a new dynamic pressure source for the calibration of

39
\ O? \ *) y NBS TECHNICAL NOTE 914 35 p HF $2.25; SOD HC $0.85 N76-28532 Unclas A New Dynamic Pressure Source for the Calibration of Pressure Transducers RECEIVED NASA sn rACiuit INPUT BRANCH

Upload: others

Post on 13-May-2022

6 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: A New Dynamic Pressure Source for the Calibration of

\O?

\*)

y NBS TECHNICAL NOTE 914

35 p HF $2.25; SOD HC $0.85

N76-28532

Unclas

A New DynamicPressure Source

for the Calibration ofPressure Transducers

RECEIVEDNASA sn rACiuit

INPUT BRANCH

Page 2: A New Dynamic Pressure Source for the Calibration of

NATIONAL BUREAU OF STANDARDS

The National Bureau of Standards' was established by an act of Congress March 3, 1901.The Bureau's overall goal is to strengthen and advance the Nation's science and technologyand facilitate their effective application for public benefit. To this end, the Bureau conductsresearch and provides: (1) a basis for the Nation's physical measurement system, (2) scientificand technological services for industry and government, (3) a technical basis for equity in trade,and (4) technical services to promote public safety. The Bureau consists of the Institute forBasic Standards, the Institute for Materials Research, the Institute for Applied Technology,the Institute for Computer Sciences and Technology, and the Office for Information Programs.

THE INSTITUTE FOR BASIC STANDARDS provides the central basis within the UnitedStates of a complete and consistent system of physical measurement; coordinates that systemwith measurement systems of other nations; and furnishes essential services leading to accurateand uniform physical measurements throughout the Nation's scientific community, industry,and commerce. The Institute consists of the Office of Measurement Services, the Office ofRadiation Measurement and the following Center and divisions:

Applied Mathematics — Electricity — Mechanics — Heat — Optical Physics — Centerfor Radiation Research: Nuclear Sciences; Applied Radiation — Laboratory Astrophysics *— Cryogenics2 — Electromagnetics2 — Time and Frequency2.

THE INSTITUTE FOR MATERIALS RESEARCH conducts materials research leading toimproved methods of measurement, standards, and data on the properties of well-characterizedmaterials needed by industry, commerce, educational institutions, and Government; providesadvisory and research services to other Government agencies; and develops, produces, anddistributes standard reference materials. The Institute consists of the Office of StandardReference Materials, the Office of Air and Water Measurement, and the following divisions:

Analytical Chemistry — Polymers — Metallurgy — Inorganic Materials — ReactorRadiation — Physical Chemistry.

THE INSTITUTE FOR APPLIED TECHNOLOGY provides technical services to promotethe use of available technology and to facilitate technological innovation in industry andGovernment; cooperates with public and private organizations leading to the development oftechnological standards (including mandatory safety standards), codes and methods of test;and provides technical advice and services to Government agencies upon request. The Insti-tute consists of the following divisions and Centers:

Standards Application and Analysis — Electronic Technology — Center for ConsumerProduct Technology: Product Systems Analysis; Product Engineering — Center for BuildingTechnology: Structures, Materials, and Life Safety; Building Environment; Technical Evalua-tion and Application — Center for Fire Research: Fire Science; Fire Safety Engineering.

THE INSTITUTE FOR COMPUTER SCIENCES AND TECHNOLOGY conducts researchand provides technical services designed to aid Government agencies in improving cost effec-tiveness in the conduct of their programs through the selection, acquisition, and effectiveutilization of automatic data processing equipment; and serves as the principal focus withinthe executive branch for the development of Federal standards for automatic data processingequipment, techniques, and computer languages. The Institute consists of the followingdivisions:

Computer Services — Systems and Software — Computer Systems Engineering — Informa-tion Technology.

THE OFFICE FOR INFORMATION PROGRAMS promotes optimum 'dissemination andaccessibility of scientific information generated within NBS and other agencies of the FederalGovernment; promotes the development of the National Standard Reference Data System anda system of information analysis centers dealing with the broader aspects of the NationalMeasurement System; provides appropriate services to ensure that the NBS staff has optimumaccessibility to the scientific information of the world. The Office consists of the followingorganizational units:

Office of Standard Reference Data — Office of Information Activities — Office of TechnicalPublications — Library — Office of International Relations — Office of InternationalStandards.

1 Headquarters and Laboratories at Gaithersburg, Maryland, unless otherwise noted; mailing addressWashington, D.C. 20234.

3 Located at Boulder, Colorado 80302.

Page 3: A New Dynamic Pressure Source for the Calibration of

A New Dynamic Pressure Source for theCalibration of Pressure Transducers

Carol F. VezzettiJohn S. HiltenJ. Franklin Mayo-Wells, andPaul S. Lederer

Electronic Technology DivisionInstitute for Applied TechnologyNational Bureau of StandardsWashington, D.C. 20234

Sponsored by:

NASA Langley Research CenterHampton, Va. 23365

U.S. DEPARTMENT OF COMMERCE, Elliot L Richardson, Secretary

Dr. Betsy Ancker-Johnson, Assistant Secretary for Science and Technology

NATIONAL BUREAU OF STANDARDS, Ernest Ambler, Ac//ng Director

Issued June 1976

Page 4: A New Dynamic Pressure Source for the Calibration of

National Bureau of Standards Technical Note 914Nat. Bur. Stand. (U.S.), Tech. Note 914, 35 pages (June 1976)

CODEN: NBTNAE

U.S. GOVERNMENT PRINTING OFFICEWASHINGTON: 1976

For sale by the Superintendent of Documents, U.S. Government Printing Office, Washington, D.C. 20402(Order by SD Catalog No. C13.46:914). '. Price 85 cents (Add 25 percent additional for other than U.S. mailing).

Page 5: A New Dynamic Pressure Source for the Calibration of

FOREWORD

The work described was performed as a task within the Electro-mechanical Transducer Project. This is a continuing project for thedevelopment of calibration and evaluation methods for electromechani-cal transducers, supported by a number of Government agencies. Thistask was funded by the National Aeronautics and Space Administrationthrough the Lang ley Research Center under order number L-88319.

Paul S. LedererActing ChiefComponents and Applications Section

Page 6: A New Dynamic Pressure Source for the Calibration of

A NEW DYNAMIC PRESSURE SOURCE FORTHE CALIBRATION OF PRESSURE TRANSDUCERS

CONTENTS

Page

1. Introduction 12. Experimental Development of Source 2

2.1 Theoretical Considerations . . 22.1.1 Pressure Amplitude 32.1.2 Natural Frequency 3

2.2 Preliminary Investigations 52.3 Background for Use of Steel Balls as Column Packing ... 62.4 Design and Description of Dynamic Pressure Source . . . . 6

2.A.I Column Design 72.4.2 Column Description 72.4.3 Tests to Determine Column Packing

and Source Repeatability . 83. Calibration 9

3.1 Calibration Procedure 93.2 Error Analysis 113.3 Results from the Calibration of Six Transducers 11

4. Conclusions 145. Recommendations 156. Acknowledgements 167. References 16

LIST OF TABLES

Table 1: Working Liquids Used in Preliminary Investigationsof Damping -17

Table 2: Reference Transducer E Output 18Table 3: Factors Contributing to Transducer Calibration Error . . 19Table 4: Transducer Characteristics 20Table 5: Transducer Experimental Calibration Data 21

LIST OF FIGURES

Figure 1: Columns and Filters of Various Geometries Used inPreliminary Experiments 22

Figure 2: The Dynamic Pressure Source in Use to Calibratea Pressure Transducer 23

Figure 3: Cross Section of Dynamic Pressure Source 24Figure 4: Oscilloscope Traces Showing Peak-to-Peak Output (mV)

from Reference Transducer E as a Function of Frequency(kHz) for Three Quantities of Balls, as Described inthe Text 25

i PAGE BLANK NOT FiL

Page 7: A New Dynamic Pressure Source for the Calibration of

Figure 5: Schematic of Apparatus Used to F i l l the Sourcewith Liquid Under Weak Vacuum .

Figure 6: Oscilloscope Traces Showing CorrespondingFrequency-Response Curves for Test Transducer(Left) and Reference Transducer E (Right) . . . . ,

Figure 7- Oscilloscope Traces Showing Frequency-ResponseCurves for the Undamped Source with TestTransducer A-l (Trace A), B (Trace B),C (Trace C) , and D (Trace D)-

Figure 8: Test Transducer B Output (mV rms) as a Functionof Frequency (Log Hz) with the Pressure Level HeldConstant, as Described in Text

Page

26

27

28

29

VI

Page 8: A New Dynamic Pressure Source for the Calibration of

A New Dynamic Pressure Source for theCalibration of Pressure Transducers

Carol F. Vezzetti, John S. Hilten,J. Franklin Mayo-Wells and Paul S. Lederer

A dynamic pressure source is described for producing sinusoidallyvarying pressures of up to 3^ kPa zero-to-peak, over the frequencyrange of approximately 50 Hz to 2 kHz. The source is intended forthe dynamic calibration of pressure transducers and consists of aliq u i d - f i l l e d cylindrical vessel, 11 cm in height, mounted uprighton the armature of a vibration exciter which is driven by an ampli-fied sinusoidally varying voltage. The transducer to be calibratedis mounted near the base of the thick-walled aluminum tube formingthe vessel so that the pressure-sensitive element is in contact withthe l i q u i d in the tube. A section of the tube is filled with smallsteel balls to damp the motion of the 10-St dimethyl siloxane work-ing fluid in order to extend the useful frequency range to higherfrequencies than would be provided by an undamped system.

