Page 1: Katrin Dahl  for the AEI 10 m Prototype team

Katrin Dahl for the AEI 10 m Prototype teamSeptember 2009 –AEI seminar

Suspension Platform Interferometer for the AEI 10 m Prototype Interferometer

Introductory talk

AEI 10m Prototype

Page 2: Katrin Dahl  for the AEI 10 m Prototype team

2

Outline

1. AEI 10 m Prototype Interferometer– > Why is an SPI needed?

2. Relative distance measurement experiments

3. Experimental setup design

4. Test setup

5. Future plans

Page 3: Katrin Dahl  for the AEI 10 m Prototype team

3

The AEI 10 m Prototype

Goals:

• Train people for GEO600

• Prove new techniques (PSL, digital control system, Khalili cavity…)

• Provide ultra low displacement noise testing environment– To probe at (and later go beyond) the SQL– Entanglement of macroscopic test masses– For geodesy/LISA related experiments – ...

Page 4: Katrin Dahl  for the AEI 10 m Prototype team

4

Noise spectrum

Page 5: Katrin Dahl  for the AEI 10 m Prototype team

5

Vacuum system

Tubes:1.5 m diameter

Volume ca. 100 m3 22 t stainless steel

Tanks:3 m diameter, 3.4 m tall

Tank centers separated by 11.65m

Roughing: One 170 l/s screw pump Main pumps: Two 2000 l/s turbo-molecular pumps Backing and differential pumping: Two scroll pumps

10-6 mbar after about 12 hours , 10-7 mbar after about 2 weeks of pumping

Page 6: Katrin Dahl  for the AEI 10 m Prototype team

6

Seismic Attenuation System

• One SAS per vacuum tank, optical table goes on top of SAS

• Improved version of HAM-SAS

• Resonance frequency around 0.1 Hz

• Up to 80 dB attenuationin both vertical and

horizontal directions

• Angstrom residual motion above 1 Hz

Page 7: Katrin Dahl  for the AEI 10 m Prototype team

7

Why a Suspension Platform Interferometer?

• Ease lock acquisition of cavities by reducing residual test mass motion

• Reduction of burden to actuators on the mirrors

• Testbed for GRACE follow-on and LISA related experiments Sets requirements on SPI

Page 8: Katrin Dahl  for the AEI 10 m Prototype team

8

Horizontal table actuation

Page 9: Katrin Dahl  for the AEI 10 m Prototype team

9

Vertical table actuation

Page 10: Katrin Dahl  for the AEI 10 m Prototype team

10

Vertical table actuation

Page 11: Katrin Dahl  for the AEI 10 m Prototype team

11

A little bit of history

LIGO-T870001-00-R

Page 12: Katrin Dahl  for the AEI 10 m Prototype team

12

Suspension Point Interferometer

• monitors differential motion of the suspension points of input and end test masses

LIGO-T070209-00-Z

Page 13: Katrin Dahl  for the AEI 10 m Prototype team

13

Suspension Point Interferometer

LIGO-T070209-00-Z

Page 14: Katrin Dahl  for the AEI 10 m Prototype team

14

Digital Interferometry

LIGO-T080139-00-I

Advantages:

• no specific lock point• continuous sensing• track the position of mirrors over many µm

Page 15: Katrin Dahl  for the AEI 10 m Prototype team

15

Stabilised metrology testbed

K Numata, J Camp, Proc. of SPIE Vol. 6265

around 1 m

Page 16: Katrin Dahl  for the AEI 10 m Prototype team

16

Stabilised metrology testbed

K Numata, J Camp, Proc. of SPIE Vol. 6265

Control bandwidth 10 Hz

Yaw angle motion of 20 nrad/sqrt(Hz) at 10 mHz leading to about 50 times worse result for only one controlled degree of freedom

Page 17: Katrin Dahl  for the AEI 10 m Prototype team

17

THE AEI SPI

• Requirements:– No specific lock point– Control bandwidth 100 Hz– 100 pm/sqrt(Hz) and 10 nrad/sqrt(Hz) @ 10 mHz

• Heterodyne Mach-Zehnder interferometry– Suits our needs best– In-house knowledge ....Thanks LTP/LISA folks!

Page 18: Katrin Dahl  for the AEI 10 m Prototype team

18

Heterodyne Mach-Zehnder IFO

Page 19: Katrin Dahl  for the AEI 10 m Prototype team

19

Optical layout

Page 20: Katrin Dahl  for the AEI 10 m Prototype team

20

Measurement bench

• Beam height 45 mm• Overall height below 65 mm

Page 21: Katrin Dahl  for the AEI 10 m Prototype team

21

Phase determination

Phase is extracted from heterodyne signal by use of an hardware Phasemeter based on FPGA chips

1. Preamplifier and A/D conversion– Photocurrent converted to voltage– Digitising signals– results in time series

2. Single bin discrete Fourier transform – Fourier transform at only one frequency– complex amplitude of PD signal at fhet

3. Signal combination of each QPD quadrant leads to phase, DC, Differential Wavefront Sensing (DWS) and contrast information

Illustration of DWS

Page 22: Katrin Dahl  for the AEI 10 m Prototype team

22

Choice of parameters

• Due to the arm length mismatch (20 m optical path length) a highly stabilised laser is necessary:

• 280 Hz/sqrt(Hz) @ 10 mHz for green light (532 nm) or• 140 Hz/sqrt(Hz) @ 10 mHz for IR light (1064 nm)• decision made for 1064 nm

Page 23: Katrin Dahl  for the AEI 10 m Prototype team

23

Iodine stabilised Nd:YAG laser

Michael Tröbs

outputpower: 1 W

Stabilisation via Modulation Transfer Spectroscopy

Page 24: Katrin Dahl  for the AEI 10 m Prototype team

24

Choice of parameters

• According to the arm length mismatch (20 m optical path length) a highly stabilised laser is necessary:

• 280 Hz/sqrt(Hz) @ 10 mHz for green light (532 nm) or• 140 Hz/sqrt(Hz) @ 10 mHz for IR light (1064 nm)• decision made for 1064 nm

• Control bandwidth 100 Hz heterodyne frequency around 20 kHz new phasemeter interface needed

Page 25: Katrin Dahl  for the AEI 10 m Prototype team

25

Phasemeter Interface

Page 26: Katrin Dahl  for the AEI 10 m Prototype team

26

Choice of parameters

• According to the arm length mismatch (20 m optical path length) a highly stabilised laser is necessary:

• 280 Hz/sqrt(Hz) @ 10 mHz for green light (532 nm) or• 140 Hz/sqrt(Hz) @ 10 mHz for IR light (1064 nm)• decision made for 1064 nm

• Control bandwidth 100 Hz heterodyne frequency around 20 kHz

• Thermal drifts requires components to be monolithically bonded to plate with low CTE (ClearCeram, CTE=0.4*10-7/K)

Page 27: Katrin Dahl  for the AEI 10 m Prototype team

27

UV curing epoxy

• Advantage: almost infinite alignment time• Optocast 3553-LV-UTF

– CTE = 55 PPM/°C, viscosity @ 25 °C = 500 cps– cps = centipoise, water = 1 cps, castrol oil = 1,000 cps, honey = 10,000 cps,

ketchup = 50,000 cps

• Disadvantage: layer thickness is not reproducible, e.g. 10 µm, 98 µm, 79 µm, 70 µm, 37 µm misalignment in pitch

• stick to hydroxide-catalysis bonding technique

Page 28: Katrin Dahl  for the AEI 10 m Prototype team

28

Optical layout

Page 29: Katrin Dahl  for the AEI 10 m Prototype team

29

Expected transversal signalsco

ntra

stDC

Phas

e di

ffere

nce

[rad

]DW

S [r

ad]

-2 -1 0 1 2Transveral displacement of MW1 or MS1 [mm]

-2 -1 0 1 2Transveral displacement of MW1 or MS1 [mm]

-2 -1 0 1 2Transveral displacement of MW1 or MS1 [mm]

-2 -1 0 1 2Transveral displacement of MW1 or MS1 [mm]

Red curve: PDCW1Black curve: PDCS1

Page 30: Katrin Dahl  for the AEI 10 m Prototype team

30

Expected longitudinal signalsco

ntra

stDC

Phas

e di

ffere

nce

[rad

]DW

S [r

ad]

-2 -1 0 1 2Longitudinal displacement of MW1 or MS1 [mm]

-2 -1 0 1 2Longitudinal displacement of MW1 or MS1 [mm]

-2 -1 0 1 2Longitudinal displacement of MW1 or MS1 [mm]

-2 -1 0 1 2Longitudinal displacement of MW1 or MS1 [mm]

Red curve: PDCW1Black curve: PDCS1

Page 31: Katrin Dahl  for the AEI 10 m Prototype team

31

Expected rotational signalsco

ntra

stDC

Phas

e di

ffere

nce

[rad

]DW

S [r

ad]

-10 -5 0 5 10Rotation of MW1 or MS1 [mdeg]

-10 -5 0 5 10Rotation of MW1 or MS1 [mdeg]

-10 -5 0 5 10Rotation of MW1 or MS1 [mdeg]

-10 -5 0 5 10Rotation of MW1 or MS1 [mdeg]

Red curve: PDCW1Black curve: PDCS1

Page 32: Katrin Dahl  for the AEI 10 m Prototype team

32

Test setup

• Use of vacuum compatible components (free of grease)

Page 33: Katrin Dahl  for the AEI 10 m Prototype team

33

Longitudinal displacement

Page 34: Katrin Dahl  for the AEI 10 m Prototype team

34

Pitch

Page 35: Katrin Dahl  for the AEI 10 m Prototype team

35

Yaw

Page 36: Katrin Dahl  for the AEI 10 m Prototype team

36

Blind test

Page 37: Katrin Dahl  for the AEI 10 m Prototype team

37

Blind test

Page 38: Katrin Dahl  for the AEI 10 m Prototype team

38

Next steps

• Stabilisation loops– Amplitude stabilisation @ 20 kHz– Optical pathlenght difference stabilisation

• Bond optics– Build calibrated QPD

• Use CDS via phasemeter interface• Install final setup inside vacuum envelope• Calibrate signals• Table actuation• Reach design sensitivity

Page 39: Katrin Dahl  for the AEI 10 m Prototype team

39

THE END