  • Continuous Gas Analysis

    ULTRAMAT 6, OXYMAT 6 Safety Manual

    Operating Instructions

    Supplement to instruction Manuals ULTRAMAT 6, OXYMAT 6 OXYMAT 61

    12/2014 A5E00695577-05

    Introduction 1

    General description of functional safety

    2

    Device-specific safety instructions

    3

    List of abbreviations A

  • Siemens AG Industry Sector Postfach 48 48 90026 NÜRNBERG GERMANY

    Order number: A5E00695577 Ⓟ 11/2014 Subject to change

    Copyright © Siemens AG 2005 - 2014. All rights reserved

    Legal information Warning notice system

    This manual contains notices you have to observe in order to ensure your personal safety, as well as to prevent damage to property. The notices referring to your personal safety are highlighted in the manual by a safety alert symbol, notices referring only to property damage have no safety alert symbol. These notices shown below are graded according to the degree of danger.

    DANGER indicates that death or severe personal injury will result if proper precautions are not taken.

    WARNING indicates that death or severe personal injury may result if proper precautions are not taken.

    CAUTION indicates that minor personal injury can result if proper precautions are not taken.

    NOTICE indicates that property damage can result if proper precautions are not taken.

    If more than one degree of danger is present, the warning notice representing the highest degree of danger will be used. A notice warning of injury to persons with a safety alert symbol may also include a warning relating to property damage.

    Qualified Personnel The product/system described in this documentation may be operated only by personnel qualified for the specific task in accordance with the relevant documentation, in particular its warning notices and safety instructions. Qualified personnel are those who, based on their training and experience, are capable of identifying risks and avoiding potential hazards when working with these products/systems.

    Proper use of Siemens products Note the following:

    WARNING Siemens products may only be used for the applications described in the catalog and in the relevant technical documentation. If products and components from other manufacturers are used, these must be recommended or approved by Siemens. Proper transport, storage, installation, assembly, commissioning, operation and maintenance are required to ensure that the products operate safely and without any problems. The permissible ambient conditions must be complied with. The information in the relevant documentation must be observed.

    Trademarks All names identified by ® are registered trademarks of Siemens AG. The remaining trademarks in this publication may be trademarks whose use by third parties for their own purposes could violate the rights of the owner.

    Disclaimer of Liability We have reviewed the contents of this publication to ensure consistency with the hardware and software described. Since variance cannot be precluded entirely, we cannot guarantee full consistency. However, the information in this publication is reviewed regularly and any necessary corrections are included in subsequent editions.

  • Safety Manual Operating Instructions, 12/2014, A5E00695577-05 5

    Table of contents

    1 Introduction ............................................................................................................................................. 7

    1.1 Purpose of this document ......................................................................................................... 7

    1.2 Device variants ......................................................................................................................... 7

    1.3 Additional documentation ......................................................................................................... 8

    1.4 History ....................................................................................................................................... 8

    1.5 Further information .................................................................................................................... 9

    2 General description of functional safety ................................................................................................. 11

    2.1 Safety-instrumented system ................................................................................................... 11

    2.2 Safety integrity level ................................................................................................................ 12

    3 Device-specific safety instructions ......................................................................................................... 15

    3.1 Applications ............................................................................................................................. 15

    3.2 Safety function ........................................................................................................................ 15

    3.3 Settings ................................................................................................................................... 16

    3.4 Behavior in case of faults ........................................................................................................ 17

    3.5 Maintenance and checks ........................................................................................................ 18

    3.6 Safety characteristics .............................................................................................................. 19

    3.7 OXYMAT 61 ............................................................................................................................ 19

    A List of abbreviations .............................................................................................................................. 21

    Glossary ............................................................................................................................................... 23

  • Table of contents

    Safety Manual 6 Operating Instructions, 12/2014, A5E00695577-05

  • Safety Manual Operating Instructions, 12/2014, A5E00695577-05 7

    Introduction 1 1.1 Purpose of this document

    This document contains all information and safety instructions required when using an ULTRAMAT 6, OXYMAT 6, ULTRAMAT/OXYMAT 6 or OXYMAT 61 gas analyzer in safety-instrumented systems.

    It addresses system planners, constructors, service and maintenance engineers and any personnel commissioning and operating the device.

