The Role of LC and LC/MS in the Environmental Laboratory:

An Overview of Recent Technology Trends

Chris Denicola, Richard Lake, Joe Konschnik, Michelle Misselwitz, Jack Cochran, Julie Kowalski, Ty Kahler

Restek Corporation, 110 Benner Circle, Bellefonte, PA 16823

Page 2: The Role of LC and LC/MS in the Environmental Laboratory ...nemc.us/docs/2011/presentations/Presentation... · in the Environmental Laboratory: An Overview of Recent Technology Trends

Outline

• Recent trends• Comparison of GC and LC• Examples of technological advantages with

emerging environmental pollutants

Page 3: The Role of LC and LC/MS in the Environmental Laboratory ...nemc.us/docs/2011/presentations/Presentation... · in the Environmental Laboratory: An Overview of Recent Technology Trends

Analytical Trends – An Overview

UCMR 3 (Unregulated Contaminant Monitoring Rule)Proposed Contaminants and Corresponding Analytical Methods

Page 4: The Role of LC and LC/MS in the Environmental Laboratory ...nemc.us/docs/2011/presentations/Presentation... · in the Environmental Laboratory: An Overview of Recent Technology Trends

New Methods for Drinking Water

ICP - AES

ICP - MS

IC-ESI/MS

GC/MS/MS

GC/ECD

HPLC - UV

GC/MS

LC/MS/MS

• 20 % of new methods from ODW are LC/MS/MS, same percentage as GC/MS

• 8 LC/MS/MS methods have been identified in the Standardized Analytical Methods for the Environmental Restoration Following Homeland Security Events (SAM)

Analytical Trends – An Overview

Page 5: The Role of LC and LC/MS in the Environmental Laboratory ...nemc.us/docs/2011/presentations/Presentation... · in the Environmental Laboratory: An Overview of Recent Technology Trends

Analytical Trends – An Overview

Page 6: The Role of LC and LC/MS in the Environmental Laboratory ...nemc.us/docs/2011/presentations/Presentation... · in the Environmental Laboratory: An Overview of Recent Technology Trends

Analytical Comparison

Analytical Step Advantages of Gas Chromatography

Advantages of Liquid Chromatography

Sample Extraction • Can reduce sample preparation steps and extraction times• Can lessen the need for derivitization (Example 1 - phenoxy acid herbicides)

Sample Introduction • Analyte solubility and compatibility

• Not prone to thermal degradation (Example 2 - BFRs)

Analyte / Matrix Separation

• High efficiency and peak capacity – significant for multi-component anlayses

• Can analyze a wider range of chemical species (Example 3 - PPCPs)• UHPLC is lessening run times (Example 4 - PAH)• Multiple class analysis is common (Example 5 - Multi pesticides)• Alternate separation mechanisms (Example 6 - PAH and Iodated X-ray media)

Analyte Ionization / Detection

• Robust and reproducible electron ionization (EI) and available libraries • Relatively inexpensive

• Specific and sensitive detection with LC/MS/MS and MRM - significant for multi-component anlayses

Page 7: The Role of LC and LC/MS in the Environmental Laboratory ...nemc.us/docs/2011/presentations/Presentation... · in the Environmental Laboratory: An Overview of Recent Technology Trends

Sample Preparation - Derivitization

Page 8: The Role of LC and LC/MS in the Environmental Laboratory ...nemc.us/docs/2011/presentations/Presentation... · in the Environmental Laboratory: An Overview of Recent Technology Trends

Sample Preparation - Derivitization

Page 9: The Role of LC and LC/MS in the Environmental Laboratory ...nemc.us/docs/2011/presentations/Presentation... · in the Environmental Laboratory: An Overview of Recent Technology Trends

Sample Preparation - Derivitization

Page 10: The Role of LC and LC/MS in the Environmental Laboratory ...nemc.us/docs/2011/presentations/Presentation... · in the Environmental Laboratory: An Overview of Recent Technology Trends

Sample Introduction - Injection Port Discrimination

Heating Elements

Heated Zone

GC/MS is a suitable technique for the determination of TOTAL Hexabromocyclododecane (HBCD)

Problem with HBCD isomers undergoing thermal isomerization in the elevated temperatures of a GC injector port

Br Br

Br Br

BrBr

γ

Br Br

Br Br

BrBr

Br Br

Br Br

BrBr

α β

[ α + β + γ ](78%) (13%) ( 9%)

Page 11: The Role of LC and LC/MS in the Environmental Laboratory ...nemc.us/docs/2011/presentations/Presentation... · in the Environmental Laboratory: An Overview of Recent Technology Trends

Viva C18 wide pore (300Å) column (5um 100mm x 2.1mm id).

