development of a dual-mode laminar flow ion source for ... · volume inlet is based on the gc-appi...

1
sample from GC make-up gas inlet to MS radiation source GC-transfer line Kai Kroll ; Walter Wissdorf; Hendrik Kersten; Thorsten Benter Physical & Theoretical Chemistry Wuppertal, Germany Institute for Pure and Applied Mass Spectrometry Experimental Setup Introduction Development of a dual-mode laminar flow ion source for APPI- and APLI-GC-MS Methods Conclusions Acknowledgement Literature 24/7 heating of all setups at 300 °C for several weeks without loss of performance GC measurements reveal a high level of linearity in the pg on-column range Transfer of the analytes into the MS without significant loss of separation performance Peak widths down to 0.7 s (FWHM) Simple variation of geometric parameters Ion source is interfaced to MS control software APLI Mode: As expected, a larger inner diameter and a longer dwell time leads to a better ionization efficiency and gain in sensitivity Limit of detection for PAHs down to the fg range APPI Mode: Setup 1 shows a significant decrease in sensitivity compared to setup 2. The longer tube length presumably leads to more pronounced ion-wall losses Gain of sensitivity with longer dwell times in the conical ionization inlet (vortex) The simulations show a longer residence time in the cone with 5 mm tube diameter. These results are in accordance with the measurements A GC column position slightly off axis leads to an increased ionization efficiency Table 1: Overview of the different setup modifications; ion dwell times calculated from make-up gas volume flow Setup Length [mm] Inner diameter [mm] Ion dwell time [ms] 1 150 5 240 2 85 5 140 3 150 2.5 70 4 85 2.5 50 APLI APPI Peak width (FWHM) [s] Limit of Detection* [pg/μL] Slope Linear range** [pg/μL] Setup Linear range** [pg/μL] Slope Limit of Detection* [pg/μL] Peak width (FWHM) [s] 0.8 – 1 0.6 34119 0.8 – 100 1 10 – 100 113 8 1 – 1.3 0.7 – 1.1 6 35 10 – 100 2 10 – 100 265 6 0.8 – 1.1 0.8 – 1.3 3 2797 6 – 600 3 100 – 600 16 60 0.8 – 1.3 0.8 – 1.3 40 65 60 – 600 4 100 – 600 12 80 0.7 – 1.3 Figure 1a: The custom dual-mode laminar flow ion source setup S/N = 3 ** for a dilution series of Naphthalene in n-hexane Table 2: Analytical performance of the four laminar flow ion source versions for APLI- and APPI-GC-MS. Performance Crucial design aspects Tightly sealed ionization volume Chemically and photo-physically inert matrix gas Conical ionization volume inlet Asymmetrical make-up gas inlet Vortex inside ionization volume for higher ionization efficiency Injection of the GC flow into the major make-up gas flow (The concept of the ionization volume inlet is based on the GC-APPI source presented by Kersten et al. in 2014 [3]) Use of non outgassing materials Controlled operation temperature up to 300 °C Figure 3e: Determination of the linear ranges of the four versions of the laminar flow ion source setup, APPI, for naphthalene (m/z 128) Figure 3a: Linear ranges for the four modifications of the laminar flow ion source setup in APLI mode for naphthalene (m/z 128). Figure 3b: Dilution series of naphthalene 0.8 – 1000 pg/μL (m/z 128). Setup 1, APLI Figure 3d: Dilution series of naphthalene 0.8 – 1000 pg/μL (m/z 128). Setup 1, APPI Figure 3c: Chromatogram of the EPA 8270 LCS Mix 1 (100 pg/μL). Ion source setup 3, APLI mode. Figure 3f: Chromatogram of the EPA 8270 LCS Mix 1 (100 pg/μL). Setup 1, APPI mode. Note: The pronounced signal deviation around 200 pg/μL is observed in all four setups and for both ionization methods. The reasons for this behavior are yet unknown and subject to current investigations. CFD Simulations Figure 2a: CFD simulation of the ionization volume inlet. Make-up gas flow 760 mL/min; GC column flow 2 mL/min. The red flow lines indicate the