXX Initials: XXX Date: XXX
6. Saving and Signing File
6.1. Determine if all tests passed
All Test Cases for Computer Pass? yes no
6.2. Save File and show name and date.
Name of File:
Data of File:
Signature: ________________________________ Date: XXX
TE Taylor Enterprises, Inc. www.variation.com
21
Appendix B: Test Matrix Table 2 shows for each requirement the tests that demonstrate the requirement is meet with justification the test cases cover the requirement.
Table 2: Test Cases Matching User Requirements ID Description
1
Input of Data:
a. Data can be typed or pasted into the data column of the spreadsheet
3.9
b. A header or name for the data can be typed into the top cell of the data column, which is displayed at the top of the summary data section
3.3
c. The type of specification limit can be typed into the specification cell of the spreadsheet
3.2
d. Only the data column, including header, and the type of specification cell can be changed
3.1
2
Data Checking:
a. Sample size is zero when there are no valid values. Otherwise it is the number of valid values
No valid values – 3.3, 3.4 covers no values and invalid value
Otherwise – 3.5 to 3.9 cover full range of n=1 to 200
b. Average is blank when there are no valid values. Otherwise it is the average of the valid values.
No valid values – 3.3, 3.4 covers no values and invalid value
Otherwise – 3.5 to 3.9, 4.1, 4.2, 4.19 cover full range of n=1 to 200
c. Standard deviation is blank when there are 1 or fewer valid values. Otherwise it is the standard deviation of the valid values.
1 or fewer valid values – 3.3 to 3.5 covers no values, invalid value, n=1
Otherwise – 3.6 to 3.9, 4.1, 4.2, 4.19 cover full range of n=2 to 200
d. Skewness is blank when there are 2 or fewer valid values or the standard deviation is 0. Otherwise it is skewness of the valid values.
2 or fewer valid values– 3.3 to 3.6 covers no values, invalid value, n=1, n=2
Standard deviation is 0 – 3.7 to 3.9 cover full range of n=3 to 200
Otherwise –4.1, 4.2, 4.19 cover range of n=4 to 200
e. Kurtosis is blank when there are 3 or fewer valid values or the standard deviation is 0. Otherwise it is the kurtosis of the valid values.
3 or fewer valid values– 3.3 to 3.7 covers no values, invalid value, n=1, n=2, n=3
Standard deviation is 0 – 3.3 to 3.9 cover full range of n=4 to 200.
Otherwise –4.1, 4.2, 4.19 cover full range of n=4 to 200
TE Taylor Enterprises, Inc. www.variation.com
22
3
Skewness Kurtosis All test:
a. If there are less than 15 valid values or the standard deviation is 0, no p-value is displayed and an error message is displayed
3.3 to 3.9, 4.1, 4.2 for n=0 to 14
b. Otherwise, the p-value is displayed
4.3 to 4.21 for n=15 to n=200
c. If the p-value is displayed and is > 0.05, a message that the data passes SK All normality test is displayed
4.3 to 4.9, 4.19 to 4.21
d. If the p-value is displayed and is ≤ 0.05, a message that the data fails SK All normality test" is displayed
4.10 to 4.18,
4
Skewness Kurtosis Specific test:
a. If there are less than 15 valid values, the standard deviation is 0 or the type of specification limit is not a “T”, “U” or “L”, an error message is displayed
3.3 to 3.9, 4.1, 4.2. 4.3
b. Otherwise, the pass/fail conclusion is displayed along with the type of specification limit.
4.4 to 4.21 – 18 test cases cover 3 type of specification limits for six cases n=20, normal, n=200 normal, positive skewness, negative skewness, positive kurtosis and negative kurtosis.
5
Tabs:
a. Tabs are independent of each other so tab can be copied and all other tab deleted with tab results remaining the same.
5.1
TE Taylor Enterprises, Inc. www.variation.com
23
Appendix C: Verification of Results 4.1 From Minitab version 18.0.
1st Quartile 0.12492Median 0.125353rd Quartile 0.12570Maximum 0.12570
0.12463 0.12602
0.12490 0.12570
0.00025 0.00162
A-Squared 0.49P-Value 0.093
Mean 0.12532StDev 0.00043Variance 0.00000Skewness -0.04558Kurtosis -5.73743N 4
Minimum 0.12490
Anderson-Darling Normality Test
95% Confidence Interval for Mean
95% Confidence Interval for Median
95% Confidence Interval for StDev
0.12560.12540.12520.1250
Median
Mean
0.126000.125750.125500.125250.125000.124750.12450
95% Confidence Intervals
Summary Report for Diameter n=4
4.2 From Minitab version 18.0
1st Quartile 0.12472Median 0.125353rd Quartile 0.12620Maximum 0.12690
0.12473 0.12587
0.12478 0.12620
0.00072 0.00159
A-Squared 0.17P-Value 0.913
Mean 0.12530StDev 0.00099Variance 0.00000Skewness -0.366071Kurtosis -0.264172N 14
Minimum 0.12340
Anderson-Darling Normality Test
95% Confidence Interval for Mean
95% Confidence Interval for Median
95% Confidence Interval for StDev
0.1270.1260.1250.124
Median
Mean
0.12640.12600.12560.12520.1248
95% Confidence Intervals
Summary Report for Diameter n= 14
TE Taylor Enterprises, Inc. www.variation.com
24
4.3 From Minitab version 18.0
1st Quartile 0.12480Median 0.125303rd Quartile 0.12620Maximum 0.12690
0.12477 0.12582
0.12484 0.12601
0.00070 0.00150
A-Squared 0.18P-Value 0.907
Mean 0.12529StDev 0.00095Variance 0.00000Skewness -0.351195Kurtosis -0.063338N 15
Minimum 0.12340
Anderson-Darling Normality Test
95% Confidence Interval for Mean
95% Confidence Interval for Median
95% Confidence Interval for StDev
0.1270.1260.1250.124
Median
Mean
0.12600.12580.12560.12540.12520.12500.1248
95% Confidence Intervals
Summary Report for Diameter n=15
4.4-6 From R version 3.2.4 - Normal
> install.packages("fBasics") > library("fBasics") > Diameter <-scan("D:\\Diameter.txt") Read 20 items > dagoTest(Diameter) Title: D'Agostino Normality Test Test Results: STATISTIC: Chi2 | Omnibus: 0.4345 Z3 | Skewness: -0.659 Z4 | Kurtosis: -0.0156 P VALUE: Omnibus Test: 0.8047 Skewness Test: 0.5099 Kurtosis Test: 0.9875
Skewness: p-value of 0.5099 is for two-sided case. For the other cases:
Upper spec limit =NORMDIST(--0.65897,0,1,TRUE) 0.745042483 Lower spec limit =NORMDIST(-0.65897,0,1,TRUE) 0.254957517
Kurtosis: The upper limit has p-value: Positive kurtosis =NORMDIST(--0.01564,0,1,TRUE) 0.506239203
TE Taylor Enterprises, Inc. www.variation.com
25
4.7-9 From R version 3.2.4 – Positive Kurtosis
> install.packages("fBasics") > library("fBasics") >PositiveKurtosis <-scan("D:\\Positive Kurtosis.txt") Read 20 items >dagoTest(PositiveKurtosis)
Title: D'Agostino Normality Test Test Results: STATISTIC: Chi2 | Omnibus: 4.4951 Z3 | Skewness: -0.1734 Z4 | Kurtosis: 2.1131 P VALUE: Omnibus Test: 0.1057 Skewness Test: 0.8624 Kurtosis Test: 0.0346
Skewness: p-value of -2.0806 is for two-sided case. For the other cases:
Upper spec limit =NORMDIST(--0.1734,0,1,TRUE) 0.568831487 Lower spec limit =NORMDIST(-0.1734,0,1,TRUE) 0.431168513
Kurtosis: The upper limit has p-value: Positive kurtosis =NORMDIST(-2.1131,0,1,TRUE) 0.017296105 4.10-12 From R version 3.2.4 – Negative Kurtosis
> install.packages("fBasics") > library("fBasics") >NegativeKurtosis <-scan("D:\\Negative Kurtosis.txt") Read 20 items >dagoTest(NegativeKurtosis)
Title: D'Agostino Normality Test Test Results: STATISTIC: Chi2 | Omnibus: 25.9043 Z3 | Skewness: 0.3238 Z4 | Kurtosis: -5.0793 P VALUE: Omnibus Test: 2.371e-06 Skewness Test: 0.7461 Kurtosis Test: 3.788e-07
Skewness: p-value of 0.7461 is for two-sided case. For the other cases:
Upper spec limit =NORMDIST(-0.3238,0,1,TRUE) 0.373044728 Lower spec limit =NORMDIST(0.3238,0,1,TRUE) 0.626955272
TE Taylor Enterprises, Inc. www.variation.com
26
Kurtosis: The upper limit has p-value:
Positive kurtosis =NORMDIST(--5.079,0,1,TRUE) 0.99999981 4.13-15 From R version 3.2.4 – Positive Skewness
> install.packages("fBasics") > library("fBasics") >PositiveSkewness <-scan("D:\\Positive Skewness.txt") Read 20 items >dagoTest(PositiveSkewness)
