absorption microspectroscopy, theory and applications in the case of the photosynthetic compartment
TRANSCRIPT
www.elsevier.com/locate/micron
Micron 38 (2007) 197–213
Absorption microspectroscopy, theory and applications in
the case of the photosynthetic compartment
Laura Barsanti a, Valtere Evangelista a, Anna Maria Frassanito a,Nicoletta Vesentini b, Vincenzo Passarelli a, Paolo Gualtieri a,*
a Istituto di Biofisica, Area della Ricerca CNR, via Moruzzi 1, 56124 Pisa, Italyb Istituto di Fisiologia Clinica, Area della Ricerca CNR, via Moruzzi 1, 56124 Pisa, Italy
Abstract
We performed microspectroscopic evaluation of the pigment composition of the photosynthetic compartments of both algae and higher plants.
The feasibility of microspectroscopy for discriminating among species and/or phylogenetic groups was tested on laboratory cultures. Gaussian
bands decompositions, and a fitting algorithm, together with fourth-derivative transformation of absorbance spectra, provided a reliable
discrimination among chlorophylls, phycobiliproteins and carotenoids. Comparative analysis of absorption spectra highlighted the evolutionary
grouping of the algae into three main lineages in accordance with the most recent endosymbiotic theories.
# 2006 Elsevier Ltd. All rights reserved.
Keywords: Microspectroscopy; Algae; Pigment; Endosymbiosis; Gaussian decomposition; Fourth-derivative transformation
1. Introduction
Nature evolved a very limited number of chromophores in
the different evolutionary branches of the tree of life because of
the closely similar needs of organisms for the detection of the
external world and the interaction with it. Two cellular
compartments are the main location of endogenous chromo-
phores, the photoreceptors and the photosynthetic compart-
ments, i.e. the thylakoid membranes. The study of pigments and
their distribution in these cellular structures is usually based on
extractive procedures, followed by biochemical or spectro-
scopic assays. Extractive techniques often include disadvan-
tages: they can modify the nature of the components and they
are not successful in isolating the pigments. Direct spectro-
scopy on intact samples (microspectroscopy) has the advantage
of preserving the integrity of biological structures or
substructures during the measurement of absorption spectra;
if it is joined with successive mathematical computation, it can
discriminate between the different pigment contributions to the
spectral envelope.
In this review we will analyse the performance of
microspectroscopy, describe absorption spectra of thylakoid
compartments belonging to different algal specimens measured
* Corresponding author. Tel.: +39 050 3153026; fax: +39 050 3152760.
E-mail address: [email protected] (P. Gualtieri).
0968-4328/$ – see front matter # 2006 Elsevier Ltd. All rights reserved.
doi:10.1016/j.micron.2006.07.015
under laboratory conditions and compare them with those
present in the literature, which have been recorded under
natural environmental conditions. Spectroscopic techniques
such as fluorescence or reflectance spectroscopy will be
reviewed successively.
Photosynthetic compartments belong to both algae and plants.
Absorption spectra measured in vivo on these compartments, can
give us very precise and accurate information about the spectral
range in which pigment molecules capture photons in their
natural environment. Therefore, the use of microspectrophoto-
metry can expand the perception of taxonomists, who identify
species in relation to natural pigmentation to supplement
classification based on morphology. Pigment distribution is
almost constant in plants, while it shows a high degree of
variation in algae. As consequence it is possible to predict the
presence of a specific pigment in an alga, to give an unknown alga
a plausible taxonomic framing, and to give support to the
phylogenetic tree of endosymbiotic events.
We will examine methods and theory behind a microspe-
trophotometer, and those that are used to identify the pigment
contributions in the absorption spectra of the photosynthetic
compartments of algae and plants.
We will review the following topics:
- B
iological background- D
esign and performance of a microspectrophotometerL. Barsanti et al. / Micron 38 (2007) 197–213198
- T
ransfer of light energy through the optical system- G
eometrical features of a microspectrophotometer- T
heory of detection of photosensitive substances- S
tatistics of photon flux and detection- M
athematical computation of absorption spectra- R
esults of microspectrophotometric investigations on photo-synthetic compartments
- C
omparison of absorption spectra obtained by differentmethods
- C
onclusions2. Biological background
The term algae has no formal taxonomic standing,
nevertheless it is routinely used to indicate a polyphyletic,
non-cohesive and artificial assemblage, of O2-evolving,
photosynthetic organisms. No easily definable classification
system acceptable to all exists for algae, since taxonomy is
under constant and rapid revision at all levels following every
day new genetic and ultrastructural evidence. Keeping in mind
that the polyphyletic nature of the algal group is somewhat
inconsistent with traditional taxonomic groupings, though they
are still useful to define the general character and level of
organization, and aware of the fact that taxonomic opinion may
change as information accumulates, we have adopted a scheme
of classification mainly based on that of Van Den Hoek et al.
(1995). Prokaryotic members of this assemblage are grouped
into two divisions, i.e. Cyanophyta and Prochlorophyta, while
eukaryotic members are grouped into nine divisions, i.e.
Glaucophyta, Rhodophyta, Heterokontophyta, Haptophyta,
Cryptophyta, Dinophyta, Euglenophyta, Chlorarachniophyta
and Chlorophyta.
According to the most recent theories, different evolutionary
lineages can be recognized within the algal world. Three major
eukaryotic photosynthetic groups have descended from a
common prokaryotic ancestor through an endosymbiotic event.
The result is a set of nested cellular compartments one inside
the other and information about the evolutionary history of the
organism can be gleaned from the study of the membranes
surrounding these compartments, the genes that they express
and their function. The three lineages of primary plastids were
found in the Glaucophyta, in the green algae and plants, and in
the red algae. The other algal groups have acquired their
plastids via secondary (or tertiary) endosymbiosis, where a
eukaryote already equipped with plastids was preyed upon by a
second eukaryotic cell.
Photosynthetic compartments contain the pigments for
absorbing light and channeling the energy of the excited
pigment molecules into a series of photochemical and
enzymatic reactions. All those pigments are organized in
supra-molecular structures of pigment–protein complexes
embedded in the membrane of sac-like flat compressed
vesicles, the thylakoids. In prokaryotic algae the thylakoids
are free within the cytoplasm, while in eukaryotic algae they are
enclosed within bounding membranes to form the chloroplast.
The fact that algae of different divisions have different
colors, due to the presence of many different pigments in the
photosynthetic membrane system represents an important
diagnostic element at the taxonomic level of class. The pigments
present in algal cells (i.e. different type(s) of chlorophylls,
different type(s) of carotenoids and different type(s) of
phycobiliproteins) also provide a convenient paradigm to
explain evolutionary development involving endosymbiotic
acquisition of photosynthetic cellular organelles.
Algal pigmentations have taken two protective routes:
(1) U
V absorption by screening pigments or quenching of UV-generated free radicals (mainly by carotenoids).
