Landscan LSP-HD Infrared Linescanners i QUALITY CUSTOMER SOLUTIONS LANDSCAN LSP-HD SYSTEM INSTALLATION OVERVIEW PUBLICATION N O 811372 LANGUAGE: ENGLISH

Upload: others

Post on 11-Oct-2020

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

Landscan LSP-HD Infrared Linescannersi

Q U A L I T Y C U S T O M E R S O L U T I O N S

LANDSCAN LSP-HDSYSTEM INSTALLATION OVERVIEWPUBLICATION NO 811372LANGUAGE: ENGLISH

Page 2: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

Health and Safety InformationRead all of the instructions in this booklet - including all the WARNINGS and CAUTIONS - before using this product. If there is any instruction which you do not understand, DO NOT USE THE PRODUCT.

Safety SignsWARNINGIndicates a potentially hazardous situation which, if not avoided, could result in death or personal injury.

CAUTIONIndicates a potentially hazardous situation which, if not avoided, could result in minor or moderate injury to the user or users, or result in damage to the product or to property.

NOTEIndicates a potentially hazardous situation which, if not avoided, could result in damage or the loss of data.

Signs and Symbols used on equipment and Documentation

Caution, risk of electric shock.

Caution, attention to possibility of risk of damage to the product, process or surroundings. Refer to instruction manual.

Caution, hot surface.

Protective Conductor Terminal.

Observe precautions for handling electrostatic discharge sensitive devices.

Equipment OperationUse of this instrument in a manner not specified by Land Instruments International may be hazardous. Read and understand the user documentation supplied before installing and operating the equipment.The safety of any system incorporating this equipment is the responsibility of the assembler.

Protective Clothing, Face and Eye ProtectionIt is possible that this equipment is to be installed on, or near to, machinery or equipment operating at high temperatures and high pressures. Suitable protective clothing, along with face and eye protection must be worn. Refer to the health and safety guidelines for the machinery/equipment before installing this product. If in doubt, contact Land Instruments International.

Electrical Power SupplyBefore working on the electrical connections, all of the electrical power lines to the equipment must be isolated. All the electrical cables and signal cables must be connected exactly as indicated in these operating instructions. If in doubt, contact Land Instruments International.

StorageThe instrument should be stored in its packaging, in a dry sheltered area.

UnpackingCheck all packages for external signs of damage. Check the contents against the packing note.

Lifting InstructionsWhere items are too heavy to be lifted manually, use suitably rated lifting equipment. Refer to the Technical Specification for weights. All lifting should be done as stated in local regulations.

IMPORTANT INFORMATION - PLEASE READ

Page 3: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

Health and Safety InformationRead all of the instructions in this booklet - including all the WARNINGS and CAUTIONS - before using this product. If there is any instruction which you do not understand, DO NOT USE THE PRODUCT.

Safety SignsWARNINGIndicates a potentially hazardous situation which, if not avoided, could result in death or personal injury.

CAUTIONIndicates a potentially hazardous situation which, if not avoided, could result in minor or moderate injury to the user or users, or result in damage to the product or to property.

NOTEIndicates a potentially hazardous situation which, if not avoided, could result in damage or the loss of data.

Signs and Symbols used on equipment and Documentation

Caution, risk of electric shock.

Caution, attention to possibility of risk of damage to the product, process or surroundings. Refer to instruction manual.

Caution, hot surface.

Protective Conductor Terminal.

Observe precautions for handling electrostatic discharge sensitive devices.

Equipment OperationUse of this instrument in a manner not specified by Land Instruments International may be hazardous. Read and understand the user documentation supplied before installing and operating the equipment.The safety of any system incorporating this equipment is the responsibility of the assembler.

Protective Clothing, Face and Eye ProtectionIt is possible that this equipment is to be installed on, or near to, machinery or equipment operating at high temperatures and high pressures. Suitable protective clothing, along with face and eye protection must be worn. Refer to the health and safety guidelines for the machinery/equipment before installing this product. If in doubt, contact Land Instruments International.

Electrical Power SupplyBefore working on the electrical connections, all of the electrical power lines to the equipment must be isolated. All the electrical cables and signal cables must be connected exactly as indicated in these operating instructions. If in doubt, contact Land Instruments International.

StorageThe instrument should be stored in its packaging, in a dry sheltered area.

UnpackingCheck all packages for external signs of damage. Check the contents against the packing note.

Lifting InstructionsWhere items are too heavy to be lifted manually, use suitably rated lifting equipment. Refer to the Technical Specification for weights. All lifting should be done as stated in local regulations.

IMPORTANT INFORMATION - PLEASE READ Contact UsUK - DronfieldLand Instruments InternationalTel: +44 (0) 1246 417691Email: [email protected]: www.landinst.com

USA - PittsburghAMETEK Land, Inc.Tel: +1 412 826 4444Email: [email protected] Web: www.ametek-land.com

ChinaAMETEK Land China Service Tel: +86 21 5868 5111 ext 122Email: [email protected]: www.landinst.com

IndiaAMETEK Land India ServiceTel: +91 - 80 67823240 Email: [email protected]: www.landinst.com

For further details on all AMETEK Land offices, distributors and representatives, please visit our websites.

Return of Damaged GoodsIMPORTANT If any item has been damaged in transit, this should be reported to the carrier and to the supplier immediately. Damage caused in transit is the responsibility of the carrier not the supplier.DO NOT RETURN a damaged instrument to the sender as the carrier will not then consider a claim. Save the packing with the damaged article for inspection by the carrier.

Return of Goods for RepairIf you need to return goods for repair please contact our Customer Service Department. They will be able to advise you on the correct returns procedure.Any item returned to Land Instruments International should be adequately packaged to prevent damage during transit.You must include a written report of the problem together with your own name and contact information, address, telephone number, email address etc.

Design and Manufacturing StandardsThe Quality Management System of Land Instruments International is approved to BS EN ISO 9001 for the design, manufacture and on-site servicing of combustion, environmental monitoring and non-contact temperature measuring instrumentation.

Approvals apply in the USA

Operation of radio transmitters, telephones or other electrical/electronic devices in close proximity to the equipment while the enclosure doors of the instrument or its peripherals are open, may cause interference and possible failure where the radiated emissions exceed the EMC directive. The protection provided by this product may be invalidated if alterations or additions are made to the structural, electrical, mechanical or pneumatic parts of this system. Such changes may also invalidate the standard terms of warranty.

CopyrightThis manual is provided as an aid to owners of Land Instruments International’s products and contains information proprietary to Land Instruments International. This manual may not, in whole or part, be copied, or reproduced without the expressed written consent of Land Instruments International Ltd.

Copyright © 2016 Land Instruments International.

MARCOM0311, Issue 7, 01 November 2016

Approvals apply in India

Page 4: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

Infrared Linescanning System

Installation Overview

Landscan LSP-HD

Blank

Page 5: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

Landscan LSP-HDInfrared Linescanning System

Installation Overview

Contents1 Introduction 1-1

1.1 About this Guide 1-11.2 System Installation Sequence 1-1

2 TypicalSystemConfigurations 2-12.1 BasicSingleLandscanConfigurationwithProcessorConfiguration

SoftwareforTCP/IPcommunication 2-12.2 BasicSingleLandscanConfigurationwithProcessorConfiguration

SoftwareandLSP-HDIOProcessor 2-22.3 BasicSingleLandscanConfigurationwithWCASoftware 2-32.4 BasicSingleLandscanConfigurationwithWCASoftwareand

LSP-HDIOProcessor 2-42.5 TwoScannerLandscanConfigurationwithWCASoftwareand

oneLSP-HDIOProcessor 2-62.6 TwoScannerLandscanConfigurationwithWCASoftwareand

twoLSP-HDIOProcessors 2-82.7 MultipleScannerLandscanConfigurationwithWCASoftware

andoneLSP-HDIOProcessor 2-102.8 MultipleScannerLandscanConfigurationwithWCAServerand