Title: D'Agostino Normality Test Test Results: STATISTIC: Chi2 | Omnibus: 8.3954 Z3 | Skewness: 2.6691 Z4 | Kurtosis: 1.1275 P VALUE: Omnibus Test: 0.01503 Skewness Test: 0.007605 Kurtosis Test: 0.2595
Skewness: p-value of 0.7461 is for two-sided case. For the other cases:
Upper spec limit =NORMDIST(-2.6691,0,1,TRUE) 0.00380274 Lower spec limit =NORMDIST(2.6691,0,1,TRUE) 0.99619726
Kurtosis: The upper limit has p-value: Positive kurtosis =NORMDIST(-1.1275,0,1,TRUE) 0.12976557 4.16-18 From R version 3.2.4 – Negative Skewness
> install.packages("fBasics") > library("fBasics") > NegativeSkewness <-scan("D:\\Negative Skewness.txt") Read 20 items >dagoTest(NegativeSkewness)
Title: D'Agostino Normality Test Test Results: STATISTIC: Chi2 | Omnibus: 8.3954 Z3 | Skewness: -2.6691 Z4 | Kurtosis: 1.1275 P VALUE: Omnibus Test: 0.01503 Skewness Test: 0.007605 Kurtosis Test: 0.2595
TE Taylor Enterprises, Inc. www.variation.com
27
Skewness: p-value of 0.7461 is for two-sided case. For the other cases:
Upper spec limit =NORMDIST(--2.6691,0,1,TRUE) 0.99619726 Lower spec limit =NORMDIST(-2.6691,0,1,TRUE) 0.00380274
Kurtosis: The upper limit has p-value: Positive kurtosis =NORMDIST(-1.1275,0,1,TRUE) 0.12976557 4.19-21 From R version 3.2.4 – n=200, normal
> install.packages("fBasics") > library("fBasics") > Normal200 <-scan("D:\\Normal n=200.txt") Read 20 items >dagoTest(Normal200)
Title: D'Agostino Normality Test Test Results: STATISTIC: Chi2 | Omnibus: 2.6555 Z3 | Skewness: -1.1567 Z4 | Kurtosis: 1.1478 P VALUE: Omnibus Test: 0.2651 Skewness Test: 0.2474 Kurtosis Test: 0.251
Skewness: p-value of 0.2474 is for two-sided case. For the other cases:
Upper spec limit =NORMDIST(--1.1567,0,1,TRUE) 0.876302525 Lower spec limit =NORMDIST(-1.1567,0,1,TRUE) 0.123697475
Kurtosis: The upper limit has p-value: Positive kurtosis =NORMDIST(-1.1478,0,1,TRUE) 0.125525569
TE Taylor Enterprises, Inc. www.variation.com
28
4.19-21 From Minitab version 18 – n=200, normal
1st Quartile -0.50000Median 0.100003rd Quartile 0.80000Maximum 2.50000
-0.03899 0.26199
-0.10000 0.30000
0.98282 1.19679
A-Squared 0.55P-Value 0.153
Mean 0.11150StDev 1.07923Variance 1.16474Skewness -0.196780Kurtosis 0.386000N 200
Minimum -3.40000
Anderson-Darling Normality Test
95% Confidence Interval for Mean
95% Confidence Interval for Median
95% Confidence Interval for StDev
210-1-2-3
Median
Mean
0.30.20.10.0-0.1
95% Confidence Intervals
Summary Report for n=200
TE Taylor Enterprises, Inc. www.variation.com
Appendix B
File: STAT-18 -Windows 10
TE Taylor Enterprises, Inc. www.variation.com
1
Appendix A: Test Script
1. Document System Used
1.1. Capture a screen shot showing the computer make and model
>
1.2. Capture a screen shot showing the operating system including version number
TE Taylor Enterprises, Inc. www.variation.com
2
1.3. Capture a screen shot showing the version of Excel used
2. Download Spreadsheet and Open
2.1. Download the spreadsheet from http://www.variation.com/procedures/SK_normality_tests.html. Download the protected version. Save it to the hard drive. Capture a screen shot showing the file downloaded.
2.2. Double click the spreadsheet to open it. Capture a screen shot of the section of the first tab showing the
version and date.
TE Taylor Enterprises, Inc. www.variation.com
3
Acceptance Criteria: The following is displayed “Revision 1: 8/29/2017”
Pass or Fail: Pass Initials: WT Date: 8/31/17
3. Input of Data When Normality Test Cannot Be Performed
3.1. Double click on at least 10 non-yellow cells on each of the three sheets (Req. 1d).
Acceptance Criteria: Each time a message should be displayed that the cell is protected.
Pass or Fail: Pass Initials: WT Date: 8/31/17
3.2. Go to the Skewness Kurtosis Norm Tab. Change cell D7 to “1” (Req. 1c).
Acceptance Criteria: The cell should change to “1”.
Pass or Fail: Pass Initials: WT Date: 8/31/17
3.3. Change cell B11 to “Diameter” (Req. 1b).
Acceptance Criteria: Both B11 and D11 should change to “Diameter”.
The sample size should be 0. The average, SD, skewness, kurtosis and p-value should be blank. The conclusions should indicate there is insufficient data.
Pass or Fail: Pass Initials: WT Date: 8/31/17
3.4. Change cell B12 to “A” (2-4 for invalid values. n=0)
TE Taylor Enterprises, Inc. www.variation.com
4
Acceptance Criteria: The sample size should be 0.
The average, SD, skewness, kurtosis and p-value should be blank. The conclusions should indicate there is insufficient data.
Pass or Fail: Pass Initials: WT Date: 8/31/17
3.5. Change cell B14 to “1” (Req. 2-4 for n=1)
Acceptance Criteria: The sample size and average should be 1.
The standard deviation, skewness, kurtosis and p-value should be blank. The conclusions should indicate there is insufficient data.
Pass or Fail: Pass Initials: WT Date: 8/31/17
3.6. Change cell B16 to “1” (Req. 2-4 for n=2)
TE Taylor Enterprises, Inc. www.variation.com
5
Acceptance Criteria: The sample size should be 2.
The average should be 1. The standard deviation should be 0. The skewness, kurtosis and p-value should be blank. The conclusions should indicate there is insufficient data.
Pass or Fail: Pass Initials: WT Date: 8/31/17
3.7. Change cell B17 to “1” (Req. 2-4 for n=3 and SD=0)
Acceptance Criteria: The sample size should be 3.
The average should be 1. The standard deviation should be 0. The skewness, kurtosis and p-value should be blank. The conclusions should indicate there is insufficient data.
Pass or Fail: Pass Initials: WT Date: 8/31/17
3.8. Change cell B18 to “1” (Req. 2-4 for n=4 and SD=0)
TE Taylor Enterprises, Inc. www.variation.com
6
Acceptance Criteria: The sample size should be 4.
The average should be 1. The standard deviation should be 0. The skewness, kurtosis and p-value should be blank. The conclusions should indicate there is insufficient data.
Pass or Fail: Pass Initials: WT Date: 8/31/17
3.9. Copy cell B18 to clipboard. Select cells B12 to B211 and then select paste “1” (Req. 1a, 2-4 for n=200 and SD=0)
TE Taylor Enterprises, Inc. www.variation.com
7
TE Taylor Enterprises, Inc. www.variation.com
8
Acceptance Criteria: The data column should be filled with “1”. The sample size should be 200. The average should be 1. The standard deviation should be 0. The skewness, kurtosis and p-value should be blank. The conclusions should indicate there is insufficient data.
Pass or Fail: Pass Initials: WT Date: 8/31/17
4. Input of Data When Normality Test Can Be Performed
4.1. Clear Cells B12:B211. Then paste the following into cells B12:B15. (Req. 2-4 for n=4 and SD>0).
0.1257 0.1249 0.1257 0.125
Acceptance Criteria: The sample size should be 4.
The average should round to 0.12532. The standard deviation should round to 0.00043. The skewness should round to -0.04558. The kurtosis should round to -5.73743. The p-value should be blank. The conclusions should indicate there is insufficient data.
Pass or Fail: Pass Initials: WT Date: 8/31/17
4.2. Paste the following into cells B16:B25 (Req. 2-4 for n=14 and SD>0).
0.1264
0.1238
0.1254
TE Taylor Enterprises, Inc. www.variation.com
9
0.1248
0.1253
0.1234
0.1262
0.1262
0.1269
0.1245
Acceptance Criteria: The sample size should be 14.
The average should round to 0.12530. The standard deviation should round to 0.00099. The skewness should round to -0.366071. The kurtosis should round to -0.264172. The p-value should be blank. The conclusions should indicate there is insufficient data.
Pass or Fail: Pass Initials: WT Date: 8/31/17
4.3. Type 0.1252 into cell B26 (Req. 3-4 for n>=15, Normal data, Invalid type of specification).
TE Taylor Enterprises, Inc. www.variation.com
10
Acceptance Criteria: The sample size should be 15.
A p-value should be given. The conclusions for the SK All test should be it passes. The conclusion for the SK Spec test should be incorrect type of specification.