    1.2 Device variants There are the following types of analyzers:

    ● Stand alone ULTRAMAT 6 analyzers

    ● Stand alone OXYMAT 6 analyzers

    ● Stand alone OXYMAT 61 analyzers

    ● Units with two analyzers ULTRAMAT 6

    ● Units with two analyzers ULTRAMAT 6/OXYMAT 6

    The term 'Gas analyzer' is used for all these types of devices. Features which apply to only one or some of these types are accordingly marked.

    This safety manual applies only to the analyzer types listed in the following table with the additional order code 'Z+C20'.

    Table 1- 1 ULTRAMAT and OXYMAT 6 types

    Analyzer name Design Standard analyzer Special analyzer ULTRAMAT 6F 1 channel (IR) 7MB2111 ULTRAMAT 6F 2 channels (1 x 2 IR) 7MB2112 ULTRAMAT 6E 1 channel (IR) 7MB2121 ULTRAMAT 6E 2 channels (2 x 1 IR) 7MB2123 ULTRAMAT 6E 2 channels (1 x 2 IR)

    3 channels (1 x 1 IR, 1 x 2 IR) 7MB2124

    ULTRAMAT 6E 4 channels (2 x 2 IR) ULTRAMAT/OXYMAT 6E 2 channels (1 IR, 1 O2) 7MB2023 ULTRAMAT/OXYMAT 6E 3 channels (2 IR,1 O2) 7MB2024 OXYMAT 6F 1 channel (O2) 7MB2011 7MB2017 OXYMAT 6E 1 channel (O2) 7MB2021 7MB2027 OXYMAT 61 1 channel (O2) 7MB2001

  • Introduction 1.3 Additional documentation

    Safety Manual 8 Operating Instructions, 12/2014, A5E00695577-05

    1.3 Additional documentation This document deals with the ULTRAMAT 6, OXYMAT 6 and OXYMAT 61 gas analyzers exclusively as part of a safety function. This document is valid only in conjunction with one of the following documents: No. Designation Order no. /1/ Instruction Manual ULTRAMAT 6 and OXYMAT 6 C79000-G5276-C143 (English)

    C79000-G5200-C143 (German) C79000-G5277-C143 (French) C79000-G5272-C143 (Italian) C79000-G5278-C143 (Spanish)

    /2/ Operating Instructions OXYMAT 61 A5E00123067 (English) A5E00123066 (German) A5E00123068 (French) A5E00123069 (Spanish) A5E00123070 (Italian)

    1.4 History The following table shows the released versions and the changes in the documentation compared to each preceding edition: Edition Remark 01 10/2005

    First edition

    02 07/2010

    Certificate of conformity updated

    03 10/2013

    OXYMAT 61 introduced

    04 07/2014

    ULTRAMAT 6 and OXYMAT 6 Safety Manuals united in one document

    05/ 12/2014

    SIL 2 relevant ULTRAMAT device numbers (MLFB) specified

  • Introduction 1.5 Further information

    Safety Manual Operating Instructions, 12/2014, A5E00695577-05 9

    1.5 Further information

    Information The contents of these instructions shall not become part of or modify any prior or existing agreement, commitment or legal relationship. All obligations on the part of Siemens AG are contained in the respective sales contract which also contains the complete and solely applicable warranty conditions. Any statements contained herein do not create new warranties or modify the existing warranty.

    The content reflects the technical status at the time of printing. We reserve the right for technical changes in the course of further development.

    References If further information on an aspect described here is referenced to, the corresponding reference will always be found at the end of a section under "See also".

  • Introduction 1.5 Further information

    Safety Manual 10 Operating Instructions, 12/2014, A5E00695577-05

  • Safety Manual Operating Instructions, 12/2014, A5E00695577-05 11

    General description of functional safety 2 2.1 Safety-instrumented system

    This chapter describes the functional safety in general and not specific to a device. The devices in the examples are selected as representative examples.

    System description The sensor (analyzer), automation system and final controlling element combine to form a safety-instrumented system, which executes a safety function.

    Functional principle The transmitter (analyzer) generates a process-related measured value that is transferred to the automation system. The automation system monitors this measured value. If the measured value exceeds the range of the high or low limit, the automation system generates a shutdown signal for the connected final controlling element, which switches the associated valve to the specified safety position.

