Micro Motion® Menu Maps for HART Device Descriptions

Configuration and Use Manual Supplement

Instruction ManualP/N MMI-20017185, Rev. ABApril 2011

(艾默生Micro Motion的质量流量计)(仪器仪表网:HART375手操器专题)

© 2011 Micro Motion, Inc. All rights reserved. The Micro Motion and Emerson logos are trademarks and service marks of Emerson Electric Co. Micro Motion, ELITE, MVD, ProLink, MVD Direct Connect, and PlantWeb are marks of one of the Emerson Process Management family of companies. All other trademarks are property of their respective owners.

About this supplementThe Field Communicator uses HART device descriptions to communicate with Micro Motion transmitters. Micro Motion is updating its device descriptions to the new “dashboard” style. The latest dashboards are available as Device Installation Kits on the Emerson web site.

This supplement provides menu maps for the dashboards. These menu maps replace the Field Communicator menu maps in your transmitter’s Configuration and Use Manual.

Configuration and Use Manual Supplement i

Contents

Chapter 1 RFT9739 Transmitters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

Chapter 2 9739 MVD Transmitters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

Chapter 3 Model 1500 Transmitters with Analog Outputs . . . . . . . . . . . . . . . . 27

Chapter 4 Model 1700 Transmitters with Analog Outputs . . . . . . . . . . . . . . . . 41

Chapter 5 Model 1700 Transmitters with Intrinsically Safe Outputs . . . . . . . . . 55

Chapter 6 Model 2200S Transmitters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69

Chapter 7 Model 2400S Transmitters with Analog Outputs . . . . . . . . . . . . . . . 81

Chapter 8 Model 2500 Transmitters with Configurable Input/Outputs . . . . . . . . 95

Chapter 9 Model 2700 Transmitters with Analog Outputs . . . . . . . . . . . . . . . 109

Chapter 10 Model 2700 Transmitters with Intrinsically Safe Outputs . . . . . . . . 125

Chapter 11 Model 2700 Transmitters with Configurable Input/Outputs . . . . . . . 141

Configuration and Use Manual Supplement 1

9739 MV

DM

od

el 1700 AN

Mo

del 1500 A

NR

FT

9739

Chapter 1RFT9739 Transmitters

Figure 1-1 On-Line menu

Configure

1 Manual Setup2 Alert Setup

Service Tools

1 Alerts2 Variables3 Trends4 Maintenance5 Simulate

On-Line Menu

2 3

Overview

1 Check Status2 Primary Purpose Variables3 Shortcuts

1

2 Micro Motion Menu Maps for HART Device Descriptions

RFT9739 Transmitters

Figure 1-2 Overview menu

Check Status

1 Refresh Alerts2 Dev Status:3 Comm Status:

Primary Purpose Variables

Mass Flow RateVolume Flow RateDensity

Shortcuts

1 Device Information2 Totalizer Control3 Zero Calibration4 Variables5 Trends

1 32

Device Information

1 Product Information2 Mat. of Construction3 Security Events

1

Product Information

1 Tag2 Model3 Xmtr Software Rev4 Final Asmbly Num5 Sensor Serial Num6 HART DD Information

Security Events

1 Number of Configuration Writes2 Number of Calibration Parameters

3

On-Line Menu >1 Overview

1

Configuration and Use Manual Supplement 3

RFT9739 Transmitters

9739 MV

DM

od

el 1700 AN

Mo

del 1500 A

NR

FT

9739

Figure 1-3 Configure menu: Top level

Figure 1-4 Configure menu: Manual Setup

Manual Setup

1 Characterize2 Measurements3 Inputs/Outputs4 Info Parameters5 Communications

21

On-Line Menu >1 Configure

Alert Setup

1 DO Fault Action2 Control Output3 Slug Duration4 Set Up Events

On-Line Menu >2 Configure >

1 Manual Setup

Characterize

1 Flow Cal2 D13 D24 TC5 K16 K27 FD

Measurements

1 Flow2 Density3 Temperature4 Set Up Special Units5 Set Up External Compensation

1 2

Inputs/Outputs

1 Set Up mA Output 12 Set Up mA Output 23 Set Up Frequency Output4 Map Variables

3

Info Parameters

1 Transmitter Info2 Sensor Information

Communications

1 Polling Address2 Tag3 Device Identification4 Comm Status:

54

4 Micro Motion Menu Maps for HART Device Descriptions

RFT9739 Transmitters

Figure 1-5 Configure menu: Manual Setup: Measurements

On-Line Menu >2 Configure >

1 Manual Setup >2 Measurements

Flow

1 Flow Direction2 Flow Damping3 Mass Flow Cutoff4 Mass Flow Unit5 Volume Flow Cutoff6 Volume Flow Unit7 Mass Factor8 Volume Factor

Density

1 Density Unit2 Density Damping3 Density Factor4 Slug Low Limit5 Slug High Limit6 Slug Duration

Temperature

1 Temperature Unit2 Temp Damping

1

2

3

4

Set Up Special Units

1 Mass Special Units2 Volume Special Units

Set Up External Compensation

1 Pressure Configuration2 External Pressure

5

Pressure Configuration

1 Pressure Unit2 Flow Cal Pressure3 Flow Press Factor4 Dens Press Factor

1

2

External Pressure

1 Field Device Tag2 Pressure Polling3 4 mA Pressure4 20 mA Pressure

Configuration and Use Manual Supplement 5

RFT9739 Transmitters

9739 MV

DM

od

el 1700 AN

Mo

del 1500 A

NR

FT

9739

Figure 1-6 Configure menu: Manual Setup: Inputs/Outputs

On-Line Menu >2 Configure >

1 Manual Setup >4 Inputs/Outputs

Set Up mA Output 1

1 Primary Variable2 Primary Output

Set Up Frequency Output

1 Frequency Output

1

3

Primary Output

1 PV LRV2 PV URV3 PV Min Span4 PV USL5 PV LSL6 PV A01 Cutoff7 PV Added Damping

Frequency Output

1 Third Variable2 Max Pulse Width3 TV Frequency Factor4 TV Rate Factor5 Display K Factor

Set Up mA Output 2

1 Secondary Variable2 Secondary Output

Secondary Output

1 SV LRV2 SV URV3 SV Min Span4 SV USL5 SV LSL6 SV A01 Cutoff7 SV Added Damping

Map Variables

1 Primary Variable2 Secondary Variable3 Third Variable4 Fourth Variable

22

1

2

4

6 Micro Motion Menu Maps for HART Device Descriptions

RFT9739 Transmitters

Figure 1-7 Configure menu: Alert Setup

On-Line Menu >2 Configure >

2 Alert Setup

4

Set Up Events

1 Event 12 Event 2

Alert Setup

1 DO Fault Action2 Control Output3 Slug Duration4 Set Up Events

Event 1

1 Event 1 Variable2 Event 1 Type3 Event 1 Setpoint

Event 2

1 Event 2 Variable2 Event 2 Type3 Event 2 Setpoint

1 2

Configuration and Use Manual Supplement 7

RFT9739 Transmitters

9739 MV

DM

od

el 1700 AN

Mo

del 1500 A

NR

FT

9739

Figure 1-8 Service Tools menu: Top level

On-Line Menu >3 Service Tools

Alerts

1 Refresh Alerts2 Alert Name3 Additional Information forAbove

Variables

1 Variable Summary2 Process Variables3 Mapped Variables4 Totalizer Control5 Outputs

1 2

Trends

1 Process Variables2 Diagnostic Variables

3

Maintenance

1 Routine Maintenance2 Zero Calibration3 Density Calibration4 Temperature Calibration5 Diagnostic Variables

Simulate

1 Simulate Outputs

54

8 Micro Motion Menu Maps for HART Device Descriptions

RFT9739 Transmitters

Figure 1-9 Service Tools menu: Variables

On-Line Menu >3 Service Tools >

2 Variables

Variable Summary

Process Variables

1 Mass Flow Rate2 Volume Flow Rate3 Density4 Temperature

Diagnostic Variables

1 Drive Gain2 LPO Amplitude3 RPO Amplitude4 Tube Frequency5 Live Zero

1

2

3

5

Totalizer Control

1 All Totalizers2 Mass3 Volume

Outputs

1 Current (mA Output 1)2 Current (mA Output 2)3 Frequency

6

All Totalizers

1 Start Totalizers2 Stop Totalizers3 Reset All Totals4 Mass Total5 Volume Total

1

2

Mass

1 Mass Flow Rate2 Mass Total3 Inventory

Volume

1 Volume Flow Rate2 Volume Total3 Inventory

3

Mapped Variables

1 PV Mass Flow Rate2 SV Mass Flow Rate3 TV Mass Flow Rate4 QV Mass Flow Rate

4

Frequency

1 Frequency2 Present Freq Output

Current (MAO1)

1 Current2 PV AO3 PV % Range

Current (MAO2)

1 Current2 SV AO3 SV % Range

21 3

Configuration and Use Manual Supplement 9

RFT9739 Transmitters

9739 MV

DM

od

el 1700 AN

Mo

del 1500 A

NR

FT

9739

Figure 1-10 Service Tools menu: Maintenance

Figure 1-11 Service Tools menu: Simulate

On-Line Menu >3 Service Tools >

4 Maintenance

Routine Maintenance

1 Trim mA Output 12 Trim mA Output 2

1

Zero Calibration

1 Mass Flow Rate2 Volume Flow Rate3 Zero Time4 Zero Value5 Standard Deviation6 Perform Auto Zero

Density Calibration

1 Density2 Dens Pt1 (Air)3 Dens Pt2 (Water)4 Dens Pt3 T-Series

2

Temperature Calibration

1 Temperature2 Temp Cal Factor3 Offset Calibration4 Slope Calibration

4 5

Diagnostic Variables

1 Drive Gain2 LPO Amplitude3 RPO Amplitude4 Tube Frequency5 Live Zero

3

Simulate

1 Simulate Outputs

Simulate Outputs

1 mA Output 1 Loop Test2 mA Output 2 Loop Test3 Frequency Output Test

On-Line Menu >3 Service Tools >

5 Simulate

1

10 Micro Motion Menu Maps for HART Device Descriptions

Configuration and Use Manual Supplement 11

9739 MV

DM

od

el 1700 AN

Mo

del 1500 A

NR

FT

9739

Chapter 29739 MVD Transmitters

Figure 2-1 On-Line menu

Configure

1 Manual Setup2 Alert Setup

Service Tools

1 Alerts2 Variables3 Trends4 Maintenance5 Simulate

On-Line Menu

2 3

Overview

1 Check Status2 Primary Purpose Variables3 Shortcuts

1

12 Micro Motion Menu Maps for HART Device Descriptions

9739 MVD Transmitters

Figure 2-2 Overview menu

Check Status

1 Refresh Alerts2 Dev Status:3 Comm Status:

Primary Purpose Variables

Mass Flow RateVolume Flow RateDensity

Shortcuts

1 Device Information2 Totalizer Control3 Zero Calibration4 Variables5 Trends

1 32

Device Information

1 Identification2 Mat. of Construction3 Revisions4 Licenses

1

Indentification

1 Tag2 Model3 Final Asmbly Num4 Transmitter Serial Num5 Date6 Descriptor7 Message

Revisions

1 Universal2 Field Device3 DD Revision4 Xmtr Software Rev5 ETO Number

3

On-Line Menu >1 Overview

1

Configuration and Use Manual Supplement 13

9739 MVD Transmitters

9739 MV

DM

od

el 1700 AN

Mo

del 1500 A

NR

FT

9739

Figure 2-3 Configure menu: Top level

Manual Setup

1 Characterize2 Measurements3 Display4 Inputs/Outputs5 Info Parameters

21

On-Line Menu >1 Configure

Alert Setup

1 Inputs/Outputs Fault Actions2 Alert Severity3 Discrete Events

14 Micro Motion Menu Maps for HART Device Descriptions

9739 MVD Transmitters

Figure 2-4 Configure menu: Manual Setup

On-Line Menu >2 Configure >

1 Manual Setup

Characterize

1 Sensor Type2 Sensor TagParameters

Measurements

1 Flow2 Density3 Temperature4 Special Units5 External Compensation6 Petroleum Measurement *7 Gas Standard Volume8 Conc Measure (CM) **

1 2

Inputs/Outputs

1 mA Output 12 mA Output 23 Frequency Output4 Discrete Output5 Discrete Input6 Milliamp Input7 Communications8 Map Variables

3

Info Parameters

1 Transmitter Info2 Sensor Information

5

Display

1 Update Period2 Display Variables3 Decimal Places

4

* Displayed only if the petroleum measurementapplication is enabled on your transmitter. Menunumbers are adjusted as required.

** Displayed only if the concentration measurementapplication is enabled on your transmitter.

Configuration and Use Manual Supplement 15

9739 MVD Transmitters

9739 MV

DM

od

el 1700 AN

Mo

del 1500 A

NR

FT

9739

Figure 2-5 Configure menu: Manual Setup: Characterize

On-Line Menu >2 Configure >

1 Manual Setup >1 Characterize

Characterize

1 Sensor Type2 Sensor Tag Parameters

Flow Parameters

1 Flow FCF2 FTG3 FFQ

21

Density Parameters

1 D12 D23 K14 K25 FD

Sensor Tag Parameters

1 Flow Parameters2 Density Parameters

2

Sensor Type

1 Curved Tube2 Straight Tube

1

Sensor Type Straight TubeCurved Tube

Sensor Tag Parameters

1 FlowCal2 D13 D24 TC5 K16 K27 FD

16 Micro Motion Menu Maps for HART Device Descriptions

9739 MVD Transmitters

Figure 2-6 Configure menu: Manual Setup: Measurements

On-Line Menu >2 Configure >

1 Manual Setup >2 Measurements

Flow

1 Flow Direction2 Flow Damping3 Mass Flow Cutoff4 Mass Flow Unit5 Volume Flow Cutoff *6 Volume Flow Unit *7 Mass Factor8 Volume Factor

Density

1 Density Unit2 Density Damping3 Density Cutoff4 Density Factor5 Slug Low Limit6 Slug High Limit7 Slug Duration

Temperature

1 Temperature Unit2 Temp Damping

1

2

3

4

Special Units

1 Mass Special Units2 Volume Special Units

External Compensation

1 Pressure Unit2 Enable Press Comp3 Flow Cal Pressure4 Static Pressure5 Flow Press Factor6 Dens Press Factor7 Enable Ext Temp8 External Temperature9 External Polling

5

* Displayed only if Volume Flow Type =Liquid

Additional options

Configuration and Use Manual Supplement 17

9739 MVD Transmitters

9739 MV

DM

od

el 1700 AN

Mo

del 1500 A

NR

FT

9739

Figure 2-7 Configure menu: Manual Setup: Measurements (continued)

Petroleum Measurement *

1 2540 CTL Table type2 TEC3 Ref Temperature4 Petroleum Msmt Setup

Gas Standard Volume

1 Volume Flow Type2 Gas Density3 GSV Cutoff4 Gas Vol Flow Unit5 Base Volume Unit6 Base Volume Time7 Gas Vol Flow Conv8 Gas Vol Flow Label9 Gas Vol Total Label

Conc Measure (CM) **

1 CM Configuration2 Matrix Configuration3 Reset Matrix Data4 Enter Matrix5 Trim CM Process Variables

6

7

8

On-Line Menu >2 Configure >

1 Manual Setup >2 Measurements

* Displayed only if the petroleum measurementapplication is enabled on your transmitter.Menu numbers are adjusted as required.

** Displayed only if the concentrationmeasurement application is enabled on yourtransmitter.

CM Configuration

1 Enable CM2 Lock/Unlock CM Matrices3 Active Matrix4 Matrix Name5 Derived Variable

Matrix Configuration

1 Matrix Being Configured2 Matrix Name3 Concentration Units4 Extrapolation Alarm5 Reference Temperature6 Water Reference Temperature7 Water Reference Density

1

2

18 Micro Motion Menu Maps for HART Device Descriptions

9739 MVD Transmitters

Figure 2-8 Configure menu: Manual Setup: Display

On-Line Menu >2 Configure >

1 Manual Setup >3 Display

Update Period Display Variables

1 Display Variables (1-5)2 Display Variables (6-10)3 Display Variables (11-15)

Decimal Places

1 For Process Variables2 For Totalizer Variables3 For Diagnostic Variables

1 2 3

Configuration and Use Manual Supplement 19

9739 MVD Transmitters

9739 MV

DM

od

el 1700 AN

Mo

del 1500 A

NR

FT

9739

Figure 2-9 Configure menu: Manual Setup: Inputs/Outputs

On-Line Menu >2 Configure >

1 Manual Setup >4 Inputs/Outputs

mA Output 2

1 Secondary Variable2 Secondary Output3 MA02 Fault Settings

Frequency Output

1 FO Settings2 FO Fault Parameters3 FO Scaling

2 3

mA Fault Settings

1 MAO Fault Action2 MAO Fault Level

mA Output Settings

1 PV LRV2 PV URV3 PV Min Span4 PV LSL5 PV USL6 PV MAO Cutoff7 PV Added Damping

FO Settings

1 Third Variable2 Max Pulse Width3 FO Polarity

FO Fault Parameters

1 Third Variable2 FO Fault Action3 FO Fault Level

FO Scaling *

1 FO Scaling Method2 Set FO Scaling

Additionaloptions

2

3

1

mA Output 1

1 Primary Variable2 Primary Output3 MA01 Fault Settings

1

2

3

mA Fault Settings

1 MAO Fault Action2 MAO Fault Level

mA Output Settings

1 PV LRV2 PV URV3 PV Min Span4 PV LSL5 PV USL6 PV MAO Cutoff7 PV Added Damping

2

3

* Options vary depending onFO Scaling Method.

