aquametry apparatus - cole-parmer

28
Aquametry Apparatus OPERATION MANUAL AND PARTS LIST 8-1185 • 3/3/05 Serial Number ______________________________________

Upload: khangminh22

Post on 19-Jan-2023

1 views

Category:

Documents


0 download

TRANSCRIPT

Aquametry ApparatusOPERATION MANUAL

AND PARTS LIST

8-1185 • 3/3/05 Serial Number ______________________________________

2

IF THERE IS BREAKAGE, DO NOT discard the packaging.

Contact your supplier. Supplier should initiate any claim for damages. BarnsteadInternational cannot be held responsible for damages to shipments not shipped directlyfrom Barnstead International.

FOLLOW INSTRUCTIONS BELOW ONLY IF THE SHIPPER IS Barnstead International.

UPS (UNITED PARCEL SERVICE), TRUCK LINE, RAIL, OR AIR FREIGHT SHIPMENTS

1. LEAVE instrument in the same shipping carton in which it was received. HOLD it in yourreceiving department.

2. CONTACT the delivering carrier to report the damage. Request the carrier to inspect thedamage and to return the shipment to Barnstead International. The packing list should bereturned with the shipment.

3. DO NOT RETURN the shipment yourself. Claims are invalid without an inspection by thecarrier.

4. A REPLACEMENT will be shipped as soon as we receive the damaged instrument andthe claim. For an immediate replacement, contact Barnstead International.

PARCEL POST SHIPMENTS

1. FILE a claim with your Post Office for “partly damaged” goods only. The Post Office willforward the instrument and the claim to Barnstead International.

2. NOTIFY Barnstead International by telephone or mail.

3. A REPLACEMENT will be shipped as soon as we receive the damaged instrument andthe claim. For an immediate replacement, contact Barnstead International.

NOTE: BARNSTEAD INTERNATIONAL CANNOT BE HELD RESPONSIBLE FOR DAM-AGES IN SHIPMENT, IF THESE INSTRUCTIONS ARE NOT FOLLOWED. For furtherinformation call toll free (800) 227-0128 in U.S.A.

ImportantCheck shipment for breakage!

3

Table of Contents

Check Shipment for Breakage ............................................................................................................................2Table of Contents ................................................................................................................................................3Introduction..........................................................................................................................................................4

Important Information....................................................................................................................................4Introduction ..................................................................................................................................................4Water Determinations by the Karl Fischer (Aquametry) Method Dead-Stop Technique ..............................4General Usage..............................................................................................................................................5

Safety Information ..............................................................................................................................................6Unpackaging ......................................................................................................................................................7

Contents of Package ....................................................................................................................................7Breakage in Shipment ..................................................................................................................................7

Specifications ......................................................................................................................................................8Accuracy ......................................................................................................................................................8Reproducibility ..............................................................................................................................................8Moisture Determination ................................................................................................................................8Environmental Conditions ............................................................................................................................8

Installation ..........................................................................................................................................................9Assembly, Aquametry I Apparatus................................................................................................................9Assembly, Aquametry II Apparatus ............................................................................................................11Checking Assembly ....................................................................................................................................12Checking the Aquametry Meter-Electrode for Proper Performance ..........................................................12

Operation ..........................................................................................................................................................14Karl Fischer Aquametry Dead-Stop Technique ..........................................................................................14Achieving Highest Precision ......................................................................................................................16

Adding the Sample ..............................................................................................................................16Titrating ................................................................................................................................................17

Troubleshooting ................................................................................................................................................19Special Problems........................................................................................................................................20

Pretreatment of Materials ....................................................................................................................20Suitable Solvent ..................................................................................................................................20Heating to Release Water ..................................................................................................................20

Maintenance......................................................................................................................................................21Aquametry Meter-Electrode........................................................................................................................21Battery ........................................................................................................................................................21Karl Fischer Reagent ..................................................................................................................................21General Cleaning Instructions ....................................................................................................................21

Parts List ..........................................................................................................................................................22Views ................................................................................................................................................................25

Aquametry I Apparatus ..............................................................................................................................25Aquametry II Apparatus ..............................................................................................................................26

Ordering Procedures ........................................................................................................................................27Warranty ............................................................................................................................................................28

4

Introduction

Congratulations on your purchase of aBarnstead|Labindustries Aquametry Apparatus.

