  • 1Instruction Manual

    INDICATOR StandardIndustrial ElectronicWeight IndicatorModel I5S

    Ohaus CorporationOhaus CorporationOhaus CorporationOhaus CorporationOhaus Corporation29 Hanover RoadFlorham Park NJ07932-0900

  • 2Unauthorized changes or modifications to this equipment are notUnauthorized changes or modifications to this equipment are notUnauthorized changes or modifications to this equipment are notUnauthorized changes or modifications to this equipment are notUnauthorized changes or modifications to this equipment are notpermitted. See manual for replacement parts.permitted. See manual for replacement parts.permitted. See manual for replacement parts.permitted. See manual for replacement parts.permitted. See manual for replacement parts.

    NOTENOTENOTENOTENOTE: This equipment has been tested and found to comply with the limits for a ClassA digital device, pursuant to part 15 of the FCC rules.

    These limits are designed to provide reasonable protection against harmful interfer-ence when the equipment is operated in a commercial environment. This equipmentgenerates, uses, and can radiate radio frequency energy and, if not installed andused in accordance with the instruction manual, may cause harmful interference toradio communications. Operation of this equipment in a residential area is likely tocause harmful interference in which case the user will be required to correct theinterference at his own expense.

    This Class A, digital apparatus meets all requirements of the Canadian Interference.Causing Equipment Regulations.

    Cet appareil numerique de la classe A respecte toutes les exigences du Reglementsur le material broilleur du Canada.

    This device corresponds to requirements stipulated in 90/384/EEC andfeatures radio interference suppression in compliance with valid ECRegulation 89/336/EEC. Note: The displayed value may be adverselyaffected under extreme electromagnetic influences, eg. when using a

    ra- dio unit in the immediate vicinity of the device. Once the interference hasbeen rectified, the product can once again be used for its intended purpose. Thedevice may have to be switched on again.

    Cet appareil correspond aux exigences selon la norme 90/384/CEE et est dparasitconformment la directive de la CE 89/336/CEE en vigueur. Remarque: Dans desconditions dinfluences lectromagntiques extrmes, par exemple en casdexploitation dun appareil radio proximit immdiate de Iappareil la valeurdaffichage risque dtre influence. Une fois que linfluence parasite est termine,le produit peut tre de nouveau utilis de manire conforme aux prescriptions; le caschant, il est ncessaire de le remettre en marche.

    Dieses Gert entspricht den Anforderungen nach 90/384/EWG und ist funkentstrtentsprechend der geltenden EG-Richtlinie 89/336/EWG. Hinweis: Unter extremenelektromagnetischen Einflssen z.B. bei Betreiben eines Funkgertes in unmittelbarerNhe des Gertes kann eine Beeinflussung des Anzeigewertes verursacht werden.Nach Ende des Streinflusses ist das Produkt wieder bestimmungsgemss benutzbar,ggfs. ist ein Wiedereinschalten erforderlich.

    CLASSIFICATIONSCLASSIFICATIONSCLASSIFICATIONSCLASSIFICATIONSCLASSIFICATIONS The I5S has been tested and found to comply with the followingstandards:

    National Institute of Standards and Technology Handbook 44 Class III and IIIL

  • 3TABLE OF CONTENTSTABLE OF CONTENTSTABLE OF CONTENTSTABLE OF CONTENTSTABLE OF CONTENTS

    Chapter 1 Introduction and Installation 5Introduction ........................................................................................ 6Description ......................................................................................... 6Unpacking .......................................................................................... 7Assembly ............................................................................................ 7Scale Base Connections .................................................................... 7Capacity Card .................................................................................... 8Closing and Setup Protection ............................................................ 8Power Options .................................................................................... 9AC Adapter ......................................................................................... 9Battery Installation ............................................................................. 9Turning the I5S ON ............................................................................ 10Display Indicator Functions ................................................................ 10To turn the I5S OFF ........................................................................... 10

    Chapter 2 Setup and Calibration....................................................... 11Setup Mode and the I5S Menu Structure ........................................... 12To Access the Main Menu and Change Settings ............................... 12To Exit a Submenu, Save Settings and Return to the Main Menu .... 12To Exit the Main Menu and Return to the Weighing Mode ................ 12I5S Menu Structure Chart .................................................................. 13Setup Submenu ................................................................................. 14Legal for Trade ................................................................................... 14Zero .................................................................................................... 15Calibration Unit .................................................................................. 16Full Scale Capacity ............................................................................ 17Graduation Size ................................................................................. 18Calibration Point ................................................................................. 19Menu Lock-out Switch Setting ........................................................... 20End/Store Settings ............................................................................. 20Scale Submenu .................................................................................. 21Reset to Factory Settings ................................................................... 21Averaging Level ................................................................................. 22Stability Level ..................................................................................... 24Unit Selection ..................................................................................... 25Auto-Zero Tracking ............................................................................ 26Auto-Off Timer .................................................................................... 27Retain Zero Data ................................................................................ 28End/Store Settings ............................................................................. 28Calibration Submenu ......................................................................... 29Procedure ........................................................................................... 29

  • 4Chapter 2 Setup and Calibration (Cont.)RS232 Submenu ................................................................................ 30Menu Protection ................................................................................. 30Legal for Trade Sealing ...................................................................... 31Audit Trail Seal ................................................................................... 32Paper seal .......................................................................................... 32Lead and Wire Seal ........................................................................... 32

    Chapter 3 Operation .......................................................................... 34Selecting a Weighing Unit .................................................................. 35Weighing ............................................................................................ 35Re-Zeroing ......................................................................................... 36

    Chapter 4 Troubleshooting ................................................................ 37Troubleshooting ................................................................................. 38Error Codes ........................................................................................ 39Care and Maintenance ....................................................................... 40Service Information ............................................................................ 40

    Chapter 5 Specifications, Replacement Parts and Accessories ........ 41Specifications ..................................................................................... 42Replacement Parts ............................................................................. 43Accessories ........................................................................................ 43

  • 5Chapter 1Chapter 1Chapter 1Chapter 1Chapter 1IntrIntrIntrIntrIntroduction andoduction andoduction andoduction andoduction andInstallationInstallationInstallationInstallationInstallation

    If the I5S indicator has alreadybeen installed, you may skip mostof this chapter. You may howeverneed to refer back to it if you changescale bases or the power supply,or need to access the menu lock-out switch.

