Page 1: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

1

Nam K. Tran, PhD, HCLD, (ABB), FACB Associate Clinical Professor Director of Clinical Chemistry, Special Chemistry/Toxicology and POCT

Interference and Point-of-Care Testing Devices

Page 2: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

2

Learning Objectives • Identify common interferences affecting POC

testing • Describe cases where interfering substances

affected patient care. • Describe solutions to mitigate the impact of

interfering substances on POC testing.

Page 3: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

3

POCT Device Formats Definition: POCT is defined as testing at or near the site of patient care

Page 4: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

4

POCT Device Formats Examples: • Disposable

• Handheld

• Portable

• Transportable

• Benchtop

• Monitoring

Page 5: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

5

POCT Device Formats Examples: • Disposable

• Handheld

• Portable

• Transportable

• Benchtop

• Monitoring

Being FDA approved as a POCT device does not mean it is not susceptible to interfering substances!!!

Page 6: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

6

Pre-Analytical

Total Testing Process: Lab testing occurs over three critical phases:

Total Testing Process: Difference Phases

Page 7: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

7

Pre-Analytical Analytical

Total Testing Process: Difference Phases Total Testing Process: Lab testing occurs over three critical phases:

Page 8: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

8

Pre-Analytical Analytical Post-Analytical

Total Testing Process: Difference Phases Total Testing Process: Lab testing occurs over three critical phases:

Page 9: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

9

Pre-Analytical Analytical Post-Analytical TREATMENT

Total Testing Process: Difference Phases Total Testing Process: Lab testing occurs over three critical phases:

Page 10: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

10

Pre-Analytical Analytical Post-Analytical TREATMENT

Errors in the Pre-Analytical Phase: Most frequent source of errors (up to 70%). Incorrect

Patient preparation Sample collection Transportation Accessioning Processing

Total Testing Process: Sources of Error C

ompo

nent

s

Page 11: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

11

Pre-Analytical Analytical Post-Analytical TREATMENT

Errors in the Pre-Analytical Phase: Most frequent source of errors (up to 70%). Incorrect

Total Testing Process: Sources of Error

Patient preparation Sample collection Transportation Accessioning Processing

Incorrect patient ID Mislabeling of specimens Hemolysis Wrong specimen type Improper specimen collection Interfering substances

Com

pone

nts

Sour

ces

of E

rror

Page 12: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

12

Pre-Analytical Analytical Post-Analytical TREATMENT

Testing

Total Testing Process: Sources of Error

Patient preparation Sample collection Transportation Accessioning Processing

Incorrect patient ID Mislabeling of specimens Hemolysis Wrong specimen type Improper specimen collection Interfering substances

Errors in the Analytical Phase: Infrequent in laboratory tests, however may be higher in POCT due to non-lab trained personnel operating devices.

Com

pone

nts

Sour

ces

of E

rror

Page 13: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

13

Pre-Analytical Analytical Post-Analytical TREATMENT

Testing

Total Testing Process: Sources of Error

Patient preparation Sample collection Transportation Accessioning Processing

Incorrect patient ID Mislabeling of specimens Hemolysis Wrong specimen type Improper specimen collection Interfering substances

QC/calibration Operator error Bad reagents

Errors in the Analytical Phase: Infrequent in laboratory tests, however may be higher in POCT due to non-lab trained personnel operating devices.

Com

pone

nts

Sour

ces

of E

rror

Page 14: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

14

Pre-Analytical Analytical Post-Analytical TREATMENT

Results interpretation Entry to LIS/EMR Contacting providers Sample archiving

Total Testing Process: Sources of Error

Testing

Patient preparation Sample collection Transportation Accessioning Processing

Incorrect patient ID Mislabeling of specimens Hemolysis Wrong specimen type Improper specimen collection Interfering substances

QC/calibration Operator error Bad reagents

Errors in the Post-Analytical Phase: Second most common among laboratory-based results.

Com

pone

nts

Sour

ces

of E

rror

Page 15: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

15

Pre-Analytical Analytical Post-Analytical TREATMENT

Results interpretation Entry to LIS/EMR Contacting providers Sample archiving

Total Testing Process: Sources of Error

Testing

Patient preparation Sample collection Transportation Accessioning Processing

Incorrect patient ID Mislabeling of specimens Hemolysis Wrong specimen type Improper specimen collection Interfering substances

QC/calibration Operator error Bad reagents

Misinterpretation of results IT problems

Errors in the Post-Analytical Phase: Second most common among laboratory-based results.

