Model 480 and 481 Orbital Shaker* Operating and Maintenance Manual 7020480 Rev. 3

Upload: others

Post on 21-Jan-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

Model 480 and 481Orbital Shaker*Operating and Maintenance Manual 7020480 Rev. 3

Page 2: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

Thermo ScientificThermo Scientific

MANUAL NUMBER 7020480

3 26178 7/23/10 Removed CSA from Model 481 as N/A for 50hz units ccs

2 25244/OS-340 2/13/09 Corrected stray line on 480-70-0-D to Rev. 2 ccs

1 24855 8/7/08 Re-inserted platform installation instructions ccs

0 19233/SI-9436 10/5/05 Condensing unit P/N from 205064 to 205142, unit to Release 12 ccs

REV ECR/ECN DATE DESCRIPTION BY

Preface

Orbital Shaker i

* Triple counter-balanced, single eccentric drive mechanism (U.S. Patent#5,558,437)

* Horizontal, HEPA-filtered airflow design (U.S. Patent #5,577,837)* Test Tube Rack (U.S. Patent #5,632,388)

Page 3: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

Thermo Scientificii Orbital Shaker Thermo Scientific

Preface

Important Read this instruction manual. Failure to read, understand and follow the instructions in this manualmay result in damage to the unit, injur y to operating personnel, and poor equipment per formance. s

Caution All internal adjustments and maintenance must be per formed by qualified service personnel. s

Material in this manual is for information purposes only . The contents and the product it describes are subject tochange without notice. Thermo Scientific makes no representations or warranties with respect to this manual. Inno event shall Thermo be held liable for any damages, dir ect or incidental, arising out of or r elated to the use ofthis manual.

©2005 Thermo Scientific. All rights reserved.

Page 4: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

Thermo Scientific Orbital Shaker iiiThermo Scientific

Preface

Important operating and/or maintenance instructions. Read the accompanying text carefully.

Potential electrical hazards. Only qualified persons should per form procedures associated with thissymbol.

Equipment being maintained or ser viced must be turned off and locked off to pr event possible injury.

Hot surface(s) present which may cause burns to unpr otected skin, or to materials which may bedamaged by elevated temperatures.

Marking of electrical and electronic equipment, which applies to electrical and electr onic equipmentfalling under the Directive 2002/96/EC (WEEE) and the equipment that has been put on the mar ketafter 13 August 2005.

This product is required to comply with the European Union’s Waste Electrical & ElectronicEquipment (WEEE) Directive 2002/96/EC. It is marked with the WEEE symbol. Thermo Scientifichas contracted with one or more recycling/disposal companies in each EU Member State EuropeanCountry, and this product should be disposed of or r ecycled through them. Further information onThermo’s compliance with this directive, the recyclers in your country and information on Thermoproducts will be available at www.thermofisher.com.

4 Always use the proper protective equipment (clothing, gloves, goggles, etc.)

4 Always dissipate extreme cold or heat and wear protective clothing.

4 Always follow good hygiene practices.

4 Each individual is responsible for his or her own safety.

Page 5: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

Thermo Scientificiv Orbital Shaker Thermo Scientific

Preface

Do You Need Information or Assistance on

Thermo Scientific Products?

If you do, please contact us 8:00 a.m. to 6:00 p.m. (Eastern Time) at:

1-740-373-4763 Direct

1-800-438-4851 Toll Free, U.S. and Canada

1-877-213-8051 FAX

http://www.thermoscientific.com Internet Worldwide Web Home Page

Service E-Mail Address

Thermo Fisher Scientific

401 Millcreek Road, Box 649

Marietta, OH 45750

Our staff can provide information on pricing and give you quotations. We can

take your order and provide delivery information on major equipment items or make

arrangements to have your local sales representative contact you. Our products are listed on the

Internet and we can be contacted through our Internet home page.

Our staff can supply technical information about proper setup, operation or

troubleshooting of your equipment. We can fill your needs for spare or replacement parts or

provide you with on-site service. We can also provide you with a quotation on our Extended

Warranty for your Thermo Scientific products.

Whatever Thermo Scientific products you need or use, we will be happy to discuss your

applications. If you are experiencing technical problems, working together, we will help you

locate the problem and, chances are, correct it yourself...over the telephone without a service

call.

When more extensive service is necessary, we will assist you with direct factory trained

technicians or a qualified service organization for on-the-spot repair. If your service need is

covered by the warranty, we will arrange for the unit to be repaired at our expense and to your

satisfaction.

Regardless of your needs, our professional telephone technicians are available to assist you

Monday through Friday from 8:00 a.m. to 6:00 p.m. Eastern Time. Please contact us by

telephone or fax. If you wish to write, our mailing address is:

International customers, please contact your local Thermo Scientific distributor.

Sales Support

Service Support

[email protected]

Page 6: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

Orbital Shaker vThermo Scientific

Table of Contents

Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1Pallet Hold-down Shipping Bracket . . . . . . . . . . . . . . . . . . . . . . . . . .1-1Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1Installing the Cabinet Stand-off Bolts . . . . . . . . . . . . . . . . . . . . . . . .1-2Chamber Drain . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2Condensate Drain . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3Installing the Platform . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3Foot Pedal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6Assembling Flask Clips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6Installing the Flask Clips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6Installing the Test Tube Holders . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-8RS-232 Interface Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-9Connecting the Remote Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-11Lid Security Lock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-12Connecting the Unit to Electrical Power . . . . . . . . . . . . . . . . . . . . .1-12

Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1Operating the Control Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2Quick Start-Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3Factory Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3Other Factory Defaults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4Changing Temp, Speed & Time Settings . . . . . . . . . . . . . . . . . . . . . .2-4

Changing the Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4Changing the RPM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-5Changing the Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-5

Shaker Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6Overtemp Shutdown . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-7Undertemp Shutdown . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-7Cycle Complete . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-7Power Failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-7RPM Tracking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-8Check Belt . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-8Sensor Fault . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-8Temperature High or Low . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-9Platform Stalled . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-9Check Fuse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-9

Changing the Shaker Configuration . . . . . . . . . . . . . . . . . . . . . . . . .2-10Turning the Audible Alarm On and Off . . . . . . . . . . . . . . . . . . . .2-11Setting Alarm Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-11

Section 1

Section 2

Page 7: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

vi Orbital Shaker Thermo Scientific

Calibrating the RPM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-16Calibrating the Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-17Remote Alarm System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-17

Cycle Complete . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-18Power Failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-18RPM Tracking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-18Check Belt . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-19Sensor Fault . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-19Temperature High or Low . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-19Viewing the Total Operating Hours . . . . . . . . . . . . . . . . . . . . . . .2-19

Heat % . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-20Software Version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-21Temperature Sensor Readings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-21

Overtemp Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-21Temperature Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-22Defrost Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-22

Turning the Defrost System On and Off . . . . . . . . . . . . . . . . . . . .2-23Setting the Defrost Temperature . . . . . . . . . . . . . . . . . . . . . . . . . .2-23

Hold Temperature Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-24MENU MAP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-26

Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1Shaker Platform and Cabinet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1Control Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1Cleaning/Replacing the Cabinet Air Filter . . . . . . . . . . . . . . . . . . . . .3-1Servicing the Drive Belt . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2PREVENTIVE MAINTENANCE . . . . . . . . . . . . . . . . . . . . . . . . . .3-3

Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-1Alarms and Alarm Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-1Changing the HEPA Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2If the Shaker Will Not Operate . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2Circuit Boards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-4Temperature Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-5Blower Fan Motors and Heating Elements . . . . . . . . . . . . . . . . . . . . .4-5Heater Element Circuit Breaker . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-6Tuning the Cabinet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-6

Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-1Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-1Refrigeration Schematics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-1Electrical Schematics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-1Warranty Information

Table of Contents

Section 2 (continued)

Section 3

Section 4

Section 5Section 6

Section 8Section 7

Page 8: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

Orbital Shaker 1-1Thermo Scientific

Pallet Hold-downShipping Bracket

Section 1 Installation

The following materials are supplied with Model 480/481 Console Shaker:

2 - Keys for the lid lock (packaged and attached to outside of unit)1 - T-handle 5/32” hex socket wrench2 - Platform alignment studs 1/4-201 - Shaker Platform 6 - Grade 8, 5/32” hex socket flat head scr ews (provided with platform)1 - 3/4” open end wr ench2 - 1/4-20 x 7” stand-off bolts with r ubber caps1 - Screwdriver for flask clip installation and r emoval

To secure the console shaker to theshipping pallet, hold-down brackets areattached to slots in both sides of thecabinet. The brackets are fastened to thewood pallet with lag screws (Figure 1-1).

Locate the shaker on a firm, lev el surface in an area free of dust and dir t.To allow for lid opening, the back of the shaker must be at least 4½ inchesfrom the wall. Allow 6 inches clearance on both sides for v entilationpurposes.

Figure 1-1. Bracket

Location

Figure 1-2. Front and Side View

Page 9: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

1-2 Orbital Shaker Thermo Scientific

Section 1Installation

The 480/481 console shaker features front-to-back air circulation forcooler compressor temperatures, increased performance and reliability andlonger compressor life. Therefore, the fan area on the back of the cabinetmust be at least 4 inches fr om the back wall or other obstr uction. Tomaintain this distance, two 1/4-20 x 7” bolts ar e included in the par tsbag.

Screw the bolts in the threaded holes on the back of the shaker cabinet(Figure 1-2). The bolts should be screwed in by hand to the limit of thethreads. Further tightening is unnecessary. Put the protective rubber capsalso supplied in the par ts bag over the heads of the bolts.

A drain is provided in the bottom of the chamber for conv enience whencleaning or removing spills. (Figure 1-3) A clear vinyl hose and plasticvalve is connected to the drain and is accessed b y removing the front grilleassembly and the lower front panel. The grille is removed by gentlypulling it off. It is held in place by six push-in type retainers.

To remove the cabinet panel located in back of the grille, r emove sixPhillips screws, three on the top and three on the bottom. It may also benecessary to loosen the two left side P hillips screws holding the kick panelassembly to the shaker frame.

Installing the CabinetStand-off Bolts

Threaded holes for

stand-off bolts

Figure 1-2. Bolt Locations

Chamber Drain

Figure 1-3. Chamber Drain Hose Location

Page 10: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

Orbital Shaker 1-3Thermo Scientific

Section 1Installation

A 3/8” O.D. stainless steel condensate drain is located on the back of theshaker to remove any water which may collect in the air duct wor k. Referto Figure 1-4.

Note If your platform is already installed, skip this section and continue tonext section. s

Caution Remove the shipping bracket and install the shaker platformbefore plugging in or attempting to operate the unit. s

After removing the orbital mechanism shipping bracket and installing theplatform, remove this protective decal from the control panel to beginshaker operation.

Figure 1-6. Shipping Bracket

To protect the shaker’s orbital mechanism during shipment, a sheet metalshipping bracket (Figure 1-6) is installed and must be r emoved before theunit can be operated. Using a 7/16” and 9/16” hex wr ench, remove thethree 1/4” and the single 3/8” scr ews. Retain this hardware for futureshipping.

Condensate Drain

Condensate Drain

Figure 1-4. Drain Location

Installing the Platform

Figure 1-5. Warning Decal

Page 11: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

This shaker accommodates either a 5/16” heavy-duty, 29.5” wide x 18”front-to-back platform.