Important InformationThis manual contains important operating andsafety information. You must carefully read andunderstand the contents of this manual prior tousing this equipment.

IntroductionThe unit is designed to do a number of jobs with-in your laboratory. Please read the instructionscarefully to ensure that you receive maximumbenefit from it. Also, be sure to fill out and returnthe enclosed warranty registration card. Wewould like to receive the information requested,and it will help us assure you of proper warrantycoverage.

Water Determinations by the KarlFischer (Aquametry) MethodDead-Stop TechniqueKarl Fischer Reagent (KFR) consists of a mixtureof pyridine, iodine, sulfur dioxide and methanol.KFR reacts with water on an exact volume-for-volume ratio. In an excess of KFR, the reactionchanges from yellow to dark brown because ofthe presence of free iodine. In the presence offree iodine, the conductivity of the solutionincreases dramatically. In an excess of water, thereaction changes from dark brown to yellow indi-cating the absence of free iodine. In the absenceof free iodine, the conductivity of the solutiondecreases.

5

The dead-stop aquametry technique is a rapidmethod of water determination which does notrequire a back titration with water. The solventis first titrated with KFR to the KF end point toeliminate all of the water from the system. Asample containing water is added to the sol-vent/KFR mixture in the reaction vessel. Themixture is then titrated with KFR to bring themixture back to the KF end point. The volumeof titrant used is proportional to the amount ofwater in the sample.

The Labindustries Aquametry Apparatus is aclosed, odorless, moisture-proof system for thedetermination of water by titration with KFR.The Labindustries Meter-Electrode is a self-powered conductivity meter which preciselyindicates the water and KF end points.

KFR Titre:mg (µl) H20

F = ml KFR

Mass (volume) of water in sample:mg (µl) H20 = A x F

Percent water in sample:A x F x 0.1

% H20 =S

Where:A = Milliliters of KFR required for sample

titrationF = Water equivalence of KFRS = Grams of sample

General UsageDo not use this product for anything other thanits intended usage.

INTRODUCTION

6

Safety Information

Your Barnstead|Labindustries AquametryApparatus has been designed with function, relia-bility, and safety in mind. It is your responsibilityto install it in conformance with local electricalcodes. For safe operation, please pay attentionto the alert boxes throughout the manual.

WarningWarnings alert you to a possibility of personal injury.

CautionCautions alert you to a possibility of damage to theequipment.

NoteNotes alert you to pertinent factsand conditions.

7

Remove components from the shipping cartonsvery carefully, examining them for breakage,defects or shortages. Do not discard packagingif breakage is found.

Contents of PackageExamine the contents of the shipping cartoncarefully. Each package includes the following:

Aquametry I1 Meter-Electrode1 Dual micro/macro Pyrex® glass buret

with 3-way stopcock and clamp1 Glass adapter and filling device3 6 oz. reaction vessels3 Calcium Sulfate drying tubes1 110 volt stirrer and stirring bar1 Lab stand and 2 clamps1 Teflon® tip

Aquametry II1 Meter Electrode1 Zero-adjusting buret with 2-way glass

stopcock1 Fluorocarbon adapter1 Filling device3 Calcium Sulfate drying tubes1 110 volt stirrer and stirring bar1 Teflon tip1 6 oz. reaction vessel1 Lab stand and 2 clamps

Breakage in ShipmentNotify your supplier at once if you find break-age, defects, or shortages upon opening pack-ages. DO NOT RETURN INSTRUMENTSUNTIL INSTRUCTED BY YOUR SUPPLIER.

Unpackaging

PYREX® is a registered trademark of Corning Inc.

TEFLON® is a registered trademark of Dupont.

8

Accuracy1% of full scale volume

Reproducibility0.5% of full scale volume

Moisture DeterminationRanges from 5 ppm to 100% water

Environmental ConditionsOperating: 17°C - 27°C; 20% - 80% relative humidity, non-condensing.

