calibration reference leaks · volume ≈ 1 ml, step < 1µl (resolution) ... 2459 2. k. jousten...
TRANSCRIPT
+
Calibration of reference leaks in a wide pressure range Ana L. Fonseca, A. Marta Barreto, A. Sofia Matos, A.M.C.Moutinho and Orlando M.N.D.Teodoro
+ Outline
Motivation Reference leaks for sniffers, R134a leaks
Calibration set-up Syringe and stepper motor User interface Measuring range Temperature effects Repeatability Time needed for calibration
Typical uncertainty
Method validation
Typical results R134a and He leaks
2
+ Motivation: reference leaks for sniffers 3
Inside-out He leak detection technique (sniffer) He reference leaks in the range 10-6 to 10-5 mbar.L/s
842/2006 regulation and subsequent statements R134a reference leaks of 5 g/year
+ Motivation 4
Calibration by comparison using a MS leak detector is not suitable Calibration pressure should be atmosphere, not vacuum
EN 13192 method C: calibration of leaks from overpressure to atmosphere Capillary with a moving slug
Although vey simple, is of limited application (slug friction, pressure control, etc.)
However, p∆V method is well suited
+ Calibration method: p∆V 5
The leak is connected to a closed volume
A piston is actuated every time a pressure increase is detected to keep the pressure constant Volume ≈ 1 mL, step < 1µL (resolution)
!"#!$
#"!!$
#!"!!$
#!!"!!$
#!!!"!!$
!$ %!$ #!!$ #%!$ &!!$
!"#$%&'()*+'
,-%&'(%-.+'
'(#!)*+$,-./"012$
'(#!)*%$,-./"012$
'(#!)*3$,-./"012$
1. M. Bergoglio e D. Mari, “INRIM primary standard for microgas-flow measurements with reference to atmospheric pressure” Measurement 45 (2012) 2459
2. K. Jousten e U. Becker, “A primary standard for the calibration of sniffer test leak devices,” Metrologia, 46 (2009) 560-568 3. F. Boineau, “Characterization of the LNE constant pressure flowmeter for the leak flow rates measurements with reference to vacuum
and atmospheric pressure,” PTB-Mitteilungen, 121 (2011) 313
• Absolute gauge, no further volume is required
• Differential gauge, reference volume is required
total volume displaced
+ Calibration set-up 6
Computer & Labview software
thermocouples
stepper motor (syringe pump)
leak under calibration
absolute pressure gauge
double wall box (water inside)
Bath, temperature control and
pump syringe
venting valve
+ Calibration set-up
Piston: 1 mL SGE gas tight syringe with PTFE plunger tip
Piston actuator: syringe pump NE-500 800 steps/revolution; 1 step≈ 1.7 µm
Gauge: Druck DPI 142 precision Barometer Resolution ≈ 3×10-3 mbar; piezo sensor
Software: Labview platform
Tubing: 1/8” swagelok fittings and valves
Stainless stell double wall box with 2 computer fans inside to increase temperature homogeneity
Water circulator Julabo F25-ED
7
+ User interface 8
Pressure (≈10-2 mbar maximum variation)
Flow rate (integrated measurement), converges with time
Temperature drift (3 thermocouples), critical to achieve low uncertainty
Actual measurement
+ Before calibration
∆pV method, for preliminary evaluation
temperature stability confirmation
9
+
!"#$$%&
$#$$%&
"#$$%&
'#$$%&
(#$$%&
)#$$%&
*#$$%&
$& *$$$& "$$$$& "*$$$& '$$$$& '*$$$& ($$$$&
!"#$
%&'()*
#%&+,+-.'/
%012%
3-!"'/%$'!-.'/%0!"#$!2%
!""#
!$"#