Note If installing the larger platform, RPM is limited to 400. The 194012counterweight kit must also be installed b y qualified service personnel. s

All shaker platforms are attached to their orbital mechanisms with six 1/4-20 hex socket flathead screws Grade 8. These screws are hardened andshould not be exchanged with any other screw type. The 5/32” hex socketwrench, included with the shaker, must be used when attaching theplatform. Refer to Figure 1-7.

Caution Do not attempt to use a P hillips head screwdriver. s

Figure 1-7. T-Handle Wrench and Hex Socket Head Screw

1-4 Orbital Shaker Thermo Scientific

Section 1Installation

Installing thePlatform (continued)

2) Platform mounting

studs (2). Insert alignment

pins.

1) Platform mounting

holes on drive

mechanism (4)

3) Chamber drain

1

2

3

Figure 1-8. Hole Locations

Page 12: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

1. Insert the two 1/4-20 alignment pins into the two mounting holesidentified in Figures 1-8 and 1-9.

2. Rotate the drive mechanism until the four mounting holes generallymatch the holes in the platform.

3. Place the platform onto the shaker and o ver the alignment pins.

4. Move the platform in an orbital motion until one or mor e of thecenter mounting holes are located.

5. Insert the hex socket head scr ews as the four holes ar e located. Do nottighten the screws.

6. Remove the 1/4-20 alignment pins and r eplace them with theremaining two hex socket screws.

7. Tighten all screws using the T-handle wrench.

Caution Use only the hex socket flat scr ews to fasten the platform, andonly the T-handle wrench to tighten the screws. Torque these screws to 10ft/lbs. Check these screws monthly if the unit is operated at or nearmaximum RPM (525 RPM). s

Orbital Shaker 1-5Thermo Scientific

Section 1Installation

Installing thePlatform (continued)

= Platform Mounting Holes

= Mounting Holes for Off-Brand Equipment

Alignment Pins

Figure 1-9. Hole Locations

Page 13: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

The lid of the console shaker is counterbalanced for ease of opening andclosing. A foot pedal on the lo wer right of the front of the cabinet is alsoprovided for operator convenience and ready access to the chamber.

Each flask clip up to 6.0 liters in siz eincludes a metal spring which must beinstalled onto the clip. For flask clipsthrough 500 ml, insert the end of eachspring into the holes on the top of the clipleg (Figure 1-10).

The 2 liter, 2.8 liter, 4 liter, 5liter, and 6 liter flask clipsinclude two metal springs andrubber spring tubes. On theselarger clips, the springs areinstalled by hooking their endstogether as illustrated in Figure1-6. The upper spring andspring tubes should be installedprior to mounting the clip tothe platform. The lower springand spring tubes, however, areplaced around the bottom ofthe clip legs after the flask clipis fastened to the platform.

Note that the rubber spring tubes are placed between the clip legs.

The Model 480/481 accommodates glassware in numbers and sizes fromninety-one 25 ml flasks to four 6 liter flasks. All Thermo platforms havemounting holes for flask clips and test tube racks made b y othermanufacturers.

1-6 Orbital Shaker Thermo Scientific

Section 1Installation

Figure 1-10. Insert Springs

Upper spring and

spring tubes

Lower spring and

spring tubes

Clip leg

Figure 1-11. Spring Installation

Installing the FlaskClips

Foot Pedal

Assembling FlaskClips

Page 14: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

Table 1-1 lists the flask clip assemblies and kits av ailable for Thermoshakers.

Table 1-1. Flask Clip Capacity and Kit Assemblies Dedicated Platform

Flask clips can be attached anywhere on the shaker platform. The counter-balanced design of these shakers accommodates the worst unbalanced load.

The flask clips are supplied with the proper screws and are attached to theplatform with a standard Phillips screwdriver or with the screwdriverprovided with the unit.

Figures 1-10 and 1-11 illustrate the installation of the flask clips. N ote thatclips for 1, 2, 2.8, 4, 5, and 6 liter flasks use fiv e screws. The 250/300lflask clip has an adhesive-backed flask cushion pad which is installed onthe flat base of the clip body. A hole is provided in the pad for themounting screw.

Orbital Shaker 1-7Thermo Scientific

Section 1Installation

Dedicated platform no.

No. ofClips Flask Size Springs per Clip Screws per

Clip

238017 91 25ml 1 1

238018 91 50ml 1 1

238019 40 125ml 1 1

238020 40 250/300ml 1 w/ lg pad 1

238021 24 500ml 1 1

238022 15 1l 1 5

238023 12 2l 1 5

238024 6 4l 2 w/10 tubes 5

238025 4 6l 2 w/12 tubes 5

238026 6 2.8l 2 w/10 tubes 5

Installing the FlaskClips (continued)

Page 15: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

The accessory test tube racks and test tube rack holders ar e available infour sizes and are listed in Table 1-2. All the test tube rack holders ar eadjustable in seven positions, swinging and locking at 15°, 30° and 45° ineither direction. Figure 1-12 illustrates the Test Tube Rack Holder with therack in place.

To remove the rack, spread the metal tabs on either end of the holder andlift it out.

1-8 Orbital Shaker Thermo Scientific

Section 1Installation

..

.

.

.

.

..

.

.

.

.

.

..

.

.

.

.

..

.

.

.

.

.

..

.

.

.

.

..

.

.

.

.

.

..

.

.

.

.

..

.

.

.

.

.

..

.

.

.

.

..

.

.

.

.

.

..

.

.

.

.

..

.

.

.

.

.

....

....

....

....

....

....

....

....

....

....

....

....

....

....

....

....

....

....

..

.

.

....

.

.

.

...

.

.

....

.

.

.

.

..

.

.

....

.

.

.

...

.

.

....

.

.

.

.

..

.

.

....

.

.

.

...

.

.

....

.

.

.

.

Spread metal tabs outwardto remove test tube rack

Locking Pin to swing/tilttest tube rack 15°, 30°, 45°

Installing the TestTube Holders

Figure 1-12. Holder with Rack

Part No. Description

950040 Test tube rack 10-13mm size

950060 Test tube rack 16-20mm size

600074 Test tube rack 21-25mm size

600075 Test tube rack 26-33mm size

600076 Adjustable-angle test tube rack holder w/rack 10-13mm

600077 Adjustable-angle test tube holder w/rack 16-20mm

600078 Adjustable-angle test tube holder w/rack 21-25mm

600079 Adjustable-angle test tube holder w/rack 26-30mm

600089 2 tier microplate rack

600090 3 tier microplate rack

194024 #10-24 Pan head Phillips screws for mounting test tube holders to Thermoorbital shaker platforms

185062 Pan head Phillips screws, washers and nuts for mounting test tube holders toThermo 2568 and 2569 shaker baths

Figure 1-2. Racks and Holders

Page 16: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

To install the Test Tube Rack Holder ontothe shaker platform, remove the rack androtate the swing-bed of the holder 90° b ypulling the knobs of the locking pins oneither end of the holder outward. Thepins are locked in the outward position byturning the knobs 1/4-turn (Figure 1-13).

The Model 480/481 Orbital Shaker is equipped with an RS-232 S erialCommunication Interface for the remote transmission of data. An RJ-11telephone style connector is located on the back of the incubator . A cablewith RJ-11 plugs and an RJ-11 to DB-25 adapter ar e required. Refer toFigure 1-15 for connector locations on the shaker back panel. F igure 1-14identifies the pin connections.

Orbital Shaker 1-9Thermo Scientific

Section 1Installation

RS-232 InterfaceConnector

Installing the TestTube Holders (cont.)

Figure 1-13. Rotate Bed

Figure 1-14. RS-232 and Remote Alarm Connectors

Figure 1-15. Pins

Page 17: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

The data is “dumb terminal” formatted, which permits inter facing witheither a computer or a serial printer .

Three wires are used for the RS-232 inter face:1. Transmit data (/TXD) - pin 2 DB-25 connections2. Receive data (/RXD) - pin 3DB-25 connections3. Signal ground (GND) - pin 7DB-25 connections

The data format is:Baud . . . . . 1200 (9600 baud w/ jumper at J2 on M ain Control Board)Data bits . . . . . .8 (7 bit ASCII with leading z ero)Start bits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1Stop bits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1Parity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .none

The data transfer sequence is transmitted in the follo wing format. X refersto the numerical time, RPM and temperatur e.

(NUL)XX.XX(H)(SP)(SP)XXXRPM(SP)(SP)XX.XC(SP)(LF)(CR)(EOT)

NUL - Null character (0)

SP - Space

LF - Line feed

CR - Carriage r eturn

EOT - End of text (4)

H - Hold Mode

The Model 480/481 transmits time, RPM and temperatur e informationone minute after power is first applied to the unit and then ev ery 60minutes.

The shaker’s microprocessor responds to two ASCII commands from theremote: DC1 (XON), and DC3 (XOFF)

1-10 Orbital Shaker Thermo Scientific

Section 1Installation

RS-232 InterfaceConnector (cont.)

Page 18: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

The shaker will transmit Time, Temperature and RPM data upon r eceiving“DC1” (XON) and will restart 60 minute interval transmissions unlessinhibited by a “DC3” (XOFF).

Receiving a “DC3” (XOFF) from the remote inhibits the shaker fromsending serial data indefinitely until a “DC1” (X ON) is received.

An internal, remote alarm SPDT relay is provided to monitor alarms andis connected by an RJ-11 (telephone style) jack located on the r ear of thecabinet. The relay provides NO (normally open) and NC (normallyclosed) output and may be wir ed to a central remote alarm location or toan independent alarm system.

Figure 1-14 identifies the pin contacts. F igure 1-16 shows the location ofthe Remote Alarm Connector.

A modular to modular cable, Stock No. 190388 and an RJ-11 telephonestyle terminal converter, Stock No. 190392 or equivalent may be used toconvert to a screw terminal connection. Refer to Figures 1-15 and 1-16.

Orbital Shaker 1-11Thermo Scientific

Section 1Installation

DC1 (17, 11 Hexadecimal)

DC3 (19, 13 Hexadecimal)

Connecting theRemote Alarm

IMPORTANT USER INFORMATIONIMPORTANT USER INFORMATION

�������� ���� ����� ������ �� ��������� � �������� �� ������� ��������������� ���� ���� �� �������� �� ��������������� ��� �������� �� �� ����������� ���������� ��������� ������ ��!���������� �����

Figure 1-16. Junction Box

Page 19: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

Connecting the Unitto Electrical Power

To protect the contents of the shaker or pr event tampering or unauthorizedaccess, a security lock is located on the right side of the lid. S ee theillustration below. Two keys for this lock ar e in the parts package attachedto the outside of this unit when shipped.

See the serial tag on the side of the unit for electrical specifications or r eferto the electrical schematics at the end of this manual.

