ueno 2016

Upload: taufik-jeboed

Post on 06-Jul-2018

214 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/17/2019 Ueno 2016

    1/6

    Brief 

    Reports

    TRACHEAL TUBE POSITION SHIFT DURING INFANT RESUSCITATION BY CHEST

    COMPRESSION: A SIMULATION COMPARISON BY FIXATION METHOD AND WITH

    OR WITHOUT CUFF

    Takeshi Ueno,   MD, Nobuyasu Komasawa,   MD,   PHD, Nozomi Majima,   MD,   PHD, Ryosuke Mihara,   MD, and

    Toshiaki Minami,   MD,   PHD

    Department of Anesthesiology, Osaka Medical College, Osaka, Japan

    Reprint Address: Nobuyasu Komasawa,  MD,  PHD, Department of Anesthesiology, Osaka Medical College, 2-7 Daigaku-machi, Takatsuki,

    Osaka 569-8686, Japan

    , Abstract—Background: Tracheal tube placementduring

    infant resuscitation is essentialfor definite airway protection.

    Accidental extubation due to tracheal tube displacement is a

    rareevent, but it results in severe complications, especially in

    infants. Objective: The present study evaluated how infant

    tracheal tube displacement is affected by tape vs.tube holder

    fixation using a manikin. Methods: A tracheal tube with in-

    ternal diameter of 3.5 mm was placed 10 cm from the gum

    ridge in an advanced life support (ALS) Baby

    simulator(Laerdal, Stavanger, Norway). In the first trial, cuff pressure

    was set at15, 20, and 25cmH2O and trials were performed at

    each setting with no fixation, Durapore (3M, St Paul, MN)

    tape fixation, Multipore

    (3M) tape fixation, and Thomas

    Tube Holder (Laerdal) fixation. After 5 min of chest

    compression, the tracheal tube shift was measured. In the

    second trial, we compared the tube shift by chest compres-

    sion with or without cuff in the same way. Results: Relative

    to no fixation, tracheal tube shift was significantly less in

    the Durapore, Multipore, and tube holder groups

    ( p < 0.05) at all cuff settings. Of the three fixation methods,

    the tube holder showed significantly less shift ( p < 0.05) rela-

    tive to tape, regardless of the initial cuff pressure. The posi-tional shift after chest compressions was significantly larger

    in the trials with cuff than in those without cuff in Durapore

    or Multipore fixation ( p  < 0.05), but did not in tube holder

    fixation. Conclusions: There is less tracheal tube displace-

    ment with tube holder fixation than with tape during

    continuous infant chest compression simulation. The tube

    cuff can contribute to the positional shift of the tube during

    infant chest compression.   2016 Elsevier Inc.

    , Keywords—tracheal tube; position shift; chest compres-

    sion; infant; tube fixation; simulation

    INTRODUCTION

    Asphyxial cardiac arrest is more common than ventricular

    fibrillation cardiac arrest in infants and children, and

    definite tracheal intubation is extremely important in

    infant resuscitation   (1,2). Once tracheal intubation is

    completed, continuous chest compression is possible

    because there is definite separation and protection of the

    trachea from the esophagus and stomach and the quality

    of chest compressions can be measured by assessing

    end-tidal CO2   values   (3). Endotracheal tube placement

    during resuscitation is important for definite tracheal pro-

    tection, especially in infants and children. After successfulendotracheal intubation, the infant endotracheal tube must

    be secured to prevent movement of the tube that can result

    in extubation  (4). Accidental extubation due to endotra-

    cheal tube displacement results in severe complications.

    Unplanned infant extubation has been reported

    frequently in the context of critical care, but very littleThis work was supported by institutional and departmental

    funding.

    RECEIVED: 7 August 2015; FINAL SUBMISSION RECEIVED: 3 November 2015;ACCEPTED: 20 November 2015

    1

    The Journal of Emergency Medicine, Vol.-, No.-, pp. 1–6, 2016Copyright 2016 Elsevier Inc.