Pass or Fail: Pass Initials: WT Date: 8/31/17
4.4. Change cell D7 to “T” and paste cells D10:29 of Example Data Sets tab into cells B12:31 (Req. 3-4 for n>=15, normal data, two-sided specification)
TE Taylor Enterprises, Inc. www.variation.com
11
Skewness = -0.29569 Y = -0.62532 B2 = 3.577778 W2 = 1.270585 Delta = 2.8899 Alpha = 2.718712 Z = -0.65897 p-value <> = 0.509914 p-value > = 0.745043 p-value < = 0.254957 Kurtosis = 2.549802 Eb2 = 2.714286 VarB2 = 0.579237 x = -0.21612 3StdMom = 1.737522 A = 18.32027 0.969103 Z= -0.01564 p-value <> = 0.987519 p-value > = 0.50624 p-value < = 0.49376 K^2 0.434488 p-value 0.804734
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.8047. The conclusions for the SK All test should be it passes. The conclusion for the SK Spec test should be it passes for a two-sided specification. Cell J17 should round to -0.659 (ZS). Cell J18 should round to 0.5099 (p-value 2-sided skewness). Cell J29 should round to -0.0156 (ZK). Cell J31 should round to 0.5062 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: Pass Initials: WT Date: 8/31/17
4.5. Change cell D7 to “U” (Req. 3-4 for n>=15, normal data, upper specification)
TE Taylor Enterprises, Inc. www.variation.com
12
Skewness = -0.29569 Y = -0.62532 B2 = 3.577778 W2 = 1.270585 Delta = 2.8899 Alpha = 2.718712 Z = -0.65897 p-value <> = 0.509914 p-value > = 0.745043 p-value < = 0.254957 Kurtosis = 2.549802 Eb2 = 2.714286 VarB2 = 0.579237 x = -0.21612 3StdMom = 1.737522 A = 18.32027 0.969103 Z= -0.01564 p-value <> = 0.987519 p-value > = 0.50624 p-value < = 0.49376 K^2 0.434488 p-value 0.804734
TE Taylor Enterprises, Inc. www.variation.com
13
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.8047. The conclusions for the SK All test should be it passes. The conclusion for the SK Spec test should be it passes for an upper specification. Cell J17 should round to -0.659 (ZS). Cell J19 should round to 0.7450 (p-value positive skewness). Cell J29 should round to -0.0156 (ZK) Cell J31 should round to 0.5062 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: Pass Initials: WT Date: 8/31/17
4.6. Change cell D7 to “L” (Req. 3-4 for n>=15, normal data, lower specification)
Skewness = -0.29569 Y = -0.62532 B2 = 3.577778 W2 = 1.270585 Delta = 2.8899 Alpha = 2.718712 Z = -0.65897 p-value <> = 0.509914 p-value > = 0.745043 p-value < = 0.254957 Kurtosis = 2.549802 Eb2 = 2.714286 VarB2 = 0.579237
TE Taylor Enterprises, Inc. www.variation.com
14
x = -0.21612 3StdMom = 1.737522 A = 18.32027 0.969103 Z= -0.01564 p-value <> = 0.987519 p-value > = 0.50624 p-value < = 0.49376 K^2 0.434488 p-value 0.804734
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.8047. The conclusions for the SK All test should be it passes. The conclusion for the SK Spec test should be it passes for a lower specification. Cell J17 should round to -0.659 (ZS). Cell J20 should round to 0.2550 (p-value negative skewness). Cell J29 should round to -0.0156 (ZK). Cell J31 should round to 0.5062 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: Pass Initials: WT Date: 8/31/17
4.7. Change cell D7 to “T” and cell B12:B31 to the values below (Req. 3-4 for n>=15, positive kurtosis data, two-sided specification)
0.1 -1 0.2 0.2 -2.9 2.6 0.1 1.1 -0.9 -0.2 -0.6 -0.3 1.3 0.4 0
-0.3 -0.1 -0.2 0.1 -1.4
TE Taylor Enterprises, Inc. www.variation.com
15
Skewness = -0.07718 Y = -0.16321 B2 = 3.577778 W2 = 1.270585 Delta = 2.8899 Alpha = 2.718712 Z = -0.17338 p-value <> = 0.862353 p-value > = 0.568824 p-value < = 0.431176 Kurtosis = 4.890742 Eb2 = 2.714286 VarB2 = 0.579237 x = 2.859708 3StdMom = 1.737522 A = 18.32027 0.430621 Z= 2.113056 p-value <> = 0.034596 p-value > = 0.017298 p-value < = 0.982702 K^2 4.495066 p-value 0.10566
TE Taylor Enterprises, Inc. www.variation.com
16
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.1057. The conclusions for the SK All test should be it passes. The conclusion for the SK Spec test should be it fails for a two-sided specification. Cell J17 should round to -0.1734 (ZS). Cell J18 should round to 0.8624 (p-value 2-sided skewness). Cell J29 should round to 2.1131 (ZK). Cell J31 should round to 0.0173 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: Pass Initials: WT Date: 8/31/17
4.8. Change cell D7 to “U” (Req. 3-4 for n>=15, normal data, upper specification)
Skewness = -0.07718 Y = -0.16321 B2 = 3.577778 W2 = 1.270585 Delta = 2.8899 Alpha = 2.718712 Z = -0.17338 p-value <> = 0.862353 p-value > = 0.568824 p-value < = 0.431176 Kurtosis = 4.890742 Eb2 = 2.714286 VarB2 = 0.579237
TE Taylor Enterprises, Inc. www.variation.com
17
x = 2.859708 3StdMom = 1.737522 A = 18.32027 0.430621 Z= 2.113056 p-value <> = 0.034596 p-value > = 0.017298 p-value < = 0.982702 K^2 4.495066 p-value 0.10566
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.1057. The conclusions for the SK All test should be it passes. The conclusion for the SK Spec test should be it fails for an upper specification. Cell J17 should round to -0.1734 (ZS). Cell J19 should round to 0.5688 (p-value positive skewness). Cell J29 should round to 2.1131 (ZK). Cell J31 should round to 0.0173 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: Pass Initials: WT Date: 8/31/17
4.9. Change cell D7 to “L” (Req. 3-4 for n>=15, normal data, lower specification)
Skewness = -0.07718 Y = -0.16321 B2 = 3.577778 W2 = 1.270585 Delta = 2.8899
TE Taylor Enterprises, Inc. www.variation.com
18
Alpha = 2.718712 Z = -0.17338 p-value <> = 0.862353 p-value > = 0.568824 p-value < = 0.431176 Kurtosis = 4.890742 Eb2 = 2.714286 VarB2 = 0.579237 x = 2.859708 3StdMom = 1.737522 A = 18.32027 0.430621 Z= 2.113056 p-value <> = 0.034596 p-value > = 0.017298 p-value < = 0.982702 K^2 4.495066 p-value 0.10566
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.1057. The conclusions for the SK All test should be it passes. The conclusion for the SK Spec test should be it fails for a lower specification. Cell J17 should round to -0.1734 (ZS). Cell J20 should round to 0.4312 (p-value negative skewness). Cell J29 should round to 2.1131 (ZK). Cell J31 should round to 0.0173 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: Pass Initials: WT Date: 8/31/17
4.10. Change cell D7 to “T” and cell B12:B31 to the values below (Req. 3-4 for n>=15, negative kurtosis data, two-sided specification)
-1.1 -0.5 -0.4 1.1 -1 1.2 -0.8 1.2 -1.2 -1 0.6 -1.1 1.2
TE Taylor Enterprises, Inc. www.variation.com
19
-1.2 1
-0.8 -1.1
1 1.2 1.2
Skewness = 0.144366
Y = 0.3053 B2 = 3.577778 W2 = 1.270585
Delta = 2.8899 Alpha = 2.718712
Z = 0.323845 p-value <> = 0.746055 p-value > = 0.373028 p-value < = 0.626972
Kurtosis = 1.167515 Eb2 = 2.714286
VarB2 = 0.579237 x = -2.03235
3StdMom = 1.737522 A = 18.32027
TE Taylor Enterprises, Inc. www.variation.com
20
3.704331
Z= -5.07932 p-value <> = 3.79E-07 p-value > = 1 p-value < = 1.89E-07
K^2 25.90433 p-value 2.37E-06
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.0000 The conclusions for the SK All test should be it fails. The conclusion for the SK Spec test should be it passes for a two-sided specification. Cell J17 should round to 0.3238 (ZS). Cell J18 should round to 0.7461 (p-value 2-sided skewness). Cell J29 should round to -5.0793 (ZK) Cell J31 should round to 1.0000 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: Pass Initials: WT Date: 8/31/17
4.11. Change cell D7 to “U” (Req. 3-4 for n>=15, normal data, upper specification)
Skewness = 0.144366 Y = 0.3053 B2 = 3.577778 W2 = 1.270585 Delta = 2.8899
TE Taylor Enterprises, Inc. www.variation.com
21
Alpha = 2.718712 Z = 0.323845 p-value <> = 0.746055 p-value > = 0.373028 p-value < = 0.626972 Kurtosis = 1.167515 Eb2 = 2.714286 VarB2 = 0.579237 x = -2.03235 3StdMom = 1.737522 A = 18.32027 3.704331 Z= -5.07932 p-value <> = 3.79E-07 p-value > = 1 p-value < = 1.89E-07 K^2 25.90433 p-value 2.37E-06
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.0000. The conclusions for the SK All test should be it fails. The conclusion for the SK Spec test should be it passes for an upper specification. Cell J17 should round to 0.3238 (ZS). Cell J19 should round to 0.3730 (p-value positive skewness). Cell J29 should round to -5.0793 (ZK). Cell J31 should round to 1.0000 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: Pass Initials: WT Date: 8/31/17
4.12. Change cell D7 to “L” (Req. 3-4 for n>=15, normal data, lower specification)
TE Taylor Enterprises, Inc. www.variation.com
22
Skewness = 0.144366 Y = 0.3053 B2 = 3.577778 W2 = 1.270585 Delta = 2.8899 Alpha = 2.718712 Z = 0.323845 p-value <> = 0.746055 p-value > = 0.373028 p-value < = 0.626972 Kurtosis = 1.167515 Eb2 = 2.714286 VarB2 = 0.579237 x = -2.03235 3StdMom = 1.737522 A = 18.32027 3.704331 Z= -5.07932 p-value <> = 3.79E-07 p-value > = 1 p-value < = 1.89E-07 K^2 25.90433 p-value 2.37E-06
TE Taylor Enterprises, Inc. www.variation.com
23
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.0000. The conclusions for the SK All test should be it fails. The conclusion for the SK Spec test should be it passes for a lower specification. Cell J17 should round to 0.3238 (ZS). Cell J20 should round to 0.6270 (p-value negative skewness). Cell J29 should round to -5.0793 (ZK). Cell J31 should round to 1.0000 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: Pass Initials: WT Date: 8/31/17
4.13. Change cell D7 to “T” and cell B12:B31 to the values below (Req. 3-4 for n>=15, positive skewness data, two-sided specification)
-0.7 -0.4 -0.4 -0.6 1.8 -0.4 -0.4 1.1 1.5 -0.4 0.3 -0.7 -0.8 -0.8 -0.8 -0.8 -0.8 -0.8 -0.3 0.7
TE Taylor Enterprises, Inc. www.variation.com
24
Skewness = 1.363523 Y = 2.883529 B2 = 3.577778 W2 = 1.270585 Delta = 2.8899 Alpha = 2.718712 Z = 2.669104 p-value <> = 0.007605 p-value > = 0.003803 p-value < = 0.996197 Kurtosis = 3.493619 Eb2 = 2.714286 VarB2 = 0.579237 x = 1.023988 3StdMom = 1.737522 A = 18.32027 0.644279 Z= 1.127517 p-value <> = 0.259524 p-value > = 0.129762 p-value < = 0.870238 K^2 8.395412 p-value 0.01503
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.0150.