Installation Category II (over-voltage) in accordancewith IEC 664. Pollution Degree 2 in accordance withIEC 664.

Altitude limit: 2,000 meters.Storage: -25°C - 65°C; 10% - 85% relative humidity.

Specifications

9

Assembly, Aquametry IApparatusThe Aquametry I Apparatus, Catalog No. 13216,is supplied with the following: Aquametry Meter-Electrode. This five-ounce,self-powered, tiny conductance meter, sensitiveto three micrograms of water, has a plug-in plat-inum electrode. The face of the 1-1/2-inchsquare meter has a yellow area to indicate thewater end point and a brown area to indicate theKFR end point. The heavy platinum ball elec-trode tips are protected by glass. A double plat-inum-to-glass seal eliminates leakage of reagentinto the electrode.

The Meter-Electrode is a complete instrument initself. It is available separately to your specifica-tions for those who already have a titration appa-ratus but who wish to achieve higher precision.

Dual Buret. The Labindustries Dual Buret con-sists of a one ml buret with 0.01 ml divisionsjoined to a 25 ml buret with 0.1 ml divisions. Atthe base of the dual buret is a 3-way stopcock.Either one or both burets may be used in adetermination and the unused KFR returned tothe bottle.

Adapter. There are two 10/18 joints in the topof the adapter and one 13/18 sample port. Oneaccepts the Dual Buret, another the Meter-Electrode. The third is recessed and is used forthe introduction of sample. It is kept closed witha ground glass stopper during titration.

Three jars.

Filling Device. Transfer of KFR from stock bottleto buret is made through a Teflon connector withdry air under pressure generated by squeezing arubber bulb. The Filling Device is furnished with

Installation

TsTs

10

five sets of caps and washers (28-400, 28-430, 33-400, 33-430 and 38-430) for fitting most bottles ofKFR available on the market.

Three Drying Tubes. BE CERTAIN TO REMOVEPLUGS FROM BOTH ENDS OF THE DRYINGTUBES OR THE INSTRUMENT WILL NOT WORK.

The Aquametry I Apparatus, Catalog No. 13226, isthe same as Catalog No. 13216, but includes inaddition:

Magnetic Stirrer, 110 VAC, with a TEFLON-coatedstirring bar. For 220 VAC installations, this is a cus-tomer-supplied stirrer.

Lab Stand

Two Clamps

Assembly, Aquametry IIApparatusThe Aquametry II Apparatus is supplied with the fol-lowing:Aquametry II Meter-Electrode. This five-ounce,self-powered, tiny conductance meter, sensitive tothree micrograms of water, has a plug-in platinumelectrode. The face of the1-1/2-inch square meter has a yellow area to indi-cate the water end point and a brown area to indi-cate the KFR end point. The heavy platinum ballelectrode tips are protected by glass. A double plat-inum-to-glass seal, developed by Labindustries,eliminates leakage of reagent into the electrode.

INSTALLATION

NoteBe certain to remove plugfrom both ends of the drying tubes or the instrument will not work.

11

The Meter-Electrode is a complete instrument initself. It is available separately to your specifica-tions for those who already have a titration appa-ratus but who wish to achieve higher precision.

Self-Zeroing Buret. The LabindustriesSelf-Zeroing Buret consists of one 25 ml buretwith 0.1 ml divisions. The buret automaticallyzeroes itself upon filling by returning any excessKFR to the bottle.

Adapter. There are three ports in the top of theadapter and one air vent. One port accepts theSelf-Zeroing Buret, one theMeter-Electrode, and one the sample to be titrat-ed. The sample port is kept closed with a groundglass stopper during titration.

One jar.

Filling Device. Transfer of KFR from stock bottleto buret is made through a TEFLON connectorwith dry air under pressure generated by squeez-ing a rubber bulb. The Filling Device is furnishedwith five sets of caps and washers(28-400, 28-430, 33-400, 33-430 and 38-430) forfitting most bottles of KFR available on the mar-ket.

Three Drying Tubes. BE CERTAIN TO REMOVEPLUGS FROM BOTH ENDS OF THE DRYINGTUBES OR THE INSTRUMENT WILL NOTWORK.

