the globalfiler(tm) kit: going further for casework
TRANSCRIPT
1 Life Technologies™ |
The GlobalFiler™ Kit: Going Further for Casework Jennifer Schroeder Forensic Specialist Bode Technology 10th Annual Advanced DNA Technical Workshop March 24-28, 2013 San Diego, CA
2 Life Technologies™ |
What is the GlobalFiler™ Kit?
1. D.R. Hares, Expanding the CODIS core loci in the United States, Forensic Sci. Int. Genet. 6 (2012), e52-‐e54.
2. D.R. Hares, Addendum to expanding the CODIS core loci in the United States, Forensic Sci. Int. Genet. 6 (2012), e135.
The GlobalFiler™ Kits are our response to the worldwide expansion of forensic locus sets, recently captured in the CODIS Core Loci Working Group’s recommendation to expand the CODIS Core Loci from 13 to 20 “required” loci and 3 “highly recommended” loci
Locus CODIS (Current) Europe CODIS
(Proposed) GlobalFiler™ Kit
D13S317 X X X
D7S820 X X X
D5S818 X X X
CSF1PO X X X
D1S1656 X X X
D12S391 X X X
D2S441 X X X
D10S1248 X X X
D18S51 X X X X
FGA X X X X
D21S11 X X X X
D8S1179 X X X X
VWA X X X X
D16S539 X X X X
TH01 X X X X
D3S1358 X X X X
AMEL X X X X
D2S1338 X X X X
D19S433 X X X X
DYS391 X X
TPOX X X X
D22S1045 X X X
SE33 X X X
INTERPOL HIT EXEMPLIFIES FUTURE!OF INTERNATIONAL MATCHING!
DNA Gateway
Suspect arrested for rape in 2009 in Austria DNA profile sent to Interpol Gateway and matches profile in CODIS
3 rapes in 2004 linked by Cal DOJ CODIS Suspect idenEfied – flees country – alias assumed Profile sent to Interpol in 2009
Ali Achekzai extradited to California Convicted on all counts
Laguna Beach/San Diego, California
5 Life Technologies™ |
What do you think today?
“Large STR multiplexes are not well suited for casework. Sensitivity and robustness can suffer with the addition of more loci.”
A. True
B. False
6 Life Technologies™ |
5-Dye Configuration Containing All Required and Highly Recommended Loci
1 2
3
4
Multiple loci > 400 bp à poor performance for degraded DNA à compromised sizing and resolution
1 2
3
4 5
6
7 8
Fewer Mini-STRs à poor performance for
degraded DNA
Primer Redesigns à less concordant
7 Life Technologies™ |
The GlobalFiler™ Kits: 6-Dye Configuration 1
2
3
4 5 6
7 8
9 10
GlobalFiler™ Kit 10 Mini-STRs
PI* = 9.2 x 10-12
MiniFiler ™ Kit 8 Mini-STRs
PI* = 1.05 x 10-10
11 12
13 14
15
16 17
18
18 Mini-STRs Combination
PI*= 5.73 x 10-21
97% of alleles in allelic ladder <400bp
Only 2 loci require primer redesign
* Preliminary calcula.ons based on a US Caucasian Database
8 Life Technologies™ |
What is an Indel? • Indels are short length polymorphisms, consisEng of the presence (INserEon)
or absence (DELeEon) of a short sequence (1-‐50 nucleoEdes) • In the GlobalFiler™ Kits, an Indel marker on the Y chromosome was
incorporated into the mulEplex
DeleEon InserEon
9 Life Technologies™ |
• Amelogenin Y null alleles occur
• AddiEonal Y tesEng may be required to confirm sample gender
• AddiEonal gender idenEficaEon markers are on the long arm of the Y chromosome
to minimize the risk of a double deleEon
• DYS391 is a stable locus with a relaEvely narrow allele range • The Y-‐indel is very short therefore more likely to perform even if DYS391 drops out
in degraded samples
p q
Amelogenin Y-‐indel/DYS391
Why Include Additional Y Markers?