Separation of HBCD diastereoisomers

Linear relationship between 10 to 5000 pg (r2 = 0.998)

Limit of detection (on-column): 4-10 pg (AB Sciex 2000)

Limit of detection (on-column): 50 - 100 fg (AB Sciex 4000)

- Detection of β-isomer

HBCD isomers accumulate differently in biota compared to sediment

Tomy et al., ES&T., 38 (2004) 2298-2303

α

γ

Courtesy of

Page 12: The Role of LC and LC/MS in the Environmental Laboratory ...nemc.us/docs/2011/presentations/Presentation... · in the Environmental Laboratory: An Overview of Recent Technology Trends

Viva C18 wide pore (300Å) column (5um 100mm x 2.1mm id).

Separation of HBCD diastereoisomers

Linear relationship between 10 to 5000 pg (r2 = 0.998)

Limit of detection (on-column): 4-10 pg (AB Sciex 2000)

Limit of detection (on-column): 50 - 100 fg (AB Sciex 4000)

- Detection of β-isomer

HBCD isomers accumulate differently in biota compared to sediment

Tomy et al., ES&T., 38 (2004) 2298-2303

α

γ

Courtesy of

Page 13: The Role of LC and LC/MS in the Environmental Laboratory ...nemc.us/docs/2011/presentations/Presentation... · in the Environmental Laboratory: An Overview of Recent Technology Trends

Analysis of Halogenated Flame Retardants

Page 14: The Role of LC and LC/MS in the Environmental Laboratory ...nemc.us/docs/2011/presentations/Presentation... · in the Environmental Laboratory: An Overview of Recent Technology Trends

Analysis of Brominated Flame Retardants

Some compounds like BDE 99 and BDE 100 have the same ionization patterns and therefore must be chromatographically separated

BDE 209 and DBDPE are troublesome compounds when it comes to solubility. Decabromodiphenyl ether has <0.1 μg/L solubility in water [1]. These compounds are typically soluble in toluene, which is not readily amenable to LCMS analysis. Compounds such as HBCD and its individual diastereoisomers degrade rapidly in acetonitrile [2].

1.Agency for Toxic Substances and Disease Registry (ATSDR). 2004. “Toxicological profile for Polybrominated Biphenyls and Polybrominated Diphenyl Ethers.” Atlanta, GA: U.S.

Department of Health and Human Services, Public Health Service.

2. Tomy, Gregg; Halldorson, Thor; Danell, Robert; Law, Kerri; Arsenault, Gilles; Alaee, Mehran. “Refinements to the diastereoisomer-specific method for the analysis of

hexabromocyclododecane.” Rapid Commun. Mass Spectrom. 2005;19: 2819-2826

Page 15: The Role of LC and LC/MS in the Environmental Laboratory ...nemc.us/docs/2011/presentations/Presentation... · in the Environmental Laboratory: An Overview of Recent Technology Trends

1. Min Yang , Steven Fazio, David Munch and Patrick Drumm , Journal of Chromatography A 1097 (2005) 124

Solvent Compatibility – Methanol vs. Acetonitrile

Methanol greatly enhances the retention and selectivity of aromatic, conjugated, and

compounds containing electron withdrawing functionality

Page 16: The Role of LC and LC/MS in the Environmental Laboratory ...nemc.us/docs/2011/presentations/Presentation... · in the Environmental Laboratory: An Overview of Recent Technology Trends