make up gas flow, the yellow flow lines indicate the GC column effluent make-up gas make-up gas make-up gas sample from GC sample from GC sample from GC sample from GC make-up gas Figure 2b: Variation of GC column position for setup 1, APPI. 100 pg/μL naphthalene GC column on-axis GC column 1 mm off-axis At the annual ASMS meeting in 2010, we introduced a novel laminar flow ion source (LFIS) for API-MS [1]. In the following years, we presented the capabilities of the LFIS with the analysis of in-situ samples from atmospheric smog chamber experiments [2]. Based on the original LFIS, we designed a novel interface for GC-API-MS. Challenges: Design of a novel GC-MS interface, suitable for atmospheric pressure photoionization (APPI) and atmospheric pressure laser ionization (APLI) Design with sustained chromatographic fidelity → Prevention of cold spots → Prevention of dead volumes Swift change between APPI and APLI operation Variable ionization chamber volume (length and inner diameter) to optimize for the ionization efficiency and ion transport in APPI and APLI Controllable via MS-software iGenTraX UG, Haan, Germany Bruker Daltonics GmbH, Bremen, Germany 5 mm tube 2.5 mm tube [1] Barnes, I.; Kersten, H.; Wissdorf, W.; Pöhler, T.; Hönen, H.; Klee, S.; Brockmann, K. J.; Benter, T., Novel laminar flow ion sources for LC- and GC-API MS, 58 th ASMS Conference on Mass Spectrometry and allied topics, Salt Lake City, UT, USA (2010) [2] Barnes, I.; Kersten, H.; Bejan, J.; Benter, T., In- situ MS monitoring of atmospheric degradation product studies of aromatic hydrocarbons with APPI and APLI, 59 th ASMS Conference on Mass Spectrometry and allied topics, Denver, CO, USA (2011) [3] Kersten, H.; Haberer, K.; Kroll, K.; Benter, T., Progress in the development of a GC-APPI source with femto gram sensitivity, 62 nd ASMS Conference on Mass Spectrometry and allied topics, Baltimore MD, USA (2014) MS Ion Source Radiation Source GC Transfer Line Samples GC injection Make-up gas CFD simulations amaZon speed ETD, Bruker Daltonik GmbH Custom-made dual-mode laminar flow ion source APPI: VUV Kr discharge RF lamp emitting 10.0 and 10.6 eV photons (Syagen, Morfo) APLI: ATL Atlex KrF*-excimer laser (248 nm, 5 mJ, 50 Hz) GC 7890 A, Agilent Technologies Inc. Custom temperature-controlled GC- transfer line EPA 8270 LCS Mix 1, Supelco 78 compounds Dilutions: 1 fg/μL – 100 pg/μL Naphthalene diluted in n-hexane Dilutions: 10 fg/μL – 1 ng/μL Split ratio 1/100 (already included in the concentration values) Injection volume 1 μL Nitrogen 5.0, Messer Industriegase GmbH Flow rate: 760 mL/min Comsol Multiphysics 4.4 0.0E+00 2.0E+05 4.0E+05 6.0E+05 8.0E+05 0 2 4 6 8 10 12 14 intensity [a. u.] retention time [min] y = 113.21x + 1011.7 R² = 0.9945 y = 265.93x – 1257.2 R² = 0.9992 y = 16.013x + 436.48 R² = 0.9999 y = 12.068x – 3.7542 R² = 0.9964 0.0E+00 5.0E+03 1.0E+04 1.5E+04 2.0E+04 2.5E+04 3.0E+04 0 100 200 300 400 500 600 700 peak area [a. u.] concentration [pg/μL] Setup 1 Setup 2 Setup 3 Setup 4 0.0E+00 2.0E+04 4.0E+04 6.0E+04 8.0E+04 1.0E+05 1.2E+05 0 200 400 600 800 1000 1200 peak area [a. u.] concentration [pg/μL] Setup 2 0.0E+00 2.0E+06 4.0E+06 6.0E+06 8.0E+06 1.0E+07 1.2E+07 1.4E+07 1.6E+07 0 200 400 600 800 1000 1200 peak area [a. u.] concentration [pg/μL] Setup 1 y = 34119x + 31893 R² = 0.9991 y = 352.27x + 1296.5 R² = 0.9964 y = 2796.7x – 1165.3 R² = 0.9993 y = 65.173x – 831.11 R² = 0.9974 0.0E+00 5.0E+05 1.0E+06 1.5E+06 2.0E+06 2.5E+06 3.0E+06 3.5E+06 4.0E+06 0 100 200 300 400 500 600 700 peak area [a. u.] concentration [pg/μL] Setup 1 Setup 2 Setup 3 Setup 4 0.0E+00 1.0E+07 2.0E+07 3.0E+07 4.0E+07 5.0E+07 6.0E+07 0 2 4 6 8 10 12 14 intensity [a. u.] retention time [min.] 0.0E+00 5.0E+03 1.0E+04 1.5E+04 2.0E+04 2.5E+04 off axis on axis peak area [a. u.] GC column position