    This chapter contains the follow-ing information:

    Introduction and description.

    Unpacking the indicator.

    Scale base connections.

    Power connections.

    Turning the I5S indicator ON.

    Introduction and Installation 11111

  • 6IntroductionIntroductionIntroductionIntroductionIntroductionThis manual covers installation, operation and troubleshooting for the Ohaus I5SIndicator Standard. To ensure proper operation of the indicator, please read thismanual completely.

    DescriptionDescriptionDescriptionDescriptionDescriptionThe Ohaus I5S Indicator Standard is a rugged, reliable electronic weight indicatorthat is made in the U.S.A. and is designed to be easy to operate. It can be easilyconnected to Ohaus or other brands of Scale Bases. The I5S Indicator has beenNTEP tested and found to comply with Class III and IIIL requirements of NIST HB44for 1:5000 performance in pounds, kilograms, ounces or grams. The I5S Indicatorcan be wall, bench or tower mounted. The I5S Indicator can drive up to four 350ohm load cells and provides capacity selections up to 10,000 lbs/kgs with amaximum resolution of 1:20,000. The indicator is constructed of extrudedaluminum with removable plastic end caps and is extremely durable. The I5Soperates from six Alkaline "C" batteries and can also be powered externally usingthe AC adapter supplied. A six and a half digit LCD display which is 0.6 inches/1.5 centimeters in height provides easy visibility when working at distances fromthe indicator. Two switches mounted on the front panel enable simple set up ofconfiguration and feature parameters. A menu lock-out switch can be set to lockout various functions of the indicator to prevent settings from being changed.Accessories include: an RS232 Interface Kit, a Wall Mount Accessory kit, and aTower Mounting Accessory Kit.

    I5S INDICATOR Standard

    11111 Introduction and Installation

  • 7UnpackingUnpackingUnpackingUnpackingUnpackingUnpack the I5S indicator at the installation site and verify that the following itemswere included in the carton:

    an AC Adapter 2 security screws a capacity card 2 washers a warranty card instruction manual

    It is recommended to save the carton and packing material for storing, transportingthe indicator or returning it for service.

    AssemblyAssemblyAssemblyAssemblyAssemblyAssembly consists of making the necessary electrical connections, and applyingpower to the unit.

    Scale Base ConnectionsScale Base ConnectionsScale Base ConnectionsScale Base ConnectionsScale Base Connections1. Remove two screws from the right

    end cap using a phillips screwdriver.

    2. Slide the printed circuit board par-tially out of the housing and discon-nect the battery cable connector.

    3. Pass the scale base cable throughthe liquid tight connector on the endcap.

    Housing

    End Cap

    Liquid TightConnector

    Terminal Block forScale Base WireConnections(Bottom)

    PowerConnector

    Battery CableConnector

    Introduction and Installation 11111

  • 8Scale Base Wire Connections

    Bottom ViewPC Board

    + EXCITATION+ SIGNAL

    - EXCITATION- SIGNAL

    GROUND

    4. Connect the scale base wires to theterminal block located on the bot-tom of the printed circuit board asshown in the illustration.

    Lock-outSwitch

    Battery CableConnector

    Slot for BatteryCable

    Top ViewPC Board

    5. Reconnect the battery connector tothe circuit board.

    6. Push the circuit board back into thehousing.

    Capacity CardCapacity CardCapacity CardCapacity CardCapacity CardType or write the capacity and readabil-ity data for the scale base on one of theperforated cards and slide it under thefront panel label near the display.

    Closing and Setup ProtectionClosing and Setup ProtectionClosing and Setup ProtectionClosing and Setup ProtectionClosing and Setup Protection If the setup information (explained in

    Chapter 2) does not need to be pro-tected from unauthorized changes,position the battery cable coming fromthe end cap into the slot on the end ofthe PC board, then replace the endcap.

    If you intend to lock out changes to thesetup selections you make, leave theend cap in position but do not tightenthe screws. You will need to accessthe menu lockout switch located onthe circuit board following the setupprocedure.

    If the unit is being powered by batter-ies during setup rather than the ACadapter, and you want to lock outsetup changes, replace the end capnow and remove it after setup to setthe menu lock-out switch.

    OffOn

    11111 Introduction and Installation

  • 9Power OptionsPower OptionsPower OptionsPower OptionsPower OptionsThe I5S may be operated using the ACAdapter supplied, or 6 Alkaline C bat-teries (not supplied). It is recommendedthat when the I5S is operated from bat-teries, the Auto-Off Timer feature maybe turned on to extend battery life.

    AC AdapterAC AdapterAC AdapterAC AdapterAC AdapterPlug the AC adapter connector into thereceptacle in the right end cap and plugthe adapter into a convenient AC outlet.

    Battery InstallationBattery InstallationBattery InstallationBattery InstallationBattery Installation

    1. Remove the two screws from theleft end cap using a phillips screw-driver.

    2. Remove the end cap.

    3. Insert 6 Alkaline "C" batteries in thebattery compartment orienting thepositive (+) and negative (-) ends asshown on the label at the rear of theindicator.

    4. Replace the end cap and

    two screws.

    6 "C" Size AlkalineBatteries

    End Cap

    Introduction and Installation 11111

  • 10

    kg Lights to indicate currentweighing unit is kilograms.

    g Lights to indicate currentweighing unit is grams.

    oz Lights to indicate currentweighing unit is avoirdupoisounces.

    Indicates displayed value isstable.

    Lights when scale is at centerof zero within + 1/4 d.

    lobAt Indicates low battery condi-tion.

    Indicates negative value.

    lb Lights to indicate currentweighing unit is pounds.

    Turning The Turning The Turning The Turning The Turning The IIIII5S ONONONONONAfter all connections have been made,

    turn the I5S ON by pressing .

    When first turned on, all display seg-ments and enabled indicators will lightbriefly, and then clear. This is followedby the software revision Sr n.n and ifLegal for Trade (LFT) is ON, a calibra-tion event counter (AC xxx) and afeature event counter (FC xxx).

