multiple echo diffusion tensor acquisition technique
TRANSCRIPT
Magnetic Resonance
Multiple echo diffusion tensor acquisition technique
Eric E. Sigmund4, Yi-Qiao SongSchlumberger-Doll Research, Ridgefield, CT 06877, USA
Received 9 August 2005; accepted 21 October 2005
Abstract
The standard method of diffusion tensor imaging (DTI) involves one diffusion-sensitizing gradient direction per acquired signal. This
paper describes an alternative method in which the entire direction set required for calculating the diffusion tensor is captured in a few scans.
In this method, a series of radiofrequency (RF) pulses are applied, resulting in a train of spin echoes. A pattern of applied magnetic field
gradients between the RF pulses generates a different diffusion weighting in both magnitude and direction for each echo, resulting in a dataset
sufficient to determine the tensor. This significantly reduces the time required for a full DTI scan and potentially allows a tradeoff of this time
for image quality. In the present work, this method is demonstrated in an anisotropic diffusion phantom (asparagus).
D 2006 Elsevier Inc. All rights reserved.
Keywords: DTI; MMME; Multiple echo; Diffusion tensor
1. Introduction
1.1. Standard DTI
Diffusion tensor imaging (DTI) [1,2] is a specific type of
diffusion-weighted MRI that maps out anisotropically
restricted diffusion in three dimensions. In this type of
scan, the diffusion-weighting gradient is typically varied in
both orientation and strength to determine the apparent
diffusion coefficient along many different directions. This
information is then used to determine the diffusion tensor,
which is the 3D generalization of the diffusion coefficient to
anisotropic motion. Mathematically, the diffusion tensor D
is a 3�3 matrix that enters the spin magnetization
attenuation in a quadratic form:
M
M0
¼ exp �Xij
bijDij
! ð1Þ
or written in the form of a set of linear equations,
� ln M=M0ð Þnumn ¼Xij
bij;nDij ð2Þ
where n labels acquisitions corresponding to different
applied gradient directions, and bij,n are the diffusion-
0730-725X/$ – see front matter D 2006 Elsevier Inc. All rights reserved.
doi:10.1016/j.mri.2005.10.015
4 Corresponding author. Tel.: +1 203 431 5561; fax: +1 203 438 3819.
E-mail address: [email protected] (E.E. Sigmund).
weighting factors for each experiment, determined by the
particular radiofrequency (RF) and gradient pulse sequence
[3–5]. The eigenvalues of the diffusion tensor are the
principal diffusivities ki, and the eigenvectors are the
principal orientations ei. These parameters are used to
calculate scalar properties of the tensor such as the fractional
anisotropy (FA) and the average diffusivity or Trace(D),
according to Ref. [1].
FA ¼
ffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi3
2
P3i¼1
ðki�hkiÞ2
P3i¼1
k2i
vuuuuuut ð3Þ
TraceðDÞ=3 ¼ hki ¼ k1 þ k2 þ k33
: ð4Þ
The FA and Trace parameters offer complementary scalar
information in that the former quantifies the orientational
anisotropy in the object, while the latter averages over it. Both
are clinically useful in tissue identification and diagnosis.
In order to measure the diffusion tensor elements Dij in
the laboratory frame, at least six linearly independent
vectors b=[bxx, byy, bzz, bxy, byz, bxz] must be generated
and an amplitude collected for each. In each voxel of the
image, these amplitudes are divided by that from another
experiment with no diffusion weighting (b=0). The range of
diffusion weighting is often covered by choosing one
Imaging 24 (2006) 7–18
Fig. 1. MMME diffusion tensor acquisition technique. The four RF pulses applied generate a series of 13 separate echoes, each of which carries different
diffusion weighting in up to three dimensions. Each gradient direction (Gx, Gy, Gz) serves a dual role of imaging resolution (read, phase, slice gradients) and
diffusion weighting.
E.E. Sigmund, Y.-Q. Song / Magnetic Resonance Imaging 24 (2006) 7–188
gradient direction and one weighting magnitude per
acquisition, and executing as many acquisitions as direc-
tions needed. This is true for most DTI modalities employed
today, including the most common echo-planar imaging
(EPI) [7] as well as steady-state free precession [8–10].
There are some notable exceptions employing multiple spin
echoes for multiple diffusion directions [11]. The minimal
direction set is six acquisitions, although most research or
clinical DTI measurements employ many more (12, 20, 100,
etc.) [12–14] for either enhanced SNR of processed images
or the measurement of anatomical structures that are not
describable by the tensor model (Eq. (1)), such as crossing
white matter fibers in the brain [15]. Much research
attention has been devoted to optimizing the set of applied
gradient directions [13,14,16–19]. While the standard EPI
modality is by far the fastest imaging method, it also carries
well-known image artifacts such as blurring, signal dropout,
and ghosting due to either its sensitivity to magnetic
susceptibility contrast or the eddy currents generated by
its rapid gradient switching. Alternative DTI modalities
without these artifacts are thus highly desirable. However, to
make up for the loss of speed compared to EPI, other
innovations are required to shorten the overall scan time.
One candidate is to improve directional efficiency, i.e., the
number of different diffusion directions sampled within one
excitation. Particularly for scans involving many directional
samples, this efficiency is quite important.
Recent work [20,21] has demonstrated the utility of a
new diffusometry modality with potential applications to
both spectroscopy and imaging. In this technique, termed
multiple modulation multiple echo (MMME), diffusion
weightings of multiple magnitudes and directions are
applied and measured within a single scan. For cases where
the directional phase space is a significant contributor to
scan time, this technique provides an improvement in
efficiency. In the present work, we show the first diffusion
tensor images acquired with this pulse sequence, termed
MMME Diffusion Tensor Acquisition Technique (MEDI-
TATE). We also elaborate on the optimization of this
method with regard to minimizing correlated errors between
different elements of the diffusion tensor.
Numerous methods have been previously developed that
successfully use multiple echoes for relaxation, diffusion,
image encoding or contrast, or 2D spectroscopy [11,22–30].
In the diffusion tensor case, previous work has employed
multiple spin echoes from multiple inversion pulses [11]
combined with a cylindrical symmetry approximation to
accelerate the DTI scan. We emphasize that the present
work, as in Ref. [21], focuses upon using the full 13 echoes
of the four-pulse MMME sequence to image multidimen-
sional diffusion. The implementation is described in the next
section, followed by a description of the sequence’s
optimization. Finally, we present experimental results on
an asparagus phantom.
