retention behavior of polymer solute systems via … · b) retention index system it was not until...
TRANSCRIPT
RETENTION BEHAVIOR RETENTION BEHAVIOR RETENTION BEHAVIOR RETENTION BEHAVIOR OF POLYMEROF POLYMEROF POLYMEROF POLYMER––––SOLUTE SOLUTE SOLUTE SOLUTE
SYSTEMS VIA INVERSE GAS SYSTEMS VIA INVERSE GAS SYSTEMS VIA INVERSE GAS SYSTEMS VIA INVERSE GAS
CHROMATOGRAPHY CHROMATOGRAPHY CHROMATOGRAPHY CHROMATOGRAPHY
By
Prof. Dr. Ashraf Yehia Z. El-Naggar
and
Dr. Adil A. Gobouri
and
Eid H. Al–osaimi
International E – Publication www.isca.me , www.isca.co.in
INTERACTION CHARACTERISTICS OF POLYMERINTERACTION CHARACTERISTICS OF POLYMERINTERACTION CHARACTERISTICS OF POLYMERINTERACTION CHARACTERISTICS OF POLYMER––––
SOLUTE SYSTEMS VIA INVERSE GAS SOLUTE SYSTEMS VIA INVERSE GAS SOLUTE SYSTEMS VIA INVERSE GAS SOLUTE SYSTEMS VIA INVERSE GAS
CHROMATOGRAPHY CHROMATOGRAPHY CHROMATOGRAPHY CHROMATOGRAPHY
By
Prof. Dr. Ashraf Yehia Zaki El-naggar
Prof. of Analytical and Petroleum Chemistry
Chemistry Department, Science Faculty, Taif University,
Taif-Al-Haweiah-P.O. Box 888 Zip Code 21974, KSA.
Egyptian Petroleum Research Institute, Nasr City, Cairo, Egypt.
Dr.Adil A. Gobouri
Assistant Professor of Organic Chemistry, Science Faculty, Taif University,
Kingdom of Saudi Arabia
.
Eid Hamed Al–osaimi
MSc in Analytical Chemistry from Chemistry Department, Science Faculty,
Taif University, Kingdom of Saudi Arabia.
2013
International E – Publication www.isca.me , www.isca.co.in
International E – Publication 427, Palhar Nagar, RAPTC, VIP-Road, Indore-452005 (MP) INDIA
Phone: +91-731-2616100, Mobile: +91-80570-83382, E-mail: [email protected], Website: www.isca.me , www.isca.co.in
© Copyright Reserved
2013 All rights reserved. No part of this publication may be reproduced, stored, in a retrieval system or transmitted, in any form or by any means, electronic, mechanical, photocopying, reordering or otherwise, without the prior permission of the publisher. ISBN: 978-93-83520-29-9
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association iii
PREFACE
Inverse gas chromatography is a relatively rabid and reliable method for the
determination of polymer- solute interaction coefficient.
Gas Chromatographic characterization of polymers as stationary phases was
studied through retention behavior, polarity through Rohrschneider scheme,
selectivity and resolution via inverse gas chromatography. This
characterization was shown to be very useful, leading to materials with
suitable properties for advanced applications, especially in the field of
mixture separation.
Thermodynamic parameters (ΔH, ΔG and ΔS) were determined using inverse
gas chromatography in order to help us to investigate the interaction
characteristics of the solutes of different polarities with the studied polymers
as stationary phases in gas chromatography.
Gas chromatographic evaluation of these polymers was done mainly through
their use in the analysis of selected mixtures of different polarities including
paraffinic hydrocarbons, aromatic hydrocarbons and alcohols. The
efficiency of separation for the studied mixtures was evaluated in terms of
resolution, and separation factor.
Dr. Ashraf Yehia El-Naggar Prof. of Analytical and Petroleum Chemistry
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association iv
TABLE OF CONTENT
Preface iii
Table of Content iv
Chapter 1: Introduction 1
(1-1) Retention parameters 3
(1-2) Types of Stationary Phases 7
(1-3) Characterization of Stationary Phases 21
(1-4) Efficiency of Gas Chromatographic Separation 25
Chapter 2: Solute- Stationary Phase Interaction 28
(2-1) Thermodynamic properties 28
(2-2) GC detectors 31
(2-3) Application of gas chromatography 38
Chapter 3: The common used materials and Methods 43
(3-1) Chemicals and symbols 43
(3-2) Optimum flow rate method 44
(3-3) Condition of gas chromatograph 44
(3-4) Stationary phase characterization 45
(3-4) Condition of thermodynamic parameters 46
Chapter 4: Characterizations of Stationary Phases 47
(4-1) Determination of optimum flow rate 47
(4-2) Rohrschneiders scheme 50
(4-3) Selectivity index 53
Chapter 5: Thermodynamic parameters 55
Chapter 6: Separation mechanism 63
Chapter 7: Chromatographic Application 65
(7-1) Separation of normal paraffinic hydrocarbons 67
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association v
(7-2) Separation of aromatic hydrocarbons 69
(7-3) Separation of alcohols 72
Conclusion 75
Acknowledgement 77
References 78
About Authors 89
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
1
CHAPTER 1:Introduction
Chromatography is an analytical method for separating
components of mixtures. This process involves distribution of a solute
component between two phases: a mobile phase and a stationary phase. Gas
chromatography (GC) plays, nowadays, a vital role in many analytical
laboratories. This is due to the elegance and the versatility of such technique
which find so many applications in different areas such as petroleum industry
for assessing petroleum and derivatives quality, environment,
pharmaceuticals, chemicals, food science and many more.
The main driver for GC, as a separation technique, is the column and
the stationary phase i.e. the packing. Thus, using packing that is characterized
by high selectivity and effectiveness is a base requirement for gas
chromatographic separation of complex mixtures. So, the demand for even
more stable and efficient phases gave impetus to further developments. This
may be realized by discoveries of new selective stationary phases and by
modifying known stationary phases and column surfaces. Modifications of
stationary phases continue with advances in deactivating and coating
techniques such as coating process (surface modification of adsorbents). This
permits for many applications to be optimized and realized, which had been
earlier thought to be impossible by GC.
In gas liquid chromatography (GLC), the stationary phase is a liquid dispersed
on an inert support, while the mobile phase is an inert gas. The gas
chromatographic separation of a given mixture of components depends on
their distribution coefficient between the two phases. By using selective
stationary phases, gas liquid chromatography provides good separation of
components with nearly the same molecular weight and boiling points. The
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
2
separated components are detected with a suitable detector at the end of the
column. Thus, gas chromatography was used to identify and determine the
quantity of each component of the mixture to be analyzed.
On the other hand, although capillary columns have introduced new
methodologies in GC, Packed columns still offer some preferable
characteristics compared to capillaries. Primarily, allowing the entire samples
to be injected without the need of a splitter. This yields greater sample
sensitivity. Thus, the advantages of packed column are due to volume of
stationary phase per unit column length, in addition to the ease of column
preparation, resistance to contamination, short retention time.
Solid adsorbents such as molecular sieves, porous polymer beads,
graphitized carbon black, alumina, and silica gel are widely used as stationary
phases for packed column. This shows highly selective separation especially
for gases, volatile polar and non polar compounds, sulfur gases and
hydrocarbon impurities air pollutants e.g. CO, CO2 in air, solvents…..etc.
Moreover, silica-bonded phase chromatographic columns are the most widely
employed for the analysis of a great diversity of compounds by different
chromatographic techniques.
In gas solid chromatography (GSC), the stationary phase is a solid that
acts as adsorbent and packed in the column. The components of the mixture
distribute themselves between the gas phase and the adsorbed phase on the
surface of the solid (1).
Silica gel and alumina can be easily prepared and modified by
calcinations and special chemical treatments such as coating by a Carbowax
polymeric material (2, 3). These modifications offer the additional advantages
of improving separation and reducing peak asymmetry. The effect of different
concentrations of silica and alumina in aluminosilicates and the correlations
between the types of modification and the surface characteristics (e.g., surface
area, pore volume, pore radii, number of surface –OH groups) and gas
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
3
chromatographic separation are studied in order to select the adequate support
or stationary phase to elute the components of the mixture.
In order to evaluate the modified stationary phases, one should follow
some gas chromatographic parameters such as column efficiency, polarity of
stationary phases, thermodynamic parameters, etc. In this context, some
theoretical aspects will be presented.
(I-1) Retention parameters
Over the history of gas-liquid chromatography, thousands of substances
have been used as stationary phases. General considerations that influence the
choice of a particular liquid for use as a stationary phase in gas-liquid
chromatography include the liquid should be unreactive, have low vapor
pressure, good coating characteristics (wet the materials used in column
fabrication) and reasonable solubility in some common volatile organic
solvent.
The most popular liquid phases for gas chromatography are poly
siloxanes. The properties of these phases include a wide temperature
operating range, acceptable diffusion and solubility properties for different
solute types, chemical inertness, good film forming properties and ease of
synthesis with a wide range of chromatographic selectivity. The poly
siloxanes used in gas chromatography are generally linear polymers. In
addition, high molecular weight hydrocarbons such as hexadecane, squalane
(C30H62), Apolane-87 (C87H176) and Apiezon greases have long been used as
nonpolar stationary phases in gas chromatography. Moreover, they are used as
nonpolar reference phases in various schemes proposed to measure the
selectivity of polar phases. On the other hand ether and ester phases are
almost used in packed column gas chromatography for the separation of polar
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
4
compounds (e.g. succinate and adipate esters of ethylene glycol, diethylene
glycol and butanediol).
a) Retention time and retention volume
The chromatogram is the graphical representation of the elution process into
the chromatographic column. Retention times or retention volumes are the
most common parameters used to measure retention within the column.
Compounds are eluted from the column in the inverse order of the magnitude
of their distribution coefficients with respect to the stationary phase (4), thus,
peaks for less-retained solutes emerge before the peaks for highly-retained
solutes.
The mobile phase is assumed to be completely inert to the column, so no
interactions between mobile phase and stationary phase take place. The
required time for the mobile phase to be eluted from the column is the so-
called dead (hold-up) time (tm), which is the time for any solute not
interacting with the stationary phase to be eluted. As solutes interact with the
stationary phase, they are eluted after the mobile phase and the time between
solute injection and the emerging the solute from the column is called
retention time, the difference between mobile phase retention time and solute
retention time is so-called adjusted retention time (t*r ). The adjusted retention
time is defined by Eq. 1
t*r = tr – tm (1)
The volume of the mobile phase in the column is called Hold-up volume
(or dead volume) Vm and the volume of mobile phase necessary to enable the
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
5
solute migration throughout the column from the moment of entrance to the
moment in which it leaves is called retention volume Vr
the difference between mobile phase retention volume and solute retention
volume is so-called adjusted retention volume (V*r). The adjusted retention
volume is defined by Eq.
V*r = Vr – Vm (2)
Times and volumes in GC are related by the flow rate, F, at the pressure of the
column outlet. If the hold-up and retention time, tm and tr. are multiplied by F,
the gas hold-up volume Vm and the total retention volume of the solute, Vr
can be determined respectively.
Vm = tm . F (3)
Vr = tr . F (4)
The adjusted retention volume, V*r, can be obtained in the same way
V*r = t*
r . F (5)
b) Retention index system
It was not until the late 50’s when Kováts (5) proposed his well-known
retention index system in terms of isothermal retention indices. This system
uses n-alkanes as reference substances, which are non-polar, chemically inert
and soluble in most common stationary phases.
In the Kováts system, the retention index of a substance I is equivalent to 100
times the number of carbon atoms of a hypothetical n-alkane with the same
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
6
adjusted retention time, adjusted retention volume, specific retention volume,
etc. The retention index values for the n-alkanes, used as the fixed points on
the retention index scale, are defined as 100 times the carbon number of the n-
alkane on all phases. Consequently, the retention index of any substance “X”
is calculated from Eq. (6) by co-injection of X with two n-alkanes usually
differing by one carbon number,
Ix = 100 n + 100 Error! (6)
where t*n’ and t*
n+1 are adjusted retention times (or adjusted retention volume,
specific retention volume, etc.) of the reference n-alkane hydrocarbons eluting
immediately before and after substance “X” and t*x is the adjusted retention
time for substance “X”.
C). Retention Factor ( k )
The Retention Factor is another measure of retention. It is the ratio of the
amount of time a solute spends in the stationary phase and mobile phase
(carrier gas). It is calculated using Equation 7. The retention factor is also
known as the partition ratio or capacity factor. Since all solutes spend the
same amount of time in the mobile phase, the retention factor is a measure of
retention by the stationary phase. For example, a solute with a k value of 6 is
twice as retained by the stationary phase (but not the column) as a solute with
a k value of 3. The retention factor does not provide absolute retention
information; it provides relative retention information. An unretained
compound has k = 0.
k = Error! = Error! (7)
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
7
(I-2) Types of Stationary Phases
The number of stationary phases suitable for gas chromatography is
quite large, and the choice is dictated largely by the nature of the sample.
Stationary phases are described as non-polar or polar according to their
structure and separating abilities. In general, the most suitable stationary
phase for a given sample is that which is chemically similar to it. There are
three types of stationary phase: solid stationary phase, liquid stationary phase
and liquid-solid stationary phase.
a) Solid stationary phases (GSC):
Bare porous solids have been used as GC stationary phases in the gas–
solid chromatography (GSC) mode. These solid adsorbents are generally
more stable over a wider temperature range and less sensitive to oxygen than
their coated counterparts. GSC often affords much better selectivity for the
separation of geometric and isotopic isomers, and is also well suited for the
separation of permanent gases and small hydrocarbons. However, the
development of this promising technique has been slowed by its intrinsic
difficulties. For example, adsorption isotherms in GSC are often non-linear,
leading to retentions that vary with sample volume, to asymmetric peaks and
to incomplete resolutions. Additionally, the very large surface area typical of
some solids have lead to excessively long retention times, thus limiting the
broader use of GSC. Despite this drawback, the specific advantages of GSC
and its unique separation abilities in some applications have recently lead to
increased interest in the technique.
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
8
The most important adsorbents that can be used as solid stationary
phase are molecular sieves, carbosieves, porous polymers, silica gel and
alumina.
1) Molecular sieves
Molecular sieves have a pore size defined by the geometric structure
of the zeolite (2). Typical pore size is 4–10 Ǻ (0.4–1 nm) which accounts for
the unusually large specific surface area. It has been used mainly for the
separation of permanent gases. The separation on such adsorbent is based on
multiple-retention mechanisms. The first selection will be done on size.
