memo report: membrane separator to: professor ravindra …the purpose of this experiment was to use...
TRANSCRIPT
1
Memo Report: Membrane Separator
TO: Professor Ravindra Datta
FROM: Martin Burkardt ____________________
Project Leader
Michael Bodanza ___________________
Brandon Clark _____________________
Sotirios Filippou ___________________
Marcus Lundgren __________________
DATE: September 9th, 2017
SUBJECT: Enriched Oxygen-Supply System Design Proposal
Introduction
Membrane separation processing has become a common, reliable practice for separating gases, especially in
industrial settings. In comparison with other equipment such as distillation, membrane separators require less
energy, minimal size, and mild conditions. Separation occurs in these units using a selective barrier that operates
based on different permeabilities through the membrane. In the case of nonporous membranes, such as the Pilot
Scale Prism© Membrane unit in Goddard Hall, the mechanism for separation is determined by the dissolution
and diffusion of each gas species within the membrane. The driving force for this separation is the partial pressure
difference of the two components on the two sides and not volatility differences of the feed. Permeance increases
proportionally with pressure.
The purpose of this experiment was to use the Monsanto Company's Pilot Scale Prism© gas permeation
unit to gather experimental data that can then be used to scale-up an oxygen enriching process. The U.O Lab
Supervisor for the Oxymoron Chemical Company has requested a method to produce a continuous supply of
oxygen-enriched air, at 250 SCFM of at least 43% oxygen, from clean dry air at ambient temperature and pressure
up to 110 psig, in order to renovate one of our processing plants. For our purposes, ideally, outlet pressure should
be high.
2
The pilot unit in the lab is composed of four 1-inch (inner diameter) aluminum tubes mounted parallel in
series, and contain a proprietary design of hollow polysulfone fibers that acts as an ideal nonporous membrane
through which the feed will separate the oxygen by solubility-based diffusion (Figure 1a). In this scenario, there
are two streams: the permeate stream and the nonpermeate stream. The nonpermeate stream is located within the
shell of the aluminum tubes, and the hollow channels of the polysulfone fibers contain the permeate stream (Figure
1b). There are thousands of these hollow fibers placed in a compact module in order to provide a higher surface
area for a small volume unit. Under normal operating conditions, the permeate is oxygen-rich, while the
nonpermeate is nitrogen-rich.
Figure 1: Membrane Separator. The aluminum tubes are filled with thousands of fibers (left, a). The feed passes through four of these
tubes. Depending on which of the valves are opened in the permeate stream, will determine the direction of separation (countercurrent
or cocurrent).
Methodology
The oxygen analyzers were calibrated by allowing a feed of compressed air to bypass the separator and
enter the nonpermeate and permeate streams. The system was maintained at 40 psig and all instruments were
allowed to stabilize before corrections were made to the known oxygen concentration, 21%.
Following calibration, the system was operated separately with pure oxygen and pure nitrogen feeds under
various pressures, ranging from 40 to 100 psig, to calculate the membrane permeability of each species. The
selectivity is determined by the ratio of these permeabilities.
The system was run with a feed of air under various operating conditions, including various stage cuts,
pressures, and flow configurations. The flow can be altered between cocurrent and countercurrent flow by
selectively opening two of the valves from the permeate streams (Figure 1). For countercurrent flow, the feed
pressure was varied for four pressures, ranging from 40 to 100 psig. For each pressure, the nonpermeate needle
valve was opened at various levels to vary the stage cut over the range of 0.75 to 0.22 (Figure 2). Once the system
had equilibrated, the outlet flow rates, pressures and oxygen compositions for each stream were recorded. An
additional run was conducted under cocurrent flow at 100 psig with the same range of stage cuts.
3
Figure 2: Pilot Scale P&ID. Control in this system is maintained by a single needle valve on the nonpermeate stream.
Results and Discussion
Our team conducted a feasibility study via COMSOL Multiphysics and Mathcad. We then collected
preliminary data for a scale-up via a pilot scale Prism membrane. We were able to determine the separation factor,
α, of the membrane-separation unit under these operating conditions by measuring the flux of oxygen and nitrogen
separately at several operating pressures (see Appendix A). The selectivity was found to be 6.05 by dividing the
flux trend line for oxygen by the trend line for nitrogen (Figure 3). We also studied the effect of operating pressure
and stage cut on the separation process. Using Fick’s law of diffusion, we are able to study concentration, velocity,
and pressure profiles.
Figure 3: Species Permeance. The slope of the curve plotting the driving force vs. the flux is the permeance of the individual species.
