Instruments for Radiation Detection and Measurement Lab # 4

Upload: zubeda

Post on 06-Jan-2016

72 views

Category:

Documents


1 download

DESCRIPTION

Instruments for Radiation Detection and Measurement. Lab # 4. In nuclear medicine it is important to ascertain the Presence Type Intensity Energy of radiations emitted by radionuclides Two commonly used devices Gas-filled detectors Scintillation detectors. - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: Instruments for Radiation Detection and Measurement

Instruments for Radiation Detection

and Measurement

Lab # 4

Page 2: Instruments for Radiation Detection and Measurement

• In nuclear medicine it is important to ascertain the

– Presence– Type– Intensity– Energy of radiations emitted by radionuclides

• Two commonly used devices

– Gas-filled detectors – Scintillation detectors

Page 3: Instruments for Radiation Detection and Measurement

Scintillation Detecting Instruments

• g-ray detecting equipment• Most commonly used:

– well counters– Thyroid probes– g or scintillation

• All these instruments are g-ray detecting devices • Consist of:• Collimator (excluding well counter) • Sodium iodide detector• Photomultiplier tube• Preamplifier• Pulse height analyzer• Display or Storage

Page 4: Instruments for Radiation Detection and Measurement

• Scintillation detectors consist of scintilator emitting flashes of light after absorbing gamma or x radiation. The light photons produced are then converted to an electrical pulse by means of a photomultiplier tube. The pulse is amplified by a linear amplifier, sorted by a pulse-height analyzer and then registred as a count. Different solid or liquid scintillators are used for different types of radiation. In nuclear medicine, sodium iodide solid crystals with a trace of thallium NaI(Tl) are used for gamma and x ray detection.

Page 5: Instruments for Radiation Detection and Measurement

g rays from a source interact in the sodium iodide detector and light photons are emitted.

The light photons will strike the photocathode of a photomultiplier(PM) tube and a pulse is generated at the end of the PM tube.

The pulse is first amplified by a preamplifier and then by a linear amplifier

Page 6: Instruments for Radiation Detection and Measurement

Scintillation Camera

• also known as a gamma camera• consists of :• Collimator• Detector• X, Y positioning circuit• PM tubes • Preamplifiers• Linear amplifiers• PHA • Display or storage

Page 7: Instruments for Radiation Detection and Measurement
Page 8: Instruments for Radiation Detection and Measurement

Collimator

• classification of collimators used in scintillation cameras depends primarily on – The type of focusing– The thickness of the holes

• Depending on the type of focusing– parallel hole– Pinholet – Converging – Diverging type

Page 9: Instruments for Radiation Detection and Measurement

Pinhole collimators are used in imaging small

organs such as thyroid glands

Converging collimators are employed when

the target organ is smaller than the size of the detector

Page 10: Instruments for Radiation Detection and Measurement

divergingcollimators are used in imaging

organs such as lungs that are larger than the

size of the detector

Parallel hole collimators are most commonly used in

nuclear medicine procedures.

Page 11: Instruments for Radiation Detection and Measurement

• Parallel hole collimators are classified as high-resolution, all-purpose, and high-sensitivity types.

• The size and number of holes the same for all these collimator

• The only change is in the thickness.• High sensitivity collimators are made with

smaller thickness than all-purpose collimators• High-resolution collimators are made thickest of

all.

Page 12: Instruments for Radiation Detection and Measurement

Detector

• NaI(Tl) crystals used as the detector

• Rectangular in shape

• Have the dimension between 33 X 43 cm and 37 X 59 cm with thickness varying between 0.64 cm and 1.9 cm

• • The most common thickness is 0.95 cm

• The 0.64-cm thick detectors are usually used in portable cameras for nuclear cardiac studies

Page 13: Instruments for Radiation Detection and Measurement

Detector

• Increasing the thickness of a crystal increases the probability of complete absorption of g rays and hence the sensitivity of the detector

Page 14: Instruments for Radiation Detection and Measurement

X, Y Positioning Circuit

• When a g ray interacts in the crystal, its exact location is determined by the X, Y positioning circuit

• Many PM tubes are mounted on the NaI(Tl) crystal in scintillation cameras

• After g-ray interaction in the crystal, a maximum amount of light will be received by the PM tube nearest to the point of interaction

Page 15: Instruments for Radiation Detection and Measurement

Pulse Height Analyzer

• circuit that sums up the output of all PM tubes to produce a pulse known as the Z pulse that represents the energy of a g ray

• The SCA analyzes the amplitude of the Z pulses and selects only those of desired energy by the use of appropriate peak energy and percent window settings

• In many scintillation cameras, the energy selection is made automatically by pushbutton type isotope selectors designated for different radionuclides such as 131I, 99mTc

Page 16: Instruments for Radiation Detection and Measurement

Pulse Height Analyzer

• In some scintillation cameras, two or three SCAs are used to select simultaneously two or three g rays of different energies

Page 17: Instruments for Radiation Detection and Measurement

Display and Storage

• most cameras employ digital computers in acquiring, storing, and processing of image data

Page 18: Instruments for Radiation Detection and Measurement

Tomographic Imagers

• limitation of the scintillation cameras is that they depict images of three-dimensional activity distributions in two-dimensional displays

• One way to solve this problem is to obtain images at different angles around the patient such as anterior, posterior, lateral, and oblique projections

• Success of the technique is limited because of the complexity of structures surrounding the organ of interest

– Single Photon Emission Computed Tomography– Positron Emission Tomography

Page 19: Instruments for Radiation Detection and Measurement

Tomographic Imagers

• mathematical algorithms, to reconstruct the images at distinct focal planes (slices).