Com

pone

nts

Sour

ces

of E

rror

Page 16: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

16

Pre-Analytical Analytical Post-Analytical TREATMENT

Results interpretation Entry to LIS/EMR Contacting providers Sample archiving

Total Testing Process: Sources of Error

Testing

Patient preparation Sample collection Transportation Accessioning Processing

Incorrect patient ID Mislabeling of specimens Hemolysis Wrong specimen type Improper specimen collection Interfering substances

QC/calibration Operator error Bad reagents

Misinterpretation of results IT problems

Errors in the Post-Analytical Phase: Second most common among laboratory-based results.

Com

pone

nts

Sour

ces

of E

rror

What is the significance of testing error in POCT?

Page 17: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

17

Glucose Meter Paradigm to Highlight the Role of Testing Errors

Page 18: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

18

Glucose Meter Paradigm to Highlight the Role of Testing Errors

>12,000 glucose meter errors are reported to the FDA each year – highest number of adverse events for any in vitro diagnostic device.

Page 19: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

19

Glucose Meter Paradigm to Highlight the Role of Testing Errors

>12,000 glucose meter errors are reported to the FDA each year – highest number of adverse events for any in vitro diagnostic device.

12,672 serious injuries reported from 2004-2008 to the FDA.

Page 20: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

20

Glucose Meter Paradigm to Highlight the Role of Testing Errors

>12,000 glucose meter errors are reported to the FDA each year – highest number of adverse events for any in vitro diagnostic device.

12,672 serious injuries reported from 2004-2008 to the FDA. Most of these reported errors are due to erroneous results from interfering substances and operator error.

Page 21: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

21

Glucose Meter Paradigm to Highlight the Role of Testing Errors

>12,000 glucose meter errors are reported to the FDA each year – highest number of adverse events for any in vitro diagnostic device.

12,672 serious injuries reported from 2004-2008 to the FDA. Most of these reported errors are due to erroneous results from interfering substances and operator error.

Page 22: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

22

Glucose Meter Paradigm to Highlight the Role of Testing Errors

>12,000 glucose meter errors are reported to the FDA each year – highest number of adverse events for any in vitro diagnostic device.

12,672 serious injuries reported from 2004-2008 to the FDA. Most of these reported errors are due to erroneous results from interfering substances and operator error.

Page 23: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

23

Common Confounding Factors for Glucose Meters

Anemia and polycythemia causes falsely high or falsely low results respectively.

Page 24: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

24

-80

-60

-40

-20

0

20

40

60

0 10 20 30 40 50 60 70

BIA

S (m

g/dL

)

Hematocrit (%)

y = -0.9465x + 30.951 R² = 0.4171

Hematocrit Effects on BGMS Measurements

Note: Bias = BGMS – Plasma Glucose

Sample Size: 60 Hematocrit Range: 19 - 60% Glucose Range: 90 - 296 mg/dL

Page 25: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

25

Common Confounding Factors for Glucose Meters

Page 26: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

26

Common Confounding Factors for Glucose Meters

Oxidizing and reducing substances interfere with electrochemical sensors causing falsely high or low results.

Page 27: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

Drug Interferents (Oxidizing Substances)

Tang Z, et al. Am J Clin Pathol 2000;113:75-86

Page 28: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

28

The role of drug interferences in critical care BGMS accuracy

Tran NK, et al. J Burn Care Res

2014;35:72-79

Page 29: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

50

100

150

200

250

300

0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 36

Time Following Admission (Hours)

Blo

od G

luco

se (m

g/dL

)

Started Ascorbic Acid (66 mg/kg/hr) Stopped Ascorbic Acid

BGMS B*** History: Patient is a 21 y/o

woman with 90% TBSA burns from MVA.

CASE EXAMPLE: ASCORBIC ACID INTERFERENCE

Page 30: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

50

100

150

200

250

300

0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 36

Time Following Admission (Hours)

Blo

od G

luco

se (m

g/dL

)

Stopped Ascorbic Acid

BGMS B PLASMA GLUCOSE (Lab)

Started Ascorbic Acid (66 mg/kg/hr)

CASE EXAMPLE: ASCORBIC ACID INTERFERENCE

Page 31: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

50

100

150

200

250

300

0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 36

Time Following Admission (Hours)

Blo

od G

luco

se (m

g/dL

)

Stopped Ascorbic Acid

BGMS A BGMS B*** PLASMA GLUCOSE (Lab)

Started Ascorbic Acid (66 mg/kg/hr)

CASE EXAMPLE: ASCORBIC ACID INTERFERENCE

Page 32: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

32

Common Confounding Factors for Glucose Meters

Page 33: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

33

Common Confounding Factors for Glucose Meters

Specimen temp alters biosensor enzyme kinetics. Hypotension/shock affect capillary specimens.