TE Taylor Enterprises, Inc. www.variation.com
25
The conclusions for the SK All test should be it fails. The conclusion for the SK Spec test should be it fails for a two-sided specification. Cell J17 should round to 2.6691 (ZS). Cell J18 should round to 0.0076 (p-value 2-sided skewness). Cell J29 should round to 1.1275 (ZK). Cell J31 should round to 0.1298 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: Pass Initials: WT Date: 8/31/17
4.14. Change cell D7 to “U” (Req. 3-4 for n>=15, normal data, upper specification)
Skewness = 1.363523 Y = 2.883529 B2 = 3.577778 W2 = 1.270585 Delta = 2.8899 Alpha = 2.718712 Z = 2.669104 p-value <> = 0.007605 p-value > = 0.003803 p-value < = 0.996197 Kurtosis = 3.493619 Eb2 = 2.714286 VarB2 = 0.579237 x = 1.023988 3StdMom = 1.737522 A = 18.32027
TE Taylor Enterprises, Inc. www.variation.com
26
0.644279 Z= 1.127517 p-value <> = 0.259524 p-value > = 0.129762 p-value < = 0.870238 K^2 8.395412 p-value 0.01503
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.0150. The conclusions for the SK All test should be it fails. The conclusion for the SK Spec test should be it fails for an upper specification. Cell J17 should round to 2.6691 (ZS). Cell J19 should round to 0.0038 (p-value positive skewness). Cell J29 should round to 1.1275 (ZK). Cell J31 should round to 0.1298 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: Pass Initials: WT Date: 8/31/17
4.15. Change cell D7 to “L” (Req. 3-4 for n>=15, normal data, lower specification)
Skewness = 1.363523 Y = 2.883529 B2 = 3.577778 W2 = 1.270585 Delta = 2.8899
TE Taylor Enterprises, Inc. www.variation.com
27
Alpha = 2.718712 Z = 2.669104 p-value <> = 0.007605 p-value > = 0.003803 p-value < = 0.996197 Kurtosis = 3.493619 Eb2 = 2.714286 VarB2 = 0.579237 x = 1.023988 3StdMom = 1.737522 A = 18.32027 0.644279 Z= 1.127517 p-value <> = 0.259524 p-value > = 0.129762 p-value < = 0.870238 K^2 8.395412 p-value 0.01503
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.0150. The conclusions for the SK All test should be it fails. The conclusion for the SK Spec test should be it passes for a lower specification. Cell J17 should round to 2.6691 (ZS). Cell J20 should round to 0.9962 (p-value negative skewness). Cell J29 should round to 1.1275 (ZK). Cell J31 should round to 0.1298 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: Pass Initials: WT Date: 8/31/17
4.16. Change cell D7 to “T” and cell B12:B31 to the values below (Req. 3-4 for n>=15, negative skewness data, two-sided specification)
0.7 0.4 0.4 0.6 -1.8 0.4 0.4 -1.1 -1.5 0.4 -0.3
TE Taylor Enterprises, Inc. www.variation.com
28
0.7 0.8 0.8 0.8 0.8 0.8 0.8 0.3 -0.7
Skewness = -1.36352 Y = -2.88353 B2 = 3.577778 W2 = 1.270585 Delta = 2.8899 Alpha = 2.718712 Z = -2.6691 p-value <> = 0.007605 p-value > = 0.996197 p-value < = 0.003803 Kurtosis = 3.493619 Eb2 = 2.714286
TE Taylor Enterprises, Inc. www.variation.com
29
VarB2 = 0.579237 x = 1.023988 3StdMom = 1.737522 A = 18.32027 0.644279 Z= 1.127517 p-value <> = 0.259524 p-value > = 0.129762 p-value < = 0.870238 K^2 8.395412 p-value 0.01503
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.0150. The conclusions for the SK All test should be it fails. The conclusion for the SK Spec test should be it fails for a two-sided specification. Cell J17 should round to -2.6691 (ZS). Cell J18 should round to 0.0076 (p-value 2-sided skewness). Cell J29 should round to 1.1275 (ZK). Cell J31 should round to 0.1298 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: Pass Initials: WT Date: 8/31/17
4.17. Change cell D7 to “U” (Req. 3-4 for n>=15, normal data, upper specification)
Skewness = -1.36352
TE Taylor Enterprises, Inc. www.variation.com
30
Y = -2.88353 B2 = 3.577778 W2 = 1.270585 Delta = 2.8899 Alpha = 2.718712 Z = -2.6691 p-value <> = 0.007605 p-value > = 0.996197 p-value < = 0.003803 Kurtosis = 3.493619 Eb2 = 2.714286 VarB2 = 0.579237 x = 1.023988 3StdMom = 1.737522 A = 18.32027 0.644279 Z= 1.127517 p-value <> = 0.259524 p-value > = 0.129762 p-value < = 0.870238 K^2 8.395412 p-value 0.01503
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.0150. The conclusions for the SK All test should be it fails. The conclusion for the SK Spec test should be it passes for an upper specification. Cell J17 should round to -2.6691 (ZS). Cell J19 should round to 0.9962 (p-value positive skewness). Cell J29 should round to 1.1275 (ZK). Cell J31 should round to 0.1298 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: Pass Initials: WT Date: 8/31/17
4.18. Change cell D7 to “L” (Req. 3-4 for n>=15, normal data, lower specification)
TE Taylor Enterprises, Inc. www.variation.com
31
Skewness = -1.36352 Y = -2.88353 B2 = 3.577778 W2 = 1.270585 Delta = 2.8899 Alpha = 2.718712 Z = -2.6691 p-value <> = 0.007605 p-value > = 0.996197 p-value < = 0.003803 Kurtosis = 3.493619 Eb2 = 2.714286 VarB2 = 0.579237 x = 1.023988 3StdMom = 1.737522 A = 18.32027 0.644279 Z= 1.127517 p-value <> = 0.259524 p-value > = 0.129762 p-value < = 0.870238 K^2 8.395412 p-value 0.01503
TE Taylor Enterprises, Inc. www.variation.com
32
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.0150. The conclusions for the SK All test should be it fails. The conclusion for the SK Spec test should be it fails for a lower specification. Cell J17 should round to -2.6691 (ZS). Cell J20 should round to 0.0038 (p-value negative skewness). Cell J29 should round to 1.1275 (ZK). Cell J31 should round to 0.1298 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: Pass Initials: WT Date: 8/31/17
4.19. Change cell D7 to “T” and cell B12:B211 to the values below (Req. 3-4 for n=200, normal data, two-sided specification)
2.1 0.1 -0.5 0.1 -0.6
1 -0.4 0.1 -2.4 -1
0.6 0.1 -0.3 2.3 -1
0.3 -0.3
2 -0.9 0.1 0.8 -1.5
1 2.1 0.7 -0.1 -0.4 0.6 -1
-0.5 0.4 0.2 -0.1
TE Taylor Enterprises, Inc. www.variation.com
33
-0.3 -2.5 -0.2 0.3 -0.8 -0.2 -3
1.9 0.7 -0.2 0.4 -3.4 0.9 -1.8 0.8 0
-0.2 0
-0.1 0.8 -1.1 1.4 2.5 0.5 -0.1 -1.8 0.9 0.1 0.2 1.3 0.1 -0.1 0.7 -0.9 0.3 -0.2 -0.1 1.3 2.3 0.5 0.6 1.1 1
-2.2
TE Taylor Enterprises, Inc. www.variation.com
34
2.5 1.6 1.2 1.7 -0.9 2.1 0.6 0.6 0.3 1.1 0.5 -0.9 0.7 -0.4 -0.7 0.9 0.1 0.5 -0.2 -1.3 -1
-2.3 1.3 -0.5 -0.3 1.9 1
-0.2 -0.1 1.2 -0.7 2.1 1.1 1.4 -0.5
2 -0.3 0.3 -0.2 1.3 1.1 -1.3 -0.6 -0.8
TE Taylor Enterprises, Inc. www.variation.com
35
-0.2 1.2 -0.7 -2.7 0.1 -1.1 -0.8 -0.8 -0.3 1.1 0
-0.5 1.6 0.5 0.5 -1.6 2.4 -0.3 0.6 -0.5
0 0.8 1.9 -0.4 0.3 -0.2 1.6 -0.5 -0.8 -1.2 -1.6 0.4 1.3 0.4 -1.4 -0.6 -0.5 -0.4 0.2 1.4 -1.5 1.3 -0.2 -0.1
TE Taylor Enterprises, Inc. www.variation.com
36
-0.8 0.7 -0.3 -1.2 0.5 -0.7 -0.7 0.7 -0.3 -1
0.5 0.6 1.6 1 1
0.7 -0.1 -0.9 -0.1
1 0.7 0.1 0.6 0.2 1.6 -0.5 -0.7 -1.1 0.4 -0.6 1.9 -0.7 -0.2 -0.5 -0.7
TE Taylor Enterprises, Inc. www.variation.com
37
TE Taylor Enterprises, Inc. www.variation.com
38
Skewness = -0.1953 Y = -1.14457 B2 = 3.159828 W2 = 1.078378 Delta = 5.14828 Alpha = 5.051471 Z = -1.15675 p-value <> = 0.247376 p-value > = 0.876312 p-value < = 0.123688 Kurtosis = 3.346567 Eb2 = 2.970149 VarB2 = 0.11136 x = 1.127987 3StdMom = 0.967701 A = 42.06667 0.756788 Z= 1.147805 p-value <> = 0.251049 p-value > = 0.125525 p-value < = 0.874475 K^2 2.655517 p-value 0.265071