    Printed in the USA. All rights reserved0736-4679/$ - see front matter

    http://dx.doi.org/10.1016/j.jemermed.2015.11.034

    http://dx.doi.org/10.1016/j.jemermed.2015.11.034http://dx.doi.org/10.1016/j.jemermed.2015.11.034

  • 8/17/2019 Ueno 2016

    2/6

    research has examined its frequency of this in the emer-

    gency department or in the context of resuscitation

    (4–6). Activities that involve patient movement are

    known to increase the risk of unintentional extubation,

    and it seems likely that patients who are intubated

    during chest compression will exhibit a high risk of 

    accidental extubation.Tracheal tube displacement leading to accidental

    extubation during infant resuscitation is a rare event, but

    can lead to severe complications. Therefore, the present

    study evaluated tracheal tube displacement, as it is affected

    by various fixation methods in a manikin and auto–chest

    compression machine simulation. Specifically, we

    compared adhesive extensive tape, nonextensive tape,

    and tube holder fixation, with no fixation set as a control.

    We hypothesized that the infant tracheal tube would

    shift with continuous chest compression, and that the fix-

    ation method would affect this shift. As a clinical evalu-

    ation of such fixation during resuscitation would beunethical, we decided to use the infant manikin to

    compare fixation methods.

    MATERIALS AND METHODS

    The advanced life support (ALS) Baby manikin (Laer-

    dal, Stavanger, Norway), which is designed to represent

    a 3-month-old infant (11 lb), was used for tracheal tube

    placement and continuous chest compression (7,8). The

    cuffed internal diameter of the tracheal tube

    (Mallinkrodt; Covidien, MA, USA) was 3.5 mm. We

    placed the tracheal tube at 10 cm from the gum ridge inthe baby manikin. The manikin or tracheal tube was

    well prepared with attached lubricant before the study.

    For the first experiment, cuff pressure was adjusted to

    15, 20, and 25 cmH2O. Cuff pressure adjustments were

    performed with an automated cuff pressure controller

    (Mallinckrodt Pressure Control; Covidien, Dublin,

    Ireland), which perform automated calibration and

    contains accuracy down to 1 decimal place, according to

    the manufacturer. After tracheal tube placement and cuff 

    pressure adjustment, the following four fixation methods

    were compared: 1) no fixation; 2) adhesive nonextensive

    tape (Durapore; 3M, St Paul, MN); 3) adhesiveextensive tape (Multipore; 3M); and 4) pediatric tube

    holder (Thomas Tube Holder; Laerdal, Stavanger,

    Norway) (Figure 1A–D).

    Next, 5 min of continuous chest compression was per-

    formed by two-thumb method according to the 2010

    European Resuscitation Council Guidelines   (2). After

    5 min of chest compressions, the incidence of accidental

    extubation and the extent of tube shift from gum ridge to

    the point of displacement were measured with the same

    ruler, which contains scale down to 1 decimal millimeter.

    Trials were performed five times in each setting.

    For the second experiment, tracheal tubes with or

    without cuff were used among the three fixation methods.

    The cuff pressure in the tube with cuff was set at 20

    cmH2O. Tracheal tube position shift was measured as

    the first experiment. All of the experimental procedures

    were conducted by authors.

    Statistical Analysis

    Differences in infant tracheal tube shift were analyzed by

    two-way repeated measures analysis of variance followed

    by Tukey’s multiple comparisons. Results are expressed

    as mean   6   standard deviation. A   p  value < 0.05 was

    considered statistically significant.

    RESULTS

    Tracheal Tube Shift After 5-Min Chest Compression

    Relative to that with no fixation, infant tracheal tube shift

    was significantly less with Durapore tape fixation, Multi-

    pore tape fixation, and tube holder fixation ( p < 0.05) at

    all cuff pressure settings (15 cmH2O: no fixation

    0.56  6  0.23 mm, Durapore 0.22  6  0.04 mm; Multipore

    0.18   6  0.08 mm, and tube holder 0.04   6  0.05 mm; 20

    cmH2O: no fixation 0.66   6   0.19 mm, Durapore

    0.20   6   0.07 mm, Multipore 0.20   6   0.10 mm, and

    tube holder 0.02   6   0.04 mm; 25 cmH2O: no fixation

    0.56   6  0.21 mm, Durapore 0.22  6  0.04 mm, Multipore

    0.18   6   0.07 mm, and tube holder 0.04   6   0.04 mm).

    Among the three fixation methods, the tube holder

    showed significantly less shift relative to both tapes,

    regardless of the initial cuff pressures ( p   < 0.05)

    (Figure 2). There were no significant differences among

    the cuff pressures in each fixation methods.