Magnetic Stirrer, 110 VAC, with a TEFLON-coat-ed stirring bar. For 220 VAC installations, this is acustomer-supplied stirrer.

Lab Stand

One Clamp

Ts

INSTALLATION

12

Checking AssemblyAfter assembly is completed and before checkingfor airtightness, allow the Aquametry Apparatusto stand for a few hours with KFR in the reactionvessel. This will dry out the moisture in the instru-ment itself. Check for airtightness. Remove theplug from the drying tube at the top of the buret.For this check, be certain that the plug is in thedrying tube which is attached to the reaction ves-sel.Pump KFR into the buret with the stop cock inthe closed position. Turn the buret stopcock to letthe KFR into the reaction vessel. The reagentshould soon stop flowing, assuring you that thereaction vessel is properly sealed and airtight.Now remove the plug from the drying tube whichis attached to the reaction vessel. KFR shouldflow into the vessel.

The above check for airtightness is necessary,because, should the apparatus not be airtight,ordinary moist air in the reaction vessel mightrequire as much as 0.1 ml to react with the mois-ture before an accurate determination can bemade.

Checking the Aquametry Meter-Electrode for ProperPerformanceRefer to the views of Aquametry I Apparatus(page 26) and Aquametry II Apparatus (page 27)in this manual.

1. Remove the Meter-Electrode from theapparatus and gently wipe dry the endwith the platinum balls.

2. The meter should read close to zero. If itis off more than a division from zero, readjust closer to zero by turning the zero set — the small, clear plastic, slot-ted screw head in the transparent plastic

INSTALLATION

NoteBe certain to remove plugs from both ends of the drying tubes or the instrument will not work.

13

face of the meter. The number of divisionsthe zero set can be adjusted is limited toone or two.

3. Touch a metal paper clip across the plat-inum balls on the electrode. The meter should read full scale 5, but the needle must not go beyond 5. Use the full scale locking adjustment in the rear of theMeter-Electrode to position the needle at 5 while the platinum tips are shorted with thepaper clip. If the meter needle reads 5 when the electrode is shorted, the instru-ment is OK.

If when shorting the platinum balls, the needle readsless than 5, and the full scale locking adjustmentcannot quite bring the needle to 5, the battery, RM-IR, should be replaced.

If the meter needle reads zero when the balls areshorted, remove the electrode by unscrewing theelectrode sleeve and pulling the electrode out of itssocket. Open the paper clip and short the electrode socket in the meter with the ends of thepaper clip. If the meter now reads 5, the meter isOK, but the electrode is defective and should bereturned to Barnstead International for repair.

INSTALLATION

14

Karl Fischer AquametryDead-Stop TechniqueThe “dead-stop” technique described below is rec-ommended by Barnstead International as an accu-rate, far more rapid and simpler method for waterdeterminations than others requiring back titrationwith water.

1. Place the TEFLON-coated stirring bar in the reaction vessel.

2. Screw the adapter assembly in place.

3. Add approximately 50 ml of the selected solvent to the reaction vessel through the sample port.

4. Turn on the magnetic stirrer.

5. Aquametry I — Turn the buret stopcock to the filling position. Hold one finger on the air-release hole in the rubber bulb and pump the bulb to fill the buret. Close the stopcock when the Karl Fischer Reagent reaches the desired level in the buret.

5a. Aquametry II — Turn the buret stopcock to the closed position. Hold one finger onthe air-release hole in the rubber bulb andpump the bulb to free the buret. Fill theburet past the self-zeroing point andremove finger from the air-release hole.This will return the excess KFR to the bot-tle.

6. Remove the water in the solvent byadding enough KFR to just turn the solu-tion color from yellow to dark brown.This is known as the KF end point. Note

Operation

15

and record the meter indication on theAquametry Meter-Electrode conductancemeter. You may titrate to any point in thebrown KF zone, provided that you alwaystitrate to that same end point in all subse-quent samplings in the series. Allow thesolution to stabilize at end point for at leastone minute before proceeding to the nextstep.