Acceptance Criteria: The sample size should be 200. The average should round to 0.11150. The standard deviation should round to 1.07923. The skewness should round to -0.196780. The kurtosis should round to 0.386000. The p-value should round to 0.2651. The conclusions for the SK All test should be it passes. The conclusion for the SK Spec test should be it passes for a two-sided specification. Cell J17 should round to -1.1567 (ZS). Cell J18 should round to 0.2474 (p-value 2-sided skewness). Cell J29 should round to 1.1478 (ZK). Cell J31 should round to 0.1255 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: Pass Initials: WT Date: 8/31/17
4.20. Change cell D7 to “U” (Req. 3-4 for n=200, normal data, upper specification)
TE Taylor Enterprises, Inc. www.variation.com
39
Skewness = -0.1953 Y = -1.14457 B2 = 3.159828 W2 = 1.078378 Delta = 5.14828 Alpha = 5.051471 Z = -1.15675 p-value <> = 0.247376 p-value > = 0.876312 p-value < = 0.123688 Kurtosis = 3.346567 Eb2 = 2.970149 VarB2 = 0.11136 x = 1.127987 3StdMom = 0.967701 A = 42.06667 0.756788 Z= 1.147805 p-value <> = 0.251049 p-value > = 0.125525 p-value < = 0.874475 K^2 2.655517 p-value 0.265071
Acceptance Criteria: The sample size should be 200. The p-value should round to 0.2651. The conclusions for the SK All test should be it passes.
TE Taylor Enterprises, Inc. www.variation.com
40
The conclusion for the SK Spec test should be it passes for an upper specification. Cell J17 should round to -1.1567 (ZS). Cell J19 should round to 0.8763 (p-value positive skewness). Cell J29 should round to 1.1478 (ZK). Cell J31 should round to 0.1255 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: Pass Initials: WT Date: 8/31/17
4.21. Change cell D7 to “L” (Req. 3-4 for n=200, normal data, lower specification)
Skewness = -0.1953 Y = -1.14457 B2 = 3.159828 W2 = 1.078378 Delta = 5.14828 Alpha = 5.051471 Z = -1.15675 p-value <> = 0.247376 p-value > = 0.876312 p-value < = 0.123688 Kurtosis = 3.346567 Eb2 = 2.970149 VarB2 = 0.11136 x = 1.127987 3StdMom = 0.967701 A = 42.06667 0.756788
TE Taylor Enterprises, Inc. www.variation.com
41
Z= 1.147805 p-value <> = 0.251049 p-value > = 0.125525 p-value < = 0.874475 K^2 2.655517 p-value 0.265071
Acceptance Criteria: The sample size should be 200. The p-value should round to 0.2651. The conclusions for the SK All test should be it passes. The conclusion for the SK Spec test should be it passes for a lower specification. Cell J17 should round to -1.1567 (ZS). Cell J20 should round to 0.1237 (p-value negative skewness). Cell J29 should round to 1.1478 (ZK). Cell J31 should round to 0.1255 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: Pass Initials: WT Date: 8/31/17
5. Independence of tab
5.1. Copy the Skewness Kurtosis Tests tab. Delete all tab but the copy.
TE Taylor Enterprises, Inc. www.variation.com
42
Skewness = -0.1953 Y = -1.14457 B2 = 3.159828 W2 = 1.078378 Delta = 5.14828 Alpha = 5.051471 Z = -1.15675 p-value <> = 0.247376 p-value > = 0.876312 p-value < = 0.123688 Kurtosis = 3.346567 Eb2 = 2.970149 VarB2 = 0.11136 x = 1.127987 3StdMom = 0.967701 A = 42.06667 0.756788 Z= 1.147805
TE Taylor Enterprises, Inc. www.variation.com
43
p-value <> = 0.251049 p-value > = 0.125525 p-value < = 0.874475 K^2 2.655517 p-value 0.265071
Acceptance Criteria: The sample size should be 200. The average should round to 0.11150. The standard deviation should round to 1.07923. The skewness should round to -0.196780. The kurtosis should round to 0.386000. The p-value should round to 0.2651. The conclusions for the SK All test should be it passes. The conclusion for the SK Spec test should be it passes for a lower specification. Cell J17 should round to -1.1567 (ZS). Cell J20 should round to 0.1237 (p-value negative skewness). Cell J29 should round to 1.1478 (ZK). Cell J31 should round to 0.1255 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: Pass Initials: WT Date: 8/31/17
6. Saving and Signing File
6.1. Determine if all tests passed
All Test Cases for Computer Pass? X yes no
6.2. Save File and show name and date.
Name of File: STAT-18 – Windows 10
Data of File: August 31, 2017
Signature: ________________________________ Date: August 31, 2017
TE Taylor Enterprises, Inc. www.variation.com
Appendix C
File: STAT-18 -Windows 7
TE Taylor Enterprises, Inc. www.variation.com
1
Appendix A: Test Script
1. Document System Used
1.1. Capture a screen shot showing the computer make and model
1.2. Capture a screen shot showing the operating system including version number
TE Taylor Enterprises, Inc. www.variation.com
2
1.3. Capture a screen shot showing the version of Excel used
2. Download Spreadsheet and Open
2.1. Download the spreadsheet from http://www.variation.com/procedures/SK_normality_tests.html. Download the protected version. Save it to the hard drive. Capture a screen shot showing the file downloaded.
2.2. Double click the spreadsheet to open it. Capture a screen shot of the section of the first tab showing the
version and date.
TE Taylor Enterprises, Inc. www.variation.com
3
Acceptance Criteria: The following is displayed “Revision 1: 8/30/2017”
Pass or Fail: Pass Initials: WT Date: 9/1/17
3. Input of Data When Normality Test Cannot Be Performed
3.1. Double click on at least 10 non-yellow cells on each of the three sheets (Req. 1d).
Acceptance Criteria: Each time a message should be displayed that the cell is protected.
Pass or Fail: Pass Initials: WT Date: 9/1/17
3.2. Go to the Skewness Kurtosis Norm Tab. Change cell D7 to “1” (Req. 1c).
TE Taylor Enterprises, Inc. www.variation.com
4
Acceptance Criteria: The cell should change to “1”.
Pass or Fail: Pass Initials: WT Date: 9/1/17
3.3. Change cell B11 to “Diameter” (Req. 1b).
Acceptance Criteria: Both B11 and D11 should change to “Diameter”.
The sample size should be 0. The average, SD, skewness, kurtosis and p-value should be blank. The conclusions should indicate there is insufficient data.
TE Taylor Enterprises, Inc. www.variation.com
5
Pass or Fail: Pass Initials: WT Date: 9/1/17
3.4. Change cell B12 to “A” (2-4 for invalid values. n=0)
Acceptance Criteria: The sample size should be 0.
The average, SD, skewness, kurtosis and p-value should be blank. The conclusions should indicate there is insufficient data.
Pass or Fail: Pass Initials: WT Date: 9/1/17
3.5. Change cell B14 to “1” (Req. 2-4 for n=1)
TE Taylor Enterprises, Inc. www.variation.com
6
Acceptance Criteria: The sample size and average should be 1. The standard deviation, skewness, kurtosis and p-value should be blank. The conclusions should indicate there is insufficient data.
Pass or Fail: Pass Initials: WT Date: 9/1/17
3.6. Change cell B16 to “1” (Req. 2-4 for n=2)
Acceptance Criteria: The sample size should be 2.
The average should be 1. The standard deviation should be 0. The skewness, kurtosis and p-value should be blank. The conclusions should indicate there is insufficient data.
Pass or Fail: Pass Initials: WT Date: 9/1/17
3.7. Change cell B17 to “1” (Req. 2-4 for n=3 and SD=0)
TE Taylor Enterprises, Inc. www.variation.com
7
Acceptance Criteria: The sample size should be 3.
The average should be 1. The standard deviation should be 0. The skewness, kurtosis and p-value should be blank. The conclusions should indicate there is insufficient data.