10 Life Technologies™ |
• Gender identification redundancy helps remove the need for laboratories to maintain a second gender confirmation test protocol
• In addition, the Y-indel position (<90 bp) in the GlobalFiler™ Kit helps improve data recovery in degraded samples
♂
♀
Improved Accuracy of Gender Determination
11 Life Technologies™ |
What percentage of evidence profiles you handle on a regular basis that are 2 or 3 contributor mixture
profiles, with at least one female contributor?
A. <25%
B. 25-40%
C. 40-60%
D. 60-80%
E. >80%
12 Life Technologies™ |
The GlobalFiler™ Kits: Configuration
DYS391 Amel
All gender markers located in the
same dye channel for ease of
interpretation
Y indel
13 Life Technologies™ |
The GlobalFiler™ Kits: Allelic Ladder
New virtual bins added to
improve genotyping efficiency
343 alleles 246 virtual bins 589 total alleles
31 alleles added to 9
GlobalFiler™ Kit marker ladders
14 Life Technologies™ |
Maximizing Discrimination Kit
D8S
1179
D21
S11
D7S
820
CSF
1PO
D3S
1358
TH01
D13
S317
D16
S539
D2S
1338
D19
S433
vWA
TPO
X
D18
S51
D5S
818
FGA
D10
S124
8
D22
S104
5
D2S
441
D1S
1656
D12
S391
SE33
PI
Identifiler® Kit X X X X X X X X X X X X X X X 5.01 x 10-18
NGM SElect™ Kit X X X X X X X X X X X X X X X X 2.35 x 10-21
GlobalFiler™ Kit X X X X X X X X X X X X X X X X X X X X X *7.12 x 10-26
GlobalFiler™ Mini’s X X X X X X X X X X *9.20 x 10-12
* Preliminary calculations based on a US Caucasian Database
§ The GlobalFiler™ Kits: − Deliver a new level of discrimination − Reduce the risk of adventitious matches − Facilitate international DNA profile comparisons
The GlobalFiler™ Kits offer a power of discrimination 1014 times greater than the total population that has ever lived on Earth!
15 Life Technologies™ |
3130 Series Genetic Analyzers
GeneMapper® ID-X Software v1.4
GeneAmp® 9700 Thermal
Cycler with Silver or Gold-Plated Silver 96-Well Block
Veriti® Thermal Cycler
3500 Series Genetic Analyzers
3730 Genetic Analyzer (GlobalFiler™ Express
Kit only) The GlobalFiler™ Kits utilize the DS36 (J6) Matrix
Standard Set
GlobalFiler™ Kits: Instrument Compatibility
16 Life Technologies™ |
Improved Reagent Traceability Packaging Improvements for the GlobalFilerTM Kits
§ New “Peel and Stick” kit labels can be removed from kit sleeves and placed on component boxes or recorded in laboratory logbooks
§ Improves ability to link kit components with kit lot numbers
18 Life Technologies™ |
Kit Development Strategy
GlobalFiler™
PCR Amplification Kit GlobalFiler ™ Express PCR Amplification Kit
The GlobalFiler™ Kits
A high performance, 6-dye chemistry for casework
samples
An ultra-high throughput, 6-dye chemistry for single
source samples
19 Life Technologies™ |
The Diverse Demands of Forensic Testing
− Predictable − Standardized collection − High yield, high quality DNA − Single source − Fresh
− Unpredictable − Highly variable substrates − Low yield, low quality DNA − Mixed source − Aged
The Samples
− Medium - high throughput − High, 1st pass success rate − Minimal sample transfer and
