internal filtration in dialyzers with different membrane permeabilities
TRANSCRIPT
BRIEF COMMUNICATION
Internal filtration in dialyzers with different membranepermeabilities
Yuichi Sato • Kenjiro Kimura • Tatsuya Chikaraishi
Received: 23 October 2009 / Accepted: 23 March 2010 / Published online: 5 June 2010
� The Japanese Society for Artificial Organs 2010
Abstract Over the last decade, hemodialysis with
enhanced internal filtration (IF) has been investigated as an
alternative to conventional dialysis. Several factors affect
IF, including the geometry and permeability of hollow-
fiber dialyzers. Although various studies have been
performed, the association between IF and membrane
permeability has not been fully examined because of the
difficulty in measuring IF. Therefore, in this study, we set
up an experimental circuit and attempted to directly mea-
sure IF as well as membrane permeability in five dialyzers.
In the circuit, we placed two dialyzers of the same type in
series, and a special sampling port between them, thereby
making it possible to determine IF by measuring the extent
to which blood was concentrated between the two dialyz-
ers. We showed that a significant amount of IF occurred in
this tandem-dialyzer circuit, ranging from 23.5 to 100 ml/min,
which increased linearly with increasing membrane per-
meability. We also showed that membrane permeability
was reduced in the first dialyzer to a greater extent than in
the second one after four hours of circulation, suggesting
that filtration caused substantial membrane fouling. In this
study we practically demonstrated that membrane perme-
ability is highly relevant to the phenomenon of IF.
Keywords Convective solute removal �Internal filtration � Dialysis
Introduction
Internal filtration (IF) in a hemodialyzer has been vigor-
ously investigated because hemodialysis with enhanced IF
dialyzers may serve as a clinical alternative to conventional
dialysis [1–6]. Previous studies have identified several
factors affecting IF, including the geometry and perme-
ability of hollow-fiber dialyzers. In theory, hollow fibers
with greater membrane permeability induce larger amounts
of IF, yet the relationship between them has not been fully
examined because of the difficulty involved in measuring
IF. In this study, we set up an experimental circuit
involving two dialyzers in series with a blood sampling
port between them. The circuit enabled us to directly
quantify IF by determining the extent to which blood was
concentrated by the IF phenomenon. We report the amount
of IF in several dialyzers with different membrane per-
meabilities but with almost identical geometries.
Materials and methods
Membrane permeability in dialyzers
We employed five cellulose triacetate dialyzers, FB-150G,
FB-150U, FB-150F, FB-150UH, and FB-150FH (Nipro,
Osaka, Japan). These dialyzers have nearly identical
geometries in their hollow fibers, as shown in Table 1. We
determined the pre-circulation ultrafiltration rate (UFR) of
each dialyzer as a marker of membrane permeability. We
placed each dialyzer in a circuit filled with reverse osmosis
Y. Sato (&) � T. Chikaraishi (&)
Department of Urology,
St. Marianna University School of Medicine,
2-16-1 Sugao, Miyamae-ku, Kawasaki 216-8511, Japan
e-mail: [email protected]
T. Chikaraishi
e-mail: [email protected]
K. Kimura
Department of Hypertension and Nephrology,
St. Marianna University School of Medicine, Kawasaki, Japan
123
J Artif Organs (2010) 13:113–116
DOI 10.1007/s10047-010-0506-z
(RO) water. While performing straightforward ultrafiltra-
tion at three different rates (Qf = 900, 1,800, 2,700 ml/h),
with two dialysate ports open to the atmosphere, we
measured the blood-side and dialysate-side hydraulic
pressures by a manometer, before and after the dialyzer, to
calculate the transmembrane pressure (TMP). We then
determined UFR using the following equation:
UFR (ml=h mmHg) ¼ Qf (ml/h)=TMP (mmHg)
TMP : ðPBinþ PBoutÞ=2 � ðPDinþ PDoutÞ=2;
PBin: blood-side pressure at the blood inlet, PBout: blood-
side pressure at the blood outlet, PDin: dialysate-side
pressure at the dialysate inlet, PDout: dialysate-side pres-
sure at the dialysate outlet.
After obtaining three sets of data for each dialyzer, UFR
was determined by a linear approximation using Microsoft
Excel X for Mac.
Determination of IF in the experimental circuit
To quantify the amount of IF in the dialyzers, we set up a
special circuit containing two dialyzers of identical type
that were placed in series with a blood sampling port
between them. There we circulated bovine blood at
200 ml/min from a 4-l reservoir and delivered dialysate at a
rate of 500 ml/min in the countercurrent manner with zero
net filtration. Anticoagulation was achieved by adding
heparin at 1,000 unit/h, and the hematocrit of bovine blood
was adjusted to 30–34% using normal saline.