(2) A
voidance strategy entailing growth at very low light levelsin stratified habitats. This requires accessory pigments,
mainly phycobiliproteins and the other chlorophylls, to
harvest scarce photons in support of chlorophyll a that is
central to the terrestrial photosynthesis, i.e. the processes
and absorption of light at wavelengths other than those
absorbed by chlorophyll a and transfer of energy to the
reaction centers of the two photosystems. Both strategies
entail stratification of communities so that the spatial
distribution of pigments has a cumulative effect of
ecological significance.
3. Design and performance of a microspectrophotometer
A microspectrophotometer is a modified microscope that
can measure the absorption spectrum of a very small area inside
the cell. Basic set-ups can be described according to the way of
measuring transmitted light from sample and reference points
and wavelength control devices.
3.1. Single optics system with stage-object switching
In this system the geometry of both the photometric and the
illuminating beam must remain unchanged when specimen
and reference material are interchanged. For the interchange,
the specimen and the reference material must be bodily
moved, without modifying the set-up of the microscope
(Fig. 1).
This instrument has the following advantages: it can be
conveniently attached to all types of microscope; it has equal
train and transmittance in the light train for both specimen and
reference material, and it is not very costly (Gualtieri et al.,
1989).
3.2. Single optics system with beam switching
In this system the geometry of the photometric beam in
front of the photocathode remains almost unchanged when
the measuring spot in the image field or in the stage-object
field is switched from the specimen to the reference material,
but the direction of the photometric beam is changed, moving
either the condenser (Fig. 2) or the objective (Fig. 3). The
specimen and the reference material must be sufficiently
small for them to lie together in the full image field of the
microscope; the photometric beam is switched so that the
photocathode receives light first from the specimen and then
L. Barsanti et al. / Micron 38 (2007) 197–213 199
Fig. 1. Schematic diagram of a single optics system with stage-object switching:
(a) photomultiplier, (b) eyepiece, (c) objective, (d) stage-object and (e) condenser.
from the reference material. This type of photometric system
does not allow the luminous field to be adjusted individually
and properly with respect to the photometric field, hence it
allows glare. The optical trains for the measurement of the
Fig. 2. Schematic diagram of a single optics system with beam switching
obtained by moving the condenser: (a) photomultiplier, (b) eyepiece, (c)
objective, (d) stage-object and (e) condenser.
Fig. 3. Schematic diagram of a single optics system with beam switching
obtained by moving the objective: (a) photomultiplier, (b) eyepiece, (c)
objective, (d) stage-object and (e) condenser.
specimen and the reference material are not strictly equal,
and systematic errors may occur according to the cos4-law.
This law expresses the modification of the optical flux in an
oblique light tube with respect to that in a normal light tube.
The optical flux in the oblique tube is diminished by the
factor cos4 a, where a is the angle subtended between the
axis of the microscope and the axis of the oblique tube,
because light rays entering the microscope have its axis
oblique to the axis of the microscope, and not perpendicular
to the plane of the object. Notwithstanding, this system has
the advantage of high speed measurements (Benedetti et al.,
1976).
3.3. Double optics system
In this system, in addition to the optical train of the
microscope, there is a control train which supplies the radiant
flux needed as reference. The original optical train is split off
behind the condenser and is then recombined with the control
train immediately in front of the photocathode (Fig. 4). In spite
of the stringent condition about equality of optical path and
transmittance in the two trains, this system has the advantage of
high speed in the measurement and there is no restriction in size
and position of specimen or reference material (Liebman and
Entine, 1964).
L. Barsanti et al. / Micron 38 (2007) 197–213200
Fig. 4. Schematic diagram of a double optics system: (a) photomultiplier, (b)
eyepiece, (c) mirror, (d) rotating mirror, (e) objective, (f) stage-object and (g)
condenser.
Fig. 5. Schematic diagram of a polychromator based instrumentation: (a)
grating polychromator, (b) CCD camera, (c) light-guide, (d) objective, (e)
stage-object and (f) condenser.
3.4. Reference-sample interchanging at each wavelength
This system requires the interchanging of the reference
material and the specimen in order to cover the entire visible
range (from 400 to 700 nm at 5 nm resolution); the main
drawback of this procedure is that after each repositioning the
photometric beam has to be positioned exactly in its previous
position on the specimen (Benedetti et al., 1976).
3.5. Wavelength change for sample and reference in
sequence
The interchanging of specimen and reference material is
made only once during the measurement. This involves changes
of the wavelengths from 400 to 700 nm for sample and
reference in sequence, and the procedure requires the stability
of both the lamp and the photoelectric (Gualtieri et al., 1989).
3.6. Polychromator based instrumentation
In this system a flat field imaging concave grating polychro-
mator is connected to a slit-shaped exit pupil of a light-guide
mounted onto the microscope. The polychromator dispersion
image is focused onto a digital slow scan cooled CCD camera
(Fig. 5). The images must be elaborated by specific software. The
main drawback of this procedure is the high cost of the apparatus,
which is however very fast.
3.7. A simple microspectrophotometer
A microspectrophotometer is a modified microscope with a
number of special sub-assemblies developed in house (Fig. 6).
The apparatus we used consists of a common microscope with
a 100 W tungsten-filament lamp (color temperature 3400 K)
and a stabilized 12 V/100 W power supply unit, equipped with
an holographic grating monochromator, two circular pinhole
diaphragms and a photomultiplier connected to a personal
computer. The wavelength is changed by a computer-
controlled stepping motor inserted on the axis of the
monochromator. One of the two diaphragms takes the place
of the field diaphragm, and the diameter of its image onto the
object plane is 8 mm. The second diaphragm is placed in the
plane of the real and inverted image, and its image onto the
object plane has a diameter of 0.5 mm. An oil-immersion
condenser (NA = 1.40) and a 100� oil-immersion planapo-
chromatic objective (NA = 1.25) are used for the measure-
ments. A photomultiplier is mounted on the binocular
phototube by means of a 40 mm diameter adapter ring,
beyond the measuring diaphragm and an eyepiece. The output
signal of the photomultiplier is amplified by a current amplifier
and transmitted to the AD card plugged into the computer bus.
The computer controls the position of the monochromator,
L. Barsanti et al. / Micron 38 (2007) 197–213 201
Fig. 6. Schematic diagram of a simple microspectrophotometer: (a) photo-
multiplier, (b) eyepiece, (c) diaphragm, (d) objective, (e) stage-object, (f)
condenser, (g) diaphragm, (h) holographic grating monochromator, (i) mono-
chromator slit and (j) lamp.
acquires the data, performs the calculation of optical density,
produces the graphic output and stores the data.
All the absorption spectra were recorded from 400 to
700 nm, with a step size of 0.5 nm and scan speed of
100 nm s�1. For each wavelength 10,000 values of optical
density were averaged. The resolution achieved with this step
size and wavelength width was sufficient to distinguish
between the major pigment classes present in the photosyn-
thetic compartments.
The first step of the procedure is the calibration. This is
performed by inserting a dark shutter in the optical path, and by
setting the dark-current reading of the current amplifier to zero.