Molecules that are smaller than the pore size will diffuse inside the pores.
Large molecules are difficult to enter the pore and will elute earlier. The
second mechanism of separation involves adsorption at active sites. This
adsorption mechanism is responsible for the high retention of permanent
gases. The retention of components with dipole interaction and hydrogen-
bonding formation is very high. As a result, carrier gas and samples should be
as dry as possible. Water is absorbed by the molecular sieve and will cause a
reduction of retention time. However, the adsorbed water can be removed by
heating the molecular sieve for a few hours at 300oC.
2) Porous polymers
The porous polymers, like the Porapaks, HayeSep, GasChrom,
Tenax-GC, and the Chromosorb Century Series, are used in special fields of
applications such as the analyses of solvents, acids, and alcohols in water.
Porous polymers are prepared by the co-polymerization of styrene and divinyl
benzene or other related monomers. The pore size and specific surface area
can be varied by the amount of monomer added to the polymer. Porous
polymers are also available with different selectivity. By incorporating vinyl
pyridine or methacrylate groups, the general selectivity can be changed and
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
9
the porous polymer can be made much more polar. Polar porous polymers
have been reported, such as the Pora PLOT S (medium polarity) and Pora
PLOT U (polar) type.
3) Deposited carbon adsorbents
Carbon materials are highly selective for hydrocarbons(3). Because of
the non-polar characteristics of its surface, carbon will retain unsaturated
hydrocarbons. Carbon materials show a high tolerance for acidic matrices so
this type of material is ideal for analyzing volatile compounds in a strongly
acidic matrix. Carbopack and Graphpac graphitized carbons are ideal for
analyzing many kinds of C1 and C10 compounds, including alcohols, free
acids, amines, ketones, phenols, and aliphatic hydrocarbons. Carbopack B
(Graphpac-GB) and Carbopack C (Graphpac-GC) carbons have surface areas
of about 100 and 12 m2/g, respectively. Therefore, Carbopack B packings will
have a larger sample capacity than Carbopack C packings. Carbopack B HT
or C HT is hydrogen treated for deactivation.
4) Silica gel
Silica gel is a common sorbent for the chromatographic separation of
organic compounds like C1–C8 hydrocarbons, volatile sulfur compounds, and
halogenated compounds in the C1–C6 range. The structure of silica gel is a
matrix of particles consists of silicon atoms joined together with oxygen
atoms by siloxane bonds (Silicon-oxygen-silicon bonds). On the surface of
silica, some residual, uncondensed hydroxyl groups from the original
polymeric silicic acid remain. These residual hydroxyl groups confer upon
silica gel its polar properties. The silica surface is quite complex and contains
more than one type of hydroxyl group, strongly bound or 'chemically'
adsorbed water, loosely bound or 'physically adsorbed' water and hydrogen
bonded silanol group. There are three types of hydroxyl group. The first is a
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
10
single hydroxyl group attached to a silicon atom which has three siloxane
bonds joining it to the gel matrix. The second is one of two hydroxyl groups
attached to the same silicon atom which, in turn, is joined to the matrix by
only two siloxane bonds. These twin hydroxyl groups are called Geminal
hydroxyl groups. The third is one of three hydroxyl groups attached to a
silicon atom which is now only joined to the silica matrix by only a single
siloxane bond.
The chemical and adsorption properties of silica gel depend on number
and reactivity of the surface silanol groups. Two forms of adsorption centers
may exist over the surface of silica gel, 1) Hydroxyl group, 2) Coordinated
unsaturated atoms of silicon and surfaced electronegativity atoms of oxygen
nascent at dehydration of silica gel. The surface of dehydrated silica gel
begins slightly to chemisorb oxygen, most likely on the centers of -Si type.
The main purpose of silica surface modification is shielding of the
active silanol groups and attachment to the accessible adsorbent surface
organic ligands which are responsible for specific surface interactions.
5) Alumina
Alumina (Al2O3) is known as a good adsorbent for gas-solid
chromatographic separation of all C1–C4 hydrocarbons and some geometric
isomers can be separated as baseline. But practical uses of alumina are
restricted because of its large adsorption capacity. However, this capacity can
be controlled by introducing an organic group into the alumina surface. It is
known that some organic groups can be easily fixed chemically on the
alumina surface.
Alumina exists in eight different polymorphs, seven metastable phases
(γ, δ, κ, ρ, η, θ, and χ) and the thermally stable α-phase. γ-alumina, called
active alumina which is formed by low temperature dehydration (< 4500C) of
boehmite (γ-AlOOH). Upon heating, γ-Al2O3 undergoes a series of
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
11
polymorphic phase transformations from a highly disordered cubic close-
packed lattice to the more ordered cubic close-packed θ-Al2O3. At higher
temperatures, e.g., 1100oC–1200oC, θ-Al2O3 undergoes a reconstructive
transformation to form the thermodynamically stable hexagonal close-packed
α-phase. Generally, this transformation sequence may be illustrated as
follows:
γ-Al-OOH450oC γ-Al2O3
900oC θ1100-1200oC α
b) Liquid stationary phases (GLC)
Over the history of gas-liquid chromatography, thousands of substances have
been used as stationary phases. General considerations that influence the
choice of a particular liquid for use as a stationary phase in gas-liquid
chromatography include the liquid should be unreactive, have low vapor
pressure, good coating characteristics (wet the materials used in column
fabrication) and reasonable solubility in some common volatile organic
solvent.
The most popular liquid phases for gas chromatography are poly siloxanes.
The properties of these phases include a wide temperature operating range,
acceptable diffusion and solubility properties for different solute types,
chemical inertness, good film forming properties and ease of synthesis with a
wide range of chromatographic selectivity. The poly siloxanes used in gas
chromatography are generally linear polymers. In addition, high molecular
weight hydrocarbons such as hexadecane, squalane (C30H62), Apolane-87
(C87H176) and Apiezon greases have long been used as nonpolar stationary
phases in gas chromatography. Moreover, they are used as nonpolar reference
phases in various schemes proposed to measure the selectivity of polar
phases. On the other hand ether and ester phases are almost used in packed
column gas chromatography for the separation of polar compounds (e.g.
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
12
succinate and adipate esters of ethylene glycol, diethylene glycol and
butanediol).
Advantage of Liquid Stationary Phase
For the great majority of analytical problems, gas liquid
chromatography has proved to be a very efficient technique.
The use of liquid phases has the following advantages (6):
1) Liquid phases are available in great variety. Thus, adequately selective
phases can be found for a particular separation.
2) The amount of liquid phase in the column can be varied easily.
Therefore, both preparative and high efficiency columns can be made
with the same liquid phase.
3) Liquid phases are available in great purity and in well-defined quality.
Thus, retention values are reproducible.
4) Both packed and open tubular column can be packed with liquid phases
in a simple manner.
In some cases, the volatility of stationary phase could lead to undesirable
bleeding which can disturb the proper function of the detector. The bleeding
of stationary phase causes baseline shift in programmed operation,
contaminates the trapped solutes, and changes the separation character of the
column until an unacceptable deterioration occurs.
In 1961 Martin (7) suggested that adsorption of solute at the gas-liquid
interface might make an important contribution to solute retention. He studied
hydrocarbon solutes (non-polar) on three stationary phases, n-hexadecane
(non-polar), 1-chloronaphthalene (moderately polar) and B,B`-thiodipropio-
nitrile (TDPN, strongly polar). He found that the strength of adsorption
relative to solution increased with increasing the polarity of the stationary
phase. Thus, with non-polar hydrocarbon solutes, liquid surface adsorption
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
13
occurred for polar but not non-polar stationary phases. This general
conclusion was reproduced and extended by Pecsok et al (8) for a wide range
of hydrocarbon solutes with the same stationary phases, and directly
substantiated by the static measurements of Martire et al (9) on benzene and
cyclohexane in TDPN. The results of Martin (7) and Pecsok (8) show that
saturated hydrocarbon solutes are more subjected to adsorption at the liquid
surface of TDPN than aromatic or polar molecules. Thus, saturated
hydrocarbons eluted rapidly with respect to aromatic since their mechanism
of retention is adsorption. Similar findings were observed by Pecsok and
Gump (10).
Choice of the particular liquid stationary phase depends on the column
temperature to be used and on the chemical nature of the compounds to be
separated. If tailing occurs it usually means that adsorption is taking place on
the solid support and it can be often controlled by the use of more polar
stationary phase or by the pretreatment of the support with such substances
such as dimethyldichlorosilane (DMCS) or hexamethyl-disilazane (HMDS).
Polymers as stationary phases
A wide variety of polymeric liquid-phases have been developed as the
stationary phases in gas chromatography (11, 12). Various deactivation and
polymer-coating processes have been also proposed for the preparation of GC
columns with good separation performance and stability.(13, 14). In contrast to
the successful applications of these polymer-coated columns, such as wall-
coated, support-coated and pellicular-coated . The polymer-packed stationary
phases, however, have been limited, except for characterizing the surface of
the stationary phase in inverse GC (IGC)(15, 16). In general, the surface
characteristics of the stationary phase have been analyzed with IGC technique
by injecting the standard samples into a polymeric liquid-phases stationary
phase -packed column and measuring the elution behavior.
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
14
This characterization method is based on the observation of the specific
interactions between the surface of the polymeric liquid-phases stationary
phase and the standard solutes injected as the probes. Consequently, it is quite
natural that a synthetic polymeric liquid-phases stationary phase can be
employed as a separation medium in GC, if the polymeric liquid-phases
stationary phase possesses both the thermal stability for the operation at an
elevated temperature and the resistance to the gaseous chemical species
through the column during the separation.
A. Y. El-naggar and G. Turky (17) evaluated three types of polyethylene
glycols of different molecular weights namely, 600, 4000 and 20000 as liquid
stationary phases in gas liquid chromatography. They investigated the
retention mechanism for the studied polymer stationary phases of 15% by
weight on chromosorb PAW and their thermodynamic parameters via inverse
gas chromatography. They found that the separation of n-alcohols is obtained
toward polyethylene glycol (PEG20000) of relatively higher molecular
weight. Saturated hydrocarbons can be separated very efficiently using low
molecular weight polyethylene glycol (PEG600).
A. Y. El-naggar and G. Turky (18) studied the effect of polymer-layer
thickness on the polarity dielectrically. The dependence of specific retention
volume, enthalpy, and entropy upon the loading of polymer on support was
also studied. In all investigated polymers, the loads 5 and 10% deactivate the
support surface, resulting in a decrease in its capacitances, which reflects the
polarity. The loads 15% PEG20,000 and 25% PEG4000 can elute n-alcohols
and exhibit high efficiency of separation; however, in the case of PEG600,
loading has to exceed 25% to be sufficient for eluting n-alcohols. The lower
thickness of the coated polymers was preferred for good separation of
saturated hydrocarbons and also for cyclohexane and aromatic hydrocarbons.
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
15
J.K.Suh.etal(19) prepared Poly(p-tert-butyltrimethoxymonopropyl -
oxycalix[4larene-methylsiloxane) as a stationary phase in isothermal capillary
gas chromatography exhibiting good separation of some positional isomers
which can not easy to separated by other technique.
R.P.Corent. et al(20) synthesed Highly crosslinked functional polymer particles
with narrow size distribution by precipitation copolymerization of
divinylbenzene, ethylene glycol dimethacrylate and vinylbenzyl chloride
using a simple reflux protocol. These particles were tested as stationary phase
in high-performance liquid chromatography for the separation of polycyclic
aromatic hydrocarbons in reversed-phase mode.
X.Sun . et al (21) synthesed a novel ionic liquid (IL) bonded polysiloxane and
another one with chloride anion ([PSOMIMI[Cl1) was also prepared and
applied them in the separation of fatty acid methyl esters exhibiting good
selectivity.
C.Y.Liu. et al (22) Compared and evaluated of copper complex-containing
siloxane polymers as stationary phases for capillary gas chromatography, and
applied these columns to the analysis of phthalate esters.
Solid Support
It is known that solids are more thermally stable than liquids, and
possess lower vapor pressures. Therefore, columns packed with active solids
are not subjected to bleeding at high temperatures. The most obvious
disadvantage of solid adsorbents is that they frequently give rise at low
concentrations of adsorbent to strongly curved isotherm; this leads to tailing
with consequent loss of efficiency in elution analysis. In addition, they are
difficult to prepare with consistent properties from batch to batch, so that
retention data may have to be continually checked if the column packing is
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
16
frequently damaged. Such disadvantages could be prevented by appropriate
pretreatment.
Many active solids are efficient catalysts, whereas catalytic activity is
rare in gas–liquid chromatography. Catalysis may result in the irreversible
adsorption of a substance on the column, or the complete or partial
conversion of substances giving rise to the artifacts in the chromatogram.
The function of a solid support is to hold the liquid phase used for packed
column in gas liquid chromatography immobile during the separation process.
It should be inert, easily packed and has a large surface area. The surface must
have sufficient energy to hold the stationary phase in place and to cause it to
wet the surface in the form of a thin film. In addition, it should be a good
conductor of heat and mechanically and thermally stable to avoid changes in
properties on handling and while in use.
In general, the support influences both the retention time and the width
of the peak. the nature of the support is a main factor controlling peak width,
because it determines both the liquid distribution and gas phase flow path. On
the other hand, the presence of active sites on its surface leads to undesirable
adsorption effects which cause tailing and may result in catalyzed
decomposition or rearrangements of the solutes passing through the column.
The adsorption phenomena have the greatest effect when the solutes and the
stationary phase greatly differ in their polarities, this illustrated as following,
adsorption of analytes on the support is most pronounced for polar substances
chromatographed on non polar stationary phases. In contrast, if the stationary
phase is polar, the effect of the support is smaller because the support active
sites are blocked by the functional group of the stationary phase.
There are different materials suitable for used as solid support in gas
chromatography such as silica, alumina, and other new materials. In the
following a brief review of the supports:
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
17
1- Silica Support
One of the great attractions of the use of silica as support for
chromatographic columns is the ease with its surface can be modified with
different groups or polymers by reaction with the silanol groups.
Diatomite
Diatomite (diatomaceous earth) is composed of the skeletons of
diatoms; the skeletal material is essentially amorphous silica with small
amounts of alumina and metallic oxide impurities. There are two type of
diatomite supports prepared and used in gas liquid chromatography. The first
one is a Chromosorb P (pink supports) which contains complex iron oxides
give this material its characteristic pink color, calcination at temperature
exceed of 900°C is used to agglomerate and strengthen this natural material.
