Response Checks of PortableRadiation Detection Instruments
![Page 2: Response Checks of Portable Radiation Detection Instruments](https://reader030.vdocuments.mx/reader030/viewer/2022032703/56649d225503460f949f81bc/html5/thumbnails/2.jpg)
Enabling Objectives
Name Instrument Control Procedure
State Calibration Organization
State Frequency of Response Checks
Identify Response Check Equipment
Identify Shepherd Components
![Page 3: Response Checks of Portable Radiation Detection Instruments](https://reader030.vdocuments.mx/reader030/viewer/2022032703/56649d225503460f949f81bc/html5/thumbnails/3.jpg)
Enabling Objectives
State Isotopes & Activity for Shepherd Sources
Identify Response Window Criteria Identify Instruments Response
Checked with Plutonium State Purpose of Scintrex Model
309A
![Page 4: Response Checks of Portable Radiation Detection Instruments](https://reader030.vdocuments.mx/reader030/viewer/2022032703/56649d225503460f949f81bc/html5/thumbnails/4.jpg)
Instrumentation Controls
“Radiological Instrument/ Equipment Controls”
![Page 5: Response Checks of Portable Radiation Detection Instruments](https://reader030.vdocuments.mx/reader030/viewer/2022032703/56649d225503460f949f81bc/html5/thumbnails/5.jpg)
Instrument Calibration
Portable Radiation Detection Instruments Calibrated by:
Environmental RadiologicalMonitoring & Instrumentation (ERM&I)
Western Area Radiological Laboratory,(WARL) Muscle Shoals, AL
![Page 6: Response Checks of Portable Radiation Detection Instruments](https://reader030.vdocuments.mx/reader030/viewer/2022032703/56649d225503460f949f81bc/html5/thumbnails/6.jpg)
Response Check Frequency
Teletector & RSO-7 Daily Beta-Gamma Instruments Weekly Alpha & Beta Friskers Weekly Neutron Dose Instruments Weekly On-Site REP Instruments Weekly Electronic Dosimeters
Quarterly
![Page 7: Response Checks of Portable Radiation Detection Instruments](https://reader030.vdocuments.mx/reader030/viewer/2022032703/56649d225503460f949f81bc/html5/thumbnails/7.jpg)
Response Checks ofBeta-Gamma Dose Instruments
Shepherd Model 89 Irradiator
- Source Rod Assembly
- Shielded Calibration Range
Assembly
- Attenuator Assembly
- Table Assembly
![Page 8: Response Checks of Portable Radiation Detection Instruments](https://reader030.vdocuments.mx/reader030/viewer/2022032703/56649d225503460f949f81bc/html5/thumbnails/8.jpg)
Shepherd Source Rod Assembly
Cs-137130 mCi400 Ci
![Page 9: Response Checks of Portable Radiation Detection Instruments](https://reader030.vdocuments.mx/reader030/viewer/2022032703/56649d225503460f949f81bc/html5/thumbnails/9.jpg)
Shepherd Shielded Calibration Range Assembly
![Page 10: Response Checks of Portable Radiation Detection Instruments](https://reader030.vdocuments.mx/reader030/viewer/2022032703/56649d225503460f949f81bc/html5/thumbnails/10.jpg)
Shepherd Attenuator Assembly
![Page 11: Response Checks of Portable Radiation Detection Instruments](https://reader030.vdocuments.mx/reader030/viewer/2022032703/56649d225503460f949f81bc/html5/thumbnails/11.jpg)
Shepherd Table Assembly
![Page 12: Response Checks of Portable Radiation Detection Instruments](https://reader030.vdocuments.mx/reader030/viewer/2022032703/56649d225503460f949f81bc/html5/thumbnails/12.jpg)
Shepherd Precautions
Plugs in Unused Access Ports Hands & Fingers Clear of Door Never Lubricate Source Rod Don’t Twist Attenuator Handles Move Table Slowly Utilize Survey Meter
![Page 13: Response Checks of Portable Radiation Detection Instruments](https://reader030.vdocuments.mx/reader030/viewer/2022032703/56649d225503460f949f81bc/html5/thumbnails/13.jpg)
Shepherd Performance Checks
Test the Interlock System- Attempt to Raise Source or Open Door with Power Off- Attempt to Raise Source with Door Open- Attempt to Open Door with Source Raised
![Page 14: Response Checks of Portable Radiation Detection Instruments](https://reader030.vdocuments.mx/reader030/viewer/2022032703/56649d225503460f949f81bc/html5/thumbnails/14.jpg)
Establishing Response Windows
Expose at Least Four Instruments
Exposure Values Near Midscale
Set Window for Each Scale
![Page 15: Response Checks of Portable Radiation Detection Instruments](https://reader030.vdocuments.mx/reader030/viewer/2022032703/56649d225503460f949f81bc/html5/thumbnails/15.jpg)
Response Window Criteria
Arithmetic Mean of all Instruments Calculate Highest Percent Difference Calculate Lowest Percent Difference Discard Highest or Lowest Percent
Difference if Exceed 10 Percent Response Window = Mean ± 20 % Expressed as Range
![Page 16: Response Checks of Portable Radiation Detection Instruments](https://reader030.vdocuments.mx/reader030/viewer/2022032703/56649d225503460f949f81bc/html5/thumbnails/16.jpg)
Instrument Pre-Operational Checks
Inspect for Damage Inspect Switches, Alarms, Controls Inspect Cables & Power Cords Check Battery Strength Instrument in Calibration
