lecture 9 - quality assurance

Upload: linda

Post on 06-Jul-2018

221 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/17/2019 Lecture 9 - Quality Assurance

    1/41

    FRCR: Physics Lectures

    Diagnostic Radiology

    Lecture 9Quality Assurance (QA) of radiographic

    systems

    Dr im !ood

    Clinical "cientist

  • 8/17/2019 Lecture 9 - Quality Assurance

    2/41

    #$er$ie%

    & !hat is QA'

    & he life cycle of ray imaging systems

     * he critical e+amination

     * Acceptance testing * Commissioning

     * Routine performance testing

    & ,P-. Report 9/

     * Recommended tests 0y modality * a %histle stop tour 

    & he role of .edical Physics in Diagnostic

    Radiology

  • 8/17/2019 Lecture 9 - Quality Assurance

    3/41

    !hat is QA'

    &   Quality as·sur·ance

    &  A programme for the systematic

    monitoring and e$aluation of the $arious

    aspects of a pro1ect2 ser$ice or facility to

    ensure that standards of 3uality are 0eing

    met

    .erriam!e0ster4s .edical Dictionary2 5 6778 .erriam!e0ster2 ,nc

  • 8/17/2019 Lecture 9 - Quality Assurance

    4/41

    !hat is QA'

    & ,t is a re3uirement under the Ionising Radiations

    Regulations 1999 (Reg 6()(;)) to<

     * ‘… make arrangements for a suitable quality

    assurance programme to be provided in respect of

    the equipment or apparatus for the purpose of

    ensuring that it remains capable of restricting so far

    as is reasonably practicable exposure to the extent

    that this is compatible with the intended clinical

     purpose or research obective!" 

  • 8/17/2019 Lecture 9 - Quality Assurance

    5/41

    !hat is QA'

    & Regulation 6 also re3uires=

     * Ade3uate testing 0efore entering clinical ser$ice

    (commissioning)

     * Ade3uate testing at appropriate inter$als and afterany ma1or maintenance procedure (routine)

     * .easurements at suita0le inter$als to ena0le

    assessment of representati$e doses to persons

    undergoing medical e+posures

  • 8/17/2019 Lecture 9 - Quality Assurance

    6/41

    !hat is QA'

    & he regulations use lots of $ague terms li>e?suita0le@2 ?ade3uate@ and ?appropriate@ * Deli0erate due to range of e3uipment the regulations

    ha$e to co$er 

    & he ?Appro$ed Code of Practice@ gi$es slightly  more detail * Depends on the nature and range of e3uipment in use * he QA programme should specify the fre3uency of

    any testing and appropriate actions le$els

     *  In establishing these levels the employer shouldtake into account guidance established byrelevant professional bodies about criteria ofacceptability

  • 8/17/2019 Lecture 9 - Quality Assurance

    7/41

    !hat is QA'

    & ,n de$ising the QA programme2 pay specialattention to e3uipment for= * Children * ealth screening programmes

     * igh dose procedures2 such as inter$entionalradiology2 C or radiotherapy

    & he employer should consult their RPA %hende$ising the QA programme

    & Bote QA2 does not 1ust co$er the e3uipment2 0utalso procedures2 etc (as re3uired under,R(.-)R)

  • 8/17/2019 Lecture 9 - Quality Assurance

    8/41

    he life cycle of ray imaging

    systems

    & here are essentially four stages of chec>s

    applica0le to ray imaging systems=

     * Critical e+amination

     * Acceptance * Commissioning

     * Routine performance testing

    & .aintenance is $ital throughout

  • 8/17/2019 Lecture 9 - Quality Assurance

    9/41

    he life cycle of ray imaging

    systems

  • 8/17/2019 Lecture 9 - Quality Assurance

    10/41

    he Critical -+amination

    & nder ,RR 99 Reg /(6)2 the installer  isre3uired to perform a critical examination ofany ne% installation

    & he purpose of the ?crite+@ is to ensure all safety

    features and %arning de$ices %or> correctly * Also includes tu0e lea>age2 total filtration2 etc