The second type is Chromosorb W (white support) which prepared by
calcining diatomite in the presence of a small amount of sodium carbonate
results in the formation of a white material in which metal impurities are
converted to colorless sodium silicates. The bulk chemical composition of the
two types of diatomaceous earth is nearly similar but there is a great different
in their surface area (18). The Chromosorb W has a smaller surface area and
slightly basic character when compared to the Chromosorb P. Therefore, pink
support gives higher column efficiency but white support has a lower
adsorpitivity (19, 20). Chromosorb G is a specially prepared white support that is
harder, more robust, and inert and has a low loading capacity but higher
density than the regular white supports.
The existences of a wetting transition are closely associated phenomena
on silanised diatomaceous earth supports on gas liquid chromatography. these
phenomena occur only when the surface tension of the non-wetting liquid
exceeds the critical surface tension of the support by an amount small enough
to give a contact angle which, though not zero, is still low. However, Liquids
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
18
such as PEG-400 with a much higher surface tension have a large contact
angle and small surface area. PEG-400 is not distributed throughout the
interior and exterior of each support but is largely restricted to the outer
surface of the particle as a thick non-uniform layer. The existence of a sharp
wetting transition depends not only on surface tension relationship but also on
the pore structure of the silanised support. The pink support with the high
density possess the necessary combination of porosities: a small porosity in
pores sufficiently fine to tie up a quantity of liquid by capillary action, and a
large porosity in larger pores, characteristic of the diatomite structure having a
narrow spread of pore size but still small enough to exhibit capillary action. In
contrast, the pores of the white support are too large for a wetting transition to
exist.
In gas liquid chromatography, to hold the stationary phase in the
column as a liquid of high surface area, a solid support is required. Therefore,
the support must be a porous material of high surface area and low surface
activity. In fact the nature of the support is a main factor controlling peak
width because it determines both the liquid distribution and gas phase paths (23).
Adsorption of a solute by the solid support contributes to retention in
the majority of GLC systems, but to very different extent, depending on the
nature of the solute, solvent and support. This adsorption may be due to
incomplete coverage which can occur either because the liquid does not wet
the surface or because it is coated at too low a liquid loading to form a
monolayer on a wetted surface (24).
In the case of wettability of the support by the stationary phase, the
way in which the liquid is distributed on the support depends on the structure
of the solid and on the relative surface tensions of liquid and solid surfaces
and of the solid-liquid interface. In this context, one should point out the
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
19
classification of the solid supports into unwetted supports and wetted
supports.
Coating of solid support
This treatment involves the coverage of the support with polymers or
metals to forms a stable layer on the surface, with the result that the adsorbent
becomes more homogenous and the chromatographic peaks more
symmetrical. In the following a brief review of the coating technique:
Coating with polymers
Supports are usually coated with a small proportion of non-volatile
polymers as stationary liquid phase such as PEG, SE-30...and etc using one of
several evaporation methods. In outline, the stationary phase in a suitable
solvent is mixed with the support, the solvent removed, and the dried packing
added to the empty column with the aid of pressure or suction (24-27).
Coating of alumina support with moderately polar stationary phase
(polypropylene glycol 2000) system was studied (28). It was found that, the
solid support surface is assumed to be successively covered with three
different types of liquid phase layers, first with a monolayer, second with a
double layer and finally with a bulk liquid layer, as the liquid loading
increases.
Hayrapetyan et al(29), studied the influence of the coating with
several polymers polyoctadecyl-methacrylate-methylmethacrylate co-
polymer, polyacrylamide gel, polyvinyl alcohol, poly-2-
hydroxyethylmethacrylate on the surface characteristic of silica gel
having different pore diameter. It was found that:
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
20
i) A large decrease in specific surface area was observed in the case of
polyoctadecyl-methacrylate-methylmethacrylate co-polymer.
ii) It is more appropriate to use wide-pore silica with a higher pore
volume for the modification with a polymer.
iii) In the silica gel with small pores and a lower pore volume, a
significant decrease of the specific surface are observed.
Coating with metals
Several studies were done dealing with coating of silica and
alumina with various inorganic salts in order to obtain interesting
chromatographic characteristics. For example, Kopecni et al(30),
proved that modification of silica with the alkali metal ions resulted in
decreased partition coefficients than that of unmodified silica. On the
other hand great selectivity for alumina metal modifications is
proposed e.g. Alkali-metal fluorides to give good separations of high
boiling point hydrocarbons solutes and geometric isomers (31), trialkali
metal phosphate and heat treatment before and after coating to give
good resolution for geometric isomers of aromatic hydrocarbons and
their halogenated derivatives (32), also sodium chloride modified
alumina showed good peak resolution for separation of geometric
isomers of 2-olefins and polychloro derivatives of benzene (33).
In another study, alumina surface was modified with various
alkali salts to improve the separation of alkenes/alkanes pair (34). The
study examined the change in retention properties on modified
alumina and the specific effects of the halide anions and metal
cations. It is postulated that the modifier anion may have a role in
reduction of active sites with the consequently elimination of the peak
tailing of unsaturated compounds. However, it is believed that this
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
21
process occurs readily with no predominant group trend. The metal
cations have a group trend for increasing alkenes/alkanes separation
selectivity with decreasing cation size is in the following order Li+ >
Na+ > K+ > Cs+. Strgelczyk et al(35), modified alumina by
impregnation with palladium as small particles to improve the
selectivity of hydrogen isotopes separation. It was found that, at the
same palladium loading alumina is the most efficient because of the
higher Pd mass content per unit volume of the column. Also, Lee et
al(36), found that column packed with palladium–platinum alloy on
alumina (PPA) exhibits good separation efficiency of hydrogen
isotopes separation.
Recently, Khan et al(37), found a good column packing material
for separation of light hydrocarbon gases by coating alumina with
Fe2O3. The selectivity for alkenes/ alkanes pair has been improved
due to the small ionic radius (0.64Å) and high electronegativity (1.8)
of Fe3+, which increased the interaction of π electrons of alkenes with
dipole Lewis acid site of the aluminum ions and Fe3+ on the modified
alumina surface(38).
(I-3) Characterization of Stationary Phases
The chromatographic retention is based on the interactions between the
solute and the stationary phase, which include directional force, induction
force, dispersion force, hydrogen bond and others. The difference of the
retention is due to the molecular structure of the solute and the properties of
the stationary phase. Thus, the characterization of stationary phases fall into
two types: the first is estimated by chromatographic tests for selected
compounds, these tests include polarity, selectivity and thermodynamic
parameters. The second is the surface characterization for understanding
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
22
adsorption and desorption phenomena including surface texture (surface area,
average pore size and pore size distribution) and surface chemistry (functional
groups and their distribution) of porous solids. Generally, the characterization
of stationary phases is useful to elucidate retention mechanisms and predict
the retention parameters (39-46).
(a) Polarity of Stationary Phase
Polarity of stationary phase is determined by stationary phase structure,
polarity of functional groups and amount of each group. In order to compare
stationary phase polarities, a systematic method must be used to normalize
instrumental variables; only then can retention data be effectively compared
for different chromatographic system.
Polarity of stationary phases has been assessed by considering the
retention time data for selective solutes of different polarities with the aim of
estimating what so called a retention index. The retention index of an
arbitrary substance X is found by logarithmic interpolation between two
neighbouring solutes: equation (6) .
If the alkanes are chromatographed simultaneously with substance X or, at
least, are chromatographed under identical condition, then Vg in equation (6)
can be replaced by any adjusted retention quantity, i.e. the adjusted retention
time or volume, capacity ratio k, etc.
The idea of the retention index was firstly introduced by Kovats (5).
The Retention Index (I), called Kovat's retention index (5), indicates where
compounds will appear on a chromatogram with respect to straight-chain
alkanes injected with the compound. Kovat's retention index of an arbitrary
sample component X is found by logarithmic interpolation between two
neighbouring solutes as given in formula 6.
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
23
Each alkane assigned an arbitrary I value 100 time its number of carbon
atoms. The advantages of retention index system attributed to, the constant of
I values for n-alkanes on each stationary phase which determined by actual
injection of the n-alkanes. Accordingly, I values provide a way of comparing
the retention characteristics of various stationary phases and also to compare
relative elution orders of various compounds for a given column and
conditions or for two different columns.
Kovats considered the retention data of n-alkanes. The retention index is a
constant characterizing the intensity of interaction between a solute and a
stationary phase at a given temperature (47). If the less polar stationary phase is
chosen as the reference phase, (the phase that gives the smallest index for the
chosen solute), the polarity is measured by the difference,
ΔI = I b – I
a (8)
Where, a is the less polar stationary phase, b is the more polar stationary
phase, and the solute used is solute other than an alkane.
On the basis of Kovàts retention indices, Rohrschneider(48) originally
proposed that the differences among the indices should be-expressed in terms
of three parameters. Later, it was found that, to take into account the basic
types of interaction, ΔI must be expressed by the sum of fine products of
factors characterizing the polarity of the substance (a to e) and that of
stationary phase (x to s) (49-52).
ΔI = ax + by + cz + du + es (9)
Five standard compounds were selected, that would reflect the main
types of interactions. The orientation forces are represented by factor c (ethyl
methyl ketone), the charge transfer forces by factor a (donor, benzene) and d
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
24
(acceptor, nitro methane) and hydrogen bonding by factor b (hydrogen donor,
ethanol) and e (hydrogen acceptor, pyridine). In order to characterize the
polarity of a particular stationary phase, the differences between the retention
indices for these five standard solutes obtained on this phase and on non polar
phase must thus be determined.
(b) Selectivity of Stationary Phase:
The selectivity of a given stationary phase is the ability of the
stationary phase to separate a pair of solutes having very similar properties
such as boiling points, molecular weights, number of carbon atom in the
molecule and vapor pressures. The selectivity is not an intrinsic property of
stationary phase: it depends on the choice of solutes and can be expressed in
different terms such as the value of separation factor for estimating the degree
of separation of two component, and uniformity criterion for estimating the
degree of separation of multi component mixture. Early attempts to define
selectivity scales were based on the systems of characteristic phase constants
introduced by Rohrschneider (48) and subsequently modified by McReynolds
(53), Synder ̉’s(54) and Kersten(55) studied solvent selectivity triangle, selectivity
indices (56) Hawakes polarity indices (57) , solubility parameters (58-60), and the
partial molar Gibbs free energy of solution for functional groups or specified
solutes (61-67). The selectivity of the stationary phase is defined as its relative
capacity to enter into specific intermolecular interactions such as dispersion,
induction, orientation, hydrogen-bond formation, and charge transfer
complexation (68).
Selectivity can be expressed in terms of uniformity criterion which
introduced by T. G. Andronikiashvili et.al (69). He used the uniformity
criterion for estimating the degree of separation of multi component mixture;
the uniformity criterion Δwas given according to the following equation:
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
25
t
τKn
Δeffk
= (10)
where,k
n is the number of peaks on the chromatogram,τ is the base width of
the narrowest peak,eff
K is the separation factor for the worst
separated pair of components, and t is the duration of analysis.
The values of Δ varied from 0 to 1; Δ=1 corresponds to the best separation.
(I-4) Efficiency of Gas Chromatographic Separation:
(a) Column Efficiency
After conditioning the column a few preliminary tests are performed to
ensure that the performance of the column is adequate. A measure of
column efficiency is made with a test sample. In this section, a brief
overview of relevant relationships and equations is presented.
Number and height equivalent of theoretical plates:
The number of theoretical plates (N) is a measure of column efficiency.
A theoretical plate is defined as the average distance traveled in one
distribution step or partition of the analyte from the mobile phase into the
stationary phase and vice versa(70). The plate number can be directly
calculated from the retention time
N = 16 (tr /wb)2 (11)
N = number of theoretical plates (efficiency of a column)
tr = retention time of the peak
wb = is peak width at the base
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
26
Both retention time and plate number are directly proportional to the
length of the column. The longer the column, the longer the retention time
and the higher the plate count. Since the total number of plates for a particular
column is based on its length, it is useful to express column efficiency in
terms of height equivalent to a theoretical plate (HETP or H). This height,
expressed in micrometers, is the height of one theoretical plate. A small
HETP allows for more plates per column and greater efficiency. HETP can be
calculated as follow:
H = L / N (12)
H = height equivalent to a theoretical plate.
L = length of the column.
N = number of theoretical plates.
(b) Stationary phase efficiency
(1) Separation Factor
The degree to which two components are separated is a function of the
ratio of their retention times and the sharpness of the peaks. The ratio of the
adjusted retention times of two components, 1 and 2 is termed the separation
factor α
αααα = Error! (13)
The separation factor can also be expressed as,
(14)αααα = Error!
Where, k is the retention factor, which is equal to Error!, tm is the gas hold up
time or the retention time of an unretained solute.
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
27
With co-eluting solutes k2 = k1 and α = 1.0. By convention α is never less
than 1.0.
Solute pairs with large α values can be separated at high resolution
columns, but at low resolution α approaches unity, columns with increasingly
larger numbers of theoretical plates are required to achieve separation(71,72).
(2) Resolution Factor between Two Peaks (R)
The degree to which two components are resolved is termed resolution
R (73-75).
R = 2 (tr(2) – tr(1)) / (Wb(1) + Wb(2) ) (15)
When R = 1, then there is a considerable degree of overlap, and when
R = 1.5, the base line separation usually achieved but asymmetry, tailing, or
gross discrepancies between the sizes of the two peaks can cause
compilations.
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
28
CHAPTER 2. Solute-Stationary Phase Interaction
In gas chromatography, solutes elute in an order of their net vapor
pressures. This order depends on the interaction between the solute and the
stationary phases under a given set of conditions. For liquid stationary phase,
such interactions are a function of the solubility of the solute in the stationary
phase (76). For solid stationary phase, the adsorption of molecules on the
surface of the solid is influenced principally by the chemistry and textural
properties of the adsorbent surface.
Three types of solute – stationary phase interactions are of concern in gas
chromatography. These interactions include dispersion, dipole and proton
interactions (77, 78). The dispersion interactions constitute the major part of the
total attractive force between a hydrocarbon solute and the stationary phase,
and they are significant for any dissolved solute molecules and the
surrounding stationary phase solvent. Dipole interactions arise from the
alignment of the dipoles of the solute and the stationary phase. The base-acid
interaction (hydrogen bonding) could be attributed to hydroxyl groups, which
may be present in the solutes leading to the formation of hydrogen bonds with
the stationary phase. The ability of a stationary phase to participate in
hydrogen bonding is usually a measure of its basicity. The hydroxyl group
can however also behave as a base and is retained by Lewis and Brønsted
acids.