In this tandem-dialyzer circuit, where the effective
length of the hollow fibers is doubled, the resultant increase
in TMP induces greater amounts of IF. We assume that a
substantial amount of filtration occurs in the first dialyzer,
which is counterbalanced by the same amount of backfil-
tration in the second one. Owing to the special blood
sampling port we provided, we were able to measure how
much the blood was concentrated between the two dia-
lyzers, thereby quantifying IF. After circulating for an
hour, we drew blood and determined the amount of IF (QIF)
using the following equation:
QIF ¼ 200ð1� HctPre=HctMidÞ;
Table 1 Characteristics of five
hollow-fiber dialyzers and
membrane permeabilities before
circulation, represented by UFR
Length
(mm)
Diameter
(lm)
Density
(%)
Number Thickness
(lm)
UFR
(ml/h mmHg)
FB-150G 227 200 50 10,300 15 35
FB-150U 227 200 50 10,300 15 116
FB-150F 227 200 50 10,300 15 160
FB-150UH 227 200 50 10,300 15 288
FB-150FH 227 185 47 11,000 15 322
where HctPre: Hct before the first dialyzer, HctMid: Hct
between the two dialyzers.
Change in UFR after circulation
After circulating for 4 h, we measured the post-circulation
UFRs in the first and second dialyzers in the same manner
as the pre-circulation UFRs were measured in a circuit
filled with RO water.
Results
The membrane permeabilities in five dialyzers, represented
by the pre-circulation UFRs, are shown in Table 1. In the
tandem-dialyzer circuit, the hematocrit value substantially
increased between the two dialyzers. In FB-150F, for
instance, it increased from 33 to 42%, and thus the amount
of IF was calculated to be 43 ml/min. As shown in Fig. 1,
IF increased with increasing membrane permeability.
In three types of dialyzer (FB-150U, FB-150F, FB-
150UH), circulation continued for 4 h, although we found
varying degrees of residual blood inside the dialyzers upon
completion. We measured the post-circulation UFRs in the
first and second dialyzers. In FB-150U, the UFRs were
52.9 and 68.6 ml/h mmHg; in FB-150F, they were 69.2
and 105.9 ml/h mmHg; in FB-150UH, they were 116.1 and
153.2 ml/h mmHg. In all three dialyzers, the post-circula-
tion UFRs were lower than the pre-circulation UFRs. In
addition, the post-circulation UFR was lower in the first
dialyzer than in the second one for all three types of
dialyzer. In the remaining two types (FB-150G and
FB-150FH), circulation did not last longer than 3 h because
of a remarkable increase in the venous pressure due to
blood clotting inside the dialyzer.
Discussion
It has been suggested that convective treatment such as
hemodiafiltration is associated with improved conditions of
114 J Artif Organs (2010) 13:113–116
123
patients with end-stage renal disease [7–12]. However,
such treatment has not been widely practiced because it
requires additional processes and equipment, especially in
the preparation of substitution fluids. Over the past decade
hemodialysis with enhanced IF has been investigated [1–6]
because it can be an efficient and readily applicable alter-
native in the treatment of dialysis patients.
Theoretical analysis has shown that the structure and
permeability of hollow fibers play crucial roles in pro-
ducing IF [3]. By increasing the total length, or decreasing
the inner diameter, the blood-side pressure drops rapidly
from the inlet to the outlet, thus increasing the local
transmembrane pressure, leading to an enhancement of IF.
In this study, we intended to focus on the relation between
IF and membrane permeability. Although membrane per-
meability, in theory, is highly relevant to IF, data on it are
scarce, mainly because it is difficult to directly measure IF.
Previously, some methods have been introduced to deter-
mine IF [5, 6], although they are somewhat cumbersome or
inaccurate. In order to directly measure IF and investigate
the association between IF and membrane permeability, we
set up a special circuit containing two dialyzers in series
with a blood sampling port between them. Using this
experimental circuit, we were able to measure IF by
determining the extent to which blood was concentrated
between the two dialyzers, which was a phenomenon
caused by IF.
We found that the amount of IF increased with
increasing measured UFR. It was noted that one dialyzer
(FB-150FH) seemed to have disproportionately larger
amounts of IF (Fig. 1), which could be explained by the
slightly narrower fiber (185 lm) of this specific dialyzer,
which further enhanced IF compared with its 200-micron
counterparts. In three types of dialyzer (FB-150U, FB-
150F, FB-15UH), circulation continued for more than 4 h
in the tandem circuit. In all of the dialyzers, the post-
circulation UFR was lower than the pre-circulation UFR,
which indicates that protein adhesion to the membrane
(i.e., membrane fouling) during the circulation has a neg-
ative impact on UFR. In addition, post-circulation UFR
was decreased in the first dialyzer, which suggests that IF
in the first dialyzer causes more membrane fouling than in
the second dialyzer, where backfiltration mainly occurred.