20 Micro Motion Menu Maps for HART Device Descriptions

9739 MVD Transmitters

Figure 2-10 Configure menu: Manual Setup: Inputs/Outputs (continued)

On-Line Menu >2 Configure >

1 Manual Setup >4 Inputs/Outputs

Discrete Output

1 DO Assignment2 DO Polarity3 DO Fault Action4 Flow Switch Source5 Flow Switch Setpoint

Discrete Input

1 Polarity

Communications

1 HART Communications2 HART Burst Mode3 Set Up RS-485 Port

4

8

Map Variables

1 Primary Variable2 Secondary Variable3 Third Variable4 Fourth Variable

5

7

HART Communications

1 HART Address2 Tag3 Device Identification

Set Up RS-485 Port

1 Protocol2 Modbus Slave Address

31

Milliamp Input

1 mA Input URV2 mA Input LRV3 mA Input Variable Assignment4 mA Input Level

6

Configuration and Use Manual Supplement 21

9739 MVD Transmitters

9739 MV

DM

od

el 1700 AN

Mo

del 1500 A

NR

FT

9739

Figure 2-11 Configure menu: Alert Setup

On-Line Menu >2 Configure >

2 Alert Setup

Input/Output Fault Action

1 MA Output 12 MA Output 23 FO Fault Parameters4 Discrete Output5 Digital Communications

Alert Severity

1 Fault Timeout2 Set Alert Severity3 View Alert Severity

Discrete Events

1 Discrete Event 12 Discrete Event 23 Discrete Event 34 Discrete Event 45 Discrete Event 56 Assign Discrete Action7 Read Discrete Action8 Review Discrete Actions

1 2 3

1, 2, 3, 4, 5

Discrete Event x

1 Discrete Event Var2 Discrete Event Type3 Setpoint A4 Setpoint B

22 Micro Motion Menu Maps for HART Device Descriptions

9739 MVD Transmitters

Figure 2-12 Service Tools menu: Top level

On-Line Menu >3 Service Tools

Alerts

1 Refresh Alerts2 Alert Name3 Additional Information forAbove

Variables

1 Variable Summary2 Process Variables3 Mapped Variables4 External Variables5 Totalizer Control6 Outputs

1 2

Trends

1 Process Variables2 Diagnostic Variables

3

Maintenance

1 Routine Maintenance2 Zero Calibration3 Density Calibration4 Temperature Calibration5 Diagnostic Variables

Simulate

1 Simulate Outputs2 Simulate Sensor

54

Configuration and Use Manual Supplement 23

9739 MVD Transmitters

9739 MV

DM

od

el 1700 AN

Mo

del 1500 A

NR

FT

9739

Figure 2-13 Service Tools menu: Variables

On-Line Menu >3 Service Tools >

2 Variables

Process Variables

1 Mass Flow Rate2 Volume Flow Rate *3 GSV Flow Rate4 Density5 Temperature

Mapped Variables

1 PV2 SV3 TV4 QV

1

3

5

Totalizer Control

1 All Totalizers2 Mass3 Volume *

Outputs

1 Current (mA output 1)2 Current (mA output 2)3 Frequency4 DO State

6

All Totalizers

1 Start Totalizers2 Stop Totalizers3 Reset All Totals4 Mass Total5 Volume Total *6 GSV Total

1

2

Mass

1 Mass Flow Rate2 Mass Total3 Mass Inventory4 Reset Total

Volume *

1 Flow Rate2 Total3 Inventory4 Reset Total

3

Variable Summary

External Variables

1 External Temperature2 External Pressure

4

2

* If Volume Flow Type = GSV,GSV variables are displayed.

24 Micro Motion Menu Maps for HART Device Descriptions

9739 MVD Transmitters

Figure 2-14 Service Tools menu: Maintenance

On-Line Menu >3 Service Tools >

4 Maintenance

Routine Maintenance

1 Trim mA Output 12 Trim mA Output 2

1

Zero Calibration

1 Mass Flow Rate2 Volume Flow Rate3 Zero Time4 Zero Value5 Standard Deviation6 Perform Auto Zero7 Restore Factory Zero

Density Calibration

1 Density2 Dens Pt1 (Air)3 Dens Pt2 (Water)4 Dens Pt3 T-Series5 Dens Pt4 T-Series6 Flowing Dens (FD)

2

Temperature Calibration

1 Temperature2 Temp Cal Factor

4 5

Diagnostic Variables

1 Sensor Model2 Drive Gain3 Input Voltage4 LPO Amplitude5 RPO Amplitude6 Board Temperature7 Tube Frequency8 Live Zero

3

Configuration and Use Manual Supplement 25

9739 MVD Transmitters

9739 MV

DM

od

el 1700 AN

Mo

del 1500 A

NR

FT

9739

Figure 2-15 Service Tools menu: Simulate

Simulate

1 Simulate Outputs2 Simulate Sensor

Simulate Outputs

1 mA Output 1 Loop Test2 mA Output 2 Loop Test3 Frequency Output Test4 Discrete Output Test

On-Line Menu >3 Service Tools >

5 Simulate

1

Simulate Sensor

1 Simulate Primary Purpose Variables2 Mass Flow Rate *3 Density *4 Temperature *

2

* Displayed only if sensor simulation isenabled.

26 Micro Motion Menu Maps for HART Device Descriptions

Configuration and Use Manual Supplement 27

9739 MV

DM

od

el 1700 AN

Mo

del 1500 A

NR

FT

9739

Chapter 3Model 1500 Transmitters with Analog Outputs

Figure 3-1 On-Line menu

Configure1 Manual Setup2 Alert Setup

Service Tools1 Alerts2 Variables3 Trends4 Maintenance5 Simulate

On-Line Menu

2

3

Overview1 Check Status2 Primary Purpose Variables3 Shortcuts

1

28 Micro Motion Menu Maps for HART Device Descriptions

Model 1500 Transmitters with Analog Outputs

Figure 3-2 Overview menu

Identification1 Tag2 Model3 Xmtr Serial Num4 Sensor Serial Num5 Date6 Descriptor7 Message

1

Revisions1 Universal2 Field Device3 DD Revision4 Transmitter Software5 CP Software6 ETO Number

Mat. of Construction1 Tube Wetted Mat.2 Tube Lining3 Sensor Flange

Check Status1 Refresh Alerts2 Dev Status:3 Comm Status:

Primary Purpose VariablesMass Flow RateVolume Flow RateDensity

Shortcuts1 Device Information2 Totalizer Control3 Zero Calibration4 Variables5 Trends6 Meter Verification *

13

2

1

2

3

On-Line Menu >1 Overview

Device Information1 Identification2 Revisions3 Mat. of Construction4 Licenses

* Displayed only if meterverification is enabled.

Configuration and Use Manual Supplement 29

Model 1500 Transmitters with Analog Outputs

9739 MV

DM

od

el 1700 AN

Mo

del 1500 A

NR

FT

9739

Figure 3-3 Configure menu: Top level

Manual Setup1 Characterize2 Measurements3 Inputs/Outputs4 Info Parameters5 Communications

2

1

Alert Setup1 Configure Alerts2 Discrete Output3 Discrete Events

On-Line Menu >1 Configure

30 Micro Motion Menu Maps for HART Device Descriptions

Model 1500 Transmitters with Analog Outputs

Figure 3-4 Configure menu: Manual Setup

Characterize1 Sensor Type2 Sensor Tag Parameters

1

2

Inputs/Outputs1 Set Up Channels2 Set Up mA Output3 Set Up Frequency Output4 Set Up Discrete Output5 Set Up RS-485 Port6 Map Variables

Info Parameters1 Transmitter Info2 Sensor Information

4

3

Transmitter Info1 Tag2 Final Assmbly Num3 Message4 Descriptor5 Date

Sensor Information1 Sensor Type2 Sensor Serial Num3 Sensor Material4 Sensor Lining5 Flange

1

2

On-Line Menu >2 Configure >

1 Manual Setup

Communications1 HART Address2 Tag3 Device Identification4 Dev ID (CP)5 Set Up Burst Mode

5

Measurements1 Flow2 Density3 Temperature4 Update Rate5 Set Up Special Units6 Set Up External Compensation7 Set Up GSV

Configuration and Use Manual Supplement 31

Model 1500 Transmitters with Analog Outputs

9739 MV

DM

od

el 1700 AN

Mo

del 1500 A

NR

FT

9739

Figure 3-5 Configure menu: Manual Setup: Characterize

On-Line Menu >2 Configure >

1 Manual Setup >1 Characterize

21

2

1

Sensor Type Straight TubeCurved Tube

Density Parameters1 D12 D23 DT4 DTG5 K16 K27 FD8 DFQ19 DFQ2

Flow Parameters1 Flow FCF2 FTG3 FFQ

Sensor Tag Parameters1 Flow Parameters2 Density Parameters

Sensor Tag Parameters1 FlowCal2 D13 D24 TC5 K16 K27 FD

Sensor Type1 Curved Tube2 Straight Tube

32 Micro Motion Menu Maps for HART Device Descriptions

Model 1500 Transmitters with Analog Outputs

Figure 3-6 Configure menu: Manual Setup: Measurements

* Displayed only if Volume Flow Type =Liquid. Menu numbers are adjustedas required.

On-Line Menu >2 Configure >

1 Manual Setup >2 Measurements

Flow1 Flow Direction2 Flow Damping3 Mass Flow Unit4 Mass Flow Cutoff5 Volume Flow Unit *6 Volume Flow Cutoff *7 Mass Factor8 Volume Factor

Density1 Density Unit2 Density Damping3 Density Cutoff4 Density Factor5 Slug High Limit6 Slug Low Limit7 Slug Duration

Temperature1 Temperature Unit2 Temp Damping

1

2

3

5Set Up Special Units1 Mass Special Units2 Volume Special Units

Update Rate4

Set Up External Compensation1 Pressure Unit2 Enable Press Comp3 Flow Cal Pressure4 Static Pressure5 Flow Press Factor6 Dens Press Factor7 Enable Ext Temp8 External Temperature9 External Polling

Set Up GSV1 Volume Flow Type2 Gas Density3 Gas Vol Flow Cutoff4 Gas Vol Flow Unit5 Gas Density Unit

6

7

Configuration and Use Manual Supplement 33

Model 1500 Transmitters with Analog Outputs

9739 MV

DM

od

el 1700 AN

Mo

del 1500 A

NR

FT

9739

Figure 3-7 Configure menu: Manual Setup: Inputs/Outputs

On-Line Menu >2 Configure >

1 Manual Setup >4 Inputs/Outputs

Set Up mA Output1 Primary Variable2 mA Output Settings3 mA Fault Settings

Set Up Frequency Output1 FO Settings2 FO Fault Parameters3 FO Scaling

2

3

mA Fault Settings1 MAO Fault Action2 MAO Fault Level

mA Output Settings1 PV LRV2 PV URV3 PV Min Span4 PV LSL5 PV USL6 PV MAO Cutoff7 PV Added Damping

FO Settings1 Third Variable2 Max Pulse Width3 FO Polarity

FO Fault Parameters1 Third Variable2 FO Fault Action3 FO Fault Level

FO Scaling *1 FO Scaling Method2 TV Frequency Factor3 TV Rate Factor4 Set FO Scaling

2

3

1

Set Up Channels1 Channel A2 Channel C

1

2

3

* Options vary dependingon FO Scaling Method.

Additionaloptions

Channel C1 Frequency Output2 Discrete Output

2

34 Micro Motion Menu Maps for HART Device Descriptions

Model 1500 Transmitters with Analog Outputs

Figure 3-8 Configure menu: Manual Setup: Inputs/Outputs (continued)

Discrete Output1 DO Assignment2 DO Polarity3 DO Fault Action4 Flow Switch Source5 Flow Switch Setpoint6 Hysteresis (0.1-10.0)

46

Map Variables1 Primary Variable2 Secondary Variable3 Third Variable4 Fourth Variable

5

Set Up RS-485 Port1 Protocol2 Baud Rate3 Stop Bits4 Dev ID (CP)5 Modbus Slave Address

On-Line Menu >2 Configure >

1 Manual Setup >4 Inputs/Outputs

Configuration and Use Manual Supplement 35

Model 1500 Transmitters with Analog Outputs

9739 MV

DM

od

el 1700 AN

Mo

del 1500 A

NR

FT

9739

Figure 3-9 Configure menu: Alert Setup

On-Line Menu >2 Configure >

2 Alert Setup

Configure Alerts1 Fault Timeout2 MAO Fault Action3 MAO Fault Level4 FO Fault Action5 FO Fault Level6 Comm Fault Action7 Set Up Alert Severity8 View Alert Severity

Discrete Output1 DO Assignment2 DO Polarity3 DO Fault Action4 Flow Switch Source5 Flow Switch Setpoint6 Hysteresis (0.1.10.9)

Discrete Events1 Discrete Event 12 Discrete Event 23 Discrete Event 34 Discrete Event 45 Discrete Event 56 Assign Discrete Action7 Read Discrete Action8 Review Discrete Actions

1

2

3

1, 2, 3, 4, 5

Discrete Event x1 Discrete Event Var2 Discrete Event Type3 Setpoint A4 Setpoint B

36 Micro Motion Menu Maps for HART Device Descriptions

Model 1500 Transmitters with Analog Outputs

Figure 3-10 Service Tools menu: Top level

2

Process Variables1 Mass Flow Rate2 Volume Flow Rate3 Gas Standard Volume4 Density5 Temperature

Diagnostic Variables1 Drive Gain2 LPO Amplitude3 RPO Amplitude4 Tube Frequency5 Live Zero6 Input Voltage7 Board Temperature8 Fld Verification Zero

1

On-Line Menu >3 Service Tools

Alerts1 Refresh Alerts

Alert NameAdditional Informationfor Above

Variables1 Variable Summary2 Process Variables3 Mapped Variables4 External Variables5 Totalizer Control6 Outputs

1

2

Trends1 Process Variables2 Diagnostic Variables

3

Maintenance1 Routine Maintenance2 Zero Calibration3 Density Calibration4 Temperature Calibration5 Diagnostic Variables

Simulate1 Simulate Outputs2 Simulate Sensor

4

5

* If Volume Flow Type = GSV,GSV variables are displayed.

Configuration and Use Manual Supplement 37

Model 1500 Transmitters with Analog Outputs

9739 MV

DM

od

el 1700 AN

Mo

del 1500 A

NR

FT

9739

Figure 3-11 Service Tools menu: Variables

* If Volume Flow Type = GSV,GSV variables are displayed.

On-Line Menu >3 Service Tools >

2 Variables

Process Variables1 Mass Flow Rate2 Volume Flow Rate *3 Density4 Temperature

Mapped Variables1 PV Mass Flow Rate2 SV Mass Flow Rate3 TV Mass Flow Rate4 QV Mass Flow Rate

1

3

5

Totalizer Control1 All Totalizers2 Mass3 Volume *

All Totalizers1 Start Totalizers2 Stop Totalizers3 Reset All Totals4 Mass Total5 Volume Total *

1

2

Mass1 Mass Flow Rate2 Mass Total3 Mass Inventory4 Reset Total

Volume *1 Volume Flow Rate2 Volume Total3 Volume Inventory4 Reset Total

3

Variable Summary

External Variables1 External Temperature2 External Pressure

4

2

6Outputs1 Current2 Frequency/DO State

Current1 Current2 PV AO3 PV % Range

Frequency1 Frequency2 Present Freq Output

1

2

38 Micro Motion Menu Maps for HART Device Descriptions

Model 1500 Transmitters with Analog Outputs

Figure 3-12 Service Tools menu: Maintenance

On-Line Menu >3 Service Tools >

4 Maintenance

Routine Maintenance1 Trim mA Output2 Meter Verification *

1

Zero Calibration1 Mass Flow Rate2 Volume Flow Rate3 Zero Time4 Zero Value5 Standard Deviation6 Perform Auto Zero7 Restore Factory Zero

Density Calibration1 Density2 Dens Pt1 (Air)3 Dens Pt2 (Water)4 Dens Pt3 T-Series5 Dens Pt4 T-Series6 Flowing Dens (FD)

2

Temperature Calibration1 Temperature2 Temp Cal Factor

4

Diagnostic Variables1 Sensor Model2 Drive Gain3 Input Voltage4 LPO Amplitude5 RPO Amplitude6 Board Temperature7 Tube Frequency8 Live Zero9 Fld Verification Zero

3

Meter Verification **1 Run Meter Verification2 View Test Results3 Schedule Meter

Verification

2

* Displayed only if meter verificationis enabled.

** Displayed only if Smart MeterVerification is enabled. For earlierversions, the Meter VerificationMethod is launched.

Additional1 Core Processor Input Voltage2 Board Temperature3 Power On Time

8

Modbus Data1 Read Modbus Data2 Write Modbus Data

6

5

Configuration and Use Manual Supplement 39

Model 1500 Transmitters with Analog Outputs

9739 MV

DM

od

el 1700 AN

Mo

del 1500 A

NR

FT

9739

Figure 3-13 Service Tools menu: Simulate

Simulate Outputs1 mA Output Loop Test2 Frequency Output Test3 Discrete Output Test

On-Line Menu >3 Service Tools >

5 Simulate

1

Simulate Sensor1 Simulate Primary Purpose Variables2 Mass Flow Rate3 Density4 Temperature

2

40 Micro Motion Menu Maps for HART Device Descriptions

Configuration and Use Manual Supplement 41

9739 MV

DM

od

el 1700 AN

Mo

del 1500 A

NR

FT

9739

Chapter 4Model 1700 Transmitters with Analog Outputs

Figure 4-1 On-Line menu

Configure1 Manual Setup2 Alert Setup

Service Tools1 Alerts2 Variables3 Trends4 Maintenance5 Simulate

2

3

Overview1 Check Status2 Primary Purpose Variables3 Shortcuts

1

On-Line Menu

42 Micro Motion Menu Maps for HART Device Descriptions

Model 1700 Transmitters with Analog Outputs

Figure 4-2 Overview menu

Identification1 Tag2 Model3 Xmtr Serial Num4 Sensor Serial Num5 Date6 Descriptor7 Message

1

Revisions1 Universal2 Field Device3 DD Revision4 Transmitter Software5 CP Software6 ETO Number

Mat. of Construction1 Tube Wetted Mat.2 Tube Lining3 Sensor Flange

Check Status1 Refresh Alerts2 Dev Status:3 Comm Status:

Primary Purpose VariablesMass Flow RateVolume Flow RateDensity

Shortcuts1 Device Information2 Totalizer Control3 Zero Calibration4 Variables5 Trends6 Meter Verification *

13

21

2

3

On-Line Menu >1 Overview

Device Information1 Identification2 Revisions3 Mat. of Construction4 Licenses

* Displayed only if meterverification is enabled.

Configuration and Use Manual Supplement 43

Model 1700 Transmitters with Analog Outputs

9739 MV

DM

od

el 1700 AN

Mo

del 1500 A

NR

FT

97399739 M

VD

Mo

del 1700 A

NM

od

el 1500 AN

RF

T9739

9739 MV

DM

od

el 1700 AN

Mo

del 1500 A

NR

FT

97399739 M

VD

Mo

del 1700 A

NM

od

el 1500 AN

RF

T9739

Figure 4-3 Configure menu: Top level

Manual Setup1 Characterize2 Measurements3 Display4 Inputs/Outputs5 Info Parameters

2

1

Alert Setup1 I/O Fault Actions2 Alert Severity3 Discrete Events

On-Line Menu >1 Configure

44 Micro Motion Menu Maps for HART Device Descriptions

Model 1700 Transmitters with Analog Outputs

Figure 4-4 Configure menu: Manual Setup

Characterize1 Sensor Type2 Sensor Tag Parameters

Measurements1 Flow2 Density3 Temperature4 Update Rate5 LD Optimization6 Special Units7 External Pressure/Temperature8 GSV

1

2

Inputs/Outputs1 Channels2 mA Output3 Frequency Output4 Discrete Output5 Communications6 Variable Mapping

3

Info Parameters1 Transmitter Info2 Sensor Information

Communications1 HART Address2 Tag3 Device Identification4 Dev ID (CP)5 Set Up Burst Mode6 mA Output Mode7 Set Up RS-485 Port

5

Display1 Language2 Display Variable Menu Features3 Offline Variable Menu Features4 Backlight5 Display Variables6 Decimal Places

4

Transmitter Info1 Tag2 Xmtr Serial Num3 Message4 Descriptor5 Date6 Write Protect

Sensor Information1 Sensor Type2 Sensor Serial Num3 Tube Wetted Mat.4 Tube Lining5 Sensor Flange

1

2

On-Line Menu >2 Configure >

1 Manual Setup

5

Configuration and Use Manual Supplement 45

Model 1700 Transmitters with Analog Outputs

9739 MV

DM

od

el 1700 AN

Mo

del 1500 A

NR

FT

97399739 M

VD

Mo

del 1700 A

NM

od

el 1500 AN

RF

T9739

9739 MV

DM

od

el 1700 AN

Mo

del 1500 A

NR

FT

97399739 M

VD

Mo

del 1700 A

NM

od

el 1500 AN

RF

T9739

Figure 4-5 Configure menu: Manual Setup: Characterize

On-Line Menu >2 Configure >

1 Manual Setup >1 Characterize

21

2

1

Sensor Type Straight TubeCurved Tube

Density Parameters1 D12 D23 DT4 DTG5 K16 K27 FD8 DFQ19 DFQ2

Flow Parameters1 Flow FCF2 FTG3 FFQ

Sensor Tag Parameters1 Flow Parameters2 Density Parameters

Sensor Tag Parameters1 FlowCal2 D13 D24 TC5 K16 K27 FD

Sensor Type1 Curved Tube2 Straight Tube

46 Micro Motion Menu Maps for HART Device Descriptions

Model 1700 Transmitters with Analog Outputs

Figure 4-6 Configure menu: Manual Setup: Measurements

* Displayed only if Volume Flow Type =Liquid. Menu numbers are adjustedas required.

