sjm mri activator™

15
SJM-EDTR-1116-0065 | Item approved for U.S. use only. SJM MRI ACTIVATOR™ HANDHELD DEVICE WORKFLOW Model: EX4000 1

Upload: others

Post on 15-Oct-2021

22 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: SJM MRI ACTIVATOR™

SJM-EDTR-1116-0065 | Item approved for U.S. use only.

SJM MRI ACTIVATOR™

HANDHELD DEVICE WORKFLOWModel: EX4000

1

Page 2: SJM MRI ACTIVATOR™

SJM-EDTR-1116-0065 | Item approved for U.S. use only. 2

APPROVED MRI SCAN CONDITIONS

1.5 TESLA

FULL BODY

Page 3: SJM MRI ACTIVATOR™

SJM-EDTR-1116-0065 | Item approved for U.S. use only. 3

The SJM MRI Activator™ handheld device must be enabled prior to using the Merlin™ Patient Care System (PCS).

SETUP SJM MRI ACTIVATOR™ HANDHELD DEVICE

Page 4: SJM MRI ACTIVATOR™

SJM-EDTR-1116-0065 | Item approved for U.S. use only. 4

MRI PATIENT WORKFLOW WITH SJM MRI ACTIVATOR™ HANDHELD DEVICE

Saves time, effort,

and inconvenience

4

Patient needs an MRI

scan

Patient has MRI scan

MR

Patient goes homePatient goes to

cardiology center for

device check

Page 5: SJM MRI ACTIVATOR™

SJM-EDTR-1116-0065 | Item approved for U.S. use only.

SJM MRI Activator™ Device Compatibility

The SJM MRI Activator™ device is not paired with a specific device and can be used for all

Assurity MRI™ pacemakers that have the SJM MRI Activator™ device enabled.

Shelf Life

3-year battery life from manufacturing date.

Activators are powered with a non-standard, non-rechargeable battery,

and must be returned to St. Jude Medical for replacement.

SJM MRI ACTIVATOR™ HANDHELD DEVICE

3 years

5

Page 6: SJM MRI ACTIVATOR™

SJM-EDTR-1116-0065 | Item approved for U.S. use only. 6

BUTTONS AND LED INDICATOR LIGHTS

MRI Settings Status

Request Button

Disable MRI Settings Button

Green LED indicator lights

Enable MR Settings Button

Red LED indicator lights

MRDo not use

in MRI

scan room

Page 7: SJM MRI ACTIVATOR™

SJM-EDTR-1116-0065 | Item approved for U.S. use only. 7

Color of LED Can the patient be put in MRI scanner? Cause

Green Yes The pacemaker is in MRI Settings

Red No

SJM MRI Activator™ device has not been enabled

Lead impedance readings may be out of programmed ranges so the pacemaker remains in permanent settings

Notify the device clinic, physician, orSt. Jude Medical representative

No Lights

+

Two-tone Beep

No

The SJM MRI Activator™ device is unable to communicate with the pacemaker

Reposition the SJM MRI Activator™ device and press the button again

Notify the device clinic, physician or St. Jude Medical representative if the communication cannot be resolved

LED INDICATOR LIGHTS

Page 8: SJM MRI ACTIVATOR™

SJM-EDTR-1116-0065 | Item approved for U.S. use only. 8

OVERVIEW

Page 9: SJM MRI ACTIVATOR™

SJM-EDTR-1116-0065 | Item approved for U.S. use only.

Step One: Verify that the patient has a St. Jude Medical™ MRI Conditional pacing system.

Step Two: Position the SJM MRI Activator™ device over the implanted pacemaker.

9

SJM MRI ACTIVATOR™ DEVICE – WORKFLOW TO SCAN PATIENTS

1 2

Page 10: SJM MRI ACTIVATOR™

SJM-EDTR-1116-0065 | Item approved for U.S. use only.

STEP 3: PRESS THE BUTTON TO ACTIVATE MRI SETTINGS

If green SJM MRI Activator™ device LED illuminates continuously for 5s

MRI settings are ON

Safe to scan the patient

If red SJM MRI Activator™ device LED illuminates continuously for 5s

DO NOT scan the patient

MRI settings are OFF

3

10

Page 11: SJM MRI ACTIVATOR™

SJM-EDTR-1116-0065 | Item approved for U.S. use only.

STEP 4: PRESS THE BUTTON TO DISABLE MRI MODE

If red SJM MRI Activator™ device LED illuminates continuously for 5s

MRI settings are OFF

Safe to discharge patient

If green SJM MRI Activator™ device LED illuminates continuously for 5s

DO NOT discharge the patient

MRI settings are ON

4

11

Page 12: SJM MRI ACTIVATOR™

SJM-EDTR-1116-0065 | Item approved for U.S. use only.

BUTTON

Use the MR Status button on the

SJM MRI Activator™ device at any

time to check whether the MRI

settings are on or off.

12

Page 13: SJM MRI ACTIVATOR™

SJM-EDTR-1116-0065 | Item approved for U.S. use only.

NO COMMUNICATION WITH DEVICE

If the SJM MRI Activator™ device makes a two-tone beep, one of the following conditions may apply:

It cannot communicate with the Assurity MRI™ pacemaker; reposition the SJM MRI Activator™ device.

The pacemaker is not MRI conditional; contact the physician (on the ID card) or St. Jude Medical Technical Support.

13

Page 14: SJM MRI ACTIVATOR™

SJM-EDTR-1116-0065 | Item approved for U.S. use only.

