q922+log+l07 v1

46
Well Log ging Course ( 2 nd Ed.)

Upload: hossein-alaminia

Post on 13-Jan-2015

124 views

Category:

Education


0 download

DESCRIPTION

 

TRANSCRIPT

Page 2: Q922+log+l07 v1

1. Early Electric Log Interpretation

2. Formation Factor

3. Water saturation

4. The Porosity Exponent, m

5. The Saturation Exponent, n

6. A Thought Experiment For A Logging Application

Page 3: Q922+log+l07 v1

1. Unfocused Devices:A. The Short Normal

B. Estimating the Borehole Size Effect

2. Focused Devices:A. Laterolog Principle

B. The Dual Laterolog

Page 4: Q922+log+l07 v1
Page 5: Q922+log+l07 v1

electrode devices

electrode devices, Is one type of electrical logging toolso named because

the measurement elements are simply metallic electrodes

These devices utilize low-frequency current sources, in most cases below 1,000 Hz.

The historical progression from the normal device to traditional focused dual laterologsThe traditional focused dual laterolog was the main device

used for electrode measurements of resistivity for many years, even though it had several known shortcomings.

• Many of these shortcomings were solved by the introduction of array devices, a development that was made possible partly by the availability of fast inversion software.

Spring14 H. AlamiNia Well Logging Course (2nd Ed.) 5

Page 6: Q922+log+l07 v1

Electrode Devices

Unfocused DevicesThe Short Normal

Focused DevicesLaterolog Principle

Spherical Focusing

The Dual Laterolog

Further DevelopmentsReference Electrodes

Thin Beds and Invasion

Array Tools

Spring14 H. AlamiNia Well Logging Course (2nd Ed.) 6

Page 7: Q922+log+l07 v1

A schematic representation of the short normalThe earliest

commercial device, the short normal, is illustrated in the figure.

A 16 in. [41cm] (thus the designation “short.”) spacing is indicated between current

electrode A and measure

electrode M (voltage electrode)

Spring14 H. AlamiNia Well Logging Course (2nd Ed.) 7

Page 8: Q922+log+l07 v1

the short normal problems:1: in a very conductive borehole mud

Two basic problems are associated with the short normal, bothrelated to the presence of the borehole,

which is normally filled with a conductive fluid.

1st: There is a sensitivity of the measurement to the mud resistivity and hole size In a borehole filled with very conductive

mud, the current tends to flow in the mud rather than the formation. In this case,

the apparent resistivity as deduced from the injected current and resultant voltage will not reflect the formation resistivity very accurately.

Idealized current pathsSpring14 H. AlamiNia Well Logging Course (2nd Ed.) 8

Page 9: Q922+log+l07 v1

the short normal problems:2: in front of a thin resistive bed

The second difficulty with this measuring technique:Once again,

the conductive borehole fluid provides an easy current path for the measure current into adjacent shoulder beds of much lower resistivity (Rs ) than the formation (Rt ) directly opposite the current electrode.

In this case, the apparent resistivity (from the measurement of

the voltage of electrode M and the current I, in combination with the tool constant)

will again be representative not of the resistive bed, but, more likely, of the less resistive shoulder bed.

Idealized current pathsSpring14 H. AlamiNia Well Logging Course (2nd Ed.) 9

Page 10: Q922+log+l07 v1
Page 11: Q922+log+l07 v1

Normal Devices estimation

The value obtained from the normal device has to be corrected for mud resistivity and borehole diameter, before true formation resistivity (RT) can be obtained.

Common Normal Devices are called 16” Normal and

64” Normal logs depending on the electrode spacing.

Normal log only gives reliable estimates of RT for homogeneous beds

that are much thicker than the electrode spacing.

Spring14 H. AlamiNia Well Logging Course (2nd Ed.) 11

Page 12: Q922+log+l07 v1

Borehole correction chart for the 16 in. short normal.for

an 8 in. hole size, the short normal does a fairly good job of measuring the correct formation resistivity, except

for very large mud/formation resistivity contrasts.

However, for the 16 in. borehole size, this is not the case.