ClientSoftwareandoneLSP-HDIOProcessor 2-122.9 MultipleScannerLandscanConfigurationwithWCAServerand

ClientSoftwareandmultipleLSP-HDIOProcessors 2-14

3 MountingAccessories 3-13.1 BasicMountingPlate-PartNº031.520 3-23.2 WaterCooled,AirPurgedBasePlate-PartNº031.496 3-33.3 RightAngleMountingPlate-PartNº031.438 3-43.4 EnhancedWaterCooledMountingSystem-PartNº800095 3-53.5 LightIndustrialHousing-PartNº802664 3-63.6 HeavyIndustrialHousing-PartNº802665 3-7

4 Blowers 4-1

5 PowerSupplies 4-15.1 LSP-HDDesktopPowerSupply-PartNº805024 5-15.2 LSP-HDIndustrialPowerSupplyonaDINRail-PartNº807105 5-25.3 LSP-HDIndustrialPowerSupplyinaTerminalBox-PartNº805037 5-3

6 Cables 6-1

7 SoftwareSetup 7-17.1 SoftwareRegistration 7-17.2 BasicSoftwareSetup 7-27.3 DataProcessing 7-127.4 Display 7-167.5 StorageManagement 7-197.6 Diagnostics 7-227.7 SaveConfiguration 7-247.8 ChangeConfiguration 7-247.9 SaveLPU/LSCConfiguration 7-257.10ChangeLPU/LSCConfiguration 7-25

8 Maintenance&Service 8-1

Page 6: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

INTRODUCTION

1

Page 7: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

1 - 1

Landscan LSP-HDInfrared Linescanning System

Installation Overview

1 Introduction

1.1 About this GuideThisdocumentprovidesanoverviewofthecomponentsavailableinaLandscanLSP-HDInfraredLinescanningSystemandisintendedasaguidetotheinstallationandcommissioningprocedure.Notethat,duetothevarietyofsystemconfigurationsavailable,thereisnodefinitive‘standard’installationandcommissioningprocess.However,thisdocumentprovidesanoverviewofthetypicalstepsrequiredtocompleteinstallationandcommissioning.Adescriptionofthemostcommonconfigurationsavailable,frombasicsinglescannersystemstomultipleScannersystemswithI/Oprocessors,isgiveninSection2.Readthisdocumentthoroughlypriortoinstallingyoursystem,referringalsototheuserdocumentationprovidedwiththecomponentsofyourLSP-HDsystem.

1.2 System Installation SequenceAtypicalinstallationandcommissioningsequenceisoutlinedbelow:1) Install the MountingandverifythattheScannerisalignedcorrectlyonthetargeti.e.

thereisnothinginthefieldofviewwhichobscuresthesighting:• Makesurethatthereisnoframeworkorsimilarinthefieldofview;• Inordertoavoidreflections,installthemountingsuchthatthescannerdoesnot

lookontorollers;• Ifheatshieldsareinstalledthecutoutmustgivesufficientspaceforthescan

angle(Maximum scan angle is 80°);• Verifythatthemountingheightallowsthemaximumproductwidthtobecovered

bythemaximumscanangleof80°.2) MakesurethatyouhavethecorrectCooling Water and Purge Air(withassociated

Blower)suppliessetupandoperational.Connectthesetothescannermountingsbeforeinstallingit.

3) Choose an installation location for the Power Supply Unit such that it is shielded from theprocessheatatthegivendistancetothescannerlocation(Thestandardlengthofconnectioncableis15m/48ft).Applythenecessarypowerconnections.

4) Connect the scanner Cabletothemounting.5) ConnectthePoweroverEthernetSwitch(PoE Switch) to the PC or Network Switch.

• A Patch Cablecanbeusedforupto100m/330ftdistance:ifalongerdistanceisrequired,thenaFibrenetworkmustbeused.

6) IfaBeckhoff IOunitispartofthesystem,itmustalsobeconnectedtothenetworkswitch.• Allcomponents,scannerandBeckhoffmustbeconnectedphysicallytothe

Networkbeforethesystemispoweredup.7) InstalltheDataServerInstall(WCA Software) on the Server PC.8) ApplyPower to all the system components.9) Verify that all scanners and

Beckhoffprocessorsarevisibleinthenetwork.Toverifytheconnectionusethe‘ping’commandfrom the command prompt as shown.TherecommendednetworkIPAddressofthePCrunningWCASoftwareis10.1.10.25,subnet255.255.0.0.

IfyouwishtocommunicatetothescannerPCfromyourplantnetwork,werecommendfittingasecondNetworkInterfaceCardandaddressingittocomplywithyourplantnetwork.

Inthisway,thescannerlivedatatrafficdoesnotaffectyournetwork.

Page 8: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

Infrared Linescanning System

Installation Overview

Landscan LSP-HD

1 - 2

10) SetupthecommunicationtotheScannerandBeckhoffIO:• All ScannersdeliveredfromAMETEKLandareshippedwiththesamedefault

IP Address: 10.1.10.100;• All Beckhoff IOunitshaveadefaultIP Address: 10.1.10.110;• Therefore,ifyouaresettingupamultiplescannersystem,thescannersmustbe

connectedtothenetworkinsequence.YouwillneedtochangetheIPaddress*ofthefirstscannerconnectedtoavoidanyaddressconflictinthenetwork.ItisrecommendedtochangetheIPforthefirstscannerto10.1.10.101,andthesecond scanner to 10.1.10.102,andsoon.ThisalsoavoidsanyaddressconflictproblemsifaspareorrepairedscannerwithadefaultIPaddresshastobeintegratedintoanexistingsystem.

• IfyouhavemultipleBeckhoffIOunits,youwillalsoneedtochangetheirIPaddressesinasimilarmanner.i.e.setthefirstIPaddressto10.1.10.111 etc.

* InstructionsonhowtosettheIPaddressofeachcomponentinyourLSP-HDsystemaregiveninSection7.2.4ofthisGuide.

Page 9: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

TYPICAL SYSTEM CONFIGURATIONS

2

Page 10: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

Infrared Linescanning System

Installation Overview

Landscan LSP-HD

Blank

Page 11: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

2 - 1

Landscan LSP-HDInfrared Linescanning System

Installation Overview

2 Typical System ConfigurationsThissectiongivesanoverviewofthecomponentsandoptionsavailableintypicalconfigurationsofaLandscanLSP-HDsystem.Furtherinformationoneachcomponentisgiveninsubsequentsectionsofthisguide.Identifywhichconfigurationmatchesyoursystemandrefertotherelevantsectionsofthisguideforinstallationandcommissioninginformation.

2.1 Basic Single Landscan Configuration with Processor Configuration Software for TCP/IP communication

Item Component Option (IllustratedoptionsshowninBold text) Land Part Nº1 Scanner LSP-HD xx 80xxxx2 Mounting Basic Mounting Plate 031.520

•OptionalRightAngleBracket(Item2a) 031.438WaterCooled,AirPurgedBasePlate 031.496•OptionalRightAngleBracket(Item2a) 031.438EnhancedWaterCooledMountingSystem 800095LightIndustrialHousing(withAirPurging/WaterCooling) 802664LightIndustrialHousing(withnon-reflectiveheatshield) 806578HeavyIndustrialHousing(withAirPurging/WaterCooling) 802665WaterCooledPlate(SealedtoProcess) 805154

3 Cable LSP-HD Cable (Low Temperature) 15 m 805038LSP-HDCable(LowTemperature)30m 807087LSP-HDCable(HighTemperature)15m 804977

3a Adapter Cable M12 to RJ45 Cable 0.5 m* 8050234 PSU LSP-HD Desktop Power Supply 805024

LSP-HDIndustrialPowerSupplyonaDINRail 807105LSP-HDIndustrialPowerSupplyinaTerminalBox**Adapter cable 805023 not required for use with 805037