BDE 47

BDE 100BDE 99

BB 153 BDE 153

480.00 485.00 490.00 495.000.0

20.0

40.0

60.0

80.0

100.0

Calculated Isotope Pattern Four Bromine Atoms

Calculated Isotope Pattern Five Bromine Atoms

560.00 565.00 570.000.0

20.0

40.0

60.0

80.0

100.0

Calculated Isotope Pattern Six Bromine Atoms

635.00 640.00 645.00 650.00 655.000.0

20.0

40.0

60.0

80.0

100.0

BDE 47

BDE 100BDE 99

BB 153 BDE 153

480.00 485.00 490.00 495.000.0

20.0

40.0

60.0

80.0

100.0

Calculated Isotope Pattern Four Bromine Atoms

480.00 485.00 490.00 495.000.0

20.0

40.0

60.0

80.0

100.0

Calculated Isotope Pattern Four Bromine Atoms

Calculated Isotope Pattern Five Bromine Atoms

560.00 565.00 570.000.0

20.0

40.0

60.0

80.0

100.0

Calculated Isotope Pattern Five Bromine Atoms

560.00 565.00 570.000.0

20.0

40.0

60.0

80.0

100.0

Calculated Isotope Pattern Six Bromine Atoms

635.00 640.00 645.00 650.00 655.000.0

20.0

40.0

60.0

80.0

100.0

Calculated Isotope Pattern Six Bromine Atoms

635.00 640.00 645.00 650.00 655.000.0

20.0

40.0

60.0

80.0

100.0

Page 17: The Role of LC and LC/MS in the Environmental Laboratory ...nemc.us/docs/2011/presentations/Presentation... · in the Environmental Laboratory: An Overview of Recent Technology Trends

Comparison of Separations – Gas - Liquid

Deactivation

Stationary Phase

CapillaryColumn

Increase Efficiency by decreasing column inner diameter (ID)

Gas – Liquid diffusion is relatively fast, highly temperature dependent, and utilizes “thick”

Page 18: The Role of LC and LC/MS in the Environmental Laboratory ...nemc.us/docs/2011/presentations/Presentation... · in the Environmental Laboratory: An Overview of Recent Technology Trends

H H

Diffusion is between two liquids (slow) and phase “thickness” must remain small < 30 Å

Comparison of Separations – Liquid- Liquid

Increase Efficiency by decreasing column

particle diameter (dp)

Complex separation mechanisms that give

rise to modes of separation

Page 19: The Role of LC and LC/MS in the Environmental Laboratory ...nemc.us/docs/2011/presentations/Presentation... · in the Environmental Laboratory: An Overview of Recent Technology Trends

Efficiency

0

2000

4000

6000

8000

10000

12000

14000

16000

1 2 3 4 5 6 7 8 9 10 11

Particle Size (µm)

Pla

tes

Theoretical Plot of Efficiency and Pressure versus Particle Size

N α Ldp

1. L. R. Snyder and J. J. Kirkland “Introduction to Modern Liquid Chromatography” John Wiley & Sons, Inc. (1979) p.37

Pressure

0

1000

2000

3000

4000

5000

1 2 3 4 5 6 7 8 9 10 11

Particle Size (µm)

Pre

ssu

re (

psi

)

22cp ddFLP ηΦ

=

Ultra High Performance Liquid Chromatography

Page 20: The Role of LC and LC/MS in the Environmental Laboratory ...nemc.us/docs/2011/presentations/Presentation... · in the Environmental Laboratory: An Overview of Recent Technology Trends

0

50000

100000

150000

200000

250000

1 2 3 4

Particle Diameter (dp)

Actual Theoretical Plates by Particle Size

Comparison of Particle Size Efficiencies

10 µm

n/m Range30,000 –

50,000

n/m Range190,000 –

230,000

n/m Range60,000 –

100,000

n/m Range100,000 –

160,000

H = 0.020mm H = 0.010mm H = 0.006mm H = 0.004mm5 µm 3 µm 1.9 µm

Page 21: The Role of LC and LC/MS in the Environmental Laboratory ...nemc.us/docs/2011/presentations/Presentation... · in the Environmental Laboratory: An Overview of Recent Technology Trends

Efficiency Comparison

Peak capacity (Pc) is the number of peaks that are resolved within a given retention time window. Isocratic peak capacity :

Pc = (t2- tr1) / W [2]

Where:t2 and tr1 = retention time of last and first eluting peaksW = average peak width at baseline

Rxi®-5Sil MS 30m x 0.25mm x 0.25µmPC = 9

Pinnacle®

DB Aqueous C18 50 mm x 2.1 mm ID, 1.9 µmPC = 5

Page 22: The Role of LC and LC/MS in the Environmental Laboratory ...nemc.us/docs/2011/presentations/Presentation... · in the Environmental Laboratory: An Overview of Recent Technology Trends

Multi Pesticide Analysis

Separation of ~ 300 pesticides on a polar modified reversed phase column with a generic gradient of water/methanol + volatile buffer (total run time 20 min)