Before the SJM MRI Activator™ device can be used, it must be enabled using the

Merlin™ Patient Care System

The SJM MRI Activator™ device is device agnostic, meaning it can be used to enable

MRI settings in any Assurity MRI™ pacemaker that has the MRI Activator enabled in the device

settings

When the activator has been enabled, a clinician has already verified MRI settings that

are appropriate for the patient.

The activator can be used prior to the scan to enable the MRI settings stored within the device

and then again after the MRI scan to return the device to the original parameter set.

SUMMARY

14

Page 15: SJM MRI ACTIVATOR™

SJM-EDTR-1116-0065 | Item approved for U.S. use only.

DISCLAIMER

15

Rx Only

Brief Summary: Prior to using these devices, please review the Instructions for Use for a complete listing of indications, contraindications , warnings, precautions, potential adverse events and directions for use.

Indications: Dual-Chamber Pacing (Dual-chamber pulse generators, CRT-Ps) is indicated for those patients exhibiting: sick sinus syndrome, chronic, symptomatic second- and third-degree AV block, recurrent Ad¬ams-Stokes syndrome,

symptomatic bilateral bundle branch block when tachyarrhythmia and other causes have been ruled out. Atrial Pacing is indicated for patients with sinus node dysfunction and normal AV and intraventricular conduction systems. Ventricular

Pacing is indicated for patients with significant bradycardia and: normal sinus rhythm with only rare episodes of A-V block or sinus arrest, chronic atrial fibrillation, severe physical disability.

MR Conditional System: MR Conditional pacemakers are conditionally safe for use in the MRI environment when used in a complete MR conditional system and according to instructions in the MRI procedure document for the St. Jude

Medical™ MR conditional system.

Contraindications: Dual-Chamber Pacing (Dual-chamber pulse generators, CRT-Ps), though not contraindicated for patients with chronic atrial flutter, chronic atrial fibrillation, or silent atria, may provide no benefit beyond that of single-

chamber pacing in such patients. Single-Chamber Ventricular Demand Pacing is relatively contraindicated in patients who have demonstrated pacemaker syndrome, have retrograde VA conduction, or suffer a drop in arterial blood

pressure with the onset of ventricular pacing. Single-Chamber Atrial Pacing is relatively contraindicated in patients who have demonstrated compromise of AV conduction. For specific contraindications associated with individual modes,

refer to the programmer’s on-screen help.

Warnings: To prevent permanent damage to the device and tissue damage at the electrode/tissue interface: Electrosurgery: Do not use electrosurgical devices in the vicinity of an implanted device. If electrocautery is necessary, use a

bipolar cauterizer or place the indifferent electrode as far from the device as possible, . Lithotripsy: Do not focus a litho tripsy beam within 6 inches of the device. Program the device to Sensor Off prior to lithotripsy to prevent inappropriate

increases in pacing rate. A thorough assessment of device function with special attention to the sensor should be performed following exposure to lithotripsy. Therapeutic Radiation: Do not use ionizing radiation in the vicinity of an

implanted device. Radiation therapy may damage the electronic circuitry of the device. Ultrasound Treatment: Do not use therapeutic ultrasound within 6 inches of the device. Ventricular Sensing: In CRT-Ps, Ventricular Sensitivity should

be programmed to the highest setting (lowest sensitivity) that will provide ventricular sensing with adequate sensing margin. Left ventricular lead dislodgement, to a position near the atria, can result in atrial oversensing and ventricular

inhibition. Perform a thorough assessment of device function following exposure to any of the above.

Backup VVI Operation: In rare instances, the device may revert to Backup VVI operation settings. These values are not programmable. When the device has reverted to Backup VVI operation, the programmer displays a pop-up message

indicating that the device is operating at the Backup VVI values. Press [Continue] and follow the on-screen instructions.

Elective Replacement Indicator (ERI): At ERI, the nominal life of the device is three or six months. When the device exhibits signs of ERI it should be replaced expeditiously. Patient follow-up visits should be scheduled at an appropriate

frequency so that ERI can be detected well before End-of-Life (EOL).

Noninvasive Programmed Stimulation (NIPS): Life -threatening ventricular tachycardia or fibrillation may occur during NIPS, there fore: (1) closely monitor the patient, and (2) make defibrillation and resuscitation equipment, and trained

personnel, readily available during testing. Only physicians trained in tachycardia induction and reversion protocols should use NIPS. For more information on NIPS, refer to the programmer's on-screen help.

Ventricular Support Pacing during NIPS testing (Dual-chamber pulse generators, CRT-Ps) is delivered in the VOO mode. The specific indications and contraindications for VOO mode can be found on the programmer’s on -screen help.

Precautions: For single use only. Device Communication: Communication with the device can be affected by electrical interference and strong magnetic fields. If this is a problem, turn off nearby electrical equipment or move it away from

the patient and the programmer. If the problem persists, contact St. Jude Medical. Suboptimal RF Communication: For devices w ith RF telemetry capability, the Merlin™ PCS indicates the quality of the RF communication by the telemetry

strength indicator LEDs on both the programmer and the Merlin™ Antenna. CT Scans: CT scans, due to their increased power leve ls and long exposure times, have the remote possibility of interfering with implanted devices. The potential

interference is transient and occurs only when the X-ray signal is present. Continuous exposure may cause a temporary sensor rate increase. In addition, there is a remote possibility for a device to intermittently oversense while the CT

scanning beam is directly over the implanted device.

Unless otherwise noted, ™ indicates that the name is a trademark of, or licensed to, St. Jude Medical or one of its subsidiar ies. ST. JUDE MEDICAL and the nine-squares symbol are trademarks and service marks of St. Jude Medical, Inc.

and its related companies. © 2016 St. Jude Medical, Inc. All Rights Reserved.