Pass or Fail: Pass Initials: WT Date: 9/1/17
3.8. Change cell B18 to “1” (Req. 2-4 for n=4 and SD=0)
Acceptance Criteria: The sample size should be 4.
The average should be 1. The standard deviation should be 0.
TE Taylor Enterprises, Inc. www.variation.com
8
The skewness, kurtosis and p-value should be blank. The conclusions should indicate there is insufficient data.
Pass or Fail: Pass Initials: WT Date: 9/1/17
3.9. Copy cell B18 to clipboard. Select cells B12 to B211 and then select paste “1” (Req. 1a, 2-4 for n=200 and SD=0)
Acceptance Criteria: The data column should be filled with “1”.
The sample size should be 200. The average should be 1.
TE Taylor Enterprises, Inc. www.variation.com
9
The standard deviation should be 0. The skewness, kurtosis and p-value should be blank. The conclusions should indicate there is insufficient data.
Pass or Fail: Pass Initials: WT Date: 9/1/17
4. Input of Data When Normality Test Can Be Performed
4.1. Clear Cells B12:B211. Then paste the following into cells B12:B15. (Req. 2-4 for n=4 and SD>0).
0.1257 0.1249 0.1257 0.125
Acceptance Criteria: The sample size should be 4.
The average should round to 0.12532. The standard deviation should round to 0.00043. The skewness should round to -0.04558. The kurtosis should round to -5.73743. The p-value should be blank. The conclusions should indicate there is insufficient data.
Pass or Fail: Pass Initials: WT Date: 9/1/17
4.2. Paste the following into cells B16:B25 (Req. 2-4 for n=14 and SD>0).
0.1264
0.1238
0.1254
0.1248
0.1253
0.1234
TE Taylor Enterprises, Inc. www.variation.com
10
0.1262
0.1262
0.1269
0.1245
Acceptance Criteria: The sample size should be 14.
The average should round to 0.12530. The standard deviation should round to 0.00099. The skewness should round to -0.366071. The kurtosis should round to -0.264172. The p-value should be blank. The conclusions should indicate there is insufficient data.
Pass or Fail: Pass Initials: WT Date: 9/1/17
4.3. Type 0.1252 into cell B26 (Req. 3-4 for n>=15, Normal data, Invalid type of specification).
TE Taylor Enterprises, Inc. www.variation.com
11
Acceptance Criteria: The sample size should be 15.
A p-value should be given. The conclusions for the SK All test should be it passes. The conclusion for the SK Spec test should be incorrect type of specification.
Pass or Fail: Pass Initials: WT Date: 9/1/17
4.4. Change cell D7 to “T” and paste cells D10:29 of Example Data Sets tab into cells B12:31 (Req. 3-4 for n>=15, normal data, two-sided specification)
TE Taylor Enterprises, Inc. www.variation.com
12
Skewness = -0.29569 Y = -0.62532 B2 = 3.577778 W2 = 1.270585 Delta = 2.8899 Alpha = 2.718712 Z = -0.65897 p-value <> = 0.509914 p-value > = 0.745043 p-value < = 0.254957 Kurtosis = 2.549802 Eb2 = 2.714286 VarB2 = 0.579237 x = -0.21612 3StdMom = 1.737522 A = 18.32027 0.969103 Z= -0.01564 p-value <> = 0.987519 p-value > = 0.50624 p-value < = 0.49376
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.8047. The conclusions for the SK All test should be it passes. The conclusion for the SK Spec test should be it passes for a two-sided specification. Cell J17 should round to -0.659 (ZS). Cell J18 should round to 0.5099 (p-value 2-sided skewness). Cell J29 should round to -0.0156 (ZK). Cell J31 should round to 0.5062 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: Pass Initials: WT Date: 9/1/17
4.5. Change cell D7 to “U” (Req. 3-4 for n>=15, normal data, upper specification)
TE Taylor Enterprises, Inc. www.variation.com
13
Skewness = -0.29569 Y = -0.62532 B2 = 3.577778 W2 = 1.270585 Delta = 2.8899 Alpha = 2.718712 Z = -0.65897 p-value <> = 0.509914 p-value > = 0.745043 p-value < = 0.254957 Kurtosis = 2.549802 Eb2 = 2.714286 VarB2 = 0.579237 x = -0.21612 3StdMom = 1.737522 A = 18.32027 0.969103 Z= -0.01564 p-value <> = 0.987519 p-value > = 0.50624 p-value < = 0.49376
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.8047. The conclusions for the SK All test should be it passes. The conclusion for the SK Spec test should be it passes for an upper
TE Taylor Enterprises, Inc. www.variation.com
14
specification. Cell J17 should round to -0.659 (ZS). Cell J19 should round to 0.7450 (p-value positive skewness). Cell J29 should round to -0.0156 (ZK) Cell J31 should round to 0.5062 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: Pass Initials: WT Date: 9/1/17
4.6. Change cell D7 to “L” (Req. 3-4 for n>=15, normal data, lower specification)
Skewness = -0.29569 Y = -0.62532 B2 = 3.577778 W2 = 1.270585 Delta = 2.8899 Alpha = 2.718712 Z = -0.65897 p-value <> = 0.509914 p-value > = 0.745043 p-value < = 0.254957 Kurtosis = 2.549802 Eb2 = 2.714286 VarB2 = 0.579237 x = -0.21612 3StdMom = 1.737522 A = 18.32027 0.969103 Z= -0.01564
TE Taylor Enterprises, Inc. www.variation.com
15
p-value <> = 0.987519 p-value > = 0.50624 p-value < = 0.49376 K^2 0.434488 p-value 0.804734
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.8047. The conclusions for the SK All test should be it passes. The conclusion for the SK Spec test should be it passes for a lower specification. Cell J17 should round to -0.659 (ZS). Cell J20 should round to 0.2550 (p-value negative skewness). Cell J29 should round to -0.0156 (ZK). Cell J31 should round to 0.5062 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: Pass Initials: WT Date: 9/1/17
4.7. Change cell D7 to “T” and cell B12:B31 to the values below (Req. 3-4 for n>=15, positive kurtosis data, two-sided specification)
0.1 -1 0.2 0.2 -2.9 2.6 0.1 1.1 -0.9 -0.2 -0.6 -0.3 1.3 0.4 0
-0.3 -0.1 -0.2 0.1 -1.4
TE Taylor Enterprises, Inc. www.variation.com
16
Skewness = -0.07718 Y = -0.16321 B2 = 3.577778 W2 = 1.270585 Delta = 2.8899 Alpha = 2.718712 Z = -0.17338 p-value <> = 0.862353 p-value > = 0.568824 p-value < = 0.431176 Kurtosis = 4.890742 Eb2 = 2.714286 VarB2 = 0.579237 x = 2.859708 3StdMom = 1.737522 A = 18.32027 0.430621 Z= 2.113056 p-value <> = 0.034596 p-value > = 0.017298 p-value < = 0.982702 K^2 4.495066
TE Taylor Enterprises, Inc. www.variation.com
17
p-value 0.10566
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.1057. The conclusions for the SK All test should be it passes. The conclusion for the SK Spec test should be it fails for a two-sided specification. Cell J17 should round to -0.1734 (ZS). Cell J18 should round to 0.8624 (p-value 2-sided skewness). Cell J29 should round to 2.1131 (ZK). Cell J31 should round to 0.0173 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: Pass Initials: WT Date: 9/1/17
4.8. Change cell D7 to “U” (Req. 3-4 for n>=15, normal data, upper specification)
Skewness = -0.07718 Y = -0.16321 B2 = 3.577778 W2 = 1.270585 Delta = 2.8899 Alpha = 2.718712 Z = -0.17338 p-value <> = 0.862353 p-value > = 0.568824 p-value < = 0.431176 Kurtosis = 4.890742 Eb2 = 2.714286 VarB2 = 0.579237
TE Taylor Enterprises, Inc. www.variation.com
18
x = 2.859708 3StdMom = 1.737522 A = 18.32027 0.430621 Z= 2.113056 p-value <> = 0.034596 p-value > = 0.017298 p-value < = 0.982702 K^2 4.495066 p-value 0.10566
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.1057. The conclusions for the SK All test should be it passes. The conclusion for the SK Spec test should be it fails for an upper specification. Cell J17 should round to -0.1734 (ZS). Cell J19 should round to 0.5688 (p-value positive skewness). Cell J29 should round to 2.1131 (ZK). Cell J31 should round to 0.0173 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: Pass Initials: WT Date: 9/1/17
4.9. Change cell D7 to “L” (Req. 3-4 for n>=15, normal data, lower specification)
Skewness = -0.07718 Y = -0.16321 B2 = 3.577778 W2 = 1.270585
TE Taylor Enterprises, Inc. www.variation.com
19
Delta = 2.8899 Alpha = 2.718712 Z = -0.17338 p-value <> = 0.862353 p-value > = 0.568824 p-value < = 0.431176 Kurtosis = 4.890742 Eb2 = 2.714286 VarB2 = 0.579237 x = 2.859708 3StdMom = 1.737522 A = 18.32027 0.430621 Z= 2.113056 p-value <> = 0.034596 p-value > = 0.017298 p-value < = 0.982702 K^2 4.495066 p-value 0.10566