On-Line Menu >2 Configure >

1 Manual Setup >2 Measurements

Flow1 Flow Direction2 Flow Damping3 Mass Flow Unit4 Mass Flow Cutoff5 Volume Flow Unit *6 Volume Flow Cutoff *7 Mass Factor8 Volume Factor

Density1 Density Unit2 Density Damping3 Density Cutoff *4 Density Factor5 Slug Duration6 Slug Low Limit7 Slug High Limit

Temperature1 Temperature Unit2 Temp Damping

1

2

3

6Special Units1 Mass Special Units2 Volume Special Units

Update Rate1 Update Rate2 100 Hz Variable

4

External Pressure/Temperature1 Pressure2 Temperature3 External Polling

GSV1 Volume Flow Type2 Gas Ref Density3 GSV Cutoff4 GSV Flow Unit5 Gas Density Unit

7

8

Pressure1 Pressure Unit2 Pressure Compensation3 Compensation Pressure4 Flow Cal Pressure5 Flow Press Factor6 Dens Press Factor

Temperature1 Temperature Unit2 External Temperature3 Correction Temperature

External Polling1 Poll Control2 Ext Dev Tag 13 Polled Variable 14 Ext Dev Tag 25 Polled Variable 2

1

2

3

Configuration and Use Manual Supplement 47

Model 1700 Transmitters with Analog Outputs

9739 MV

DM

od

el 1700 AN

Mo

del 1500 A

NR

FT

97399739 M

VD

Mo

del 1700 A

NM

od

el 1500 AN

RF

T9739

9739 MV

DM

od

el 1700 AN

Mo

del 1500 A

NR

FT

97399739 M

VD

Mo

del 1700 A

NM

od

el 1500 AN

RF

T9739

Figure 4-7 Configure menu: Manual Setup: Display

On-Line Menu >2 Configure >

1 Manual Setup >3 Display

LanguageEnglishGermanFrenchSpanish

Display Variable MenuFeatures1 Totalizer Reset2 Start/Stop Totals3 Auto Scroll4 Scroll Time *5 Refresh Rate6 Status LED Blinking

Offline Variable MenuFeatures1 Offline Menu2 Alert Menu3 Acknowledge All4 Offline Passcode5 Alert Passcode6 Offline Passcode

1

2

3

5

Display Variables1 Display Variables (1-5)2 Display Variables (6-10)3 Display Variables (11-15)

Decimal Places1 For Process Variables2 For Totalizer Variables3 For Diagnostic Variables

6

Backlight1 Control2 Intensity (0-63)

4

* Displayed only if Auto Scroll isenabled. Menu numbers areadjusted as required.

48 Micro Motion Menu Maps for HART Device Descriptions

Model 1700 Transmitters with Analog Outputs

Figure 4-8 Configure menu: Manual Setup: Inputs/Outputs

On-Line Menu >2 Configure >

1 Manual Setup >4 Inputs/Outputs

mA Output1 Primary Variable2 mA Output Settings3 mA Fault Settings

Frequency Output1 FO Settings2 FO Fault Parameters3 FO Scaling

2

3

mA Fault Settings1 mAO Fault Action2 mAO Fault Level

mA Output Settings1 PV LRV2 PV URV3 PV Min Span4 PV LSL5 PV USL6 PV MAO Cutoff7 PV Added Damping

FO Settings1 Third Variable2 Max Pulse Width3 FO Polarity

FO Fault Parameters1 Third Variable2 FO Fault Action3 FO Fault Level

FO Scaling *1 FO Scaling Method2 TV Pulses/Unit3 Set FO Scaling

2

3

1

Channels1 Channel A2 Channel B3 Channel C

1

2

3

* Options vary dependingon FO Scaling Method.

Additionaloptions

Channel B1 Frequency Output2 Discrete Output

2

Configuration and Use Manual Supplement 49

Model 1700 Transmitters with Analog Outputs

9739 MV

DM

od

el 1700 AN

Mo

del 1500 A

NR

FT

97399739 M

VD

Mo

del 1700 A

NM

od

el 1500 AN

RF

T9739

9739 MV

DM

od

el 1700 AN

Mo

del 1500 A

NR

FT

97399739 M

VD

Mo

del 1700 A

NM

od

el 1500 AN

RF

T9739

Figure 4-9 Configure menu: Manual Setup: Inputs/Outputs (continued)

Discrete Output1 DO Assignment2 DO Polarity3 DO Fault Action4 Flow Switch Source5 Flow Switch Setpoint6 Hysteresis (0.1-10.0)

46

Map Variables1 Primary Variable2 Secondary Variable3 Third Variable4 Fourth Variable

5

Communications1 HART Address2 Tag3 Device Identification4 Dev ID (CP)5 Set Up Burst Mode6 mA Output Mode7 Set Up RS-485 Port

Set Up Burst Mode1 Burst Mode2 Burst Option3 Field Device Var 14 Field Device Var 25 Field Device Var 36 Field Device Var 4

Set Up RS-485 Port1 Protocol2 Baud Rate3 Parity4 Stop Bits5 Modbus Slave Address

7

5

On-Line Menu >2 Configure >

1 Manual Setup >4 Inputs/Outputs

50 Micro Motion Menu Maps for HART Device Descriptions

Model 1700 Transmitters with Analog Outputs

Figure 4-10 Configure menu: Alert Setup

On-Line Menu >2 Configure >

2 Alert Setup

I/O Fault Actions1 mAO Fault Action2 mAO Fault Level3 FO Fault Action4 FO Fault Level5 Comm Fault Action

Alert Severity1 Fault Timeout2 Set Alert Severity3 View Alert Severity

Discrete Events1 Discrete Event 12 Discrete Event 23 Discrete Event 34 Discrete Event 45 Discrete Event 56 Assign Discrete Action7 Read Discrete Action8 Review Discrete Actions

1

2

3

1, 2, 3, 4, 5

Discrete Event x1 Discrete Event Var2 Discrete Event Type3 Setpoint A4 Setpoint B

Configuration and Use Manual Supplement 51

Model 1700 Transmitters with Analog Outputs

9739 MV

DM

od

el 1700 AN

Mo

del 1500 A

NR

FT

97399739 M

VD

Mo

del 1700 A

NM

od

el 1500 AN

RF

T9739

9739 MV

DM

od

el 1700 AN

Mo

del 1500 A

NR

FT

97399739 M

VD

Mo

del 1700 A

NM

od

el 1500 AN

RF

T9739

Figure 4-11 Service Tools menu: Top level

On-Line Menu >3 Service Tools

Alerts1 Refresh Alerts

Alert NameAdditional Informationfor Above

Variables1 Variable Summary2 Process Variables3 Mapped Variables4 External Variables5 Totalizer Control6 Outputs

1

2

Trends1 Process Variables2 Diagnostic Variables

3

Maintenance1 Routine Maintenance2 Zero Calibration3 Density Calibration4 Temperature Calibration5 Diagnostic Variables6 Modbus Data

Simulate1 Simulate Outputs2 Simulate Sensor

4

5

52 Micro Motion Menu Maps for HART Device Descriptions

Model 1700 Transmitters with Analog Outputs

Figure 4-12 Service Tools menu: Variables

* If Volume Flow Type = GSV,GSV variables are displayed.

On-Line Menu >3 Service Tools >

2 Variables

Process Variables1 Mass Flow Rate2 Volume Flow Rate *3 Density4 Temperature

Mapped Variables1 PV Mass Flow Rate2 SV Mass Flow Rate3 TV Mass Flow Rate4 QV Mass Flow Rate

1

3

5

Totalizer Control1 All Totalizers2 Mass3 Volume *

All Totalizers1 Start Totalizers2 Stop Totalizers3 Reset All Totals4 Mass Total5 Volume Total *

1

2

Mass1 Mass Flow Rate2 Mass Total3 Mass Inventory4 Reset Total

Volume *1 Volume Flow Rate2 Volume Total3 Volume Inventory4 Reset Total

3

Variable SummaryMass Flow RateVolume Flow RateDensity

External Variables1 External Temperature2 External Pressure

4

2

6

Outputs1 mA Output2 Frequency Output/

DO State

Frequency Output1 Present Freq Gauge2 Present Freq Output

mA Output1 Present MAO Gauge2 PV MAO3 PV % Range

1

2

Additionaloptions

Configuration and Use Manual Supplement 53

Model 1700 Transmitters with Analog Outputs

9739 MV

DM

od

el 1700 AN

Mo

del 1500 A

NR

FT

97399739 M

VD

Mo

del 1700 A

NM

od

el 1500 AN

RF

T9739

9739 MV

DM

od

el 1700 AN

Mo

del 1500 A

NR

FT

97399739 M

VD

Mo

del 1700 A

NM

od

el 1500 AN

RF

T9739

Figure 4-13 Service Tools menu: Maintenance

On-Line Menu >3 Service Tools >

4 Maintenance

Routine Maintenance1 Trim mA Output2 Meter Verification *

1

Zero Calibration1 Mass Flow Rate2 Volume Flow Rate3 Zero Time4 Zero Value5 Standard Deviation6 Perform Auto Zero7 Restore Factory Zero

Density Calibration1 Density2 Dens Pt1 (Air)3 Dens Pt2 (Water)4 Dens Pt3 T-Series5 Dens Pt4 T-Series6 Flowing Dens (FD)7 Flowing Dens (FD)

2

Temperature Calibration1 Temperature2 Temp Cal Factor

4

3

Meter Verification **1 Run Meter Verification2 View Test Results3 Schedule Meter

Verification

2

* Displayed only if meter verificationis enabled.

** Displayed only if Smart MeterVerification is enabled. For earlierversions, the Meter VerificationMethod is launched.

Additional1 Core Processor Input

Voltage2 Board Temperature3 Power On Time

8

Modbus Data1 Read Modbus Data2 Write Modbus Data

6

5

Diagnostic Variables1 Sensor Model2 Drive Gain3 LPO Amplitude4 RPO Amplitude5 Tube Frequency6 Live Zero7 Fld Verification Zero8 Additional

54 Micro Motion Menu Maps for HART Device Descriptions

Model 1700 Transmitters with Analog Outputs

Figure 4-14 Service Tools menu: Simulate

On-Line Menu >3 Service Tools >

5 Simulate

1

Simulate Sensor1 Simulate Primary Purpose Variables2 Mass Flow Rate3 Density4 Temperature

2

Simulate Outputs1 mA Output Loop Test2 Frequency Output Test/

Discrete Output Test *

* Options vary dependingon channel settings.

Configuration and Use Manual Supplement 55

Mo

del 2200S

Mo

del 2500 C

IOM

od

el 2400S A

NM

od

el 1700 IS

Chapter 5Model 1700 Transmitters with Intrinsically Safe Outputs

Figure 5-1 On-Line menu

Configure1 Manual Setup2 Alert Setup

Service Tools1 Alerts2 Variables3 Trends4 Maintenance5 Simulate

2

3

Overview1 Check Status2 Primary Purpose Variables3 Shortcuts

1

On-Line Menu

56 Micro Motion Menu Maps for HART Device Descriptions

Model 1700 Transmitters with Intrinsically Safe Outputs

Figure 5-2 Overview menu

Product Information1 Tag2 Model3 Xmtr Software Rev4 CP Software Rev5 Option Board6 ETO Number7 Final Asmbly Num8 Sensor Serial Num9 HART DD Information

1

Checksums1 Core Firmware2 Xmtr Firmware

Mat. of Construction1 Sensor Material2 Sensor Liner3 Flange

Check Status1 Refresh Alerts2 Dev Status:3 Comm Status:

Primary Purpose VariablesMass Flow RateVolume Flow RateDensity

Shortcuts1 Device Information2 Totalizer Control3 Zero Calibration4 Variables5 Trends6 Meter Verification *

13

21

4

2

On-Line Menu >1 Overview

Device Information1 Product Information2 Mat. of Construction3 Licenses4 Checksums

* Displayed only if meterverification is enabled.

Licenses1 Features3

Configuration and Use Manual Supplement 57

Model 1700 Transmitters with Intrinsically Safe Outputs

Mo

del 2200S

Mo

del 2500 C

IOM

od

el 2400S A

NM

od

el 1700 IS

Figure 5-3 Configure menu: Top level

Manual Setup1 Characterize2 Measurements3 Display4 Inputs/Outputs5 Info Parameters6 Communications

2

1

Alert Setup1 Configure Alerts2 Discrete Output3 Discrete Events

On-Line Menu >1 Configure

58 Micro Motion Menu Maps for HART Device Descriptions

Model 1700 Transmitters with Intrinsically Safe Outputs

Figure 5-4 Configure menu: Manual Setup

Characterize1 Sensor Type2 Sensor Tag Parameters

Measurements1 Flow2 Density3 Temperature4 Update Rate5 Set Up Special Units6 Set Up External Compensation7 Set Up GSV

1

2

Inputs/Outputs1 Set Up Channels2 Set Up mA Output3 Set Up Frequency Output4 Set Up Discrete Output5 Map Variables

3

Info Parameters1 Transmitter Info2 Sensor Information

5

Display1 Language2 Display Variable Menu Features3 Offline Variable Menu Features4 Backlight5 Set Up Display Variables6 Set Up Decimal Places

4

Transmitter Info1 Tag2 Xmtr Serial Num3 Message4 Descriptor5 Date6 Write Protect

Sensor Information1 Sensor Type2 Sensor Serial Num3 Tube Wetted Mat.4 Tube Lining5 Sensor Flange

1

2

On-Line Menu >2 Configure >

1 Manual Setup

Communications1 HART Address2 Tag3 Device Identification4 Dev ID (CP)5 Set Up Burst Mode

6

Configuration and Use Manual Supplement 59

Model 1700 Transmitters with Intrinsically Safe Outputs

Mo

del 2200S

Mo

del 2500 C

IOM

od

el 2400S A

NM

od

el 1700 IS

Figure 5-5 Configure menu: Manual Setup: Characterize

On-Line Menu >2 Configure >

1 Manual Setup >1 Characterize

21

2

1

Sensor Type Straight TubeCurved Tube

Density Parameters1 D12 D25 K16 K27 FD

Flow Parameters1 Flow FCF2 FTG3 FFQ

Sensor Tag Parameters1 Flow Parameters2 Density Parameters

Sensor Tag Parameters1 FlowCal2 D13 D24 TC5 K16 K27 FD

Sensor TypeCurved TubeStraight Tube

60 Micro Motion Menu Maps for HART Device Descriptions

Model 1700 Transmitters with Intrinsically Safe Outputs

Figure 5-6 Configure menu: Manual Setup: Measurements

* Displayed only if Volume Flow Type =Liquid. Menu numbers are adjustedas required.

On-Line Menu >2 Configure >

1 Manual Setup >2 Measurements

Flow1 Flow Direction2 Flow Damping3 Mass Flow Cutoff4 Mass Flow Unit5 Volume Flow Cutoff *6 Volume Flow Unit *7 Mass Factor8 Volume Factor

Density1 Density Unit2 Density Damping3 Density Cutoff *4 Density Factor5 Slug High Limit6 Slug Low Limit7 Slug Duration

Temperature1 Temperature Unit2 Temp Damping

1

2

3

5Set Up Special Units1 Mass Special Units2 Volume Special Units

Update Rate1 Update Rate4

Set Up External Compensation1 Pressure Unit2 Enable Press Comp3 Flow Cal Pressure4 Static Pressure5 Flow Press Factor6 Dens Press Factor7 Enable Ext Temp8 External Temperature9 External Polling

Set Up GSV1 Volume Flow Type2 Gas Density3 Gas Vol Flow Cutoff4 Gas Vol Flow Unit5 Gas Density Unit

6

7

External Polling1 Poll Control2 Ext Dev Tag 13 Polled Variable 14 Ext Dev Tag 25 Polled Variable 2

9

Configuration and Use Manual Supplement 61

Model 1700 Transmitters with Intrinsically Safe Outputs

Mo

del 2200S

Mo

del 2500 C

IOM

od

el 2400S A

NM

od

el 1700 IS

Figure 5-7 Configure menu: Manual Setup: Display

On-Line Menu >2 Configure >

1 Manual Setup >3 Display

LanguageEnglishGermanFrenchSpanish

Display Variable MenuFeatures1 Totalizer Reset2 Start/Stop Totals3 Auto Scroll4 Scroll Time *5 Update Period6 Status LED Blinking

Offline Variable MenuFeatures1 Offline Menu2 Alert Menu3 Acknowledge All4 Offline Passcode5 Alert Passcode

1

2

3

5

Set Up Display Variables1 Display Variables (1-5)2 Display Variables (6-10)3 Display Variables (11-15)

Set Up Decimal Places1 For Process Variables2 For Totalizer Variables3 For Diagnostic Variables

6

Backlight1 Control2 Intensity (0-63)

4

* Displayed only if Auto Scroll isenabled. Menu numbers areadjusted as required.

62 Micro Motion Menu Maps for HART Device Descriptions

Model 1700 Transmitters with Intrinsically Safe Outputs

Figure 5-8 Configure menu: Manual Setup: Inputs/Outputs

On-Line Menu >2 Configure >

1 Manual Setup >4 Inputs/Outputs

Set Up mA Output1 Primary Variable2 mA Output Settings3 mA Fault Settings

Set Up Frequency Output1 FO Settings2 FO Fault Parameters3 FO Scaling

2

3

mA Fault Settings1 mAO Fault Action2 mAO Fault Level

mA Output Settings1 PV LRV2 PV URV3 PV Min Span4 PV LSL5 PV USL6 PV MAO Cutoff7 PV Added Damping

FO Settings1 Third Variable2 Max Pulse Width3 FO Polarity

FO Fault Parameters1 Third Variable2 FO Fault Action3 FO Fault Level

FO Scaling *1 FO Scaling Method2 TV Frequency Factor3 TV Rate Factor4 Set FO Scaling

2

3

1

Set Up Channels1 Channel A2 Channel B

1

2

3

* Options vary dependingon FO Scaling Method.