Restek Ultra Aqueous C18 (100 x 2.1mm) 3μmWater/methanol + 10mM ammonium formate

Ultra C18

0

2

4

6

8

10

Hydrop

hobic

Aromati

cAcid

Base

Ultra Aqueous C18

0

2

4

6

8

10

Hydrop

hobic

Aromati

cAcid Bas

e

Comparison of Retention profiles using the Hydrophobic Subtraction Model

Courtesy of

Page 23: The Role of LC and LC/MS in the Environmental Laboratory ...nemc.us/docs/2011/presentations/Presentation... · in the Environmental Laboratory: An Overview of Recent Technology Trends

Compound CAS Q1 Ions Group

Acetaminophen 103-90-2 152.2 1

PositiveESI

AcidExtract

Caffeine 58-08-2 195 1

1,7-dimethylxanthine 611-59-6 181.2 1

Carbamazepine 298-46-4 237.4 1

Codeine 76-57-3 300 1

Cotinine 486-56-6 177 1

Erythromycin 114-07-8 734.4 1

Fluoxetine 54910-89-3 310.3 1

Sulfamethoxazole 723-46-6 254 1

Thiabendazole 148-79-8 202.1 1

Trimethoprim 738-70-5 291 1

Gemfibrozil 25812-30-0 249 3

NegativeESI

Ibuprofen 15687-27-1 205.1 3

Naproxen 22204-53-1 228.9 3

Triclosan 3380-34-5 286.8 3

Albuterol 18559-94-9 240 4Positive

ESIBase

ExtractCimetidine 51481-61-9 253.1 4

Ranitidine 66357-35-5 315 4

Standard Preparation

Solvent Preparation

Groups 1 & 4A: 0.3% Formic Acid, 0.1%

Ammonium Formate in water

B: Acetonitrile:Methanol (1:1)

Group 3A: 0.1% Ammonium Acetate,

0.1% Acetic Acid in waterB: Acetonitrile:Methanol (1:1)

Extract Solvent0.1% Formic Acid in Methanol: Water (75:25)

25 µg/mL in Extract Solvent

Analysis of PPCPs

Page 24: The Role of LC and LC/MS in the Environmental Laboratory ...nemc.us/docs/2011/presentations/Presentation... · in the Environmental Laboratory: An Overview of Recent Technology Trends

Group 3Group 1 Group 4

Analysis of PPCPsIonization

suppression

Page 26: The Role of LC and LC/MS in the Environmental Laboratory ...nemc.us/docs/2011/presentations/Presentation... · in the Environmental Laboratory: An Overview of Recent Technology Trends

2 3 5

7

8

10

11

12

13

14

15

1617

18

6

14

9

Time % ACN

0 50

1 60

3 100

5 100

EPA Targeted PAHsPinnacle DB PAH 1.9μm 50 x 2.1mm

Mobile phase: Water/Acetonitrile, Flow: 0.6 mL/min, Temp: 30°C, UV detector: 254nm

1. Naphthalene2. Acenaphthylene3. 1-methylnaphthalene4. 2-methylnaphthalene5. Acenaphthene6. Fluorene7. Phenanthrene8. Anthracene9. Fluoranthene10. Pyrene11. Benzo(a)anthracene12. Chrysene13. Benzo(b)fluoranthene14. Benzo(k)fluoranthene15. Benzo(a)pyrene16. Dibenzo(a,h)anthracene17. Benzo(ghi)perylene18. Indeno(1,2,3-cd)pyrene

Specifically Designed Phase – L/B ratio

Page 27: The Role of LC and LC/MS in the Environmental Laboratory ...nemc.us/docs/2011/presentations/Presentation... · in the Environmental Laboratory: An Overview of Recent Technology Trends

Chemical structures of X-ray contrast media

HN OH

OH

I I

I

HN OH

O

OH

N

CH3

HO

O

O

Chemical Formula:C17H22I3N3O8Exact Mass: 776.85

Iomeprol

N OH

OH

I I

I

HN OH

O

OH

NH

O

CH3

H3CO

O

Chemical Formula: C18H24I3N3O8ExactMass: 790.87

Iopromide

HN OHOH

I I

I

HN OH

O

OH

NHOOH

H3C O

O

Chemical Formula: C19H26I3N3O9ExactMass: 820.88

Iohexol

COO NaI I

INH CH3NHH3C

O O

Chemical Formula: C11H8I3N2NaO4Exact Mass: 635.75

Diatriazoate sodium

Iopamidol and iomeprol are positional isomers and difficulty to separate

Analysis of X-Ray Agents in Wastewater

Page 28: The Role of LC and LC/MS in the Environmental Laboratory ...nemc.us/docs/2011/presentations/Presentation... · in the Environmental Laboratory: An Overview of Recent Technology Trends