    Display Indicator FunctionsDisplay Indicator FunctionsDisplay Indicator FunctionsDisplay Indicator FunctionsDisplay Indicator Functions

    To turn the To turn the To turn the To turn the To turn the IIIII5S OFF,5S OFF,5S OFF,5S OFF,5S OFF,

    Press and hold until the displayindicates OFF, then release it.

    Introduction and Installation 11111

  • 11

    Chapter 2Chapter 2Chapter 2Chapter 2Chapter 2Setup andSetup andSetup andSetup andSetup andCalibrationCalibrationCalibrationCalibrationCalibration

    The Setup mode enables you toprogram and calibrate the I5S.

    When you first install the I5S, youmust use the Setup mode to enterinformation about the scale basebeing used and to perform calibra-tion. You may also use it to pro-gram options that will customizethe I5S for your specific weighingrequirements.

    This Chapter explains how to:

    Access and use the Setupmode.

    Calibrate the indicator.

    Prevent unauthorized setupchanges using the menu lock-out switch.

    Seal the indicator for Legal forTrade installations.

    Setup and Calibration 22222

  • 12

    Setup Mode and the Setup Mode and the Setup Mode and the Setup Mode and the Setup Mode and the IIIII5S Menu Structure5S Menu Structure5S Menu Structure5S Menu Structure5S Menu StructureProgrammable features of the I5S are contained in menus which are accessed inthe Setup mode. The chart on page 2-3 shows how these menus are structured.

    To access the main menu and change settings:To access the main menu and change settings:To access the main menu and change settings:To access the main menu and change settings:To access the main menu and change settings:1. Make sure the indicator is OFF and the menu protection lockout switch on the

    PC board is in the OFF position.

    2. Press and hold until SETUP is displayed, then release it.

    For first time setup, you must go through all of the submenus to configure theindicator. After the indicator is in use, you can make changes to individual settingsand bypass others.

    The following rules apply when using the setup mode:

    Press to enter the displayed submenu or select a displayed setting.

    Press to change the displayed submenu or setting.

    To exit a submenu, save settings and return to the main menu:To exit a submenu, save settings and return to the main menu:To exit a submenu, save settings and return to the main menu:To exit a submenu, save settings and return to the main menu:To exit a submenu, save settings and return to the main menu:

    Repeatedly press until END is displayed, then press .

    To exit the main menu and return to the weighing mode:To exit the main menu and return to the weighing mode:To exit the main menu and return to the weighing mode:To exit the main menu and return to the weighing mode:To exit the main menu and return to the weighing mode:You must be in the main menu with either SETUP, SCALE, CAL or RS232displayed.

    Repeatedly press until END is displayed, then press to exit and returnto the weighing mode.

    22222 Setup and Calibration

  • 13

    SCALE

    Resetto FactorySettings

    Averaging LevelHI or LO

    Stability Level0.5, 1 or 3

    Units

    lb ON or OFF

    kg ON or OFF

    oz ON or OFF

    g ON or OFF

    End

    Auto Zero Tracking0.5, 1, 3 or OFF*

    Auto Off TimerON (2 min.) or OFF

    Retain Zero DataON or OFF

    End/Store Settings

    SETUP

    Legal For TradeON or OFF

    Zero2%, 100%

    Calibration Unitlb or kg

    Full Scale Capacity5, 7, 10, 15, 20, 25,30, 40, 50, 60, 75,100, 125, 150, 200,250, 300, 400, 500,600, 700, 750, 1000,1500, 2000, 2500,3000, 5000, 7500,10000

    Graduation Size.0005, .001, .002,.005, .01, .02, .05,.1, .2, .5, 1, 2, 5, 10

    Calibration Point5, 7, 10, 15, 20, 25,30, 40, 50, 60, 75,100, 125, 150, 200,250, 300, 400, 500,600, 700, 750, 1000,1500, 2000, 2500,3000, 5000, 7500,10000

    Menu Lockout Switch

    Setup ON or OFFScale ON or OFFCal ON or OFFRS232 ON or OFFEnd

    End/Store Settings

    END

    IIIII5S Menu Structure Chart5S Menu Structure Chart5S Menu Structure Chart5S Menu Structure Chart5S Menu Structure ChartMAIN MENU

    NOTE:RS232 menu appears onlywhen RS232 option isinstalled.

    RS232CAL

    CALIBRATIONperform

    Press to enter the displayed submenu

    or select a displayed setting.

    Press to change the displayed submenu

    or setting.* Not available with LFT set to ONFactory default settings are shown in underlined and boldface type.

    Setup and Calibration 22222

  • 14

    Setup SubmenuSetup SubmenuSetup SubmenuSetup SubmenuSetup Submenu1. With the indicator OFF, press and

    hold until SETUP is dis-played, then release it.

    2. When SETUP is displayed, press to enter the submenu. The

    first parameter, Legal For Trade, willbe displayed.

    Legal For Trade (LFT)Legal For Trade (LFT)Legal For Trade (LFT)Legal For Trade (LFT)Legal For Trade (LFT)This parameter is used to configure theI5S for Legal for Trade or non Legal forTrade use. Setting this parameter to ONautomatically sets certain parameters forlegal for trade requirements, disablesothers, and limits overall display resolu-tion to 1:5,000.

    To access the current setting (ON orOFF), press when LFT is dis-played.

    1. To change the displayed setting,press .

    2. To accept the displayed setting,press . To proceed to the

    next parameter (Zero), press when LFT is displayed.

    To exit the Setup submenu and storeyour settings, repeatedly press

    until END is displayed, then press to return to the main menu. To exit mainmenu and return to a weighing mode,repeatedly press until END is dis-

    played then press . The scale isnow in a weighing mode.

    Legal for Trade OFFZero 100% CalibrationUnit lb Full ScaleCapacity 5000Graduation Size 1Calibration Point 5000Menu Lock-out Switch

    Setup ONScale ONCalibration ONRS232 ON

    SETUP MENUSETUP MENUSETUP MENUSETUP MENUSETUP MENUFACTORY DEFAULTSFACTORY DEFAULTSFACTORY DEFAULTSFACTORY DEFAULTSFACTORY DEFAULTS

    22222 Setup and Calibration

  • 15

    ZeroZeroZeroZeroZero

    Zero specifies the percentage of fullcapacity load (2% or 100%) that may becleared by pressing .