805037

2

13

3a

4

Customer’sPC

ProcessorConfigurationSoftwaresuppliedwithLSP-HD

Customer’s RJ45 Ethernet cable

Page 12: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

Infrared Linescanning System

Installation Overview

Landscan LSP-HD

2 - 2

2.2 Basic Single Landscan Configuration with Processor Configuration Software and LSP-HD IO Processor

Item Component Option (IllustratedoptionsshowninBold text) Land Part Nº1 Scanner LSP-HD xx 80xxxx2 Mounting BasicMountingPlate 031.520

•OptionalRightAngleBracket 031.438WaterCooled,AirPurgedBasePlate 031.496•OptionalRightAngleBracket 031.438Enhanced Water Cooled Mounting System 800095LightIndustrialHousing(withAirPurging/WaterCooling) 802664LightIndustrialHousing(withnon-reflectiveheatshield) 806578HeavyIndustrialHousing(withAirPurging/WaterCooling) 802665WaterCooledPlate(SealedtoProcess) 805154

3 Cable LSP-HD Cable (Low Temperature) 15 m 805038LSP-HDCable(LowTemperature)30m 807087LSP-HDCable(HighTemperature)15m 804977

3a Adapter Cable M12toRJ45Cable0.5m 8050234 PSU LSP-HDIndustrialPowerSupplyonaDINRail 807105

LSP-HD Industrial Power Supply in a Terminal Box 8050375 IO Processor LSP-HD I/O Processor 2/4/4/4 805773

LSP-HDI/OProcessor4/12/8/12 8057745a IO PSU PSU for LSP-HD I/O Processor 805953

2

1

3

5a

4

OptionalPCforConfiguration/DataDisplay

Customer’s RJ45 Ethernet cables

5

Page 13: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

2 - 3

Landscan LSP-HDInfrared Linescanning System

Installation Overview

2.3 Basic Single Landscan Configuration with WCA Software

Item Component Option (IllustratedoptionsshowninBold text) Land Part Nº1 Scanner LSP-HD xx 80xxxx2 Mounting Basic Mounting Plate 031.520

• Optional Right Angle Bracket (Item 2a) 031.438WaterCooled,AirPurgedBasePlate 031.496•OptionalRightAngleBracket(Item2a) 031.438EnhancedWaterCooledMountingSystem 800095LightIndustrialHousing(withAirPurging/WaterCooling) 802664LightIndustrialHousing(withnon-reflectiveheatshield) 806578HeavyIndustrialHousing(withAirPurging/WaterCooling) 802665WaterCooledPlate(SealedtoProcess) 805154

3 Cable LSP-HD Cable (Low Temperature) 15 m 805038LSP-HDCable(LowTemperature)30m 807087LSP-HDCable(HighTemperature)15m 804977

3a Adapter Cable M12 to RJ45 Cable 0.5 m* 8050234 PSU LSP-HD Desktop Power Supply 805024

LSP-HDIndustrialPowerSupplyonaDINRail 807105LSP-HDIndustrialPowerSupplyinaTerminalBox**Adapter cable 805023 not required for use with 805037

805037

5 WCA Software Single Head License 8041096 Workstation DELL Precision for 1 Scanner 806152

2a

2

1

3

3a

4

6

5

Page 14: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

Infrared Linescanning System

Installation Overview

Landscan LSP-HD

2 - 4

2.4 Basic Single Landscan Configuration with WCA Software and LSP-HD IO Processor

2

13

3a

4

8

55a

6

7

Customer’s RJ45 Ethernet cables

Page 15: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

2 - 5

Landscan LSP-HDInfrared Linescanning System

Installation Overview

Item Component Option Land Part Nº1 Scanner LSP-HD xx 80xxxx2 Mounting BasicMountingPlate 031.520

•OptionalRightAngleBracket 031.438WaterCooled,AirPurgedBasePlate 031.496•OptionalRightAngleBracket 031.438EnhancedWaterCooledMountingSystem 800095LightIndustrialHousing(withAirPurging/WaterCooling) 802664LightIndustrialHousing(withnon-reflectiveheatshield) 806578Heavy Industrial Housing (with Air Purging/Water Cooling)

802665

WaterCooledPlate(SealedtoProcess) 8051543 Cable LSP-HDCable(LowTemperature)15m 805038

LSP-HDCable(LowTemperature)30m 807087LSP-HD Cable (High Temperature) 15 m 804977

3a Adapter Cable M12 to RJ45 Cable 0.5 m* 8050234 PSU LSP-HD Desktop Power Supply 805024

LSP-HDIndustrialPowerSupplyonaDINRail 807105LSP-HDIndustrialPowerSupplyinaTerminalBox**Adapter cable 805023 not required for use with 805037

805037

5 IO Processor LSP-HDI/OProcessor2/4/4/4 805773LSP-HD I/O Processor 4/12/8/12 805774

5a IO PSU PSU for LSP-HD I/O Processor 8059536 WCA Single Head License 8041097 Workstation DELL Precision for 1 Scanner 8061528 Switch Ethernet Switch 800983

Page 16: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

Infrared Linescanning System

Installation Overview

Landscan LSP-HD

2 - 6

2.5 Two Scanner Landscan Configuration with WCA Software and one LSP-HD IO Processor

2a

1

3a

4

8

5

5a

6

7

1

2b

3b

4

Cus

tom

er’s

RJ4

5 Et

hern

et c

able

s

Page 17: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

2 - 7

Landscan LSP-HDInfrared Linescanning System

Installation Overview

Item Component Option Land Part Nº1 Scanner LSP-HD xx 80xxxx

BasicMountingPlate 031.520•OptionalRightAngleBracket 031.438WaterCooled,AirPurgedBasePlate 031.496•OptionalRightAngleBracket 031.438

2a Mounting Enhanced Water Cooled Mounting System 800095LightIndustrialHousing(withAirPurging/WaterCooling) 802664LightIndustrialHousing(withnon-reflectiveheatshield) 806578

2b Mounting Heavy Industrial Housing (with Air Purging/Water Cooling)

802665

WaterCooledPlate(SealedtoProcess) 8051543a Cable LSP-HD Cable (Low Temperature) 15 m 8050383b Cable LSP-HD Cable (Low Temperature) 30 m 807087

LSP-HDCable(HighTemperature)15m 8049774 PSU LSP-HDDesktopPowerSupply 805024

LSP-HDIndustrialPowerSupplyonaDINRail 807105LSP-HD Industrial Power Supply in a Terminal Box 805037

5 IO Processor LSP-HDI/OProcessor2/4/4/4 805773LSP-HD I/O Processor 4/12/8/12 805774

5a IO PSU PSU for LSP-HD I/O Processor 8059536 WCA Server Installation 8024277 Workstation DELL Precision for 1 Scanner 8061528 Switch Ethernet Switch 800983

Page 18: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

Infrared Linescanning System

Installation Overview

Landscan LSP-HD

2 - 8

2.6 Two Scanner Landscan Configuration with WCA Software and two LSP-HD IO Processors

2b

1

3a

4

8

5

5a

6

7

1

2a

3b

4

Cus

tom

er’s

RJ4

5 Et

hern

et c

able

s

5

5a

Page 19: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

2 - 9

Landscan LSP-HDInfrared Linescanning System

Installation Overview

Item Component Option Land Part Nº1 Scanner LSP-HDxx 80xxxx2a Mounting Basic Mounting Plate 031.520

•OptionalRightAngleBracket 031.4382b Mounting Water Cooled, Air Purged Base Plate 031.496

•OptionalRightAngleBracket 031.438EnhancedWaterCooledMountingSystem 800095LightIndustrialHousing(withAirPurging/WaterCooling) 802664LightIndustrialHousing(withnon-reflectiveheatshield) 806578HeavyIndustrialHousing(withAirPurging/WaterCooling) 802665

WaterCooledPlate(SealedtoProcess) 8051543a Cable LSP-HD Cable (Low Temperature) 15 m 8050383b Cable LSP-HD Cable (Low Temperature) 30 m 807087

LSP-HDCable(HighTemperature)15m 8049774 PSU LSP-HDDesktopPowerSupply 805024

LSP-HDIndustrialPowerSupplyonaDINRail 807105LSP-HD Industrial Power Supply in a Terminal Box 805037