A few drops of cell suspension are put onto a microscope slide
and Brownian movements are eliminated using 10% gelatine in
culture medium as a mounting medium since this solution gels
at room temperature and entraps the cells without causing
apparent changes in absorption. The specimen is also protected
from early evaporation by means of a gelatin edge-sealed
coverslip.
The absorbing structures are detected under microscope
examination, using a monochromatic light of wavelength
longer than 650 nm, or in phase contrast microscopy and the
selected structure is centred on the cross-hair of the finder.
The orientation of the structure with respect to the
electrical vector of light train has no practical importance,
except when we have to measure dichroic structures. In this
case, we can polarize the light beam, inserting a linear
polarizing filter just above the condenser. The 100� oil-
immersion planapochromatic objective (NA = 1.25) allows
the measurement of both the transmitted and the forward
scattered rays. These rays recombine in the plane of the real
image where the measuring diaphragm is located, and
therefore the scattering of the structure being measured does
not decrease the light impinging onto the photomultiplier.
Actually, the microspectrophotometer is and performs as a
confocal microscope.
Thereafter, the two diaphragms are aligned on an empty
bright field, maximizing the photocathode output value, and the
light transmitted for each wavelength is measured. Then, the
cell or the subcellular component on which the spectrum has to
be performed, is placed in the microscope field and finely
adjusted, and the light transmitted by the sample is measured.
The radiant flux impinging on the stage-object field is around
10�11 W for each measurement. This flux is regulated using the
optimal exposure value reported in the successive sections, in
order to equate the bleached pigment to the instrumental noise.
The value of the radiant flux corresponds to about 10�10 J of
energy striking the sample during the 5 s necessary to perform
the spectrum.
In the following we will describe the factors which influence
the light beam in the optical system. Knowledge of these factors
is required in order to enable the operator to control the beam in
such a way as to obtain the highest signal-to-noise ratio (SNR)
for photometric procedures (Piller, 1997).
4. Transfer of light energy in the optical system
The photoelectric sensor (phototube) transforms light
energy into an electric current. But our quantity, i.e. the
radiant flux, is really a power, i.e. energy per unit time. It is
symbolized by F and expressed in Watts (W).
Five factors determine the radiant flux impinging on the
surface of the photocathode:
(I) T
he initial power (IP) of the lamp, called spectral radiance(W cm�2 sr�1 nm�1), that is the radiant flux per unit area,
unit solid angle and unit spectral bandwidth.
(II) T
he spectral bandwidth (SB) in nanometers (nm).(III) T
he volume of the optical system (OF) through which thebeam passes, named optical flux or light tube by Zimmer
(1970). It is expressed as the ability of an optical system to
transfer light energy, but without considering loss of light
by absorption, reflection, scattering and diffraction. It is a
purely geometric quantity applied to the volume through
which the light is transferred and is expressed in cm2 sr.
The OF factor is
OF ¼ Fp NA2 (1)
L. Barsanti et al. / Micron 38 (2007) 197–213202
where F is the area of aperture diaphragm (or of the image
of the diaphragm), and NA is the effective numerical
aperture.
The maximum optical flux in the microscope is
OF ¼ ðNAobj prÞ2 (2)
where r is the radius of the maximum field area at the
ocular level (Piller, 1997).The smallest optical flux giving
a reliable measuring value is about 10�8 cm2 sr. With a
circular pinhole of 0.5 mm in the object field, the value of
the effective numerical aperture has to be at least 1.0. With
a pinhole of 10 mm and an aperture of 0.1, we can obtain a
more favourable optical flux of 10�6 cm2 sr. In a normal
microscope the OF is about 10�3 cm2 sr.
(IV) T
he total transmittance (TR) of the optical system, whichis a measure of the fraction of light that remains after loss
by absorption (monochromatising device), reflection (lens
surfaces), scattering and diffraction.
(V) T
he interactions of the light with the sample (IF)consisting in absorption, specular and diffuse reflection,
fluorescence, interference, refraction and diffraction;
these interactions obviously can influence each other.
The fundamental law of microspectrophotometry can be set out
as follows:
I ¼ IP � SB � OF � TR � IF (3)
As we have described before, measurements of absorption
microspectroscopy are based on the comparison of two radiant
fluxes density It and Ii. The first one It results from the
interaction of light with the specimen being measured (IF is
related to absorption cross-section of the molecules and the
number of absorbing molecules) (Cantor and Schimmel,
1980), and the second Ii results from the interaction of light
with the reference material. Therefore, we can consider the
absorbance of a specimen (Asp) as derived from the measures as
follows:
Asp ¼ logðIiÞ � logðItÞ (4)
This equation is known as the Lambert-Beer’s law, which is
valid for non-scattering substances, and for dilute solutions.
Knowing Asp, the path of light in the absorbing material, and the
extinction coefficient, we can calculate the concentration of a
chemical compound.
5. Geometrical features
In conventional microscopy, Abbe demonstrated how light
diffraction determines image resolution, and established the
role of numerical aperture (NA) of objective and condenser
lenses on image resolution (1/d), when the objective lenses used
are free from significant aberration, and when the field of view
is not extremely small (Born and Wolf, 1975).
d ¼ 1:22l
NAobj þ NAcond
(5)
The axial resolution, z, i.e. perpendicular to the plane of focus,
increases with the square of the NA and with the square of
lateral magnification. The distance from the centre of the
diffraction image to the first minimum along the microscope
axis turns out to be approximately twice the distance from the
centre of the diffraction image to the first minimum in the plane
of focus.
z ¼ 2l
n sin2ðaÞ(6)
The microspectroscope 3D response function constitutes an
optical probe with lateral (d) and axial (z) spatial dimension.
The illumination distribution and detection sensitivity distribu-
tion can be controlled by pinholes in the respective illumination
and detector paths. As indicated above, the optical probe for
confocal image formation is the convolution of both distribu-
tions.
The 3D shape of illuminating distribution may be controlled
by varying the effective NA of the used confocal lens. If the
illumination pinhole is set at a diameter di, the beam divergence
is equal to l/(2di). If the illumination lens has a (de)magnifica-
tion factor equal to M, Eqs. (5) and (6) can be written as
(Brakenhoff et al., 1979, 1989; Wilson, 1990):
d ¼ 1:2di
nM(7)
z ¼ 8d2i
nlM2(8)
Using the lens at its full maximum aperture u and in the case
that the detection pinhole diameter has intermediate values, the
effect of the spatial extension of the detection sensitivity
distribution depends on the detection pinhole diameter (dd) as:
d ¼ dd
M(9)
z ¼ 1:41dd
M tanðuÞ (10)
With very small detection pinholes the dimension of the
detection sensitivity function is determined by diffraction as
in Eqs. (7) and (8).