    + A 2% Zero setting is used for hop-per scales or other large scaleswhere an accidental zero wouldlose the current weight. A 2% Zerois also required for Canadian Le-gal for Trade applications. Benchand counter scales normally use100% Zero setting.

    To access the current setting, press when ZERO is displayed.

    1. To change the displayed setting,press .

    2. To accept the displayed setting,press .

    To proceed to the next parameter (Cali-bration Unit), press when ZERO isdisplayed.

    To exit the Setup submenu and storeyour settings, repeatedly press

    until END is displayed, then press to return to the main menu.

    To exit main menu and return to a weigh-ing mode, repeatedly press until

    END is displayed then press . Thescale is now in a weighing mode.

    Setup and Calibration 22222

  • 16

    Calibration UnitCalibration UnitCalibration UnitCalibration UnitCalibration UnitThe I5S may be calibrated using poundsor kilograms. Select the desired calibra-tion unit as follows:

    To access the current setting (lb or kg),press when CAL UN is displayed.

    1. To change the displayed setting,press .

    2. To accept the displayed setting,press .

    To proceed to the next parameter (FullScale Capacity), press when CALUN is displayed.

    To exit the Setup submenu and storeyour settings, repeatedly press

    until END is displayed, then press to return to the main menu.

    To exit main menu and return to a weigh-ing mode, repeatedly press until

    END is displayed then press . Thescale is now in a weighing mode.

    22222 Setup and Calibration

  • 17

    Full Scale CapacityFull Scale CapacityFull Scale CapacityFull Scale CapacityFull Scale CapacityThe maximum capacity of the load re-ceiver is entered for Full Scale Capacityusing the same weighing unit selectedfor the Calibration Unit.

    + If Legal for Trade is selected, reso-lution will be limited to a maximumof 5,000 divisions.

    To access the current setting, press when FS CAP is displayed.

    1. To change the displayed setting,press .

    2. To accept the displayed setting,press .

    To proceed to the next parameter (Gradu-ation Size), press when FS CAP isdisplayed.

    To exit the Setup submenu and storeyour settings, repeatedly press

    until END is displayed, then press to return to the main menu.

    To exit main menu and return to a weigh-ing mode, repeatedly press until

    END is displayed then press . Thescale is now in a weighing mode.

    Available CapacitiesAvailable CapacitiesAvailable CapacitiesAvailable CapacitiesAvailable Capacities

    5, 7, 10, 15, 20, 25, 30, 40,50, 60, 75, 100, 125,150, 200,250, 300, 400, 500, 600, 700,750, 1000, 1500, 2000, 2500,3000, 50005000500050005000, 7500, 10000

    Setup and Calibration 22222

  • 18

    Graduation SizeGraduation SizeGraduation SizeGraduation SizeGraduation SizeThis parameter specifies the incrementsin which the display will count.

    If Legal For Trade is set to ON, resolu-tion (Capacity Graduations) is limitedto a maximum of 5000 divisions. Thegraduation selections available are lim-ited by the capacity entered under FullScale Capacity to prevent a combinationwhich exceeds 5,000 divisions.

    If the graduation size selected does notmeet the resolution requirement, theSetup menu cannot be exited. An errormessage is displayed and you are auto-matically returned to the GRAD param-eter. A valid graduation value must beselected.

    To access the current setting, press when GRAD is displayed.

    1. To change the displayed setting,press .

    2. To accept the displayed setting,press .

    To proceed to the next parameter (Cali-bration Point), press when GRADis displayed.

    To exit the Setup submenu and storeyour settings, repeatedly press

    until END is displayed, then press to return to the main menu.

    To exit main menu and return to a weigh-ing mode, repeatedly press until

    END is displayed then press . Thescale is now in a weighing mode.

    Complete List of GraduationComplete List of GraduationComplete List of GraduationComplete List of GraduationComplete List of GraduationSelectionsSelectionsSelectionsSelectionsSelections

    .0005, .001, .002,

    .005, .01, .02, .05,

    .1, .2, .5, 11111, 2, 5, 10

    22222 Setup and Calibration

  • 19

    Calibration PointCalibration PointCalibration PointCalibration PointCalibration PointThe calibration point is the weight valuethat will be used to calibrate the I5S. Theavailable values are listed in the adja-cent table. A value equal to or less thanthe full scale capacity value must beselected.

    To access the current setting, press when CAL PT is displayed.

    1. To change the displayed setting,press .

    2. To accept the displayed setting,press .

    To proceed to the next parameter (MenuLockout Switch Setting), press when CAL PT is displayed.

    To exit the Setup submenu and storeyour settings, repeatedly press

    until END is displayed, then press to return to the main menu.

    To exit main menu and return to a weigh-ing mode, repeatedly press until

    END is displayed then press . Thescale is now in a weighing mode.

    Available ValuesAvailable ValuesAvailable ValuesAvailable ValuesAvailable Values

    5, 7, 10, 15, 20, 25, 30, 40,50, 60, 75, 100, 125, 150,200, 250, 300, 400, 500,600, 750, 1000, 1500, 2000,2500, 3000, 50005000500050005000, 7500,10000

    Setup and Calibration 22222

  • 20

    Menu Lockout Switch SettingMenu Lockout Switch SettingMenu Lockout Switch SettingMenu Lockout Switch SettingMenu Lockout Switch SettingThe menu lock-out switch must be in theOFF position to change settings in thismenu. This menu specifies which sub-menus will be locked out when the menulockout switch is enabled. An ON settinglocks out the selected menus. To accessthe current settings (ON or OFF), press

    when LOCSW is displayed.NOTENOTENOTENOTENOTE: When Legal for Trade is set ON,you are required to lock the Setup and Calmenus.1. The first submenu -SETUP- will be displayed.

    2. Press to access the currentsetting.

    3. To change the displayed setting,press .

    4. To accept the displayed setting, press . To proceed to the next

    submenu, press .