    Special characteristics

    ● Four freely-programmable measuring ranges per component, also with suppressed zero

    ● All measuring ranges are linear

    ● The isolated analog output is set to 4 to 20 mA (NAMUR)

  • General description of functional safety 2.2 Safety integrity level

    Safety Manual 12 Operating Instructions, 12/2014, A5E00695577-05

    2.2 Safety integrity level

    Definitions SIL, Safety Integrity Level

    The international standard IEC 61508 defines four discrete Safety Integrity Levels (SIL) from SIL 1 to SIL 4. Each level corresponds to the probability range for the failure of a safety function. The higher the SIL of the safety-instrumented system, the higher probability that the required safety function will work.

    The achievable SIL is determined by the following safety characteristics:

    ● Average probability of dangerous failure of a safety function in case of demand (PFDAVG)

    ● Hardware fault tolerance (HFT)

    ● Safe failure fraction (SFF)

    Description The following table shows the dependency of the SIL on the average probability of dangerous failures of a safety function of the entire safety-instrumented system" (PFDAVG) The table deals with "Low demand mode", i.e. the safety function is required a maximum of once per year on average.

    Table 2- 1 Safety Integrity Level

    SIL Interval 4 10-5 ≤ PFDAVG < 10-4 3 10-4 ≤ PFDAVG < 10-3 2 10-3 ≤ PFDAVG < 10-2 1 10-2 ≤ PFDAVG < 10-1

    The "average probability of dangerous failures of the entire safety-instrumented system" (PFDAVG) is normally split between the three sub-systems in the following figure.

    Figure 2-1 PFD distribution

  • General description of functional safety 2.2 Safety integrity level

    Safety Manual Operating Instructions, 12/2014, A5E00695577-05 13

    The following table shows the achievable Safety Integrity Level (SIL) for the entire safety-instrumented system for type B subsystems depending on the proportion of safe failures (SFF) and the hardware fault tolerance (HFT). Type B subsystems include sensors with positioners, actuators with complex components, e.g. microprocessors (see also IEC 61508, Section 2).

    Table 2- 2 Achievable Safety Integrity Level (type B subsystems)

    SFF HFT 0 1 (0) * 2 (1) *

    < 60 % Not allowed SIL 1 SIL 2 60 ... 90 % SIL 1 SIL 2 SIL 3 90 ... 99 % SIL 2 SIL 3 SIL 4

    > 99 % SIL 3 SIL 4 SIL 4 * As per IEC 61511-1, Section 11.4.4

    Operational reliability According to IEC 61511-1, Section 11.4.4, the hardware fault tolerance (HFT) can be reduced by one (values in brackets) for transmitters and final controlling elements with complex components if the following conditions are applicable to the device:

    ● The device is based on prior use.

    ● The user can configure only the process-related parameters, e.g. control range, signal direction in case of a fault, limiting values, etc.

    ● The configuration level of the firmware is blocked against unauthorized operation.

    ● The function requires SIL of less than 4.

    The device fulfills these conditions.

  • General description of functional safety 2.2 Safety integrity level

    Safety Manual 14 Operating Instructions, 12/2014, A5E00695577-05

  • Safety Manual Operating Instructions, 12/2014, A5E00695577-05 15

    Device-specific safety instructions 3 3.1 Applications

    The gas analyzers are suitable for use in a safety-instrumented function of Safety Integrity Level (SIL) 2 in low demand mode within a 1oo1 architecture.

    The proven in-use-assessment was carried out by RISKNOWLOGY according to IEC 61508 / IEC 61511. Product revisions will be carried out by the manufacturer in accordance with IEC 61508. The gas analyzers are usable in safety applications to monitor limits.

    3.2 Safety function The gas analyzers are mainly used for user defined threshold monitoring. Only the 4 to 20 mA analog output (NAMUR) may be used as part of a safety-instrumented function (SIF). In addition to the application-specific measurement error, an additional safety accuracy of 5 % of the maximum measurement range must be added. The total tolerance (safety function) is calculated as follows:

    Total tolerance (safety function) = ± [application-specific measuring error + 5% safety accuracy]

    Safety Accuracy of the gas analyzers: the maximum effect of a single failure on the measured value, which is classified to have no effect.

    NOTICE

    Loss of safety function

    The mandatory settings and conditions are listed in the "Settings (Page 16)" and "Safety characteristics (Page 19)" sections.

    These conditions must be adhered to in order to fulfill the safety function.

    When the safety function has been executed the safety-instrumented systems shall be brought to a safe state. The Mean Time To Restoration (MTTR) is 8 hours.