5 IO Processor LSP-HDI/OProcessor2/4/4/4 805773LSP-HD I/O Processor 4/12/8/12 805774

5a IO PSU PSU for LSP-HD I/O Processor 805953•Either805773or805774tobeallocatedtoeachindividualscanner

6 WCA Server Installation 8024277 Workstation DELL Precision for 2-4 Scanners 8061518 Switch Ethernet Switch 800983

Page 20: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

Infrared Linescanning System

Installation Overview

Landscan LSP-HD

2 - 10

2.7 Multiple Scanner Landscan Configuration with WCA Software and one LSP-HD IO Processor

2b

3a

4

8

6

7

1

2a

3b

4

Cus

tom

er’s

RJ4

5 Et

hern

et c

able

s

5

5a

3b

1

1

2c

4

Page 21: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

2 - 11

Landscan LSP-HDInfrared Linescanning System

Installation Overview

Item Component Option Land Part Nº1 Scanner LSP-HD xx 80xxxx2a Mounting Basic Mounting Plate 031.520

•OptionalRightAngleBracket 031.4382b Mounting Water Cooled, Air Purged Base Plate 031.496

•OptionalRightAngleBracket 031.4382c Mounting Enhanced Water Cooled Mounting System 800095

LightIndustrialHousing(withAirPurging/WaterCooling) 802664LightIndustrialHousing(withnon-reflectiveheatshield) 806578HeavyIndustrialHousing(withAirPurging/WaterCooling) 802665WaterCooledPlate(SealedtoProcess) 805154

3a Cable LSP-HD Cable (Low Temperature) 15 m 8050383b Cable LSP-HD Cable (Low Temperature) 30 m 807087

LSP-HDCable(HighTemperature)15m 8049774 PSU LSP-HDDesktopPowerSupply 805024

LSP-HDIndustrialPowerSupplyonaDINRail 807105LSP-HD Industrial Power Supply in a Terminal Box 805037

5 IO Processor LSP-HDI/OProcessor2/4/4/4 805773LSP-HD I/O Processor 4/12/8/12 805774

5a IO PSU PSU for LSP-HD I/O Processor 805953•Either805773or805774tobeallocatedtoeachindividualscanner

6 WCA Server Installation 8024277 Workstation DELL Precision for 2-4 Scanners 8061518 Switch Ethernet Switch 800983

Page 22: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

Infrared Linescanning System

Installation Overview

Landscan LSP-HD

2 - 12

2.8 Multiple Scanner Landscan Configuration with WCA Server and Client Software and one LSP-HD IO Processor

2b

3a

4

8

6a

7a

1

2a

3b

4

Cus

tom

er’s

RJ4

5 Et

hern

et c

able

s

5

5a

3b

1

1

2c

4

6b

7b

Page 23: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

2 - 13

Landscan LSP-HDInfrared Linescanning System

Installation Overview

Item Component Option Land Part Nº1 Scanner LSP-HD xx 80xxxx2a Mounting Basic Mounting Plate 031.520

•OptionalRightAngleBracket 031.4382b Mounting Water Cooled, Air Purged Base Plate 031.496

•OptionalRightAngleBracket 031.4382c Mounting Enhanced Water Cooled Mounting System 800095

LightIndustrialHousing(withAirPurging/WaterCooling) 802664LightIndustrialHousing(withnon-reflectiveheatshield) 806578HeavyIndustrialHousing(withAirPurging/WaterCooling) 802665WaterCooledPlate(SealedtoProcess) 805154

3a Cable LSP-HD Cable (Low Temperature) 15 m 8050383b Cable LSP-HD Cable (Low Temperature) 30 m 807087

LSP-HDCable(HighTemperature)15m 8049774 PSU LSP-HDDesktopPowerSupply 805024

LSP-HDIndustrialPowerSupplyonaDINRail 807105LSP-HD Industrial Power Supply in a Terminal Box 805037

5 IO Processor LSP-HD I/O Processor 4/12/8/12 8057745a IO PSU PSU for LSP-HD I/O Processor 8059536a WCA Server Installation 8024276b Client License 8041107a Workstation DELL Precision for 2-4 Scanners for Server

Installation806151

7b DELL Precision for Client License 8061528 Switch Ethernet Switch 800983

Note

TheLiveClientmustbeconnectedtotheLandscanNetwork,nottotheCustomer’sNetwork

Page 24: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

Infrared Linescanning System

Installation Overview

Landscan LSP-HD

2 - 14

2.9 Multiple Scanner Landscan Configuration with WCA Server and Client Software and multiple LSP-HD IO Processors

2b

3a

4

8

6a

7a

1

2a

3b

4

Cus

tom

er’s

RJ4

5 Et

hern

et c

able

s

5a

5c

3b

1

1

2c

4

6b

7b

5a

5b

5c

5c

Page 25: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

2 - 15

Landscan LSP-HDInfrared Linescanning System

Installation Overview

Item Component Option Land Part Nº1 Scanner LSP-HD xx 80xxxx2a Mounting Basic Mounting Plate 031.520

•OptionalRightAngleBracket 031.4382b Mounting Water Cooled, Air Purged Base Plate 031.496

•OptionalRightAngleBracket 031.4382c Mounting Enhanced Water Cooled Mounting System 800095

LightIndustrialHousing(withAirPurging/WaterCooling) 802664HeavyIndustrialHousing(withAirPurging/WaterCooling) 802665

3a Cable LSP-HD Cable (Low Temperature) 15 m 8050383b Cable LSP-HD Cable (Low Temperature) 30 m 807087

LSP-HDCable(HighTemperature)15m 8049774 PSU LSP-HDDesktopPowerSupply 805024

LSP-HDIndustrialPowerSupplyonaDINRail 807105LSP-HD Industrial Power Supply in a Terminal Box 805037

5a IO Processor LSP-HD I/O Processor 4/12/8/12 8057745b IO Processor LSP-HD I/O Processor 4/12/8/12 8057745c IO PSU PSU for LSP-HD I/O Processor 8059536a WCA Server Installation 8024276b Client License 8041107a Workstation DELL Precision for 2-4 Scanners for Server

Installation806151

7b DELL Precision for Client License 8061528 Switch Ethernet Switch 800983

Note

TheLiveClientmustbeconnectedtotheLandscanNetwork,nottotheCustomer’sNetwork

Page 26: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

MOUNTING ACCESSORIES

3

Page 27: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

3 - 1

Landscan LSP-HDInfrared Linescanning System

Installation Overview

3 Mounting AccessoriesThissectionoftheguidecontainsinstallationinformationforthefollowingMountingAccessories:• BasicMountingPlate(031.520) • OptionalRightAngleMountingPlate(031.438)• WaterCooled,AirPurgedBasePlate(031.496) • OptionalRightAngleMountingPlate(031.438)• EnhancedWaterCooledMountingSystem(800095)• LightIndustrialHousing(withAirPurging/WaterCooling)(802664)• HeavyIndustrialHousing(withAirPurging/WaterCooling)(802665)

Page 28: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

Infrared Linescanning System

Installation Overview

Landscan LSP-HD

3 - 2

3.1 Basic Mounting Plate - Part Nº 031.520TheBasicMountingPlateisusedinapplicationswhereanLSP-HDLinescannerheadistobeinstalledinlightindustrialenvironmentswheretheambienttemperaturesiswithintherecommendedoperatingrangeoftheinstrumentandtheatmosphereisrelativelyfreefromdustandairborne particles.

Thebasicbaseplateprovidesaquickreleasemountingarrangementforthescanner,makingrealignmentasimpleandeasyoperation.