Caution Connect the orbital shaker to a gr ounded, dedicated circuit.The power ON/OFF switch is the mains disconnect device for the orbitalshaker. Position the shaker so the switch is easily accessible. s

1-12 Orbital Shaker Thermo Scientific

Section 1Installation

Lid Security Lock

Figure 1-17. Rear and Side Panels

Figure 1-18. Front View with Grille and Kick Plate Removed

Security

Key Lock

Foot

Pedal

Condensate Drain

Access

Port

Alarm

Contacts

30V max/

1A max

RS-232

Interface

Power Switch/Circuit Breaker

Power Inlet

Page 20: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

Orbital Shaker 2-1Thermo Scientific

Section 2 Operation

The Console Model 480/481 is a microprocessor controlled, incubator/refrigerated orbital shaker designed to accommodate a wide v ariety offlasks, test tubes and other glassware. The control system is easilyprogrammed and stores the user-defined time, temperature and speedsettings which remain in memory even when the shaker is turned off andunplugged.

The computer-based speed controller continuously adjusts for line voltagefluctuations and provides smooth start-ups and consistent RPM control.The circuitry is designed to slowly bring the platform up to speed anddown to a stop to prevent liquid splashing from flasks or test tubes.

The insulated lid with viewing port is counter-balanced for easy openingby hand or foot pedal. A conv enience interlock requires that the lid beclosed for the drive motor, circulating fans and refrigeration system tooperate.

Caution The microprocessor speed control system may take up to oneminute to bring the platform up to speed. N ever leave the shakerunattended when starting. Make sure all flasks and test tube racks ar efirmly seated in the clips and check the security of the flask clip andplatform attachment screws monthly. Air circulating fans will only operatewith the lid closed. Do not operate the shaker at maximum RPM withouta load. s

Figure 2-1. Front View

Page 21: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

2-2 Orbital Shaker Thermo Scientific

Section 2Operation

Operating the ControlPanel

The Model 480 control panel has a liquid cr ystal display and eightoperating keys or buttons which ar e identified by word and symbol.During programming activities, the up and down arrows increase anddecrease the numerical values of time, platform speed, or temperatur e.Pressing and holding either arrow will cause the values to scroll in thatdirection. Pressing and holding for about fiv e seconds will increase thescrolling speed.

When programming the system configuration, the Up arrow returns thedisplay to the system’s Operating Screen showing the Time, RPM andTemperature, while the Down arrow advances the display to the nextprogramming screen.

The alarm indicator and alarm silence button complete the shaker contr olpanel. When in alarm, the unit sounds an audible warning and flashes thethree red indicators. Pressing the Silence button turns off the audibleportion of the alarm. However, the three flashing indicators continue toflash until the alarm condition is corr ected. The audible warning willsound again in about fifteen minutes if the condition continues.

The alarm features are discussed in detail later in this section.

Figure 2-2. Control Panel

Page 22: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

Orbital Shaker 2-3Thermo Scientific

Section 2Operation

Quick Start-Up The Model 480 Orbital Shaker may be operated as soon as the platform isinstalled, the unit is plugged in, and turned on.

Note At power up, the screen at the right willbriefly appear. s

Pressing Start and Stop will operate the shaker at the factor y settingsshown in Figure 2-2. When starting, the Actual numbers along the top ofthe liquid crystal display will differ from the Setpoint values shown alongthe bottom. These numbers will change as the unit begins to operate.

Time - With the time set at Hold, the time display in the upper left por-tion of the screen will begin to count upward, showing the total operat-ing hours and minutes. The system will reset to 00:00 whenever the unitis stopped and restarted, using the Stop and Start buttons. The unit willnot reset if the unit is turned off and on using the po wer switch, or if theconsole lid is repeatedly opened and closed.

Speed - The Actual speed will display zero RPM’s and will gradually riseto the setpoint as the platform begins its motion.

Temperature - The temperature, shown in the upper right por tion of theliquid crystal display, will indicate the actual ambient temperatur e insidethe cabinet and will gradually move toward the 37° setpoint.

The values shown in Figure 2-2 are set at the factory and are considereddefault values. Other factory settings are shown in the table below.

The Model 480/481 Shaker has been shipped from the factory with thefollowing default settings:

Time: When the shaker is turned on for the first time, the liquid cr ystaldisplay will show 00:00H. (Hold time) This means the unit is set torecord accumulated operating time. Any pr ogramming changes in theTime settings are made in increments of five minutes.

RPM: The display shows the unit ready to operate at 100 RPM.Programming changes in the RPM are made in increments of 1 RPM.However, if the Up or Down buttons are held for about two seconds, thedisplay will scroll in that direction.

Temperature: The display shows the operating temperature set at 37°C.Changes to the Temperature program settings are made in increments of0.1°C. However, if the Up or Down buttons are held for about two sec-onds, the display will scroll in that direction.

Factory Settings

Page 23: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

To get the most from the shaker’s microprocessor-based technology, theModel 480 can be easily programmed to meet the most demandinglaboratory requirements. The following sections outline the procedures forchanging the settings and for programming the shaker control system.

When the unit is turned on or when the shaker is operating, Temperature,Time and RPM values are displayed on the LCD. A typical screen isillustrated in Figure 2-2. For convenience, this is called the OperatingScreen throughout the operating manual. All programming or settingchanges start from this screen.

The instructions to program the Model 480 are written in a step-by-stepformat. For convenience, these instructions begin and end at the O peratingScreen.

Caution At any time during programming or changing settings, if nocontrol panel buttons are pressed for about fifteen seconds, the display willautomatically return to the Operating Screen, storing any settings made.New settings are stored immediately when either arrow key is pressed. s

Press the button beneath temperature setpoint (Temp°C). The temperaturevalue will begin to flash.

Press the up or down arrows to set the new temperature in 0.1°Cincrements. Hold either button to scroll.

Press the temperature button twice again to r eturn to the OperatingScreen.

2-4 Orbital Shaker Thermo Scientific

Section 2Operation

Other FactoryDefaults

Function Default

Audible Alarm ON

RPM Tracking Limit (fixed) 5

Temperature Tracking Limit 10°C

Over Temperature Shut-down 63°C to 65°C

Under Temperature Shut-down -1°C to +2°C

All Remote Alarms ON

Changing Temp, Speed& Time Settings

Changing the Temperature

Page 24: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

Orbital Shaker 2-5Thermo Scientific

Section 2Operation

Changing the RPM Press the button beneath the Speed setpoint. RPM value begins to flash.

Press the up or down arrows to set the new speed in 1 RPM increments.Hold either button to scroll.

Press the Speed button again to return to the Operating Screen.

The Model 480 manages operating time in two ways:

Hold - When Time is set to Hold, the value shown in the Actual portionof the display represents total operating time and may be r eset at theoperator’s convenience. The shaker will continue to count up wards evenif the console lid has been r epeatedly opened and closed, or turned offand on with the power switch. The Time will, however, reset to 00:00when the Stop button is pressed and the unit then r estarted by pressingthe Start button.

Countdown - When the Hold setpoint is changed to Countdown, enter-ing a time value in hours and minutes programs the shaker to operate forthat period of time and automatically shut do wn. The display will showthe total time in the Setpoint segment and the operating time r emainingin the Actual part of the display, as the microprocessor counts down tozero. An additional feature of the Model 480 is Hold Temperature. Thisallows the operator to program the shaker to operate at one temperatur eand then hold the cabinet at another temperatur e when the countdownreaches zero. The platform will stop, but the fans and the temperatur econtrol systems will continue to operate.

1. Press the button beneath the Timesetpoint. Hold will begin to flash.

2. Press either arrow to access the Countdown Time setpoint. The presettime setpoint will begin to flash.

3. Press the up or down arrows to set the desired operating time in fiveminute increments. Hold either arrow to scroll in that direction.

4. When the desired elapsed time is set (6 hours, 30 minutes in thisexample), press the Time button to return to the Operating Screen.Pressing the Start button will start the shaker and begin the countdo wnsequence. As it counts down, the Actual time shown will decrease.When 00:00 is reached, the shaker will automatically shut off and theCycle Complete alarm will sound.

Changing the Time

Changing from Hold toCountdown

Page 25: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

The Model 480 Orbital Shaker control system monitors and providesalarms for nine operating parameters.

Parameter . . . . . . . . . . . . . . . . . . . . . .Alarm MessageOvertemp Setpoint Status . . . . .Overtemp ShutdownCycle Status . . . . . . . . . . . . . . . . . . .Cycle CompleteLoss of Input Power . . . . . . . . . . . . . .Power Failure*RPM versus Setpoint . . . . . . . . . . . . . .RPM TrackingDrive Belt Integrity . . . . . . . . . . . . . . . . . .Check Belt Temp Sensor Integrity . . . . . . . . . . . . . . .Sensor FaultTemp Control Status . . . . . . . . . . . .Temp High/LowPlatform Movement Status . . . . . . . .Platform StalledMotor Fuse Integrity . . . . . . . . . . . . . . . .Check Fuse

Both audible and visual alarm warnings for these nine parameters ar eprovided by the orbital shaker. Visual flashing of the three diagonalindicator lights on the control panel, a progression of text messages on thedisplay, and an audible tone aler ts the operator that an alarm condition hasoccurred or currently exists.

For convenience, the audible tone is silenced b y pressing the Silencebutton, but will ring back in about 15 minutes. H owever, the alarmwarning indicator lights and alarm messages continue until the alarmcondition is corrected by the operator. Then, pressing the Silence buttonclears the message from the display. (The Check Belt and Check Fusealarm messages clear from the display when the unit is turned back on aftercorrecting the alarm condition.)

As discussed in ‘Changing the Shaker Configuration’ in this section, theaudible alarm feature may be turned off to suit operator or laborator yneeds. Refer also to the alarms and corr ective actions chart in the Servicesection of this manual.

If power is interrupted for two hours or more while the Shaker is turnedon but not shaking, a Power Fail alarm will occur. The purpose of thealarm in this case is to aler t the user that an extended duration po werfailure occurred during the Hold interval after a timed shaking operation,or during a period of incubation only. This alarm will also occur any timethe Shaker is turned on after an extended off period gr eater than 2 hours(such as when the unit is shipped fr om the factory, or when it is r eturnedto use after a period of storage).

* The alarm will not occur if the po wer failure is less than 15 seconds induration.

2-6 Orbital Shaker Thermo Scientific

Section 2Operation

Shaker Alarms

Page 26: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

Overtemp Shutdown alerts the operatorthat the overtemp setpoint has beenexceeded by a few tenths of a degree.

The Overtemp Shutdown message will be displayed and the heaters turnedoff, but the platform and blowers continue to operate.

In the alarm state, the audible alarm is silenced b y pressing the Silencebutton, but will ring back in about 15 minutes. The screen message andwarning lights, however, will continue until the fault is corr ected. Then thealarm message is cleared by pressing the Silence button.

Undertemp Shutdown alerts the operatorthat the chamber temperature has gonebelow the Undertemp setpoint by a fewtenths of a degree.

The Undertemp Shutdown message will be displayed and the refrigerationwill be turned off, but the platform and blowers continue to operate.

In the alarm state, the audible alarm is silenced b y pressing the Silencebutton, but ring back in about 15 minutes. The screen message andwarning lights, however, continue until the fault is corr ected. Then thealarm message is cleared by pressing the Silence button.

Cycle Complete alerts the operator thatthe end of the count-down running timehas been reached.

The Cycle Complete message shown is displayed and the shaker stops.Pressing the Silence button clears the message from the display screen.

Power Failure alerts the operator thatelectrical power to the shaker has beeninterrupted, then restored during shaking.