1.2. MEDITATE sequence — basic theory
The principle of the MEDITATE technique is described
as follows and illustrated in Fig. 1. A series of four RF
pulses are rapidly applied in uneven time intervals. This
preparation is similar to others such as BURST [25],
DANTE and DUFIS [27]. This action generates a large
number of independently evolving echo pathways, with
magnitudes determined by the flip angles of the RF pulses
[5,31–33]. Different pathways include differing portions of
longitudinal and transverse evolution. We denote each echo
or coherence pathway q by a series of values �1, 0 or 1,
corresponding to co-rotating, longitudinal and counter-
rotating magnetization periods, respectively. The pathways
giving rise to the 13 echoes in the MMME train are listed in
the first column of Table 1.
Between and after the RF pulses, magnetic field gradient
pulses are applied in different combinations. Each of the
three gradients (Gx, Gy, Gz) plays a dual role of position
encoding (read, phase, slice) and diffusion weighting. Along
Table 1
Coherence pathways and relaxation coefficients for the 13 echoes of the
four-pulse MMME sequence
Echo # { q} c1q c2q c1qV c2qV
0 +00� 12 2 4 0
1 0+0� 10 5 3 2
2 ��+0� 9 7 3 2
3 ++0� 9 8 3 2
4 �+� 0 18 0 6
5 0�+� 1 18 0 6
6 +�+� 0 20 0 6
7 �0+� 3 18 1 6
8 00+� 4 18 1 6
9 +0+� 3 20 1 6
10 �++� 0 24 0 8
11 0++� 1 24 0 8
12 +++� 0 26 0 8
The pathways are described by a sequence of values corresponding to the
magnetization states in the four time intervals between the RF pulses of the
sequence (see Fig. 1).
E.E. Sigmund, Y.-Q. Song / Magnetic Resonance Imaging 24 (2006) 7–18 9
one direction (x), a constant field gradient is applied
between and after the RF pulses. This gradient provides a
gradually increasing diffusion weighting along one direction
to the echo train and also serves as a readout gradient for its
detection. This gradient is deactivated during the RF pulses
to avoid slice selection effects. Along the other two
directions ( y and z), pulsed gradients of duration d are
applied in a pattern that preserves the formation of each
echo in the train. The first two gradient pulses are arbitrary,
but once they are fixed the last four are derived from the
echo formation conditionR TE0
q tð ÞG tð Þdt ¼ 0, where TE is
the echo time, q(t) is the coherence pathway and G(t) is the
gradient waveform. The gradient pulses provide diffusion
weighting along the two remaining directions in an amount
and direction that vary from echo to echo. The diffusion-
weighting factors bij for each echo can be calculated in the
standard way [3,4]:
bij ¼ZTE0
ki tð Þkj tð Þdt ð5Þ
where i=x,y,z and TE is the echo time, and the vector k is
calculated from
*k tð Þ ¼
Zt0
cq tVð Þ*G tVð ÞdtV: ð6Þ
c is the proton gyromagnetic ratio. Thus, the directional
phase space in the MMME diffusion sequence is covered
with a combination of gradient waveforms G and magne-
tization histories q. Only those echo pathways that are
transverse ( q=F1) during the pulsed Gy and Gz gradients
will suffer their diffusion weighting. For example, the
pathway generating Echo 2 is exposed to the first gradient
pulse but not the second, whereas the pathway for Echo 5 is
exposed to both, etc. Each echo will have a different
diffusion weighting bij given by Eqs. (5) and (6). Once the
diffusion weightings are determined and the echo magni-
tudes M measured, the diffusion tensor elements Dij are
calculated exactly as is done in a standard diffusion tensor
experiment, by solving the linear set of equations in Eq. (2).
Since there are 13 echoes in this sequence and only six
independent elements of the diffusion tensor, this problem is
overdetermined. However, flip angle and relaxation effects
must be separated out as follows. The full expression of the
amplitude of the qth echo can be written
Mq ¼ M0 fq að Þexp�� c1q
sT1
� cV1qsVT1
� c2qsT2
� cV2qsVT2
�
� exp
��Xij
bij;qDij
�; ð7Þ
where M0 is the total magnetization, fq(a) is a magnetiza-
tion fraction (0b fq(a)b1) depending on the RF flip angles aand the coherence pathway q [20], {c1q, c2q, c1qV , c2qV } are
relaxation coefficients that depend on the coherence
pathway and are listed in Table 1, T1 and T2 are relaxation
times, bij,q is the diffusion-weighting matrix for echo q and
Dij are the diffusion tensor elements. The nondiffusive
terms in this expression can be separated from the data using
analytical calculations or empirical measurements. For
example, the fractions fq(a) can be calculated for any set
of flip angles a [20]. However, if the RF field is spatially
inhomogeneous, this correction will be imperfect. In this
case, a reference experiment without diffusion weighting is
preferable to divide out the terms fq(a). In this spirit, each
measurement in this work involved the ratio of two MMME
scans with the same excitation pulses a: one including both
the steady gradient Gx and the pulsed gradients Gy and Gz,
and one with the steady gradient only and (in general)
different pulse timing. The ratio of the echo magnitudes for
the two cases thus eliminates the coherence pathway factors
fq(a). The pulse timing must be different between the two
scans to avoid canceling the diffusion weighting along the
constant gradient (x) direction, since the gradient magnitude
was kept constant to produce equal readout fields-of-view
(FOVs). This unequal timing means that relaxation con-
tributions must be incorporated in the analysis. This can be
done through a separate measurement of the relaxation rates
to divide out their contributions. Alternatively, the solution
for the diffusion tensor can be expanded to include the
relaxation times as independent variables, since they enter
the problem linearly (Eq. (7)). In practice, this combined
solution was found to be poorly conditioned since the
T2 weighting and diffusion weighting along the constant
gradient axis have similar trends. The conditioning of the
diffusion tensor calculation will be discussed further in the
next section.