    Set each submenu in the same manner.When END is displayed, press toreturn to the Setup submenu.

    To exit the Setup submenu and store yoursettings, press to display END, thenpress to return to the main menu.

    To exit main menu and return to a weigh-ing mode, repeatedly press until ENDis displayed then press . The scaleis now in a weighing mode.

    End/Store SettingsEnd/Store SettingsEnd/Store SettingsEnd/Store SettingsEnd/Store SettingsYou must use END to exit store Setupsubmenu settings. If you do not use ENDto exit the submenu, settings you havemade will not be stored.

    (when option is installed)

    22222 Setup and Calibration

  • 21

    Scale SubmenuScale SubmenuScale SubmenuScale SubmenuScale Submenu

    If you are already in the main menu, youcan repeatedly press until SCALEis displayed. If you are not already in themain menu, proceed as follows:

    1. With the indicator OFF, press andhold until SETUP is dis-played, then release it.

    2. Press to display SCALE.

    3. When SCALE is displayed, press to enter the Scale submenu.

    The first parameter, Reset to Fac-tory Settings, will be displayed.

    Reset to Factory SettingsReset to Factory SettingsReset to Factory SettingsReset to Factory SettingsReset to Factory SettingsThis parameter is used to reset all scalesubmenu parameters to the factory de-fault settings outlined in the adjacenttable.

    To reset all scale submenu settings,press when RESET is displayed.

    When is released, DONE will bedisplayed briefly, indicating RESET iscompleted.

    To proceed to the next parameter (Av-eraging Level), press when RE-SET is displayed.

    To exit the Scale submenu and store

    your settings, repeatedly press

    until END is displayed, then press

    to return to the main menu.

    SCALE MENUSCALE MENUSCALE MENUSCALE MENUSCALE MENUFACTORY DEFAULTSFACTORY DEFAULTSFACTORY DEFAULTSFACTORY DEFAULTSFACTORY DEFAULTS

    Averaging Level LOStability Level 0.5dUnits:

    lb ONkg ON g ONoz ON

    Auto Zero Tracking 0.5dAuto Off Timer OFFRetain Zero Data OFF

    Setup and Calibration 22222

  • 22

    Reset to Factory Settings (Cont.)Reset to Factory Settings (Cont.)Reset to Factory Settings (Cont.)Reset to Factory Settings (Cont.)Reset to Factory Settings (Cont.)To exit main menu and return to a weigh-ing mode, repeatedly press until

    END is displayed then press . Thescale is now in a weighing mode.

    Averaging LevelAveraging LevelAveraging LevelAveraging LevelAveraging LevelAveraging level compensates for vibra-tion or excessive air currents on thescale base. During operation, the indi-cator continually takes weight readingsfrom the load cell. Successive readingsare then digitally processed to achieve astabilized display. Using this feature,you specify how much processing youneed.

    HI and LO settings are available.

    HI setting:More processing, greater stability andslower stabilization time.

    LO setting:Less processing, less stability and fasterstabilization time.

    To access the current setting, press when AL is displayed.

    1. To change the displayed setting,press .

    2. To accept the displayed setting,press .

    22222 Setup and Calibration

  • 23

    To proceed to the next parameter (Sta-bility Level), press when AL isdisplayed.

    To exit the Scale submenu and storeyour settings, repeatedly press

    until END is displayed, then press to return to the main menu.

    To exit main menu and return to a weigh-ing mode, repeatedly press until

    END is displayed then press . Thescale is now in a weighing mode.

    Setup and Calibration 22222

  • 24

    Stability LevelStability LevelStability LevelStability LevelStability LevelThe stability level specifies how much adisplayed weight may change (in gradu-ations) while the stability indicator re-mains ON. When the displayed weightchanges beyond the allowable range,the stability indicator turns OFF. Stabil-ity values of 0.5, 1 or 3 graduations maybe selected.

    To access the current setting, press when STAB is displayed.

    1. To change the displayed setting,press .

    2. To accept the displayed setting,press .

    To proceed to the next parameter (UnitSelection), press when STAB isdisplayed.

    To exit the Scale submenu and storeyour settings, repeatedly press

    until END is displayed, then press to return to the main menu.

    To exit main menu and return to a weigh-ing mode, repeatedly press until

    END is displayed then press . Thescale is now in a weighing mode.

    22222 Setup and Calibration

  • 25

    Unit SelectionUnit SelectionUnit SelectionUnit SelectionUnit SelectionThe I5S permits you to display weight inpounds, kilograms, ounces and grams.To enable or disable various weighingunits, use the following procedure.

    To access the current settings, press when UNITS is displayed.

    The first unit, lb will be displayed.

    1. Press to access the currentsetting.

    2. To change the displayed setting,press .

    3. To accept the displayed setting,press .

    To proceed to the next unit, press .

    Set each unit in the same manner.

    When END is displayed, press toreturn to the Scale submenu.

    To proceed to the next parameter (AutoZero Tracking), press when UNITSis displayed. To exit the Scale submenuand store your settings, repeatedly press

    until END is displayed, then press

    to return to the main menu.

    To exit main menu and return to a weigh-ing mode, repeatedly press until

    END is displayed then press . Thescale is now in a weighing mode.

    NOTE: ON is the default settingfor lb, kg, oz, and g.

    Setup and Calibration 22222

  • 26

    Auto Zero TrackingAuto Zero TrackingAuto Zero TrackingAuto Zero TrackingAuto Zero TrackingAuto Zero Tracking compensates forchanges in zero reading due to temper-ature changes or build up of smallamounts of debris on the scale baseplatform. Select (0.5, 1, or 3 divisions orOFF), the value of weight change thatthe indicator will disregard in order tomaintain a zero reading.

    Auto Zero Tracking occurs only whenthe display indicates zero regardless ofthe load on the scale base.

    NOTENOTENOTENOTENOTE: If LFT is set ON, the Auto-ZeroTracking OFF setting is not available.

    To access the current setting, press when Auto-0 is displayed.

    1. To change the displayed setting,press .

    2. To accept the displayed setting,press .

    To proceed to the next parameter (AutoOff Timer), press when Auto-0 isdisplayed.