Page 34: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

34

Common Confounding Factors for Glucose Meters

Some glucose meters cannot differentiate between certain non-glucose sugars (e.g., maltose, galactose)

Page 35: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

35

Non-Glucose Sugar Interferences

• Icodextrin is a dialysis drug. It is metabolized by the body to maltose. In some glucose biosensors, maltose is indistinguishable from glucose.

Page 36: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing
Page 37: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

37

FDA MAUDE Database website: http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfmaude/ search.cfm, Accessed on August 20, 2014

BGMS A BGMS B BGMS C

Timeframe 1997-14 2013-14 2007-11

Adverse Events (Deaths)

28 (13) 5 (0) 0 (0)

Erroneous Results 557 168 15

Non-Clinical Event 387 59 21

TOTAL 1094 232 36

Maltose Related Deaths

Page 38: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

38

Similar sensor designs so susceptible to similar interferences (will vary based on manufacturer).

CGM based on interstitial fluid measurements and not plasma or whole blood.

Potential for many other sources of interferences.

CGM does not fall under CLIA and most devices compared against obsolete or poor reference methods such as the YSI.

Use WITH caution!

Continuous Glucose Monitors?

Page 39: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

39

INTERFERENCES IN WHOLE BLOOD ANALYSIS

Page 40: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

40

INTERFERENCES IN WHOLE BLOOD ANALYSIS

Air Contamination

Delayed Testing

Hemolysis

Hemodilution/Hemoconcentration

Page 41: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

41

INTERFERENCES IN WHOLE BLOOD ANALYSIS

Air Contamination

Delayed Testing

Hemodilution/Hemoconcentration

Hemolysis

Page 42: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

42

Air Contamination of Blood Specimens

Background: Anesthesia reports “impossible venous blood gas values” in one patient where end tidal CO2 was greater than the venous blood gas (VBG).

Page 43: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

43

Air Contamination of Blood Specimens

Background: Anesthesia reports “impossible venous blood gas values” in one patient where end tidal CO2 was greater than the venous blood gas (VBG). • POC Venous Blood Gas: pH = 7.54, pCO2 = 17.5, pO2 = 168.5

• POC VBG#2: pH = 7.56, pCO2 = 12.7, pO2 = 165.9

• End tidal CO2 = 28

Page 44: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

44

Air Contamination of Blood Specimens

Background: Anesthesia reports “impossible venous blood gas values” in one patient where end tidal CO2 was greater than the venous blood gas (VBG). • POC Venous Blood Gas: pH = 7.54, pCO2 = 17.5, pO2 = 168.5

• POC VBG#2: pH = 7.56, pCO2 = 12.7, pO2 = 165.9

• End tidal CO2 = 28

• Lab Venous Blood Gas: pH 7.54, pCO2 = 19.2, pO2 = 161.5

Page 45: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

45

Air Contamination of Blood Specimens

Blood Gas Laboratory identified “air bubbles” in syringe

Background: Anesthesia reports “impossible venous blood gas values” in one patient where end tidal CO2 was greater than the venous blood gas (VBG). • POC Venous Blood Gas: pH = 7.54, pCO2 = 17.5, pO2 = 168.5

• POC VBG#2: pH = 7.56, pCO2 = 12.7, pO2 = 165.9

• End tidal CO2 = 28

• Lab Venous Blood Gas: pH 7.54, pCO2 = 19.2, pO2 = 161.5

Page 46: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

46

Air Contamination of Blood Specimens

Background: Anesthesia reports “impossible venous blood gas values” in one patient where end tidal CO2 was greater than the venous blood gas (VBG). • POC Venous Blood Gas: pH = 7.54, pCO2 = 17.5, pO2 = 168.5

• POC VBG#2: pH = 7.56, pCO2 = 12.7, pO2 = 165.9

• End tidal CO2 = 28

• Lab Venous Blood Gas: pH 7.54, pCO2 = 19.2, pO2 = 161.5

• Air bubbles can quickly (<5 mins) cause the specimen to

equilibrate atmospheric air (1 atm = 760 mmHg = 0.21 x 760 = 150 mmHg for pO2!!!)

Page 47: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

47

INTERFERENCES IN BLOOD GAS ANALYSIS

Air Contamination

Delayed Testing

Hemolysis

Hemodilution/Hemoconcentration

Page 48: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

48

Pre-Analytical • Transportation delays Analysis should be performed within 20 to

30 minutes—Faster is better!