(2.1) Thermodynamic properties
Gas chromatography (GC) is a powerful and accurate method for measuring
the thermodynamic properties and investigating the properties of solvents and
the behavior of solutes in stationary phases. Polymers as stationary phases
demonstrate excellent chromatographic properties, such as a wide range of
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
29
operating temperatures, high column efficiency, good thermal stability, and
inertness resulting in low bleed at high temperatures and low detection limits.
Also, it is considered to be of unique selectivity in the separation of nonpolar,
moderate and polar compounds.
In order to study the retention and separation mechanism of this
stationary phases, the relation between retention volume and carbon
number(79, 80) and the thermodynamic parameters of solution, the enthalpy of
solution (ΔH), the free energy of solution (ΔG), and the entropy of solution
(ΔS) were measured.
Many others (81-85) studied the feasibility of GLC for the thermodynamic
investigation of different stationary phases and hydrocarbon solutes. Seifert
and Kraus (86) investigated the solution behavior of liquid crystal phases in
capillary GC throw their thermodynamic parameters.
The aim of this investigation is to study the thermodynamics of solution in the
three studied stationary phases using n-hexane , benzene and ethanol as model
solutes representing their families paraffinic and aromatic hydrocarbons and
to correlate the measured thermodynamic quantities and structures of studied
the stationary phases and solutes.
Specific retention volumes, Vg, were calculated from the corrected peak
retention times and the column operating condition by using the well-known
equation derived by Littlewood et al (87). The specific retention volume Vg is
given as:
Vg = (Vr − Vm) /w (16)
Where Vr is the elution volume of the probe under conditions of instantaneous
equilibrium between the stationary and mobile phases; Vm is the void volume
of the column (which is equal to the elution volume of an ideal marker; i.e., of
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
30
a probe that does not interact with the column at all); w is the mass of the
polymer on the column.
The elution volumes, Vr, were calculated from the flow rate, F, measured
at the column outlet and the elution time te as (James and Martin,(88)):
Vr = j tr F (17)
j is a correction factor to account for the compressibility of the gas.
The partial molar excess enthalpies ΔH of mixing at infinite dilution
were calculated from an equation formulated by Greene and Pust (89).
for the direct calculation of ΔH from the relation between the logarithm
of retention time and reciprocal of temperature. Accordingly, ΔH can be
determined from the slope of this relation:
tr = LAB/F exp(−ΔH/RT ) (18)
where tr = retention time, R = linear velocity of zone, A = intestinal area
of column, B = constant, F = gas velocity, and ΔH = heat of solution.
The excess partial molar free energy of solution can be expressed as
(Kessaissa et al.) (90):
ΔG = -2.3 RT ln Vg (19)
where T is the column temperature and Vg is the retention volume.
The entropy of solution can be calculated by knowing ΔH and ΔG from
the relation:
ΔS = (ΔH − ΔG)/T (20)
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
31
(2.2) GC detectors (91)
(a) Detector characteristics
The detector produces an electrical signal which is proportional to either the
concentration or the mass of the analyte molecules in the effluent stream. The
signal is displayed as a chromatogram on a chart recorder, or on the screen of
a desktop computer data system.
The general characteristics of GC detectors which need to be considered are
the following:
(1) Universality vs. selectivity: If a detector responds with similar sensitivity
to a very wide variety of analytes in the effluent it is said to be universal
In the other extreme, a selective detector may give a significant response
to only a limited class of compounds or possessing certain types of
functional groups or substituents which posses certain affinitie or
reactivities.
(2). Destructive vs. nondestructive: Some detectors destroy the analyte as part of
the process of their operation (e.g., by burning it in a flame or fragmenting it
in the vacuum of a mass spectrometer). Others leave it intact and in a state
where it may be passed on to another type of detector for additional
characterization.
(3) Mass flow vs. concentration response: In general, destructive detectors are
mass flow detectors. If the flow of analyte in effluent gas stops, the signal
drops to zero. A nondestructive detector does not affect the analyte, and the
concentration measurement depends on the volume of the effluent
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
32
(4) Requirement for auxiliary gases: Some detectors do not function well with the
carrier gas composition or flow rates from a capillary column effluent. Makeup
gas, sometimes the same as the carrier gas, may be required to increase flow
rates through the detector to levels at which it responds better and/or to
suppress detector dead volume degradation of resolution achieved on the
column. Some detectors require both air and hydrogen supplied at different
flow rates than the carrier to support an optimized flame for their operation.
Makeup flow dilutes the effluent but does not change the detection mechanism
from concentration to mass-flow detection.
(5) Sensitivity and linear dynamic range: Detectors (both universal, and selective
ones) vary in their sensitivity to analytes. Sensitivity refers to the lowest
concentration of a particular analyte that can be measured with a specified
signal-to-noise ratio. The more sensitive the detector, the lower this
concentration. The range over which the detector’s signal response is linearly
proportional to the analyte’s concentration is called the linear dynamic range.
(b) Types of detectors
1- The Thermal Conductivity Detector (TCD) (91)
TCD Characteristics
Universal (except for H2 and He); non-destructive; concentration detector; no
auxiliary (aux.) gas; works better with a parallel column; insensitive; limited
dynamic range.
The TCD was the first widely commercially available GC detector. It
measured differences in the thermal conductivity and/or specific heat of
highly thermally conductive (either H2 or He) carrier gas when diluted by
small concentrations of much less conductive analyte vapor (anything else). A
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
33
schematic of a typical TCD is shown in Figure 1, A current through a thin
resistive wire heated the wire in the detector flow cell. The thermal
conductivity of the flowing carrier gas cooled the wire. When analytes were
in the stream, their lower thermal conductivity produced less cooling, which
caused the wire’s temperature to rise and its resistance to increase. This wire
resistor was in a “Wheatstone bridge” circuit (an arrangement of four resistors
on the sides of a square). One of the other resistors was in a matching TCD
cell connected to a matching column and flow with no analyte passing
through. In isothermal GC, carrier flows and temperatures would remain
constant, but with temperature programming of the column the temperature
would increase and the flow would decrease, independently affecting the
conductivity.
Fig. 1. Thermal Conductivity Detector
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
34
2- Flame Ionization Detector (FID) (91)
FID Characteristics :
Nearly universal (all carbon compounds except CO, CO2, HCN, but not many
inorganic gases); destructive; mass flow detector; needs H2 and air or O2 aux.
gas; sensitive with wide dynamic range.
The FID is the most commonly employed detector, as it gives a response to
almost all organic compounds. A schematic of a typical FID is shown in
Figure 2, on a molecular basis the signal is roughly proportional to the
number of carbon atoms in the molecule. Hydrogen and oxygen (or air) must
be separately provided to fuel a flame in the detector cell. The H2 is
introduced and mixed with the carrier effluent from the GC column. Even if
H2 is used as capillary carrier gas, an additional separately controlled H2
supply is necessary to adjust the appropriate fuel supply for the flame. The
mixed gas enters the cell through a jet, where air or O2 flows past to serve as
the flame oxidizer supply. An electrical glow plug in the cell (not illustrated)
can be pulsed to ignite the flame. The fuel and oxidizer flows are adjusted
with needle valves to achieve a stable flame with optimal FID response, often
by bleeding a volatile unretained hydrocarbon into the carrier stream to
provide a reference signal. The jet tip is charged by several hundred volts
positive relative to several “collector electrodes” or a “collector ring”
surrounding the flame. In the absence of eluting organic analytes, no current
flows in the jet collector circuit. When a carbon-containing analyte elutes into
the flame, the molecule breaks up into smaller fragments during the cascade
of oxidation reactions. Some of these are positively charged ions, and they
can carry current across the flame in the circuit. Although the ionization
efficiency of the FID is low, its base
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
35
current is also very low. Hence its signal-to-noise ratio is very high. Against
such a low background, even very small ionization currents can be accurately
measured using modern electronics which draw very low currents (high input
impedance, voltage measurement circuits). Hence the good sensitivity and
extraordinary dynamic range of this detector, often exceeding six orders of
magnitude.
FID exhibits high efficiency of separation of petroleum crude oil especially
paraffinic hydrocarbons and aromatic hydrocarbons in addition to the polar
compounds such as alcohols, ketons, esters, etc. depending on the used
selective column.
Fig. 2. Flame Ionization Detector
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
36
3- The Electron Capture Detector (ECD) (91)
ECD Characteristics
Very selective (for organic compounds with halogen substituents like
pesticides, nitro and some other oxygen-containing functional groups); non-
destructive; concentration detector; needs to use N2 or argon/CH4 as carrier,
or as makeup gas if used with H2 or He capillary carrier flow; extreme but
highly variable sensitivity but with limited dynamic range.
The invention of the ECD is generally attributed to Lovelock, based on his
publication in 1961(92). It is a selective detector that provides very high
sensitivity for those compounds that “capture electrons.” It is an ionization-
type detector, but unlike most detectors of this class, samples are detected by
causing a decrease in the level of ionization. When no analytes are present,
the radioactive 63Ni emits beta particles as follow .
ß- 63Ni
These negatively charged particles collide with the nitrogen carrier gas and
produce more electrons as follow
ß- +N2—,2e+N2*
The electrons formed by this combined process result in a high standing
current (about 10-8 a) when collected by a positive electrode. When an
electronegative analyte is eluted from the column and enters the detector, it
captures some of the free electrons and the standing current is decreased
giving a negative peak: as follow .
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
37
A+e- —A* (7)
The negative ions formed have slower mobilities than the free electrons and
are not collected by the anode.
The mathematical relationship for this process is similar to Beers Law (used
to describe the absorption process for electromagnetic radiation).
Thus, the extent of the absorption or capture is proportional to the
concentration of the analyte.
The carrier gas used for the ECD can be very pure nitrogen (as indicated
in the mechanism presented) or a mixture of 5% methane in argon. When
used with a capillary column some make-up gas is usually needed, and it is
convenient to use inexpensive nitrogen as make-up and helium as the carrier
gas.
A schematic of a typical ECD is shown in Figure 3. 63Ni is shown as the beta
emitter although tritium has also been used; nickel is usually preferred
because it can be used at a higher temperature (up to 400°C) and it has a
lower activity (and is safer).
The ECD is one of the most easily contaminated detectors and is adversely
affected by oxygen and water. Ultrapure, dry gases, freedom from leaks, and
clean samples are necessary. Evidence of contamination is usually a noisy
baseline or peaks that have small negative dips before and after each peak.
Cleaning can sometimes be accomplished by operation with hydrogen carrier
gas at a high temperature to burn off impurities, but dismantling is often
required. ECD is a sensitive and selective detector for halogenated materials
but one which is easily contaminated and more prone to problems.
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
38
.Fig. 3. Electron Capture Detector
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
39
(2.3).Application of gas chromatography
Gas chromatography has been applied successfully to numerous compounds
in variety of fields . it is applied to both organic and inorganic compounds .
Even metals as volatile chelates has been successfully handled . The only
requirement is reasonable vapor pressure at the temperature of operation . we
have chosen only a few examples from the major fields
1- : Clinical
Gas chromatography now routinely handles many compounds of Clinical
interest. Methods; for amino acid (93,94); carbohydrate(95); CO2 and O2 in blood (96,97); fatty acid and derivatives (98) ; plasma triglycerides (99) steroids(100);
barbiturates(101) ; and vitamins (102) , have been described in recent reviews or
articles.
2- pharmaceuticals and Drugs :
Gas chromatography plays an essential role in the analysis of
pharmaceutical products and drugs . It is used in quality control, analysis of
new products and in monitoring metabolites in biological fluids.H. A. Lioyd
et al(103) first demonstrated the G.C. analysis of alkaloids. Gudzinowicsa and
S. J. Clark (104) provided dat for sensitivity of barbituratates by ECD and FID.
K,Langel .et al (105) described fully validated GC-MS method 50 different
drug compounds can be identified and quantitated simultaneously from a
single oral fluid sample.
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
40
.
3- Food The analytical tool gas chromatography was applied in food in order to
evaluate an preferred minor amount in the food to check for food quality. Gas
chromatography coupled with FID or MS or ECD or ASD is the more
efficient tool used food maintaining The high purity and high quality of food
.T. Nishimoto and M. Uyeta(106) described a GC., procedure for antioxidants
and preservatives. GC usually in combination with TLC and column
chromatography, is used to study adulteration, contamination and
decomposition of foods : olive oil (107) lard(108) dairy products (109) ; and
plasticizers in food(110). L. D. Metcalfe (111) describes a rabid esterification for
fatty acid and J. G. Niklly (112) uses glas beads as the solid support to separate
them. H.Kataoka .et al (113) determine of the contents of isophorone in food
samples which is cancer causing possibility using headspace solid-phase
microextraction (HS-SPME) coupled with gas chromatography- mass
spectrometry (GC- MS). T.Cajka .et al (114) analyze the pesticide in fruit-
based baby food which are registered for use in agriculture to meet food
supply demands using GC-MS. Others (115-118) Determine trace food-derived
hazardous compounds in with selective detectors.
4- Petroleum
It was in the petroleum industry that gas chromatography initially saw its
widest application. It has been successfully used to separate the entire rang of
products from light hydrocarbon gases up to waxes, asphalts and crude oil .
The composition and heat value of natural gas has been determined by GC(119-
120). E. M. Krupp, et al(121) investigate the determination of trimethylarsine in
natural gas and its partitioning into gas and condensate phases using
(cryotrapping)/gas chromatography coupled to inductively coupled plasma
mass spectrometry and liquid/solid sorption techniques. R.Gras,et al(122)
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
41
investegate Practical method for the measurement of Alkyl mercaptans in
natural gas by multi-dimensional gas chromatography, capillary flow
technology, and flame ionization detection. Nagy. E Moustafa And Jan.
Andersson(123)analyze the polycyclic aromatic sulfur heterocycles in Egyptian
petroleum condensate and volatile oils by gas chromatography with atomic
emission detection. C. T. Tran et al(124) Using the comprehensive two-
dimensional gas chromatography/time-of-flight mass spectrometry for the
characterization of biodegradation and unresolved complex mixtures in
petroleum .T. Dutries , et al (125) Improved hydrocarbons analysis of heavy
petroleum fractions by high temperature comprehensive two-dimensional gas
chromatography. E. Saari. et al(126)evaluate the occurrence of a matrix effect
in the gas chromatographic determination of petroleum hydrocarbons in soil.