The dynamic response of six transducers provided by the sponsorwas evaluated using the pressure sources; the results of these cali-brations are given.

Key words: calibration; dynamic; dynamic calibration; dynamicpressure; dynamic pressure source; 1i q u i d .column; pressure; pressuresource; pressure transducer; sinusoidal pressure; transducer.

1. INTRODUCTION

Increased use of electromechanical transducers for measure-ment of dynamic pressures in complex processes arising in diversefields has increased the importance of determining the dynamiccharacteristics of such transducers. Dynamic pressure-calibrationsystems now in existence do not meet all the amplitude or fre-quency range requirements of present technology.

Page 9: A New Dynamic Pressure Source for the Calibration of

In 1971 the National Bureau of Standards InstrunlentationApplications Section* developed a method for generating sinusoi-dally varying pressures constant in amplitude within +5% over afrequency range of a few Hertz to about 1 kHz [I]-*. In thecurrent work, this method has been modified and extended [2, >3] in both amplitude and frequency capability to result in asource of sinusoidally varying pressures of up to 3^ kPa zero-to-peak (about 5 psi zero-to-peak), over the frequency range ofapproximately 50 Hz to 2 kHz. The source is intended for thedynamic calibration of pressure transducers and consists of aliquid-filled cylindrical vessel, 11 cm in height, mounted up-right on the armature of a vibration exciter which is driven byan amplified sinusoidally varying voltage. The transducer tobe calibrated is mounted near the base of the thick-walled alumi-num tube forming the vessel so that the pressure-sensitive elementis in contact with the liquid in the tube. A section of the tubeis filled with small steel balls to damp the motion of the 10-Stdimethyl siloxane working fluid in order to extend the useful fre-quency range to higher frequencies than would be provided by anundamped system. Both the natural frequency and the degree ofdamping of the combined 1iquid-column-and-transducer structuredetermine the useful upper frequency li m i t .

2. EXPERIMENTAL DEVELOPMENT OF SOURCE

2.1 Theoretical Considerations

For the geometry described, both the amplitude of the sinu-soidal ly varying pressure and the natural frequency of the l i q u i dcolumn may be calculated to first approximations by simple re-lations to permit comparison with experimental results.

* The Section is now the Components and Applications Section.**Figures in brackets indicate the literature references listed in

section 7.

Page 10: A New Dynamic Pressure Source for the Calibration of

Damping is also an important consideration because a dampedsystem may have a higher useful frequency range than the samesystem undamped. With an undamped, single-degree-of-freedomsystem, the amplitude-frequency response remains constant within+5% up to approximately 20% of the natural frequency. In a sys-tem with optimum damping (a damping ratio of about 0.6 of criti-cal), the amplitude-frequency response remains constant within+5% to about 80% of the natural frequency [3,4] . To maintaina flat response to 2 kHz thus requires an undamped system with anatural frequency of 10 kHz, an optimally damped system with anatural frequency of 2.5 kHz, or a non-optimally damped systemwith a natural frequency between 2.5 and 10 kHz, with the exactnatural frequency required dependent on the degree of damping.

2.1.1 Pressure Amplitude - The factors that in combinationdetermine the pressure levels attainable over.the frequency rangeof interest are (1) the force and displacement capabilities ofthe vibration exciter, (2) the density of available working fluids,(3) the height of the liquid column, and (k) the degree of damping.This last factor interacts with the others and depends on geometry,on the bulk modulus and viscosity of the fluid. As noted above,damping need not be considered at frequencies below about 20% ofthe system natural frequency because the amplitude-frequency re-sponse is flat up to this limit. Up to the stated limit, theamplitude of the sinusoidal pressure generated within the liquidcolumn and acting on the transducer is given by [1],

ahcp

where

P = pressure (Pa zero-to-peak), j.a = acceleration amplitude (gn zero-to-peak ),hc= liquid-column height above the center of the

transducer diaphragm (m), andp = density of liquid (kg/m3).

As an example, for the source to be described in this paper, hc =O.lm and p = 972 kg/m3. At an acceleration of 36 gn zero-to-peak,the pressure is calculated to be 3^ kPa zero-to-peak, and at 20 gnzero-to-peak, 19 kPa.

2.1.2 Natural Frequency - The natural frequency of a liquidcolumn is directly proportional to the square root of the velocityof sound in the 1iquid and inversely proportional to the height of

*The symbol gn represents the unit of acceleration equal to thestandard value of the acceleration of gravity at the earth's surface,

Page 11: A New Dynamic Pressure Source for the Calibration of

the column. For an infinitely stiff, open-topped vessel containingthe liquid, the system has one degree of freedom with a natural fre-quency given by [1]

f = 0 21 — ( — )%l w * £>J ^~ \ ^^ / •n n o

where

fn = natural frequency (Hz),h = total height of liquid column (m), andB = bulk modulus (Pa).

The quotient of B and p is proportional to the velocity of sound inthe liquid. For the dynamic pressure source, h = 0.11 m and B =9.̂ 7 x 1Q8 Pa, the value given by the supplier for 10-St dimethylsiloxane. The undamped natural frequency is calculated to be 2.2kkHz.

In practice the tube is not infinitely stiff, but has someelasticity. For other than thin-walled vessels, a volume change inthe tube itself is usually small enough compared to other volumechanges that it may be neglected for practical calibrations. Moreimportantly, the transducer diaphragm forms part of the tube wall,and since this diaphragm is displaced in response to pressure, anoverall volume change occurs. The effect of this change is tolower the natural frequency as the effective bulk modulus is reduced.The effective bulk modulus may be defined by the relation [1]

B'

where

B1 = effective bulk modulus (Pa),V = volume of liquid in the tube (m3), anddVt = change of volume resulting from unit pressure change

(m3/Pa).

As an example, transducer B introduces a volume change of 9-11* x 10~15

m3/Pa, as calculated from information supplied by the manufacturer.With this transducer installed in the source, the effective bulkmodulus is calculated to be 7.03 x 108 Pa and the natural frequencyis lowered to 1.93 kHz.

In the source, steel balls are used for damping and the volumeof the liquid is reduced by an amount equal to the total volume ofthe balls, although the column height remains unchanged. As an ex-ample, consider that 15% of the volume of the liquid is replaced bythe balls (as was done in some of the experiments). The effective

ORIGINAL PAGE IS POOi

Page 12: A New Dynamic Pressure Source for the Calibration of

bulk modulus would then be given by 6.73 x 108 Pa, and the undampednatural frequency would become 1.89 kHz. If the balls were to occupy50% of the available volume, the calculated natural frequency wouldbe 1.72 kHz, which is significantly lower than the 2-kHz goal.

Combining the results of the above discussion, the natural fre-quency is given by

f =n

0. 25i

h

B(V - Vuc b

P ( Vc ' Vb

)

+ BdVt)

where

Vc = volume available for li q u i d with no balls present (m3),Vb = volume occupied by balls (m3) , and

h, B, p, and dV are as given above.

2.2 Preliminary Investigations

Preliminary experiments described in the progress reports [ 3 ]were conducted to investigate approaches for improving the frequencyrange of the method developed earlier. These experiments involveda variety of working liquids, including water, tetrabromoethane,petroleum oiIs, a fluorocarbon li q u i d , dimethyl siloxane liquid ofvarious viscosities, glycerine, and mercury. These candidate liquidswere chosen because of specific physical properties given in table 1and on the basis of experience. Chemical reactivity and the criterionof ready availability also governed the choice of liquids. For example,water was chosen because it has a relatively high bulk modulus and be-cause its use in the earlier work permitted intercomparison of experi-mental results. Tetrabromoethane, the fluorocarbon liquid, and es-pecially mercury have high densities and therefore should confer high-pressure capability with short columns of liquid. Mercury has beenused in related work with a closed-tube dynamic pressure source* [5],Dimethyl siloxane liquids and petroleum oils are available in a widerange of viscosities; glycerine possesses a combination of high vis-cosity and high bulk modulus.

The damping provided by a number of column geometries was alsoinvestigated experimentally with the various liquids, as shown intable 1. Among these experiments, which were designed to increasethe available wetted surface compared to that of the interior of asimple tube, were trials with columns of various cross-sections (in-cluding columns with vertical fins, spiral fins, and multiple channels)and with columns packed with various materials (including bundles ofsmal1-diameter tubes, sintered metal filters, and sea sand). A se-lection of columns is shown in figure 1. Damping of a magnetic l i q u i dby means of a coil electromagnet wound around the outside of the columnwas also attempted.

