04 temperature measurement methods

3
Experiment A: Temperature Measurement Methods Objective: To demonstrate the application of temperature scales and fixed points using mercury-glass thermometer, bimetallic thermometer, thermocouple, resistance temperature detector, and thermistor and vapor pressure thermometer. Experiment Procedures: 1. Fill the water heater with clean water and connect the power cord. 2. Half fill the vacuum flask with a mixture of crushed ice and pure water (the ice being made from pure water and crushed into fine particles). 3. Ambient air temperature measurement: Take out the mercury-glass thermometer, close inspection will reveal a column of mercury protrude from the bulb. Temperature measurement is achieved by relating the length of this column to an engraved scale on the glass. Read the temperature indicated by the column at ambient air temperature. Take out the mercury in bi-metal thermometer, close inspection will reveal a metal rod at the end of the indicator. Temperature measurement is achieved by transferring heat to the metal rod. Take out a Type K thermocouple. Connect the blue and yellow plugs to the corresponding sockets of the Type K thermocouple temperature indicator. Place the thermocouple on the baseboard and allow the readings to stabilize at the ambient air temperature. Read the temperature indicated on the temperature indicator. Take out a resistance temperature detector (RTD). Connect the RTD plug to the corresponding sockets of the RTD indicator. Place the RTD on the baseboard and allow the readings to stabilize at the ambient air temperature. Read the temperature indicated on the RTD temperature indicator. Take out a thermistor. Connect the thermistor plugs to the corresponding sockets of the thermistor indicator. Place the thermistor on the baseboard and allow the readings to stabilize at the ambient air temperature. Read the temperature indicated on the thermistor indicator. Take out the vapor pressure thermometer. Place the vapor pressure thermometer on the baseboard and allow the readings to stabilize at the ambient air temperature. Read the temperature indicated on the indicator 4. Ice-point temperature measurement Insert the bulb of the thermometer into the water-ice mixture; stir gently to ensure intimate contact with the mixture. Observe the reading on the thermometer.

Upload: ahmadaliakbarphambra

Post on 01-Oct-2015

5 views

Category:

Documents


1 download

DESCRIPTION

Temprature measurement experiment

TRANSCRIPT

  • Experiment A: Temperature Measurement Methods

    Objective:

    To demonstrate the application of temperature scales and fixed points using mercury-glass

    thermometer, bimetallic thermometer, thermocouple, resistance temperature detector, and

    thermistor and vapor pressure thermometer.

    Experiment Procedures:

    1. Fill the water heater with clean water and connect the power cord.

    2. Half fill the vacuum flask with a mixture of crushed ice and pure water (the ice being

    made from pure water and crushed into fine particles).

    3. Ambient air temperature measurement:

    Take out the mercury-glass thermometer, close inspection will reveal a column of mercury

    protrude from the bulb. Temperature measurement is achieved by relating the length of this

    column to an engraved scale on the glass. Read the temperature indicated by the column at

    ambient air temperature.

    Take out the mercury in bi-metal thermometer, close inspection will reveal a metal rod at the

    end of the indicator. Temperature measurement is achieved by transferring heat to the metal

    rod.

    Take out a Type K thermocouple. Connect the blue and yellow plugs to the corresponding

    sockets of the Type K thermocouple temperature indicator. Place the thermocouple on the

    baseboard and allow the readings to stabilize at the ambient air temperature. Read the

    temperature indicated on the temperature indicator.

    Take out a resistance temperature detector (RTD). Connect the RTD plug to the

    corresponding sockets of the RTD indicator. Place the RTD on the baseboard and allow the

    readings to stabilize at the ambient air temperature. Read the temperature indicated on the

    RTD temperature indicator.

    Take out a thermistor. Connect the thermistor plugs to the corresponding sockets of the

    thermistor indicator. Place the thermistor on the baseboard and allow the readings to

    stabilize at the ambient air temperature. Read the temperature indicated on the thermistor

    indicator.

    Take out the vapor pressure thermometer. Place the vapor pressure thermometer on the

    baseboard and allow the readings to stabilize at the ambient air temperature. Read the

    temperature indicated on the indicator

    4. Ice-point temperature measurement

    Insert the bulb of the thermometer into the water-ice mixture; stir gently to ensure intimate

    contact with the mixture. Observe the reading on the thermometer.

  • Insert the metal rod of the bi-metal thermometer into the water-ice mixture; stir gently to

    ensure intimate contact with the mixture. Observe the reading on the bimetallic

    temperature indicator.

    Insert the thermocouple probe into the water-ice mixture; stir gently to ensure intimate

    contact with the mixture. Observe the reading on the thermocouple temperature indicator.

    Insert the RTD probe into the water-ice mixture; stir gently to ensure intimate contact with

    the mixture. Observe the reading on the RTD temperature indicator.

    Insert the thermistor probe into the water-ice mixture; stir gently to ensure intimate contact

    with the mixture. Observe the reading on the thermistor temperature indicator.

    Insert the metal rod of the vapor pressure indicator into the water- ice mixture; stir gently

    to ensure intimate contact with the mixture. Observe the reading on the indicator.

    5. Boiling-point temperature measurement Insert the bulb of the thermometer into boiling water. Observe the reading on the

    thermometer.

    Insert the metal rod of the bimetallic indicator into boiling water. Observe the reading on

    the bimetallic indicator.

    Insert the thermocouple probe into boiling water. Observe the reading on the temperature

    indicator.

    Insert the RTD probe into boiling water. Observe the reading on the resistance indicator.

    Insert the thermistor probe into boiling water. Observe the reading on the thermistor

    indicator.

    Insert the metal rod of the vapor pressure indicator into boiling water.

    Observe the reading on the thermometer.

    Repeat the readings in ice and boiling water and observe that the readings are consistently

    0C and 100C respectively.

    Note:

    The boiling point of water occurs at 100C for standard atmosphere pressure only(760mmHg).

    At lower atmospheric pressure, boiling will occur at lower temperature and, conversely, at

    higher atmospheric pressure, boiling will occur at high temperature. For example, a rise in

    atmosphere of 27 mmHg above the standard pressure will result in a rise of 1C in the boiling

    water point of water. Where boiling occurs at a temperature other than 100C, a barometer

    should be consulted to confirm the deviation.

    Pure water must be used where accurate calibration is required, paying attention to

    atmospheric pressure at the time of calibration. On completion of the experiment, switch off

    the water heater.

  • Observations:

    Apparatus Ambient (C) Ice Point (C) Boiling Point (C)

    Glass Thermometer (10C to 110C)

    Glass Thermometer (5C to 350C)

    Bimetal Thermometer

    RTD

    Thermistor

    Type K Thermocouple

    Vapor Pressure Thermometer