    & Primarily related to radiation protection featuresthat affect staff and $isitors2 0ut some impact on

    patient safety too& -mployer must not allo% the e3uipment to gointo clinical ser$ice until the results of the crite+are satisfactory

  • 8/17/2019 Lecture 9 - Quality Assurance

    11/41

     Acceptance testing

    & erify that the contractor has supplied all thee3uipment specified and has performedade3uate tests to demonstrate specifiedre3uirements are met

    & .ay 0e a simple chec> list

    & .echanical and electrical safety chec>s alsore3uired

  • 8/17/2019 Lecture 9 - Quality Assurance

    12/41

    Commissioning

    & "et of tests performed 0y a representati$e of the

    employer (usually .edical Physics)2 to ensure

    the e3uipment is ready for clinical use2 and to

    esta0lish 0aseline $alues against %hich routineQA can 0e compared

    & Commissioning tests may need to 0e repeated

    during the life of the system if any ma1or

    maintenance is underta>en eg ne% ray tu0e * Be% 0aselines may need to 0e esta0lished

  • 8/17/2019 Lecture 9 - Quality Assurance

    13/41

    Routine testing

    & Regular tests throughout the lifetime of thee3uipment

    & Loo>ing for changes in performance thatindicates remedial action re3uired

    & Eenerally2 routine are a su0set of thecommissioning tests

  • 8/17/2019 Lecture 9 - Quality Assurance

    14/41

    ,P-. Report 9/

    & he ,nstitute of Physics and-ngineering in .edicine (,P-.)produce a series of reports relatedto medical e3uipment QA

    & he most useful for DiagnosticRadiology is ,P-. Report 9/ * Recommended standards for the

    routine performance testing ofdiagnostic #$ray imaging systems 

    (677)& Relates primarily to imaging

    performance and radiation safetychec>s

  • 8/17/2019 Lecture 9 - Quality Assurance

    15/41

    ,P-. Report 9/

    & he report is split into chapters on differentmodalities

    & -ach chapter starts %ith a summary ta0le of= *  The physical parameter to be tested

     *   Frequency * $aries from daily to three yearly *   Priority * le$el / is essential for ?good practice@= le$el

    6 is not essential2 0ut considered ?0est practice@ *  Level of expertise required * le$el A applies to

    fre3uent and relati$ely 0asic tests performed 0y

    Radiographers= le$el G tests are less fre3uent2 0utre3uire greater e+pertise and more comple+e3uipment (.edical Physics tests)

     *   ction levels * split into ?Remedial@ and ?"uspension@(see later)

  • 8/17/2019 Lecture 9 - Quality Assurance

    16/41

    ,P-. Report 9/

    & -ach test is then descri0ed 0riefly2%ith appropriate references to otherdocuments2 such as the ,P-. Report6 series * hese are a series of reports that gi$e

    much greater detail on the method fortesting systems eg ray tu0es andgenerators2 fluoroscopy2 C2 etc

  • 8/17/2019 Lecture 9 - Quality Assurance

    17/41

    ,P-. Report 9/

    &   !emedial Level:

     * A le$el of performance at %hich remedial

    action is re3uired2 0ut the unit may continue to

    0e used in the meantime * he action %ill 0e 0ased on a ris> assessment

    of the e3uipments performance and the ris>

    arising should it continue to 0e used

     * Follo%ing assessment2 a timescale must 0e

    agreed and restrictions on its use follo%ed

  • 8/17/2019 Lecture 9 - Quality Assurance

    18/41

    ,P-. Report 9/

    &   "uspension Level:

     * A le$el of performance at %hich it is

    recommended the e3uipment is remo$ed from

    clinical use immediately * Bot all tests ha$e suspension le$els set due

    to the su01ecti$e nature of the test eg image

    3uality

  • 8/17/2019 Lecture 9 - Quality Assurance

    19/41

    ,P-. Report 9/

    & ,P-. 9/ also emphasises that= * A senior radiography or other suita0le person should