7. Standardize the water equivalence of the KFR. Normally, this needs to be done only once a day or when changing the KFR sup-ply.

a. Into the reaction vessel, still at the end point, pipet a weighed or accu-rately measured volume of pure water,usually 15 mg (15 µl), through the sample port. For greatest accuracy, weigh the pipet before and after dispensing to determine the exact weight dispensed.

b. Titrate the water to the same end point as in Step 6 (1 ml of fresh KFR equivalent to approximately 5 mg (µl) water).

c. Compute the KFR water equivalence, F. F = mg (µl) water added per ml KFR to return to the KF end point (meter reading before the water was added).

8. To the reaction vessel, still at the end point,add a weighed amount of your sample whichis of sufficient size to contain 20 to 50 mg(µl) water. A sufficiency of water is indicatedby the solution color change to light yellowand/or the meter registering below the KFzone.

OPERATION

16

9. Titrate the water as in Step 6 to the original end point.

10. CALCULATION: Calculate the water content of the sample as follows:

(A x F x 0.1)Water, percent= S where:

A = Milliliters of KFR required for titration of sample

F = Water equivalence of KFR as determined in Step 7(expressed in mg (µl) of water per ml of KFR) and

S = Grams of sample

Re Aquant®, a proprietary product of J.T. Baker Company, may be used instead of KFR. It has the advantage of having no disagreeable odor by the disadvantage of requiring two solutions. The second solution contains the sulfur component, and enough of the second solution must be used to provide adequate buffering capacity, i.e. 1 ml for each 5 mg (µl) of water in the sample.

Achieving Highest Precision andEfficiency

Adding the SampleA liquid may be introduced into the reaction vesselthrough the sample port by removing the pennyhead stopper. Be certain to keep the openingclosed securely at all times when it is not in use.

A powdered sample may be introduced into thevessel by unscrewing the vessel from the adapter,

OPERATION

17

exposing the conveniently large opening of thejar. Add the sample to the solution and rapidlyscrew the jar back into the adapter to avoid theentrance of moisture into the apparatus.Powdered samples may also be added by dis-solving them in solvent and introducing themthrough the sample port in the adapter.When working with several parts per million ofwater, one cannot expose the solution withoutadding several micrograms of moisture from theair. For such required sensitivity, the jar must notbe opened. Replace the stopper of the sampleport in the adapter with a rubber capsule coverand add the sample or standard using a syringeand attached needle. An easily available capsulecover is the stopper in any blood collecting tube.

OPERATION

NoteFOR AQUAMETRY I APPARATUS OWNERS:

During a titration, if the dual buret has been filled to the zero line of the 1 ml buret and more than 1 ml is required, close the stopcock when the Karl Fischer Reagent reaches the 1 ml calibration line. Turn the stopcock to empty the Karl Fischer Reagent back into its bottle, lowering the of level of Karl Fischer Reagent in the buret to the zero line of the 25 ml buret. Resume the titration, using the final reading of the 25 ml buret plus 1 ml for the value of A in the calculations. Thus, either the micro or the standard buret may be chosen without sacrificing the airtight integrity of the apparatus, extending the high precision and speed of the determina-tion throughout the entire range of the instrument.

18

TitratingAdd solvent to the reaction vessel until the sol-vent touches the platinum balls of the AquametryMeter-Electrode while the magnetic stirring bar isrotating steadily beneath the electrode. (Irregularstirring will cause erratic indication of the endpoint.) Add KFR until the KFR end point isreached. Add more KFR in a few minutes if theend point is not held. When the true end point isreached, it will hold for hours. To titrate anothersample, refill the buret to the zero line.

When titrating, either the reaction must be stirredor the electrode gently moved to remove polariza-tion products. The Labindustries meter is alwaysready to operate. No switch is required.

Add KFR slowly to prevent errors caused by rapidheating of the reaction. Keep the stirrer motorrunning at a constant speed. A change in stirrerspeed may shift the end point. Titrate to the samenumber of the meter dial, e.g. 44, and allow themeter to indicate that number for at least 15 sec-onds before considering the titration completed. Ifthe meter indication drops back to the excesswater side of the end point, the sample should beleft in the jar for a longer period before titrating. Ashifting end point may indicate that the apparatusis not quite dry. Letting the apparatus stand withKFR in the jar will dry it out, since KFR is a veryeffective desiccant. A shifting end point might alsoindicate an interference with a substance otherthan water. Sometimes cooling the reaction ves-sel will eliminate the shifting end point.