If the SNR is too low, some resolution can be sacrificed by
opening the detector pinhole to increase the signal. Then,
illumination distribution width can be increased to the detected
width, without additional loss of resolution. The probe intensity
is kept constant; hence the local irradiation load on the
specimen does not change. To maintain the statistical accuracy
of the measurements in the case the desired resolution doubles,
an irradiation dose four times greater for imaging a
bidimensional sample and a dose eight times greater in the
case of a three-dimensional sample will be necessary. Clearly
these conditions contain an inherent contradiction: in order to
achieve more quantitative temporal or spatial accuracy, more
light must pass through the sample but this will produce more
fading and cytotoxicity, which in turn will reduce the biological
reliability of the measurements. As the pinhole becomes
L. Barsanti et al. / Micron 38 (2007) 197–213 203
smaller, approaching a point detector, less light reaches the
detector, but transverse and axial resolution increase to
a maximum of 1.4 times the resolution of a wide-field
microscope.
Other microspectrophotometric instrumentations bridge
the gap between cells and molecules. At first glance, one
might wish to challenge this assertion by invoking the above
stated optical resolution limit imposed by the diffraction
phenomena. However, recent advancements in microscope
technology and data processing have extended the application
of microscopy to the study of dynamic properties of
molecules. The ‘‘ultramicroscopy effect’’ or ultraresolution
effect can be obtained by increasing the generalized Rayleigh
resolution limit without extending the spatial cutoff frequency
demanded by the diffraction (Cox and Sheppard, 1986). The
use of a very strong lateral illumination together with a dark
field condenser is the simplest way to obtain an ultraresolu-
tion effect (Passarelli et al., 1990). Moreover, by adding an
Epi-fluorescence illumination to the dark field illumination,
isolated objects such as colloidal particles as small as 20 nm
in diameter can be detected (Hirschberg et al., 1989). Among
the several possibilities, we want to emphasize the use of
Fluorescence Resonance Energy Transfer Microscopy (Jovin
and Arndt-Jovin, 1989), Fluorescence Photobleaching Recov-
ery Microscopy (Ware, 1989) and Differential Polarization
Imaging Microscopy (Finzi et al., 1989). These complex set-
ups prove advantageous because they incorporate both
sophisticated image-processing computers and confocal
microscope which greatly enhances the axial resolution of
the microscope for optical sectioning and true three-
dimensional imaging (Sheppard and Wilson, 1981). The
confocal microscope allows the selective excitation of only
that region of the specimen lying in the focal plane of the
objective, i.e. the specimen is not excited in its whole
thickness.
6. Theory of detection of photosensitive substances
In order to identify a substance by its light absorption
characteristics, we should be able to determine that the
difference between the average number of photon transmitted
by the sample and the average number of photon incident on the
sample (the signal) exceeds the random fluctuation (noise) by a
sufficient amount in the combined measuring and measured
system. The noise may arise from dark noise and photoelectron
shot noise in the photomultiplier (see the next section), from
amplifier or electrical circuit noise, from mechanical vibration,
and from biological motion due to Brownian movements or to
convection currents in the preparation. The last of these is easily
eliminated by adding gelatine to the culture medium for the
slide preparation; as already stated, the solution gels at room
temperature and entraps the cells without causing apparent
changes in the absorption properties of the cells. For the light
intensities used in Microspectrophotometry, the relative
contribution of dark current, i.e. the irreducible intrinsic limit
of any photoelectric measurement, is larger than that of other
possible noise sources.
The mean-square photoelectron noise or shot noise current,
In, at the level of the cathode of a photomultiplier is given by
(Liebman, 1972).
I2n ¼
e2kIiR
t(11)
where e is the charge of the electron, k the ratio of photoelec-
trons generated to photons incident at the cathode surface, Ii the
incident photon flux density at the object plane, R the area of
object illuminated and t is the detection time.
The signal to be detected is Is = Ii � It where Ii is the radiant
flux incident on the sample and It the flux density after
attenuation by the sample. Since the fraction absorbed by the
sample is A = 1 � T, where T is the fraction transmitted, it
follows that the signal is also Is = Ii � It = AIi.
This produces an average photoelectric signal current
difference Is, given by
Is ¼ ekAI iR (12)
The total mean-square shot noise involved in a measurement is
contributed essentially by the incident beam, if the absorption is
small (20% or less). We can calculate the root-mean-square of
signal-to-noise ratio of a shot-noise-limited photometer using
the determined values of both signal and noise. Therefore,
SNR ¼ Is
In
¼ AðkIitRÞ1=2(13)
Detectivity is quantitatively proportional to the signal-to-noise
ratio. The chances of detecting a substance that absorbs a
fraction A of the incident light are improved by selecting a
detector with a high quantum efficiency, k; and by using a high
flux density, Ii, and a long exposure time t; and by collecting
light from a larger area R, e.g. by using a larger sample. When
we attempt to apply these principles to the detection of
photoreceptor pigment in a single cell, however, several diffi-
culties arise. Both path-length and concentration of photore-
ceptor pigments in single cells are small and unalterable.
Again, if the light exposure E (Iit) is increased by increasing
either Ii or t, bleaching of the photoreceptor pigment could
occur. Moreover, the rate of pigment bleaching increases in
direct proportion to the exposure, whereas detectivity is
improved only as a square root of the exposure time. When
bleaching due to the measuring light alters a significant frac-
tion of the molecules originally present, distortion of the
recorded absorption spectrum may be expected and this can
occur in two ways. If the bleached molecules are quickly
replaced by product molecules that absorb little or no visible
light, the effect on the recorded curve will be to shift its
absorption maximum toward that end of the spectrum from
which the scan originates (Marks, 1965). If the bleached
molecules form colored photoproducts whose spectra differ
from those of the unbleached pigments, another kind of dis-
tortion occurs, which is more difficult to correct. Spectra of the
products as well as the kinetics constant of their decay are
required. Therefore, in order to obtain good spectral resolution
and detectivity for weakly-absorbing cells we need to use high
light exposure to reduce the relative noise fluctuations, but the
L. Barsanti et al. / Micron 38 (2007) 197–213204
more light we use, the more pigment is bleached with increas-
ing danger of spectral distortion. Thus, a suitable compromise
between detectivity and bleaching should be found.
For small densities, the bleaching equation is (Liebman,
1972)
Pt
P0
¼ expð�IiteaÞ (14)
where P0 and Pt are respectively the amount (mol/cm2) of
pigment initially present and the amount present after a time, t,
of light exposure Ii that bleaches with an efficiency of e mmol of
pigment lost per mmol of photons absorbed; and a is the
naperian (logarithm to the base e) extinction coefficient.
At the end of the spectral scan, most of the pigment
originally present should be still present so that
Pt
P0
¼ expð�IiteaÞ � 1 (15)
This result is approached as exp(�Iitea) approaches zero.
The exponential function exp(�Iitea) can be expanded and
considered equal to (1 � Iitea) to a first approximation, that is
expð�IiteaÞ � 1� Iitea (16)
However, if Pb is the amount of pigment bleached, we can write
the following equation:
Pt þ Pb ¼ P0!Pt
P0
þ Pb
P0
¼ 1! Pb
P0
¼ 1� Pt
P0
(17)
By combining Eq. (17) with both Eqs. (15) and (16), we
have:
Pb
P0
¼ 1� Pt
P0
¼ Iitea (18)
When Pb is a small quantity, which is the experimentally
desirable condition, the approximation expressed in formula
(16), gives the exposure value (E = Iit) that must not be
exceeded for a given majority of pigment molecules to remain
unbleached.