Upload: others

Post on 23-Sep-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Development of a dual-mode laminar flow ion source for ... · volume inlet is based on the GC-APPI source presented by Kersten et al. in 2014 [3]) •Use of non outgassing materials

sample from GC

make-up gas inlet

to MS radiation source

GC-transfer line

Kai Kroll; Walter Wissdorf; Hendrik Kersten; Thorsten BenterPhysical & Theoretical Chemistry

Wuppertal, Germany

Institute for Pure and Applied Mass Spectrometry

Experimental SetupIntroduction

Development of a dual-mode laminar flow ion source for APPI-

and APLI-GC-MS

Methods

Conclusions

Acknowledgement

Literature

• 24/7 heating of all setups at 300 °C for several

weeks without loss of performance

• GC measurements reveal a high level of linearity

in the pg on-column range

• Transfer of the analytes into the MS without

significant loss of separation performance

Peak widths down to 0.7 s (FWHM)

• Simple variation of geometric parameters

• Ion source is interfaced to MS control software

APLI Mode:

• As expected, a larger inner diameter and a longer

dwell time leads to a better ionization efficiency

and gain in sensitivity

• Limit of detection for PAHs down to the fg range

APPI Mode:

• Setup 1 shows a significant decrease in sensitivity

compared to setup 2. The longer tube length

presumably leads to more pronounced ion-wall

losses

• Gain of sensitivity with longer dwell times in the

conical ionization inlet (vortex)

• The simulations show a longer residence time in

the cone with 5 mm tube diameter. These results

are in accordance with the measurements

• A GC column position slightly off axis leads to an

increased ionization efficiency

Table 1: Overview of the different setup modifications; ion

dwell times calculated from make-up gas volume flow

SetupLength

[mm]

Inner

diameter

[mm]

Ion dwell time

[ms]

1 150 5 240

2 85 5 140

3 150 2.5 70

4 85 2.5 50

APLI APPIPeak width

(FWHM)

[s]

Limit of

Detection*

[pg/µL]

Slope

Linear

range**

[pg/µL]

Setup

Linear

range**

[pg/µL]

Slope

Limit of

Detection*

[pg/µL]

Peak width

(FWHM)

[s]

0.8 – 1 0.6 34119 0.8 – 100 1 10 – 100 113 8 1 – 1.3

0.7 – 1.1 6 35 10 – 100 2 10 – 100 265 6 0.8 – 1.1

0.8 – 1.3 3 2797 6 – 600 3 100 – 600 16 60 0.8 – 1.3

0.8 – 1.3 40 65 60 – 600 4 100 – 600 12 80 0.7 – 1.3

Figure 1a: The custom dual-mode laminar flow ion source setup

• S/N = 3 ** for a dilution series of Naphthalene in n-hexane

Table 2: Analytical performance of the four laminar flow ion source versions for APLI- and APPI-GC-MS.

Performance

Crucial design aspects

• Tightly sealed ionization volume

• Chemically and photo-physically

inert matrix gas

• Conical ionization volume inlet

• Asymmetrical make-up gas inlet

Vortex inside ionization volume

for higher ionization efficiency

• Injection of the GC flow into the

major make-up gas flow

(The concept of the ionization

volume inlet is based on the GC-APPI

source presented by Kersten et al. in

2014 [3])

• Use of non outgassing materials

• Controlled operation temperature

up to 300 °C

Figure 3e: Determination of the linear ranges of the four

versions of the laminar flow ion source setup, APPI, for

naphthalene (m/z 128)

Figure 3a: Linear ranges for the four modifications of the

laminar flow ion source setup in APLI mode for naphthalene

(m/z 128).

Figure 3b: Dilution series of naphthalene 0.8 –

1000 pg/µL (m/z 128). Setup 1, APLI

Figure 3d: Dilution series of naphthalene 0.8 –

1000 pg/µL (m/z 128). Setup 1, APPI

Figure 3c: Chromatogram of the EPA 8270 LCS Mix 1

(100 pg/µL). Ion source setup 3, APLI mode.

Figure 3f: Chromatogram of the EPA 8270 LCS Mix 1

(100 pg/µL). Setup 1, APPI mode.

Note: The pronounced signal deviation around 200 pg/µL is observed in all four setups and for both ionization

methods. The reasons for this behavior are yet unknown and subject to current investigations.

CFD Simulations

Figure 2a: CFD simulation of the ionization volume inlet. Make-up gas flow 760

mL/min; GC column flow 2 mL/min. The red flow lines indicate the make up gas

flow, the yellow flow lines indicate the GC column effluent

make-up gasmake-up gas

make-up gas

sample from GC sample from GC

sample from GC sample from GC

make-up gas

Figure 2b: Variation of GC

column position for setup 1,

APPI. 100 pg/µL naphthalene

GC column on-axisGC column 1 mm off-axisAt the annual ASMS meeting in 2010, we introduced

a novel laminar flow ion source (LFIS) for API-MS [1].

In the following years, we presented the capabilities

of the LFIS with the analysis of in-situ samples from

atmospheric smog chamber experiments [2]. Based

on the original LFIS, we designed a novel interface

for GC-API-MS.