processing steps − Eliminate rate limiting steps − Automated processing
− Low - medium throughput − Maximum results from
compromised samples − High sensitivity − Maximum specificity − Simplified interpretation
Key Workflow Requirements
Databasing Casework
21 Life Technologies™ |
Optimizing Reaction Time for the GlobalFiler™ Express Kit
§ Protocols developed on Veriti® thermal cyclers using 100% ramp speed
§ Performance verified on the 9700 thermal cyclers using MAX mode
Identifiler® Direct Kit
NGM SElect™ Express Kit
GlobalFiler™ Express Kit
Total Time 2 hr 35 min 45 min 39 - 40 min
PCR Conditions 95°C/ 1 min 94°C/ 3 sec 60°C/ 30 sec 60°C/ 8 min
25-27 x
22 Life Technologies™ |
GlobalFiler™ Express Kit Population Study Sample Type N PCR Cycle
Number 1st Pass Success Rate
(175 RFU)
Blood on FTA® 200 25 100%
Blood on Bloodstain Card (903 Paper) 45 25 100%
Buccal on FTA® (EasiCollect™ Device) 56 26 96.4%
Buccal on Bode Buccal Collector™ 90 26 96.6%
Buccal on Copan 4N6FLOQSwab™ 46 24 95.7%
Test Site Sample Type N PCR Cycle Number
1st Pass Success Rate (175 RFU)
A Blood & Buccal on FTA® 80 27 90.0%
B Buccal on FTA® 86 26 93.0%
C Buccal on FTA® 80 27 97.5%
D Buccal on FTA® 64 26 93.8%
E Buccal on FTA® and Swabs 40 26/25 97.5%
Total 350 94.0%
GlobalFiler™ Express Kit External Testing
24 Life Technologies™ |
GlobalFiler™ Kit: Casework Design Considerations
Casework Kit Performance Requirements:
§ High sensitivity for trace/low level DNA samples § Good intracolor balance for mixture samples § Sufficient tolerance to PCR inhibitors § Maintain high species specificity § Efficient thermal cycling conditions
25 Life Technologies™ |
Optimization of GlobalFiler™ Kit Reaction Conditions
Reagent Volumes
3.3X Master Mix 7.5 µL 10X Primer Set 2.5 µL Sample Input ≤15.0 µL
25.0 µL
Greater flexibility of sample input volume to facilitate analysis of low level samples
Identifiler® Plus & NGM SElect™ Kits GlobalFiler™ Kit
Total Time 2 hr 25 min 1 hr 20 min
PCR Conditions
95°C/ 1 min 94°C/ 20 sec 59°C/ 90 sec 60°C/ 10 min
29 x
Optimized thermal cycling conditions shorten amplification conditions by > 1 hr
26 Life Technologies™ |
GlobalFiler™ Kit Performance Example: Sensitivity
Greater sensiEvity was observed with the GlobalFiler™ Kit compared to the IdenEfiler® Plus Kit
GlobalFiler™ Kit Identifiler® Plus [DNA] D8
D21 D7 CSF D3 TH01 D13 D16
D2(S1) D19 vW
A TPO
X D18 Am
el D5 FGA D1 D10 D12
D2(S4) D22 DY3 SE33 Yind D8
D21 D7 CSF D3 TH01 D13 D16 D2 D19 vW
A TPO
X D18 Am
el D5 FGA
1 ng
500 pg
250 pg
125 pg
62.5 pg
31.25 pg
15.63 pg
The GlobalFiler™ Kit exhibits slightly greater sensiEvity than the IdenEfiler® Plus Kit
27 Life Technologies™ |
GlobalFiler™ Kit Performance Example: Inhibitor Tolerance
The GlobalFiler™ Kit buffer has been optimized to deliver enhanced inhibition tolerance compared to the Identifiler® Plus Kit
Humic Acid Inhibition Hematin Inhibition
Identifiler® Plus Kit GlobalFiler™ Kit
28 Life Technologies™ |
GlobalFiler™ Kit Performance on Casework Samples
Intralocus Balance
GlobalFiler™ Kit
GlobalFiler™ Kit
IdenZfiler® Plus Kit
