understanding the nu-430 / 435 - nuaire · class ii, type b2 biological safety cabinet / fume hood...
TRANSCRIPT
Class II, Type B2 Biological Safety Cabinet / Fume Hood
Understanding the NU-430 / 435
A Guide For
Planning
Installation
Operation
21HEPA [High Efficiency Particulate Air]
filters capture only particulate.
Chemicals, fumes, and vapors will pass
through a HEPA filter and enter the
exhaust system.
100% of the cabinet’s contaminated
airflow passes through the Exhaust
HEPA filter.
60% of the cabinet’s total airflow passes
through the Supply HEPA filter creating
a unidirectional downflow.
40% of the total cabinet’s airflow pass
through the front access opening creating
an air barrier.
NU-430 • UL listed, UL-C listed (115 VAC)
• CE (230 VAC)
• SupplyandExhaustHEPAfilter standardULClassII
• Canusestandard,internal servicevalves
• Canuseultravioletlight
• Mustuselabel“DoNotUseToxic orExplosivesinCabinet”
• NSFListed(115VAC)
NU-435 • UL Classified (115 VAC)
• SupplyHEPAfilterULClassI
• ExhaustHEPAfiltertoUL586
• Noultravioletgermicidallightavailable
• Onlyuseremoteservicevalves
• Nolabelrestrictions
• Availablein115V,60Hz.only
• NSFListed(115VAC)
Personnel, Product, and Environmental Protection
Description of Class II, Type B2 Safety Cabinet
TheLabGard NU-430biologicalsafetycabinetisaClassII,TypeB2
TotalExhaustcabinetfeaturingflow-throughconstructionwithno
internalrecirculationofairwithinthecabinet.Requiredairtooperate
thecabinetistakeneitherfromtheroomthroughtheworkaccess
openingorthroughthesupplyairintakeductdirectlyintothesupply
airblower.
TheLabGard NU-435biological/chemicalfumehoodisaClassII,
TypeB2TotalExhaustcabinetfeaturingthesamebasicdesign
parametersastheNU-430.
USE: Type B2 cabinets may be used for
work with volatile toxic chemicals and
radionuclides required as an adjunct to
microbiological studies.*
* Definition from NSF / ANSI 49
Understanding the NU-430 airflow system is critical to the safe operation of the cabinet
TheNU-430biologicalsafetycabinetprovidespersonnel,product,
andenvironmentalprotection.Personnelprotectionisprovidedfor
byhaving40%ofthetotalcabinet’sairflowpassthroughthefront
accessopeningcreatinganairbarrier.Theremaining60%ofthe
cabinet’stotalairflowpassesthroughtheSupplyHEPAfilterand
intotheworkareacreatingunidirectional,turbulence-freedownflow
thatminimizesthepossibilityofcross-contaminationwithinthe
workzone.Thetotalcabinet’sairflowisthenpulledthroughthe
ExhaustHEPAfilterandintotheexhaustsystemprovidingexcellent
environmentalprotection.
21HEPA [High Efficiency Particulate Air]
filters capture only particulate.
Chemicals, fumes, and vapors will pass
through a HEPA filter and enter the
exhaust system.
100% of the cabinet’s contaminated
airflow passes through the Exhaust
HEPA filter.
60% of the cabinet’s total airflow passes
through the Supply HEPA filter creating
a unidirectional downflow.
40% of the total cabinet’s airflow pass
through the front access opening creating
an air barrier.
NU-430 • UL listed, UL-C listed (115 VAC)
• CE (230 VAC)
• SupplyandExhaustHEPAfilter standardULClassII
• Canusestandard,internal servicevalves
• Canuseultravioletlight
• Mustuselabel“DoNotUseToxic orExplosivesinCabinet”
• NSFListed(115VAC)
NU-435 • UL Classified (115 VAC)
• SupplyHEPAfilterULClassI
• ExhaustHEPAfiltertoUL586
• Noultravioletgermicidallightavailable
• Onlyuseremoteservicevalves
• Nolabelrestrictions
• Availablein115V,60Hz.only
• NSFListed(115VAC)
Personnel, Product, and Environmental Protection
Description of Class II, Type B2 Safety Cabinet
TheLabGard NU-430biologicalsafetycabinetisaClassII,TypeB2
TotalExhaustcabinetfeaturingflow-throughconstructionwithno
internalrecirculationofairwithinthecabinet.Requiredairtooperate
thecabinetistakeneitherfromtheroomthroughtheworkaccess
openingorthroughthesupplyairintakeductdirectlyintothesupply
airblower.