This spectroscopic sequence is converted to a 2D
imaging sequence with an additional phase encoding pulse
(GPE), just after the last RF pulse that encodes transverse
position for all of the detected echoes. Although the above
sequence does not contain any slice selection, it could be
-400
-200
0
200
400
b (s
/mm
2 )
121086420
Echo number
bxx
byy
bxy
-1000
-500
0
500
1000
b (s
/mm
2 )
121086420
Echo number
byy
bzz
byz
Fig. 2. Diffusion-weighting patterns and vector visualizations for various MEDITATE sequences used to acquire the images in this work. These vectors are
shown along with an example diffusion ellipsoid with a 33% higher axial than transverse diffusivity. Top row: pattern used for scans 1/2 and 1/3 (top and
bottom rows in Fig. 6); middle row: pattern used for scans 4/5; bottom row: bcombined Q pattern showing all diffusion sensitizing directions employed to
generate the 3D tensor data; each MEDITATE-2D sequence covered one plane of sensitivity in q-space.
E.E. Sigmund, Y.-Q. Song / Magnetic Resonance Imaging 24 (2006) 7–1810
straightforwardly introduced through Gz pulses during the
RF excitation pulses, which could be shaped if necessary.
Multislice acquisitions, i.e., acquiring signal from new slices
while others relaxed, would also be possible in this
sequence. In the present work, all of the RF pulses were
bhardQ and the spin density along the z-axis (a thickness of
approximately 8 mm) was projected onto the x–y plane. The
imaging modality employed in this work is a standard spin-
warp technique. The diffusion weighting, however, is
executed with multiple directions and magnitudes sufficient
for a diffusion tensor measurement in 2D, or potentially 3D,
in a few scans. This is in sharp contrast to most other
diffusion imaging techniques demonstrated to date.
There are several limitations to the MEDITATE
sequence that must also be noted. The analysis used to
extract the tensor assumes a time-independence of the
physical parameters of the system over the course of the
echo train, and there are several reasons in practice why
this might not be the case. First, a spin system undergoing
restricted diffusion or multi-compartmental exchange
will possess a time-dependent diffusion coefficient D(t)
[6,34–38], and if this time-dependence is significant over
the time scale of the echo train, both the magnitude and
the anisotropy in the diffusion tensor may be compro-
mised. Another well-known parasitic effect in NMR
diffusometry is that of internal susceptibility-driven, or
E.E. Sigmund, Y.-Q. Song / Magnetic Resonance Imaging 24 (2006) 7–18 11
bbackgroundQ, gradients whose contribution to the spin
dephasing artificially enhances the measured diffusivity. A
variety of pulse sequences have been developed that
compensate for this effect through tailored gradient wave-
forms [39–50], sometimes involving inversions of the
applied gradient that are not duplicated by the internal
gradient. For the MEDITATE sequence, every echo
originates from a different coherence pathway and diffusion
gradient waveform; some are spin-echo pathways, some
stimulated echo pathways, etc. This means that different
echoes will have different susceptibility gradient compen-
sation, which can distort the tensor calculation. Also,
multiple environments, such as intracellular/extracellular
spaces, intra-axonal/extra-axonal spaces, etc., can lead to
multi-exponential relaxation behavior that will confuse the
diffusion information. Finally, unlike the standard EPI
imaging modality, the spin-warp modality employed here
may suffer ghosting artifacts if subject motion occurs
between successive phase encodes. These artifacts would
also vary from echo to echo since each has a different
formation time. All of these effects are each discussed
individually as they apply to the present study in Methods.
1.3. MEDITATE sequence — optimization
Much research attention has been devoted to the
optimization of gradient direction sets for DTI experiments
[12–14,16–19]. Many different direction sets have been
generated and applied; some are motivated by symmetry
arguments, others by minimization of noise propagation and
still others by maximal angular resolution of higher order
structures. Several scalar indices have been proposed to
quantify the quality of one direction set vs. another. While
the MMME sequence affords less flexibility in direction
selection, we will see that it nevertheless compares
favorably with existing DTI direction schemes.
A visual representation of the diffusion weighting of the
MEDITATE-2D sequences used in the present work is
shown in Fig. 2. On the left-hand side of this figure are
shown the numerical values of the coefficient matrix bij for
several scans used in this article. On the right-hand side
are vectors proportional to tvb ¼ffiffiffiffiffiffibxx
p;
ffiffiffiffiffiffibyy
p;
ffiffiffiffiffiffibzz
p� �for
the corresponding patterns at left. The bottom figure
shows the combined direction set for all data contributing
to the 3D tensor measurement described in Results; note
that each of the three MEDITATE-2D sequences contributes
a range of vectors within one plane. This representation
provides a general sense of the angular coverage of a
MEDITATE-2D sequence.
We now proceed to describe the optimization of the
gradient patterns used in the MEDITATE sequence. In the
following discussion, we adopt the following conventional
notation for the processing required to extract the diffusion
tensor. The linear equation set in Eq. (2) can be written in
matrix form:
Bdtd ¼ t
m ð8Þ
where
B ¼
bxx1 byy1 bzz1 2bxy1 2byz1 2bxz1bxx2 byy2 bzz2 2bxy2 2byz2 2bxz2v v v v v v
bxxN byyN bzzN bxyN byzN bxzN
1CCA
0BB@ ð9Þ
is the diffusion-weighting coefficient matrix,
td ¼
Dxx
Dyy
Dzz
Dxy
Dyz
Dxz
1CCCCCCA
0BBBBBB@
ð10Þ
is the diffusion tensor written in vector form, and
tm ¼
� ln M=M0ð Þ1� ln M=M0ð Þ2
v� ln M=M0ð ÞN
1CCA
0BB@ ð11Þ
is the set of measured signals.
We briefly review a quantitative measure of sequence
quality employed for optimization. The goal of this
optimization is the extraction of all the desired elements
Dij, with a minimum of (a) absolute variance and (b)
covariance. In other words, we wish to prepare the
coefficient matrix bij so as to extract each measured value
Dij with the least error and the least confusion with other
tensor elements. An ideal figure of merit for this purpose is
the condition number [13,51,52], which relates the relative
error in the input data to that in the output parameters:
yd
d¼ condðBÞ ym
mð12Þ
and is also equal to the ratio of the maximum and minimum
singular values of B. The condition number can be
calculated from the coefficient matrix according to
condðBÞ ¼ kBkkB�1k ð13Þ
where ||B|| equals the largest singular value of B, and B�1
may represent the pseudoinverse of B if it is not a square
matrix. It is clear that lower condition numbers are more
favorable, with a minimum ideal condition number of 1. In
practice, condition numbers for DTI experiments have been
shown to have a slightly higher limit of 1.3 [13].