Specimen Processing Delays and Lactate

Page 49: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

49

Pre-Analytical • Transportation delays

Seymour CW, et al. BMC Research Notes 2011;4:169

Specimen Processing Delays and Lactate

Page 50: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

50

Pre-Analytical • Transportation delays • Inadequate inhibition of

glycolysis

If delays are expected, using a grey top tube may be appropriate, however it may take up to 15 minutes to achieve inhibition!

Specimen Processing Delays and Lactate

Page 51: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

51

Pre-Analytical • Transportation delays • Inadequate inhibition of

glycolysis

Astles R, et al. Clin Chem 1994;404:1327

Specimen Processing Delays and Lactate

Page 52: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

52

Pre-Analytical • Transportation delays • Inadequate inhibition of

glycolysis

If delays are expected, using a grey top tube may be appropriate, however it may take up to 15 minutes to achieve inhibition!

Astles R, et al. Clin Chem 1994;404:1327

Specimen Processing Delays and Lactate

Page 53: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

53

Pre-Analytical • Transportation delays • Inadequate inhibition of

glycolysis • Specimens not placed on ice

False elevations of lactate could be mitigated by placing samples on ice. Iced samples exhibit similar results to those tested immediately at up to 6 hours.

Specimen Processing Delays and Lactate

Page 54: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

54

Pre-Analytical • Transportation delays • Inadequate inhibition of

glycolysis • Specimens not placed on ice

Seymour CW, et al. BMC Research Notes 2011;4:169

Specimen Processing Delays and Lactate

Page 55: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

55

INTERFERENCES IN WHOLE BLOOD ANALYSIS

Air Contamination

Delayed Testing

Hemolysis

Hemodilution/Hemoconcentration

Page 56: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

56

• Spectrophotometric (Non-Cyanohemoglobin)

Contemporary Hemoglobinometric Techniques

• Measurement of hemoglobin is based on the absorption spectra

• Oxy- and deoxyhemoglobin exhibit

different absorption in the red to IR wavelengths.

• Measurement based on Beer’s Law (A = elc).

• Some methods require lysis and reacting with non-cyanide-based reagents.

Abso

rban

ce

HHb O2Hb COHb MetHb

Prism

White light

500 550 600 650 700 nm

Page 57: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

57

Conductance (Impendance)

Contemporary Hemoglobinometric Techniques

Electrode

VS.

• Red blood cell membranes are not conductive.

High Resistance Low Resistance

Page 58: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

58

Conductance (Impendance)

Contemporary Hemoglobinometric Techniques

Electrode

VS.

Hematocrit (%)

Res

ista

nce

(Ω)

• Red blood cell membranes are not conductive.

Page 59: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

59

Conductance (Impendance)

Contemporary Hemoglobinometric Techniques

Electrode

VS.

• Red blood cell membranes are not conductive.

• The number of red blood cells is proportional to the change in conductance and conforms to Ohm’s Law (V = IR)

Page 60: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

60

Conductance (Impendance)

Contemporary Hemoglobinometric Techniques

Electrode

VS.

• Red blood cell membranes are not conductive.

• The number of red blood cells is proportional to the change in conductance and conforms to Ohm’s Law (V = IR)

• Conductance-based methods measure hematocrit. The hematocrit can then be used to calculate hemoglobin based on a conversion factor (estimated hemoglobin = hematocrit / 3.4)*

Page 61: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

61

Conductance (Impendance)

Contemporary Hemoglobinometric Techniques

Electrode

VS.

Page 62: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

62

Case Study 2: Hemoconcentration

Background: Patient with suspected Ebola Virus symptoms admitted for evaluation. Isolation protocols were in effect.

Page 63: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

63

Case Study 2: Hemoconcentration

Background: Patient with suspected Ebola Virus symptoms admitted for evaluation. Isolation protocols were in effect.

Page 64: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

64

Case Study 2: Hemoconcentration

Background: Patient with suspected Ebola Virus symptoms admitted for evaluation. Isolation protocols were in effect. A handheld blood gas chemistry analyzer served as the primary chemistry analyzer. 0853 hrs – Specimens collected for chemistry and CBC testing.

Page 65: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

65

Case Study 2: Hemoconcentration

Background: Patient with suspected Ebola Virus symptoms admitted for evaluation. Isolation protocols were in effect. A handheld blood gas chemistry analyzer served as the primary chemistry analyzer. 0853 hrs – Specimens collected for chemistry and CBC testing.

Page 66: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

66

Case Study 2: Hemoconcentration

Background: Patient with suspected Ebola Virus symptoms admitted for evaluation. Isolation protocols were in effect. A handheld blood gas chemistry analyzer served as the primary chemistry analyzer. 0853 hrs – Specimens collected for chemistry and CBC testing.