TheLabGard NU-435biological/chemicalfumehoodisaClassII,
TypeB2TotalExhaustcabinetfeaturingthesamebasicdesign
parametersastheNU-430.
USE: Type B2 cabinets may be used for
work with volatile toxic chemicals and
radionuclides required as an adjunct to
microbiological studies.*
* Definition from NSF / ANSI 49
Understanding the NU-430 airflow system is critical to the safe operation of the cabinet
TheNU-430biologicalsafetycabinetprovidespersonnel,product,
andenvironmentalprotection.Personnelprotectionisprovidedfor
byhaving40%ofthetotalcabinet’sairflowpassthroughthefront
accessopeningcreatinganairbarrier.Theremaining60%ofthe
cabinet’stotalairflowpassesthroughtheSupplyHEPAfilterand
intotheworkareacreatingunidirectional,turbulence-freedownflow
thatminimizesthepossibilityofcross-contaminationwithinthe
workzone.Thetotalcabinet’sairflowisthenpulledthroughthe
ExhaustHEPAfilterandintotheexhaustsystemprovidingexcellent
environmentalprotection.
3
4
5
LOCATION: Thebiologicalsafetycabinet
shouldbelocatedoutofthetrafficpattern,
andawayfromroomaircurrentsthat
coulddisruptthecontainmentprovided
bytheworkaccessopeningairbarrier.
Windowsshouldremainedclosedatall
timesandroomventilationairinletsshould
notbedirectedacrossthefrontopening.
Clearanceshouldbeprovidedbehindand
oneachsideofthecabinet.
Both supply air and exhaust air have to be controlled for proper operation of the NU-430
Supplyair,typicallytakendirectlyfromtheroomorcanbetakenfrom
aremotesourceifconditioned(i.e.70°F(21°C)30to50%RH).Ifitis
desirabletotakesupplyairfromaremotesource(i.e.adjacenthallway)
morethan5feet(1.5m)fromthecabinet,itwillbenecessarytoaddan
auxiliaryblowertogainadequateairfow.Itwouldalsobenecessaryto
interlocktheauxiliaryblowertotheinternalsupplyblowerofthecabinet
toavoidpressurizationofthecabinetwhennotinuse.Asupplydamper
mayalsoberequireddependinguponsystemdesignasameansto
interlocktheremotesourcetoagainavoidcabinetpressurization.
MAKE UP AIR: Propersizingoftheexhaust
andmakeupsupplysystemsarecritical
tothesuccessfulinstallationofthecabinet.
Adequateroommakeupairinflowmust
replaceexhaustedair.Airdiffusionrate
nottoexceedvelocityof100LFPM[.51m/s]
tominimizedisruptiveaircurrents.
Supply air is typically taken from the room, but can be supplied from a remote source
Theroomorlaboratorywherethecabinetresidesshouldmeetnecessary
ventilationrequirementsforitsintendeduseasspecifiedinguideline
documents(i.e.ANSI/AIHIZ9.5,USP797).UsetheConcurrentBalance
Value(CBV)forcalculatingHVACrequirements.Allductlosses(i.e.duct
diameter,length,numberofelbows,reductions,etc.)mustbeconsidered
andaddedtothecabinetstaticlossrequirementfortheselectionofthe
exhaustblower.Thecabinetshouldneverbethededicatedexhaustfor
theroomorlaboratory.Thereshouldalwaysbeasecondaryroomexhaust
(couldbetiedtothecabinetexhaust)withitsownVariableAirVolume
(VAV)valvetohandlesystemfluctuationsandbetheprimarydynamic
controlleroftheroomorlaboratory.Thecabinetshouldalwaysremainas
aconstantairvolumedevice.ThecabinetcanbeinterfacedtotheBASfor
blowerfanorrelaycontacts,nightsetbackcontacts,andremoteoverride
cabinetshutdowncontacts.
The exhaust system must provide conditions under which the cabinet was certified to meet its stated performance
The biological safety cabinet will not function properly until the exhaust requirements are met.
NU-430 / 435-400 NU-430 / 435-600
ConcurrentBalanceValue
(CFM/CMH): 829/1409 1221/2075
CertificationExhaustValue
(CFM/CMH): 754/1281 1100/1869
PlantDuctStatic
Pressure(Eng/Metric): 1.7”w.g./43mmw.g. 2.0”w.g./51mmw.g.
[A] Measurestatic
pressureand
requiredexhaust.