    Tracheal Tube Shift in Tracheal Tube With Cuff and 

    Without Cuff 

    Figure 3 shows the positional shift of the infant tracheal

    tube with or without cuff. The amount of tracheal tube

    shift was significantly higher with cuff than without

    cuff when the tube was fixed with Durapore or Multi-

    pore (Durapore; with cuff: 0.20   6   0.06 mm, withoutcuff: 0.02   6   0.04 mm. Multipore; with cuff:

    0.20   6   0.09 mm, without cuff: 0.02   6   0.04 mm;

     p   < 0.05 each). In contrast, the amount of positional

    shift did not differ significantly between with or without

    cuff in the tube holder fixation trials (with cuff 

    0.02   6   0.04 mm, without cuff 0.02   6   0.04 mm). In

    the comparison of fixation methods, the degree of posi-

    tional shift did not differ among the fixation methods,

    although, as mentioned, it was significantly larger in

    the cuffed tube with tape fixation compared with the

    cuffed tube with tube holder fixation ( p  < 0.05).

    2 T. Ueno et al.

  • 8/17/2019 Ueno 2016

    3/6

    DISCUSSION

    Airway management is considered an essential elementfor both in-hospital and out-of-hospital infant cardiopul-

    monary resuscitation (CPR)   (9,10). Current European

    Resuscitation Council Guidelines for pediatric

    resuscitation emphasize the importance of airway

    management with as few interruptions as possible   (2).

    After confirmation of infant tracheal tube placement,

    tracheal tube fixation is highly critical because even a

    small tracheal tube shift can result in accidental extuba-

    tion or one-lung ventilation. In adult studies, securing

    the tracheal tube during the use of a tube holder, or

    wire anchoring the tube to the oral cavity, can reduce

    the incidence of unplanned extubation   (11,12).

    However, there are few studies regarding the best tube

    fixation method for infant resuscitation.In our study, the infant tracheal tube was found to shift

    from the trachea to the oral cavity during chest compres-

    sions. Even with Durapore or Multipore tape fixation,

    non-negligible tracheal tube shift was observed. These

    results are compatible with those of adult simulations,

    and it might be becauseof the repetitive intratracheal pres-

    sure caused by chest compressions, which causes a contin-

    uous air stream to flow from the trachea to the mouth, as

    has been demonstrated by end-tidal CO2   volumes

    measured by capnography (2,4). Chest compression can

    also cause a slight increase in the intratracheal pressure,

    Figure 1. Fixation methods. (A) No fixation of the tracheal tube. (B) Tube fixation by Durapore adhesive nonextensive tape. (C)Tube fixation by Multipore adhesive extensive tape. (D) Thomas Tube Holder.

     Tracheal Tube Shifts with Infant Chest Compression 3

  • 8/17/2019 Ueno 2016

    4/6

    leading to bronchospasm or laryngospasm (13,14). Justas

    chest compression is recommended to alleviate choking in

    infants and adults, the same mechanism of intratracheal

    pressure increase can occur through infant chest

    compression during infant cardiorespiratory arrest   (2).

    The finding that the positional shift is greater in the cuffed

    tube (cuff inflated to 20 cmH2O) than in the cuffless tube

    with tape fixation (Durapore or Multipore) indicates that

    the intratracheal pressure increase is transmitted to the

    cuffed tube, but released in the cuffless tube.

    In contrast, the tracheal tube shift was significantlydiminished with tube holder compared with the other

    three fixations that were also compatible with the result

    of adult simulation (4). A tube holder is expected to be

    a more reliable method for securing the infant tracheal

    tube, which could otherwise be dislodged by the repeti-

    tive extubation force of the chest compressions.

    Other studies on adult cadavers and manikins have

    also indicated that the tube holder provides more effective

    resistance to extubation than adhesive tape (15,16). The

    present study demonstrated that although adhesive tape

    provided more resistance to extubation than no fixation,

    tube movement was still detected. Consistent with

    previous studies conducted in intensive care settings,we would recommend using a tube holder to stabilize

    the infant tracheal tube during chest compression (5,6).

    LIMITATIONS

    Our study has several limitations. First, the plastic

    manikin would perform differently than human airway

    tissue in various points. Specifically, air pressure for the

    cuffed tube might have caused more significant move-

    ment in the plastic airway than in the human airway,

    where greater elasticity is present. Second, because we

    used an infant manikin, certain factors encountered inclinical use, such as blood, vomit, and sputum in the

    oropharynx, were not mimicked in our simulations (17).