5

Page 13: A New Dynamic Pressure Source for the Calibration of

The results of these investigations tended to single out 10-Stdimethyl siloxane as the li q u i d possessing the best combination ofproperties, including handling quality, of those liquids tried. Theprincipal reasons for rejecting the other liquids may be summarizedas follows: water and tetrabromoethane -- low viscosity precludesachievement of effective damping; petroleum oils -- viscosity ishighly temperature dependent; fluorocarbon liquid — bulk modulustoo low; glycerine -- presence of dissolved gas (which could not beremoved reliably by the vacuum f i l l i n g procedures described in anearlier report [3} reduced bulk modulus to too low a value; andmercury— viscosity too low.

2.3 Background for Use of Steel Balls as Column Packing

At an early stage of the work, the use of a material such assea sand had been proposed as a means of increasing the column wettedsurface, and hence the degree of damping, by a very large factor.However, experimental difficulties and inconclusive results discour-aged further work along these lines at the' time. The comparativesuccess achieved with a column packed with small tubes (inside diam-eter 0.16 cm) and filled with 10-St dimethyl siloxane encouraged asearch for other geometries offering s t i l l greater wetted surface(without the insuperable f i l l i n g problems .which had precluded theuse of tubes substantially smaller than 0.16 cm i.d.). The use ofsmall objects as a column packing was again considered, with the re-quirement that the objects not absorb or be soluble in the dimethylsiloxane. If suitable objects could be found in a selection of sizes,a range of wetted areas would be available for experiment. Thepolished steel balls produced for use in ball bearings are readilyavailable in a range of sizes and do not interact with dimethyl silox-ane, and were therefore tried as a column packing. These trials (de-scribed in 2) were successful in that the project goals for the sourcewere met (sinusoidally varying pressures of at least 3k kPa zero-to-peak, flat to within +5% from approximately 50 Hz to 2 kHz), and de-velopment of the source was considered complete for the intended pur-poses .

2.A Design and Description of Dynamic Pressure Source

An overall view of the dynamic pressure source and associated 'instrumentation is shown in figure 2. Mounted on the armature of thevibration exciter (C) is the source column (A) with the reference"transducer (B) screwed into the base. The position for the transducerto be calibrated is on the opposite side of the column from (B) and isnot shown. The controller (D) for the vibration exciter supplies thedriving signal and can be programmed to sweep over the frequency rangeof interest at constant acceleration, displacement, or velocity. Thevibration exciter system has a displacement capability of 1.3 cm peak-to-peak and can impart accelerations of 25 gn to an 11-kg mass atfrequencies of 30 Hz and above.

Page 14: A New Dynamic Pressure Source for the Calibration of

A digital voltmeter \E) may be switched to read the output ofthe test transducer, the reference transducer, or the accelerometermounted in the armature of the vibration exciter. Transducer excita-tion power supplies are denoted by (F) and (G), and (H) is the accel-erometer amplifier. Preissure transducer output as a function of fre-quency Is displayed on the oscilloscope (1), which is equipped with arecording camera.

2.k.1 Column Design - In order to accommodate the ball packing,the column required modification with respect to earlier columns used.The following considerations governed the design: (1) the balls mustnot be allowed to contact and thus damage the transducer diaphragm,(2) the balls must not be permitted to move in relation to one anotheror to the column to any significant degree (ball movement would resultin uncontroiled changes in damping), (3) a method for convenientlyvarying the ball loading (ball size and number) is required, and (k)the amount of liquid displaced by the balls should not be so great asto lower the natural frequency below acceptable limits, that is, asshown in 2.1.2, the balls should occupy less than half of the avail-able volume.

The design that was selected uti1ized pierced circular platesto clamp the balls in a section of the bore near the bottom of thecolumn. The geometry of the plates was subject to the following con-siderations: (1) the plates should have the maximum practicableamount of open area so as not to impede the movement of the dimethylsiloxane liquid (thus not to change the damping characteristics orthe natural frequency appreciably), (2) the holes should be smallenough to retain the smallest balls envisaged, (3) the distance be-tween the holes should be less than the diameter of the smallest ballenvisaged (to prevent balls from blocking all the holes), and (4) theplates should be thick enough to provide adequate rigidity.

2.k.2 Column Description - The construction that was chosen tosatisfy the varying requirements is shown in a cross-sectional sketch,figure 3. Steel balls (K) are clamped tightly in place by two 0.48-cm-thick pierced steel plates (L). The outer diameter of the platesloosely fits the inner diameter of the aluminum column; forty-five0.13~cm-diameter holes drilled through each plate perpendicular to thefaces account for approximately one-third of the face area and providefor free movement of the liquid. The bottom plate rests on a machinedshoulder 1.6 cm above the bottom of the column bore. The top plateclamps down on the ball packing, the clamping force being transmittedto the plate from a vented, threaded plug (M) by means of a brass re-tainer tube (N) which Is a loose fit in the main bore (J). This tubehas an outside diameter of 1.6 cm and a O.l6-cm wall thickness. Theplug engages threads machined into an upper section of the bore. Thevolume of the column bore available for ball packing may be adjustedthrough the use of retainer tubes of various lengths combined with

Page 15: A New Dynamic Pressure Source for the Calibration of

various positions of the threaded plug. Threaded holes for mount-ing the reference transducer (B) and the transducer to be calibrated(0) are provided in two machine flats near the base of the column.The centerline of these holes is 1 cm above the bottom of the bore.The column is fastened to the vibration exciter armature by means offour machine screws passing through holes (not shown) in the columnflange.

2.̂ .3 Tests to Determine Column Packing and Source Repeatability -The output of reference transducer E was used to determine an accept-able column packing in terms of ball diameter and length of sectionpacked with balls. Balls with diameters of 0.2̂ , 0.28, 0.32, O.AO,and 0.1*8 cm were available for these tests with selected section lengthsof 2.0, 2.7, 3.0, 6.0, and 8.0 cm.

The procedure used was as follows. Transducer E was mounted onthe column and the controller for the vibration exciter set to sweepthe excitation frequency from 35 Hz to 3 kHz at an arbitrary constantacceleration. Three swept-frequency response curves of transducer Eoutput were generated on the oscilloscope screen and photographicallyrecorded: (1) with the column empty of liquid, plates, etc., (2) withthe column filled with 10-St dimethyl siloxane to the 11-cm level, and(3) with the column filled with dimethyl siloxane as in (2) and theplates, retainer tube, and plug in place (the plates separated by meansof a spacer tube, since no balls were used). The curve from run (1) isthe transverse vibrational response of transducer E. From the run (2)curve, the resonance frequency and damping characteristics of the basicsystem may be determined. Resonance was at 2.35 kHz, and the dampingratio approximately 0.03 of critical. The curve from run (3) shows aresonance frequency of 2.20 kHz and a damping ratio of 0.10 of criti-cal, which was considered to be acceptable.

The experimental value of resonance frequency from run (2) ishigher than the calculated value of 2.2*» kHz obtained in 2.1.2. Thedifference may result from several sources, including an inexact knowl-edge of the bulk modulus for the 10-St dimethyl siloxane used and im-precision in the determination of resonance frequency. It is alsopossible that the theory is deficient. Another factor, not consideredin the analysis of 2.1.2, that can affect the resonance frequency isthe presence in the li q u i d of absorbed gases and other impurities.The presence of gases always results in a lowered resonance frequency,and the presence of impurities usually does.

Other swept-frequency response curves of transducer E outputwere then generated and recorded with various sized balls and varioussection heights. These tests showed that damping could be varied overa wide range (from very under damped to very over damped) and that,regardless of ball size, the resonance frequency is lowered signifi-cantly when either 6.0- or 8.0-cm section lengths are used. The probablereason for this lowered resonance frequency is that with these quantitiesof balls the volume available for liquid (filled to the 11-cm level) is

8

Page 16: A New Dynamic Pressure Source for the Calibration of

reduced enough to produce a significant reduction in the value of theeffective bulk modulus, as discussed In 2.1.2. .

Examples of the results obtained using 0.24-cm balls are shownin figure 4. With a ball section 3-0 cm in length, the pressure am-plitude is seen to decrease with increased frequency (top trace), in-dicating that the calibrator system is overdamped. With a 2.0-cmsection, the pressure amplitude is seen to rise to a maximum between2.1 and 2.A kHz (center trace), indicating that the column is under-damped (estimated to be 0.38 of. critical). The most nearly flat re-sponse was obtained with the 2.7~cm section, for which the pressureamplitude is seen to remain almost constant to approximately 2.4 kHz(bottom trace). The small resonances near 1.1 and 1.4 kHz are be-lieved to result from mechanical resonances of the column assembly,and data near these frequencies are not used.