The variables for optimization of the MEDITATE-2D
sequence include the pulse delay s and the diffusion
gradient amplitudes Gx, Gy1, Gy2. However, not all of these
parameters are equally important. The determining factor in
the structure of the diffusion-weighting matrix is the ratio of
the two pulsed gradient amplitudes, Gy2/Gy1. Variation of
s and Gx produces a faster or slower diffusion-weighting
Fig. 4. Asparagus phantom sketch.
E.E. Sigmund, Y.-Q. Song / Magnetic Resonance Imaging 24 (2006) 7–1812
variation along the x-axis, but little qualitative change in the
pattern structure. We thus focus upon variations of the
sequence with the following parameter:
h ¼ tan�1 Gy2
Gy1
��ð14Þ
This angle parametrizes the various combinations of the
two pulsed field gradients applied in the preparation stage of
a MMME2D experiment. For example, h=08 corresponds
to only the first pulsed gradient being activated, h=908corresponds to only the second being activated, h=458corresponds to both activated at equal value, etc. Given this
parameter, we may ask two optimization questions: (1) how
does the condition number vary with h? and (2) how do the
accuracies of experiments performed with different hdepend on the condition number? Fig. 3 illustrates these
two issues. The panel shows the condition number as a
function of h. The calculations were made for the following
parameter set: s=1.5 ms, sV=1.5 ms, d=1.0 ms,
Gx=2.5 G/cm, jGyj ¼ffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiG2
y1 þ G2y2
q=15 G/cm. We see
small bfavorableQ values near h=1358 and a broad
Fig. 3. Variations of MMME2D sequence. Top: dependence of condition
number on h parameter describing sequence variation (see text). Bottom:
average deviations in asparagus sample diffusion tensor elements measured
in MMME2D sequence from values measured by standard PFG sequence
as a function of condition number.
bunfavorableQ peak maximizing at h=658. Another param-
eter that we expect to correlate with the condition number
is the angular range Db covered by the diffusion vectorstvb ¼
ffiffiffiffiffiffibxx
p;
ffiffiffiffiffiffibyy
p;
ffiffiffiffiffiffibzz
p� �in the x–y plane. For the range
of sequences considered in Fig. 3, Db shows a similar trend
(not shown) to cond(B), ranging from Db=468 at h=728 toDb=718 at h=1448. The bottom panel shows the accuracy
of a range of MMME2D experiments with different h on an
asparagus sample as compared to results obtained from the
same sample with a traditional pulsed-field-gradient stimu-
lated echo sequence [4,53]. This accuracy (or bbiasQ)is calculated as jyDj ¼
ffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiyD2
xx þ yD2yy þ yD2
xy
q, where
yDij=Dij,MMME�Dij,PFG. The error bar of each data set
represents the variation in bias with two repeated measure-
ments. The correlation of the bias with the condition number
is fairly good (r=0.92). This is consistent with the ubiquitous
use of the condition number to evaluate DTI schemes. The
optimal sequence pattern corresponds to h=1358, near a
condition number minimum of 3.5, and has nearly the
smallest bias. This pattern is used for much of the data
collection in this article. Similar arguments to these can be
applied to the MMME3D case, in which another direction of
pulsed gradients is applied. The ideal configuration in this
case is likely an extension of the ideal MMME2D case.
However, the third dimension’s design cannot correspond to
the same h as derived above if the two directions’ diffusivities
are to be distinguished. Work is in progress on this higher-
dimensional optimization, which is crucial for obtaining
the maximum advantage of the MEDITATE protocol.
2. Methods
The sample chosen to demonstrate the MEDITATE
sequence is a set of asparagus stalk portions arranged in
three orientations. Asparagus has a well-known cell
structure that leads to an axial diffusion anisotropy, with a
larger diffusivity along the stalk than transverse to it by
approximately 30%. This anisotropy has been noted in
numerous NMR/MRI studies [21,54]. The present work’s
sample, sketched in Fig. 4, was chosen to include preferred
diffusion directions along all three laboratory axes (in
different regions) to fully test the 3D capability of the
MEDITATE sequence. For clarity, the stalk axis on each
sample portion is labeled with a vector pointing either
Table 2
MEDITATE Pulse sequence parameters
Scan Gx Gy1 Gy2 Gz1 Gz2 GPE s sV d sPE
1 3.0 20 �20 0 0 3.0 1.6 1.5 1.0 0.5
2 3.0 0 0 0 0 3.0 1.1 1.5 1.0 0.5
3 3.0 0 0 20 �20 3.0 1.6 1.5 1.0 0.5
4 3.0 20 �20 20 40 3.0 1.6 1.5 1.0 0.5
5 3.0 0 0 0 0 3.0 1.6 1.5 1.0 0.5
Gradients are listed in gauss per centimeter and time intervals are
in milliseconds.
Fig. 5. Raw images acquired from a MEDITATE scan with the parameters
in row 1 of Table 2.
E.E. Sigmund, Y.-Q. Song / Magnetic Resonance Imaging 24 (2006) 7–18 13
horizontally (x), vertically ( y) or out-of-page (z). In the
following discussion, we refer to the corresponding regions
as the bmouthQ, bnoseQ and beyesQ, respectively.The images in this work were acquired on a 2-T
horizontal bore superconducting magnet (Nalorac Cryogen-
ics, Walnut Creek, CA) operating at a proton frequency of
85.1 MHz with a Bruker Biospec spectrometer (Bruker
Biospin, Billerica, MA) and a 44-mm-diameter RF probe
(Nova Medical, Wilmington, MA). Pulse programming was
performed on the Bruker XWINNMR software, and image
processing and analysis were performed with homegrown
code using Igor software (Wavemetrics, Portland, OR) on a
2-GHz Dell PC workstation.
Several measurements were performed to address the
limitations of the MEDITATE sequence mentioned in the
Introduction. First, the diffusion coefficient was measured
over a range of diffusion times with a standard stimulated
echo PFG diffusion sequence. The longitudinal diffusivity
along the stalk axis was found to be time-independent, and
the transverse diffusivity varied by ~10% over the time
range covered by the echo train (40 msb tb80 ms).
Relaxation times were measured separately via standard
inversion recovery (for T1) and spin-echo (for T2) sequen-
ces. The values T1=1.4 s and T2=200 ms were used as
representative averages from four different asparagus
samples, measured in non-spatially resolved measurements.