SpecificationsMaximumAmbientTemperature 60°C/140°FWeight 1.2kg/2.6lbDimensions(LxWxD) 276x176x52.5mm/10.87x6.93x2.07in

MountingHoles Ø8.5mm/0.34intoacceptM8screws

Dimensions Alldimensionsinmm/in

52.5/2.07

22.5/0.89

152/5.98

276/10.87

240/9.45

HoleØ8.5/0.34toacceptM8screws

176/6.93

Page 29: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

3 - 3

Landscan LSP-HDInfrared Linescanning System

Installation Overview

Dimensions Alldimensionsinmm/in

3.2 Water Cooled, Air Purged Base Plate - Part Nº 031.496TheWaterCooled,AirPurgedBasePlateisusedinapplicationswhereanLSP-HDLinescanneristobeinstalledintoenvironmentswheretheambienttemperaturesareintherangeof60to100°C/140to212°F.Itisideallysuitedtogeneralindustrial applications.Theairpurgefacilityshouldbeutilisedwheretheoperatingenvironmentisdustyandairborneparticlescouldcontaminatethescannerviewingwindow.Theplateprovidesaquickreleasemountingarrangementforthescanner,makingrealignmentasimpleandeasyoperation.

SpecificationsMaximumAmbientTemperature 100°C/212°FCoolingWaterRequirements 1.5to2litrepermin/0.4to0.53USgalpermin@

30°C/86°FMaximumWaterPressure 5.88Bar/85PSI

CoolingWaterConnection ¼inBSPfor9.5mm/3/8inborewaterhoseAirPurgeRequirements 300litrepermin/12.36cuftperminAirPurgeConnection G3/8suitablefor38mm/1.5ininnerdiameterneo-

prene hoseWeight 2.2kg/4.8lbDimensions(lxwxd) 375x176x100mm/14.76x6.93x3.94inMountingHoles Ø8.5mm/0.34intoacceptM8screws

375/14.76Max

240/9.45100/3.94MaxScanner Clamp

152/5.98

176/6.93

HoleØ8.5/0.34toacceptM8screws

AirpurgeinletG3/8thread.38/1.5I/D

neoprene hose

Waterinlet/outlet1/4BSP.9.5/3/8IDreinforcedPVChose

Page 30: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

Infrared Linescanning System

Installation Overview

Landscan LSP-HD

3 - 4

240/9.45

152/5.98

52/2.05

222/8.74

Stainless Steel

Alldimensionsinmm/in

Scannermountingface

HoleØ8.5/0.34toacceptM8screws

3.3 Right Angle Mounting Plate - Part Nº 031.438TherightanglemountingplateisasimpledesignintendedforbasicLSP-HDLinescannerHeadmounting.Itisdesignedforwallmountingandispre-drilledtoaccepteithertheBasicMountingPlate(031.520)ortheWaterCooled,AirPurgedBasePlate(031.496).Allfixingsaresuppliedwiththeplate.

Dimensions Alldimensionsinmm/in

Page 31: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

3 - 5

Landscan LSP-HDInfrared Linescanning System

Installation Overview

Dimensions All dimensions in mm/in

3.4 Enhanced Water Cooled Mounting System - Part Nº 800095The Water Cooled, Air Purged Base Plate is used in applications where an LSP-HD Linescanner is to be installed into environments where the ambient temperatures are in the range of 60 to 120 °C / 140 to 248 °F. It is ideally suited to general industrial applications.The air purge facility should be utilised where the operating environment is dusty and airborne particles could contaminate the scanner viewing window. The plate provides a quick release mounting arrangement for the scanner, making realignment a simple and easy operation.

SpecificationsMaximum Ambient Temperature 120 °C / 248 °FCooling Water Requirements 1.5 to 2 litre per min / 0.4 to 0.53 US gal per min @

30 °C / 86 °FMaximum Water Pressure 5.88 Bar / 85 PSI

Cooling Water Connection ¼” BSP for 9.5 mm / 3/8 in bore water hoseAir Purge Requirements 300 litre per min / 12.36 cu ft per minAir Purge Connection G3/8 suitable for 38 mm / 1.5 in inner diameter neo-

prene hoseWeight 8.4 kg / 18.5 lbDimensions (l x w x d) 393 x 275 x 240 mm / 15.47 x 10.83 x 9.45 inMounting Holes Ø 8.5 mm / 0.34 in to accept M8 screws

Page 32: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

Infrared Linescanning System

Installation Overview

Landscan LSP-HD

3 - 6

3.5 Light Industrial Housing - Part Nº 802664The Light Industrial Housing is used in applications where an LSP-HD scanner needs to be enclosed for additional protection due to airborne particulates. It can be installed into environments where the ambient temperatures are up to 150°C / 302°F. The air purge facility should be utilised where the operating environment is dusty and airborne particles could contaminate the scanner viewing window. A quick release mounting arrangement is included for the scanner, making realignment a simple and easy operation.

SpecificationsMaximum Ambient Temperature 150 °C / 302 °FCooling Water Requirements 1.5 to 2 litre per min / 0.4 to 0.53 US gal per min @

30 °C / 86 °FMaximum Water Pressure 5.88 Bar / 85 PSI

Cooling Water Connection ¼ in BSP for 9.5 mm / 3/8 in bore water hoseAir Purge Requirements 300 litre per min / 12.36 cu ft per minAir Purge Connection G3/8 suitable for 38 mm / 1.5 in inner diameter neo-

prene hoseWeight 14.4 kg / 32 lbDimensions (L x W x H) 310 x 290 x 300 mm / 12.2 x 11.42 x 11.81 inMounting Holes Ø 8.5 mm / 0.34 in to accept M8 screws

Dimensions All dimensions in mm/in

Page 33: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

3 - 7

Landscan LSP-HDInfrared Linescanning System

Installation Overview

3.6 Heavy Industrial Housing - Part Nº 802665TheHeavyIndustrialHousingisusedinapplicationswhereanLSP-HDscannerneedstobeenclosed for additional protection due to airborne particulates or for outdoor operation. It can be installedintoenvironmentswheretheambienttemperaturesareupto150°C/302°F.Theairpurgefacilityshouldbeutilisedwheretheoperatingenvironmentisdustyandairborneparticlescouldcontaminatethescannerviewingwindow.Aquickreleasemountingarrangementisincludedforthescanner,makingrealignmentasimpleandeasyoperation.

SpecificationsMaximumAmbientTemperature 150°C/302°FCoolingWaterRequirements 1.5to2litrepermin/0.4to0.53USgalpermin@

30°C/86°FMaximumWaterPressure 5.88Bar/85PSI

CoolingWaterConnection ¼inBSPfor9.5mm/3/8inborewaterhoseAirPurgeRequirements 300litrepermin/12.36cuftperminAirPurgeConnection G3/8suitablefor38mm/1.5ininnerdiameterneo-

prene hoseWeight 29kg/54lbDimensions(LxWxH) 511x320x296.2mm/20.12x12.8x11.67inMountingHoles Ø8.5mm/0.34intoacceptM8screws

MaterialofHousingCover SUS304

Dimensions Alldimensionsinmm/in

WaterIn/Outconnection1/4inBSPincoolingplatetoaccommodatewaterhoseØ9.5mm(3/8in)bore

GlandplatefittedwithM20cableglandforSTDcables031.402&802406.Forhightemperaturecable(031.794)replaceM20cableglandwithM16cablegland&adaptor(suppliedwithmountings),fixedinplaceusingM20cableglandnut

Fixingbolts&washers(notsuppliedbyLandInstruments)M16x35hexheadscrewc/wplainandspringwashers

PurgeairinletG3/8inthreadinpurgeplatePurgeadaptortofitrecommendedNeopreneairhoseØ38mm(1.5in)

511(20.1)472(18.6)436(17.2)

16.1

27.9

Scan line

Page 34: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

BLOWERS

4

Page 35: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

4 - 1

Landscan LSP-HDInfrared Linescanning System

Installation Overview

4 BlowersThePlateMountedBlowerunitprovidesairtotheLSP-HD.Eachspecificapplicationmayrequireacontrolledairsupply.Opticalairpurgingmayberequiredinordertomaintainaclearsightingwindow.Eachblowercansupportuptotwoscanners.

