power analyzer ar5-l & ar5 series leak-meter software...

21
POWER ANALYZER AR5-L & AR5 series Leak-Meter Software (Code M80229) INSTRUCTION MANUAL (M98204701-03-06B) CIRCUTOR S.A.

Upload: vudiep

Post on 23-Apr-2018

220 views

Category:

Documents


0 download

TRANSCRIPT

POWER ANALYZER AR5-L & AR5 series

Leak-Meter Software

(Code M80229)

INSTRUCTION MANUAL

(M98204701-03-06B)

CIRCUTOR S.A.

2

Table of contents 1 BASIC INSTRUCTIONS. ...........................................................................................3 1.1 CHECKS ON RECEIPT.................................................................................3 1.2 SAFETY ADVICE. .........................................................................................3 1.3 CONNECTION INSTRUCTIONS...................................................................3

2 ANALYZER START-UP. ............................................................................................4 2.1 NEW PROGRAM RECORDING PROCEDURE. ...........................................4 2.2 WORKING PROGRAM SELECTION. ...........................................................5

3 PRESENTATION OF DATA ON THE DISPLAY. .......................................................5 3.1 DATA SCREEN.............................................................................................5 3.2 TRIGGER SCREEN. .....................................................................................5 3.3 SETUP VISUALIZATION...............................................................................6 3.3.1 Filter activation. .........................................................................................7

4 MEASURING MODULE CONNECTION. ...................................................................7 5 PROGRAMMING THE ANALYZER. ..........................................................................8 5.1 SETUP MENU...............................................................................................9 5.1.1 SETUP MENU. ..........................................................................................9 5.1.1.1 Analysis. .......................................................................................... 10 5.1.1.2 Sensor selection. ............................................................................. 10 5.1.1.3 RCD: Residual Current Device setting. ............................................ 10 5.1.1.4 RANGE: Measuring Range selection. .............................................. 11 5.1.1.5 MODE.............................................................................................. 12 5.1.1.6 TRIGGER. ....................................................................................... 13 5.1.1.7 File Menu......................................................................................... 14

5.2 COMM: COMMUNICATION PARAMETERS. .............................................. 14 5.3 CLOCK: INTERNAL CLOCK. ...................................................................... 15 5.4 PASSWORD: SAFETY SETTING. .............................................................. 15 5.4.1 Change of Setup access password:......................................................... 15

5.5 RECALL: READ CONFIGURATION............................................................ 16 5.6 DISPLAY MENU: SCREEN CONTRAST..................................................... 16 5.7 RUN: DATA RECORDING PROCESS STATUS ......................................... 17 5.8 FILES MENU............................................................................................... 17 5.8.1 DIR: Directory. ......................................................................................... 18 5.8.2 DELETE: Deleting a file. .......................................................................... 18 5.8.3 FORMAT: Formatting the analyzer internal memory. ............................... 18

5.9 LANGUAGE MENU. .................................................................................... 19 6 ANALYZER COMMUNICATIONS............................................................................ 19 7 SAFETY WARNINGS. ............................................................................................. 20 8 MAINTENANCE....................................................................................................... 20 9 MEASURING MODULE’S TECHNICAL FEATURES. .............................................. 20 10 TECHNICAL SERVICE. ......................................................................................... 21

3

1 BASIC INSTRUCTIONS. This manual assists in the installation and use of the "LEAK-METER" program, designed for the AR5 series analyzer. The operation of the networks analyze with program “LEAK-METER” allows the detection and analysis, of the leakage current in the installations.

1.1 CHECKS ON RECEIPT. On receipt of your order check the following:

1. The material received corresponds to the specifications in your order. 2. Check that the material received has not been damaged during delivery. 3. Check that the following accessories have been received:

• 1 Programming cartridges for the analyzer setting with the required program.

• 1 Programming cartridges for the analyzer with the coprocessor program.

• 1 Instruction manual.

• 1 measuring module.