To External Exhaust System
Foratypicalconfiguration,supplyairwouldbetakenfromtheroom.The
NuAireTouchLinkcontrolsystemwouldautomaticallyregulatetheinternal
supplyblowerviaadigitaldualthermistorairflowsensorlocatedinthe
downflowairstreamtomonitorandcontroltosetpoint.TheTouchLink
controlsystemautomaticallycompensatesforfilterloading,voltage
variances,andotherenvironmentaleffects.
AConstantAirVolume(CAV)valveordampershouldbeplaceddownstream
ofthecabinettoassureasteadyrateofexhaustflow.Theworkaccess
barrier105±5fpm(.53±.025m/s)ismaintainedbytheCAVvalve.The
CAVvalveinmostcasesiscontrolledbytheBuildingAutomationSystem
(BAS),butalsomaybecontrolledbytheTouchLinkcontrolsystemvia
thedigitaldifferentialvelocitypressureflowgridthatmonitorsexhaust
airvolumeandoptionallycontrolsutilizinganactuatedgas-tightdamper
(i.e.NU-951-012).NOTE:Itisalsorecommendedtouseamanualgas-tight
damper(i.e.NU-940-012)mountedjustabovethecabinetforsealingofthe
exhaustductduringthecabinetdecontaminationprocessifastandard
BAScontrolledCAVisused.
Supplyairistypicallytaken
fromtheroom,butcanbe
remotelysuppliedasshown
ifinterlockedproperly
Note:TheExhaustsystemmustbedesignedtoprovidethestaticpressureand
requiredconcurrentbalancevalueatthelocationindicated[A]above.
3
4
5
LOCATION: Thebiologicalsafetycabinet
shouldbelocatedoutofthetrafficpattern,
andawayfromroomaircurrentsthat
coulddisruptthecontainmentprovided
bytheworkaccessopeningairbarrier.
Windowsshouldremainedclosedatall
timesandroomventilationairinletsshould
notbedirectedacrossthefrontopening.
Clearanceshouldbeprovidedbehindand
oneachsideofthecabinet.
Both supply air and exhaust air have to be controlled for proper operation of the NU-430
Supplyair,typicallytakendirectlyfromtheroomorcanbetakenfrom
aremotesourceifconditioned(i.e.70°F(21°C)30to50%RH).Ifitis
desirabletotakesupplyairfromaremotesource(i.e.adjacenthallway)
morethan5feet(1.5m)fromthecabinet,itwillbenecessarytoaddan
auxiliaryblowertogainadequateairfow.Itwouldalsobenecessaryto
interlocktheauxiliaryblowertotheinternalsupplyblowerofthecabinet
toavoidpressurizationofthecabinetwhennotinuse.Asupplydamper
mayalsoberequireddependinguponsystemdesignasameansto
interlocktheremotesourcetoagainavoidcabinetpressurization.
MAKE UP AIR: Propersizingoftheexhaust
andmakeupsupplysystemsarecritical
tothesuccessfulinstallationofthecabinet.
Adequateroommakeupairinflowmust
replaceexhaustedair.Airdiffusionrate
nottoexceedvelocityof100LFPM[.51m/s]
tominimizedisruptiveaircurrents.
Supply air is typically taken from the room, but can be supplied from a remote source
Theroomorlaboratorywherethecabinetresidesshouldmeetnecessary
ventilationrequirementsforitsintendeduseasspecifiedinguideline
documents(i.e.ANSI/AIHIZ9.5,USP797).UsetheConcurrentBalance
Value(CBV)forcalculatingHVACrequirements.Allductlosses(i.e.duct
diameter,length,numberofelbows,reductions,etc.)mustbeconsidered
andaddedtothecabinetstaticlossrequirementfortheselectionofthe
exhaustblower.Thecabinetshouldneverbethededicatedexhaustfor
theroomorlaboratory.Thereshouldalwaysbeasecondaryroomexhaust
(couldbetiedtothecabinetexhaust)withitsownVariableAirVolume
(VAV)valvetohandlesystemfluctuationsandbetheprimarydynamic
controlleroftheroomorlaboratory.Thecabinetshouldalwaysremainas
aconstantairvolumedevice.ThecabinetcanbeinterfacedtotheBASfor
blowerfanorrelaycontacts,nightsetbackcontacts,andremoteoverride
cabinetshutdowncontacts.
The exhaust system must provide conditions under which the cabinet was certified to meet its stated performance
The biological safety cabinet will not function properly until the exhaust requirements are met.