    Third, in clinical situations, the homogeneity of CPR

    techniques cannot be assured. Fourth, we evaluated

    tracheal tubes with only one cuff type. The amount of 

    tracheal tube positional shift can also differ, depending

    on the cuff type (18).

    Future studies to obtain clinical data on infant tube

    displacement, as well as ethically sound randomized clin-

    ical trials comparing tracheal tube fixation methods dur-

    ing chest compression are required.

    CONCLUSIONS

    A lesser degree of infant tracheal tube displacement

    occurred with tube holder fixation than with tape fixation

    during simulation of continuous infant chest compres-

    sion. The tracheal tube cuff can contribute to the posi-

    tional shift of the infant tracheal tube during chest

    compression.

     Acknowledgments— Takeshi Ueno and Nobuyasu Komasawa

    contributed to the study design, study implementation, data

    analysis, and manuscript preparation. Nozomi Majima contrib-uted to the study design, data collection, data analysis, and

    manuscript preparation. Ryosuke Mihara contributed to the

    data collection, and Toshiaki Minami contributed to the study

    design and manuscript preparation. All authors discussed the

    results and approved the final manuscript.

    REFERENCES

    1.  Monsieurs KG, Nolan JP, Bossaert LL, et al. European Resuscita-tion Council Guidelines for Resuscitation 2015: Section 1. Execu-tive summary. Resuscitation 2015;95:1–80.

    Figure 3. Comparison of tracheal tube shift during infantchest compression with or without cuff (mean ± standard de-viation). Durapore, Durapore adhesive nonextensive tape;Multipore, Multipore adhesive extensive tape; Tube holder,Thomas Tube Holder. Differences were analyzed by two-way analysis of variance. NS, no significant difference;* p < 0.05 compared with tube without cuff.

    0

    0.1

    0.2

    0.3

    0.4

    0.5

    0.6

    0.7

    0.8

    0.9

    No fixat ion Durapore Multipore Tube Holder

    15cmH2O

    20cmH2O

    25cmH2O

    ## ###

    #

    **  *

       T  r  a  c   h  e  a   l   t  u   b  e  m  o  v  e  m  e  n   t   (  c  m   )

    Figure 2. Tracheal tube movement by infant chest compres-sion (mean ± standard deviation) at initial cuff pressure at 15,20, and 25 cmH2O. Durapore, Durapore

    adhesive nonexten-

    sive tape; Multipore, Multipore

    adhesive extensive tape;Tube holder, Thomas Tube Holder. Differences wereanalyzed by two-way analysis of variance. NS, no significantdifference; * p  < 0.05 compared with the other three simula-

    tions;  #

     p < 0.05 compared with no fixation.

    4 T. Ueno et al.

    http://refhub.elsevier.com/S0736-4679(15)01373-6/sref1http://refhub.elsevier.com/S0736-4679(15)01373-6/sref1http://refhub.elsevier.com/S0736-4679(15)01373-6/sref1http://refhub.elsevier.com/S0736-4679(15)01373-6/sref1http://refhub.elsevier.com/S0736-4679(15)01373-6/sref1http://refhub.elsevier.com/S0736-4679(15)01373-6/sref1

  • 8/17/2019 Ueno 2016

    5/6

    2.  Maconochie IK, Bingham R, Eich C, et al. European ResuscitationCouncil Guidelines for Resuscitation 2015: Section 6. Paediatriclife support. Resuscitation 2015;95:223–48.

    3.  Komasawa N, Atagi K, Ueki R, Nishi S, Kaminoh Y, Minami T.Comparison of optic laryngoscope Airtraq  and Miller laryngo-scope for tracheal intubation during infant cardiopulmonary resus-citation. Resuscitation 2011;82:736–9.

    4.   Komasawa N, Fujiwara S, Miyazaki S, Ohchi F, Minami T. Shifts in

    endotracheal tube position due to chest compressions: a simulationcomparison by fixation method. J Emerg Med 2015;48:241–6.

    5.   Babl FE, Vinci RJ, Bauchner H, Mottley L. Pediatric pre-hospitaladvanced life support care in an urban setting. Pediatr EmergCare 2001;17:5–9.

    6.  Bouza C, Garcia E, Diaz M, Segovia E, Rodriguez I. Unplannedextubation in orally intubated medical patients in the intensivecare unit: a prospective cohort study. Heart Lung 2007;36:270–6.