SpecificationsOption 1 Option 2 Option 3

PowerSupply 230Va.c. 110Va.c. 220va.c.Frequency 50Hz 60Hz 60HzPowerConsumption 400W 500W 500W

Circuit Breaker (TypeC)

10A 16A 10A

OperatingTemperature:

-20to50°C/-4to122°F(-40to50°C/-40to122°Fwithoptionalairheaterfitted)

EnvironmentalRating

IP55/NEMA4(IP55/NEMA4X(withweatherproofcoverfitted)

CEMarking 89/336/CEE;73/23/CEEWeatherProofing InaccordancewithEN60034-5

Dimensions Alldimensionsinmm/in

Page 36: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

POWER SUPPLIES

5

Page 37: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

5 - 1

Landscan LSP-HDInfrared Linescanning System

Installation Overview

5 Power SuppliesThissectionoftheguidecontainsinstallationinformationforthefollowingPowerSupplies:• LSP-HDDesktopPowerSupply(805024)• LSP-HDIndustrialPowerSupplyonaDINRail(807105)• LSP-HDIndustrialPowerSupplyinaTerminalBox(805037)

5.1 LSP-HD Desktop Power Supply - Part Nº 805024TheDesktopPowerSupplyisanunsealedpowersupplyforuseincleanenvironmentstoprovidepowerfortheLSP-HD.Note:ToconnecttheLSP-HDpowerandsignalcabletotheDesktopPowerSupply(PartNº805024)youwillneedtousetheM12toRJ45Cable(0.5m)PartNº805023.

SpecificationsInputPowerSupply 100to240Va.c.Instrument Connection 1xRJ45terminationDownstreamConnections 1xRJ45termination

Suppliedwith: M12 to RJ45 Adaptor CableMaximumAmbientTemperature 45°C/113°FDimensions(lxwxd) 166x80x43.6mm/6.48x3.15x1.72in

Page 38: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

Infrared Linescanning System

Installation Overview

Landscan LSP-HD

5 - 2

5.2 LSP-HD Industrial Power Supply on a DIN Rail - Part Nº 807105ThisisaDINrailmountedpowersupplyunitforfittinginsidecabinets.

SpecificationsInputPowerSupply 100to240Va.c.LSP-HDConnection 1xRJ45terminationDownstreamConnections 1xMulti-ModeFibreOpticSC

4xRJ45TerminationsMaximumAmbientTemperature 45°C/113°FDimensions(hxwxd) 170x200x200mm/6.8x7.9x7.9in

Note:Tocomplywithsafetyregulationsthesystemmust be mounted inside an enclosure

Page 39: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

5 - 3

Landscan LSP-HDInfrared Linescanning System

Installation Overview

5.3 LSP-HD Industrial Power Supply in a Terminal Box - Part Nº 805037ThisisaterminalboxwhichallowsconnectionoftheincomingLSP-HDconnectioncableviaanM12industrialEthernetport,incomingmainspowerandthedownstreamdataconnectionviaeitherCAT5/6cableormulti-modefibreoptic.Note: This unit is for indoor use only.

SpecificationsInputPowerSupply 100to240Va.c.Instrument Connection M124pinD-CodedIndustrialEthernetSocketDownstreamConnections 1xMulti-ModeFibreOpticSC

4xRJ45TerminationsEnvironmentalRating IP65MaximumAmbientTemperature 50°C/120°FDimensions(hxwxd) Seebelow

Dimensions Alldimensionsinmm/in

Page 40: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

CABLES

6

Page 41: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

6 - 1

Landscan LSP-HDInfrared Linescanning System

Installation Overview

6 CablesConnectioncablesbetweentheLSP-HDscannerandtheLSP-HDPowerSupplyunitarepre-terminatedwithM12industrialEthernetconnectors.ThissinglecableprovidesbothdataandPower-over-Ethernetconnections.

SpecificationsLSP-HD 15m Low Temperature

LSP-HD 30m Low Temperature

LSP-HD 15m High Temperature

PartNumber 805038 807087 804977Length 15m/49ft 30m/98ft 15m/49ftMaximumAmbientTemperature

80°C/176°F 80°C/176°F 200°C/390°F

Note:ToconnectanyoftheabovecablestotheDesktopPowerSupply(PartNº805024)youwillneedtheM12toRJ45Cable(0.5m)PartNº805023.

Page 42: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

SOFTWARE SETUP

7

Page 43: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

7 - 1

Landscan LSP-HDInfrared Linescanning System

Installation Overview

7 Software SetupTheinstallationoftheDataServerrequiresalicencekey.Toobtainthekeythefollowinginformationisrequired:• Copy Control Version• Product Name• Serial Number• Update NumberFromtheLandscandirectoryrun Register.exe to open the Change or Remove Registrationdialog.ThisregistrationinformationisusuallyavailabletoAMETEKLandengineersduringthecommissioningprocess.If you need to obtain your license information,contactAMETEKLandviatheemailfollowingaddress: [email protected].

7.1 Software RegistrationThefollowingadditionalinformationisrequiredtogeneratetheappropriatelicencekey:• Number of Scanners connected (max. 8)• Number of Live Clients connected (max. 4)• Type of external communication• Any Active X Licence for live dataThelicencecanbeinputtedtothesystemfromafile.CopythefileintotheLandscandirectoryand run Register.exe.Ifthekeyiscorrect,asmallpop-upboxappearsandasksforconfirmation.Oncedone,thesoftwareregistrationiscompleted.Anexistinglicencecanbeextendedatanytime.

Page 44: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

Infrared Linescanning System

Installation Overview

Landscan LSP-HD

7 - 2

7.2 Basic Software SetupThissectiongivesanoverviewaboutthebasicsystemsettingstosetuptheLandscansoftwareforvisualisation,dataprocessinganddatastorage• DoubleclickontheStart LandscandesktopicontostarttheSoftware:

7.2.1 Start ScreenAfterinitialsetup,theblankstartscreenisshown.

7.2.2 Basic ConfigurationFromtheConfiguremenu,selectSystem to enter the Basic Configurationdialog:

Page 45: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

7 - 3

Landscan LSP-HDInfrared Linescanning System

Installation Overview

7.2.3 Basic Configuration – Process tab1) Select the Units of Temperature,Units of Distance,Units of Product Width and

Units of Longitudunal Measurement for display setup and orientation.2) Definenamesforleftandrightsideoftheprocesstobemonitored.3) Definethenumberofscannersconnectedtothesoftware.Achangehererequiresa

rebootofthesoftware.

Page 46: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

Infrared Linescanning System

Installation Overview

Landscan LSP-HD

7 - 4

Orientation NamingTointerpretthescanprofileproducedbytheScanner,thescandirectionmustbeknown.ThisisindicatedbyanembossedarrowcastintothetopoftheScanner.TheScanneralwaysscansfromthelefttotheright,wherethearrowheadindicatestherightsideofthescan.TheProfiledisplayindicatestheLeftsidewith(-) and the Rightsidewith(+).

Page 47: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

7 - 5

Landscan LSP-HDInfrared Linescanning System

Installation Overview

7.2.4 Basic Configuration – Equipment tab1) Allocate the Network IP AddressoftheconnectedScannerandIOProcessor.2) Access the IO ConfigurationoftheconnectedIOprocessor.

3) Allocate the Network IP AddressoftheconnectedScannerandIOProcessor.4) Access the IO ConfigurationoftheconnectedIOprocessor.

Page 48: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

Infrared Linescanning System

Installation Overview

Landscan LSP-HD

7 - 6

5) ApplythenecessaryIPAddresschangeandconfirmwithOK.DoubleclickonthehighlightedlinetoseethenewIPAddressintheNetwork IP Addressbox.

6) DeactivateMaintenancetoactivatetheStation. The same action can be taken in the IO IP AddressboxtoperformanIPAddress

changeontheBeckhoffIOProcessor.

Page 49: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

7 - 7

Landscan LSP-HDInfrared Linescanning System

Installation Overview

7) ToallocatetheIOs,selecttheIO Configuration option. 8) ClicktheLogin button.