This result is in accordance with the results of Yamamoto
et al. [2], who found that membrane fouling caused by IF in
a dialyzer tends to occur near the dialysate outlet port.
There are obvious limitations to the present study. Dif-
ferent bovine blood samples were used in each experiment,
with only the hematocrit values adjusted; a blood sampling
port with a short tubing between the two dialyzers could
modify the hydrodynamics in the circuit, which potentially
affects QIF. Moreover, although QIF was simulated to take
the largest value near the center of a dialyzer [3], it might
not necessarily do so between the two dialyzers of this
experimental set-up.
Taking all of these shortcomings into account, the
results are admittedly qualitative rather than quantitative.
The experimental setting is different from a clinical one.
However, given the difficulty involved in quantifying IF,
we believe that our direct measurements of IF have
advanced knowledge in this research field.
Conclusion
We have demonstrated a strong correlation between IF and
membrane permeability; the measured IF increased linearly
with increasing membrane permeability in five different
dialyzers. After the circulation, UFR was decreased,
especially in the first dialyzer, suggesting that the IF caused
significant membrane fouling, which led to decreased water
permeability of the membrane.
References
1. Mineshima M, Ishimori I, Sakiyama R. Validity of internal
filtration-enhanced hemodialysis as a new hemodiafiltration
therapy. Blood Purif. 2009;27:33–7.
2. Yamamoto K, Hiwatari M, Kohori F, Sakai K, Fukuda M,
Hiyoshi T. Membrane fouling and dialysate flow pattern in
an internal-filtration enhancing dialyzer. J Artif Organs.
2005;8:198–205.
3. Mineshima M, Ishimori I, Ishida K, Hoshino T, Kaneko I, Sato Y,
Agishi T, Tamamura N, Sakurai H, Masuda T, Hattori H. Effects
of internal filtration on the solute removal efficiency of a dialyzer.
ASAIO J. 2000;46:456–60.
4. Dellanna F, Wuepper A, Baldamus CA. Internal filtration—
advantage in haemodialysis? Nephrol Dial Transpl. 1996;11
(Suppl 2):83–6.
5. Ronco C, Brendolan A, Feriani M, Milan M, Conz P, Lupi A,
Berto P, Bettini M, La Greca G. A new scintigraphic method to
80
100
120
20
40
60
QIF
(m
l/min
)
00 50 100 150 200 250 300 350
UFR (ml/hr mmHg)
Fig. 1 Relationship between membrane permeability and internal
filtration in five dialyzers (FB-150G, FB-150U, FB-150F, FB-150UH,
FB-150FH, from left to right)
J Artif Organs (2010) 13:113–116 115
123
characterize ultrafiltration in hollow fiber dialyzers. Kidney Int.
1992;41:1383–93.
6. Sato Y, Mineshima M, Ishimori I, Kaneko I, Akiba T, Teraoka S.
Effect of hollow fiber length on solute removal and quantification
of internal filtration rate by Doppler ultrasound. Int J Artif
Organs. 2003;26:129–34.
7. Leypoldt JK, Cheung AK, Carroll CE, Stannard DC, Pereira BJ,
Agodoa LY, Port FK. Effect of dialysis membranes and middle
molecule removal on chronic hemodialysis patient survival. Am J
Kidney Dis. 1999;33:349–55.
8. Kuchle C, Fricke H, Held E, Schiffl H. High-flux hemodialysis
postpones clinical manifestation of dialysis-related amyloidosis.
Am J Nephrol. 1996;16:484–8.
9. Koda Y, Nishi S, Miyazaki S, Haginoshita S, Sakurabayashi T,
Suzuki M, Sakai S, Yuasa Y, Hirasawa Y, Nishi T. Switch from
conventional to high-flux membrane reduces the risk of carpal
tunnel syndrome and mortality of hemodialysis patients. Kidney
Int. 1997;52:1096–101.
10. Locatelli F, Marcelli D, Conte F, Limido A, Malberti F, Spotti D.
Comparison of mortality in ESRD patients on convective and
diffusive extracorporeal treatments. Kidney Int. 1999;55:286–93.
11. Locatelli F, Manzoni C, Di Filippo S. The importance of con-
vective transport. Kidney Int. 2002;61:s115–20.
12. Canaud B, Bragg-Gresham JL, Marshall MR, Desmeules S,
Gillespie BW, Depner T, Klassen P, Port FK. Mortality risk for
patients receiving hemodiafiltration versus hemodialysis: Euro-
pean results from the DOPPS. Kidney Int. 2006;69:2087–93.
116 J Artif Organs (2010) 13:113–116
123