    7.   Komasawa N, Ueki R, Yamamoto N, Nishi S, Kaminoh Y,Minami T. Comparison of Pentax-AWS Airwayscope, Airtraqand Miller laryngoscope for tracheal intubation by novice doctorsduring infant cardiopulmonary resuscitation; A randomized cross-over trial. J Anesth 2013;27:778–80.

    8.  Komasawa N, Ueki R, Yamamoto N, et al. Comparison of air-Qand Soft Seal laryngeal mask for airway management by novicedoctors during infant chest compression: a manikin study. Resusci-

    tation 2012;83:365–8.9.   Nishisaki A, Marwaha N, Kasinathan V, et al. Airway management

    in pediatric patients at referring hospitals compared to a receivingtertiary pediatric ICU. Resuscitation 2011;82:386–90.

    10.  Suominen P, Olkkola KT, Voipio V, Korpela R, Palo R, Räsänen J.Utstein style reporting of in-hospital paediatric cardiopulmonaryresuscitation. Resuscitation 2000;45:17–25.

    11.   ChiangAA, LeeKC, LeeJC, Wei CH.Effectiveness of a continuousquality improvement program aiming to reduce unplanned extuba-tion: a prospective study. Intensive Care Med 1996;22:1269–71.

    12.   Gardner A, Hughes D, Cook R, Henson R, Osborne S, Gardner G.Best practice in stabilisation of oral endotrachealtubes: a systematic

    review. Aust Crit Care 2005;18:158. 160–165.13.   Eason J, Tayler D, Cottam S, et al. Manual chest compression for

    total bronchospasm. Lancet 1991;337:366.14.   Al-Metwalli RR, Mowafi HA, Ismail SA. Gentle chest compression

    relieves extubation laryngospasm in children. J Anesth 2010;24:854–7.

    15.   Owen R, Castle N, Hann H, Reeves D, Naidoo R, Naidoo S. Extu-bation force: a comparison of adhesive tape, non-adhesive tape anda commercial endotracheal tube holder. Resuscitation 2009;80:1296–300.

    16.  Carlson J, Mayrose J, Krause R, Jehle D. Extubation force: tapeversus endotracheal tube holders. Ann Emerg Med 2007;50:686–91.

    17.  Komasawa N, Ueki R, Itani M, Nishi S, Minami T. Validation of Pentax-AWS Airwayscope utility for intubation device during car-diopulmonary resuscitation on the ground. J Anesth 2010;24:582–6.

    18.   Komasawa N, Fujiwara S, Miyazaki S, SoenM, MinamiT. Compar-ison of fluid leakage from four different cuffed pediatric endotra-cheal tubes using a pediatric airway simulation model. Pediatr Int2014;56:634–6.