NU-430 / 435-400 NU-430 / 435-600
ConcurrentBalanceValue
(CFM/CMH): 829/1409 1221/2075
CertificationExhaustValue
(CFM/CMH): 754/1281 1100/1869
PlantDuctStatic
Pressure(Eng/Metric): 1.7”w.g./43mmw.g. 2.0”w.g./51mmw.g.
[A] Measurestatic
pressureand
requiredexhaust.
To External Exhaust System
Foratypicalconfiguration,supplyairwouldbetakenfromtheroom.The
NuAireTouchLinkcontrolsystemwouldautomaticallyregulatetheinternal
supplyblowerviaadigitaldualthermistorairflowsensorlocatedinthe
downflowairstreamtomonitorandcontroltosetpoint.TheTouchLink
controlsystemautomaticallycompensatesforfilterloading,voltage
variances,andotherenvironmentaleffects.
AConstantAirVolume(CAV)valveordampershouldbeplaceddownstream
ofthecabinettoassureasteadyrateofexhaustflow.Theworkaccess
barrier105±5fpm(.53±.025m/s)ismaintainedbytheCAVvalve.The
CAVvalveinmostcasesiscontrolledbytheBuildingAutomationSystem
(BAS),butalsomaybecontrolledbytheTouchLinkcontrolsystemvia
thedigitaldifferentialvelocitypressureflowgridthatmonitorsexhaust
airvolumeandoptionallycontrolsutilizinganactuatedgas-tightdamper
(i.e.NU-951-012).NOTE:Itisalsorecommendedtouseamanualgas-tight
damper(i.e.NU-940-012)mountedjustabovethecabinetforsealingofthe
exhaustductduringthecabinetdecontaminationprocessifastandard
BAScontrolledCAVisused.
Supplyairistypicallytaken
fromtheroom,butcanbe
remotelysuppliedasshown
ifinterlockedproperly
Note:TheExhaustsystemmustbedesignedtoprovidethestaticpressureand
requiredconcurrentbalancevalueatthelocationindicated[A]above.
86
7
The TOUCHLINK Electronic Airflow Control System
isamicroprocessormodulethatestablishesandmaintains
anoptimumbalancebetweenthesupplyairsystemand
exhaustairsystemthroughconstantmonitoringofairflow
usingadigitaldualthermistorairflowsensorandadigital
differentialvelocitypressureflowgrid.Thefirstsensor
monitorsdownflowairwithintheworkzone,thepressure
flowgridmonitorsexhaustflow.Theinformationisreceivedbythe
ElectronicControlSystemtocontrolbothdownflowandexhaustflow
[ifoptionalactuatedvalveisinstalled]accordingtocalibratedsetpoints
ascalibratedbytheservicetechnician.Thesystemmaintainsa
completediagnosticfunctionandhasremotecontactsforremote
sensingofvariousfunctions.
Plumbing Services:Serviceballvalveswiththetypeofservice
specifiedbytheremovablebuttononthehandle,arelocatedinthe
workzone.Theserviceballvalvesarenotrecommendedforpressure
over75p.s.i.[5.2BAR].Reducingvalvesshouldbeinstalledexternalto
thecabinetifnecessary.Servicevalvesshouldneverbeusedforoxygen
service.Aspecialneedlevalveforoxygenserviceisrequiredand
available.Externalon/offvalvesarerecommendedforallinstallations.
TOUCHLINK Electronic Airflow Control System
NU-430-400 / 435-400 NU-430-600/ 435-600
ConcurrentBalanceValueCFM/CMH: 829/1409 1221/2075
CertificationExhaustValueCFM/CMH: 754/1281 1100/1869
PlantDuctStaticPressureEng/Metric 1.7”w.g./43mmw.g. 2.0”w.g./51mmw.g.
HeatRejected,BTU,PerHour 474 584
Electrical:Volts,AC,Hz 115,60(230,50/60) 115,60(230,50/60)
Amps:Blower/Lights 5/(3) 7/(4)
Amps:Duplex 3/(3) 3/(3)
Amps:Total 8/(6) 10/(7)
12ft.PowerCord(one) 14GA-3Wire,15A 14GA-3Wire,15A
CratedShippingWeight: 570lbs./259kg. 760lbs./345kg.
NetWeight: 520lbs./236kg. 710lbs./322kg.
Specifications
Note:The NU-430 / 435 must be balanced and certified to meet the performance requirements.
Common Errors
Error Indicator
Cabinet turns on but internal
blower does not work.
High level of noise - over 65 dbA.
Air flows out front of cabinet.
High level of vibration.