%""#
%$"#
&"""#
&"$"#
"# $"""# &""""# &$"""# '""""# '$"""# (""""# ($"""#
!"#$$%"#&'(
)*"+&
!,$-./&0.$,1./&'$-#0$+&
Syringe characterization
Gravimetric volume calibration Step volume Vs=28.0 ± 0.22 nL
Syringe expansion without leak Boyle-Mariotte relation
Full volume range ≈ 32 000 steps
Step volume Vs and dead volume V0 can be calculated by fitting these parameters in the experimental data
10
p = p0V0nVs +V0
Vs =V0n
p0p!1
"
#$
%
&'
Best fit: Vs = 28.18 nL V0 = 3.862 mL
Step volume as function of position
most used position = larger diameter
fitting parameters
Dark line, average of 2000 steps
+ Measuring range, time needed for calibration
2000 steps averages thread defects
11
total steps Q *me
s min h
2000
10-‐4 mbar.L/s
560 9.3
10-‐5 mbar.L/s 5600 93.3 1.6
10-‐6 mbar.L/s 56000 933.3 15.6
5 g/year R134 a 1474 24.6
+ Repeatibility
Effect of the position of the syringe plunger tip (piston)
12
!"!#$%#&'
!"!&$%#&'
!"(#$%#&'
!"(&$%#&'
!")#$%#&'
!")&$%#&'
!"&#$%#&'
!"&&$%#&'
!"*#$%#&'
#' +#' !#' (#' )#' &#' *#'
!"#$%&#'"%()*#&+!,-.%
-/&012"%34-0541%%()).%
average
He leak, nominal leak rate = 2.60 × 10-5 mbar.L/s calibrated to vacuum
+ Repeatability
4 different leaks were measured 4 times each
Repeatability depends on:
1. Number of steps
(acquisition time)
2. Temperature drift has a
major impact
3. Leak stability
On the graph 1000 to
10000 steps were used.
13
1.0E-07
1.0E-06
1.0E-05
1.0E-04
1.0E-03
0 1 2 3 4 5
Lea
k r
ate
(mb
ar.L
/s)
Leak number #
std=14.35%
std=1.13%
std=0.68%
std=0.23%
+ Typical uncertainty
R134a leak, Q=3.48 g/year
14
Quantity, xi Estimate u(xi) ci u(xi)/d (g/year)
relative contribution %
pressure 1000 mbar 0.015 mbar 1.54E-‐05 0.027
volume 22 µL 0.17 µL 7.79E-‐03 13.7
temperature 296.9 K 0.06 K 5.38E-‐04 0.4
Eme 827 s 9.6 x 10-‐3 s 1.16E-‐05 0.02
repeatability (∆T≈0.1K) 0.17 g/year 0.049 85.9
Total (k=2) 0.098
100 2.82%
Q = 3.48 ± 0.10 g/year
+ Temperature effects
The use of an absolute gauge turns this calibration set-up more sensitive to temperature drifts A drift in the temperature is equivalent to a temporary virtual leak (positive or
negative)
This effect is slightly cancelled by tubing dilatation
15
1.00E-08
1.00E-07
1.00E-06
1.00E-05
0.01 0.1 1
equ
ival
ent l
eak
rat
e (m
bar
.L/s
)
Temperature drift (K)
Qeq =p0V0t
T0 +!TT0
"1#
$%
&
'(
Gay-Lussac law
Dead volume V0 ≈ 4 mL
+ Temperature effects
Total temperature drift is more important than temperature uniformity. If temperature changes slowly, what matters is the difference between the starting temperature and
the final temperature
Experiment: induced temperature step Temperature effects may be cancelled by positive and negative drifts
16
!"#$%
!"#&%
!"#'%
!!#"%
!!#(%
!!#$%
!!#&%
!!#'%
!(#"%
!(#(%
!(#$%
)%
)#$%
$%
$#$%
*%
*#$%
&%
+% "+++% !+++% (+++% )+++% $+++%
!"#$%&#'(#)
*#$%+,$#'-./0'
1#%2'$%
+#'-&
34#%$0'
5%678$%9:;'9)#'-<0'
Temperature step in the PID controller: t=550s, ∆T=-0.5 K t=2200s, ∆T=+0.5 K
Nominal leak rate
+ Method validation
A series of R134a leaks were built with leak rates ranging 4 to 24 g/year
17
Caracterização de microfluxos: medição, geração e calibração
72
Figura 4.5 – Constituintes das fugas de permeação de R134a.