IdenZfiler® Plus Kit
500 pg 1 ng
Intracolor Balance (%)
The GlobalFiler™ Kit maintains the high levels of intracolor and intralocus balance established by other Applied Biosystems next generaEon STR kits
29 Life Technologies™ |
Identifiler® Plus Kit GlobalFiler™ Kit
Species Specificity Stutter Percentage 70bp 470bp
10 ng Cat DNA
10 ng Dog DNA
10 ng Horse DNA
10 ng Bacteria DNA
400 RFU
GlobalFiler™ Kit Performance on Casework Samples
Reduced mean stutter percentage for the majority of loci and high specificity delivered by optimized buffer chemistry and established primer design methods
simplify interpretation of GlobalFiler™ Kit data
007 DNA
31 Life Technologies™ |
TS1 TS2 TS3 TS4 TS5
ExtracZon Method Organic or EZ1 Organic MagneEc
Bead Qiagen
InvesEgator Organic
ComparaZve Kit
IdenEfiler Plus, PP 16HS, PP Fusion
NGM SElect IdenEfiler Plus
IdenEfiler Plus NGM SElect
Thermal Cycler 9700 Gold VeriE VeriE 9700 Gold 9700 Gold
CE Instrument 3500 3500 3500xL 3500xL 3500
GlobalFiler™ Kit Test Site Overview
32 Life Technologies™ |
300 Test Site #1
GlobalFiler™ Kit Baseline Quality
300
300
Test Site #2
Test Site #3
Read Region 74 444
Non-template Controls
33 Life Technologies™ |
GlobalFiler™ Kit Performance Example: Sensitivity
GlobalFiler™ Kit [DNA] D8
D21
D7
CSF
D3
TH01
D13
D16
D2(S1)
D19
vWA
TPOX
D18
Amel
D5
FGA
D1
D10
D12
D2(S4)
D22
DY3
SE33
Yind
1 ng
500 pg
250 pg
125 pg
62.5 pg
31.25 pg
15.63 pg
Increased allele recovery with low input samples.
34 Life Technologies™ |
Sensitivity Results <50pg of DNA using 10uL Sample Input Volume
TS 1 TS2 TS3 TS4 TS 5
46.88 pg 4 5 16 13 33 38 36 38 18 15 15 26 19 12 16 26 29 35 30 10
31. 25 pg 4 3 1 2 25 24 24 22 6 8 7 12 20 10 6 18 21 18 22 26
23.43 pg 4 2 4 4 12 21 15 12 3 11 7 3 4 5 6 8 10 17 12 19
15.62 pg 0 0 1 1 15 12 14 12 0 1 1 0 1 6 5 2 6 4 7 8
• Variability in allele recovery from site to site
35 Life Technologies™ |
[DNA] = 1.5 pg/uL [DNA] = 3.1 pg/ uL
N = 4 Sample Input Volume Increase in
Recovery
Sample Input Volume Increase in Recovery 10 uL 15 uL 10 uL 15 uL
Test Site 1 0.5 3.5 7 2.5 9.5 3.8
Test Site 2 13.25 15 1.11 23.75 36.25 1.53
Test Site 3 0.5 6 12 8.25 18.5 2.24
Test Site 4 3.5 5.75 1.64 13.5 18.25 1.35
Test Site 5 6.25 14.5 2.32 21.75 26 1.2
Total 24 44.75 1.86 69.75 108.5 1.56
Allele Recovery: Extended Sample Volume Input
• Test sites ran the same samples varying only the volume of DNA added to the reaction. • Adding an additional 5uL of DNA to the reaction delivered approximately 1.5 -2.0 times more alleles in low concentration samples
36 Life Technologies™ |
Test Site Sample Success Rates and Allele Recovery GlobalFiler™ Kit vs. Identifiler® Plus Kit
TS1 TS3 TS4
Kit GF ID+ GF ID+ GF ID+
Full Profile 8 8 13 11 5 4
ParEal Profile -‐ -‐ 2 4 9 9
No Profile -‐ -‐ -‐ -‐ -‐ 1
Total Alleles Recovered 337 240 578 393 461 212
40% 47% 117%
• More full profiles recovered with GlobalFiler resulting in an increased number off alleles called
• 10 samples with partial profiles in both kits: Total alleles recovered is greater with GlobalFiler than with Identifiler Plus.