Then, using the relaxation coefficients {c1q, c2q, c1qV , c2qV }
from Table 1 and the measured relaxation times, the
relaxation term was divided out of each measured echo
amplitude. Both the spin–spin and spin–lattice relaxation
times (averaged over a 5-mm-diameter sample) were found
to be single-exponential.
The first MEDITATE-2D experiment was performed
with diffusion sensitivity in the (x, y) plane. For all data
shown in this work, the RF pulse flip angles were a1/a2/a3/a4=558/718/718/1108. This set has been previously deter-
mined [20] to equalize the magnitudes of the unweighted
echoes as much as possible. A pair of scans was run, with
pulse sequence parameters shown in the first two rows of
Table 2. Sixty-four phase encode lines were acquired at a
sampling rate (or readout bandwidth) of fs=100 kHz.
Each transient in Scan 1 contained Nx= fss=160 points,
compared to 110 points in Scan 2, which was smaller
because the echoes were closer together. In Scan 1, the echo
separation (within each group of three) was s=1.6 ms, and
for Scan 2 it was s=1.1 ms. Echo overlap was empirically
minimized by increasing the readout gradient Gx until no
overlap was visible. The resolution and FOVs for both
directions were thus Dx ¼ 1Nx
2pcGx
fs ¼ 0:5 mm; Dy ¼ 12�
2pcGPEsPE
¼ 0:78 mm; FOV ¼ NxDx ¼ 2pcGx
f s ¼ 54 mm;
FOVy¼ NPEDy ¼ NPE
22p
cGPEsPE¼ 53mm. While Scan 2 pos-
sessed a shorter Nx=110, zero filling was used to equalize
the scans’ grids, allowing a voxelwise ratio to be calculated.
To avoid line artifacts introduced from the echo-truncation
effect of zero filling, a Gaussian filter was applied in k-space
to all echoes. The corresponding blurring from the increase
of the point spread functions in the image domain (2.0�1.7 mm) was small compared to all relevant dimensions
of the phantom (see Fig. 4). The remaining scans (3–5)
were similarly executed with the parameters shown in
Table 2. The gradient values were chosen to be the optimal
2D combination determined in the previous section
(Gy2=�Gy1, or h=1358) when the constant gradient was
included and a similarly determined optimum (Gz2=2Gz1)
when it was not. Scans 2 and 5 were breferenceQ scans as
previously described. Finally, while a ratio of Scans 4 and 2,
in principle, provides a 3D tensor measurement in two
scans, in practice the condition number of that experiment
was too high to give accurate results. Work is in progress to
Fig. 6. Ratio images acquired from a pair of MEDITATE scans with
parameters shown in the first two rows of Table 2.
E.E. Sigmund, Y.-Q. Song / Magnetic Resonance Imaging 24 (2006) 7–1814
determine a better conditioned MEDITATE-3D sequence;
the proceeding data were all acquired with different
MEDITATE-2D scans.
Fig. 7. Amplitude profiles for two ROIs in the asparagus phantom and the
ratio of scans 1 and 2. Upper panel: diffusion-weighting pattern for this
case. Middle panel: raw image and ROI sketch. Lower panel: amplitude
profiles for the two ROIs shown in the middle panel. Circles: bmouthQregion. Triangles: bnoseQ region.
3. Results
Fig. 5 shows the 13 raw images acquired from Scan 1.
The amplitude variation in these images includes effects due
to diffusion, relaxation and excitation flip angle. To
highlight the diffusion effects of interest, Fig. 6 shows
images of the amplitude ratio of Scans 1 and 2, which
eliminates all flip angle dependence, as discussed in
Methods. The diffusion-weighting factors bij corresponding
to these ratio data (the difference of the weighting factors for
Scans 1 and 2) are shown in the upper panel of Fig. 7. This
pattern affords a gradually increasing weighting for diffu-
sion along the x-axis, and a bsteppedQ weighting along the
y-axis (Echoes 4, 5 and 6 receive larger weighting than all of
the others). Inspection of the ratio images in Fig. 6 shows a
qualitative correspondence to this pattern. Namely, images
4, 5 and 6 clearly experience more diffusion weighting than
adjacent images in the sequence; furthermore, the bnoseQregion of the phantom shows more decay than the other
regions, consistent with its less restricted diffusivity along
the y-axis. Similarly, the bmouthQ region gradually darkens
more than the other regions for Images 10, 11 and 12; this is
consistent with the gradually increasing diffusion weighting
along the x-axis and the larger Dxx value in the bmouthQregion. More quantitative examples of the diffusion behav-
ior of the asparagus phantom are shown in Fig. 7. The upper
panel shows the diffusion-weighting factors (bxx, byy, bxy)
for this sequence. The middle panel shows a raw image of
the first echo on which regions-of-interest (ROIs) have been
drawn. The lower panel shows the amplitudes, taken from
the full set of ratio images shown in Fig. 6, averaged over
each color-coded ROI. Qualitatively, we observe deeper
response to the bsteppedQ byy pattern in the bnoseQ ROI, anda steeper response to the gradual rise in the bxx pattern in the
bmouthQ ROI. The diffusion tensors calculated from these
two ROIs are as follows: (nose) Dxx/Dyy/Dxy=1.5/2.4/
�0.1*10�3 mm2/s; (mouth) Dxx /Dyy /Dxy = 2.0/1.6/
Dxx Dyy Dzz Directivity
3.0
2.5
2.0
1.5
1.0
0.5
0.0
D (1
0-3 m
m2/s)
Fig. 8. Diffusion tensor element and directivity results from three separate MEDITATE-2D experiments and their average. In directivity plot, color-coding
represents the direction of the primary diffusion eigenvector at each voxel. Red corresponds to diffusion along the x-axis, green to the y-axis and blue to the
z-axis. The top three rows each measure diffusion in a different plane, while the bottom row contains the average of all three 2D tensors.
Raw Image FA
Trace (D) Directivity
1.0
0.8
0.6
0.4
0.2
0.0
FA
2.0
1.5
1.0
0.5
0.0
<λ>
(10 m
m2/s)
-3
Fig. 9. Diffusion tensor image parameters obtained from the MEDITATE
scans in the asparagus phantom.
E.E. Sigmund, Y.-Q. Song / Magnetic Resonance Imaging 24 (2006) 7–18 15
�0.1*10�3 mm2/s. In both cases, the expectation of the
diffusion anisotropy is borne out.