The system will return to normal operation when power is restored,however, the alarm message and audible tone continue to aler t theoperator. Both the display and audible tone ar e cleared by pressing Silence.

Orbital Shaker 2-7Thermo Scientific

Section 2Operation

Overtemp Shutdown

Undertemp Shutdown

Cycle Complete

Power Failure

Page 27: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

RPM Tracking alerts the operator byeither alarm message shown below thatthe platform speed has varied ±5 RPM.

Note A two minute alarm delay is builtinto the software. s

In the alarm state, the audible alarm is muted b y pressing the Silencebutton, but rings back in about 15 minutes. The screen message andwarning lights, however, continue until the fault is corr ected. Then, thealarm message can be cleared by pressing the Silence button.

Check Belt alerts the operator that thedrive belt may have broken orsomething is slowing or preventing platform movement.

In the alarm state, the audible alarm is muted b y pressing the Silencebutton, but rings back in about 30 minutes. The screen message andwarning lights, however, continue until the fault is corr ected. The alarmmessage is cleared by cycling power to the unit OFF, then ON.

Sensor Fault alerts the operator that eitherof the shaker’s two temperature sensorshave failed. An alarm message similar tothose shown at right indicates whichsensor has failed.

When in the alarm state, the audible alarm is muted b y pressing theSilence button, but rings back in about 15 minutes if the fault conditionpersists. The screen message and warning lights continue until the fault iscorrected.

When the sensor problem is corrected, the alarm message is clear ed bypressing the Silence button.

2-8 Orbital Shaker Thermo Scientific

Section 2Operation

RPM Tracking

Check Belt

Sensor Fault

Page 28: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

Temperature High or Temperature Lowalerts the operator that the operatingtemperature of the shaker has risen abo veor fallen below the programmedtemperature tracking limit control point.Therefore, either of the alarm messages sho wn at the right are displayed.

In the alarm state, the audible alarm is muted b y pressing the Silencebutton, but rings back in about 15 minutes if the fault condition persists.The screen message and warning lights continue until the fault iscorrected.

When the temperature problem is corrected, the alarm message is clear edby pressing the Silence button.

Platform Stalled alerts the operator thatsomething is preventing free platformmovement. The motor automatically shutsoff and the audible alarm, scr een message and warning lights ar e initiated.The motor will attempt restart after approximately 15-20 seconds. Themotor will continue to cycle on and off until the obstr uction is removed,or the unit is turned off . On motor restart, the audible alarm and warninglights are automatically cleared. The screen message will remain untilcleared by the operator.

Check Fuse alerts the operator thatprimary drive motor fuse has blown.The audible alarm, screen message andwarning lights are initiated. When the unit is turned on after fusereplacement, all alarm indicators are automatically cleared.

Orbital Shaker 2-9Thermo Scientific

Section 2Operation

Temperature High or Low

Platform Stalled

Check Fuse

Page 29: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

To access the system Configuration menu, pr ess the down arrow, the uparrow, and the Silence button in that sequence.

This screen will appear on the display.

Pressing the down arrow continues with system configuration.

Pressing the up arrow returns to the Operating Screen.

Note During the following configuring procedures, menu options are givento either modify a setting as it appears in sequence or scr oll past to the nextitem. If no selection is made by pressing a button or arrow, the display willrevert to the Operating Screen in about fifteen seconds. The completeconfiguration menu is shown in the chart at the end of this section. s

Note In these procedures, values and settings for time, temperature, speeds,alarms, and so forth are shown on the display screens. These numbers arefor example only and may not be the v alues encountered whenprogramming your shaker. s

2-10 Orbital Shaker Thermo Scientific

Section 2Operation

Changing the ShakerConfiguration

Page 30: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

Begin by pressing the down arrow, the up arrow, andthe Silence button. The screen following will appearon the display:

When this screen opens, press the down arrow once.

At this screen (above), press the Time button beneath Audible. Thefollowing screen will appear and the curr ent setting will flash.

Press the up or down arrow to turn theaudible alarm function on or off .Pressing any of the three buttons(Time, Speed, or Temp) will return thedisplay to the previous screen. Notpressing anything for about fifteen seconds r eturns the display to theOperating Screen.

When the audible alarm is disabled, a warning message is placed in theSetpoint portion of the OperatingScreen display as illustrated at right.

Three temperature alarms are programmed into Model 480 OrbitalShaker; Overtemperature, Undertemperature, and Tracking Limits.

Orbital Shaker 2-11Thermo Scientific

Section 2Operation

Turning the Audible AlarmOn and Off

Setting Alarm Limits

Page 31: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

The Overtemperature alarm activates whenever the operating temperaturegoes above the Overtemp setpoint temperature by a few tenths of °C. Thisadjustable limit is set at the factor y at approximately64°C. To change this value, open the Configurationmenu by pressing the down arrow, up arrow, and theSilence button in the sequence shown at the right.

When this screen (below) opens....Press the down arrow once ....

And then press the Temperature button beneath Alarms.

To change the overtemperature alarm setting, press the Time buttonbeneath Over. The following screen will appear and the curr ent over-temperature alarm setting will flash.

Change the temperature setting by pressing the up or down arrow. Whenset, press the Temperature button to return to the previous screen or pressnothing for about fifteen seconds. The display will change to theOperating Screen, saving the new settings into memory.

2-12 Orbital Shaker Thermo Scientific

Section 2Operation

Setting the OvertemperatureAlarm

Page 32: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

When the overtemperature setpoint is exceeded by a few tenths of adegree, the control system will shut the heating system do wn by turningoff the heaters.

The Overtemp Shutdown warning shown above will be displayed, thewarning lights will flash and the audible warning (if not turned off ) willsound.

Pressing the Silence button will turn off the audible alarm. H owever, thewarning lights will continue to flash and the alarm message will continueto be displayed until the over-temperature condition is corrected. Theaudible warning will also sound again in about 15 minutes if theovertemperature condition persists.

When the fault is corrected, press the Silence button to clear the alarmmessage from the display.

The Undertemperature alarm activates whenever the operating temperaturegoes below the undertemp setpoint temperature by a few tenths of a °C.This adjustable limit is set at the factor y atapproximately -1°C. To change this value, open theConfiguration menu by pressing the down arrow,up arrow, and the Silence button in the sequenceshown at the right.

When this screen (below) opens....Press the down arrow once ....

And then press the temperature button beneath Alarms.

Orbital Shaker 2-13Thermo Scientific

Section 2Operation

Setting the OvertemperatureAlarm (continued)

Setting the UndertemperatureAlarm

Page 33: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

To change the undertemperature alarm setting, press the Speed buttonbeneath Under. The following screen will appear and the curr entundertemperature alarm setting will flash.

Change the temperature setting by pressing the up or down arrow. Whenset, press the Temperature button to return to the previous screen or pressnothing for about fifteen seconds. The display will change to theOperating Screen, saving the new settings into memory.

When the undertemperature setpoint is exceeded by a few tenths of adegree, the control system will shut the cooling system do wn by turningoff the refrigeration system.

The Undertemp Shutdown warning shown above will be displayed, thewarning lights will flash and the audible warning (if not turned off ) willsound.

Pressing the Silence button will turn off the audible alarm. H owever, thewarning lights continue to flash and the alarm message r emains until theundertemperature condition is corrected. The audible warning will soundagain in about 15 minutes if the under temperature condition persists.

When the fault is corrected, press the Silence button to clear the alarmmessage from the display.

2-14 Orbital Shaker Thermo Scientific

Section 2Operation

Setting the UndertemperatureAlarm (continued)

Page 34: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

The Temperature Tracking alarm activates whenever theoperating temperature goes above or below the setpointtemperature by a predetermined value. This adjustablelimit is set at the factor y as 10° above and below thetemperature setpoint.

Note The above and below limits are always the same value. s

To change this limit, open the configuration menu as in the pr evious alarmprocedures, by pressing the buttons in the sequence sho wn.

When this screen (below) opens, press the down arrow once ....

then press the temperature button beneath Alarms.

Then, from the screen below, press the temperature button beneathTracking.

The following screen will appear and the present temperature trackingalarm limit setting will flash.

Change the temperature tracking limit by pressing the up or down arrow.When set, press the Temperature button to save the setting and return tothe previous screen. If no buttons are pressed for about fifteen seconds, thedisplay will change to the Operating Screen, saving the new setting tomemory.

Orbital Shaker 2-15Thermo Scientific

Section 2Operation

Setting the Temperature AlarmTracking Limit

Page 35: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

When the chamber temperaturerises above, or falls below, thetemperature tracking limit, theappropriate message is displayed,the warning lights flash and the audible warning (if not turned off )sounds.

Pressing the Silence button turns off the audible alarm. H owever, thewarning lights continue to flash and the alarm message r emains until thehigh or low temperature condition is corrected. The audible warningsounds again in about fifteen minutes if the o ver or under temperaturetracking condition persists.

When the fault is corrected, press the Silence button to clear the alarmmessage from the display.

From the Operating screen, press the down arrow, uparrow and Silence button in that sequence (shown atthe right) to open the Configuration menu.

From the screen below, press the down arrow twice to bring up thefollowing screen.

Press the Speed button beneath RPM.

The value shown on this screen is the present speed setpoint. Using the upand down arrows, increase or decrease the platform speed until the r eadingon an independent, accurate speed measuring device matches the shakerspeed setpoint.

When finished, press the Speed button to save the setting. The display willreturn to the Calibrate - RPM Temp screen. Or, if nothing is pressed forabout fifteen seconds, the display will r evert to the Operating Screen andthe setting will be automatically sav ed to memory.

2-16 Orbital Shaker Thermo Scientific

Section 2Operation

Calibrating the RPM

Setting the Temperature AlarmTracking Limit (continued)

Page 36: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

Begin by pressing the down arrow, theup arrow, the Silence button, and thenthe down arrow twice. The screen belowwill appear on the display:

Press the Temperature button beneath Temp.

Using the up and down arrows, increase or decrease the temperature valueto match an independent, accurate temperature measuring device. Whenselected, press the Time, Speed, or Temp button to save the setting. Thedisplay will return to the Calibrate - RPM Temp screen. (Or, if nothing ispressed for about fifteen seconds, the display will r evert to the OperatingScreen and the setting will be automatically sav ed to memory.)

Any of the alarm states described per viously can alert a remote alarmmonitoring system through an internal SPDT relay connected to an RJ-11jack on the rear of the shaker cabinet. For the convenience of thelaboratory, these remote alarms can be individually turned on or off . Anyor all of the remote alarms which are set to On will activate the internalrelay.

If no buttons are pressed, the display will automatically r eturn to theOperating Screen after about fifteen seconds, saving the selection tomemory.

Note The Remote Overtemp Shutdown and Check Fuse alarms cannot bedeactivated. s

Orbital Shaker 2-17Thermo Scientific

Section 2Operation

Calibrating theTemperature

Remote Alarm System

Page 37: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

To set the remote alarms to On or Off, open the Remote AlarmConfiguration menu by pressingthe down arrow, up arrow, andsilence buttons in that sequence.Then, press the down arrowthree times until the screenbelow is showing.

Press Rmte (remote). The alarms will be shown in the sequence following.