Profile GF / ID+ GF ID+ % Inc alleles ParZal / ParZal (10
samples) 302 107 182%
37 Life Technologies™ |
Alleles Recovered Using the GlobalFiler™ and Identifiler® Plus Kits : Partial Profiles
0
5
10
15
20
25
30
35
40
45
Fat Trace Swab /310pg
Key Swab/ 60pg
Booleneck Swab / 50pg
Drawer Swab/60pg
Swab on Safe / 50pg
Wheel Swab /80pg
Cigareoe Buo / 40pg
Cigareoe Buo / 30pg
Cap Swab / 165 pg
Stone / Undet
No. Alleles R
ecovered
ParZal Profile Samples
GF ParEal
ID+ ParEal
The increased robustness of the GlobalFiler Kit delivered 182% more alleles than Identifiler Plus with partial profiles
• Deliver more information per sample • Increases the likelihood of a meaningful database match
38 Life Technologies™ |
GlobalFiler™ Kit Test Site Results
NGM SElect™ Kit: Full Profile (26 alleles) GlobalFiler™ Kit: Full Profile (36 alleles)
Cigarette Butt, Organic Extraction NGM SElect™ Kit: 50 RFU PAT (3130xl) GlobalFiler™ Kit: 175 RFU PAT (3500xL)
39 Life Technologies™ |
GlobalFiler™ Kit: Full (40 alleles) Identifiler® Plus Kit: Full (29 alleles)
GlobalFiler™ Kit Test Site Results
Bone Sample; Magnetic Bead Extraction (1ng) Identifiler® Plus Kit: 50 RFU PAT (3130xl) GlobalFiler™ Kit: 175 RFU PAT (3500xL)
40 Life Technologies™ |
GlobalFiler™ Kit Test Site Results
Identifiler® Plus Kit: Partial Profile (21 alleles) GlobalFiler™ Kit: Partial Profile (28 alleles)
Bone Sample; Organic Extraction Identifiler® Plus Kit: 50 RFU PAT (3130xl) GlobalFiler™ Kit: 175 RFU PAT (3500xL)
41 Life Technologies™ |
GlobalFiler Kit: Full (40 alleles) Identifiler® Plus Kit: Partial (21 alleles)
Door Handle Swab, Qiagen Investigator (310pg) Identifiler® Plus Kit: 50 RFU PAT (3130xl) GlobalFiler™ Kit:175 RFU PAT (3500xL)
GlobalFiler™ Kit Test Site Results
42 Life Technologies™ |
Identifiler® Plus: Partial (21 alleles) GlobalFiler Kit: Partial (34 alleles)
GlobalFiler™ Kit Test Site Results
Hair Root, Magnetic Bead (460pg) Identifiler® Plus Kit: 50 RFU PAT (3130xl) GlobalFiler™ Kit: 175 RFU PAT (3500xL)
43 Life Technologies™ |
GlobalFiler™ Kit Test Site Results
NGM SElect™ Kit: Partial/Mixture (24 alleles) GlobalFiler™ Kit: Partial/Mixture (37 alleles)
Underwear, Organic Extraction NGM SElect™ Kit: 50 RFU PAT (3130xl) GlobalFiler™ Kit: 175 RFU PAT (3500xL)
44 Life Technologies™ |
GlobalFiler Kit: Full (42 alleles)
Sample Type: Unknown GlobalFiler Kit: 175 RFU PAT (3500) PowerPlex Fusion:175 RFU PAT (3500)
PowerPlex® Fusion Kit: Partial (33 alleles)
GlobalFiler™ Kit Test Site Results
45 Life Technologies™ |
Test Site Results: 900 Year Old Bone Sample
§ Medieval Austrian, Agnes of Waiblingen was the daughter of Henry IV, Holy Roman Emperor and was married to Leopold III, one of the most important Babenberg Margraves, who ruled the Austrian territory
§ Born in 1072, she died in
1143 (aged 71) and is reported to have had >20 children
46 Life Technologies™ |
After hearing more about the GlobalFiler System, how do you feel now when you
think about implementing and analyzing a 24 locus multiplex?