OPERATION

19

SYMPTOM POSSIBLE CAUSE REMEDY

Difficult or impossible to fillburet.

Drying tube caps still attached todrying tubes.

TYGON® tubing attaching dryingtube to buret may be crimped orcollapsed.

Bottom washer within the KFRbottle cap is either split or missing.

Screw cap on the KFR bottle maynot be tightly closed.

Remove drying tube caps (plugs).

Cut away crimped section and reattachdrying tube, or use a new piece ofTYGON tubing.

Bottom washer must be a whole,unsplit washer to seal the bottle andglass-filling device. Either rearrangewasher correctly or obtain a properwasher, if the washer is missing.

Tighten screw cap.

Shifting end point. Change in stirrer speed.

Rapid heating of the reactionvessel.

Interference by substance otherthan water.

Do not change stirrer speed. Keep thestirrer motor running at a constantspeed.

Let the apparatus stand with KFR inthe reaction vessel to dry out theapparatus.

Add KFR slowly. Cool reaction vessel.

Material may need pretreatment. (SeeSpecial Problems, Pretreatment ofMaterials, page 21).

End point cannot be reached. Unsuitable solvent. Some materialswill not dissolve in methanol.

Moisture released with difficulty.

KFR has deteriorated.

KFR has yellow or blackprecipitate.

Platinum balls of Meter-Electrodehave lost sensitivity after long(500) hours of use.

RM-1R battery in meter is dead.

RM-1R battery mounting terminalsare corroded.

Meter is out of adjustment.

Select a solvent in which material issoluable. (See Special Problems,Suitable Solvent, page 21).

Heating may be required to releasewater.

Use fresh KFR.

Use fresh KFR.

Wash platinum balls in HCl to increasesensitivity.

Replace battery.

Remove battery, wipe contacts clean.Press contacts closer together toincrease contact pressure. Reinstallbattery.

Check Aquametry Meter-Electrodeaccording to instructions on page 13.CHECKING AQUAMETRY METER-ELECTRODE FOR PROPERPERFORMANCE.

Troubleshooting

20

Special Problems

Pretreatment of MaterialsAlthough water in most materials may be titratedwith KFR without difficulty, some materials reactwith KFR and require some pretreatment to permitan accurate water determination. Some of thesesubstances which require pretreatment are metalhydroxides and oxides, carbonates and bicarbon-ates, strong oxidants and reductants, boric acid,some metallic chlorides, ammonia, hydrogen sul-fide and sodium sulfide. To minimize side reac-tions, dilute the KFR to a titre of 1 or 2 mg (µl) percc. Methyl cellosolve and methanol are suitablediluents, depending upon the brand of KFR used.

Suitable SolventAlthough methanol is the most widely used solventin aquametry, some materials will not dissolve inmethanol and release the water to react with KFR.Select a solvent in which the material is soluble.We suggest methylene chloride for petroleumproducts, dimethyl formamide for biologicals,tetrahydrofuran for plastics, dioxane for bicarbon-ates.

Heating to Release WaterSometimes moisture may be released with difficul-ty and heating is required to speed up the releaseof water.

TROUBLESHOOTING

21

Aquametry Meter-ElectrodeAfter every 500 hours of use, wash the platinumballs in hydrochloric acid to increase the elec-trode sensitivity.

Check the Meter-Electrode for proper perfor-mance. See page 13.

BatteryThe Mallory RM-IR battery in the AquametryMeter-Electrode should last for one year withordinary usage. The back may be removed toexpose this battery for replacement by removingthe two sheet metal screws that secure it in posi-tion.

Karl Fischer ReagentKFR deteriorates with time. When measuring thetitre of the KFR, observe the mixture. Thereshould be no precipitate. If a yellow powdery pre-cipitate or a very fine black precipitate forms, donot use the reagent. To remove the precipitateleft on the glassware, rinse with methanol.Occasionally a few deposits of yellow will flakeout of the KFR and deposit on the walls of theburet. When excessive, rinse out the buret withmethanol.