Page 67: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

67

Case Study 2: Hemoconcentration

Background: Patient with suspected Ebola Virus symptoms admitted for evaluation. Isolation protocols were in effect. A handheld blood gas chemistry analyzer served as the primary chemistry analyzer. 0853 hrs – Specimens collected for chemistry and CBC testing.

Handheld Results Hct = 68% Hb = 21.9 g/dL

Page 68: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

68

Case Study 2: Hemoconcentration

Background: Patient with suspected Ebola Virus symptoms admitted for evaluation. Isolation protocols were in effect. A handheld blood gas chemistry analyzer served as the primary chemistry analyzer. 0853 hrs – Specimens collected for chemistry and CBC testing.

Handheld Results Hct = 68% Hb = 21.9 g/dL

CBC Results Hct = 41% Hb = 13.2 g/dL

Page 69: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

69

Case Study 2: Hemoconcentration

Background: Patient with suspected Ebola Virus symptoms admitted for evaluation. Isolation protocols were in effect. A handheld blood gas chemistry analyzer served as the primary chemistry analyzer. 0853 hrs – Specimens collected for chemistry and CBC testing.

Hct = 43% Hb = 13.8 g/dL

Handheld Results Hct = 68% Hb = 21.9 g/dL RE-MIXING!

CBC Results Hct = 41% Hb = 13.2 g/dL

Page 70: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

70

Case Study 2: Hemoconcentration

Background: Patient with suspected Ebola Virus symptoms admitted for evaluation. Isolation protocols were in effect. A handheld blood gas chemistry analyzer served as the primary chemistry analyzer. 0853 hrs – Specimens collected for chemistry and CBC testing.

Hct = 43% Hb = 13.8 g/dL

Inadequate mixing may result in artificial changes in total hemoglobin measurements.

Handheld Results Hct = 68% Hb = 21.9 g/dL RE-MIXING!

CBC Results Hct = 41% Hb = 13.2 g/dL

Page 71: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

71

Electrode

Conductance (Impendence)

Contemporary Hemoglobinometric Techniques

• Plasma protein content contributes to hematocrit measurements for conductance-based systems.

= Plasma Protein

High Resistance

Page 72: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

72

Electrode

Conductance (Impendence)

Contemporary Hemoglobinometric Techniques

= Plasma Protein

• Plasma protein content contributes to hematocrit measurements for conductance-based systems.

• Conductance-based systems assumes a relatively fixed protein concentration. Therefore, during hemodilution, hematocrit may be falsely lower and causing an underestimation of total hemoglobin.

Low Resistance from low plasma protein concentration!

Page 73: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

73

Electrode

Conductance (Impendence)

Contemporary Hemoglobinometric Techniques

= Plasma Protein

• Plasma protein content contributes to hematocrit measurements for conductance-based systems.

• Conductance-based systems assumes a relatively fixed protein concentration. Therefore, during hemodilution, hematocrit may be falsely lower and causing an underestimation of total hemoglobin.

• UCDMC Study: Comparison of a handheld blood gas analyzer using conductance-based measurement of hemoglobin versus a benchtop blood gas analyzer using a spectrophotometric-based method for hemoglobinometry.

Page 74: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

74

• Sixty patients requiring cardiac surgery were evaluated.

• Paired specimens were tested using a handheld POC analyzer and spectrophotometric methods through the core laboratory.

• Mean (SD) bias was -1.4 (1.1) g/dL, P = 0.011.

• Based on core laboratory results 12 patients would have received unnecessary transfusions.

Clinical Impact of Hemodilution for Point-of-Care Hemoglobin Measurements

Page 75: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

75

• Sixty patients requiring cardiac surgery were evaluated.

• Paired specimens were tested using a handheld POC analyzer and spectrophotometric methods through the core laboratory.

• Mean (SD) bias was -1.4 (1.1) g/dL, P = 0.011.

• Based on core laboratory results 12 patients would have received unnecessary transfusions.

Clinical Impact of Hemodilution for Point-of-Care Hemoglobin Measurements

Page 76: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

76

• Sixty patients requiring cardiac surgery were evaluated.

• Paired specimens were tested using a handheld POC analyzer and spectrophotometric methods through the core laboratory.

• Mean (SD) bias was -1.4 (1.1) g/dL, P = 0.011.

• Based on core laboratory results 12 patients would have received unnecessary transfusions.