Data Collected from the Pilot Plant
When measuring oxygen mole percentage in the retentate (non-permeate) and permeate against stage cut, they
both tend to decrease as stage cut increases. A larger stage cut allows less air to release into the retentate stream,
4
creating more time for oxygen to find the membrane and lowering the retentate oxygen mole percentage.
Conversely, the permeate stream begins at a high oxygen mole percentage because as the inlet stream is readily
pushed towards the retentate, the nitrogen can easily be swept past, and the membrane’s oxygen selectivity takes
over.
Figure 4: Effect of Stage Cut on Outlet Oxygen Percentage. Generally, as the stage cut decreases the outlet oxygen concentration
increases. This is especially so for the permeate stream. Additionally, as the feed pressure increases, the permeate oxygen concentration
increases.
When measuring oxygen mole percentage against inlet pressure, it seems that higher pressures force more
oxygen through the membrane compared to nitrogen, resulting in higher percentages in the permeate. Partial
pressure differences across the membrane are the driving force for mass transfer in this system. This difference
in effect on permeate percentage between low and high pressure becomes less pronounced as stage cut increases,
since less nitrogen can escape into the retentate.
Furthermore, it is interesting to note that at higher pressures such as 80 and 100 psig (countercurrent
operation) the oxygen mole percentage in the retentate increases sharply at high stage cuts. This might be due to
the fact that, at a certain turning point, a large stage cut and high partial pressure allows enough time and driving
force for nitrogen to permeate more through the membrane, thus counteracting the membrane selectivity exhibited
at lower stage cuts. Curiously, the co-current setup, operating at a high pressure of 100 psig, shows no sign of a
drastic oxygen mole percentage increase. This can either suggest a difference in permeability at high pressures
based upon stream orientation or a potential error in measurement, as this trend is not observed in any of our other
data.
Figure 5: Effect of Stage Cut on Outlet Flow Rates. Interestingly, the permeate flow rate is not affected by stage cut, but responds
more to feed pressure, while the nonpermeate responds more dramatically.
5
When measuring flow rates compared to stage cuts, it seems that as the stage cut increases the retentate flow rate
decreases exponentially, as shown in Figure 5 (left). However, as the stage cut increases the permeate flow rate
sees little to no effect (Figure 5, right). This is because the rate-limiting step for mass transfer to the permeate is
the permeability of oxygen and nitrogen through the membrane, which is determined by the partial pressure
difference of these gas components on both sides. The retentate flow rate itself does not affect the permeate flow
rate; it only lowers the selectivity for oxygen through the membrane. The retentate flow rate mainly affects the
inlet flow rate. Therefore, by shutting the needle valve, the retentate flow decreases, forcing the inlet flow rate to
decrease, thus increasing the stage cut. This idea is why the permeate pressure is relatively constant with respect
to stage cut as well.
Figure 6: Effect of Stage Cut on Permeate Pressure. The stage cut is ineffective on the permeate pressure.
When measuring permeate pressure against inlet pressure, there seems to be no appreciable trend. Manometer
inches of water are a pretty small unit of measurement, so any trends that might occur would be hard to pick up.
Regardless, this data might be due to the membrane holding most of the inlet pressure on the retentate side.
Another curious observation was that the retentate flow meter sign said that it measured flow rate in SCFM
(standard cubic feet per minute), while the permeate flow meter measured flow in SCCM (standard cubic
centimeters per minute). The former unit is 28,317 times smaller than the latter, which is quite large of a difference
as shown in the Figure 6 (right). This is most likely a mistake on the retentate sign, since the two flows should be
a similar order of magnitude.
Verification of Data with Theory
After all data was collected, the selectivity calculated previously was used in partial differential equations that
modeled the pilot separation system for countercurrent and co-current flow (see Appendix B). These
calculations are the differential versions of the perfect mixing model, which uses two mass balances and two
flux expressions, as seen below. All calculations were performed in PTC Mathcad. This is necessary because
the permeability driving force changes continuously along the length of the membrane tube.
6
Both theoretical models seem to match up very well with actual permeate flow rate data. What looks more
problematic is that both models also consistently predict higher permeate oxygen percentages than what was
measured. However, this can be explained by the oxygen meter performance in the lab. Since, they took an
extremely long time to equilibrate and come to a final reading, the team decided to record oxygen levels after a
period specified by the instructor in the interest of time. Since the oxygen percentage trend looks correct, it can
be safely assumed that both models predict percentages relatively well.
Figure 7: Comparison of Theoretical to Data Calculations. On comparison, the model accurately predicts the data collected. Some
deviations do occur, but are not appreciable.