Obviously, the optimum value of the exposure E is that
which equates the pigment bleached to the root-mean-square
photoelectron shot noise.
Thus, equating Eq. (13) with Eq. (18), and solving for the
exposure E we obtain this optimum value:
AðkIitRÞ1=2 ¼ P0
Pb
¼ ðIiteaÞ�1(19)
E ¼ ðA2a2e2kRÞ�1=3(20)
Therefore, E must be adjusted to be in inverse proportion to the
two-thirds power of the fraction A of the light absorbed by that
pigment. A pigment absorbing 10% of the incident light can
thus be recorded at a signal-to-noise ratio only 4.6 times higher
than that of a pigment absorbing 1% when the bleaching is
made equal to the noise.
7. Statistic of photon flux detection
With an ideal device, detection would be limited only by
the noise or background of the radiant source, and no
additional noise should be added by the detector itself or
subsequent amplification and conversion electronics. It is
possible to design ideal amplification and conversion stages,
but the detector scheme is still limited by the noise of the
detector stage itself. Internal detection with semiconductor
devices (photoconductive and photovoltaic) is associated with
two general classes of noise. The first class arises from
thermal motion of the charges within a material (Johnson
noise); the second class of noise is specific of the type of
device. In photoconductors, it is the generation–recombina-
tion noise, due to fluctuation in the rate of thermal generation
and recombination of charge carriers, which in turn generates
a fluctuation in the average carrier concentration, and
therefore a variation in average resistance. In diodes it is
the shot noise, due to the quantum nature of current carrier,
which generates a statistical variation of the amplitude of the
measured current. The final form of noise is the 1/f noise, so
named because the power varies more or less inversely with
frequency. In order to evaluate the SNR achievable by a
device, for example a photomultiplier, the behavior of such a
detector has to be analyzed in terms of SNR degradation
(Robben, 1971).
The detection process is limited by the SNR of the photon
flux. If the average rate of photon emission is Ip, then during the
observation interval t the average number of photons observed
is:
np ¼ Ipt (21)
the variance is given by
s2p ¼ Ipt ¼ np (22)
and the SNR is given by
SNR ¼ np
sp
¼ n1=2p (23)
For any photon incident on the photocathode, the quantum
efficiency, h, is the product of the probability that the photon is
absorbed and the probability that this photon produces a free
electron, and the probability that this free electron reaches the
surface of the material. Since the probability of releasing one
photoelectron per incident photon is related to the quantum
efficiency, we have
npe ¼ hIpt (24)
while the variance is given by
s2pe ¼ npe (25)
As a consequence, the SNR at the stage of emission of photo-
electrons from the photocathode is
SNRpe ¼ n1=2pe (26)
L. Barsanti et al. / Micron 38 (2007) 197–213 205
Fig. 7. Decomposition of chlorophyll a absorption spectrum obtained using
Gaussian (a), Lorenzian (b) and mixed Gaussian–Lorentians (c) profiles.
During any recording period, thermoionic electrons will be
emitted, and the average of these emissions and their variance
are related
s2pe ¼ nd (27)
Because the photoevent and the thermal event are independent,
the square values of their variance add,
s2pe ¼ nd þ npe (28)
and when dark noise is present, the SNR of a photomultiplier is
given by
SNRpe ¼npe
ðnpe þ ndÞ1=2(29)
8. Mathematical computation of absorption spectra
Absorbance attributable to accessory pigments is often
difficult to quantify. For this reason, the ability to discriminate
among distinct phylogenetic groups and potential species with
different thylakoidal pigment compositions will depend upon
the robustness of the technique chosen to differentiate the
diverse components within the portion of the spectrum where
carotenoids, phycobiliproteins and other chlorophylls absorb.
The relative contribution of the individual pigment classes to
absorption spectra can be estimated by curve-fitting routine
using templates of the absorption spectra of the pigments found
in the different divisions of algae, namely chlorophylls a, b and
c; cytochromes; carotenoids; and phycobiliproteins. Resolu-
tions of the different forms of chlorophyll c pigments that differ
only of 1 or 2 nm are not achievable with this analytical method.
The generation of templates of pigment absorption spectra is
based on the representation of the spectral dependence of the
pigment absorption in terms of a small number of symmetric
Gaussian bands (these bands are not attributable to a single
bond in the complex pigment molecules, but express the bulk
properties of the molecule) as in the following equation:
pigmentðlÞ ¼XL
j¼1
a j exp
�� ðl� l jÞ2
2s2j
�(30)
where aj is the specific absorption coefficient for pigment in the
jth Gaussian band, and s2j and lj are the widths and centers of
the corresponding band, and pigment(l) gives the absorption as
a function of wavelength.
The spectral decomposition could be equally performed
using Lorentzian or Voigt profiles (convolution of Lorentzian
and Gaussian bands).
Fig. 7a–c shows the decomposition of chlorophyll a
obtained using Gaussian, Lorenzian and mixed Gaussian–
Lorentians profiles; if the same number of bands and the same
position of the centers of the curves are used for each
decomposition, we can reconstruct the chlorophyll a spectra
without significative differences.
The number of Gaussian bands and their maxima can be
determined using the peaks of the fourth-derivative curve of
the absorption spectrum of extracted pigments. Fourth
derivative resolves the position of the absorbance maxima
attributable to pigment absorption bands (Bidigare et al., 1989;
Smith and Alberte, 1994). Fourth-derivative curve was adopted
because it shows maxima which could not be resolved by the
second-derivative curve. Since microspectroscopy has a high
L. Barsanti et al. / Micron 38 (2007) 197–213206
signal-to-noise ratio the derivative operation does not produce
artefacts, i.e. the maxima resulting from derivation do
correspond to pigment absorption maxima. The widths of
the Gaussian bands, and their specific absorption coefficient aj
can be estimated on the basis of both the spectra of extracted
pigments and/or data present in the literature.
The Gaussian bands were fed into a spreadsheet. Band
parameters were varied automatically to find the smallest
difference between the pigment absorption obtained by Eq. (30)
[pigment(l)] and the absorbance of the pigment measured at
that wavelength; the resulting templates matched the absorption
recording without significant difference ( p < 0.01).
The pigment spectra templates can be then combined
together; the model that describes the spectrum for each
photosynthetic compartment is:
spectrumðlÞ ¼Xn
i¼1
Ci
XL
j¼1
ai j exp
�� ðl� li jÞ2
2s2i j
�(31)
where Ci is the concentration of the ith pigment, aij the specific
absorption coefficient for the ith pigment in the jth Gaussian
band, and lij are the widths of the corresponding band.
The centers s2i j of these Gaussian bands can be screened for
their correspondence with the peaks present in the fourth-
derivative curves of the absorption spectrum measured on the
photosynthetic compartments and shifted in the presumptive
real position, since the positions of the pigment absorption
bands in the proteic moieties are different from those in the
solvent solution.