Toggle the Cycle Complete alarm with either the up (O n) arrow or thedown (Off ) arrow. Pressing the Next (temperature) button advances thedisplay to the next alarm, saving the sho wn Cycle Complete alarm settingto memory.

Toggle the Power Failure alarm with either the up (ON) arr ow or the down(OFF) arrow. Pressing the Next (temperature) button advances the displayto the next alarm, saving the Power Failure setting to memory.

Toggle the RPM Tracking alarm with either the up arr ow (ON) or thedown arrow (OFF). Pressing the Next (temperature) button advances thedisplay to the next alarm, saving the RPM Tracking setting to memory.

2-18 Orbital Shaker Thermo Scientific

Section 2Operation

Cycle Complete

Remote Alarm System(continued)

Actual

Setpoints RunHrs Rmte

for more

Heat%

Power Failure

RPM Tracking

Page 38: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

Toggle the Check Belt alarm with either the up (on) arr ow or the down(off ) arrow. Pressing the Next (temperature) button advances the display tothe next alarm, saving the Check B elt setting to memory.

Toggle the Sensor Fault alarm with either the up (ON) arr ow or the down(OFF) arrow. Pressing the Next (temperature) button advances the displayto the next alarm, saving the on/off setting to memor y.

Toggle the Temperature High/Low alarm with either the up (ON) arr owor the down (OFF) arrow. Pressing the Return (temperature) buttonreturns the display to the previous screen.

Whether the shaker has been operated in the H old or Countdown modesand/or has been turned off and unplugged many times, the micr oprocessorcontrol system maintains a running total operating hours.

To view this information, accessthe Configuration menu bypressing the down and up arrowsand the Silence button insequence, then pressing thedown arrow three times.

Orbital Shaker 2-19Thermo Scientific

Section 2Operation

Check Belt

Sensor Fault

Temperature High or Low

Viewing the TotalOperating Hours

Page 39: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

Pressing RunHrs shows the total accumulated run hours as displayed in theillustration below. In about fifteen seconds, the display will r eturn to theOperating Screen.

Caution Heat percentages are intended for factory use only, and can behelpful in troubleshooting the heat control system. s

To view this information, pressHeat % from the previous screenor access the Configurationmenu by pressing the downarrow, the up arrow and theSilence button in that sequence, then pr ess the down arrow three times.

Press Heat %.

Main Heat % is the percentage of time that the temperature controlsystem’s heaters are turned on during a fiv e secondperiod. Example: If the heaters are being cycled onfor two seconds and off for thr ee seconds, the Heat %value is 40 percent.

Refrig Heat % is the percentage of time that the r efrigeration systemcapillary tube heater is operating during a fiv e second period.

Return to the RunHrs Rmte Heat% screen by pressing any of the threebuttons beneath the display. If no buttons are pressed, the display willautomatically return to the Operating Screen in about fifteen seconds.

2-20 Orbital Shaker Thermo Scientific

Section 2Operation

Heat %

Viewing the TotalOperating Hours (cont.)

Actual

o

Setpoints RunHrs Rmte

for more

Heat%

Actual

Speed Temp Co

Time

Setpoints RunHrs Rmte

for more

Heat%

Page 40: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

Software Version is for factory use only and will be impor tant iftroubleshooting the microprocessor programming is ever necessary.

To access this screen, pressthe down arrow, the uparrow, the Silence button,and then the down arrowbutton four more times.The screen below will appear on the display:

Press the Time button beneath SwVers and the following screen will appearshowing the Model 480 software version in the control system memory.

To return to the previous screen, press the Timebutton. To return to the Operating Screen, waitabout fifteen seconds.

Temperature Sensor Readings is for factory use only and will be impor tantif troubleshooting the microprocessor programming is ever necessary.

To access this screen, press the down arrow, the up arrow, the Silencebutton, then the downarrow button four moretimes. The screen belowappears on the display:

Press the Speed button beneath O-tempand the screen at the right will appear,showing the temperatures being measuredor read by the Overtemperature sensor.

To return to the previous screen, press the Time, Speed, or Temperaturebutton. To return to the Operating Screen, wait about fifteen seconds.

Orbital Shaker 2-21Thermo Scientific

Section 2Operation

Software Version

Actual

Setpoints

return

SwVers O-temp

Temperature SensorReadings

Overtemp Sensor

Actual

Setpoints

return

SwVers O-temp

Page 41: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

This control allows the shaker to be operated without heat or r efrigeration.

To access this screen, pressthe down arrow, the uparrow, the Silence button,then the down arrow fourmore times.

The screen shown below will appear on the display:

Press the temperature button beneath T-ctrl and the screen shown belowwill appear.

Toggle the Temperature Control with either the up (ON) arr ow or thedown (OFF) arrow.

Pressing the Time, Speed, or Temp buttons returns the screen to theprevious (SwVers O-temp T-ctrl) display. If no buttons are pressed, thedisplay will automatically return to the Operating Screen after about fifteenseconds, saving selection to memory.

Defrosting of this shaker automatically takes place about ev ery eight hours.When turned on, the system increases the cabinet temperature until itreaches the 12°C or 14°C defrost setpoint. At that time, the defrostcontinues for about three minutes, then shuts off.

2-22 Orbital Shaker Thermo Scientific

Section 2Operation

Defrost Control

Temperature Control

Actual

Setpoints

to return

SwVers O-temp T-ctrl

Page 42: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

To turn the defrost system on and off, press the down arrow, the up arrow,the Silence button, then the down arrow button five more times.

The screen shown below will appear on the display:

Press the Speed button beneath On/Off and the following screen willappear:

Toggle the Defrost Control with either the up (ON) arr ow or the down(OFF) arrow.

Pressing the Time, Speed or Temp buttons will return the screen to theprevious (Defrost on/off, Temp) display.

If no buttons are pressed, the display will automatically r eturn to theOperating Screen after about fifteen seconds, saving selection to memor y.

To toggle the defrost temperature between 12°C and 14°C, press the downarrow, the up arrow, the Silence button, then the down arrow button fivemore times.

The screen shown below will appear on the display:

Orbital Shaker 2-23Thermo Scientific

Section 2Operation

Turning the DefrostSystem On and Off

Setting the DefrostTemperature

Page 43: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

Press the temperature button beneath Temp and the following screen willappear:

Toggle between the two defrost temperatures using the up and downarrows.

Press Time, Speed, or Temp to return to the previous screen or pressnothing and the display will automatically r eturn to the Operating Screenafter about fifteen seconds, also saving the setpoint to memor y.

Another useful feature of the Model 480 Series Shaker is the ability to holdthe product at a specific temperature after operating for a pr edeterminedtime.

A typical scenario is shown below. The illustration shows the shakeroperating at 250 RPM with the cabinet temperatur e at 37°C. The 06:30time value in the lower left corner of the scr een indicates that the shaker iscounting down and will stop in 6 hours and 30 minutes.

The next illustration shows that the time has counted do wn to zero andthe shaker has stopped orbiting (z ero RPM’s). The temperature value of5.0° in the lower right corner of the scr een is the temperature at which theproduct is being held. This hold will continue at this temperatur e until theStart button is pressed or a new hold or operating temperature is entered.

2-24 Orbital Shaker Thermo Scientific

Section 2Operation

Hold TemperatureControl

Setting the DefrostTemperature (continued)

Page 44: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

To set a new hold temperature, press the temperature button twice. Thescreen below will appear. The Hold temperature will flash.

Using the up and down arrows, program a new hold temperature. Press thetemperature button to save the new setting and return to the OperatingScreen.

Orbital Shaker 2-25Thermo Scientific

Section 2Operation

Hold TemperatureControl (continued)

Page 45: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

2-26 Orbital Shaker Thermo Scientific

Page 46: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

Orbital Shaker 2-27Thermo Scientific

Page 47: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

2-28 Orbital Shaker Thermo Scientific

Page 48: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

Orbital Shaker 2-29Thermo Scientific

Page 49: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

2-30 Orbital Shaker Thermo Scientific

Page 50: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

Orbital Shaker 2-31Thermo Scientific

Page 51: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation
Page 52: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

Orbital Shaker 3-1Thermo Scientific

Section 3 Maintenance

The Model 480/481 Shaker uses a brushless DC motor and oversized,permanently lubricated bearings which require no maintenance.

The anodized brushed aluminum platform and powder-coated steel cabinetsurfaces can be cleaned with common laborator y materials. However,liquids should not be allowed to enter the shaker cabinet fr om under theplatform. All spills should be cleaned up immediately. If necessary, removethe platform. Follow ‘Installing the Platform’ procedure in Section 1 whenre-installing the platform.

The microprocessor control panel uses sealed push buttons and liquidcrystal display and may be cleaned with laborator y detergents and driedwith a soft cloth.

The air filter (9.5” x 23.625”) is located behind the grille on the fr ont ofthe cabinet. The grille is held in place b y six press-in type retainers and iseasily removed by grasping it by its edges and pulling it off .

The air filter is held in place b y four retaining springs (Figure 3-1) and iseasily removed. It may be washed in water with a mild detergent and driedbetween two lint-free towels.

Shaker Platform andCabinet

Control Panel

Cleaning/Replacingthe Cabinet Air Filter

Figure 3-1. Interior View of Front Grille with Filter

Page 53: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

3-2 Orbital Shaker Thermo Scientific

Section 3Maintenance

Servicing the DriveBelt

The motor drive pulley, the large mechanism pulley, belt and motormounting bolts will be visible, after r emoving the grille (Figure 3-2).

Loosening the three 1/4-20 bolts at the base of the driv e motor allows thebelt to be changed or tension applied to the belt. Tighten the three boltsand torque to 120 in/lbs.

To remove the pulleys from their shafts, use a 1/8” Allen wr ench to removetwo set screws from the belt groove of the larger pulley; use a 5/32” Allenwrench to remove the single set screw from the belt groove of the smallerpulley.

When replacing the pulleys, seat the larger pulley completely against theshoulder of the mechanism drive shaft. The smaller pulley, however, isinstalled with 0.300” space between it and the baseplate.

Figure 3-2. Front View of Cabinet with Grille Removed

Figure 3-3. Motor/Belt Drive System

Page 54: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

Orbital Shaker 3-3Thermo Scientific

PR

EV

EN

TIV

E M

AIN

TEN

AN

CE

Sha

kers

Your

equ

ipm

ent h

as b

een

thor

ough

ly te

sted

and

cal

ibra

ted

befo

re s

hipm

ent.

Reg

ular

pre

vent

ive

mai

nten

ance

is im

porta

nt to

kee

p yo

ur u

nit f

unct

ion-

ing

prop

erly

. The

ope

rato

r sho

uld

perf

orm

rout

ine

clea

ning

and

mai

nten

ance

on

a re

gula

r bas

is. F

or m

axim

um p

erfo

rman

ce a

nd e

ffic

ienc

y, it

is re

c-om

men

ded

that

the

unit

be c

heck

ed a

nd c

alib

rate

d pe

riodi

cally

by

a qu

alifi

ed s

ervi

ce te

chni

cian

.

The

follo

win

g is

a c

onde

nsed

list

of p

reve

ntiv

e m

aint

enan

ce re

quire

men

ts. S

ee th

e sp

ecifi

ed s

ectio

n of

the

inst

ruct

ion

man

ual f

or fu

rther

det

ails

.