A. Terrified, I can’t imagine how difficult it might be to use
B. Apprehensive, I need a lot of questions answered before agreeing to try it
C. Interested, I’d like to try it sooner rather than later
D. Very excited, when can I get my hands on it
47 Life Technologies™ |
GlobalFiler™ Kit: Test Site Feedback
“The baseline is perfect, no artifact was observed in
negative controls. Sample profiles have showed good
quality data with a good balance PHR.”
“The fact that we are able to use up to 15µL of DNA per PCR reaction improves the
sensitivity … very interesting for the amplification of poor DNA samples (<50pg/
µL).”
“Having 50% of the PCR products with a size less than or equal to 220pb allows better amplification of degraded DNA. In addition, with a volume of 15µL of DNA per PCR reaction, we are in the best conditions for amplification of degraded DNA with an amount of
DNA lower than100pg/µL.”
“Seems to withstand well to inhibitors…”
48 Life Technologies™ |
GlobalFiler™ Kit: Test Site Feedback
“Thanks to the integration of DYS391 and the Y-indel we can
no longer have any doubt concerning the analysis of the Y
chromosome in the case of amelogenin mutation.”
“Using a 6th dye allows to separate the markers analyzed and thus avoid allele overlap between neighboring markers.”
“The use of this kit should reduce the sample rate that has to be re-amplified whether with the same PCR kit or with another kit, while improving performance…”
“PCR duraEon (1:20 vs. 2:30 with IdenEfiler® Plus)
is also an advantage especially in the context of
custody…”
49 Life Technologies™ |
Development of a Next Generation Global Multiplex
§ Go Faster! − Up to 5x faster amplification time than previous generation kits
> GlobalFiler™ Express Kit: ~40 minutes; GlobalFiler™ Kit: ~80 minutes > Enables sample to answer in <2 hours
§ Go Further! − 24 loci, including 10 mini-STRs, increasing discrimination power by up to 9
orders of magnitude − Enhanced buffer system and sensitivity for even better performance on
inhibited and low-level samples
§ Go Global! − The only kit containing all markers recommended for inclusion by the CODIS
Core Loci Working Group, including all markers commonly used in Europe − Maximizes global database compatibility
50 Life Technologies™ |
Acknowledgements § External Test Sites
− 66 worldwide and growing!
§ R&D Julio Mulero, Siddhita Gopinath, Wilma Norona, Maohew Ludeman, Robert Lagace, Chien-‐Wei Chang, Chang Zhong, Nigel Beard
§ Tech Transfer Larry Wang, Anthony Maohews, Vanessa Chan, Yenyoung Lei
§ Soxware Allan Minn, Gloria Lam, Sharada Vijaychander, Tony Hatch, Makesh Kumar, Ravi Gupta, Doug Greig, Babu Yama
§ Product Management Karen Cormier, Jacki Benfield
§ Forensic Science ApplicaEons Jennifer Schroeder, Lisa Calandro, Jeff Sailus
§ Quality Assurance Liwei Qi, Yurah Yen
§ MarkeEng Ariana Wheaton, Jonathan Tabak, Shereen Darwish, Lisa Schade
§ Oligo Development Sarah Ruhe, Kathy Wang, Hong Ji
§ UK Manufacturing Catherine Duke
§ Project Management Wendy Coffey, Linda Kauffman
51 Life Technologies™ |
Thank You
© 2013 Life Technologies CorporaEon. All rights reserved. The trademarks menEoned herein are the property of Life Technologies CorporaEon or their respecEve owners. The content provided herein may relate to products that have not been officially released and is subject to change without noEce. “FOR FORENSIC AND PATERNITY USE ONLY”