    To exit the Auto-Zero Tracking submenuand store your settings, repeatedly press

    until END is displayed, then press

    to return to the main menu.

    To exit main menu and return to a weigh-ing mode, repeatedly press until

    END is displayed then press . Thescale is now in a weighing mode.

    22222 Setup and Calibration

  • 27

    Auto Off TimerAuto Off TimerAuto Off TimerAuto Off TimerAuto Off TimerThis parameter can be set to automati-cally turn the I5S off when it is idle formore than 2 minutes. This is especiallyuseful when the unit is being batteryoperated and will extend the battery lifesignificantly. This feature can also beswitched off

    To access the current setting, press when A OFF is displayed.

    1. To change the displayed setting,press .

    2. To accept the displayed setting,press .

    To proceed to the next parameter (Re-tain Zero Data), press whenA OFF is displayed.

    To exit the Auto-Off Timer submenu andstore your settings, repeatedly press

    until END is displayed, then press

    to return to the main menu.

    To exit main menu and return to a weigh-ing mode, repeatedly press until

    END is displayed then press . Thescale is now in a weighing mode.

    Setup and Calibration 22222

  • 28

    Retain Zero DataRetain Zero DataRetain Zero DataRetain Zero DataRetain Zero Data

    Set this parameter to ON to save zerodata in permanent storage in case theindicator is turned off or power is lost.This feature is used primarily in hopperand tank scale applications where lossof weight data will have serious conse-quences.

    To access the current setting, press when RETAIN is displayed.

    1. To change the displayed setting,press .

    2. To accept the displayed setting,press .

    To exit the Retain Zero submenu and

    store your settings, press to dis-

    play END, then press to return tothe main menu.

    To exit main menu and return to a weigh-ing mode, repeatedly press until

    END is displayed then press . Thescale is now in a weighing mode.

    End/Store SettingsEnd/Store SettingsEnd/Store SettingsEnd/Store SettingsEnd/Store SettingsYou must use END to exit store Scalesubmenu settings. If you do not use ENDto exit the submenu, settings you havemade will not be stored.

    22222 Setup and Calibration

  • 29

    Calibration SubmenuCalibration SubmenuCalibration SubmenuCalibration SubmenuCalibration SubmenuThe calibration menu is used to performcalibration.

    If you are already in the main menu, you

    can repeatedly press until CAL isdisplayed and proceed with step 3 of thefollowing procedure. If you are not al-ready in the main menu, start with step 1.

    ProcedureProcedureProcedureProcedureProcedure

    Before beginning, make sure there is noload on the scale base and calibrationweight equal to the Calibration Point isavailable.

    1. With the indicator OFF, press andhold until SETUP is dis-played, then release it.

    2. Press twice to display CAL.

    3. When CAL is displayed, press to begin calibration.

    4. C 0 will be displayed. With no loadon the platform, press .

    5. When is released, -C- will bedisplayed followed by C n indicat-ing the calibration weight value.

    6. Place the weight on the load re-ceiver and press to acceptthe calibration data.

    7. The display will show -C-, return tothe weighing mode and display thecalibration weight value with thecurrent unit indicator.

    Setup and Calibration 22222

  • 30

    RS232 SubmenuRS232 SubmenuRS232 SubmenuRS232 SubmenuRS232 SubmenuThe RS232 submenu only appears inthe main menu if the option is installed inthe indicator. Refer to the instructionsprovided with the accessory for setupand operating information.

    Menu Protection (Lockout Switch)Menu Protection (Lockout Switch)Menu Protection (Lockout Switch)Menu Protection (Lockout Switch)Menu Protection (Lockout Switch)The lockout switch located on the circuitboard can be switched to the ON posi-tion to disable unauthorized changes tothe indicator setup. Only those submenusthat are turned ON in the LOCSW pa-rameter of the SETUP submenu are dis-abled. This is indicated by -SAFE- whenentering a locked out submenu.

    If -SAFE- is displayed, submenu settingsmay be reviewed but cannot be changed.

    Set the switch as shown in the illustrationand replace the end cap.

    Menu Lock-out switch shown inON position to lock out menus.

    ON

    OFF

    22222 Setup and Calibration

  • 31

    Legal For Trade SealingLegal For Trade SealingLegal For Trade SealingLegal For Trade SealingLegal For Trade SealingThe I5S has been tested and found tocomply with the Class III and IIIL require-ments of NIST Handbook 44.

    After the indicator has been tested andfound to comply with applicable regula-tions by a local Weights and Measuresofficial, it may be sealed as follows:

    1. Remove the right end cap and makesure that the menu lockout switchon the printed circuit board is set tothe OFF position.

    2. Enter the Setup submenu and go tothe LOCSW parameter. Set theSETUP and CAL LOCSW param-eters to ON. Refer to the MenuLock-out Switch Setting on page2-10 for detailed procedures.

    3. Set the menu lockout switch on theprinted circuit board to the ON posi-tion.

    4. If an audit trail or paper seal will beused, install the End Cap and tightenthe two 6-32 phillips pan headscrews.

    5. If a lead and wire seal will be used,replace the 6-32 phillips pan headscrews with the 6-32 hex socketsecurity screws and washers en-closed in the plastic bag containingthe accessories.

    Setup and Calibration 22222

  • 32

    22222 Setup and Calibration

    Audit Trail SealAudit Trail SealAudit Trail SealAudit Trail SealAudit Trail SealCalibration Event CounterCalibration Event CounterCalibration Event CounterCalibration Event CounterCalibration Event CounterA calibration event counter incrementsthe calibration event number by oneeach time the indicator is calibrated.

    This number can be verified each timethe indicator is turned on. It will bedisplayed in the following format wherexxx is the calibration event number.