Courtesy of Schlumberger.Spring14 H. AlamiNia Well Logging Course (2nd Ed.) 12

Page 13: Q922+log+l07 v1

A schematic of the short normal in common logging situationIn the figure some

idea of the actual tool implementation is given. The electrode B

is at the surface, whereas

the electrode N, to which the potential measurement is referenced, is actually located down-hole on the measurement sonde.

Adapted from Doll et alSpring14 H. AlamiNia Well Logging Course (2nd Ed.) 13

Page 14: Q922+log+l07 v1

A schematic of the short normal response in two common situations A continuing examination of the

shortcomings of the short normal in the Figure reveals the kind of response problems encountered for large contrasts between the shoulder beds and the bed of interest.

In this particular case, the resistivity contrast between beds is 14, and the borehole diameter [d] is half the spacing between current source and voltage electrode [AM]. (AM=2d)

Even for a bed 3 ft [36 in=91cm] thick, it is seen that the central value of resistivity does not attain the desired value. (Probably

d=6 in=15 cm, AM=12 in=30cm)

If the bed is only 6 in. thick, then the behavior becomes nonintuitive, with the apparent resistivity dipping below the value of the shoulder bed.

Spring14 H. AlamiNia Well Logging Course (2nd Ed.) 14

Page 15: Q922+log+l07 v1
Page 16: Q922+log+l07 v1

the lateral device

When attempts were made to improve this bed-boundary resolution, the normal device evolved to the lateral device, illustrated in the Fig.

The lateral sonde is much like the normal sonde except that there are two voltage

electrodes, and the potential difference

between them is used to indicate the resistivity of the formation layer between them.

This will be nearly the case for beds whose thickness exceeds the spacing between the electrodes marked A and N.

Spring14 H. AlamiNia Well Logging Course (2nd Ed.) 16

Page 17: Q922+log+l07 v1

the lateral device (Cont.)

The figure shows the response to two beds, whose thickness is given in terms of the electrode spacing. It is clear that there has

been some improvement for bed resolution,

but the response is still quite complicated because of current flow through the mud to zones other than the one directly in front of the measuring points.

Spring14 H. AlamiNia Well Logging Course (2nd Ed.) 17

Page 18: Q922+log+l07 v1

The correction curves for the 18’8” Lateral DeviceThe value obtained from

the Lateral Device has to be corrected for mud resistivity and borehole diameter, before true formation resistivity can be obtained. A common Lateral Device

is the 18’8” Lateral Log. This log only gives reliable

estimates of RT for homogeneous beds that are much thicker than the electrode spacing.

Spring14 H. AlamiNia Well Logging Course (2nd Ed.) 18

Page 19: Q922+log+l07 v1
Page 20: Q922+log+l07 v1
Page 21: Q922+log+l07 v1

The Laterolog electrode

The Laterolog electrode arrangement and typical responses are shown in the Figure.

Spring14 H. AlamiNia Well Logging Course (2nd Ed.) 21

Page 22: Q922+log+l07 v1

The Laterolog

The Laterolog tries to force the electrical current into the formation

and thus reduce the effect of the mud and invaded zone on the log response.

This log in conjunction with the Induction Log provides the best readings to attempt to estimate the true formation resistivity.

When the invaded zone diameter is less than 12” (30cm) (for 8” (20cm) diameter hole), the laterolog resistivity value

is within 10% of the true formation resistivity.

Spring14 H. AlamiNia Well Logging Course (2nd Ed.) 22

Page 23: Q922+log+l07 v1

current flow in the borehole and formation from a central electrode

The next step in the evolution of electrical tools was the implementation of current focusing.

Figure illustrates, on the left half of the diagram, the current paths for the normal device

in the case of a resistive central bed. The current tends to flow around it,

through the mud, into the less resistive shoulders.

The desired current path is shown on the right half of the figure, where the measure current is somehow

forced through the zone of interest.

Idealized patternsSpring14 H. AlamiNia Well Logging Course (2nd Ed.) 23

Page 24: Q922+log+l07 v1

The principle of focusing (Laterolog-3 (LL3))Idealized

current distribution from the Laterolog-3 device in a

homogeneous formation,

with current focused into the formation.

From SerraSpring14 H. AlamiNia Well Logging Course (2nd Ed.) 24

Page 25: Q922+log+l07 v1

Laterolog-3 (LL3)

in previous slide, there are now three current emitting electrodes, A0, A1, and A 1’.