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.1057. The conclusions for the SK All test should be it passes. The conclusion for the SK Spec test should be it fails for a lower specification. Cell J17 should round to -0.1734 (ZS). Cell J20 should round to 0.4312 (p-value negative skewness). Cell J29 should round to 2.1131 (ZK). Cell J31 should round to 0.0173 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: Pass Initials: WT Date: 9/1/17
4.10. Change cell D7 to “T” and cell B12:B31 to the values below (Req. 3-4 for n>=15, negative kurtosis data, two-sided specification)
-1.1 -0.5 -0.4 1.1 -1 1.2 -0.8 1.2 -1.2 -1 0.6
TE Taylor Enterprises, Inc. www.variation.com
20
-1.1 1.2 -1.2
1 -0.8 -1.1
1 1.2 1.2
Skewness = 0.144366 Y = 0.3053 B2 = 3.577778 W2 = 1.270585 Delta = 2.8899 Alpha = 2.718712 Z = 0.323845 p-value <> = 0.746055 p-value > = 0.373028 p-value < = 0.626972 Kurtosis = 1.167515 Eb2 = 2.714286 VarB2 = 0.579237
TE Taylor Enterprises, Inc. www.variation.com
21
x = -2.03235 3StdMom = 1.737522 A = 18.32027 3.704331 Z= -5.07932 p-value <> = 3.79E-07 p-value > = 1 p-value < = 1.89E-07 K^2 25.90433 p-value 2.37E-06
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.0000 The conclusions for the SK All test should be it fails. The conclusion for the SK Spec test should be it passes for a two-sided specification. Cell J17 should round to 0.3238 (ZS). Cell J18 should round to 0.7461 (p-value 2-sided skewness). Cell J29 should round to -5.0793 (ZK) Cell J31 should round to 1.0000 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: Passes Initials: WT Date: 9/1/17
4.11. Change cell D7 to “U” (Req. 3-4 for n>=15, normal data, upper specification)
Skewness = 0.144366 Y = 0.3053 B2 = 3.577778 W2 = 1.270585
TE Taylor Enterprises, Inc. www.variation.com
22
Delta = 2.8899 Alpha = 2.718712 Z = 0.323845 p-value <> = 0.746055 p-value > = 0.373028 p-value < = 0.626972 Kurtosis = 1.167515 Eb2 = 2.714286 VarB2 = 0.579237 x = -2.03235 3StdMom = 1.737522 A = 18.32027 3.704331 Z= -5.07932 p-value <> = 3.79E-07 p-value > = 1 p-value < = 1.89E-07 K^2 25.90433 p-value 2.37E-06
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.0000. The conclusions for the SK All test should be it fails. The conclusion for the SK Spec test should be it passes for an upper specification. Cell J17 should round to 0.3238 (ZS). Cell J19 should round to 0.3730 (p-value positive skewness). Cell J29 should round to -5.0793 (ZK). Cell J31 should round to 1.0000 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: Pass Initials: WT Date: 9/1/17
4.12. Change cell D7 to “L” (Req. 3-4 for n>=15, normal data, lower specification)
TE Taylor Enterprises, Inc. www.variation.com
23
Skewness = 0.144366 Y = 0.3053 B2 = 3.577778 W2 = 1.270585 Delta = 2.8899 Alpha = 2.718712 Z = 0.323845 p-value <> = 0.746055 p-value > = 0.373028 p-value < = 0.626972 Kurtosis = 1.167515 Eb2 = 2.714286 VarB2 = 0.579237 x = -2.03235 3StdMom = 1.737522 A = 18.32027 3.704331 Z= -5.07932 p-value <> = 3.79E-07 p-value > = 1 p-value < = 1.89E-07 K^2 25.90433 p-value 2.37E-06
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.0000.
TE Taylor Enterprises, Inc. www.variation.com
24
The conclusions for the SK All test should be it fails. The conclusion for the SK Spec test should be it passes for a lower specification. Cell J17 should round to 0.3238 (ZS). Cell J20 should round to 0.6270 (p-value negative skewness). Cell J29 should round to -5.0793 (ZK). Cell J31 should round to 1.0000 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: Pass Initials: WT Date: 9/1/17
4.13. Change cell D7 to “T” and cell B12:B31 to the values below (Req. 3-4 for n>=15, positive skewness data, two-sided specification)
-0.7 -0.4 -0.4 -0.6 1.8 -0.4 -0.4 1.1 1.5 -0.4 0.3 -0.7 -0.8 -0.8 -0.8 -0.8 -0.8 -0.8 -0.3 0.7
TE Taylor Enterprises, Inc. www.variation.com
25
Skewness = 1.363523 Y = 2.883529 B2 = 3.577778 W2 = 1.270585 Delta = 2.8899 Alpha = 2.718712 Z = 2.669104 p-value <> = 0.007605 p-value > = 0.003803 p-value < = 0.996197 Kurtosis = 3.493619 Eb2 = 2.714286 VarB2 = 0.579237 x = 1.023988 3StdMom = 1.737522 A = 18.32027 0.644279 Z= 1.127517 p-value <> = 0.259524 p-value > = 0.129762 p-value < = 0.870238 K^2 8.395412 p-value 0.01503
TE Taylor Enterprises, Inc. www.variation.com
26
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.0150. The conclusions for the SK All test should be it fails. The conclusion for the SK Spec test should be it fails for a two-sided specification. Cell J17 should round to 2.6691 (ZS). Cell J18 should round to 0.0076 (p-value 2-sided skewness). Cell J29 should round to 1.1275 (ZK). Cell J31 should round to 0.1298 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: Pass Initials: WT Date: 9/1/17
4.14. Change cell D7 to “U” (Req. 3-4 for n>=15, normal data, upper specification)
Skewness = 1.363523 Y = 2.883529 B2 = 3.577778 W2 = 1.270585 Delta = 2.8899 Alpha = 2.718712 Z = 2.669104 p-value <> = 0.007605 p-value > = 0.003803 p-value < = 0.996197 Kurtosis = 3.493619 Eb2 = 2.714286 VarB2 = 0.579237 x = 1.023988
TE Taylor Enterprises, Inc. www.variation.com
27
3StdMom = 1.737522 A = 18.32027 0.644279 Z= 1.127517 p-value <> = 0.259524 p-value > = 0.129762 p-value < = 0.870238 K^2 8.395412 p-value 0.01503
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.0150. The conclusions for the SK All test should be it fails. The conclusion for the SK Spec test should be it fails for an upper specification. Cell J17 should round to 2.6691 (ZS). Cell J19 should round to 0.0038 (p-value positive skewness). Cell J29 should round to 1.1275 (ZK). Cell J31 should round to 0.1298 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: Pass Initials: WT Date: 9/1/17
4.15. Change cell D7 to “L” (Req. 3-4 for n>=15, normal data, lower specification)
Skewness = 1.363523 Y = 2.883529 B2 = 3.577778 W2 = 1.270585 Delta = 2.8899
TE Taylor Enterprises, Inc. www.variation.com
28
Alpha = 2.718712 Z = 2.669104 p-value <> = 0.007605 p-value > = 0.003803 p-value < = 0.996197 Kurtosis = 3.493619 Eb2 = 2.714286 VarB2 = 0.579237 x = 1.023988 3StdMom = 1.737522 A = 18.32027 0.644279 Z= 1.127517 p-value <> = 0.259524 p-value > = 0.129762 p-value < = 0.870238 K^2 8.395412 p-value 0.01503
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.0150. The conclusions for the SK All test should be it fails. The conclusion for the SK Spec test should be it passes for a lower specification. Cell J17 should round to 2.6691 (ZS). Cell J20 should round to 0.9962 (p-value negative skewness). Cell J29 should round to 1.1275 (ZK). Cell J31 should round to 0.1298 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: Pass Initials: WT Date: 9/1/17
4.16. Change cell D7 to “T” and cell B12:B31 to the values below (Req. 3-4 for n>=15, negative skewness data, two-sided specification)
0.7 0.4 0.4 0.6 -1.8 0.4 0.4 -1.1 -1.5 0.4 -0.3 0.7
TE Taylor Enterprises, Inc. www.variation.com
29
0.8 0.8 0.8 0.8 0.8 0.8 0.3 -0.7
Skewness = -1.36352 Y = -2.88353 B2 = 3.577778 W2 = 1.270585 Delta = 2.8899 Alpha = 2.718712 Z = -2.6691 p-value <> = 0.007605 p-value > = 0.996197 p-value < = 0.003803 Kurtosis = 3.493619 Eb2 = 2.714286 VarB2 = 0.579237 x = 1.023988 3StdMom = 1.737522
TE Taylor Enterprises, Inc. www.variation.com
30
A = 18.32027 0.644279 Z= 1.127517 p-value <> = 0.259524 p-value > = 0.129762 p-value < = 0.870238 K^2 8.395412 p-value 0.01503
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.0150. The conclusions for the SK All test should be it fails. The conclusion for the SK Spec test should be it fails for a two-sided specification. Cell J17 should round to -2.6691 (ZS). Cell J18 should round to 0.0076 (p-value 2-sided skewness). Cell J29 should round to 1.1275 (ZK). Cell J31 should round to 0.1298 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: Pass Initials: WT Date: 9/1/17