The aforementioned analysis was carried out for every
voxel in the image, with a sliding average box of 5�5
voxels or 2.5�4 mm, thus providing a measure of the 2D
diffusion tensor in the x–y plane for this image. The results
are shown in the upper row of Fig. 8. The first three
columns show the maps of the diagonal elements of the
diffusion tensor. We see the relative brightness of the
bmouthQ in the Dxx image and of the bnoseQ in the Dyy
image, consistent with the orientation of the asparagus fibers
for those regions.
The rightmost column displays the color-coded directiv-
ity. This map shows the local orientation of the principal
eigenvector as a color, with (red, green, blue) components
proportional to the three components (ex, ey, ez) of the
eigenvector [55]. For the first row, we see the expected
result of predominant x-diffusion in the bmouthQ and
predominant y-diffusion in the bnoseQ, and some competi-
tion in the beyesQ. A similar experiment and analysis was
carried out for two more pairs of data, with each pair
measuring the diffusion tensor in another plane ((x,y), (y,z)).
The pulse sequence parameters used are listed in Table 1.
The results are shown as rows 2 and 3 in Fig. 8; row 2 was
obtained through a ratio of Scans 3 and 2, while row 3 was
obtained through a ratio of Scans 4 and 5. To combine the
information from the three rows, first an average diagonal
element Dii was determined in each column. These three
diagonal elements were combined with the off-diagonal
element Dij measured in each row (not shown) to produce
the full (six-element) diffusion tensor. The diagonal
E.E. Sigmund, Y.-Q. Song / Magnetic Resonance Imaging 24 (2006) 7–1816
elements and directivity maps of this tensor are shown in the
bottom row. The 3D directivity map compactly displays the
complete directional information. The standard DTI modal-
ity of directional coverage requires at least seven acquis-
itions (scans) to obtain the same result, compared to the five
acquired here.
In Fig. 9, we show a set of images representing several
common outputs of the diffusion tensor measurement. The
upper left-hand panel shows a raw image. The upper right-
and lower left-hand panels show, respectively, the FA and
Trace(D) parameters, calculated from the diffusion eigen-
values ki at each voxel using Eqs. (3) and (4). We see that
the Trace(D) parameter is fairly uniform in all regions of the
phantom at a value of 1.6*10�3 mm2/s. The FA value is also
mostly uniform in the nose and mouth at about FA=0.3,
with a slightly lower value in the eyes of FA=0.15.
4. Discussion
Overall, the results of the diffusion tensor images are
reasonable. In each portion of the asparagus phantom the
stalk axis orientation is correctly identified, while the
average diffusivity (Trace(D)) is correctly equal for all
portions. The reason for the slight discrepancy in FA values
in different regions of the phantom is unclear; two
possibilities are (1) the partial volume averaging in the eyes
is along the longitudinal stalk axis compared to along the
radial axis for the other two, or (2) some free water may
have been resting above and below the beyeQ sections, thuslowering the average anisotropy.
The optimization described in this work for the MED-
ITATE-2D sequence generated a viable sequence for 2D
DTI, as the phantom results show. While several such
sequences can be combined as in this work to measure a 3D
tensor, a much more efficient modality would employ all
three gradient directions together in a MEDITATE-3D
sequence. However, as mentioned previously, the optimiza-
tion of this sequence is nontrivial. The minimum condition
number attained by a similar optimization as in the present
work is a factor of 10 higher than the 2D minimum
condition number of 3.5. Standard DTI direction sets have
been optimized according to additional parameters beyond
condition number, such as rotational invariance [16]. This
property may be useful in generating viable MEDITATE-3D
sequences as well.
The MEDITATE technique provides the opportunity to
obtain diffusion tensor images with fewer artifacts than
standard methods while maintaining a reasonable scan time.
This advantage comes at some cost in versatility, since
systems demonstrating time-dependent diffusion, relaxation
or exchange behavior are not compatible with its multiple
echo structure. Similarly, diffusion weighting from suscep-
tibility gradients is a potential problem. However, while
susceptibility-induced gradients are significant in many
geological porous media [56–63], their importance for in
vivo tissue diffusion imaging depends greatly on the
anatomical region of interest (brain, muscle, heart, bone,
etc.) [45,64–66], so there may be regimes (particularly in the
case of brain imaging) for which the susceptibility gradients
are not overly problematic for the MEDITATE sequence.
Although the present stationary phantom study does not
address motion artifacts, we can propose variations on the
MEDITATE sequence that may reduce this problem. Since
the diffusion tensor calculation is overdetermined with
13 echo amplitudes, one of the echoes (e.g., Echo 0) can
instead be used as a navigator echo to correct for small
motions between scans. In the limit of constant velocity
motion, each echo can be corrected according to its own
gradient history. A variation of the MMME sequence has, in
fact, been previously employed using phase information from
an MMME echo train to measure flow [67]; thus, correcting
for constant velocity motion is feasible. Given each of these
issues, the MEDITATE sequence possesses a finite range of
applicability. However, applications that require fast aniso-
tropy imaging such as diffusion-weighted functional MRI
[68,69] or the monitoring of chemical reactions may be
especially suited to the MEDITATE technique.
5. Conclusions
A new methodology for DTI termed MMME Diffusion
Tensor Acquisition Technique (MEDITATE) has been
demonstrated in the anisotropic diffusion phantom aspara-
gus. Through the use of 13 independent spin echoes
generated by a series of four RF pulses, diffusion weighting
of multiple intensities and directions sufficient to calculate
the diffusion tensor are applied and measured within a few
scans. The 2D variety of this sequence (MEDITATE-2D) is
successfully demonstrated in an asparagus phantom. A
commonly used quality index, condition number, is
employed to optimize this sequence for minimal systematic
error and can potentially be useful in optimizing the more
efficient MEDITATE-3D variety.
Acknowledgments
We thank X.-P. Tang, D. Madio, T. Brown and P.
Batchelor for useful discussions.
References
[1] Basser PJ. Inferring microstructural features and the physiological
state of tissues from diffusion-weighted images. NMR Biomed
1995;8(7–8):333–44.
[2] Basser PJ, Mattiello J, Lebihan D. Estimation of the effective self-
diffusion tensor from the Nmr Spin-Echo. J Magn Reson B 1994;
103(3):247–54.
[3] Kiselev VG. Calculation of diffusion effect for arbitrary pulse
sequences. J Magn Reson 2003;164(2):205–11.