This type of array is known as a guard focusing device and

is commonly referred to as a Laterolog-3 (LL3), device.

The potential of electrodes A1 and A 1’ is held constant and

at the same potential as the central electrode A0. Since current flows only if a potential difference exists, there

should, in principle, be no current flow in the vertical direction.

Spring14 H. AlamiNia Well Logging Course (2nd Ed.) 25

Page 26: Q922+log+l07 v1

Laterolog-3 (LL3) (Cont.)

The sheath of current therefore emanates horizontally

from the central measurement electrode.

The current emitted from the focusing, or “guard” electrodes is often referred to as the “bucking” current, as its function is to impede the measure current

from flowing in the borehole mud.

Spring14 H. AlamiNia Well Logging Course (2nd Ed.) 26

Page 27: Q922+log+l07 v1

LL3 difficulties

Despite these good intentions, the LL3 device still showed some difficulty with bed boundaries. The effects of shoulder

bed resistivity on the behavior of an LL3 device. The top sketch indicates

current passing through the mud into a highly conductive shoulder.

The bottom sketch indicates the effect of a thin conductive bed.

Spring14 H. AlamiNia Well Logging Course (2nd Ed.) 27

Page 28: Q922+log+l07 v1

The electrode configuration of the Laterolog-7

Another approach to focusing the measure current is the seven electrode device, or LL7. The electrode configuration of

one such device is sketched here.

Monitor electrodes drive the bucking current in the guard electrode to maintain a differential voltage of zero.

The array is symmetric with A0 in the center.

Adapted from SerraSpring14 H. AlamiNia Well Logging Course (2nd Ed.) 28

Page 29: Q922+log+l07 v1

The Laterolog-7

The guard electrodes A1 and A1’ are no longer elongated: instead, additional monitoring electrodes have been

introduced in order to impede the flow of current parallel to the sonde though the borehole mud.

This is achieved by varying the bucking current of the guard electrodes so that the potential drop between the pairs of monitor electrodes (i.e.,M1–M1’ and M2–M2’) is zero.

Since the potential drop is zero along this vertical direction, the current will be focused into the formation.

Spring14 H. AlamiNia Well Logging Course (2nd Ed.) 29

Page 30: Q922+log+l07 v1
Page 31: Q922+log+l07 v1

spherical focusing

Another approach to compensating for the effect of the borehole is the concept of spherical focusing. In this technique,

which has been adopted for medium and shallow resistivity measurements, bucking currents attempt to establish the spherical equipotential surfaces that would exist if no borehole were present.

Spring14 H. AlamiNia Well Logging Course (2nd Ed.) 31

Page 32: Q922+log+l07 v1

Approximate current lines and equipotential surfacesThe Figure is

a rough sketch of the equipotential surfaces

which surround the current electrode in a normal device, as a result of the presence of the conductive mud in the borehole.

Instead of spherical surfaces, they are of elongated shape.

the short normal in a boreholeSpring14 H. AlamiNia Well Logging Course (2nd Ed.) 32

Page 33: Q922+log+l07 v1

The electrode configuration of the spherically focused array.

The objective of the spherical focusing is to provide a bucking current to force

the equipotential lines to become spherical once again.

Then the potential difference at two points along the sonde will be determined by the resistivity of a slice of formation in a spherical shell with radii equal to the two spacings.

The depth of investigation can be controlled by the size of the shell.

Spring14 H. AlamiNia Well Logging Course (2nd Ed.) 33

Page 34: Q922+log+l07 v1
Page 35: Q922+log+l07 v1

dual laterologs

The most common traditional electrode devices use a dual focusing system. Those known as dual laterologs combine

the features of the LL3 and LL7 arrays, in an alternating sequence of measurements.

By rapidly changing the role of various electrodes, a simultaneous measurement of deep and shallow resistivity is achieved.

Spring14 H. AlamiNia Well Logging Course (2nd Ed.) 35

Page 36: Q922+log+l07 v1

the current paths computed for dual laterologsOn the left side, the

electrodes are in the deep configuration. The length of

the guard electrodes, which use parts of the sonde, is about 28 ft [8.5 m] to achieve deep penetration of a current beam of 2 ft [61cm] nominal thickness.