We

have

qua

lifie

d se

rvic

e te

chni

cian

s, u

sing

NIS

T tra

ceab

le in

stru

men

ts, a

vaila

ble

in m

any

area

s. F

or m

ore

info

rmat

ion

on P

reve

ntiv

e M

aint

enan

ceor

Ext

ende

d W

arra

ntie

s, p

leas

e co

ntac

t us

at th

e nu

mbe

r bel

ow.

Cle

anin

g an

d ca

libra

tion

adju

stm

ent i

nter

vals

are

dep

ende

nt u

pon

use,

env

ironm

enta

l con

ditio

ns a

nd a

ccur

acy

requ

ired.

Tip

s fo

r al

l sha

kers

:• U

se o

nly

Ther

mo

stan

dard

flat

-hea

d sc

rew

s fo

r fla

sk c

lips.

• Use

onl

y Th

erm

o st

anda

rd ro

und-

head

scr

ews

for t

est t

ube

rack

s, h

olde

rs a

nd u

tility

tray

s.

•401

Mill

cree

k R

oad,

Box

649

•Mar

ietta

, Ohi

o 4

5750

USA

•740

-373

-476

3•U

SA a

nd C

anad

a 8

88-2

13-1

790

•Tel

efax

: 740

-373

-418

9 •h

ttp://

ww

w.s

ervi

ce.le

d.m

arie

tta@

ther

mof

ishe

r.com

Page 55: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

3-4 Orbital Shaker Thermo Scientific

Prev

entiv

e M

aint

enan

ce fo

r M

odel

s 48

0 an

d 48

1 Sh

aker

s

Ref

er to

Man

ual S

ectio

nA

ctio

nD

aily

Mon

thly

Year

ly

--C

lean

the

unit

with

mild

det

erge

nt a

nd w

ipe

dry

as n

eede

d.þ

--C

lean

the

win

dow

with

a m

ild d

eter

gent

and

wip

e dr

y.þ

--C

heck

und

er th

e pl

atfo

rm fo

r bro

ken

glas

s or

oth

er d

ebris

--In

spec

t and

/or c

lean

the

cond

ense

r.þ

4.1c

Insp

ect a

ir fil

ter.

Cle

an a

s ne

eded

3.8

* C

heck

and

doc

umen

t cal

ibra

tion

of te

mpe

ratu

re, a

larm

s, s

peed

and

tim

e, a

s ap

plic

able

--*

Verif

y op

erat

ion

of c

ircul

atio

n fa

n m

otor

5.2

Cha

nge

the

HEP

A fi

lter a

s ne

eded

* Q

ualif

ied

serv

ice

tech

nici

ans

only

Page 56: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

Orbital Shaker 4-1Thermo Scientific

Section 4 Service

Warning The procedures outlined in this section should be carried outonly by qualified service personnel or people trained and cer tified inelectrical/ mechanical repair of laboratory equipment. s

With the exception of the HEPA filter, the Model 480/481 Orbital Shakercontains no user-serviceable components. The following table lists displaymessages which may help diagnose abnormal conditions.

If the microprocessor control system senses a fault, malfunction orabnormal operating condition, alarm messages will appear on the liquidcrystal display. These messages will be helpful should ser vice or repairassistance be necessary. Refer to the table below and to the alarm matrixfollowing in this section.

Table 4-1. Alarm Messages and Fault Conditions

Alarms and AlarmConditions

Alarm Message Fault Condition

Overtemp Shutdown System shutdown due to overtemp condition

Main Temp Sensor Temperature sensor has failed

Over Temp Sensor Temperature sensor has failed

Temperature is High Temp tracking has sensed higher temperature than setting

Temperature is Low Temp tracking has sensed lower temperature than setting

RPM High RPM tracking has sensed shaker speed higher than setting

RPM Low RPM tracking has sensed shaker speed lower than setting

Power Failure Power has failed during shaker operation

Cycle Complete End of countdown cycle has been reached

Check Belt Motor V-belt has broken or slipped

Audible is Disabled! Continuously notifies operator that audible alarm has been disabled

Platform Stalled Free movement of the platform is obstructed

Check Fuse Primary drive motor fuse has blown

Page 57: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

4-2 Orbital Shaker Thermo Scientific

Section 4Service

Changing the HEPAFilter

The HEPA filter (P/N 760164) is located on the left side of the chamberand is accessed by pulling up on the four, black press-in fasteners andsliding the cover off. Refer to the illustration below.

Figure 4-1. Lid Open Showing Locations

If the shaker platform will not operate with the unit plugged in and thepower switch turned on, the following conditions may be present:

• The lid may be open - Lo wer the lid to its fully closed position.

• Time countdown reached - Reset the time or change to a pr e-setprogram.

Three fuses are located on the left side of the r elay tray located in the lowerpart of the console cabinet. Figure 4-2 shows the location of the fuses.

Access to the relay tray is made by removing the grille on the front of thecabinet. It is held in place by six press-in type retainers and is easilyremoved by grasping the edges of the grille and pulling.

To remove the panel beneath the grille, r emove six Phillips screws; three onthe bottom of the panel and thr ee on the top. The two Phillips screws onthe left side of the foot pedal will need to be loosened to allo w the panelto be slid outward.

If the Shaker Will NotOperate

Fuses

Page 58: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

Orbital Shaker 4-3Thermo Scientific

Section 4Service

Table 5-2. Fuses

Caution Do not substitute! Replace these fuses with fuses of identicalelectrical ratings only. s

Figure 5-2. Fuse and Connector Locations

Warning Never use the recorder plug-in outlet with the groundingconductor disconnected. s

Fuses, Model 480

Rating Application Part Number

0.25 amp Main Power Relay Board 230144

1.6 amp Drive Motor 230145

0.1 amp Recorder (opt.) 30107

Fuses, Model 481

Rating Application Part Number

0.15 amp Main Power Relay Board 230142

0.8 amp Drive Motor 230141

0.1 amp Optional Recorder 230107

Fuses (continued)

Page 59: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

Five circuit boards control the Orbital Console Shaker. Four boards arelocated in the relay tray compartment, the fifth is behind the liquid cr ystaldisplay. Figure 4-3 identifies the circuit boards and other majorcomponents in the relay tray.

Refer also to the relay tray illustrations on Page 4-7 & 4-8.

Figure 4-3. Electrical Component Locations, Relay Tray Compartment

4-4 Orbital Shaker Thermo Scientific

Section 4Service

Circuit Boards

Page 60: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

Two temperature sensors are located behind a per forated cover plate on theright side of the chamber (r efer to Figure 4-4). To access these sensors,pull outward on the two black press-in fasteners on the top edge of thecover and lift the cover upward. The cover is held in place by four metalclips.

Figure 4-4. Right Side Chamber - Temperature Sensor Cover Plate Removed

To replace the cover, make sure all four clips engagethe metal edges of the chamber and the two fastenersare firmly seated in their holes. P ress the top of thefastener inward until a “click” is heard. (Refer toFigure 4-5.) Figure 4-5. Press-in

Sheet Metal Fastener

Locations of the blower motor and ambient fan motor ar e shown in Figure4-6 (refrigeration unit removed).

Removal of the relay tray is necessary to service the blower motor or theheating elements.

Figure 5-6. Air Plenum with Heating and Air Moving Components

Orbital Shaker 4-5Thermo Scientific

Section 4Service

Temperature Sensors

Blower Fan Motorsand Heating Elements

Page 61: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

A manual reset circuit breaker is located between the heating elementelectrical connectors on the side of the air plenum. (F igures 4-6 and 4-7)To access this circuit breaker, remove the front grille and remove the fourPhillips screws securing the relay tray. Move or tip the relay tray farenough away from the air plenum to r each the reset button.

Warning Disconnect and lock-out electrical power when working on ornear the relay control tray and heating element connectors. Allo w sufficienttime for the heating elements to cool befor e reaching into that area. s

After the console cabinet is in place, lev eled and the platform installed,turn the unit on and set the RPM to 300. Kneeling in fr ont of theconsole, lightly touch the lower left and right corners of the cabinet. I f oneside seems to vibrate more then the other, raise or lower the corner supportleg using the 3/4” open end wr ench supplied in the par ts bag. Continuethis “fine tuning” until the vibrations are lowered as far as possible.

4-6 Orbital Shaker Thermo Scientific

Section 4Service

Heater ElementCircuit Breaker

Figure 4-7. Heating Element Connectors and Manual Reset Circuit Breaker

Tuning the Cabinet

Page 62: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation
Page 63: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation
Page 64: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation
Page 65: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation
Page 66: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

Orbital Shaker 5-1Thermo Scientific

Section 5 Specifications

ShakingRange . . . . . . . . . . . . . . . . . . . . . . . . .25-525 RPMAccuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 RPMMotion . . . . . . . . . . . . . . . . . . . . .One inch/orbitalIndicator . . . . . . . . . . .LCD in 1 RPM increments

TemperatureRange . . . . . . . . . . . . .4°C (39°F) to 60°C (140°F)Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0.1°CUniformity . . . . . . . . . . . . . . . . . .±0.2°C (in flask)Indicator . . . . . . . . . . . .LCD, in 0.1°C increments

TimerPeriods . . . Programmable from 5 minutes to 200 hours or continuousoperationIndicator . . . . . . . . . .LCD in 5 minute incrementsRun Time. . . Display counts down for a timed run or counts up whenin “hold” function

AlarmsTemperature . .Adjustable tracking high/low tempsRPM . . . . . . . . .Adjustable tracking high/low RPMTime . . . . . . . . . . . . . . . . . . . . . . . . .Run completePower Fail . . . . .Message displayed on LCD screen

SafetyTemperature . . .Independent over and under tempRPM . . . . . . . . . . . .Independent platform motion

Page 67: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

5-2 Orbital Shaker Thermo Scientific

Section 5Specifications

LCD DisplayRun Time, RPM, Temperature, User Program, Alarm Conditions and Power Failure indicated by messages

Mechanical Drive SystemTriple counterbalanced. Handles unbalanced platform loads regardless offlask placement

Drive Motor1/3 HP brushless DC, permanently lubricated ball bearing

LidCounterbalanced, hand or foot operated, with temper ed thermal panewindow and key lock.

Automatic RestartMicroprocessor retains all programming in non-volatile memory. In theevent of a power outage, the shaker restarts automatically.