Calculating Required Number of Membrane Tube Modules
From the determined correlation that a decrease in stage-cut is proportional to an increase in permeate
concentration, we decided to investigate the lowest stage cut at each operating pressure/system orientation. Of all
of the feed pressures, the 100 psig countercurrent set up provided above the required oxygen percentage (44%)
with the highest permeate flow rate (72.5 SCCM), which would minimize the number of tubes needed for the
final separation system design. Therefore, this set-up is recommended. The area of one pilot plant tube is smaller
than that of the ones specified by a factor of 4.7. However, increasing both the area and flow rate by the same
factor scales the permeate flow by that exact same factor, while preserving all outlet parameters in the Mathcad
countercurrent flow model. This means that scaling up the area and feed, and therefore the permeate flow to the
necessary 250 SCFM, doesn’t require complex calculation. Scaling up the two countercurrent pilot tubes by 4.7
creates two tubes that can provide 341 SCCM’s. Dividing this number by 28,316, the conversion factor mentioned
earlier, gives a SCFM value of 0.01204. Dividing 250 SCFM by 0.01204 SCFM gives the total number of pairs
of 12 cm, 1.5 m length tubes needed to provide 250 SCFM of flow. This number is quite large, 20,760 pairs, or
41,520 modules.
7
Conclusion and Recommendations
Based upon the preliminary dataset collected through testing of the pilot scale membrane system, it is
recommended that 41,520 polysulfone membrane modules be used; where each module is 12 centimeters in
diameter and 1.5 meters in length. Operating at a system pressure of 100 psig would be adequate at producing
250 SCFM of 44% pure oxygen (1% above the design requirements). Under these conditions, we can produce the
highest flow and pressure of product, and the second-highest purity oxygen compared to the other potential
operating conditions that were investigated.
While this seems like a lot of modules, keep in mind that this is a scaled up, industrialized process that
requires an extremely large flow rate compared to that of the pilot system. Each module takes up only a volume
of 0.017 m2. Meaning that 41,520 would take up a minimum of 706 m3. This space can be provided at a large
scale, industrialized plant.
We always recommend that rapid changes to the system be avoided to prevent damage to the membrane.
High pressures can result in cracks or leaks in the system, or overload and damage the rotameters. Chemical
hazards such as flammability, explosions, and toxicity may be produced in the event that oxygen or nitrogen
escapes the system. Precautionary measures should be taken in order to make this process as safe for users as
possible.
References
Turton, R., et al. (2018). Analysis, Synthesis, and Design of Chemical Processes: Fourth Edition.
Prentice Hall. Print.
S. Weller, W.A. Steiner, Separation of gases by fractional permeation through membranes, Journal
of Applied Physics 21(4) (1950) 279–283.
C.T., Blaisdell, K. Kammermeyer, Counter-current and co-current gas separation, Chemical
Engineering Science 28(6) (1973) 1249-1255.
C.Y. Pan, H.W. Habgood, An analysis of the single-stage gaseous permeation process, Industrial
& Engineering Chemistry Fundamentals 13(4) (1974) 323–331.
W.P. Walawender, S.A. Stern, Analysis of membrane separation parameters. II. Counter-current
and cocurrent flow in a single permeation stage, Separation Science 7(5) (1972) 553-584.
C.J. Geankoplis, Transport Processes and Separation Process Principles (Includes Unit
Operations), Ch. 13, Prentice Hall Press, 2003.
R.A. Davis, O.C. Sandall, A simple analysis for gas separation membrane experiments,
Chemical Engineering Education 37(1) (2003) 74-80.
H. Binous, Gas permeation computations with Mathematica, Chemical Engineering Education,
40(2) (2006) 140-144.
8
Appendix
Appendix A: Excel Data-Pilot Plant Data and Selectivity Calculations
9
Appendix B: Mathcad Calculations on Countercurrent Model
10
Appendix C: Mathcad Calculations on Co-current Model
11
Appendix D: Nomenclature
Am - membrane surface area, Am,t = total membrane surface area
a - dimensionless membrane area, Am/Am,t
B - defined in equation 8
Lf - total inlet gas flow rate
Lo - total nonpermeate/retentate flow rate
l - dimensionless non-permeate flow rate, L/Lf
pf - nonpermeate/retentate pressure
pp - permeate pressure
Q’1 - ratio of oxygen permeability to membrane thickness
Q’2 - ratio of nitrogen permeability to membrane thickness
VP - total permeate flow
v - dimensionless permeate flow rate, V/Lf
12
xf - mole fraction oxygen in feed
xo - mole fraction oxygen in nonpermeate/retentate
yP - mole fraction oxygen in permeate