The results indicate that solid phase extraction does not appear to be effective
enough in removing interfering matrix components from the extract. As a
result the effect of the matrix-induced chromatographic response prevents the
accurate determination of total petroleum hydrocarbon concentrations in soil
by GC- FID. D. Mao, et al(127) assessment of aqueous solubility (leaching
potential) of soil contaminations with petroleum hydrocarbons (PH) is
important in the context of the evaluation of (migration) risks and
soil/groundwater remediation. Field measurements using monitoring wells
often overestimate real PH concentrations in case of presence of pure oil in
the screened interval of the well.
5- Environment
The environmental pollution has been, and will continue to be of great public
concern. Environmental pollution is factor contributing to many chronic
diseases – lung cancer, emphysema, bronchitis and asthma. Many analytical
methods are used to measure environmental pollution, but gas
chromatography is of ever increasing importance. The separating or
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
42
concentrating power of column, combined with the sensitivity and selectivity
of detector, make GC an ideal technique to determine many compounds of
interest in pollution work.
G.C. has been used to concentrate trace impurities (128) ; to detect and size
organic aerosols (129,130) ; and to separate and determine polycyclic
hydrocarbons (131) . A sulfur specific microcoulometric detector capable of
seeing 5 ppm of SO2 and 1 ppm of H2S is describe (132) . Total organics in air
can be determined by a portable gas chromatograph (133) .
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
43
CHAPTER 3. The common used Materials and Methods
3.1. Chemicals and symbols
All chemicals were of reagent grade and were used without further
purification.
• Analytical grade paraffinic hydrocarbons pentane, n-hexane,
and n-heptane were purchased from BDH (99%), n-octane and
n-nonane from BDH (98%) and n-decane from Aldresh
Company (98%).
• Analytical grade aromatic hydrocarbons, benzene, toluene,
ethyl benzene, propyl benzene and butyl benzene were
purchased from BDH company of purities 95%, 98.5%, 98%,
98%, 98.5% respectively.
• The series of alcohols, methanol, ethanol, propanol , butanol ,
pentanol and hexanol. from BDH grade of purities ranged from
98% to 99.5%.
• nitromethane was purchased from Aldresh grade of purity
98.5%.
• pyridine and methyl ethyle ketone were purchased from BDH
grade of purities 97% and 98% respectively.
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
44
Symbols of the studied stationary phases
1. Dimethyl poly siloxane OV-101
2. Polyethylene glycol 20,000 Carbowax-20M
3. Caynopropyl poly siloxane OV-275
3.2.Optimum flow rate method
The optimum flow rat depends on the column length, column diameter and
column packing. The studied packing columns have the same length, same
diameter, same solid support and thicknessfilm of stationary phase. So,
they would have the same optimum flow rat, two organic solvents n-
hexane and benzene were injected in the split splitles injector under oven
temperature 120 oC at five different flow rates, 15, 20, 25, 30 and 40 ml
min-1. The height equivalent to theoretical plates (H) was determined for
each injected sample from the results obtained from the chromatogram
(retention time tr and peak width (W). From the correlation between the
height equivalent to theoretical plates and carrier gas flow rats we can
obtain the required optimum flow rate at the minimum H value.
3.3. Condition of gas chromatograph
The gas chromatograph used is shimmadzu 2014 equipped with flame
ionization detector (FID) and split/splitless injector. Helium was used as the
mobile phase at optimum flow rate 30 ml min-1, the carriers hydrogen and air
were used for FID at flow rate 45 and 450 ml min-1 respectively. The
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
45
chromatographic data were recorded using HP Chemstation software, which
was controlled by Microsoft Windows NT.
The gas chromatographic analysis was achieved on three studied
polymers of different polarities as stationary phases inside stainless steel tubes
of 4m in length and 1/8inch internal diameter. The thickness film of the
studied polymer as stationary phases is 3% on chromosorb P.A.W 60-80
mish.
The solutes used for chromatographic characterization were selected to cover
the wide range of polarity such as n-parrafins, aromatics, and alcohols. The
polarity indices were assessed with respect to the reference non-polar column
squalane (20% squalane on chromosorb WAW, 60-80 mesh).
3.4. Stationary phase characterization
Polarity index (Rohrschnieder)
The Roschnieders components, nitro methane, benzene, pyridine,
methyle ethyle ketone and ethanol in addition to the paraffinic series from
pentane to decane were injected at oven temperature 120oC in order to
obtain the retention index of each Roschnieders component. The helium
flow rat was maintained at 30 ml min-1.
Selectivity and applications
The paraffinic hydrocarbons from pentane to decane, the alcohols from
methanol to hexanol and the aromatics from benzene to butyl benzene,
were separated at different oven temperature programming starting from
50oC to 150 oC at a rate of 10 oC min-1. The helium flow rat was maintained
at 30 ml min-1.
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
46
3.5. Conditions of thermodynamic parameters
The determination of thermodynamic parameters (ΔH, ΔG and ΔS)
were achieved by injecting normal hexane, benzene and ethanol at different
oven temperatures: 70oC, 80oC, 90oC, 100oC and 110oC. these solutes were
selected as representative compounds of their families, paraffines, aromatics
and alcohols respectively. The helium flow rat was maintained at 30 ml min-
1.
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
47
CHAPTER 4. Characterizations of Stationary Phases
4.1. Determination of optimum flow rate
The present study should be occurred at the optimum flow rat of helium
carrier gas in order obtain high efficiency of gas chromatographic separation.
The studied three columns have the same column dimensions (column length
and column diameter), same support (Type & diameter) and the same
thickness film, so the experiment of optimum flow rat can be done on one
column representing the other columns. The probes n-hexane and benzene
were selected for this test at five different flow rates, 15, 20, 25, 30 and 40 ml
min-1. From the results of the gas chromatographic analysis given in Tables
(4&5), the height equivalent to theoretical plates (H) can be calculated froh
the last equation (11) , (12) .
Table 1. Retention data for determining the optimum flow rate using n-hexane
as a prob.
Flow
rates
tr W N H
15 7.35 1.35 474.272 0.843
20 5.65 1 510.760 0.783
25 4.75 0.8 564.063 0.709
30 3.85 0.6 658.778 0.607
40 3 0.5 576.000 0.694
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
48
Table 2. Retention data for determining the optimum flow rate using benzene
as a prob.
Flow
rates
tr W N H
15 9.88 1.80 482.046 0.830
20 8.18 1.35 587.372 0.681
25 6.87 1.05 684.932 0.584
30 5.54 0.8 767.754 0.521
40 4.19 0.65 665.557 0.601
15 20 25 30 35 40
0.60
0.65
0.70
0.75
0.80
0.85
H (
mm
)
Flow Rate (ml min-1)
n-Hexane
Fig. 4 The optimum flow rate using of the studied column dimensions using
n-hexane as a prob.
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
49
15 20 25 30 35 40
0.50
0.55
0.60
0.65
0.70
0.75
0.80
0.85
H (
mm
)
Flow Rate (ml min-1)
Benzene
Fig. 5 The optimum flow rate using of the studied column dimensions using
benzene as a prob.
From the results given in Tables 1and 2 and shown the Figures 4 and 5, it has
been found that the optimum flow rate at the minimum height equivalent to
theoretical plates was 30 ml min-1 for the probes n-hexane and benzene at all
the studied chromatographic columns.
S/R), but in ordinary gas chromatography the first term is negligible
and can be ignored (138).
The free energies of adsorption were calculated by the formula (139)
:
ΔG = -2.3 RT ln Vg (19)
where, R is the gas constant, T is the absolute temperature (K) and Vg is the
specific retention volume.
Figures (9-11) show the relationships between lnVg and 1000/T for the
studied polymers as stationary phases in gas chro
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
50
4.2.Rohrschneiders scheme
Characterization and evaluation the behavior of the studied polymer
stationary phases according to Rohrshneiders method is based on the
determination of retention indices for five solutes namely benzene, methyl
ethylketone, nitromethane and pyridine (X, Y, Z, U and S respectively)
belonging to different structural classes on the studied phases as well as on
squalane, chosen as the reference standard phase for this calculation.
The five Rohrschneider constants for the studied stationary phases are
obtained by calculating the differences observed for each of the substances
tested between their retention index on sqalane (Isqalane) and that corresponding
to the stationary phase being studied (Iphase)
Rohrschneider constants (∆I) = Iphase - Isqalane
The polarity of the studied chromatographic columns depends according to
Rohrschneider conceptions, not only on the stationary phase but also on the
solute to be analyzed. Thus each solute from the five selected solutes for this
characterization scheme may refer to a certain type of interaction between the
polymer stationary phase and solutes.
Generally, The Rohrschneider constants which are related to molecular
structures, allow an appreciation of the interactive forces between stationary
phase and solute as a function of compound class.
The orderly classification of polymer stationary phases with numerical data is
very valuable for comparison and selection of columns for particular type of
analysis. The sum of the five calculated values has been used to define the
overall polarity of the phase under test study.
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
51
In the present work, the polymers studied were characterized using
Rohrschneiders method for more investigation concerning the correlation
between the molecular structure and the polarities of these polymers towards
each type of solutes. Rohrschneider constants of the studied three polymers
are given in Table 3. .
The data obtained were determined at the same conditions (oven temperature
and mobile phase flow rate) on the studied polymers as stationary phases.
From the obtained data the following observations can be outlined:
The OV-101 has the lowest values of the Rohrschneider constants among the
studied polymers. These low values suggest that the dimethyl polysiloxane
stationary phase poorly retains the compounds that contain the corresponding
organic functions. This leads to low selectivity for this type of compound.
The overall polarity of the OV-101 stationary phase has the lowest value
reflecting low interactive forces between it and Rohrschneider solutes as a
function of compound class.
In case of OV-275 all its Rohrschneider constants have highest values than
those of the rest studied polymers. Among the Rohrschneider standards,
nitromethane and Pyridine exhibit the highest values compared with the other
five standards, followed by methyl ethyl ketone and benzene. Reflecting the
good ability of the OV-275 stationary phase to separate nitro compounds, the
aromatics including nitrogen compounds, ketones and aromatic hydrocarbons.
The high value of the overall polarity of OV-275 stationary phase indicating
the high interactive forces between it and Rohrschneider solutes as a function
of their families. Also, suggesting that the caynoprpyle stationary phase
strongly retains the compounds that contain the corresponding organic
functions. This leads to an improve selectivity for this type of compound.
The Rohrschneider constants of the poly ethylene glycol 20,000 (Carbowax-
20M) exhibit moderate values below OV-275 and above OV-101 indicating
their moderate polarity. These values decrease in the following order:
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
52
Pyridine, nitromethane, benzene, methyl ethyl ketone and Ethanol reflecting
the order of its efficiency of gas chromatographic separation toward the
corresponding families of theses constants.
Table 3 The difference in retention index of Rohrschneider constants of the
studied polymers
OV-275 Carbowax-
20M
OV-101 Squalane Symbole Compounds
819 722 16 0 X Benzene
1806 917 55 0 Y nitromethane
1709 1019 52 0 Z Pyridine
1011 624 39 0 U MEK
807 623 54 0 S Ethanol
6152 3905 216 0 Sum ∆I Summation
Table 4 The retention index for the five reference compounds given above (X,
Y, Z, U and S) on squalane are:-
I X Y Z U S
squalane 648 550 869 550 483
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
53
4.3.Selectivity index
Selectivity can be expressed in terms of uniformity criterion Δ (78) it is used
for estimating the degree of separation of multi component mixture; the
uniformity criterion was given according to the following equation:
t
τKnΔ
effk= (10)
where,k
n is the number of peaks on the chromatogram,τ is the base width of
the narrowest peak,eff
K is the separation factor for the worst separated pair of
components, and t is the duration of analysis.
Table 5 shows the calculated criterion for model systems comprising saturated
hydrocarbons, aromatic hydrocarbons and normal alcohols. It has been found
generally that all studied polymer stationary phases differ in their efficiencies
of separation according to the separated solutes. Saturated hydrocarbons can
be separated by the three studied polymers but in different degree of
efficiency, the dimethyl poly siloxane is the most efficient stationary phase as
illustrated from its criterion may be due to their low polarity which is
compatible with the polarity of saturated hydrocarbons.
For separation of aromatic hydrocarbons, the most polar caynopropyl poly
siloxane has the highest criterion number(Δ = 802 ). So it is the most
efficient stationary phase for these purpose followed by the polyethylene
glycol 20,000 (Δ = 702 ).
For separation of normal alcohols, only caynopropyl poly siloxane stationary
phase is success in separation of them giving peak sharp and prevent tailing
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
54
and broadening may be due to its high polarity is compatible with that of
normal alcohols.
This result agrees well and confirms the previous data obtained on the polarity
measurements of the studied polymers.
Therefore, the higher the polarity of the polymer, the higher the value of its
selectivity index. This means that the most polar stationary phase exhibits
high uniformity criterion number reflecting its high efficiency of separation of
polar compounds.
Table 5 Uniformity criterion Δ for different mixtures depending upon the
studied polymers as stationary phases
Families OV-101 Carbowax-20M OV-275
Paraffines 0.959 0.693 0.574
Aromatics 0.475 0.702 0.802
Alcohols 0.409 0.518 0.890
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
55
CHAPTER 5. Thermodynamic parameters
Gas chromatography affords a general technique for studying the
interactions of gases with solids or liquids and determination of the
thermodynamic parameters of the interaction: enthalpy (∆H), free energy
(∆G) and entropy (∆S).
Thermodynamic parameters of normal hexane, benzene and ethanol as
representative paraffin, aromatic hydrocarbons and alcohols respectively,
were calculated and listed in Table (6). The retention, selectivity and
interaction of the studied stationary phases are discussed in view of these
various thermodynamic parameters.
The thermodynamic functions (∆H and ∆S) can be estimated by
measuring the specific retention volume, Vg of probes at different
temperatures using the following equation (134-136):
lnVg = ln(R trs) + ∆S/R – ∆H/R . 1/T (21)
where, R is the gas constant (8.314 Jmol-1K-1), ns (mole) is the total amount of
solute in the adsorbed state and T is the absolute temperature (K). Vg can be
calculated from the following relation (137):
Vg= (tr-tm) FaT/Ta j (22)
where, tr is the probe retention time, tm dead time (min) is the retention time
of unretained solute (methane), Fa is the flow-rate (ml/min) measured at
column outlet and at ambient temperature Ta, and j is the compressibility
correction factor.