As described in 3-3, reference transducer E was used in 16 cali-brations of other transducers. In particular, 30 measurements weremade with a constant acceleration of 20 gn at two frequencies, 35 and50 Hz, for which the output should be the same. 'Two computed valuesare also available. The data and statistics are given in table 2.The computed coefficient of variation of 1.1% is considered to demon-strate the acceptable repeatability of the source. Sources of bothsystematic and random errors are discussed in 3-2.

3- CALIBRATION

3.1 Calibration Procedure

Before calibration of a pressure transducer can be carried outusing the source, the size of the contribution to zero shift from thetransducer transverse acceleration response is checked. The transduceris mounted on the empty column, and the vibration exciter driven at theintended test level, usually 20 gn zero-to-peak. For all transducerstested in the study, the transverse response was not significant, i.e.,contributed to less than 0.02% of the transducer response.

For the calibration itself, the column must be loaded with ballsand filled. The procedure is as follows:

1. Mount the transducer to be calibrated and the referencetransducer (or, if not used, a plug) in the holes providedat the base of the column.

2. Position the restraining plates, steel balls, and retainertube in the column.

3. Place the column in a suitable bell jar, as shown in figure5. The bell jar should have a tube passing through its topand be equipped with a ball valve (0_).

Page 17: A New Dynamic Pressure Source for the Calibration of

4. Attach a suitable flask to the ball valve with the upperopening unstpppered. One type of suitable flask is aleveling bulb. The flask should have a volume of at leasttwice the volume of the empty column.

5. Close the valve.

6. Pour at least enough 10-St dimethyl siloxane into the flaskto f i l l the empty column.

7. Connect the bell jar lower port and the upper opening of theflask to a mechanical vacuum pump and operate the pump untilthe liquid in the flask does not bubble (fifteen minutes isusually sufficient).

8. Open the valve until the column is filled approximately tothe desired level; then close the valve.

9. Slowly release the vacuum in the bell jar.

10. Remove the column from the jar.

11. Mount the column on the armature of the vibration exciter.

12. Use a suitable depth micrometer to measure the liquid level.The lowest setting of the micrometer at which a drop ofliquid adheres to the end of the micrometer probe is takenas the distance between the liquid surface and the top ofthe column.

13. Adjust the level by adding or removing a few drops with asuitable instrument, such as an eye dropper, until the de-sired level is attained.

14. Screw the threaded plug into the fixture and torque to therecommended value of from 16 to 19 N-m (12 to 14 pound-force•ft). This completes the f i l l i n g procedure.

Following f i l l i n g , the controller of the Vibration exciter is set tosweep the frequency of the driving signal from 35 Hz to 3 kHz at agiven acceleration, usually 20 gn zero-to-peak.

An acceleration of 36 gn zero-to-peak is required to generatepressures of 34 kPa zero-to-Peak with a 10-cm column of 10-St dimethylsiloxane. However, as 36 gn zero-to-peak is close to the upper acceler-ation level capability of the vibration exciter with the column as load,and operation of the exciter at that level resulted in occasional over-travel alarms,.the majority of the calibrations reported in this paperwere made using an acceleration of 20 gn zero-to-peak, resulting inpressures of 19 kPa zero-to-peak over the frequency range of 35 Hz to3 kHz. One test was run at an acceleration of 36 gn zero-to-peak overa frequency range of 50 Hz to 3 kHz to demonstrate the higher pressurecapability of the method. In this test, the lower frequency limit of50 Hz was imposed by the displacement limits of the armature of thevibration exciter.

10

Page 18: A New Dynamic Pressure Source for the Calibration of

During a dynamic calibration run, readings on the digitalvoltmeter of the transducer outputs are recorded at selected f re-•quencies, and curves of transducer response vs frequency are re-corded by photographing the oscilloscope display of the test (andreference, if desired) transducer outputs as the fixture is vibratedat a constant acceleration over the frequency range of interest.

Static calibrations of d-c responding transducers are made withthe transducer mounted in the liquid-filled fixture by imposing knownpneumatic pressures on the top of the l i q u i d column and measuring theresulting transducer output with the digital voltmeter. A liquid-head correction is made. Static pressures are produced with a pre-cision bellows and read with a dial gage.

3.2 Error Analysis

The dynamic pressure source provides capabilities not beforeavailable. There is no "standard transducer" that can be used toevaluate the source performance. The best that can be done is tocarry out repetitive measurements on a transducer for which theorypredicts that the frequency response should be flat beyond the fre-quency range of interest. Such a series of measurements was carriedout with the piezoelectric quartz crystal reference transducer E;the results are given in 2.^.3; Because there is no "standard trans-ducer," it is instructive to analyze the potential sources of error.

Factors contributing to the uncertainty of the pressure avail-able from the source include (1) uncertainty in the height of theliquid column resulting from meniscus effects and from machining in-accuracies, (2) uncertainty in the knowledge of the true density ofthe liquid resulting from variations in temperature, (3) uncertaintyin the acceleration produced by the vibration exciter and measuredby its accelerometer, (k) uncertainties in various voltage measure-ments, and (5) electrical noise. For the dynamic calibration as awhole, uncertainty in the amount of transducer output resulting fromtransverse vibrational response, additional voltage-measurement un-certainties, and additional electrical noise should be added to theuncertainties listed above. For static calibrations, uncertainty inthe knowledge of the additional pressure applied to the liquid inthe column must be considered.

Estimated values for these uncertainties are given in table 3.The overall estimated uncertainty for dynamic calibration measurementsis ±k.]% of the true value and for static calibration measurements+0.4% of the true value.

3-3 Results from the Calibration of Six Transducers

Six pressure transducers were supplied by the project sponsor fordynamic calibration using the dynamic pressure source. The character-

11

Page 19: A New Dynamic Pressure Source for the Calibration of

isties of these test transducers and of reference transducer E aregiven in table k. Three of the transducers (A1, A2, and A3) are ofthe same model and have a silicon diaphragm into which a Wheatstonebridge circuit has been diffused. The other three (B, C, and D) arebidirectional unbonded strain-gage types. Reference transducer E,used as a cpntrol in all tests, is a piezoelectric quartz crystaltransducer, as has been noted.

Three calibration runs each were made of transducers Al, A2, A3,C, and D using a 20 gn zero-to-peak acceleration level from 35 Hz to3 kHz. Transducer B was calibrated in only one run (calibration No. 10)and at a level of only 10 gn zero-to-peak. The reason for the changein procedure was that transducer B was accidentally overranged inpreliminary resonance tests and damaged as a result. After the damagehad occurred, the output waveform of the transducer was distorted whenfull-scale sinusoidal pressure was applied. Calibration at approximatelyone-half the full-scale pressure was attempted and resulted in good wave-form; therefore, calibration was completed at that level.* The testtransducer and the reference transducer E outputs were monitored on anoscilloscope whose horizontal deflection system was driven by a swept-frequency sine-wave generator from 35 Hz to 3 kHz. In addition, theoutputs were monitored with a digital voltmeter at 35, 50, 75, 100, 200,300, and AOO Hz for all calibrations and at 500 Hz and 1, 1.5, 2, and2.5 kHz for selected calibrations. Reference transducer E was includedas a control in all 16 of the calibration runs.

The frequency-response curve for transducer Al, shown in figure 6A,is different from that of transducer E and shows evidence of overdamp-ing. The curves for transducers A2 and A3 are similar. The manufacturerwas asked to supply details of the design. It was reported that thismodel of transducer has a protective diaphragm 0.015-cm thick in frontof the sensing diaphragm, with an air space of 0.015 to 0.033 cm betweenthe two diaphragms. Small holes, 0.013 to 0.015 cm in diameter, aroundthe periphery of the protective diaphragm admit the pressure to thesensing diaphragm. It seems likely that this arrangement significantlyaffects the damping characteristics of the transducer.

For a given transducer, it is expected that all measurements madeat 35 Hz and 50 Hz would agree. For the three transducers Al, A2, andA3, the mean values of the six measurements (two measurements in eachof the three calibration runs for each transducer) are 21.232, 21.483,-and 23.427 mV rms, respectively. The maximum deviations from these meansare 0.3%, 0.2%, and 0.7%, with sample standard deviations of 0.059, pi031,and 0.111 mV rms. Comparison of the static calibration output measurementwith the above mean values showed differences of 3-7%, 2.3%, and 2.k% fortransducers Al, A2, and A3, respectively. The data are given in table 5.

*The oscilloscope trace shown in figure 6B was recorded using an accelera-tion of 18 <7n zero-to-peak before the damage occurred. (18 gn gives full-scale pressure variations for this transducer.)