    0e appointed to super$ise the QA programme

     * ime should 0e allocated to staff and e3uipment for

    testing *  ll Q tests should be documented as part of the

    Q programme * may 0e re3uired as e$idencepresented to the "- inspectors

     * Results and remedial actions must 0e follo%ed uppromptly

     * est e3uipment should 0e a$aila0le and %ithincali0ration (annual recali0ration usually re3uired fordosemeters)

  • 8/17/2019 Lecture 9 - Quality Assurance

    20/41

     A %histle stop tour of ,P-. 9/

  • 8/17/2019 Lecture 9 - Quality Assurance

    21/41

    ray tu0es and generators(Chapter )

    &   Level

     * rayHlight 0eam alignment and centring

     * Light 0eamH0uc>y centring

     * Field siIe cali0ration

     * Distances and scales

     * Radiation output repeata0ility and

    reproduci0ility (small range of settings)

  • 8/17/2019 Lecture 9 - Quality Assurance

    22/41

    ray tu0es and generators(Chapter )

    &   Level #

     * Radiation output repeata0ility and

    reproduci0ility (larger range of settings)

     * -+posure time * u0e potential

  • 8/17/2019 Lecture 9 - Quality Assurance

    23/41

    FilmHscreen radiography2

    processors and A-Cs

    (Chapter ;)

    &   Level

     * De$eloper temperature2 fog2 film speed2

    contrast inde+2 replenishment2 p2 sil$ercontent

     * ,ntensifying screens and dar>room chec>s

     * A-C guard timer and resulting film #D

  • 8/17/2019 Lecture 9 - Quality Assurance

    24/41

    FilmHscreen radiography2

    processors and A-Cs

    (Chapter ;)

    &   Level #

     * #nly A-C tests

     * Consistency 0et%een cham0ers

     * Repeata0ility and reproduci0ility

     * Receptor dose

  • 8/17/2019 Lecture 9 - Quality Assurance

    25/41

    CR and DR(Chapters and J)

    &   Level $

     * Detector dose indicator monitoring

     * niformity

     * Condition of image plates (CR only)

     * Lo% contrast sensiti$ity

     * Limiting spatial resolution

  • 8/17/2019 Lecture 9 - Quality Assurance

    26/41

    CR and DR(Chapters and J)

    &   Level #$ (in con1unction %ith ,P-. 6 Part ,,) * DD, repeata0ility and reproduci0ility

     * .easured uniformity

     * hreshold contrast detail detecta0ility

     * -rasure cycle efficiency (CR only)

     * Limiting spatial resolution

     * "caling errors

     * Dar> noise

     * .odulation transfer function (.F)

     * Bormalised noise po%er spectrum (BBP")

     * A-C tests

  • 8/17/2019 Lecture 9 - Quality Assurance

    27/41

    ,mage display(Chapter 8)

    &   Film vie%ers *  Level  * general condition *  Level # * luminance2 uniformity2

    $ariation 0et%een $ie%ers2 room

    illumination&   &onitors *  Level  * general condition2

    greyscale and resolution %ith testpatterns

     *  Level # * D,C#. greyscalecali0ration2 test patterns2 uniformity2$ariation 0et%een monitors2 roomillumination

  • 8/17/2019 Lecture 9 - Quality Assurance

    28/41

    .ammography(Chapter K)

    & "ee also ,P-. Report K9 ?%he commissioningand routine testing of mammographic #$ray

    systems@ and the latest B"G"P reports

    & B"G"P -3uipment Report 7J7; for full fielddigital testing

    & Lots of testing re3uired

     * Radiographer testing daily2 %ee>ly2 monthly

     * Physics testing e$ery J months

  • 8/17/2019 Lecture 9 - Quality Assurance

    29/41

    .ammography(Chapter K)

    & -+ample tests include= * Processing2 %here applica0le

     * A-C

     * Limiting spatial resolution

     * ,mage 3uality %ith test phantoms

     * .ED to standard 0reast

     * ray 0eam alignment

     * Compression force

     * > accuracy (specific cali0rations)