    Reference

    Instruction manuals ULTRAMAT 6/OXYMAT 6 and OXYMAT 61 (see section Additional documentation (Page 8) for order nos.)

    See also Settings (Page 16)

    Safety characteristics (Page 19)

    Additional documentation (Page 8)

  • Device-specific safety instructions 3.3 Settings

    Safety Manual 16 Operating Instructions, 12/2014, A5E00695577-05

    3.3 Settings After assembly and commissioning in line with the device manual, the following parameter settings shall be made when the devices is used as part of a SIF:

    Safety parameters Enter the following parameter via user menu: Func-tion number

    Function Action

    24 AUTOCAL/-check Submenu 'AUTOCAL/-check mode': Parameter AUTOCAL/-check on/off' must be set to 'off'

    70 Analog output Select setting 'NAMUR/4 - 20 mA' 77 Store analog output

    Set 'analog output to 21 mA' Set 'store on/off' to 'on''

    Reference ● ULTRAMAT 6/OXYMAT 6 Manual, OXYMAT 61 Operating instructions

    Protection against configuration changes After configuration, the analyzer access codes (function 79) shall be changed such that the device is protected against unintentional and unauthorized changes/operation.

    Checking the safety function after installation Following installation and commissioning of the analyzer a safety function test has to be carried out as described in chapter 5 "Operation" and chapter 6 "Maintenance" of the instruction manual.

    ULTRAMAT 6: In versions with purged reference gas side the same gas should be used as zero gas and reference gas, e. g N2. Using zero gas, 4 mA must be measured at analog output. For the test of the safety function it is fundamental to use a second calibration gas with a defined proportion of the span gas. The results of the measurement must be within a range of ±5 % (full span) of the expected result.

    OXYMAT 6: Use zero gas as reference gas. You can also use as zero gas from the same source as the reference gas. In that case make sure that the reference gas pressure remains stable during the zero gas calibration. Using zero gas, 4 mA must be measured at analog output. For the test of the safety function it is fundamental to use a second calibration gas with a defined proportion of oxygen. The results of the measurement must be within a range of ±5 % (full span) of the expected result.

  • Device-specific safety instructions 3.4 Behavior in case of faults

    Safety Manual Operating Instructions, 12/2014, A5E00695577-05 17

    3.4 Behavior in case of faults

    Faults The procedure in case of faults is described in the manual or the operating Instructions of the device.

    Repairs Defective devices should be sent to the repair department stating details and cause of any fault. When ordering replacement devices, specify also the serial number of the original device. The serial number can be found on the nameplate.

    Reference Addresses of the responsible repair center, contact partners, spare parts lists etc. can be obtained from the ULTRAMAT 6 and OXYMAT 6 Manual and from the OXYMAT 61 Operating instructions as well as from the following web address: Services & Support (http://www.siemens.com/automation/service&support).

    The SIL declaration and other manuals can be obtained from the following web address: SIL (www.siemens.com/SIL).

    http://www.siemens.com/automation/service&supporthttp://www.siemens.com/SIL

  • Device-specific safety instructions 3.5 Maintenance and checks

    Safety Manual 18 Operating Instructions, 12/2014, A5E00695577-05

    3.5 Maintenance and checks

    Checking the analyzer functions We strongly recommend to check the function capability of the gas analyzers at regular intervals of at least three months.

    Such a test shall comprise at least the basic functions as described in the ULTRAMAT 6 and OXYMAT 6 manual and in the OXYMAT 61 operating instructions.

    Functional safety proof test You should regularly check the safety function of the entire safety circuit according to IEC 61508/61511.

    The test interval is determined during calculation of each individual safety circuit in a system (PFDAVG). The recommended testing interval depends largely on the application but should never exceed three months. We strongly recommend to execute

    ● proof test 1 every three months

    ● proof test 2 every month.

    To detect dangerous faults the analyzer analog output and analog input shall be checked with the following tests:

    Table 3- 1 Proof test 1

    Step Action 1 Bypass the safety PLC or take another appropriate action to avoid a false trip. 2 Generate or simulate an alarm condition to force the gas analyzer to exceed the high alarm current output

    limit and verify the analog current value (e.g. by disconnecting pressure / temperature input). 3 Generate or simulate an alarm condition to force the gas analyzer to exceed the low alarm current output

    limit and verify the analog current value (e.g. by disconnecting pressure / temperature input). 4 Restore the loop to full operation. 5 Remove the bypass from the safety PLC and restore to normal operation.