Clinical Impact of Hemodilution for Point-of-Care Hemoglobin Measurements

= $219

$219 x 12 = $2,628 POTENTIALLY WASTED

Toner RW, et al. Appl Health Econ Health Policy 2011;9:29-37

Page 77: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

77

Case Study 3: Hemodilution

Page 78: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

78

Case Study 3: Hemodilution

Background: FDA MAUDE database reports a case (03P76-25) of a neonatal patient with discrepant point-of-care (POC) hemoglobin values compared to the laboratory. The POC device used a conductance-based method of hemoglobin measurement, while the laboratory used a spectrophotometric method.

Page 79: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

79

Case Study 3: Hemodilution

Background: FDA MAUDE database reports a case (03P76-25) of a neonatal patient with discrepant point-of-care (POC) hemoglobin values compared to the laboratory. The POC device used a conductance-based method of hemoglobin measurement, while the laboratory used a spectrophotometric method. • POC device reported a hematocrit of 22%. Physician administered 7 mL of

blood based on the POC result.

Page 80: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

80

Case Study 3: Hemodilution

Background: FDA MAUDE database reports a case (03P76-25) of a neonatal patient with discrepant point-of-care (POC) hemoglobin values compared to the laboratory. The POC device used a conductance-based method of hemoglobin measurement, while the laboratory used a spectrophotometric method. • POC device reported a hematocrit of 22%. Physician administered 7 mL of

blood based on the POC result.

• Transfusion was stopped halfway after the laboratory reported a hematocrit of 40% and hemoglobin of 11.7 g/dL.

Page 81: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

81

Case Study 3: Hemodilution

Background: FDA MAUDE database reports a case (03P76-25) of a neonatal patient with discrepant point-of-care (POC) hemoglobin values compared to the laboratory. The POC device used a conductance-based method of hemoglobin measurement, while the laboratory used a spectrophotometric method. • POC device reported a hematocrit of 22%. Physician administered 7 mL of

blood based on the POC result.

• Transfusion was stopped halfway after the laboratory reported a hematocrit of 40% and hemoglobin of 11.7 g/dL.

• Post-transfusion POC and lab hematocrit values were 45 and 50% respectively.

Page 82: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

82

Analytical Performance of Optical vs. Conductance-Based Hemoglobinometry

Dev

ice

#1 H

b (g

/dL)

Central Laboratory Hb (g/dL)

y = 0.5092x + 4.0176 R² = 0.5253

4

5

6

7

8

9

10

11

4 5 6 7 8 9 10 11 12

Page 83: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

83

Analytical Performance of Optical vs. Conductance-Based Hemoglobinometry

Dev

ice

#2 H

b (g

/dL)

Central Laboratory Hb (g/dL)

y = 0.5249x + 3.9443 R² = 0.5407

4

5

6

7

8

9

10

11

4 5 6 7 8 9 10 11 12

Page 84: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

84

Analytical Performance of Optical vs. Conductance-Based Hemoglobinometry

Dev

ice

#3 H

b (g

/dL)

y = 0.9345x + 0.4057 R² = 0.9205

4

5

6

7

8

9

10

11

4 5 6 7 8 9 10 11

Central Laboratory Hb (g/dL)

Page 85: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

85

Analytical Performance of Optical vs. Conductance-Based Hemoglobinometry

Notes: Reference Method = Beckman LH hematology analyzer

Page 86: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

86

Analytical Performance of Optical vs. Conductance-Based Hemoglobinometry

Notes: Reference Method = Beckman LH hematology analyzer

Median (IQR) Bias: 0.78 (0.78) g/dL P < 0.001 N = 50

Page 87: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

87

Analytical Performance of Optical vs. Conductance-Based Hemoglobinometry

Notes: Reference Method = Beckman LH hematology analyzer

Median (IQR) Bias: 0.73 (0.60) g/dL P < 0.001 N = 50

Page 88: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

88

Analytical Performance of Optical vs. Conductance-Based Hemoglobinometry

Notes: Reference Method = Beckman LH hematology analyzer

Median (IQR) Bias: 0.22 (0.20) g/dL P = 0.510 N = 50

Page 89: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

89

Analytical Performance of Optical vs. Conductance-Based Hemoglobinometry

Notes: *** P<0.001, Central Lab = Spectrophotometric Method, n = 20 patients

6

6.5

7

7.5

8

8.5

9

9.5

10

1 2 3 4 5

Tota

l Hem

oglo

bin

(g/d

L)

Time Point

Serial Testing Performance at 7 and 8 g/dL • Serial testing revealed significant analytical

bias between spectrophotometry vs. conductance-based measurements.