• 1 connection wire to connect the measuring module to the analyzer.

1.2 SAFETY ADVICE. For the safe use of the analyzer it is fundamental that the people who install it or manipulate, follow the usual safety measures, as well as the different warnings indicated in manual saying.

1.3 CONNECTION INSTRUCTIONS. Before connecting the equipment, check the following:

1. Power supply voltage: from external supply.

• Mains input 100 V AC. - 240 V AC.

• Supplier output to AR5/AR5-L 12 V DC. 2. Mains frequency

• 45...65 Hz. 3. Maximum current in the voltage measurement circuit:

• 500 V AC. phase-neutral.

4

2 ANALYZER START-UP.

2.1 NEW PROGRAM RECORDING PROCEDURE. The analyzer has an internal memory to store different programs. Once these programs have been loaded, the equipment can be used as: Standard analyzer, display of the data stored, harmonic analyzer, disturbance logger ... Before starting this process, ensure that the equipment is supplied from a supplier. To load this program, the following procedure has to be followed:

1. Switch off the equipment 2. Connect the program cartridge to the cartridge port located on the

supplier. 3. Switch on the equipment.

4. With [5555] and [6666] keys select the loading program option (LOAD PROGRAM). Press [ENTER] or wait for a few moments to confirm that this is the required operation.

5. Select where the program is to be stored. An empty position is marked as EMPTY PROGRAM. If a position is selected where there is already a program such as Harmonics, the program will be replaced

6. The analyzer will test to check if the cartridge has been connected. 7. If it detects that a cartridge has been connected, it will proceed with

loading the program. The process ends when LOAD OK appears on the screen.

8. Once the program is loaded, reset the equipment clicking on OFF. 9. If no cartridge is detected or there has been any error with the program

loading process, reset the equipment and try again.

One single loader will be operational with the Analyzer once the program has been loaded for the first time. Input the equipment series number into the loader

5

2.2 WORKING PROGRAM SELECTION. The equipment can store different operating programs in its memory. The required program is chosen on start up. Switch on the equipment.

1. A list of available programs will appear on the screen.

2. Use the [5555] and [6666] keys to select the program the analyzer will use. 3. Press [ENTER] or wait for a few moments to confirm that this is the

required operation.

3 PRESENTATION OF DATA ON THE DISPLAY. The PROGRAM Leak-Meter has different screens to program the data to be displayed and subsequently shown on the screen.

3.1 DATA SCREEN The first screen allows to visualize the measured voltage, the earth leakage current and the maximum value of this that has been detected.

LEAK-METER I1.0

V1

IL

IMAX

Vp-n

mA

mA

230

150

330

RECORD08/02/06 12:15:00INST MAX

Using the [6] and [5] keys, the user can change between the instantaneous values or the maximums that the equipment has detected. The selected option is indicated in bold. If key ESC is pressed it will visualize the setup.

3.2 TRIGGER SCREEN. If you have selected a single record mode (see 5.1.1.5 MODE.), when the trigger conditions are complied, appears a window with the characteristics of the trigger.

6

Each parameter indicates the following: RCD Sensit: Programmed sensitivity Delay: Programmed retardation Leak Max: maximum current detected Int: leakage average current Time: Time between the trigger and the voltage falls down.

3.3 SETUP VISUALIZATION. This screen permits to check the most important setup parameters in the analyzer. The screen on the left is the one shown on the analyzer’s display. The screen on the right explains the meaning of each term.

LEAK-METER I1.0

1V/1V

CLAMP 1AFILTER ON

SENSOR

SENSOR

30mA 20msRCD

RECORDSTD-PRG.STDContinuos 60s

COMM9600 / NO / 8 / 1

08/02/06 12:15:00RECORD

LEAK-METER I1.0

Voltage ratio transformesSort of clamp or CT usedFilter activated or deactivated

SENSOR

SENSOR

Sensity and triping time of the residual current device

DDR

RECORDFile namerecording type

COMMCommunications setup

RECORDTime and date 12:15:00

7

3.3.1 Filter activation.

The module has built-in a lowpass filter, that allows to discard the leakage currents of high frequency, very frequent in the facilities. In order to activate or to deactivate the filter it must press key ENTER and will appear the indication FILTER in the inferior part of display.