General Cleaning InstructionsWipe exterior surfaces with a dampened clothcontaining a mild soap solution.

Maintenance

22

Burets

Aquametry I Aquametry IICatalog No. Catalog No. Description1125L — Burets, dual, 1 and 25 ml, with balljoint and clamp, 3-way stopcock.1125A — Lower half of dual buret, 25 ml, with

stopcock.1125B — Upper half of dual buret, 1 ml.1125C — Clamp for dual buret.1125G — 3-way glass stopcock plug.1125T — 3-way TEFLON stopcock plug.1125R 1135R Stopcock plug retainer.11266 11266 TEFLON tip with lock nut buret.1130 1130 Lab Clamp— 1135 Buret, single, 25 ml, zero-adjustingwith 2-way stopcock.— 1135G Glass #2 stopcock plug.

Aquametry Meter-Electrode

Aquametry I Aquametry IICatalog No. Catalog No. Description1002A6 — Meter assembly with straight electrode.13326 — Electrode only (straight).— 1003-6 Meter assembly with bent electrode.— 13336 Electrode only (bent).1131 1131 Meter only.1328 1328 Meter Battery.

Parts List

23

Vessels

Aquametry I Aquametry IICatalog No. Catalog No. Description11626 11626 Jars, 6 oz., with caps, 63-400. Case of 24.1162S6 1162S6 Jar, 6 oz., with cap, 63-400. Single.

Adapter

Aquametry I Aquametry IICatalog No. Catalog No. Description— 1145 HALAR® adapter, only, with three

openings and tubulation for drying tube.1158 — TEFLON screw cap, 63-400, and gasket.1158A — TEFLON screw cap, 63-400 only.1159 — PYREX glass adapter, only, with two

integral 10/18 joints and one 12/18joint.

11606 — Adapter kit: Glass adapter, cap and gaskets, three 6 oz. jars, and twoclosed-jar lids.

1159P 1159P Stopper, pennyhead.1161 — TEFLON gasket.1161S 1161S Silicone gasket.

Ts Ts

PARTS LIST

HALAR® is a registered trademark of AUSIMONT U.S.A. Inc.

24

Filling Device and Connectors

Aquametry I Aquametry IICatalog No. Catalog No. Description1165 — Filling device kit: Davol bulb, drying

tube, 12” inlet tube, 12” connector tube, two clamps, filling device and cap set.

— 11652 Filling device kit: Davol bulb, drying tube, 12” inlet tube, 18” connector tube, two clamps, filling device and cap set.

1165A 1165A Filling device, 12” inlet tube and cap set.1127 1127 Davol bulb.1128 — 12” connector tube and two clamps.— 11282 18” connector tube and two clamps.131 131 Clamp, nickel plated for 5 mm tube.

Air Moisture Absorbers

Aquametry I Aquametry IICatalog No. Catalog No. Description1140 1140 Labindustries calcium sulfate drying

tube.

Stirrer

Aquametry I Aquametry IICatalog No. Catalog No. Description1269 1269 Synchrostir Magnetic Stirrer, 110 VAC.1271 1271 AC/DC Magnetic stirrer.ZSX122A ZSX122A TEFLON-coated stirring bar.

PARTS LIST

25

Views

Aquametry I Apparatus

Upper Half ofDual Buret, 1ml

Clamp

Lower Half ofDual Buret, 25ml

Meter Assembly

Cap AssemblyGlass Adapter

Drying Tube

Stirring BarMagnetic Stirrer

(120V model only)

DryingTube

Lab Stand

Lab Clamp (2 required)

Connector Tubeand Clips

Filling Device

Platinum Ball Electrode Tips

10/18 jointTs

Zero Set MeterFull ScaleLockingAdjustment

Aquametry I Meter-Electrode

Karl Fischer Reagent

Drying Tube

Davol Bulb with hole

SamplePort

26

VIEWS

Aquametry II Meter-Electrode

Aquametry II Apparatus

Self-ZeroingBuret

Lab Clamp

Meter-Electrode

Drying Tube

Stirring Bar

Magnetic Stirrer

Karl FischerReagent

Davol Bulb

Drying Tube

Filling Device

Connector Tubeand Clips

Lab Stand

Drying Tube

Zero Set Meter

Full ScaleLockingAdjustment

Platinum BallElectrode Tips

10/18 joint Ts

SamplePort

Adapter

Jar

27

Ordering Procedures

Please refer to the Specification Plate for the complete model number, serial number, andseries number when requesting service, replacement parts or in any correspondence con-cerning this unit.