ConstructionInterior . . . . . . . .Stainless steel with coved cornersExterior . . . . . . . . . . . . . . . . . . . . .Cold rolled steelFinish. . .Powder coated for a durable, easily maintained sur facePlatform . . . . . . . . . .Anodized brushed aluminum

DimensionsExterior . . . . . . . . .45.0” W x 39.0” H x 30.0” F-B

. . . . . . . . . . . . . . .(114.3cm x 99.1cm x 76.2cm)Exterior (lid open). . 45.0” W x 39.0” H x 30.5” F-B

. . . . . . . . . . . . . . .(114.3cm x 99.1cm x 77.5cm)Interior . . . . . . . . .34.3” W x 18.8” H x 21.1” F-B

. . . . . . . . . . . . . . . .(87.1cm x 47.8cm x 53.6cm)

Page 68: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

Electrical480 . .Nominal 120 VAC, 60 Hz, 1 PH, 15.6 FLA

. . . . . . . . . . . . .Operating Range: 108-125 VAC481 . . .Nominal 230 VAC, 50 Hz, 1 PH, 8.0 FLA

. . . . . . . . . . . . .Operating Range: 207-253 VACData Output . . . . . . . . . . . . . . . . .RS-232 standardRemote Alarm Contacts . . .Time, RPM, Temperature and Loss of PowerAlarms

Certifications480: . . . . . . . . . . . . . .UL Standard No. 61010A-1

CSA . . . . . . . . . . . .Standard C22.2 No. 1010.1

481: . . . . . . CE Mark Electromagnetic and Low Voltage Directives

CapacityFlasks . . . . . . . . . . . .From (91) 25ml up to (4) 6L

WeightsNet . . . . . . . . . . . . . . . . . . . . . .568 lbs. (257.9kg)Shipping . . . . . . . . . . . . . . . . . . .645 lbs. (292kg)

Optional PlatformsSize . . . . . . . . . . . .29.5” x 18” (74.9cm x 45.7cm)Clips . . 25ml, 50ml, 125ml, 250/300ml, 500ml, 1L, 2L, 4L, 5L, 6L,and 2800ml Fernbach sizes availableRacks . . Adjustable angle test tube holder with rack, 10-30mm

FilterHEPA . . . . . . . . . . . .Rated efficient at 0.3 micronsSize . . .18” x 16” x 2” (45.7cm x 40.6cm x 5.1cm)

Orbital Shaker 5-3Thermo Scientific

Section 5Specifications

Page 69: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

Ambient Operating ConditionsIndoor use onlyTemperature . . . . . . . 18°C (64°F) to 40°C (104°F)Humidity . .80% RH at or below 31°C, decreasing linearly to 50% RHat 40°C

Sound Level - Not to exceed 85db

Safety SpecificationsAltitude . . . . . . . . . . . . . . . . . . . . . .2,000 metersTemperature . . . . . . . . . . . . . . . . . . .5°C to 40°CHumidity . . 80% RH at or belo w 31°C, decreasing linearly to 50% RHat 40°C

Mains Supply Fluctuations Operating Voltage RangeInstallation Category II 1

Pollution Degree 2 2

Class of Equipment IClimatic Condition - ST (EN 60335, S ubtropical)

_________________________________________

1 Installation category (overvoltage category) defines the level of transient over-voltage which the instrument is designed to withstand safely. It depends onthe nature of the electricity supply and its o vervoltage protection means. Forexample, in CAT II which is the categor y used for instruments in installa-tions supplied from a supply comparable to public mains such as hospital andresearch laboratories and most industrial laboratories, the expected transientovervoltage is 2500V for a 230V supply and 1500V for a 120V supply .

2 Pollution Degree describes the amount of conductiv e pollution present in theoperating environment. Pollution Degree 2 assumes that normally only non-conductive pollution such as dust occurs with the ex ception of occasional con-ductivity caused by condensation.

5-4 Orbital Shaker Thermo Scientific

Section 5Specifications

Page 70: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

Section 6 Parts List

Model 480Part No. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Description

129024 . . . . . . . . . . . . . . . . . . . .(2) Pneumatic Spring, 80 lbs.138009 . . . . . . . . . . . . . .Heater, Wirewound 450W 115/230V156089 . . . . . . . . . . . . . . . . . . . . . .Motor, Brushless - DC 24V 191535 . . . . . . . . . . . . . . . . . . . . . . . . . . . .Motor Drive Board 191545 . . . . . . . . . . . . . . . . . . . . . . . . . . .Temp Control Board 190525 . . . . . . . . . . . . . . . . . . . . . . .Triple Outlet Relay Board 190818 . . . . . . . . . . . . . . . . . . . . . . . . . . . . LCD Display Panel 191538 . . . . . . . . . . . . . . . . . . . . . . . . . . . .Shaker Micro Board 400113 . . . . . . . . . . . . . . . . . . . .Thermostat - Opens at 200°F420064 . . . . . . . . . . . . . . . . . . . . . . . . . . .Transformer, 130VA 420085 . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Transformer, 25VA 600600 . . . . . . . . . . . . . . . . . . .Cartridge Heater, 180W 120V900092 . . . . . . . . . . . . . . . . . . . .230 CFM Blower 115V 60Hz900105 . . . . . . . . . . . . . . . . . . .Tubeaxial Fan, 235CFM, 115V991003* . . . . . . . . . . . .Dryer 1/4 SAE Non-CFC, Unit Size 8 212036* . . . . . . . . . . . . . . . . . . . . .Evaporator 8-3/4 x 11-3/8 205142* . . . . . . . . .1/3 HP Condensing Unit 115V w/R134A 220309* . . . . . . . . . . . . . . . . . . . . . .Access Valve Assembly, 2” 250111* . . . . . . . . . . . . . . . .1/4 ODF NC Ref. Solenoid Valve 300275 . . . . . . . . . . . . . . . . . . . . . . . .20A Relay, DPDT, 120V 230107 . . 100 mA Fuse (for optional recorder) T.D. 5mm x 20mm 230144 . . . . . . . . . . . . . . . . .250 mA Fuse T.D. 5mm x 20mm 230145 . . . . . . . . . . . . . . . . . . .1.6 A Fuse T.D. 5mm x 20mm 800040 . . . . . . . . . . . . . . . . . . . . . . . . . . . .V-Belt, 1/2” x 45” 760167 . . . . . . . . . . . . . . . . . . . . . .Air Filter 10.5” x 23.625” 760164 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .HEPA Filter 194015 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Shipping Bracket 443021 . . . . . . . . . . . . . . . . . . . . . . . .3/4” Open End Wrench 435051 . . . . . . . . . . . . . . . . . . . . . . . . . . . .Screwdriver, Phillips 443020 . . . . . . . . . . . . . . . .Wrench, 5/32” Hex with T-handle 194046 . . . . . . . . . . . . . . .Spare Parts Bag, (platform and clips) 270127 . . . . . . . . . . . . . . . . . . . . .EMI Power Line Filter, 20A 107003 . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Lid Glass Window 990024 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Lid Gasket 194580 . . . . . . . . . .Condensing Unit Evaporator Assy 1/3 HP * part of P/N 194580

Orbital Shaker 6-1Thermo Scientific

Page 71: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

Model 481

Part No. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Description

129024 . . . . . . . . . . . . . . . . . . . .(2) Pneumatic Spring, 80 lbs. 138009 . . . . . . . . . . . . . .Heater, Wirewound 450W 115/230V 156089 . . . . . . . . . . . . . . . . . . . . . .Motor, Brushless DC 24V 191535 . . . . . . . . . . . . . . . . . . . . . . . . . . . .Motor Drive Board 191545 . . . . . . . . . . . . . . . . . . . . . . . . . . .Temp Control Board 190525 . . . . . . . . . . . . . . . . . . . . . . .Triple Outlet Relay Board 190818 . . . . . . . . . . . . . . . . . . . . . . . . . . . .LCD Display Panel 191538 . . . . . . . . . . . . . . . . . . . . . . . . . . . .Shaker Micro Board 400113 . . . . . . . . . . . . . . . . . . . .Thermostat - Opens at 200°F 420064 . . . . . . . . . . . . . . . . . . . . . . . . . . .Transformer, 130VA 420085 . . . . . . . . . . . . . . . . . . . . . . . . . . . .Transformer, 25VA 133009 . . . . . . . . . . . . . . . . . . .Cartridge Heater 200W, 240V 900093 . . . . . . . . . . . . . . . . . . . . . . . . .230 CFM Blower 220V 900107 . . . . . . . . . . . . . . . . . .Tubeaxial Fan, 235 CFM, 230V 991003* . . . . . . . . . . . . .Dryer 1/4 SAE Non-CFC Unit Size 8 212036* . . . . . . . . . . . . . . . . . . . . . .Evaporator 8-3/4 x 11-3/8 205079* . . . . . . . . .1/3 HP Condensing Unit 220V, w/R134A 220309* . . . . . . . . . . . . . . . . . . . . . .Access Valve Assembly, 2” 250115* . . . . . . . . . . . . . .1/4 ODF Ref. Solenoid Valve, 220V 300276 . . . . . . . . . . . . . . . . . . . . . . . .20A Relay, DPDT, 220V 230141 . . . . . . . . . . . . . . . . .800 mA Fuse T.D. 5mm x 20mm 230142 . . . . . . . . . . . . . . . . .150 mA Fuse T.D. 5mm x 20mm 230107 . . 100 mA Fuse (for optional recorder) T.D. 5mm x 20mm 800040 . . . . . . . . . . . . . . . . . . . . . . . . . . . .V-Belt, 1/2” x 45” 760167 . . . . . . . . . . . . . . . . . . . . . .Air Filter 10.5” x 23.625” 760164 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .HEPA Filter 194015 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Shipping Bracket 443021 . . . . . . . . . . . . . . . . . . . . . . . . . .3/4” open end wrench 435051 . . . . . . . . . . . . . . . . . . . . . . . . . . .Screwdriver, Phillips 443020 . . . . . . . . . . . . . . . .Wrench, 5/32” Hex with T-handle 194046 . . . . . . . . . . . . . . .Spare Parts Bag, (platform and clips) 270126 . . . . . . . . . . . . . . . . . . . . .EMI Power Line Filter, 10A 107003 . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Lid Glass Window 990024 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Lid Gasket194582 . . . . . . . . . .Condensing Unit Evaporator Assy 1/3 HP * part of P/N 194582

6-2 Orbital Shaker Thermo Scientific

Section 6Parts List

Page 72: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

Orbital Shaker 7-1Thermo Scientific

Page 73: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

7-2 Orbital Shaker Thermo Scientific

Page 74: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

Orbital Shaker 8-1Thermo Scientific

Page 75: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

8-2 Orbital Shaker Thermo Scientific

Page 76: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

Orbital Shaker 8-3Thermo Scientific

Page 77: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation
Page 78: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

Orbital Shaker 8-4Thermo Scientific

Page 79: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

8-5 Orbital Shaker Thermo Scientific

Page 80: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

Orbital Shaker 8-6Thermo Scientific

Page 81: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation
Page 82: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

Orbital Shaker 8-7Thermo Scientific

Page 83: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

8-8 Orbital Shaker Thermo Scientific

Page 84: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

TH

ER

MO

FIS

HE

R S

CIE

NT

IFIC

OR

BIT

AL

SH

AK

ER

WA

RR

AN

TY

US

A

Th

e W

arr

an

ty P

erio

d s

tart

s t

wo

we

eks f

rom

th

e d

ate

yo

ur

eq

uip

me

nt

is s

hip

pe

d f

rom

ou

r fa

cili

ty.

Th

is a

llow

s s

hip

pin

g t

ime

so

th

e w

arr

an

ty w

ill g

o in

to e

ffe

ct a

t a

pp

roxim

ate

ly th

e s

am

e tim

e y

ou

r e

qu

ipm

en

t is

de

live

red

. T

he

wa

rra

nty

pro

tectio

n e

xte

nd

s

to a

ny s

ub

se

qu

en

t o

wn

er

du

rin

g t

he

wa

rra

nty

pe

rio

d.