4 MEASURING MODULE CONNECTION. The measuring module allows measuring the leakage current in the main. It has connectors to connect the CF’s clamps, and also allows to connect the protection transformer that already maybe exist in the installation, using a connection cable, which can be the same one that are used to measure the voltage. With the also two plugs are supplied, to connect the cable to the connection terminals of the transformer. The module also has a DB-9 connector to be connected to entrance AUX of AR5/AR-L and IL1 (black connector)

Once the module it’s connected and the analyzer recognizes it, it will appear an indication in the inferior part of the screen.

8

5 PROGRAMMING THE ANALYZER. The LEAK-METER program has a setup menu to configure the analyzer. To access into the setup, press SET key in the file selection screen.

Password:SET SET

SETUPDISPLAYRUNFILES

CLEAROFFLANGUAGE

The analyzer will then inquiry for a password that consists of a key sequence to be pressed (the user has 15 seconds to press this sequence). If the default password has not been modified, then the following key sequence must be followed:

PASSWORD

[3] [SET] [5] [SET]

Once this password is entered, the analyzer will permit the user to modify any Setup parameters. All programs have an independent setup, therefore, the setup must be always check to ensure a proper operation, since any modification will only affect the active operation program. If any parameter has been changed, when you exit from SETUP, it ask for save the new configuration or exit without save the changes.

Setup ChangedSure

< Yes > < > No

• <Yes>: Save the changes and exit.

• <No>: Exit from SETUP without save the changes.

9

There are the following menus in setup:

SETUPDISPLAYRUNFILES

CLEAROFFLANGUAGE

• Select one option with keys [6] & [5].

• To access any menu option use [8] or [ENTER].

• To close the menu press [3] or [ESC]. If this key is used when only the main menu is open, this is closed. If any modification over any setup parameter was done, before closing a confirmation of setup change is requested on display.

5.1 SETUP MENU. The analyzer can be user-configured to different performances involving its data analysis and recording modes, as it is followed shown:

5.1.1 SETUP MENU.

This option allows setting the conditions of analyzer, in particular the connection type and the transformer relation of voltage and currents, among others.

10

SETUPDISPLAYRUNFILES

CLEAROFFLANGUAGE

SETUP.

COMM.CLOCKPASSWORDRECALL

ANALYSIS

5.1.1.1 Analysis.

SETUPDISPLAYRUNFILES

CLEAROFFLANGUAGE

SETUP.

COMM.CLOCKPASSWORDRECALL

ANALYSISANALYSIS

RCDRANGE

MODETRUGGERNAME

SENSOR

5.1.1.2 Sensor selection.

The measuring module has different connector in order to connect the following sort of sensors:

1. WN: Transformer series WN (see transformer label). 2. WG: Transformer serie WG (see transformer label). 3. CLAMP: Leakage current clamp CF-10 or CF-5 4. OTHER: If another transformer type its used, you should select this

option, then the analyzer will enquire for the number of turns of the winding.

In this menu, you should select the same sensor you connect in the measuring module.

5.1.1.3 RCD: Residual Current Device setting.

In this menu, the installation residual current device specifications should be set in the analyzer.

11

SETUPDISPLAYRUNFILES

CLEAROFFLANGUAGE

SETUP.

COMM.CLOCKPASSWORDRECALL

ANALYSISANALYSISSENSOR

RANGE

MODETRUGGERNAME

RCD

SENSIT. (mA)

100mA

7894 6

1230

5

20msDELAY (ms)

ESC ENT

SET

50 HzFrequency

ESC ENT

SET

The parameters of sensitivity, delay and frequency, determine the behavior of a residual current device. The analyzer will measure the earth leakage current similarly would do a residual current device with the programmed characteristics. With the activation of the filter or no, an “immunized” residual current device can be simulated “or no.