All parts listed herein may be ordered from the Barnstead International dealer fromwhom you purchased this unit or can be obtained promptly from the factory. When serviceor replacement parts are needed we ask that you check first with your dealer. If the dealercannot handle your request, then contact our Customer Service Department at 563-556-2241 or 800-553-0039.

Prior to returning any materials to Barnstead International, please contact ourCustomer Service Department for a “Return Goods Authorization” number (RGA). Materialreturned without a RGA number will be refused.

28

Barnstead International (“BARNSTEAD”) warrants that if a product manufactured by Barnstead shall be free ofdefects in materials and workmanship for one (1) year from the first to occur of (i) the date the product is sold byBARNSTEAD or (ii) the date the product is purchased by the original retail customer (the “CommencementDate”). Except as expressly stated above, BARNSTEAD MAKES NO OTHER WARRANTY, EXPRESSED ORIMPLIED, WITH RESPECT TO THE PRODUCTS AND EXPRESSLY DISCLAIMS ANY AND ALL WARRANTIES,INCLUDING BUT NOT LIMITED TO, WARRANTIES OF DESIGN, MERCHANT ABILITY AND FITNESS FOR APARTICULAR PURPOSE.

An authorized representative of BARNSTEAD must perform all warranty inspections. In the event of a defectcovered by BARNSTEAD’s warranty, BARNSTEAD shall, as its sole obligation and exclusive remedy, provide freereplacement parts to remedy the defective product. In addition, for products sold by BARNSTEAD within the con-tinental United States or Canada, BARNSTEAD shall provide provide free labor to repair the products with thereplacement parts, but only for a period of ninety (90) days from the Commencement Date.

BARNSTEAD’s warranty provided hereunder shall be null and void and without further force or effect if there isany (i) repair made to the product by a party other than BARNSTEAD or its duly authorized service representa-tive, (ii) misuse (including use inconsistent with written operating instructions for the product), mishandling, conta-mination, overheating, modification or alteration of the product by any customer or third party or (iii) use ofreplacement parts that are obtained from a party who is not an authorized dealer of BARNSTEAD.

Heating elements, because of their susceptibility to overheating and contamination, must be returned to theBARNSTEAD factory and if, upon inspection, it is concluded that failure is due to factors other than excessivehigh temperature or contamination, BARNSTEAD will provide warranty replacement. As a condition to the returnof any product, or any constituent part thereof, to BARNSTEAD’s factory, it shall be sent prepaid and a prior writ-ten authorization from BARNSTEAD assigning a Return Goods Number to the product or part shall be obtained.

IN NO EVENT SHALL BARNSTEAD BE LIABLE TO ANY PARTY FOR ANY DIRECT, INDIRECT, SPECIAL, INCI-DENTAL, OR CONSEQUENTIAL DAMAGES, OR FOR ANY DAMAGES RESULTING FROM LOSS OF USE ORPROFITS, ANTICIPATED OR OTHERWISE, ARISING OUT OF OR IN CONNECTION WITH THE SALE, USEOR PERFORMANCE OF ANY PRODUCTS, WHETHER SUCH CLAIM IS BASED ON CONTRACT, TORT(INCLUDING NEGLIGENCE), ANY THEORY OF STRICT LIABILITY OR REGULATORY ACTION.

The name of the authorized Barnstead International dealer nearest you may be obtained by calling 1-800-446-6060 (563-556-2241) or writing to:

One Year Limited Warranty

2555 Kerper BoulevardDubuque, Iowa 52001-9918 Phone: 563-556-2241 or 800-553-0039 Fax: 563-589-0516E-mail: [email protected]