[4] Callaghan PT. Principles of nuclear magnetic resonance microscopy.
Oxford7 Oxford University Press; 1993. p. 492.
[5] Sodickson A, Cory DG. A generalized k-space formalism for treating
the spatial aspects of a variety of NMR experiments. Prog Nucl Magn
Reson Spectrosc 1998;33:77–108.
E.E. Sigmund, Y.-Q. Song / Magnetic Resonance Imaging 24 (2006) 7–18 17
[6] Inglis BA, Bossart EL, Buckley DL, Wirth III ED, Mareci TH.
Visualization of neural tissue water compartments using biexponential
diffusion tensor MRI. Magn Reson Med 2001;45(4):580–7.
[7] StehlingMK, Turner R,Mansfield P. Echo-planar imaging—magnetic-
resonance-imaging in a fraction of a second. Science 1991;254(5028):
43–50.
[8] Deoni SCL, Peters TM, Rutt BK. Quantitative diffusion imaging
with steady-state free precession. Magn Reson Med 2004;51(2):
428–33.
[9] Wu EX, Buxton RB. Effect of diffusion on the steady-state
magnetization with pulsed field gradients. J Magn Reson 1990;
90(2):243–53.
[10] Zur Y, Bosak E, Kaplan N. A new diffusion SSFP imaging technique.
Magn Reson Med 1997;37(5):716–22.
[11] Gulani V, Iwamoto GA, Jiang H, Shimony JS, Webb AG, Lauterbur
PC. A multiple echo pulse sequence for diffusion tensor imaging and
its application in excised spinal cords. Magn Reson Med 1997;
38:868–73.
[12] Papadakis NG, Xing D, Huang CL-H, Hall LD, Carpenter TA. A
comparative study of acquisition schemes for diffusion tensor imaging
using MRI. J Magn Reson 1999;137(2):67–82.
[13] Skare S, Hedehus M, Moseley ME, Li TQ. Condition number as a
measure of noise performance of diffusion tensor data acquisition
schemes with MRI. J Magn Reson 2000;147(2):340–52.
[14] Jones DK, Horsfield MA, Simmons A. Optimal strategies for
measuring diffusion in anisotropic systems by magnetic resonance
imaging. Magn Reson Med 1999;42(3):515–25.
[15] Tuch DS, Reese TG, Wiegell MR, Makris N, Belliveau JW, Wedeen
VJ. High angular resolution diffusion imaging reveals intravoxel
white matter fiber heterogeneity. Magn Reson Med 2002;48:577–82.
[16] Batchelor PG, Atkinson D, Hill DLG, Calmante F, Connelly A.
Anisotropic noise propagation in diffusion tensor MRI sampling
schemes. Magn Reson Med 2003;49(6):1143–51.
[17] Hasan KM, Parker DL, Alexander AL. Comparison of gradient
encoding schemes for diffusion-tensor MRI. J Magn Reson Imaging
2001;13(5):769–80.
[18] Ozcan A. (Mathematical) Necessary conditions for the selection of
gradient vectors in DTI. J Magn Reson 2004;172:238–41.
[19] Poonawalla AH, Zhou XHJ. Analytical error propagation in
diffusion anisotropy calculations. J Magn Reson Imaging 2004;
19(4):489–98.
[20] Song Y-Q, Tang X-P. A one-shot method for measurement of
diffusion. J Magn Reson 2004;170:136–48.
[21] Tang X-P, Sigmund EE, Song Y-Q. Simultaneous measurement of
diffusion along multiple directions. J Am Chem Soc 2004;
126(50):16336–7.
[22] Counsell CJ. PREVIEW: a new ultrafast imaging sequence requiring
minimal gradient switching. Magn Reson Imaging 1993;11:603–16.
[23] Heid O, Deimling M, Huk W. QUEST—A quick echo split NMR
imaging technique. Magn Reson Med 1993;29:280–3.
[24] Ong H, Chin C-L, Wehrli SL, Tang X-P, Wehrli FW. A new approach
for simultaneous measurement of ADC and T_2 from echoes
generated via multiple coherence pathways. J Magn Reson 2005;
173:153–9.
[25] Doran SJ, Decorps M. A robust, single-shot method for measuring
diffusion coefficients using the bBurstQ sequence. J Magn Reson, Ser
A 1995;117:311.
[26] Doran SJ, Jakob P, Decorps M. Rapid repetition of the bBurstQsequence: the role of diffusion and consequences for imaging. Magn
Reson Med 1996;35(4):547–53.
[27] Lowe IJ, Wysong RE. DANTE ultrafast imaging sequence (DUFIS).
J Magn Reson B 1993;101:106.
[28] Stamps JP, Ottink B, Visser JM, van Duynhoven JPM, Hulst R.
Difftrain: a novel approach to a true spectroscopic single-scan
diffusion measurement. J Magn Reson 2001;151:28–31.
[29] Chin CL, Tang XP, Bouchard LS, Saha PK, Warren WS, Wehrli FW.
Isolating quantum coherences in structural imaging using intermolec-
ular double-quantum coherence MRI. J Magn Reson 2003;165(2):
309–14.
[30] Shannon KL, Branca RT, Galiana G, Cenzano S, Bouchard LS,
Soboyejo W, et al. Simultaneous acquisition of multiple orders of
intermolecular multiple-quantum coherence images in vivo. Magn
Reson Imaging 2004;22(10):1407–12.
[31] Haacke EM, Brown RW, Thompson MR, Venkatesan R. Magnetic
resonance imaging: physical principles and sequence design. New
York7 Wiley-Liss; 1999.
[32] Hurlimann MD. Diffusion and relaxation effects in general stray field
NMR experiments. J Magn Reson 2001;148:367–78.
[33] Song YQ. Categories of coherence pathways for the CPMG sequence.
J Magn Reson 2002;157(1):82–91.
[34] Hurlimann MD, Helmer KG, Latour LL, Sotak CH. Restricted
diffusion in sedimentary rocks. Determination of surface-to-volume
ratio and surface relaxivity. J Magn Reson, Ser A 1994;111:169–78.
[35] McCall KR, Johnson DL, Guyer RA. Magnetization evolution in
connected pore system: II. Pulse field gradient NMR and pore-space
geometry. Phys Rev, B 1993;48:5997–6006.
[36] Mitra PP, Sen PN, Schwartz LM. Short-time behavior of the diffusion
coefficient as a geometrical probe of porous media. Phys Rev, B
1993;47:8565.
