Laser Speckle ExtensometerME 53
Class 1 Device
MESSPHYSIKMATERIALS TESTING
-8 -6 -4-2
02
46
8 -8-6
-4-2
02
46
8
© MESSPHYSIK Materials Testing
Version 2.3
![Page 2: Laser Speckle Extensometer ME 53 Class 1 Device MESSPHYSIK MATERIALS TESTING © MESSPHYSIK Materials Testing Version 2.3](https://reader036.vdocuments.mx/reader036/viewer/2022062515/56649c7c5503460f9492fe74/html5/thumbnails/2.jpg)
Features:Non-contacting measurement
Works with most materials
High temperature measurement
1 Laser diodes
2 Cameras (CCD)
3 Speckle Patterns
4 Specimen
Micro Tensile Specimen measurement
![Page 3: Laser Speckle Extensometer ME 53 Class 1 Device MESSPHYSIK MATERIALS TESTING © MESSPHYSIK Materials Testing Version 2.3](https://reader036.vdocuments.mx/reader036/viewer/2022062515/56649c7c5503460f9492fe74/html5/thumbnails/3.jpg)
Principles of Displacement Measurement with Laser Speckle CorrelationThe optically rough surface of a specimen reflects coherent light.Laser diode modules with a wave length of λ = 660 nm are directedtowards the specimen.The reflected light passing through the laser beam creates so-called „Speckle Patterns“.The image size of the CCD camera is called „Field of View“.
Speckle = spot, dab or point
Speckle Pattern Image Size = Field of View (FOV)
![Page 4: Laser Speckle Extensometer ME 53 Class 1 Device MESSPHYSIK MATERIALS TESTING © MESSPHYSIK Materials Testing Version 2.3](https://reader036.vdocuments.mx/reader036/viewer/2022062515/56649c7c5503460f9492fe74/html5/thumbnails/4.jpg)
Requirements for displacement measurement with standard application:
2 Cameras = 2 Speckle Patterns
Distance between cameras = 50mm
Lenses … 75f x2 / 3,9 F
with 2x extender we receivef=150mm
With these lenses a distance of 480 mm between specimen and cameras is possible
Principles of Displacement Measurement with Laser Speckle Correlation
![Page 5: Laser Speckle Extensometer ME 53 Class 1 Device MESSPHYSIK MATERIALS TESTING © MESSPHYSIK Materials Testing Version 2.3](https://reader036.vdocuments.mx/reader036/viewer/2022062515/56649c7c5503460f9492fe74/html5/thumbnails/5.jpg)
Principles of Laser Speckle Correlation
To measure the displacement of the surface, it‘s necessary to follow themovements of the Speckle Patterns.
This is achieved by means of a cross-correlation algorithm comparing two successive patterns.
Master Slave
While cameras and laser modules remain in a fixed position, the surface of the specimen moves when extended.
This comparison is carried out up to twenty times per second.Measuring rate = 20 Hz
![Page 6: Laser Speckle Extensometer ME 53 Class 1 Device MESSPHYSIK MATERIALS TESTING © MESSPHYSIK Materials Testing Version 2.3](https://reader036.vdocuments.mx/reader036/viewer/2022062515/56649c7c5503460f9492fe74/html5/thumbnails/6.jpg)
The Speckle Pattern
The size of the evaluated patterns can be set to up to 128x128 pixels.
The cross-correlation (FFT) requires high computing capacity. Therefore PCs with a minimum CPU speed of 500 MHz are needed to reach measuring rates of 20 Hz.In order to measure up to 200Hz a 3 GHz Processor and High Speed DigitalCameras are required.
![Page 7: Laser Speckle Extensometer ME 53 Class 1 Device MESSPHYSIK MATERIALS TESTING © MESSPHYSIK Materials Testing Version 2.3](https://reader036.vdocuments.mx/reader036/viewer/2022062515/56649c7c5503460f9492fe74/html5/thumbnails/7.jpg)
The Correlation Peak
-8 -6 -4-2
02
46
8 -8-6
-4-2
02
46
8
The result of the cross-correlation is a function, where the position of the maximum (Correlation Peak) corresponds with the displacement of the pattern.
A special interpolation algorithmdetermines the position of thecorrelation peak in the subpixel range and compensates for changesof the surface in real-time.
Parameters for this interpolationalgorithm can be defined by the user.
![Page 8: Laser Speckle Extensometer ME 53 Class 1 Device MESSPHYSIK MATERIALS TESTING © MESSPHYSIK Materials Testing Version 2.3](https://reader036.vdocuments.mx/reader036/viewer/2022062515/56649c7c5503460f9492fe74/html5/thumbnails/8.jpg)
-8 -6 -4-2
02
46
8 -8-6
-4-2
02
46
8
The peak is intersected by several planes parallel to thexy – plane.
The results are sectional planes which are evaluated in two dimensions.The centres of gravity are regarded as the position of the peak.
With this procedure it is possibleto reach a very accurate displacement measurements while saving a lot of calculation time.
S
![Page 9: Laser Speckle Extensometer ME 53 Class 1 Device MESSPHYSIK MATERIALS TESTING © MESSPHYSIK Materials Testing Version 2.3](https://reader036.vdocuments.mx/reader036/viewer/2022062515/56649c7c5503460f9492fe74/html5/thumbnails/9.jpg)
Calibration
Calibrating the LSE, the Speckle pattern must be moved for a defined stretch. This way the size of the FOVs (Field of Views) and theCalibration factors can be determined.