12

Page 20: A New Dynamic Pressure Source for the Calibration of

Figures 6B, 6C, and 6D are of the frequency response curvesfor transducers B, C, and D. The mean values of the six measure-ments made at 35 Hz'and 50 Hz for transducers C and D are 10.67and 3.556 mV rms, respectively. The maximum percent deviations fromthese means are O.k% and 0.8%, with sample standard deviations of0.027 and 0.015 mV rms. No statistics are given for transducer B,which was-calibrated in only one run. Comparison of the static cali-bration output measurement with the above mean values showed differ-ences of 2.6% and 2.k% for transducers C and D, respectively. Com-parison of the mean of the two measurements at 35 Hz and 50 Hz fortransducer B with the static calibration output measurement for thistransducer showed a difference of 0.8%. These reported differencesbetween static and low-frequency dynamic calibrations probably resultfrom the uncertainties involved in determining the height of theliquid column. It is not possible to perform a static calibration onreference transducer E, as this transducer has no response to staticpressure.

If the test transducer natural frequency is high, the undampedresonance frequency of the source-transducer combination w i l l not beappreciably affected: this is the case for transducer Al (and A2 andA3), as is shown by the frequency-response curve in figure 7A. Thevertical axis scale (transducer output) is arbitrary. If the testtransducer natural frequency is relatively low, as it is for trans-ducers B, C, and D (3.5 kHz, 5 kHz, and 8.5 kHz, respectively), thecalibrator system resonance w i l l be lowered. The frequency-responsecurves in figures 7B, 7C, and 7D show the resonance frequency of thesource-transducer combination as being about 1.3, '-7, and 2.2 kHz,respectively. These curves and values were obtained with the columnfilled with dimethyl siloxane only, balls and plates having been re-moved. For calibrations of transducers with low natural frequencies,not only is the resonance frequency lowered, but the degree of damp-ing is also altered; 0.28-, 0.32-, and 0.48-cm-diameter balls wererequired for transducers B, C, and D, respectively, when the lengthof the section of balls was held to 2.7 cm. (For transducer B, itappears that balls of about 0.44-cm diameter would have given a moreuniform response but were not available; the height of the section ofballs could have been changed but for the sake of consistency was not.)

Consideration of test transducer natural frequencies is useful inan evaluation of transducer dynamic performance as shown in curves suchas those of figure 6. Pressure transducers are commonly not used tomeasure pressures varying at frequencies beyond approximately 20% ofthe natural frequency of the instrument. For transducers B, C, and D,this 20% l i m i t is approximately 700 Hz, 1.0 kHz, and 1.7 kHz, respec-tively. As may be seen by examination, the response of transducers Cand D is reasonably flat up to their respective limits. The responseof transducer B up to its 20% l i m i t is somewhat less flat.

in a final experiment, the output of the reference transducerwas both recorded and used to provide an input signal for the control-ler for the vibration exciter with the power to the exciter continuouslyadjusted to maintain the reference transducer output at a constant level.

13

Page 21: A New Dynamic Pressure Source for the Calibration of

The output of the test transducer under these conditions was recorded.A plot is shown in figure 8 for transducer B. The curve suggests agree-ment with the manufacturer's specified natural frequency of '3.5 kHz andshould represent the frequency response of the transducer with no inter-pretation required. Further work is required to validate this technique.

*f. CONCLUSIONS

The useful frequency range of the liquid-column sinusoidal pres-sure calibration source has been increased considerably by damping thel i q u i d column. The use of steel balls with 10-St dimethyl siloxaneli q u i d to achieve this damping is simpler and more versatile thanother damping schemes considered. The result is a dynamic pressure,source capable of generating pressure levels of up to 3** kPa over afrequency range from approximately 50 Hz to 2 kHz, flat to within+5%. At frequencies below 50 Hz the pressure amplitude is limitedby the displacement capability of the vibration exciter used. Theestimated total calibration measurement error using the source witha typical transducer is +k.]%. . .

It is important to note that a flat frequency response can beachieved over a frequency range of up to a maximum of about 80% of thenatural frequency of the l i q u i d column-transducer combination, andthat this natural frequency is lower than that of the component withthe lowest natural frequency.

The.wetted surface area required to damp adequately a particularl i q u i d column and transducer combination is dependent on the naturalfrequency of the combination. Combinations with high natural frequenciesrequire larger wetted surface areas than those with low natural frequen-cies. The wetted surface area may be adjusted easily by changing ballsize, ball quantity, or both.

Over a frequency range of up to 20% of the resonance frequency .of the liquid-column transducer combination, the dynamic pressure sourceprovides an absolute calibration to within 5% of the true pressure sup-plied in the sense that this pressure is calculable on the basis ofmechanical parameters that can be measured and on the basis of 'knowl-edge of the acceleration imparted to the column, which acceleration canbe measured with instruments that may be traced in calibration to basicstandards. For the remainder of the frequency range, a reference trans-ducer with a flat response over that range is required both for settingand monitoring the degree of damping.

Another calibration technique using the source has been the s.ub-ject of several experiments and requires further development. In thistechnique, a signal derived from the reference-transducer output issupplied as a control input to the vibration-exciter control, and theacceleration of the column is continuously adjusted to maintain the pres-sure amplitude at a constant level as the frequency is swept over thedesired range. In principle, the only frequency 1imitat ions should be

14

Page 22: A New Dynamic Pressure Source for the Calibration of

imposed by the frequency responses of the reference transducer andthe exciter system.

5. RECOMMENDATIONS

The following recommendations for future work are based on theexperience gained during the development of the dynamic pressuresource and during the various calibrations.

1. A number of different high-natural-frequency, flush-diaphragmpressure transducers should be calibrated with the dynamic pressuresource to determine if the degree of damping is approximately constant.

2. The number and size of balls should be more widely variedand curves generated showing the relationship between these parametersand column resonance frequency and damping characteristics.

3. The effect of temperature changes 6n the degree of dampingshould be investigated.

A. Calibrations using the source should be attempted on trans-ducers with low natural frequencies, with a constant-level-controlservo used to maintain the reference transducer output constant withfrequency. This technique should be developed and validated.

5. The use of columns with larger bores to reduce the percentcontribution of dV^ (see 2.1.2) should be investigated. This recom-mendation is particularly aimed at calibrations of transducers withlarge dV^ values, usually transducers with low natural frequencies.

6. The use of short columns (with high resonance frequencies)should be investigated without damping over a frequency range of upto 2Q% of the resonance frequency of the column-transducer combination.

7. The presence of the retainer tube used to transmit forcefrom the threaded plug to the upper plate has been found to complicatemeasurement of the height of the liquid column. A modified designeliminating the retainer tube would be desirable. It has been sug-gested that the upper plate could itself be threaded into the bore.

<8. Any new columns should be fabricated without the flats

machined near the base into which the transducer mounting holes arebored. The presence of the slots formed by the flats reduces columnstiffness.

9. The discrepancy between the theoretical predictions fornatural frequency and the experimental values achieved requires fur-ther examination. Experiments with several fluids, dimensions, anddegrees of damping should provide data for further analysis.

15

Page 23: A New Dynamic Pressure Source for the Calibration of

6. ACKNOWLEDGEMENTS

Kurt Muhlberg designed and supervised construction of themechanical components of the various development versions of cali-brators .

7. REFERENCES

[1] Hilten, J. S., Lederer, P. S., and Sethian, J. D., A Simple Hy-draulic Sinusoidal Pressure Calibrator, NBS Tech. Note 720 (April1972).

[2] Lederer, P. S., Development of a Dynamic Pressure Calibration Tech-nique - A Progress Report for the Period June 15, 1973 to Septem-ber 15, 1973, NBSIR 73-290 (October 1973).(Available as COM 74-10974from the National Technical Information Services, Springfield, VA 22161.)

[3] Vezzetti, C. F. , Lederer, P. S., and Hilten, J. S., Development ofa Dynamic Pressure Calibration Technique - A Progress Report forthe Period February 15, 1971* to August 15, 197*», NBSIR 75-708(June 1975). (Available as COM 75-10817 from the National TechnicalInformation Services, Springfield, VA 22161.)

[4] Bradley, W., (and Eller, E. E., Introduction to Shock and VibrationMeasurements, Book, Shock and Vibration Handbook, Vol. 1, Eds.C. M. Harris and C. E. Crede, pp. 12-1 -- 12-2*» (McGraw-Hill BookCompany, New York, NY, 1961).

[5] Hilten, J. S. , Lederer, P. S., Vezzetti, C. F., and Mayo-Wells,J. F., Development of Dynamic Calibration Methods for Pogo Pres-sure Transducers, NBS Tech. Note (to be published in 1976).

