Results of the first BBA tests at FERMI@Elettra, Trieste A. Latina for the CERN BBA Team Thanks to Simone Di Mitri, Eduardo Ferrari, Giulio Gaio (Elettra) CLIC Beam Physics Meeting – May 28, 2014

Upload: gibson

Post on 21-Jan-2016

27 views

Category:

Documents


0 download

DESCRIPTION

Results of the first BBA tests at FERMI@Elettra , Trieste. A. Latina for the CERN BBA Team Thanks to Simone Di Mitri , Eduardo Ferrari, Giulio Gaio ( Elettra ) CLIC Beam Physics Meeting – May 28, 2014. Plan of the Elettra Laboratory. FERMI electron beamlines. - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: Results of the first BBA tests at  FERMI@Elettra , Trieste

Results of the first BBA testsat FERMI@Elettra, Trieste

A. Latina for the CERN BBA Team

Thanks to Simone Di Mitri, Eduardo Ferrari, Giulio Gaio (Elettra)

CLIC Beam Physics Meeting – May 28, 2014

Page 2: Results of the first BBA tests at  FERMI@Elettra , Trieste

2

Plan of the Elettra Laboratory

Page 3: Results of the first BBA tests at  FERMI@Elettra , Trieste

3

FERMI electron beamlines

Linac energy: from 100 MeV to ~1.2 GeVBunch charge: 800 pCBunch length: from 1.5 mm to 60 micron 10 Hz bunch repetition rateEmittance at gun: ~1.2 mm mrad

Page 4: Results of the first BBA tests at  FERMI@Elettra , Trieste

4

FERMI linacs

Page 5: Results of the first BBA tests at  FERMI@Elettra , Trieste

5

The Fermi control room

Page 6: Results of the first BBA tests at  FERMI@Elettra , Trieste

6

BBA parameters• 27 X correctors• 27 Y correctors• 30 BPMs

• 1.5 betatron oscillations• Effective bpm resolution = 5 - 10 micron• 10 Hz repetition rate• SysID time required:

– About 3 minutes to measure 1 response matrix (can be reduced to 30 seconds, with deeper integration in their control system)

Measured horizontal orbit response matrix

Page 7: Results of the first BBA tests at  FERMI@Elettra , Trieste

7

Orbit correction – 1:1

• First test of orbit correction:– Excited an oscillation – BBA algorithm is able to cancel it

Used 18 singular values in each axis (out of 27)

Page 8: Results of the first BBA tests at  FERMI@Elettra , Trieste

8

Emittance before BBAat the end of L4

HV

Page 9: Results of the first BBA tests at  FERMI@Elettra , Trieste

9

1:1 + DFS

The dispersion was measured reducing the amplitude of klystron 2 to 98% of its nominal value

This operation occurs in 1 cycle @ 10 Hz

Weight for DFS = 7 (tuned for ~10 micron BPM resolution)

Page 10: Results of the first BBA tests at  FERMI@Elettra , Trieste

10

Page 11: Results of the first BBA tests at  FERMI@Elettra , Trieste

11Vertical dispersion reduced by a factor ~40 ! Horizontal dispersion by factor ˜4 (!)

Page 12: Results of the first BBA tests at  FERMI@Elettra , Trieste

12

Emittance measurement after DFS

Horizontal emittance from 4.21 mm mrad to 3.30 mm mrad. Vertical emittance wasn’t very good though : we had many klystron failures / the machine was drifting away

Page 13: Results of the first BBA tests at  FERMI@Elettra , Trieste

13

1:1 + WFS

The wakefields were measured by changing the bunch charge to 650 pC(tests with 900 pC were foreseen, but there was not time for performing them)

Changing the charge takes about 10 seconds (time for charge feedback to converge)

Weight for WFS = 7-20 (tuned for ~10 micron BPM resolution)

Page 14: Results of the first BBA tests at  FERMI@Elettra , Trieste

14

Horizontal and vertical beam sizes during WFS

WFS !

Page 15: Results of the first BBA tests at  FERMI@Elettra , Trieste

15

Horizontal and vertical beam sizes during WFS

WFS !

Page 16: Results of the first BBA tests at  FERMI@Elettra , Trieste

16

1:1 + DFS + WFS

Page 17: Results of the first BBA tests at  FERMI@Elettra , Trieste

17

Emittance measurement after1:1 + DFS + WFS

HV

Page 18: Results of the first BBA tests at  FERMI@Elettra , Trieste

18

Conclusions

• Looks like an ideal machine for tests– Very fast control system– Excellent diagnostics

• Debug – debug – debug. Only three hours of effective beam-time

• Everything worked really well. Emittance (H; V):– Before correction: 4.31 mm mrad; 3.21 mm mrad– After DFS: 3.30 mm mrad; -– After DFS+WFS: 2.75 mm mrad; 2.57 mm mrad

• Future– Fully explore the the space of parameters: e.g.

• 1 axis at time | both axes decoupled | both axes coupled• DFS weight scan• WFS weight scan• Dependence of the result on the number of singular values

– Tests of on-line parasitic DFS / WFS ?