     * niformity

     * Radiation output repeata0ilityHreproduci0ility

     * And much more<

  • 8/17/2019 Lecture 9 - Quality Assurance

    30/41

    Dental Radiography(Chapter 9)

    &   Processing tests * emperature2 solutions2 step

    %edge2 light proofing

    &   '(ray)detector tests * u0e $oltage2 e+posure time2collimation2 dose at collimator tip

    for ,#2 DAP for #PE2 detector

    condition2 clinical image 3uality

    compared %ith reference

  • 8/17/2019 Lecture 9 - Quality Assurance

    31/41

    Fluoroscopy(Chapter /7)

    &   Level

     * Dose rate reproduci0ility under AGCHA-C

     * Display monitor setup

     * Limiting spatial resolution

     * hreshold contrast

     * RadiationHimage field siIe

  • 8/17/2019 Lecture 9 - Quality Assurance

    32/41

    Fluoroscopy(Chapter /7)

    &   Level #

     * Dose rate at the entrance surface of a

    phantom under AGCHA-C

     * Dose rate at the input face of the detectorunder AGCHA-C

     * Limiting spatial resolution

     * hreshold contrast

  • 8/17/2019 Lecture 9 - Quality Assurance

    33/41

    Fluorography(Chapter //)

    &   Level

     * Dose per image under A-C

     * Limiting spatial resolution

     * hreshold contrast

  • 8/17/2019 Lecture 9 - Quality Assurance

    34/41

    Fluorography(Chapter //)

    &   Level #

     * Dose per image at the input face of the

    detector under A-C

     * Limiting spatial resolution * hreshold contrast

     * Dynamic range

  • 8/17/2019 Lecture 9 - Quality Assurance

    35/41

    Patient dose measurements(Chapter /)

    &   Perform dose audits and)or Q measurements * ,ndi$idual radiographic e+posures

    & -ntrance surface dose or DAP M Diagnostic Reference Le$els (DRLs)

     * Complete e+aminations

    & DAP M DRLs

     * .ammography& .ED to standard 0reast M6 mEy per film

    & .ED to patients M mEy per film

     * Dental Radiography& ,# dose at collimator tip M DRL

    & #PE dose M DRL

     * Fluoroscopy& Dose at entrance surface of phantom M 7 mEyHmin

    & DAP M DRL

  • 8/17/2019 Lecture 9 - Quality Assurance

    36/41

    C(Chapter /6)

    & Craig to co$er<

  • 8/17/2019 Lecture 9 - Quality Assurance

    37/41

    QA is important 0ecause<

    & QA is important 0ecause< * Can identify e3uipment deterioration

     * Can 0e used to achie$e ALARP

     *-nsures patient (and staff) safety

     * -nsures legislati$e compliance

    & Final thought<

     * Although it@s not your responsi0ility to chec>

    e3uipment2 you should ensure that the ray sets you

    use ha$e 0een chec>ed for your o%n safety2 as %ell

    as the patient@s

    h l f . di l Ph i i

  • 8/17/2019 Lecture 9 - Quality Assurance

    38/41

    he role of .edical Physics in

    Diagnostic Radiology

    &   Q * Le$el G tests * Ad$ising on QA programme and monitoring performance

    &   !adiation protection and physics advice

    *!P)&P+, * -nsuring compliance %ith rele$ant regulations eg ris>

    assessments2 controlled areas2 dose monitoring2 etc * ,RR and ,R(.-)R audits * ma>e sure %e@re ready if the

    inspector calls

     * Be% installations shielding2 testing2 etc * Botification of incidents to the CQC2 "-2 -A2 Police2 etc