     Tracheal Tube Shifts with Infant Chest Compression 5

    http://refhub.elsevier.com/S0736-4679(15)01373-6/sref2http://refhub.elsevier.com/S0736-4679(15)01373-6/sref2http://refhub.elsevier.com/S0736-4679(15)01373-6/sref2http://refhub.elsevier.com/S0736-4679(15)01373-6/sref3http://refhub.elsevier.com/S0736-4679(15)01373-6/sref3http://refhub.elsevier.com/S0736-4679(15)01373-6/sref3http://refhub.elsevier.com/S0736-4679(15)01373-6/sref3http://refhub.elsevier.com/S0736-4679(15)01373-6/sref3http://refhub.elsevier.com/S0736-4679(15)01373-6/sref4http://refhub.elsevier.com/S0736-4679(15)01373-6/sref4http://refhub.elsevier.com/S0736-4679(15)01373-6/sref4http://refhub.elsevier.com/S0736-4679(15)01373-6/sref5http://refhub.elsevier.com/S0736-4679(15)01373-6/sref5http://refhub.elsevier.com/S0736-4679(15)01373-6/sref5http://refhub.elsevier.com/S0736-4679(15)01373-6/sref6http://refhub.elsevier.com/S0736-4679(15)01373-6/sref6http://refhub.elsevier.com/S0736-4679(15)01373-6/sref6http://refhub.elsevier.com/S0736-4679(15)01373-6/sref7http://refhub.elsevier.com/S0736-4679(15)01373-6/sref7http://refhub.elsevier.com/S0736-4679(15)01373-6/sref7http://refhub.elsevier.com/S0736-4679(15)01373-6/sref7http://refhub.elsevier.com/S0736-4679(15)01373-6/sref7http://refhub.elsevier.com/S0736-4679(15)01373-6/sref7http://refhub.elsevier.com/S0736-4679(15)01373-6/sref8http://refhub.elsevier.com/S0736-4679(15)01373-6/sref8http://refhub.elsevier.com/S0736-4679(15)01373-6/sref8http://refhub.elsevier.com/S0736-4679(15)01373-6/sref8http://refhub.elsevier.com/S0736-4679(15)01373-6/sref8http://refhub.elsevier.com/S0736-4679(15)01373-6/sref9http://refhub.elsevier.com/S0736-4679(15)01373-6/sref9http://refhub.elsevier.com/S0736-4679(15)01373-6/sref9http://refhub.elsevier.com/S0736-4679(15)01373-6/sref10http://refhub.elsevier.com/S0736-4679(15)01373-6/sref10http://refhub.elsevier.com/S0736-4679(15)01373-6/sref10http://refhub.elsevier.com/S0736-4679(15)01373-6/sref10http://refhub.elsevier.com/S0736-4679(15)01373-6/sref10http://refhub.elsevier.com/S0736-4679(15)01373-6/sref10http://refhub.elsevier.com/S0736-4679(15)01373-6/sref10http://refhub.elsevier.com/S0736-4679(15)01373-6/sref11http://refhub.elsevier.com/S0736-4679(15)01373-6/sref11http://refhub.elsevier.com/S0736-4679(15)01373-6/sref11http://refhub.elsevier.com/S0736-4679(15)01373-6/sref12http://refhub.elsevier.com/S0736-4679(15)01373-6/sref12http://refhub.elsevier.com/S0736-4679(15)01373-6/sref12http://refhub.elsevier.com/S0736-4679(15)01373-6/sref13http://refhub.elsevier.com/S0736-4679(15)01373-6/sref13http://refhub.elsevier.com/S0736-4679(15)01373-6/sref14http://refhub.elsevier.com/S0736-4679(15)01373-6/sref14http://refhub.elsevier.com/S0736-4679(15)01373-6/sref14http://refhub.elsevier.com/S0736-4679(15)01373-6/sref15http://refhub.elsevier.com/S0736-4679(15)01373-6/sref15http://refhub.elsevier.com/S0736-4679(15)01373-6/sref15http://refhub.elsevier.com/S0736-4679(15)01373-6/sref15http://refhub.elsevier.com/S0736-4679(15)01373-6/sref16http://refhub.elsevier.com/S0736-4679(15)01373-6/sref16http://refhub.elsevier.com/S0736-4679(15)01373-6/sref16http://refhub.elsevier.com/S0736-4679(15)01373-6/sref17http://refhub.elsevier.com/S0736-4679(15)01373-6/sref17http://refhub.elsevier.com/S0736-4679(15)01373-6/sref17http://refhub.elsevier.com/S0736-4679(15)01373-6/sref18http://refhub.elsevier.com/S0736-4679(15)01373-6/sref18http://refhub.elsevier.com/S0736-4679(15)01373-6/sref18http://refhub.elsevier.com/S0736-4679(15)01373-6/sref18http://refhub.elsevier.com/S0736-4679(15)01373-6/sref18http://refhub.elsevier.com/S0736-4679(15)01373-6/sref18http://refhub.elsevier.com/S0736-4679(15)01373-6/sref18http://refhub.elsevier.com/S0736-4679(15)01373-6/sref18http://refhub.elsevier.com/S0736-4679(15)01373-6/sref17http://refhub.elsevier.com/S0736-4679(15)01373-6/sref17http://refhub.elsevier.com/S0736-4679(15)01373-6/sref17http://refhub.elsevier.com/S0736-4679(15)01373-6/sref16http://refhub.elsevier.com/S0736-4679(15)01373-6/sref16http://refhub.elsevier.com/S0736-4679(15)01373-6/sref16http://