    Table 3- 2 Proof test 2

    Step Action 1 Bypass the safety PLC or take another appropriate action to avoid a false trip. 2 Perform a two-point calibration of the gas analyzer. 3 Perform a reference measurement with at least one measuring point between minimum and maximum

    concentration. To do so use a calibration gas with a well-known gas concentration. The expected result shall not have a tolerance of more than 2 %.

    4 Restore the loop to full operation. 5 Remove the bypass from the safety PLC and restore to normal operation.

    Both proof test parts will detect more than 90% of possible "du" failures of the gas analyzers.

  • Device-specific safety instructions 3.6 Safety characteristics

    Safety Manual Operating Instructions, 12/2014, A5E00695577-05 19

    3.6 Safety characteristics The safety characteristics necessary for use of the system are listed in the SIL declaration of conformity. These values apply under the following conditions:

    ● The user shall take care that during the measuring procedure the reference gas overpressure complies to the specified values.

    ● The user is responsible for a sufficient measuring gas and reference gas flow according to the values specified in the manuals.

    ● The gas analyzer is only used in safety-related systems in low demand mode for the safety function.

    ● The safety-related parameters/settings as described in section Settings (Page 16) have been input by local operation and checked before starting safety-instrumented operation.

    ● The gas analyzer is blocked against unintentional and unauthorized changes/operations.

    ● The average temperature observed over a long period did not exceed 40 °C (104 °F).

    ● All used materials are compatible with process conditions.

    ● The MTTR after a device fault is 8 hours.

    ● The best time to react on a dangerous detected failure is 1 hour.

    ● The logic solver (PLC) has to be configured to detect low range (21 mA) failures of the gas analyzer (fail low and fail high) and will recognize these as an internal failure of the devices and not cause a spurious trip.

    ● The AUTOCAL/-check functionality must be deactivated by setting the 'AUTOCAL/-check on/off' switch to 'off'

    3.7 OXYMAT 61 When operating an OXYMAT 61 gas analyzer consider additionally that the measuring features of the OXYMAT 61 compared to the OXYMAT 6 and ULTRAMAT 6 analyzers are inferior in the following aspects:

    ● higher influence of the ambient temperature on measured values

    ● no compensation of vibrations

  • Device-specific safety instructions 3.7 OXYMAT 61

    Safety Manual 20 Operating Instructions, 12/2014, A5E00695577-05

  • Safety Manual Operating Instructions, 12/2014, A5E00695577-05 21

    List of abbreviations A

    Abbreviations Abbreviation Full term Meaning > Greater than Mathematical sign for inequality < Less than Mathematical sign for inequality ± Plus-minus sign Mathematical sign that indicates a tolerance in both directions of the

    referenced value % Percent Mathematical sign for the hundredth part of a whole DC Diagnostic Coverage Parameter which describes the ratio of the failure rate of detected

    dangerous failures to the failure rate of all dangerous failures. "du" Dangerous undetected Dangerous undetected failures FMEDA Failure Modes, Effects and Diag-

    nostic Analysis Method to determine out of an overall failure rate of a system the proportion of failures without the potential to bring the safety-instrumented system into a dangerous or impermissible functional status (SFF) and the diagnostic coverage (DC) according to the re-quirements to IEC 61508.

    HFT Hardware Fault Tolerance Capability of a function unit to continue execution of a required func-tion in the presence of faults or deviations.

    IEC International Electrotechnical Commission

    An international standards organization dealing with electrical, elec-tronic and related technologies

    I/O Input/Output I/O refers to the communication between an information processing system (such as a computer), and the outside world, such as a hu-man, or another information processing system. Inputs are the sig-nals or data received by the system, and outputs are the signals or data sent from it.