On the right side, they are in

the shallow (or medium) configuration.

Spring14 H. AlamiNia Well Logging Course (2nd Ed.) 36

Page 37: Q922+log+l07 v1

Dual laterologs

The current distributions computed for the dual laterolog in its two modes of operation.

The central electrode is the source of measure current for both shallow and deep modes.

In the deep mode, both the two long electrodes and

the smaller electrodes next to them are sources of bucking current.

In the shallow mode, the bucking current is sent from the small

to the long electrodes to provide a type of spherical focusing.

Spring14 H. AlamiNia Well Logging Course (2nd Ed.) 37

Page 38: Q922+log+l07 v1

apparent resistivity

For purposes of comparison of the different electrical measuring devices, it is convenient to think of the signal measured as being the result of the influence of three distinct regions of the measuring environment:

the borehole, the invaded zone, and the undisturbed formation.

Each of these zones is attributed its own characteristic resistivity: Rm, Rxo, and Rt . • Generally the mud resistivity Rm is much less than either Rxo or

Rt .

In this model, the response of an electrode device can be conveniently thought of as an approximately linear combination of the invaded zone and the true resistivity.

Spring14 H. AlamiNia Well Logging Course (2nd Ed.) 38

Page 39: Q922+log+l07 v1

apparent resistivity

This is expressed as:

Ra is the apparent resistivity.

The pseudogeometric factor J is a normalized weighting factor which gives the relative contributions of the invaded zone (of diameter, di ) and virgin zone, to the final answer. It is referred to as the pseudogeometric factor

(as opposed to a pure geometric factor, as will be seen later with the induction tool) since the weighting function will actually be influenced by the contrast between Rxo and Rt.

Spring14 H. AlamiNia Well Logging Course (2nd Ed.) 39

Page 40: Q922+log+l07 v1

the pseudogeometric factor for several devicesThe comparison of

calculated pseudogeometric factors for a number of common electrode devices.

LLd and LLs refer respectively to the deep and shallow arrays of a dual laterolog device.

Courtesy of SchlumbergerSpring14 H. AlamiNia Well Logging Course (2nd Ed.) 40

Page 41: Q922+log+l07 v1

the pseudogeometric factor

The pseudogeometric factors can be used to estimate the influence of the invaded zone on the measurement of resistivity when there is a contrast between Rt and Rxo. The shallow curve (marked LLs) rises steeply and

indicates that in the case of a more conductive invasion zone (Rxo = 0.1Rt ), half of the shallow response comes from the first 8 in. of invasion and 90% comes from within a diameter of about 80 in.

The deep measurement (marked LLd) shows less sensitivity to the invaded zone since only about 15% of its response comes from a diameter of 20 in. (or the first 6 in. of invasion in this calculation for an 8 in. borehole).

Spring14 H. AlamiNia Well Logging Course (2nd Ed.) 41

Page 42: Q922+log+l07 v1

laterolog’s sensitivity to the borehole

It is important to note the laterolog’s sensitivity to the borehole. Next slide shows the correction chart for the deep and

shallow measurement of a particular dual laterolog device, plotted in a manner similar to chart for the short normal.

This chart is for a centered tool. Other charts are available for an eccentered tool,

the eccentricity being characterized by the standoff between tool and borehole wall.

Spring14 H. AlamiNia Well Logging Course (2nd Ed.) 42

Page 43: Q922+log+l07 v1

A borehole correction chart for deep & shallow laterolog measurements.It is to be

compared to correction chart for the short normal, to appreciate the improved response due to focusing.

Adapted from SchlumbergerSpring14 H. AlamiNia Well Logging Course (2nd Ed.) 43

Page 44: Q922+log+l07 v1

1. Ellis, Darwin V., and Julian M. Singer, eds. Well logging for earth scientists. Springer, 2007. Chapter 5

2. “Formation Evaluation” Master of Petroleum Engineering. Curtin University of Technology,. Chapter 9

Page 45: Q922+log+l07 v1

1. Introduction

2. Microelectrode Devices

3. Uses For Rxo

4. Azimuthal Measurements

5. Resistivity Measurements While Drilling

Page 46: Q922+log+l07 v1