Du

rin

g t

he

first

two

ye

ars

or

12

,00

0 h

ou

rs o

f u

se

, w

hic

he

ve

r co

me

s l

ate

r, c

om

po

ne

nt

pa

rts p

rove

n t

o b

e n

on

-co

nfo

rmin

g i

n

ma

teria

ls o

r w

ork

ma

nsh

ip w

ill b

e r

epa

ire

d o

r re

pla

ce

d a

t T

he

rmo

’s e

xp

en

se

, la

bo

r in

clu

de

d.

In

ad

ditio

n,

afte

r th

e in

itia

l w

ar-

ran

ty p

erio

d o

f tw

o y

ea

rs o

r 1

2,0

00

ho

urs

, th

e O

rbita

l S

ha

ke

r m

ech

an

ism

is w

arr

an

ted

fo

r U

nit P

rod

uctio

n L

ife

, pa

rts o

nly

,

F.O

.B. fa

cto

ry. T

he

me

ch

an

ism

is d

efin

ed

as th

e b

ea

rin

g a

sse

mb

lies. U

nit P

rod

uctio

n L

ife

is d

efin

ed

as th

e p

erio

d o

f tim

e th

e

cu

rre

nt m

ech

an

ism

de

sig

n is in

pro

du

ctio

n, p

lus s

eve

n y

ea

rs. T

he

wa

rra

nty

will

be

vo

id if th

e e

qu

ipm

en

t is

alte

red

with

ou

t w

rit-

ten

au

tho

riza

tio

n fro

m T

he

rmo

. T

he

12

,00

0 h

ou

rs u

se

mu

st b

e v

alid

ate

d b

y th

e in

tern

al tim

er

an

d/o

r o

bje

ctive

do

cu

me

nta

tio

n.

If v

alid

atio

n o

r d

ocu

me

nta

tio

n o

f ru

n h

ou

rs a

re n

ot

ava

ilab

le,

un

it w

arr

an

ty d

efa

ults t

o 3

ye

ars

, pa

rts a

nd

la

bo

r. I

nsta

llatio

n a

nd

ca

libra

tio

n i

s n

ot

co

ve

red

by t

his

wa

rra

nty

ag

ree

me

nt.

T

he

Te

ch

nic

al

Se

rvic

es D

epa

rtm

en

t m

ust

be

co

nta

cte

d f

or

wa

rra

nty

de

term

ina

tio

n a

nd

dire

ctio

n p

rio

r to

pe

rfo

rma

nce

of

an

y r

epa

irs.

Exp

en

da

ble

ite

ms,

i.e

., g

lass,

filte

rs,

ligh

t b

ulb

s a

nd

lid

ga

s-

ke

ts a

re e

xclu

de

d f

rom

th

is w

arr

an

ty.

Re

pla

ce

me

nt o

r re

pa

ir o

f co

mp

on

en

ts p

art

s o

r e

qu

ipm

en

t u

nd

er

this

wa

rra

nty

sh

all

no

t e

xce

ed

th

e w

arr

an

ty to

eith

er

the

eq

uip

-

me

nt

or

to t

he

co

mp

on

en

t pa

rt b

eyo

nd

th

e o

rig

ina

l w

arr

an

ty p

erio

d.

Th

e T

ech

nic

al

Se

rvic

es D

epa

rtm

en

t m

ust

giv

e p

rio

r

ap

pro

va

l fo

r re

turn

of

an

y c

om

po

ne

nts

or

eq

uip

me

nt.

TH

IS W

AR

RA

NT

Y IS

E

XC

LU

SIV

E A

ND

IN

L

IEU

O

F A

LL

O

TH

ER

W

AR

RA

NT

IES

, W

HE

TH

ER

W

RIT

TE

N,

OR

AL

, O

R

IMP

LIE

D.

NO

WA

RR

AN

TIE

S O

F M

ER

CH

AN

TA

BIL

ITY

OR

FIT

NE

SS

FO

R A

PA

RT

ICU

LA

R P

UR

PO

SE

SH

AL

L A

PP

LY

.

Th

erm

o s

ha

ll n

ot b

e lia

ble

fo

r a

ny in

dire

ct o

r co

nse

qu

en

tia

l d

am

ag

es in

clu

din

g, w

ith

ou

t lim

ita

tio

n, d

am

ag

es r

ela

tin

g to

lo

st p

rof-

its o

r lo

ss o

f p

rod

ucts

.

Yo

ur

loca

l Th

erm

o S

ale

s O

ffic

e is

re

ad

y to

he

lp w

ith

co

mp

reh

en

siv

e s

ite

pre

pa

ratio

n in

form

atio

n b

efo

re y

ou

r e

qu

ipm

en

t a

rriv

es.

Prin

ted

in

str

uctio

n m

an

ua

ls c

are

fully

de

tail

eq

uip

me

nt

insta

llatio

n,

op

era

tio

n,

an

d p

reve

ntive

ma

inte

na

nce

.

If e

qu

ipm

en

t se

rvic

e is r

eq

uire

d,

ple

ase

ca

ll yo

ur

Te

ch

nic

al S

erv

ice

s D

epa

rtm

en

t a

t 1

-80

0-4

38

-48

51

(U

SA

an

d C

an

ad

a)

or

1-

74

0-3

73

-47

63

. W

e’re

re

ad

y t

o a

nsw

er

an

y q

ue

stio

ns o

n e

qu

ipm

en

t w

arr

an

ty,

op

era

tio

n,

ma

inte

na

nce

, se

rvic

e a

nd

sp

ecia

l

ap

plic

atio

ns.

Ou

tsid

e t

he

US

A,

co

nta

ct

yo

ur

loca

l d

istr

ibu

tor

for

wa

rra

nty

in

form

atio

n.

Rev.

3

4/0

9

ISO

9001

REGI

STER

ED

Page 85: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

THER

MO F

ISHE

R SC

IENT

IFIC

INTE

RNAT

IONA

L OR

BITA

L SH

AKER

WAR

RANT

YT

he W

arr

anty

Period s

tart

s t

wo m

onth

s f

rom

the d

ate

your

equip

ment

is s

hip

ped f

rom

our

facility.

This

allow

s s

hip

pin

g t

ime s

o t

he

warr

anty

will go into

effect at appro

xim

ate

ly the s

am

e tim

e y

our

equip

ment is

delivere

d. T

he w

arr

anty

pro

tection e

xte

nds to a

ny s

ub-

sequent

ow

ner

during t

he w

arr

anty

period.

During t

he f

irst

two y

ears

or

12,0

00 h

ours

of

use,

whic

hever

com

es late

r, c

om

ponent

part

s p

roven t

o b

e n

on-c

onfo

rmin

g in m

ate

ri-

als

or

work

manship

will be r

epaired o

r re

pla

ced a

t T

herm

o’s

expense,

exclu

din

g labor.

The o

rbital shaker

mechanis

m is w

arr

ante

d

for

life

, part

s o

nly

, F.O

.B.

facto

ry.

The m

echanis

m i

s d

efined a

s t

he b

earing a

ssem

blies.

Life i

s d

efined a

s t

he p

eriod o

f tim

e t

he

curr

ent

mechanis

m d

esig

n is in p

roduction,

plu

s s

even y

ears

. T

he w

arr

anty

will be v

oid

if

the e

quip

ment

is a

ltere

d w

ithout

the

writ-

ten a

uth

orization fro

m T

herm

o. T

he 1

2,0

00 h

our

use m

ust be v

alidate

d b

y the inte

rnal tim

er

and/o

r obje

ctive d

ocum

enta

tion. If

val-

idation o

r docum

enta

tion o

f ru

n h

ours

are

not

available

, unit w

arr

anty

defa

ults t

o 3

years

, part

s o

nly

. I

nsta

llation a

nd c

alib

ration is

not covere

d b

y this

warr

anty

agre

em

ent. E

xpendable

ite

ms, i.e., g

lass, filters

, light bulb

s a

nd lid

gaskets

are

exclu

ded fro

m this

war-

ranty

.

Repla

cem

ent

or

repair o

f com

ponent

part

s o

r equip

ment

under

this

warr

anty

shall n

ot

exceed t

he w

arr

anty

to e

ither

the e

quip

ment

or

to the c

om

ponent part

beyond the o

rigin

al w

arr

anty

period. T

he T

echnic

al S

erv

ices D

epart

ment m

ust giv

e p

rior

appro

val fo

r re

turn

of

any c

om

ponents

or

equip

ment.

TH

IS W

AR

RA

NT

Y I

S E

XC

LU

SIV

E A

ND

IN

LIE

U O

F A

LL

OT

HE

R W

AR

RA

NT

IES

, W

HE

TH

ER

WR

ITT

EN

, O

RA

L,

OR

IM

PL

IED

.

NO

WA

RR

AN

TIE

S O

F M

ER

CH

AN

TA

BIL

ITY

OR

FIT

NE

SS

FO

R A

PA

RT

ICU

LA

R P

UR

PO

SE

SH

AL

L A

PP

LY.

Therm

o s

hall n

ot be

liable

for

any indirect

or

consequential dam

ages inclu

din

g,

without

lim

itation,

dam

ages r

ela

ting t

o lost

pro

fits

or

loss o

f pro

ducts

.

Therm

o I

nte

rnational S

ale

s O

ffic

e is r

eady t

o h

elp

with c

om

pre

hensiv

e s

ite p

repara

tion info

rmation b

efo

re y

our

equip

ment

arr

ive

s.

Printe

d instr

uction m

anuals

care

fully d

eta

il e

quip

ment

insta

llation,

opera

tion,

and p

reventa

tive m

ain

tenance.

If equip

ment

serv

ice is

re

quired,

ple

ase call your

Technic

al

Serv

ices D

epart

ment

at

740-3

73-4

763,

or

Fax 740-3

73-4

189.

W

e’re

ready to a

nsw

er

your

questions o

n e

quip

ment w

arr

anty

, opera

tion, m

ain

tenance, serv

ice a

nd s

pecia

l applications. O

uts

ide the U

SA

,

conta

ct

your

local dis

trib

uto

r fo

r w

arr

anty

info

rmation.

ISO

9001

REGI

STER

EDR

ev.

2

2/0

7

Page 86: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation

Manufacturer'S Name: Thermo Fisher Scientific (Asheville) LLC

Manufacturer'S Address: 401 Millcreek Road Marietta, Ohio 45750 U.S.A.

Product Description: Forma® Orbital Shaker

Product Designations: 481

Year of Initial Marking (CE): 1998

Affected Serial Numbers: Release 11 (Release Level [REL#] shown on Serial Tag)

This product conforms to the following European Union Directive(s):

EMC: 89/336/EEC LVD: 73/23/EEC

This product conforms to the following Harmonized, International and National Standards:

EMC: LVD: EN 61326-1 :1997 EN 61010-1:1993

EN 50081-1 :92 IEC 1010-1 Amendment 2 EN 50082-1 :97 EN 60335-2-24 (applicable sections)

CSAC22.2 No. 1010.1-92

Richard L. Miller, CQE Regulatory Compliance Manager

ThermoFisher SCIENTIFIC

05 February 2007

Rev. 4

Page 87: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation
Page 88: Model 480 and 481 - Thermo Fisher Scientific.pdfInternet and we can be contacted through our Internet home page. Our staff can supply technical information about proper setup, operation