16

Page 24: A New Dynamic Pressure Source for the Calibration of

TABLE 1

Working Liquids Used in Preliminary Investigations of Damping

Liquid Viscosity Density BulkModulus

(St) (kg/m3) (Pa)

Hater 0.009 1.00 x 103 20.9 x 10'

Tetrabromo- 0.080 2.95 x ID3 29.3 x 109

ethane

Fluorocarbon 0.026 1.88 x 103 9.8 x 109

Mercury 0.001 13.6 x 103 265 x 109

Velocityof

Sound(m/s)

1460

1007

730

1410

Advantages

High bulkmodulus, andvelocity ofsound;non-toxic;inexpensive

High density,non-toxic

High density;non-toxic

High density,bulk modulus,and velocityof sound

Disadvantages

Low viscosity,high vaporpressure;corrosive tosteel

Low viscosity;high vaporpressure;corrosive tosteel , aluminumand other metalsand rubber

Low viscosity.low bulk modulus

Low viscosity;dissolves mostmetals, healthhazard

ExperimentsColumn Description

I.D.(cm)

2.21.31.31.2

2.21.2

1.2

2.2

1.1

Height(cm)

35353535

35

35

3535

35

35

35

5

Material

BrassBrassCopperBrass

Copper

Copper

BrassBrass

Brass

• Copper

Brass

Stainlesssteel

Damping

Smooth wallSmooth wallSpiral finsWoven -metalmesh filterMultiplechannelsInterlacing

Smooth wallWoven-metalmesh filterSintered-metfilter

AmplitudeRatio*

1412158.0

9.0

fins **

5.0**

:al **

Multiple channels 3.7

Smooth wall

Smooth wall

9.0

5.0

Glycerine 7.50 1 . 2 6 x l 0 3 44.8 x ID9 1904 High viscos- Hygroscopicity, bulkmodulus, andvelocity ofsound; non-toxic

10 Brass Smooth wall 5.2

Petroleum 0.5 0.86 x 103 15.2 x 109 1340Oils 925°C

15G25°C

Dinethyl 600 0.97 x 103 9.28 x 109 987

300125

10

!

High viscos- Low density;ity and viscosityvelocity of temperaturesound dependent

High Low bulk modulusviscosity; and densitynon-toxic

2.22.2

2.2 .

2.41.10.50

2.42.42.4 '1.10.50

0.40

0.20

0.15

0.16

3535

353535

101010

1010101010

10

10

10

10

BrassBrass

CopperBrassCopper

BrassBrassAluminumand glassBrassBrassBrassBrassAluminumand glassAluminumand glassAluminumand glassAluminumand glassBrass

Smooth wall 16Woven- metal mesh 3.8fl.lterMultiple channels 4.5Smooth wall 12Multiple channels 1 .1

Smooth wall 13Smooth wall 8.0Smooth wall 3.3

Smooth wall 13.0Smooth wall 20Smooth wall 33Smooth wall 27Smooth wall 7.4

Smooth wall 2.8

Smooth wall 1.4

Smooth wall »*

Multi-tube Responsefixture fiat to

±5% from40 Hz to750 Hz

*For a given value of acceleration, amplitude ratio 1s here defined as the ratio of measured maximum amplitude (at resonance) to the average of theamplitudes measured over 35-50 Hz.

**0ata not obtainable.

17

Page 25: A New Dynamic Pressure Source for the Calibration of

CU.0E3Z

CO

Ca

ll b

raF

requ

ency

i—

r—

•—

r—

i—

*

• i—

OO

r*.

vo

"".

rslX

CO

r^

mCO

r~

§r*.

r.

oo

CM

CO

CM

CO

8r^

in

COin

-

m

c\jr*.r-

10

r-

o"

CO

VO

03

*

CO

inCO

COCO

r-.

8r—

C\Jcnr~

CMO

OO

,_

CO

CO0

00

CsJenr-~

£

<\jin

CO

00

10

r-.

U3r*.r~

0

CO

o00

OJ

CO

0if)

MCO

r~

8r~

Cn

t-

COoCO

ID

CO

r-.0

00

CMenr-.

£

,—

CO

CO

inCO

r-*

CSJCO

r^

0

CO

oCO

CMCM

CO

CD0

CM0

CO

CM0

CO

COoCO

CM

CO

r-.

CO

r-»CM

CO

CM

CO

cnCO

CM

COCO

in

inenr-.

r-*

i —

oCO

0

CO

roCM

CO

o0CM

O

CO

in

CO

(0

CO

r^CO

CO

CO

CO

CO

CO

CMr*.CO

oen

oen

COCO

£

0oCO

ooCO

CO

oo

en0

CO

CM

CO

OoCO

5CO

CM«3-

CO

in

CO

#CO

in i — co en ID

co co r-. CM co

inCO

CO

CM

cn

CM0

CMoCO

voov

0

CO

0

CO

o

CO

2 § $ § S §co co co en en co

r-

CO

COCM

CO

o .0 o o o oCD O O O CD O«a- in o m o LO

i — i — CM CM

CM 40

-O QJC -r-

fO 0

(/> £+-><*-o5 o<u U

H"toOl Chs"~.

°°.01 C

5-2+J

*o£</) QJO T3

+j-otn i.•r- 03 •

rtJ C C

t/l 4J CUtfl E

OJx: <D «+-J— •— o

• § ^x .«™

.c cn E

o^co£ »O4-> C 1

tJm cu «

E

ll1fl3 t*-

CJ (/> O

S- tf> .

cu o +

5 >OJ> Wl I —

4-> W O

^3~CU 'I- l-

C 4-»

« — 3 fd3 0- <JO CJ tn

(V C

C to T-

O.-C <d0) *>t- -M O>

"3 -O

£ g.1CT 3'xd) O US

£ uj ETJ

- -C QJ tO

•M T3 »«

.C C *~IW 2-

O O OO

O- W i—

US"•— c

_o > ••-

"3 -w t!O 3 (OS Q. >

4-» O 4-•-• (J 0

cE3

O

°

mE(J

1.—cno<4-0

51

(/)c0)-o

4->

3

0)C

X

(/)

4-> Ci.

E 0•r- +J•o i

o4-> i.CO CU

1 M

' — C

. °*•o o<U CM(/)3 *

•o o3 4->

<U.. o

s«81

roWro •»

4-> 01

S" £U J=X Q.<u ra*-o

§ £•— w< — 3o -oM- y>

Ct/> (O

4->

S_ (4-QJ O

S_4-> <UC 4->

E O)

mea

sure

e th

e c

M- 0

toI/)

|B4*30

T3

O •»->O .C

cnCD -i-

f-2

•— O

r- ^£ 4J

18

Page 26: A New Dynamic Pressure Source for the Calibration of

TABLE 3

Factors Contributing to Transducer Calibration Error a

Factor Systematic Error Random Error(% of measured value) (% of measured value)

DYNAMIC

Liquid Column

Height of liquid column ±0.1 ±1.0

Variation of liquid density with temperature - ±0.2

Acceleration Applied

Calibration of exciter control accelerometer ±1.0 ±0.1 °with reference accelerometer and precision ofreading

Transducer Output

Voltmeter calibration of precision of reading ±0.3 ±0.05 e

fNoise - ±0.1

Transverse acceleration response of transducer ±0.02 3 —

TOTAL RMS ±1.05 ±1.03

Estimated Error = RMS systematic + 3 RMS random = ±4.14%

STATIC

Pressure Applied

Pressure gage calibration and precision of reading ±0.07 ±0.06 *"

Transducer Output•7 kVoltmeter calibration and precision of reading ±0.07 J ±0.03

Noise - ±0.07 l

TOTAL RMS ±0.099 ±0.097

Estimated Error = RMS systematic + 3 RMS random = ±0.39%

a Typical calibration values for the source are acceleration, 20 gn zero-to-peak and liquid columnheight above centerline of transducer, 10 cm using 10-St dimethyl siloxane with a bulk modulus of9.47 x 108 m3/Pa.

b Estimate based on manufacturer's value of 0.098%/°C and variation of ±2°C.0 Estimate based on least count of ±1 mV for measurement of 1000 mV.

Manufacturer's value for range of 20 mV.e Estimate based on least count of ±0.01 mV for measurement of 20 mV.

^ This figure corresponds to ±0.02 mV for a typical calibration.9 This value represents a maximum value for any of the transducers listed in table 4.

Manufacturer's value for range of 17 kPa (5 psi).

^ Estimate based on least count of ±0.2 kPa (±0.03 psi) for measurement of 17 kPa.3 Manufacturer's value for range of 30 mV.T,

Estimate based on least count of ±0.1 mV for measurement of 30 mV.

This figure corresponds to ±0.02 mV for a typical calibration.

19

Page 27: A New Dynamic Pressure Source for the Calibration of

TABLE 4

Transducer Characteristics*

Transducer

A-l,A-2,A-3

B

C

D

E(Reference)

Type

Silicondiaphragmwith bondedWheatstonebridge

Unbondedstraingage

Unbondedstraingage

Unbondedstraingage

Piezo-electricquartzcrystal

Range(kPa zero-to peak)

34

17

34

102

54,400

NaturalFrequency

(kHz)

45

3.5

5.0

8.5

100

DiaphragmDiameter

(cm)

0.22

1.26

1.26

1.26

0.953

Full -RangeDeflection of

Diaphragmat Center**

(cm)

2.5 x 10-"

3.8 x 10-1*

3.8 x ID'1*

3.8 x 10-"

Data notavailablefrommanufacturer

*From manufacturer's data.