refhub.elsevier.com/S0736-4679(15)01373-6/sref15http://refhub.elsevier.com/S0736-4679(15)01373-6/sref15http://refhub.elsevier.com/S0736-4679(15)01373-6/sref15http://refhub.elsevier.com/S0736-4679(15)01373-6/sref15http://refhub.elsevier.com/S0736-4679(15)01373-6/sref14http://refhub.elsevier.com/S0736-4679(15)01373-6/sref14http://refhub.elsevier.com/S0736-4679(15)01373-6/sref14http://refhub.elsevier.com/S0736-4679(15)01373-6/sref13http://refhub.elsevier.com/S0736-4679(15)01373-6/sref13http://refhub.elsevier.com/S0736-4679(15)01373-6/sref12http://refhub.elsevier.com/S0736-4679(15)01373-6/sref12http://refhub.elsevier.com/S0736-4679(15)01373-6/sref12http://refhub.elsevier.com/S0736-4679(15)01373-6/sref11http://refhub.elsevier.com/S0736-4679(15)01373-6/sref11http://refhub.elsevier.com/S0736-4679(15)01373-6/sref11http://refhub.elsevier.com/S0736-4679(15)01373-6/sref10http://refhub.elsevier.com/S0736-4679(15)01373-6/sref10http://refhub.elsevier.com/S0736-4679(15)01373-6/sref10http://refhub.elsevier.com/S0736-4679(15)01373-6/sref9http://refhub.elsevier.com/S0736-4679(15)01373-6/sref9http://refhub.elsevier.com/S0736-4679(15)01373-6/sref9http://refhub.elsevier.com/S0736-4679(15)01373-6/sref8http://refhub.elsevier.com/S0736-4679(15)01373-6/sref8http://refhub.elsevier.com/S0736-4679(15)01373-6/sref8http://refhub.elsevier.com/S0736-4679(15)01373-6/sref8http://refhub.elsevier.com/S0736-4679(15)01373-6/sref8http://refhub.elsevier.com/S0736-4679(15)01373-6/sref7http://refhub.elsevier.com/S0736-4679(15)01373-6/sref7http://refhub.elsevier.com/S0736-4679(15)01373-6/sref7http://refhub.elsevier.com/S0736-4679(15)01373-6/sref7http://refhub.elsevier.com/S0736-4679(15)01373-6/sref7http://refhub.elsevier.com/S0736-4679(15)01373-6/sref7http://refhub.elsevier.com/S0736-4679(15)01373-6/sref6http://refhub.elsevier.com/S0736-4679(15)01373-6/sref6http://refhub.elsevier.com/S0736-4679(15)01373-6/sref6http://refhub.elsevier.com/S0736-4679(15)01373-6/sref5http://refhub.elsevier.com/S0736-4679(15)01373-6/sref5http://refhub.elsevier.com/S0736-4679(15)01373-6/sref5http://refhub.elsevier.com/S0736-4679(15)01373-6/sref4http://refhub.elsevier.com/S0736-4679(15)01373-6/sref4http://refhub.elsevier.com/S0736-4679(15)01373-6/sref4http://refhub.elsevier.com/S0736-4679(15)01373-6/sref3http://refhub.elsevier.com/S0736-4679(15)01373-6/sref3http://refhub.elsevier.com/S0736-4679(15)01373-6/sref3http://refhub.elsevier.com/S0736-4679(15)01373-6/sref3http://refhub.elsevier.com/S0736-4679(15)01373-6/sref3http://refhub.elsevier.com/S0736-4679(15)01373-6/sref2http://refhub.elsevier.com/S0736-4679(15)01373-6/sref2http://refhub.elsevier.com/S0736-4679(15)01373-6/sref2

  • 8/17/2019 Ueno 2016

    6/6

    ARTICLE SUMMARY

    1. Why is this topic important?Tracheal tube placement during infant resuscitation is

    essential for definite airway protection. Accidental extu-bation due to tracheal tube displacement is a rare event,but it results in severe complications, especially in infants.2. What does this study attempt to show?

    The present study evaluated infant tracheal tube shiftwith continuous chest compression and how trachealtube displacement is affected by tape vs. tube holder fix-ation using a manikin. We also hypothesized the cuff af-fects the infant tracheal tube shift and evaluated thedisplacement with or without cuff.3. What are the key findings?

    Relative to no fixation, tracheal tube shift was signifi-cantly less in the Durapore, Multipore, and tube holdergroups ( p  < 0.05) at all initial cuff settings. Of the threefixation methods, the tube holder showed significantlyless shift ( p < 0.05) relative to tape regardless of the initialcuff pressure. The positional shift after chest compres-sions was significantly larger in the trials with cuff thanin those without cuff.4. How is patient care impacted?

    Tube holder fixation is effective for tracheal tube fixa-tion during infant chest compression. The tube cuff cancontribute to the positional shift of the tube during infantchest compression.

    6 T. Ueno et al.