[37] Parsons EC, Does MD, Gore JC. Modified oscillating gradient pulses
for direct sampling of the diffusion spectrum suitable for imaging
sequences. Magn Reson Imaging 2003;21(3–4):279–85.
[38] Sen PN. Time-dependent diffusion coefficient as a probe of geometry.
Concepts Magn Reson 2004;23:1–21.
[39] Williams WD, Seymour EFW, Cotts RM. A pulsed-gradient multiple-
spin-echo NMR technique for measuring diffusion in the presence
of background magnetic field gradients. J Magn Reson 1978;31:
271–82.
[40] Karlicek JRF, Lowe IJ. A modified pulsed gradient technique for
measuring diffusion in the presence of large background gradients.
J Magn Reson (1969) 1980;37(1):75.
[41] Corns RM, Hoch MJR, Sun T, Markert JT. Pulsed field gradient
stimulated echo methods for improved NMR diffusion measurements
in heterogeneous systems. J Magn Reson 1989;83(2):252.
[42] Latour LL, Li LM, Sotak CH. Improved PFG stimulated-echo method
for the measurement of diffusion in inhomogeneous fields. J Magn
Reson B 1993;101(1):72.
[43] Lian J, Williams DS, Lowe IJ. Magnetic resonance imaging of
diffusion in the presence of background gradients and imaging of
background gradients. J Magn Reson, Ser A 1994;106(1):65.
[44] Trudeau JD, Dixon WT, Hawkins J. The effect of inhomogeneous
sample susceptibility on measured diffusion anisotropy using NMR
imaging. J Magn Reson B 1995;108(1):22.
[45] Clark CA, Barker GJ, Tofts PS. An in vivo evaluation of the effects
of local magnetic susceptibility-induced gradients on water
diffusion measurements in human brain. J Magn Reson 1999;141(1):
52–61.
[46] Johnson CS. Diffusion ordered nuclear magnetic resonance spectros-
copy: principles and applications. Prog Nucl Magn Reson Spectrosc
1999;34(3–4):203–56.
[47] Seland JG, Sorland GH, Zick K, Hafskjold B. Diffusion measure-
ments at long observation times in the presence of spatially variable
internal magnetic field gradients. J Magn Reson 2000;146(1):14–9.
[48] Mohoric A. A modified PGSE for measuring diffusion in the pre-
sence of static magnetic field gradients. J Magn Reson 2005;174(2):
223–8.
[49] Galvosas P, Stallmach F, Karger J. Background gradient suppression
in stimulated echo NMR diffusion studies using magic pulsed field
gradient ratios. J Magn Reson 2004;166(2):164.
[50] Sun PZ, Seland JG, Cory D. Background gradient suppression in
pulsed gradient stimulated echo measurements. J Magn Reson 2003;
161(2):168.
[51] Press WH, Teukolsky SA, Vetterling WT, Flannery BP. Numerical
recipes in C. Cambridge7 Cambridge University Press; 1997.
E.E. Sigmund, Y.-Q. Song / Magnetic Resonance Imaging 24 (2006) 7–1818
[52] Strang G. Introduction to linear Algebra. Wellesley, MA7 Wellesley-
Cambridge Press; 1998.
[53] Stejskal EO, Tanner JE. Spin diffusion measurements: spin echoes in
the presence of a time-dependent field gradient. J Chem Phys 1965;
42:288–92.
[54] Boujraf S, Luypaert R, Eisendrath H, Osteaux M. Echo planar
magnetic resonance imaging of anisotropic diffusion in asparagus
stems. Magn Reson Mater Phys Biol Med 2001;13:82–90.
[55] Pajevic S, Pierpaoli C. Color schemes to represent the orientation of
anisotropic tissues from diffusion tensor data: Application to white
matter fiber tract mapping in the human brain. Magn Reson Med
1999;42(3):526–40.
[56] Audoly B, Sen PN, Ryu S, Song Y-Q. Correlation functions for
inhomogeneous magnetic field in random media with application to a
dense random pack of spheres. J Magn Reson 2003;164:154–9.
[57] Borgia GC, Brown RJS, Fantazzini P. The effect of diffusion and
susceptibility differences on T2 measurements for fluid in porous
media and biological tissues. Magn Reson Imaging 1996;14:731–6.
[58] Hurlimann MD. Effective gradients in porous media due to
susceptibility differences. J Magn Reson 1998;131:232–40.
[59] Petkovic J, Huinink HP, Pel L, Kopinga K. Diffusion in porous
building materials with high internal magnetic field gradients. J Magn
Reson 2004;167:97–106.
[60] Seland JG, Washburn KE, Anthonsen HW, Krane J. Correlations
between diffusion, internal magnetic field gradients, and transverse
relaxation in porous systems containing oil and water. Phys Rev, E
2004;70(5):051305.
[61] Sen PN, Axelrod S. Inhomogeneity in local magnetic field due to
susceptibility contrast. J Appl Phys 1999;86:4548–54.
[62] Song Y-Q. Using internal magnetic fields to obtain pore size
distributions of porous media. Concepts Magn Reson 2003;18A(2):
97–110.
[63] Song Y-Q, Ryu S, Sen PN. Determining multiple length scales in
rocks. Nature (London) 2000;406:178–81.
[64] Kiselev VG. Effect of magnetic field gradients induced by microvas-
culature on NMR measurements of molecular self-diffusion in
biological tissues. J Magn Reson 2004;170(2):228–35.
[65] Majumdar S, Gore JC. Studies of diffusion in random fields produced
by variations in susceptibility. J Magn Reson 1988;78:41–55.
[66] Zhong J, Kennan RP, Gore JC. Effects of susceptibility variations on
NMR measurements of diffusion. J Magn Reson 1991;95(2):267.
[67] Song YQ, Scheven UM. An NMR technique for rapid measurement of
flow. J Magn Reson 2005;172(1):31–5.
[68] Song AW, Harshbarger T, Li TL, Kim KH, Ugurbil K, Mori S, et al.
Functional activation using apparent diffusion coefficient-dependent
contrast allows better spatial localization to the neuronal activity:
evidence using diffusion tensor imaging and fiber tracking. Neuro-
Image 2003;20(2):955–61.
[69] Song AW, Wong EC, Tan SG, Hyde JS. Diffusion weighted fMRI at
1.5 T. Magn Reson Med 1996;35(2):155–8.