Additionaloptions

Channel BFrequency OutputDiscrete Output

2

Configuration and Use Manual Supplement 63

Model 1700 Transmitters with Intrinsically Safe Outputs

Mo

del 2200S

Mo

del 2500 C

IOM

od

el 2400S A

NM

od

el 1700 IS

Figure 5-9 Configure menu: Manual Setup: Inputs/Outputs (continued)

Set Up Discrete Output1 DO Assignment2 DO Polarity3 DO Fault Action4 Flow Switch Source5 Flow Switch Setpoint6 Hysteresis (0.1-10.0)

46

Map Variables1 Primary Variable2 Secondary Variable3 Third Variable4 Fourth Variable

5

Communications1 HART Address2 Tag3 Device Identification4 Dev ID (CP)5 Set Up Burst Mode6 mA Output Mode7 Set Up RS-485 Port

Set Up Burst Mode1 Burst Mode2 Burst Option3 Field Device Var 14 Field Device Var 25 Field Device Var 36 Field Device Var 4

Set Up RS-485 Port1 Protocol2 Baud Rate3 Parity4 Stop Bits5 Modbus Slave Address

7

5

On-Line Menu >2 Configure >

1 Manual Setup >4 Inputs/Outputs

64 Micro Motion Menu Maps for HART Device Descriptions

Model 1700 Transmitters with Intrinsically Safe Outputs

Figure 5-10 Configure menu: Alert Setup

On-Line Menu >2 Configure >

2 Alert Setup

Configure Alerts1 Fault Timeout2 MAO Fault Action3 MAO Fault Level4 FO Fault Action5 FO Fault Level6 Comm Fault Action7 Set Alert Severity8 View Alert Severity

Discrete Output1 DO Assignment2 DO Polarity3 DO Fault Action4 Flow Switch Source5 Flow Switch Setpoint6 Hysteresis (0.1-10.0)

Discrete Events1 Discrete Event 12 Discrete Event 23 Discrete Event 34 Discrete Event 45 Discrete Event 56 Assign Discrete Action7 Read Discrete Action8 Review Discrete Actions

1

2

3

1, 2, 3, 4, 5

Discrete Event x1 Discrete Event Var2 Discrete Event Type3 Setpoint A4 Setpoint B

Configuration and Use Manual Supplement 65

Model 1700 Transmitters with Intrinsically Safe Outputs

Mo

del 2200S

Mo

del 2500 C

IOM

od

el 2400S A

NM

od

el 1700 IS

Figure 5-11 Service Tools menu: Top level

On-Line Menu >3 Service Tools

Alerts1 Refresh Alerts2 (Alert Name)3 Additional Information

for Above

Variables1 Variable Summary2 Process Variables3 Mapped Variables4 External Variables5 Totalizer Control6 Outputs

1

2

Trends1 Process Variables2 Diagnostic Variables

3

Maintenance1 Routine Maintenance2 Zero Calibration3 Density Calibration4 Temperature Calibration5 Diagnostic Variables

Simulate1 Simulate Outputs2 Simulate Sensor

4

5

66 Micro Motion Menu Maps for HART Device Descriptions

Model 1700 Transmitters with Intrinsically Safe Outputs

Figure 5-12 Service Tools menu: Variables

* If Volume Flow Type = GSV,GSV variables are displayed.

On-Line Menu >3 Service Tools >

2 Variables

Process Variables1 Mass Flow Rate2 Volume Flow Rate *3 Density4 Temperature

Mapped Variables1 PV Mass Flow Rate2 SV Volume Flow Rate3 TV Mass Flow Rate4 QV Volume Flow Rate

1

3

5

Totalizer Control1 All Totalizers2 Mass3 Volume *

All Totalizers1 Start Totalizers2 Stop Totalizers3 Reset All Totals4 Mass Total5 Volume Total *

1

2

Mass1 Mass Flow Rate2 Mass Total3 Mass Inventory4 Reset Total

Volume *1 Volume Flow Rate2 Volume Total3 Volume Inventory4 Reset Total

3

Variable SummaryMass Flow RateVolume Flow RateDensity

External Variables1 External Temperature2 External Pressure

4

2

6

Outputs1 Current2 Frequency/

DO State

Frequency Output1 Frequency2 Present Freq Output

Current1 Current2 PV AO3 PV % Range

1

2

Additionaloptions

Configuration and Use Manual Supplement 67

Model 1700 Transmitters with Intrinsically Safe Outputs

Mo

del 2200S

Mo

del 2500 C

IOM

od

el 2400S A

NM

od

el 1700 IS

Figure 5-13 Service Tools menu: Maintenance

On-Line Menu >3 Service Tools >

4 Maintenance

Routine Maintenance1 Trim mA Output2 Meter Verification *

1

Zero Calibration1 Mass Flow Rate2 Volume Flow Rate3 Zero Time4 Zero Value5 Standard Deviation6 Perform Auto Zero7 Restore Factory Zero

Density Calibration1 Density2 Dens Pt1 (Air)3 Dens Pt2 (Water)4 Dens Pt3 T-Series5 Dens Pt4 T-Series6 Flowing Dens (FD)7 Flowing Dens (FD)

2

Temperature Calibration1 Temperature2 Temp Cal Factor

4

3

Meter Verification **1 Run Meter Verification2 View Test Results3 Schedule Meter

Verification

2

* Displayed only if meter verificationis enabled.

** Displayed only if Smart MeterVerification is enabled. For earlierversions, the Meter VerificationMethod is launched.

5

Diagnostic Variables1 Sensor Model2 Drive Gain3 Input Voltage4 LPO Amplitude5 RPO Amplitude6 Board Temperature7 Tube Frequency8 Live Zero

68 Micro Motion Menu Maps for HART Device Descriptions

Model 1700 Transmitters with Intrinsically Safe Outputs

Figure 5-14 Service Tools menu: Simulate

On-Line Menu >3 Service Tools >

5 Simulate

1

Simulate Sensor1 Simulate Mode2 Mass Flow Rate3 Density4 Temperature

2

Simulate Outputs1 mA Output Loop Test2 Frequency Output Test3 Discrete Output Test

Configuration and Use Manual Supplement 69

Mo

del 2200S

Mo

del 2500 C

IOM

od

el 2400S A

NM

od

el 1700 IS

Chapter 6Model 2200S Transmitters

Figure 6-1 On-Line menu

Configure

1 Manual Setup2 Alert Setup

Service Tools

1 Alerts2 Variables3 Trends4 Maintenance5 Simulate

On-Line Menu

2 3

Overview

1 Check Status2 Primary Purpose Variables3 Shortcuts

1

70 Micro Motion Menu Maps for HART Device Descriptions

Model 2200S Transmitters

Figure 6-2 Overview menu

Check Status

1 Refresh Alerts2 Dev Status:3 Comm Status:

Primary Purpose Variables

Mass Flow RateVolume Flow RateDensity

Shortcuts

1 Device Information2 Totalizer Control3 Zero Calibration4 Variables5 Trends

1 32

Device Information

1 Product Information2 Mat. of Construction

1

Product Information

1 Device Tag2 Model3 Xmtr Software Rev4 Option Board5 ETO Number6 Final Asmbly Num7 Sensor Serial Num8 HART DD Information

On-Line Menu >1 Overview

1

Configuration and Use Manual Supplement 71

Model 2200S Transmitters

Mo

del 2200S

Mo

del 2500 C

IOM

od

el 2400S A

NM

od

el 1700 IS

Figure 6-3 Configure menu: Top level

Manual Setup

1 Characterize2 Measurements3 Display4 Inputs/Outputs5 Info Parameters6 Communications

21

On-Line Menu >1 Configure

Alert Setup

1 Fault Timeout2 mAO Fault Action3 Fault Level (Direct)4 Flt Lvl (Adapter-Barrier)5 Comm Fault Action6 Set Alert Severity7 View Alert Severity

72 Micro Motion Menu Maps for HART Device Descriptions

Model 2200S Transmitters

Figure 6-4 Configure menu: Manual Setup

On-Line Menu >2 Configure >

1 Manual Setup

Characterize

1 Sensor Type2 Sensor TagParameters

Measurements

1 Flow2 Density3 Temperature4 Set Up Special Units5 Set Up External Compensation6 Set Up GSV

1 2

Display

1 Display Variable MenuFeatures2 Offline Variable MenuFeatures3 Set Up Display Variables4 Set Up Decimal Places

3

Inputs/Outputs

1 Set Up mA Output2 Map Variables

Info Parameters

1 Transmitter Info2 Sensor Information

Communications

1 HART Address2 Device Tag3 Device Identification4 Set Up Burst Mode

654

Configuration and Use Manual Supplement 73

Model 2200S Transmitters

Mo

del 2200S

Mo

del 2500 C

IOM

od

el 2400S A

NM

od

el 1700 IS

Figure 6-5 Configure menu: Manual Setup: Characterize

On-Line Menu >2 Configure >

1 Manual Setup >1 Characterize

Characterize

1 Sensor Type2 Sensor Tag Parameters

Flow Parameters

1 Flow FCF2 FTG3 FFQ

21

Density Parameters

1 D12 D23 DT4 DTG5 K16 K27 FD8 DFQ19 DFQ2

Sensor Tag Parameters

1 Flow Parameters2 Density Parameters

2

Sensor Type

1 Curved Tube2 Straight Tube

1

Sensor Type Straight TubeCurved Tube

Sensor Tag Parameters

1 FlowCal2 D13 D24 TC5 K16 K27 FD

74 Micro Motion Menu Maps for HART Device Descriptions

Model 2200S Transmitters

Figure 6-6 Configure menu: Manual Setup: Measurements

On-Line Menu >2 Configure >

1 Manual Setup >2 Measurements

Flow

1 Flow Direction2 Flow Damping3 Mass Flow Cutoff4 Mass Flow Unit5 Volume Flow Cutoff *6 Volume Flow Unit *7 Mass Factor8 Volume Factor

Density

1 Density Unit2 Density Damping3 Density Cutoff4 Density Factor5 Slug High Limit6 Slug Low Limit7 Slug Duration

Temperature

1 Temperature Unit2 Temp Damping

1

2

3

4

Set Up Special Units

1 Mass Special Units2 Volume Special Units

Set Up External Compensation

1 Pressure Unit2 Enable Press Comp3 Flow Cal Pressure4 Static Pressure5 Flow Press Factor6 Dens Press Factor7 Enable Ext Temp8 Static Temperature

5

Set Up GSV

1 Volume Flow Type2 Gas Density3 GSV Flow Cutoff4 GSV Flow Unit

6

* Displayed only if Volume Flow Type =Liquid. Menu numbers are adjusted asrequired.

Configuration and Use Manual Supplement 75

Model 2200S Transmitters

Mo

del 2200S

Mo

del 2500 C

IOM

od

el 2400S A

NM

od

el 1700 IS

Figure 6-7 Configure menu: Manual Setup: Display

On-Line Menu >2 Configure >

1 Manual Setup >3 Display

Display Variable MenuFeatures

1 Totalizer Reset2 Start/Stop Totals3 Auto Scroll4 Scroll Time *5 Update Period

Offline Variable MenuFeatures

1 Offline Menu2 Alert Menu3 Acknowledge All4 Offline Passcode5 Alert Passcode

1

2

4

Set Up Decimal Places

1 For Process Variables2 For Totalizer Variables3 For Diagnostic Variables

Set Up Display Variables

1 Display Variables (1-5)2 Display Variables (6-10)3 Display Variables (11-15)

3

* Displayed only if Auto Scroll is enabled.Menu numbers are adjusted asrequired.

76 Micro Motion Menu Maps for HART Device Descriptions

Model 2200S Transmitters

Figure 6-8 Configure menu: Manual Setup: Inputs/Outputs

On-Line Menu >2 Configure >

1 Manual Setup >4 Inputs/Outputs

Set Up mA Output

1 Enter mA Msmt Pt2 Primary Variable3 mA Output Settings4 mA Fault Settings

1 2

mA Fault Settings

1 MAO Fault Action2 Fault Level (Direct)3 Flt Lvl (Adapter-Barrier)

mA Output Settings

1 PV LRV2 PV URV3 PV Min Span4 PV LSL5 PV USL6 PV MAO Cutoff7 PV Added Damping

3

Map Variables

1 Primary Variable2 Secondary Variable3 Third Variable4 Fourth Variable

4

Configuration and Use Manual Supplement 77

Model 2200S Transmitters

Mo

del 2200S

Mo

del 2500 C

IOM

od

el 2400S A

NM

od

el 1700 IS

Figure 6-9 Service Tools menu: Top level

On-Line Menu >3 Service Tools

Alerts

1 Refresh Alerts2 Alert Name3 Additional Information forAbove

Variables

1 Variable Summary2 Process Variables3 Mapped Variables4 External Variables5 Totalizer Control6 Outputs

1 2

Trends

1 Process Variables2 Diagnostic Variables

3

Maintenance

1 Routine Maintenance2 Zero Calibration3 Density Calibration4 Temperature Calibration5 Diagnostic Variables

Simulate

1 Simulate Outputs2 Simulate Sensor

54

78 Micro Motion Menu Maps for HART Device Descriptions

Model 2200S Transmitters

Figure 6-10 Service Tools menu: Variables

On-Line Menu >3 Service Tools >

2 Variables

Process Variables

1 Mass Flow Rate2 Volume Flow Rate *3 Density4 Temperature

Mapped Variables

1 PV2 SV3 TV4 QV

1

3

5

Totalizer Control

1 All Totalizers2 Mass3 Volume *

Outputs

1 Current (Direct)2 Current (Adapter-Barrier)3 PV MAO (Direct)4 PV MAO (Adapter-Barrier)5 PV % Range6 Fault Level (Direct)7 Flt Lvl (Adapter-Barrier)

6

All Totalizers

1 Start Totalizers2 Stop Totalizers3 Reset All Totals4 Mass Total5 Volume Total *

1

2

Mass

1 Mass Flow Rate2 Mass Total3 Mass Inventory4 Reset Total

Volume *

1 Volume Flow Rate2 Volume Total3 Volume Inventory4 Reset Total

3

Variable Summary

External Variables

1 External Temperature2 External Pressure

4

2

* If Volume Flow Type = GSV, GSVvariables are displayed.

Configuration and Use Manual Supplement 79

Model 2200S Transmitters

Mo

del 2200S

Mo

del 2500 C

IOM

od

el 2400S A

NM

od

el 1700 IS

Figure 6-11 Service Tools menu: Maintenance

On-Line Menu >3 Service Tools >

4 Maintenance

Routine Maintenance

1 Trim mA Output

1

Zero Calibration

1 Mass Flow Rate2 Volume Flow Rate3 Zero Time4 Zero Value5 Standard Deviation6 Perform Auto Zero

Density Calibration

1 Density2 Dens Pt1 (Air)3 Dens Pt2 (Water)4 Dens Pt3 T-Series5 Dens Pt4 T-Series6 Flowing Dens (FD)

2

Temperature Calibration

1 Temperature2 Temp Cal Factor

4 5

Diagnostic Variables

1 Sensor Model2 Drive Gain3 Input Voltage4 LPO Amplitude5 RPO Amplitude6 Board Temperature7 Tube Frequency8 Live Zero

3

80 Micro Motion Menu Maps for HART Device Descriptions

Model 2200S Transmitters

Figure 6-12 Service Tools menu: Simulate

Simulate

1 Simulate Outputs2 Simulate Sensor

Simulate Outputs

1 mA Output Loop Test

On-Line Menu >3 Service Tools >

5 Simulate

1

Simulate Sensor

1 Simulation Mode2 Mass Flow Rate3 Density4 Temperature

2

Configuration and Use Manual Supplement 81

Mo

del 2200S

Mo

del 2500 C

IOM

od

el 2400S A

NM

od

el 1700 IS

Chapter 7Model 2400S Transmitters with Analog Outputs

Figure 7-1 On-Line menu

Configure

1 Manual Setup2 Alert Setup

Service Tools

1 Alerts2 Variables3 Trends4 Maintenance5 Simulate

On-Line Menu

2 3

Overview

1 Check Status2 Primary Purpose Variables3 Shortcuts

1

82 Micro Motion Menu Maps for HART Device Descriptions

Model 2400S Transmitters with Analog Outputs

Figure 7-2 Overview menu

Primary Purpose Variables

Mass Flow RateVolume Flow RateDensity

Shortcuts

1 Device Information2 Totalizer Control3 Zero Calibration4 Variables5 Trends6 Meter Verification *

1

On-Line Menu >1 Overview

3

Device Information

1 Product Information2 Mat. of construction3 Licenses

1

Product Information

1 Tag2 Model3 Xmtr Software Rev4 Option Board5 ETO Number6 Final Asmbly Num7 Sensor Serial Num8 HART DD information

Check Status

1 Refresh Alerts2 Dev Status:3 Comm Status:

2

1

* Displayed only if meterverification is enabled.

Configuration and Use Manual Supplement 83

Model 2400S Transmitters with Analog Outputs

Mo

del 2200S

Mo

del 2500 C

IOM

od

el 2400S A

NM

od

el 1700 IS

Figure 7-3 Configure menu: Top level

Figure 7-4 Configure menu: Manual Setup

Manual Setup

1 Characterize2 Measurements3 Display4 Inputs/Outputs5 Info Parameters6 Communications

21

On-Line Menu >1 Configure

Alert Setup

1 I/O Fault Actions2 Alert Severity3 Discrete Events

On-Line Menu >2 Configure >

1 Manual Setup

Characterize

1 Sensor Type2 Sensor TagParameters

Measurements

1 Flow2 Density3 Temperature4 LD Optimization5 Special Units6 External Pressure/Temperature7 GSV

1 2

Display

1 Language2 Display Variable MenuFeatures3 Offline Variable MenuFeatures4 Backlight5 Display Variables6 Decimal Places

3

Inputs/Outputs

1 Channels2 mA Output3 Frequency Output4 Discrete Output5 Communications6 Variable Mapping

Info Parameters

1 Transmitter Info2 Sensor Information

Communications

1 HART Address2 Tag3 Device Identification4 Set Up Burst Mode

654

84 Micro Motion Menu Maps for HART Device Descriptions

Model 2400S Transmitters with Analog Outputs

Figure 7-5 Configure menu: Manual Setup: Characterize

On-Line Menu >2 Configure >

1 Manual Setup >1 Characterize

Characterize

1 Sensor Type2 Sensor Tag Parameters

Flow Parameters

1 Flow FCF2 FTG3 FFQ

21

Density Parameters

1 D12 D23 DT4 DTG5 K16 K27 FD8 DFQ19 DFQ2

Sensor Tag Parameters

1 Flow Parameters2 Density Parameters

2

Sensor Type

1 Curved Tube2 Straight Tube

1

Sensor Type Straight TubeCurved Tube

Sensor Tag Parameters

1 FlowCal2 D13 D24 TC5 K16 K27 FD

Configuration and Use Manual Supplement 85

Model 2400S Transmitters with Analog Outputs

Mo

del 2200S

Mo

del 2500 C

IOM

od

el 2400S A

NM

od

el 1700 IS

Figure 7-6 Configure menu: Manual Setup: Measurements

On-Line Menu >2 Configure >

1 Manual Setup >2 Measurements

Flow

1 Flow Direction2 Flow Damping3 Mass Flow Unit4 Mass Flow Cutoff5 Volume Flow Unit *6 Volume Flow Cutoff *7 Mass Factor8 Volume Factor

Density

1 Density Unit2 Density Damping3 Density Cutoff4 Density Factor5 Slug High Limit6 Slug Low Limit7 Slug Duration

Temperature

1 Temperature Unit2 Temp Damping

1

2

3

4

Special Units

1 Mass Special Units2 Volume Special Units

External Pressure/Temperature

1 Pressure2 Temperature3 External Polling

6

GSV

1 Volume Flow Type2 Gas Ref Density3 GSV Cutoff4 GSV Flow Unit5 Gas Density Unit

* Displayed only if Volume Flow Type = Liquid.Menu numbers are adjusted as required.