The real centers of the bands of chlorophyll a (the most
important absorbing pigment) were determined using the
absorption spectrum of algae that contain only this chlorophyll
as main pigment. The peaks present in the red region and not
belonging to chlorophyll a were assigned to chlorophyll b or c
depending upon the algae where the determination was
straightforward.
The templates of the different pigment components were fed
into a spreadsheet, then summed and compared with the in vivo
absorption spectrum. Template concentrations were automa-
tically modulated to find the smallest difference between the
spectrum(l) and the in vivo absorbance at that wavelength. The
decomposition obtained by this procedure matched the in vivo
recording without significant difference ( p < 0.01).
9. Results of microspectrophotometric investigations on
photosynthetic compartments
In the following we will describe in detail the different
absorption spectra of the algae measured with the microspec-
trophotometric set-up described above.
Figs. 8–10 show the algae used for the measurements, the
color of each culture, the relative spectrum, and the main
pigments present in the cell and revealed by the decomposition
procedure.
The spectrum recorded on the trichomes of prokaryote
Leptolyngbya (Cyanophyta) shows that the main pigments of
this cyanobacterium are chlorophyll a, b-carotene, and the
phycobiliproteins phycoeritrocyanin, c-phycocyanin and allo-
phycocyanin (Fig. 8).
In the case of Prochlorotrix hollandica (Prochlorophyta), the
other prokaryotic alga, chlorophyll a is supported by
chlorophyll b, and the main carotenoid is zeaxanthin.
Phycobiliproteins are absent (Fig. 8).
The spectral patterns of both Cyanophora paradoxa
(Glaucophyta) (Fig. 8) and Porphyridium cruentum (Rhodo-
phyta) (Fig. 8) differ in that only chlorophyll a is present,
together with carotenoids, and phycobiliproteins. In the case of
Cyanophora, zeaxanthin is the main carotenoid, and the
phycobiliproteins are represented by c-phycocyanin and
allophycocyanin. In Porphyridium, lutein is the main carote-
noid, and r-phycocyanin, b-phycoerythrin and allophycocyanin
are the phycobiliproteins present.
In Ochromonas danica (Chrysophyceae) (Fig. 8) and
Phaeodactylum tricornutum (Bacillariophyceae) (Fig. 9) both
belonging to the division of Heterokontophyta, and in Pavlova
lutherii (Haptophyceae) (Fig. 9) belonging to the division of
Haptophyta, both chlorophylls a and c are present together with
fucoxanthin, which is the main carotenoid in all the three
species. In Nannochloropsis sp. (Eustigmatophyceae, Hetero-
kontophyta) (Fig. 9) only chlorophyll a and violaxantin are
present.
In Cryptomonas ovata (Cryptophyta) (Fig. 9) both
chlorophylls a and c are present, and the main carotenoid is
alloxanthin. Among the phycobiliproteins, phycoerithrin is the
only one detected, with the absorption maximum at 545 nm.
Phacus triqueter (Euglenophyta) (Fig. 9), Gymnochlora
stellata (Chlorarachniophyta) (Fig. 10) and Dunaliella salina
(Chlorophyta) (Fig. 10) share both chlorophylls a and b, and
differ in the main carotenoid, which is diadinoxanthin in
Phacus, neoxanthin in Gymnochlora and lutein in Dunaliella.
A very similar spectral pattern is present in the lawn daisy
Bellis perennis L., characterized by chlorophylls a and b, with
b-carotene as main carotenoid (Fig. 10).
The spectral patterns of three different dinoflagellates were
measured. Two of them share both chlorophylls a and c but
differ in the main carotenoid, which is peridinin in
Prorocentrum micans (Fig. 10), and fucoxanthin in the
undetermined sample (Fig. 10).
We measured the absorption spectra in algae cultivated
under laboratory conditions, since we considered the growth
conditions we adopted for our cultures as standard for
laboratory routines and small laboratory collections. Hence,
the pigment compositions we measured were interpreted as
typical for all members in individual algal classes, and the
identified pigments were taken as class-specific marker
pigments. Therefore, these pigment signatures can be used
as reference for identification purposes. In natural ecosystems,
factors such as light quality and quantity, nutrient availability,
salinity and temperature variations, and growth phase may
affect pigmentation dynamics. The total irradiance and
spectral quality of the available light undergo more or less
predictable variations; diurnal and seasonal cycles, atmo-
spheric and/or more localized shading phenomena can change
the flux of photosynthetically active radiation, while shading
L. Barsanti et al. / Micron 38 (2007) 197–213 207
Fig. 8. Results of microspectrophotometric measurement on photosynthetic compartments; each series shows the sample used, the color of the culture on filter, the
absorption spectrum, and the main pigments after spectrum band decomposition.
L. Barsanti et al. / Micron 38 (2007) 197–213208
Fig. 9. Results of microspectrophotometric measurement on photosynthetic compartments; each series shows the sample used, the color of the culture on filter, the
absorption spectrum, and the main pigments after spectrum band decomposition.
L. Barsanti et al. / Micron 38 (2007) 197–213 209
Fig. 10. Results of microspectrophotometric measurement on photosynthetic compartments; each series shows the sample used, the color of the culture on filter, the
absorption spectrum, and the main pigments after spectrum band decomposition; in the case of the lawn daisy, the image of the plant substitutes for the image of the
filter showing the color of the culture.
L. Barsanti et al. / Micron 38 (2007) 197–213210
Fig. 11. Comparison between absorption spectra obtained by different methods on diatoms belonging to different genera (a) and on green algae belonging to the same
genus (b); refer to the text for details. (€) Spectrophotometry on cell suspension on filter; (|) spectrophotometry on cell suspension in cuvette; (k)
microspectrophotometry on cells on slide; (^) microspectrophotometry on cells on filter.
by other algae, and selective absorption by various materials
suspended or dissolved in the water column can alter the
spectral distribution of the irradiance. Both these phenomena
are known to influence the photosynthetic machinery of algal
Fig. 12. Comparison between absorption spectra performed on algae belonging to dif
and Reynaud, 1977).
cells (Cunningham et al., 1989; Marquardt and Hanelt, 2004).
On the basis of these assumptions, now we will compare the
other spectra measured in the natural external conditions or
with other microspectrophotometers.
ferent genera of cyanobacteria (a–c) and red algae (d) (Hewes et al., 1998; Roger
L. Barsanti et al. / Micron 38 (2007) 197–213 211
Fig. 13. Absorption spectra of algae grown under different environmental conditions and habitats, performed by different methods, and present in the literature;
reference is reported inside the frame (Hewes et al., 1998; Roger and Reynaud, 1977).
L. Barsanti et al. / Micron 38 (2007) 197–213212
10. Comparison of microspectrophotometric data with
other spectroscopic investigations on photosynthetic
compartments
In the following we will describe the different absorption
spectra of the algal measured in different environmental
conditions and with different spectrophotometric instruments
as they are reported in the literature.