*** Spectrophotometric-based Methods

Page 90: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

90

Analytical Performance of Optical vs. Conductance-Based Hemoglobinometry

Notes: *** P<0.001, Central Lab = Spectrophotometric Method, n = 20 patients

6

6.5

7

7.5

8

8.5

9

9.5

10

1 2 3 4 5

Tota

l Hem

oglo

bin

(g/d

L)

Time Point

Serial Testing Performance at 7 and 8 g/dL • Serial testing revealed significant analytical

bias between spectrophotometry vs. conductance-based measurements.

*** Spectrophotometric-based Methods

Conductance-based Methods

Page 91: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

91

Analytical Performance of Optical vs. Conductance-Based Hemoglobinometry

6

6.5

7

7.5

8

8.5

9

9.5

10

1 2 3 4 5

Tota

l Hem

oglo

bin

(g/d

L)

Time Point

Serial Testing Performance at 7 and 8 g/dL • Serial testing revealed significant analytical

bias between spectrophotometry vs. conductance-based measurements.

• Conductance-based devices would have prompted unnecessary transfusions at time point #5 for patients using the 7 g/dL cutoff.

Unnecessary Transfusion Risk

***

Notes: *** P<0.001, Central Lab = Spectrophotometric Method, n = 20 patients

Page 92: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

92

Analytical Performance of Optical vs. Conductance-Based Hemoglobinometry

6

6.5

7

7.5

8

8.5

9

9.5

10

1 2 3 4 5

Tota

l Hem

oglo

bin

(g/d

L)

Time Point

Serial Testing Performance at 7 and 8 g/dL • Serial testing revealed significant analytical

bias between spectrophotometry vs. conductance-based measurements.

• Conductance-based devices would have prompted unnecessary transfusions at time point #5 for patients using the 7 g/dL cutoff.

• All serial conductance measurements were at risk for potential transfusions if the 8 g/dL cutoff was used.

Unnecessary Transfusion Risk

***

Notes: *** P<0.001, Central Lab = Spectrophotometric Method, n = 20 patients

Page 93: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

93

Analytical Performance of Optical vs. Conductance-Based Hemoglobinometry

6

6.5

7

7.5

8

8.5

9

9.5

10

1 2 3 4 5

Central Laboratory Method

Tota

l Hem

oglo

bin

(g/d

L)

Time Point

Serial Testing Performance at 7 and 8 g/dL • Serial testing revealed significant analytical

bias between spectrophotometry vs. conductance-based measurements.

• Conductance-based devices would have prompted unnecessary transfusions at time point #5 for patients using the 7 g/dL cutoff.

• All serial conductance measurements were at risk for potential transfusions if the 8 g/dL cutoff was used.

Unnecessary Transfusion Risk

***

Notes: *** P<0.001, Central Lab = Spectrophotometric Method, n = 20 patients

Page 94: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

94

Manufacturer and User Facility Device Experience (MAUDE) Database Summary

Device 1 Device 2 Device 3 Timeframe 2011-2016 2011-2016 2014-2016* Erroneous Results 8 0 0 Improper Transfusions

5 0 0

https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfmaude/results.cfm, Accessed on July 19, 2016

Page 95: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

95

INTERFERENCES IN WHOLE BLOOD ANALYSIS

Air Contamination

Delayed Testing

Hemolysis

Hemodilution/Hemoconcentration

Page 96: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

96

INTERFERENCES IN WHOLE BLOOD ANALYSIS

Air Contamination

Delayed Testing

Hemolysis

Hemodilution/Hemoconcentration Pseudohyperkalemia

Page 97: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

97

INTERFERENCES IN WHOLE BLOOD ANALYSIS

Air Contamination

Delayed Testing

Hemolysis

Hemodilution/Hemoconcentration Pseudohyperkalemia “Pseudonormokalemia”

Page 98: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

98

INTERFERENCES IN WHOLE BLOOD ANALYSIS

Air Contamination

Delayed Testing

Hemolysis

Hemodilution/Hemoconcentration Pseudohyperkalemia “Pseudonormokalemia”

No current FDA approved integrated solutions for detecting hemolysis at the point-of-care

Page 99: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

99

Biotin: The “Snake Oil” of 2018?

Page 100: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing
Page 101: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

101

Biotin and Cardiac Troponin Testing

Page 102: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

102

Mumma B, et al. AACC Poster Presentation 2018

• 1,443 Gen 5 troponin T samples tested (0-hour, n = 797; 3-hour, n=646) from 850 patients.

Estimating the Probability of Biotin Interference

Page 103: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

103

Mumma B, et al. AACC Poster Presentation 2018

• 1,443 Gen 5 troponin T samples tested (0-hour, n = 797; 3-hour, n=646) from 850 patients.