By determining Vg at different temperatures, one can calculate ΔH from
the slope and ΔS value from the intercept of the plot of lnVg versus 1/T. The
intercept is equal to ln(Rtrs) + (Δ matography using n-hexane, benzene and
ethanol. These plots give linear relationship. ∆H & ∆S could be derived from
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
56
the slope and intercept of the straight line, The thermodynamic data of various
stationary phases are summarized in Table (6). ∆G was calculated at different
temperature (from 343K to 393K) and given in Table 7.
The advantages of the gas chromatographic technique compared to classical
static equilibrium methods for obtaining thermodynamic data for mixing
polymers and solutes lies mainly in the speed of obtaining data. The GC
separation of the studied probes could be illustrated by considering the
thermodynamic parameters of these solutes.
It is evident from the data that negative ∆H and ∆S values for all
studied polymers using normal hexane increase in the sequence
Dimethylpolysiloxane > polyethylene glycol > cyanopropylpolysiloxane. The
more negative the ∆H and ∆S values the greater the interaction between the
probe and stationary phase. A similar study was also done by Inel et al(140),
they evaluated the thermodynamic parameters (∆H, ∆S and ∆G) of some
probes, each representing a class of organic compounds (n-hexane,
cyclohexane and benzene) on 4A and 13X Zeolites, it was found that
thermodynamic parameter increase in the sequence cyclo-hexane < n-hexane
< Benzene. Also, Bilgic and Askin(141), obtained the same result for activated
alumina stationary phase. Although, n-heptane and methylcyclohexane
interact non-specifically, but n-heptane interacts more intensively than
methylcyclohexane, this result indicates a better contact of an open chain
structure molecules with the surface of stationary phases (142).
The entropy values of normal hexane at different positions using dimethyl
polysiloxane show relatively minute differences in negative values. This is
reflected on the GC separation. Since the polymethylsiloxane is considered to
be non-polar or of relatively negligible polarity, the separation appears to be
depending on probes’ boiling points rather than their polarity.
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
57
The negative ∆H and ∆S values for all studied stationary phases using
benzene as a probe increase in the sequence cyanopropylpolysiloxane >
polyethylene glycol > Dimethylpolysiloxane. This reflects the greater
interaction of aromatics with the cyano and hydroxyl groups of
cyanopropylpolysiloxane and polyethylene glycol respectively exhibiting
good separation. The stronger elution of benzene on cyanopropylpolysiloxane
than n-hexane most probably attributed to the contribution of the specific
interaction between the cyano group and the π-electrons of the benzene ring.
So, cyanopropyl polysiloxane can interact specifically with the three π-
electrons of benzene molecules. Therefore, polyethylene glycol interacts with
aromatics but with less degree compared with the high polar cyanopropyl
polysiloxane stationary phase giving good separation of aromatics but with
long duration time of analysis. This may be due to the nearly matched polarity
of the polymer and benzene molecule.
The negative ΔH and ∆S values for all studied polymers using ethanol as a
probe increase in the sequence cyanopropylpolysiloxane > polyethylene
glycol > Dimethylpolysiloxane. The high negative ΔH value of
cyanopropylpolysiloxane with ethanol reflecting the higher energies required
for the interaction between them, reflecting on the high efficiency of
separation of alcohols as shown in Figure 12. On the opposite,
dimethylpolysiloxane has bad separation of alcohols due to its low negative
∆H value and low polarity, this behavior agrees well with the gas
chromatographic separation.
The free energy ΔG of the substances is the most important thermodynamic
characteristic of the reaction. Table 7 shows the values of the partial molar
excess free energy ΔG of the selected solutes in the studied stationary phases
at different temperatures from 70oC to 120oC. Generally, the free energy
reveals that free energy differences exist in minute values, and the free energy
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
58
shows an increase as a function of temperature for all solutes on the studied
stationary phases.
Table 6. Thermodynamic parameters of the studied polymers as stationary
phases using n-hexane, benzene and ethanol as probs
Columns Solutes -ΔH KJmol-1
-ΔS Jmol-1
degree-1
OV-101
n-Hexane 319.498 822.735 Benzene 135.246 318.347 Ethanol 35.407 47.793
Carbowax-
20M
n-Hexane 260.190 669.198 Benzene 157.292 344.583 Ethanol 68.320 101.717
OV-275
n-Hexane 238.973 615.499 Benzene 167.689 393.522 Ethanol 78.685 122.348
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
59
Figure 6. Plot of Ln retention volume (Vg) of n-hexane versus reciprocal of temperature for dimethylpolysiloxane (OV-101) as stationary phase
2.80 2.85 2.90 2.95 3.00 3.05 3.10 3.15 3.20
4.25
4.30
4.35
4.40
4.45
4.50
4.55
4.60
1000/TK
n-Hexane on OV-101
Ln
vg
2.50 2.55 2.60 2.65 2.70 2.75 2.80 2.85 2.90 2.95
2.6
2.8
3.0
3.2
3.4
3.6
1000/TK
n-Hexane on Carbowax-20M
Ln
Vg
2.50 2.55 2.60 2.65 2.70 2.75 2.80 2.85 2.90 2.95 1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.0 3.2 3.4
1000/TK
n-Hexane on OV-275
Ln
Vg
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
60
Figure 7. Plots of lnVg versus 1000/T of n-hexane, benzene and ethanol
for polyehylene glycol (carbowax-20M) as stationary phase.
2.80 2.85 2.90 2.95 3.00 3.05 3.10 3.15 3.20
3.3
3.4
3.5
3.6
3.7
3.8
3.9
1000/TK
Benzene on OV-101
Ln
Vg
2.50 2.55 2.60 2.65 2.70 2.75 2.80 2.85 2.90 2.95
5.16
5.18
5.20
5.22
5.24
5.26
5.28
5.30
5.32
5.34
1000/TK
Benzene on Carbowax-20M
Ln
Vg
2.50 2.55 2.60 2.65 2.70 2.75 2.80 2.85 2.90 2.95 3.6
3.7
3.8
3.9
4.0
4.1
4.2
4.3
4.4
1000/TK
Benzene on OV-275
Ln
Vg
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
61
Figure 8. Plots of lnVg versus 1000/T of n-hexane, benzene and ethanol for
cyanopropyle polysiloxane (OV-275) as stationary phase.
2.80 2.85 2.90 2.95 3.00 3.05 3.10 3.15 3.20
2.5
2.6
2.7
2.8
2.9
3.0
3.1
3.2
3.3
3.4
1000/TK
Etnanol on OV-101
Ln
Vg
2.50 2.55 2.60 2.65 2.70 2.75 2.80 2.85 2.90 2.95
4.50
4.55
4.60
4.65
4.70
4.75
4.80
4.85
4.90
1000/TK
Ethanol on Carbowax-20M
Ln
Vg
2.50 2.55 2.60 2.65 2.70 2.75 2.80 2.85 2.90 2.95 5.12
5.14
5.16
5.18
5.20
5.22
5.24
5.26
5.28
5.30
5.32
1000/TK
Ethanol on OV-275
Ln
Vg
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
62
Table 7. ΔG (KJmol-1) for n-hexane, benzene and ethanol on the studied
polymer stationary phases
Columns
Solutes 70 oC 80 oC
90 oC
100 oC
110 oC 120
oC
OV-101
n-Hexane
28.790 29.071
29.490
29.641
29.687 29.780
Benzene
22.772 22.868
22.980
23.016
23.090 23.161
Ethanol
18.521 18.536
18.638
18.679
18.717 18.774
Carbowax-
20M
n-Hexane
23.580 23.964
24.048
24.293
24.326 24.397
Benzene
34.964 35.651
36.500
37.314
38.115 38.907
Ethanol
31.836 32.417
32.885
33.206
33.543 33.854
OV-275
n-Hexane
21.460 21.701
21.860
21.910
21.925 22.024
Benzene
28.692 28.776
28.961
29.013
29.163 29.284
Ethanol
34.768 35.508
36.325
37.079
37.850 38.640
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
63
CHAPTER 6. Separation mechanism
In order to elucidate the mechanism of gas chromatographic separation,
via either adsorption or partition mechanisms, ∆H of the studied polymer
stationary phases has been plotted versus ∆S (143) as shown in Fig. (9). This
plot was done using three solutes of different polarities, namely n-hexane;
benzene and ethanol. In all cases, the solutes exhibit the phenomenon of
enthalpy-entropy compensation and a linear dependence between ∆H and ∆S
is fulfilled. This indicates the predominance of one mechanism, named
partition mechanism of the selected solutes on the nonpolar, moderate polar
and high polar stationary phases.
In order to clarify the mechanism of gas chromatographic separation of non-
polar molecules on the studied stationary phases, continuing the ideas
presented by Gurevich and Roshchina (144), test molecules which have the
same carbon number n-hexanee and benzene were chosen. It should be
recalled here that the aromatic hydrocarbons adsorb more weakly on the
surface of dimethyl polysiloxane stationary phase than their linear analogues
because of the smaller number of units directly interacting with the surface of
polymer.
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
64
Figure (9): Plots of ∆S Vs ∆H for indicated solutes of different
polarity on the studied polymers as stationary phasesin gas
chromatography
60 80 100 120 140 160 180 200 220 240 260
100
200
300
400
500
600
700
-Δ H KJmol -1
-1
OV-275
---- Δ S Jmol S Jmol S Jmol S Jmol ----1 1 1 1 degredegredegredegre----1111
50 100 150 200 250 300 100
200
300
400
500
600
700
-Δ H KJmol-1
-1
Carbowax-20M
---- Δ S Jmol S Jmol S Jmol S Jmol ----1 1 1 1 degredegredegredegre----1111
0 50 100 150 200 250 300 350 0
100 200 300 400 500 600 700 800 900
-Δ H KJmol -
1
OV-101
---- Δ S Jmol S Jmol S Jmol S Jmol ----1 1 1 1 degredegredegredegre----1111
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
65
CHAPTER 7. Chromatographic Application
Study of the physico-chemical characteristics of the paraent and
modified stationary phases gives information about the behavior towards their
applications. The chromatographic evaluation of the phases was done mainly
through their employment in the analysis of selected mixtures of different
polarities. The substances used for this purpose are, n-Alkanets, Aromatic,
Alcohols, Ketones, Nitro compounds, Ethers, Esters and Poly aromatics.
Efficiency of polymers as stationary phases for gas
chromatographic separation
As can be seen from the literature survey, polymers are commonly used as
stationary phases for gas chromatographic analysis of hydrocarbons families
such as paraffinic hydrocarbons, aromatics hydrocarbons, alcohols and esters.
This is due to the differentiation degrees of their polarities and the high value
of their constants on Rohrschneider,s scale.
The efficiency of gas chromatographic separation for the studied polymer
samples can be evaluated in terms of resolution and separation factor.
Resolution could be considered as an important chromatographic parameter, it
is used as a quantitative measure of the degree of separation between two
adjacent peaks.
R = (2Δt / W1+W2) (15)
R = degree of separation between two adjacent peaks
Δt = the time interval between peaks
W1 = width of the first peak at the base
W2 = width of the second peak at the base
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
66
Separation factor is the ratio of the adjusted retention times of two
components; it measures only the separation of the centers of mass of the two
peaks:
α α α α = tr(2) / tr(1) (13)
Results of these two parameters for the studied polymer stationary phases
using the selective solutes are summarized in Tables (8-10) and portrayed in
Figures (10-12). It has been found that gas chromatographic separations using
polymers as stationary phases have different separation efficiencies.
7.1. Separation of normal paraffinic hydrocarbons
The separation of normal paraffinic mixture (C6- C10) is capable of testing
specific interaction between solute and studied stationary phases. The
resolution and separation factor of paraffinic mixture were given in Table (8),
and shown in Fig. (10). By definition the separation factor is greater than
unity because species 1 elutes faster than species 2. it has been found that αααα
values of paraffines on all studied polymers can separate paraffines but with
different degrees of separation efficiencies.
They depict the separation of n-Alkanes mixture on the selected
columns. It has been found that; dimethyl polysiloxane stationary phase gives
suitable surface for eluting paraffines as the previous works (145, 146). The
resolution values of normal paraffines on OV 101 is the highest followed by
carbowax-20M and OV-275 stationary phases as given in Table (8). So, the
mixture of n-alkanes can be separated with high efficiency by the order: OV-
101, carbowax-20M and OV-275 columns which compatible with their
unformaty criterion values (Δ = 0.959, 0.693 and 0.574) respectively. The
dimethyl polysiloxane (OV-101) exhibits the most efficient separation of n-
alkanes with sharp and symmetrical peaks. Unfortunately, OV-275 column
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
67
has bad separation of paraffin, this may be related to its high polarity and its
low uniformity criterion value toward paraffines. With respect to polyethylene
glycol (carbowax-20M), paraffins can separated, but with slightly lower
efficiency of separation (Δ = 0.693) as compared with the OV-101, this is due
to its moderate polarity and selectivity values toward paraffinic hydrocarbons.
It was postulated that, the elution was directly correlated not only to
stationary phases polarity but also to polarity of solutes. Accordingly, the
polarity and selectivity are the main factors affecting the separation.
Table 8. Resolution and Separation factor of paraffins on the studied polymers
as stationary phases
No. Carbon Number
OV-101 Carbowax-20M OV-275
α R α R α R 1 C6-C7 1.3086 1.450 1.1204 1.05 1.549 0.284 2 C7-C8 1.3091 1.521 1.1951 1.12 1.593 0.329 3 C8-C9 1.2567 1.568 1.2689 1.06 1.452 0.293 4 C9-C10 1.2112 1.506 1.3101 1.16 1.342 0.267
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
68
OV-275
Carbwax-20M
OV-101
Fig.10. Gas chromatographic Separation of normal paraffins mixture from
n-pentane to n- hexane on the studied polymer stationary phases at 50OC
to 150OC at rate 10
oC min
-1anf flow rate 30 ml min
1.
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
69
7.2.Separation of aromatic hydrocarbons
The aromatic hydrocarbons (Benzene, toluene, ethylbenzene,
propylbenzene and butylbenzene) are important industrial chemicals. They
generally coexist in the catalytic reforming process in aromatic production.
Furthermore they were used as probes to investigateπ-complex formations.
The resolution and separation factor of aromatic hydrocarbons were given in
Table (9), and the gas chromatographic separation was shown in Fig. (11).
They depict the performance of the studied polymers on the separation of
aromatic mixture. Cyanoprpyl polysiloxane is the best studied stationary
phase for separating aromatic hydrocarbons obtaining good separation with
high resolution value. This could be directly linked with its polarity and
selectivity which elute specifically the molecules containing π-electrons.