7

86 Micro Motion Menu Maps for HART Device Descriptions

Model 2400S Transmitters with Analog Outputs

Figure 7-7 Configure menu: Manual Setup: Measurements (continued)

On-Line Menu >2 Configure >

1 Manual Setup >2 Measurements >

7 External Pressure/Temperature

Pressure

1 Pressure Unit2 Pressure Compensation3 Compensation Pressure4 Flow Cal Pressure5 Flow Press Factor6 Dens Press Factor

Temperature

1 Temperature Unit2 External Temperature3 Correction Temperature

External Polling

1 Poll Control2 Ext Dev Tag 13 Polled Variable 14 Ext Dev Tag 25 Polled Variable 2

31 2

Configuration and Use Manual Supplement 87

Model 2400S Transmitters with Analog Outputs

Mo

del 2200S

Mo

del 2500 C

IOM

od

el 2400S A

NM

od

el 1700 IS

Figure 7-8 Configure menu: Manual Setup: Display

On-Line Menu >2 Configure >

1 Manual Setup >3 Display

Display Variable Menu Features

1 Totalizer Reset2 Start/Stop Totals3 Auto Scroll4 Scroll Time *5 Refresh Rate6 Status LED Blinking

Offline Variable Menu Features

1 Offline Menu2 Alert Menu3 Acknowledge All4 Offline Passcode5 Alert Passcode

2

3

4

6

Display Variables

1 Display Variables (1-5)2 Display Variables (6-10)3 Display Variables (11-15)

Decimal Places

1 For Process Variables2 For Totalizer Variables3 For Diagnostic Variables

Backlight

1 Control2 Intensity (0-63)

5

* Displayed only if Auto Scroll is enabled.Menu numbers are adjusted asrequired.

88 Micro Motion Menu Maps for HART Device Descriptions

Model 2400S Transmitters with Analog Outputs

Figure 7-9 Configure menu: Manual Setup: Inputs/Outputs

On-Line Menu >2 Configure >

1 Manual Setup >4 Inputs/Outputs

mA Output

1 Primary Variable2 mA Output Settings3 mA Fault Settings

Frequency Output

1 FO Settings2 FO Fault Parameters3 FO Scaling

3

Milliamp Fault Settings

1 MAO Fault Action2 MAO Fault Level

mA Output Settings

1 PV LRV2 PV URV3 PV Min Span4 PV LSL5 PV USL6 PV AO Cutoff7 PV Added Damping

FO Settings

1 Third Variable2 Max Pulse Width3 FO Polarity

FO Fault Parameters

1 Third Variable2 FO Fault Action3 FO Fault Level

FO Scaling *

1 FO Scaling Method2 TV Frequency Factor3 TV Rate Factor4 Set FO Scaling

3

2

3

1

2

Channels

1 Channel A2 Channel A Power3 Channel B4 Channel B Power

1 2

* Options vary depending onFO Scaling Method.

Additionaloptions

Configuration and Use Manual Supplement 89

Model 2400S Transmitters with Analog Outputs

Mo

del 2200S

Mo

del 2500 C

IOM

od

el 2400S A

NM

od

el 1700 IS

Figure 7-10 Configure menu: Manual Setup: Inputs/Outputs (continued)

On-Line Menu >2 Configure >

1 Manual Setup >4 Inputs/Outputs

5

Discrete Output

1 DO Assignment2 DO Polarity3 DO Fault Action4 Flow Switch Source5 Flow Switch Setpoint

Communications

1 HART Address2 Tag3 Device Identification4 Set Up Burst Mode5 mA Output Mode

4

Set Up Burst Mode

1 Burst Mode2 Burst Option3 Field Device Var 14 Field Device Var 25 Field Device Var 36 Field Device Var 4

4

90 Micro Motion Menu Maps for HART Device Descriptions

Model 2400S Transmitters with Analog Outputs

Figure 7-11 Configure menu: Alert Setup

On-Line Menu >2 Configure >

2 Alert Setup

I/O Fault Action

1 MAO Fault Action2 MAO Fault Level3 FO Fault Action4 FO Fault Level5 Comm Fault Action

Alert Severity

1 Fault Timeout2 Set Alert Severity3 View Alert Severity

Discrete Events

1 Discrete Event 12 Discrete Event 23 Discrete Event 34 Discrete Event 45 Discrete Event 56 Assign Discrete Action7 Read Discrete Action8 Review Discrete Actions

1 2 3

1, 2, 3, 4, 5

Discrete Event x

1 Discrete Event Var2 Discrete Event Type3 Setpoint A4 Setpoint B

Configuration and Use Manual Supplement 91

Model 2400S Transmitters with Analog Outputs

Mo

del 2200S

Mo

del 2500 C

IOM

od

el 2400S A

NM

od

el 1700 IS

Figure 7-12 Service Tools menu: Top level

Simulate

1 Simulate Outputs2 Simulate Sensor

On-Line Menu >3 Service Tools

Alerts

1 Refresh Alerts2 Alert Name3 Additional Information forAbove

Variables

1 Variable Summary2 Process Variables3 Mapped Variables4 External Variables5 Totalizer Control6 Outputs

1 2

Trends

1 Process Variables2 Diagnostic Variables

3

Maintenance

1 Routine Maintenance2 Zero Calibration3 Density Calibration4 Temperature Calibration5 Diagnostic Variables6 Modbus Data

54

92 Micro Motion Menu Maps for HART Device Descriptions

Model 2400S Transmitters with Analog Outputs

Figure 7-13 Service Tools menu: Variables

On-Line Menu >3 Service Tools >

2 Variables

Process Variables

1 Mass Flow Rate2 Volume Flow Rate *3 Density4 Temperature

Mapped Variables

1 PV Mass Flow Rate2 SV Mass Flow Rate3 TV Mass Flow Rate4 QV Mass Flow Rate

1

3

5

Totalizer Control

1 All Totalizers2 Mass3 Volume *

All Totalizers

1 Start Totalizers2 Stop Totalizers3 Reset All Totals4 Mass Total5 Volume Total *

1

2

Mass

1 Mass Flow Rate2 Mass Total3 Mass Inventory4 Reset Total

Volume *

1 Volume Flow Rate2 Volume Total3 Volume Inventory4 Reset Total

3

Variable Summary

External Variables

1 External Temperature2 External Pressure

4

2

* If Volume Flow Type =GSV, GSV variables aredisplayed.

Outputs

1 mA Output2 Frequency Output3 DO State

Frequency Output

1 Present Freq Gauge2 Present Freq Output

mA Output

1 Present MAO Gauge2 PV MAO3 PV % Range

6

1 2

Configuration and Use Manual Supplement 93

Model 2400S Transmitters with Analog Outputs

Mo

del 2200S

Mo

del 2500 C

IOM

od

el 2400S A

NM

od

el 1700 IS

Figure 7-14 Service Tools menu: Maintenance

On-Line Menu >3 Service Tools >

4 Maintenance

Routine Maintenance

1 mA Output Trim2 Meter Verification *

1

Zero Calibration

1 Mass Flow Rate2 Volume Flow Rate3 Zero Time4 Zero Value5 Standard Deviation6 Perform Auto Zero7 Restore Factory Zero

Density Calibration

1 Mass Flow Rate2 Density3 Dens Pt1 (Air)4 Dens Pt2 (Water)5 Dens Pt3 T-Series6 Dens Pt4 T-Series7 Flowing Dens (FD)

2

Temperature Calibration

1 Temperature2 Temp Cal Factor3 Offset Calibration4 Slope Calibration

4 5

Diagnostic Variables

1 Sensor Model2 Drive Gain3 LPO Amplitude4 RPO Amplitude5 Tube Frequency6 Live Zero7 Core Processor InputVoltage8 Board Temperature9 Power On Time

3

Meter Verification **

1 Run Meter Verification2 View Test Results3 Schedule MeterVerification

2

* Displayed only if meter verification isenabled.

** Displayed only if Smart MeterVerification is enabled. For earlierversions, the Meter VerificationMethod is launched.

Modbus Data

1 Read Modbus Data2 Write Modbus Data

6

94 Micro Motion Menu Maps for HART Device Descriptions

Model 2400S Transmitters with Analog Outputs

Figure 7-15 Service Tools menu: Simulate

Simulate

1 Simulate Outputs2 Simulate Sensor

Simulate Outputs

1 mA Output Loop Test2 Frequency Output Test3 Discrete Output Test

On-Line Menu >3 Service Tools >

5 Simulate

1

Simulate Sensor

1 Simulation Primary Purpose Variables2 Mass Flow Rate3 Density4 Temperature

2

Configuration and Use Manual Supplement 95

Mo

del 2200S

Mo

del 2500 C

IOM

od

el 2400S A

NM

od

el 1700 IS

Chapter 8Model 2500 Transmitters with Configurable Input/Outputs

Figure 8-1 On-Line menu

Configure

1 Manual Setup2 Alert Setup

Service Tools

1 Alerts2 Variables3 Trends4 Maintenance5 Simulate

On-Line Menu

2 3

Overview

1 Check Status2 Primary Purpose Variables3 Shortcuts

1

96 Micro Motion Menu Maps for HART Device Descriptions

Model 2500 Transmitters with Configurable Input/Outputs

Figure 8-2 Overview menu

Check Status

1 Refresh Alerts2 Dev Status:3 Comm Status:

Primary Purpose Variables

Mass Flow RateVolume Flow RateDensity

Shortcuts

1 Device Information2 Totalizer Control3 Zero Calibration4 Variables5 Trends6 Meter Verification *

1 32

Device Information

1 Product Information2 Mat. of Construction3 Licenses4 Weights and Measures **5 Checksums

Product Information

1 Tag2 Model3 Xmtr Software Rev4 CP Software Rev5 Option Board6 ETO Number7 Final Asmbly Num8 Sensor Serial Num9 HART DD Information

1

On-Line Menu >1 Overview

1

Weights and Measures

1 Cstdy Trnsfr Approval2 W&M Software Version

4

* Displayed only if meterverification is enabled.

** Displayed only if Weights &Measures application is enabled.Menu numbers are adjusted asrequired.

Configuration and Use Manual Supplement 97

Model 2500 Transmitters with Configurable Input/Outputs

Mo

del 2200S

Mo

del 2500 C

IOM

od

el 2400S A

NM

od

el 1700 ISM

od

el 2200SM

od

el 2500 CIO

Mo

del 2400S

AN

Mo

del 1700 IS

Mo

del 2200S

Mo

del 2500 C

IOM

od

el 2400S A

NM

od

el 1700 ISM

od

el 2200SM

od

el 2500 CIO

Mo

del 2400S

AN

Mo

del 1700 IS

Figure 8-3 Configure menu: Top level

Figure 8-4 Configure menu: Manual Setup

Manual Setup

1 Characterize2 Measurements3 Inputs/Outputs4 Info Parameters5 Communications

21

On-Line Menu >1 Configure

Alert Setup

1 Configure Alerts2 Discrete Output3 Discrete Events

On-Line Menu >2 Configure >

1 Manual Setup

Characterize

1 Sensor Type2 Sensor TagParameters

Measurements

1 Flow2 Density3 Temperature4 Update Rate5 Set Up Special Units6 Set Up External Compensation7 Set Up Petroleum *8 Set Up GSV

1

Inputs/Outputs

1 Set Up Channels2 Set Up mA Output 13 Set Up mA Output 24 Set Up Frequency Output5 Set Up Discrete Output6 Set Up Discrete Input7 Set Up RS-485 Port8 Map Variables

3

Info Parameters

1 Transmitter Info2 Sensor Information

Communications

1 HART Address2 Tag3 Device Identification4 Device ID (CP)5 Set Up Burst Mode

4 5

2

* Displayed only if the petroleum measurementapplication is enabled on your transmitter. Menunumbers are adjusted as required.

98 Micro Motion Menu Maps for HART Device Descriptions

Model 2500 Transmitters with Configurable Input/Outputs

Figure 8-5 Configure menu: Manual Setup: Characterize

On-Line Menu >2 Configure >

1 Manual Setup >1 Characterize

Characterize

1 Sensor Type2 Sensor Tag Parameters

Flow Parameters

1 Flow FCF2 FTG3 FFQ

21

Density Parameters

1 D12 D23 DT4 DTG5 K16 K27 FD8 DFQ19 DFQ2

Sensor Tag Parameters

1 Flow Parameters2 Density Parameters

2

Sensor Type

1 Curved Tube2 Straight Tube

1

Sensor Type Straight TubeCurved Tube

Sensor Tag Parameters

1 FlowCal2 D13 D24 TC5 K16 K27 FD

Configuration and Use Manual Supplement 99

Model 2500 Transmitters with Configurable Input/Outputs

Mo

del 2200S

Mo

del 2500 C

IOM

od

el 2400S A

NM

od

el 1700 ISM

od

el 2200SM

od

el 2500 CIO

Mo

del 2400S

AN

Mo

del 1700 IS

Mo

del 2200S

Mo

del 2500 C

IOM

od

el 2400S A

NM

od

el 1700 ISM

od

el 2200SM

od

el 2500 CIO

Mo

del 2400S

AN

Mo

del 1700 IS

Figure 8-6 Configure menu: Manual Setup: Measurements

On-Line Menu >2 Configure >

1 Manual Setup >2 Measurements

Flow

1 Flow Direction2 Flow Damping3 Mass Flow Cutoff4 Mass Flow Unit5 Volume Flow Cutoff *6 Volume Flow Unit *7 Mass Factor8 Volume Factor

Density

1 Density Unit2 Density Damping3 Density Cutoff4 Density Factor5 Slug High Limit6 Slug Low Limit7 Slug Duration

Temperature

1 Temperature Unit2 Temp Damping

1

2

3

5

Set Up Special Units

1 Mass Special Units2 Volume Special Units

Set Up External Compensation

1 Pressure Unit2 Enable Press Comp3 Flow Cal Pressure4 Static Pressure5 Flow Press Factor6 Dens Press Factor7 Enable Ext Temp8 External Temperature9 External Polling

6

Update Rate

1 Update Rate2 100 Hz Variable

4Set Up GSV

1 Volume Flow Type2 Gas Density3 Gas Vol Flow Cutoff4 Gas Vol Flow Unit5 Gas Density Unit

8

* Displayed only if Volume Flow Type = Liquid. Menunumbers are adjusted as required.

** Displayed only if the petroleum measurementapplication is enabled on your transmitter. Menunumbers are adjusted as required.

Set Up Petroleum **

1 2540 CTL Table type2 TEC3 Ref Temperature4 Petroleum Msmt Setup

7

100 Micro Motion Menu Maps for HART Device Descriptions

Model 2500 Transmitters with Configurable Input/Outputs

Figure 8-7 Configure menu: Manual Setup: Inputs/Outputs

On-Line Menu >2 Configure >

1 Manual Setup >3 Inputs/Outputs

Set Up mA Output 1

1 Primary Variable2 mA Output Settings3 MA01 Fault Settings

2 3

MA01 Fault Settings

1 MAO1 Fault Action2 MAO1 Fault Level

mA Output Settings

1 PV LRV2 PV URV3 PV Min Span4 PV LSL5 PV USL6 PV MAO1 Cutoff7 PV Added Damping

Additionaloptions

2

3

Set Up Channels

1 Channel A2 Channel A Power Source3 Channel B4 Channel B Power Source5 Channel C6 Channel C Power Source

1

Set Up mA Output 2

1 Secondary Variable2 mA Output Settings3 MA02 Fault Settings

MA02 Fault Settings

1 MAO2 Fault Action2 MAO2 Fault Level

mA Output Settings

1 SV LRV2 SV URV3 SV Min Span4 SV LSL5 SV USL6 SV MAO2 Cutoff7 SV Added Damping

2

3

Configuration and Use Manual Supplement 101

Model 2500 Transmitters with Configurable Input/Outputs

Mo

del 2200S

Mo

del 2500 C

IOM

od

el 2400S A

NM

od

el 1700 ISM

od

el 2200SM

od

el 2500 CIO

Mo

del 2400S

AN

Mo

del 1700 IS

Mo

del 2200S

Mo

del 2500 C

IOM

od

el 2400S A

NM

od

el 1700 ISM

od

el 2200SM

od

el 2500 CIO

Mo

del 2400S

AN

Mo

del 1700 IS

Figure 8-8 Configure menu: Manual Setup: Inputs/Outputs (continued)

Set Up Discrete Output

1 DO 1 Assignment2 DO 1 Polarity3 DO 1 Fault Action4 DO 2 Assignment5 DO 2 Polarity6 DO 2 Fault Action7 Flow Switch Source8 Flow Switch Setpoint9 Hysteresis (0.1-10.0)

Map Variables

1 Primary Variable2 Secondary Variable3 Third Variable4 Fourth Variable

5

On-Line Menu >2 Configure >

1 Manual Setup >3 Inputs/Outputs

Set Up Frequency Output

1 FO Settings2 FO Fault Parameters3 FO Scaling

FO Settings

1 Third Variable2 Max Pulse Width3 FO Polarity

FO Fault Parameters

1 Third Variable2 FO Fault Action3 FO Fault Level

FO Scaling *

1 FO Scaling Method2 Set FO Scaling

1

2

3

4

Set Up Discrete Input

1 DI Assignment2 DI Polarity

6

Set Up RS-485 Port

1 Protocol2 Baud Rate3 Parity4 Stop Bits5 Modbus Slave Address

7 8

* Options vary depending onFO Scaling Method.

102 Micro Motion Menu Maps for HART Device Descriptions

Model 2500 Transmitters with Configurable Input/Outputs

Figure 8-9 Configure menu: Alert Setup

On-Line Menu >2 Configure >

2 Alert Setup

Configure Alerts

1 Fault Timeout2 MAO1 Fault Settings3 MAO2 Fault Settings4 FO Fault Action5 FO Fault Level6 Comm Fault Action7 Set Alert Severity8 View Alert Severity

Discrete Output

1 DO 1 Assignment2 DO 1 Polarity3 DO 1 Fault Action4 DO 2 Assignment5 DO 2 Polarity6 DO 2 Fault Action7 Flow Switch Source8 Flow Switch Setpoint9 Hysteresis (0.1-10.0)

Discrete Events

1 Discrete Event 12 Discrete Event 23 Discrete Event 34 Discrete Event 45 Discrete Event 56 Assign Discrete Action7 Read Discrete Action8 Review Discrete Actions

1 2 3

1, 2, 3, 4, 5

Discrete Event x

1 Discrete Event Var2 Discrete Event Type3 Setpoint A4 Setpoint B

Configuration and Use Manual Supplement 103

Model 2500 Transmitters with Configurable Input/Outputs

Mo

del 2200S

Mo

del 2500 C

IOM

od

el 2400S A

NM

od

el 1700 ISM

od

el 2200SM

od

el 2500 CIO

Mo

del 2400S

AN

Mo

del 1700 IS

Mo

del 2200S

Mo

del 2500 C

IOM

od

el 2400S A

NM

od

el 1700 ISM

od

el 2200SM

od

el 2500 CIO

Mo

del 2400S

AN

Mo

del 1700 IS

Figure 8-10 Service Tools menu: Top level

On-Line Menu >3 Service Tools

Alerts

1 Refresh Alerts2 Alert Name3 Additional Information forAbove

Variables

1 Variable Summary2 Process Variables3 Mapped Variables4 External Variables5 Totalizer Control6 Variables (PM) *7 Variables (CM) **8 Outputs

1 2

Trends

1 Process Variables2 Diagnostic Variables

3

Maintenance

1 Routine Maintenance2 Zero Calibration3 Density Calibration4 Temperature Calibration5 Diagnostic Variables

Simulate

1 Simulate Outputs2 Simulate Sensor

54

* Displayed only if the petroleum measurementapplication is enabled on your transmitter. Menunumbers are adjusted as required.