    AC xxx

    Feature Event CounterFeature Event CounterFeature Event CounterFeature Event CounterFeature Event CounterThe feature event counter incrementsthe feature event number by one eachtime any of the following features (pa-rameters) are changed. Changing oneor any combination of these features willconstitute one event. The list of featuresmonitored is as follows:

    LFTLFTLFTLFTLFT ZEROZEROZEROZEROZERO CAL UNITSCAL UNITSCAL UNITSCAL UNITSCAL UNITSRETAINRETAINRETAINRETAINRETAIN STB LEVELSTB LEVELSTB LEVELSTB LEVELSTB LEVEL AVG LEVELAVG LEVELAVG LEVELAVG LEVELAVG LEVELAZT LEVELAZT LEVELAZT LEVELAZT LEVELAZT LEVEL AUTO PRINTAUTO PRINTAUTO PRINTAUTO PRINTAUTO PRINT STABLE PRINTSTABLE PRINTSTABLE PRINTSTABLE PRINTSTABLE PRINTFULL SCALEFULL SCALEFULL SCALEFULL SCALEFULL SCALE CAL POINTCAL POINTCAL POINTCAL POINTCAL POINT GRADGRADGRADGRADGRAD

    This number can be verified each timethe indicator is turned on. It will bedisplayed in the following format wherexxx is the calibration event number.

    FC xxx

    These numbers should be recorded onaccompanying documentation or on apaper seal that is affixed to the balancefor future verification.

  • 33

    Setup and Calibration 22222

    Paper SealPaper SealPaper SealPaper SealPaper SealApply the paper seal over one of the 6-32phillips pan head screws on the right endcap.

    Lead and Wire SealLead and Wire SealLead and Wire SealLead and Wire SealLead and Wire SealAfter the indicator has been inspected bya Weights and Measures official, it maybe sealed using a lead and wire seal asfollows:

    1. Replace the 6-32 phillips pan headscrews in the right end cap with thewashers and security screws provid-ed in the I5S packaging.

    2. Pass the wire through the holes in thesecurity screws on the right end cap.

    3. Pass the end of the wire through thelead seal and fasten the seal.

    LEAD AND WIRE SEAL

    PAPER SEAL

  • 34

    Chapter 3Chapter 3Chapter 3Chapter 3Chapter 3OperationOperationOperationOperationOperation

    Once the indicator has been setup and scale base connectionshave been made, you can performweighing operations.

    If the RS232 interface accessoryis being used, refer to the operat-ing instructions supplied with thataccessory.

    33333 Operation

  • 35

    Operation 33333

    Selecting a Weighing UnitSelecting a Weighing UnitSelecting a Weighing UnitSelecting a Weighing UnitSelecting a Weighing Unit

    Repeatedly pressing will causethe displayed unit indicator to change.Only the units which are set to ON in thesetup mode will appear. If the desiredunit indicator does not appear, refer toChapter 2 for setup instructions.

    WeighingWeighingWeighingWeighingWeighing1. Make sure there is no load on the

    scale base.

    2. Press to obtain a zero read-ing if necessary.

    3. Add the load to the scale base andread the weight on the display.

    When the stability indicator ( )ap-pears in the display, the reading isstable.

    *

  • 36

    33333 Operation

    Re-ZeroingRe-ZeroingRe-ZeroingRe-ZeroingRe-ZeroingUse the RE-ZERO switch to reset thedisplay to zero. When weighing materialor objects that must be held in a contain-er, re-zeroing enables the indicator tosave the containers weight in memoryand subtract it from the gross load on thescale base.

    1. Press to obtain a zero read-ing.

    2. Place an empty container on thescale base and read its weight onthe display.

    3. Press . The display will showzero and the container weight willbe stored in memory.

    4. Add the material to the container.As material is added its net weightwill be displayed.

    5. When the loaded container is re-moved, the weight of the containerwill be displayed as a negative (-)value.

  • 37

    Chapter 4Chapter 4Chapter 4Chapter 4Chapter 4TTTTTrrrrroubouboubouboubleshootingleshootingleshootingleshootingleshooting

    The information in this chapter isintended to help identify and cor-rect errors that may be made ininstalling or operating the I5S. Thischapter includes:

    A troubleshooting chart.

    A description of messages thatmay appear on the display.

    Information on proper care andmaintenance.

    Troubleshooting 44444

  • 38

    TroubleshootingTroubleshootingTroubleshootingTroubleshootingTroubleshooting

    PROBABLE CAUSE(S)Adapter not plugged inor properly connected.

    Batteries dead or notproperly connected.

    Load on scale base ex-ceeds allowable zero %entered in ZERO pa-rameter of Setup menu(see Chapter 2).

    Retain Zero Data is en-abled in Scale menu.

    Scale base motion or dis-turbances exceed cen-ter of zero criteria.

    Desired unit not set toON in Scale menu.

    RS232 communicationparameters set up incor-rectly.

    Cable connections.

    Scale base discon-nected.

    LOCSW set to ON andmenu protection switchset to ON position.

    REMEDY

    Check power cord con-nections.

    Make sure adapter con-nector is plugged all theway into the indicator.

    Check battery connector.

    Check orientation of thebatteries.

    Make sure end caps aresecure on indicator.

    Replace batteries.

    Remove load on scalebase to less than enteredzero %.

    Change allowable zero% in ZERO parameterof Setup menu (seeChapter 2).Normal operation whenthis feature is enabled.

    Remove disturbances orreduce motion.

    Increase averaging levelin AL menu.

    Enable desired unit (seeChapter 2).Verify communicationparameters.

    Check cable connec-tions.

    Check connections.

    Set menu protectionswitch to OFF position.

    SYMPTOM

    Unit will not turn on.

    Cannot rezero indicator,or will not re-zero whenturned on.

    Center of Zero displayindicator erratic or doesnot appear with no loadon scale base.

    Cannot display weight indesired weighing unit.

    RS232 not working.

    Unable to calibrate unit.

    44444 Troubleshooting

  • 39

    Troubleshooting 44444

    Error CodesError CodesError CodesError CodesError Codes

    DISPLAYEDMESSAGE DESCRIPTION

    Flashes at 10 second intervals when batteries are weak. Approxi-mately 2 minutes of operating time remain.

    Underload condition.

    Overload condition.

    Err 0: Invalid memory checksum, on either cal data, parameters, or both.Checksum is verified every time I5S is turned on. Check setupparameters in menus and recalibrate. If error persists, a hardwareproblem may exist and unit must be serviced.

    Err 1: Overload error. (Interface only). This error shows as all uppersegments on the display.