• Biotin not detectable in 471 (59%) and 399 (62%) 3-hour samples.

Estimating the Probability of Biotin Interference

Page 104: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

104

• 1,443 Gen 5 troponin T samples tested (0-hour, n = 797; 3-hour, n=646) from 850 patients.

• Biotin not detectable in 471 (59%) and 399 (62%) 3-hour samples.

• Only one 0-hour sample and one 3-hour sample had biotin >20 ng/mL (0.13% [95% CI: 0-0.7%]).

Mumma B, et al. AACC Poster Presentation 2018

Estimating the Probability of Biotin Interference

Page 105: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

105

Mumma B, et al. AACC Poster Presentation 2018

Estimating the Probability of Biotin Interference

Page 106: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

106

Estimating the Probability of Biotin Interference

Page 107: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

107

Adult ED Patients with Unknown Biotin Status: 540 Average Plasma Biotin: 1.15 (0.97) ng/mL

Specimens collected as part of clinical validation

UC Davis Cardiac Troponin Patients

Page 108: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

108

Adult ED Patients with Unknown Biotin Status: 540 Average Plasma Biotin: 1.15 (0.97) ng/mL

0

20

40

60

80

100

120

140

160

Num

ber o

f Pat

ient

s

Biotin Concentration (ng/mL)

Gen 5 TnT Biotin Interference Threshold is 20 ng/mL

Biotin quantified by GC-TOF-MS

UC Davis Cardiac Troponin Patients

Page 109: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

109

Adult ED Patients with Unknown Biotin Status: 540 Average Plasma Biotin: 1.15 (0.97) ng/mL

0

20

40

60

80

100

120

140

160

Num

ber o

f Pat

ient

s

Biotin Concentration (ng/mL)

Gen 5 TnT Biotin Interference Threshold is 20 ng/mL

Biotin quantified by GC-TOF-MS

BIOTIN IS LESS LIKELY TO BE A PROBLEM IN CARDIAC TROPONIN TESTING AND IS POPULATION SPECIFIC!

UC Davis Cardiac Troponin Patients

Page 110: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

110

Biotin and Urine Pregnancy Testing

Page 111: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

111

Biotin Interference with Urine Pregnancy Tests

• Recent studies show some point-of-care urine pregnancy tests were affected by biotin.

• Biotin is cleared by the kidneys.

• In this study, the QuickVue urine pregnancy test exhibited interference as low as 6 microgram/mL of urine biotin!

Williams G, et al. Clin Biochem 2018;53:168-170

Page 112: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

112

Best POCT Practices for Mitigating Interfering Substances

Page 113: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

113

• Education: The laboratory must be the leader in educating providers and patients of potential test interferences. Go to grand rounds, build partnerships, and provide multi-modality means to disseminate knowledge.

POCT Best Practices for Interferences

Page 114: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

114

• Education: The laboratory must be the leader in educating providers and patients of potential test interferences. Go to grand rounds, build partnerships, and provide multi-modality means to disseminate knowledge.

POCT Best Practices for Interferences

Page 115: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

115

• Education: The laboratory must be the leader in educating providers and patients of potential test interferences. Go to grand rounds, build partnerships, and provide multi-modality means to disseminate knowledge.

• Surveillance: Know your population! Collect data and determine if your local population may be be at risk for certain interferences (e.g., biotin, vitamin C, etc). MAUDE database is also helpful!

POCT Best Practices for Interferences

Page 116: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

116

• Education: The laboratory must be the leader in educating providers of potential test interferences. Go to grand rounds, build partnerships, and provide multi-modality means to disseminate knowledge.

• Surveillance: Know your population! Collect data and determine if your local population may be be at risk for certain interferences (e.g., biotin, vitamin C, etc). MAUDE database is also helpful!

• Electronic Early-Warning Systems: Leverage electronic solutions. Ordering of susceptible tests could flag both on the provider and laboratory side certain substances are identified.

POCT Best Practices for Interferences

Page 117: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

117

Conclusions • Interfering substances are out there and

impact POC testing as much as traditional lab testing!

• Interferences in common POC devices such as glucose meters have resulted in injury and death.

• Interferences in whole blood analysis have resulted in inappropriate treatment decisions.

• Medications and supplements may also affect POC immunoassays such as urine pregnancy tests.

• Education and awareness is critical to minimizing errors associated with interfering substances.

Page 118: Interference and Point-of-Care Testing Devices€¦ · Interference and Point-of-Care Testing Devices . 2 . Learning Objectives • Identify common interferences affecting POC testing

118

Questions?