9) DoubleclickintheDevice Communication Partner IP Address and select the Scanner.

10) TicktheEnable Logincheckbox.11) EntertherequirednumberofIOsinChannels in Use by this partner.12) ConfirmanychangewithApply to IO System,thenLogout,thenOK. TheBeckhoffIOProcessorisnowsetupforcommunicationwiththeScanner.

Page 50: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

Infrared Linescanning System

Installation Overview

Landscan LSP-HD

7 - 8

7.2.5 Basic Configuration - IconsAsuccessfullyestablishedcommunicationbetweenWCASoftware,ScannerandBeckhoffIOisindicatedbythecolourofthestationiconsinthestatusbar:

Connection established

NoconnectionbetweenPCandScanner

NoconnectionbetweenScannerandBeckhoffIO

StationinMaintenance(BasicConfiguration:Equipmenttab)

ScannerinMaintenance(BasicConfiguration:LineScannertab)

Page 51: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

7 - 9

Landscan LSP-HDInfrared Linescanning System

Installation Overview

7.2.6 Basic Configuration – Product tab1) Select Date and Time for the Product Name.2) Select Automatic (Percentage of Peak) andavalueof40% for the Hot Edge

Detection.3) SettheappropriatevalueforEmissivity.4) Select Hot Edges for Product Determination to determine a product once the hot

edgesaredetected.5) SetanappropriatevaluefortheProduct Speed.IfanIOProcessorisconnectedandan

analoguespeedinputisapplied,theAnalogue speed input... option can be selected.

Page 52: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

Infrared Linescanning System

Installation Overview

Landscan LSP-HD

7 - 10

7.2.7 Basic Configuration – Processor Input tabThistabshowstheactiveAnalogue Inputs and Digital InputsofaconnectedBeckhoffIOProcessor.Ifananaloguespeedsignalisused,selecttheProcess SpeedFunction,themA Range(either0or4to20mA)andtherelevantScale.Thisexampleshows4to20mAfor0to8000mm/sec.AdigitalinputcanbeactivatedtotriggertheproductdeterminationbyselectingtheProcess Input function.ThisrequirestheactivationofDigital Input at LPU/LSC and allocation of the input channel on the Product tab.

Page 53: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

7 - 11

Landscan LSP-HDInfrared Linescanning System

Installation Overview

7.2.8 Basic Configuration – Line Scanner tab1) Select the required Scanner Speedinarangebetween10and150Hz.2) Enter the Distance from product(thisisthedistancebetweenScannerWindowand

product surface).3) The Scan Anglecanbealteredbetween40and80degreesandwillbeshowninthe

relevantdisplays.4) Select the Enable Internal Ambient Temperature Sensor option to display the

ambient temperature of the Scanner.

Page 54: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

Infrared Linescanning System

Installation Overview

Landscan LSP-HD

7 - 12

7.3 Data Processing

7.3.1 Control Zone Processing1) FromtheData Processingmenu,selectControl Processing to apply Control Zones.2) Double-clickintothelineofeachZonetodisplaythedialog,whichallowsyoutosetthe

relevantZoneparameters: • Units:either%ormm • Reference:Start,CentreorEndofScanline • Size:StartandEndPosition • Tracking:withinHotEdgesorfixedposition • Processing:e.g.Peak,Minetc.

Page 55: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

7 - 13

Landscan LSP-HDInfrared Linescanning System

Installation Overview

7.3.2 Data Processing - Zone AlarmsZone AlarmscanbeappliedtoeachindividualControlZone.1) Select the required Level.2) Select the required Type,eitherHigh or Low.

Page 56: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

Infrared Linescanning System

Installation Overview

Landscan LSP-HD

7 - 14

7.3.3 Data Processing - Analogue Output Processing1) ToallocatethesetControlZonestoananalogueOutput,therelevantChannelshaveto

beactivated:Channel On.2) Select Zone Value as the Function for the selected Channel.3) The Function Parameter:1,2,3etc.indicatethereferencetotheindividualControl

Zone.4) The Rangecanbeselectedaseither0or4to20mAoutputsignal.

Page 57: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

7 - 15

Landscan LSP-HDInfrared Linescanning System

Installation Overview

7.3.4 Data Processing - Digital Output Processing• DigitalOutputsarecommonlyusedforZone Alarms,eitherHigh/LowAlarmsortogive

awarningsignaliftheScannerAmbienttemperaturesexceedthespecifiedlimits.• The reference for the Zone Alarm is the Function Parameterwhichindicatesthe

allocated Control Zone.• If Ambient Hi Warningisselected,therelaytripsat65°C:theAmbient Hi Alarm

tripsat70°C.

Page 58: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

Infrared Linescanning System

Installation Overview

Landscan LSP-HD

7 - 16

7.4 Display

7.4.1 Display - New Display1) FromtheDisplaymenu,selectNew Display.2) FromtheSelectdatasource,selecttheMeasurement Station and the option Real

Time Scanner.ThiswillprovideadisplaywhichshowscontinuouslythelivedatageneratedbytheScanner.

3) Oncethedatasource(MeasurementStation)isset,theDisplay Type can be selected. The most common display types are Profile and Map.

Page 59: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

7 - 17

Landscan LSP-HDInfrared Linescanning System

Installation Overview

7.4.2 Display - ProfileThe ProfileDisplayisasimplelineplotofthescanlineoftemperature.

ProfileDisplay

ProfileDisplayshowingpositionsofControlZones:RightmouseclickintotheDisplayandselectShow Control Zones.

Page 60: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

Infrared Linescanning System

Installation Overview

Landscan LSP-HD

7 - 18

7.4.3 Display - MapThe Mapdisplaysthetemperaturesamplesinascanaspointsofcolourarrangedinaverticalline.Ascaleonthesideofthemapdefinesthetemperaturerangerepresentedbyeachindividualcolour.

Page 61: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

7 - 19

Landscan LSP-HDInfrared Linescanning System

Installation Overview

7.5 Storage Management

7.5.1 General tabFromtheStoragemenu,selectManagementtoactivatedatastorageanddatabasesettings.

1) OntheGeneraltabselecttherequiredDatabasesizetodefinethelimitoftheDatabase2) WhenDelete the oldest productsisselected,theoldestproductfilewillbedeleted

fromtheDatabasewhenthesetlimitisreached.3) Select a Storage Location.ItisrecommendedtosettheDatabaseonadifferentdrive

or a partition.

Page 62: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

Infrared Linescanning System

Installation Overview

Landscan LSP-HD

7 - 20

7.5.2 Station tab1) OntheStationtab,selecttheStorage Mode. If Store Productsisselected,products

willbestoredautomaticallyassoonasdetectedbytheScanner.2) In Storage Data Reductionachoiceofoptionsisavailabletoreducethesizeof

productfiles,e.g. • Product Reduction • Line Reduction • Sample Reduction3) ActivateStorage EnabledforeachStation.StoragewillstartassoonasSystem

StorageisactivatedontheGeneral tab.

Page 63: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

7 - 21

Landscan LSP-HDInfrared Linescanning System

Installation Overview

7.5.3 LandscanDatabase1) OntheStationtab,selecttheStorage Mode. If Store Productsisselected,products

willbestoredautomaticallyassoonasdetectedbytheScanner.2) In Storage Data Reductionachoiceofoptionsisavailabletoreducethesizeof

productfiles,e.g. • Product Reduction • Line Reduction • Sample Reduction3) ActivateStorage EnabledforeachStation.StoragewillstartassoonasSystem

StorageisactivatedontheGeneral tab.

Page 64: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

Infrared Linescanning System

Installation Overview

Landscan LSP-HD

7 - 22

7.6 Diagnostics

7.6.1 ScannerFromtheDiagnosticsmenu,selectScannertodisplaytheScannerStatuswindow.

The Scanner Statuswindowdisplaystheactualscanspeedandtheambienttemperatureofthe Scanner.