    Err 2: Underload error. (Interface only). This error shows as all lowersegments on the display.

    Err 4: Invalid calibration error. This error usually occurs during thecalibration procedure when there is insufficient signal for the ScaleBase as compared to the indicator set up parameters.

    In LFT operation, this error will occur if an attempt to calibrate theindicator when the active range of the load cell is less than1.2 mv/v for the active full-scale load. Recheck the load cell rating,set up parameters, and calibration weights.

    Err 14: Attempt to zero balance when zero criteria are not met. Zerocriteria can change depending on user settings and whether or notthe indicator is doing a power-on zero or a normal zero. See power-on zero operation for more details.

    Err 21: Calibration data does not match current Full Scale, Grad and CalPoint settings. Settings must be restored or the indicator must berecalibrated using the current settings.

    Err 22: The current combination of Grad and Full Scale settings exceedsrated resolution (1:5,000 LFT, 1:20,000 non-LFT). This can occurif the user modifies the Full Scale setting, then attempts to exit theSETUP menu without entering the Grad submenu. Once the user

  • 40

    44444 Troubleshooting

    entered the Grad submenu, graduation selection will be limited to choices that donot exceed rated resolution.

    Err 23: Retain is enabled in the Scale menu, but valid Retain data does notexist. Normal power-on zero operation will occur. The new power-on zero, since Retain is enabled, will be written to the EEPROM andused as valid retain data the next time the indicator is turned on.

    Err 24: An attempt was made to process an interface command that iscurrently locked by the lock switch.

    Err 25: Attempt to leave the SEL submenu with all units turned off.

    Care and MaintenanceCare and MaintenanceCare and MaintenanceCare and MaintenanceCare and MaintenanceTo keep your I5S operating properly, observe the following:

    Remove the power pack before cleaning.

    Use only a damp cloth and mild detergent for cleaning. Abrasives may damagethe display window.

    Be careful not to scratch the display window.

    Do not allow any liquid to flow inside the I5S.

    Do not store long periods of time with the batteries installed.

    Service InformationService InformationService InformationService InformationService InformationIf the troubleshooting section does not resolve or describe your problem, you willneed to contact an authorized Ohaus Service Agent. For Service assistance in theUnited States, call Ohaus Corporation toll-free at (800) 526-0659. An OhausProduct Service Specialist will be available to help you.

  • 41

    Chapter 5Specifications,ReplacementPartsand Accessories

    Specifications, Replacement Parts and Accessories 55555

  • 42

    55555 Specifications, Replacement Parts and Accessories

    SpecificationsSpecificationsSpecificationsSpecificationsSpecificationsCapacity (Ib or kg) 5 to 10,000*Graduations (readability) (lb or kg) 0.0005 to 10*Displayed resolution 1:5,000 LFT or 1:20,000 maximum* (Non LFT)Weighing modes lb, kg, oz, g*Over range capacity Capacity plus 9d Stab i l i za -tion time 3 seconds Auto-zero tracking

    capture rangeOFF, 0.5, 1 or 3 divisions* Zeroing range2% or 100% of capacity* Internal resolution 1:200,000

    Operating temperature range 32 to104F/0 to 40C (Non LFT) Operating humidity range10% to 95% RH Span calibrationpush-button (selectable up to 100% of scale basecapacity) Weighing system Strain gauge

    Load cell excitation voltage 5 V dcLoad cell input sensitivity Up to 3 mV/V (Non LFT)Load cell drive 60mA at 5V dc (drives up to four 350 ohm

    load cells)Display (in/cm) LCD (0.6/1.5) highPower requirements AC adapter: 100, 120, 220, 240 V ac,

    50/60 Hz or 6 alkaline "C" batteriesTypical battery life 125 hours with one 350 ohm load cell(without RS232) 55 hours with four 350 ohm load cellsDimensions(WxHxD) (in/cm) 7.5 x 6 x 3/19.1 x 15.2 x 7.6Net weight (Ib/kg) 2.7/1.2Shipping weight (lb/kg) 4/2

    NIST/NTEP NIST HB44 Class III and IIIL nmax

    =5000dCertificate of conformance No. 93-112A1Operating temperature50 to 104F/10 to 40CLoad cell input sensitivity 1.2 to 3 mV/V

    Electrical ratings (for power 100/120 V UL1910, CSA C22.2 no. 66 andadapters only) 107, 230 V (Europe)/ 240 V (UK) EN60950,

    240 V (Australian) IEC 950CE electromagnetic compatiblity Directive No. 89/336/EEC

    EN50081-1 and EN50082-1

    * User selectable

  • 43

    Specifications, Replacement Parts and Accessories 55555

    Replacement PartsReplacement PartsReplacement PartsReplacement PartsReplacement PartsOHAUS

    DESCRIPTION PART NO.

    AC Adapters:100/120 V ac North American Plug 90524-77230 V ac, 50 Hz, Euro Plug 90524-75230/240 V ac, 50 Hz, UK Plug 90524-79240 V ac Australian Plug 90524-76

    AccessoriesAccessoriesAccessoriesAccessoriesAccessoriesOHAUS

    DESCRIPTION PART NO.

    Wall Mount Accessory Kit 77688-01Tower Accessory Kit 77689-01RS232 Interface Kit 77694-01

  • 44

    Limited WarrantyLimited WarrantyLimited WarrantyLimited WarrantyLimited WarrantyOhaus products are warranted against defects in materials andworkmanship from the date of delivery through the duration of thewarranty period. During the warranty period Ohaus will repair, or,at its option, replace any component(s) that proves to be defectiveat no charge, provided that the product is returned, freight prepaid,to Ohaus.

    This warranty does not apply if the product has been damaged byaccident or misuse, exposed to radioactive or corrosive materials,has foreign material penetrating to the inside of the product, or asa result of service or modification by other than Ohaus. In lieu of aproperly returned warranty registration card, the warranty periodshall begin on the date of shipment to the authorized dealer. Noother express or implied warranty is given by Ohaus Corporation.Ohaus Corporation shall not be liable for any consequentialdamages.

    As warranty legislation differs from state to state and country tocountry, please contact Ohaus or your local Ohaus dealer forfurther details.