Challenges:

• Design of a novel GC-MS interface, suitable for

atmospheric pressure photoionization (APPI) and

atmospheric pressure laser ionization (APLI)

• Design with sustained chromatographic fidelity

→ Prevention of cold spots

→ Prevention of dead volumes

• Swift change between APPI and APLI operation

• Variable ionization chamber volume (length and

inner diameter) to optimize for the ionization

efficiency and ion transport in APPI and APLI

• Controllable via MS-software

iGenTraX UG, Haan, Germany

Bruker Daltonics GmbH, Bremen, Germany

5 m

m t

ub

e2

.5 m

m t

ub

e

[1] Barnes, I.; Kersten, H.; Wissdorf, W.; Pöhler, T.; Hönen, H.; Klee, S.; Brockmann, K. J.; Benter, T., Novel laminar flow ion sources for LC- and GC-API MS, 58th ASMS Conference on Mass Spectrometry and allied topics, Salt Lake City, UT, USA (2010)

[2] Barnes, I.; Kersten, H.; Bejan, J.; Benter, T., In- situ MS monitoring of atmospheric degradation product studies of aromatic hydrocarbons with APPI and APLI, 59th ASMS Conference on Mass Spectrometry and allied topics, Denver, CO, USA (2011)

[3] Kersten, H.; Haberer, K.; Kroll, K.; Benter, T., Progress in the development of a GC-APPI source with femtogram sensitivity, 62nd ASMS Conference on Mass Spectrometry and allied topics, Baltimore MD, USA (2014)

MS

Ion Source

Radiation

Source

GC

Transfer

Line

Samples

GC

injection

Make-up

gas

CFD

simulations

amaZon speed ETD, Bruker Daltonik

GmbH

Custom-made dual-mode laminar

flow ion source

APPI: VUV Kr discharge RF lamp

emitting 10.0 and 10.6 eV photons

(Syagen, Morfo)

APLI: ATL Atlex KrF*-excimer laser

(248 nm, 5 mJ, 50 Hz)

GC 7890 A, Agilent Technologies Inc.

Custom temperature-controlled GC-

transfer line

EPA 8270 LCS Mix 1, Supelco

78 compounds

Dilutions: 1 fg/µL – 100 pg/µL

Naphthalene diluted in n-hexane

Dilutions: 10 fg/µL – 1 ng/µL

Split ratio 1/100 (already included in

the concentration values)

Injection volume 1 µL

Nitrogen 5.0, Messer Industriegase

GmbH

Flow rate: 760 mL/min

Comsol Multiphysics 4.4

0.0E+00

2.0E+05

4.0E+05

6.0E+05

8.0E+05

0 2 4 6 8 10 12 14

inte

nsi

ty [

a.

u.]

retention time [min]

y = 113.21x + 1011.7

R² = 0.9945

y = 265.93x – 1257.2

R² = 0.9992

y = 16.013x + 436.48

R² = 0.9999

y = 12.068x – 3.7542

R² = 0.9964

0.0E+00

5.0E+03

1.0E+04

1.5E+04

2.0E+04

2.5E+04

3.0E+04

0 100 200 300 400 500 600 700

pe

ak a

rea

[a

. u

.]

concentration [pg/µL]

Setup 1 Setup 2 Setup 3 Setup 4

0.0E+00

2.0E+04

4.0E+04

6.0E+04

8.0E+04

1.0E+05

1.2E+05

0 200 400 600 800 1000 1200

pe

ak a

rea

[a

. u

.]

concentration [pg/µL]

Setup 2

0.0E+00

2.0E+06

4.0E+06

6.0E+06

8.0E+06

1.0E+07

1.2E+07

1.4E+07

1.6E+07

0 200 400 600 800 1000 1200

pe

ak a

rea

[a

. u

.]

concentration [pg/µL]

Setup 1

y = 34119x + 31893

R² = 0.9991

y = 352.27x + 1296.5

R² = 0.9964

y = 2796.7x – 1165.3

R² = 0.9993

y = 65.173x – 831.11

R² = 0.9974

0.0E+00

5.0E+05

1.0E+06

1.5E+06

2.0E+06

2.5E+06

3.0E+06

3.5E+06

4.0E+06

0 100 200 300 400 500 600 700

pe

ak a

rea

[a.

u.]

concentration [pg/µL]

Setup 1 Setup 2 Setup 3 Setup 4

0.0E+00

1.0E+07

2.0E+07

3.0E+07

4.0E+07

5.0E+07

6.0E+07

0 2 4 6 8 10 12 14

inte

nsi

ty [

a.

u.]

retention time [min.]

0.0E+00

5.0E+03

1.0E+04

1.5E+04

2.0E+04

2.5E+04

off axis on axis

pe

ak a

rea

[a

. u

.]

GC column position