    IR Infrared Gas detection method using an infrared source to detect and meas-ure infrared-sensitive gases.

    mA Milliampere A unit of electrical current MLFB Maschinenlesbare Fabrikatebez-

    eichnung German for Machine-readable product identification, a product code

    MTBF Mean Time Between Failures Average period between two failures MTTR Mean Time To Restoration Average period between the occurrence of a fault on a device or

    system and the restoration of a flawless state N2 Nitrogen (molecular formula) Chemical element with atomic number 7 and represented by the

    symbol N. Its name derives from the Greek roots νίτρον (sodium carbonate) and -γενής (producer, literally begetter), because at the time of naming, the word 'nitre' was mistakenly used for sodium salts. At standard temperature and pressure, two atoms of the element bind to form dinitrogen, a colorless, odorless, tasteless diatomic gas with the formula N2. In this case: a component used as reference gas

  • List of abbreviations

    Safety Manual 22 Operating Instructions, 12/2014, A5E00695577-05

    Abbreviation Full term Meaning NAMUR Normenarbeitsgemeinschaft für

    Mess- und Regeltechnik NAMUR is an international user association of automation technology in process industries

    O2 Oxygen (molecular formula) Chemical element with atomic number 8 and represented by the symbol O. Its name derives from the Greek roots ὀξύς (acid, literally "sharp", referring to the sour taste of acids) and -γενής (producer, literally begetter), because at the time of naming, it was mistakenly thought that all acids required oxygen in their composition. At stand-ard temperature and pressure, two atoms of the element bind to form dioxygen, a colorless, odorless, tasteless diatomic gas with the for-mula O2. In this case: a measuring component

    PFD Probability of Failure on Demand Probability of dangerous failures of a safety function on demand PFDAVG Probability of Failure on Demand

    (AVERAGE) Average probability of dangerous failures of a safety function on demand

    PLC Programmable Logic Controller A digital computer used for automation of industrial processes, such as machinery control in factories

    SFF Safe Failure Fraction Proportion of safe failures: Proportion of failures without the potential to bring the safety-instrumented system into a dangerous or impermissible functional status.

    SIF Safety Instrumented Function Function to be implemented by a safety-related system or an external risk reduction facility, which is intended to achieve or maintain a safe state for the process, with respect to a specific hazardous event

    SIL Safety Integrity Level The international standard IEC 61508 defines four discrete Safety Integrity Levels (SIL 1 to SIL 4). Each level corresponds to a proba-bility range for failures of a safety function. The higher the Safety Integrity Level of the safety-instrumented system, the lower the prob-ability that it will not execute the required safety functions.

    XooY X out of Y voting Classification and description of the safety-instrumented system in terms of redundancy and the selection procedures used in which - Y specifies how often a safety function is executed (redundancy) and - X determines how many channels have to work correctly. Example: Pressure measurement in a 1oo2 architecture: A safety instrument-ed-system decides that a specified pressure limit has been exceeded if one out of two pressure sensors reaches this limit. In a 1oo1 archi-tecture, there is only one pressure sensor.

  • Safety Manual Operating Instructions, 12/2014, A5E00695577-05 23

    Glossary

    Dangerous failure Failure with the potential to bring the safety-instrumented system into a dangerous or non-functional status

    Low Demand Mode The frequency of demands for operation made on a safety-related system is not more than one per year and not more than twice the proof-test frequency.

    Safety function Defined function executed by a safety-instrumented system with the objective of achieving or maintaining a safe system status which takes into account a defined occurrence of dangerous failures.

    Example: Monitoring of user-defined limit values.

    Safety Integrity Level See SIL

    Safty-instrumented system A safety-instrumented system excludes the safety functions that are required to achieve or maintain a safe status in a system. It consists of a sensor, a logic unit/control system and a final controlling element.

    Example: A safety-instrumented system is made up of a pressure transmitter, a limit signal sensor and a control valve.

    SIL The international standard IEC 61508 defines four discrete Safety Integrity Levels (SIL) from SIL 1 to SIL 4. Each level corresponds to a probability range for the failure of a safety function. The higher the SIL of the safety-instrumented system, the higher the probability that the required safety function will work.

    The achievable SIL is determined by the following safety characteristics:

    ● Average probability of dangerous failure of a safety function in case of demand (PFDAVG)

    ● Hardware fault tolerance (HFT)

    ● Safe failure fraction (SFF)

  • Glossary

    Safety Manual 24 Operating Instructions, 12/2014, A5E00695577-05

  • 1 Introduction1.1 Purpose of this document1.2 Device variants1.3 Additional documentation1.4 History1.5 Further information2 General description of functional safety2.1 Safety-instrumented system2.2 Safety integrity level3 Device-specific safety instructionsA List of abbreviations

    3.1 Applications3.2 Safety function3.3 Settings3.4 Behavior in case of faults3.5 Maintenance and checks3.6 Safety characteristics3.7 OXYMAT 61