Figure (11) illustrates the chromatographic separation of aromatic
hydrocarbons mixture on OV-101, Carbowax-20M and OV-275. The low
polarity and low selectivity of the polymer OV-101 as liquid stationary phase
make its surface unsuitable for separating aromatics, this bad separation was
evidenced with its lower uniformity criterion value (Δ = 0.475).
Consequently, the separation of aromatic hydrocarbons were improved on the
polymer OV-275 which exhibit high resolution values along each two
consecutive aromatic components in the mixture. This result was evidenced
with its high selectivity value towards aromatics (Δ = 0.890). Moreover, the
separation of aromatics on the carbowax-20M was achieved giving good
separation but accompanied by long duration time of analysis. this was
evidenced with the moderate uniformity criterion values (Δ= 0.702).
Also, good separation of aromatics was achieved on cyanoprpyle
polysiloxane having high separation factor, high resolution and high
uniformity criterion toward aromatics. this suggested that the polarity,
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
70
selectivity, resolution and unformaty criterion are the main factors affecting
the efficiency of separation.
On the other hand, reasonable separation of aromatic hydrocarbons was
achieved using polyethyleneglycol- 20M which has moderate polarity,
moderate resolution values and moderate uniformity criterion value. This was
reflected on the chromatogram shown in fig 11. having long duration time of
analysis.
Table 9 Resolution and Separation factor of aromatic hydrocarbons (benzene,
toluene, ethylbenzene, propylbenzene and butylbenzene) on the
studied polymers as stationary phases.
No. Carbon
Number OV-101 Carbowax-20M OV-275
α R α R α R 1 C6-C7 1.642 0.440 1.240 0.950 1.266 1.221 2 C7-C8 1.516 0.415 1.178 0.733 1.204 1.236 3 C8-C9 1.333 0.339 1.154 0.623 1.157 1.352 4 C9-C10 1.308 0.374 1.229 0.804 1.186 1.245
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
71
0V-275
Carbowax-20M
0V-101
Fig.11 Gas chromatographic Separation of aromatic hydrocarbons (1-
benzene, 2-toluene, 3-ethylbenzene, 4-propylbenzene and 5-
butylbenzene) on the studied polymer stationary phases at optimum
conditions (50oC; 10
oC/min to 150
oC)
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
72
Separation of alcohols .3.7
Alcohols (methanol, ethanol, propanol, butanol, pentanol and hexanol)
are the most polar organic compounds which interact with electron-donor
groups of the surface with a hydrogen bond evolving, reflecting on their bad
resolution and their highly tailing peaks.
The resolution and separation factor of alcohols on the studied polymer
stationary phases were given in Table (10). The gas chromatographic
separation of alcohols was shown in Fig. (12). The resolution of dimethyl
polysiloxane exhibits the lowest value compared with other studied polymers.
So, OV-101 can elute alcohols with unsymmetrical peaks and much skewed
peaks with a long tail (Fig. 12 ), due to its low polarity as discussed above.
This data was matched with its low criterion value and its low negativity of
enthalpy toward alcohols.
The resolution of OV-275 using alcohols as props exhibits the highest
value than the other polymers. Also, it has the highest selectivity and highest
negativity of enthalpy than the other studied polymer stationary phases.
These data reflecting on their separation of alcohols as shown in Fig 12.
Generally, the high polar cyanopropyl polysiloxane stationary phase
successes for eluting alcohols because of the similarity of their high polarities
coming from their hydroxyl groups. Thus, the better separation of light
alcohols till hexaneol was achieved on cyanopropyl polysiloxane with sharp,
symmetrical peaks and through an acceptable duration time of analysis.
The Resolution of alcohols on carbowax-20M column has values
lower than OV-275 and higher than OV-101 (Table 10 ). This reflected on the
separation of alcohols giving peaks having some unsymmetrical and some
tailing. This result was matched with polarity, selectivity and negativity of
enthalpy of carbowax-20M stationary phase using alcohols as probes.
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
73
From Fig. (11), the OV-275 was preferred than Carbowax-20M for
separating alcohols may be due to the high polarity of the polymer OV-275
resulting from cyanopropyle groups. This was evidenced by their differences
in uniformity criterion values.
Also alcohols are severely tailed on dimethylpolysiloxane, this
potentially due to the interaction of all oxygen groups on the surface of
stationary phase with the hydroxyl groups of alcohols producing hydrogen
bonds which is the reason of tailing peaks and bad separation. This result was
evidenced by its lower polarity value and lower negativity of enthalpy.
Table 10: Resolution and Separation factor of alcohols (1-methanol, 2-
ethanol, 3-propanol, 4-butanol, 5-pentanol) on the studied polymers
as stationary phases.
No. Carbon Number
OV-101 Carbowax-20M OV-275
α R α R α R 1 C1-C2 1.108 0.019 0.000 0.000 1.084 1.320 2 C2-C3 1.348 0.056 1.195 0.511 1.209 1.520 3 C3-C4 1.562 0.101 1.208 0.622 1.190 1.621 4 C4-C5 1.599 0.124 1.167 0.649 1.192 1.652 5 C5-C6 1.464 0.125 1.131 0.566 1.216 1.596
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
74
OV-275
Carbowax-20M
OV-101
Fig.12 Gas chromatographic Separation of alcohols (1-methanol, 2-
ethanol, 3-propanol, 4-butanol, 5-pentanol) on the studied polymer
stationary phases at optimum conditions (100oC; 5
oC/min to 220
oC)
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
75
Conclusion
In the present study, polymers gained a great attention since such
material can be easily prepared and their different polarities offer the
additional advantages of improving separation and reducing peak asymmetry.
Gas chromatographic characterization of the studied polymer was
followed by polarity assessment, selectivity and determination of the
thermodynamic parameters of the interaction of compounds as representatives
of their corresponding families with the investigated stationary phases.
Rohrschneider scheme of polarity helps to identified and arranged
polymer stationary phases according to their polarities. Paraffinic
hydrocarbons show higher selectivity values on the dimethyl polysiloxane
stationary phase than the cyanopropyl polysiloxane and polyethyleneglycol.
The opposite occurs using alcohols and aromatic hydrocarbons, OV 275
followed by carbowax- 20M exhibit higher selectivity than OV-101, these
data compatible with their polarities and reflecting on their gas
chromatographic separations.
Thermodynamic parameters assisted to study interaction between the
stationary phases and solutes of different polarities and elucidate the
separation mechanism. The high negative enthalpy and entropy values of
normal hexane in case of dimethyl polysiloxane reflect the high interaction
between normal paraffines and the nonpolar polysiloxane stationary phase.
The negative values of enthalpy and entropy on dimethyl polysiloxane
increase in the order normal paraffinic hydrocarbons > aromatic hydrocarbons
> alcohols. This indicates that the interactions on the non polar stationary
phase depends firstly on its polarity rather than on polarity of solutes. Also,
the GC separation of the studied probes on the dimethyl polysiloxane depends
on their boiling point rather than their polarity.
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
76
There is a slight difference in the free energy of the mentioned probes on
the studied stationary phases and ΔG is a function of temperature.
Using the three studied polymer stationary phases, the relation between
ΔH and ΔS of all studied probes is a linear relationship. This indicates the
predominance of the partition mechanism.
Chromatographic evaluation of the studied polymers as stationary phases
in gas chromatography was done mainly through their use in the analysis of
selected mixtures of different polarities including n-alkanes, aromatics and
alcohols. The efficiency of separation for the studied samples was evaluated
in terms of resolution, separation factor and uniformity criterion.
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
77
Acknowledgment
I would like to express deepest sincere gratitude to my Egyptian Petroleum
Research Institute (EPRI) the largest scientific organization and thanks to the
director of EPRI: Prof. Dr. Ahmed El-Sabagh and the vice director Prof.
Dr. Yasser Mostafa for their addition administratively and scientifically in
our institute.
Many thanks to my Prof. Dr. Abd-Elaziz Mostafa El-Fadly the more
human has a good impact in my life and for his continuous guidance, advice
and support, My prayers to God that keeps him and give him long old with
good health.
The author expresses his sincere thanks to Prof. Dr. Matar Mesehal Al-
Esaimi prof. of Physical Chemistry, Chemistry Department, Faculty of
Science, Umm AL-Qura University, KSA for his guidance advice and
invaluable help to Provide a gas chromatographic device throughout the work.
Also, It is a pleasant duty of the author to express his gratitude to all of his
colleagues at the chemical division in Umm Elkora University
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
78
References
(1) Stock R. and Rice C. B., "Chromatographic Methods", John Wiley&
Sons Inc., New York, Chapter 1 (1979).
(2) Faramawy S., El-Fadly A. M., El-Naggar A. Y. and Youssef A. M., Surf.
Coat. Technol., 90, 53 (1997).
(3) El-Fadly A. M., Faramawy S., El-Naggar A. Y. and Youssef A. M., Sep.
Sci. Technol., 32 (12), 2993 (1997).
(4) Scott R.P.W, Introduction to Analytical Gas Chromatography. (Marcel
Dekker, Inc). (1998).
(5) Kováts E.sz., Helvetica Chimica Acta; 41, 1915,.(1958).
(6) Horvàth C., in Ettre L. S., and Zlatkis A., "The Practice of Gas
Chromatography", Interscience, New York, 129 (1967).
(7) Martin R. L., Anal. Chem., 33, 347 (1961). (8) Pecsok R. L., De Yllana A., and Abdul-Karim A., Anal. Chem., 36, 452
(1964). (9) Martire D. E., Pecsok R. L., and Purnell J. H., Nature, 203,1279 (1964);
Trans. Faraday Soc., 61, 2496 (1965).
(10) Pecsok R. L., and Gump B. H., J. Phys. Chem., 71, 2202 (1967). (11) Eiceman G., Hill H. H., and Gardea-Torresday J., AnaL. Chem., 72,
137R.(2000).
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
79
(12) Abraham H. H., Poole C. F. and Poole S. K., J. Chromatogr. A, 842,
79,(1999).
(13) Grob K. and Grob G., J. Chromatogr., 125, 471, (1976).
(14) Welsch T., J. High Resolut. Chrornatogr. Chrornatogr. Commun., 11,
471, (1988).
(15) Voelkel A., Crit. Rev. AnaL Chern., 22, 411, (1991).
(16) Al-Saigh Z. Y., Tre. Folyrn. Sd., 5, 97, (1997).
(17) A. Y. El-naggar; G. Turky, Journal of Applied Polymer Science, Vol. 82,
1709–1717 (2001).
(18) A. Y. El-naggar; G. Turky, International Journal of Polymeric Materials,
Volume 50, Issue 2 August 2001 , pages 129 – 140.
(19) Jung Kee Suh,’ In Whan Kim, Seung Hyun Chang, Byoung Eog Kim,
Jae Wook Ryu, and Jung Hag Park . Korean Chem. Soc. 22, 4
409,(2001).
(20) PerrierCornet R., Heroguez V., Thienpont’A., Babot 0. and Toupance T.
Journal of Chromatography A, 28,1179,(2008).
(21) Xiaojie Sun, Yuling Zhu, Peng Wang,Jing Li, Caiying Wu.aAAAAAA
and Jun Xing . Journal of Chromatography A, 1218, 833, (2011).
(22) Chuen-Ying Liu*, Sheng-Hsiung Yang, Mao-Hsing Chau and
ChiaChann Shiue . Journal of Chromatography A, 933, 117, (2001).
(23) Conder J. R. and Young C. L., "Physicochemical Measurement by
Gas Chromatography", John Wiley & Sons Inc., Chapter 1 (1979).
(24) Conder J. R. and Young C. L., "Physicochemical Measurement by Gas
Chromatography", John Wiley & Sons Inc., Chapter 11 (1979).
(25) Kiselev A.V., Kovaleva N.V. and Nikitin Yu.S.: J. Chromatogr. A 58,
19, (1971).
(26) Corcia A.Di., Liberti A. and Samperi R.: Anal. Chem. 45, 1228, (1973).
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
80
(27) Naito K., Sagara T. and Takei S.: J. Chromatogr. A 503, 25, (1990).
(28) Melendez S. and Parker W.C.: J. High Resolut. Chromatogr. 2, 580,
(1979).
(29) Hayrapetyan S.S., Neue U.D. and Khachatryan H.G.: J. Sep. Sci. 29, 810,
(2006).
(30) Kopecni M.M., Milonjic S.K. and laub R.J.: Anal. Chem. 52, 1032,
(1980).
(31) Moriguchi S., Naito K., and Take1 S.: J. Chromatogr. A 131, 19, (1977).
(32) Naito K., Lmoriguchi S. and Take1 S.: J. Chromatogr. A 246, 199,
(1982).
(33) Fukazawa A., Moriguchi S., Naito K. and Takel S.: J. Chromatogr. A
295, 63, (1984).
(34) Braithwaite A. and Coper M.: Chromatographia 42, 77, (1996).
(35) Strzelczyk F., Leterq D., Wilhelm A.M. and Steinbrunn A.: J.
Chromatogr. A 822, 326, (1998).
(36) Lee M., Paek S., Ahn D.H., Kim K.R., Yim S.P. and Chung H.: J. Alloys
and Compd. 441, 368, (2007).
(37)-Khan A., Rinz M., Butt S.B. and Zaidi J.H.: J. Sep. Purif. Techol. 55,
396, (2007).
(38) Stock R. and Rice C. B., "Chromatographic Methods", John Wiley& Sons Inc., New York, Chapter 4 (1979).
(39) Olivero J. and Kannan K.: J. Chromatogr. A 849, 621, (1999).
(40) Katritzky A.R. and Chen K.: Anal. Chem. 72, 101, (2000).
(41) Kortvelyesi T., Gorgenyi M. and Heberger K.: Anal. Chim. Acta 428, 73,
(2001).
(42) Fatemi M.H.: J. Chromatogr. A 955, 273, (2002).
(43) Junkes B.S., Amboni R.D.M.C., Yunes R.A. and Heinzen V.E.F.: Anal.
Chim. Acta 477, 29, (2003).
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
81
(44) Nejad Z.G., Karlovits M., Demuthb W., Stimpfl T., Vycudilik W.,
Heravi M.J. and Varmuza K.: J. Chromatogr. A 1028, 287, (2004).
(45) Luan F., Xue C.X., Zhang R.S., Zhao C.Y., Liu M.C., Hu Z.D. and Fan
B.T.: Anal. Chim. Acta 537, 101, (2005).
(46) Song Y., Zhou J., Zi S., Xie J. and Y. Ye: Bioorg. Med. Chem. 13, 3169,
(2005).