The spectra were recorded using commercial spectro-
photometers very often equipped with integrating sphere on cell
suspension in cuvettes (Canto de Loura et al., 1987; Enriquez
et al., 2005; Gerloff-Elias et al., 2005; http://www.ioccg.org/
. . .; Janssen et al., 2001; Johnsen et al., 1994, 1999; Razi Naqvi
et al., 2004; Yoshii et al., 2002) on suspension filtered onto glass
fibre filters (Astoreca et al., 2005; http://www.ioccg.org/. . .;Lutz et al., 2001; Millie et al., 2002; Payri et al., 2001; Stæhr
and Cullen, 2003; Ting et al., 2002), on cell suspension dried to
give films of 0.5 mm (Erokhina et al., 2002), or using different
microspectrophotometer set-ups (Nultsch et al., 1983; Barsanti
et al., 1989, 1990; Graham, 2000; Albertano, 1991; Evangelista
et al., 2006).
Fig. 11a and b shows the comparison between different
methods. All the different experimental methods were tested on
diatoms (Bacillariophyceae) belonging to different genera
(Fig. 11a); though the spectra are comparable, still the
measurement performed by microspectrophotometry results
in a more resolved spectrum (in term of number of detected
pigment bands). Fig. 11b shows different methods applied to
the same species in the same environmental conditions;
therefore the differences between spectra must be due only
to the methods; also in this case the more resolved and detailed
spectrum has been obtained by microspectroscopy.
Fig. 12 shows spectra performed on algae belonging to
different genera of cyanobacteria (Fig. 12a–c) and red algae
(Fig. 12d), containing accessory pigments organized in
phycobilisomes. The spectra look very different, but the
difference between them is due to the environmental conditions
in which these algae dwell, supporting the hypothesis that
environmental parameters influence pigment distribution and
content.
Fig. 13 shows other absorption spectra present in the
literature, which have been performed by different methods, on
algae grown under different environmental conditions and
habitats. To extrapolate the specific experimental and environ-
mental parameters relative to each spectrum, the reader can
refer to corresponding paper cited under the references section.
11. Conclusions
Absorbance attributable to accessory photopigments often is
difficult to quantify and routinely discern for great number of
these. For that reason, the ability to discriminate among distinct
phylogenetic groups (and potential species) will be dependent
upon the effectiveness of the techniques chosen to finding out
‘‘real’’ Gaussian bands and their successive groupings.
Fourth-derivative spectra transformation of the microspec-
trophotometric absorbance spectra and the successive Gaussian
bands decomposition and their fitting into the spectral envelope
of pigment templates provides a realistic discrimination
between the different algal divisions.
For each division sensu Van Den Hoek et al. (1995) the
contribution and the spectral position of each pigment category
were well defined using the computation algorithms herein
described (see Section 8). Even taking into account the
inevitable approximations we made in the template reconstruc-
tions, the list of spectra presented here is the complete
description of pigment distribution in the thylakoid compart-
ments of the different algal divisions grown laboratory
condition (Figs. 8–10), and the microspectrophotometric
determination results the most detailed and accurate in terms
of detected pigments bands (Fig. 11a and b).
Therefore, these spectra can be used as a reference for the
comparison of the photosynthetic pigment present in the algae
growing under different conditions and diverse natural habitats,
which as shown in Figs. 12 and 13 clearly influence pigment
distributions and content.
References
Albertano, P., 1991. Effects of monochromatic light on four species of Lepto-
lyngbya. Arch. Hydrobiol. Algol. Stud. 64, 199–214.
Astoreca, R., Rousseau, V., Ruddick, K., Van Mol, B., Parent, J.Y., Lancelot, C.,
2005. Optical properties of algal blooms in an eutrophicated coastal area
and its relevance to remote sensing. http://www.mumm.ac.be/OceanColour/
Publications/Astoreca-SPIE2005.pdf.
Barsanti, L., Passarelli, V., Gualtieri, P., 1989. Spectra of euglena gracilis and
haematococcus pluvialis pigmented subcellular structures by in situ micro-
spectroscopy. Med. Biol. Environ. Int. J. 17 (2), 847–851.
Barsanti, L., Passarelli, V., Gualtieri, P., 1990. Pigment distribution in cyano-
bacteria: an in vivo microspectroscopy investigation. Experentia 46, 255–
256.
Benedetti, P.A., Bianchini, G., Chiti, G., 1976. Fast scanning microspectro-
scopy: an electrodynamic moving-condenser method. Appl. Opt. 15, 2554–
2558.
Bidigare, R.R., Morrow, J.H., Kiefer, D.A., 1989. Derivative analysis of spectral
absorption by photosynthetic pigments in the western Sargasso Sea. J. Mar.
Res. 47, 323–341.
Born, M., Wolf, E., 1975. Principles of Optics. Pergamon Press, Oxford.
Brakenhoff, G.J., Blom, P.J., Barends, P., 1979. Confocal scanning light
microscopy with high aperture immersion lenses. J. Microsc. 117, 219–
232.
Brakenhoff, G.J., Visscher, K., Van der Voort, H.T.M., 1989. Size and shape of
the confocal spot: control and relation to 3D imaging and image processing.
In: Pawley, J.B. (Ed.), Handbook of Biological Confocal Microscopy.
Plenum Press, New York, pp. 87–91.
Cunningham Jr., F.X., Dennenberg, R.J., Mustardy, L., Jursinic, P.A., Gantt, E.,
1989. Stoichiometry of Photosystem I, Photosystem II, and phycobilisomes
in the red alga Porphyridium cruentum as a function of growth irradiance.
Plant Physiol. 91, 1179–1187.
Canto de Loura, I., Dubacq, J.P., Thomas, J.C., 1987. The effects of nitrogen
deficiency on pigments and lipids of cyanobacteria. Plant Physiol. 83, 838–
843.
Cantor, C.R., Schimmel, P.R., 1980. Biophysical Chemistry, Part II. Freeman,
New York, pp. 394–405.
Cox, I., Sheppard, C.J.R., 1986. Information capacity and resolution in a optical
system. J. Opt. Soc. Am. A 3, 1152–1158.
Enriquez, S., Mendez, E.R., Igliesas-Prieto, R., 2005. Multiple scattering on
coral skeletons enhances light absorption by symbiotic algae. Limnol.
Oceanogr. 50 (4), 1025–1032.
Erokhina, L.G., Shatilovich, A.V., Kaminskaya, O.P., Gilichinskii, D.A., 2002.
The absorption and fluorescence spectra of the cyanobacterial phycobionts
L. Barsanti et al. / Micron 38 (2007) 197–213 213
of cryptoendolithic lichens in the high-polar regions of Antarctica. Micro-
biology 71 (5), 601–607.
Evangelista, V., Frassanito, A.M., Passarelli, V., Barsanti, L., Gualtieri, P., 2006.
Microspectroscopy of the photosynthetic compartment of algae. Photo-
chem. Photobiol. 82, 1039–1046.
Finzi, L., Bustamante, C., Garab, G., Juang, C., 1989. Direct observation of
large chiral domains in chloroplast thylakoid membranes by differential
polarization microscopy. Proc. Natl. Acad. Sci. U.S.A. 86, 8748–8752.