![Page 17: Response Checks of Portable Radiation Detection Instruments](https://reader030.vdocuments.mx/reader030/viewer/2022032703/56649d225503460f949f81bc/html5/thumbnails/17.jpg)
Performing Response Checks with Shepherd Irradiator
Position Instrument on Guide
- Beta Gamma Survey Meter
- Teletector
- Electronic Dosimeter
![Page 18: Response Checks of Portable Radiation Detection Instruments](https://reader030.vdocuments.mx/reader030/viewer/2022032703/56649d225503460f949f81bc/html5/thumbnails/18.jpg)
Beta-Gamma Survey Meter
![Page 19: Response Checks of Portable Radiation Detection Instruments](https://reader030.vdocuments.mx/reader030/viewer/2022032703/56649d225503460f949f81bc/html5/thumbnails/19.jpg)
Teletector
![Page 20: Response Checks of Portable Radiation Detection Instruments](https://reader030.vdocuments.mx/reader030/viewer/2022032703/56649d225503460f949f81bc/html5/thumbnails/20.jpg)
Electronic Dosimeter
![Page 21: Response Checks of Portable Radiation Detection Instruments](https://reader030.vdocuments.mx/reader030/viewer/2022032703/56649d225503460f949f81bc/html5/thumbnails/21.jpg)
Performing Response Checks with Shepherd Irradiator Position Attenuators and Platform Close Door Raise Source to make Exposure Lower Source to “Off” Repeat for Each Scale Document Results
![Page 22: Response Checks of Portable Radiation Detection Instruments](https://reader030.vdocuments.mx/reader030/viewer/2022032703/56649d225503460f949f81bc/html5/thumbnails/22.jpg)
Observation Window
![Page 23: Response Checks of Portable Radiation Detection Instruments](https://reader030.vdocuments.mx/reader030/viewer/2022032703/56649d225503460f949f81bc/html5/thumbnails/23.jpg)
Response Check of AlphaSurvey Instruments Alpha Sources, ie Pu-239 Inspect Mylar Windows Perform Light Test Expose Detector Compare Results with Response Log Check Each Scale or Decade Document Results
![Page 24: Response Checks of Portable Radiation Detection Instruments](https://reader030.vdocuments.mx/reader030/viewer/2022032703/56649d225503460f949f81bc/html5/thumbnails/24.jpg)
Alpha Survey Instrument
![Page 25: Response Checks of Portable Radiation Detection Instruments](https://reader030.vdocuments.mx/reader030/viewer/2022032703/56649d225503460f949f81bc/html5/thumbnails/25.jpg)
Neutron Survey Instrument
![Page 26: Response Checks of Portable Radiation Detection Instruments](https://reader030.vdocuments.mx/reader030/viewer/2022032703/56649d225503460f949f81bc/html5/thumbnails/26.jpg)
Response Check ofNeutron Instruments
Set up the Instrument Scale Instrument Over Neutron Source Check Other Scales/Decades with
Pulse Generator Compare Results with Response
Log Document Results
![Page 27: Response Checks of Portable Radiation Detection Instruments](https://reader030.vdocuments.mx/reader030/viewer/2022032703/56649d225503460f949f81bc/html5/thumbnails/27.jpg)
Frisker
![Page 28: Response Checks of Portable Radiation Detection Instruments](https://reader030.vdocuments.mx/reader030/viewer/2022032703/56649d225503460f949f81bc/html5/thumbnails/28.jpg)
Response Check ofFriskers
Sources Identified in Response Log Set Frisker to Appropriate
Range/Scale Position Probe over Source Repeat for Other Scales Compare Results with Response Log Document Results
![Page 29: Response Checks of Portable Radiation Detection Instruments](https://reader030.vdocuments.mx/reader030/viewer/2022032703/56649d225503460f949f81bc/html5/thumbnails/29.jpg)
Tritium in Air Monitor
![Page 30: Response Checks of Portable Radiation Detection Instruments](https://reader030.vdocuments.mx/reader030/viewer/2022032703/56649d225503460f949f81bc/html5/thumbnails/30.jpg)
Response Check ofScintrex Model 309A
Inspect Filter Paper Ensure Battery >3.0 volts Set Function Switch to ‘Display On’ Press ‘Test’ Button Observe Display < 50
![Page 31: Response Checks of Portable Radiation Detection Instruments](https://reader030.vdocuments.mx/reader030/viewer/2022032703/56649d225503460f949f81bc/html5/thumbnails/31.jpg)
Response Check ofScintrex Model 309A
Position Gamma Source on Right Side
Wait 1-2 Minutes Observe Reading Between 8 & 24 Depress ‘Test’ Button Document Results
![Page 32: Response Checks of Portable Radiation Detection Instruments](https://reader030.vdocuments.mx/reader030/viewer/2022032703/56649d225503460f949f81bc/html5/thumbnails/32.jpg)
Summary Instrument Calibration Critical Response Checks Confirm Operation Shepherd Model 89 Irradiator Other Response Checks:
- Alpha Survey Instruments- Neutron Instruments- Friskers- Tritium-in-Air
![Page 33: Response Checks of Portable Radiation Detection Instruments](https://reader030.vdocuments.mx/reader030/viewer/2022032703/56649d225503460f949f81bc/html5/thumbnails/33.jpg)
Prevent Errors -Reduce Contamination