Possible Cause
Inadequateexhaust.Makesure
exhaustsystemcanhandleairflow
requirements.Staticpressureand
requiredexhaustmustbemeasured
atconnectionbetweentheNU-430/
435andfacilityexhaustsystem.[A]
Ductdiametertoosmall/
lessthan12”(305mm).
Ductsystemnotsecurelyanchored
tobuilding.
Ductsheetmetalgaugesandseams
arelessthan20gaugeforround
duct,thisisrequiredtopreventduct
collapseduetohighstaticpressure
conditions.
TheSupplyAir,ifintegrally
designed,mustbeinterlocked
withtheinternalfanrelayto
assureproperoperations.
Ductsystemisdirectlyattached
tocabinet.Useaflexiblesoft
connectionbetweenductand
cabinet.[i.e.NU-940-001]
For additional informationconcerningtheplanning,installation,
availableoptions,orcorrectoperationalprocedures,pleasecontact
NuAire.1.800.328.3352 or www.nuaire.com
[A]
86
7
The TOUCHLINK Electronic Airflow Control System
isamicroprocessormodulethatestablishesandmaintains
anoptimumbalancebetweenthesupplyairsystemand
exhaustairsystemthroughconstantmonitoringofairflow
usingadigitaldualthermistorairflowsensorandadigital
differentialvelocitypressureflowgrid.Thefirstsensor
monitorsdownflowairwithintheworkzone,thepressure
flowgridmonitorsexhaustflow.Theinformationisreceivedbythe
ElectronicControlSystemtocontrolbothdownflowandexhaustflow
[ifoptionalactuatedvalveisinstalled]accordingtocalibratedsetpoints
ascalibratedbytheservicetechnician.Thesystemmaintainsa
completediagnosticfunctionandhasremotecontactsforremote
sensingofvariousfunctions.
Plumbing Services:Serviceballvalveswiththetypeofservice
specifiedbytheremovablebuttononthehandle,arelocatedinthe
workzone.Theserviceballvalvesarenotrecommendedforpressure
over75p.s.i.[5.2BAR].Reducingvalvesshouldbeinstalledexternalto
thecabinetifnecessary.Servicevalvesshouldneverbeusedforoxygen
service.Aspecialneedlevalveforoxygenserviceisrequiredand
available.Externalon/offvalvesarerecommendedforallinstallations.
TOUCHLINK Electronic Airflow Control System
NU-430-400 / 435-400 NU-430-600/ 435-600
ConcurrentBalanceValueCFM/CMH: 829/1409 1221/2075
CertificationExhaustValueCFM/CMH: 754/1281 1100/1869
PlantDuctStaticPressureEng/Metric 1.7”w.g./43mmw.g. 2.0”w.g./51mmw.g.
HeatRejected,BTU,PerHour 474 584
Electrical:Volts,AC,Hz 115,60(230,50/60) 115,60(230,50/60)
Amps:Blower/Lights 5/(3) 7/(4)
Amps:Duplex 3/(3) 3/(3)
Amps:Total 8/(6) 10/(7)
12ft.PowerCord(one) 14GA-3Wire,15A 14GA-3Wire,15A
CratedShippingWeight: 570lbs./259kg. 760lbs./345kg.
NetWeight: 520lbs./236kg. 710lbs./322kg.
Specifications
Note:The NU-430 / 435 must be balanced and certified to meet the performance requirements.
Common Errors
Error Indicator
Cabinet turns on but internal
blower does not work.
High level of noise - over 65 dbA.
Air flows out front of cabinet.
High level of vibration.
Possible Cause
Inadequateexhaust.Makesure
exhaustsystemcanhandleairflow
requirements.Staticpressureand
requiredexhaustmustbemeasured
atconnectionbetweentheNU-430/
435andfacilityexhaustsystem.[A]
Ductdiametertoosmall/
lessthan12”(305mm).
Ductsystemnotsecurelyanchored
tobuilding.
Ductsheetmetalgaugesandseams
arelessthan20gaugeforround
duct,thisisrequiredtopreventduct
collapseduetohighstaticpressure
conditions.
TheSupplyAir,ifintegrally
designed,mustbeinterlocked
withtheinternalfanrelayto
assureproperoperations.
Ductsystemisdirectlyattached
tocabinet.Useaflexiblesoft
connectionbetweenductand
cabinet.[i.e.NU-940-001]
For additional informationconcerningtheplanning,installation,
availableoptions,orcorrectoperationalprocedures,pleasecontact
NuAire.1.800.328.3352 or www.nuaire.com
[A]