**The manufacturers specify these deflections as maximum values.

20

Page 28: A New Dynamic Pressure Source for the Calibration of

s

r*. o F— in o m-̂ <± -̂ »— CM Olin in in in m ^

O <j- ^ rn CM3 in in in in inT3 *— . . . .tit po co co co ro fo o

en o r- o in o*m in m tn m v

CM 4O CO - r- CM CM\o «o r** to so 10o o o o o o

•pi-; 0 0 0 0 0 * 0 0 * . ' - ^ ,"r— i— •— r- i— t— F— »— f— i— O

CO Ul fO f- CO ^

o" o d o o" CD

O in co

*j- ^»- in m ^ to

CM CM CM CM CM CM

CM «3 <3- O «O CMU CO CM PO ^ CM i—

(U fc 3 COu C ~o3 I / ) C M C M C M C M C M C M

^^— h.

in in in in 4- ro

CM CM CM CM . CM CM

S i— \o m *p o*iri in m en CM

CM CM CM CM CM CM

•i__ m <T> in<u - --u m

T3 CM CM CM CM CM CM

S ^» CO C* ^~ CO10 in in <f rO

CM CM CM CM CM CM

i - ^ t - ^ » — « - ^ o ' d o o oC M C M C M C M C M C M C M f —

•iVO tO 00 CMCM »- O O

U CM . . . .

CM CM CM CM CM CM

CM CO CM r- CO \O

r— •— r— r- O OCM CM CM CM CM CM

*— »— O

O •—

O •—

^ — o

in a*in CM^- ocs t—

r~ O

o •—

o» i/></l«^^ C ^^

U) O O O OPO m o o o

r- CM CO

O O CO Os s § a-̂ r- CM CM

f- U

51C C

to o

-O -M

0|

§4:•*- trt

5o;

CD ON

O- ID

-*9 O

4-> V) CUII $j•*- 4->

•Su Tg

(A c/1

£ §1

-g

S

^ E f 3

g «J

21

Page 29: A New Dynamic Pressure Source for the Calibration of

4-> Ic — o) X-N0) 4J XI <TvE — 3 >-"

— 3 t->t- E m(I) 0) uQ. .* U 0)Xx-^ O 4J0) CM X) —

u in 4->(TJ c O •—C — O TO

— «4- E 4->E </> <l>

— •— E— (0 •- I0) C x-. T)u O vO 0)Q. O)~-^ »-

ID 0)C — "O 4->

— T> C C(0 ._

-ax: mQJ 4-> -in — .—* T33 2 LA C

^^ fOin a)

u 4-1 -a- oo<u .-E -—- in 4JO ?- 0) 0)d) •—' xiO) 3V-

m 4-1 a)in c 4->3 .- <u —O <4- »- —

O M-l- —• XI(D <D I XI> i- x: in

— 4-1 0)>4- D. o EO <n O i

in x: in (DI- 4-> 4->(U .- . -0)4J 2— E— ro i— 0)^-' C14- XI 0)

3 <U >-O 4-> XI OC 32(D tt> 4-> .

V- ^.^in ID — 4-> oC <U — i-E c c 2-^3 2 C

— O ID --^ T3O .̂ .̂ t~~- C.

O 00 O —' (D

I!CD

22

Page 30: A New Dynamic Pressure Source for the Calibration of

0) 0)U 4J3 —-O Oin X

0)

— ' OT3 3u o) -a<u in </>4-> 3 C«)E

0)ro — 4->

O>

0) 4J — l_ 4JI- (T3 (D dj3 »- 4J — 0)

O) O »—— oT3 (D

in —0) >uo. -ro oo u

0)(U 4-1

(D 0>U U

-D 3— T3p— </>(D CU 03

0)a.OUin

0) OCo — —

(/> C3 0)

C 0)

0}0)ou3

a) ^^u S.in(/> 0)a) uu t-O- 3

OO U)

c ro

(U.C V)

0> — U>- O. in.0 E O— to> -D

• C

O 13

u. u •*~*.*~* (U C)o -MO

0) U 3

— — D C / )O 0-in o. c <L)C 3 O -CO in — 4JU 4->

I- <X> TJI- 0) U C

— Q. > 0)O jQ

4J 0) O CC U 4-i V0 3 ^ - 0O "O C "O

• c E JE-̂« 10o u o in

CM

V

301

Z3REPRODUCIBILITY OF THEORIGINAL PAGE IS POOR

Page 31: A New Dynamic Pressure Source for the Calibration of

Figure 3: Cross section of dynamic pressure source. Shown are referencetransducer port (B), liquid-column chamber with 1.7-cm-diameterbore (J) , steel balls (K) , pierced steel plates (L), threadedplug (M) with vent, retainer tube (N), and test transducerport (0).

24

Page 32: A New Dynamic Pressure Source for the Calibration of

^^^^wwn ̂ ^^^w wwwww ••••• w^^^mw ̂ ^^^^^H••••••••• mm^mm mmmmm mmmmm m^mmm BM^^BM iI •*••• mm^mm mmtmmft ••••• MBMMB Ba^^^B i

iSSSS5liil

B••••••••9SSSBB!InI

Figure 4: Oscilloscope traces showing peak-to-peak output (mV) fromreference transducer E as a function of frequency (kHz) forthree quantities of balls, as described in the text. Thetop trace shows the source-transducer combination as beingoverdamped; the center trace, underdamped; and the bottomtrace, near optimum damping. The vertical scale is 20 mVpeak-to-peak per division, and the horizontal scale is 0.3kHz per division. The left-hand edge of each trace beginsat 35 Hz; the sweep range is 3 kHz.

25

Page 33: A New Dynamic Pressure Source for the Calibration of

Figure 5: Schematic of apparatus used to f i l l the source with liquidunder a weak vacuum. Q. is a ball valve in the line betweenbell jar and flask. Line R is connected to a source of vacuum.

26REPRODUCIBILITY OF THECEI0INAL PAGE IS POOR

Page 34: A New Dynamic Pressure Source for the Calibration of

Figure 6: Oscil'oscope traces showing corresponding frequency-responsecurves for test transducer (left) and reference transducer E(right). Of interest is the frequency range over whichtransducer output is at a relatively constant level. Trace A:test transducer A-1; vertical scale for both transducers is20 mV, peak-to-peak, per division. Trace B: test transducerB; vertical scale for both transducers is 10 mV, peak-to-peak,per division. Trace C: test transducer C; vertical scale forboth transducers is 10 mV, peak-to-peak, per division. TraceD: test transducer D; vertical scale for D is 2 mV, peak-to-peak,per division and for E, 10 mV, peak-to-peak, per division. Thehorizontal scale for all traces is 0.3 kHz per division.

27

Page 35: A New Dynamic Pressure Source for the Calibration of

.C 0)4-> 4-1~ C

4) »4-O O

DOin

incroi_4-> C

O4) ._U inC i— •0) > N

--^ 4> TJ

-o0) 0)O- OE m

4> i- 00u v

O- in

cD

0).c

O ^to

Cmi-

in4) 4)> O

O Oin a.

in 4)^ .- 4)TO 5u 4) in4-1 ID 4)«- o :̂4) in 4->

O 4-> 4) (0 N

4)I/) UCo -Q-^Jl CD4)U 4)I U>~ (0O 1-C -M4) •—

O u~»N CO

in L. 4->4) O <0

UC 0)0) -CDcr4)

•— C. (0 3

014)

a>c 4)

•— oS (DO ^>

-C -Min •—in r-0) iO <IDi- i-4-> 4)

U0) 3Q.-0O ino cI/I (DO i-

O inin 4)O -i-1

3 >-4J U _C(D ID UC I- (D

4-> 4)4) •—> -Q H-

— u o4J IDU 4)D C D)O..— -oin 4)4) 4J1- 3 T3

a. c4) 4J (D^: 3 _c4-1 O I

0)

4)4-1 U 0)in 3 _c4) X) I—

CT,

Page 36: A New Dynamic Pressure Source for the Calibration of

o

IM

>-O

c0)</>0)

<u 1.-C Q.

4J -a

'2 "3o

3: '4->

o> oo —

— CL

u _cc H-0)

CT0)

Xa)

o a).c

C 4-)O

— C4-1 .—OC T33 (0

M-

ro

(1)O

o( / ) ( / ><D 0)

•a(/I IPe TO

- o> 4JE C (U

4-> C

4-> in O3 C Q.0- O in4-1 O <\)3 1-O -a

co a) o_c c

1- <u0) — 3o a) cr3 > <U

"O <D l-in •— M-cro a) a>1-1-3

4 J 3 1 -in 4J

4-> inin (U (U0) i- -C

I— Q_ 4-1

CO

(U

3O)

AW) indj.no a «3onasi\ivai

Page 37: A New Dynamic Pressure Source for the Calibration of

NBS-114A (REV. 7-73)

U.S. DEPT. OF COMM.