5.1.1.4 RANGE: Measuring Range selection.

SETUPDISPLAYRUNFILES

CLEAROFFLANGUAGE

SETUP.

COMM.CLOCKPASSWORDRECALL

ANALYSISANALYSISSENSORRCD

MODETRIGGERNAME

RANGE

1ARANGE

ESC ENT

SET

12

This option allows setting the measuring range, less than 1 Amperes or up to 10 Amperes.

5.1.1.5 MODE.

To two ways of operation or taking of measures can be selected. 1. Continuous mode: the analyzer periodically do records with a period

defined in the menu Period. 2. Single: the device do a record in memory, only when there is a leakage

current higher than programmed one. With this way the following options are due to select.

SETUPDISPLAYRUNFILES

CLEAROFFLANGUAGE

SETUP.

COMM.CLOCKPASSWORDRECALL

ANALYSIS

ANALYSISSENSORRCDRANGE

TRIGGERNAME

MODE

SingleMODE

ESC ENT

SET

ContinuousMODE

ESC ENT

SET

5.1.1.5.1 Trigger

It allows selecting if the equipment does a record in memory according to the following conditions.

• V1 Trigger: The analyzer does a record when the voltage reaches an level lower than 50V. The record file store 4 seconds before the trigger and 1 second after it

• IL trigger: When there is a earth leakage current of a higher value than programmed one. The analyzer records a file, with values that include form 1 second before the trigger and 4 seconds after it.

V1p_nTRIGGER

ESC ENT

SET

IL1TRIGGER

ESC ENT

SET

13

5.1.1.6 TRIGGER.

You can program here certain conditions (Trigger) so that values are saved in memory only when these conditions are met.

SETUPDISPLAYRUNFILES

CLEAROFFLANGUAGE

SETUP.

COMM.CLOCKPASSWORDRECALL

ANALYSISANALYSISSENSORRCDRANGE

MODE

NAMETRIGGER

SETUPDISPLAYRUNFILES

CLEAROFFLANGUAGE

SETUP.

COMM.CLOCKPASSWORDRECALL

ANALYSISANALYSISSENSORRCDRANGE

MODE

NAMETRIGGER

TRIGGER

OFFTIME

5.1.1.6.1 Time.

Lets to set the time trigger, that is, to define the period to perform the data storage.

TIME ON

SET

ESC ENT

30/ 05 / 05 23:00:00

TIME OFF

ESC ENT

SET

31 / 05 / 05 23:00:00

When a time is set, will appear in the main screen the indication D/T?, meanwhile the conditions are not complied.

mA330

D/T ?

INST MAX

5.1.1.6.2 OFF.

It allows to deactivate any programming of trigger that has been defined previously.

OFF

ESC ENT

SET

Yes

Yes: deactivate the trigger

14

5.1.1.7 File Menu.

This section allows programming the name of the file with 8 characters and without extension

SETUPDISPLAYRUNFILES

CLEAROFFLANGUAGE

SETUP.

COMM.CLOCKPASSWORDRECALL

ANALYSISANALYSISSENSORRCDRANGE

MODETRIGGERNAME

SETUPDISPLAYRUNFILES

CLEAROFFLANGUAGE

SETUP.

COMM.CLOCKPASSWORDRECALL

ANALYSISANALYSISSENSORRCDRANGE

MODETRIGGERNAME

NAMESTD-PROG

QWERTYUIOPASDFGHJ LZXCVBNMLOCK SPAC

K7894561230

• Pressing [ENTER]: Text on display is directly validated.

• To modify: (alphanumeric input)..

• Select with keys [6], [5], [4] o [3]the character to modify and pres [SET] to validate each character.

• Select “�“on display + [SET] to delete a character.

