longitudinal diagnostics of electron bunches using coherent transition radiation

22
Fermilab, Jan. 16, 2007 Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation Daniel Mihalcea Northern Illinois University Department of Physics

Upload: baker-vang

Post on 03-Jan-2016

40 views

Category:

Documents


2 download

DESCRIPTION

Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation. Daniel Mihalcea. Northern Illinois University Department of Physics. Outline:. Fermilab/NICADD overview Michelson interferometer Bunch shape determination Experimental results Conclusions. - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation

Fermilab, Jan. 16, 2007

Longitudinal Diagnostics of Electron Bunches Using

Coherent Transition Radiation

Daniel Mihalcea

Northern Illinois University

Department of Physics

Page 2: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation

•Fermilab/NICADD overview

• Michelson interferometer

• Bunch shape determination

• Experimental results

• Conclusions

Outline:

Page 3: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation

Fermilab NICADD Photo-injector Laboratory

FNPL is a collaborative effort of several institutes and universities to operate a high-brightness photo-injector dedicated to accelerator and beam physics research.

Collaborators include:

U. of Chicago, U. of Rochester, UCLA, U. of Indiana, U. of Michigan, LBNL, NIU, U. of Georgia, Jlab, Cornell University

DESY, INFN-Milano, IPN-Orsay, CEA-Saclay

Page 4: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation

FNPL layout

Michelson interferometer for longitudinal diagnostics

Page 5: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation

Michelson Interferometer

(University of Georgia & NIU)

Autocorrelation = I1/I2 Detectors: - Molectron pyro-electric

- Golay cell (opto-acoustic)

Page 6: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation

Data Flow

Interferometer

ICT Detectors Stepping motor

Scope

Controller

Get Q Get I1 and I2

LabView code: Advances stepping motor between x1 and x2 with adjustable step size (50 m)

At each position there are N readings (5)

A reading is valid if bunch charge is within some narrow window (Ex: 1nC 0.1 nC)

Position, average values of I1 and I2 and their ’s are recorded.

Autocorrelation function is displayed.

Page 7: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation

Basic Principle (1)

222

22

03

22

)cos1(

sin

4

c

e

dd

UdIe

Ginsburg-Franck:

)(fIe

Detector aperture 1 cm

Backward transition radiation

Page 8: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation

Basic Principle (2)

2)()1()( fINNI eIntensity of Optical Transition Radiation:

Coherent part N2

Form factor )(f related to longitudinal charge distribution: dzczizf )/exp()()(

To determine (z) need to know I() and the phase of f()

dttEctEI2

1 )()/(

dttE

dtctEtE

I

IS 2

2

1

)(

)/()(Re)(

dE

deES

ci

2

/2

)(

)(Re)( deSEI ci

/2)()()(

Kramers-Kröning technique

Page 9: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation

Coherence condition

z mm3max

Due to detector sensitivity:

Acceptable resolution: mmz 6.0 Need bunch compression !

Page 10: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation

Bunch Compression RF field in booster cavity

Electron bunch before compression

Energy-Position correlation

Head

Tail

After compression

mm1

mm15.0

2566 c

Tt f

Page 11: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation

Measurement Steps

Ideal apparatus

mmz 19.0

2/

FT

K-K

mmz 18.0

Kramers-Kröning method:

dczIc

z

x

IxIdx

/)(cos)(1

)(

)(/)(ln)(

0

022

Page 12: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation

Experimental results (1)

Path difference (mm)

Frequency (THz)c

ndE

22 sin||

Molectron pyro-electric detectors

Interference effect Missing frequencies

Page 13: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation

Experimental results (2)

Golay detectors: no problem with interference !

Still need to account for:

low detector sensitivity at low frequencies

diffraction at low frequencies

absorption at large frequencies

Page 14: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation

Experimental results (3)

Low detector sensitivity

DiffractionAbsorption

Interference

Apparatus response function:

)(

)()(

simulation

measured

I

IR

Beam conditions:

Q = 0.5 nC

maximum compression

Page 15: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation

Experimental results (4)

Auto-correlation function:

Q = 3nC

9-cell phase was 3 degrees from maximum compression

Power spectrum:

Asymptotic behavior

low frequencies:

high frequencies:

Least square fit.

21)( aI

)(I

Molectron pyroelectric detectors

Page 16: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation

Experimental results (5)

• Molectron pyroelectric detectors

• Kramers-Kroning method

Head

Tail

Parmela simulation

Head-Tail ambiguity

Page 17: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation

Experimental results (6)

FT

Spectrum correction with R()

Spectrum completion for:

andzc /0 0

Beam conditions:

Q = 3.0 nC

moderate compression

K-K

Golay cell

Start point

z 1ps

Page 18: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation

Complicated bunch shapes

Stack 4 laser pulses

Select 1st and 4th pulses (t 15ps)

Before compression

After compression

(Parmela simulations)

Page 19: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation

Experimental results (7)

Double-peaked bunch shapes

Beam conditions:

Q = 0.5 nC each pulse

15 ps initial separation between the two pulses

both pulses moderately compressed

K-K method may not be accurate for complicate bunch shapes !

Page 20: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation

K-K method accuracy

R. Lai and A. J. Sievers, Physical Review E, 52, 4576, (1995)

Generated

Reconstructed

K-K method accurate if:

• Simple bunch structure

• Stronger component comes first

Calculated widths are still correct !

Page 21: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation

Other approaches

Major problem: the response function is not flat.

1. Complete I() based on some assumptions at low and high frequencies.

R. Lai, et al. Physical Review E, 50, R4294, (1994).

S. Zhang, et al. JLAB-TN-04-024, (2004).

2. Avoid K-K method by assuming that bunches have a predefined shape and make some assumptions about I() at low frequencies.

A. Murokh, et al. NIM A410, 452-460, (1998).

M. Geitz, et al. Proceedings PAC99, p2172, (1999).

This work:

D. Mihalcea, C. L. Bohn, U. Happek and P. Piot, Phys. Rev. ST Accel. Beams 9, 082801 (2006).

Page 22: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation

Conclusions:

Longitudinal profiles with bunch lengths less than 0.6 mm can be measured.

Systematic uncertainties dominated by approximate knowledge of response function and completion procedure.

Golay cells are better because the response function is more uniform.

Some complicate shapes (like double-peaked bunches) can be measured.