BIBLIOGRAPHIC DATASHEET ,

1. PUBLICATION OR REPORT NO.

BS Technical Note 914

2. Gov't AccessionNo.

3. Recipient's Accession No.

4. TITLE AND SUBTITLE

A New Dynamic Pressure Source for the Calibration ofPressure Transducers

5. Publication Date

June 19766. Performing Organization Code

7. AUTHOR(S) Carol F. Vezzetti, John S. Hi Hen,J. Franklin Mavo-Nells. and Paul S. Lederer

8. Performing Organ. Report No.

9. PERFORMING ORGANIZATION,NAME AND ADDRESS

NATIONAL BUREAU OF STANDARDSDEPARTMENT OF COMMERCEWASHINGTON, D.C. 20234

10. Project/Task/Work Unit No.

425345611. Contract/Grant No.

NASA Order L-8831912. Sponsoring Organization Name and Complete Address (Street, City, State, ZIP)

NASA Langley Research CenterHampton, VA 23365

13. Type of Report & PeriodCovered

Final14. Sponsoring Agency Code

15. SUPPLEMENTARY NOTES

16. ABSTRACT (A 200-word or less [actual summary of most significant information. If document includes a significantbibliography or literature survey, mention it here.)

A dynamic pressure source is described for producing sinusoidally varyingpressures of up to 34 kPa zero-to-peak, over the frequency range of approximately50 Hz to 2 kHz. The source is intended for the dynamic calibration of pressuretransducers and consists of a liquid-filled cylindrical vessel, 11 cm in height,mounted upright on the armature of a vibration exciter which is driven by anamplified sinusoidally varying voltage. The transducer to be calibrated ismounted near the base of the thick-walled aluminum tube forming the vessel sothat the pressure-sensitive element is in contact with the liquid in the tube.A section of the tube is filled with small steel balls to damp the motion of the10-St dimethyl siloxane working fluid in order to extend the useful frequencyrange to higher frequencies than would be provided by an undamped system.

The dynamic response of six transducers provided by the sponsor wasevaluated using the pressure sources; the results of these calibrations aregiven.

17. KEY WORDS (six to twelve entries; alphabetical order; capitalize only the first letter of the first key word .unless a proper

name,-separated &y semicolons; Calibration; dynamic; dynamic calibration; dynamic pressure;dynamic pressure source; liquid column; pressure; pressure source; pressuretransducer; sinusoidal pressure; transducer.

18. AVAILABILITY pT] Unlimited

Q^] For Official Distribution. Do Not Release to NTIS

2D Order From Sup. of Doc., U.S. Government Printing OfficeWashington, D.C. 20402, SD Cat. No. C13 . 46:914

L7 |̂ Order From National Technical Information Service (NTIS)Springfield, Virginia 22151

19. SECURITY CLASS(THIS REPORT)

UNCLASSIFIED

20. SECURITY CLASS(THIS PAGE)

UNCLASSIFIED

21. NO. OF PAGES

3522. Price

$0.85

USCOMM-DC 29042-P74

Page 38: A New Dynamic Pressure Source for the Calibration of

NBS TECHNICAL PUBLICATIONS

PERIODICALS

JOURNAL OF RESEARCH reports National Bureauof Standards research and development in physics,mathematics, and chemistry. It is published in two sec-tions, available separately:

• Physics and Chemistry (Section A)Papers of interest primarily to scientists working inthese fields. This section covers a broad range of physi-cal and chemical research, with major emphasis onstandards of physical measurement, fundamental con-stants, and properties of matter. Issued six times ayear. Annual subscription: Domestic, $17.00; Foreign,$21.25.

• Mathematical Sciences (Section B)Studies and compilations designed mainly for the math-ematician and theoretical physicist. Topics in mathe-matical statistics, theory of experiment design, numeri-cal analysis, theoretical physics and chemistry, logicaldesign and programming of computers and computersystems. Short numerical tables. Issued quarterly. An-nual subscription: Domestic, $9.00; Foreign, $11.25.

DIMENSIONS/NBS (formerly Technical News Bul-letin)—This monthly magazine is published to informscientists, engineers, businessmen, industry, teachers,students, and consumers of the latest advances inscience and technology, with primary emphasis on thework at NBS. The magazine highlights and reviews suchissues as energy research, fire protection, building tech-nology, metric conversion, pollution abatement, healthand safety, and consumer product performance. In addi-tion, it reports the results of Bureau programs inmeasurement standards and techniques, properties ofmatter and materials, engineering standards and serv-ices, instrumentation, and automatic data processing.

Annual subscription: Domestic, $9.45; Foreign, $11.85.

NONPERIOOICALS

Monographs—Major contributions to the technical liter-ature on various subjects related to the Bureau's scien-tific and technical activities.

Handbooks—Recommended codes of engineering andindustrial practice (including safety codes) developedin cooperation with interested industries, professionalorganizations, and regulatory bodies.

Special Publications—Include proceedings of confer-ences sponsored by NBS, NBS annual reports, and otherspecial publications appropriate to this grouping suchas wall charts, pocket cards, and bibliographies.

Applied Mathematics Series—Mathematical tables,manuals, and studies of special interest to physicists,engineers, chemists, biologists, • mathematicians, com-puter programmers, and others engaged in scientificand technical work.

National Standard Reference Data Series—Providesquantitative data on the physical and chemical proper-ties of materials, compiled from the world's literatureand critically evaluated. Developed under a world-wide

program coordinated by NBS. Program under authorityof National Standard Data Act (Public Law 90-396).

NOTE: At present the principal publication outlet forthese data is the Journal of Physical and ChemicalReference Data (JPCRD) published quarterly for NBSby the American Chemical Society (ACS) and the Amer-ican Institute of Physics (AIP). Subscriptions, reprints,and supplements available from ACS, 1155 SixteenthSt. N. W., Wash. D. C. 20056.

Building Science Series—Disseminates technical infor-mation developed at the Bureau on building materials,components, systems, and whole structures. The seriespresents research results, test methods, and perform-ance criteria related to the structural and environmen-tal functions and the durability and safety character-istics of building elements and systems.

Technical Notes—Studies or reports which are completein themselves but restrictive in their treatment of asubject. Analogous to monographs but not so compre-hensive in scope or definitive in treatment of the sub-ject area. Often serve as a vehicle for final reports ofwork performed at NBS under the sponsorship of othergovernment agencies.

Voluntary Product Standards—Developed under pro-cedures published by the Department of Commerce inPart 10, Title 15, of the Code of Federal Regulations.The purpose of the standards is to establish nationallyrecognized requirements for products, and to provideall concerned interests with a basis for common under-standing of the characteristics of the products. NBSadministers this program as a supplement to the activi-ties of the private sector standardizing organizations.

Federal Information Processing Standards Publications(FIPS PUBS)—Publications in this series collectivelyconstitute the Federal Information Processing Stand-ards Register. Register serves as the official source ofinformation in the Federal Government regarding stand-ards issued by NBS pursuant to the Federal Propertyand Administrative Services Act of 1949 as amended,Public Law 89-306 (79 Stat. 1127), and as implementedby Executive Order 11717 (38 FR 12315, dated May 11,1973) and Part 6 of Title 15 CFR (Code of FederalRegulations).

Consumer Information Series—Practical information,based on NBS research and experience, covering areasof interest to the consumer. Easily understandablelanguage and illustrations provide useful backgroundknowledge for shopping in today's technologicalmarketplace.

NBS Interagency Reports (NBSIR)—A special series ofinterim or final reports on work performed by NBS foroutside sponsors (both government and non-govern-ment). In general, initial distribution is handled by thesponsor; public distribution is by the National TechnicalInformation Service (Springfield, Va. 22161) in papercopy or microfiche form.

Order NBS publications (except NBSIR's and Biblio-graphic Subscription Services) from: Superintendent ofDocuments, Government Printing Office, Washington,D.C. 20402.

BIBLIOGRAPHIC SUBSCRIPTION SERVICESThe following current-awareness and literature-surveybibliographies are issued periodically by the Bureau:Cryogenic Data Center Current Awareness Service

A literature survey issued biweekly. Annual sub-scription: Domestic, $20.00; foreign, $25.00.

Liquefied Natural Gas. A literature, survey issued quar-terly. Annual subscription: $20.00.

Superconducting Devices and Materials. A literature

survey issued quarterly. Annual subscription: $20.00.Send subscription orders and remittances for thepreceding bibliographic services to National Bu-reau of Standards, Cryogenic Data Center (275.02)Boulder, Colorado 80302.

Electromagnetic Metrology Current Awareness ServiceIssued monthly. Annual subscription: $24.00. Sendsubscription order and remittance to ElectromagneticsDivision, National Bureau of Standards, Boulder,Colo. 80302.

Page 39: A New Dynamic Pressure Source for the Calibration of

U.S. DEPARTMENT OF COMMERCENational Bureau of StandardsWashinflton, D.C. 20234

OFFICIAL BUSINESS

Penalty for Private Use. £300

POSTAGE AND FEES PAIDU.S. DEPARTMENT OF COMMERCE

COM-21S

SPECIAL FOURTH-CLASS RATE

BOOK

YEARSS

13O1-1S76