** Displayed only if the concentration measurementapplication is enabled on your transmitter. Menunumbers are adjusted as required.

104 Micro Motion Menu Maps for HART Device Descriptions

Model 2500 Transmitters with Configurable Input/Outputs

Figure 8-11 Service Tools menu: Variables

On-Line Menu >3 Service Tools >

2 Variables

Process Variables

1 Mass Flow Rate2 Volume Flow Rate *3 Density4 Temperature

Mapped Variables

1 PV2 SV3 TV4 QV

1

3

5

Totalizer Control

1 All Totalizers2 Mass3 Volume *4 CM Volume at ReferenceTemperature5 CM Net Mass ***6 CM Net Volume ***7 Temp Corrected Volume (PM) **8 Custody Transfer *****

All Totalizers

1 Start Totalizers2 Stop Totalizers3 Reset All Totals4 Mass Total5 Volume Total *

1

2

Mass

1 Mass Flow Rate2 Mass Total3 Mass Inventory4 Reset Total

Volume *

1 Volume Flow Rate2 Volume Total3 Volume Inventory4 Reset Total

3

Variable Summary

External Variables

1 External Temperature2 External Pressure

4

2

* If Volume Flow Type = GSV, GSV variables aredisplayed.

** Displayed only if the petroleum measurementapplication is enabled on your transmitter. Menunumbers are adjusted as required.

*** Displayed only if the concentration measurementapplication is enabled on your transmitter. Menunumbers are adjusted as required.

**** Displayed only if the Weights & Measuresapplication is installed on your transmitter.

Additional options

Configuration and Use Manual Supplement 105

Model 2500 Transmitters with Configurable Input/Outputs

Mo

del 2200S

Mo

del 2500 C

IOM

od

el 2400S A

NM

od

el 1700 ISM

od

el 2200SM

od

el 2500 CIO

Mo

del 2400S

AN

Mo

del 1700 IS

Mo

del 2200S

Mo

del 2500 C

IOM

od

el 2400S A

NM

od

el 1700 ISM

od

el 2200SM

od

el 2500 CIO

Mo

del 2400S

AN

Mo

del 1700 IS

Figure 8-12 Service Tools menu: Variables (continued)

Outputs

1 Current (MA01)2 Current (MA02)3 Frequency4 DO State

On-Line Menu >3 Service Tools >

2 Variables

Variables (PM) *

1 Density at Reference Temperature2 Average Observed Density3 Volume Flow at Reference Temperature4 Average Observed Temperature5 CTL

Variables (CM) **

1 Standard Volume Flow Rate2 Standard Net Volume Flow Rate3 Net Mass Flow Rate4 Density at Reference5 Concentration6 Density (Fixed SG Units)

6

8

7

Frequency

1 Frequency2 Present Freq Output

1

Current (MAO1)

1 Current2 PV AO3 PV % Range

Current (MAO2)

1 Current2 SV AO3 SV % Range

32

* Displayed only if the petroleum measurementapplication is enabled on your transmitter. Menunumbers are adjusted as required.

** Displayed only if the concentration measurementapplication is enabled on your transmitter. Menunumbers are adjusted as required.

106 Micro Motion Menu Maps for HART Device Descriptions

Model 2500 Transmitters with Configurable Input/Outputs

Figure 8-13 Service Tools menu: Maintenance

On-Line Menu >3 Service Tools >

4 Maintenance

Routine Maintenance

1 Trim mA Output 12 Trim mA Output 23 Meter Verification *

1

Zero Calibration

1 Mass Flow Rate2 Volume Flow Rate3 Zero Time4 Zero Value5 Standard Deviation6 Perform Auto Zero7 Restore Factory Zero

Density Calibration

1 Density2 Dens Pt1 (Air)3 Dens Pt2 (Water)4 Dens Pt3 T-Series5 Dens Pt4 T-Series6 Flowing Dens (FD)

2

Temperature Calibration

1 Temperature2 Temp Cal Factor3 Offset Calibration4 Slope Calibration

4 5

Diagnostic Variables

1 Sensor Model2 Drive Gain3 Input Voltage4 LPO Amplitude5 RPO Amplitude6 Board Temperature7 Tube Frequency8 Live Zero9 Fld Verification Zero

3

Meter Verification **

1 Run Meter Verification2 View Test Results3 Schedule MeterVerification

3

* Displayed only if meter verification isenabled.

** Displayed only if Smart MeterVerification is enabled. For earlierversions, the Meter VerificationMethod is launched.

Configuration and Use Manual Supplement 107

Model 2500 Transmitters with Configurable Input/Outputs

Mo

del 2200S

Mo

del 2500 C

IOM

od

el 2400S A

NM

od

el 1700 ISM

od

el 2200SM

od

el 2500 CIO

Mo

del 2400S

AN

Mo

del 1700 IS

Mo

del 2200S

Mo

del 2500 C

IOM

od

el 2400S A

NM

od

el 1700 ISM

od

el 2200SM

od

el 2500 CIO

Mo

del 2400S

AN

Mo

del 1700 IS

Figure 8-14 Service Tools menu: Simulate

Simulate

1 Simulate Outputs2 Simulate Sensor

Simulate Outputs

1 mA Output 1 Loop Test2 mA Output 2 Loop Test3 Frequency Output Test4 Discrete Output 1 Test5 Discrete Output 2 Test

On-Line Menu >3 Service Tools >

5 Simulate

1

Simulate Sensor

1 Simulation Mode2 Mass Flow Rate3 Density4 Temperature

2

108 Micro Motion Menu Maps for HART Device Descriptions

Configuration and Use Manual Supplement 109

Mo

del 2700 IS

M

od

el 2700 CIO

Mo

del 2700 A

N

Chapter 9Model 2700 Transmitters with Analog Outputs

Figure 9-1 On-Line menu

Configure1 Manual Setup2 Alert Setup

Service Tools1 Alerts2 Variables3 Trends4 Maintenance5 Simulate

2

3

Overview1 Check Status2 Primary Purpose Variables3 Shortcuts

1

On-Line Menu

110 Micro Motion Menu Maps for HART Device Descriptions

Model 2700 Transmitters with Analog Outputs

Figure 9-2 Overview menu

Identification1 Tag2 Model3 Xmtr Serial Num4 Sensor Serial Num5 Date6 Descriptor7 Message

1

Revisions1 Universal2 Field Device3 DD Revision4 Transmitter Software5 CP Software6 ETO Number

Mat. of Construction1 Tube Wetted Mat.2 Tube Lining3 Sensor Flange

Check Status1 Refresh Alerts2 Dev Status:3 Comm Status:

Primary Purpose VariablesMass Flow RateVolume Flow RateDensity

Shortcuts1 Device Information2 Totalizer Control3 Zero Calibration4 Variables5 Trends6 Meter Verification *

13

1

2

3

On-Line Menu >1 Overview

Device Information1 Identification2 Revisions3 Mat. of Construction4 Licenses5 Weights and Measures **

* Displayed only if meterverification is enabled.

** Displayed only if the Weights& Measures application isenabled. Menu numbers areadjusted as required.

4Weights and Measures1 Cstdy Trnsfr Approval2 W&M Software Version

2

Configuration and Use Manual Supplement 111

Model 2700 Transmitters with Analog Outputs

Mo

del 2700 IS

M

od

el 2700 CIO

Mo

del 2700 A

N

Figure 9-3 Configure menu: Top level

Manual Setup1 Characterize2 Measurements3 Display4 Inputs/Outputs5 Info Parameters

2

1

Alert Setup1 I/O Fault Actions2 Alert Severity3 Discrete Events4 CM Alerts

On-Line Menu >1 Configure

112 Micro Motion Menu Maps for HART Device Descriptions

Model 2700 Transmitters with Analog Outputs

Figure 9-4 Configure menu: Manual Setup

Characterize1 Sensor Type2 Sensor Tag Parameters

Measurements1 Flow2 Density3 Temperature4 Update Rate5 LD Optimization6 Special Units7 External Pressure/Temperature8 Petroleum Measurement *9 GSV10 Conc Measure (CM) **

1

2

Inputs/Outputs1 Channels2 mA Output3 Frequency Output4 Discrete Output5 Communications6 Variable Mapping

3

Info Parameters1 Transmitter Info2 Sensor Information

Communications1 HART Address2 Tag3 Device Identification4 Dev ID (CP)5 Set Up Burst Mode6 mA Output Mode7 Set Up RS-485 Port

5

Display1 Language2 Display Variable Menu Features3 Offline Variable Menu Features4 Backlight5 Display Variables6 Decimal Places

4

Transmitter Info1 Tag2 Xmtr Serial Num3 Message4 Descriptor5 Date6 Write Protect

Sensor Information1 Sensor Type2 Sensor Serial Num3 Tube Wetted Mat.4 Tube Lining5 Sensor Flange

1

2

On-Line Menu >2 Configure >

1 Manual Setup

5

* Displayed only if petroleummeasurement is enabled on yourtransmitter. Menu numbers areadjusted as required.

** Displayed only if concentrationmeasurement is enabled on yourtransmitter.

Configuration and Use Manual Supplement 113

Model 2700 Transmitters with Analog Outputs

Mo

del 2700 IS

M

od

el 2700 CIO

Mo

del 2700 A

N

Figure 9-5 Configure menu: Manual Setup: Characterize

On-Line Menu >2 Configure >

1 Manual Setup >1 Characterize

21

2

1

Sensor Type Straight TubeCurved Tube

Density Parameters1 D12 D23 DT4 DTG5 K16 K27 FD8 DFQ19 DFQ2

Flow Parameters1 Flow FCF2 FTG3 FFQ

Sensor Tag Parameters1 Flow Parameters2 Density Parameters

Sensor Tag Parameters1 FlowCal2 D13 D24 TC5 K16 K27 FD

Sensor Type1 Curved Tube2 Straight Tube

114 Micro Motion Menu Maps for HART Device Descriptions

Model 2700 Transmitters with Analog Outputs

Figure 9-6 Configure menu: Manual Setup: Measurements

On-Line Menu >2 Configure >

1 Manual Setup >2 Measurements

Flow1 Flow Direction2 Flow Damping3 Mass Flow Unit4 Mass Flow Cutoff5 Volume Flow Unit *6 Volume Flow Cutoff *7 Mass Factor8 Volume Factor

Density1 Density Unit2 Density Damping3 Density Cutoff4 Density Factor5 Slug Duration6 Slug Low Limit7 Slug High Limit

Temperature1 Temperature Unit2 Temp Damping

1

2

3

6Special Units1 Mass Special Units2 Volume Special Units

* Displayed only if Volume Flow Type = Liquid.Menu numbers are adjusted as required.

Update Rate1 Update Rate2 100 Hz Variable

4

Additionaloptions

Configuration and Use Manual Supplement 115

Model 2700 Transmitters with Analog Outputs

Mo

del 2700 IS

M

od

el 2700 CIO

Mo

del 2700 A

N

Figure 9-7 Configure menu: Manual Setup: Measurements (continued)

On-Line Menu >2 Configure >

1 Manual Setup >2 Measurements

1

2

3

87

9

10

Pressure1 Pressure Unit2 Pressure Compensation3 Compensation Pressure4 Flow Cal Pressure5 Flow Press Factor6 Dens Press Factor

Temperature1 Temperature Unit2 External Temperature3 Correction Temperature

External Polling1 Poll Control2 Ext Dev Tag 13 Polled Variable 14 Ext Dev Tag 25 Polled Variable 2

External Pressure/Temperature1 Pressure2 Temperature3 External Polling

Petroleum Measurement *1 2540 CTL Table type2 TEC3 Ref Temperature4 Petroleum Msmt Setup

GSV1 Volume Flow Type2 Gas Ref Density3 GSV Cutoff4 GSV Flow Unit5 Gas Density Unit

Conc Measure (CM) **1 CM Configuration2 Matrix Configuration3 Enter Matrix4 Trim CM Process

Variables5 Reset Matrix Data

* Displayed only if petroleummeasurement is enabled on yourtransmitter. Menu numbers areadjusted as required.

** Displayed only if concentrationmeasurement is enabled on yourtransmitter.

116 Micro Motion Menu Maps for HART Device Descriptions

Model 2700 Transmitters with Analog Outputs

Figure 9-8 Configure menu: Manual Setup: Display

On-Line Menu >2 Configure >

1 Manual Setup >3 Display

LanguageEnglishGermanFrenchSpanish

Display Variable Menu Features1 Totalizer Reset2 Start/Stop Totals3 Auto Scroll4 Scroll Time *5 Refresh Rate6 Status LED Blinking

Offline Variable Menu Features1 Offline Menu2 Alert Menu3 Acknowledge All4 Offline Passcode5 Alert Passcode6 Offline Passcode

1

2

3

5

Display Variables1 Display Variables (1-5)2 Display Variables (6-10)3 Display Variables (11-15)

Decimal Places1 For Process Variables2 For Totalizer Variables3 For Diagnostic Variables

6

Backlight1 Control2 Intensity (0-63)

4

* Displayed only if Auto Scroll isenabled. Menu numbers areadjusted as required.

Configuration and Use Manual Supplement 117

Model 2700 Transmitters with Analog Outputs

Mo

del 2700 IS

M

od

el 2700 CIO

Mo

del 2700 A

N

Figure 9-9 Configure menu: Manual Setup: Inputs/Outputs

On-Line Menu >2 Configure >

1 Manual Setup >4 Inputs/Outputs

mA Output1 Primary Variable2 mA Output Settings3 mA Fault Settings

Frequency Output1 FO Settings2 FO Fault Parameters3 FO Scaling

2

3

Milliamp Fault Settings1 MAO Fault Action2 MAO Fault Level

mA Output Settings1 PV LRV2 PV URV3 PV Min Span4 PV LSL5 PV USL6 PV MAO Cutoff7 PV Added Damping

FO Settings1 Third Variable2 Max Pulse Width3 FO Polarity

FO Fault Parameters1 Third Variable2 FO Fault Action3 FO Fault Level

FO Scaling *1 FO Scaling Method2 TV Frequency Factor3 TV Rate Factor4 Set FO Scaling

2

3

1

1

2

3

* Options vary dependingon FO Scaling Method.

Additionaloptions

Channels1 Channel A2 Channel B3 Channel C

Channel B1 Frequency Output2 Discrete Output

2

118 Micro Motion Menu Maps for HART Device Descriptions

Model 2700 Transmitters with Analog Outputs

Figure 9-10 Configure menu: Manual Setup: Inputs/Outputs (continued)

Discrete Output1 DO Assignment2 DO Polarity3 DO Fault Action4 Flow Switch Source5 Flow Switch Setpoint6 Hysteresis (0.1-10.0)

46

Map Variables1 Primary Variable2 Secondary Variable3 Third Variable4 Fourth Variable

5

Communications1 HART Address2 Tag3 Device Identification4 Dev ID (CP)5 Set Up Burst Mode6 mA Output Mode7 Set Up RS-485 Port

Set Up Burst Mode1 Burst Mode2 Burst Option3 Field Device Var 14 Field Device Var 25 Field Device Var 36 Field Device Var 4

Set Up RS-485 Port1 Protocol2 Baud Rate3 Parity4 Stop Bits5 Modbus Slave Address

7

5

On-Line Menu >2 Configure >

1 Manual Setup >4 Inputs/Outputs

Configuration and Use Manual Supplement 119

Model 2700 Transmitters with Analog Outputs

Mo

del 2700 IS

M

od

el 2700 CIO

Mo

del 2700 A

N

Figure 9-11 Configure menu: Alert Setup

On-Line Menu >2 Configure >

2 Alert Setup

I/O Fault Action1 MAO Fault Action2 MAO Fault Level3 FO Fault Action4 FO Fault Level5 Comm Fault Action

Alert Severity1 Fault Timeout2 Set Alert Severity3 View Alert Severity

Discrete Events1 Discrete Event 12 Discrete Event 23 Discrete Event 34 Discrete Event 45 Discrete Event 56 Assign Discrete Action7 Read Discrete Action8 Review Discrete Actions

1

2

3

1, 2, 3, 4, 5 Discrete Event x1 Discrete Event Var2 Discrete Event Type3 Setpoint A4 Setpoint B

120 Micro Motion Menu Maps for HART Device Descriptions

Model 2700 Transmitters with Analog Outputs

Figure 9-12 Service Tools menu: Top level

On-Line Menu >3 Service Tools

Alerts1 Refresh Alerts

Alert NameAdditional Information

Variables1 Variable Summary2 Process Variables3 Mapped Variables4 External Variables5 Totalizer Control6 Variables (PM) *7 Variables (CM) **8 Outputs

1

2

Trends1 Process Variables2 Diagnostic Variables

3

Maintenance1 Routine Maintenance2 Zero Calibration3 Density Calibration4 Temperature Calibration5 Diagnostic Variables6 Modbus Data

Simulate1 Simulate Outputs2 Simulate Sensor

4

* Displayed only if petroleum measurement isenabled on your transmitter. Menu numbersare adjusted as required.

** Displayed only if concentration measurementis enabled on your transmitter. Menu numbersare adjusted as required.

5

Configuration and Use Manual Supplement 121

Model 2700 Transmitters with Analog Outputs

Mo

del 2700 IS

M

od

el 2700 CIO

Mo

del 2700 A

N

Figure 9-13 Service Tools menu: Variables

On-Line Menu >3 Service Tools >

2 Variables

Process Variables1 Mass Flow Rate2 Volume Flow Rate *3 Density4 Temperature

Mapped Variables1 PV Volume Flow Rate2 SV Volume Flow Rate3 TV Mass Flow Rate4 QV Volume Flow Rate

1

3

5

Totalizer Control1 All Totalizers2 Mass3 Volume *4 CM Volume at Reference Temperature5 CM Net Mass ***6 CM Net Volume ***7 Temp Corrected Volume (PM) **8 Custody Transfer ****

All Totalizers1 Start Totalizers2 Stop Totalizers3 Reset All Totals4 Mass Total5 Volume Total *

1

2

Mass1 Mass Flow Rate2 Mass Total3 Mass Inventory4 Reset Total

Volume *1 Volume Flow Rate2 Volume Total3 Volume Inventory4 Reset Total

3

Variable SummaryMass Flow RateVolume Flow Rate *Density

External Variables1 External Temperature2 External Pressure

4

2

* If Volume Flow Type = GSV, GSV variablesare displayed.