(47) Stock R. and Rice C. B., "Chromatographic Methods", John Wiley&
Sons Inc., New York, Chapter 1 (1979).
(48) Rohrschneider L., J. Chromatogr., 17, 1 (1965). (49) Rohrschneider L., J. Chromatogr. Sci., 11, 160 (1973). (50) Rohrschneider L., J. Chromatogr., 22, 6 (1966). (51) Rohrschneider L., J. Chromatogr., 4, 333 (1967). (52) Rohrschneider L., J. Chromatogr., 39, 383 (1969). (53) McReynolds W. O., J.Chromatogr. Sci., 8, 685 (1970). (54) Synder L. R., J. Chromatogr. Sci., 16, 223 (1978). (55) Kersten B. R. and Poole C. F., J. Chromatogr., 452, 191 (1988). (56) Evans M. B. and Haken J. K., J. Chromatogr., 406, 105 (1987). (57) Chong E., de Bricero B., G. Miller, and S. J. Hawkes, Chromatographia,
20, 293 (1985). (58) Keller R. A., J. Chromatogr. Sci., 11, 49 (1973). (59) Voelkel A. and Janas J., J. Chromatogr., 645, 141 (1993). (60) Voelkel A. and Borowiak-Restena A., J. Chromatogr. A, 740, 253
(1996). (61) Poole S. K. and Poole C. F., J. Chromatogr., 500, 329 (1990).
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
82
(62) Kersten B. R., Poole S. K., and Poole C. F., J. Chromatogr., 468, 235 (1989).
(63) Figgins C. E., Risby T. H., and Jurs P. C., J. Chromatogr. Sci., 14, 453
(1976). (64) Golovnya R. V. and Misharina T. M., J. High Resolut. Chromatogr.,
Chromatogr. Commun., 3, 15 (1980).
(65) Poole C. F., Pomaville R. M., and Dean T. A., Anal. Chem. Acta, 225,
193 (1989).
(66) Kollie T. O. and Poole C. F., J. Chromatogr., 550, 213 (1991). (67) Kollie T. O. and Poole C. F., J. Chromatogr., 556, 457 (1991). (68) Abraham M. H., Poole C. F., and Poole S. K., J. Chromatogr. A, 842, 79
(1999). (69) Andronikashvili T. G., Tsitsishviti G. V., Kvernadze T. K., and
Okudzhava N. G., J. Chromatogr., 406, 145 (1987).
(70) Stock R. and Rice C. B., "Chromatographic Methods", John Wiley&
Sons Inc., New York, Chapter 4 (1979).
(71) Rouessac F. and Rouessac A. "Chemical Analysis Modern Instrumental
Methods and Techniques", English Edition, John Wiley & Sons, LTD,
Chapter 1 (2000).
(72) Jennings W., Mittlefehldt E., and Stremple P., "Analytical Gas
Chromatography", 2nd Edition, Academic Press, California, Chapter 1
(1997).
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
83
(73) Ambrose D., James A. T., Keulemans A. I. M., and Kovàts sz. E., Gas
Chromatography (1960), R. P. W. Scott ed., Butterworth, London, 184
(1960).
(74) Jones W. L. and Kieselbach R., Anal. Chem., 30, 1590 (1958). (75) Littelwood A. B., "Gas Chromatography Principles, Techniques and
Applications", Academic Press, New York and London, 138 (1962).
(76) Jennings W., Mittlefehldt E., and Stremple P., "Analytical Gas
Chromatography", 2nd Edition, Academic Press, California, Chapter 4
(1997).
(77) Burns W. and Hawakes J., J.Chromatogr. Sci., 15, 185 (1977). (78) Stark T. J., Larson P. A., and Dandeneau R. D., Proc.5th Int. Symp.
Capillary Chromatogr., 65 (1983).
(79) Dominguez, J. A. G., Lopez, J. E. Q., and Aguilar, R. L
J. Chromatogr. A803(1+2):197, (1998). (80) Aguilar, R. L., Dominguez, J. A. G., and Lopez, J. E. Q.
J. Chromatogr. A805(1+2):161. (1998). (81) A. Y. El-naggar; G. Turky, Journal of Applied Polymer Science, Vol.
82, 1709–1717 (2001)
(82) A. Y. El-naggar; G. Turky, International Journal of Polymeric Materials,
Volume 50, Issue 2 August 2001 , pages 129 - 140
(83) Zielinski W. L., Freeman D. H., Martire D. E., and Chow L. C.
Anal. Chem. 42, 176 (1970). (84) Chow L. C., and Martire D. E. (1971). J. Phys. Chem. 75:2005. (85) Jeknavorian A. A., Barrett P., Watterson A. C., and Barry E. F.
J. Chromatogr. 107, 317 (1975).
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
84
(86) Seifert K., and Kraus G. J. Chromatogr. 192, 97 (1980). (87) Littlewood A. B., Phillips C. G. and Price D. T. J. Chem. Soc.1480.
(1955).
(88) James A. T. and Martin A. J. P. Biochem. J. 50. 679 (1952). (89) Green S. A., and Pust H. J. Phys. Chem. 62, 55 (1958). (90) Kessaissa Z., Papirer E. and Donnet J. B. J. Chromatogr. 196, 481
(1980).
(91) Raymond P. W. Scott, , chromatographic Detectors, Design,Function,
and Operation, 73, 83, (1996).
(92) Lovelock J. E. and Lipsky S. R., J. Amer. Chem. Soc. 82, 431 (1960).
(93) Gebrke C.W., Stalling D.L.. Sep. Sci . 2 (1) 101 (1976) . (94) Varian Aerograph, Research Notes, SUMMER(1976). (95) Wells W. W., Sweelley C. C., Bentely R.. Biomedical application of
gas chromatography, 5th lecture Ann. gas chromatography Institute,
Buffalo, Herman A, Szymanski, ed. , p. 169, Plenum Press, New
York, (1964).
(96) Yee H. Y. Anal. chem. 37, 924 (1963) . (97) Lankford P. G. Am J. med . Techenol,31, 225(1965). (98) Supina W. R.. Biomedical application of gas chromatography, 5th
lecture Ann. gas chromatography Institute, Buffalo, Herman A,
Szymanski, ed. , p. 271, Plenum Press, New York, (1964).
(99) Bowers C. Y., Hamilton J. G., Muldrey J. E., Mayamasu W. T.,
Raynolds G. A. , Schally A. V. , J. Am. Oil chemists Soc. 43, 2
(1966).
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
85
(100) Vanden Heuvel W. J. A., Horning E. C.. , 5th Buffalo, Biomedical
application of gas chromatography Institut, Herman A. Szymanski,
ed. , p. 89, Planum Press, New York, (1963).
(101 ) Andres M. W.. Anal. Chem.. 38, 1945 (1966) (102 ) Nair P. P., Turner D. A. , Am. Oil chemists Soc. , 40, 535 (1963). (103) Lioyd H. A., Fales H. M., Higbet P. F. , Vanden Heuvel W. J. A. and
Wildman W. C.. J. Am Chem. Soc. 82, 3791 (1960)
(104) Gudzinowicz B. J. and Clerk S. J. J.ofG.C. 3, 147 (1965). (105) Kaarina Langel*, Teernu Gunnar, Kari Arinierni, Outi Rajarnäki,
Pirjo Lilisunde . Journal ot Chromatography B, 879, 859 (2011)
(106) Nishimoto T. and Uyeta M.. J.FOOD. Hyg, soc. Japan 5,287(1964), Anal, Abst13,409 (1966). (107) Esner J. , Lverson J. L., Mozenco A .K. and Firestone D.. J,Assoc, Offic,Agr, Chimists. 48, 417 (1965). (108) E, Pascucci and F, Paolini. Rif, ital, Sotanze, Geasse.41, 315 (1964). (109) Cannon J.H.. J.Asosoc. Offic. Agr. Chemists. 47, 577 (1964). (110) Wandel M. and Tengler H.. Deut. Lebensm. Randscbau. 60, 355 (1964). (111) Metcalfe L. D.. Anal, chem.38, 514 (1966). (112) Niklly J. G.. Anal. chem. 36, 2244 (1964). (113) Hiroyuki Kataoka, Yohei Terada. Reiko Tnoue, Kurie Mitani. Journal of Chromatography A, 1155 (2007) 100 -104 (114)Tomas Cajka , Jana Hajslovaa, Ondrej Lacina’, Katerina
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
86
Mastovska’. Steven J.Lehotay. Journal of Chromatography A, 1186, 281 (2008). (115)Yonghong Zhu., Genrong Li., Yunpeng Duan., Shiqi Chen., Chun Zhang. and Yanfei Li. Food Chemistry 109, 899 (2008). 116- Afidah A. Rahirn , Bahruddin Saad , Hasnah Osman , NoorHasani
Hashirn , Soihan Yahy, Khairuddin Md Talib . Food Chemistry 126 (2011) 1412—1416.
117- Jérémy Rate!, Erwan Enge .Journal of Chromatography A,
1216(2009) 7889-789 (118) Feng Zhang, Xiaogang Chu., Li Sun., Yansheng Zhao., Yun Ling., Xiujuan Wang., Wei Yong., Mmli Yang. and Xiuqin Li. Journal of Chromatography A, 1209, 220 (2008) (119) Thempson B., and Cavenab R. C.. Proc. Natl. Anal.Inster. Symb. 10th San Francisco, 261-5(1964) (120) Blakemore G. and Hillman G. E.. Analyst 90, 703 (1965). (121) Krupp E.M. , Johnson C. , Rechsteiner C. , Moir M. , Leong D. , Feldmann j. Spectrochimica Acta Part B 62, 970 (2007). (122) Gras R., Luong J., Carter V., Sieben L. and Cortes H. Journal of Chromatography A, 1216, 2776 (2009). (123) Nagy E., Moustafa J., and Andersson T. Fuel Processing Technology 92, 547 (2011). (124) Tin C. Tran, Graham A. Logan, Emmanuelle Grosjean, Danielle Ryan , Philip J. Marriott. Geochimica et Cosmochimica Acta 74 6468 (2010). (125) Thomas Dutriez , Marion Courtiade, Didier Thiébaut, Hugues Dulo, Marie- Claire Hennion1. Fuel 89, 2338 (2010). (126) Eija Saari, Paavo Perämäki , Jorma . Jalonen Microchemical Journal 87, 113 (2007). (127) Debin Mao, Richard Lookman, Hendrik Van De Weghe, Guido
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
87
Vanermena, Nicole De Brucker’, Ludo Diels. Journal of Chromatography A, 1216, 2873 (2009). (128) Novak J., Vasak V.. Janak J.. Anal chem..37, 660 (1966). (129) Frostling H. and Lindgren P. H.. J. Gas chromatography. 4, 243 (1966). (130) Obline R. W.. Anal. Chem.. 37, 93 (1963). (131) Demaio L. and Corn M. . Anal. Chem.. 38, 131 (1966). (132) Adams D. F., Jensen G. A., Steadman J. P., Koppe R. K. and Roberstone T. J. Anal. Chem. 38, 1094 (1966). (133) Guerrant G. O. Anal. Chem. 37, 516 (1965).
(134) Katsanos N.A., Lycourghiotis A., and Tsiatsios A.: J. Chem. Soc.
(Faraday Trans.) 74 (1978) 575.
(135) Katsanos N.A, Thede R. and Kalantzopoulou F.R.: J. Chromatogr. A,
795 (1998) 133.
(136) Katsanos N.A., Rakintzis N., Kalantzopoulou F.R., Arvanitopoulou E.
and Kalantzopoulos A.: J. Chromatogr. A, 845 (1999) 103.
(137) Grob R.L. (Ed.), Modern Practice of Gas Chromatography, Chapter 9,
Wiley- Iterscience, New York, (1995).
(138) Gurkan R: Adsorption 11 (2005) 603.
(139) Golovnya R, Terenina M.B. and Samusenko A.L.: Russian Chemical
Bulletin, 48 (1999) 1472.
(140) Inel O., Topaloglu D., Askin A. and F. Tumsek: Chem. Eng. J., 88,
(2002) 255.
(141) Bilgic C. and Askin A.: J. Chromatogr. A, 1006 (2003) 281.
(142) Milonjic S.K., Ceranic T.S. and Petkovic M.D.: Chromatographia 27
(1989) 306.
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
88
(143) Denayer J.F., Baron G.V., Martens J.A. and Jacobs P.A.: J. Phys.
Chem. B 102 (1998) 3077.
(144) Gurevich K.B. and Roshchina T.M.: J. Chromatogr. A, 1008 (2003) 97.
(145) El-Naggar A. Y. "Studies on GC separation in relation to some
characteristics of solid support and stationary phase" Ph.D Egypt
Petroleum Research Institute, 1994.
(146) Ebrahim A. A. "Surface modified aluminosilicates: Structural
characteristics, acidity, and application in GC MSc. Egypt Petroleum
Research Institute, 2005.
Publication – E Internatioanl in.co.isca.www ,me.isca.www
International Science Congress Association
89
ABOUT AUTHORS
Prof. Dr. Ashraf Yehia El-naggar
Prof. of Analytical and Petroleum Chemistry. Dr. Ashraf Yehia El-naggar is erudite scholar and seasoned chemist scientist. He obtained his M. Sc degree and Ph. D. degree in analytical chemistry from Cairo university and Mansoura university, Egypt respectively. He obtained his Ass. Prof. and Prof. degrees in analytical and petroleum chemistry from Egyptian Petroleum Research Institute (EPRI). He was a head of chromatographic Department and vice director of Central Analytical Lab., EPRI. His area of research interests include petroleum chemistry, chromatographic science, biochemical degradation of wastes. fuels and recycle of wastes to useful compounds. Dr A. Y. El-naggar has several teaching and research experience. Now he is vice director of chemistry Department, Science Faculty, Taif University, KSA. He has published over ninety research articles and reviews in peer-reviewed journals.
Dr.Adil A. Gobouri Dr. Adil A. Gobouri has PhD in Organic Chemistry, The University of Manchester, United Kingdom. Assistant Professor of Organic Chemistry and Head of Chemistry Department, Science Faculty, Taif University, Kingdom of Saudi Arabia. His research interests in Physical Organic Chemistry, Dendrimers, Heterocyclic and Click Chemistry.
Eid Hamed Al–osaimi
Eid Hamed Al–osaimi has MSc in analytical chemistry from Chemistry Department Science Faculty, Taif /University, KSA. He is a teacher in Ministry of Education, Makkah, KSA. He worked now in Ph. D. in the same field.