Structure of Iopamidol – isomer of Iomeprol

Iopamidol and iomeprol are positional isomers - need chromatographic separation

Iomeprol, iopromide and iohexol have chiral centers -need to sum LC peaks

Analysis of X-Ray Agents in Wastewater

Page 29: The Role of LC and LC/MS in the Environmental Laboratory ...nemc.us/docs/2011/presentations/Presentation... · in the Environmental Laboratory: An Overview of Recent Technology Trends

• Shimadzu Prominence HPLC System• AB SCIEX API 5000TM LC/MS/MS system• Column: Restek Ultra Aromax 3µm, 100 x 2.1 mm• 40°C

• Solvent A = water + 2 mM ammonium acetate• Solvent B = methanol + 2 mM ammonium acetate

LC/MS/MS Separation Method Parameters

Time (min) %A %B0.01 98.0 2.02.00 98.0 2.0

18.00 2.0 98.018.50 98.0 2.030.00 stop

• Flow rate = 0.2 mL/min• 20µL injection volume• All the analytes eluted within 12 minutes• Additional time added to remove any heavy residues that may be present in

sewage water

Page 30: The Role of LC and LC/MS in the Environmental Laboratory ...nemc.us/docs/2011/presentations/Presentation... · in the Environmental Laboratory: An Overview of Recent Technology Trends

The HPLC Separation:iohexol, iomeprol and iopromide result in 2–3 peaks each

2 4 6 8 10 12 14 16 18 20Time, min

0.0

5.0e4

1.0e5

1.5e5

2.0e5

2.5e5

3.0e5

3.5e5

4.0e5

4.5e5

5.0e5

5.5e5

6.0e5

diatriazoate

iopamidol

iohexoliomeprol

iopromide

Removal of heavy molecules

Courtesy of

• Shimadzu Prominence HPLC System• AB SCIEX API 5000TM LC/MS/MS system• Column: Restek Ultra Aromax 3µm, 100 x 2.1 mm• 40°C

• Solvent A = water + 2 mM ammonium acetate• Solvent B = methanol + 2 mM ammonium acetate

Page 31: The Role of LC and LC/MS in the Environmental Laboratory ...nemc.us/docs/2011/presentations/Presentation... · in the Environmental Laboratory: An Overview of Recent Technology Trends

Multiple reaction monitoring (MRM) Table

Q1 Q3 Dwell ID CE Retention Time(m/z) (m/z) (msec) (V) (min)614.8 361.0 50 diatriazoic acid 28.3 3.41631.8 361.0 50 diatriazoic acid + NH4 37.0777.9 558.7 50 iopamidol 32.0 7.30794.9 558.7 50 iopamidol+ NH4 35.8777.9 686.9 50 Iomeprol 30.0 9.00 + 9.56794.9 686.9 50 Iomeprol + NH4 36.0791.9 558.9 50 Iopromide 1 38.5 10.69 + 10.81791.9 572.9 50 Iopromide 2 34.9821.9 803.9 50 Iohexol 37.0 8.00 + 8.36838.9 803.9 50 Iohexol + NH4 35.3

IS=5000V; GS1=60; GS2=40; Temp=680C

Page 32: The Role of LC and LC/MS in the Environmental Laboratory ...nemc.us/docs/2011/presentations/Presentation... · in the Environmental Laboratory: An Overview of Recent Technology Trends

Summary

• UHPLC and LC/MS/MS in particular are adding a powerful tool to multiple industries

• Liquid Chromatography offers separation and detection advantages

• Liquid Chromatography is well suited for emerging pollutants

Page 33: The Role of LC and LC/MS in the Environmental Laboratory ...nemc.us/docs/2011/presentations/Presentation... · in the Environmental Laboratory: An Overview of Recent Technology Trends

Acknowledgements

Special thanks to our collaborating researchers:

• Andre Schreiber and Takeo Sakuma - AB Sciex• Chris Marvin, Simon Zhou Eric Reiner - Environment

Canada, Burlington, Ontario• Nicole Riddell – Wellington Laboratories• Frank Dorman – Pennsylvania State University