** Displayed only if petroleum measurement isenabled on your transmitter. Menu numbersare adjusted as required.

*** Displayed only if concentration measurementis enabled on your transmitter. Menunumbers are adjusted as required.

**** Displayed only if the Weights & Measuresapplication is enabled on your transmitter.

Additionaloptions

122 Micro Motion Menu Maps for HART Device Descriptions

Model 2700 Transmitters with Analog Outputs

Figure 9-14 Service Tools menu: Variables (continued)

Outputs1 mA Output (MA0)2 Frequency Output/

DO State

On-Line Menu >3 Service Tools >

2 Variables

Variables (PM) *1 Density at Reference

Temperature2 Average Observed Density3 Volume Flow at Reference

Temperature4 Average Observed

Temperature5 CTL

Variables (CM) **1 Standard Volume Flow Rate2 Standard Net Volume Flow

Rate3 Net Mass Flow Rate4 Density at Reference5 Concentration6 Density (Fixed SG Units)

6

8

7

Frequency Output1 Present Freq Gauge2 Present Freq Output

1

mA Output1 Present MAO Gauge2 PV MAO3 PV % Range

2

* Displayed only if petroleum measurement is enabledon your transmitter. Menu numbers are adjusted asrequired.

** Displayed only if concentration measurement isenabled on your transmitter. Menu numbers areadjusted as required.

Configuration and Use Manual Supplement 123

Model 2700 Transmitters with Analog Outputs

Mo

del 2700 IS

M

od

el 2700 CIO

Mo

del 2700 A

N

Figure 9-15 Service Tools menu: Maintenance

On-Line Menu >3 Service Tools >

4 Maintenance

Routine Maintenance1 Trim mA Output2 Meter Verification *

1

Zero Calibration1 Mass Flow Rate2 Volume Flow Rate3 Zero Time4 Zero Value5 Standard Deviation6 Perform Auto Zero7 Restore Factory Zero

Density Calibration1 Density2 Dens Pt1 (Air)3 Dens Pt2 (Water)4 Dens Pt3 T-Series5 Dens Pt4 T-Series6 Flowing Dens (FD)7 Flowing Dens (FD)

2

Temperature Calibration1 Temperature2 Temp Cal Factor

4

3

Meter Verification **1 Run Meter Verification2 View Test Results3 Schedule Meter

Verification

2

* Displayed only if meter verificationis enabled.

** Displayed only if Smart MeterVerification is enabled. For earlierversions, the Meter VerificationMethod is launched.

Additional1 Core Processor Input

Voltage2 Board Temperature3 Power On Time

8

Modbus Data1 Read Modbus Data2 Write Modbus Data

6

5

Diagnostic Variables1 Sensor Model2 Drive Gain3 LPO Amplitude4 RPO Amplitude5 Tube Frequency6 Live Zero7 Fld Verification Zero8 Additional

124 Micro Motion Menu Maps for HART Device Descriptions

Model 2700 Transmitters with Analog Outputs

Figure 9-16 Service Tools menu: Simulate

On-Line Menu >3 Service Tools >

5 Simulate

1

Simulate Sensor1 Simulate Primary Purpose Variables2 Mass Flow Rate3 Density4 Temperature

2

Simulate Outputs1 mA Output 1 Loop Test2 Frequency Output Test/

Discrete Output Test *

* Options vary dependingon channel settings.

Configuration and Use Manual Supplement 125

Mo

del 2700 IS

M

od

el 2700 CIO

Mo

del 2700 A

N

Chapter 10Model 2700 Transmitters with Intrinsically Safe Outputs

Figure 10-1 On-Line menu

Configure1 Manual Setup2 Alert Setup

Service Tools1 Alerts2 Variables3 Trends4 Maintenance5 Simulate

2

3

Overview1 Check Status2 Primary Purpose Variables3 Shortcuts

1

On-Line Menu

126 Micro Motion Menu Maps for HART Device Descriptions

Model 2700 Transmitters with Intrinsically Safe Outputs

Figure 10-2 Overview menu

Identification1 Tag2 Model3 Xmtr Serial Num4 Sensor Serial Num5 Date6 Descriptor7 Message

1

Revisions1 Universal2 Field Device3 DD Revision4 Transmitter Software5 CP Software6 ETO Number

Mat. of Construction1 Tube Wetted Mat.2 Tube Lining3 Sensor Flange

Check Status1 Refresh Alerts2 Dev Status:3 Comm Status:

Primary Purpose VariablesMass Flow RateVolume Flow RateDensity

Shortcuts1 Device Information2 Totalizer Control3 Zero Calibration4 Variables5 Trends6 Meter Verification *

13

21

2

3

On-Line Menu >1 Overview

Device Information1 Identification2 Revisions3 Mat. of Construction4 Licenses5 Weights and Measures **

* Displayed only if meterverification is enabled.

** Displayed only if the Weights& Measures application isenabled. Menu numbers areadjusted as required.

5Weights and Measures1 Cstdy Trnsfr Approval2 W&M Software Version

Configuration and Use Manual Supplement 127

Model 2700 Transmitters with Intrinsically Safe Outputs

Mo

del 2700 IS

M

od

el 2700 CIO

Mo

del 2700 A

N

Figure 10-3 Configure menu: Top level

Manual Setup1 Characterize2 Measurements3 Display4 Inputs/Outputs5 Info Parameters

2

1

Alert Setup1 I/O Fault Actions2 Alert Severity3 Discrete Events

On-Line Menu >1 Configure

128 Micro Motion Menu Maps for HART Device Descriptions

Model 2700 Transmitters with Intrinsically Safe Outputs

Figure 10-4 Configure menu: Manual Setup

Characterize1 Sensor Type2 Sensor Tag Parameters

Measurements1 Flow2 Density3 Temperature4 Update Rate5 LD Optimization6 Special Units7 External Pressure/Temperature8 Petroleum Measurement *9 GSV10 Conc Measure (CM) **

1

2

Inputs/Outputs1 Channels2 mA Output 13 mA Output 24 Frequency Output5 Discrete Output6 Communications7 Variable Mapping

3

Info Parameters1 Transmitter Info2 Sensor Information

Communications1 HART Address2 Tag3 Device Identification4 Dev ID (CP)5 Set Up Burst Mode6 mA Output Mode7 Set Up RS-485 Port

5

Display1 Language2 Display Variable Menu Features3 Offline Variable Menu Features4 Backlight5 Display Variables6 Decimal Places

4

Transmitter Info1 Tag2 Xmtr Serial Num3 Message4 Descriptor5 Date6 Write Protect

Sensor Information1 Sensor Type2 Sensor Serial Num3 Tube Wetted Mat.4 Tube Lining5 Sensor Flange

1

2

On-Line Menu >2 Configure >

1 Manual Setup

5

* Displayed only if petroleummeasurement is enabled on yourtransmitter. Menu numbers areadjusted as required.

** Displayed only if concentrationmeasurement is enabled on yourtransmitter.

Configuration and Use Manual Supplement 129

Model 2700 Transmitters with Intrinsically Safe Outputs

Mo

del 2700 IS

M

od

el 2700 CIO

Mo

del 2700 A

N

Figure 10-5 Configure menu: Manual Setup: Characterize

On-Line Menu >2 Configure >

1 Manual Setup >1 Characterize

21

2

1

Sensor Type Straight TubeCurved Tube

Density Parameters1 D12 D23 DT4 DTG5 K16 K27 FD8 DFQ19 DFQ2

Flow Parameters1 Flow FCF2 FTG3 FFQ

Sensor Tag Parameters1 Flow Parameters2 Density Parameters

Sensor Tag Parameters1 FlowCal2 D13 D24 TC5 K16 K27 FD

Sensor Type1 Curved Tube2 Straight Tube

130 Micro Motion Menu Maps for HART Device Descriptions

Model 2700 Transmitters with Intrinsically Safe Outputs

Figure 10-6 Configure menu: Manual Setup: Measurements

On-Line Menu >2 Configure >

1 Manual Setup >2 Measurements

Flow1 Flow Direction2 Flow Damping3 Mass Flow Unit4 Mass Flow Cutoff5 Volume Flow Unit *6 Volume Flow Cutoff *7 Mass Factor8 Volume Factor

Density1 Density Unit2 Density Damping3 Density Cutoff4 Density Factor5 Slug Duration6 Slug Low Limit7 Slug High Limit

Temperature1 Temperature Unit2 Temp Damping

1

2

3

6Special Units1 Mass Special Units2 Volume Special Units

* Displayed only if Volume Flow Type = Liquid.Menu numbers are adjusted as required.

** Displayed only if petroleum measurement isenabled on your transmitter. Menu numbersare adjusted as required.

Update Rate1 Update Rate2 100 Hz Variable

4

Additionaloptions

Configuration and Use Manual Supplement 131

Model 2700 Transmitters with Intrinsically Safe Outputs

Mo

del 2700 IS

M

od

el 2700 CIO

Mo

del 2700 A

N

Figure 10-7 Configure menu: Manual Setup: Measurements (continued)

On-Line Menu >2 Configure >

1 Manual Setup >2 Measurements

1

2

3

87

9

10

Pressure1 Pressure Unit2 Pressure Compensation3 Compensation Pressure4 Flow Cal Pressure5 Flow Press Factor6 Dens Press Factor

Temperature1 Temperature Unit2 External Temperature3 Correction Temperature

External Polling1 Poll Control2 Ext Dev Tag 13 Polled Variable 14 Ext Dev Tag 25 Polled Variable 2

External Pressure/Temperature1 Pressure2 Temperature3 External Polling

Petroleum Measurement *1 2540 CTL Table type2 TEC3 Ref Temperature4 Petroleum Msmt Setup

GSV1 Volume Flow Type2 Gas Ref Density3 GSV Cutoff4 GSV Flow Unit5 Gas Density Unit

Conc Measure (CM) **1 CM Configuration2 Matrix Configuration3 Enter Matrix4 Trim CM Process

Variables5 Reset Matrix Data

* Displayed only if petroleummeasurement is enabled on yourtransmitter. Menu numbers areadjusted as required.

** Displayed only if concentrationmeasurement is enabled on yourtransmitter.

132 Micro Motion Menu Maps for HART Device Descriptions

Model 2700 Transmitters with Intrinsically Safe Outputs

Figure 10-8 Configure menu: Manual Setup: Display

On-Line Menu >2 Configure >

1 Manual Setup >3 Display

LanguageEnglishGermanFrenchSpanish

Display Variable Menu Features1 Totalizer Reset2 Start/Stop Totals3 Auto Scroll4 Scroll Time *5 Refresh Rate6 Status LED Blinking

Offline Variable Menu Features1 Offline Menu2 Alert Menu3 Acknowledge All4 Offline Passcode5 Alert Passcode6 Offline Passcode

1

2

3

5

Display Variables1 Display Variables (1-5)2 Display Variables (6-10)3 Display Variables (11-15)

Decimal Places1 For Process Variables2 For Totalizer Variables3 For Diagnostic Variables

6

Backlight1 Control2 Intensity (0-63)

4

* Displayed only if Auto Scroll isenabled. Menu numbers areadjusted as required.

Configuration and Use Manual Supplement 133

Model 2700 Transmitters with Intrinsically Safe Outputs

Mo

del 2700 IS

M

od

el 2700 CIO

Mo

del 2700 A

N

Figure 10-9 Configure menu: Manual Setup: Inputs/Outputs

On-Line Menu >2 Configure >

1 Manual Setup >4 Inputs/Outputs

4

21

* Options vary dependingon FO Scaling Method.

Additionaloptions

Channels1 Channel A2 Channel B3 Channel C

Channel BFrequency OutputDiscrete Output

2

mA Output 11 Primary Variable2 mA Output Settings3 mA Fault Settings

MAO 1 Fault Settings1 MAO1 Fault Action2 MAO1 Fault Level

mA Output Settings1 PV LRV2 PV URV3 PV Min Span4 PV LSL5 PV USL6 PV MAO1 Cutoff7 PV Added Damping

2

3

mA Output 21 Secondary Variable2 mA Output Settings3 mA Fault Settings

MAO2 Fault Settings1 MAO2 Fault Action2 MAO2 Fault Level

mA Output Settings1 SV LRV2 SV URV3 SV Min Span4 SV LSL5 SV USL6 SV MAO2 Cutoff7 SV Added Damping

2

3

Frequency Output1 FO Settings2 FO Fault Parameters3 FO Scaling

FO Settings1 Third Variable2 Max Pulse Width3 FO Polarity

FO Fault Parameters1 Third Variable2 FO Fault Action3 FO Fault Level

FO Scaling *1 FO Scaling Method2 TV Frequency Factor3 TV Rate Factor4 Set FO Scaling

1

2

3

3

134 Micro Motion Menu Maps for HART Device Descriptions

Model 2700 Transmitters with Intrinsically Safe Outputs

Figure 10-10 Configure menu: Inputs/Outputs (continued)

Discrete Output1 DO Assignment2 DO Polarity3 DO Fault Action4 Flow Switch Source5 Flow Switch Setpoint6 Hysteresis (0.1-10.0)

46

Variable Mapping1 Primary Variable2 Secondary Variable3 Third Variable4 Fourth Variable

5

Communications1 HART Address2 Tag3 Device Identification4 Dev ID (CP)5 Set Up Burst Mode6 mA Output Mode

Set Up Burst Mode1 Burst Mode2 Burst Option3 Field Device Var 14 Field Device Var 25 Field Device Var 36 Field Device Var 4

5

On-Line Menu >2 Configure >

1 Manual Setup >4 Inputs/Outputs

Configuration and Use Manual Supplement 135

Model 2700 Transmitters with Intrinsically Safe Outputs

Mo

del 2700 IS

M

od

el 2700 CIO

Mo

del 2700 A

N

Figure 10-11 Service Tools menu: Alert Setup

On-Line Menu >2 Configure >

2 Alert Setup

I/O Fault Action1 MAO1 Fault Action2 MAO1 Fault Level3 MAO2 Fault Action4 MAO2 Fault Level5 FO Fault Action6 FO Fault Level7 Comm Fault Action

Alert Severity1 Fault Timeout2 Set Alert Severity3 View Alert Severity

Discrete Events1 Discrete Event 12 Discrete Event 23 Discrete Event 34 Discrete Event 45 Discrete Event 56 Assign Discrete Action7 Read Discrete Action8 Review Discrete Actions

1

2

3

1, 2, 3, 4, 5 Discrete Event x1 Discrete Event Var2 Discrete Event Type3 Setpoint A4 Setpoint B

136 Micro Motion Menu Maps for HART Device Descriptions

Model 2700 Transmitters with Intrinsically Safe Outputs

Figure 10-12 Service Tools menu: Top level

On-Line Menu >3 Service Tools

Alerts1 Refresh Alerts

Alert NameAdditional Information

Variables1 Variable Summary2 Process Variables3 Mapped Variables4 External Variables5 Totalizer Control6 Variables (PM) *7 Variables (CM) **8 Outputs

1

2

Trends1 Process Variables2 Diagnostic Variables

3

Maintenance1 Routine Maintenance2 Zero Calibration3 Density Calibration4 Temperature Calibration5 Diagnostic Variables6 Modbus Data

Simulate1 Simulate Outputs2 Simulate Sensor

4

* Displayed only if petroleum measurement isenabled on your transmitter. Menu numbersare adjusted as required.

** Displayed only if concentration measurementis enabled on your transmitter. Menu numbersare adjusted as required.

5

Configuration and Use Manual Supplement 137

Model 2700 Transmitters with Intrinsically Safe Outputs

Mo

del 2700 IS

M

od

el 2700 CIO

Mo

del 2700 A

N

Figure 10-13 Service Tools menu: Variables

On-Line Menu >3 Service Tools >

2 Variables

Process Variables1 Mass Flow Rate2 Volume Flow Rate *3 Density4 Temperature

Mapped Variables1 PV Volume Flow Rate2 SV Volume Flow Rate3 TV Mass Flow Rate4 QV Volume Flow Rate

1

3

5

Totalizer Control1 All Totalizers2 Mass3 Volume *4 CM Volume at Reference Temperature5 CM Net Mass ***6 CM Net Volume ***7 Temp Corrected Volume (PM) **8 Custody Transfer ****

All Totalizers1 Start Totalizers2 Stop Totalizers3 Reset All Totals4 Mass Total5 Volume Total *

1

2

Mass1 Mass Flow Rate2 Mass Total3 Mass Inventory4 Reset Total

Volume *1 Volume Flow Rate2 Volume Total3 Volume Inventory4 Reset Total

3

Variable SummaryMass Flow RateVolume Flow Rate *Density

External Variables1 External Temperature2 External Pressure

4

2

* If Volume Flow Type = GSV, GSV variablesare displayed.

** Displayed only if petroleum measurement isenabled on your transmitter. Menu numbersare adjusted as required.

*** Displayed only if concentration measurementis enabled on your transmitter. Menunumbers are adjusted as required.

**** Displayed only if the Weights & Measuresapplication is enabled on your transmitter.

Additionaloptions

138 Micro Motion Menu Maps for HART Device Descriptions

Model 2700 Transmitters with Intrinsically Safe Outputs

Figure 10-14 Service Tools menu: Variables (continued)

Outputs1 mA Output 1 (MA01)2 mA Output 2 (MA02)3 Frequency Output/

DO State

On-Line Menu >3 Service Tools >

2 Variables

Variables (PM) *1 Density at Reference

Temperature2 Average Observed Density3 Volume Flow at Reference

Temperature4 Average Observed

Temperature5 CTL

Variables (CM) **1 Standard Volume Flow Rate2 Standard Net Volume Flow

Rate3 Net Mass Flow Rate4 Density at Reference5 Concentration6 Density (Fixed SG Units)

6

8

7

Frequency Output1 Present Freq Gauge2 Present Freq Output

1

mA Output 1 (MAO1)1 Present MAO1 Gauge2 PV AO3 PV % Range

3* Displayed only if petroleum measurement is enabled

on your transmitter. Menu numbers are adjusted asrequired.

** Displayed only if concentration measurement isenabled on your transmitter. Menu numbers areadjusted as required.

2

mA Output 2 (MAO2)1 Present MAO2 Gauge2 SV AO3 SV % Range

Configuration and Use Manual Supplement 139

Model 2700 Transmitters with Intrinsically Safe Outputs

Mo

del 2700 IS

M

od

el 2700 CIO

Mo

del 2700 A

N

Figure 10-15 Service Tools menu: Maintenance

On-Line Menu >3 Service Tools >

4 Maintenance

Routine Maintenance1 Trim mA Output 12 Trim mA Output 23 Meter Verification *

1

Zero Calibration1 Mass Flow Rate2 Volume Flow Rate3 Zero Time4 Zero Value5 Standard Deviation6 Perform Auto Zero7 Restore Factory Zero

Density Calibration1 Mass Flow Rate2 Density3 Dens Pt1 (Air)4 Dens Pt2 (Water)5 Dens Pt3 T-Series6 Dens Pt4 T-Series7 Flowing Dens (FD)

2

Temperature Calibration1 Temperature2 Temp Cal Factor

4

3

Meter Verification **1 Run Meter Verification2 View Test Results3 Schedule Meter

Verification

3

* Displayed only if meter verificationis enabled.