A taxa de fuga, no caso de uma fuga de permeação, é descrita pela equação:
! " " equação 4.1
Onde é a taxa de fuga em mol/s, é o coeficiente de permeabilidade da membrana de permeação
em mol/(Pa.m.s), é a área de permeação em m2, é o gradiente de pressão entre a pressão interna
da fuga e a exterior (pressão atmosférica) em Pa e é a espessura da membrana de permeação em m.
Dezenas de frascos foram testados, constituídos por diferentes tipos de vidros, dimensões e gargalos.
Os ensaios preliminares revelaram que alguns tipos de frascos não suportavam a diferença de pressão
(dentro do frasco a pressão, em condições atmosféricas normais era de aproximadamente 5 bar);
R134a
Elemento de
permeação
Elemento de
vedação
Tampa
Anilha metálica
furada
Área de permeação
Elemento de permeação
Tampa
Frasco de vidro Elemento de vedação
+ Method validation
Leaks were weighted over several days (U=± 15%) and leak rate calculated.
Then, all leaks were calibrated in the p∆V set-up and compared
18
!"
#"
$!"
$#"
%!"
%#"
&!"'()*"+"$,#--".$"
'()*"+"$,#--".%"
'()*"+"$,/"-
-"
'()*"+"%,!"-
-".$"
'()*"+"%,!"-
-".%"
'()*"+"%,!"-
-".&"
'()*"+"%,!"-
-".0"
'()*"+"%,%"-
-"
'()*"+"%,#"-
-".$"
'()*"+"%,!"-
-".%"
'()*"+"&,!"-
-".$"
'()*"+"&,!"-
-".%"
!"#$%&#
'"%()
*+"#&,%
1(2345"'677"
89:"
;%#<="
+ R134a reference leaks
Dependence on the delivering pressure 2000 steps measurements ≈ 25 min
19
5.20
5.25
5.30
5.35
5.40
5.45
5.50
5.55
5.60
940 960 980 1000 1020 1040 1060
Lea
k r
ate
(g/y
ear)
delivering pressure (mbar)
-0.033%/mbar@23°C
+ R134a reference leaks
Dependence on the temperature
20
0 1 2 3 4 5 6 7 8 9
10
10 15 20 25 30 35 40
Leak
rat
e(g
/yea
r)
Temperature (ºC)
R134a ø1.5 mm #1
0 1 2 3 4 5 6 7 8 9
10
10 15 20 25 30 35 40 Le
ak r
ate(
g/y
ear)
Temperature (ºC)
R134a ø1.8mm
α =3.1%/K α =2.8%/K
+ He reference leaks (permeation)
Dependence on the delivering pressure 2000 steps measurements ≈ 35 min
21
2.40E-05
2.50E-05
2.60E-05
2.70E-05
2.80E-05
940 960 980 1000 1020 1040 1060
Lea
k r
ate
(mb
ar.L
/s)
delivering pressure (mbar)
-0.027%/mbar@23°C
+ Conclusions
p∆V calibration method was assembled with a syringe pump and is properly working at Metrovac laboratory in Lisbon.
One absolute pressure gauge is used This gauge is suitable to measure both the calibration pressure and the
pressure change
Temperature stability is critical to achieve good uncertainty
Uncertainties ≈3% are easily achieved for the range of 10–5 mbar.L/s or 5g/year
R134a reference leaks were developed and fully characterized. This leaks were used to validate the method
22
+ 23
Acknowledgements: The financial support of the Portuguese Foundation for Science and
Technology is gratefully acknowledged (PTDC/EME- MFE/098738/2008)
Thanks for your attention