• [ENTER] to validate the total text or [ESC] to exit with no modification.

If the typed name already exists in memory, when exiting setting actions, the display will show: “Overwrite file Sure? “ --> Do you really overwrite the file?

• If yes is answered, the previous file is deleted.

• If no is answered, setup menu is not exited, thus a new name can be typed for the file, or the setting actions can be cancelled.

5.2 COMM: COMMUNICATION PARAMETERS. Program here the parameters of the built-in RS-232 serial output. When selecting this option the present parameters are shown on display:

9600 NO 8 1COMM

ESC ENT

SET

• Pressing [ENTER]: values on display are directly validated.

• Select with keys [4] or [3] the position to modify.

• Though keys [6], [5] the value of the selected position is increased or decreased.

• [ENTER] to validate the total value or [ESC] to exit with no modification.

15

5.3 CLOCK: INTERNAL CLOCK. Use this section to set the analyzer internal clock in time: date / time and displaying format. When this option is selected, presently programmed values are shown by display:

DD / MM / AA HH:MM:SSCLOCK TYPE

ESC ENT

SET

After validating the date format, the present date and time will be displayed according to the selected arrangement:

CLOCK

ESC ENT

SET

31/ 05 / 05 13:21:14

Proceed similarly than for the former section.

5.4 PASSWORD: SAFETY SETTING. Set here the analyzer password configuration. This password will be required to access the Setup menu. Thus, the manipulation of the analyzer by not-authorized people can be avoided. The default password is:

PASSWORD

[3333] [SET] [5555] [SET]

5.4.1 Change of Setup access password:

Firstly enter the old password:

OLD PASSWORD * * * *

Then enter the new password:

NEW PASSWORD * * * *

Finally, confirm the new password:

CHECK PASSWORD * * * *

16

During this password modification process, the following error messages can be shown in screen during 5 s:

Incorrect Old password

Wrong password

Incorrect Check password

The confirmation of the new password has failed

5.5 RECALL: READ CONFIGURATION

You can here recall a standard configuration.

SETUPDISPLAYRUNFILES

CLEAROFFLANGUAGE

SETUP.ANALYSISCOMM.CLOCKPASSWORDRECALL

A confirmation is requested: “Recall Setup sure <yes> or <no>“. With keys [4] & [3] select yes or no, and then press [ENTER]. A “Standard” operation program for the analyzer is available to be user-recalled. Its features are: Sensity RCD = 100mA Delay RCD = 20ms Frequency = 50Hz Mode = continuo Períod = 60 seconds Filter = ACTIVATED Range = 1A Sensor = WG

5.6 DISPLAY MENU: SCREEN CONTRAST. The user can here vary the contrast of the analyzer’s display.

With the [4] you can intensify the display contrast and with the key [3] this can be lowered.

17

SETUP

RUNFILES

CLEAROFFLANGUAGE

DISPLAYSETUP

RUNFILES

CLEAROFFLANGUAGE

DISPLAY

DISPLAYCONTRAST

CONTRAST

DEC INC

5.7 RUN: DATA RECORDING PROCESS STATUS You can here enable or disable the data collection and logging action in the analyzer.

SETUPDISPLAY

FILES

CLEAROFFLANGUAGE

RUN

RUN

ESC ENT

SET

OFF

• With keys [6], [5] RUN or STOP is selected.

• [ENTER] to validate or [ESC] to exit with no modification.

5.8 FILES MENU. Non-volatile analyzer internal memory is storing data up to its maximum capacity. Once full, neither new data will not be saved in, nor stored data will be lost (provided no incorrect operation is done). When memory is full, the display will show: "MEMORY FULL".

FILES FILES MENU è DIR DELETE FORMAT

18

5.8.1 DIR: Directory.

This option shows on display a directory of all files saved in memory.

DIR .