** Displayed only if Smart MeterVerification is enabled. For earlierversions, the Meter VerificationMethod is launched.

Additional1 Core Processor Input

Voltage2 Board Temperature3 Power On Time

8

5

Diagnostic Variables1 Sensor Model2 Drive Gain3 LPO Amplitude4 RPO Amplitude5 Tube Frequency6 Live Zero7 Fld Verification Zero8 Additional

Modbus Data1 Read Modbus Data2 Write Modbus Data

6

140 Micro Motion Menu Maps for HART Device Descriptions

Model 2700 Transmitters with Intrinsically Safe Outputs

Figure 10-16 Service Tools menu: Simulate

On-Line Menu >3 Service Tools >

5 Simulate

1

Simulate Sensor1 Simulate Primary Purpose Variables2 Mass Flow Rate3 Density4 Temperature

2

Simulate Outputs1 mA Output 1 Loop Test2 mA Output 2 Loop Test3 Frequency Output Test/

Discrete Output Test *

Configuration and Use Manual Supplement 141

Mo

del 2700 IS

M

od

el 2700 CIO

Mo

del 2700 A

N

Chapter 11Model 2700 Transmitters with Configurable Input/Outputs

Figure 11-1 On-Line menu

Configure1 Manual Setup2 Alert Setup

Service Tools1 Alerts2 Variables3 Trends4 Maintenance5 Simulate

2

3

Overview1 Check Status2 Primary Purpose Variables3 Shortcuts

1

On-Line Menu

142 Micro Motion Menu Maps for HART Device Descriptions

Model 2700 Transmitters with Configurable Input/Outputs

Figure 11-2 Overview menu

Identification1 Tag2 Model3 Xmtr Serial Num4 Sensor Serial Num5 Date6 Descriptor7 Message

1

Revisions1 Universal2 Field Device3 DD Revision4 Transmitter Software5 CP Software6 ETO Number

Mat. of Construction1 Tube Wetted Mat.2 Tube Lining3 Sensor Flange

Check Status1 Refresh Alerts2 Dev Status:3 Comm Status:

Primary Purpose VariablesMass Flow RateVolume Flow RateDensity

Shortcuts1 Device Information2 Totalizer Control3 Zero Calibration4 Variables5 Trends6 Meter Verification *

13

21

2

3

On-Line Menu >1 Overview

Device Information1 Identification2 Revisions3 Mat. of Construction4 Licenses5 Weights and Measures **

* Displayed only if meter verification is enabled.** Displayed only if Weights & Measures

application is enabled. Menu numbers areadjusted as required.

Weights and Measures1 Cstdy Trnsfr Approval2 W&M Software Version3 Core Firmware4 Xmtr Firmware

4

Configuration and Use Manual Supplement 143

Model 2700 Transmitters with Configurable Input/Outputs

Mo

del 2700 IS

M

od

el 2700 CIO

Mo

del 2700 A

N

Figure 11-3 Configure menu: Top level

Manual Setup1 Characterize2 Measurements3 Display4 Inputs/Outputs5 Info Parameters

2

1

Alert Setup1 I/O Fault Actions2 Alert Severity3 Discrete Events

On-Line Menu >1 Configure

144 Micro Motion Menu Maps for HART Device Descriptions

Model 2700 Transmitters with Configurable Input/Outputs

Figure 11-4 Configure menu: Manual Setup

Characterize1 Sensor Type2 Sensor Tag Parameters

Measurements1 Flow2 Density3 Temperature4 Update Rate5 LD Optimization6 Special Units7 External Pressure/Temperature8 Petroleum Measurement*9 GSV10 Conc Measure (CM)**

1

2

Inputs/Outputs1 Channels2 mA Output 13 mA Output 24 Frequency Output5 Discrete Output6 Discrete Input7 Communications8 Variable Mapping

3

Info Parameters1 Transmitter Info2 Sensor Information

5

Display1 Language2 Display Variable Menu Features3 Offline Variable Menu Features4 Backlight5 Display Variables6 Decimal Places

4

Transmitter Info1 Tag2 Xmtr Serial Num3 Message4 Descriptor5 Date6 Write Protect

Sensor Information1 Sensor Type2 Sensor Serial Num3 Tube Wetted Mat.4 Tube Lining5 Sensor Flange

1

2

On-Line Menu >2 Configure >

1 Manual Setup

* Displayed only if petroleummeasurement is enabled on yourtransmitter. Menu numbers areadjusted as required.

** Displayed only if concentrationmeasurement is enabled on yourtransmitter.

Configuration and Use Manual Supplement 145

Model 2700 Transmitters with Configurable Input/Outputs

Mo

del 2700 IS

M

od

el 2700 CIO

Mo

del 2700 A

N

Figure 11-5 Configure menu: Manual Setup: Characterize

On-Line Menu >2 Configure >

1 Manual Setup >1 Characterize

21

2

1

Sensor Type Straight TubeCurved Tube

Density Parameters1 D12 D23 DT4 DTG5 K16 K27 FD8 DFQ19 DFQ2

Flow Parameters1 Flow FCF2 FTG3 FFQ

Sensor Tag Parameters1 Flow Parameters2 Density Parameters

Sensor Tag Parameters1 FlowCal2 D13 D24 TC5 K16 K27 FD

Sensor Type1 Curved Tube2 Straight Tube

146 Micro Motion Menu Maps for HART Device Descriptions

Model 2700 Transmitters with Configurable Input/Outputs

Figure 11-6 Configure menu: Manual Setup: Measurements

On-Line Menu >2 Configure >

1 Manual Setup >2 Measurements

Flow1 Flow Direction2 Flow Damping3 Mass Flow Unit4 Mass Flow Cutoff5 Volume Flow Unit *6 Volume Flow Cutoff *7 Mass Factor8 Volume Factor

Density1 Density Unit2 Density Damping3 Density Cutoff4 Density Factor5 Slug Duration6 Slug Low Limit7 Slug High Limit

Temperature1 Temperature Unit2 Temp Damping

1

2

3

6Special Units1 Mass Special Units2 Volume Special Units

Update Rate1 Update Rate2 100 Hz Variable

4

* Displayed only if Volume Flow Type = Liquid.Menu numbers are adjusted as required.

Additionaloptions

Configuration and Use Manual Supplement 147

Model 2700 Transmitters with Configurable Input/Outputs

Mo

del 2700 IS

M

od

el 2700 CIO

Mo

del 2700 A

N

Figure 11-7 Configure menu: Manual Setup: Measurements (continued)

On-Line Menu >2 Configure >

1 Manual Setup >2 Measurements

1

2

3

8

7

9

10

Pressure1 Pressure Unit2 Pressure Compensation3 Compensation Pressure4 Flow Cal Pressure5 Flow Press Factor6 Dens Press Factor

Temperature1 Temperature Unit2 External Temperature3 Correction Temperature

External Polling1 Poll Control2 Ext Dev Tag 13 Polled Variable 14 Ext Dev Tag 25 Polled Variable 2

External Pressure/Temperature1 Pressure2 Temperature3 External Polling

Petroleum Measurement *1 2540 CTL Table type2 TEC3 Ref Temperature4 Petroleum Msmt Setup

GSV1 Volume Flow Type2 Gas Ref Density3 GSV Cutoff4 GSV Flow Unit5 Gas Density Unit

Conc Measure (CM)**1 CM Configuration2 Matrix Configuration3 Enter Matrix4 Trim CM Process

Variables5 Reset Matrix Data

* Displayed only if petroleum measurementis enabled on your transmitter. Menunumbers are adjusted as required.

** Displayed only if concentrationmeasurement is enabled on yourtransmitter.

148 Micro Motion Menu Maps for HART Device Descriptions

Model 2700 Transmitters with Configurable Input/Outputs

Figure 11-8 Configure menu: Manual Setup: Display

On-Line Menu >2 Configure >

1 Manual Setup >3 Display

LanguageEnglishGermanFrenchSpanish

Display Variable MenuFeatures1 Totalizer Reset2 Start/Stop Totals3 Auto Scroll4 Scroll Time *5 Refresh Rate6 Status LED Blinking

Offline Variable MenuFeatures1 Offline Menu2 Alert Menu3 Acknowledge All4 Offline Passcode5 Alert Passcode6 Offline Passcode

1

2

3

5

Display Variables1 Display Variables (1-5)2 Display Variables (6-10)3 Display Variables (11-15)

Decimal Places1 For Process Variables2 For Totalizer Variables3 For Diagnostic Variables

6

Backlight1 Control2 Intensity (0-63)

4

* Displayed only if Auto Scroll isenabled. Menu numbers areadjusted as required.

Configuration and Use Manual Supplement 149

Model 2700 Transmitters with Configurable Input/Outputs

Mo

del 2700 IS

M

od

el 2700 CIO

Mo

del 2700 A

N

Figure 11-9 Configure menu: Manual Setup: Inputs/Outputs

2

3

MA01 Fault Settings1 MA01 Fault Action2 MA01 Fault Level

mA Output Settings1 PV LRV2 PV URV3 PV Min Span4 PV LSL5 PV USL6 PV MAO1 Cutoff7 PV Added Damping

2

3

Channels1 Channel A2 Channel A Power3 Channel B4 Channel B Power5 Channel C6 Channel C Power

1

mA Output 21 Secondary Variable2 mA Output Settings3 MA02 Fault Settings

MA02 Fault Settings1 MA02 Fault Action2 MA02 Fault Level

mA Output Settings1 SV LRV2 SV URV3 SV Min Span4 SV LSL5 SV USL6 SV MAO2 Cutoff7 SV Added Damping

2

3

On-Line Menu >2 Configure >

1 Manual Setup >4 Inputs/Outputs

mA Output 11 Primary Variable2 mA Output Settings3 MAO1 Fault Settings

Additionaloptions

Channel B1 Frequency Output2 AO23 Discrete Output 1

3

Channel B Power1 External2 Internal

4

Channel C1 Frequency Output2 Discrete Output 23 Discrete Input

5

Channel C Power1 External2 Internal

6

150 Micro Motion Menu Maps for HART Device Descriptions

Model 2700 Transmitters with Configurable Input/Outputs

Figure 11-10 Configure menu: Manual Setup: Inputs/Outputs (continued)

Discrete Output1 DO 1 Assignment2 DO 1 Polarity3 DO 1 Fault Action4 DO 2 Assignment5 DO 2 Polarity6 DO 2 Fault Action7 Flow Switch Source8 Flow Switch Setpoint9 Hysteresis (0.1-10.0)

5

Discrete Input1 DI Assignment2 DI Polarity

6

On-Line Menu >2 Configure >

1 Manual Setup >4 Inputs/Outputs

Frequency Output1 FO Settings2 FO Fault Parameters3 FO Scaling

4

FO Settings1 Third Variable2 Max Pulse Width3 FO Polarity

FO Fault Parameters1 Third Variable2 FO Fault Action3 FO Fault Level

FO Scaling *1 FO Scaling Method2 TV Pulses/Unit3 Set FO Scaling

1

2

3 * Options vary dependingon FO Scaling Method.

Additionaloptions

Configuration and Use Manual Supplement 151

Model 2700 Transmitters with Configurable Input/Outputs

Mo

del 2700 IS

M

od

el 2700 CIO

Mo

del 2700 A

N

Figure 11-11 Configure menu: Manual Setup: Inputs/Outputs (continued)

8

Variable Mapping1 Primary Variable2 Secondary Variable3 Third Variable4 Fourth Variable

7

Communications1 HART Address2 Tag3 Device Identification4 Dev ID (CP)5 Set Up Burst Mode6 mA Output Mode

Set Up Burst Mode1 Burst Mode2 Burst Option3 Field Device Var 14 Field Device Var 25 Field Device Var 36 Field Device Var 4

5

On-Line Menu >2 Configure >

1 Manual Setup >4 Inputs/Outputs

152 Micro Motion Menu Maps for HART Device Descriptions

Model 2700 Transmitters with Configurable Input/Outputs

Figure 11-12 Configure menu: Alert Setup

On-Line Menu >2 Configure >2 Alert Setup

Discrete Events1 Discrete Event 12 Discrete Event 23 Discrete Event 34 Discrete Event 45 Discrete Event 56 Assign Discrete Action7 Read Discrete Action8 Review Discrete Actions

1

2

3

1, 2, 3, 4, 5

Discrete Event x1 Discrete Event Var2 Discrete Event Type3 Setpoint A4 Setpoint B

I/O Fault Actions1 MAO1 Fault Action2 MAO1 Fault Level3 MAO2 Fault Action4 MAO2 Fault Level5 FO Fault Action6 FO Fault Level7 Comm Fault Action

Alert Severity1 Fault Timeout2 Set Alert Severity3 View Alert Severity

Configuration and Use Manual Supplement 153

Model 2700 Transmitters with Configurable Input/Outputs

Mo

del 2700 IS

M

od

el 2700 CIO

Mo

del 2700 A

N

Figure 11-13 Service Tools menu: Top level

On-Line Menu >3 Service Tools

Alerts1 Refresh Alerts

Alert NameAdditional Informationfor Above

Variables1 Variable Summary2 Process Variables3 Mapped Variables4 External Variables5 Totalizer Control6 Variables (PM) *7 Variables (CM) **8 Outputs

1

2

Trends1 Process Variables2 Diagnostic Variables

3

Maintenance1 Routine Maintenance2 Zero Calibration3 Density Calibration4 Temperature Calibration5 Diagnostic Variables6 Modbus Data

Simulate1 Simulate Outputs2 Simulate Sensor

4

* Displayed only if the petroleum measurementapplication is enabled on your transmitter.Menu numbers are adjusted as required.

** Displayed only if the concentrationmeasurement application is enabled on yourtransmitter. Menu numbers are adjusted asrequired.

5

154 Micro Motion Menu Maps for HART Device Descriptions

Model 2700 Transmitters with Configurable Input/Outputs

Figure 11-14 Service Tools menu: Variables

On-Line Menu >3 Service Tools >

2 Variables

Process Variables1 Mass Flow Rate2 Volume Flow Rate *3 Density4 Temperature

Mapped Variables1 PV Mass Flow Rate2 SV Temp3 TV Volume Flow Rate4 QV Volume Flow Rate

1

3

5

Totalizer Control1 All Totalizers2 Mass3 Volume *4 CM Volume at Reference Temperature ***5 CM Net Mass ***6 CM Net Volume ***7 Temp Corrected Volume (PM) **8 Custody Transfer ****

All Totalizers1 Start Totalizers2 Stop Totalizers3 Reset All Totals4 Mass Total5 Volume Total *

1

2

Mass1 Mass Flow Rate2 Mass Total3 Mass Inventory4 Reset Total

Volume *1 Volume Flow Rate2 Volume Total3 Volume Inventory4 Reset Total

3

Variable SummaryMass Flow RateVolume Flow RateDensity

External Variables1 External Temperature2 External Pressure

4

2

* If Volume Flow Type = GSV, GSV variablesare displayed.

** Displayed only if petroleum measurement isenabled on your transmitter. Menu numbersare adjusted as required.

*** Displayed only if concentration measurementis enabled on your transmitter. Menunumbers are adjusted as required.

**** Displayed only if the Weights & Measuresapplication is installed on your transmitter.

Additionaloptions

Configuration and Use Manual Supplement 155

Model 2700 Transmitters with Configurable Input/Outputs

Mo

del 2700 IS

M

od

el 2700 CIO

Mo

del 2700 A

N

Figure 11-15 Service Tools menu: Variables (continued)

Outputs1 mA Output 1 (MA01)2 mA Output 2 (MAO2)3 Frequency Output4 DO State 15 DO State 2

On-Line Menu >3 Service Tools >

2 Variables

Variables (PM) *1 Density at Reference Temperature2 Average Observed Density3 Volume Flow at Reference

Temperature4 Average Observed Temperature5 CTL

Variables (CM) **1 Standard Volume Flow Rate2 Standard Net Volume Flow Rate3 Net Mass Flow Rate4 Density at Reference5 Concentration6 Density (Fixed SG Units)

6

8

7

Frequency Output1 Present Freq Gauge2 Present Freq Output

1

mA Output 1 (MAO1)1 Present MAO1 Gauge2 PV MAO13 PV % Range

3* Displayed only if petroleum measurement is

enabled on your transmitter. Menu numbersare adjusted as required.

** Displayed only if concentration measurementis enabled on your transmitter. Menunumbers are adjusted as required.

mA Output 2 (MAO2)1 Present MAO2 Gauge2 SV MAO23 SV % Range

2

156 Micro Motion Menu Maps for HART Device Descriptions

Model 2700 Transmitters with Configurable Input/Outputs

Figure 11-16 Service Tools menu: Maintenance

On-Line Menu >3 Service Tools >

4 Maintenance

Routine Maintenance1 Trim mA Output 12 Trim mA Output 23 Meter Verification *

1

Zero Calibration1 Mass Flow Rate2 Volume Flow Rate3 Zero Time4 Zero Value5 Standard Deviation6 Perform Auto Zero7 Restore Factory Zero

Density Calibration1 Mass Flow Rate2 Dens Pt1 (Air)3 Dens Pt2 (Water)4 Dens Pt3 T-Series5 Dens Pt4 T-Series6 Flowing Dens (FD)

2

Temperature Calibration1 Temperature2 Temp Cal Factor

4

Diagnostic Variables1 Sensor Model2 Drive Gain3 LPO Amplitude4 RPO Amplitude5 Tube Frequency6 Live Zero7 Fld Verification Zero8 Additional

3

* Displayed only if meter verification isenabled.

** Displayed only if Smart Meter Verificationis enabled. For earlier versions, theMeter Verification Method is launched.

Additional1 Core Processor Input

Voltage2 Board Temperature3 Power On Time

8

Modbus Data1 Read Modbus Data2 Write Modbus Data

6

5

Configuration and Use Manual Supplement 157

Model 2700 Transmitters with Configurable Input/Outputs

Mo

del 2700 IS

M

od

el 2700 CIO

Mo

del 2700 A

N

Figure 11-17 Service Tools menu: Simulate

On-Line Menu >3 Service Tools >

5 Simulate

1

Simulate Sensor1 Simulate Primary Purpose Variables2 Mass Flow Rate3 Density4 Temperature

2

Simulate Outputs1 mA Output 1 Loop Test2 mA Output 2 Loop Test3 Frequency Output Test4 Discrete Output 1 Test5 Discrete Output 2 Test

Micro Motion Inc. USAWorldwide Headquarters7070 Winchester CircleBoulder, Colorado 80301T +1 303-527-5200

+1 800-522-6277F +1 303-530-8459

Micro Motion EuropeEmerson Process ManagementNeonstraat 16718 WX EdeThe NetherlandsT +31 (0) 318 495 555F +31 (0) 318 495 556

Micro Motion JapanEmerson Process Management1-2-5, Higashi ShinagawaShinagawa-kuTokyo 140-0002 JapanT +81 3 5769-6803F +81 3 5769-6844

Micro Motion AsiaEmerson Process Management1 Pandan CrescentSingapore 128461Republic of SingaporeT +65 6777-8211F +65 6770-8003