The standard stretch for the calibration is 2 mm. The sample is mounted in one grip only and shifted for exactly 2 mm.
This stretch can be as long as needed and adjusted in the LSE Software.
The size of the FOVs is calculated, the result saved and the LSE isready for measuring.
If the distance between specimen and camera has changed, theLSE has to be re-calibrated.
![Page 10: Laser Speckle Extensometer ME 53 Class 1 Device MESSPHYSIK MATERIALS TESTING © MESSPHYSIK Materials Testing Version 2.3](https://reader036.vdocuments.mx/reader036/viewer/2022062515/56649c7c5503460f9492fe74/html5/thumbnails/10.jpg)
Features• 2 - dimensional strain measurement (longitudinal and transversal)
![Page 11: Laser Speckle Extensometer ME 53 Class 1 Device MESSPHYSIK MATERIALS TESTING © MESSPHYSIK Materials Testing Version 2.3](https://reader036.vdocuments.mx/reader036/viewer/2022062515/56649c7c5503460f9492fe74/html5/thumbnails/11.jpg)
Features• Displacement & 3 Point – Deflection Measurement
Features
![Page 12: Laser Speckle Extensometer ME 53 Class 1 Device MESSPHYSIK MATERIALS TESTING © MESSPHYSIK Materials Testing Version 2.3](https://reader036.vdocuments.mx/reader036/viewer/2022062515/56649c7c5503460f9492fe74/html5/thumbnails/12.jpg)
FeaturesTriangulation Sensor Head ME53-33
This application turning the cameras allows L0 setups from 0-250mm
Features
![Page 13: Laser Speckle Extensometer ME 53 Class 1 Device MESSPHYSIK MATERIALS TESTING © MESSPHYSIK Materials Testing Version 2.3](https://reader036.vdocuments.mx/reader036/viewer/2022062515/56649c7c5503460f9492fe74/html5/thumbnails/13.jpg)
Features
Two dimensional strain, multiple strain
and deflection measurement
Testing all kind of materials:
Metals
Plastics
Textiles
Rubbers
Triangulation Sensor Head ME53-33
applied to a Midi 10-20 Machine
Features
![Page 14: Laser Speckle Extensometer ME 53 Class 1 Device MESSPHYSIK MATERIALS TESTING © MESSPHYSIK Materials Testing Version 2.3](https://reader036.vdocuments.mx/reader036/viewer/2022062515/56649c7c5503460f9492fe74/html5/thumbnails/14.jpg)
Features
Works with most materials
Steel Aloy
Rubber Plastics
Features
![Page 15: Laser Speckle Extensometer ME 53 Class 1 Device MESSPHYSIK MATERIALS TESTING © MESSPHYSIK Materials Testing Version 2.3](https://reader036.vdocuments.mx/reader036/viewer/2022062515/56649c7c5503460f9492fe74/html5/thumbnails/15.jpg)
Features• 800° C high temperature measurement done in Australia, University of
Technology in Brisbane
Features
![Page 16: Laser Speckle Extensometer ME 53 Class 1 Device MESSPHYSIK MATERIALS TESTING © MESSPHYSIK Materials Testing Version 2.3](https://reader036.vdocuments.mx/reader036/viewer/2022062515/56649c7c5503460f9492fe74/html5/thumbnails/16.jpg)
Features
• compression test of all kind of material
• compression tests at high temperatures up to 1600 °C
Features
![Page 17: Laser Speckle Extensometer ME 53 Class 1 Device MESSPHYSIK MATERIALS TESTING © MESSPHYSIK Materials Testing Version 2.3](https://reader036.vdocuments.mx/reader036/viewer/2022062515/56649c7c5503460f9492fe74/html5/thumbnails/17.jpg)
Features
illuminated specimenLaser Speckle Extensometer looking throughtemperature chamber glas window
Features
![Page 18: Laser Speckle Extensometer ME 53 Class 1 Device MESSPHYSIK MATERIALS TESTING © MESSPHYSIK Materials Testing Version 2.3](https://reader036.vdocuments.mx/reader036/viewer/2022062515/56649c7c5503460f9492fe74/html5/thumbnails/18.jpg)
Features
• 0.1 µm resolution measurement of micro specimen• Calibration by marks on the surface of the specimen
Features
![Page 19: Laser Speckle Extensometer ME 53 Class 1 Device MESSPHYSIK MATERIALS TESTING © MESSPHYSIK Materials Testing Version 2.3](https://reader036.vdocuments.mx/reader036/viewer/2022062515/56649c7c5503460f9492fe74/html5/thumbnails/19.jpg)
FeaturesMicro Specimen curve
Features
![Page 20: Laser Speckle Extensometer ME 53 Class 1 Device MESSPHYSIK MATERIALS TESTING © MESSPHYSIK Materials Testing Version 2.3](https://reader036.vdocuments.mx/reader036/viewer/2022062515/56649c7c5503460f9492fe74/html5/thumbnails/20.jpg)
MESSPHYSIKMATERIALS TESTING
Credits
Phillip WinkelmayerDipl. Ing. Oliver Spinka
© Messphysik Materials Testing, Oktober 2005