    &   +ducation * FRCR2 pdate raining2 etc

    h l f . di l Ph i i

  • 8/17/2019 Lecture 9 - Quality Assurance

    39/41

    he role of .edical Physics in

    Diagnostic Radiology

    &   -ptimisation * !e aren@t 1ust here to interfere and police the

    regulations

     * "ignificant part of the Radiation Physics groups %or>

    is loo>ing at ho% %e can get the most out of our ray

    imaging systems

     * Rely on feed0ac> and cooperation from Radiology to

    ensure %e are optimising e+posures

     * !e@re here to help<

  • 8/17/2019 Lecture 9 - Quality Assurance

    40/41

    Recent pu0lications

    &   #eavis2 . "aunderson2 N !ard 2!-EGRA7J: Cali0rating an ,onisation Cham0er: Eaining -+perience sing a Dosimetry ?flight "imulator@2 .ed Phys 92 987 (67/6)

    &   /" &oore2 E A$ery2 " Galcam2 L Beedler2 A "%ift2 0 #eavis1 .!"aunderson 2 se of a digitally reconstructed radiograph 0ased computersimulation for the optimisation of chest radiographic techni3ues for

    computed radiography imaging systems 2 Gr N Radiol 67/6=K eJ7eJ9&   T. 0ood1 0 2avis1 /" &oore1 0 #eavis 3 .! "aunderson 2

    alidation of a Large"cale Audit echni3ue for C Dose #ptimisation 2Radiat Prot Dosimetry (67/6) /7(;): ;68;

    &   /" &oore1 4P Liney1 0 #eavis1 .! "aunderson2 A method to produceand $alidate a digitally reconstructed radiograph0ased computer simulationfor optimisation of chest radiographs ac3uired %ith a computed radiography

    imaging system2 Gritish Nournal of Radiology2 K; (67//)2 K97976&   &L 0ilson1 0P /olley 3 0 #eavis 2 he effect of the car0on fi0re insert

    for the arian -+actO couch on the attenuation and 0uildup of high energyphoton 0eams2 Nournal of Radiotherapy in Practice (67//) /72 88K

    http://www.aapm.org/meetings/2012AM/PRAbs.asp?mid=68&aid=18282http://www.aapm.org/meetings/2012AM/PRAbs.asp?mid=68&aid=18282http://online.medphys.org/resource/1/mphya6/v39/i6/p3970_s2?isAuthorized=nohttp://online.medphys.org/resource/1/mphya6/v39/i6/p3970_s2?isAuthorized=nohttp://bjr.birjournals.org/content/85/1017/e630.abstracthttp://bjr.birjournals.org/content/85/1017/e630.abstracthttp://bjr.birjournals.org/content/85/1017/e630.abstracthttp://rpd.oxfordjournals.org/content/150/4/427.abstracthttp://rpd.oxfordjournals.org/content/150/4/427.abstracthttp://bjr.birjournals.org/cgi/content/abstract/84/1006/890http://journals.cambridge.org/action/displayAbstract?fromPage=online&aid=8316032&fulltextType=RA&fileId=S1460396910000373http://journals.cambridge.org/action/displayAbstract?fromPage=online&aid=8316032&fulltextType=RA&fileId=S1460396910000373http://bjr.birjournals.org/cgi/content/abstract/84/1006/890http://rpd.oxfordjournals.org/content/150/4/427.abstracthttp://rpd.oxfordjournals.org/content/150/4/427.abstracthttp://bjr.birjournals.org/content/85/1017/e630.abstracthttp://bjr.birjournals.org/content/85/1017/e630.abstracthttp://bjr.birjournals.org/content/85/1017/e630.abstracthttp://online.medphys.org/resource/1/mphya6/v39/i6/p3970_s2?isAuthorized=nohttp://www.aapm.org/meetings/2012AM/PRAbs.asp?mid=68&aid=18282http://www.aapm.org/meetings/2012AM/PRAbs.asp?mid=68&aid=18282

  • 8/17/2019 Lecture 9 - Quality Assurance

    41/41

    Current Pro1ects

    & Further de$elopment of our dose audit soft%are

    & C A-C optimisation %ith a custom 0uilt phantom

    & De$elopment of a high s>in dose reporting system

    for inter$entional procedures& De$elopment of a ne% test o01ect for routine

    image 3uality assessment in digital

    mammography

    &  And many more< %%%hullradorgu> 

    http://www.hullrad.org.uk/http://www.hullrad.org.uk/