Gerloff-Elias, A., Spijkerman, E., Schubert, H., 2005. Light acclimation of
Chlamydomonas acidophila accumulating in the hypolimnion of an acidic
lake (pH 2.6). Freshwater Biol. 50, 1301–1314.
Graham, M.H., 2000. Planktonic patterns and processes in the giant kelp
Macrocystis pyrifera. A dissertation submitted in partial satisfaction for
the degree Doctor of Philosophy in Oceanography by Michael Hall Graham.
http://phycology.mlml.calstate.edu/PDFs/Graham00.pdf.
Gualtieri, P., Passarelli, V., Barsanti, L., 1989. A simple instrument to perform
in vivo absorption spectra of pigmented cellular organelles. Micron
Microsc. Acta 20, 107–110.
Hewes, C.D., Mitchell, B.G., Moisan, T.A., Vernet, M., Reid, F.M.H., 1998. The
phycobilin signatures of chloroplasta from three dinoflagellate species: a
microanlytical study of Dinophysis caudata, D. fortii and D. acuminata
(Dinophysiales, Dinophyceae). J. Phycol. 34, 945–951.
Hirschberg, J.G., Kohen, E., Kohen, C., 1989. High resolution fluorescence and
phase microscopy in conjunction with micromanipulation for in situ study
of metabolism in living cells. In: Kohen, E., Hirshberg, J.G. (Eds.), Cell
Structure and Function by Microspectrofluorometry. Academic Press, New
York, pp. 87–98.
Janssen, M., Bathke, L., Marquardt, J., Krumbein, W.E., Rhiel, E., 2001.
Changes in photosynthetic apparatus of diatoms in response to low and
high light intensities. Int. Microbiol. 4, 27–33.
Johnsen, G., Nelson, N.B., Jovine, R.V.M., Prezelin, B.B., 1994. Chromopro-
tein- and pigment-dependent modeling of spectral light absorption in two
dinoflagellates, Prorocentrum minimun and Heterocapsa pygmaea. Mar.
Ecol. Progr. Ser. 114, 245–258.
Johnsen, G., Dalløkken, R., Eikrem, W., Legrand, C., Aure, J., Skjoldsal, H.R.,
1999. Eco-physiology, bio-optics and toxicity of the ichthyotoxic Chryso-
chromulina leadbeateri (Prymnesiophyceae). J. Phycol. 35, 1465–1476.
Jovin, T.M., Arndt-Jovin, D.J., 1989. Luminescence digital imaging micro-
scopy. Annu. Rev. Biophys. Biophys. Chem. 18, 271–308.
Liebman, P.A., Entine, G., 1964. Sensitive low-light level microspectrophot-
ometer detection of photosensitive pigments of retinal cones. J. Opt. Soc.
Am. 54, 1451–1459.
Liebman, P.A., 1972. Microspectrophotometry of photoreceptors. In: Dartnall,
H.J.A. (Ed.), Handbook of Sensory Physiology, vol. 7. Springer-Verlag,
Berlin.
Lutz, V.A., Sathyendaranath, S., Head, E.J.H., Li, W.K.W., 2001. Changes in
vivo absorption and fluorescence excitation spectra with growth irradiance
in three species of phytoplankton. J. Plankton Res. 23 (6), 555–569.
Marks, W.B., 1965. Visual pigments of single goldfish cone. J. Physiol. 178, 14–
32.
Marquardt, J., Hanelt, D., 2004. Carotenoid composition of Delesseria lanci-
folia and other marine red algae from polar and temperate habitats. Eur. J.
Phycol. 39, 285–292.
Millie, D.F., Schofield, O.M.E., Kirkpatrick, G.J., Johnsen, G., Evens, T.J.,
2002. Using absorbance and fluorescence spectra to discriminate micro-
algae. Eur. J. Phycol. 37, 313–322.
Nultsch, W., Benedetti, P.A., Gualtieri, P., 1983. Microspectrophotometric
investigation of photobleaching in Anabaena variabilis cell and heterocysts
and its prevention by sodium azide. Z. Pflanzen Physiol. Bd. 111, 327–332.
Passarelli, V., Barsanti, L., Gualtieri, P., 1990. A simple device for ultramicro-
scopy study of motile microorganisms. Micron Microsc. Acta 21, 101–104.
Payri, C.E., Maritorena, S., Bizeau, C., Rodiere, M., 2001. Photoacclimation in
the tropical coralline alga Hydrolithon onkodes (Rhodophyta, Corallina-
ceae) from a French Polynesian reef. J. Phycol. 37, 223–234.
Piller, H., 1997. Microscope Photometry. Springer-Verlag, Berlin.
Razi Naqvi, K., Merzlyakb, M.N., Melo, T.B., 2004. Absorption and scattering
of light by suspensions of cells and subcellular particles: an analysis in terms
of Kramers–Kronig relations. Photochem. Photobiol. Sci. 3, 132–137.
Robben, F., 1971. Noise in the measurement of light with photomultipliers.
Appl. Opt. 10 (4), 776–796.
Roger, P.A., Reynaud, P.A., 1977. Correction de la diffusion pour l’etablisse-
ment de spectres d’absorption par des cultures d’algues microscopiques.
Cah. ORSTOM, ser. Biol. XII (2), 129–131.
Sheppard, C.J.R., Wilson, T., 1981. The theory of the direct-view confocal
microscope. J. Microsc. 124, 107–117.
Smith, C.M., Alberte, R.S., 1994. Characterization of in vivo absorption
features of chlorophyte, phaeophyte and rhodophyte algal species. Mar.
Biol. 118, 511–521.
Stæhr, P.A., Cullen, J.J., 2003. Detection of Karenia mikimotoi by spectral
absorption signatures. J. Plankton Res. 25 (10), 1237–1249.
Ting, C.S., Rocap, G., King, J., Chisholm, S.W., 2002. Cyanobacterial photo-
synthesis in the oceans: the origins and significance of divergent light-
harvesting strategies. Trends Microbiol. 10 (3), 134–142.
Van Den Hoek, C., Mann, D.G., Jahns, H.M., 1995. Algae—An Introduction to
Phycology. Cambridge University Press.
Ware, B.R., 1989. Fluorescence photochemical techniques for the study of
transport in cytoplasm and cytoplasm models. In: Kohen, E., Hirshberg,
J.G. (Eds.), Cell Structure and Function by Microspectrofluorometry. Aca-
demic Press, New York, pp. 145–167.
Wilson, T., 1990. The role of detector geometry in confocal imaging. J. Microsc.
158 (2), 133–144.
Yoshii, Y., Takaichi, S., Maoka, T., Hanada, S., Inouye, I., 2002. Characteriza-
tion of two unique carotenoid fatty acid exters from Pterosperma crisatum
(Prasinphyceae, Chlorophyta). J. Phycol. 38, 297–303.
Zimmer, H.G., 1970. Geometrical Optic, Springer-Verlag, Berlin. http://
www.ioccg.org/training/turkey/DrClementson_lect.pdf.