4.17. Change cell D7 to “U” (Req. 3-4 for n>=15, normal data, upper specification)
Skewness = -1.36352 Y = -2.88353 B2 = 3.577778 W2 = 1.270585
TE Taylor Enterprises, Inc. www.variation.com
31
Delta = 2.8899 Alpha = 2.718712 Z = -2.6691 p-value <> = 0.007605 p-value > = 0.996197 p-value < = 0.003803 Kurtosis = 3.493619 Eb2 = 2.714286 VarB2 = 0.579237 x = 1.023988 3StdMom = 1.737522 A = 18.32027 0.644279 Z= 1.127517 p-value <> = 0.259524 p-value > = 0.129762 p-value < = 0.870238 K^2 8.395412 p-value 0.01503
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.0150. The conclusions for the SK All test should be it fails. The conclusion for the SK Spec test should be it passes for an upper specification. Cell J17 should round to -2.6691 (ZS). Cell J19 should round to 0.9962 (p-value positive skewness). Cell J29 should round to 1.1275 (ZK). Cell J31 should round to 0.1298 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: Pass Initials: WT Date: 9/1/17
4.18. Change cell D7 to “L” (Req. 3-4 for n>=15, normal data, lower specification)
TE Taylor Enterprises, Inc. www.variation.com
32
Skewness = -1.36352 Y = -2.88353 B2 = 3.577778 W2 = 1.270585 Delta = 2.8899 Alpha = 2.718712 Z = -2.6691 p-value <> = 0.007605 p-value > = 0.996197 p-value < = 0.003803 Kurtosis = 3.493619 Eb2 = 2.714286 VarB2 = 0.579237 x = 1.023988 3StdMom = 1.737522 A = 18.32027 0.644279 Z= 1.127517 p-value <> = 0.259524 p-value > = 0.129762 p-value < = 0.870238 K^2 8.395412 p-value 0.01503
TE Taylor Enterprises, Inc. www.variation.com
33
Acceptance Criteria: The sample size should be 20. The p-value should round to 0.0150. The conclusions for the SK All test should be it fails. The conclusion for the SK Spec test should be it fails for a lower specification. Cell J17 should round to -2.6691 (ZS). Cell J20 should round to 0.0038 (p-value negative skewness). Cell J29 should round to 1.1275 (ZK). Cell J31 should round to 0.1298 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: Pass Initials: WT Date: 9/1/17
4.19. Change cell D7 to “T” and cell B12:B211 to the values below (Req. 3-4 for n=200, normal data, two-sided specification)
2.1 0.1 -0.5 0.1 -0.6
1 -0.4 0.1 -2.4 -1
0.6 0.1 -0.3 2.3 -1
0.3 -0.3
2 -0.9 0.1 0.8 -1.5
1 2.1 0.7 -0.1 -0.4 0.6 -1
-0.5 0.4 0.2 -0.1
TE Taylor Enterprises, Inc. www.variation.com
34
-0.3 -2.5 -0.2 0.3 -0.8 -0.2 -3
1.9 0.7 -0.2 0.4 -3.4 0.9 -1.8 0.8 0
-0.2 0
-0.1 0.8 -1.1 1.4 2.5 0.5 -0.1 -1.8 0.9 0.1 0.2 1.3 0.1 -0.1 0.7 -0.9 0.3 -0.2 -0.1 1.3 2.3 0.5 0.6 1.1 1
-2.2
TE Taylor Enterprises, Inc. www.variation.com
35
2.5 1.6 1.2 1.7 -0.9 2.1 0.6 0.6 0.3 1.1 0.5 -0.9 0.7 -0.4 -0.7 0.9 0.1 0.5 -0.2 -1.3 -1
-2.3 1.3 -0.5 -0.3 1.9 1
-0.2 -0.1 1.2 -0.7 2.1 1.1 1.4 -0.5
2 -0.3 0.3 -0.2 1.3 1.1 -1.3 -0.6 -0.8
TE Taylor Enterprises, Inc. www.variation.com
36
-0.2 1.2 -0.7 -2.7 0.1 -1.1 -0.8 -0.8 -0.3 1.1 0
-0.5 1.6 0.5 0.5 -1.6 2.4 -0.3 0.6 -0.5
0 0.8 1.9 -0.4 0.3 -0.2 1.6 -0.5 -0.8 -1.2 -1.6 0.4 1.3 0.4 -1.4 -0.6 -0.5 -0.4 0.2 1.4 -1.5 1.3 -0.2 -0.1
TE Taylor Enterprises, Inc. www.variation.com
37
-0.8 0.7 -0.3 -1.2 0.5 -0.7 -0.7 0.7 -0.3 -1
0.5 0.6 1.6 1 1
0.7 -0.1 -0.9 -0.1
1 0.7 0.1 0.6 0.2 1.6 -0.5 -0.7 -1.1 0.4 -0.6 1.9 -0.7 -0.2 -0.5 -0.7
TE Taylor Enterprises, Inc. www.variation.com
38
Skewness = -0.1953 Y = -1.14457 B2 = 3.159828
TE Taylor Enterprises, Inc. www.variation.com
39
W2 = 1.078378 Delta = 5.14828 Alpha = 5.051471 Z = -1.15675 p-value <> = 0.247376 p-value > = 0.876312 p-value < = 0.123688
Kurtosis = 3.346567 Eb2 = 2.970149 VarB2 = 0.11136 x = 1.127987 3StdMom = 0.967701 A = 42.06667
0.756788
Z= 1.147805 p-value <> = 0.251049 p-value > = 0.125525 p-value < = 0.874475
K^2 2.655517 p-value 0.265071
Acceptance Criteria: The sample size should be 200. The average should round to 0.11150. The standard deviation should round to 1.07923. The skewness should round to -0.196780. The kurtosis should round to 0.386000. The p-value should round to 0.2651. The conclusions for the SK All test should be it passes. The conclusion for the SK Spec test should be it passes for a two-sided specification. Cell J17 should round to -1.1567 (ZS). Cell J18 should round to 0.2474 (p-value 2-sided skewness). Cell J29 should round to 1.1478 (ZK). Cell J31 should round to 0.1255 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: Pass Initials: WT Date: 9/1/17
4.20. Change cell D7 to “U” (Req. 3-4 for n=200, normal data, upper specification)
TE Taylor Enterprises, Inc. www.variation.com
40
Skewness = -0.1953 Y = -1.14457 B2 = 3.159828 W2 = 1.078378 Delta = 5.14828 Alpha = 5.051471 Z = -1.15675 p-value <> = 0.247376 p-value > = 0.876312 p-value < = 0.123688 Kurtosis = 3.346567 Eb2 = 2.970149 VarB2 = 0.11136 x = 1.127987 3StdMom = 0.967701 A = 42.06667 0.756788 Z= 1.147805 p-value <> = 0.251049 p-value > = 0.125525 p-value < = 0.874475 K^2 2.655517 p-value 0.265071
TE Taylor Enterprises, Inc. www.variation.com
41
Acceptance Criteria: The sample size should be 200. The p-value should round to 0.2651. The conclusions for the SK All test should be it passes. The conclusion for the SK Spec test should be it passes for an upper specification. Cell J17 should round to -1.1567 (ZS). Cell J19 should round to 0.8763 (p-value positive skewness). Cell J29 should round to 1.1478 (ZK). Cell J31 should round to 0.1255 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: Pass Initials: WT Date: 9/1/17
4.21. Change cell D7 to “L” (Req. 3-4 for n=200, normal data, lower specification)
Skewness = -0.1953 Y = -1.14457 B2 = 3.159828 W2 = 1.078378 Delta = 5.14828 Alpha = 5.051471 Z = -1.15675 p-value <> = 0.247376 p-value > = 0.876312 p-value < = 0.123688 Kurtosis = 3.346567 Eb2 = 2.970149 VarB2 = 0.11136 x = 1.127987
TE Taylor Enterprises, Inc. www.variation.com
42
3StdMom = 0.967701 A = 42.06667 0.756788 Z= 1.147805 p-value <> = 0.251049 p-value > = 0.125525 p-value < = 0.874475 K^2 2.655517 p-value 0.265071
Acceptance Criteria: The sample size should be 200. The p-value should round to 0.2651. The conclusions for the SK All test should be it passes. The conclusion for the SK Spec test should be it passes for a lower specification. Cell J17 should round to -1.1567 (ZS). Cell J20 should round to 0.1237 (p-value negative skewness). Cell J29 should round to 1.1478 (ZK). Cell J31 should round to 0.1255 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: Pass Initials: WT Date: 9/1/17
5. Independence of tab
5.1. Copy the Skewness Kurtosis Tests tab. Delete all tab but the copy.
TE Taylor Enterprises, Inc. www.variation.com
43
Skewness = -0.1953 Y = -1.14457 B2 = 3.159828 W2 = 1.078378 Delta = 5.14828 Alpha = 5.051471 Z = -1.15675 p-value <> = 0.247376 p-value > = 0.876312 p-value < = 0.123688 Kurtosis = 3.346567 Eb2 = 2.970149 VarB2 = 0.11136 x = 1.127987 3StdMom = 0.967701 A = 42.06667 0.756788 Z= 1.147805 p-value <> = 0.251049 p-value > = 0.125525 p-value < = 0.874475
TE Taylor Enterprises, Inc. www.variation.com
44
K^2 2.655517 p-value 0.265071
Acceptance Criteria: The sample size should be 200. The average should round to 0.11150. The standard deviation should round to 1.07923. The skewness should round to -0.196780. The kurtosis should round to 0.386000. The p-value should round to 0.2651. The conclusions for the SK All test should be it passes. The conclusion for the SK Spec test should be it passes for a lower specification. Cell J17 should round to -1.1567 (ZS). Cell J20 should round to 0.1237 (p-value negative skewness). Cell J29 should round to 1.1478 (ZK). Cell J31 should round to 0.1255 (p-value positive kurtosis). Note that these cells are not visible unless selected.
Pass or Fail: Pass Initials: WT Date: 9/1/17
6. Saving and Signing File
6.1. Determine if all tests passed
All Test Cases for Computer Pass? X yes no
6.2. Save File and show name and date.
Name of File: STAT-18 – Window 7
Data of File: 9/1/17
Signature: ________________________________ Date: September 1, 2017