7.6.2 LPU/LSC –Internal Event LogThe Internal Event Log tab in the LPU/LSC DiagnosticswindowdisplaystheVersion InformationoftheconnectedLSP-HDScannerorLSCProcessor.

Page 65: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

7 - 23

Landscan LSP-HDInfrared Linescanning System

Installation Overview

7.6.3 LPU/LSC – Test OutputsThe Test Outputs tab in the LPU/LSC DiagnosticswindowdisplaysatestmodefacilitytocheckanalogueanddigitaloutputsofaconnectedBeckhoffIOProcessor.SimplyactivatetheOverride the LPU/LSC Outputsoptionandcheckthedigitalandanalogueoutputsbyactivatingthetickboxesandmovingthesliders.

Page 66: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

Infrared Linescanning System

Installation Overview

Landscan LSP-HD

7 - 24

7.7 Save ConfigurationFromtheFilemenu,selectSave Configuration.

OncethesetupofWCAsoftware(withallnecessarysettings)hasfinished,selecttheSave ConfigurationoptionfromFilemenu.TheactualsystemconfigurationwillbestoredasMasterConfig.mdb in the ConfigDatabase folder of the Landscan directory.

7.8 Change ConfigurationFromtheFilemenu,selectChange Configuration.ConfirmtheReplace on Restart MessagewithYes.Thesoftwarewillrebootthenandwillaskyoutochoosetheconfigurationdatabasetobeused.SelectUse a previously saved database,theOpenwindowwillthendisplaytheconfigurationsstoredintheConfigDatabase folder.

Page 67: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

7 - 25

Landscan LSP-HDInfrared Linescanning System

Installation Overview

7.9 Save LPU/LSC ConfigurationFromtheFilemenu,selectSave LPU/LSC Configuration.TheactualScannerConfigurationwillbestoredasLPU2Config_1.mdb in the Workspace folderoftheLandscandirectory.

7.10 Change LPU/LSC ConfigurationFromtheFilemenu,changeSave LPU/LSC Configuration.ConfirmtheCautionmessagewithYesandselecttheMeasurementStation.FromtheOpen dialog,selectthesavedconfigurationfilefromtheWorkspacefolderandconfirmwithOpen.

Page 68: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

MAINTENANCE & SERVICE

8

Page 69: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

8 - 1

Landscan LSP-HDInfrared Linescanning System

Installation Overview

8 Maintenance & ServiceThe body of the scanner and the mountingscanbecleanedbywipingwithacleanclothmoistenedwitheitherwateroramild,neutraldetergent.

The scanner windowcanbecleanedbyusinganysoftclothalongwithamildkitchentypedetergentcleaner.Thescannerwindowissapphireandsoresistsscratching,acidsorsolvents.

Hoses and pipesshouldbecheckedregularlyforleaksand/orcracks.Westronglyrecommendtouseflow meters/indicatorstomonitortheairandwaterflow.

If the cleanliness of the water supplycannotbeguaranteed,thea100~150µmparticlefiltershouldbeinstalledinthesupplyfeed.Aby-passshouldbeavailableforwhenthefilterhas to be replaced.

The recommended calibration interval for this system is 12 months.

Page 70: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

PRODUCT WARRANTYThank you for purchasing your new product from AMETEK Land Instruments International. This Land manufacturer’s ‘back-to-base’ warranty covers product malfunctions arising from defects in design or manufacture. The warranty period commences on the instrument despatch date from the AMETEK Land Instruments International Ltd. factory in Dronfield, UK.

12 MONTHS WARRANTYLandscan LSP-HD Linescanners are provided with a 12 months warranty.

EXCLUSIONS FROM WARRANTYIt should be noted that costs associated with calibration checks which may be requested during the warranty period are not covered within the warranty.Land reserve the right to charge for service/calibration checks undertaken during the warranty period if the cause is deemed to fall outside the terms of the warranty.This Land manufacturer’s warranty does not cover product malfunction arising from:-• incorrect electrical wiring.• connection to electrical power sources outside the rating of the product.• physical shock (being dropped, etc.) and impact damage.• inappropriate routing, support, physical shock & strain protection, etc. of the lightguide (Fibroptic

thermometers only).• environmental conditions exceeding the IP / NEMA rating of the product.• environmental conditions outside the Ambient Temperature, Humidity and Vibration rating of the product.• environmental contamination (solvent vapours, deposition of airborne contamination, cooling liquids of

non-neutral pH, etc.).• overheating as a result of interruption of water/air flow through cooling jackets or of incorrect installation.• inappropriate modification of product (drilling holes in thermometer bodies, etc.).• inappropriate recalibration which results in product calibration being taken outside specification.• improper resealing of thermometer following parameter adjustment.• attempted repair by a non-Land-authorised repair centre.

Land Instruments International Ltd • Dronfield S18 1DJ • England • Tel: +44 (0) 1246 417691 • Fax: +44 (0) 1246 410585 Email: [email protected] • www.landinst.comAMETEK Land, Inc . • 150 Freepor t Rd. • P i t t sburgh, PA 15238 • U.S.A. • Te l : +1 (412) 826 4444 Email: [email protected] • www.ametek-land.com

For a complete list of our international offices, please visit www.landinst.com Issue 1: 20 September 2016

Page 71: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

PRODUCT WARRANTYThank you for purchasing your new product from AMETEK Land Instruments International. This Land manufacturer’s ‘back-to-base’ warranty covers product malfunctions arising from defects in design or manufacture. The warranty period commences on the instrument despatch date from the AMETEK Land Instruments International Ltd. factory in Dronfield, UK.

18 MONTHS WARRANTYLandscan LSP-HD Linescanners are provided with an 18 months warranty.

EXCLUSIONS FROM WARRANTYIt should be noted that costs associated with calibration checks which may be requested during the warranty period are not covered within the warranty.Land reserve the right to charge for service/calibration checks undertaken during the warranty period if the cause is deemed to fall outside the terms of the warranty.This Land manufacturer’s warranty does not cover product malfunction arising from:-• incorrect electrical wiring.• connection to electrical power sources outside the rating of the product.• physical shock (being dropped, etc.) and impact damage.• inappropriate routing, support, physical shock & strain protection, etc. of the lightguide (Fibroptic

thermometers only).• environmental conditions exceeding the IP / NEMA rating of the product.• environmental conditions outside the Ambient Temperature, Humidity and Vibration rating of the product.• environmental contamination (solvent vapours, deposition of airborne contamination, cooling liquids of

non-neutral pH, etc.).• overheating as a result of interruption of water/air flow through cooling jackets or of incorrect installation.• inappropriate modification of product (drilling holes in thermometer bodies, etc.).• inappropriate recalibration which results in product calibration being taken outside specification.• improper resealing of thermometer following parameter adjustment.• attempted repair by a non-Land-authorised repair centre.

Land Instruments International Ltd • Dronfield S18 1DJ • England • Tel: +44 (0) 1246 417691 • Fax: +44 (0) 1246 410585 Email: [email protected] • www.landinst.comAMETEK Land, Inc. • 150 Freeport Rd. • Pittsburgh, PA 15238 • U.S.A. • Tel: +1 (412) 826 4444 Email: [email protected] • www.ametek-land.com

MARCOM0411 For a complete list of our international offices, please visit www.landinst.com Issue 1: 28 September 2016

Page 72: LSP-HD System Installation Overview (Document Nº 811372) Website/infrared...Infrared Linescanning System Landscan LSP-HD Installation Overview Contents 1 Introduction 1-1 1.1 About

For a full list of international offices, please visit our website www.landinst.com Copyright © 2017 LAND Instruments International. Continuous product development may make it necessary to change these details without notice. LSP-HD System Installation Overview, Issue 2, 08 February 2017

LANDSCAN LSP-HDINFRARED LINESCANNERS

Applies in the UK Applies in the US Applies in India

Land Instruments InternationalDronfield, S18 1DJ, England

Tel: +44 (0) 1246 417691Email: [email protected] www.landinst.com

AMETEK Land - Americas 150 Freeport Road, Pittsburgh, PA 15238, U.S.A.