STD-PROG. STD xxxxx dd / mm / aa hh: mm / ss PRUEBA1. STD dd / mm / aa hh: mm / ss Bytes free: xxxxxxxxxx

File name 1 / size Day/time file 1 File name 2 / size Day/time file 2

• Keys [6] or [5] allows reading more files in case that all files stored by the analyzer cannot be shown in only one screen.

• Keys [ENTER] or [ESC] to exit.

5.8.2 DELETE: Deleting a file.

You can here delete any file from the internal memory.

DELETE .

STD-PROG. STD xxxxx PRUEBA1. STD xxxxx CIRCUTOR. A5D xxxxx Bytes free: xxxxxxxxxx

File name 1 / size File name 2 / size File name 3 / size

• With keys [6] & [5] select the file to be deleted.

• [ENTER] to confirm selected file erasing. Once press, a confirmation is required.

• Press key [ESC] to exit with no modification.

5.8.3 FORMAT: Formatting the analyzer internal memory.

This option lets the user to format the internal memory.

Format Mem.Sure?

< Yes > < > No

Once the format action over the internal memory is confirmed, a confirmation is required1. Take into account that this action will mean all stored data to be deleted.

1 Do not turn the analyzer off during the memory format process, otherwise the display will show an error message and the process should be redone.

19

5.9 LANGUAGE MENU. It allows to select the language2 in which the menus will be presented.

LANGUAGE

ESC ENT

SET

English

• With keys [6] & [5] select the file to be deleted.

• [ENTER] to confirm selected file erasing. Once press, a confirmation is required.

• Press key [ESC] to exit with no modification.

6 ANALYZER COMMUNICATIONS. To connection to PC of the analyzer must be done through the power supplier set, which will be necessary connected to the main. Perform the connection using the two wires factory-delivered with the analyzer. One cable will link the analyzer to the power supplier set, and another cable is a standard RS232 connector When starting communication tasks take into account:

• Communication parameters of the analyzer and the PC must fully coincide.

• The power supplier set must be plugged into.

• The analyzer cannot be into the setup menu. CIRCUTOR has a software package that permits the user to retrieve stored data in the analyzer to PC for a further complete analysis.

2 This configuration is commune for all the programs

20

7 SAFETY WARNINGS.

The user should take into account all installation instructions referred in sections INSTALLATION AND START-UP, Connection instructions and TECHNICAL SPECIFICATIONS of the analyzer. Note that with the instrument powered on, the terminals could be

dangerous to touching, and cover opening or elements removal actions may allow accessing dangerous parts. The analyzer has been designed and tested to meet IEC 348 standard and is factory-shipped in proper conditions.

8 MAINTENANCE. The analyzer does not require any special maintenance. No adjustment, maintenance or repairing actions should be done over the instrument open and, should those are essential, high-qualified operators must perform them. Spares should be provided by manufacturer. The batteries are rechargeable and should be provided by manufacturer.

Before any adjustment, replacement, maintenance or repairing operation is carried out, the instrument must be disconnected from any power supply source.

When any protection failure is suspected to exist, the instrument must be immediately put out of service. The own instrument design permits a quick replacement in case of damage. Occasionally clean the equipment with water and detergent using a soft, damp cloth. Do not use abrasives or solvents. Completely dry the equipment before using it again.

9 MEASURING MODULE’S TECHNICAL FEATURES.

Burden Voltage 5Vdc Current 20mA

Clamps Measuring Range 10mA – 10A Models CF-5

CF-10 Other of 5000 winds

CT Models 500 – 2000 espiras Measuring Range 10mA -10A Isolation Magnetic isolation

Construct specifications Dimensions 90x90x20mm Weight 120g

21

10 TECHNICAL SERVICE. For any inquiry about the instrument operation mode or in case of malfunction, you can contact CIRCUTOR’s technical service. CIRCUTOR S.A. - Aftersales Service Vial Sant Jordi, s/n 08232 – Viladecavalls (SPAIN) Tel: (+34) 93 745 29 00 Fax: (+34) 93 745 29 14 E-mail: [email protected]